DGT-CO2BX, Digital CO 2 Controller. Manual

Size: px
Start display at page:

Download "DGT-CO2BX, Digital CO 2 Controller. Manual"

Transcription

1 DGT-CO2BX, Digital CO 2 Controller Manual Ver.02/2008

2 1 Preface DGT-CO2BX is a Digital Control Unit that generates a CO 2 -Air mixture with an adjustable CO 2 concentration in the range 0-20%, with an accuracy of ± 5% of CO 2 concentration, for instance, if CO 2 set point is 5%, accuracy is ± 0.25%. DGT-CO2BX can be used as a stand-alone device as well as in combination with any kind of microscope incubator. The air flow is regulated by a floating ball air flow meter in the range Nliters/min. A CO 2 infra red sensor continuously measures CO 2 concentration in the mixed gas stream and a PID closed loop controller gives feedback to a fine valve regulating CO 2 flow. Measured value of CO 2 concentration is displayed in real time in green (DGT-CO2BX- PLUS displays also set point value). An alarm can be set on CO 2 concentration, so that, if CO 2 concentration goes below such value, display colour turns red and optionally a buzzer turns on. In the model DGT-CO2BX-PLUS the serial RS-232 interface and the CO 2 Control Software allow the user to control the unit with a personal computer and to acquire and store concentration data in computer memory. The DGT-CO2BX, CO 2 Digital Controller, together with the Okolab Microscope Stage Incubation Systems is designed to maintain all the required environmental conditions for cell cultures right on the microscope stage, thus allowing prolonged observations of cell events. 2

3 2 Index 1 PREFACE INDEX SAFETY NOTES EQUIPMENT E Q UI P M E N T SU P P L I E D Hardware details Tubing details Cables details Accessories E Q UI P M E N T NO T SUP P L I E D D IGIT A L CO2 C O N T R O L L E R SE T T I N G Front and Rear of the DGT-CO2BX- PLUS Overview CONNECTIONS S ET TH E GA S LIN E S T UB E S C O N N E C T I O N Connect DGT-CO2BX with Humidifying Module or Specimen Chamber Connect Gas reservoirs with DGT-CO2BX Unit Fix the Filter holder to the rear panel of the DGT-CO2BX Unit PC CABLES CONNECTIONS CHECK YOUR INTERNATIONAL OPTIONS INSTALL THE SOFTWARE CHANGE THE SET POINT STARTING THE EXPERIMENT AND GETTING THE BEST OUT OF YOUR INCUBATION SYSTEM O RG A N I Z E T H E EX P E R I M E N T A IR F LO W M ET E R RE G U L A T I O N APPENDIX D G T- C O 2 B X PA R A M E T E R S Factory parameters SUPPORT W EB C O N F E R E N C E F O R AS S I S T A N C E AN D T R AI N I N G T R O U B L E S H O O T I N G T EC H N I C A L SU P P O R T M AIN T E N A N C E External surfaces maintenance T EC H N I C A L S PE C I F I C A T I O N S

4 3 Safety Notes Before operating the equipment please read carefully the instructions and the safety notes. Read carefully also the supplied on request Omega Process Controller manual. If you have any questions, please contact Okolab assistance. Equipment should only be operated by technically qualified personnel Equipment and its internal parts can be damaged by dropping and by shock Before starting, assembly the equipment far from your microscope and/or electric device Do not start up the equipment if the supply cable is damaged After the use, disassembly the equipment far from your microscope and/or electric device Connect the equipment only to grounded mains power socket Prevent throttling and kinking of tubing Check gases tubes time to time for possible material usage Check the tubes are well fixed so they cannot slip off Use distilled treated water for gas stream humidification. For further informations please contact us before to use it. The equipment must only be used as intended and as described in this Manual. This includes operation by suitably instructed qualified personnel This device is not designed for use under medical conditions Do not use the flow meters as valves. Once regulated the flow rate (at P=1bar) use the flow meters only for fine regulations. Install safety valves on gas lines. We reserve the right to make technical variations. Change the Filters periodically as following described. International caution symbol marks this device. It is important to read the Safety Notes before installing, using and commissioning this device, as the notes contain important information relating to safety and EMC. IN NO EVENT SHALL OKOLAB S.R.L. BE LIABLE FOR ANY DIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY NATURE, OR LOSSES OR EXPENSES RESULTING FROM ANY DEFECTIVE PRODUCT OR THE USE OF ANY PRODUCT. CAUTION- Risk of electric shock! Do not open the Box. To prevent the risk of electric shock, do not remove cover or back. No user serviceable parts inside. 4

5 4 Equipment 4.2 Equipment supplied The DGT-CO2BX (-PLUS), CO 2 Digital Controller, is comprised of the following components: Hardware: #1 device, DGT-CO2BX (-PLUS-S) Software (Optional): Plastic connection tubings kit + accessories Electric Cables Hardware details #1 CD-ROM with DGT-CO2BX-PLUS-S Software 1) DGT-CO2BX Version, Digital CO 2 Controller. CO 2 can be regulated in the range 0-20% with great accuracy. Measured CO 2 value is displayed in real time. 2) DGT-CO2BX-PLUS Version, Digital CO 2 Controller Plus. Same as DGT-CO2BX + RS232 serial port to control and read CO 2 levels via software. Measured CO 2 and set point values are displayed in real time. 3) External Air Pump (Upon request) Tubing details 4) #2 (or more depending on model) white rigid tubes of 6mm OD, 3 m long. 5) #1 tube composed by 1 meter of blue rigid tube of 6mm OD at each end, 50cm of silicon tube of 4mm I.D. with a filter, degree of filtration 0.2 micron. 6) Accessory to connect together the blue rigid tube and the silicon tube of item 5) Cables details 7) #1 power supply cable. 8) #1 RS232 communication cable (Optional) included in the DGT-CO2BX-PLUS system. 9) #1 RS232/USB converter cables and its installation CD (Optional) included in the DGT- CO2BX-PLUS system Accessories 10)CD with control software (-PLUS-S). 11)Fixing-tubes kit (to fix tubes on the microscope table). 12)Pressure Gauge for CO 2 with regulator, degree of filtration 20 μm and condensate drain (supplied). 13)Pressure Gauge for pressurised Air line with regulator, degree of filtration 20 μm and condensate drain (not supplied, upon request). 14)Gas filter, degree of filtration 0.2 micron with #2 silicon tubing 10cm long (supplied). 15)Gas filter holder. 5

