What is ultrasonic plethysmography and how can we use it in airway disease?

Size: px
Start display at page:

Download "What is ultrasonic plethysmography and how can we use it in airway disease?"

Transcription

1 What is ultrasonic plethysmography and how can we use it in airway disease? Christian Buess ndd Medizintechnik AG, Zürich, Switzerland ERS /19

2 Background Until approx. the year 2000 mainly the following flow sensor principles were used in lung function testing: Differential pressure (variable or fixed resistance) Turbines (rotation speed) Heated wires (cooling effect) All these principles have potential limitations: Hygiene: Difficult to clean Unable to measure very low flows Need of regular calibration Dependency on gas composition and/or temperature ERS /19

3 Operating Principle Ultrasonic flow measurement is commonly used for liquid flow. It is now also used for air flow measurement in lung function diagnostics. Integrated Mouthpiece Ultrasonic Transducer Sound Transparent Windows Exchangeable Flow Tube ERS /19

4 Operating Principle Ultrasonic transit-time measurement. Please note that the principle differs to the Doppler principle used for blood flow measurement. Sound Receiver Transmission of Ultrasonic Pulse Sound Transmission Path ERS /19

5 Operating Principle Transmission of Ultrasonic Pulse Sound Receiver Sound transmission and reception is repeated every 1 to 5 ms. Sampling rates 200 to 1000 Hz. ERS /19

6 Measurement principle (simple) New York Zürich 9 Hours 7 Hours Time difference is proportional to speed of Jet stream Ultrasonic flow sensor: Time difference up to 1 μs ERS /19

7 Measurement principle: Flow upstream transit-time (Zürich NY) downstream transit-time (NY Zürich) t d t u = = L c u cosϕ L c + u cosϕ u = Flow L 2 cosϕ t t d d t t u u Speed of sound depends on temperature and gas composition Flow measurement depends only on flow and geometry No dependency on temperature, gas composition, humidity or pressure ERS /19

8 Measurement principle: Molar Mass upstream transit-time t d = L c u cosϕ Molar Mass downstream transit-time t u = L c + u cosϕ M 4κR = 2 T L t t u u t + t d d 2 speed of sound c R T = κ M Molar mass measurement depends on transit-times and temperature along the sound transmission path Molar mass measurement for free no additional sensor required Molar mass measurement is not specific ERS /19

9 Strengths and Weaknesses No calibration since flow measurement depends only on time measurement and geometry of sensor Completely open, unobstructed tube Good hygienic solution High resolution and dynamic range Molar mass measurement for free Relatively high cost of sensor ERS /19

10 Spirometry: Devices Examples of devices using ultrasound technology Standalone spirometer for office use Flow sensors for spirometry and exercise testing Flow sensor for Intensive Care application ERS /19

11 Application The technology can be used for a large range Neonatology Resolution: < 1 ml/s Range: 500 ml/s Adult Resolution: < 10 ml/s Range: 18 l/s Horses Resolution: < 80 ml/s Range: 100 l/s ERS /19

12 Spirometry: Studies Performance of 70 devices over 3 to 6 month period R. Pérez-Padilla et al.: Long-term stability of portable spirometers. (Respir Care 2006) Performance of 6 devices over 200 days J. Walters et al.: Stability of the EasyOne ultrasonic spirometer for use in general practice. (Respirology 2006) Ultrasonic spirometer used by world-wide BOLD study (Burden of Obstructive Lung Disease). Ultrasonic spirometer used by Platino study in Latin America. ERS /19

13 FRC Measurement: Background Measurement of FRC using multiple breath wash-out / wash-in of a tracer gas by measuring molar mass of in- and expired gas flow. expiratory sidestream MM trace start of tracer wash-in wash-in of tracer gas Computation of FRC, LCI Main stream molar mass signal is synchronous with flow signal Simple instrument setup compared to mass spectrometry Best results are only achieved with side-stream MM sensor ERS /19

14 FRC Measurement: Setup Valve in open position Valve in closed position Test gas supply Ultrasonic Flow / MM Sensor Side-stream sampling tube ERS /19

15 FRC Measurement: Studies Direct comparison of ultrasonic system with mass spectrometer Fuchs et al., Pediatr Pulmol, 2006, 41. r 2 = 1.00, mean difference -0.2% Setup Problems: Gas leaks etc. Most devices and most software are in prototype stage! The above results were obtained using a side-stream MM sensor If applied carefully results compare well with mass spectrometers Additional studies: J. Pillow et al., Schibler et al. etc. ERS /19

16 Tidal Breathing Analysis: Background Comparable to capnography the molar mass analysis of tidal breathing shows the expiratory molar mass curve vs. expired volume S2 S Vd Volumen [l] Trial 1 Trial 2 Trial healthy control Volumen [l] Trial 1 Trial 2 Trial patient with COPD Several parameters can be computed from the curve data: Slope of phase 2 S 2 Slope of phase 3 S 3 Dead Space volume V d ERS /19

