The Challenges Of Maintaining And Improving The Uncertainty Of An Industrial Humidity Calibration Laboratory

Size: px
Start display at page:

Download "The Challenges Of Maintaining And Improving The Uncertainty Of An Industrial Humidity Calibration Laboratory"

Transcription

1 The Challenges Of Maintaining And Improving The Uncertainty Of An Industrial Humidity Calibration Laboratory Speaker Jack Herring, Michell Instruments, Inc., 11 Old Sugar Hollow Road, Danbury CT 06810, USA phone (203) , fax (203) , Author Andrew M. V. Stokes Michell Instruments Limited, Lancaster Way, Ely, CB6 3NW, UK, phone , fax , Abstract The paper describes the issues associated with maintaining a busy industrial humidity calibration laboratory and the challenges faced in trying to maintain the laboratory measurement uncertainty whilst achieving a throughput of thousands of sensors per year. The humidity calibration facility of Michell instruments is essentially split into two parts a UKAS (EA) accredited laboratory for high level measurements with direct traceability and audit path to NPL and NIST standards, and a commercial laboratory providing lower level tertiary calibration of tens of thousands of dewpoint sensors per annum. The UKAS section focuses on excellence rather than on volume hence the processes and procedures are largely manual and quite time-consuming, whilst the commercial section is there to handle large volumes of sensors automatically and with the minimum of human intervention. The two therefore require very different approaches in terms of equipment, procedures and analysis and the uncertainty levels achieved reflect the type of operating model used in each case. Humidity calibration systems used by secondary laboratories have tended to be constructed either as clones of National Standards, using a two-pressure or two-temperature generation method, or as simple divided flow systems utilising calibrated vertical tube flow meters. The former are very expensive to produce and have certain limitations in terms of usability. The latter are cheaper to produce, but also suffer from inflexibility and difficulty in automation. Furthermore, these systems tend to offer varying flow rates dependent on the generated humidity level. The humidity calibration system described in this paper provides accurate and highly repeatable humidity generation using a combination of liquid and gas mass flow controllers. It allows automated use through the integration of a precision chilled mirror dew-point hygrometer that provides both the control feedback to the generator and traceability to National Humidity Standards. The paper describes the two processes, provides a practical consideration of the component uncertainties and explores ways in which these uncertainties can be refined and minimised through improved procedures, better equipment and careful operation. Also provided are detailed calculations of the liquid and gas mass flow ratios used to derive appropriate humidity levels in the measurement chamber. A novel technique to ensure sensitivity and stability of the generated humidity is described, along with the techniques employed to ensure homogeneity of the humidified air. The paper also describes the physical design and construction challenges that were overcome in producing a fully integrated system. An uncertainty budget for the whole system is provided, indicating the key contributory factors and suggesting ways in which the measurement uncertainty can be minimised.

2 1 Background Calibration is the comparison of a measured value with the true value. It has a formal definition under ISO standards: Calibration: the set of operations which establish, under specified conditions, the relationship between values indicated by a measuring instrument.. And the corresponding values of a quantity realised by a reference standard Adjustment is the setting of a minimum deviation from measured value and true value. Again, according to ISO: Adjustment: the operation intended to bring a measuring instrument into a state of performance and freedom from bias suitable for its use National Standard: A standard recognised by an official national decision to serve, in a country, as the basis for fixing the value of all other standards of the quantity concerned. The national standard in a country is often a primary standard. Details of several prominent national standards can be seen in figure 1, below. Figure 1. Prominent international standards laboratories. Traceability: The property of the result of a measurement standard, generally international or national standard, by an uninterrupted chain of comparisons All definitions taken from ISO : (1992)

3 Regular calibration of humidity sensors is important not only to maintain traceability, but also to ensure the correct operation of the process concerned. Most humidity instruments need calibration at least every year to ensure correct performance - some need calibration every few months. Of course, calibration can be taken to mean either verification or adjustment, followed by certification Figure 2. Calibration traceability hierarchy. The term uncertainty is defined as the tolerance band in which (at k=2) a measured value will be in agreement with the ACTUAL value for 95 % of the measurements made. It is effectively a confidence band for the measurement and can be calculated by combining various experimental and theoretical data on the system and instruments in use, according to well established formulae. Figure 3. Normal distribution chart. The uncertainties of the instrument in use and the calibration reference are both important. Whereas a traceable laboratory typically has an uncertainty of the order of ±0.2 C dew point (k=2 coverage) a typical field instrument may have an uncertainty ten times worse. The calibration of a hygrometer will result in two components - an off-set (often termed its accuracy) at a given point and an uncertainty value. So for example, if it has an off-set of +0.2 C dew point and the calibration uncertainty is ±0.1 C dew point, then you have 95 % confidence that the actual value = (reading -0.2) ±0.1. Michell Instruments Limited has been manufacturing and calibrating chilled mirror and capacitive (impedance type) dew-point hygrometers for more than thirty years. During most of that time, the basic method of calibration has been consistent. This process consists four component parts:

4 o Preparation of a supply of clean, dry compressed air with a moisture content below the minimum measurable range of the instrument or sensor to be calibrated o Generation of a range of dew-point temperatures across the operating range of the instrument or sensor to be tested, using a single- or multi-stage divided flow method o Provision of a suitable test chamber to house the instruments or sensors under test o Verification of the whole system performance on a continuous or sampling basis using a traceable reference hygrometer In the early years the performance of these calibration systems was limited by a number of factors, for example the quality and stability of the dry air source, the sensitivity limitations of the mixing system and the uncertainty of the reference measurement instrumentation. In more recent times individual components of the calibration system have been optimised to produce a better overall uncertainty for the particular calibration process, or product type, for which they have been designed. Figure 4. Typical calibration system overview. Although based on the same physical principles, the often disparate needs of the various types of calibration dictate very different forms. This is particularly true in the case of a laboratory set up for UKAS calibrations and also set up for volume calibration of mass produced hygrometers. The different challenges and their solutions, as implemented by Michell Instruments, are discussed in the remainder of this paper. 2 UKAS Laboratory: Quality Over Quantity Figure 5. UKAS calibration hierarchy. Referring back to our definitions, traceability is the property of the result of a measurement standard, generally international or national standard, by an uninterrupted chain of comparisons. Clearly, the longer the chain the more uncertainty is introduced. The more uncertainty there is in a measurement the greater the potential error.

