1 OPERATOR MANUAL Table Top and Convertible Anesthesia Machine Models: M3000 and M3100 M3000 M3100 (SHOWN WITH OPTIONAL EQUIPMENT) SE Hwy 212 Clackamas, Oregon (503) Local and International Toll Free
2 Table of contents Page 3 Product Examination and Preparation for Use Page 4 M3000 Machine Diagram and Part Numbers Page 5... M3100 Stand Model Diagram Page 6-8 M3100 Additional Assembly Page 9 E-Tank Manifold Installation Instructions Page 10..Machine Set-up Instructions Page Accessory Options: Multi-Station Manifolds - 12, 13 Induction Chambers 14, 15 Breathing Circuit 16, 17 Rodent Heated Surgical Station with Downdraft 18, 19 Multi-Station Surgical Board 20, 21 Waste Gas Interfaces Page 26..Warranty / Return Policy Page 27..Shipping Policy Page 28..Documentation 2
3 Examination and Preparation for Use Thank you for ordering a Scivena Scientific Anesthesia Innovations anesthesia machine! We are delighted to have you as a Scivena Scientific Anesthesia Innovations customer and want you to be completely satisfied with your purchase. Please inspect the contents of your order to see if everything is as you expected. Should anything not be exactly right or if anything was damaged in shipping, please contact your sales representative right away for help. Our goal is to make your new anesthesia machine as easy to use and care for as possible. This device is meant to be operated under the normal surveillance and control of a veterinarian trained in its use. However, you need to know more about this device than just how to operate it. Please read this manual in its entirety before using the anesthesia machine! If you have any comments or questions, we welcome the opportunity to address them. Please contact us directly at Thank you! Brian Lawson President, Scivena Scientific Anesthesia Innovations Designed and Manufactured With Pride In Oregon, USA 3
4 M3000 Diagram and Part Numbers FIGURE #1 A I F J H E B D G C FRONT BACK A. Main Flow Meter- Indicates precise amount of carrier gas entering the system. B. Flow Control Knob- adjusts precise amount of carrier gas entering the anesthesia machine. C. Flush Valve- flushes pure carrier gas (bypasses the vaporizer) through the system. Flow restricted to 30 LPM for subject safety. D. Optional Vaporizer- vaporizes the anesthetic agent from a liquid to a gaseous state. Delivers anesthetic agent at precise percentage concentration independent of temperature and flow rate. New Tec 3 (p/n VAP3000 Isoflurane, p/n VAP3100 Sevoflurane). E. Common Gas Outlet with 15mm female adaptor. Port through which carrier and anesthetic gas mixture is delivered through the anesthesia system. (See page for induction chamber and breathing circuit options). F. Carry Handle G. Carrier gas connection (O2, Med Air or other)- DISS male thread a. NOTE: optional quick disconnect fitting can be installed. Please call for options. H. Optional Waste Gas Interface (p/n EVC630) I. Accessory mounting receptacles (2). J. Optional Balanced Flow Multi-Station Manifold (p/n RES535) 4
5 M3100 Stand Model Diagram and Additional Assembly: M3000 Anesthesia Machine shown with optional Isoflurane Vaporizer (VAP3050), Waste Gas Interface (ECV630) and Balanced Flow Multi-Station Manifold (RES535) Top Platform Accessory Basket Dual E-Tank Manifold (optional) Center Post (2 round tubes) Five Leg Base 5
6 Attaching the Base Post (See FIGURE #2) 1. Unpack the 5 leg base and set it on the floor 2. Unpack both 2 round tubes 3. Assemble the base tube onto the base post. 4. Install and tighten the screws supplied into the post. NOTE: The post should fit flush onto the base, if the holes do not align, try the other end of the post. FIGURE #2 6
7 Attaching the Top Post (See FIGURE #3) 1. Assemble the top post over the bottom post pin 2. Install and tighten the screws supplied into the post. NOTE: The post should fit flush onto the base post, if the holes do not align, try the other end of the post. FIGURE #3 7
8 Attaching the Top (See FIGURE #4) 1. Assemble the top platform onto the post. 2. Install and tighten the screws supplied into the top post. FIGURE #4 8
9 Installing the Optional E-Tank Manifold (See FIGURE #5) (P/N MA2000) 1. Align the holes in the E tank manifold with the two holes located in the post. 2. Insert the 2 screws and tighten firmly. NOTE: Do not over tighten the e tank yoke knobs ( T shaped handles) * If the e tank leaks, change the seal. CAUTION! NEVER USE PETROLEUM PRODUCTS ON OR NEAR THE CARRIER GAS SUPPLY WHEN USING OXYGEN AS THE CARRIER GAS! FIGURE #5 9
10 M3000/M3100 Set Up: 1. Place M3000 on level surface or, for M3100 model, on stand and secured with included pins. 2. Connect selected carrier gas hose to DISS male thread connection on back of the machine (G) and connect hose to carrier gas supply (SEE FIGURE #1). a. Common carrier gases include Oxygen (O2), Medical Air (Med Air), and other combinations of gases. **CAUTION! NEVER USE PETROLEUM PRODUCTS ON OR NEAR THE CARRIER GAS SUPPLY WHEN USING OXYGEN AS THE CARRIER GAS! 3. Connect any optional accessories (see pages 11-25). 4. Connect waste anesthetic gas management system. 5. Fill vaporizer with appropriate anesthetic agent. 6. Make sure carrier gas supply is on. M3000/3100 is now ready for use 10
