1 An Extractive Method for Determining Particulate Contamination Levels of Wafer Carriers Using Ultrasonic Extraction: SEMASPEC Technology Transfer A-STD

2 and the logo are registered trademarks of, Inc. 1994, Inc.

3 An Extractive Method for Determining Particulate Contamination Levels of Wafer Carriers Using Ultrasonic Extraction: SEMASPEC October 8, 1994 Abstract: Keywords: This specification defines a quantitative measurement method for particulate contamination on the functional surfaces of wafer carriers. This method uses ultrasonic extraction to remove surface contamination and to suspend it in an aqueous solution. A portion of the solution is measured with a liquid-borne particle counter (LPC) to determine the concentration of particles with respect to various sizes. The particle size range that this specification addresses is 0.5 µm to 60 µm. This method will measure cleaning processes to a level that allows six sigma control. SMIF, Contamination, Wafer Transport, Measurement, Specifications, 200 mm Wafers, Particle Count Authors: Approvals: Dr. Nagarajan (IBM) Gary Knoth, Project Engineer Ray Martin, Project Manager Jackie Marsh, Manager, Standards Programs Venu Menon, Director, Contamination Free Manufacturing

4

5 iii Table of Contents 1 EXECUTIVE SUMMARY INTRODUCTION Purpose Scope Referenced Documents FACILITIES, EQUIPMENT, CHEMICALS Class 100 per Federal Standard 209 (or better) environment for performing the test(s) SAFETY PRECAUTIONS PRINCIPLES OF OPERATION AND EQUIPMENT OVERVIEW CALIBRATION TEST PROCEDURE Set-Up Determination of Background Sample Analysis DATA ANALYSIS Calculation Interpretation of Results DATA PRESENTATION PRECISION AND BIAS BIBLIOGRAPHY... 9

6 iv ACKNOWLEDGEMENTS The Project Engineer wishes to acknowledge the contributions of Dr. Nagarajan and Mr. Ron Coplen of IBM-Storage Systems Division, San Jose, CA. Dr. Nagarajan developed the measurement methodology detailed in this document. Dr. Nagarajan is also the author of this document. Mr. Coplen aided Dr. Nagarajan with the development of the measurement methodology and provided all of the equipment information, including the calibration information. The Project Engineer served as the document editor and the contact. He also facilitated the member company review of the draft document and incorporated the reviewer suggestions into the final document, where appropriate.

7 1 1 EXECUTIVE SUMMARY This specification defines a quantitative measurement method for particulate contamination on the functional surfaces of wafer carriers (200 mm standard mechanical interface [SMIF] pods, 200 mm cassettes, and 200 mm run boxes). The method uses ultrasonic extraction to remove surface contamination and to suspend it in an aqueous solution. A portion of the solution is measured with a liquid-borne particle counter (LPC) to determine the concentration of particles with respect to various sizes (for this method, 0.5 µm to 60 µm). This procedure will measure cleaning processes to a level that allows six sigma control 2 INTRODUCTION 2.1 Purpose This specification defines a quantitative measurement method for particulate contamination on the functional surfaces of wafer carriers. 2.2 Scope For the purposes of this specification, "wafer carriers" are 200 mm standard mechanical interface pods, 200 mm wafer cassettes, and 200 mm run boxes. This method uses ultrasonic extraction to remove surface contamination and to suspend it in an aqueous solution. A portion of the solution is measured with a liquid-borne particle counter (LPC) to determine the concentration of particles with respect to various sizes. The particle size range that this specification addresses is 0.5 µm to 60 µm.

8 2 3 Referenced Documents 3.1 Operating and Maintenance Manual for the PMS Micro Laser Particle Spectrometer, P/N Operating and Maintenance Manual for the PMS CLS-600 Sampler, P/N Operating and Maintenance Manual for the PMS IMOLV-.5 Sensor, P/N Instruction Manual for the Branson Ultrasonic Cleaner, Model DHA Federal Standard MIL-STD-1246B, Product Cleanliness Levels and Contamination Control Program, September 4, FACILITIES, EQUIPMENT, CHEMICALS Note: For purposes of repeatability and reproducibility of this test method, the names of specific products used in test development are included. In each case, an equivalent product may be used.

9 3 4.1 Class 100 per Federal Standard 209 (or better) environment for performing the test(s) 4.2 Model DHA-1000 Ultrasonic Cleaner (or equivalent), Branson Cleaning Equipment Company, Shelton, CT Model Micro LPS-8/16 (Micro Laser Particle Spectrometer) (or equivalent), Particle Measuring Systems, Inc., 1855 South 57th Court, Boulder, Colorado 80301, (303) Model CLS-600 Sampler, (Corrosive Liquid Sampler), (or equivalent), Particle Measuring Systems, Inc., 1855 South 57th Court, Boulder, Colorado 80301, (303) Model IMOLV-.5 (LD), Integrated Micro-Optical Liquid Volumetric, Sensor - Solid State Laser Diode), (or equivalent), Particle Measuring Systems, Inc., 1855 South 57th Court, Boulder, Colorado 80301, (303) Distilled or deionized (DI) water, preferably filtered to 0.2 µm at point of use 4.7 Timer (Graylab Universal timer or equivalent is recommended) 4.8 Centering bracket for glassware in the ultrasonic tank 4.9 Printer Paper for the PMS Micro LPS, P/N PR-22 (or equivalent) Technical Equipment Sales (510) Stopwatch 4.11 Hydrophone Type 8103, Calibrator Type 4223, Sound Level Meter Type 2231, (or equivalent), Bruel & Kjaer Instruments, Inc. 185 Forest Street, Marlborough, MA 01752, (617) P/N L Pyrex Glassware (12 1/8 in. X 12 1/8 in. X 12 in. deep), P/N Pyrex Glassware (4 l beaker), (or equivalent), VWR Scientific, San Francisco, CA 94120, (415) SAFETY The equipment listed does not present any unusual hazards. Good safety practice, however, should always be observed when handling glassware. Exercise caution with the ultrasonic cleaner and all electrical connections. Do not turn on the ultrasonic heater. The heating coil is above the water line and may cause injury if turned on. The sensor houses a Class 1 laser, which transmits output inaccessible to the user. This laser is accessible only during maintenance.