5 Therefore it follows that the aim of a secondary laboratory, such as one operating under the UKAS accreditation scheme, is to provide calibrations with as short a chain of comparisons back to the appropriate national standards as can be practically achieved. In the case of Michell Instruments UKAS laboratory this means that the comparisons are performed directly against our in-house reference standard chilled-mirror hygrometers. These in turn are periodically calibrated against the primary national standards in the UK (NPL gravimetric hygrometer) and USA (NIST). Michell first obtained UKAS (then NAMAS) accreditation in 1987 and have since developed and refined our SOP s and established a large database of calibration history on the various reference standard hygrometers, some of which have calibration history right back to the mid 1980 s. The UKAS part of our facility now has XX reference standards, Y traceable to NPL and Z to NIST. The laboratory has increased capacity from its original one to one comparison to the level now where up to five working standards or customer instruments can be compared to the reference standard simultaneously. The process is however still largely a manual one and is labour intensive, and therefore expensive. Figure 6. Michell Instruments UKAS accreditation. The performance limitations of a calibration methodology are determined by a combination of factors including: Physical facilities of the laboratory used temperature control air quality stability of electrical power supply Quality of staff training

6 attention to detail repeatability variations in staff Calibration equipment Accuracy Repeatability Traceability In any laboratory these factors can be quantified statistically and combined to give a measure of that laboratory s capability as an uncertainty budget. The methods for calculating an uncertainty budget are well documented and beyond inclusion in a general paper of this type. For illustration, the uncertainty budget at -90 C dp is comprised of the following: Components of Uncertainty Calculation at -90 C dp Parameter Divisor Uncertainty Standardised Calibration of standard Drift of standard Repeatability Bridge calibration Bridge repeatability Bridge drift Contamination Temp. gradients in condensate Temp. gradients in mirror Pressure difference Mirror temperature fluctuations Non linearity Sampling uncertainties Standard Uncertainty K= Figure 6. Uncertainty budget calculation for -90 C dp. The overall uncertainty of the UKAS section of the dew point calibration laboratory is shown most clearly in the chart 1. On the following page.

7 Chart 1. UKAS measurement uncertainty. 3 Tertiary Calibrations: Achieving High Volume While Maintaining Quality Figure 7. Tertiary calibration hierarchy Michell Instruments production output exceeds 30,000 hygrometer sales per annum, each of which requires a traceable calibration. Given the limitations of the UKAS equipment previously mentioned, this requires a different approach to be taken. What is required here are calibration systems that can deal with multiple IUT s simultaneously with minimal additional uncertainty and, preferably, minimal operator intervention. The issues associated with the design, validation and calculation of uncertainty for such a system are detailed below. A primary enabling step that allowed Michell Instruments to develop a new calibration technique was their development of a range of sensors and instruments for which the calibration process is much easier to automate. In the past low-cost sensors tended to have used simple analogue trimming circuits to enable zero, span and linearity to be adjusted. More modern transmitters and instruments have benefited from digital communications and significant on-board memory capacity that in turn facilitate automated calibration processes.

8 A further factor in the need to improve reliability and automation of the calibration process is the response speed of sensors and systems to changes in applied humidity. At the lower humidity limit of the early systems (typically -75 C frost point, equivalent to 1 part per million) the response time for a system to a dry-down from ambient conditions might have typically been 24 to 48 hours. Current demands for instrumentation calibrated down to C frost point or below (13 parts per billion) demands much longer dry down times due to molecular absorption, even on carefully prepared and optimised electro-polished stainless steel surfaces. It is common for 120 hour (or longer) dry down periods to be used in order to ensure full system equilibrium prior to a calibration. These elongated calibration periods increase the need for system reliability, not only to ensure proper calibration, but also to ensure satisfactory throughput of product to meet customer demand. Typical stabilisation times for a calibration system such as that described in this paper are shown below. Dewpoint Temperature Minimum stabilisation time days hours hours hours hours hours hour hour hour hour 0 1 hour hour hour Figure 8. Calibration system stabilisation times For tertiary calibrations the objective was to produce a calibration system with the following key criteria: overall uncertainty better than ±2 C frost point at the low end and better than ±1 C frost/dew point at the high end; greater set-point stability; fully automated; increased capacity; self diagnostic; automatic sensor verification and self-adjusting. 3.1 Available Technologies: Pro s & Con s Dry gas supply Two alternatives were considered dry compressed air and liquid nitrogen. Ultimately it was decided that dry compressed air would be used, primarily on the grounds of safety, ease of implementation and familiarity. Michell has always used dried compressed air and found the method to be extremely reliable provided a sensible preventative maintenance programme is in place for compressors and dryers. A two stage drying process of oil-free compressed air at 7 barg is used to give a dry air feed to the generator of around 13 ppb moisture content, approximately -100 C frost point when expanded to atmospheric pressure. In theory a nitrogen source, dried through a nitrogen cold trap, can generate much lower moisture levels, but this was regarded as an un-necessary additional complication to the project. Practical experience has shown that the dry air system, now implemented on eight calibration systems, has resulted in zero down-time during more than two years of continuous operation.

9 3.1.2 Humidity Generator A feasibility study was undertaken to consider three options for humidity generation: twotemperature generator, two pressure generator and mass flow mixing system. These three technologies are amongst the most widely implemented within National Standards laboratories and commercial calibration laboratories worldwide. A mass flow mixing system was chosen because of the ability to produce a system with flow rate flexibility, easy control and relatively low cost in a short time scale. The humidity generator is the main subject of this paper and will be described in more detail in the following sections of this paper Test Chamber and Sampling System Considerable applications experience and expertise were utilised in designing a sensor calibration manifold and sampling system to ensure maximum integrity of the generated humidity at the test point. The design of this part of the system is described in more detail further on in the paper. 3.2 Calibration Integrity and Traceability Although not strictly necessary, based on the nature of the humidity generation method, it was decided to use a chilled mirror dew-point hygrometer as an on-line reference and as the control element of the closed loop generation system. This hygrometer provides traceability through an un-broken chain to NPL and NIST standards. 4 Generator Design 4.1 Basic Concept: The basic design concept was to use a novel combination of cascaded air mass flow controllers in conjunction with a proprietary liquid mass flow controller with controlled vaporisation/mixing, in order to allow the generation of any humidity level between -100 and +20 C frost/dew point, equivalent 13 ppb to 23,600 ppm. This represents a dynamic concentration range of 2 x As a result, extremely careful selection of mass flow controllers was necessary to ensure an appropriate sensitivity and reproducibility of generation. 4.2 Selection Of Suitable Mass Flow Controllers: A complex mixing system comprising three stages of mixing of wet and dry air flows was developed in conjunction with the Dutch manufacturer Bronkhorst Hi-Tec, using their model EL-Flow liquid and gas mass flow controllers and a Bronkhorst CEM Controlled Evaporator and Mixer unit. Figure 9 illustrates the basic working of the system.

