Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Size: px
Start display at page:

Download "Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial"

Transcription

1 Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle, 1 William Williamson Southwest Research Institute, 6220 Culebra Road, San Antonio, TX National Oceanic and Atmospheric Administration / Earth System Research Laboratory, 325 Broadway, Boulder, CO Supporting information that follows includes: A description and schematic of the laboratory-based delivery system Experimental description and results for moisture tests A picture of the underside of the SkyWisp wing showing two micro-aircores installed for flight A figure illustrating the operational concept for in situ measurement of collected gas Results for testing with other volatile organic molecules This material is available free of charge via the Internet at 1

2 Description of Laboratory-Based Delivery System The custom gas delivery system (SI-Figure 1) consists of two separate gas cylinders each connected to a mass flow controller (MFC) capable of maintaining flow from 0.1 ml/min to 9.0 ml/min (MKS). The mass flow controllers are fed into a six-port switching valve (Valco) which allows the flow to be directed to one of two outputs. Both outputs were constructed to be identical, consisting of a low-dead volume cross and a long capillary tube (30 m, 0.53 mm ID Restek Hydroguard fused silica). The dual construction was designed to balance the pressure so that upon switching gases, there is not a disruption in flow. The result is that both tubes become filled as mirror images of one another. During typical experiments presented here, the tube on the right was analyzed and the contents of the tube on the left ( dummy ) were discarded; an alternative set-up would be to fill the dummy tube with internal standard and alternate between the tubes during analysis. This is expected to provide improved accuracy but was not performed as part of this study; a six-port switching valve was used but only four ports were needed. The extra two ports needed to be connected using a short length of capillary thereby creating a small amount of dead volume for the dummy not present on the analytical side. If the dummy side is to be used, a four-port switching valve is ideal. All tubing after the switching valve consists of 0.53 mm capillary (Restek Hydroguard) in order to minimize dead volume. The MFCs and switching valve are fully programmable allowing precise intervals of gas to be delivered into the micro-aircore. A vacuum line provides evacuation of the apparatus to approximately psia. During typical operation, the tubes are purged with test gas and evacuated. The tubes are filled with the desired pattern of gases with the valves at the end of the tubes closed; fill time is approximately 75 minutes at 0.1 ml/min to reach ambient pressure. The pressure on the dummy side is monitored throughout the experiment, with the micro-aircore side closed off to the vacuum gauge to prevent cross-contamination. After the micro-aircore tube reaches 2

3 atmospheric pressure, the micro-aircore inlet valve is closed and the gas flow stopped. The cross region just before the inlet valve is evacuated for 4 minutes to remove any residual gas that did not reach the tube. The flow path was such that the last gas delivered to the tube was the first gas detected (flow path reversed for detection). Analysis occurs by opening the micro AirCore tube to the Mass Spectrometer (MS) detector. The valve at the opposite end of the micro AirCore is also opened causing that end to remain at room pressure. The contents of the micro AirCore flow through a 3 m, 0.1 mm capillary (Restek Hydroguard fused silica) prior to reaching the MS source; this additional capillary provides restriction to lower the flow rate into the MS and control the ion source pressure. The analysis is passive, meaning that the vacuum of the MS source draws the contents through the tube, eventually ending with room air. Laboratory-Based Moisture Tests Eight cylinders were prepared to target high (600 ppm) and low (300 ppm) concentrations of 12 CO 2 using nitrogen fill gas containing 3515 ppm argon and 410 ppm CO as internal standards for microaircore/ms and GC/PDHID, respectively). Four of these cylinders were humidified (two low concentration and two high concentration, labeled hum 12 CO 2 std6 through std9). This was accomplished by adding approximately 400 µl of water to the evacuated cylinder and heating slightly to ensure saturation (100% relative humidity). The presence of an elevated level of water was confirmed by mass spectrometry. The dry cylinders are labeled dry 12 CO 2 std10 through std13. Actual concentration of 12 CO 2 in each cylinder was validated by GC/PDHID as described above. For the mass spectrometric measurement, the micro-aircore tube was filled with the test gas then analyzed as described above using argon as the internal standard. Standard deviation was calculated from triplicate measurements. 3

