Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

Size: px
Start display at page:

Download "Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting"

Transcription

1 Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose output reference dose output = calibration dose output correction for collimator setting correction for distance from source. correction for beam modifiers correction for attenuation and scatter Collimator setting Multiply calibration dose output by output factor (field size factor) Dose at d max with collimator set to specified value divided by dose at d max for calibration value If calibration done at another depth d, use ratio of doses at d 1

2 Collimator setting Factors affecting output If f calibration dose output measured in air only factor causing change in output factor is collimator scatter (S c ) If calibration dose output measured in phantom additional factor due to phantom scatter (S p ) Phantom scatter dependent on field size and not collimator setting Collimator setting Note difference between collimator setting and field size Collimator setting field size of unblocked field at source to isocenter distance Field size is actual (or effective) size of treatment field at specified distance from source Field size may be larger than or smaller than collimator setting Collimator setting If f field size significantly difference from collimator setting, phantom scatter can be accounted for by multiplying by ratio of peakscatter factors PSF PSF field coll. 2

3 Collimator setting For high-energy (>10 MV or so) linac beams, peakscatter factor essentially independent of field size, so value of 1.0 is often used without introducing serious error Collimator setting Khan approach Khan prefers to separate collimator scatter explicitly from phantom scatter and write the collimator setting correction as S c S p S c : collimator scatter factor S p : phantom scatter factor Summary MDACC approach Khan approach PSF ( ) ( fs) OF cs Sc ( cs) S p ( fs) PSF( cs) 3

4 Where to find data Data typically available in machine data book (or files) Data normally presented for square fields For rectangular fields, need to determine equivalent square The square field that has the same dosimetric properties as the rectangular field Purely an empirical quantity Equivalent square Table look up Good rule of thumb s = 4 a/p = 2 lw/(l+w) The two are equivalent for all but extremely elongated fields Calculating an output factor Determine equivalent square Look up value in table If not explicitly expressed, estimate interpolation, e.g., output factor for 9 9 should be half the difference between 8 8 and Express values to 3 decimal places, e.g.,

5 Distance from source Corrects for differences between distance of calibration point from source and distance of reference point from source Calibration point may be at SAD+d max, while reference point may be at SAD Distance from source Example: Calibrated dose output at SAD + d max from source and reference point at isocenter Distance correction factor is SAD + d SAD max 2. Distance from source Example: Calibrated dose output at isocenter and reference point at isocenter Distance correction factor is

6 Beam modifiers Accounts for effects of trays, wedges, compensators, anything else that may be placed in beam between radiation source and patient Beam modifiers Beam modifier factors frequently taken to be constant In some cases, field size and/or depth dependent Beam modifiers Care when incorporating wedge factors, especially if extracting information from treatment planning system For some systems, reference dose is dose to reference point for unwedged field reference dose output must be for unwedged field, and no wedge factor included 6

7 Beam modifier data Beam modifier data typically found in machine data book (or files) Data generally acquired as part of beam commissioning process Effect of patient Must account for attenuation of beam in patient Most common factor is tissue-maximum ratio (TMR) We will present several other factors that may be easier to understand than TMR first Effect of patient If calibration c dose output specified in air -- all calculations up to this point give dose rate to reference point in air To get dose rate in tissue, multiply by tissue- air ratio (TAR) 7

8 Effect of patient If calibration c dose output specified in tissue at depth all calculations up to this point give dose rate to reference point at depth To get dose rate in tissue, multiply by either tissue-phantom ratio (TPR), or, if specified depth is d max, tissue-maximum ratio (TMR) Let Tissue-air ratio (TAR) D n dose delivered to small mass of tissue at some depth in water phantom D air dose delivered to same mass of tissue in air at same distance from the source, then D TAR = D n air Tissue-air ratio (TAR) D TAR = D n air 8

9 Tissue-air ratio (TAR) Example: Calculate treatment time required to deliver 150 cgy to a point at 6 cm depth using 8 cm 8 cm 60 Co beam with in-air dose rate of 135 cgy/min at 80 cm SAD. Tissue-air ratio (TAR) Dose rate in air 135 cgy/min TAR for 8 cm 8 cm 60 Co beam at depth of 6 cm found from TAR table to be Dose rate in tissue: 135 cgy/min = cgy/min Tissue-air ratio (TAR) To o deliver dose of 150 cgy requires treatment time of 150 cgy = 1.31min cgy/min 9