10 Figure 9. MFC based dewpoint generator schematic. A pressurised pure water source is fed through the Liquid Mass Flow Controller, LFC, at a controlled flow rate from 0 to litres per minute. In the first stage of mixing this liquid water is mixed with a source of dry air (at approximately -100 C frost point) controlled by air mass flow controller GFC1 at a rate of between 0 and 5 litres per minute. This mixture then passes through the CEM evaporator/mixing unit and into an air receiver that acts as a further buffer, improving output stability of the total system. A controlled amount of this pre-mix air is then mixed in the second stage using GFC2, with more dry air through GFC3, the excess flow from the first mixing stage being exhausted the electronic pressure controller EPC1 and a rotary vent valve. In a similar arrangement, GFC4 and GFC 5 provide third stage mixing of pre-mix (2) and dry air, with the excess again vented through EPC2. The resultant mixed gas is then ready for delivery at a controlled flow rate to the sensors under test. Figure 10. MFC based dewpoint generator.

11 Whilst relatively simple in principle, practical realisation of a good working system relies both on the careful selection of the appropriate controllers and a sophisticated electronic control system, which in the case of this generator system is developed as a Windows based automated routine using C++ programming language. Table 1 on the following page lists the six mass flow controller elements used in the generator with their operating flow range and reproducibility. Controller Range Worst Accuracy, Nl/min LFC 0 to 12.5 g/h g/h < g/h GFC1 0 to 10 Nl/min 0.1 <0.01 GFC2 0 to 2 Nl/min 0.02 <0.002 GFC3 0 to 10 Nl/min 0.1 <0.01 GFC4 0 to 0.2 Nl/min 0.02 <0.002 GFC5 0 to 10 Nl/min 0.1 <0.01 Table 1. MFC selection. Reproducibility, Nl/min The liquid mass flow controller is calibrated in grams per hour. This can be converted to a volumetric flow rate of the evaporated water as follows: 1 gram mole occupies a volume of 22.4 litres at stp, so g occupies 22.4 litres Therefore a flow of 12.5 grams per hour is equivalent to 22.4 x12.5/ = litres per hour. So, the Controlled Evaporation method is an effective method for precisely delivering a saturation water vapour flow at a rate as low as litres per minute. 4.3 Calculation Of Theoretical Generator Settings: Whilst the design of the generator allows for either open-loop or closed-loop operation, it has been set up to operate in a semi-open loop configuration whereby the reference hygrometer is used to validate the generated humidity and control the overall calibration cycle, but not to provide direct closed loop control of the mass flow controllers. This latter feature can be considered as a design modification that would further improve the system in the future. In the open-loop configuration, mass flow controller settings are developed based on a simple theoretical model using volumetric calculation of the liquid water and multi-stage dry air mixing stages. A theoretical generated frost-point temperature is calculated from the mass flow controller settings as follows, assuming dry gas feed is at 13 ppb moisture content. In this example the target generated frost point is -90 C: Stage 1: Wet flow is g/h water = 22.4*(0.308/18.001) = litres per hour Dry flow is l/min = litres per hour at < 0.01 ppmv Resultant flow is effectively 10 6 *0.3833/ = ppmv ( C f.p.)

12 Stage 2: Wet flow is = 0.055* /5.055 = Dry flow is = 5.000*0.013/5.005 = Resultant flow is therefore ( ) = ppmv ( C f.p.) Stage 3: Wet flow is = 0.041* /5.041 = Dry flow is = 5.000*0.013/5.005 = Resultant flow is therefore ( ) = ppmv ( C f.p.) Three-stage mixing at even the lowest humidity levels provides a good repeatability, being less than 0.2% of controlled value for each flow controller. Table 2 illustrates the relevant flow settings for all six mass flowmeters required to generate a -90 C frost point. Mixing Stage Flow settings, Nl/min [g/h] Target frost point, Repeatability, C C Stage 1 GFC1=5.001, [LFC=0.308] ±0.02 Stage 2 GFC2=0.055, GFC3= ±0.01 Stage 3 GFC4=0.041, GFC5= ±0.01 Table 2. Three stage mixing of a -90 C frost point. This gives a system repeatability, excluding variations in dry air moisture content, of ±0.025 C frost point, calculated in quadrature based on maximum repeatability errors at each mixing stage. 4.4 Inclusion Of A Buffer Air Reservoir: Initial test results with the system indicated that a problem existed in achieving a stable generated humidity. It was discovered that operation of the liquid flow meter at very low flow rates, even with the addition of the Controlled Evaporator and Mixer, was giving rise to an erratic evaporation rate associated with the evaporation of individual water droplets. Whilst not expected to be a problem for the calibration of impedance type hygrometers, this cyclic response with a peak-to-peak variation of approximately 0.5 C at +10 C dew point and 1 C at -20 C frost point was large enough to cause total system uncertainty levels to exceed the required levels. It was therefore decided to incorporate a secondary mixing chamber after the first stage premix in order to provide a time averaging of this fluctuation and reduce its peak-to-peak effect from the above levels down to 0.04 C at +10 C dew point and 0.2 C at -20 C frost point.

13 The subsequent remixing ( x 2) of this first stage air mix improves the fluctuation further down to levels that are insignificant compared to the overall system uncertainty. Whilst selection of an appropriate air reservoir was important from the perspective of pressure rating, for safety reasons, the materials of construction were less important as the lowest pre-mix frost point at this stage is approximately -20 o C, rendering adsorption/wall effects insignificant. Charts 2. and 3. show the effects of the air reservoir on the output stability at -20 C frost point, plotted over a twenty minute period with no air dryer changeover, in order to de-couple any variation that may be caused by imbalance between the efficiency of the two molecular sieve desiccant columns. Chart 2. Generator output stability at -20 C frost point without air reservoir.

