Practical approach and problems in in-situ RGA calibration
|
|
- Benedict Park
- 6 years ago
- Views:
Transcription
1 Practical approach and problems in in-situ RGA calibration Oleg Malyshev and Keith Middleman Vacuum Science Group, ASTeC Accelerator Science and Technology Centre STFC Daresbury Laboratory UK Workshop on measurement characteristics and use of quadrupole mass spectrometers for vacuum applications, EMRP IND12. Bled, Slovenia, April 10 13,
2 ASTeC Vacuum Science group main interests Vacuum in particle accelerators: Achieving, measuring, modelling, designing... Vacuum related studies in the VS lab: Thermal outgassing Gauge and RGA calibration Pumping property measurements Electron stimulated desorption Surface coatings and analysis Photocathode development 2
3 What RGAs used Make: MKS (Microvision) 8 Hiden 1 SX200 (head) + VGQ 15 Dyson 1 Prisma 3 Modes used: FAR and SEM Profile, trend and MID Leak detection Total pressure range: from 10-5 down to below mbar Other requirements: Bakeability, Stability, including XHV Traceability Low outgassing A wish: No setting change after initial calibration check and necessary adjustments 3
4 Why we need the RGA calibration A need of quantitative partial pressure measurements Lack of space for two instruments: gauge + RGA Outgassing of gauge + RGA is greater than RGA only Strange experimental results received with RGA output data Example: sticking probability >1?! Pimping speed S > S ideal = A v/4 It does not take too long to find that: RGAs are not calibrated they have a number of factory set parameters I(m/e=28) usually correspond to P(N 2 ) measured with a UHV gauge RGAs are adjusted with injection of noble gases not for residual gases in UHV Generally, as-received it is a qualitative (not a quantitative) device 4
5 Definitions of RGA calibration (1) (2) Peak alignment Accurate partial pressure Width alignment measurement Use of noble gas Different type of gases mixture Questions: Influence of cracking pattern Influence of RGA s gas factory An influence of a large peak on a neighbour small peaks Influence of ESD in the ion source 5
6 Set-up for NEG pumping evaluation O.B. Malyshev and K.J. Middleman. In situ ultrahigh vacuum residual gas analyzer calibration. J. Vac. Sci. Technol. A 26 (2008), p Test chamber 1 (option) 6
7 Main steps of in-situ calibration on the research rig Choice of injected gases Cleaning of injected gases (if necessary) Filling the gas chamber with known volume and high accuracy Baratron gauge If necessary, checking the Extractor gauge calibration vs High Accuracy Baratron gauge FS=1.3 mbar; Res = 10-6 mbar Calibrating RGA vs Extractor gauge by gas injection Analysis of calibration data to obtain calibration coefficients. 7
8 Injected gases H 2, CH 4, CO, CO 2, same as in the residual gas spectrum, N 2, Ar, O 2 LN 2 trap for cleaning of injected gases Very useful to reduce an impurity of injected gases (even for class 9999 gases) H 2 O is present for calibration before a bakeout or by heating a small part of vacuum chamber, or switching on a filament 8
9 Checking the extractor gauge ex-situ The extractor gauge is calibrated on the secondary calibration facility in ASTeC Vacuum Laboratory against two primary calibrated (at PTB and NPL) extractor gauges Advantage: accurate calibration to the secondary standard Disadvantage: requires a lot of work: removing the gauge from an experimental installation, installing it on the calibration facility, perform bakeout, calibration, transfer it back to the experimental installation Traceability against N 2 only... 9
10 Checking the extractor gauge calibration in-situ If there are any doubt about the gauge error in factor 2 or more: Gas expansion from the 0.15-l chamber initially filled with N 2 at about 10-3 mbar to the 1.5-l chamber Pumping gas out from the 1.5-l chamber Expansion to the 1.5-l chamber repeated. Pressure at 0.15-l chamber is a few 10-6 mbar (not accurately measurable by Baratron gauge but well calculated). Expansion to the test chamber (valve to the pumps closed). Equilibrium pressure is in the range of 10-7 mbar Comparison of calculated and measured pressures. Advantage: Time saving Disadvantage: Low accuracy (compared to ex-situ calibration) Accuracy of volume measurement and Baratron calibration 10
11 Calibrating RGA vs Extractor gauge by gas injection Pressure in RGA port and the gauge port should be the same due to test chamber symmetry Pressure in the test chamber varied between ~10-10 mbar to 10-6 mbar. Pressure recorded for the gauge and RGA peaks At mbar to 10-8 mbar the injected gas might be not dominant in gas spectrum 11