10 Tissue-air ratio (TAR) TAR relates dose to a point to dose in air at the same point Used when patient set up isocentrically Used when calibration dose output expressed in air Typically only for low-energy photon beams TAR properties Even though TAR is no longer frequently used, we will examine some properties of the TAR, as analogous behavior is exhibited by the more commonly used TMR TAR properties Effect of depth: TAR decreases with increasing depth beam attenuation TAR evaluated at depth of central axis maximum (d max ) backscatter factor (BSF) For depths <d max, TAR increases with depth 10

11 Effect of beam quality: At low energies BSF increases with increasing energy increased penetration of scattered photons TAR properties Effect of beam quality: At high energies BSF decreases with increasing energy greater amount of forward scatter TAR properties Effect of beam quality: BSF reaches maximum value around HVL mm Cu where it may be as high as 1.5 TAR properties 11

12 TAR properties Effect of field size: TAR increases with increasing field size increased amount of scatter reaching calculation point TAR properties Effect of source distance: TAR independent of source distance both D n and D air measured at same distance from source TAR data Find TAR tables: For 60 Co: BJR, supplement 25 For specific linacs: in physics data notebook Important to use correct table for appropriate machine 12

13 Tissue-air ratio (TAR) In n order to calculate dose rate at distance other than the SAD use inverse square relationship to go from one source distance to another Important to note that inverse square relationship only holds in air Tissue-air ratio (TAR) To relate dose rate in tissue at one distance to that at another: Calculate dose rate in air by dividing by appropriate TAR Use inverse square law to go from one distance to the other Calculate dose rate in tissue at new distance by multiplying in-air dose rate by appropriate TAR 13

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt

More information

Review of fundamental photon dosimetry quantities

Review of fundamental photon dosimetry quantities Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited

More information

Dosimetric Calculations. Lonny Trestrail

Dosimetric Calculations. Lonny Trestrail Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations

More information

1. Question Answer cgy / MU cgy / MU 2. Question Answer

1. Question Answer cgy / MU cgy / MU 2. Question Answer GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom

More information

Math Review. Overview

Math Review. Overview Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY

More information

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

More information

Relative Dosimetry. Photons

Relative Dosimetry. Photons Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge

More information

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results

More information

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the

More information

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning

More information

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

The MC commissioning of CyberKnife with MLC (Tips and Tricks) The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed

More information

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11 MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction

More information

Monitor Unit Verification for Small Fields

Monitor Unit Verification for Small Fields Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the

More information

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics. FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic

More information

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical

More information

Clinical Implementation of Volumetric Modulated Arc Therapy

Clinical Implementation of Volumetric Modulated Arc Therapy Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy

More information

3D Treatment Planning and verification with hand calculations

3D Treatment Planning and verification with hand calculations 3D Treatment Planning and verification with hand calculations Randy Holt, PhD, CEO Pacific Crest Medical Physics, Inc Chico, CA Dosimetric Calculations Calibration and Measurement Conditions Patient Conditions

More information

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN 22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA

More information

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto

More information

High Speed Direct SAD Radiosurgery Beam Scanner

High Speed Direct SAD Radiosurgery Beam Scanner Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife

More information

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB

More information

Constancy checks of well-type ionization chambers with external-beam radiation units

Constancy checks of well-type ionization chambers with external-beam radiation units JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew

More information

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS

More information

Planning Assignment (3 field rectum)

Planning Assignment (3 field rectum) Megan Sullivan Clinical Lab Rectum April 25, 2016 Planning Assignment (3 field rectum) Use a CT dataset of the pelvis. Create a CTV by contouring the rectum (start at the anus and stop at the turn where

More information

Scientific Journal of Nuclear Medicine & Radiation Therapy

Scientific Journal of Nuclear Medicine & Radiation Therapy Scientific Journal of Nuclear Medicine & Radiation Therapy Research Article Radiotherapy Dose Calculation Program, for Monitoring and Commissioning Treatment Planning Systems - Ioannis Vamvakas*, Stella

More information

Zoubir Ouhib Lynn Cancer Institute

Zoubir Ouhib Lynn Cancer Institute Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy

More information

CHAPTER-2. Wedge filters are the most important beam modifiers used in

CHAPTER-2. Wedge filters are the most important beam modifiers used in CHAPTER-2 2. WEDGE FLTERS 2.1. NTRODUCTON Wedge filters are the most important beam modifiers used in radiotherapy which are placed in the path of photon beams to modify their isodose diistributions. The