14 Chart 3. Generator output stability at -20 C frost point with air reservoir. 4.5 Design Of The Sensor Calibration Chamber: In the past, Michell had used multi-sensor calibration chambers machined from a solid block of stainless steel, with drilled and plugged gas flow paths to minimise dead space and internal volume. These blocks were expensive to produce and had limited capacity due to the design, accommodating a maximum of 24 traditional sensors or 12 transmitters, due to the larger housing size required for transmitter electronics. For the new humidity calibration systems it was decided to employ a new design of sensor calibration chamber, to accommodate the larger transmitter format and allow a much larger number of sensors to be calibrated simultaneously. The general arrangement of a typical calibration manifold is shown in figure 11. Figure 11. A calibration manifold (Note the system is loaded from both sides).

15 A vertical format stainless steel manifold design was implemented, giving each of the eight calibration chambers the capacity to house up to 250 transmitters during any calibration cycle. The calibration manifold is welded construction with high integrity VCR coupling gas inlet and outlet ports. Each transmitter is secured into the calibration tube using its sensor guard thread, and is sealed to the manifold with an individual Teflon o ring. As a further enhancement, each calibration chamber is mounted within a sealed environment during the calibration process. The cowling is purged with air at -100 C frost point to minimise humidity gradients between the inside and outside of the sensor calibration chamber thus negating any leakage effects that might result from a poor seal at the sensor o ring. Tests have indicated that this can reduce the differential (air out air in) by more than 80%, so improving the calibration integrity of the sensors under test and therefore reducing the overall uncertainty of calibration. Figure 12 View of a section of the laboratory The air path from generator to sensor calibration chamber is maintained in electro-polished 10 mm OD stainless steel tubing with a minimum of bends and welded angle connections. A 5 metre tail pipe after the calibration chamber, in 6mm OD stainless steel tube, is used to reduce back diffusion effects. 5 Verification of System Performance This calibration system is essentially a practical device for the calibration of capacitive type dew-point sensors over the range -100 to +20 C dew point. As such, it is designed to give a reasonable uncertainty in context of the nature of the sensors being calibrated. The intention in design was to produce a highly reliable system that would allow an overall calibration uncertainty for the finished product (capacitive sensors) of ±1 C dew point from +20 to - 60 and ±2 C in the range -60 to -100 C frost point. Table 3 on the following page illustrates the results of a series of 95 calibration runs, from which the generation uncertainty and overall system uncertainty were calculated. It should be noted that the overall combined system uncertainty incorporates the uncertainty of the reference hygrometer, calibrated in Michell s UKAS accredited laboratory over the range -75 to +20 C dew point. The reference hygrometer uncertainty below -75 C frost point is determined from previous experimental data.

16 Target Dew Point, o C Mean Actual Dew Point, o C Standard Deviation Uncertainty (95% CI) Max Value, o C Min Value, o C Spread, o C UKAS Inst. Uncertainty (95%CI), o C Combined System Uncertainty (95% CI), o C Table 3. Summary of actual test results and calculation of combined system uncertainty. As can be seen from the results, the target combined uncertainty from -60 to +20 C dew point exceeds the design brief by an average of around 50%. Whilst the combined uncertainty currently exceeds the ±2 C target at lowest frost point (not generated), it is expected that significant improvements will result when the reference hygrometer can be calibrated against a national standard to these levels and further enhancements are made to the system to improve low end reproducibility. 6 Summary A commercial calibration laboratory exists to maintain a strong link with national standards, and by appropriate means to be able to offer calibration services with the best possible overall uncertainty. The practical challenge is to be able to meet the requirements of low volume traceable calibration and also to be able to provide high volume calibrations with minimal detriment to uncertainty. In this paper we have seen how a high volume humidity calibration system is able to provide practical and reliable calibration of capacitive dew-point sensors. Figure 13. Unloading calibrated transmitters

17 In conjunction with a fully automatic, computer control program, these systems are able to generate a range of dew point levels, verify them in real time against a traceable reference chilled mirror dew-point hygrometer, measure and record the data from the 250 sensors under test and validate correct programming of sensor memory with the appropriate calibration data. This allows the fully automatic calibration of capacitive dew-point sensors in a controlled environment with no human supervision or involvement. We have also demonstrated that these systems have, in the main, achieved their design brief and have delivered a combined uncertainty of less than ±2 C dew/frost point across an extremely wide measurement range of -100 C frost point to +20 C dew point. These systems have been fully operational for several years now, each one capable of performing in the region of 40 calibration runs per year, providing a total capacity of more than 80,000 calibrated dew-point sensors in a 12 month period. An extended validation programme has demonstrated both the reliability and long-term stability of the system. So, in conclusion, we have achieved our objective and have in one laboratory systems based on similar technologies that are capable of low volume, high labour content, traceable calibration on the one hand and high volume automated calibration on the other. Bibliography Bronkorst Hi-Tech BV, Mass Flow Meter Manual, 29 Dec 2000 Bronkhorst Hi-Tech BV, CEM Controlled Evaporator Manual, 13 Oct 1997 G W C Kaye and T H Laby, Handbook of Physical and Chemical Constants, 14 th Edition, 1973 The Institute of Measurement and Control, A Guide to the Measurement of Humidity, 1996 B Cretinon and J Merigoux, La Mesure de l humidite dans le Gaz, 2000 C Nordling, J Osterman, Physics Handbook, 1980 ISO : (1992)

AMT-Ex Dewpoint Transmitter

AMT-Ex Dewpoint Transmitter AMT-Ex Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality

More information

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc. The HumiSys RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The new HumiSys with single or dual RH probes capabilities is

More information

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013 NMISA NEW GAS FLOW CALIBRATION FACILITY Deona Jonker 7 October 2013 Introduction Flow the total volume of a fluid that flows past a fixed point in a given time Flow measurements used in industry as a means

More information

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0 High Accuracy Humidity Calibrator The role of humidity calibrations that are accurate, reproducible, and documentable is becoming more and more important. ISO quality guidelines and regulations according

More information

HumiSys HF High Flow RH Generator

HumiSys HF High Flow RH Generator HumiSys HF High Flow RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The HumiSys HF is a high flow version of the previously

More information

CORESTA RECOMMENDED METHOD N 6

CORESTA RECOMMENDED METHOD N 6 CORESTA RECOMMENDED METHOD N 6 DETERMINATION OF VENTILATION DEFINITIONS AND MEASUREMENT PRINCIPLES (2015 Revision September 2016) 1. SCOPE This CORESTA Recommended Method specifies a method for the determination

More information

Model PDT Dewpoint Transmitter

Model PDT Dewpoint Transmitter Model PDT Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk

More information

Ron Gibson, Senior Engineer Gary McCargar, Senior Engineer ONEOK Partners

Ron Gibson, Senior Engineer Gary McCargar, Senior Engineer ONEOK Partners New Developments to Improve Natural Gas Custody Transfer Applications with Coriolis Meters Including Application of Multi-Point Piecewise Linear Interpolation (PWL) Marc Buttler, Midstream O&G Marketing

More information

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities Specific Accreditation Criteria Calibration ISO IEC 17025 Annex Mass and related quantities January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected

More information

An Automated High Precision Saturation/Dilution Calibration System for Trace Moisture Sensors

An Automated High Precision Saturation/Dilution Calibration System for Trace Moisture Sensors GE Oil & Gas An Automated High Precision Saturation/Dilution Calibration System for Trace Moisture Sensors The paper discusses the physical construction and operating principles of this saturation/dilution

More information

Instruction Manual. Model DSP-Ex Portable Dewpoint Meter. Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England

Instruction Manual. Model DSP-Ex Portable Dewpoint Meter. Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Model DSP-Ex Portable Dewpoint Meter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 Email: info@amsytems.co.uk Web:

More information

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR) Procedure No: SOP-025 Revision No: 1.0 (January 21, 2011) Page No.: 1 of 8 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features ANALOG MASS FLOW CONTROLLERS Model AF mass fl ow controllers are designed to indicate fl ow rates and control set fl ow rates of gases. Each of these units incorporates an advanced straight tube sensor

More information

Generating Calibration Gas Standards

Generating Calibration Gas Standards Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and

More information

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS Precise and Stable SF 6 Gas Analyzer Simultaneous measurement of humidity, SF 6 purity and SO 2 concentration Integrated gas recovery system with automatic pump back Fully automated SF 6 gas testing Fundamental

More information

Improve Process Reliability

Improve Process Reliability Convenient Conductivity Calibration to Meet Pharmacopeia Regulations Pharmacopeias require that conductivity measurement systems be periodically calibrated to ensure their precision. The range of accessories

More information

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving flotek.g 2017- Innovative Solutions in Flow Measurement and Control - Oil, Water and Gas August 28-30, 2017, FCRI, Palakkad, Kerala, India Validation of Custody Transfer Metering Skid at Site After Laboratory

More information

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation High-precision Humidity Calibrator The role of humidity calibrations that are accurate, reproducible, and documentable is becoming more and more important. ISO quality guidelines and regulations according

More information

The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions

The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions Flow Meter with multiple functions for gases and liquids M3 The M-series Flow Meter, with its versatile and user-friendly

More information

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi

More information

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS APPLICATION NOTE LC-136 Introduction Specialized spaces often times must

More information

APPLICATION NOTE. GC Integrated Permeation Device

APPLICATION NOTE. GC Integrated Permeation Device GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords:

More information

This test shall be carried out on all vehicles equipped with open type traction batteries.

This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following

More information

Drilling Efficiency Utilizing Coriolis Flow Technology

Drilling Efficiency Utilizing Coriolis Flow Technology Session 12: Drilling Efficiency Utilizing Coriolis Flow Technology Clement Cabanayan Emerson Process Management Abstract Continuous, accurate and reliable measurement of drilling fluid volumes and densities

More information

HUMOR 20 HUMOR 20. High-precision Humidity Calibrator. Water T, p1

HUMOR 20 HUMOR 20. High-precision Humidity Calibrator. Water T, p1 High-precision Humidity Calibrator The role of humidity calibrations that are accurate, reproducible, and documentable is becoming more and more important. ISO quality guidelines and regulations according

More information

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring.

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring. Product Data Sheet CODEL Continuous Emission Monitoring GCEM4100 Extractive Gas Analyser CO, CO2, NO, NO2, SO2, HCL, CH4 & H2O Hot extractive multi-channel gas analyser for continuous process and emissions

More information

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD) Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet

More information

Humidity and Dew-Point Calibration

Humidity and Dew-Point Calibration Humidity and Dew-Point Calibration Instruments, Systems & Services Michell Instruments has 40 years' experience providing moisture measurement & calibration solutions to our customers Contents A Background

More information

The latest automotive systems require innovative leak test methods and fixturing.

The latest automotive systems require innovative leak test methods and fixturing. Leak Testing Auto Parts The latest automotive systems require innovative leak test methods and fixturing. In the past, automobile manufacturers only required basic forms of leak testing to check standard

More information

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring.

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring. Product Data Sheet CODEL Continuous Emission Monitoring GCEM4100 Extractive Gas Analyser CO, CO2, NO, NO2, SO2, HCL, CH4 & H2O Hot extractive multi-channel gas analyser for continuous process and emissions

More information

Improving distillation tower operation

Improving distillation tower operation Improving distillation tower operation Measuring differential pressure across long sections of distillation columns has always been challenging, but purpose-built sensor systems provide a solution Fast

More information

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34) CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE (+34) 968 162 005 www.sensovant.com comercial@sensovant.com www.epluse.com CALIBRATION AT THE HIGHEST LEVEL The

More information

General Accreditation Guidance. User checks and maintenance of laboratory balances

General Accreditation Guidance. User checks and maintenance of laboratory balances General Accreditation Guidance User checks and maintenance of laboratory balances January 2018 Copyright National Association of Testing Authorities, Australia 2010 All intellectual property rights in

More information

ONSITE PROVING OF GAS METERS. Daniel J. Rudroff WFMS Inc West Bellfort Sugar Land, Texas. Introduction

ONSITE PROVING OF GAS METERS. Daniel J. Rudroff WFMS Inc West Bellfort Sugar Land, Texas. Introduction ONSITE PROVING OF GAS METERS Daniel J. Rudroff WFMS Inc. 13901 West Bellfort Sugar Land, Texas Introduction With the increased use of Natural Gas as a fuel, and higher natural gas prices buyers and sellers

More information

Gas mixtures. Individual solutions specifically for your application

Gas mixtures. Individual solutions specifically for your application Gas mixtures Individual solutions specifically for your application For many applications besides pure gases also defined gas mixtures are required. For many routine applications, from operating gases

More information

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION. Andrzej Hantz RADWAG BALANCES AND SCALES INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION. The article presents the influence of comparison process automatization

More information

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA) Int. J. Metrol. Qual. Eng. 5, 303 (2014) c EDP Sciences 2014 DOI: 10.1051/ijmqe/2014012 Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA) D. Jonker and E.P. Tarnow

More information

CHAPTER 3 RESULTS AND DISCUSSION

CHAPTER 3 RESULTS AND DISCUSSION CHAPTER 3 RESULTS AND DISCUSSION The research included the development and verification of the earlier apparatus, and the collection and validation of moisture diffusion data under both isothermal and

More information

AUTOMATIC HOSE TEST UNIT, TYPE SPU

AUTOMATIC HOSE TEST UNIT, TYPE SPU VALVES AND FITTINGS UP TO 14,000 BAR TEST AND CONTROL EQUIPMENT H IGH PRESSURE TECHNOLOGY AUTOMATIC HOSE TEST UNIT, TYPE SPU Pressure range from 1 up to 10,000 bar User-friendly touch panel operation HIGH-PRESSURE

More information

Keeping It Simple. Gas Mixtures for Stack Emissions Monitoring.

