TEST REPORT BEAMSCAN WITH HALCYON
|
|
- Erin Chase
- 6 years ago
- Views:
Transcription
1 TEST REPORT BEAMSCAN WITH HALCYON UZ Leuven, November 20 th, = Ideal BEAMSCAN tank dimensions to fit perfectly into the Halcyon bore Fast tank positioning using the Halcyon lasers, no electronics on the couch Automatic and fast filling with the function Match water level to sensor Auto-correction of couch tilt using the function Leveling in less than 2 minutes No manual adjustment due to BEAMSCAN s unique AutoSetup features Easy and fast setup in less than 20 minutes Convincing and fast measurement results due to continuous scanning
2 BEAMSCAN measurements at Halcyon (UZ Leuven, Belgium) A few weeks after the installation of the first Halcyon in Europe PTW got the possibility to perform measurements with BEAMSCAN and in order to demonstrate how easy the use is with the automatic setup procedures of BEAMSCAN. The goal was to elaborate an optimal workflow for all BEAMSCAN users with Halcyon. Preparation: For the use of BEAMSCAN together with Halcyon only a longer water hose (1.6 m) is needed. Workflow: 1. Position the BEAMSCAN beside the patient couch (handle towards Halcyon). Connect the power cable and LAN cable to the PC (or use wifi access point) and switch on BEAMSCAN. 2. Put the Varian service plate onto the couch to protect the couch against scratches (with one side approx. 29 cm from the upper end of the couch). Lift the couch to a height of approx. 75 cm above the floor
3 3. Move the BEAMSCAN lift to the same height as the couch. 4. Shift the BEAMSCAN PMMA-tank to the couch and turn the PMMA-tank clockwise 180. It should be placed 40 cm from the end of the couch and with the crosshair at the bottom in the middle of the couch. Switch control pendant (ipod) on and perform a reference run. Now fix the field chamber using TRUFIX holder
4 5. Move the couch slowly into the bore and shift the PMMA-tank on the couch so that the crosshair at the bottom is aligned with the Halcyon cross laser from the top. Then move the couch up/downwards until the laser from the side is visible in the middle of the side markings of the BEAMSCAN tank (red arrow). 6. Then move the couch with BEAMSCAN into the measuring position (isocenter) and position the reference chamber using the small cross at the top of the bore
5 7. Use the BEAMSCAN control pendant (ipod) to move the detector in B to position 100 mm and set a new zero point. Then perform the function match water level to sensor. Now the water is pumped automatically into the tank to the desired filling height within 4 minutes. 8. The total weight of the (not completely) filled water phantom (approx. 192 kg) leads to an elastic deflection of the couch of a few millimeters. Use the control pendant and perform the function Leveling. 9. Use the MV imager to check that the water surface is exactly at SSD 90 cm. Move the MV imager to 84.3, open the upper jaw to 5 cm and the lower jaw to 14 cm so you get an orthogonal beam parallel to the water surface. Use the Halcyon measuring tool to measure the distance between isocenter (yellow cross) and water surface. This should be 100 mm. Correct a deviation by changing the couch height
6 10. For your information, you can use the MV imager to check the position of the detector. Using a brass build-up cap increases the visibility of the chamber. It is not necessary to correct a deviation from the beam center manually. This will be done automatically using the BEAMSCAN AutoSetup function BeamCenter Adjustment. 11. Start BEAMSCAN software, switch Halcyon beam on and perform the functions BeamCenterAdjustment and AutoFieldAlignment with 10 x 10 cm field size. Hint: Due to the elastic deflection of the couch an inclination angle of approx. 0.3 in C is expected. The Leveling function has this measured and a coordinate transformation has been performed. An inclination angle of 0.3 corresponds to an incline of the 500 mm total scanning range of approx. 2.5 mm
