1 Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle, 1 William Williamson Southwest Research Institute, 6220 Culebra Road, San Antonio, TX National Oceanic and Atmospheric Administration / Earth System Research Laboratory, 325 Broadway, Boulder, CO Supporting information that follows includes: A description and schematic of the laboratory-based delivery system Experimental description and results for moisture tests A picture of the underside of the SkyWisp wing showing two micro-aircores installed for flight A figure illustrating the operational concept for in situ measurement of collected gas Results for testing with other volatile organic molecules This material is available free of charge via the Internet at 1
2 Description of Laboratory-Based Delivery System The custom gas delivery system (SI-Figure 1) consists of two separate gas cylinders each connected to a mass flow controller (MFC) capable of maintaining flow from 0.1 ml/min to 9.0 ml/min (MKS). The mass flow controllers are fed into a six-port switching valve (Valco) which allows the flow to be directed to one of two outputs. Both outputs were constructed to be identical, consisting of a low-dead volume cross and a long capillary tube (30 m, 0.53 mm ID Restek Hydroguard fused silica). The dual construction was designed to balance the pressure so that upon switching gases, there is not a disruption in flow. The result is that both tubes become filled as mirror images of one another. During typical experiments presented here, the tube on the right was analyzed and the contents of the tube on the left ( dummy ) were discarded; an alternative set-up would be to fill the dummy tube with internal standard and alternate between the tubes during analysis. This is expected to provide improved accuracy but was not performed as part of this study; a six-port switching valve was used but only four ports were needed. The extra two ports needed to be connected using a short length of capillary thereby creating a small amount of dead volume for the dummy not present on the analytical side. If the dummy side is to be used, a four-port switching valve is ideal. All tubing after the switching valve consists of 0.53 mm capillary (Restek Hydroguard) in order to minimize dead volume. The MFCs and switching valve are fully programmable allowing precise intervals of gas to be delivered into the micro-aircore. A vacuum line provides evacuation of the apparatus to approximately psia. During typical operation, the tubes are purged with test gas and evacuated. The tubes are filled with the desired pattern of gases with the valves at the end of the tubes closed; fill time is approximately 75 minutes at 0.1 ml/min to reach ambient pressure. The pressure on the dummy side is monitored throughout the experiment, with the micro-aircore side closed off to the vacuum gauge to prevent cross-contamination. After the micro-aircore tube reaches 2
3 atmospheric pressure, the micro-aircore inlet valve is closed and the gas flow stopped. The cross region just before the inlet valve is evacuated for 4 minutes to remove any residual gas that did not reach the tube. The flow path was such that the last gas delivered to the tube was the first gas detected (flow path reversed for detection). Analysis occurs by opening the micro AirCore tube to the Mass Spectrometer (MS) detector. The valve at the opposite end of the micro AirCore is also opened causing that end to remain at room pressure. The contents of the micro AirCore flow through a 3 m, 0.1 mm capillary (Restek Hydroguard fused silica) prior to reaching the MS source; this additional capillary provides restriction to lower the flow rate into the MS and control the ion source pressure. The analysis is passive, meaning that the vacuum of the MS source draws the contents through the tube, eventually ending with room air. Laboratory-Based Moisture Tests Eight cylinders were prepared to target high (600 ppm) and low (300 ppm) concentrations of 12 CO 2 using nitrogen fill gas containing 3515 ppm argon and 410 ppm CO as internal standards for microaircore/ms and GC/PDHID, respectively). Four of these cylinders were humidified (two low concentration and two high concentration, labeled hum 12 CO 2 std6 through std9). This was accomplished by adding approximately 400 µl of water to the evacuated cylinder and heating slightly to ensure saturation (100% relative humidity). The presence of an elevated level of water was confirmed by mass spectrometry. The dry cylinders are labeled dry 12 CO 2 std10 through std13. Actual concentration of 12 CO 2 in each cylinder was validated by GC/PDHID as described above. For the mass spectrometric measurement, the micro-aircore tube was filled with the test gas then analyzed as described above using argon as the internal standard. Standard deviation was calculated from triplicate measurements. 3
