To provide adequate disinfection for reusable semi-critical equipment other than scopes. Processing of scopes is covered in
|
|
- Cecily Spencer
- 6 years ago
- Views:
Transcription
1 Purpose To provide adequate disinfection for reusable semi-critical equipment other than scopes. Processing of scopes is covered in Audience Policy Guidelines for High- Level All those in the UTMB Health System who clean, disinfect, sterilize or store patient care instruments and other healthcare items. Reusable semi-critical equipment will be processed as described below and stored in a manner to prevent re-contamination. Equipment must always be cleaned per manufacturer s instructions for use (MFG IFU) prior to high-level disinfection. Transportation of contaminated equipment will be in impermeable containers labeled biohazard. 1. For those departments that perform high-level disinfection with chemicals, the area for reprocessing should be separated from the area where items are stored. If only two sinks other than the handwashing sink are available, one should be designated clean and the other dirty. 2. Cidex ortho-phthalaldehyde (OPA) is used for high level disinfection. OPA does not irritate the eyes and nasal passages, does not require exposure monitoring, has a barely perceptible odor and requires no activation. It will stain proteins grey and must be thoroughly rinsed to prevent discoloration of the patient s skin and inflammation of mucous membranes. 3. Cidex OPA solution should be stored in its original sealed container at controlled room temperature C (59-86 F) in a well-ventilated, low traffic area. The expiration date of the Cidex OPA solution is found on the original container as received from the manufacturer. Once opened the unused portion of the solution may be stored in the original container for up to 75 days until used. Calculate 75 days from when the original container was opened and record this date on the bottle. Verify that the expiration date on the original container is longer than 75 days. 4. Reprocessing should be performed away from patient care areas, and the area should be well-ventilated. The area must have sufficient air changes to prevent build up of vapor. Healthcare workers who process instruments in Cidex OPA must wear PPE including face shield, fluid-resistant gown and gloves. If latex gloves are used, don 2 pair of gloves. A single pair of gloves may be used if made from 100% synthetic copolymer, nitrate rubber or butyl rubber. 5. The bucket and tray systems used for OPA disinfection must be made from polypropylene, acrylonitrile-butadiene-styrene, polyethylene, glass-filled polypropylene and/or polycarbonate plastics. Prior to placement of instruments into the disinfectant solution, they must be thoroughly cleaned with a suitable detergent by brushing the surfaces to remove all blood, body fluids, tissue and any other foreign matter. Hinged instruments must be opened to permit thorough removal of all organic material. Lumens in instruments must be thoroughly brushed and irrigated until clean. 6. The soaking bucket or tray containing OPA solution must be labeled with name of Page 1 of 12
2 Processing Ultrasound Probes with Trophon EPR the solution, date of first use, and date of expiration. The instruments must be fully immersed, hinged instruments opened, instrument lumens filled with disinfectant and the cover closed. High-level disinfection requires 12 minutes for OPA. 7. Checking the MEC of the OPA solution must be monitored with a chemical test strip and results documented each time the solution is used. 8. The temperature of OPA solution must be checked and documented each time the solution is used. Minimum temperature required is 20ºC or 68ºF. The sole purpose of the Trophon EPR is to high level disinfect validated ultrasound probes (see section B2.1) according to the specified processes outlined in this manual. It is not intended for any other use. Do NOT use this device for any application other than its expressed purpose. The Trophon EPR together with the Sonex-HL is a high-level instrument grade disinfectant system. The Trophon EPR is NOT intended to reprocess single use devices The Trophon EPR is NOT intended to pre-clean ultrasound probes The cable management system is an accessory designed for use with the Trophon EPR (see section A4.2) Chemical Indicator use is required with every disinfection cycle. More information can be found in the Chemical Indicator Instructions for use, provided with the Chemical Indicator. The Trophon Printer, Trophon Connect, Trophon Wall Mount and the Trophon Cart are accessories designed for use with the Trophon EPR; contact your customer service representative or visit the Nanoionics website for additional information. Refer to Part D for information on the Trophon Print At the beginning of the cycle, the Trophon EPR creates an aerosol of concentrated hydrogen peroxide. This is quickly and evenly distributed over the surface of the probe, including very small crevices. This process provides thorough, high level disinfection of the shaft and the handle of the probe. The device breaks down the hydrogen peroxide into small amounts of water and oxygen and safely vents them into the external environment Pre-cleaning and transportation of probe from procedure room 1. Perform hand hygiene and don clean gloves. 2. Remove 2 appropriate MFG recommended wipes from container and place on clean counter. 3. Place 2 paper towels side-by-side on clean counter. 4. After procedure, remove condom from probe and discard condom in regular trash in exam room. 5. Use one wipe to remove excess ultrasound gel from probe. Use a cotton tipped applicator to remove any particles from grooves in the probe. Place probe on clean paper towels on counter. 6. Use second wipe and start with probe end and follow through to end of power cord. Inspect probe to ensure all residual ultrasound gel has been removed. Repeat step if necessary. 7. Place probe on clean paper towel. 8. Remove gloves and perform hand hygiene. 9. Don clean glove. 10. Transport clean trans-vaginal probe to Trophon room in a covered impermeable container labeled biohazard. High-level disinfection I. Load disinfectant cartridge Page 2 of 12