Keeping It Simple. Gas Mixtures for Stack Emissions Monitoring. Keeping It Simple Gas Mixtures for Stack Emissions Monitoring www.uk.airliquide.com Stack Emissions Monitoring The Stack Emissions Monitoring market has undergone substantial legislative and quality changes

More information

Title: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator

Title: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator Procedure No: SOP-012 Revision No: 1.2, January 27, 2011 Page No.: 1 of 7 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Sabio model 2010 and 4010 dilution calibrators.

More information

Humidity and Dew-Point Calibration Instruments, Systems & Services

Humidity and Dew-Point Calibration Instruments, Systems & Services www.nbn.at Humidity and Dew-Point Calibration Instruments, Systems & Services Michell Instruments has 40 years' experience providing moisture measurement & calibration solutions to our customers Contents

More information

Dead-weight tester High-pressure version Model CPB3800HP

Dead-weight tester High-pressure version Model CPB3800HP Calibration Dead-weight tester High-pressure version Model CPB3800HP WIKA data sheet CT 31.07 for further approvals see page 5 Applications Primary standard for calibrating the pressure scale in a hydraulic

More information

NMISA New Gas Flow Calibration Facility

NMISA New Gas Flow Calibration Facility NMISA New Gas Flow Calibration Facility Speaker / Author: D. Jonker Co-author: E.P. Tarnow National Metrology Institute of South Africa (NMISA) Private Bag X34, Lynnwood Ridge, Pretoria, 0040, South Africa

More information

Best Practice Guide, Servomex 2700

Best Practice Guide, Servomex 2700 For full installations details refer to the. Best Practice Guide, Servomex 2700 Mounting: General Guidelines: Servomex 2700 Control Units and air supplies (utilities units) should, ideally, be mounted

More information

800 Series Mass Flow Meters and Controllers. High Performance Mass Flow Meters and Controllers

800 Series Mass Flow Meters and Controllers. High Performance Mass Flow Meters and Controllers 800 Series Mass Flow Meters and Controllers High Performance Mass Flow Meters and Controllers WHY MASS FLOW? In most processes it is mass, not volume, which is the critical variable. Volumetric flow measurements

More information

Vehicle- or rack-mounted liquefied gas meters, pump supplied

Vehicle- or rack-mounted liquefied gas meters, pump supplied Purpose This inspection procedure outline (IPO) defines the minimum tests which must be performed to ensure that basic volumetric measuring devices comply with the legislation. Application Rack/fixed metering

More information

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW AC 2010-2145: MEASUREMENT OF HYDROGEN IN HELIUM FLOW Randy Buchanan, University of Southern Mississippi Christopher Winstead, University of Southern Mississippi Anton Netchaev, University of Southern Mississippi

More information

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING Blair Chalpin Charles A. Matthews Mechanical Design Engineer Product Support Manager Scanivalve Corp Scanivalve Corp Liberty Lake, WA

More information

Design Features. General Description

Design Features. General Description Model thermal Mass Flow Controllers are designed to indicate and control set flow rates of gases. The combines the characteristics and accuracy of conventional mass flow devices into a unique compact design

More information

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1250 METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Total Organic Compounds N/A

More information

Moisture levels in compressed breathing air. Prepared by QinetiQ Limited for the Health and Safety Executive 2006 RESEARCH REPORT 427

Moisture levels in compressed breathing air. Prepared by QinetiQ Limited for the Health and Safety Executive 2006 RESEARCH REPORT 427 Moisture levels in compressed breathing air Prepared by QinetiQ Limited for the Health and Safety Executive 2006 RESEARCH REPORT 427 Moisture levels in compressed breathing air T G Anthony & P R Clarke

More information

ISO/TR TECHNICAL REPORT. Natural gas Calibration of chilled mirror type instruments for hydrocarbon dewpoint (liquid formation)

ISO/TR TECHNICAL REPORT. Natural gas Calibration of chilled mirror type instruments for hydrocarbon dewpoint (liquid formation) TECHNICAL REPORT ISO/TR 12148 First edition 2009-02-15 Natural gas Calibration of chilled mirror type instruments for hydrocarbon dewpoint (liquid formation) Gaz naturel Étalonnage des instruments du type

More information

Vehicle-mounted meters, pump supplied

Vehicle-mounted meters, pump supplied Purpose This inspection procedure outline (IPO) defines the minimum tests which must be performed to ensure that basic volumetric measuring devices comply with the legislation. Application Vehicle-mounted

More information

Measurement And Control Appreciation

Measurement And Control Appreciation Measurement And Control Appreciation Course Content Aim To provide an appreciation of the operation and application of process plant instrumentation used for the measurement and display of the main process

More information

Hydraulic pressure balance Model CPB5800

Hydraulic pressure balance Model CPB5800 Calibration technology Hydraulic pressure balance Model CPB5800 WIKA data sheet CT 31.11 for further approvals see page 8 Applications Primary standard for calibrating the pressure scale in a hydraulic

More information

Pneumatic dead-weight tester Model CPB3500

Pneumatic dead-weight tester Model CPB3500 Calibration technology Pneumatic dead-weight tester Model CPB3500 WIKA data sheet CT 31.22 Applications Primary standard for calibrating the scale in a pneumatic range from -1... 120 bar / -14.5... 1,600

More information

Mitos Fluika Pressure and Vacuum Pumps Datasheet

Mitos Fluika Pressure and Vacuum Pumps Datasheet Unit 1, Anglian Business Park, Orchard Road, Royston, Hertfordshire, SG8 5TW, UK T: +44 (0)1763 242491 F: +44 (0)1763 246125 E: sales@dolomite-microfluidics.com W: www.dolomite-microfluidics.com Dolomite