10 4 6 PRECAUTIONS 6.1 This test method may involve hazardous materials, operations, and equipment. This test method does not purport to address the safety considerations associated with its use. It is the responsibility of the user to establish appropriate safety and health practices and to determine the applicability of regulatory limitations before using this method. 7 PRINCIPLES OF OPERATION AND EQUIPMENT OVERVIEW 7.1 An ultrasonic generator, located beneath the ultrasonic tank, converts 50/60 Hz line voltage to high frequency (40 khz) electrical energy. This energy is fed to transducers connected externally to the underside of the ultrasonic tank, where it is converted from electrical to mechanical energy. The transducers vibrate longitudinally and transfer this motion as a pressure wave to the cleaning, or extraction, medium. This causes the liquid to cavitate. Ultrasonic waves in the liquid establish a pattern of pressures below and above the vapor pressure of the liquid. Bubbles formed in the low pressure regions explode when subjected to high pressures. Shock waves emitted by exploding bubbles act to scrub particles from immersed surfaces. 7.2 The material removed by ultrasonic extraction and placed in solution is detected by the particle counter. To maintain a constant energy level of extraction, it is recommended that only the Branson ultrasonic unit called out in the equipment section be used. If other ultrasonic tanks are proposed for use instead, a B & K hydrophone or other ultrasonic energy measurement devices may be used to verify that the energy level in the proposed tank is 220 +/- 5 dbs (or equivalent in other units). It is necessary, however, that particle counts obtained be correlated with those obtained using the reference standard, Branson DHA-1000 ultrasonic cleaner, to confirm equivalence. Also to be noted is that the test vessel (or substrate) must always be centered in the ultrasonic tank. 7.3 The IMOLV sensor operates on the principle that laser light scattered by a particle is a direct function of its size. Particles produce pulses of radiant energy during transit through the laser beam. These light pulses are sensed by a detector and are then sized with a sixteen channel pulse height analyzer to determine particle size. 7.4 The Micro LPS accepts particle data from the sensor and the flowmeter, and controls the sampler. The sampler automatically draws solution from the extraction vessel, compresses it to remove any bubbles, passes it through the view volume of the sensor, then drains it into the user s drain or waste container. The test described in this document measures only a portion of the contamination found on the parts tested. Sixty ml is drawn from the sample container and the actual volume of solution tested is 45 ml. 7.5 The B & K hydrophone used to measure ultrasonic intensity in the tank is a piezoelectric transducer; i.e., it uses piezoelectric ceramics as sound-pressure sensing

11 elements. It has a flat frequency response and is omnidirectional over a wide frequency range. 5 8 CALIBRATION 8.1 The IMOLV sensor is calibrated using a standard fluid consisting of polystyrene latex microspheres suspended in water. This calibration is carried out by the manufacturer prior to shipment. The system must be recalibrated annually. To calibrate, follow the instructions in the Operating and Maintenance Manual for the sensor. 9 TEST PROCEDURE 9.1 Set-Up At the initial start-up, pour one inch of water into the ultrasonic tank (the tank water depth is important and should be accurately measured). Then turn the ultrasonic on for five minutes minimum to degas the water Prior to using the system, the Micro LPS and sampler must be adjusted to the following parameters (refer to the operating manuals for details on how to do the adjustments). Micro LPS: Counts/N/mL Compression Delay = Counts (the word counts is displayed on screen) = 10 seconds Sampler: 1. Operating Pressure: The gauge on the front of the sampler displays the operating pressure when the sample solution is being compressed and when it is moving through the sensor during the sample cycle. The gauge must read 20 psig ± 2 psig during these cycles. Adjust the compression pressure knob if an adjustment is necessary. 2. Fill Cycle Time: (Follow the instructions in the operating manual) Rotate the vacuum knob if an adjustment is necessary. 3. Flow Rate: During the sample cycle the flow rate must be 80 ml/min +/- 10 ml/min. Adjust the flow control knob if an adjustment is necessary. 9.2 Determination of Background All surfaces that contact the test fluid must be kept meticulously clean. Care must be exercised to eliminate possible sources of contamination to the sample since the technique is sensitive to environmental contributions. When glassware is used in the form of extraction vessels, clean the insides with a sponge and an unscented liquid detergent. Rinse with overflow water until the background particle count, obtained by the procedure outlined below, meets the stated criteria. Before using any beakers,

12 6 examine the insides visually for signs of chips, pits, cracks, or other physical defects. It is advisable to discard the glassware after about six months of regular use Perform the following steps to create a background solution: 1. Pour 4 L of filtered water into the beaker described in Step Place the suction tube from the sampler into the beaker. 3. Test the background with the PMS system a minimum of three times. The first test result is inaccurate because only a portion of the solution tested is from the contents of the beaker. The second and third tests must be within 10% of each other. The cumulative counts >/= 0.5 µm are to be used for this comparison. If the third reading is not within 10% of the second, then a fourth reading should be taken, and checked for ± 10% agreement with the third. This process may be repeated up to six times, after which the solution must be discarded, and the procedure repeated. 4. Average the two readings that are within 10% of each other. This represents the background count. 5. The background count must not exceed 1000 particles >/= 0.5 µm in 45 ml of solution. Note: If the background solution does not meet the above requirement, additional cleaning or replacement of the glassware, or purging of contaminants from the sampler by running fresh filtered water may be necessary. 9.3 Sample Analysis mm SMIF pods 1. Pour DI water into a 4 L beaker to a level of 3.5 L. 2. Place the test pod into the ultrasonic tank. 3. Pour the DI water into the test pod. 4. Make up three more batches of 3.5 L of filtered water from the same 4 L beaker and pour the contents into the pod. The total volume will be 14 L. Note: Background testing of the latter batches of solution is unnecessary provided that the operator does not introduce any contamination to the solution. 5. Turn the ultrasonic on for two minutes ± two seconds. 6. Place the suction tube into the pod, and test the solution with the PMS system. Test the solution with the PMS system three times minimum. The first test result is inaccurate since only a portion of the solution tested is from the contents of the pod. The second and third tests must be within 10% of each other. The cumulative counts 0.5 µm are to be used for this comparison. If the third reading is not within 10% of the second, then a fourth reading should be taken, and checked for ±