6 4.3 Equipment NOT supplied Personal Computer (OS: Windows 2000 /Xp) with #2 RS232 Serial Communication Ports or #1 USB Port. CO 2 reservoir. Air reservoir (It can be substituted by a compressed air line or External Air Pump) Pressure Gauge for Air with regulator, degree of filtration 20μm and condensate drain (provided by Okolab on request). 6

7 4.4 Digital CO 2 Controller setting Depending on you chose of the Okolab Digital CO 2 Controller version, the Front and Rear will be different. Tip DGT-CO2BX and DGT-CO2BX-PLUS versions only differ on the Display Screen and on the presence of the RS232 serial communication port on the Rear. IMPORTANT This manual is referred to the version DGT-CO2BX-PLUS Front and Rear of the DGT-CO2BX-PLUS Overview 1. Display (measured and set point value) 2. CO 2 fine valve controller 3. Floating ball Air flow meter µm filter 5. RS232 or infra red sensor calibration 6. RS232 to PC 7. Air inlet 8. CO 2 inlet 9. Mixed gas outlet 10. Power switch with #2 120mA-T Fuses Figure 1: Front and Rear of the DGT-CO2BX-PLUS Overview. 7

8 5 Connections 5.2 Set the gas lines Provide your system of CO 2 reservoir with safety valve and pressure gauge with pressure rigid tubes 6 mm OD exiting (Figure 3.number 1). Provide your system of Air reservoir (Okolab suggests to use a compressed Air line) with safety valve and pressure fine gauge with pressure rigid tubes 6 mm OD exiting (Figure 3.number 4). If a compressed air line is not present in the lab an Air Pump with filter inside can be used. It is not included in the system, upon request (Figure 3.number 5). Provide your system of O 2 instead of Air (optional configuration) reservoir, or compressed line, with safety valve and pressure gauge with pressure rigid tubes 6 mm OD exiting (Figure 3.number 4). Mount the Pressure Gauge for CO 2 with regulator, degree of filtration 20μm and condensate drain (supplied) on the wall vertically-(figure 3.number 2). Be sure that Air stream is FILTRATED and with a STABLE Pressure Gauge otherwise mount another Pressure Gauge for air with regulator, degree of filtration 20μm and condensate drain (non supplied, if you need ask to Okolab for a prompt shipping)- (Figure 3.number 6). If there is no pressurised air line in the lab an Air Pump with filter inside can be used. Follow the arrow on the Pressure Gauge for the correct gas In-Out. (see Figure 2) Figure 2: Pressure Gauge: Correct Gas Input Output. 8

9 1. CO2 reservoir with Pressure Fine Gauge 2. Pressure Gauge for CO 2 with regulator, 20 micron filter inside 3. White rigid tube with 6mm O.D., 3m long 4. Compressed air line with its Pressure Gauge 5. Air pump with filter inside, upon request instead of pressurised air line 6. Pressure Gauge for Air with regulator, 20 micron filter inside 7. White rigid tube with 6mm O.D., 3m long 8. Blue rigid tube with 6mm O.D., 3m long Figure 3: Gas lines scheme from reservoirs to Control Unit 5.3 Tubes connection Connect DGT-CO2BX with Humidifying Module or Specimen Chamber Connect the Blue marked tubing #1 tube composed by 1 meter of blue rigid tube of 6mm OD at each end, 50cm of silicon tube of 4mm I.D. with a filter, degree of filtration 0.2 micron. listed before to the push-in connector labelled as Air + CO2 out on the rear panel of the DGT-CO2BX and to the preheating module or to the humidifying module or directly to the chamber, depending on the incubating system model. Figure 4: Zoom of the blue tube and silicon tube connection. Tip Look at the instructions written on the labels attached on the transparent plastic bags. Tip In order to avoid gas leak and correctly insert the tube, please, strongly push the tube into the Swift-Fit connector. 9

10 Tip Connect the 4mm I.D. tube to the humidifying Module or directly to the Incubating Chamber if the preheating module is not present. Tip Look at the colour of the tubes and of the adhesive stripes attached on these devices. Tip This tubing moves the Air and CO 2 gas mixture to the Specimen Chamber Connect Gas reservoirs with DGT-CO2BX Unit White rigid tubes (4): Connect the controlled (filtered and at P=1atm) flows of air and CO 2 (O 2 optional configuration) from Pressure Gauges provided by to the DGT-CO2BX. Insert the tubes into the Swift-Fit connectors (Air IN and CO 2 IN on the rear of the Unit, see Figure 3. Tip If you want to disconnect the tube, please, pull the tube while pushing the black ring on the Swift-Fit connector (Figure 5). Tip In order to avoid gas leak and correctly insert the tube, please, strongly push the tube into the Swift-Fit connector Fix the Filter holder to the rear panel of the DGT-CO2BX Unit Tip The DGT-CO2BX Unit is equipped with two additional Hydrophobic PTFE filters with degree of filtration 0.2 microns for the protection against bacteria and condensations. One of them has to be mounted on the rear panel of the control box. It filters the Air and CO 2 flows coming from their tanks before entering into the carbon dioxide sensing cell. It prevents sample contamination from bacteria and infra red CO 2 sensor from damages caused by dirty gas streams. It is connected with two 4 I.D. mm silicon tubes with the Filter In and Filter Out connectors on the Rear of the DGT-CO2BX. It is fixed by a support, called "Filter Holder" to the Control Unit. This "Filter Holder" is fixed to the rear of the control box with two screws. Refer to Figure 6 for the instructions about how to mount this Filter holder to the rear panel of the control box. Another Hydrophobic PTFE filter is present along the Blue marked tubing #1 tube composed by 1 meter of blue rigid tube of 6mm OD at each end, 50cm of silicon tube of 4mm I.D. with a filter, degree of filtration 0.2 micron. This filter avoid contamination coming from the Humidifying module to the sensing cell, done by a small back pressure, that may occur during the CO 2 or Air pressurized lines closing. Tip Replace the gas filters each 2 months of continuous use (Figure 1, number 4; Figure 6). Okolab has these filters at stock. Contact Okolab to purchase them. Do not remove the filters, absolutely. This may cause damages to the sensing cell and sample contaminations. Avoid narrowing of the tubes and formation of air bubbles inside them. Please, periodically check fluids are correctly flowing and tubes are properly connected. 10