11 OPTIONAL ACCESSORY INSTRUCTIONS Balanced Flow Multi-Station Manifolds Induction Chambers Breathing Circuits Multi-Station Procedure Board Heated Rodent Surgical Station Waste Gas Interfaces 11
12 Balanced Flow Multi-Station Manifold User Instructions RES535-2 Port RES536-4 Port RES537-6 Port The Scivena Scientific Balanced Flow Multi-Station Manifold System is designed so that every port on the manifold gets equal fresh gas delivery independent of the accessory being used. The manifold can accommodate up to six accessory ports. C B D A E F Shown in four port configuration RES356 A. Accessory Receptacle (M3000/M3100) or Mounting Bracket (all other machines) B. Manifold post C. Manifold body with ports. D. Port on/off stopcocks a. Port is ON when toggle is in vertical position. E. Main fresh gas inlet with 15mm female adaptor. F. 15mm Male common outlets (up to six) 12
13 Set-up: Use: 1. Mount the manifold to the anesthesia machine utilizing the included bracket (A) by inserting the post of the manifold (B) into the bracket and tightening. 2. Connect the main fresh gas inlet adaptor (E) to the common outlet of the vaporizer. 3. Connect your accessories (induction chamber, breathing circuits etc.) to the ports on the manifold body (F) noting which number on the manifold body coordinates with an accessory. **DO NOT TURN ON THE MAIN ANESTHESIA FLOW METER UNTIL AT LEAST ONE STOPCOCK ON THE MULTI-STATION MANIFOLD IS IN THE ON POSITION. Otherwise, pressure can build up in the system which can damage the vaporizer.** 1. Connect and activate the waste anesthetic gas system to your accessories. 2. When ready to use any accessory, open the stopcock to that accessory by turning the toggle to the ON position. 3. With one or more toggle switches in the ON position, it is now safe to turn on the main anesthesia flow meter. 4. The Multi-Port Manifold will now deliver the same rate of fresh gas flow to each open port. a. Example: The main flow meter on the anesthesia machine is set to 3 LPM, and two ports are open on the Multi-Port Manifold, each port will deliver 1.5 LPM fresh gas flow. It is recommended that when the anesthesia machine is not in use, that at least one port on the Multi-Station Manifold remains open. This will prevent accidental pressure build up in the anesthesia system if the main flow meter is turned on. 13
14 Sealed Induction Chamber User Instructions RES644 3 Liter RES645 5 Liter RES646 7 Liter The Scivena Scientific Sealed Induction Chamber allows for safe and effective anesthetic induction of rodents while protecting the user from waste anesthetic gas exposure. A B D E C A. Chamber lid with airtight seal gasket B. Chamber body with seal clamps C. 15mm female gas inlet (O2/gas or CO2) D. 22mm male waste gas evacuation port a. 19mm EVAC ports available E. Five foot extension hose with 15mm female adaptors 14
15 Set-Up: Use: 1. Place chamber on level surface. 2. Connect one end of the extension hose to the common outlet of the anesthesia machine, Scivena Scientific Multi-Port Manifold outlet port, or any diverter system with a male common outlet on the end of the tubing, utilizing the 15mm male adaptor on the extension hose (E). 3. Connect opposite end of the extension hose to the 15mm gas inlet on the induction chamber (C). 4. Connect 22mm EVAC tubing to the 22mm male waste gas evacuation port on the induction chamber (D). a. Connect 22mm tubing to waste gas management system. b. **The Sealed Induction Chamber must be used with either a passive waste gas management system, or if using an active waste gas vacuum, a Scivena Scientific Waste Gas Interface (RES505 and EVC630). 5. Remove the lid of the chamber by pulling the blue secure clamps outward and down to release. 1. Make sure the waste anesthetic gas system is in place and functioning. 2. Before turning on any anesthetic gas, place the subject into the chamber and secure the lid using the blue clamps- the chamber is now sealed. 3. Turn on the fresh gas flow using the main flow meter on the anesthesia machine. a. Recommended fresh gas flow for induction chambers is 2 LPM. 4. Turn the vaporizer on to desired concentration. a. Recommended vaporizer concentration for induction chambers is 2-3%. 5. Once subject is anesthetized, turn off the vaporizer leaving the main flow meter on to 2LPM. 6. Using the O2 flush valve on the anesthetic machine, flush pure O2 through the chamber to remove remaining anesthetic gas. Recommended purge times: RES644-3 Liter- 90 seconds RES645-5 Liter- 105 seconds RES646-7 Liter-115 seconds 7. Once chamber is sufficiently purged, turn off fresh gas flow to the chamber, open chamber lid, remove subject and replace chamber lid. 15