13 10% agreement with the third. This process may be repeated until convergence to ± 10% agreement between any two readings is obtained. 7. Average the two readings that are within 10% of each other. This represents the reading for the pod. 8. Subtract the background count (Section 8.2) from the pod reading in each channel. Minus values should be recorded as zeros. 9. Record the results mm run boxes, 200 mm wafer cassettes 1. Use the 4 L beaker to fill a clean 12 1/8 in. X 12 1/8 in. X 12 in. deep Pyrex glassware with 20 L of DI water. 2. Take a background reading on the water in the Pyrex container. The cumulative count per 45 ml > 0.5 µm must not exceed 1,000. Note: Background counts less than 10% of the reading for the part are acceptable in most cases. 3. Immerse the 200 mm wafer cassette or either half of the 200 mm run box in the solution until it is fully covered. Note: It is recommended that if only one half of the run box is tested, it should be the top half since it is more likely to shed particles on to the product. 4. Ultrasonic clean, as before, for 2 minutes ± 2 seconds, and obtain an average particle count for the sample solution. 5. Subtract the background count from the averaged reading for the part, and record the data DATA ANALYSIS At the conclusion of the test, particle counts are obtained in the sampling volume of 45 ml. The monitor displays both differential and cumulative counts in the 0.5 to 60 µm size range. It is recommended that the cumulative count > 0.5 µm in 45 ml of solution be used as the primary cleanliness parameter. Distribution of particle counts in various size buckets may be examined when attempting process investigation or improvement Calculation The counts in 45 ml may be converted to particle counts per part by scaling the number to the total extraction volume. For example, in the case of the SMIF pods, use the following: Particles per SMIF pod > 0.5 µm = Particles per 45 ml > 0.5 µm * 14,000/45 As an additional refinement, the particles-per-part number may be divided by the total wetted surface area of the part to yield counts per square centimeter. The particle size

14 8 distribution (in the 1-60 µm size range) obtained in this manner may then be plotted on a log-normal graph according to MIL-STD-1246B [1]. MIL-STD-1246 levels of surface cleanliness may then be defined. These levels are useful when comparing different materials or different processes for producing the same product Interpretation of Results When cleanliness criteria are being established, a minimum of 30 parts should be measured for cleanliness. [Note: The actual number of parts to be tested will vary depending on the normality and tightness of the cleanliness distribution.] From this population, a mean and a standard deviation may be established. The upper control limit for the process is typically set as mean + 3 X standard deviation; the acceptance criterion is typically set as mean X standard deviation. This should yield a process that is 6 sigma capable and should be acceptable provided the criterion established according to this procedure satisfies functional requirements. 11 DATA PRESENTATION Cleanliness data may either be presented as differential or cumulative counts relative to individual sizes or as size-concentration distributions, depending on product requirements. Once the critical parameters have been defined, they should be charted using statistical process control techniques. At a minimum, the raw data, calculated averages, standard deviations, and ranges must be charted and used to establish and to respond to trends. Presenting the > 1.0 µm data in MIL-STD-1246B plots offers the added benefit of being able to compare against a reference ( normal ) population. 12 PRECISION AND BIAS The gauge capability of the indirect surface cleanliness measurement method involving ultrasonic extraction and the use of a liquid-borne particle counter has previously been demonstrated to be acceptable. The natural variability in the ultrasonic field is the primary limiting factor in determining the repeatability and reproducibility of this method. A cumulative count of 2,000 particles > 0.5 µm in 45 ml sample volume represents the approximate threshold limit above which the procedure outlined in this specification is clearly valid. Sizing and counting accuracy of the particle counter apparatus are documented [2]. Although ultrasonic extraction will tend to lose some efficiency in the sub-µm size range, the efficiency is nevertheless repeatable and reproducible the primary requirements for a measurement method.

15 9 13 BIBLIOGRAPHY 1. MIL-STD-1246B, "Product Cleanliness Levels and Contamination Control Program, September 4, Welker, R.L. Weaver and R.P. Coplen, "Gauge Capability, Correlation, And Other Analyses of Particle Cleanliness Measurement Methods," IBM San Jose Internal Report, 1991 (unclassified) 3. Nagarajan, "Estimating Surface Cleanliness and Cleanability, in Proceedings, CleanRooms 93 West, September 27-30, 1993, Santa Clara, CA, pp Nagarajan, " Cavitation Erosion of Substrates in Disk Drive Component Cleaning: An Exploratory Study," Wear J., 152 (1992) 75-89

16 10 A. APPENDIX: ALTERNATIVE EXTRACTION PROCEDURES A.1 Low pressure spray extraction may be used in troubleshooting or investigative mode to selectively measure portions of the surface instead of the entire surface. While also offering the advantage of less sample dilution, the method is more subject to operatorinduced variability and is therefore not as capable a gauge as ultrasonic extraction. A.2 Multiple ultrasonic extractions may be performed to assess the cleanability of the surface being measured. The procedure and the definition of cleanability indices are contained in [3]. It should be noted that multiple extraction analyses of the plastic materials involved in this study did not indicate the onset of cavitation-induced erosive damage [4]. While prolonged exposure to sustained sonication can cause many polymers to degrade, a single two-minute extraction or 6-8 stages of extraction that are spaced to allow relaxation of the material will not, in this case, cause detectable surface deterioration. A.3 This procedure uses DI water without surfactant as the extraction medium. This is based on data indicating that fewer particles are extracted with a surfactant than without. Surfactants promote particle removal by wetting the particle/surface interface, but simultaneously lower the ultrasonic energy level in the tank by introducing microbubbles and by lowering the surface tension of water. The interaction between these two conflicting effects is complex and must be understood on a case-by-case basis. If an application requires the use of a surfactant, a low concentration (0.02 percent by volume) of a non-foaming, non-ionic surfactant, (e.g., Alfonic R made by Vista Chemical Co., Chicago, IL) is recommended. The background count in this case must be limited to 2000 particles per 45 ml 0.5 µm, or 10% of the reading for the part, whichever is achievable. NOTICE: DISCLAIMS ALL WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. MAKES NO WARRANTIES AS TO THE SUITABILITY OF THE METHOD FOR ANY PARTICULAR APPLICATION. THE DETERMINATION OF THE SUITABILITY OF THIS METHOD IS SOLELY THE RESPONSIBILITY OF THE USER.