4 Results of the 12 CO 2 concentration measurement by GC/PDHID as compared to the measurement of the same cylinder by mass spectrometry are shown in SI-Figure 2. As observed for the 13 CO 2 measurements (Figure 6), the concentration as measured by GC/PDHID and micro-aircore/ms correlated well. The average percent deviation between the two measurement techniques was 2.0%. GC/PDHID and micro-aircore/ms each yielded average percent relative deviations of 1.5% and 1.3%, respectively, for replicate measurements. The humidified air did not affect the measurement of 12 CO 2 by micro-aircore/ms because the observed concentration agreed with that measured by GC/PDHID. GC/PDHID is known to not be sensitive to the presence of water vapor. Applications to Other Volatile Chemicals The micro-aircore was demonstrated for select EPA TO-15 compounds (1 ppm each of chloroform, benzene, hexane, and toluene). Using the laboratory-based validation system, 2 minute intervals of test gas separated by 10 minutes of nitrogen were delivered to the micro- AirCore device. Raw results are shown in Figure 8. An internal standard was not used to correct for detector drift. Note the tube was filled in one direction and then the contents analyzed in the opposite direction; the gas delivered to the device first was analyzed last. This is contributing to the band broadening observed as length along the tube increases. The results also demonstrate that a small chromatographic effect is occurring (separation of toluene from chloroform); in practice, results could be corrected by application of a retention factor. 4
5 Supporting Information Figures SI-Figure 1. Laboratory-based validation system 5
6 SI-Figure 2. GC/PDHID and micro-aircore/ms results for humidified and dry 12 CO 2 standards 6
7 SI-Figure 3. Underside of SkyWisp wing showing two micro-aircores installed for flight 7
8 SI-Figure 4. Raw mass spectrometric data demonstrating the operational concept for in situ measurement of collected gas with one valve (top) and both valves (bottom) open 8
9 SI-Figure 5. Volatile organic compounds are recovered from the micro-aircore tube at a concentration of 1 ppm (2 min intervals separated by 10 min of nitrogen 9
Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator
Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.
Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions
More informationDETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY
DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY Nengbing Xu, Hongmei Ying and Libo Zhu Ningbo Environmental Monitoring
More informationGases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.
Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to
More informationCommercial Practice Test Method Internal Vapor Analysis of Hermetic Devices
Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The
More informationMajor Design Features
Major Design Features 3Flex Features Design Benefits Discussion 316 SS fittings and pneumatically actuated, hard-seal valves Provides virtually leak-free gas management with the lowest outgassing rate
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 9 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 3.1 Canister Receipt 3.2 Canister Storage 4.0 INTERFERENCE AND POTENTIAL PROBLEMS
More informationApplications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples
Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Outline of Presentation Introduction Process Gas Analysis Magnetic Sector Mass Spectrometer Standard
More informationComparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS
TECHNICAL NOTE 1598 Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS Authors Amit Gujar, 1 Tim Anderson, 1 Daniela
More informationUsing the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer
application Note Gas Chromatography/ Mass Spectrometry Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer Overview A mass spectrometer
More informationA New GC/MS System Designed for Helium Carrier Gas Conservation
A New GC/MS System Designed for Helium Carrier Gas Conservation Approaches for the Smarter Use of Helium Dale R. Walker GCMS Applications Specialist August, 7, 2013 1 Helium Facts How it's produced Where
More informationRetention Time Locking: Concepts and Applications. Application
Retention Time Locking: Concepts and Applications Application Gas Chromatography December 1997 Authors Vince Giarrocco Bruce Quimby Matthew Klee Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More informationTitle: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator
Procedure No: SOP-012 Revision No: 1.2, January 27, 2011 Page No.: 1 of 7 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Sabio model 2010 and 4010 dilution calibrators.
More informationAddressing the World Shortage of Helium
Addressing the World Shortage of Helium Introducing the Programmable Helium Conservation Module GPD Solutions June 2013 Market Situation Unreliable supply of helium worldwide and increasing prices have
More informationVIC offers a variety of Calibrated Gas Leaks
CALIBRATED GAS LEAKS VIC offers a high quality line of calibrated gas leaks traceable to National Institute of Standards Technology (NIST). Calibrated gas leaks are used to provide a controlled delivery
More informationFlare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1
Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC
More informationThe HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.
