1 HPICAL Operation & Data Logging Procedures Click spacebar to advance through slides 1
2 WARNING Always wear proper safety equipment when using high pressure equipment. Do not exceed 125 psi air pressure. Never exceed the maximum pressure rating of the gauge being tested. Be sure pressures on air and water systems are relieved before disconnecting any hoses or fittings. Pressurization of any materials is dangerous follow all OSHA procedures for stored energy and pressurization precautions pertaining to the fluids used. Read and understand all instructions before performing a test/calibration. 2
3 Control & Feature Overview Air Regulator System Pressure & Data Logging Gauge Auxiliary Test Port Pressure Adjust Valve Air On-Off Valve Bleed Valve Water On-Off Valve Main Test Port Air Pressure Gauge 3
4 Before introducing air or water to the system, turn air and water valves to OFF Rotate Air Regulator Knob Counter Clock-wise to Off = Open Position Turn Both Air and Water Valves Clock-wise to Off Position 4
5 Water, Air and Power Connections Connect Air Line Here (3/8 NPT)* 90 PSI Air Supply Required No lubrication is required, and should be avoided Plug gauge AC adapter into power strip Connect Water Line Here (1/2 NPT)* Clean water supply required at normal street pressure. NOTE: System contains a 40 micron in-line strainer (PC-1039), which can be replaced as necessary. Connect to a drain if desired, or use supplied drain bottle** *NPT fittings should be installed with Teflon (PTFE) tape. Use 1 ½ - 2 wraps of tape, then tighten 1 ½ - 3 turns past hand tight. **If using the supplied drain bottle, be sure to empty it periodically Plug AC cord into standard 120V outlet 5
6 Additel Digital Logger Gauge Overview PEAK Button POWER Button ZERO Button ANALOGUE DIAL Button UNITS Button BACKLIGHT Button 6
7 Setting Date and Time 1. Press Power button to turn on gauge. 2. Press and hold the PEAK button to enter main menu. 3. Press the ZERO button to enter Logging menu (Option 1). 4. Press the ZERO button again to enter date and time settings (option 1). 5. Press the ZERO button again to begin setting year. 6. Press the UNITS, PEAK and BACKLIGHT buttons to change the year, month, day and time. 7. When all are set correctly, press the ZERO button to accept the changes. 8. Press the POWER button to escape to the main screen. 9. NOTE: If the gauge loses power (unplugged from AC power), the date and time will reset to the factory set date and time, and must be re-set. 7
8 Connecting to Remote Device Using Main Test Port Remove Adapter Cap (PC-1032) from Main Test Port Adapter (PC-1029) and store safely. Attach green Test Hose (PC-1019) to Main Test Port Adapter (PC-1029) and tighten firmly. Test Hose (PC-1019) Main Test Port Adapter (PC-1029) Adapter Cap (PC-1032) Install Auxiliary Test Port Plug (PC-1031) and Gland Nut (PC ) into Auxiliary Test Port and tighten firmly. Gland Nut (PC ) Auxiliary Test Port Plug (PC-1031) 8
9 Connecting to Remote Device Using Main Test Port (Cont d) Be sure Isolation Valve (if present) is fully closed. Install appropriate adapter for Test Hose to remote device being tested and attach Test Hose (PC-1019).* (HPIC D unit shown being tested for reference.) *The Test Hose (PC-1019) has standard 9/16 -M hose fittings at each end. A ½ NPT male to 9/16 -M Adapter (PC-1030) is supplied. Other adapters are available upon request. NPT fittings should be installed with Teflon (PTFE) tape. Use 1 ½ - 2 wraps of tape, then tighten 1 ½ - 3 turns past hand tight. ½ NPT to 9/16-M Adapter (PC-1030) 9
10 Connecting to Gauge Using Auxiliary Test Port Remove Auxiliary Test Port Plug (PC-1031) and Gland Nut (PC ) from Auxiliary Test Port and store safely. Install appropriate adapter for the gauge being tested.* Main Test Port Adapter (PC-1029) NOTE: A ¼ NPT Female Adapter* (PC /4NPT) is supplied. Other adapters are available upon request. Remove Test Hose (PC-1019) from Main Test Port and store safely. Install Adapter Cap (PC-1032) onto Main Test Port Adapter (PC-1029) and tighten firmly. Install gauge to be tested. Adapter Cap (PC-1032) *NPT fittings should be installed with Teflon (PTFE) tape. Use 1 ½ - 2 wraps of tape, then tighten 1 ½ - 3 turns past hand tight. 10
11 Bleeding the System When Testing a Gauge directly Using the Auxiliary Test Port Be sure the Pressure Adjust Valve is backed all the way off (counter clockwise) 1. Open the Bleed Valve (only 1/4-1/2 of a turn is necessary) 2. Zero the Calibration Station gauge (Press and hold ZERO for 3 seconds) 3. Zero the gauge being tested (if applicable) 4. DATA LOGGING should be started now (see page 14) 5. Turn the Water Valve to the ON position 6. Let the water run through the system for about 5-10 seconds, until full. 7. Close the Bleed Valve firmly 8. Open the Bleed valve again and repeat steps 6 &7 to ensure all air is removed from the system 11