11 Figure 5: Avoid gas leak: strongly push the tube into the Swift-Fit connector Figure 6: Gas Filter 1. Filter holder. Metal sheet with two clamps. 2. #2 M3x8 screws. 3. #2 Silicon tubes 8-10 cm long 4. PTFE Hydrophobic Filter, degree of filtration 0.2 microns.!. Connect the Inlet of the the Filter to the Filter In connector on the rear panel of the control box. The inlet of the filter is labelled on the transparent plastic filter housing. 11

12 6 PC cables connections I. Connect the DGTCO2BX-PLUS to your PC using the RS232 Serial Cable supplied. II. Use the RS232/USB converter cable if your PC does not have RS232 serial communication port available. Please, install the driver on your PC using RS232/USB converter cable installation CD and re-start your PC. 7 Check your International options IMPORTANT For a proper work of the Okolab Read Temperature software, you MUST select the point as decimal separator in the International options of your PC, following this path: Start > Control Panel > Regional and Language Options > Regional Options > Customize Click Start. Select Control Panel. Select Regional and Language Options in the Control Panel. Click Customize in the Regional Options window. Select or write the point as decimal symbol in the text box. Select or write the space or comma as digit grouping symbol in the text box Click Apply than OK 2 times.. 12

13 Figure 7: Regional and Language Options Panel 8 Install the Software Insert the Installation CD-rom on your PC. Launch the setup.exe and follow the instructions on the screen. Once you have completed the installation, the program will appear on your PC Programs Menu. Re-start the Pc. 9 Change the Set Point The electronic controller has two different Set Points. Te Set Point 1 (SP1) represents the controlled CO 2 value and the Set Point 2 (SP2) represents the desired steady state value for Alarm functions. The Unit normally displays the current CO 2 concentration and the Set 13

14 Point 1. The SP1 is 5.00%, as default. SP2 has to be equal to SP1. If you wish to change manually the two Set Point values, please read the following instructions: Figure 8. Electronic controller display. IMPORTANT Do NOT change others parameters already adjusted to optimize the performances of the Unit. Press 1) Press, if necessary until SP1 prompt appears. Press 2) Display shows previous value of Setpoint 1. Press & 3) Press and to increase or decrease Setpoint 1 respectively. Holding & buttons down for approximately 3 seconds will speed up the rate at which the Setpoint value increments or decrements. Press & 4) Continue to use and to enter your 4-digit Setpoint 1 value. Press 5) Display shows STRD stored message momentarily and then advances to SP2 only, if a change was made, otherwise press a to advance to SP2 Setpoint 2 Menu. Do the same for SP2. The software supplied by Okolab allows to change the Set point during an experiment and to change the alarm configurations. It reads the Carbon Dioxide percentage in time and stores the data in the computer memory. Ask to Okolab for further informations. 14

15 10 Starting the experiment and Getting the best out of your incubation system 10.2 Organize the experiment Read carefully the manual of the Incubating System you are using to control the specimen temperature. If the temperature controller has been supplied by Okolab, ask to us how to get the best from the incubating system equipped with the DGT-CO2BX. IMPORTANT Overloading the Humidifying Module could induce condensation inside the tubes, on the windows surface of the incubating chamber and on the surfaces of plastic supports, not depended on the DGT-CO2BX. Avoid gases back flow. Check that all tubes are properly connected and prevent narrowing, throttling, kinking of tubing, and formation of air bubbles. Check the tubes are well fixed so they cannot slip off using the kit supplied. Switch on the Temperature Controller Air Flow Meter regulation IMPORTANT Do not use the flow meters as closing-opening valves. Once regulated the flow rate (at P=1bar, see Figure 3, use the flow meters carefully only for fine regulations. Open the Air gas valves slowly, regulate the pressure at 1 bar. Otherwise turn on the Air Pump with filter inside. Regulate the floating ball of the Air Flow Meter within the range Litres/min. Target Gas Percentage [%] Air flow rate values and CO2 consumption [Nl/min] Air CO 2 Air CO 2 Total Table 1. CO 2 consumption for different air flow rates. [Nl = Normal litre] Tip If you need, ask Okolab for personalized Gas Percentage Tables. 15

16 'Target Gas Percentage indicates the stream gas percentage that the operator wants during the experiment. Air flow rate values and CO2 consumption, in Normal litre per minute [NL/min], are the values of air flow rate set on the graduated scale and the corresponding CO 2 consumption. 'Total indicates the measured total gas stream flow rate. Regulate the Floating ball Air Flow Meter at the desired value within the range NL/min. Open the pressure gauge of the CO 2 reservoir and set it at bar. Regulate the Pressure Gauge for CO 2 with regulator, 20 micron filter inside, supplied by Okolab at bar. Be sure that no pressure drop happens. Use this pressure regulator as closing valve when you finish the experiment too. Turn on the DGT-CO2BX. After few second the CO 2 percentage is displayed on the Unit. After half an hour the CO 2 reach the steady state value (5.00% as default). You can start the experiment and carry on the specimen into the incubating chamber. For further informations and especially use of the Unit, please feel free to contact the Okolab people. System has been calibrated at the factory. It is maintenance free. Sensor is stable up to five years of work. We suggest to calibrate the Unit each two years of continuous use. Contact the Okolab Assistance to do the calibration. 11 Appendix 11.2 DGT-CO2BX parameters The PID closed loop controller reads the CO 2 level within the gas stream and gives feedback to a fine valve regulating CO 2 flow with an accuracy of ± 5% of CO 2 percentage. The PID parameters were stored to obtain the maximum performances from the system in terms of stability and fast recovery, after a gas flow changing due to quickly changed of pressures. The controller needs half an hour to reach the steady state after turning on the unit. Avoid modifying the controller factory parameters. Some parameters alterations could make the system bad-functioning or could induce the system failure. Please, contact Okolab assistance in case, some control parameters, in PID display, were accidentally garbled Factory parameters Setpoint1: 5.00%; Setpoint1: 5.00%; Anti integral: Disable; Proportional: 50.0; Reset: 150; 16