7 12. AutoSetup is closed and a task list for Halcyon measurements is loaded. 13. With the task list PDD s and in-plane and cross-plane profiles for the field sizes 2 x 2 cm, 4 x 4 cm, 6 x 6 cm, 8 x 8 cm, 10 x 10 cm, 20 x 20 cm and 28 x 28 cm in the depths 13 mm, 50 mm, 100 mm and 200 mm have been measured within about 30 minutes
8 14. We analyzed for example the PDD for the field size 10 x 10 cm. An analysis according to IAEA 398 showed R100 at mm depth. 15. We analyzed also the in-plane profile at 100 mm depth (isocenter). The analyze is automatically performed according to Fogliata resulting in a CAX deviation of mm, a field size of 9.957cm and a penumbra of 6.21 mm left and 5.95 mm right
9 Conclusion: Following this workflow allows a quick and easy setup of BEAMSCAN at Halcyon in about 20 minutes. The white marks and lines on the BEAMSCAN tank allow an easy adjustment using the Halcyon lasers. The BEAMSCAN tank fits perfectly into the 1 m bore of the Halcyon allowing the use of the whole scanning range of BEAMSCAN in in-plane and cross-plane direction. The big advantages of BEAMSCAN are the AutoSetup functions like Match water level to sensor which allows an automatic filling to the desired filling height, Leveling which performs the auto-correction of the couch tilt in less than 2 minutes. PDD s have been measured with 5 mm/s scanning speed, profiles with 10 mm/s scanning speed and all curves show very smooth raw data. The results of these test measurements are very good. Handling of the BEAMSCAN PMMA-tank (shift onto the couch) could probably be improved with an additional transfer plate with handles. Also, the contrast of the Semiflex 3D at the MV imager could be improved when using a brass build-up cap or a special marker-cap for the chamber. Thanks to its unique patented automatic functions BEAMSCAN constitutes a perfect solution for measurements at the Halcyon machine. We would like to express our special thanks to the Leuven Halcyon team for their sincere guidance, helpful information, and all friendly support they provided us during our tests
13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.
13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical
More informationProton Therapy QA Tools
Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially
More informationRelative Dosimetry. Photons
Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge
More informationClinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS
More informationVariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator
VariSource High Dose Rate Afterloader Procedures For Performing Breast Brachytherapy with The SAVI TM Applicator Contents I. CT Evaluation (~1 week prior to insertion) II. SAVI Applicator Implantation
More informationZoubir Ouhib Lynn Cancer Institute
Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy
More informationOnly 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51
More informationCHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS
47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the
More informationUser Manual MP3-S Therapy Beam Analyzer
User Manual MP3-S Therapy Beam Analyzer D176.131.0/2 2006-02 Hn/zi General Information General Information The product bears the CE-mark "CE-0124" in accordance with the Council Directive 93/42/EEC about
More informationEN-MOTION TREADMILL. EN-Motion: the ultimate PHYSIO treadmill
The EN-Motion treadmill is a valuable asset to the physiotherapy and rehabilitation world. An innovative treadmill, ideal for physiotherapy and rehabilitation purposes. EN-MOTION GETS THE WORLD MOVING
More informationSELF-LEVELLING PLANERS
PL SELF-LEVELLING PLANERS For planing asphalt and cement in pre-set depths. 4 Designed for removing the entire layer of asphalt or cement in preparation for trenching, or for milling deteriorated sections
More information1. Question Answer cgy / MU cgy / MU 2. Question Answer
GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom
More informationThermo K-Alpha XPS Standard Operating Procedure
Thermo K-Alpha XPS Standard Operating Procedure Quick Guide Draft v.0.1 Procedure overview 1. Vent the loadlock 2. Secure your sample to the stage using clips, check the height of the final assembly. 3.