12 Bleeding the System When Testing a Remote Device Using the Main Test Port Be sure the Pressure Adjust Valve is backed all the way off (counter clockwise) 1. Open the Bleed Valve (1/4-1/2 turn) on the Calibration Station 2. Open the Bleed Valve on the device being tested 3. Zero the Calibration Station gauge (Press and hold ZERO for 3 seconds) 4. Zero the gauge being tested (if applicable) 5. DATA LOGGING can also be started now (see page 14) 6. Turn the Water Valve to the ON position 7. Let the water run through the system for 5-10 seconds, until full 8. Close the Bleed Valve firmly on the device being tested 9. Close the Bleed Valve firmly on the Calibration Station 10. Open both Bleed Valves again and repeat steps 7,8 & 9 to ensure all air is removed from the system (HPIC-1000-D unit shown for reference) 12
13 Building Pressure 1. Follow all previous instructions on connecting air and water inputs, connecting a gauge directly or a remote device and bleeding the system. 2. Be sure the regulator is backed all the way off (counter clockwise) 3. Turn the Air Valve to the ON position 4. SLOWLY turn the air regulator clockwise to increase the system pressure 5. Increase the pressure slowly until it reaches PSI BELOW the desired test pressure 6. Back the regulator off COMPLETELY (counter clockwise) NOTE: this will cause the system pressure to drop slightly due to the check valves seating 7. Turn the Pressure Adjust Valve clockwise to raise the pressure to the exact desired test pressure 8. Turn the Pressure Adjust Valve counter clockwise to reduce pressure 9. To relieve system pressure: Back the Pressure Adjust Valve all the way out (counterclockwise), then open Bleed Valve slightly to relieve pressure. Close Bleed Valve firmly and repeat steps 4 8 to build pressure to next desired test pressure. 10. When testing is complete, back Pressure Adjust Valve all the way out (counter clockwise) Turn air and water valves to the OFF position and open bleed valve(s) to relieve system pressure 13
14 General Data Logging Instructions 1. Press and hold the PEAK button to reach the main menu. 2. Press the ZERO button to enter Logging menu (Option 1). 3. Press the BACKLIGHT button to scroll down to the GAP setting (Option 5). The GAP is the time interval between readings recorded by the gauge. The default GAP is 1 second. If a different GAP is desired, it must be set every time a new session is started by following steps Press the UNITS button to move over until the correct digit is flashing. 5. Press the PEAK and BACKLIGHT buttons to move the interval up or down. 6. Press the ZERO button to accept your set interval. 7. Press the BACKLIGHT button to scroll down to the OFF setting (Option 6). If LOGGING is active, this option will read ON 8. Press the ZERO button to turn on Logging. (At this point the % should be flashing to indicate Logging is active) 9. Press the POWER button twice to escape to the main screen. 10. To turn off Logging, repeat steps 1,2,7 and 8. 14
15 Inspection vs. Calibration Inspection is performed to verify that the gauge is working properly and is within the accuracy percentage range. In the case of the Ashcroft gauge in the HPIC D, the accuracy is ±0.25% of full scale, or ±25 PSI. Calibration is performed if the tested gauge is outside of its specified range (verified by the Calibration Station). 15
16 General Gauge Inspection Procedure Follow instructions for attaching a device or gauge to the Calibration Station (see pages 8-10). Follow instructions for bleeding the system (see pages 11 & 12). Build the system pressure to half of the full range of the gauge (5000 PSI for a PSI gauge) following the Building Pressure instructions (see page 13). Compare the pressure on device or gauge being tested to the gauge on the Calibration station. If it is within the manufacturer s intended accuracy percentage range, it does not need to be calibrated. If the gauge is not within range, it will need to be calibrated according to the manufacturer s instructions. 16
17 General Gauge Calibration Procedure Follow instructions for attaching a device or gauge to the Calibration Station (see pages 8-10). Follows instructions for bleeding the system (see pages 11 & 12). Be sure DATA LOGGING is active (flashing % on Additel gauge; see page 14). Follow the manufacturer s instructions for calibration of the gauge. See the following slides for instructions on calibrating the Ashcroft gauge standard in the HPIC D. After calibration is complete the DATA LOGGING can be turned off (see page 14). Turn Air and Water valves to the OFF positions. Be sure regulator is backed all the way off (counterclockwise). Open the Bleed Valve and relieve the system of all pressure At this point the data obtained can be uploaded to the laptop supplied using the Additel/Land software and used to create a calibration report (see pages 24-28). 17
18 Ashcroft Gauge Calibration ON/OFF ZERO/CLEAR MENU ENTER ARROW DOWN ARROW UP 18
19 HPIC D Unit/Ashcroft Gauge Calibration- Step 1 1. Press the MENU button 2. Press the (up arrow key) or (down arrow key) on the keypad until the word CONFIG appears. 3. Press ENTER 19