The HumiSys RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The new HumiSys with single or dual RH probes capabilities is
More informationMETHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER
1250 METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Total Organic Compounds N/A
More informationDetector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane
Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,
More informationGas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)
Int. J. Metrol. Qual. Eng. 5, 303 (2014) c EDP Sciences 2014 DOI: 10.1051/ijmqe/2014012 Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA) D. Jonker and E.P. Tarnow
More informationElimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU.
Elimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU. Control Analytics, Inc 6017 Enterprise Dr Export, PA 15632 (724) 387 CEMS (2367)
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCE AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More informationThe HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator
The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Temperature controlled, multi-technique volumetric
More informationDemystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.
Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 5.1 Subatmospheric
More informationOptimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler
Optimizing Vial Pressurization Parameters for the Analysis of Residual Solvents Using the 7697A Headspace Sampler Application Note Pharmaceuticals Author Roger L Firor Agilent Technologies, Inc.
More informationMultiple Gas#5 GC configuration Jan 2016
History: Unfortunately there is no single column that can separate: Hydrogen Oxygen Nitrogen Methane CO CO2 Ethane Water Propane Butane Pentane Over the years SRI Instruments has devised several solutions
More informationBest Practice for Identifying Leaks in GC and GC/MS Systems
Best Practice for Identifying Leaks in GC and GC/MS Systems Technical Overview Maintaining a leak-free GC or GC/MS system is critical for obtaining optimal system performance with reliable, reproducible,
More informationZebron Gas Management
Zebron www.phenomenex.com/gasmanagement What are the key benefits of Zebron filters and traps? The high adsorption capacity of the Zebron Filters removes contaminants from mobile phase gases, reducing
More informationMeasuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability
Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability R. C. Kullberg*, D. J. Rossiter** Oneida Research Services, Inc., 8282 Halsey Road, Whitesboro, NY13492 Abstract Many
More informationLINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER
LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER Michael T Tang, Ph.D. Grace Feng Greg Merideth Rui Huang Matheson Gas Applied Lab
More informationImproving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer
Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer Ben D. Gardner, Phillip M. Erwin, Wai Tak Lee, Amber M. Tissandier, Souzan M. Thoresen Hamilton Sundstrand
More informationHumiSys HF High Flow RH Generator
HumiSys HF High Flow RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The HumiSys HF is a high flow version of the previously
More informationGenerating Calibration Gas Standards
Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and
More informationThe HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator
The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Universal pycnometer, no
More informationAdvantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types
19 th World Conference on Non-Destructive Testing 2016 Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types Klaus HERRMANN 1, Daniel WETZIG
More informationMETHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the
1151 METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS 1.0 Scope and Application. 1.1 Analytes. Analyte Volatile Organic Compounds (VOC) CAS No. No CAS number assigned 1.2 Scope. This method
More informationDevelopment of a high vacuum sample preparation system for helium mass spectrometer
Journal of Physics: Conference Series Development of a high vacuum sample preparation system for helium mass spectrometer To cite this article: P Kumar et al 2012 J. Phys.: Conf. Ser. 390 012056 View the
More informationEAS Basic Soil Gas Sampler
EAS Basic Soil Gas Sampler Please Read The Following Important Information before Starting The EAS Basic Soil Gas Sampler has the traditional design where a pump or evacuated canister can be to purge the
More informationH 2. Sensor Series Take no risk! Install a Hydrogen Sensor in your Gaschromatograph
Sensor Series 9000 Take no risk! Install a Hydrogen Sensor in your Gaschromatograph Carrier Gases for GC Probably more than 90% of the present GC instruments run with helium as carrier gas. Some people