12 NEG-1 rig test chamber Pumping RGA Gas injection Extractor Gauge 12
13 Analysis of the measurements RGA spectrum: there is always a mixture of gases Assuming that the calibration coefficients a i for the gauge and b i for the RGA, the measured currents are: Then I ( i) a P; I ( i) b P g i i RGA i i a I a P i I () i g i i RGA b i i i Details of calibration coefficient matrix calculation was covered by B. Jenninger in his talk in Session 2 and in our paper in J. Vac. Sci. Technol. A 26 (2008), p
14 Typical results The RGA calibration coefficients are normalised to a Nitrogen coefficient and compared with referenced coefficients for an ionisation gauge Gas Type H 2 CH 4 CO N 2 Ar CO 2 Gauge a i /a N2 RGA-FAR b i /b N2 RGA-SEM b i /b N
15 RGA calibration check vs. extractor gauge Injection: H 2 CO CO 2 CH 4 H2 CO CO2 CH P ext measured P ext calculated H2 mas s15 mass 16 CO Ar CO2 mass 12 mass 14 WS-63, September 2010, Ávila, Spain 15/40
16 Calibration of Extractor Gauge and RGA s Probability of Ionisation for ionisation gauges H 2 = 0.44 CO = 1.05 CH 4 = 1.6 CO 2 = 1.4 Ar = 1.2 Compared to our measurements H 2 differs by a factor of 4.3 CO differs by a factor of 2.4 CH 4 differs by a factor of 3.0 CO 2 differs by a factor of 4.7 Ar differs by a factor of
17 Linearity over pressure ranges 2 RGAs and extractor gauge measured simultaneously Injection of H 2, CH 4, CO and CO 2 Relative sensitivity normalised to one measured at 10-7 mbar +5% -10% -10% +60% 17
18 SEM vs. FAR Calibration with a Faraday cap is quite stable no drift over ~3 years detected. SEM is calibrated against a Faraday cap at P > mbar Checked during each experiment when both SEM and FAR used Residual sensitivity coefficients need to be corrected every ~3 months WS-63, September 2010, Ávila, Spain 18
19 Electron Stimulated Desorption (ESD) Another factor to consider with RGA data at low pressure is the influence of ESD from the ion source. Typical ESD generated peaks include: H +, O +, F +, 35 Cl + and 37 Cl + If unaccounted for it can lead to false conclusions in interpretation of RGA data. This is particularly important when considering the influence of Oxygen containing species when activating GaAs photocathodes. These species are considered a contaminant and can kill the QE of a GaAs surface. Suggestions are that partial pressures of < mbar for such species is required. 19
20 Partial Pressure (mbar) Partial Pressure (mbar) Influence of ESD Peaks in RGA Data Gas phase and ESD species have different energies which allow separation between the two. RGA Scan from Outgassing System RGA Scan from Outgassing System Mass 19 is the dominant peak ESD Peaks only Mass Mass (amu) 20
21 Partial Pressure (mbar) 10-9 RGA Scan showing the influence of ESD peaks Mass (amu)
22 P [mbar] GaAs photocathode studies 3E-12 Before injection 2.5E-12 2E-12 RGA scan showing ideal vacuum system 1.5E-12 1E-12 5E
23 QE (a.u.) Pressure (mbar) QE (a.u.) Pressure (mbar) QE (a.u.) Pressure (mbar) GaAs lifetime studies, purposely poisoning the cathode QE Pressure Elapsed time (s) CO exposure 3.0x x x x x x10-11 CO injection 7E-12 6E-12 5E-12 4E-12 3E-12 2E-12 1E E QE Pressure CO 2 exposure 3.0x x x x x10-10 CO 2 injection 2.5E-12 2E E-12 1E-12 5E x E Elapsed time (s) QE Pressure Elapsed time (s) O 2 exposure 3.0x x x x x x10-11 O 2 8E-12 6E-12 4E-12 2E E-12 23
24 Conclusions Vacuum science requires: a quantitative RGA: P i = f (I i ) = I i /C(I i,p tot,i e,e ion,...) Stabile, including XHV With high traceability Low outgassing In-situ RGA calibration can be performed against a total pressure UHV/XHV gauge when gas injection is available In-situ RGA calibration check might be performed without a gas injection Ex-situ RGA calibration would bring more confidence in the RGA performance 24
Development of Cleaning and Processing Techniques for UHV/XHV at Daresbury Laboratory Keith Middleman
Development of Cleaning and Processing Techniques for UHV/XHV at Daresbury Laboratory Keith Middleman ASTeC Vacuum Science Group STFC Daresbury Laboratory Accelerators + Vacuum Particle accelerators come
More informationTotal Pressure Measurement Using Residual Gas Analyzers
Total Pressure Measurement Using Residual Gas Analyzers M. Maskell, Old Dominion University; G. Myneni, P. Adderley, Jefferson Lab; G. Brucker, Stanford Research Systems; C. Day, Forschungszentrum, Karlsruhe
More informationGun Vacuum Performance Measurements February 2003
JLAB-TN-03-013 Gun Vacuum Performance Measurements February 2003 Marcy Stutzman Abstract. The vacuum in the polarized photoguns at Jefferson Lab is crucial to long lifetime operation of the polarized source.