20 Ashcroft Gage Calibration- Step 2 4. Enter user password if it has been programmed. 5. Press (up arrow key) or (down arrow key) until the word RECAL appears. 6. Press ENTER. 7. The gauge will show 0 on the top line and flash between INPUT and unit of measure on the lower line. Apply zero pressure to the gauge. 8. Press ENTER. Zero pressure is now set. 20
21 Ashcroft Gage Calibration- Step 3 9. The gauge will display full-scale pressure (10000 PSI). Apply full-scale pressure to the gauge. (see page 13) 10. Press ENTER. Full-scale pressure is now set. 11. The gauge will now display mid-scale pressure (5000 PSI). Apply mid-scale pressure to the gauge. (see page 13) 12. Press ENTER. Mid-scale pressure is now set. 21
22 Ashcroft Gage Calibration - Finish After zero, full-scale and mid-scale calibration have been set, the word SAVE appears on the gauge display. Press ENTER to finalize calibration. 22
23 General System Shutdown Procedure After all desired tests have been performed: Be sure Data Logging is turned off. (no flashing %) Turn Air and Water valves OFF. Be sure regulator is backed all the way off (counter clockwise). Open the Bleed Valve to relieve the system of all pressure. 23
24 Uploading Data to Laptop 1. Connect the gauge to the USB port using the included adapter cable. 2. Open the Additel/Land software on the laptop. 3. Click the scan button to search for the gauge. 4. Click on the gauge when it appears. 5. Click the DATA LOGGING button to access the data 6. Select desired units for pressure and temperature. 7. Click UPLOAD. 24
25 Exporting Data The data that has been uploaded can now be viewed in a graph in the Additel/Land software or exported to an Excel Spreadsheet. The only way to save the data is to export it to Excel. To do this, simply click the export button. You will be prompted to save the file. After saving, the file will automatically open in Excel. Do not close the Additel/Land window. 25
26 Creating a Certificate of Calibration Open the Certificate of Calibration Template. Only certain fields in this file can be edited. Input the Additel gauge serial number, your name, and the model and serial number of the gauge/device being tested. 26
27 Certificate Continued Go back to the exported Excel sheet and copy the points at 0 PSI, PSI and 5000 PSI as shown. Go back to the Certificate and click on the first cell under Results. Paste the values using Cntrl+C. The graph will automatically update to the values pasted. You can now print and SAVE AS. 27
28 Deleting Data on Gauge After all data is exported AND SAVED, the data can now be deleted from the gauge. Go back to the Additel/Land window. Click the Delete button. Input the password (211) for the Additel gauge and select OK. You can now start another logging session. 28
29 If there are any questions, concerns or comments do not hesitate to contact: HPI Manufacturing, Inc. (203)
WIDDER TOOLS. HPICAL Calibration Station 1 (203)
WIDDER TOOLS HPICAL-15000 Calibration Station WWW.WIDDERTOOLS.COM 1 1. TEST SET-UP a. Be sure air and water supply lines are installed per basic set-up & installation instructions supplied with calibration
More informationPegas 4000 MF Gas Mixer InstructionManual Columbus Instruments
Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...
More informationHPIC D Hydrostatic Test Pump PRODUCT INFORMATION AND OPERATING INSTRUCTIONS
WIDDER TOOLS HPIC-10000-D Hydrostatic Test Pump PRODUCT INFORMATION AND OPERATING INSTRUCTIONS Description: The WIDDER Hydrostatic Test Systems C Series is a portable, selfcontained, air driven hydrostatic
More informationAir Bubbler Depth Gauge DG2200 Installation and Reference Manual
Air Bubbler Depth Gauge DG2200 Installation and Reference Manual Rev. 3.2.2 06/2011 Installation and Reference Manual Page 2 Contents Chapter 1: Installation and Overview... 3 1.1 DG2200 Location Diagram...
More informationNITROGEN INFLATION SYSTEM NG-4, NG-6, NG-12, NG-18
NITROGEN OPERATING INSTRUCTIONS NITROGEN INFLATION SYSTEM NG-4, NG-6, NG-12, NG-18 READ INSTRUCTIONS THOROUGHLY BEFORE OPERATING 3451 S. 40th Street Phoenix, AZ 85040 602.437.5020 800.223.4540 www.tsissg.com
More informationELECTRONIC PLC SERIES Test Benches
Whether you are producing a single assembly or a production run, CustomCrimp has the test equipment to give both you and your customer the assurance that the product has been assembled correctly. A programmable
More informationUser Manual for the Mars Calibration Bench
User Manual for the Mars Calibration Bench Fall 2013 Table of Contents Table of Contents Table of Contents... iii Introduction... v Chapter 1: The Mars Calibration Bench... 1 What Is the Mars Calibration
More informationInstructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems
Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the
More informationSection 1: Overview... 6 Features... 6 Supported Software Part 1: Getting Started 7
Table of Contents About this Manual... 3 Intended Audience... 3 How this Manual is Organized... 4 Conventions Used in this Manual... 5 Related Information... 5 Where To Get Help... 5 Section 1: Overview...