3 II. III. 1. Perform hand hygiene and don clean gloves and safety glasses. 2. Always check expiration date on the cartridge before use. If the cartridge has expired, contact EHS for disposal (ext Do not attempt to open or load a damaged cartridge or a distorted cartridge. Do not manually pierce the cartridge. 3. For a list of probes that may be disinfected in the Trophon EPR, refer to the following: Nanosonics Note: only validated probes may be placed in the Trophon EPR. 4. Visually inspect probe to ensure probe is free of all organic material and is dry BEFORE proceeding with the high-level disinfection process. If probe still contains organic material, repeat step 5 under cleaning. 5. Load cartridge 6. Cartridge door will automatically open when cartridge needs replacing. Do NOT force cartridge door open as this may damage and render the device nonfunctional. 7. Screen message: CARTRIDGE EMPTY. REPLACE CARTRIDGE NOW is displayed. 8. Press the button under OK to open the cartridge door. 9. Do not insert empty cartridges into the device. Remove gloves and safety glasses and perform hand hygiene. Note: refer to IFU enclosed with Trophon Sonex-HL for detailed instructions on installation and/or replacement of Trophon Sonex-HL cartridges. Note: cartridges will last for approximately one month from date of installation, depending on usage and whether the device has been switched off. 10. If the Trophon device is in Sleep Mode, press the restart button. The screen should display message WAKING UP FROM SLEEP. 11. When the device is ready, screen message will be LOAD PROBE AND INDICATOR. 12. Open the chamber door and load the probe, ensuring the probe is straight and not touching the walls or the bottom. The tip of the probe must be above the embossed line. 13. Refer to diagram in user s manual for loading the probe Load the probe wearing gloves, follow diagram in user s manual Load the Chemical indicator 14. See diagram in users manual. PPE required: gloves 15. A Chemical Indicator must be used for each disinfection cycle and can only be used once. Place the chemical indicator in the holder on the floor of the device chamber, red side up. (Note: verify expiration on indicator prior to use. 16. Follow the diagram for loading the chemical indicator in the MFG: IFU IV Process 1. Closing the Chamber Door If the door is not properly closed, screen message will be displayed CLOSE CHAMBER DOOR. 2. The next Screen message will be: IS THE PROBE CLEAN AND DRY? Respond YES if the probe has been cleaned and dried. 3. Once the device is ready, screen message: PRESS START TO BEGIN Press the START button to initiate the cycle. Screen message will be DISINFECTIONG. Remove gloves and perform hand hygiene.. High Level cycle will take 7 minutes to complete 4. When the cycle has been successfully completed, the device will sound an audible alarm. Screen message: CYCLE COMPLETE REMOVE AND WIPE PROBE. Perform hand hygiene and don clean gloves NOTE: If the probe is not removed within 1 minute of the cycle being complete, the warm-up time will increase for the subsequent cycle. IV. Check chemical indicator1. 1. Remove the used Chemical Indicator from the device and verify the chemical indicator color change from red to orange or lighter to validate successful high-level disinfection before discarding. Follow diagram in the user s manual Page 3 of 12
4 2. If the chemical indicator passes (refer to Chemical Indicator Insert), remove probe and wipe the probe with a lint-free cloth. Visually inspect the probe and ensure any peroxide residue is removed. Cove the probe with a peel pack. 3. Document the results of the chemical indicator and other required information on the log after each use. 4. Transport probe to probe rack for storage or store probe in probe rack located next to the Trophon EPR. Remove gloves and perform hand hygiene 5. Required information Printer labels: 2 labels will print after each cycle. Items a-c will automatically print on the labels. Items d-g must be recorded manually on the labels. Place one label on the Trophon Log Sheet and the other on the peel pack palced over the probe. a. Date and time of cleaning b. Cycle number c. Trophon LCD Status (pass or fail) d. Chemical indictor status (pass or fail) e. Operator s initials f. Probe # g. MRN of patient probe was used on 6. If chemical indicator fails, refer to section B2.7.2 in MFG IFU. 7. Incomplete or failed cycle: refer to MFG IFU section B5. Document corrective action next to notes section on printed label below patient s medical record number. 8. Routine Care and maintenance: refer to MFG IFU Section B6 9. Maintenance issues: contact CES and and notify supervisor. V. Remove disinfectant cartridge 1. Perform hand hygiene. Don clean gloves and safety glasses. 2. The device will automatically prompt to run a purge cycle if the cartridge has expired. 3. Screen message will say REPLACE THE CARTRIDGE AND CLOSE CARTRIDGE DOOR 4. Cartridge door opens automatically. Do not use excessive force to open. 5. Lift cartridge out by touching the areas exposed while the bottle is in the holder and avid touching pierced areas. 6. Empty used cartridge should be disposed of in nearest waste receptacle. TEE- High-level disinfection by immersion is the accepted method of infection control for Echocardiog transesophageal transducers. Philips recommends that a protective transducer cover be raphy Lab used during studies. This procedure covers cleaning and disinfecting the various parts of the transducer. Different solutions may be used on different parts. For a list of compatible disinfectants, see Choosing Disinfectants and Cleaning Solutions. Before performing this procedure, read Warnings and Cautions, Transducer Care and Operator Safety, and Warnings and Cautions for TEE Transducer Care. Cleaning and Disinfecting TEE Transducers 1. Disconnect the transducer from the system, and remove any accessories attached to or covering the transducer. 2. Clean the transducer distal tip and flexible shaft using any of the following: enzymatic cleaner, mild soap solution, or an approved wipe. Follow the manufacturer's instructions for use. Do not use iodine-based soaps. 3. Use water to gently rinse the distal tip and flexible shaft thoroughly. Do not rinse or Page 4 of 12
5 Other equipment not specified Off-Site Transportatio n of Sterile Items immerse the handle, steering mechanism, cable, or connector. 4. Choose a high-level disinfection solution compatible with your transducer. For a list of compatible disinfectants, see Choosing Disinfectants and Cleaning Solutions. Follow the label instructions for preparation and solution strength. If a pre-mixed solution is used, be sure to observe the solution expiration date. 5. Disinfect the distal tip and flexible shaft by placing them into the appropriate disinfectant as shown in the illustration. Follow the instructions on the disinfectant label for the duration of transducer immersion. Do not immerse transducers longer than the minimum time needed for your level of disinfection 6. Remove the distal tip and flexible shaft from the disinfectant and thoroughly rinse with water according to the instructions from the disinfectant manufacturer. 7. Check the transducer for any residual organic material. If any is present, remove it and disinfect the transducer again. 8. Gently dry the distal tip and flexible shaft with a sterile cloth or pad, or allow it to air dry. 9. Lightly wipe only the steering mechanism and handle with a pad moistened with rubbing alcohol (70% isopropyl alcohol), or use approved cleaners as directed on the handle and steering mechanism. 10. Use a soft cloth lightly dampened in a mild soap or detergent solution to clean the cable, the strain reliefs, and the connector. A soft-bristled brush can be used to clean only the metal surfaces of the connector. 