More informationGas Sampling in the DST
UCRL-ID-129732 Gas Sampling in the DST L. DeLoach M. Chiarappa R. Martinelli B. Glassley January 12, 1998 Lawrence Livermore National Laboratory This is an informal report intended primarily for internal
More informationTitle: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)
Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet
More informationThe instrument should have a mass range from 2 to 2000 m/z or better Interface
Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for
More informationPredicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010
Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600 November 2010 Table of Contents ``Abstract...3 ``Introduction...4 ``Methods & Results...6 ``Data Analysis...9 ``Conclusion...12
More informationVapor Intrusion. Sampling Solutions VAPOR INTRUSION. Sampling Solutions
Vapor Intrusion Sampling Solutions VAPOR INTRUSION Sampling Solutions The All-In-One Soil Gas Sampling Solution Solutions for all 1 2 3 4 ZONES Indoor Air Quality Soil gas levels inside structures. Sub
More informationJaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation
Using Nitrogen instead of Helium as carrier gas, while maintaining exact the same separation efficiency, retention times, peak elution order, without changing the oven temperature conditions. Jaap de Zeeuw
More information29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies
IMPROVED CALIBRATION AND OPERATION OF A STABLE-GAS QUANTIFICATION MANIFOLD AS PART OF A RADIOXENON COLLECTION SYSTEM ABSTRACT Richard M. Williams, James C. Hayes, and Randy R. Kirkham Pacific Northwest
More informationThe New PHD 4 Portable Helium Detector
The New PHD 4 Portable Helium Detector SPECIAL FEATURES: High Sensitivity to Helium Easy to Use Truly Portable Versatile Dependable The system is designed to allow easy replacement of sampling line components
More informationThermo Scientific Model 146i Principle of Operation
Thermo Scientific Model 146i Principle of Operation The Model 146i Multi-gas Calibrator dilutes calibration gases to precise concentrations. The diluted gases are used to perform zero, precision and Level
More informationIn-Water Mass Spectrometry for Characterization of Light Hydrocarbon Seeps and Leaks
In-Water Mass Spectrometry for Characterization of Light Hydrocarbon Seeps and Leaks R. T. Short, S. K. Toler, A. M. Cardenas-Valencia, S. Untiedt, C. Cullins, M. Ryder, and J. Kloske HEMS Workshop 2015
More informationWeiyi Zheng, Xiaoxia Dong, Evan Rogers, Jimmie C. Oxley, and James L. Smith*
Improvements in the Determination of Decomposition Gases from 1,3,3-Trinitroazetidine and 5-Nitro-2,4-dihydro-3H-1,2,4,-trizol-3-one using Capillary Gas Chromatography-Mass Spectrometry Weiyi Zheng, Xiaoxia
More informationBasic & Controlled Atmosphere Glove Boxes
I N T R O D U C T I O N Precise Basic and Controlled Atmosphere Glove Boxes provide a leak-tight environment for work with contamination-sensitive materials. Precise Basic Glove Boxes are simple, economical
More informationAir Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions
Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions Introduction: There is a bit of explanation of the health effects of benzene in the Introduction section of the first part
More informationASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist
ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking
More informationSetting a New Standard in Flash Chromatography Performance
Setting a New Standard in Flash Chromatography Performance What s New Faster Flow rates, up to 300 ml/min Higher pressures, up to 300 psi (20 bar) Bigger touchscreen, 12 or 15 inches Wider Dynamic Range
More informationMETHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES
METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide
More informationHiQ laboratory gas generators.
Product information HiQ laboratory gas generators. The convenient alternative for laboratory environments. 02 HiQ laboratory gas generators HiQ laboratory gas generators. Designed with safety and convenience
More informationExperiment GC : Analysis of BTEX by GC-FID
46 Experiment GC : Analysis of BTEX by GC-FID Learning Goals: Familiarity with gas chromatography Gain experience in temperature programming and method development Correctly use an internal standard, and
More informationProtocol for the verification tests on ethylene oxide treatment technology
Pilot project for the environmental technology verification In the field of ethylene oxide treatment technology Protocol for the verification tests on ethylene oxide treatment technology Sept. 11, 2003
More informationThe Helium Leak Detector
The Helium Leak Detector Helium Leak Detector Main Components The main components of a helium leak detector are: 1. The analyzed, which enables to separate the tracer gas from other gases inside leak detector.