17 Rate: 3.0; Damping: 4; Thermocouple type: k; Color Display: Green; Temperature Unit: C Filter Value: Support 12.2 Web conference for assistance and training In case you requested on line support and/or web conference assistance for system assembling and experiment support, you received a web cam. Please follow these rules to receive a good service: Webcam installation (instruction and Cdrom included) Last MSN Messenger installed ( Register yourself on to have an account (MSN passport) Set the audio and video and test them using MSN Messenger. If you have problem with MSN Messenger you can choose Skype software ( Contact our technical support (sibillo@oko-lab.com) by to take an appointment for the web assistance Troubleshooting Incorrect operations are often mistaken for trouble and malfunctions. If you think that there is something wrong with a component, check the points below. Sometime the trouble may lie in another component. Investigate the others components, electrical appliance and pneumatic system being used. If the trouble cannot be rectified even after exercising the checks listed below, ask Okolab service. Symptom Probable cause Remedy No power supplied to Unit No CO 2 and/or Air flow rate stability Power plug is disconnected from outlet Blown fuse Pressure drop Air Flow too low Insert plug securely into outlet Replace the fuse, 120mA-Time Lag Fuses Check reservoir and/or compressed line Change the PTFE filter with degree of filtration 0.2 micron Check the air flow with the Floating ball flow meter. Increase air flow. 17

18 Symptom Probable cause Remedy No PC - Unit communication RS 232 cable not plugged in USB-RS232 converter cable not plugged in USB-RS232 converter cable driver not installed on this USB port Serial Port communication numbers not inserted Insert plug securely into outlet No CO 2 and/or Air flow streams Regulators closed Open regulators Fast sample drying No CO 2 read The Okolab Software does not work properly The Okolab Software does not read the correct gas percentage displayed on the Unit Error 1 at open file An input parameter is invalid. Gas leak on the roar of the control box Humidifying Module empty The Unit is off The PTFE filter with degree of filtration 0.2 micron is dirty The selected decimal separator is not the point Probable PC compatibility problem Saving in an inexistent Directory Directory Path not accessible Anti-Virus or Firewall make invalid an input parameter Not properly inserted tubes Use CD rom to install the driver on this USB port Click on the Settings button of Okolab Software and follow the instructions Fill it with distilled water until 50 mm below the top of glass column. Switch on the Control Unit Change the PTFE filter with degree of filtration 0.2 micron Select the point as decimal separator in the International options of your PC Contact Okolab to receive assistance Verify the Directory Path Change your account type. Ask to computer Administrator Disable Anti-Virus or Firewall activities on Okolab Software Strongly push the tube into the Swift- Fit connector I check the previous troubleshooting but the system does NOT reach a stable temperature Water Bath does not work properly The Temperature Meter does not work properly Contact Okolab to receive assistance Contact Okolab to receive assistance 12.4 Technical support. Please, do not hesitate to contact Okolab should you need any further commercial information or technical support. Please, check Okolab web site for news, events, new products and general FAQ. For COMMERCIAL SUPPORT: lanzaro@oko-lab.com Phone Fax: Mobile: For TECHNICAL SUPPORT: sibillo@oko-lab.com Phone Fax: Mobile: For SOFTWARE SUPPORT: caserta@oko-lab.com 18

19 Phone Fax: Mobile: Published by Okolab S.r.l.,Via G. di Prisco, 152, Ottaviano (NA), ITALY 12.5 Maintenance External surfaces maintenance Use a polishing cloth or dry cloth to wipe off dust and dirt. Never use thinners, benzene, solvents on or near the devices, since these could corrode their surfaces. Change the silicon tubes after one year of continuous use. 19

20 12.6 Technical Specifications Specification Table Product code Sensing technology CO 2 Set Point value range 0 20% Accuracy DGT-CO2BX (-PLUS) Repeatability of CO 2 concentration Below 0.1% Air Flow rate Filtering device Digital interface (RS232) Communication Software (-S) Dimensions Weight Voltage/Frequency Compatibility Non dispersive infra red (NDIR) dual wave length detector ± 5% of CO2 concentration, i.e. ± 0.25% if CO 2 is 5% From 0.2 Nl/min to 0.8 Nl/min Reusable PTFE membrane with 0.2 micron pores RS232 to PC Stand alone version and Plug-in version for all Okolab Software 206x178x310 mm 3.5 kg V / Hz Any microscope incubator Unit: Input Vac ± 10%; Hz, 6 W; 2 Fuse Time-Lag 125mA. Air Flow Meters: Flow Rate range Nl/min, 5%full scale error; Pmax 6.0 bar, Operating pressure 1.0 bar at 20 C. 20

21 Figures FIGURE 1: FRONT AND REAR OF THE DGT-CO2BX-PLUS OVERVIEW...7 FIGURE 2: PRESSURE GAUGE: CORRECT GAS INPUT OUTPUT...8 FIGURE 3: GAS LINES SCHEME FROM RESERVOIRS TO CONTROL UNIT...9 FIGURE 4: ZOOM OF THE BLUE TUBE AND SILICON TUBE CONNECTION...9 FIGURE 5: AVOID GAS LEAK: STRONGLY PUSH THE TUBE INTO THE SWIFT-FIT CONNECTOR...11 FIGURE 6: GAS FILTER...11 FIGURE 7: REGIONAL AND LANGUAGE OPTIONS PANEL...13 FIGURE 8. ELECTRONIC CONTROLLER DISPLAY...14 Tables TABLE 1. CO2 CONSUMPTION FOR DIFFERENT AIR FLOW RATES. [NL = NORMAL LITRE]...15

22 WARRANTY Okolab S.r.l. warrants its products to be free of defects in materials and workmanship for a period of one year starting from invoice date. If the unit malfunctions, it must be returned to the factory for evaluation. If the equipment has to be returned to the factory, please ensure that is carefully and properly packed. Okolab S.r.l. accepts no responsibility for damage due to unsatisfactory packing. Upon examination of Okolab S.r.l., if the unit is found to be defective, it will be repaired or replaced at no charge. This warranty does not apply to defects resulting from any actions of the purchaser. Components which wear are not warranted. Okolab S.r.l. neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by Okolab S.r.l. Okolab S.r.l. warrants only the parts manufactured by it will as specified and free of defects. Okolab S.r.l. makes no other warranties or representations of any kind whatsoever, express or implied, except that of title, and all implied warranties including any warranty of merchantability and fitness for a particular purpose are hereby disclaimed. LIMITATION OF LIABILITY: the total liability of Okolab S.r.l. shall not exceed the purchase price of the component upon which liability is based. In NO event shall Okolab S.r.l. be liable for consequential, incidental or special damages.