11. Choose the disinfection solution compatible with your cable and connector. For a list of compatible disinfectants, see Choosing Disinfectants and Cleaning Solutions. Follow the label instructions for preparation and solution strength. If a pre-mixed solution is used, be sure to observe the solution expiration date. 12. Wipe or spray the cable, strain relief, and connector with the disinfectant, following disinfectant label instructions for wipe durations, solution strengths, and duration of disinfectant contact with the cable. Ensure that the solution strength and duration of contact are appropriate for the intended clinical use of the device. Ensure that the disinfectant solution does not enter the device. 13. Hang the transducer on a wall-mounted rack, and let it air dry. 14. Examine the transducer for damage, such as cracks, splitting, fluid leaks, or sharp edges or projections. If damage is evident, discontinue use of the transducer and contact your Philips representative Always follow MFG IFU. Minimum PPE when cleaning and disinfecting equipment: gloves. Perform hand hygiene and don new gloves when moving from dirty to clean part of process. Item must be clean prior to soaking in OPA. Rinse with filtered water, dry and store in accordance with MFG IFU and department protocol. Off-site transportation from SPD on the UTMB campus to the hospital/clinics will be carried out using closed containers, and closed containers will be used for items being returned to SPD. Page 5 of 12
6 References 1. Rutala WA, Weber DJ. Selection and Use of Disinfectants in Healthcare, In Mayhall, CG, Ed. Hospital Epidemiology & Infection Control. Fourth Edition, Lippincott Williams and Wilkins, Association for the Advancement of Medical Instrumentation. Quality Control 2010:10: Young M. Quality control of table-top steam sterilizers. Managing Infection Control 2007; Page 6 of 12
7 QUALITY CONTROL PROCEDURE FOR CIDEX OPA TEST STRIPS Processing Steps: STORAGE AND USE OF OPA TEST STRIPS CIDEX OPA TEST STRIPS must be discarded 90 days after the bottle is opened or on the expiration date of the bottle, whichever comes first. The test strip bottle must not be left open for more than 30 minutes. Keep the bottle tightly closed after removing the test strip. Do not refrigerate or freeze the bottle of test strips Protect the strips from heat, light and moisture PREPARATION OF CONTROL SOLUTIONS Use full strength CIDEX OPA Solution as a positive control. To prepare a negative control, dilute 1 part CIDEX OPA Solution with 1 part of water. TESTING PROCEDURE Submerge 3 test strips in each freshly prepared solution for 1 second each. Remove. The 3 strips dipped in full strength positive control solution should exhibit a complete purple color on the indicating pad at 90 seconds. The 3 strips dipped in the diluted negative control solution should either remain completely blue or exhibit an incomplete color change to purple at 90 seconds. Refer to the color chart on the test strip bottle for interpretation of results. TESTING FREQUENCY The testing of positive and negative controls should be performed on each newly opened bottle of CIDEX OPA Solution Test Strips. Testing of freshly prepared positive and negative controls should be performed when test strips are improperly stored or handled. UNSATISFACTORY QC TEST PERFORMANCE If the results obtained from using positive and negative controls indicate the test strip is not functioning properly, discard strips. DO NOT USE STRIPS. Chec k () Comments/follow-up Surveyor s Name: Surveyed HCW s Initials: Survey Date: Surveyed Area: Page 7 of 12
8 CIDEX OPA TEST STRIPS QUALITY CONTROL LOG Location/Department: Purpose: Document performing quality control of the Cidex OPA Quality Control strips when a new bottle of strips is opened. Definitions: a Expiration date is printed on the label by the manufacturer b Discard date is calculated by adding 90 days to the open date of the strips c Quality control means both positive and negative controls are performed each time a new bottle of QC strips is opened. Procedure to Prepare the Positive & Negative Controls: POSITIVE Control - use full strength, activated Cidex OPA (30 ml); NEGATIVE Control is one part activated Cidex OPA (15 ml) and one part water (15 ml). Timing is critical. Insert 3 strips into each of the two control solutions for (1) second. Remove excess solution by standing up the strips. Read results in exactly ninety (90) seconds. Controls Results: To PASS on the positive controls, the dipped QC strip(s) must turn purple; if any blue appears on indicating pad apart from the top line, the solution did not pass ( FAILS ); to PASS on the negative controls, the strip color is blue or blue/purple. *Corrective Action Key: use the following key to denote action if either control did not pass: D: Cidex OPA Solution Test Strips discarded today because test failed; D1: Cidex OPA Solution Test Strips discarded today because bottle was left open; D2: Cidex OPA Solution Test Strips discarded today because bottle was not dated when opened. Test Date Year Cidex OPA Solution Lot# and Manufacturer a Expiration Date Cidex OPA Strips Lot# and Manufacturer a Expiration Date Cidex OPA Strips: Date Opened and Calculated b Discard Date c Positive Quality Controls All 3 results:(pass/fail) Pass = only purple color c Negative Quality Controls All 3 results:(pass/fail) Pass = Blue or blue/purple Print Employee Name performing Quality Controls *Corrective Action Key (use reverse side for comments) Page 8 of 12
9 PROTOCOL FOR HIGH-LEVEL DISINFECTION CIDEX OPA Chec Processing Steps: k () CIDEX OPA CONTAINER Label with chemical name (if not official Cidex container). Cover fluid at all times. MSDS sheets for Cidex OPA should be easily accessible. Label container with expiration date and initial each time the chemical is changed. Cidex OPA must be discarded at 14 days even if the OPA Test Strips indicate that there is a minimally effective concentration (MEC) in the Cidex OPA solution. TEST STRIPS Date bottle of test strips when opened (expiration at 90 days or the expiration date on the bottle of test strips, whichever comes first). Cidex OPA working solution must be tested EACH TIME when medical instruments are reprocessed. Dip test strip into solution for 1 second. Do not shake strip after removal. Remove excess solution from the pad by standing the strip upright on a paper towel. Read the results of the chemical reaction 90 seconds after the test strip has been removed from the solution. The entire indicating pad must be completely purple to pass the test indicating an effective concentration of the solution. Compare this color reaction to the color chart on the side of the test strip bottle. If any blue appears on the indicating pad apart from the top line, this is a failure, verifying the solution is below the MEC and should be discarded. Document the results of the test strip. If the color of the test strip matches the fail panel on the color chart: Check the test strips expiration date. If the test strips are out of date, discard and retest with new strips. If test strips are in-date, discard solution, rinse out container, and refill with fresh solution. Retest new solution and document test results. Cidex OPA working solution temperature must be checked each time medical instruments are reprocessed. SOAKING IN CIDEX OPA Cidex OPA is a high-level disinfectant. Do not soak instruments in Cidex OPA prior to steam or plasma gas sterilization Do not soak single-use disposables. They must be discarded after use All instruments that cut or biopsy must be sterilized (do not soak in Cidex OPA). Instruments that are soaked in Cidex OPA must be pre-washed in enzymatic solution prior to high-level disinfection. ALL particulate matter must be removed (brushes and enzyme cleaners are available for this process) Instruments must be TOTALLY SUBMERGED in the solution. All lumens must be filled and air pockets eliminated. Instruments must soak for 12 MINUTES using a timer at a minimum of 68ºF or 20ºC. After removal of instruments from the Cidex OPA solution, they must be Comments/follow-up Page 9 of 12