More informationGaseous oxygen technical grade Specification
CDC 09 (5010) P3 Rev. TZS 217:1984 FINALIZED TANZANIA STANDARD Gaseous oxygen technical grade Specification TANZANIA BUREAU OF STANDARDS. 2 nd edition TBS 2017 0. FOREWORD 0.1. This Tanzania Draft Standard
More informationAFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features
ANALOG MASS FLOW CONTROLLERS Model AF mass fl ow controllers are designed to indicate fl ow rates and control set fl ow rates of gases. Each of these units incorporates an advanced straight tube sensor
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople
More informationTitle: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator
Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality
More informationGas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems
Inorganic Analysis Chromatography Molecular Spectroscopy Thermal/Elemental Analysis Informatics Gas Chromatography MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass
More informationNMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013
NMISA NEW GAS FLOW CALIBRATION FACILITY Deona Jonker 7 October 2013 Introduction Flow the total volume of a fluid that flows past a fixed point in a given time Flow measurements used in industry as a means
More informationGases and welding distributors continue
Specialty Gas Instrumentation By Ray Borzio, Mike Lee, and Scott Varian Gases and welding distributors continue to have a significant future in the production, delivery, and related services for specialty
More informationOMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative
More informationAPPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT
APPLICATION NOTE Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT The INFICON 3000 Micro GC provides two fast and accurate solutions for the analysis of coal mine gas components.
More informationTrace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS
Abstract Highly sensitive analysis of metanephrines in plasma is critical in the diagnosis and treatment of pheochromocytoma and paraganglioma. Here, a HILIC LC-MS/MS method was developed using a Raptor
More informationID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #:
ID-Cube Installation and Operations Manual For ID-Cube Source and Power Supply Revision 2 Document #: 7.1.002 IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906 Copyright 2014 by IonSense Inc. All rights
More informationAutomated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:
Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion
More informationREFINERY GAS ANALYSIS BY GAS CHROMATOGRAPHY
REFINERY GAS ANALYSIS BY GAS CHROMATOGRAPHY UOP Method SCOPE This automated method is for determining the composition of refinery gas samples or expanded liquefied petroleum gas (LPG) samples obtained
More informationNMISA New Gas Flow Calibration Facility
NMISA New Gas Flow Calibration Facility Speaker / Author: D. Jonker Co-author: E.P. Tarnow National Metrology Institute of South Africa (NMISA) Private Bag X34, Lynnwood Ridge, Pretoria, 0040, South Africa
More informationBEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION
BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the
More informationCHAPTER 3 RESULTS AND DISCUSSION
CHAPTER 3 RESULTS AND DISCUSSION The research included the development and verification of the earlier apparatus, and the collection and validation of moisture diffusion data under both isothermal and
More informationPre-installation Checklist
Pre-installation Checklist Varian 1200/1200L Mass Spectrometer Use this checklist as a guide while preparing your site for the installation of the Varian 1200 or 1200L MS system. Make sure that your site
More informationCHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV
CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV 16-1 INTRODUCTION 16-1.1 Purpose. The purpose of this chapter is to familiarize divers with the techniques used to mix divers breathing gas. 16-1.2 Scope. This chapter
More informationUser s Guide. Vacuum Controller for liquid delivery systems
Flow Control User s Guide Vacuum Controller for liquid delivery systems Precise Vacuum Control throughout the experiment Flow control Compatible with any perfusion system Ideal for Small Volume Delivery
More informationAPPLICATION NOTE. GC Integrated Permeation Device
GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords:
More informationBonding Reliability Testing for Wafer Level Packaged MEMS Devices