17 Tidal Breathing Analysis: Method 90 seconds of tidal breathing recording flow and molar mass {S 2,S 3,V d, {S 2,S 3,V d, {S 2,S 3,V d, {S 2,S 3,V d, {S 2,S 3,V d, {S 2,S 3,V d, {S 2,S 3,V d, {S 2,S 3,V d, {S 2,S 3,V d, V i,v e,f, } 1 V i,v e,f, } 2 V i,v e,f, } 3 V i,v e,f, } 4 V i,v e,f, } 5 V i,v e,f, } 6 V i,v e,f, } 7 V i,v e,f, } 8 V i,v e,f, } 9 Determination of mean expiratory MM curve S 2,S 3,V d Statistical Analysis Diagnosis ERS /19

18 Tidal Breathing Analysis: Studies Studies analyzing the Molar Mass Index in side-stream MMI = S 3 / S MMI Normal controls Very severe to moderate COPD Restricted Beckert, Buess, Magnussen: Not yet published Correlation of MMI and LCI, both determined by ultrasonic flow sensor Fuchs, Gaultier, Buess, Gappa: ERS ERS /19

19 Conclusion Ultrasonic flow measurement is a new technology with many advantages: No calibration, good hygienic solution and high accuracy Ultrasonic molar mass measurement is the mass spectrometer for everybody its use may have a large potential for measuring multi-breath FRC. Analysis of expiratory molar mass curve during quiet breathing: Potentially interesting for analysis of gas distribution in the lungs. Further investigations are required. ERS /19

EasyOne Pro LAB Measurement Technology Background

EasyOne Pro LAB Measurement Technology Background Application Note CB, June 2017 EasyOne Pro LAB Measurement Technology Background Content EasyOne Pro LAB Measurement Technology Background... 1 1 Introduction... 2 2 EasyOne Pro LAB System Layout... 2

More information

EasyWarePro Message Numbers

EasyWarePro Message Numbers Application Note CB 06/2016 EasyWarePro Message Numbers Important note: this application note is only valid beginning from EOPC V1.9.0.18 and EOP / Lab V1.9.0.19. Compared to the previous released version

More information

EasyWarePro Message Numbers

EasyWarePro Message Numbers Application Note CB 10/2018 EasyWarePro Message Numbers Important note: this application note is only valid beginning from EOPC V1.9.0.18 and EOP / Lab V1.9.0.19. Compared to the previous released version

More information

RESPIRATORY PHYSIOLOGY, PHYSICS AND

RESPIRATORY PHYSIOLOGY, PHYSICS AND Course n : Course 3 Title: RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE Sub-category: Techniques Topic: Respiratory monitoring Date: May 05-07, 2016 Language:

More information

Stikstofuitwas en Heliumdilutie. Eric Derom Universitair Ziekenhuis Gent

Stikstofuitwas en Heliumdilutie. Eric Derom Universitair Ziekenhuis Gent Stikstofuitwas en Heliumdilutie Eric Derom Universitair Ziekenhuis Gent Background Overview Helium-dilution technique Principle of Technique Equipment Procedure and Calculations Quality Control Nitrogen

More information

(12) United States Patent

(12) United States Patent (12) United States Patent USOO8352206B2 () Patent No.: Buess (45) Date of Patent: Jan. 8, 2013 (54) METHOD FOR THE SIGNAL (56) References Cited LINEARIZATION OF A GAS SENSOR OUTPUT SIGNAL U.S. PATENT DOCUMENTS

More information

February 22, Dear Jerry Masiello:

February 22, Dear Jerry Masiello: DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 22, 2017 Ndd

More information

Rotameter. 7. Respiration. Fleisch Pneumotachograph. Volumeter. Venturi Tube. Turbulent Flowmeter. Biomedical Information Technology Lab 1

Rotameter. 7. Respiration. Fleisch Pneumotachograph. Volumeter. Venturi Tube. Turbulent Flowmeter. Biomedical Information Technology Lab 1 7. Respiration Type Principle Rotameter Rotation of a rotor Volumeter Rotation of a rotor Pneumotachograph Air resistance Venturi tube Dynamic pressure Hot-wire anemometer Heat dissipation Time-of-flight

More information

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers Slemeyer, A. University of Applied Sciences of Giessen, Dept. of Electrical Engineering 1 Wiesenstr. 14, D-35390

More information

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance Introduction Expired Gas Analysis or indirect calorimetry, can be used to measure ventilation and the fractions of oxygen and carbon dioxide in expired air. From these measurements, the body's oxygen consumption