More information

SnapShot IMRT with compensators and FFF beams

SnapShot IMRT with compensators and FFF beams SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy

More information

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08. Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51

More information

OUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD

OUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD RAPHEX 2001 T1. All of the following are true regarding Percentage Depth Dose (PDD) except: A. Increases with increasing energy. B. Depends on field size. C. Is the dose at depth expressed as a % of the

More information

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and

More information

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the

More information

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61: Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca

More information

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE

More information

IMRT in clinical practice at the UMC-Utrecht. Utrecht

IMRT in clinical practice at the UMC-Utrecht. Utrecht 3 IMRT in clinical practice at the UMC-Utrecht Utrecht Clinical use of IMRT Improvement of local control without increased incidence of normal tissue complications + Dose escalation in the prostate Step-and

More information

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/

More information

Radiotherapy physics & Equipments

Radiotherapy physics & Equipments Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education

More information

Introduction. Materials and Methods

Introduction. Materials and Methods STANDARD WEDGE AND TRAY TRANSMISSION VALUES FOR VARIAN, SEIMENS, ELEKTA/PHILIPS ACCELERATORS; A QUALITY ASSURANCE TOOL David S. Followill, Nadia Hernandez, and William F. Hanson Introduction The Radiological

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Verification of Monitor Unit Calculations for Eclipse Treatment Planning System by in-house Developed Spreadsheet Athiyaman Mayivaganan 1*, Hemalatha

More information

Commissioning and periodic tests of the Esteya electronic brachytherapy system

Commissioning and periodic tests of the Esteya electronic brachytherapy system Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,

More information

A beam-matching concept for medical linear accelerators

A beam-matching concept for medical linear accelerators Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,

More information

THE development of more advanced techniques in radiotherapy,

THE development of more advanced techniques in radiotherapy, 1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,

More information

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007 The Effect of Immobilisation Devices on Radiotherapy Dose Distributions A thesis submitted in fulfilment of the requirements for the Degree of Master of Applied Science (Medical and Health Physics) By

More information

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS Acta Radiologica Oncology 23 (1984) Fusc. 6 FROM THE RADIATION THERAPY DEPARTMENT, ANTON1 VAN LEEUWENHOEK HOSPITAL, THE NETHERLANDS CANCER INSTITUTE, AMSTERDAM, THE PHILIPS MEDICAL SYSTEMS DIVISION, BEST,

More information

Investigation of Thermal Effects of CO 2 on Earth-Atmosphere System

Investigation of Thermal Effects of CO 2 on Earth-Atmosphere System Investigation of Thermal Effects of CO 2 on Earth-Atmosphere System Design Team Michel Beguin, Tim Bevins Dan Jakiela, Frank Kuchinski Design Advisor Prof. Yiannis Levendis Abstract This Capstone Project

More information

TG-51: Experience from 150 institutions, common errors, and helpful hints

TG-51: Experience from 150 institutions, common errors, and helpful hints TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of

More information

Abstract: Introduction:

Abstract: Introduction: Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston

More information

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Received: 21 November 2016 Revised: 7 April 2017 Accepted: 5 June 2017 DOI: 10.1002/acm2.12160 RADIATION ONCOLOGY PHYSICS Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Jeffrey

More information

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Radiotherapy and Oncology 71 (2004) 235 241 Technical Note A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Nadine Linthout*,

More information

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions Wendel Dean Renner a) Math Resolutions, LLC, 5975 Gales Lane, Columbia,

More information

Ultrasound Dose Calculations

Ultrasound Dose Calculations Ultrasound Dose Calculations One of the advantages of ultrasound therapy remains the reasonably broad range of trials from which effective treatment doses can be established. In principle, there is no

More information

How to Design Medical Accelerator Systems for Reliability: IBA PT System

How to Design Medical Accelerator Systems for Reliability: IBA PT System How to Design Medical Accelerator Systems for Reliability: IBA PT System Yves Jongen Founder & Chief Research Officer Ion Beam Applications s.a. Belgium 1 Outline A short introduction of a proton therapy

More information

Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation

Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation By Ruth Ann M. Good, CMD. R.T. (R)(T) Virginia Commonwealth University, Richmond,

More information

3 1 PRESSURE. This is illustrated in Fig. 3 3.