16 Rat and Mouse Non-Rebreathing Circuit User Instructions RES500 The Rat and Mouse Non-Rebreathing Circuit is specially designed to both deliver fresh anesthetic gas to rats and mice, and effectively remove waste anesthetic gases- all in one, complete system. When used properly, the nosecone diaphragm minimizes waste anesthetic gas leaks into the environment- even when using a passive waste anesthetic gas removal system. B C D A A. Fresh anesthetic gas delivery tube with 15mm male adaptor. Connects to common outlet of the vaporizer on the anesthesia machine. B. Fresh anesthetic delivery nosecone body. C. Nosecone diaphragm with ring. D. Waste anesthetic gas outlet for 19mm or 22mm EVAC tubing. 16
17 Set-Up: Use: 1. Connect the fresh anesthetic gas delivery tube to the common outlet of the vaporizer or Diverter System utilizing the 15mm male adaptor (A). 2. Secure the latex diaphragm (C) onto the nosecone (B) utilizing the included O-ring. a. Stretch the diaphragm material over the end of the face mask, holding it in place with the thumb and forefinger of one hand and reinstall the O-ring over the diaphragm material. b. The O-ring will only fit over the latex material when oriented properly. If the O-ring does not slide on easily, or fit snug, remove the O-ring, flip it over and replace over the latex material. c. The diaphragm is made of 12 mil thickness latex sheet material. This material is thicker and more resistant to tearing and oxidation than surgery glove material (2-3 mil thickness). We do not recommend that you replace the diaphragm with surgery glove material. Surgery gloves are potentially permeable to anesthetic gases allowing the waste gases to escape into the environment. 10 packs of replacement diaphragms are available (P/N RES501). 3. Cut an appropriate sized hole in diaphragm. a. The diaphragm can be cut appropriately using a pair of delicate sharp/sharp scissors. It is vital to the proper operation of the Rodent Nosecone that the diaphragm be cut with the proper sized orifice--appropriate for the size of the subject and the position of the subject. The circular hole in the diaphragm needs to be small enough such that the diaphragm forms a tight seal around the subject's muzzle, but large enough so that it is not too tight and occludes the nares preventing proper spontaneous breathing. Do not cut a cross (+) or an X in the diaphragm for the subject s nose because diaphragm will not seal properly around the subject s muzzle and anesthetic gases will escape into work space. 4. Connect either 19mm or 22mm EVAC tubing to the waste gas outlet of the breathing circuit (D). 5. Connect the opposite end of the EVAC tubing to your waste gas management system. 1. Make sure waste anesthetic gas management system is in place and functioning. 2. Place anesthetized subject s muzzle into the nosecone/diaphragm, making sure the diaphragm seals around the subject s muzzle. 3. Turn on the fresh anesthetic gas flow to the circuit. a. Recommended fresh gas flow rate is LPM at 1-2% isoflurane concentration. 4. When finished with procedure, turn vaporizer and fresh gas flow to zero, and remove subject from the nosecone/diaphragm to recovery area. 17
18 Rodent Heated Surgical Station with Downdraft Waste Gas Management RES4000 Provides body temperature regulation, fresh anesthetic gas delivery, waste anesthetic gas removal and subject security all in one convenient product. B C D E A G F A. 22mm Waste gas outlet B. Non-Rebreathing Circuit fresh gas inlet C. Non-rebreathing circuit holding block D. Rat and Mouse Non-Rebreathing Circuit E. Procedure platform F. Tie-down attachment G. Downdraft 18
19 Set-Up and Use: 1. Place Surgical Station on level surface 2. Connect warm water circulating pump to included attachments and turn on- allow platform to warm for approximately 5 minutes. 3. Insert breathing circuit waste gas outlet (D) into holding block (C) as shown in picture above. 4. Attach active vacuum source to waste gas outlet (A) with 22mm EVAC tubing and turn on. 5. Attach Rat and Mouse Non-Rebreathing Circuit fresh gas inlet (B) to fresh gas outlet of anesthesia machine using included 15mm male adaptors. 6. The Surgical Station is now ready to use 19
20 Multi-Station Surgical Board User Instructions RES3200 The Multi-Station Surgical Board delivers fresh anesthetic gas to up to six rodent subjects and provides a means to effectively remove waste anesthetic gases, all while keeping the anesthetized subjects securely in place. C B A F E FIGURE #1 D A. Six Port Supply Manifold B. Manifold waste gas exit port C. Coaxial Rat and Mouse Non-Rebreathing Circuit (up to six) D. Procedure platform E. Movable tie-down anchors F. Tie-down material 20
21 Set-up and Use: 1. Place the Surgical Board on a level surface. 2. Place the nosecones (C) into the manifold (A) with the fresh gas inlet tubing on the top surface (as shown in FIGURE #1). 3. Connect the nosecones to the anesthesia machine vaporizer common outlet using Scivena Scientific s Flow Balanced Multi-Station Manifold or any other diverter system that accepts a 15mm male adaptor. 4. Connect 22mm EVAC tubing to the manifold waste gas exit port (B) and connect to an activated charcoal canister or Scivena Scientific Waste Gas Interface if using with an active vacuum source. 5. Place the tie-down anchors (E) in the desired positions. 6. Use the tie-down materials (F) to secure the anesthetized animals in place once they are on the surgical board. 7. If not using all six stations, use included plugs to occlude unused stations. 21