Bonding Reliability Testing for Wafer Level Packaged MEMS Devices Michael Shillinger Vice President of Operations Innovative Micro Technology Santa Barbara, CA, USA mjs@imtmems.com October 20, 2011 Innovative
More informationVERIFYING GAS CHROMATOGRAPH OPERATION AT CUSTODY TRANSFER LOCATIONS Murray Fraser Daniel Measurement & Control
VERIFYING GAS CHROMATOGRAPH OPERATION AT CUSTODY TRANSFER LOCATIONS Murray Fraser Daniel Measurement & Control Houston, Texas ABSTRACT The on-line gas chromatograph (GC) has been widely used for natural
More informationCHEM 355 EXPERIMENT 7. Viscosity of gases: Estimation of molecular diameter
CHEM 355 EXPERIMENT 7 Viscosity of gases: Estimation of molecular diameter Expressed most simply, the viscosity of a fluid (liquid or gas) relates to its resistance to flow. The viscosity of a gas is determined
More informationApplication Note AN-107
SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance
More information1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor)
1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor) 1. According to NFPA 99, SCFM is an acronym for. a) Surface Conditions at Fahrenheit Mercury b) Standard Conditions
More informationGCMSD-Headspace Analysis SOP
Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to
More informationReal-time PCB Gas Monitoring System (MOHMS-21GP)
Real-time PCB Gas Monitoring System (MOHMS-21GP) SHINSAKU DOBASHI* 1 YOSHIHIRO DEGUCHI* 1 MASAHARU KIRA* 2 RYUICHIRO TANAKA* 3 YOSHINORI IZAWA* 3 This report describes the technical characteristics and
More informationCarrier Gases in Capillary GC
Carrier Gases in Capillary GC LC Columns and Consumables Mark Sinnott Application Engineer January 15, 2009 CARRIER GAS Mobile Phase Carries the solutes down the column Selection and velocity influences
More informationPractical approach and problems in in-situ RGA calibration
Practical approach and problems in in-situ RGA calibration Oleg Malyshev and Keith Middleman Vacuum Science Group, ASTeC Accelerator Science and Technology Centre STFC Daresbury Laboratory UK Workshop
More informationA NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES
A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES F. Papa 1, J. Brindley 2, T. Williams 2, B. Daniel 2, V. Bellido-Gonzalez 2, Dermot Monaghan
More informationAppendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation
Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly
More informationColumbus Instruments
0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011
More informationWhat I have learned about SF 6 gas testing.a Practical explanation
What I have learned about SF 6 gas testing.a Practical explanation Over the past few years, testing of SF6 gas quality has become increasingly important and with the EPA in USA, setting guidelines for
More information1 Design of the Portable mass spectrometric sensor for vascular and endocrine disease diagnostics
1 Design of the Portable mass spectrometric sensor for vascular and endocrine disease diagnostics Andrei Antonov, Viktor Kogan, Aleksandr Krutikov A.F. Ioffe Physical Technical Institute, Academy of Science,
More informationStandard Operating Procedure for H 2 S and SO 2 Converter
Procedure No: SOP-001 Revision No: 1.0 January 24, 2011 Page No.: 1 of 6 1. INTRODUCTION AND SCOPE This procedure describes the operation and configuration of heated converters used in continuous ambient
More informationMethods Used in Measuring Surface Seawater pco 2 Aboard RVIB Nathaniel B. Palmer
F:\NBPalmer\Palmer_methods.doc Page 1 09/16/02 11:55 AM 1. Introduction: Methods Used in Measuring Surface Seawater pco 2 Aboard RVIB Nathaniel B. Palmer Taro Takahashi, Colm Sweeney, Stewart C. Sutherland
More informationSINGULUS TECHNOLOGIES
Fast Vacuum for Optical Disc Metallization and Photovoltaic Coatings B. Cord, M. Hoffmann, O. Hohn, F. Martin SINGULUS Technologies AG, Kahl am Main SINGULUS TECHNOLOGIES June 2014 SINGULUS Overview 25-June-2014-2
More informationTotal Pressure Measurement Using Residual Gas Analyzers
Total Pressure Measurement Using Residual Gas Analyzers M. Maskell, Old Dominion University; G. Myneni, P. Adderley, Jefferson Lab; G. Brucker, Stanford Research Systems; C. Day, Forschungszentrum, Karlsruhe
More informationMarc McMullen. Physics Detector Support Group
Marc McMullen Physics Detector Support Group Hall B Gaseous Detectors 1. Drift Chamber (DC) 2. Low Threshold Cherenkov Counter (LTCC) 3. High Threshold Cherenkov Counter (HTCC) 4. Silicon Vertex Tracker
More information