More informationRejuvenation Instructions
Rejuvenation Instructions #401 Air Systems UPR This NRI covers the following: Understanding the applications and operation of flow meters. Understand the application and operation of test pressure gauges.
More informationMANUAL BE SERIES Test Benches
The CustomCrimp Manual BE Series Test Benches are designed with features that make proof and burst testing of hydraulic hose assemblies a quick and easy procedure. CUSTOMIZED AND SPECIAL DESIGN BENCHES
More informationFireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS
FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS WARNING THE WARRANTIES MADE BY MSA WITH RESPECT TO THE PRODUCT ARE VOIDED IF THE PRODUCT IS NOT USED AND MAINTAINED IN ACCORDANCE
More informationPressure Differential Automatic Manifold For LPG. PDH1000 Series. LPG Installation & Operations Instructions
Pressure Differential Automatic Manifold For LPG PDH1000 Series LPG Installation & Operations Instructions Precise Equipment Company 3822 Market St., STE 101 Denton, Texas Phone : 800.795.8388 Precise
More informationSuperconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)
Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures
More informationNB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE
NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE INSTALLATION, OPERATION & MAINTENANCE MANUAL INTERFACE DEVICES, INC. 230 Depot Road, Milford, CT 06460 Ph: (203) 878-4648, Fx: (203) 882-0885, E-mail: info@interfacedevices.com
More informationGETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP
GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP LIMITED WARRANTY Getz Equipment Innovators warrants its products, and component parts of any product manufactured by
More informationChromat Calibration Updated October 27th, 2017
Chromat Calibration Updated October 27th, 2017 Calibrating the Chromatograph Performing the Chromat Calibration is highly recommended when starting a new well. The MLogger already has a default calibration
More informationIn Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure
PPMS Service Note 1099-412 In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure Introduction: Loss of electricity to the PPMS EverCool II should not cause damage to
More informationKEM Scientific, Inc. Instruments for Science from Scientists
KEM Scientific, Inc. Instruments for Science from Scientists J-KEM Scientific, Inc. 6970 Olive Blvd. St. Louis, MO 63130 (314) 863-5536 Fax (314) 863-6070 E-Mail: jkem911@jkem.com Precision Vacuum Controller,
More informationUser Instruction Manual
User Instruction Manual 4500 psi Air Compressor Ver 2, 1.18 Contents Parts Included...3 Assembly Instructions...3-5 Operation Instructions...6-7 Oil Change Intervals...8 Air Filter Replacement...9 Setting
More informationInstallation Operation Maintenance. Bermad Level Control Valve with Modulating Horizontal Float Pilot valve One Way Flow IOM.
Bermad Level Control Valve with Modulating Horizontal Float Pilot valve One Way Flow Model: FP 450-80 Installation Operation Maintenance PAGE 1 OF 5 1. Safety First BERMAD believes that the safety of personnel
More informationRunning the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached
Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached Note: You must read the Primer manual Sections 1-6 before using the IRGA. The manual is located in the wall cabinet
More informationTR Electronic Pressure Regulator. User s Manual
TR Electronic Pressure Regulator Page 2 of 13 Table of Contents Warnings, Cautions & Notices... 3 Factory Default Setting... 4 Quick Start Procedure... 5 Configuration Tab... 8 Setup Tab... 9 Internal
More informationINSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y
INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50 Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y OFFICE: (507) 843-4700 P.O. BOX 7079 Toll
More information3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289
3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS Item #31289 The EASTWOOD 3 GALLON, OILLESS PANCAKE COMPRESSOR, with an Integral Air Regulator, efficiently supplies all compressed air requirements for
More informationHand-held pressure calibrator with integrated pump Model CPH6600
Calibration technology Hand-held pressure calibrator with integrated pump Model CPH6600 WIKA data sheet CT 16.01 Applications Calibration service companies and service industry Measurement and control
More informationTouch Screen Guide. OG-1500 and OG Part # T011
Touch Screen Guide OG-1500 and OG-2000 Part # 9000000.T011 Effective 11/2010 External View Internal View 1. Transducer Banks 2. Oxygen Sensor 3. PLC These are the two manifolds with three (3) transducers
More informationOperation Manual Piston Sensed Gas Pressure Regulators
687 Technology Way Napa, CA 94558 Phone: (707) 259-0102 FAX: (707) 259-0117 www.aptech-online.com Operation Manual Piston Sensed Gas Pressure Regulators (Models KT9, KT10, Welded KT10, KT12) Table of Contents:
More informationSENSUS PRO MANAGER (for SENSUS or SENSUS PRO devices) User s Guide -- Windows. Version 2.0 Published October 17, ReefNet Inc.