More informationGo/no-go test for the acceptance of components for unbaked vacuum sectors
Go/no-go test for the acceptance of components for unbaked vacuum sectors Paolo Chiggiato (Thanks to Sophie Meunier for the preparation of the presentation) Introduction CERN policy: All vacuum components
More informationResidual Gas Analysis Systems for Industry
HIDEN RC SYSTEMS QUADRUPOLE MASS SPECTROMETERS FOR RGA, GAS ANALYSIS AND PROCESS MONITORING The HAL RC systems are designed for RGA, gas analysis and process monitoring applications including leak detection,
More informationLeaks, contamination and closed loop control how RGAs make coating processes more profitable
Leaks, contamination and closed loop control how RGAs make coating processes more profitable Backup: What s an RGA Residual Gas Analyzer An instrument that is measuring the gas composition in a chamber
More informationThe Principles of Vacuum Technology
The Principles of Vacuum Technology Vacuum Terminology Vacuum units Vacuum regimes How to measure vacuum. Gauge designs. How to create vacuum Pump classifications and designs UHV compatibility considerations
More informationHigh Performance Mass Spectrometers
High Performance Mass Spectrometers For Vacuum and Semiconductor Process Monitoring Detailed product information / introduction MOCVD semiconductor processing Applications: XXCVD/MOCVD XXPECVD XXALD XXSputtering
More informationBEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION
BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the
More informationLCLS Heavy Met Outgassing Tests * K. I. Kishiyama, D. M. Behne Lawrence Livermore National Laboratory
UCRL-TR-223160 LCLS-TN-06-12 July 12, 2006 LCLS Heavy Met Outgassing Tests * K. I. Kishiyama, D. M. Behne Lawrence Livermore National Laboratory Abstract A Heavy Met that is 95% tungsten, 3% nickel and
More informationISO/TC 112/WG 2 N 49. ISO TS QMS WD v6. Date of document Expected action Comment Due Date Background
ISO/TC 112/WG 2 N 49 ISO/TC 112/WG 2 Vacuum instrumentation E-mail of Secretary: juergen.eisenreich_vdma.org@livelink.din.de Secretariat: DIN ISO TS 20175 QMS WD v6 Date of document 2016-05-02 Expected
More informationMETHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL PRESSURE AND COMPOSITIONAL ANALYSIS
METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL PRESSURE AND COMPOSITIONAL ANALYSIS Robert E. Ellefson REVac Consulting Dayton, OH 45459 USA robert.ellefson@sbcglobal.net ERMP IND12 QMS Workshop Bled,
More informationChoosing a Quadrupole Gas Analyzer Application Note #9
Choosing a Quadrupole Gas Analyzer Application Note #9 Modern-day contamination control requirements for gas phase processes are constantly pushing the limits of performance of quadrupole gas analyzers.