4 Results of the 12 CO 2 concentration measurement by GC/PDHID as compared to the measurement of the same cylinder by mass spectrometry are shown in SI-Figure 2. As observed for the 13 CO 2 measurements (Figure 6), the concentration as measured by GC/PDHID and micro-aircore/ms correlated well. The average percent deviation between the two measurement techniques was 2.0%. GC/PDHID and micro-aircore/ms each yielded average percent relative deviations of 1.5% and 1.3%, respectively, for replicate measurements. The humidified air did not affect the measurement of 12 CO 2 by micro-aircore/ms because the observed concentration agreed with that measured by GC/PDHID. GC/PDHID is known to not be sensitive to the presence of water vapor. Applications to Other Volatile Chemicals The micro-aircore was demonstrated for select EPA TO-15 compounds (1 ppm each of chloroform, benzene, hexane, and toluene). Using the laboratory-based validation system, 2 minute intervals of test gas separated by 10 minutes of nitrogen were delivered to the micro- AirCore device. Raw results are shown in Figure 8. An internal standard was not used to correct for detector drift. Note the tube was filled in one direction and then the contents analyzed in the opposite direction; the gas delivered to the device first was analyzed last. This is contributing to the band broadening observed as length along the tube increases. The results also demonstrate that a small chromatographic effect is occurring (separation of toluene from chloroform); in practice, results could be corrected by application of a retention factor. 4

5 Supporting Information Figures SI-Figure 1. Laboratory-based validation system 5

6 SI-Figure 2. GC/PDHID and micro-aircore/ms results for humidified and dry 12 CO 2 standards 6

7 SI-Figure 3. Underside of SkyWisp wing showing two micro-aircores installed for flight 7

8 SI-Figure 4. Raw mass spectrometric data demonstrating the operational concept for in situ measurement of collected gas with one valve (top) and both valves (bottom) open 8

9 SI-Figure 5. Volatile organic compounds are recovered from the micro-aircore tube at a concentration of 1 ppm (2 min intervals separated by 10 min of nitrogen 9

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY Nengbing Xu, Hongmei Ying and Libo Zhu Ningbo Environmental Monitoring

More information

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE. Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to

More information

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The

More information

Major Design Features

Major Design Features Major Design Features 3Flex Features Design Benefits Discussion 316 SS fittings and pneumatically actuated, hard-seal valves Provides virtually leak-free gas management with the lowest outgassing rate

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 9 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 3.1 Canister Receipt 3.2 Canister Storage 4.0 INTERFERENCE AND POTENTIAL PROBLEMS

More information

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Outline of Presentation Introduction Process Gas Analysis Magnetic Sector Mass Spectrometer Standard

More information

Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS

Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS TECHNICAL NOTE 1598 Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS Authors Amit Gujar, 1 Tim Anderson, 1 Daniela

More information

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer application Note Gas Chromatography/ Mass Spectrometry Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer Overview A mass spectrometer

More information

A New GC/MS System Designed for Helium Carrier Gas Conservation

A New GC/MS System Designed for Helium Carrier Gas Conservation A New GC/MS System Designed for Helium Carrier Gas Conservation Approaches for the Smarter Use of Helium Dale R. Walker GCMS Applications Specialist August, 7, 2013 1 Helium Facts How it's produced Where

More information

Retention Time Locking: Concepts and Applications. Application

Retention Time Locking: Concepts and Applications. Application Retention Time Locking: Concepts and Applications Application Gas Chromatography December 1997 Authors Vince Giarrocco Bruce Quimby Matthew Klee Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

Title: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator

Title: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator Procedure No: SOP-012 Revision No: 1.2, January 27, 2011 Page No.: 1 of 7 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Sabio model 2010 and 4010 dilution calibrators.

More information

Addressing the World Shortage of Helium

Addressing the World Shortage of Helium Addressing the World Shortage of Helium Introducing the Programmable Helium Conservation Module GPD Solutions June 2013 Market Situation Unreliable supply of helium worldwide and increasing prices have

More information

VIC offers a variety of Calibrated Gas Leaks

VIC offers a variety of Calibrated Gas Leaks CALIBRATED GAS LEAKS VIC offers a high quality line of calibrated gas leaks traceable to National Institute of Standards Technology (NIST). Calibrated gas leaks are used to provide a controlled delivery

More information

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1 Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC

More information

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc. The HumiSys RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The new HumiSys with single or dual RH probes capabilities is

More information

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1250 METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Total Organic Compounds N/A

More information

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,

More information

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA) Int. J. Metrol. Qual. Eng. 5, 303 (2014) c EDP Sciences 2014 DOI: 10.1051/ijmqe/2014012 Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA) D. Jonker and E.P. Tarnow

More information

Elimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU.

Elimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU. Elimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU. Control Analytics, Inc 6017 Enterprise Dr Export, PA 15632 (724) 387 CEMS (2367)

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCE AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Temperature controlled, multi-technique volumetric

More information

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 5.1 Subatmospheric

More information

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler Optimizing Vial Pressurization Parameters for the Analysis of Residual Solvents Using the 7697A Headspace Sampler Application Note Pharmaceuticals Author Roger L Firor Agilent Technologies, Inc.

More information

Multiple Gas#5 GC configuration Jan 2016

Multiple Gas#5 GC configuration Jan 2016 History: Unfortunately there is no single column that can separate: Hydrogen Oxygen Nitrogen Methane CO CO2 Ethane Water Propane Butane Pentane Over the years SRI Instruments has devised several solutions

More information

Best Practice for Identifying Leaks in GC and GC/MS Systems

Best Practice for Identifying Leaks in GC and GC/MS Systems Best Practice for Identifying Leaks in GC and GC/MS Systems Technical Overview Maintaining a leak-free GC or GC/MS system is critical for obtaining optimal system performance with reliable, reproducible,

More information

Zebron Gas Management

Zebron Gas Management Zebron www.phenomenex.com/gasmanagement What are the key benefits of Zebron filters and traps? The high adsorption capacity of the Zebron Filters removes contaminants from mobile phase gases, reducing

More information

Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability

Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability R. C. Kullberg*, D. J. Rossiter** Oneida Research Services, Inc., 8282 Halsey Road, Whitesboro, NY13492 Abstract Many

More information

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER Michael T Tang, Ph.D. Grace Feng Greg Merideth Rui Huang Matheson Gas Applied Lab

More information

Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer

Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer Ben D. Gardner, Phillip M. Erwin, Wai Tak Lee, Amber M. Tissandier, Souzan M. Thoresen Hamilton Sundstrand

More information

HumiSys HF High Flow RH Generator

HumiSys HF High Flow RH Generator HumiSys HF High Flow RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The HumiSys HF is a high flow version of the previously

More information

Generating Calibration Gas Standards

Generating Calibration Gas Standards Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and

More information

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Universal pycnometer, no

More information

Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types

Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types 19 th World Conference on Non-Destructive Testing 2016 Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types Klaus HERRMANN 1, Daniel WETZIG

More information

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the 1151 METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS 1.0 Scope and Application. 1.1 Analytes. Analyte Volatile Organic Compounds (VOC) CAS No. No CAS number assigned 1.2 Scope. This method

More information

Development of a high vacuum sample preparation system for helium mass spectrometer

Development of a high vacuum sample preparation system for helium mass spectrometer Journal of Physics: Conference Series Development of a high vacuum sample preparation system for helium mass spectrometer To cite this article: P Kumar et al 2012 J. Phys.: Conf. Ser. 390 012056 View the

More information

EAS Basic Soil Gas Sampler

EAS Basic Soil Gas Sampler EAS Basic Soil Gas Sampler Please Read The Following Important Information before Starting The EAS Basic Soil Gas Sampler has the traditional design where a pump or evacuated canister can be to purge the

More information

H 2. Sensor Series Take no risk! Install a Hydrogen Sensor in your Gaschromatograph

H 2. Sensor Series Take no risk! Install a Hydrogen Sensor in your Gaschromatograph Sensor Series 9000 Take no risk! Install a Hydrogen Sensor in your Gaschromatograph Carrier Gases for GC Probably more than 90% of the present GC instruments run with helium as carrier gas. Some people

More information

Gas Sampling in the DST

Gas Sampling in the DST UCRL-ID-129732 Gas Sampling in the DST L. DeLoach M. Chiarappa R. Martinelli B. Glassley January 12, 1998 Lawrence Livermore National Laboratory This is an informal report intended primarily for internal

More information

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD) Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet

More information

The instrument should have a mass range from 2 to 2000 m/z or better Interface

The instrument should have a mass range from 2 to 2000 m/z or better Interface Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for

More information

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010 Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600 November 2010 Table of Contents ``Abstract...3 ``Introduction...4 ``Methods & Results...6 ``Data Analysis...9 ``Conclusion...12

More information

Vapor Intrusion. Sampling Solutions VAPOR INTRUSION. Sampling Solutions

Vapor Intrusion. Sampling Solutions VAPOR INTRUSION. Sampling Solutions Vapor Intrusion Sampling Solutions VAPOR INTRUSION Sampling Solutions The All-In-One Soil Gas Sampling Solution Solutions for all 1 2 3 4 ZONES Indoor Air Quality Soil gas levels inside structures. Sub