3 1 PRESSURE. This is illustrated in Fig. 3 3. P = 3 psi 66 FLUID MECHANICS 150 pounds A feet = 50 in P = 6 psi P = s W 150 lbf n = = 50 in = 3 psi A feet FIGURE 3 1 The normal stress (or pressure ) on the feet of a chubby person is much greater than

More information

The use of the analytical balance, and the buret.

The use of the analytical balance, and the buret. 1211L Experiment 1. Density 2015 by H. Patterson Instructor Notes: Students make measurements individually then share data to make the graph. There are four volumetric measurements to be studied; 3.00

More information

TG219: IT'S USE, STRENGTHS AND WEAKNESSES

TG219: IT'S USE, STRENGTHS AND WEAKNESSES 1 CONFLICT OF INTEREST TG219: IT'S USE, STRENGTHS AND WEAKNESSES SOTIRI STATHAKIS, PHD, DABR None ROSTER STATEMENT OF THE PROBLEM Timothy C. Zhu, University of Pennsylvania, Philadelphia, PA (Co- Chair)

More information

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center Today s objectives Review the

More information

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation This intercomparison will be carried out under the lead of KRISS with the cooperation of ARPANSA. This comparison has been approved

More information

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2017, 6, 57-72 http://www.scirp.org/journal/ijmpcero ISSN Online: 2168-5444 ISSN Print: 2168-5436 Validation of Treatment

More information

MATH GRADE 6 UNIT 6 RATE ANSWERS FOR EXERCISES

MATH GRADE 6 UNIT 6 RATE ANSWERS FOR EXERCISES MATH GRADE 6 UNIT 6 RATE FOR EXERCISES LESSON 2: PRICE AS A RATE 1. $6.25 2. $.625, or $.63 3. $5.25 4. $.3125, or $.31 5. a. $2.5 b. $13.75 6. a. Amount (pt) 1 2 3 4 5 6 Cost non-organic ($) $.75 $1.5

More information

INTRODUCTION Porosity, permeability, and pore size distribution are three closely related concepts important to filter design and filter performance.

INTRODUCTION Porosity, permeability, and pore size distribution are three closely related concepts important to filter design and filter performance. Measurement of Filter Porosity using a Custom-Made Pycnometer George Chase The University of Akron INTRODUCTION Porosity, permeability, and pore size distribution are three closely related concepts important

More information

Underwater measurement of photosynthetically active radiation

Underwater measurement of photosynthetically active radiation Underwater measurement of photosynthetically active radiation Mark Blonquist, Damon Nitzel, and Bruce Bugbee Apogee Instruments, Inc., Logan, Utah Introduction Quantum sensors (photosynthetically active

More information

Exercise 2-3. Flow Rate and Velocity EXERCISE OBJECTIVE C C C

Exercise 2-3. Flow Rate and Velocity EXERCISE OBJECTIVE C C C Exercise 2-3 EXERCISE OBJECTIVE C C C To describe the operation of a flow control valve; To establish the relationship between flow rate and velocity; To operate meter-in, meter-out, and bypass flow control

More information

Commissioning of a new total body irradiation protocol

Commissioning of a new total body irradiation protocol University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2012 Commissioning of a new total body irradiation protocol Zoe Baldwin University

More information

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media N Tyagi 1, B H Curran, P L Roberson, J M Moran, E Acosta and B A Fraass Department

More information

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex.

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex. Beamex Calibration White Paper info@beamex.com Calculating Total Uncertainty of Temperature Calibration with a Dry Block World-class calibration solutions Calculating Total Uncertainty of Temperature Calibration

More information

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy University of South Florida Scholar Commons Graduate Theses and Dissertations Graduate School 7-14-2010 Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

More information

General Accreditation Guidance. User checks and maintenance of laboratory balances

General Accreditation Guidance. User checks and maintenance of laboratory balances General Accreditation Guidance User checks and maintenance of laboratory balances January 2018 Copyright National Association of Testing Authorities, Australia 2010 All intellectual property rights in

More information

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF

More information

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,

More information

The dose distribution of medium energy electron boosts to the laryngectomy stoma

The dose distribution of medium energy electron boosts to the laryngectomy stoma JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 1, WINTER 2001 The dose distribution of medium energy electron boosts to the laryngectomy stoma Ellen D. Yorke* Memorial Sloan-Kettering Cancer