More informationCommercial Practice Test Method Internal Vapor Analysis of Hermetic Devices
Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The
More informationOutgassing. the hows and whys of measurement. Ron Reid ASTeC Vacuum Science Group STFC Daresbury Laboratory UK
Outgassing the hows and whys of measurement Ron Reid ASTeC Vacuum Science Group STFC Daresbury Laboratory UK Vacuum System Performance Vacuum Equation P = Q S Where P is the base pressure in the system
More informationA NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES
A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES F. Papa 1, J. Brindley 2, T. Williams 2, B. Daniel 2, V. Bellido-Gonzalez 2, Dermot Monaghan
More informationA Mass Spectrometry Method for Water Vapour Transmission Rate (WVTR) testing to 10-6 gm/m 2 /day, utilising Deuterium Oxide
A Mass Spectrometry Method for Water Vapour Transmission Rate (WVTR) testing to 10-6 gm/m 2 /day, utilising Deuterium Oxide Stephen Swann Development Scientist www.vgs-plasticelectronics.com WVTR Illustration
More informationQuadruple mass spectrometers (transducer type)
Quadruple mass spectrometers that meet the need for sensitive analysis and low outgassing which were specifically designed to be controlled by PCs Quadruple mass spectrometers (transducer type) M-070QA-TDF,
More informationIUVSTA WORKSHOP WS-63 Surface Phenomena Limiting Pressure in Vacuum Systems REPORT TO IUVSTA. prepared by JOSÉ L. DE SEGOVIA
IUVSTA WORKSHOP WS-63 Surface Phenomena Limiting Pressure in Vacuum Systems INTRODUCTION REPORT TO IUVSTA prepared by JOSÉ L. DE SEGOVIA on behalf of the SPANISH VACUUM SOCIETY Madrid,February 1, 2011
More informationResist round robin ELETTRA
Resist round robin outgassing @ ELETTRA Trieste - Italy M. Bertolo, P. Lacovig, L.Rumiz, D. Giuressi, R. Sergo, A. Stolfa Experimental setup (1) 100 mm linear translator Sample manipulator Mask manipulator
More informationat NIST: ultra-low outgassing rates
Traceable outgassing measurements at NIST: ultra-low outgassing rates and new materials James Fedchak Thermodynamic Metrology Group Sensor Science Division Physical Measurement Laboratory Vacuum Project:
More informationApplications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples
Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Outline of Presentation Introduction Process Gas Analysis Magnetic Sector Mass Spectrometer Standard
More informationRecent steps toward traceability for partial pressure and outgassing measurements Karl Jousten, PTB, Berlin
Recent steps toward traceability for partial pressure and outgassing measurements Karl Jousten, PTB, Berlin 1. Problems of calibrating QMS 3. Toward traceability in EMRP IND12 2. Toward standardization
More informationThe Broadest Gauging Line in the Industry
The Broadest Gauging Line in the Industry Vacuum Measurement Varian s broad line of vacuum gauge controllers and gauge tubes are the most reliable, accurate, and economical means of measuring, monitoring,
More informationOMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.
OMNISTAR / THERMOSTAR The efficient solution for gas analysis. Intelligent software. Wide range of applications. OMNISTAR /THERMOSTAR The efficient solution for gas analysis. Intelligent software. Wide
More informationChapter 6 Leak detection
Chapter 6 Leak detection 1 2 1. Leak rate q L = V p t mbar l s -1 A leak rate of 1 mbar l s -1 exists in a closed vessel having a volume of 1 liter when the pressure increases by 1 mbar within one second,
More informationVacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007
Vacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007 Vacuum increases the mean-free-path of gas molecules. Vacuum prevents chemical reaction. Vacuum removes contaminants from surfaces. Vacuum
More informationHiden QIC Series Gas Analysers Real-time gas analysis systems
Hiden QIC Series Gas Analysers Real-time gas analysis systems vacuum analysis surface science plasma diagnostics gas analysis Application Focussed Options Multistream valves 20, 40 and 80 way versions
More informationAchieving Ultra-High Vacuum without (in-situ) Bakeout
Achieving Ultra-High Vacuum without (in-situ) Bakeout Matthew Cox Diamond Light Source, UK WS 63 Ávila September 2010 1 What is Diamond Light Source? The UK national synchrotron facility 3 rd generation
More informationThe Helium Leak Detector
The Helium Leak Detector Helium Leak Detector Main Components The main components of a helium leak detector are: 1. The analyzed, which enables to separate the tracer gas from other gases inside leak detector.