More information

LUNG CLEARANCE INDEX. COR-MAN IN Issue A, Rev INNOVISION ApS Skovvænget 2 DK-5620 Glamsbjerg Denmark

LUNG CLEARANCE INDEX. COR-MAN IN Issue A, Rev INNOVISION ApS Skovvænget 2 DK-5620 Glamsbjerg Denmark LUNG CLEARANCE INDEX METHOD COR-MAN-0000-008-IN Issue A, Rev. 3 2013-07-01 INNOVISION ApS Skovvænget 2 DK-5620 Glamsbjerg Denmark Tel.: +45 65 95 91 00 Fax: +45 65 95 78 00 info@innovision.dk www.innovision.dk

More information

Human Respiration Laboratory Experiment By

Human Respiration Laboratory Experiment By Human Respiration Laboratory Experiment By Alison L., Thurow, Brittany Baierlein, Rachel C. Holsinger and Robin L. Cooper Department of Biology, University of Kentucky, Lexington, KY 40506 0225, USA. Purpose:

More information

HARD (Gas Dilution) HARDER (Plethysmography) EASIEST (MiniBox+)

HARD (Gas Dilution) HARDER (Plethysmography) EASIEST (MiniBox+) Robert Brown, MD Director PFT Lab, Massachusetts General Hospital, Boston, MA Associate Professor of Medicine, Harvard Medical School o o o HARD (Gas Dilution) HARDER (Plethysmography) EASIEST (MiniBox+)

More information

Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams)

Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams) Name Lab Partners Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams) Part 1. Lung Volumes and Capacities Objectives 1. Obtain graphical representation of lung capacities

More information

CARDIOVIT AT-104 ergospirometry

CARDIOVIT AT-104 ergospirometry E D E R I T Z M A N U F A C T U R E D I N A N D S W I G N S E D L A N D CARDIOVIT AT-104 ergospirometry Cardiopulmonary exercise tests (CPET) with touch screen by SCHILLER it s the experience that counts.

More information

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT IV - 2013 MECHANICAL VENTILATORS Prof. Yasser Mostafa Kadah Mechanical Ventilator A ventilator is a machine, a system of related elements designed to alter, transmit, and direct energy

More information

Respiration. Exercise 1A: Breathing in Resting Volunteers Aim: To measure breathing parameters in a resting individual.

Respiration. Exercise 1A: Breathing in Resting Volunteers Aim: To measure breathing parameters in a resting individual. Respiration Background The amount of air that moves in or out of the lungs during any one breathing cycle is called the tidal volume. Above and beyond normal inspiration, it is possible to breathe in additional

More information

Technical Data and Specifications

Technical Data and Specifications Technical Data and Specifications INTENDED USE Ventilator designed to provide Invasive and Non-invasive ventilation for the critical care management of adult, pediatric and neonate-infant (including premature)

More information

A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers

A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers 2011 ROCKY MOUNTAIN NASA SPACE GRANT CONSORTIUM 1 A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers Christina Long, and Joseph Orr, Ph.D. Department of Bioengineering, University

More information

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung Susan Blonshine RRT, RPFT, FAARC, AE-C Joint ATS/ERS Taskforce Recent literature reviewed Surveyed current technical

More information

Respiration Lab Instructions

Respiration Lab Instructions Respiration Lab Instructions This laboratory investigation can be performed in any order. Be sure to read all instructions for each section before performing the experiment. PART 1 STUDENT WET SPIROMETER

More information

1 Design of the Portable mass spectrometric sensor for vascular and endocrine disease diagnostics

1 Design of the Portable mass spectrometric sensor for vascular and endocrine disease diagnostics 1 Design of the Portable mass spectrometric sensor for vascular and endocrine disease diagnostics Andrei Antonov, Viktor Kogan, Aleksandr Krutikov A.F. Ioffe Physical Technical Institute, Academy of Science,

More information

PULMONARY FUNCTION TESTING CORE

PULMONARY FUNCTION TESTING CORE PULMONARY FUNCTION TESTING CORE NHLBI Pulmonary Vascular Disease Phenomics Program Funded by the National Heart, Lung, and Blood Institute of the National Institutes of Health with support from the Pulmonary

More information

Biology 347 General Physiology Lab Basic Pulmonary Functions: Respirometry and Factors that Effect Respiration

Biology 347 General Physiology Lab Basic Pulmonary Functions: Respirometry and Factors that Effect Respiration Biology 347 General Physiology Lab Basic Pulmonary Functions: Respirometry and Factors that Effect Respiration Objectives Students will measure breathing parameters in a resting subject. Students will

More information

Gas exchange measurement in module M- COVX

Gas exchange measurement in module M- COVX Gas exchange measurement in module M- COVX This document explains how the gas exchange measurement in the gas module M-COVX works. 1. Basic equations To obtain the oxygen consumption VO2 of a patient,

More information

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Lab 1c Isentropic Blow-down Process and Discharge Coefficient 058:080 Experimental Engineering Lab 1c Isentropic Blow-down Process and Discharge Coefficient OBJECTIVES - To study the transient discharge of a rigid pressurized tank; To determine the discharge coefficients

More information

8. Pulmonary Function Testing

8. Pulmonary Function Testing 8. Pulmonary Function Testing Fu 95 Rolf M. Schlegelmilch, Rüdiger Kramme Pulmonary function tests (PFT) examine the functionality of the lungs. This chapter describes the working principles of major test

More information

Figure 1. A schematic diagram of the human respiratory system.