10 rinsed in 2 gallons of *filtered tap water. The devices must be submerged for at least one minute. All lumens must be irrigated with at least 100 ml of water. After the initial rinse, the water must be used for no other purpose and discarded. This rinse process will be repeated with 2 gallons of fresh *filtered tap water 2 more times for a total of three (3) rinses. * 0.2µ filter Surveyor s Name: Surveyed HCW s Initials: Survey Date: Surveyed Area: Page 10 of 12
11 Page 11 of 12
12 QUALITY CONTROL PROCEDURE FOR 3M COMPLY RAPICIDE MONITORS PREPARATION OF CONTROL SOLUTIONS To prepare positive and negative control solutions for testing, obtain a container of Rapicide High-Level Disinfect and Sterilant solution. Verify that the solution has not met or exceeded the labeled expiration date. For the positive control, the full strength solution should be used. To prepare a negative control, dilute one part of full strength Rapicide High-Level Disinfectant and Sterilant solution with one part of room temperature water. Label each control solution appropriately. TESTING PROCEDURE Following the directions for use, dip three indicator strips in each of the above freshly prepared solutions. The three strips dipped in the full strength positive control solution should exhibit a complete red color on the indicator pad when read at 75 seconds. The three strips dipped in the diluted negative control should either remain completely yellow or exhibit an incomplete color change to red when read at 75 seconds. Refer to the visual standard on the test strip bottle for interpretation of the results. This visual is also displayed below. Chart here?: TESTING FREQUENCY It is recommended that the testing of positive and negative controls be performed on each newly opened test strip bottle of Rapicide Glutaraldehyde Indicator test strips. After this initial testing, it is recommended that testing of freshly prepared positive and negative controls be performed on a regular basis as established by your own quality control procedures and program. This testing program will serve to minimize errors between different users, use of outdated materials or product that has been improperly stored or handled. UNSATISFACTORY QC TEST PERFORMANCE If the results obtained from using the positive and negative controls indicate the test strip is not functioning properly, discard the remaining strips. DO NOT USE. For customer support, contact Medivators at Page 12 of 12
13 QUALITY CONTROL PROCEDURE NOTE: Single use disposables must not be reprocessed. Instruments that cut or biopsy must be sterilized. DEPARTMENT OF HEALTHCARE EPIDEMIOLOGY Page 13 of 12
Shukla Medical. TITLE: Manual Surgical Orthopedic Instruments Recommendations for Care, Cleaning, Maintenance and Sterilization
Page 1 of 5 1. PROCESSING INSTRUCTIONS These processing instructions are intended to assist the hospital and central supply management in developing procedures for safe and effective reprocessing of both,
More informationTJF-Q180V Cleaning and Disinfection Checklist
TJF-Q180V Cleaning and Disinfection Checklist TJF-Q180V Cleaning and Disinfection Checklist This checklist is used to evaluate and confirm if cleaning and disinfection of the TJF-Q180V has been performed
More informationTJF-Q180V Cleaning and Disinfection Checklist
TJF-Q180V Cleaning and Disinfection Checklist TJF-Q180V Cleaning and Disinfection Checklist This checklist is used to evaluate and confirm if cleaning and disinfection of the TJF-Q180V has been performed
More informationEndo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax
OPERATION MANUAL Zutron Medical, LLC 17501 W 98 th St #40-27 Lenexa, KS 66219 Phone 877-343-5873 Fax 913-967-5944 ZUT-Lab-004-30004 REV. 03312017 Table of Contents 2 Introduction 1. Intended Use 2. Labels,
More informationMaintenance and Calibration
Maintenance and Calibration PeriFlux 6000 peripheral pressure made intelligent 44-00315-01 Disclaimer The information contained in this document is intended to provide general information only. It is not
More informationCavitron DualSelect. Introduction ENGLISH. Directions For Use. Cavitron DualSelect Dispensing System ENGLISH 1
ENGLISH Cavitron DualSelect Dispensing System Directions For Use Please read carefully and completely before operating unit. Introduction Cavitron DualSelect Dispensing System Congratulations. Your purchase
More information6900 Maintenance Instruction System Flush
Equipment Required FA74005 Damper Drain Tube FA16005 Cover Removal Tool FA900005 Beaker 0.25 Litre FA900003 Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves Latex
More informationGLIDERITE RIGID STYLET. Operations & Maintenance Manual
GLIDERITE RIGID STYLET Operations & Maintenance Manual 0900 4686 00 60 GLIDERITE RIGID STYLET Operations & Maintenance Manual Effective: March 4, 2016 Caution: Federal (United States) law restricts this
More informationVCH Ebola Virus Disease Health Care Worker PPE Checklist/Sign In-Out Sheet
The trained observer must complete this document for EVERY instance of a staff member entering and exiting rooms where persons are suspected or confirmed of having EVD. The trained observer must receive
More informationTEMPLE UNIVERSITY ENVIRONMENTAL HEALTH AND RADIATION SAFETY
Page 1 of 9 ISSUED: 5/00 REVISED: 08/06 Introduction Purpose: In accordance with applicable regulations and Temple University, this policy was developed to minimize exposure to Ethylene Oxide. Applicability
More informationWorking Safely with Hydrofluoric Acid
SAFE WORK PROCEDURE Insert Reference Code: UBC-RMS-OHS-SWP-17-002 Department of www.rms.ubc.ca Effective date: November 27, 2017 Review date: NA Supersedes: NA Working Safely with Hydrofluoric Acid 1.
More informationLM-900 Cryosurgery System
LM-900 Cryosurgery System Models 50501 and 50502 Directions for Use LM-900 Cryosurgery System Directions for Use INTRODUCTION Model 50501 requires Medical Grade N 2 O (Nitrous Oxide) gas. Model 50502 requires
More informationWolfson Campus Natural Science Department. Biology Laboratory. Student Safety Contract
Wolfson Campus Natural Science Department Biology Laboratory Student Safety Contract July 2000 Wolfson Campus Student Safety Contract-Biology Laboratory Purpose The Biology laboratory is a hands-on learning
More informationEntry to these spaces must be rigorously controlled to prevent serious injury or death.