More informationAbsorption measurements for a carbon fiber couch top and its modelling in a treatment planning system
Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli
More informationDosimetric Calculations. Lonny Trestrail
Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations
More informationCommissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale
Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the
More informationRADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER
RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER Proven Rad-Check technology specifically designed to provide you with the ultimate in versatility and cost-effective operation. Fast and easy
More informationTHE development of more advanced techniques in radiotherapy,
1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,
More informationExradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers
EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,
More informationATHLETICS Track & Field Equipment
ATHLETICS Track & Field Equipment Aluminium Curbing Order No. 11260 The aluminium curbing is made from special made anodized aluminium profiles. It is a mobile marker and separator of the running track.
More informationWe measure it. For gas and water installers. testo hpa. bar
Pressure measuring instruments For gas and water installers testo 312-2 hpa testo 312-3 testo 312-4 bar C testo 312-2 / testo 312-3 Pressure meters for gas and water fitters testo 312-2 testo 312-2, fine
More informationBASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1
BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 The front door of the main body of the instrument may be open when you arrive. Take the sample chamber and slide it into position with the
More informationEarthworks X200 Quick Reference Guide
Earthworks X200 Quick Reference Guide Part No. 875-0267-000 Rev. A1 Contents Dig a 1.0 m Deep Hole Using a Flat Plane 1 Extend a Level Excavation 3 Dig a 1.0 m Deep Hole Using a Flat Plane with a 5 Laser
More informationActivity P07: Acceleration of a Cart (Acceleration Sensor, Motion Sensor)
Activity P07: Acceleration of a Cart (Acceleration Sensor, Motion Sensor) Equipment Needed Qty Equipment Needed Qty Acceleration Sensor (CI-6558) 1 Dynamics Cart (inc. w/ Track) 1 Motion Sensor (CI-6742)
More informationANZAI AZ-733VI. Respiratory Gating System. since 1976
ANZAI since 1976 Respiratory Gating System AZ-733VI Specifications AZ-733VI Dimension Configuration Options Sensor Port W:260 D:230 H:126mm Sensor Port Laser Sensor & Fixing Arm Relay Box Wave Monitor
More informationConstancy checks of well-type ionization chambers with external-beam radiation units
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew
More informationSEM LEO 1550 MANUAL. I. Introduction. Generality
SEM LEO 1550 MANUAL RESERVATION POLICY: 2 booking slots maximum per day and per person (ie. 1h). 6 booking slots maximum per week and per person (ie. 3h). Reservation names must correspond to the operators.
More informationGEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY
The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF
More informationOD plug gauge. for bores and internal profiles from dia. 2,5 mm. OD plug gauge in open design. Special designs Accessories. OD plug gauge for various
OD plug gauge for various bore diameters Cross sleeve Cross sleeve with runners OD plug gauge for bores and internal profiles from dia. 2,5 300 Oskar Schwenk GmbH & Co. Esslinger Straße 84 70736 Fellbach
More informationIntroduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components
Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto
More informationAssembly Instructions and User Manual MP3-P Therapy Beam Analyzer L981403
Assembly Instructions and User Manual MP3-P Therapy Beam Analyzer L981403 Version 0 or higher Contents Operating Manual Technical Manual Service Manual D850.131.03/01 en 2016-05 Bru General Information
More informationNew Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications
19 th World Conference on Non-Destructive Testing 2016 New Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications Christoph HENKEL 1, Markus SPERL 1, Walter
More informationTG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009
TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and
More informationAN31E Application Note
Balancing Theory Aim of balancing How an unbalance evolves An unbalance exists when the principle mass axis of a rotating body, the so-called axis of inertia, does not coincide with the rotational axis.
More informationA high flying simplicity DUAL SOFT BAG INJECTION SYSTEM
A high flying simplicity DUAL SOFT BAG INJECTION SYSTEM THE INNOVATIVE SOLUTION for contrast media and saline injection in CT medical imaging Designed for MULTI-USE Dual soft bag injector for MONO & BI-LIQUID
More informationMath Review. Overview
Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY
More informationHigh Speed Direct SAD Radiosurgery Beam Scanner
Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife
More informationCALIBRATION PRODUCTS & PHANTOMS
CALIBATION PODUCTS & PHANTOMS WATE PHANTOM MOTO DIVE SYSTEM For safety the maximum power supply to the motor is 24 volts D.C. All components of the motor control box and motor drive unit are UL approved.