More information

INSTRUMENT INSTRUMENTAL ERROR (of full scale) INSTRUMENTAL RESOLUTION. Tutorial simulation. Tutorial simulation

INSTRUMENT INSTRUMENTAL ERROR (of full scale) INSTRUMENTAL RESOLUTION. Tutorial simulation. Tutorial simulation Lab 1 Standing Waves on a String Learning Goals: To distinguish between traveling and standing waves To recognize how the wavelength of a standing wave is measured To recognize the necessary conditions

More information

1. BEAM MODIFIERS IN RADIOTHERAPY

1. BEAM MODIFIERS IN RADIOTHERAPY CHAPTER-1 1. BEAM MODIFIERS IN RADIOTHERAPY INTRODUCTION In radiotherapy, tumours under varying body contour are very often encountered. In such cases, due to the irregular body surface, the dose distribution

More information

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Comparison of ionization chambers of various volumes for IMRT absolute dose verification Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical

More information

Outlet Structures T-12

Outlet Structures T-12 Description This section provides guidance and details for outlet structures for use primarily with BMPs utilizing sedimentation, (i.e., extended detention basins (EDBs), retention ponds, and constructed

More information

WMO LABORATORY INTERCOMPARISON OF RAIN INTENSITY GAUGES

WMO LABORATORY INTERCOMPARISON OF RAIN INTENSITY GAUGES WMO LABORATORY INTERCOMPARISON OF RAIN INTENSITY GAUGES Christophe ALEXANDROPOULOS and Muriel LACOMBE Météo-France, Direction des Systèmes d Observation, BP 202-78195 Trappes France christophe.alexandropoulos@meteo.fr

More information

Characterizers for control loops

Characterizers for control loops Characterizers for control loops By: F. G. Shinskey (May 1999) Introduction Commercial controllers such as the PID series (proportional, integral, derivative, and their combinations) are linear devices

More information

Ratio & Rate Reasoning PRESENTED BY MR. LAWS 6 TH GRADE MATH

Ratio & Rate Reasoning PRESENTED BY MR. LAWS 6 TH GRADE MATH Ratio & Rate Reasoning PRESENTED BY MR. LAWS 6 TH GRADE MATH JCMS Common Core State Standard (CCSS) 6.RP.3 -Use ratio and rate reasoning to solve real-world and mathematical problems, e.g., by reasoning

More information

Development of Graphical User Interfaces (GUI) software and database for radiation therapy applications

Development of Graphical User Interfaces (GUI) software and database for radiation therapy applications The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2005 Development of Graphical User Interfaces (GUI) software and database for radiation therapy applications

More information

THE BEHAVIOR OF GASES

THE BEHAVIOR OF GASES 14 THE BEHAVIOR OF GASES SECTION 14.1 PROPERTIES OF GASES (pages 413 417) This section uses kinetic theory to explain the properties of gases. This section also explains how gas pressure is affected by

More information

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,

More information

Surface buildup dose dependence on photon field delivery technique for IMRT

Surface buildup dose dependence on photon field delivery technique for IMRT Surface buildup dose dependence on photon field delivery technique for IMRT Shigeru Yokoyama, Peter L. Roberson, Dale W. Litzenberg, Jean M. Moran, and Benedick A. Fraass Department of Radiation Oncology,

More information

Preparation for Salinity Control ME 121

Preparation for Salinity Control ME 121 Preparation for Salinity Control ME 121 This document describes a set of measurements and analyses that will help you to write an Arduino program to control the salinity of water in your fish tank. The

More information

Exposure System Selection

Exposure System Selection Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others

More information

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli

More information

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS AP PHYSICS

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS AP PHYSICS DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS AP PHYSICS LSN 11-7: WAVE MOTION LSN 11-8: TYPES OF WAVES; LONGITUDINAL AND TRANSVERSE LSN 11-9: ENERGY TRANSPORTED BY WAVES Physics of Waves Questions From Reading

More information

APPROVED FACILITY SCHOOLS CURRICULUM DOCUMENT SUBJECT: Mathematics GRADE: 6. TIMELINE: Quarter 1. Student Friendly Learning Objective

APPROVED FACILITY SCHOOLS CURRICULUM DOCUMENT SUBJECT: Mathematics GRADE: 6. TIMELINE: Quarter 1. Student Friendly Learning Objective TIMELINE: Quarter 1 i-ready lesson: Rational Numbers and Absolute Value i-ready lesson: Numerical Expressions and Order of Operations 6/16/15 1 i-ready lesson (2a, 2b and 2c): Algebraic Expressions 6/16/15