In Vivo Scientific, LLC INSTRUCTION MANUAL

In Vivo Scientific, LLC INSTRUCTION MANUAL CO 2 Controller In Vivo Scientific, LLC INSTRUCTION MANUAL CONTENTS CONTENTS...1 ABOUT THIS MANUAL...2 INTRODUCTION...2 Cautions and Warnings...2 Parts List...2 Unpacking...2 INSTRUMENT DESCRIPTION...3

More information

Bold Line Top Stage Incubator

Bold Line Top Stage Incubator H301 T UNIT BL H301 T UNIT BL Bold Line Top Stage Incubator Temperature, Humidity, CO 2 /O 2 Control, Touch Screen and Web server The systems comprises: Incubating chamber (model avaialble for any XY stage

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

H301-T-UNIT-BL-PLUS Electric Top Stage Incubation System Manual

H301-T-UNIT-BL-PLUS Electric Top Stage Incubation System Manual H301-T-UNIT-BL-PLUS Electric Top Stage Incubation System Manual Vers. 02.14 1 This page was left blank 2 Index Index 1 REVISION TABLE... 4 2 PREFACE... 5 3 SYMBOLS DESCRIPTIONS... 5 3.1.1 Symbols in this

More information

Bold Line Cage Incubator

Bold Line Cage Incubator Bold Line Cage Incubator com Bold Line Cage Incubator Bold Line Cage Incubator Large volume incubator with local CO 2 /O 2 climate chamber The systems comprises: Microscope enclosure Obscuring panels (optional)

More information

UNO-COMBINED-CONTROLLER. CO 2 Microscope Electric Top Stage Incubation System. Manual. Vers

UNO-COMBINED-CONTROLLER. CO 2 Microscope Electric Top Stage Incubation System. Manual. Vers UNO-COMBINED-CONTROLLER CO 2 Microscope Electric Top Stage Incubation System Manual Vers. 02.14 1 This page was left blank 2 Index Index 1 REVISION TABLE... 5 2 PREFACE... 6 3 SYMBOLS DESCRIPTIONS... 7

More information

Overview. Front Panel: Keypad and Display

Overview. Front Panel: Keypad and Display Overview The GA-200B is an analyzer that integrates a gas sampling system with sensors to measure and display the concentrations of oxygen and carbon dioxide in a sample as the percentage of a gas in the

More information

RAM 4021 Operation Manual

RAM 4021 Operation Manual RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

Digital Vacuum Regulator

Digital Vacuum Regulator Temperature Control for Research and Industry Digital Vacuum Regulator User s Manual Model 300 INDEX SECTION PAGE 1. QUICK OPERATING INSTRUCTIONS........................... 3 Safety Notices.................................................

More information

SPECIFICATIONS APCEPH1

SPECIFICATIONS APCEPH1 APCEPH1 ph CONTROLLER SPECIFICATIONS APCEPH1 Input voltage 120 Volts AC Maximum amperage 14.5 amps @ 120 VAC ph Accuracy +/- 0.2 ph ph Control range Adjustable 4.5 8.5 ph Weight < 1 lbs Dimensions 3" x

More information

RK-IR Sample Draw Aspirator Adapter Operator s Manual

RK-IR Sample Draw Aspirator Adapter Operator s Manual 30-0951RK-IR Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0018RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021-PR Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-PR For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4

More information

CSA Sample Draw Aspirator Adapter Operator s Manual

CSA Sample Draw Aspirator Adapter Operator s Manual 30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

UNO STAGE TOP INCUBATOR All In One Stage Top Incubator. Manual

UNO STAGE TOP INCUBATOR All In One Stage Top Incubator. Manual UNO STAGE TOP INCUBATOR All In One Stage Top Incubator Manual Vers. 3-2017 1 This page was left blank 2 Index Index PREFACE... 5 2 SYMBOLS DESCRIPTION... 6 SYMBOLS IN THIS MANUAL AND ON THE PRODUCT...

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

RAM 4021-DPX Operation Manual

RAM 4021-DPX Operation Manual RAM 4021-DPX Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS ABL 4021-DPX / RAM 4021-DPX For Your Safety... 3 Description... 3 Setup Mode... 4 Lights/Alarms... 4 Operation...

More information

RK LEL Sample Draw Aspirator Adapter Operator s Manual

RK LEL Sample Draw Aspirator Adapter Operator s Manual 30-0951RK LEL Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0017RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

Single UNO. Incubator. .com. Main Features: Incubating chamber. of a Multi Well plate. Manual Air/CO 2 mixer. operation.

Single UNO. Incubator. .com. Main Features: Incubating chamber. of a Multi Well plate. Manual Air/CO 2 mixer. operation. UNO Incubator All-in-one temperature/humidity/co2 incubator Top Stage Incubator with the footprint of a Multi Well plate Fits in any XY stage, with the addition of the Multi Well plate holder (supplied

More information

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr A DIVISION OF THE FREDERICKS COMPANY 2400 PHILMONT AVE. HUNTINGDONVALLEY, PA 19006 PARTS LIST 1 3 4 2 # QTY ITEM DESCRIPTION

More information

TR Electronic Pressure Regulator. User s Manual

TR Electronic Pressure Regulator. User s Manual TR Electronic Pressure Regulator Page 2 of 13 Table of Contents Warnings, Cautions & Notices... 3 Factory Default Setting... 4 Quick Start Procedure... 5 Configuration Tab... 8 Setup Tab... 9 Internal

More information

RAM Operation Manual

RAM Operation Manual RAM 4021-1 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-1 For Your Safety... 2 Description... 2 Setup Mode... 3 Lights/Alarms... 3 Operation... 4 Calibration...

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE11C150, VE15C150, VE18C150 PART NO: 2226005, 2226000, 2226015 OPERATION & MAINTENANCE INSTRUCTIONS LS0615 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

Digital Vacuum Regulator

Digital Vacuum Regulator Temperature Control for Research and Industry Digital Vacuum Regulator User s Manual Model DVR-380 INDEX SECTION PAGE SAFETY NOTICES................................................. 3 1. QUICK OPERATING

More information

92831 TEL: (714) FAX:

92831 TEL: (714) FAX: Document No. 1800-75 Respiration Test Chamber Copyright 2010 Terra Universal Inc. All rights reserved. Revised September 2010 Terra Universal, Inc. TerraUniversal.com 800 S. Raymond Ave. Fullerton, CA

More information

INSTALLATION & MAINTENANCE INSTRUCTIONS

INSTALLATION & MAINTENANCE INSTRUCTIONS DESCRIPTION / IDENTIFICATION The QBX series valve uses Proportion-Air closed loop technology for Pressure control. It gives an output pressure proportional to an electrical command signal input. The QB1X