17

18 Technology Transfer 2706 Montopolis Drive Austin, TX

SEMATECH Provisional Test Method for Pressure Cycle Testing Filter Cartridges Used in UPW Distribution Systems

SEMATECH Provisional Test Method for Pressure Cycle Testing Filter Cartridges Used in UPW Distribution Systems Provisional Test Method for Pressure Cycle Testing Filter Cartridges Used in UPW Distribution Systems Technology Transfer 92010941B-STD and the logo are registered service marks of, Inc. 1996, Inc. Provisional

More information

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev - 1 W. W. Hansen Experimental Physics Laboratory STANFORD UNIVERSITY STANFORD, CALIFORNIA 94305-4085 Gravity Probe B Relativity Mission OPERATION OF LIQUID PARTICLE COUNTER GP-B P0045 Rev - May 23, 1997

More information

SEMASPEC Test Method for Determination of Cycle Life of Automatic Valves for Gas Distribution System Components

SEMASPEC Test Method for Determination of Cycle Life of Automatic Valves for Gas Distribution System Components SEMASPEC Test Method for Determination of Cycle Life of Automatic Valves for Gas Distribution System Components Technology Transfer 90120395B-STD and the logo are registered service marks of, Inc. 1996,

More information

Level MEASUREMENT 1/2016

Level MEASUREMENT 1/2016 Level MEASUREMENT 1/2016 AGENDA 2 A. Introduction B. Float method C. Displacer method D. Hydrostatic pressure method E. Capacitance method G. Ultrasonic method H. Radar method I. Laser method J. Level

More information

SPECIFICATIONS PARTICLE SENSOR KS-19F Higashimotomachi, Kokubunji, Tokyo , Japan

SPECIFICATIONS PARTICLE SENSOR KS-19F Higashimotomachi, Kokubunji, Tokyo , Japan SPECIFICATIONS PARTICLE SENSOR KS-19F 3-20-41 Higashimotomachi, Kokubunji, Tokyo 185-8533, Japan No. 13012-3E 17-06 Printed in Japan Outline The KS-19F is a sensor which uses the light scattering method

More information

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions L 100 Bubble-Tube Level System Installation, Operation and Maintenance Instructions Figure 1 Contents Section Description Page 1.0 Introduction 2 2.0 Specifications 3 3.0 Installation 3 4.0 Warranty 6

More information

T EK-SUB 4800C 19 mm Submersible Level Transmitter

T EK-SUB 4800C 19 mm Submersible Level Transmitter Technology Solutions T EK-SUB 4800C 19 mm Submersible Level Transmitter Instruction Manual Document Number: IM-4800C www.tek-trol.com Table of Contents 1 Safety Instructions... 2 1.1 Intended Use... 2

More information

SPECIFICATIONS PARTICLE SENSOR KS-18F Higashimotomachi, Kokubunji, Tokyo , Japan

SPECIFICATIONS PARTICLE SENSOR KS-18F Higashimotomachi, Kokubunji, Tokyo , Japan SPECIFICATIONS PARTICLE SENSOR KS-18F 3-20-41 Higashimotomachi, Kokubunji, Tokyo 185-8533, Japan No. 05083-5E 18-01 Printed in Japan Outline The KS-18F is a sensor which uses the light scattering method

More information

ANALOG ULTRASOUND WATTMETER USP-30

ANALOG ULTRASOUND WATTMETER USP-30 ANALOG ULTRASOUND WATTMETER USP-30 USER MANUAL BC BIOMEDICAL USP-30 ULTRASOUND WATTMETER TABLE OF CONTENTS WARNINGS, CAUTIONS, NOTICES... ii DESCRIPTION... 1 TEST SETUP... 2 TESTING... 5 MANUAL REVISIONS...

More information

Biography. Abstract. Keywords. Body box, garments, validation, calibration, clean garments, IEST-RP-CC003.3, garment. Introduction and Scope

Biography. Abstract. Keywords. Body box, garments, validation, calibration, clean garments, IEST-RP-CC003.3, garment. Introduction and Scope NEW METHOD CALIBRATION FOR CALIBRATION OF A BODY BOX [Paper presented at IEST s ESTECH 2003 Conference, Phoenix, AZ, May 18-21] Rajan Jaisinghani, Technovation Systems, Inc. Biography Rajan (Raj) Jaisinghani

More information

ISO INTERNATIONAL STANDARD. Hydraulic fluid power Filter elements Determination of resistance to flow fatigue using high viscosity fluid

ISO INTERNATIONAL STANDARD. Hydraulic fluid power Filter elements Determination of resistance to flow fatigue using high viscosity fluid INTERNATIONAL STANDARD ISO 23181 First edition 2007-08-15 Hydraulic fluid power Filter elements Determination of resistance to flow fatigue using high viscosity fluid Transmissions hydrauliques Éléments

More information

Ti-Pure TITANIUM DIOXIDE DETERMINATION OF SLURRY RHEOLOGY METHOD: T WP

Ti-Pure TITANIUM DIOXIDE DETERMINATION OF SLURRY RHEOLOGY METHOD: T WP Ti-Pure TITANIUM DIOXIDE DETERMINATION OF SLURRY RHEOLOGY METHOD: T4400.308.01.WP 1 TITANIUM DIOXIDE (Ti-Pure ) Rheology of Titanium Dioxide Slurries I. Principle Rheology is a measurement of the fluid

More information

USER S GUIDE AUTOTUNE SERIES HIGH INTENSITY ULTRASONIC PROCESSOR Watt Model TABLE OF CONTENTS

USER S GUIDE AUTOTUNE SERIES HIGH INTENSITY ULTRASONIC PROCESSOR Watt Model TABLE OF CONTENTS USER S GUIDE AUTOTUNE SERIES HIGH INTENSITY ULTRASONIC PROCESSOR 1500 Watt Model TABLE OF CONTENTS Warranty Important Safeguards Low Surface Tension Liquids - Organic Solvents SECTION 1 INSTALLATION Inspection

More information

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality

More information

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the 1151 METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS 1.0 Scope and Application. 1.1 Analytes. Analyte Volatile Organic Compounds (VOC) CAS No. No CAS number assigned 1.2 Scope. This method