More information

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation Using Nitrogen instead of Helium as carrier gas, while maintaining exact the same separation efficiency, retention times, peak elution order, without changing the oven temperature conditions. Jaap de Zeeuw

More information

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies IMPROVED CALIBRATION AND OPERATION OF A STABLE-GAS QUANTIFICATION MANIFOLD AS PART OF A RADIOXENON COLLECTION SYSTEM ABSTRACT Richard M. Williams, James C. Hayes, and Randy R. Kirkham Pacific Northwest

More information

The New PHD 4 Portable Helium Detector

The New PHD 4 Portable Helium Detector The New PHD 4 Portable Helium Detector SPECIAL FEATURES: High Sensitivity to Helium Easy to Use Truly Portable Versatile Dependable The system is designed to allow easy replacement of sampling line components

More information

Thermo Scientific Model 146i Principle of Operation

Thermo Scientific Model 146i Principle of Operation Thermo Scientific Model 146i Principle of Operation The Model 146i Multi-gas Calibrator dilutes calibration gases to precise concentrations. The diluted gases are used to perform zero, precision and Level

More information

In-Water Mass Spectrometry for Characterization of Light Hydrocarbon Seeps and Leaks

In-Water Mass Spectrometry for Characterization of Light Hydrocarbon Seeps and Leaks In-Water Mass Spectrometry for Characterization of Light Hydrocarbon Seeps and Leaks R. T. Short, S. K. Toler, A. M. Cardenas-Valencia, S. Untiedt, C. Cullins, M. Ryder, and J. Kloske HEMS Workshop 2015

More information

Weiyi Zheng, Xiaoxia Dong, Evan Rogers, Jimmie C. Oxley, and James L. Smith*

Weiyi Zheng, Xiaoxia Dong, Evan Rogers, Jimmie C. Oxley, and James L. Smith* Improvements in the Determination of Decomposition Gases from 1,3,3-Trinitroazetidine and 5-Nitro-2,4-dihydro-3H-1,2,4,-trizol-3-one using Capillary Gas Chromatography-Mass Spectrometry Weiyi Zheng, Xiaoxia

More information

Basic & Controlled Atmosphere Glove Boxes

Basic & Controlled Atmosphere Glove Boxes I N T R O D U C T I O N Precise Basic and Controlled Atmosphere Glove Boxes provide a leak-tight environment for work with contamination-sensitive materials. Precise Basic Glove Boxes are simple, economical

More information

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions Introduction: There is a bit of explanation of the health effects of benzene in the Introduction section of the first part

More information

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking

More information

Setting a New Standard in Flash Chromatography Performance

Setting a New Standard in Flash Chromatography Performance Setting a New Standard in Flash Chromatography Performance What s New Faster Flow rates, up to 300 ml/min Higher pressures, up to 300 psi (20 bar) Bigger touchscreen, 12 or 15 inches Wider Dynamic Range

More information

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide

More information

HiQ laboratory gas generators.

HiQ laboratory gas generators. Product information HiQ laboratory gas generators. The convenient alternative for laboratory environments. 02 HiQ laboratory gas generators HiQ laboratory gas generators. Designed with safety and convenience

More information

Experiment GC : Analysis of BTEX by GC-FID

Experiment GC : Analysis of BTEX by GC-FID 46 Experiment GC : Analysis of BTEX by GC-FID Learning Goals: Familiarity with gas chromatography Gain experience in temperature programming and method development Correctly use an internal standard, and

More information

Protocol for the verification tests on ethylene oxide treatment technology

Protocol for the verification tests on ethylene oxide treatment technology Pilot project for the environmental technology verification In the field of ethylene oxide treatment technology Protocol for the verification tests on ethylene oxide treatment technology Sept. 11, 2003

More information

The Helium Leak Detector

The Helium Leak Detector The Helium Leak Detector Helium Leak Detector Main Components The main components of a helium leak detector are: 1. The analyzed, which enables to separate the tracer gas from other gases inside leak detector.