22 Waste Gas Interface User Instructions EVC630 For Use with Fan Style Waste Gas Evacuation Unit (EVC3000). The interface unit balances the pressure between the anesthesia machine and the active, fan style evacuation unit. A C F B E E D FRONT SIDE A. 19mm Inlet - connects to waste gas outlet of anesthesia machine or accessory. B. Secondary inlet with included cap- connects to waste gas outlet of induction chamber. C. Interface Manifold D. Reservoir Bag E. Room Air Intake F. 22mm Outlet- connects to fan style waste gas evacuation system. 22
23 Set-up: Use: 1. Mount the interface to the anesthesia machine with included bracket and mounting hardware. 2. Connect blue 19mm EVAC tubing from the 19mm Inlet to the waste gas outlet on the pop-off valve of the anesthesia machine rebreathing head (for rebreathing machines) or the waste gas outlet of the non-rebreathing circuit (for machines without rebreathing head). 3. Connect the 19mm blue EVAC tubing from the secondary inlet to the waste gas outlet of the induction chamber. 4. Connect the clear 22mm EVAC tubing to the Fan Style Waste Gas Evacuation System. 1. Turn the waste gas evacuation unit on. 2. Use the induction chamber to anesthetize the subject. 3. When finished with the induction chamber, remove the tubing from the Secondary Inlet on the Waste Gas Interface and cap off with included cap. 4. Place subject on nosecone and maintain anesthesia as normal. 5. When finished with procedure, turn evacuation unit off. 23
24 Waste Gas Interface User Instructions RES505 For Use with Pump Style Waste Gas Evacuation Unit (EVC3100). The interface unit balances the pressure between the anesthesia machine and the active, pump style evacuation unit. A B C D A. 22mm Inlet - connects to waste gas outlet of breathing circuit or induction chamber. B. Interface Body with room air intake. C. ¼ Inch ID Hose barb Outlet- connects to pump style waste gas evacuation system or WAGD wall outlet. D. Vacuum stopcock. 24
25 Set-up: Use: 1. Connect clear 22mm EVAC tubing from the 22mm Inlet to the waste gas outlet on the breathing circuit or induction chamber. 2. Connect ¼ ID EVAC tubing/hose from pump waste gas evacuation unit to the Hose barb Outlet. 1. Turn the waste gas evacuation unit on. 2. Turn the Vacuum Stopcock (D) to the ON position 3. Use the accessory and as normal when Waste Gas Interface is connected inline. 4. When finished with the accessory, turn the anesthetic gas off to that accessory and turn the Vacuum Stopcock in the line to the OFF position. 5. When finished with procedures, turn waste gas evacuation unit off. 25
26 Warranty / Return Policy Scivena Scientific Anesthesia Innovations warrants that each product or part shall be free from defects in workmanship and materials, under normal use and with appropriate maintenance, for 10 (ten) years from the date of purchase. For plastic, rubber and disposable parts or items, Scivena Scientific Anesthesia Innovation warrants only that each such part and item shall be free from defects in workmanship and materials at the time of delivery. This warranty shall be void for any product that has been altered, defaced or removed from the original Scivena Scientific Anesthesia Innovation product. Scivena Scientific Anesthesia Innovations shall not be liable for any damage, injury or loss arising out of the use of the product, whether as a result of a defect in the product or otherwise, if, prior to such damage, injury or loss, the product was (1) damaged, misused, or misapplied; (2) repaired, altered or modified by persons other than Scivena Scientific Anesthesia Innovations; (3) not installed in strict compliance with the applicable codes and ordinances; (4) not installed by Scivena Scientific Anesthesia Innovations or an authorized Scivena Scientific Anesthesia Innovations dealer. UNDER NO CIRCUMSTANCES SHALL Scivena Scientific Anesthesia Innovations BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES AS THOSE TERMS ARE DEFINED IN THE UNIFORM COMMERCIAL CODE. All items returned for service or repairs are the responsibility of the customer. Proper packing methods should be used in returning items to Scivena Scientific Anesthesia Innovations. 26
27 SHIPPING POLICIES 1. Do not sign the Bill Of Lading until you have inspected the box or crate. 2. Examine the box AS it is delivered and BEFORE the truck/driver leaves. 3. If there is any evidence of damage when it arrives, note it in detail with the phrase possible concealed damage on the bill of lading and immediately call the office for instructions before the truck/driver leaves if at all possible. 4. If there is obvious damage such as a hole in the box, a crushed box, etc., refuse the shipment. The product will then go back to the freight company s terminal where they are entirely responsible. 5. Open and inspect your product as soon as possible. DO NOT WAIT. 6. If you find damage, take as many photographs of everything as soon as you can and them to inquiry@scivenascientific.com 7. Note: unless the following procedures are followed correctly and we are notified within ten (10) days, SCIVENA SCIENTIFIC LLC cannot accept any responsibility for the problems that may ensue. DO NOT RETURN ANY DAMAGED GOODS TO SUPERA LLC WITHOUT PRIOR AUTHORIZATION OF SCIVENA SCIENTIFIC, LLC AND THE CARRIER. KEEP ALL PACKAGING! DO NOT RETURN ANY DAMAGED ITEMS UNTIL SHIPPING INSTRUCTIONS ARE RECEIVED. ALL CLAIMS MUST BE FILED WITHIN TEN (10) DAYS OF RECEIPT OF GOODS. IF YOU HAVE ANY QUESTIONS CONTACT SCIVENA SCIENTIFIC DIRECTLY AT