Figure 1. A schematic diagram of the human respiratory system. Introduction to Respiration In this experiment, you will investigate various aspects of normal breathing, hyperventilation, rebreathing the effect of changing airway resistance and ways in which to measure

More information

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE Course n : Course 3 Title: RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE Sub-category: Intensive Care for Respiratory Distress Topic: Pulmonary Function and

More information

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Outline of Presentation Introduction Process Gas Analysis Magnetic Sector Mass Spectrometer Standard

More information

Pulmonary Function I (modified by C. S. Tritt, April 10, 2006) Volumes and Capacities

Pulmonary Function I (modified by C. S. Tritt, April 10, 2006) Volumes and Capacities I. Introduction Pulmonary Function I (modified by C. S. Tritt, April 10, 2006) Volumes and Capacities The volume of air a person inhales (inspires) and exhales (expires) can be measured with a spirometer

More information

Independent Health Facilities

Independent Health Facilities Independent Health Facilities Assessment Protocol for Pulmonary Function Studies - Technologist Assessor INSTRUCTIONS: Please complete ( ) the attached protocol during the assessment. Ensure that all the

More information

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV VENTIlogic LS VENTIlogic plus 100 % Mobility and Reliability in IV and NIV VENTIlogic LS VENTIlogic plus Your requirements for reliability and mobility are our benchmark. VENTIlogic LS and VENTIlogic plus

More information

Experiment HS-2: Breathing and Gravity

Experiment HS-2: Breathing and Gravity Experiment HS-2: Breathing and Gravity Background Over a long time, the amount of oxygen (O 2 ) taken up and carbon dioxide (CO 2 ) given off at the tissues is matched with the amount of O 2 taken up and

More information

iworx Sample Lab Experiment HE-4: Respiratory Exchange Ratio (RER)

iworx Sample Lab Experiment HE-4: Respiratory Exchange Ratio (RER) Experiment HE-4: Respiratory Exchange Ratio (RER) Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to perform the experiment

More information

fabian HFO: Modular, Powerful, Versatile.

fabian HFO: Modular, Powerful, Versatile. fabian HFO: Modular, Powerful, Versatile. Introducing the flagship of the ACUTRONIC FABIAN FAMILY. The FABIAN HFO, a ventilator developed specifically for the NICU and PICU needs, packed with many great

More information

Collin County Community College. Lung Physiology

Collin County Community College. Lung Physiology Collin County Community College BIOL. 2402 Anatomy & Physiology WEEK 9 Respiratory System 1 Lung Physiology Factors affecting Ventillation 1. Airway resistance Flow = Δ P / R Most resistance is encountered

More information

TV = Tidal volume (500ml) IRV = Inspiratory reserve volume (3,000 ml)

TV = Tidal volume (500ml) IRV = Inspiratory reserve volume (3,000 ml) By: Amin alajlouni Lec: 2nd record Date: 29/10/2017 First of all, this is my first sheet so excuse any mistakes I might make and let's start: As we said before in our last lecture about lung capacities

More information

Fashionable, don t you think?

Fashionable, don t you think? Fashionable, don t you think? 1. Passageway 2. Structure 3. Passageway 4. What is the name of the structure labeled with # 9 in the model at left? 5. What is the name of the structure labeled with # 11

More information

O 6 = 6 CO H 2 O H 2 = 16 CO 2. consumed during the oxidation of fatty acids, a ratio of 0.7.