More information

User s Guide. CO2 & O2 Controller for miniature incubators

User s Guide. CO2 & O2 Controller for miniature incubators CO2 O2 Control User s Guide CO2 & O2 Controller for miniature incubators Precise CO2 & O2 Control throughout the experiment Media ph control Compatible with any perfusion system Miniature incubators for

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE15C150, VE18C150, VE25C150 PART NO: 2226010, 2226020, 2226025 OPERATION & MAINTENANCE INSTRUCTIONS LS0715 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information

RM-80 respiration monitor

RM-80 respiration monitor RM-80 respiration monitor User Manual September 18, 2015 0025-003M 950 North Hague Avenue Columbus, Ohio 43204-2121 USA Sales: sales@colinst.com Service: service@colinst.com Phone: (614) 276-0861 Fax:

More information

Model PDT Dewpoint Transmitter

Model PDT Dewpoint Transmitter Model PDT Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk

More information

User Manual MFCS TM -EZ

User Manual MFCS TM -EZ User Manual MFCS TM -EZ Version 6A_December 2016 1. INTRODUCTION 5 2. GENERAL INFORMATION 6 2.1 General functioning 6 2.2 Different MFCS TM 6 3. PACKAGE CONTENT 8 4. MFCS TM -EZ DESCRIPTION 9 5. CONNECTIONS

More information

92831 TEL: (714) FAX:

92831 TEL: (714) FAX: Document N0. 1800-03 Copyright 2010 Terra Universal Inc. All rights reserved. Revised Sept. 2010 Terra Universal, Inc. TerraUniversal.com 800 S. Raymond Ave. Fullerton, CA 92831 TEL: (714) 578-6000 FAX:

More information

BGA244 Binary Gas Analyzer

BGA244 Binary Gas Analyzer Quick Start Guide Revision 1.0 Certification Warranty Service certifies that this product met its published specification at the time of shipment. This product is warranted against defects in materials

More information

SDM-2012 Docking Station Standalone Configuration Operator s Manual

SDM-2012 Docking Station Standalone Configuration Operator s Manual SDM-2012 Docking Station Standalone Configuration Operator s Manual Part Number: 71-0254RK Revision: P5 Released: 10/5/12 www.rkiinstruments.com Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

User s Guide. CO2 Controller for miniature incubators

User s Guide. CO2 Controller for miniature incubators CO2 Control User s Guide CO2 Controller for miniature incubators Precise CO2 Control throughout the experiment Media ph control Compatible with any perfusion system Miniature incubators for any microscope

More information

4 and 8 Station PDS Control Installation & Operators Manual

4 and 8 Station PDS Control Installation & Operators Manual 4 and 8 Station PDS Control Installation & Operators Manual 1402-15 11/2000 June 2003 MW1402C Chore-Time Warranty 4 and 8 Station PDS Control Chore-Time Warranty Chore-Time Equipment ( Chore-Time ) warrants

More information

UsER manual for Watersens ph -REDOX

UsER manual for Watersens ph -REDOX UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5

More information

User s Guide. Pressure Controller for liquid delivery systems

User s Guide. Pressure Controller for liquid delivery systems Flow Control User s Guide Pressure Controller for liquid delivery systems Precise Pressure Control throughout the experiment Flow control Compatible with any perfusion system Ideal for Small Volume Delivery

More information

MODEL CALIBRATION GAS DELIVERY SYSTEM

MODEL CALIBRATION GAS DELIVERY SYSTEM MODEL 1200-26 CALIBRATION GAS DELIVERY SYSTEM Sierra Monitor Corporation 1991 Tarob Court, Milpitas, CA 95035 (408) 262-6611 MODEL 1200-26 CALIBRATION GAS DELIVERY SYSTEM APPLICABILITY & EFFECTIVITY This

More information

PERSONAL AIR BREATHING UNIT

PERSONAL AIR BREATHING UNIT PERSONAL AIR BREATHING UNIT Operation & Maintenance Manual MARTECH SERVICES C O M P A N Y 1-800-831-1525 Table of Contents INTRODUCTION: Page 2 COMPONENTS DRAWING: Page 3 START UP: Page 4 OPERATION: Page

More information

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE INSTALLATION, OPERATION & MAINTENANCE MANUAL INTERFACE DEVICES, INC. 230 Depot Road, Milford, CT 06460 Ph: (203) 878-4648, Fx: (203) 882-0885, E-mail: info@interfacedevices.com

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0578 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

with O 2 Controller Instruction Manual

with O 2 Controller Instruction Manual 14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Hypoxic Cabinets (In-Vivo / In-Vitro) with O 2 Controller Instruction Manual Index Page Warranty 2 Warnings

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300 Technical Note GA-300 Overview The GA-300 CO2 and O2 Gas Analyzer is easy to use, robust, and adaptable to human, animal, and plant applications. The GA-300 has two analog outputs to allow recording and

More information

AWG Fittings LLC. Pressure Relief Valve Up to 250 PSI. Product Number Read this instruction manual before use.

AWG Fittings LLC. Pressure Relief Valve Up to 250 PSI. Product Number Read this instruction manual before use. AWG Fittings LLC Pressure Relief Valve Up to 250 PSI Product Number 30004033 Read this instruction manual before use. Using this device without t understanding di this products operation and care may lead

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0545 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL - Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL - Table of Contents 1. Introduction 3 2. Instrument Setup 4 2.1 Instrument Checklist 4 2.2 Optional Accessories 4 2.3 Gas Concentration Ranges Available

More information

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS 2000 Dionex Corporation Document No. 031277 Revision 03 April 2000 2000 Dionex Corporation All rights reserved worldwide. Printed in the United States of

More information

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves

More information

Series Environmental Chambers

Series Environmental Chambers 3119-600 Series Environmental Chambers Challenges in Non-Ambient Testing Testing at non-ambient temperatures adds another layer of challenges to your testing laboratory. Ensuring you get accurate and stable

More information

Digital Vacuum Regulator

Digital Vacuum Regulator Temperature Control for Research and Industry Digital Vacuum Regulator User s Manual Model DVR-280 INDEX SECTION PAGE 1. QUICK OPERATING INSTRUCTIONS............................. 3 KEM-NET DATA LOGGING

More information

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 QUICK START GUIDE P/N 6006154, REVISION C MAY 2013 Model 8120 Mask Integrity Tester is patented under U.S. Patent No. 8,312,761. Additional patents are pending.