SENSUS PRO MANAGER (for SENSUS or SENSUS PRO devices) User s Guide -- Windows Version 2.0 Published October 17, 2002 2002 ReefNet Inc. 1.0 Introduction The SENSUS PRO data recorder captures the time, depth,
More informationFencing Time Version 4.3
Fencing Time Version 4.3 Upgrading your Fencing Time Server October 2017 Copyright 2017 by Fencing Time, LLC. All rights reserved. Overview Periodically, a new version of Fencing Time is released. In most
More informationTechnology. Using Bluetooth
Bluetooth is a communication technology that makes it possible to send and receive data without using wires. Using the Bluetooth features, you can set up a wireless connection between your NXT and other
More informationInstruction Manual Updated 7/26/2011 Ver. 2.2
4-Unit Model MB HTHP Filter Press #171-50-4: 115-Volt #171-51-4: 230-Volt Instruction Manual Updated 7/26/2011 Ver. 2.2 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. Houston, Texas 77065 U.S.A. Tele:
More informationTypes 749B and R130 Changeover Manifolds
Instruction Manual MCK-1179 Types 749B and R130 June 2012 Types 749B and R130 Changeover Manifolds TYPE HSRL-749B TYPE 64SR/122 TYPE R130/21 TYPE 749B/21 Figure 1. Changeover Manifolds and Regulator Assemblies
More information256 Pneumatic Pressure Indicator
256 Pneumatic Pressure Indicator 51425699 Copyright 2002 Slope Indicator Company. All Rights Reserved. This equipment should be installed, maintained, and operated by technically qualified personnel. Any
More informationInstalling ASI Umbilical Delivery Systems
Rev A 05 Nov 2015 Page 1 of 10 PURPOSE This guide provides an overview of tubing connections into an ASI wall-mounted Junction Box that has already been plumbed with a compressed air connection including
More informationMedical Gas Control Panel
Medical Gas Control Panel INSTALLATION & MAINTENANCE MANUAL 255105 (Rev. 6) 2018-04 2 255105 (Rev. 6) 2018-04 1.0 User Responsibilities The information contained in this installation and operation maintenance
More informationSpeed control assembly model A-1
Pressure Regulation 533a 1. DESCRIPTION The Viking Speed Control Assembly provides adjustment of the opening speed of Viking Deluge Valves, and adjustment of both the opening and closing speed of Viking
More informationModel Air Caddy Orig TABLE OF CONTENTS
92-0919 Orig. 020201 Model 75-0114 Air Caddy TABLE OF CONTENTS CUSTOMER MESSAGE Inside Front Cover SAFETY PRECAUTIONS 3 GENERAL DESCRIPTION 6 SPECIFICATIONS 7 MAINTENANCE 9 INSTALLATION AND OPERATION 11
More informationOperating Instructions Model and Hydrostatic Test Pump
Operating Instructions Model 29200 and 2920 Hydrostatic Test Pump Dimension Weight Pump Pressure for 29200 Pump Pressure for 2920 Gauge for 29200 Gauge for 2920 Inlet Connection Outlet Connection Hose
More informationPrecision digital pressure gauge Model CPG1000
Calibration technology Precision digital pressure gauge Model CPG1000 WIKA data sheet CT 10.01 Applications Oil and gas industry Maintenance and service facilities Calibration service companies and service
More informationQuickstart Installation Checklist (Please refer to operation manual for complete installation instructions)
Quickstart Installation Checklist (Please refer to operation manual for complete installation instructions) 1. Uncrate blender from packaging. 2. Lift the blender and position over machine throat or floor
More informationLaboratory Mortar Mixer (Testing)
TomTom-Tools GmbH Zelgli 20 8905 Arni info@tomtom-tools.com Switzerland www.tomtom-tools.com User Manual Version February 22, 2015 Laboratory Mortar Mixer (Testing) 1 Introduction The Laboratory Mortar
More informationTechcon Systems TS1258 Pressure Pot
Techcon Systems TS1258 Pressure Pot User Guide TABLE OF CONTENT 1. SAFETY. 3 1.1 Intended Use 3 1.2 Safety Precaution 3 2. CERTIFICATE OF COMFORMANCE 4 3. FEATURES. 5 4. SPECIFICATIONS 5 5. INSTALLATION
More informationUser s Guide. CO2 & O2 Controller for miniature incubators
CO2 O2 Control User s Guide CO2 & O2 Controller for miniature incubators Precise CO2 & O2 Control throughout the experiment Media ph control Compatible with any perfusion system Miniature incubators for
More informationCell Injury Controller II (CIC II)
Cell Injury Controller II (CIC II) CIC II Operation Manual Contents Introduction 2 Quick-Start 2 Theory of Operation 3 Injury Measurement Method 3 Components 4 External Requirements 4 Front Panel 5 Rear
More informationUser Manual GRI- 1500Li
User Manual GRI- 1500Li Your Cart Tek caddy cart was thoroughly quality control checked and road tested before being shipped to your address. We do everything possible to assure that your caddy is in perfect
More informationTru Flight TRUFLIGHT INSTALLATION GUIDE TRUGOLF.COM
Tru Flight T R A C K I N G S Y S T E M TRUFLIGHT INSTALLATION GUIDE TRUGOLF.COM TruFlight Camera Tracking System Setup & Installation TruFlight : How It Works... 1 TruFlight Components... 1 TruFlight Installation...