O 6 = 6 CO H 2 O H 2 = 16 CO 2. consumed during the oxidation of fatty acids, a ratio of 0.7. Experiment HE-4: Respiratory Exchange Ratio (RER) Background The two sources of energy available for human metabolism are carbohydrates (CHO) and fats. These molecules are broken down, or catabolized,

More information

Vortex flowmeters. Product family introduction Principle of operation Product review Applications Key product features

Vortex flowmeters. Product family introduction Principle of operation Product review Applications Key product features Vortex flowmeters introduction Product review s Key product features This document should not be duplicated, used, distributed, or disclosed for any purpose unless authorized by Siemens. Page 1 Vortex

More information

Organis TestChest. Flight Simulator for Intensive Care Clinicians

Organis TestChest. Flight Simulator for Intensive Care Clinicians Organis TestChest Flight Simulator for Intensive Care Clinicians Organis TestChest Critical Care challenges Training on ventilation modes with simulation is crucial for patient safety The equipment and

More information

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands Principles of mechanical ventilation Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands Disclosure Research grant Chiesi Pharmaceuticals Research grant CareFusion GA: 27 weeks,

More information

TESTCHEST RESPIRATORY FLIGHT SIMULATOR SIMULATION CENTER MAINZ

TESTCHEST RESPIRATORY FLIGHT SIMULATOR SIMULATION CENTER MAINZ TESTCHEST RESPIRATORY FLIGHT SIMULATOR SIMULATION CENTER MAINZ RESPIRATORY FLIGHT SIMULATOR TestChest the innovation of lung simulation provides a breakthrough in respiratory training. 2 Organis is the

More information

ONSITE PROVING OF GAS METERS. Daniel J. Rudroff WFMS Inc West Bellfort Sugar Land, Texas. Introduction

ONSITE PROVING OF GAS METERS. Daniel J. Rudroff WFMS Inc West Bellfort Sugar Land, Texas. Introduction ONSITE PROVING OF GAS METERS Daniel J. Rudroff WFMS Inc. 13901 West Bellfort Sugar Land, Texas Introduction With the increased use of Natural Gas as a fuel, and higher natural gas prices buyers and sellers

More information

A Pulmonary Function Product of Rare Distinction

A Pulmonary Function Product of Rare Distinction A Pulmonary Function Product of Rare Distinction SpiroAir Fully Digital Positive Displacement PFT System How do you recognize PFT Excellence? The ultimate in volume and flow accuracy Reliable and reproducible

More information

Public Assessment Report Scientific discussion. Lung test gas CO (He) AGA, 0.28%, inhalation gas, compressed (carbon monoxide, helium) SE/H/1154/01/MR

Public Assessment Report Scientific discussion. Lung test gas CO (He) AGA, 0.28%, inhalation gas, compressed (carbon monoxide, helium) SE/H/1154/01/MR Public Assessment Report Scientific discussion Lung test gas CO (He) AGA, 0.28%, inhalation gas, compressed (carbon monoxide, helium) SE/H/1154/01/MR This module reflects the scientific discussion for

More information

Susan Blonshine, RRT, RPFT, AE-C, FAARC

Susan Blonshine, RRT, RPFT, AE-C, FAARC Susan Blonshine, RRT, RPFT, AE-C, FAARC 1 The measurement of TLC to distinguish between restrictive and obstructive processes. The evaluation of the pathophysiology of obstructive lung diseases that may

More information

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates Experiment HE-9: Resting, Active, and Exercising Metabolic Rates Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to

More information

Chapter 53: Clinical Center Single-Breath Carbon Monoxide Diffusing Capacity (DLCO) MOP 53.1 Introduction

Chapter 53: Clinical Center Single-Breath Carbon Monoxide Diffusing Capacity (DLCO) MOP 53.1 Introduction Chapter 53: Clinical Center Single-Breath Carbon Monoxide Diffusing Capacity (DLCO) MOP 53.1 Introduction The DLCO will determine the degree of pulmonary impairment in participants enrolled in the PVDOMICS

More information

New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters.

New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters. New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters. System M Spectrum Medicals total commitment to continuous product improvement is demonstrated

More information

Restrictive and Obstructive Airway Diseases

Restrictive and Obstructive Airway Diseases iworx Physiology Lab Experiment Experiment HS-8-1 Restrictive and Obstructive Airway Diseases Note: The lab presented here is intended for evaluation purposes only. iworx users should refer to the User

More information

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV VENTIlogic LS VENTIlogic plus 100 % Mobility and Reliability in IV and NIV VENTIlogic VENTIlog LS Your Requirements for Reliability and Mobility are Our Benchmark. VENTIlogic LS and VENTIlogic plus are

More information

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow Georgia Institute of Technology School of Earth and Atmospheric Sciences EAS 4641, Spring 2008 Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow Purpose of Lab 3: 1) To gain a

More information

Electric Blower Based Portable Emergency Ventilator

Electric Blower Based Portable Emergency Ventilator Electric Blower Based Portable Emergency Ventilator Mike Fogarty Department of Bioengineering, University of Utah Introduction When a patient is undergoing general anesthesia he/she will most likely be

More information

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111 INTRODUCTION Water vapor in natural gas has more than a substantial effect on the quality of the gas stream. Without quality measurement of water vapor the gas is basically not saleable. Contracts are

More information

Let s talk about Capnography

Let s talk about Capnography Let s talk about Capnography This is one of a series of articles by Keith Simpson BVSc MRCVS MIET (Electronics) discussing the practical aspects of some common monitoring techniques. Capnometry is the