28 DOCUMENTATION Date purchased: Purchased from: Machine serial number: Vaporizer serial number: Machine service information: Service Date: Service information: Service Date: Service information: Service Date: Service information: 28
Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200
Operators Manual Supera Floor Stand Anesthesia Machine Models: M1000 Basic Non-rebreathing M1200 Basic M1000 M1200 (All shown with optional equipment) Copyright Supera, LLC 2017 Examination and Preparation
More informationM5400 Operators Manual
M5400 Operators Manual For Veterinary Use Only Supera, LLC Copyright 2012 Examination and Preparation for Use Thank you for ordering an Supera Anesthesia Innovations anesthesia machine! We are delighted
More informationOperators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500
Operators Manual Supera Wall Mount Anesthesia Machines Models: M2300 M2500 (All shown with optional equipment) M2300 M2500 Examination and Preparation for Use Thank you for ordering an Supera Anesthesia
More informationURN (UNIVERSAL RODENT NOSE CONE) (17mm diameter) Figure 1
URN (UNIVERSAL RODENT NOSE CONE) 78914727 (17mm diameter) Figure 1 URN NRB SYSTEM (UNIVERSAL RODENT NOSE CONE NON-REBREATHING SYSTEM) 78909584 (17mm diameter) Figure 2 SMALL RODENT FACE MASK for use with
More informationYoke Block Instruction Manual
Yoke Block Instruction Manual ! WARNING IMPORTANT: READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE This manual contains instructions on periodically required checks to be performed by the user. These
More informationYoke Block Instruction Manual
Yoke Block Instruction Manual ! WARNING IMPORTANT: READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE This manual contains instructions on periodically required checks to be performed by the user. These
More informationPC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual
PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual ! WARNING IMPORTANT: READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE This manual contains instructions on periodically required checks to be
More informationINSTRUCTION MANUAL. (Variante 2 / System 2) Surgery table standard (Variante 1 / System 1)
Rothacher Medical GmbH Spisi 39 CH 1714 Heitenried, Switzerland Tel: + 41 31 371 12 81 Fax: + 41 31 371 53 35 email: info@rothacher-medical.ch www.rothacher-medical.ch INSTRUCTION MANUAL Surgery table
More informationCompact Triple Cabinet Outlet Station Model Installation and Operating Instructions
Compact Triple Cabinet Outlet Station Model 6258-1 Installation and Operating Instructions The Porter Compact Triple Outlet Station (6258-1) provides a quick, safe, and reliable method of connection to
More informationCompact Triple Cabinet Outlet Station Model B Installation and Operating Instructions
PORTER Parker Hannifin Corporation Porter Instrument Division 245 Township Line Rd. P.O. Box 907 Hatfield, PA 19440-0907 USA (215) 723-4000 / fax (215) 723-5106 Compact Triple Cabinet Outlet Station Model
More informationGAS ANESTHESIA SYSTEMS. Cat. No / 22100
GAS ANESTHESIA SYSTEMS Cat. No. 21100 / 22100 UGO BASILE BIOLOGICAL RESEARCH APPARATUS INSTRUCTION MANUAL Via G. Borghi 43 GAS ANESTHESIA SYSTEMS 21025 COMERIO - Varese, ITALY Cat. No. 21100 / 22100 Series
More informationIV-ANE. User s Manual. Anesthesia equipment for test animals. Petition
User s Manual IV-ANE Anesthesia equipment for test animals Petition Thank you for your purchase of Anesthesia equipment for test animals at this time. Prior to using these products, read this instruction
More information2 CYLINDER MOBILE CART USER S MANUAL / INSTRUCTIONS
Parker Hannifin Corporation Precision Fluidics Division Porter Instrument 245 Township Line Road Hatfield, PA 9440 Office 25-723-4000/Fax 25-723-506 www.porterinstrument.com 2 CYLINDER MOBILE CART USER
More informationINSTRUCTION MANUAL. Combi-vet Anesthesia system
INSTRUCTION MANUAL Combi-vet Anesthesia system CONTENTS 1 GENERAL... 4 1.1 CONNECTING DIAGRAM... 5 2 INSTALLATION... 6 2.1 Unpacking & Preliminary Check... 6 2.2 Notes on the Instruction Manual... 6 2.3
More informationGETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP
GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP LIMITED WARRANTY Getz Equipment Innovators warrants its products, and component parts of any product manufactured by
More informationSomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings
SomnoSuite FAQ V1.3 January 2015 Setup 1. How do I connect the SomnoSuite to my oxygen source? 2. Is there a way to speed up the downward movement of the pusher block when setting the empty position? 3.
More informationSDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual
SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual 8 Executive Drive P.O. Box 2105 Toms River, NJ 08754 (732) 244-0010 (800) 337-3762 fax (732) 244-8140 www.permapure.com info@permapure.com
More informationHallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting
Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting This is a quick reference for setting up, using, and troubleshooting the Hallowell EMC 2000 and 2002 Small Animal Ventilator.
More informationUser Guide and Operating Manual
User Guide and Operating Manual Table Top Laboratory Animal Anesthesia System 901806 & Mobile Laboratory Animal Anesthesia System 901807 / 901809 VetEquip, Inc. - Pleasanton, California USA - 800-466-6463
More informationBioAerosol Nebulizing Generator. Operation and Maintenance User Manual
BioAerosol Nebulizing Generator Operation and Maintenance User Manual INTRODUCTION The BANG or BioAerosol Nebulizing Generator is a unique nebulizer for the generation of aqueous aerosols at a low air
More informationNITRONOX SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS
NITRONOX SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS Examination All components should be removed from their respective shipping containers and examined. Any omissions and/or damage should be reported
More informationCHEMICAL INDUCTION KIT (Pt.No Issue 6, October 2015)
CHEMICAL INDUCTION KIT (Pt.No.2400-0740 Issue 6, October 2015) DESCRIPTION The C-Dax Chemical Induction Kit is an accessory kit for the GoldLine range of three point linkage sprayers. The system operates
More informationDFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER)
DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER) 0-3, 0-15, 0-30 & 0-70 LPM ASSEMBLY INSTRUCTIONS & INSTRUCTIONS FOR USE R219P87-400 (0-70 LPM) R219P88-400 (0-30 LPM) R219P79-400 (0-15 LPM) R219P99-400
More informationRK-IR Sample Draw Aspirator Adapter Operator s Manual
30-0951RK-IR Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0018RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment
More informationFOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM
IMI CORNELIUS INC One Cornelius Place Anoka, MN 55303-623 Telephone (800) 238-3600 Facsimile (612) 22-326 INSTALLATION INSTRUCTIONS FOR INSTALLING CO 2 BLENDER KIT (P/N 111612000 IN BEER SYSTEM SECONDARY
More informationVAPOMATIC ANESTHETIC VAPORIZER
VAPOMATIC ANESTHETIC VAPORIZER INSTRUCTION MANUAL SOLD FOR VETERINARY USE ONLY BY: A.M. BICKFORD, INC. 12318 BIG TREE ROAD WALES CENTER, NY 14169 1-800-795-3062 Fax: 716-652-2046 www.ambickford.com sales@ambickford.com
More informationMOBILE E TANK STAND INSTALLATION AND INSTRUCTIONS
Parker Hannifin Corporation Precision Fluidics Division Porter Instrument 245 Township Line Road Hatfield, PA 19440 Office 215-723-4000/Fax 215-723-5106 MOBILE E TANK STAND INSTALLATION AND INSTRUCTIONS
More informationGas and Vacuum Tubing Connection Kit for Manual Operation, No
LABCONCO CORPORATION 8811 Prospect Ave, Kansas City, MO 64132 (816) 333-8811, Fax (816) 363-0130, (800) 821-5525 Gas and Vacuum Tubing Connection Kit for Manual Operation, No. 5245100 Operation The Gas
More informationParker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref Rev.
Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA 19440 USA Tel: 215-723-4000 Ref. 10049700 Rev. L Table of Contents Warnings...2 Examination...2 Description and
More informationGETZ EQUIPMENT INNOVATORS RECOVERY SYSTEM CLEAN AGENT FE-36 / HALOTRON 1 PART NO. 4G59751 OPERATIONS MANUAL
GETZ EQUIPMENT INNOVATORS CLEAN AGENT FE-36 / HALOTRON 1 PART NO. 4G59751 OPERATIONS MANUAL GETZ EQUIPMENT INNOVATORS PEKIN, ILLINOIS, U.S.A. PHONE: (888) 747-4389 FAX: (309) 495-0625 Limited Warranty
More informationCSA Sample Draw Aspirator Adapter Operator s Manual
30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating
More informationBlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A
BlenderBuddy 2 Operating Manual & Instructions For Use ENGLISH R233M01 REV. A Maxtec 2305 South 1070 West Salt Lake City, Utah 84119 USA TEL (800) 748.5355 FAX (801) 973.6090 www.maxtec.com E-mail: sales@maxtec.com
More informationMulti-Probe Sampling System User Guide & Manual
Multi-Probe Sampling System User Guide & Manual Rev. 06/05/12 Part #22200012 Table of Contents CHAPTER 1: SYSTEM DESCRIPTION... 2 FUNCTION AND THEORY... 3 SYSTEM COMPONENTS... 4 CHAPTER 2: SYSTEM INSTALLATION...