More informationBoyle s Law: Pressure-Volume Relationship in Gases
Boyle s Law: Pressure-Volume Relationship in Gases The primary objective of this experiment is to determine the relationship between the pressure and volume of a confined gas. The gas we will use is air,
More informationPressure Calibrators. Series PC..., PM...
Pressure Calibrators Series PC..., PM... SIKA-Precision Pressure Calibrators, Series PC..., PM... We help you, if you are under pressure! Pneumatic pressure calibrators PM 25 A, PM 25 D, PM 110 PIC pressure
More informationDiver-Office. Getting Started Guide. 2007, Schlumberger Water Services
Diver-Office Getting Started Guide 2007, Schlumberger Water Services Copyright Information 2007 Schlumberger Water Services. All rights reserved. No portion of the contents of this publication may be reproduced
More informationCART FOR CENTURION 1500+
CART FOR CENTURION 1500+ Operation & Maintenance Manual www.ameriwater.com 800-535-5585 AmeriWater 3345 Stop 8 Rd. Dayton, OH 45414 98-0161 Rev A Contents 1. GENERAL INFORMATION... 1 1.1. Features...
More informationTechcon Systems TS1254 Pressure Pot
Techcon Systems TS1254 Pressure Pot User Guide Copyright OK International, Inc. TABLE OF CONTENT 1. SAFETY. 3 1.1 Intended Use 3 1.2 Safety Precaution 3 2. FEATURES. 5 3. SPECIFICATIONS 4 4. INSTALLATION
More informationInstallation Instructions
Model: ACT-4500 ACT-4500-3 ACT-4500-6 SUN ELECTRIC CORPORATION Page: 1 of 5 Installation Instructions INSTALLATION MUST BE PERFORMED BY QUALIFIED SUN PERSONNEL ONLY INSTALLATION OVERVIEW: The Installation
More informationArmfield Distillation Column Operation Guidelines
Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become
More informationWelker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker Sampler Model GSS-1 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide to operations. Correct
More informationInstallation Instructions
SUN ELECTRIC CORPORATION Model: MRC-450 MRC-450-3 MRC-450-6 MRC-450-GR Page: 1 of 5 Installation Instructions INSTALLATION MUST BE PERFORMED BY QUALIFIED SUN PERSONNEL ONLY INSTALLATION OVERVIEW: The Installation
More informationaccidents which arise due to non-observance of these instructions and the safety information herein.
3 GALLON PANCAKE COMPRESSOR Model: 50959 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement, or other masonry.
More informationSystem 6 Water Polo. For the System 6 Sports Timer Software User Guide. F872 Rev
System 6 Water Polo For the System 6 Sports Timer Software User Guide F872 Rev. 20171114 Customer Service Department www.coloradotime.com Email: customerservice@coloradotime.com Phone: 970-667-1000 Toll
More informationATD LB PRESSURE BLASTER INSTRUCTION MANUAL
ATD-8402 90LB PRESSURE BLASTER INSTRUCTION MANUAL SAVE THESE INSTRUCTIONS SAFETY INSTRUCTIONS FOR SANDBLASTER 1. Before opening the tank release the air pressure on the sand tank. To do this, turn off
More informationFAULT CODE TROUBLESHOOTING INDEX
FAULT CODE TROUBLESHOOTING INDEX 1. Display indicates Change Filters 2. Display indicates Drip Tray Full Continuous Alarm will Sound 3. Display indicates Cold Fault 4. Display indicates Hot Fault 5. Display
More informationDS05C,D,G Dial Set Pressure Regulating Valves
DS05C,D,G Dial Set Pressure Regulating Valves APPLICATION The Honeywell DS05C,D,G Dial Set Pressure Regulating Valve is a high quality pressure regulating valve that maintains a constant outlet pressure
More informationUNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008
UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method
More informationOperating Manual. Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000)
Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000) Operating Manual AIR SYSTEMS INTERNATIONAL, INC. 829 Juniper Crescent, Chesapeake, Va., 23320 Telephone (757) 424-3967 Toll Free 1-800-866-8100
More informationPneumatic high-pressure controller Model CPC7000
Calibration technology Pneumatic high-pressure controller Model CPC7000 WIKA data sheet CT 27.63 Applications Healthcare and avionics industry Industry (laboratory, workshop and production) Transmitter
More informationWARNING: READ THE GENERAL INFORMATION MANUAL INCLUDED FOR OPERATING AND SAFETY PRECAUTIONS AND OTHER IMPORTANT INFORMATION.