More information

MODEL 100 NITROGEN INFLATION CART

MODEL 100 NITROGEN INFLATION CART MODEL 100 NITROGEN INFLATION CART Installation & Operation Information Branick Industries, Inc. 4245 Main Avenue P.O. Box 1937 Fargo, North Dakota 58103 REV120106 P/N: 81-0113 TABLE OF CONTENTS SAFETY

More information

MCQ Gas Blender 100 Series 3 Channels Gas Mixer

MCQ Gas Blender 100 Series 3 Channels Gas Mixer Channels Gas Mixer Revision 1.8 October 2016 High Performance Digital Gas flow Gas Dilutor Blender & Gas for Mixing flow Mixing Gases System and Calibration Routines Technical Data Sheet Lab in a box From

More information

OPERATION MANUAL NTF-15

OPERATION MANUAL NTF-15 OPERATION MANUAL NTF-15 Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81235-00 (Rev B) TABLE

More information

INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May

INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May 2015 98-0520 S/N 2001 N. Indianwood Ave. Tulsa, Oklahoma 74012 U.S.A. TEL: (918) 250-7200 FAX: (918) 459-0165 E-mail: chandler.sales@ametek.com

More information

D10S/D20S Wall/Post Mount Inflator Quick Start Manual

D10S/D20S Wall/Post Mount Inflator Quick Start Manual PART NUMBER SERIAL NUMBER D10S/D20S Wall/Post Mount Inflator Quick Start Manual Please read and save these instructions. Read carefully before attempting to assemble, install, operate or maintain the product

More information

24L OIL FREE AIR COMPRESSOR MODEL NO: TIGER 7/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS10/13

24L OIL FREE AIR COMPRESSOR MODEL NO: TIGER 7/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS10/13 24L OIL FREE AIR COMPRESSOR MODEL NO: TIGER 7/250 PART NO: 2244030 OPERATION & MAINTENANCE INSTRUCTIONS LS10/13 INTRODUCTION Thank you for purchasing this product. Before attempting to use this product,

More information

Using the Akta Prime plus October 22, 2012

Using the Akta Prime plus October 22, 2012 Some starting precautions: 1. Vacuum filter all buffers. Removes any large particles/debris that may clog your column De-gases the buffers 2. Clarify lysates first by centrifugation and then filtration

More information

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight Technical Specifications Operating Instructions Maintenance Information Troubleshooting Guide Parts Diagrams AC1810 / AC1810-A THE EVOLUTION OF PERFECTION CAUTION: Before attempting to use or service this

More information

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

Portable Gas Monitor GX User Maintenance Manual (H4-0050) H4E-0050 Portable Gas Monitor GX-8000 User Maintenance Manual (H4-0050) Need of Maintenance and Servicing This gas monitor must be maintained in a normal state at all times to prevent accidents due to

More information

BUBBLER CONTROL SYSTEM

BUBBLER CONTROL SYSTEM BUBBLER CONTROL SYSTEM Description: The HDBCS is a fully automatic bubbler system, which does liquid level measurements in water and wastewater applications. It is a dual air compressor system with, air

More information

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual MODEL 50012 Maverick Heavy Duty Aerosol Compressor Operator s Manual This page left blank intentionally CONTENTS Page No. INSPECTION... 1 GENERAL INFORMATION... 1 SAFETY PRECAUTIONS... 2 BASIC OPERATING

More information

CCT-7320/ROC-2313 Reverse Osmosis Controller

CCT-7320/ROC-2313 Reverse Osmosis Controller CCT-7320/ROC-2313 Reverse Osmosis Controller 1 General The instrument is a combined control instrument of a reverse osmosis controller and an on-line conductivity instrument. It can perform the operation

More information

Portable oxygen concentrator

Portable oxygen concentrator Table of contents G3 1. Introduction 2 2. Important Safety Rules 3 3. Unpack your Lovego 4 4. User Controls & System Status Indicators 5 5. Quick Start 6 6. How to operate 7 7. How to operate with battery

More information

Manual Weighingblock VB2 series and Uniscale

Manual Weighingblock VB2 series and Uniscale Manual Weighingblock VB2 series and Uniscale Note: At page 8 in this manual you will find a short form instruction. Normally the only instruction shipped together with the Scale. Overview different ranges.

More information

Duplex Commercial Softeners

Duplex Commercial Softeners Duplex Commercial Softeners Installation, Operation & Maintenance Guide Manual 003.1 Contents Page 1. Unpacking Instructions 3 2. Installation 4 Pre-installation checks Fitting the bottom distribution

More information

EPP4 Pressure Regulator Basic G 1/4"

EPP4 Pressure Regulator Basic G 1/4 EPP4 Pressure Regulator Basic G 1/4" Flow Curve 1/4" Dimensions Outlet pressure - bar 9 8 7 6 5 4 3 2 1 0 0 20 40 60 70 80 100 120 Flow rate - Nm 3 /h The male connector adopted on the EPP4 is a standard

More information

User s Guide Temperature Sensor Converter TSC-599

User s Guide Temperature Sensor Converter TSC-599 User s Guide Temperature Sensor Converter TSC-599 ILX Lightwave Corporation 31950 Frontage Road Bozeman, MT, U.S.A. 59715 U.S. & Canada: 1-800-459-9459 International Inquiries: 406-556-2481 Fax 406-586-9405

More information

OC Panel High Limit Aquastat Kit, Manual Reset p/n

OC Panel High Limit Aquastat Kit, Manual Reset p/n OC Panel High Limit Aquastat Kit, Manual Reset p/n 233202 Instruction Sheet APPLICATION The OC (Option Control) Panel High Limit Aquastat Kit provides electronic temperature sensing in a UL limit-rated

More information

Dextens Thermo Conductivity Sensors. Type 52101, 54101, 55101

Dextens Thermo Conductivity Sensors. Type 52101, 54101, 55101 Dextens Thermo Conductivity Sensors Type 52101, 54101, 55101 Operating Manual May 2012 Table of Content 1 GENERAL INFORMATION...4 1.1 DISCLAIMER...4 1.2 SYMBOLS AND CONVENTIONS...4 1.3 SERVICES AND REPAIRS...4

More information

ORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual

ORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual ORB-400 BUBBLE/HAZE MACHINE Item ref: 160.462UK User Manual Caution: Please read this manual carefully before operating Damage caused by misuse is not covered by the warranty Introduction Thank you for