More informationSCHEINWORKS Measuring and Analysis Systems by
Pressure Measurement Systems for standing and walking analysis Germany since 1879 Pressure Measurement Systems for standing and walking analysis Documentation of Gait image Stance Symmetry of all parameters
More informationDate: December 2014 Staff: T Millington, A Hooper Equipment used: Leica AT401 Laser Tracker, 1.5 SMR, Spatial Analyzer software
SURVEY REPORT Project: MICE Hall Re-survey Date: December 2014 Staff: T Millington, A Hooper Equipment used: Leica AT401 Laser Tracker, 1.5 SMR, Spatial Analyzer software Reference Drawing Nos MICE-HALL-USMN-2012-06.xit
More informationGripping rotary modules
Gripping rotary modules Gripping rotary modules GRIPPING ROTARY MODULES Series Size Page Gripping rotary modules RP 314 RP 1212 318 RP 1216 322 RP 1520 326 RP 2120 330 RP 2128 334 RC 338 RC 1212 342 RC
More informationCover Page for Lab Report Group Portion. Flow Visualization in a Water Channel
Cover Page for Lab Report Group Portion Flow Visualization in a Water Channel Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 08 September 2017 Name 1: Name 2: Name 3: [Name
More informationMapping a Magnetic Field. Evaluation copy. Figure 1: Detecting the magnetic field around a bar magnet
Mapping a Magnetic Field Experiment 16 The region around a magnet where magnetic forces can be detected is called a magnetic field. All magnets, no matter what their shape, have two poles labeled north
More informationOperating manual for the pressure transducer unit CCS 1000
Operating manual for the pressure transducer unit CCS 1000 Please read the operating manual before using the device. 1 Content: 1. General description 2. Technical data 3. Operation 4. Scope of supply
More informationkvue IGRT Couch top One couch top, multiple treatment solutions For Varian, Elekta, Siemens, Proton and more The Best of Both Worlds
TM kvue IGRT Couch top One couch top, multiple treatment solutions For Varian, Elekta, Siemens, Proton and more kvue The Best of Both Worlds kvue IGRT COUCH TOP One couch top, multiple treatment solutions
More informationReview of fundamental photon dosimetry quantities
Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited
More informationPneumatic proportional controller Types M Types FM Function tested for use of the float in Ex-zone 0
Operating Manual LTIA5E Pneumatic proportional controller Types M Types FM Function tested for use of the float in Ex-zone 0 Contents 1. Safety Instructions 2. Conformity to standards 3. Technical data
More informationMPG-plus. Application example. Pneumatic 2-Finger Parallel Gripper Gripper for small components. Gripping force 38 N 175 N. Sizes
MPG-plus Sizes 25 40 Weight 0.06 kg 0.24 kg Gripping force 38 N 175 N Stroke per finger 3 mm 6 mm Workpiece weight 0.19 kg 0.7 kg Application example Pneumatically driven, dual-axis pick-and-place machine
More informationA beam-matching concept for medical linear accelerators
Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,
More informationMAINTANCE MANUAL CNC ROTARY TABLE MODEL: DMNC-5C YUASA INTERNATIONAL.