GSR500 Ram Assembly General Description The GSR500 single post lift / ram uses a 3-1/4 air powered cylinder, which is welded to a heavy gauge base. It is normally used to raise and lower a fluid handling
More informationIntroduction. Part one: Identify the Hydraulic Trainer Components
The University Of Jordan School of Engineering Mechatronics Engineering Department Fluid Power Engineering Lab Experiments No.4 Introduction to Hydraulic Trainer Objective: Students will be able to identify
More informationAnti-flood device Model B-1
December 4, 2009 Dry Systems 123a 1. DESCRIPTION The Anti-flood Device is required when Viking accelerators are installed on dry systems according to Viking Model E-1 Accelerator Trim Charts. In the SET
More informationG200 Analyzer Range. Operating Manual
G200 Analyzer Range Operating Manual G200 N 2 0 0-1,000ppm G210 N 2 0 0-100% VIASENSOR 850 South Via Lata Suite 112 Colton, CA 92324 USA Toll Free: 855 VIASENSOR Tel: +1 (909) 783-9472 Fax: +1 (909) 825-0591
More informationProduct Information News September 26, 2003
Product Information News September 26, 2003 LEVER HEIGHT INSPECTION FOR FIREHAWK MMR MSA is announcing a revision to the instructions for the Firehawk MMR Second Stage Regulator as they relate to the air
More informationThe MRL Furnaces USED FOR THIS MANUAL COVERS
The MRL Furnaces USED FOR 1. Making Silicon Oxide films from 10nm to 2um thickness 2. Annealing with temperatures from 30 800C and times from 5min to 2days 3. Diffusion with temperatures from 30 1100C
More informationPXRF- Tracer Helium Flow Control System
PXRF- Tracer Helium Flow Control System This flow control system was designed to regulate the gas flow from a helium (He) balloon tank containing 8.9 cu ft (0.25 cu m). Tank not included with the system.
More informationKIV-SACS Safe automatic calibration system:
KIV-SACS Safe automatic calibration system: LOK: g:\prosjekter\calibration_system\kiv safe calibration system.doc Page: 1 of 14 Parts description:... 4 Wash extractors:... 6 Tunnel washers:... 5 Test valve:...
More informationMicrosoft Windows Software Manual for FITstep Stream Version 4
Thank you for purchasing this product from Gopher. If you are not satisfied with any Gopher purchase for any reason at any time, contact us and we will replace the product, credit your account, or refund
More informationIndustrial pressure controller Model CPC4000
Calibration technology Industrial pressure controller Model CPC4000 WIKA data sheet CT 27.40 Applications Oil and gas industry Industry (laboratory, workshop and production) Transmitter and pressure gauge
More informationPTG100 Precision Test Gauge
PTG100 Precision Test Gauge User Manual PD1007 Rev B 03/28/2014 Palmer Instruments Inc. 234 Old Weaverville Road Asheville, NC 28804 Toll Free: 800-421-2853 Phone: 828-658-3131 Fax: 828-658-0728 Email:
More informationTechnology. Using Bluetooth
Bluetooth is a communication technology that makes it possible to send and receive data without using wires. Using the Bluetooth features, you can set up a wireless connection between your NXT and other
More informationOperator Quick Guide ORBISPHERE 3654
Operator Quick Guide ORBISPHERE 3654 Revision H - 14/03/2008 Operating Information About this Guide The information in this guide has been carefully checked and is believed to be accurate. However, Hach
More informationWELCOME TO THE REVOLUTION
USER GUIDE WELCOME TO THE REVOLUTION THANK YOU FOR CHOOSING THE GCQUAD We listened to what you wanted - and created the most accurate, versatile and game-enhancing ball and club analysis solution available
More information2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL
2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL Model: 52024 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement,
More informationHyperSecureLink V6.0x User Guide
HyperSecureLink V6.0x User Guide Note: This software works with the LS-30 Version (06.0x or later) 1, Hardware Installation: 1-1, Connection Diagram for USB or RS-232 Computer Interface To LS-30 CM1 To
More informationInstallation of Your SprayMaster System
Installation of Your SprayMaster System 1. At the installation site, remove all equipment from the corrugated box and the polyethylene drum and replace the drum lid. Check the picture to identify each
More informationUser s Guide. CO2 Controller for miniature incubators
CO2 Control User s Guide CO2 Controller for miniature incubators Precise CO2 Control throughout the experiment Media ph control Compatible with any perfusion system Miniature incubators for any microscope
More informationMASSACHUSETTS STATE POLICE CRIME LABORATORY. Certificate of Calibration Procedure for the Alcotest 9510
MASSACHUSETTS STATE POLICE CRIME LABORATORY Certificate of Calibration Procedure for the Alcotest 9510 Effective Date: May 10, 2017 Page 1 of 23 TABLE OF CONTENTS 1 INTRODUCTION... 3 2 ABBREVIATIONS...