Chapter 6 - Confined Spaces Chapter 6 Confined Spaces Definitions A confined space is defined as any structure that must be entered and that has or may contain dangerous concentrations of hazardous gases
More informationCompetency Record for Sonic Cleaning
Competency Record for Sonic Cleaning Name: Competency Achieved: (Date) Evaluator: Learner: Comments: Background information on sonic testing ST 79; 2009A section 7.5.3.3 states this on weekly testing Mechanical
More informationClient Care Items and Equipment
Client Care Items and Equipment Use disposable, single-use critical and semi-critical medical devices and equipment when possible. Single-Use a critical or semi-critical medical device, designated by the,
More informationSuction Regulators. Recommendations for Cleaning, Disinfection and Reprocessing: Boehringer 3800 &3900 Series Suction Regulators
Suction Regulators Recommendations for Cleaning, Disinfection and Reprocessing: Boehringer 3800 &3900 Series Suction Regulators Boehringer Laboratories, LLC 300 Thoms Drive Phoenixville, PA 19460 (800)
More informationInstructions for Cleaning Equipment: Low Level Disinfection. Beaumont Hospital, Dearborn Infection Prevention & Control 2017
Instructions for Cleaning Equipment: Low Level Disinfection Beaumont Hospital, Dearborn Infection Prevention & Control 2017 Low Level Disinfection Low level disinfection (LLD) is appropriate for any noncritical
More informationInstructions for Use. OptiLife Nasal Lab Mask
Instructions for Use OptiLife Nasal Lab Mask Intended Use The OptiLife Nasal Mask is intended to provide an interface for application of CPAP or bi-level therapy to patients. The mask is for multi-patient
More informationCabinet. 1.0 POLICY It is the policy of Calvin College to comply with the MIOSHA Respiratory Protection Standard Part 451.
Calvin College Respiratory Protection Program Prepared by: Date: Heather Chapman 10/7/10 Approved by: Cabinet Revision 2 Date: 1.0 POLICY It is the policy of Calvin College to comply with the MIOSHA Respiratory
More informationHemoCue Hb Procedure Template
HemoCue Hb 201 + Procedure Template PURPOSE The HemoCue Hb 201 + System is used for the quantitative determination of hemoglobin in blood using a specially designed analyzer, HemoCue Hb 201 +, and specially
More informationAMBLER SURGICAL INSTRUCTIONS FOR USE - OPHTHALMIC DUET-STYLE SCISSORS
AMBLER SURGICAL INSTRUCTIONS FOR USE - OPHTHALMIC DUET-STYLE SCISSORS Ambler Item # 6112E Duet-style MICS Packer/Chang IOL cutter, 1 3/8", 19 gauge, straight shaft, for use with Ambler #6100T INDICATIONS
More informationINDUSTRIAL - SOLVENT PARTS CLEANER
INDUSTRIAL - SOLVENT PARTS CLEANER OPERATION & INSTRUCTION MANUAL I. Introduction A. Manufacturer s Statement CM400 This manual will provide you with important information about the Solvent Parts Cleaners
More informationHydrofluoric Acid SOP
Introduction Hydrofluoric acid (HF) has a number of chemical, physical, and toxicological properties, which make handling this material especially hazardous. Anhydrous HF is a clear, colorless, fuming,
More informationEbola Virus Disease PPE Checklist/Sign In-Out Sheet Higher Transmission Risk Option 2: PAPRS
This document must be completed by the trained observer for EVERY staff member entering and exiting into a suspected or confirmed Ebola patient room. The trained observer must receive specific training
More informationE3628 THE RESPIRATORY PROTECTION PROGRAM: EMPLOYEE TRAINING. Leader s Guide ERI Safety Videos EMPLOYEE TRAINING
E3628 THE RESPIRATORY PROTECTION PROGRAM: EMPLOYEE TRAINING Leader s Guide 2007 ERI Safety Videos THE RESPIRATORY PROTECTION PROGRAM: EMPLOYEE TRAINING This easy-to-use Leader s Guide is provided to assist
More informationFrigitronics CRYO-PLUS
Frigitronics CRYO-PLUS OPERATING & MAINTENANCE MANUAL REF 2400 R x Only CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. INDICATIONS FOR USE The Cryo-Plus is
More informationSTANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA
STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1157.1 Date Issued: 05/14 Date Revised: 5/15 Page 1 of 6 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: SurgiVet
More informationLABORATORY FUME HOOD SPECIFICATION AND MAINTENANCE PROGRAM
LABORATORY FUME HOOD SPECIFICATION AND MAINTENANCE PROGRAM Last Updated: Tuesday, January 31, 2006 1. INTRODUCTION AND SCOPE 2. RESPONSIBILITIES 3. USER GUIDELINES 4. SYSTEM DESIGN SPECIFICATIONS o The
More informationINSTRUCTION MANUAL. Gas Salamander
INSTRUCTION MANUAL Gas Salamander This manual contains important information regarding your unit. Please read this manual thoroughly prior to equipment set-up, operation and maintenance. Failure to comply
More informationThe health hazards of HF are dependent upon the concentration, temperature and nature of exposure.
Hydrofluoric Acid Safe Handling and Use Hydrofluoric acid (HF) is a corrosive acid used for many processes including mineral digestion, surface cleaning, glass etching, and biological staining. HF is a
More informationTITLE: Pulse Oximetry COMPETENCY #: Resp #1 NEW COMPETENCY REVISION DATE: 10/18/12 EMPLOYEE NAME. DATE INITIAL RE-EVALUATION ANNUAL (if required) PRN
: Pulse Oximetry COMPETENCY #: Resp #1 NEW COMPETENCY REVISION : 10/18/12 INITIAL RE-EVALUATION ANNUAL (if required) PRN 1. Verifies physician order. 2. Gathers equipment and supplies. 3. Knocks on door
More informationPrestige Endoscopic Graspers Instructions for Use
Prestige Endoscopic Graspers Instructions for Use DESCRIPTION The Prestige Endoscopic Graspers are designed to grasp soft tissue during endoscopic procedures. The graspers are designed to be used through
More informationF-330 Instruction manual. Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic
F-330 Instruction manual Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic Warning Never under any circumstances attempt to open or inspect the internal components or accessories of the
More informationCCM Approval: IACUC Approval:
CCM Approval: IACUC Approval: 1.0 Purpose: The purpose of this SOP is to describe the necessary procedures for the safe administration of potentially hazardous agents 1 to rodents via injection, drinking
More informationEssential Chemical Safety & Hygiene Training 2015
Essential Chemical Safety & Hygiene Training 2015 Introduction What does this training cover? Knowledge of working with cleaning chemicals Forms part of induction training Forms part of on-going job training
More informationThe primary purpose of laboratory fume hoods is to keep toxic or irritating vapors out of the general laboratory working area.