More information

Long term stability tests of INO RPC prototypes

Long term stability tests of INO RPC prototypes Long term stability tests of INO RPC prototypes While the problem of sudden aging in the glass RPC prototypes is being investigated, a few RPCs of dimension 40 cm 30 cm were fabricated, using glass procured

More information

CORI-FLOW PRECISION MASS FLOW METER / CONTROLLER FOR LIQUIDS AND GASES

CORI-FLOW PRECISION MASS FLOW METER / CONTROLLER FOR LIQUIDS AND GASES SENSORS FOR MASS FLOW AND PRESSURE CORI-FLOW PRECISION MASS FLOW METER / CONTROLLER FOR LIQUIDS AND GASES INTRODUCTION Bronkhorst High-Tech BV is the worldwide pioneer in the field of low flow liquid metering

More information

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors Abstract Boiler water in the low parts per billion dissolved oxygen concentration range can be significantly contaminated by oxygen

More information

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology: Title: Comparison of Chemical Sensor and NDIR Technologies TSN Number: 25 File:\\MII- SRV1\Metron\Bridge_Analyzers\Customer_Service_Documentation\White_Papers\25_G en EGA NDIR vs Chemical Sensor.docx Created

More information

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

Beamex. Calibration White Paper.  Weighing scale calibration - How to calibrate weighing instruments Beamex Calibration White Paper info@beamex.com Weighing scale calibration - How to calibrate weighing instruments Weighing scale calibration - How to calibrate weighing instruments Weighing scales, weighing

More information

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 602 (2009) 845 849 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

Selecting the right pressure sensor for your application

Selecting the right pressure sensor for your application Selecting the right pressure sensor for your application This guide from Acal BFi details the key factors that you must consider when selecting a pressure sensor for your application. Selecting the best

More information

IBEDA GAS MIXERS HIGHEST PRECISION FOR YOUR WELDING AND CUTTING PROCESSES

IBEDA GAS MIXERS HIGHEST PRECISION FOR YOUR WELDING AND CUTTING PROCESSES IBEDA GAS MIXERS HIGHEST PRECISION FOR YOUR WELDING AND CUTTING PROCESSES GAS SAFETY ENGINEERING GAS MANIFOLD SYSTEMS AUTOGENOUS ENGINEERING FLAME SPRAYING GAS MIXERS IBEDA GAS MIXERS FOR MANY AREAS OF

More information

MEASUREMENT BEST PRACTICES FORIMPROVEDREFINERY SAFETY, AVAILABILITY & EFFICIENCY

MEASUREMENT BEST PRACTICES FORIMPROVEDREFINERY SAFETY, AVAILABILITY & EFFICIENCY MEASUREMENT BEST PRACTICES FORIMPROVEDREFINERY SAFETY, AVAILABILITY & EFFICIENCY Mark Menezes, PlantWeb Marketing Manager Rosemount Inc. Eric Wickberg, Pressure Applications Marketing Manager Rosemount

More information

FAS-W. Features. Applications. Automatic condensation hygrometer

FAS-W. Features. Applications. Automatic condensation hygrometer FAS-W HUMIDITY ANALYZER Automatic condensation hygrometer the «FAS-W» Moisture Analyzer automatic condensation hygrometer can measure both the volume fraction of moisture and the dew point temperature

More information

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111 INTRODUCTION Water vapor in natural gas has more than a substantial effect on the quality of the gas stream. Without quality measurement of water vapor the gas is basically not saleable. Contracts are

More information

Investigating the effects of interchanging components used to perform ripple assessments on calibrated vector network analysers

Investigating the effects of interchanging components used to perform ripple assessments on calibrated vector network analysers Abstract Investigating the effects of interchanging components used to perform ripple assessments on calibrated vector network analysers Andrew G Morgan and Nick M Ridler Centre for Electromagnetic and

More information

CO, NO, NO 2, NO x, SO 2, CH 4, HCl, CO 2 & H ² O. Extractive low cost, low maintenance devices for continuous process and emissions monitoring.

CO, NO, NO 2, NO x, SO 2, CH 4, HCl, CO 2 & H ² O. Extractive low cost, low maintenance devices for continuous process and emissions monitoring. Product Data Sheet CODEL Continuous Emission Monitoring GCEM 40 Series Extractive Gas Analyser CO, NO, NO 2, NO x, SO 2, CH 4, HCl, CO 2 & H ² O Extractive low cost, low maintenance devices for continuous

More information

MEASURING WATER CONTENT IN COMPRESSED BREATHING AIR IN COMPLIANCE WITH EN 12021:2014 ULTRA-PRECISE AND ULTRA-RELIABLE WITH POLYMER TECHNOLOGY

MEASURING WATER CONTENT IN COMPRESSED BREATHING AIR IN COMPLIANCE WITH EN 12021:2014 ULTRA-PRECISE AND ULTRA-RELIABLE WITH POLYMER TECHNOLOGY MEASURING WATER CONTENT IN COMPRESSED BREATHING AIR IN COMPLIANCE WITH EN 12021:2014 ULTRA-PRECISE AND ULTRA-RELIABLE WITH POLYMER TECHNOLOGY SPORTS & SAFETY BAUER KOMPRESSOREN WHY PRECISE MEASUREMENT

More information

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service? Pipette specifications according to ISO 8655 Chapter 5 5.1 The ISO 8655 standard gives the accuracy and precision limits as both absolute and relative values. Specifications will depend on the technique

More information

WP Eliminating Oxygen from the Purge Gas and the use of Monitoring Equipment

WP Eliminating Oxygen from the Purge Gas and the use of Monitoring Equipment Part 4 WP - 269 Eliminating Oxygen from the Purge Gas and the use of Monitoring Equipment Even using specialised weld purging equipment does not guarantee defect free welds. Control of the oxygen content

More information

OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply

OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply 4.1.1 General The primary function of a system for handling fuel oil is to transfer oil from the storage tank to the oil burner at specified conditions

More information

PURA. Pure Gas Dewpoint Transmitter. Users Guide

PURA. Pure Gas Dewpoint Transmitter. Users Guide PURA Pure Gas Dewpoint Transmitter Users Guide Issue January 2003 Page 2 Table of Contents SECTION PAGE 1 Product Overview 3 2 Preparation 3 3 Installation 3 3.1 PURA Premium & OEM Dewpoint Transmitter