More information

INTENSIVE CARE VENTILATORS

INTENSIVE CARE VENTILATORS INTENSIVE CARE VENTILATORS iternis ADV INTENSIVE CARE VENTILATION FOR ALL PATIENT CATEGORIES Design Features HEYER iternis ADV s 12-inch touch-screen display combines comprehensive, easy-to-follow user

More information

Capnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA

Capnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA Capnography in the Veterinary Technician Toolbox Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA What are Respiration and Ventilation? Respiration includes all those chemical and physical

More information

Introduction. Respiration. Chapter 10. Objectives. Objectives. The Respiratory System

Introduction. Respiration. Chapter 10. Objectives. Objectives. The Respiratory System Introduction Respiration Chapter 10 The Respiratory System Provides a means of gas exchange between the environment and the body Plays a role in the regulation of acidbase balance during exercise Objectives

More information

Blease900 Ventilator PATIENT CENTERED VENTILATION

Blease900 Ventilator PATIENT CENTERED VENTILATION Blease900 Ventilator PATIENT CENTERED VENTILATIN PATIENT CENTERED VENTILATIN From the first Blease/Manley ventilator sixty years ago, Spacelabs Healthcare has built a heritage of innovative and intuitive

More information

Part 1: Inspiratory and expiratory pressures

Part 1: Inspiratory and expiratory pressures BIOEN 327 Autumn 2013 Experimental procedures Throughout these experiments, record in your notebook the purpose of the experiments, the methods you used, and the results. Where possible, make predictions

More information

UNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services

UNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services UNDERSTANDING NEONATAL WAVEFORM GRAPHICS Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services Disclosures Purpose: To enhance bedside staff s knowledge of ventilation and oxygenation

More information

Gas Measurement Fundamentals Certification. Curriculum

Gas Measurement Fundamentals Certification. Curriculum Gas Certification Institute, LLC P.O. Box 131525 Houston, Texas 77219-1525 281-598-7200 Phone 281-598-7199 Fax contact@gascertification.com www.gascertification.com Curriculum Copyright 2007-2012 Gas Certification

More information

Level MEASUREMENT 1/2016

Level MEASUREMENT 1/2016 Level MEASUREMENT 1/2016 AGENDA 2 A. Introduction B. Float method C. Displacer method D. Hydrostatic pressure method E. Capacitance method G. Ultrasonic method H. Radar method I. Laser method J. Level

More information

Astral in AirView: Improving patient care through connectivity. ResMed.com

Astral in AirView: Improving patient care through connectivity. ResMed.com Astral in AirView: Improving patient care through connectivity ResMed.com Using Astral in AirView via the ResMed Connectivity Module (RCM) Astral is ResMed s portable, invasive and non-invasive life support

More information

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features) The Crossvent 2i+ 1. How is this ventilator classified 2. Ventilator Concept (brief theory of operation and features) -Your Two Choices with this Ventilator 3. An overview of the device (an in-service)

More information

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates Experiment HE-9: Resting, Active, and Exercising Metabolic Rates Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to

More information

Physiology of the Respiratory System

Physiology of the Respiratory System Biology 212: Anatomy and Physiology II Physiology of the Respiratory System References: Saladin, KS: Anatomy and Physiology, The Unity of Form and Function 8 th (2018). Required reading before beginning

More information

Cornell Institute for. Biology Teachers. Respirometry Part I: Lung Volumes and Capacities. Lab issue/rev. date: 12/12/96. Title:

Cornell Institute for. Biology Teachers. Respirometry Part I: Lung Volumes and Capacities. Lab issue/rev. date: 12/12/96. Title: Cornell Institute for Biology Teachers Copyright Cornell Institute for Biology Teachers, 1999. This work may be copied by the original recipient from CIBT to provide copies for users working under the

More information

HYDRAULICS. H89.8D - Hydraulic Bench

HYDRAULICS. H89.8D - Hydraulic Bench HYDRAULICS H89.8D - Hydraulic Bench 1. General The H89.8D and ancillary equipment have been developed to provide a comprehensive range of experiments in fluid mechanics. The bench is of robust construction

More information

Intensa GO Alternate Care Ventilator

Intensa GO Alternate Care Ventilator EN Intensa GO Alternate Care Ventilator 0197 Intensa GO Versatility and simplicity The all new WILAmed Ventilator INTENSA Go offers great versatility and simplicity. It is designed to ventilate (invasive

More information

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR)

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR) Experiment HE-5: Resting Metabolic Rate (RMR) Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to perform the experiment

More information

A bench study comparison of volume delivery in the presence of leak

A bench study comparison of volume delivery in the presence of leak A bench study comparison of volume delivery in the presence of leak Study objective To compare the volume delivery of the Trilogy1 ventilator to the volume delivery of the LTV 1 as a representative of