More informationSurvey Technical Support Notes October 2015
Survey Technical Support Notes October 2015 SonarMite with Trimble Access Overview: Trimble Access software, when connected to a SonarMite will store water depths associated with field measurements. Equipment:
More informationAIR COMPRESSOR. Failure to follow all instructions as listed below may result in electrical shock, fire, and/or serious personal injury.
2 GALLON AIR COMPRESSOR Model: 7517 DO NOT RETURN TO STORE. Please CALL 800-348-5004 for parts and service. CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials
More informationTECHNICAL DATA ANTI-FLOOD DEVICE MODEL B-1 1. DESCRIPTION
Page 1 of 6 1. DESCRIPTION The Model B-1 Anti-flood Device is required when Viking accelerators are installed on dry systems according to Viking Model E-1 Accelerator Trim Charts. In the SET condition,
More informationDDR Dive Data Recorder Manual - Rel /12
DDR Dive Data Recorder Manual - Rel. 1.1 2/12 TABLE OF CONTENTS 1. WHAT IS THE DDR? 4 2. TECHNICAL CHARACTERISTICS 5 3. DOWNLOADING DATA TO A PC 6 4. DDR MANAGER 7 5. CONFIGURE DEVICE BUTTON 9 6. READ
More informationMP15 Jockey Pump Controller
Setup and Operating Instructions MP15 Jockey Pump Controller This manual provides general information, installation, operation, maintenance, and system setup information for Metron Model MP15 Jockey Pump
More informationPneumatic high-pressure controller Model CPC7000
Calibration technology Pneumatic high-pressure controller Model CPC7000 WIKA data sheet CT 27.63 Applications Automotive and avionics industry Industry (laboratory, workshop and production) Transmitter
More informationMyControl Bathing System Signature Series
MyControl Bathing System Signature Series Touch Assist Touch Pad and CM Technologies Seal Pressure Regulation Instructions Acceptable seal inflation pressure setting must be varified, and adjusted if necessary,
More informationSTANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.
STANDARD OPERATING PROCEDURES Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.5 ) Sampler (RAAS) AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION
More information1)! DO NOT PROCEED BEYOND THIS MARK
Operating Instructions for X-ray Photoelectron Spectrometer: Physical Electronics Model 555 XPS/AES (John H. Thomas, III, Ph.D., Electron Spectroscopy) Sample Insertion: figure 1. Sample insertion rod
More informationWL16 WATER LEVEL LOGGERS Submersible pressure transducer and USB datalogger combination
WL16 WATER LEVEL LOGGERS Submersible pressure transducer and USB datalogger combination Warning: Non-vented water level loggers may have readings with errors of up to 250mm due to barometric pressure changes.
More informationOMNISPORT 2000 SWIMMING QUICK REFERENCE
Before During After OMNISPORT 2000 SWIMMING QUICK REFERENCE 1. Confirm that the timer is reset. 2. Set the timer to the correct event and heat. The right LCD, line 2 displays E:(event) H:(heat) R:(round)
More informationOperating Instructions for Cobra and Lynx Duct Leakage Testers
Operating Instructions for Cobra and Lynx Duct Leakage Testers Table of Contents Introduction Introduction... 2 Safety Precautions Safety Precautions.. 2 Prior to Testing System Preparation.. 4 Power Requirements...
More informationA180 DO/BOD/OUR/SOUR Meter Instruction Manual
A180 DO/BOD/OUR/SOUR Meter Instruction Manual BANTE INSTRUMENTS CO., LTD A180 DO/BOD/OUR/SOUR Meter 1 Introduction Thank you for selecting the A180 multiparameter dissolved oxygen meter. This manual provides
More informationMASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120
MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 QUICK START GUIDE P/N 6006154, REVISION C MAY 2013 Model 8120 Mask Integrity Tester is patented under U.S. Patent No. 8,312,761. Additional patents are pending.
More informationModel 130M Pneumatic Controller
Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate
More informationNew Cortex Software User Guide
New Cortex Software User Guide Initial Set Up Turn on the Cortex and leave to warm up for approximately 15 minutes before calibrating Load MSS Toolbox from the desktop From the toolbox menu select Calibration
More informationUser Manual Pressure Calibration System C and Hydraulic Pressure Pump Models T-620/T-620H Copyright 2008 AMETEK Denmark A/S
User Manual Pressure Calibration System C and Hydraulic Pressure Pump Models T-620/T-620H Copyright 2008 AMETEK Denmark A/S About this manual. The structure of the manual This user manual is aimed at users
More information