More informationVacuum. Methods for in situ QMS calibration for partial pressure and composition analysis. Robert E. Ellefson * abstract
Vacuum 101 (2014) 42e42 Contents lists available at ScienceDirect Vacuum journal homepage: www.elsevier.com/locate/vacuum Methods for in situ QMS calibration for partial pressure and composition analysis
More informationGraphite and C-C materials for UHV applications
Graphite and C-C materials for UHV applications Prepared by J.M. JIMENEZ AT/VAC/SL Section Graphite Outgassing Measurements for UHV Applications Main topics What did we learn from past experiences? Static
More informationThe use of remote plasma emission spectroscopy as an alternative method of gas sensing for direct monitoring of vacuum processes
Enlighten Conference and Vacuum Symposium 12 th October 2017 The use of remote plasma emission spectroscopy as an alternative method of gas sensing for direct monitoring of vacuum processes Joe Brindley,
More informationLOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12
LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/ This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the rates with which selected
More informationLow-Power Atmospheric Gas Sampling System Based on ART MS Sensor and NEG-Ion Pump
8 th Harsh Environment Mass Spectrometry Wokshop Low-Power Atmospheric Gas Sampling System Based on ART MS Sensor and NEG-Ion Pump Sept 22, 2011 Gerardo A. Brucker and G. Jeffery Rathbone Granville-Phillips
More informationVacuum Science and Technology in Accelerators
Vacuum Science and Technology in Accelerators Ron Reid Consultant ASTeC Vacuum Science Group (ron.reid@stfc.ac.uk) Session 5 Materials Properties Relevant to Vacuum Aims To understand which mechanical
More informationImproving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer
Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer Ben D. Gardner, Phillip M. Erwin, Wai Tak Lee, Amber M. Tissandier, Souzan M. Thoresen Hamilton Sundstrand
More informationLeak Detection. Kirsten Zapfe, DESY CAS Vacuum in Accelerators Platja D Aro, May 21, Special thanks to S. Holm, N. Mildner and A.
Leak Detection, DESY CAS Vacuum in Accelerators Platja D Aro, May 21, 2006 Special thanks to S. Holm, N. Mildner and A. Wagner Outline Introduction Leak Rate and Leak Types Leak Detection Methods Helium
More informationCombining NEG pumps and an XHV BNNT cryopump
Combining NEG pumps and an XHV BNNT cryopump Marcy L. Stutzman, Anahí Segovia Miranda, Veronica Over, Philip Adderley, Matt Poelker Thomas Jefferson National Accelerator Facility Newport News, VA 23601,
More informationAssembling NEG, Ion Pump and Bakeable BNNT Cryopump System to Reach XHV
Assembling NEG, Ion Pump and Bakeable BNNT Cryopump System to Reach XHV Anahí Segovia Miranda Universidad Autónoma de Zacatecas Unidad Académica de Física Marcy Stutzman Thomas Jefferson National Accelerator
More informationMeasuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability
Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability R. C. Kullberg*, D. J. Rossiter** Oneida Research Services, Inc., 8282 Halsey Road, Whitesboro, NY13492 Abstract Many
More informationShort Introduction to Cryo-Pumps and Refrigerators. Dr Graham Rogers Leybold UK Ltd.
Short Introduction to Cryo-Pumps and Refrigerators Dr Graham Rogers Leybold UK Ltd. Cryo Pumps Topics Physical principles of cryo-pumping Design and control of modern refrigerator cryo-pumps UHV and XHV
More informationA novel class of Non Evaporable Getter (NEG) alloys with lower hydrogen equilibrium pressure and reduced outgassing properties
A novel class of Non Evaporable Getter (NEG) alloys with lower hydrogen equilibrium pressure and reduced outgassing properties P. Manini, FabrizioSiviero, A. Gallitognotta, L. Caruso, T. Porcelli,G. Bongiorno,
More informationAGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY
AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY Agilent Vacuum Products: Our Committment to Quality Agilent s commitment to quality and Measurement s philosophy Reference Standards (PNEUROP, ISO)
More informationPressure & Vacuum Measurement. Series 370. Solutions. Stabil-Ion Vacuum gauge and Controller
w w w. m k s i n s t. c o m Pressure & Vacuum Measurement Solutions Series 370 Stabil-Ion Vacuum gauge and Controller Features & Benefits All-metal, rack-mount controller for Stabil-Ion and Convectron
More informationThis educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum.
This educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum. Specifically, today s talk will cover: Brief review of Vacuum Fundamentals Applications Using Rough Vacuum Rough
More informationMass Spec will not Autotune
Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems
More informationIntroduction of Vacuum Science & Technology. Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project.