More information

Hygea Ultrasonic Validation Kit

Hygea Ultrasonic Validation Kit Hygea Ultrasonic Validation Kit Ultrawave s Hygea Ultrasonic Validation kit contains the tools and accessories required to perform Validation and Periodic testing of your ultrasonic cleaning bath. Designed

More information

Agilent Flowmeter ADM1000

Agilent Flowmeter ADM1000 Agilent Flowmeter ADM1000 Operating Instructions Notices Agilent Technologies, Inc. 2003 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval

More information

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR) Procedure No: SOP-025 Revision No: 1.0 (January 21, 2011) Page No.: 1 of 8 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to

More information

DPC-30 DPC-100. Reference Manual

DPC-30 DPC-100. Reference Manual DPC-30 DPC-100 Reference Manual 1. Introduction 1.1 Description The Martel DPC Digital Pneumatic Calibrator improves upon traditional dial gauge pneumatic calibrators. The Martel DPC improves accuracy,

More information

Oil Diagnostic Systems. Online and bottle measurement Mobile and stationary operation Lab quality accuracy e

Oil Diagnostic Systems. Online and bottle measurement Mobile and stationary operation Lab quality accuracy e Oil Diagnostic Systems Online and bottle measurement Mobile and stationary operation Lab quality accuracy 100.90-1e The new generation of oil diagnostics Mobile and stationary operation: Compressed air

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

RADIATION PROCEDURES MANUAL Procedure Cover Sheet RADIATION PROCEDURES MANUAL Procedure Cover Sheet Procedure Title: Calibration of Eberline Portable Particulate Noble Gas Monitor Procedure Number: TSO-08-13-REV 1 Effective Date: July 1, 2008 Approved

More information

QUICK START GUIDE TO DRINKING WATER MONITORING

QUICK START GUIDE TO DRINKING WATER MONITORING APPENDIX F QUICK START GUIDE TO DRINKING WATER MONITORING Introduction: This Quick Start guide is designed to allow users who are already familiar with the basic use of YSI 6-series sondes for surface

More information

Instrumentation & Data Acquisition Systems

Instrumentation & Data Acquisition Systems Instrumentation & Data Acquisition Systems Section 3 -Level Robert W. Harrison, PE Bob@TheHarrisonHouse.com Made in USA 1 Level Section Question Which level measuring technology is the best solution when

More information

TEK-SUB 4800B. Submersible Level Transmitter. Flow Level Temperature Pressure Valves Analyzers Accessories TekValSys LEVEL

TEK-SUB 4800B. Submersible Level Transmitter.   Flow Level Temperature Pressure Valves Analyzers Accessories TekValSys LEVEL Technology Solutions TEK-SUB 4800B Submersible Level Transmitter LEVEL Flow Level Temperature Pressure Valves Analyzers Accessories TekValSys Introduction The Tek-Sub 4800B Submersible Level Transmitter

More information

Installation, Operation & Maintenance Manual

Installation, Operation & Maintenance Manual Original Instructions Installation, Operation & Maintenance Manual Sentry Cobra FV Sampler Low Emission Samplers S-MS-IOM-00543-0 08-18 Table of Contents Safety Information... 3 General Safety Precautions...

More information

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015 page 1 of 6 TNI Standard EL-V1M4 Sections 1.7.1 and 1.7.2 September 2015 Description This TNI Standard has been taken through all of the voting stages and has received consensus approval by the TNI membership.

More information

GILMONT ACCUCAL FLOWMETERS

GILMONT ACCUCAL FLOWMETERS OPERATING MANUAL GILMONT ACCUCAL FLOWMETERS 28W092 Commercial Ave. Barrington, IL U.S.A. 60010-2392 (847) 381-4888 (847) 381-7053 (Fax) 800-962-7142 www.barnant.com e-mail: barnant@barnant.com A-1299-0766

More information

Real-Time Smoothness Measurements on Concrete Pavements During Construction

Real-Time Smoothness Measurements on Concrete Pavements During Construction Recommended Practice for Real-Time Smoothness Measurements on Concrete Pavements During Construction XX-## (2017) 1. SCOPE 1.1. This document provides language that can be used by an Owner-Agency to develop

More information

BGI, INC. WALTHAM, MA 02451

BGI, INC. WALTHAM, MA 02451 BGI, INC. WALTHAM, MA 02451 JANUARY, 2002 58 Guinan Street Waltham, MA 02451 Tel: 781-891-9380 Fax: 781-891-8151 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

More information

Dust Measurement & SIBATA Products. Sibata Scientific Technology Ltd.

Dust Measurement & SIBATA Products. Sibata Scientific Technology Ltd. Dust Measurement & SIBATA Products Sibata Scientific Technology Ltd. Content Conditions of Dust Measurement in Japan Mass Concentration Method & SIBATA Products Relative Concentration Method & SIBATA Products

More information

RH800 & 2000 BASIC OPERATIONS

RH800 & 2000 BASIC OPERATIONS RH800 & 2000 BASIC OPERATIONS 2 General information This manual contains technical information regarding Bayer SeedGrowth Equipment. Please read and understand these instructions completely before proceeding

More information

PREPARED BY: Marshall K. Cheung, Ph.D., Laboratory Director. REVISED BY: Marshall K. Cheung, Ph.D., Laboratory Director

PREPARED BY: Marshall K. Cheung, Ph.D., Laboratory Director. REVISED BY: Marshall K. Cheung, Ph.D., Laboratory Director DOCUMENT TYPE: DOCUMENT CLASS: Standard Operating Procedure Physical Property Procedure TITLE: Conductivity, EPA 120.1 INSTRUMENTATON: HACH CO150 Conductivity Meter PREPARED BY: Marshall K. Cheung, Ph.D.,

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 36-751 XenoGard 133 Xe Monitor Operators Manual March 2005 Manual No. 128011 Rev. 3 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

Air Displacement Pipetting Modules. Total Integrated Liquid Handling

Air Displacement Pipetting Modules. Total Integrated Liquid Handling Air Displacement Pipetting Modules Total Integrated Liquid Handling Intelligent Liquid Handling For Better Results MOVING LIQUID IS EASY; PRECISION LIQUID HANDLING IS HARD Instrument manufacturers frequently