More information

Breathing of Humans and its Simulation

Breathing of Humans and its Simulation Breathing of Humans and its Simulation Mina Nishi LSTM-Erlangen Institute of Fluid Mechanics Friedlich-Alexander-University Erlangen-Nuremberg Cauerstr.4, D-91058 Erlangen June 14, 2004 Abstract In this

More information

Effect of noise in the performance of the transducers in an ultrasonic flow meter of natural gas

Effect of noise in the performance of the transducers in an ultrasonic flow meter of natural gas 23rd World Gas Conference, Amsterdam 2006 Effect of noise in the performance of the transducers in an ultrasonic flow meter of natural gas Main author R. Villarroel Venezuela TABLE OF CONTENTS 1. Abstract

More information

Peter Kremeier, Christian Woll. 2nd. Understanding and comparing modes of ventilation. The Kronberg List of Ventilation Modes

Peter Kremeier, Christian Woll. 2nd. Understanding and comparing modes of ventilation. The Kronberg List of Ventilation Modes Extended Edition Revised and 2nd Peter Kremeier, Christian Woll Understanding and comparing modes of ventilation The Kronberg List of Ventilation Modes Contents Preface... 3 Comparison Table - Parameters...6

More information

Stephanie. Pediatrics. The ventilation system for neonatology

Stephanie. Pediatrics. The ventilation system for neonatology Pediatrics Stephanie The ventilation system for neonatology + Conventional ventilator + High-frequency oscillator + Proportional Assist Ventilation (PAV) + Integrated patient gas humidifier + Minimum volume

More information

Instruction Manual. Pipe Friction Training Panel

Instruction Manual. Pipe Friction Training Panel Instruction Manual HL 102 Pipe Friction Training Panel 100 90 80 70 60 50 40 30 20 10 HL 102 Instruction Manual This manual must be kept by the unit. Before operating the unit: - Read this manual. - All

More information

Yokogawa Systems and PCI Training

Yokogawa Systems and PCI Training Yokogawa Systems and PCI Training 09 th December 2014 Nico Marneweck 082 883 2652 Kallie Bodenstein 083 226 2787 Marco Coccioni 072 409 5779-1 - Introduction Training customised for the South Africa Industrial

More information

MEDUMAT Transport. High-End Ventilation in Any Situation

MEDUMAT Transport. High-End Ventilation in Any Situation MEDUMAT Transport High-End Ventilation in Any Situation MEDUMAT Transport High-End Ventilation in Any Situation We worked in close cooperation with our customers and partners to develop MEDUMAT Transport

More information

Astral in AirView: Improving patient care through connectivity. ResMed.com

Astral in AirView: Improving patient care through connectivity. ResMed.com Astral in AirView: Improving patient care through connectivity ResMed.com This guide will assist you with: Setting up the ResMed Connectivity Module for Astral 2 Troubleshooting the ResMed Connectivity

More information

April KHALED MOUSA BACHA. Physiology #2. Dr. Nayef AL-Gharaibeh. Pulmonary volumes & capacities

April KHALED MOUSA BACHA. Physiology #2. Dr. Nayef AL-Gharaibeh. Pulmonary volumes & capacities 25 th April Physiology #2 Pulmonary volumes & capacities Dr. Nayef AL-Gharaibeh KHALED MOUSA BACHA We will start this lecture by explaining an important concept from the previous one: Intrapleural pressure

More information

Application Note 268 Metabolic Analysis Systems and Propane-based Verification

Application Note 268 Metabolic Analysis Systems and Propane-based Verification Application Note 268 Metabolic Analysis Systems and Propane-based Verification 8.10.11 This application note examines a conventional equipment setup for metabolic measurements. The setup is exposed to

More information

ADVANCED SUBSIDIARY GCE G622 APPLIED SCIENCE

ADVANCED SUBSIDIARY GCE G622 APPLIED SCIENCE ADVANCED SUBSIDIARY GCE APPLIED SCIENCE Unit 3: Monitoring the activity of the human body TUESDAY 5 JUNE 2007 G622 Afternoon *CUP/T21601* Additional materials: Electronic calculator Ruler (cm/mm) Time:

More information

LAB 7 HUMAN RESPIRATORY LAB. Complete the charts on pgs. 67 and 68 and read directions for using BIOPAC

LAB 7 HUMAN RESPIRATORY LAB. Complete the charts on pgs. 67 and 68 and read directions for using BIOPAC 66 LAB 7 HUMAN RESPIRATORY LAB Assignments: Due before lab: Quiz: Three Respiratory Interactive Physiology Animations pages 69 73. Complete the charts on pgs. 67 and 68 and read directions for using BIOPAC

More information

HAMILTON-C3. The compact high-end ventilator

HAMILTON-C3. The compact high-end ventilator The compact high-end ventilator We live for ventilation technology We live for ventilation technology. Technology that helps caregivers improve the lives of their critically ill patients. We believe that

More information

Rad-57 Pulse CO-Oximeter

Rad-57 Pulse CO-Oximeter Pulse co-oximeter Rad-57 Pulse CO-Oximeter Helps detect methemoglobinemia and carbon monoxide poisioning on-site in seconds PeRFoRMance Specifications measurement range Oxygen Saturation (SpO2)........................................