MAINTANCE MANUAL CNC ROTARY TABLE MODEL: DMNC-5C YUASA INTERNATIONAL. INDEX DESCRIPTION PAGE (1)SPECIFICATION 2 (2)OUTSIDE DIMENSION LAYOUT 3 (3)INSPECTION REPORT 4 (4)PREPARATION FOR OPERATION 5 4-1.LUBRICATING
More informationCommissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN
22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA
More informationOutline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface
Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities
More informationADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM. OPENING DATE & TIME: January 3, 2018 at 3:00PM
Procurement Services ADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM ITB NUMBER: 1708MC OPENING DATE & TIME: January 3, 2018 at 3:00PM ITB TITLE: X-Ray Photoelectron Spectroscopy System ADDENDUM
More informationATT-59, Part I. ATT-59/99, SMOOTHNESS OF PAVEMENTS Part I, Manual Profilograph. b) Determining the Profile Index from profilograms
ATT-59, Part I ATT-59/99, SMOOTHNESS OF PAVEMENTS Part I, Manual Profilograph 1.0 SCOPE This method describes the procedures for: a) Operating the profilograph. b) Determining the Profile Index from profilograms
More informationHydraulic hand spindle pump Models CPP1000-M, CPP1000-L
Calibration Hydraulic hand spindle pump Models CPP1000-M, CPP1000-L WIKA data sheet CT 91.05 for further approvals see page 2 Applications Calibration service companies and service industry Calibrations
More informationField size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
More informationHydraulic lift crane. Complies with ANSI/ASME B 30.5 LR enus LR
Hydraulic lift crane LR 1110 Complies with ANSI/ASME B 30.5 enus LR 1001.02 Concept and characteristics LR 1110 Standard Excellent lifting capacities thanks to optimized distribution of forces New cabin
More informationMonitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription
Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt
More informationHydro. Tab Marine Engineering. Interceptor Trim Tabs. Born to live in the sea. 2017ver.1
Hydro Tab Interceptor Trim Tabs Born to live in the sea 2017ver.1 Designed to operate... Developed under NMEA 2000 protocol Compatible with NMEA 018 Hydrotab is an innovative patented trim system designed
More informationProject Title: Pneumatic Exercise Machine
EEL 4924 Electrical Engineering Design (Senior Design) Preliminary Design Report 27 January 2011 Project Title: Pneumatic Exercise Machine Team Members: Name: Gino Tozzi Name: Seok Hyun (John) Yun Email:
More informationModel 130M Pneumatic Controller
Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate
More informationCover Page for Lab Report Group Portion. Boundary Layer Measurements
Cover Page for Lab Report Group Portion Boundary Layer Measurements Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 30 March 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section
More informationProduct Information. Gripper for small components MPG 20
Product Information MPG 20 MPG Precise. Compact. Reliable. MPG gripper for small components 2-finger parallel gripper with smooth roller guides of the base jaws Field of application Gripping and moving
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople
More informationNew product release. Universal Rebreather Monitor (URBM) Single O2 cell and Dual HP package
New product release Closed Circuit Research is pleased to announce the launch of our range of Universal Rebreather Monitors Key features and benefits include: Universal Rebreather Monitor (URBM) Single
More informationThe MC commissioning of CyberKnife with MLC (Tips and Tricks)
The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed
More informationSIGMA MASCOT. Product Hand Assembly made easy. PCB Hand Assembly made easy
We have been specialists in Hand Assembly systems for 25 years, with 1500 systems installed in: Australia Belgium China Czech Republic France India Ireland Malaysia New Zealand Russia South Africa Turkey
More informationOPERATION MANUAL DIVE CALIPER NOT SCALE / IF IN DOUBT ASK* *ALL UNITS IN MM/DEGREES UNLESS SPECIF. Dive Caliper General DC-051
NOT SCALE / IF IN DOUBT ASK* DIVE CALIPER OPERATION MANUAL *THIS INFORMATION IS THE PROPERTY OF ZEBRA-TECH LTD. ALL RIGHTS RESERVED.* *ALL UNITS IN MM/DEGREES UNLESS SPECIF SURFACE F 3.2 Ra UNLESS S Part