More information

Smart Water Application Technologies (SWAT)

Smart Water Application Technologies (SWAT) Smart Water Application Technologies (SWAT) Turf and Landscape Irrigation Equipment PRESSURE REGULATING SPRAY HEAD SPRINKLERS Equipment Functionality Test Testing Protocol Version 3.0 (May 2012) Developed

More information

EC214 EC215 - EC215R Bench Conductivity Meters

EC214 EC215 - EC215R Bench Conductivity Meters Instruction Manual EC214 EC215 - EC215R Bench Conductivity Meters http://www.hannainst.com These Instruments are in Compliance with the CE Directives Dear Customer, Thank you for choosing a Hanna Instruments

More information

ISO/TC 131/SC Hydraulic fluid power Monitoring the level of particulate contamination in the fluid

ISO/TC 131/SC Hydraulic fluid power Monitoring the level of particulate contamination in the fluid DRAFT INTERNATIONAL STANDARD ISO/DIS 21018-4 Hydraulic fluid power Monitoring the level of particulate contamination in the fluid Part 4: Use of the light extinction technique Transmissions hydrauliques

More information

1999 Microcomponents Technology SLYA017A

1999 Microcomponents Technology SLYA017A Application Brief Number 001 1999 Microcomponents Technology SLYA017A IMPORTANT NOTICE Texas Instruments and its subsidiaries (TI) reserve the right to make changes to their products or to discontinue

More information

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD) Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet

More information

ISO 3724 INTERNATIONAL STANDARD. Hydraulic fluid power Filter elements Determination of resistance to flow fatigue using particulate contaminant

ISO 3724 INTERNATIONAL STANDARD. Hydraulic fluid power Filter elements Determination of resistance to flow fatigue using particulate contaminant INTERNATIONAL STANDARD ISO 3724 Second edition 2007-08-15 Hydraulic fluid power Filter elements Determination of resistance to flow fatigue using particulate contaminant Transmissions hydrauliques Éléments

More information

Denise L Seman City of Youngstown

Denise L Seman City of Youngstown Denise L Seman City of Youngstown The control chart is one of the most important tools of quality control for laboratory data. A control chart is a specific kind of run chart that allows unusual change

More information

Standard Recommended Practice. Field Measurement of Surface Profile of Abrasive Blast-Cleaned Steel Surfaces Using a Replica Tape

Standard Recommended Practice. Field Measurement of Surface Profile of Abrasive Blast-Cleaned Steel Surfaces Using a Replica Tape NACE Standard RP0287-2002 Item No. 21035 Standard Recommended Practice Field Measurement of Surface Profile of Abrasive Blast-Cleaned Steel Surfaces Using a Replica Tape This NACE International standard

More information

FLUORESCENCE DETERMINATION OF OXYGEN

FLUORESCENCE DETERMINATION OF OXYGEN FLUORESCENCE DETERMINATION OF OXYGEN Objectives This experiment will familiarize the students with the principles of fluorescence quenching as well as the Ocean Optics FOXY Fiber Optic Sensor System. Spectra

More information

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING Blair Chalpin Charles A. Matthews Mechanical Design Engineer Product Support Manager Scanivalve Corp Scanivalve Corp Liberty Lake, WA

More information

NT High-temperature flowmeter model Simultaneous flow and pressure outputs for high-temperature applications

NT High-temperature flowmeter model Simultaneous flow and pressure outputs for high-temperature applications NT High-temperature flowmeter model 4401 Simultaneous flow and pressure outputs for high-temperature applications high-temperature flow measurement Overview Now more than ever there is a need to measure

More information

Model DPC3500 Continuous Low Range Dew Point Analyzer. Operations Manual

Model DPC3500 Continuous Low Range Dew Point Analyzer. Operations Manual Model DPC3500 Continuous Low Range Dew Point Analyzer Operations Manual Please read, understand, and follow these instructions before operating this equipment. Super Systems, Inc. is not responsible for

More information

QUARTERLY REPORT NO. 3 DE-FG22-94PC /01/95

QUARTERLY REPORT NO. 3 DE-FG22-94PC /01/95 QUARTERLY REPORT NO. nvestigation of the Effect of Coal Particle Sizes on the nterfacial and Rheological Properties of Coal-Water Slurry Fuels CONTRACT NO.: DE-FG-PC TECHNCAL MONTOR Dr. S u n g S. Kim

More information

Measuring Mass and Volume

Measuring Mass and Volume Measuring Mass and Volume Experiment 2 Expt 2 Measurement.wpd INTENT The purpose of this experiment is to introduce some fundamental aspects of the measurement making process as well as to introduce some

More information

ANL Bubble Chamber Superheated Active Target System

ANL Bubble Chamber Superheated Active Target System ANL Bubble Chamber Superheated Active Target System N 2 O and C 2 F 6 as Active Fluids B. DiGiovine Physics Division and Bubble Chambers April 2009 First Bubble Chamber Received Full Operation Authorization

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010 Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600 November 2010 Table of Contents ``Abstract...3 ``Introduction...4 ``Methods & Results...6 ``Data Analysis...9 ``Conclusion...12

More information

T EK-COR 1100A. Coriolis Mass Flowmeter. FLOW. Technology Solutions

T EK-COR 1100A. Coriolis Mass Flowmeter.   FLOW. Technology Solutions Technology Solutions T EK-COR 1100A Coriolis Mass Flowmeter FLOW www.tek-trol.com Flow Level Temperature Pressure Valves Analyzers Accessories TekValSys Introduction There can often be more than one type

More information

PERFORM Operating Document

PERFORM Operating Document PERFORM Operating Document Use and Maintenance of CO 2 Incubator PC-POD-CA-007-v03 Revision History Version Reason for Revision Date 01 New POD 30-Sep-13 02 Minor revisions for section 2.3, 3.1, 4.3. 14-April-16

More information

Engineering: Measurement Technology Pressure/Level (SCQF level 6)

Engineering: Measurement Technology Pressure/Level (SCQF level 6) National Unit Specification: general information CODE F5KT 12 SUMMARY This Unit can be delivered as part of a National Qualification Group Award but can also be taken as a free-standing Unit. This Unit