More information

Series 7250 Ruska High-Speed Digital Pressure Controller. GE Sensing. Features

Series 7250 Ruska High-Speed Digital Pressure Controller. GE Sensing. Features Features Pressure ranges from 0 to 5 and 0 to 3000 psi (0 to 400 mbar and 0 to 210 bar) xi and 7250i provide advanced precision of 0.005% of reading provides 0.003% of full scale precision Stability: 0.0075%

More information

Gerald D. Anderson. Education Technical Specialist

Gerald D. Anderson. Education Technical Specialist Gerald D. Anderson Education Technical Specialist The factors which influence selection of equipment for a liquid level control loop interact significantly. Analyses of these factors and their interactions

More information

Analogue and Digital Mass Flow Meters and Controllers for Gases MASS-STREAM

Analogue and Digital Mass Flow Meters and Controllers for Gases MASS-STREAM Analogue and Digital Mass Flow Meters and Controllers for Gases MASS-STREAM M+W Instruments Your partner Key Facts M+W Instruments was founded in 1988 and has always specialised in thermal mass flow meters

More information

Cover Page for Lab Report Group Portion. Pump Performance

Cover Page for Lab Report Group Portion. Pump Performance Cover Page for Lab Report Group Portion Pump Performance Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 March 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:

More information

COMPAFLOW. Compressed Air. Volumetric flow. Gas. Mass flow. Steam. Net volumetric flow. Liquid

COMPAFLOW. Compressed Air. Volumetric flow. Gas. Mass flow. Steam. Net volumetric flow. Liquid Volumetric flow Compressed Air Mass flow Gas Net volumetric flow Steam Liquid Universal compact orifice flow meter combines a compact orifice flow sensor and sotiphicated converter One-piece flow sensor,

More information

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter Overview C % ppm 4~20mA Analog Output ppb Power Supply 100~240 VAC mg/l Dimensions 96 x 96 x 132mm RS-485 Digital Output (for DC-5310-RS only)

More information

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010 Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600 November 2010 Table of Contents ``Abstract...3 ``Introduction...4 ``Methods & Results...6 ``Data Analysis...9 ``Conclusion...12

More information

Niltox Breathing Air Purifier Range

Niltox Breathing Air Purifier Range Breathing Air Requirements The main standards affecting breathing air are BSEN529, EN12021 and COSHH. The requirements for a breathing air system are shown in the table opposite and the requirements for

More information

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions L 100 Bubble-Tube Level System Installation, Operation and Maintenance Instructions Figure 1 Contents Section Description Page 1.0 Introduction 2 2.0 Specifications 3 3.0 Installation 3 4.0 Warranty 6

More information

Thermo Scientific Model 146i Principle of Operation

Thermo Scientific Model 146i Principle of Operation Thermo Scientific Model 146i Principle of Operation The Model 146i Multi-gas Calibrator dilutes calibration gases to precise concentrations. The diluted gases are used to perform zero, precision and Level

More information

STANDARDS FOR COMPRESSED GAS TESTING

STANDARDS FOR COMPRESSED GAS TESTING STANDARDS FOR COMPRESSED GAS TESTING APPLICATION NOTE HPD-002 (US) Introduction This publication provides excerpts from some of the many guidelines and standards that pertain to compressed gases in cleanrooms.

More information

Automatic Permeability Testing: The Challenges and Solutions Author: Alyce Hartvigsen, PBI-Dansensor A/S, Ringsted, Denmark

Automatic Permeability Testing: The Challenges and Solutions Author: Alyce Hartvigsen, PBI-Dansensor A/S, Ringsted, Denmark Automatic Permeability Testing: The Challenges and Solutions Author: Alyce Hartvigsen, PBI-Dansensor A/S, Ringsted, Denmark In recent years, numerous product developments and major changes in the distribution

More information

The Future of Hydraulic Control in Water-Systems

The Future of Hydraulic Control in Water-Systems The Future of Hydraulic Control in Water-Systems A. Heimann Manager of R&D and of Technical Support & Applications Engineering departments at Dorot Automatic Control Valves Dorot Control Valves, Kibbutz

More information

Fundamentals of Compressed Air Systems. Pre-Workshop Assignment

Fundamentals of Compressed Air Systems. Pre-Workshop Assignment Page 1 In order to ensure that the Compressed Air Challenge Fundamentals of Compressed Air Systems Training is most useful to you, it will be important for you to bring information about your plant s compressed

More information

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS Annex 3, page 2 ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS The text of existing chapter 15 is replaced by the following: "1 Application This

More information

Soil Water Characteristic Curve (SWCC) Using the WP4C

Soil Water Characteristic Curve (SWCC) Using the WP4C Energy GeoEngineering Laboratory Soil Water Characteristic Curve (SWCC) Using the WP4C Prepared by: Josbel Cordero & Geralds Useche Note: This pictorial manual is not intended to replace the official Decagon

More information

Nitrogen subtraction on reported CO 2 emission using ultrasonic flare gas meter

Nitrogen subtraction on reported CO 2 emission using ultrasonic flare gas meter Nitrogen subtraction on reported CO 2 emission using ultrasonic flare gas meter Kjell-Eivind Frøysa, Christian Michelsen Research AS, Bergen, Norway Henning Ekerhovd, StatoilHydro ASA, Kollsnes, Norway

More information

Constant Pressure Inlet (CCN) Operator Manual

Constant Pressure Inlet (CCN) Operator Manual Constant Pressure Inlet (CCN) Operator Manual DOC-0125 Revision J 2545 Central Avenue Boulder, CO 80301-5727 USA C O P Y R I G H T 2 0 1 1 D R O P L E T M E A S U R E M E N T T E C H N O L O G I E S, I

More information

PTB Ex PT Scheme. Procedure Instruction of program Explosion Pressure - Test Round 2017

PTB Ex PT Scheme. Procedure Instruction of program Explosion Pressure - Test Round 2017 Procedure Instruction of program Explosion Pressure - Test Round 2017 Content 1 General Performance... 2 2 Measurand / Characteristic of interest... 2 3 Test Item and Configurations... 2 3.1 Components...

More information

PROCESS MONITORING AND CONTROL

PROCESS MONITORING AND CONTROL Chemical Engineering Process Design PROCESS MONITORING AND CONTROL Keith Marchildon David Mody Monitoring means Measurement 1. There is no Control without Monitoring 2. Monitoring should not be done without

More information

TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION

TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION Approved By: Senior Manager: Mpho Phaloane Revised By: Field Manager: Neville Tayler Date of Approval: 2015-08-26 Date of Implementation: 2015-08-26

More information