More information

BiPAP Vision Ventilator Testing with Various Patient Interfaces Report February 18, 2002

BiPAP Vision Ventilator Testing with Various Patient Interfaces Report February 18, 2002 BiPAP Vision Ventilator Testing with Various Patient Interfaces Report 48086 February 18, 2002 Prepared for ResMed by Valley Inspired Products, LLC inspiredrc.com Valley Inspired Products, LLC Revision

More information

Flight Medical presents the F60

Flight Medical presents the F60 Flight Medical presents the F60 Reliable Ventilation Across the Spectrum of Care Adult & Pediatric Pressure/Volume Control Basic/Advanced Modes Invasive/NIV High Pressure/Low Flow O2 Up to 12 hours batteries

More information

Mass Flow Controller (MFC) for Gases

Mass Flow Controller (MFC) for Gases Mass Flow Controller (MFC) for Gases Type 8713 can be combined with... Direct flow measurement by MEMS- Technology for nominal flow rates from 1 ml N /min to 8 l N /min (N 2 ) High accuracy and repeatability

More information

EVE. The mobile Ventilation Concept from Emergency to Intensive Care. easy - light - smart. Ventilation

EVE. The mobile Ventilation Concept from Emergency to Intensive Care. easy - light - smart. Ventilation Ventilation EVE easy - light - smart The mobile Ventilation Concept from Emergency to Intensive Care + Ventilation at the push of a button + Cutting edge turbine technology + Critical care during transport

More information

Breeze Suite Quick Start Operator Manual. An Introduction to Pulmonary Function Testing

Breeze Suite Quick Start Operator Manual. An Introduction to Pulmonary Function Testing Breeze Suite Quick Start Operator Manual An Introduction to Pulmonary Function Testing The MedGraphics Breeze Suite Quick Start Operator Manual is Part No. 142143-001 Rev A. Caution: Federal (USA) law

More information

Oxylog VE300 Emergency & Transport Ventilation

Oxylog VE300 Emergency & Transport Ventilation Oxylog VE300 Emergency & Transport Ventilation The straightforward and user-friendly Dräger Oxylog VE300 is built to face your challenges in preclinical emergency services With reliable ventilation technology,

More information

A NEW METHOD FOR ASSURING ALVEOLAR-EQUILIBRATED BREATH ALCOHOL SAMPLES. A. Slemeyer, Ph.D.* SYNOPSIS

A NEW METHOD FOR ASSURING ALVEOLAR-EQUILIBRATED BREATH ALCOHOL SAMPLES. A. Slemeyer, Ph.D.* SYNOPSIS A NEW METHOD FOR ASSURING ALVEOLAR-EQUILIBRATED BREATH ALCOHOL SAMPLES A. Slemeyer, Ph.D.* SYNOPSIS Reproducibility is one of the most important figures of merit in breath alcohol analysis. A number of

More information

2. Determine how the mass transfer rate is affected by gas flow rate and liquid flow rate.

2. Determine how the mass transfer rate is affected by gas flow rate and liquid flow rate. Goals for Gas Absorption Experiment: 1. Evaluate the performance of packed gas-liquid absorption tower. 2. Determine how the mass transfer rate is affected by gas flow rate and liquid flow rate. 3. Consider

More information

Code Basic module and level control complete with optionals code

Code Basic module and level control complete with optionals code AUTOMATION AND PROCESS CONTROL MPCT - Modular Process Control Trainer Code 916960 Basic module and level control complete with optionals code 916961-916962-916963-916964-916965 1. General Didacta proposes

More information

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Background There are two main sources of energy available for animal metabolism: carbohydrates (CHO) and fats. These molecules are broken

More information

FIG: 27.1 Tool String

FIG: 27.1 Tool String Bring up Radioactive Tracer service. Click Acquisition Box - Edit - Tool String Edit the tool string as necessary to reflect the tool string being run. This is important to insure proper offsets, filters,

More information

WeatherStation Monitoring for Industrial marine Applications

WeatherStation Monitoring for Industrial marine Applications WeatherStation Monitoring for Industrial marine Applications PB200 WeatherStation Instrument Applications: Industrial marine weather monitoring Features: True and Apparent Wind Speed and Direction Barometric

More information