Introduction Welcome to the Biosafety Cabinets and Fume Hoods Training Course (OHS_BIO304). This training is required for anyone that will be conducting work and/or research that requires the use of a
More informationAir-Mate High Efficiency Powered Air Purifying Respirator (PAPR) Staff Training Module
Air-Mate High Efficiency Powered Air Purifying Respirator (PAPR) Staff Training Module PRES0389 7/05-1 Powered Air Purifying Respirators (PAPR) Reduce the risk of transmission of infectious agents spread
More informationCurrent versions of approved documents are maintained in Livelink. Printed copies are uncontrolled. Page 1 of 6
SARNIA REFINERY REPAIRING OR SERVICING ALKYLATION UNIT EQUIPMENT BY OFFSITE VENDORS PROCEDURE Procedure Type: Routine Document Owner: Manager, EH&S Department: Environmental, Health and Safety Document
More informationOperation Manual. H & R Healthcare 1750 Oak Street Lakewood, NJ Phone: Fax:
Operation Manual H & R Healthcare 1750 Oak Street Lakewood, NJ 08701 Phone: 800-801-5533 Fax: 732-367-6231 www.handrhealthcare.com Congratulations and thank you for purchasing this Relief Aire Alternating
More informationMcLane / Black Lake Fire Department Policy & Procedural Guideline PPG #
Subject: EMERGENCY DECONTAMINATION Section: PPG# 4400.55 Chapter: Operations Effective Date: 11/12/2015 1.0 POLICY 1.1 1.2 1.3 It shall be the policy of the McLane Black Lake Fire Department to provide
More informationVarithena Delivery System Instructions for Use INSTRUCTIONS FOR USE Please read the Product Monograph before using the product.
Varithena Delivery System Instructions for Use INSTRUCTIONS FOR USE Please read the Product Monograph before using the product. The Instructions for Use are for the entire Varithena TM system. There are
More information14500 Coy Drive, Grass Lake, Michigan Isolation Glove Box Instruction Manual
14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Isolation Glove Box Instruction Manual Index Page Warranty 1 1.0 Warnings 2 2.0 General Overview 3 3.0 Care
More informationRespiratory Protection. Program
Respiratory Protection Program Respiratory Protection Program Table of Contents A. General B. Selection of Respirators C. Medical Screening D. Fit Testing E. Respirator Use Instructions F. Respirator Inspection
More informationStandard Operating Procedures (v9 5/1/17)
Standard Operating Procedures (v9 5/1/17) Safe Working Practices for Cell Sorting on the Influx Table of Contents I. General Facility Information II. Facility Orientation and Training III. Approval to
More informationA. APPLICATIONS 1. WARNING TO USER
IV. MOLDEX INDUSTRIAL 42 CFR 84 DISPOSABLE RESPIRATOR USER INSTRUCTIONS Each specific model may contain additional warnings and instructions. You MUST refer to the packaging for specific instructions.
More informationOPERATION MANUAL. Hand Held Leak Tester Model # ZUTR-10002
OPERATION MANUAL ZUT-Lab-003-10002 REV. B Table of Contents Page 2 Introduction 1. Intended Use 2. Labels, Symbols and Signal Words Page 3 Page 4 Page 5 Important Safety Information and Cleaning Procedure
More informationDate Effective: Date Reviewed: Nebraska Biocontainment Unit. Doffing Biological Protective Suit and PAPR
Section: Subject: POLICIES AND PROCEDURES MANUAL Number: 1093 Biocontainment Unit Doffing PPE with PAPR (Biological Protective Suit and Powered Air Purifying Respirator) System Department Attachments:
More informationSUGGESTED POLICY FOR THE MONITORING OF THE CLEANING EFFICIENCY OF AN ULTRASONIC CLEANERS( SONIC & TOSI) WEEKLY
SUGGESTED POLICY FOR THE MONITORING OF THE CLEANING EFFICIENCY OF AN ULTRASONIC CLEANERS( SONIC & TOSI) WEEKLY SUBJECT: ULTRASONIC CLEANER MONITORING WEEKLY DEPARTMENT: Central Service APPROVED BY: EFFECTIVE:
More informationSpecimen Collection. Lead Screen Filter Paper Testing Dried Blood Spots (DBS) Laboratory Services February 2017
1 Specimen Collection Lead Screen Filter Paper Testing Dried Blood Spots (DBS) Laboratory Services February 2017 2 Agenda Why Use Lead Screen Filter Paper Testing? Before You Begin Collection Drying &
More informationMANITOBA RENAL PROGRAM
MANITOBA RENAL PROGRAM SUBJECT Fresenius AquaUNO Reverse Osmosis System; use of SECTION CODE 30.30.04 30.30 Water Treatment AUTHORIZATION Professional Advisory Committee, Manitoba Renal Program Nursing
More information6800 Maintenance Instruction System Flush Procedure
Equipment Required FA74005 FA65318 FA900005 FA900003 Damper Drain Tube 6800 Cover Removal Tool Beaker 0.25 Litre Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves
More informationNewport News Shipbuilding Contractor Environmental, Health and Safety Resource Manual Sewage
Newport News Shipbuilding Contractor Environmental, Health and Safety Resource Manual Sewage Newport News Shipbuilding Page 1 SEWAGE 1. Hazard, Reference and Requirements a. Sewage is a potential hazard
More informationUse and Care Guide. PMU and PMU w/over the Patient. Swing Mount. Over the Patient. Side Delivery
Swing Mount Use and Care Guide PMU and PMU w/over the Patient Over the Patient Side Delivery Table of Contents Page General Information 2 Operation 3-5 Dental Unit Adjustments 6-9 Cleaning, Disinfecting
More informationStandard Operating Procedure (SOP)
INTRODUCTION This SOP applies to. FORMALDEHYDE exposure in Anatomy Lab will be monitored at the beginning of each semester dissections are performed. Environmental Health and Safety will perform monitoring
More informationBREATHE SAFELY: The Proper Use of Respiratory Protection
BREATHE SAFELY: The Proper Use of Respiratory Protection This easy-to-use Leader s Guide is provided to assist in conducting a successful presentation. Featured are: INTRODUCTION: A brief description of
More informationLaboratory Safety Training. Environmental Health and Safety
Laboratory Safety Training Environmental Health and Safety Why You Need Lab Safety Training Hazardous materials and/or equipment. You need to understand the risks and how to mitigate them. OSU requires