More informationTable of Contents. Operating Instructions. Resource v.2 Conserving Regulator
Operating Instructions Table of Contents Resource v.2 Conserving Regulator Safety Information Device Precautions Introduction Product Features Product Specifications Feature Illustrations Set Up Usage
More informationPORTER INSTRUMENT COMPANY, INC.
PORTER INSTRUMENT COMPANY, INC. MXR FLOWMETER Instructions and Safety Considerations MXR Models of Porter Conscious Sedation Flowmeters and Bag Tee (Accessory) Wall Mount Models: 2000 and 2000-OS, 3000
More informationMANUAL BE SERIES Test Benches
The CustomCrimp Manual BE Series Test Benches are designed with features that make proof and burst testing of hydraulic hose assemblies a quick and easy procedure. CUSTOMIZED AND SPECIAL DESIGN BENCHES
More information92831 TEL: (714) FAX:
Document No. 1800-75 Respiration Test Chamber Copyright 2010 Terra Universal Inc. All rights reserved. Revised September 2010 Terra Universal, Inc. TerraUniversal.com 800 S. Raymond Ave. Fullerton, CA
More informationAnes Machine Use and Pre-use Check - CA1 focus
Anes Machine Use and Pre-use Check - CA1 focus BWH Anesthesia Grand Rounds August 31, 2005 James H. Philip, MEE, MD, CCE Copyright 2000-2005, James H Philip, all rights reserved BWH Anesthesia Workspace
More informationwith O 2 Controller Instruction Manual
14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Hypoxic Cabinets (In-Vivo / In-Vitro) with O 2 Controller Instruction Manual Index Page Warranty 2 Warnings
More informationACUC ۰ ROUTINE PROCEDURE
IACUC Use Only Procedure Number : LAH11-03 Genus (if other than mus): Procedure Type: Major Surgery Minor Surgery Non-Surgical Procedure Name: Isoflurane Anesthesia in Rodents 1. Procedure Description
More informationSERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL
www.modsafe.com SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device.
More informationJ Air and Water Kit Instructions Part# 02584
J Air and Water Kit Instructions Part# 02584 Unpacking Please open and inspect your package upon receipt. Your package was packed with great care and all the necessary packing materials to arrive to you
More informationinstruction manual Gas Anesthesia System Cat. No / UGO BASILE S.R.L. /
instruction manual Gas Anesthesia System Cat. No. 21100 / 22100 UGO BASILE S.R.L. sales@ugobasile.com / service@ugobasile.com www.ugobasile.com UGO BASILE S.R.L. Via G. Borghi 43 21025 COMERIO, VA, ITALY
More informationKJ4000 Operating Instructions & Parts Manual NSN
KJ4000 Operating Instructions & Parts Manual NSN 1025-01-473-7710 Mandus Group Ltd. KJ4000 Operators Manual Date: 3 Jan. 2002 TABLE OF CONTENTS General Safety Instructions...Page 1 Operator Instructions...Page
More informationSurgiVet Universal CDS 9000 Small Animal Anesthesia Machine Operation Manual
TM SurgiVet Universal CDS 9000 Small Animal Anesthesia Machine Operation Manual - English Catalog Number WW V1932 Version 0, January 2009 2009 Smiths Medical family of companies. All rights reserved. Table
More informationThis document to be used with Hurley traditional Davits
~ InstructIon Manual ~ This document to be used with Hurley traditional Davits WarnIngs WarnIng - Failure to install, maintain, protect, and operate the system properly can cause malfunction resulting
More informationACRYLIC FLOW METERS 0-30, 0-70 & LPM
ACRYLIC FLOW METERS 0-30, 0-70 & 0-120 LPM ASSEMBLY INSTRUCTIONS & INSTRUCTIONS FOR USE R219P86 (0-120LPM) R219P87 (0-70LPM) R219P88 (0-30 LPM) R138P11 Rev. D TABLE OF CONTENTS: 1.0 PRODUCT OVERVIEW...1
More informationINSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor
INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y P.O. Box 7079 OFFICE: (507)
More informationATD LB PRESSURE BLASTER INSTRUCTION MANUAL
ATD-8402 90LB PRESSURE BLASTER INSTRUCTION MANUAL SAVE THESE INSTRUCTIONS SAFETY INSTRUCTIONS FOR SANDBLASTER 1. Before opening the tank release the air pressure on the sand tank. To do this, turn off
More informationModel PSI Compressor with 3-Gallon Air Tank 12VDC
Model 6350 150 PSI Compressor with 3-Gallon Air Tank 12VDC IMPORTANT: It is essential that you and any other operator of this product read and understandd the contents of this manual before installing
More informationSTANDARD FLOWMETER USER S MANUAL
STANDARD FLOWMETER USER S MANUAL Instructions and Safety Considerations MXR Models of Porter Conscious Sedation Flowmeters and Bag Tee (Accessory) Flowmeter Models: 1000 1100 Bag Tee Model: P1407A Porter
More information2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL
2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL Model: 52024 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement,
More informationINSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y
INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50 Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y OFFICE: (507) 843-4700 P.O. BOX 7079 Toll
More informationD Series Air and Water Kit Part# 02550
D Series Air and Water Kit Part# 02550 Unpacking Please open and inspect your package upon receipt. Your package was packed with great care and all the necessary packing materials to arrive to you undamaged.