More information

A Review of the Bed Roughness Variable in MIKE 21 FLOW MODEL FM, Hydrodynamic (HD) and Sediment Transport (ST) modules

A Review of the Bed Roughness Variable in MIKE 21 FLOW MODEL FM, Hydrodynamic (HD) and Sediment Transport (ST) modules A Review of the Bed Roughness Variable in MIKE 1 FLOW MODEL FM, Hydrodynamic (HD) and Sediment Transport (ST) modules by David Lambkin, University of Southampton, UK 1 Bed roughness is considered a primary

More information

IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy

IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy Guidelines on standardized procedures at Secondary Standards Dosimetry Laboratories (SSDLs) and hospitals March 2002 The

More information

B. Radium Isotope Mass Balance Model Used to Derive Submarine Groundwater Discharge

B. Radium Isotope Mass Balance Model Used to Derive Submarine Groundwater Discharge Supplementary Materials Lingering radioactivity at the Bikini and Enewetak Atolls Ken Buesseler a *, Matthew A. Charette a, Steven Pike a, Paul Henderson a, Lauren Kipp a a Department of Marine Chemistry

More information

Chem 110 General Principles of Chemistry

Chem 110 General Principles of Chemistry CHEM110 Worksheet - Gases Chem 110 General Principles of Chemistry Chapter 9 Gases (pages 337-373) In this chapter we - first contrast gases with liquids and solids and then discuss gas pressure. - review

More information

Unit 2 Kinetic Theory, Heat, and Thermodynamics: 2.A.1 Problems Temperature and Heat Sections of your book.

Unit 2 Kinetic Theory, Heat, and Thermodynamics: 2.A.1 Problems Temperature and Heat Sections of your book. Unit 2 Kinetic Theory, Heat, and Thermodynamics: 2.A.1 Problems Temperature and Heat Sections 10.1 10.2 of your book. Convert the following to Celsius and Kelvin temperatures: 1. 80.0 o F Early E. C.:

More information

Student Manual. Lesson 4- Nuclear Density Testing. Version 1.0-3/09 4-1

Student Manual. Lesson 4- Nuclear Density Testing. Version 1.0-3/09 4-1 Version 1.0-3/09 4-1 Version 1.0-3/09 4-2 Version 1.0-3/09 4-3 What is the Density we are measuring? Can the Density be more than 100% of Maximum Dry Density reported in the Proctor test? Version 1.0-3/09

More information

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER Proven Rad-Check technology specifically designed to provide you with the ultimate in versatility and cost-effective operation. Fast and easy

More information

COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC

COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC by TAN ZI XIANG Thesis submitted in fulfillment of the requirements for the degree of Master of Science (Medical Physics) February 2015 ACKNOWLEDGEMENT

More information

Performance of the Liquid Hydrogen Target

Performance of the Liquid Hydrogen Target Performance of the Liquid Hydrogen Target Presented by Chad Gillis Indiana University at NPDGamma Collaboration Meeting July 2012, Oak Ridge, Tennessee Since upgrading the target for operation at the SNS,

More information

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

RADIATION PROCEDURES MANUAL Procedure Cover Sheet RADIATION PROCEDURES MANUAL Procedure Cover Sheet Procedure Title: Calibration of Eberline Portable Particulate Noble Gas Monitor Procedure Number: TSO-08-13-REV 1 Effective Date: July 1, 2008 Approved

More information

PX4 P L A S T I C PX4 PERFORMANCE WIL T-02

PX4 P L A S T I C PX4 PERFORMANCE WIL T-02 PX4 P L A S T I C PX4 PERFORMANCE WIL-10161-T-02 Section 5B Pro-Flo X TM Operating Principal The Pro-Flo X air distribution system with the revolutionary Efficiency Management System (EMS) offers flexibility

More information

Ozobot Bit Classroom Application: Boyle s Law Simulation

Ozobot Bit Classroom Application: Boyle s Law Simulation OZO AP P EAM TR T S BO RO VE D Ozobot Bit Classroom Application: Boyle s Law Simulation Created by Richard Born Associate Professor Emeritus Northern Illinois University richb@rborn.org Topics Chemistry,

More information