More information

Gaseous oxygen technical grade Specification

Gaseous oxygen technical grade Specification CDC 09 (5010) P3 Rev. TZS 217:1984 FINALIZED TANZANIA STANDARD Gaseous oxygen technical grade Specification TANZANIA BUREAU OF STANDARDS. 2 nd edition TBS 2017 0. FOREWORD 0.1. This Tanzania Draft Standard

More information

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features ANALOG MASS FLOW CONTROLLERS Model AF mass fl ow controllers are designed to indicate fl ow rates and control set fl ow rates of gases. Each of these units incorporates an advanced straight tube sensor

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310 www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople

More information

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality

More information

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems Inorganic Analysis Chromatography Molecular Spectroscopy Thermal/Elemental Analysis Informatics Gas Chromatography MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass

More information

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013 NMISA NEW GAS FLOW CALIBRATION FACILITY Deona Jonker 7 October 2013 Introduction Flow the total volume of a fluid that flows past a fixed point in a given time Flow measurements used in industry as a means

More information

Gases and welding distributors continue

Gases and welding distributors continue Specialty Gas Instrumentation By Ray Borzio, Mike Lee, and Scott Varian Gases and welding distributors continue to have a significant future in the production, delivery, and related services for specialty

More information

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative

More information

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT APPLICATION NOTE Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT The INFICON 3000 Micro GC provides two fast and accurate solutions for the analysis of coal mine gas components.

More information

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS Abstract Highly sensitive analysis of metanephrines in plasma is critical in the diagnosis and treatment of pheochromocytoma and paraganglioma. Here, a HILIC LC-MS/MS method was developed using a Raptor

More information

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #:

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #: ID-Cube Installation and Operations Manual For ID-Cube Source and Power Supply Revision 2 Document #: 7.1.002 IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906 Copyright 2014 by IonSense Inc. All rights

More information

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion: Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion

More information

REFINERY GAS ANALYSIS BY GAS CHROMATOGRAPHY

REFINERY GAS ANALYSIS BY GAS CHROMATOGRAPHY REFINERY GAS ANALYSIS BY GAS CHROMATOGRAPHY UOP Method SCOPE This automated method is for determining the composition of refinery gas samples or expanded liquefied petroleum gas (LPG) samples obtained

More information

NMISA New Gas Flow Calibration Facility

NMISA New Gas Flow Calibration Facility NMISA New Gas Flow Calibration Facility Speaker / Author: D. Jonker Co-author: E.P. Tarnow National Metrology Institute of South Africa (NMISA) Private Bag X34, Lynnwood Ridge, Pretoria, 0040, South Africa

More information

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the

More information

CHAPTER 3 RESULTS AND DISCUSSION

CHAPTER 3 RESULTS AND DISCUSSION CHAPTER 3 RESULTS AND DISCUSSION The research included the development and verification of the earlier apparatus, and the collection and validation of moisture diffusion data under both isothermal and

More information

Pre-installation Checklist

Pre-installation Checklist Pre-installation Checklist Varian 1200/1200L Mass Spectrometer Use this checklist as a guide while preparing your site for the installation of the Varian 1200 or 1200L MS system. Make sure that your site

More information

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV 16-1 INTRODUCTION 16-1.1 Purpose. The purpose of this chapter is to familiarize divers with the techniques used to mix divers breathing gas. 16-1.2 Scope. This chapter

More information

User s Guide. Vacuum Controller for liquid delivery systems

User s Guide. Vacuum Controller for liquid delivery systems Flow Control User s Guide Vacuum Controller for liquid delivery systems Precise Vacuum Control throughout the experiment Flow control Compatible with any perfusion system Ideal for Small Volume Delivery

More information

APPLICATION NOTE. GC Integrated Permeation Device

APPLICATION NOTE. GC Integrated Permeation Device GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords:

More information

Bonding Reliability Testing for Wafer Level Packaged MEMS Devices

Bonding Reliability Testing for Wafer Level Packaged MEMS Devices Bonding Reliability Testing for Wafer Level Packaged MEMS Devices Michael Shillinger Vice President of Operations Innovative Micro Technology Santa Barbara, CA, USA mjs@imtmems.com October 20, 2011 Innovative

More information

VERIFYING GAS CHROMATOGRAPH OPERATION AT CUSTODY TRANSFER LOCATIONS Murray Fraser Daniel Measurement & Control

VERIFYING GAS CHROMATOGRAPH OPERATION AT CUSTODY TRANSFER LOCATIONS Murray Fraser Daniel Measurement & Control VERIFYING GAS CHROMATOGRAPH OPERATION AT CUSTODY TRANSFER LOCATIONS Murray Fraser Daniel Measurement & Control Houston, Texas ABSTRACT The on-line gas chromatograph (GC) has been widely used for natural