More information

Acoustic signals for fuel tank leak detection

Acoustic signals for fuel tank leak detection Acoustic signals for fuel tank leak detection More info about this article: http://www.ndt.net/?id=22865 Wenbo Duan 1, Siamak Tavakoli 1, Jamil Kanfoud 1, Tat-Hean Gan 1, Dilhan Ocakbaşı 2, Celal Beysel

More information

Drilling Efficiency Utilizing Coriolis Flow Technology

Drilling Efficiency Utilizing Coriolis Flow Technology Session 12: Drilling Efficiency Utilizing Coriolis Flow Technology Clement Cabanayan Emerson Process Management Abstract Continuous, accurate and reliable measurement of drilling fluid volumes and densities

More information

HI 2314 HI 2315 HI 23151

HI 2314 HI 2315 HI 23151 Instruction Manual HI 2314 HI 2315 HI 23151 Multi-Range Conductivity Meters for Laboratories www.hannainst.com Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction

More information

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi

More information

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI NanoSight NS300 Operation instructions Injection/flushing brief overview: 1. Do not exceed flow of 1 ml per 20 seconds. 2. Inject two 1 ml syringes with nano-pure or DI water. 3. If the water does not

More information

Coriolis Mass Flow Meter

Coriolis Mass Flow Meter Coriolis Mass Flow Meter TMFW Series Mini Type Coriolis Mass Flow Meter Xi an Tosilon Automation Co., Ltd No.299, Daqing Rd, Lianhu District, Xi'an Shaanxi, China Tel: +86-29-8823 8550 info@tosilon.com;

More information

New power in production logging

New power in production logging New power in production logging Locating the zones where fluids enter the wellbore in a producing or injecting well is an important aspect of production logging. It is relatively straightforward to establish

More information

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2. STANDARD OPERATING PROCEDURES Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.5 ) Sampler (RAAS) AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION

More information

PERFORM Operating Document

PERFORM Operating Document PERFORM Operating Document Use and Maintenance of CO 2 Incubator PC-POD-CA-007-v02 Revision History Version Reason for Revision Date 01 New POD 30-Sep-13 02 Minor revisions for section 2.3, 3.1, 4.3. 14-April-16

More information

VLI Visible Level Interface Operation & Maintenance Manual

VLI Visible Level Interface Operation & Maintenance Manual VLI Visible Level Interface Operation & Maintenance Manual All rights reserved. Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted

More information

BS ml Bottle Sampling Unit

BS ml Bottle Sampling Unit BS500 500ml Bottle Sampling Unit User Guide www.mpfiltri.co.uk 200.307-EN Covers Model Numbers BS500 SAFETY WARNING Hydraulic systems contain dangerous fluids at high pressures and temperatures. Installation,

More information

CTG Certification Protocol -- Version 3.0 SUMMARY

CTG Certification Protocol -- Version 3.0 SUMMARY Date of Test: Test Report Number: SUMMARY Date of Prior Certification: Certified by: Re-certification Due Date: Unit Description: Make : Model: Serial Number: Location: CERTIFIER INFORMATION Name: Client

More information

Calibration Requirements for Direct Reading Confined Space Gas Detectors

Calibration Requirements for Direct Reading Confined Space Gas Detectors : Calibration Requirements for Direct Reading Confined Space Gas Detectors However, the definition of bump test has always been a little slippery. Some manufacturers differentiate between a bump test that

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

HAYWARD FLOW CONTROL TFS SERIES FLOW METER INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

HAYWARD FLOW CONTROL TFS SERIES FLOW METER INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS HAYWARD FLOW CONTROL TFS SERIES FLOW METER INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS PLEASE READ THE FOLLOWING INFORMATION PRIOR TO INSTALLING AND USING HAYWARD TFS SERIES FLOW METERS. FAILURE

More information

Fingertip Pulse Oximeter

Fingertip Pulse Oximeter Instruction Manual Fingertip Pulse Oximeter Item # 40-810-000 Item # 40-811-000 Item # 40-812-000 Item # 40-813-000 Please read this guidebook completely before operating this unit. Limited Two-Year Warranty

More information

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel: PERMA PURE INSTRUCTION MANUAL FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model 3300 Version 4.06 Perma Pure LLC Tel: 732-244-0010 P.O. Box 2105, 8 Executive Drive Tel: 800-337-3762 (toll free US)

More information

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW AC 2010-2145: MEASUREMENT OF HYDROGEN IN HELIUM FLOW Randy Buchanan, University of Southern Mississippi Christopher Winstead, University of Southern Mississippi Anton Netchaev, University of Southern Mississippi

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

R2S Standard Sampling Assembly OPERATION PROCEDURE

R2S Standard Sampling Assembly OPERATION PROCEDURE PURPOSE 1 of 10 To describe the procedure to monitor for viable airborne bacteria with the Remote-Slit- Sampler Assembly (R2S) from EMTEK, LLC. Principle A 100 mm agar based test plate is placed on the

More information

Any laboratory is equipped with specific tools, equipment,

Any laboratory is equipped with specific tools, equipment, Use of Laboratory Equipment and Supplies 3 When you have completed this exercise, you will be able to: 1. Use a balance. 2. Use pipettes and graduated cylinders to measure the volume of liquids. 3. Use

More information

Leak Checking Large Vacuum Chambers

Leak Checking Large Vacuum Chambers Leak Checking Large Vacuum Chambers Technical Overview Vacuum Technologies Introduction Understanding the pump-down characteristics of a large vacuum vessel is critical for determining whether the vacuum

More information

TR Electronic Pressure Regulator. User s Manual

TR Electronic Pressure Regulator. User s Manual TR Electronic Pressure Regulator Page 2 of 13 Table of Contents Warnings, Cautions & Notices... 3 Factory Default Setting... 4 Quick Start Procedure... 5 Configuration Tab... 8 Setup Tab... 9 Internal

More information

WIKA INSTRUMENT CORPORATION

WIKA INSTRUMENT CORPORATION WIKA INSTRUMENT CORPORATION Instruction Manual DIFFERENTIAL PRESSURE GAUGE Series 1500 Series 1000 Series 300 WIKA Instrument Corporation 1000 Wiegand Boulevard Lawrenceville, GA 30043 1-888-945-2872 http://www.wika.com