More information

Budget Range Operators Handbook

Budget Range Operators Handbook Budget Range Operators Handbook BAMBI AIR COMPRESSORS LTD 152 Thimble Mill Lane Heartlands Birmingham B7 5HT United Kingdom Tel: 0121 322 2299 Fax: 0121 322 2297 Email: sales@bambi-air.co.uk www.bambi-air.co.uk

More information

User Instruction Manual

User Instruction Manual User Instruction Manual 4500 psi Air Compressor Ver 2, 1.18 Contents Parts Included...3 Assembly Instructions...3-5 Operation Instructions...6-7 Oil Change Intervals...8 Air Filter Replacement...9 Setting

More information

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

AIR/OXYGEN BLENDER INSTRUCTION MANUAL BIO-MED DEVICES AIR/OXYGEN BLENDER INSTRUCTION MANUAL CATALOG #2120 REV 042203 BIO-MED DEVICES INC. 1445 BOSTON POST ROAD, GUILFORD, CT 06437 (203) 458-0202 FAX (203) 458-0440 www.biomeddevices.com TABLE

More information

24L AIR COMPRESSOR MODEL NO: TIGER 11/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13

24L AIR COMPRESSOR MODEL NO: TIGER 11/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13 24L AIR COMPRESSOR MODEL NO: TIGER 11/250 PART NO: 2244010 OPERATION & MAINTENANCE INSTRUCTIONS LS01/13 INTRODUCTION Thank you for purchasing this product. Before attempting to use this product, please

More information

Cocoa Patio Pond with Lit Spillway

Cocoa Patio Pond with Lit Spillway Cocoa Patio Pond with Lit Spillway REMINDER CALL 1-888-755-5641 BEFORE RETURNING TO STORE. PACKAGE CONTENTS ITEM # GQSPPB/GQSPPW Questions, problems, missing parts? Before returning to your retailer, call

More information

Research Grade Blood Pressure Transducer User's Manual

Research Grade Blood Pressure Transducer User's Manual Research Grade Blood Pressure Transducer User's Manual Research Grade Blood Pressure Transducer, 110 VAC/60 Hz MA1 60-3002 Research Grade Blood Pressure Transducer, 220 VAC/50 Hz MA1 60-3003 WEEE/RoHS

More information

Instruction Manual Dräger MSI P7 and MSI P7 plus

Instruction Manual Dräger MSI P7 and MSI P7 plus Dräger MSI GmbH Rohrstraße 32 58093 Hagen Tel.: +49-2331 / 9584-0 Fax: +49-2331 / 9584-29 e-mail: info@draeger-msi.de D 923; Edition 2011-01-01 Content 1. General Hints Page 4 2. The Instrument 2.1 Front

More information

FlashGARD HP Reverse Osmosis Filtration System Model Number FSTMO75 Part Number

FlashGARD HP Reverse Osmosis Filtration System Model Number FSTMO75 Part Number 3M Water Filtration Products FlashGARD HP Reverse Osmosis Filtration System Model Number FSTMO75 Part Number 56123-06 Installer: Please leave this manual with owner/operator. End User: Please retain for

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP LIMITED WARRANTY Getz Equipment Innovators warrants its products, and component parts of any product manufactured by

More information

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels 12.09 product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels Introduction: This manual covers the installation and operation of Humboldt FlexPanels for Triaxial and Permeability

More information

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties Quick Installation Guide Doc nr.: 9.17.052H Date: 10-12-2015 SCOPE OF THIS GUIDE mini CORI-FLOW instruments are highly accurate instruments for measuring and controlling the mass flow rate of liquids and/or

More information

24L AIR COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: RANGER 7/240 PART NO: LS0913

24L AIR COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: RANGER 7/240 PART NO: LS0913 24L AIR COMPRESSOR MODEL NO: RANGER 7/240 PART NO: 2242000 OPERATION & MAINTENANCE INSTRUCTIONS LS0913 INTRODUCTION Thank you for purchasing this CLARKE 24L Air Compressor. Please read this manual fully

More information

5 Channel Calibrator

5 Channel Calibrator 212 N. Woodwork Lane Palatine, IL 60067 800-223-3977 5 Channel Calibrator Part # 0724 User Manual Questions? Contact us at 800-223-3977 or online at http://www.cleanair.com/equipment/express/main.html

More information

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y P.O. Box 7079 OFFICE: (507)

More information

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc. The HumiSys RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The new HumiSys with single or dual RH probes capabilities is

More information

Interface Devices, Inc. Hydraulic Mini Mule

Interface Devices, Inc. Hydraulic Mini Mule Interface Devices, Inc. Hydraulic Mini Mule INSTALLATION, OPERATION & MAINTENANCE MANUAL IMPORTANT! FILE THIS MANUAL IN A SAFE PLACE FOR FUTURE SERVICE & PARTS NEEDS ALWAYS REFERENCE THE SERIAL NUMBER

More information

LRS(H)4 Pressure-Reducing Regulator User Manual

LRS(H)4 Pressure-Reducing Regulator User Manual LRS(H)4 Pressure-Reducing Regulator User Manual Read the complete manual before installing and using the regulator. 2 Safe Product Selection When selecting a product, the total system design must be considered

More information

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units Models: EPS Single Pump EPT Twin Pump EPS-HP EPT-HP Single Pump High Pressure Twin Pump High Pressure

More information

01 TM SuperModified Testing with RS485 <> USB

01 TM SuperModified Testing with RS485 <> USB Testing with RS485 USB 1. Preinstalled cables The Supermodified controller comes with several wires soldered. By default the power and RS485 cables are soldered. I2C and legacy RC servo interface cables

More information

CC Burst Test System

CC Burst Test System CC Burst Test System Series 1835 Contents Preface 4 Unpacking 4 Warranty 4 After-Sale Support 4 Section 1: Introduction 5 1.1 General Description 5 1.2 The Series 1835 Pressure Supply System 5 1.3 Specifications

More information

100L AIR COMPRESSOR MODEL NO: TIGER 16/1010 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13

100L AIR COMPRESSOR MODEL NO: TIGER 16/1010 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13 100L AIR COMPRESSOR MODEL NO: TIGER 16/1010 PART NO: 2244025 OPERATION & MAINTENANCE INSTRUCTIONS LS01/13 INTRODUCTION Thank you for purchasing this product. Before attempting to use this product, please

More information

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions L 100 Bubble-Tube Level System Installation, Operation and Maintenance Instructions Figure 1 Contents Section Description Page 1.0 Introduction 2 2.0 Specifications 3 3.0 Installation 3 4.0 Warranty 6

More information

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information