Introduction of Vacuum Science & Technology Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project. 1 What is a vacuum? 760 mm Hg Vacuum ATM A vacuum is defined as less than
More informationMASSIVE TITANIUM SUBLIMATION PUMPING IN THE CESR INTERACTION REGION *
MASSIVE TITANIUM SUBLIMATION PUMPING IN THE CESR INTERACTION REGION * N.B. Mistry, R. Kersevan and Yulin Li, Laboratory of Nuclear Studies, Cornell University, Ithaca, NY 153 USA Abstract * The residual
More informationTechnical Specifications of Hydrogen Isotope Handling and Recovery System
SECTION - C TECHNICAL SPECIFICATIONS OF STORES AND DRAWINGS. Technical Specifications of Hydrogen Isotope Handling and Recovery System INSTITUTE FOR PLASMA RESEARCH GANDHINAGAR, GUJARAT 382428 ANNEXURE-I
More informationTests for High Voltage & Vacuum of the 2 nd 500 kv DC Gun at KEK
Tests for High Voltage & Vacuum of the 2 nd 500 kv DC Gun at KEK OLAV IV @NSRRC 3 rd April 2014 Masahiro Yamamoto KEK 2 nd DC-gun Development Team T. Miyajima, Y. Honda, T. Uchiyama, M. Kobayashi, Y. Saito,
More informationSelection of Materials and Processes for Space Use. Ruben Edeson, Mechanical Engineer, SSTD.
Selection of Materials and Processes for Space Use Ruben Edeson, Mechanical Engineer, SSTD. Materials Selection for Space Mechanical properties, ie specific stiffness Thermal and optical properties Cost
More informationOutgas of Methane from NEG Coating
WS63: SURFACE PHENOMENA LIMITING PRESSURE IN VACUUM SYSTEM Outgas of Methane from NEG Coating Chia-Mu Cheng ( 鄭家沐 ) Vacuum Group NSRRC, Taiwan 18 /09/2010 e-mail : chiamucheng@nsrrc.org.tw 1 Application
More informationLab 1c Isentropic Blow-down Process and Discharge Coefficient
058:080 Experimental Engineering Lab 1c Isentropic Blow-down Process and Discharge Coefficient OBJECTIVES - To study the transient discharge of a rigid pressurized tank; To determine the discharge coefficients
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople
More informationCanadian Light Source Inc. Vacuum Component Leak Test Technical Procedure
Canadian Light Source Inc. Vacuum Component Leak Test Technical Procedure CLSI Document 8.7.33.2 Rev. 2 Date: 2006-03-08 Copyright 2006, Canadian Light Source Inc. This document is the property of Canadian
More informationTOTAL AND PARTIAL PRESSURE MEASUREMENT IN VACUUM SYSTEMS
TOTAL AND PARTIAL PRESSURE MEASUREMENT IN VACUUM SYSTEMS J.H. LEeK Emeritus Professor Department of Electrical Engineering and Electronics University of Liverpool Blackie Glasgow and London B1ackie & Son
More informationExtreme High Vacuum: The Need, Production, and Measurement
Extreme High Vacuum: The Need, Production, and Measurement, Philip Adderley, Matt Poelker (Jefferson Lab) Polarized Electron Gun Group Newport News, Virginia Run by JSA for the US DOE What is XHV Extreme
More informationLOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary
ADH 1/7/014 LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the
More informationUltrahigh Vacuum in Superconducting Synchrotrons A.Smirnov, JINR, Dubna, Russia
Ultrahigh Vacuum in Superconducting Synchrotrons A.Smirnov, JINR, Dubna, Russia Topics: - Pumping & outgasing of hydrogen - Cryogenic pumps & non evaporable getters for UHV - Superconducting accelerators:
More informationOXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:
OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle
More informationMajor Design Features
Major Design Features 3Flex Features Design Benefits Discussion 316 SS fittings and pneumatically actuated, hard-seal valves Provides virtually leak-free gas management with the lowest outgassing rate
More information7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications
GE Sensing & Inspection Technologies 7250 Sys Ruska Multi-Range Pressure Calibration System Features Fully integrated multi-range pressure test and calibration systems Select either an eight range or the
More informationAPPLICATION NOTE. GC Integrated Permeation Device
GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords:
More informationContamination Removal of EUVL masks and optics using 13.5-nm and 172-nm radiation
Contamination Removal of EUVL masks and optics using 13.5-nm and 172-nm radiation Takeo Watanabe, Kazuhiro Hamamoto and Hiroo Kinoshita University of Hyogo IEUVI Resist & Contamination TWG Oct.19, 2006
More informationAdvantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types
19 th World Conference on Non-Destructive Testing 2016 Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types Klaus HERRMANN 1, Daniel WETZIG
More informationOOC RGA. Ion Pump 400 l/s. 8" ports with baked viton o-ring seals. Welsh rotary roughing pump 650 l/s manual isolation valve.