More information

<788> PARTICULATE MATTER IN INJECTIONS

<788> PARTICULATE MATTER IN INJECTIONS PARTICULATE MATTER IN INJECTIONS This general chapter is harmonized with the corresponding texts of the European Pharmacopoeia and/or the Japanese Pharmacopoeia. These pharmacopeias have undertaken

More information

UsER manual for Watersens ph -REDOX

UsER manual for Watersens ph -REDOX UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5

More information

Truck-mounted Mass Flow Metering for LPG Delivery

Truck-mounted Mass Flow Metering for LPG Delivery Truck-mounted Mass Flow Metering for LPG Delivery Itron, Inc. 1310 Emerald Rd. Greenwood, SC 29646 Introduction The mathematician Gustave De Coriolis first described the Coriolis Effect, which is an inertial

More information

STANDARDS FOR COMPRESSED GAS TESTING

STANDARDS FOR COMPRESSED GAS TESTING STANDARDS FOR COMPRESSED GAS TESTING APPLICATION NOTE HPD-002 (US) Introduction This publication provides excerpts from some of the many guidelines and standards that pertain to compressed gases in cleanrooms.

More information

SEMI E is due for its Five Year Review. This process is required by the SEMI Regulations to ensure that the standard is still valid.

SEMI E is due for its Five Year Review. This process is required by the SEMI Regulations to ensure that the standard is still valid. Background Statement for SEMI Draft Document 4677 REAPPROVAL OF SEMI E66-1103 TEST METHOD FOR DETERMINING PARTICLE CONTRIBUTION BY MASS FLOW CONTROLLERS Note: This background statement is not part of the

More information

Technical Procedure for General Laboratory Equipment

Technical Procedure for General Laboratory Equipment Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all

More information

Best Practice for Calibrating LTH Conductivity Instruments

Best Practice for Calibrating LTH Conductivity Instruments Application Note Best Practice for Calibrating LTH Conductivity Instruments As accurate process measurement becomes an everyday requirement it is vital to be able to calibrate conductivity instruments

More information

CHM250 Calibration and Measurement Lab. Balance Calibration

CHM250 Calibration and Measurement Lab. Balance Calibration CHM250 Calibration and Measurement Lab Green Profile Balance Calibration Introduction: Balances that are properly operated, calibrated and maintained are crucial for laboratory operations. The accuracy

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 16232-9 First edition 2007-06-01 Road vehicles Cleanliness of components of fluid circuits Part 9: Particle sizing and counting by automatic light extinction particle counter

More information

Elcometer 138. Bresle Kit and Patches. Operating Instructions

Elcometer 138. Bresle Kit and Patches. Operating Instructions Elcometer 138 English Bresle Kit and Patches Operating Instructions R English The Conductivity Meter model B-173 has been tested in accordance with EU regulations governing Electro-magnetic compliance

More information

Development of a Non-Destructive Test Machine for Testing Tunnel Lining Concrete

Development of a Non-Destructive Test Machine for Testing Tunnel Lining Concrete Development of a Non-Destructive Test Machine for Testing Tunnel Lining Concrete Seiichi Matsui, Humihiro Osada Technical Research and Development Dept. Railway Operations Headquarters West Japan Railway

More information

Air Sensor. FCS Ex. Manual. AQ Elteknik AB

Air Sensor. FCS Ex. Manual. AQ Elteknik AB Air Sensor FCS Ex Manual AQ Elteknik AB Air Sensor Model FCS Ex Manual version 2.0 Oct 2010 AQ Elteknik AB AQ Elteknik AB Air Sensor Manual 2 Table of contents 1. Manufacturer information... 4 Manufacture

More information

Instruction Sheet. Calibration of Multiple 1720 and FT660 Turbidimeters

Instruction Sheet. Calibration of Multiple 1720 and FT660 Turbidimeters Instruction Sheet 26596-89 Calibration of Multiple 1720 and FT660 Turbidimeters Overview This instruction sheet describes the calibration of up to eight 1720C, 1720D, or 1720E turbidimeters, or up to three

More information

The Discussion of this exercise covers the following points: Range with an elevated or suppressed zero Suppressed-zero range Elevated-zero range

The Discussion of this exercise covers the following points: Range with an elevated or suppressed zero Suppressed-zero range Elevated-zero range Exercise 4-3 Zero Suppression and Zero Elevation EXERCISE OBJECTIVE In this exercise, you will learn the effect that mounting a pressure transmitter above or below the reference level has on the hydrostatic

More information

ICE LOADS MONITORING SYSTEMS

ICE LOADS MONITORING SYSTEMS Guide for Ice Loads Monitoring Systems GUIDE FOR ICE LOADS MONITORING SYSTEMS MAY 2011 American Bureau of Shipping Incorporated by Act of Legislature of the State of New York 1862 Copyright 2011 American

More information

Positive displacement proportioning of two-component epoxy intumescent coatings for passive fire protection

Positive displacement proportioning of two-component epoxy intumescent coatings for passive fire protection Positive displacement proportioning of two-component epoxy intumescent coatings for passive fire protection WHITE PAPER In today s protective coating industry, there is a growing demand to proportion and

More information

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers Slemeyer, A. University of Applied Sciences of Giessen, Dept. of Electrical Engineering 1 Wiesenstr. 14, D-35390

More information

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION Application Note Doc. G8108-001 Rev. A - 23-Jul-02 ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION For more information regarding Digiquartz products contact: Paroscientific,

More information

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS APPLICATION NOTE LC-136 Introduction Specialized spaces often times must

More information

ISS0065 Control Instrumentation Lecture 12

ISS0065 Control Instrumentation Lecture 12 ISS0065 Control Instrumentation Lecture 12 1 Flow measurement Flow, defined as volume per unit of time at specified temperature and pressure conditions. The principal classes of flow-measuring instruments

More information

Elcometer 138. Bresle Kit and Patches. Operating Instructions

Elcometer 138. Bresle Kit and Patches. Operating Instructions Elcometer 138 English Bresle Kit and Patches Operating Instructions R English This product meets the Electromagnetic Compatibility Directive. The product is Class B, Group 1 ISM equipment according to

More information