More informationPFH. Application example. Pneumatic 2-Finger Parallel Gripper Long-stroke Gripper. Weight 2.65 kg 12.6 kg. Gripping force 630 N 2950 N
PFH www.comoso.com Sizes 30 50 Weight 2.65 kg 12.6 kg Gripping force 630 N 2950 N Stroke per finger 30 mm 100 mm Workpiece weight 3.15 kg 13 kg Application example Assembly unit for intermediate sleeves
More information[needs to be discussed with faculty if any new material to be done]
Tool Name: EVO 18 Version 1 Authorized Users: Gayatri, Vishnu kant Materials Used: Sample holder has facility to mount 9 stubs: Si, Glass, Materials allowed : [needs to be discussed with faculty if any
More informationQuick Start Guide. A. Hardware installation B. Software installation C. Start the software for the first time D. Do your first measurement
Quick Start Guide This Quick Start Guide describes the hardware and software installation process and the Measurement feature in a simple way. Please follow the sequence of the steps to avoid problems
More informationCover Page for Lab Report Group Portion. Head Losses in Pipes
Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section
More information2990 Lorne Scots Royal Canadian Army Cadet Corps Rifle Drill
2990 Lorne Scots Royal Canadian Army Cadet Corps Rifle Drill Page 1 of 27 TABLE OF CONTENTS PARA ITEM PAGE Section 1 BASIC RIFLE DRILL 1.1 INTRODUCTION 3 1.2 ATTENTION 5 1.3 STAND AT EASE FROM ATTENTION
More informationTru Flight TRUFLIGHT INSTALLATION GUIDE TRUGOLF.COM
Tru Flight T R A C K I N G S Y S T E M TRUFLIGHT INSTALLATION GUIDE TRUGOLF.COM TruFlight Camera Tracking System Setup & Installation TruFlight : How It Works... 1 TruFlight Components... 1 TruFlight Installation...
More informationNEW: CO2 incubator with sterilizable sensor
CB series - NEW 160 C2 Incubators NEW: C2 incubator with sterilizable sensor The new CB series offers optimal growth conditions for cell cultures. The ANTI.PLENUM concept, the 180 C sterilization routine
More informationTrackMan Operator Training
TrackMan Operator Training The following setup and use information is the same for using TrackMan in the Indoor setting or Outdoor setting unless noted Extend the two bottom legs and single back leg before
More informationHumiSys HF High Flow RH Generator
HumiSys HF High Flow RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The HumiSys HF is a high flow version of the previously
More informationCLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions
CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning
More informationHeavy Duty Dissolved Oxygen Meter
User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 FAX 781.665.0780 - TestEquipmentDepot.com Introduction Congratulations
More informationSnapShot IMRT with compensators and FFF beams
SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy
More informationThe Coriolis Effect - Deflect the Arrows!
NAME: DATE: The Coriolis Effect - Deflect the Arrows Directions: The Circle below represents the Earth. The equator is present, dividing the image into the Northern and Southern hemispheres. The arrows
More informationLEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i
LEO SEM SOP Page 1 of 9 LEO 440 SEM SOP Gun (Filament) Column Manual Valves Chamber Window Chamber Stage Movement Leica Leo Stereoscan 440i 1. Scope 1.1 This document provides the procedure for operating
More informationIn Vessel Composter Installation & Commissioning Manual
In Vessel Composter Installation & Commissioning Manual Contents Customer responsibilities Transportation Lifting Delivery Pre-installation requirements Installation Drainage Ventilation Levelling Unpacking
More informationService Test Unit for ROSS Proportional Valves RESK
Service Unit for Proportional Valves Page 1 Technical Description ROSS 04.08.2015 Service Test Unit for ROSS Proportional Valves Technical Description V 1.0 / 2015 RESK 4591.40 For all Prop Valve Models
More informationScience&Motion. SAM BalanceLab. Sports. control the invisible. Most advanced pressure plate for coaching
SAM BalanceLab Science&Motion Sports control the invisible Most advanced pressure plate for coaching Balance - the invisible key to success Stop the guesswork and start to rely on facts Science and Motion