LASER INTERFEROMETER GRAVITATIONAL WAVE OBSERVATORY - LIGO - CALIFORNIA INSTITUTE OF TECHNOLOGY MASSACHUSETTS INSTITUTE OF TECHNOLOGY Technical Note LIGO-T000126-00- R 11/20/2000 Vacuum pressure at the
More informationVacuum Systems Operation EiC Training and Refresher Course 26 th April 2007
Vacuum Systems Operation EiC Training and Refresher Course 26 th April 2007 Contents 1. Introduction 2. Pumps and pumping on JET Primary Volumes Primary Pumps Secondary Pumps Vacuum Diagnostics Vessel
More informationLINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER
LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER Michael T Tang, Ph.D. Grace Feng Greg Merideth Rui Huang Matheson Gas Applied Lab
More informationResist Outgassing and its Role in Optics Contamination
Resist Outgassing and its Role in Optics Contamination Gregory Denbeaux, Rashi Garg, Kim Dean* College of Nanoscale Science and Engineering University at Albany *SEMATECH, Austin, TX Advanced Materials
More information(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.
Name: 1. You are asked to measure the percent iron in a steel sample you believe to be about 50% Fe by mass. You dissolve four replicate samples in nitric acid and dilute them by mass to a concentration
More informationThis educational seminar discusses creating, measuring, and troubleshooting Ultra-High Vacuum ( Torr) systems.
This educational seminar discusses creating, measuring, and troubleshooting UltraHigh Vacuum ( 10 8 10 12 Torr) systems. Specifically, today s talk will cover: Brief review of High Vacuum (Characteristics,
More informationUser guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection
User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind
More informationDSC 204 HP Phoenix. High-Pressure Differential Scanning Calorimetry up to 15 MPa Method, Technique and Applications
DSC 204 HP Phoenix High-Pressure Differential Scanning Calorimetry up to 15 MPa Method, Technique and Applications DSC 204 HP Phoenix The ability to apply pressure improves the capabilities of Differential
More informationLASER INTERFEROMETER GRAVITATIONAL WAVE OBSERVATORY - LIGO - CALIFORNIA INSTITUTE OF TECHNOLOGY MASSACHUSETTS INSTITUTE OF TECHNOLOGY
LASER INTERFEROMETER GRAVITATIONAL WAVE OBSERVATORY - LIGO - CALIFORNIA INSTITUTE OF TECHNOLOGY MASSACHUSETTS INSTITUTE OF TECHNOLOGY Document Type LIGO-T980008-01- W 2/18/98 Bake Oven Requirements and
More informationGas Release from Solids, Materials for Vacuum and Leak Detector
VACUUM AND PLASMA SCIENCE AND TECHNOLOGY W9 Gas Release from Solids, Materials for Vacuum and Leak Detector Ming-Show Wong Nov. 2015 Gas Release from Solids Vaporization Thermal Desorption First Order
More informationSupporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial
Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle,
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions
More informationRevision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline
Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline This training course outline is intended to cover the following: Introduction to vacuum Measurement Lubricated rotary pumps
More informationVacuum System at IUAC
Vacuum System at IUAC A. Mandal Inter-University Accelerator Centre, Aruna Asaf Ali Marg, Post Box - 10502, New Delhi 110067, Fax:011-26893666. email: mandal@iuac.res.in Abstract: Vacuum technology is
More informationBrooks TMF-Zone 1 Models 5816-Ex/5864-Ex
Data Sheet DS-TMF-586Ex-MFC-eng TMF-Zone Brooks TMF-Zone Models 586-Ex/5864-Ex Customer Benefits Assured process accuracy and repeatability Long-term reliability Reduced maintenance CE certified Improved
More informationVacuum Simulations of the KATRIN Experiment
Vacuum Simulations of the KATRIN Experiment Marcel Krause (KATRIN collaboration) 31.03.2014 KIT University of the State of Baden-Wuerttemberg and National Research Center of the Helmholtz Association www.kit.edu
More information29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies
IMPROVED CALIBRATION AND OPERATION OF A STABLE-GAS QUANTIFICATION MANIFOLD AS PART OF A RADIOXENON COLLECTION SYSTEM ABSTRACT Richard M. Williams, James C. Hayes, and Randy R. Kirkham Pacific Northwest
More informationSorbent Pen Heavy & Labile SVOC s Taints, Off-Flavors, Odors