More informationStandard Operating Procedure (SOP)
Standard Operating Procedure (SOP) For using Cryogenics (Liquid Nitrogen) #1 General Process Description Liquid nitrogen is used for many different applications in the department of Materials Science and
More informationLaboratory Safety Guideline Emergency Eyewash and Safety Showers
Introduction The purpose of this program is to ensure that all safety eyewashes and showers supply clean, potable water and are in proper working order. This program describes procedures for emergency
More informationBIOSAFETY SELF AUDIT PRINCIPAL INVESTIGATOR/PERMIT HOLDER: CONTAINMENT LEVEL: 1 2 (Shaded) RISK GROUP: 1 2 AUDIT TO BE COMPLETED BY (DATE):
Page 1 of 5 BIOSAFETY PROGRAM BIOSAFETY SELF AUDIT PRINCIPAL INVESTIGATOR/PERMIT HOLDER: SELF AUDIT COMPLETED BY: LABORATORY ROOM NUMBER: CONTAINMENT LEVEL: 1 2 (Shaded) RISK GROUP: 1 2 DATE OF AUDIT:
More informationChemical Hygiene Plan
Chemical Hygiene Plan Mid-Michigan District Health Department Page 1 of 12 Effective Date: 10/1/2011 The Chemical Hygiene Plan is required by Part 431 of the Michigan Occupational Safety and Health Act
More informationOPERATOR S MANUAL HOTLINE 3 Blood and Fluid Warmer
l OPERATOR S MANUAL - HOTLINE 3 Blood and Fluid Warmer < HL-390-38 6 s l HOTLINE 3 Blood and Fluid Warmer < HL-390-38 OPERATOR S MANUAL PN 4534013EN Rev. 002 s G e n e r a l I n f o r m a t i o n HOTLINE
More informationPhalanx Alpha Phalanx Beta and Ultra Elite
Phalanx Alpha Phalanx Beta and Ultra Elite riot control agent gas masks instructions parts list NIOSH approved for respiratory protection in atmospheres containing CN and CS and particulates under the
More informationEffective Date: 24-April-2017 Version: 6 Page: 1 of 8 TITLE: SAF-370, PERSONAL PROTECTIVE EQUIPMENT GUIDELINES. Table of Contents
24-April-2017 Version: 6 Page: 1 of 8 Table of Contents 1.0 Introduction... 2 1.1 Purpose... 2 1.2 Scope... 2 1.3 Definitions... 2 1.4 Responsibilities... 3 Environmental Health and Safety Office (EHSO),
More informationIdentify, evaluate and assess workplace improvements to reduce hazards. Provide personal protective equipment based on hazard assessment
PAGE 1 OF 7 1.0 INTRODUCTION The purpose of this program is to help employees work safely in their environments by identifying and assessing workplace hazards, eliminating or reducing the hazards where
More informationLiquid Oxygen- Helios & Helios Marathon A Guide for Patients in the Home
Liquid Oxygen- Helios & Helios Marathon A Guide for Patients in the Home Contact Name: Phone number: Table of Contents Contents Receiving Liquid Oxygen in the Home... 2 Instructions for Liquid Oxygen-Helios
More informationCentrifugation: High Speed Floor Centrifuges
Use this form to assist you to complete risk assessments for hazardous activities and processes. Any serious or ongoing hazards should be reported via RiskWare to ensure that appropriate corrective actions
More information1.0 PURPOSE The purpose of this procedure is to describe the use of basic aseptic technique.
1.0 PURPOSE The purpose of this procedure is to describe the use of basic aseptic technique. 2.0 RESPONSIBILITY 2.1 It is the responsibility of each employee working in a laboratory setting to follow this
More informationEnglish. ivolve Full Face Mask. User Manual
ivolve Full Face Mask User Manual 0123 Table of Contents ivolve Full Face Mask... 1 Intended Use... 1 Before Using the Mask... 2 Parts of the ivolve Full Face Mask... 4 Getting the Right Cushion Size...
More informationHi Flow Stroller H I F L O W S T R O L L E R
Hi Flow PAT I E N T O P E R AT I N G I N S T R U C T I O N S Stroller H I F L O W S T R O L L E R Warning Information Read this page carefully before using your unit. The unit contains liquid oxygen which
More informationRECOMBINATE. [Antihemophilic Factor (Recombinant)] (For intravenous use only)
Website for more information: http://www.recombinate.com/ INSTRUCTIONS FOR USE RECOMBINATE [Antihemophilic Factor (Recombinant)] (For intravenous use only) Do not attempt to do an infusion to yourself
More informationX. PERSONAL PROTECTIVE EQUIPMENT
X. PERSONAL PROTECTIVE EQUIPMENT 1.0 Personal Protective Equipment Defined 1.1 Personal Protective Equipment (PPE) includes all clothing and work accessories designed to protect employees from workplace
More informationCounty of Santa Clara Emergency Medical Services System
County of Santa Clara Emergency Medical Services System Policy # 700-S01 Ebola Virus Prevention and Control EBOLA VIRUS PREVENTION AND CONTROL Effective: October 30, 2014 Replaces: October 8, 2014 (Version
More informationUse of Biosafety II cabinets
RISK ASSESSMENT FORM Use this form to assist you to complete risk assessments for hazardous activities and processes. Any serious or ongoing hazards should be reported via RiskWare to ensure that appropriate
More informationKaty Independent School District Science Safety Rules
Katy Independent School District Science Safety Rules PURPOSE Science is a hands-on laboratory class. You will be doing many laboratory activities which may require the use of hazardous chemicals or potentially
More informationMONT BLANC 10400X XL XLED 10401X XL XLED 10403X XL XLED
MONT BLANC CONTR-ANGL D IMPLANTOLOGI 20:1 IMPLANTOLOGY CONTRA-ANGL 20:1 CONTRAANGULO PARA IMPLANTOLOGIA 20:1 CONTRANGOLO D IMPLANTOLOGIA 20:1 IMPLANTOLOGI WINKLSTÜCK 20:1 CONTRA-ÂNGULO PARA IMPLANTOLOGIA
More informationStandard Operating Procedure
Standard Operating Procedure Cleaning Ni Columns Department: Date SOP was written: Date SOP was approved by PI/lab supervisor: 12/29/2012 Reviewed by: Principal Investigator: Internal Lab Safety Coordinator/Lab
More informationSTANDARD OPERATING PROCEDURE TEMPLATE