More informationTroubleshooting Guide Aquarius
Low Access Pressure Access pressure has dropped below the lower alarm limit of 0 to - 250mmHg Blood pump is off Blood flow is too low Turn blood pump on Check blood flow rate. Consider increasing the blood
More informationBRONZE BUSHING REPLACEMENT PROCEDURE DN345 & NL450C
1 BRONZE BUSHING REPLACEMENT PROCEDURE V.2 12/3/2014 DN345 & NL450C 2 Safety Instructions Removing Walking Beams 3 1. Position spreader on a flat concrete surface capable of supporting weight of spreader
More informationEXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY
18 th World Conference on Non destructive Testing, 16-20 April 2012, Durban, South Africa EXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY Arno VOLKER 1 and Hendrik VOS 1 TNO, Stieltjesweg 1,
More informationNEW: CO2 incubator with sterilizable sensor
CB series - NEW 160 C2 incubators NEW: C2 incubator with sterilizable sensor The new CB series offers optimal growth conditions for cell cultures. The ANTI.PLENUM concept, the 180 C sterilization routine
More informationRIFLE DRILL. 1. General. Before commencing instruction in rifle drill, it is important that the cadets receive instruction in the parts of the rifle.
RIFLE DRILL 1. General. Before commencing instruction in rifle drill, it is important that the cadets receive instruction in the parts of the rifle. 2. Rifle Movements. Unless stated otherwise, cadets
More informationThe University of Hong Kong Department of Physics Experimental Physics Laboratory
The University of Hong Kong Department of Physics Experimental Physics Laboratory PHYS2260 Heat and Waves 2260-1 LABORATORY MANUAL Experiment 1: Adiabatic Gas Law Part A. Ideal Gas Law Equipment Required:
More informationThe HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.
The HumiSys RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The new HumiSys with single or dual RH probes capabilities is
More informationOXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:
OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle
More informationTransmitter CS 21 Operation Manual
Transmitter CS 21 Operation Manual Content Page For your Safety 3 General Description 3 Detection Principle 4 Operational Notes 4 Design 4 Mounting Position of CS21 5 Mounting 6 Installation of Electrical
More informationPneumatic switch module 3/2-way valve (On/Off) Types P Types FP Function tested for use of the float in Ex-zone 0
Operating Manual LTIA3E Pneumatic switch module 3/2-way valve (On/Off) Types P Types FP Function tested for use of the float in Ex-zone 0 Contents 1. Safety Instructions 2. Conformity to standards 3. Technical
More informationAnchor Load Cells. Among the different types that are used as anchor load cells
Chapter: 3.1 Anchor Load Cells Page No.: 1 Number of Pages: 1 In rock and base engineering anchors are applied as construction element. They stabilise the rock by absorbing axial load and shear forces.
More informationUNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS International General Certificate of Secondary Education
UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS International General Certificate of Secondary Education *5757795995* PHYSICS 0625/33 Paper 3 Extended October/November 2012 1 hour 15 minutes Candidates
More informationDPZ-plus. Application example. Pneumatic 3-Finger Centric Grippers Sealed Grippers. Gripping force 520 N N. Sizes
DPZ-plus Sizes 64.. 200 Weight 0.62 kg.. 20.1 kg Gripping force 520 N.. 16800 N Stroke per finger 3 mm.. 25 mm Workpiece weight 2.6 kg.. 60.0 kg Application example Insertion tool for assembling small
More informationFIG: 27.1 Tool String
Bring up Radioactive Tracer service. Click Acquisition Box - Edit - Tool String Edit the tool string as necessary to reflect the tool string being run. This is important to insure proper offsets, filters,
More informationEmily Winslow, MD April 1, 2014 SSAT Career Development Grant Update. Summary
Summary Emily Winslow, MD April 1, 2014 SSAT Career Development Grant Update The primary aim of this grant was to develop and pilot the use of an MRI compatible box that would allow precise radiographic-pathologic
More information