Revolutionary extraction technology quantitative by design. Say Goodbye to SPME... and hello to Sorbent Pens! A revolutionary new extraction technology. Sorbent Pen Heavy & Labile SVOC s Taints, Off-Flavors,
More informationOutgassing, photoablation and photoionization of organic materials by the electron-impact and photon-impact methods
Outgassing, photoablation and photoionization of organic materials by the electron-impact and photon-impact methods Grace H. Ho*, Yen-H. Huang, Hung-M. Lin, Chih-H. Shao, Chen-Y. Yeh, and Jia-J. Sung Department
More informationOperating Procedures for Metal Evaporator I
Operating Procedures for Metal Evaporator I Metal Evaporator I is intended as a tool and a training device. Understanding the operation of this equipment should give you a basic knowledge of vacuum and
More information159 mm 154 mm. Subject to changes 10/2012
Low-pressure calibrator LPX calibrator medium: air / up to 10 bar KELLER pressure calibrators are highly accurate calibration instruments which are available for the following pressure ranges: -0,85 10
More informationThe new PTB standard for dynamic vacuum pressures
The new PTB standard for dynamic vacuum pressures Karl Jousten, Physikalisch-Technische Bundesanstalt, Berlin, Germany 1. Idea and design of the standard 2. Development 3. Pressure approximation and experiments
More informationTitle: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator
Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi
More informationPHOENIX L300i Best-in-class Helium Leak Detectors
180.85.02 Best-in-class Helium Leak Detectors The perfect Helium leak detector matching your demands The helium leak detector opens up new dimensions of productivity and reliability for your applications.
More informationTest pumps. Pressure sources. Good reasons for proper and reliable calibration
Test pumps Pressure is one of the most commonly measured quantities in engineering, which is why exact and reliable pressure measurement is especially important. However, the characteristics of the even
More informationTitle: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator
Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality
More informationDevelopment of a high vacuum sample preparation system for helium mass spectrometer
Journal of Physics: Conference Series Development of a high vacuum sample preparation system for helium mass spectrometer To cite this article: P Kumar et al 2012 J. Phys.: Conf. Ser. 390 012056 View the
More informationSAES Packaging service. making innovation happen, together
SAES Packaging service MEMS Packaging service description Facility and operations Capabilities and future prospective Choose your type of service : Fast prototyping Process developments Packaging foundry
More informationThe instrument should have a mass range from 2 to 2000 m/z or better Interface
Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for
More informationReal-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy
Real-time Analysis of Industrial Gases with Online Near- Infrared Branch 5 Petrochemical, Biofuels Keywords Process, NIR, Gases, Gas Cell, Ethylene, Carbon Dioxide, Propane, Butane, Acetylene Summary Near-infrared
More informationCombination of a getter pump with turbomolecular pumps in UHV applications
Vacuum 55 (1999) 27}37 Combination of a getter pump with turbomolecular pumps in UHV applications Roberto Giannantonio*, Magda Bovisio, Andrea Conte SAES Getters S.p.A., Viale Italia, 77-20020-Lainate
More informationQuantitative study of sniffer leak rate and pressure drop leak rate of liquid nitrogen panels of SST-1 tokamak
Journal of Physics: Conference Series Quantitative study of sniffer leak rate and pressure drop leak rate of liquid nitrogen panels of SST-1 tokamak To cite this article: F S Pathan et al 2008 J. Phys.:
More informationADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM. OPENING DATE & TIME: January 3, 2018 at 3:00PM
Procurement Services ADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM ITB NUMBER: 1708MC OPENING DATE & TIME: January 3, 2018 at 3:00PM ITB TITLE: X-Ray Photoelectron Spectroscopy System ADDENDUM
More informationPlasma Sources and Feedback Control in Pretreatment Web Coating Applications
Plasma Sources and Feedback Control in Pretreatment Web Coating Applications Joseph Brindley, Benoit Daniel, Victor Bellido-Gonzalez, Dermot Monaghan Gencoa Ltd., Physics Rd, L24 9HP Liverpool, UK (+44)
More information