STANDARD OPERATING PROCEDURE TEMPLATE #1 CONTACT INFORMATION: Procedure Title Procedure Authors Impedance Analyzer User Protocol Yu-Hung Li & Sang Jong Kim Date of Creation/Revision 11-12-2011 Name of
More informationLaboratory Self Inspection Checklist LAB01
Laboratory Self Inspection Checklist LAB01 General Chemical/Biological Laboratories Location Assessor Date This core checklist is provided by the Health and Safety Department as a template, which provides
More informationUSER MANUAL LTH-255A Hand Piece
USER MANUAL LTH-255A Hand Piece 250 Corporate Blvd.,Suite B Newark, DE 19702, USA Tel 302-709-0408 Fax 302-709-0409 http://www.companiontherapylaser.com LBL000037C USER MANUAL LTH-255A Hand Piece DCN000049
More informationStroller. Sprint S T R O L L E R S P R I N T
Stroller S T R O L L E R PAT I E N T O P E R AT I N G I N S T R U C T I O N S Sprint S P R I N T Warning Information Read this page carefully before using your unit. The unit contains liquid oxygen which
More informationSTANDARD OPERATING PROCEDURE (SOP): URETHANE
TEMPLE UNIVERSITY A Commonwealth University Environmental Health & Radiation Safety (EHRS) STANDARD OPERATING PROCEDURE (SOP): URETHANE LABORATORY/USER INFORMATION Principal Investigator (PI): Phone #
More informationAquaSPArkle Hot Tub and Spa Guide
AquaSPArkle Hot Tub and Spa Guide Water Treatment Chemicals ESSENTIAL INFORMATION ON... Starting up Circulation and filtration Water testing Chemical water treatment Draining and refilling Working out
More informationUser Instructions Series Descent Device
User Instructions - 199 Series Descent Device This document serves as the Manufacturer s Instructions, and is to be used as part of an employee training program for the system, as required by OSHA. ATTENTION:
More informationAUTOVENT 4000 VENTILATOR
OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on
More information6 CCR : Colo. Code Regs :6-101Alternatively cited as 6 CO ADC : Employee Health.
6 CCR 1010-7:6-101 6 Colo. Code Regs. 1010-7:6-101Alternatively cited as 6 CO ADC 1010-7 1010-7:6-101. Employee Health. Employees shall be in good health and be free from communicable disease while caring
More informationExeter Clinical and Health Research
Exeter Clinical and Health Research Standard Operating Procedure: Centrifuge Operation SOP Number: NIHRexe/019/GEN Version Number & Date: V2 13/06/2017 Review Date: 13/06/2018 Superseded Version Number
More informationOrthokeratology. What is Orthokeratology? How does it work?
What is? The goal of, OrthoK or corneal refractive therapy is to use a rigid gas permeable contact lens to reshape the cornea and provides correction for refractive errors. is a non-surgical procedure
More informationPERFORM Operating Document
PERFORM Operating Document Use and Maintenance of CO 2 Incubator PC-POD-CA-007-v03 Revision History Version Reason for Revision Date 01 New POD 30-Sep-13 02 Minor revisions for section 2.3, 3.1, 4.3. 14-April-16
More informationMy CoughAssist. A patient guide to CoughAssist T70. Please visit
Philips Healthcare is part of Royal Philips Electronics Europe, Middle East, Africa +49 7031 463 2254 How to reach us www.philips.com/healthcare healthcare@philips.com Latin America +55 11 2125 0744 Asia
More informationUSER MANUAL. AirPal Air Supply. (AS-1100 North America) 1488 Limeport Pike, Coopersburg, PA airpal.com fax
USER MANUAL AirPal Air Supply (AS-1100 North America) 1488 Limeport Pike, Coopersburg, PA 18036 airpal.com 800.633.4725 fax 610.965.2460 Welcome! Thank you for purchasing an AirPal Air Supply. This User
More informationPERFORM Operating Document
PERFORM Operating Document Use and Maintenance of CO 2 Incubator PC-POD-CA-007-v02 Revision History Version Reason for Revision Date 01 New POD 30-Sep-13 02 Minor revisions for section 2.3, 3.1, 4.3. 14-April-16
More informationCompact Ball Valves ASAHI AV VALVES. Contents. User s Manual. (Page) (1) Be sure to read the following warranty clauses of our product 1
Serial No. H-V030-E-5 Compact Ball Valves User s Manual Contents (Page) (1) Be sure to read the following warranty clauses of our product 1 (2) General operating instructions 2 (3) General instructions
More informationAnatomy Lab Safety Plan
Anatomy Lab Safety Plan Fall 2007 I. Anatomy Program The Anatomy program is run by the Biology Department at SFSU to provide an in depth understanding of the human body through study and practical experience.
More informationTools to Enhance Laboratory Biosafety and Biosecurity Performing Risk Assessments
Tools to Enhance Laboratory Biosafety and Biosecurity Performing Risk Assessments Selection of Safety Practices Biosecurity in the Laboratory Safety Orientation and Training Biosafety Competencies Audits,
More informationManagement Plan for Electrical Safety ISD #535
Management Plan for Electrical Safety ISD #535 Health & Safety Office Maintenance Service Building 10 SE 9 ½ Street Rochester, MN 55904 507-328-4507 MANAGEMENT PLAN FOR ELECTRICAL SAFETY Table of Contents
More information11/7/2011. Chemical Hazards and Controls for Rehabilitation Professionals. What is a chemical hazard? Toxicology. Module 3
Chemical Hazards and Controls for Rehabilitation Professionals Module 3 What is a chemical hazard? A chemical hazard is a chemical that, because of its characteristics and effects, may cause harm to an
More informationHercules 3 Universal Stabilizer Arm. Catalog Number
Instructions for Use Reusable Device Instructions for Use Hercules 3 Universal Stabilizer Arm Catalog Number 401-161 Carefully read all instructions prior to use. Observe all contraindications, warnings
More informationChemical Spill Clean-up Protocol
Protocol Department of www.rms.ubc.ca UBC-RMS-OHS-PRO 17-002 Date created: November 3, 2015 Date revised: April 28 th, 2017 Chemical Spill Clean-up Protocol Scope Laboratory-specific safe work procedures
More informationUsER manual for Watersens ph -REDOX
UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5
More information