More information

CHEM 355 EXPERIMENT 7. Viscosity of gases: Estimation of molecular diameter

CHEM 355 EXPERIMENT 7. Viscosity of gases: Estimation of molecular diameter CHEM 355 EXPERIMENT 7 Viscosity of gases: Estimation of molecular diameter Expressed most simply, the viscosity of a fluid (liquid or gas) relates to its resistance to flow. The viscosity of a gas is determined

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor)

1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor) 1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor) 1. According to NFPA 99, SCFM is an acronym for. a) Surface Conditions at Fahrenheit Mercury b) Standard Conditions

More information

GCMSD-Headspace Analysis SOP

GCMSD-Headspace Analysis SOP Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to

More information

Real-time PCB Gas Monitoring System (MOHMS-21GP)

Real-time PCB Gas Monitoring System (MOHMS-21GP) Real-time PCB Gas Monitoring System (MOHMS-21GP) SHINSAKU DOBASHI* 1 YOSHIHIRO DEGUCHI* 1 MASAHARU KIRA* 2 RYUICHIRO TANAKA* 3 YOSHINORI IZAWA* 3 This report describes the technical characteristics and

More information

Carrier Gases in Capillary GC

Carrier Gases in Capillary GC Carrier Gases in Capillary GC LC Columns and Consumables Mark Sinnott Application Engineer January 15, 2009 CARRIER GAS Mobile Phase Carries the solutes down the column Selection and velocity influences

More information

Practical approach and problems in in-situ RGA calibration

Practical approach and problems in in-situ RGA calibration Practical approach and problems in in-situ RGA calibration Oleg Malyshev and Keith Middleman Vacuum Science Group, ASTeC Accelerator Science and Technology Centre STFC Daresbury Laboratory UK Workshop

More information

A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES

A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES F. Papa 1, J. Brindley 2, T. Williams 2, B. Daniel 2, V. Bellido-Gonzalez 2, Dermot Monaghan

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

What I have learned about SF 6 gas testing.a Practical explanation

What I have learned about SF 6 gas testing.a Practical explanation What I have learned about SF 6 gas testing.a Practical explanation Over the past few years, testing of SF6 gas quality has become increasingly important and with the EPA in USA, setting guidelines for

More information

1 Design of the Portable mass spectrometric sensor for vascular and endocrine disease diagnostics

1 Design of the Portable mass spectrometric sensor for vascular and endocrine disease diagnostics 1 Design of the Portable mass spectrometric sensor for vascular and endocrine disease diagnostics Andrei Antonov, Viktor Kogan, Aleksandr Krutikov A.F. Ioffe Physical Technical Institute, Academy of Science,

More information

Standard Operating Procedure for H 2 S and SO 2 Converter

Standard Operating Procedure for H 2 S and SO 2 Converter Procedure No: SOP-001 Revision No: 1.0 January 24, 2011 Page No.: 1 of 6 1. INTRODUCTION AND SCOPE This procedure describes the operation and configuration of heated converters used in continuous ambient

More information

Methods Used in Measuring Surface Seawater pco 2 Aboard RVIB Nathaniel B. Palmer

Methods Used in Measuring Surface Seawater pco 2 Aboard RVIB Nathaniel B. Palmer F:\NBPalmer\Palmer_methods.doc Page 1 09/16/02 11:55 AM 1. Introduction: Methods Used in Measuring Surface Seawater pco 2 Aboard RVIB Nathaniel B. Palmer Taro Takahashi, Colm Sweeney, Stewart C. Sutherland

More information

SINGULUS TECHNOLOGIES

SINGULUS TECHNOLOGIES Fast Vacuum for Optical Disc Metallization and Photovoltaic Coatings B. Cord, M. Hoffmann, O. Hohn, F. Martin SINGULUS Technologies AG, Kahl am Main SINGULUS TECHNOLOGIES June 2014 SINGULUS Overview 25-June-2014-2

More information

Total Pressure Measurement Using Residual Gas Analyzers

Total Pressure Measurement Using Residual Gas Analyzers Total Pressure Measurement Using Residual Gas Analyzers M. Maskell, Old Dominion University; G. Myneni, P. Adderley, Jefferson Lab; G. Brucker, Stanford Research Systems; C. Day, Forschungszentrum, Karlsruhe

More information

Marc McMullen. Physics Detector Support Group

Marc McMullen. Physics Detector Support Group Marc McMullen Physics Detector Support Group Hall B Gaseous Detectors 1. Drift Chamber (DC) 2. Low Threshold Cherenkov Counter (LTCC) 3. High Threshold Cherenkov Counter (HTCC) 4. Silicon Vertex Tracker

More information