More informationBlendox AII5000F Inhalation Antalgic System. User s Manual :SP-Y Series No. Version :A1. Date of Review:
AII5000F Inhalation Antalgic System User s Manual Series No. :SP-Y53564 Version :A1 Date of Review:2012-04-29 To Users: Welcome to Blendox AII5000F Inhalation Antalgic System (hereafter referred to as
More informationAssembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &
REDQ Regulators Model 411B Barstock Design Powreactor Dome Regulator OPERATION AND MAINTENANCE Contents Scope..............................1 Installation..........................1 General Description....................1
More informationModel 687 Mouse Ventilator User s Manual. Model 687 Mouse Ventilator
Model 687 Mouse Ventilator User s Manual Model 687 Mouse Ventilator 55-7066 1 Table of Contents SUBJECT PAGE NO. Table of Contents...1 General Information - Warranty and Repairs...2 Specifications...3
More informationFloWash. Industrial Filter Cart
OPERATION AND MAINTENANCE MANUAL FloWash Industrial Filter Cart Read all instructions before installation or operation of equipment. Failure to comply with these instructions could result in bodily injury
More information453 Series Steam Heated Vaporizing Regulator
ADI 0453A Certified ISO 9001:2000 453 Series Steam Heated Vaporizing Regulator INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls Corporation
More informationAIR INLINE METAL SHEAR
AIR INLINE METAL SHEAR ASSEMBLY and OPERATING INSTRUCTIONS 3491 Mission Oaks Blvd. / Camarillo, CA 93011 Copyright 1997 by Harbor Freight Tools. All rights reserved. No portion of this manual or any artwork
More informationUSER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.
USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. RECEIVING / INSPECTION Remove the Precision Medical,
More informationAC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight
Technical Specifications Operating Instructions Maintenance Information Troubleshooting Guide Parts Diagrams AC1810 / AC1810-A THE EVOLUTION OF PERFECTION CAUTION: Before attempting to use or service this
More informationRK LEL Sample Draw Aspirator Adapter Operator s Manual
30-0951RK LEL Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0017RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment
More informationSERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL
Product Manual MST Portable Air Filtration System SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the
More informationAnesthesia Machine Intro
Anesthesia Machine Intro Brigham and Women s Hospital New Residents July 9, 2010 Copyright 2000-2010, James H Philip, all rights reserved ADS1 Ready JHP Anesthesia Machine Intro James H. Philip, MEE, MD,
More informationIn Vivo Scientific, LLC INSTRUCTION MANUAL
CO 2 Controller In Vivo Scientific, LLC INSTRUCTION MANUAL CONTENTS CONTENTS...1 ABOUT THIS MANUAL...2 INTRODUCTION...2 Cautions and Warnings...2 Parts List...2 Unpacking...2 INSTRUMENT DESCRIPTION...3
More informationCocoa Patio Pond with Lit Spillway
Cocoa Patio Pond with Lit Spillway REMINDER CALL 1-888-755-5641 BEFORE RETURNING TO STORE. PACKAGE CONTENTS ITEM # GQSPPB/GQSPPW Questions, problems, missing parts? Before returning to your retailer, call
More informationCS Controlled Dissolved Gas assembly with optional CS Air Aspirator assembly. Installation and Operation information
CS-425-01 Controlled Dissolved Gas assembly with optional CS-603-01 Air Aspirator assembly Installation and Operation information Serial # CS-XXX Date of shipment:, 2005 Shipped to: Purchase order #: xxxxxxx
More information/ FLT-M Quick Clean Rotary Airlock Valve Operators Manual. Kice Industries, Inc. RAV-M RAV-M
1 Quick Clean Rotary Airlock Valve Operators Manual Kice Industries, Inc. RAV-M04-0000 RAV-M04-0000 2 INTRODUCTION When you purchased your new Kice equipment, you bought a dependable and quality-built
More informationNB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE
NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE INSTALLATION, OPERATION & MAINTENANCE MANUAL INTERFACE DEVICES, INC. 230 Depot Road, Milford, CT 06460 Ph: (203) 878-4648, Fx: (203) 882-0885, E-mail: info@interfacedevices.com
More informationSection 3- Repair and Calibration Procedures
Section 3- Repair and Calibration Procedures 1. Refrigerant Removal Procedures 2. Scale/PC Board Recalibration Procedure 3. PC Board Configuration Procedure of AR400 4. Scale Tare Procedure 5. Gauge Calibration
More informationBS Series Basket Strainer
BS Series Basket Strainer Operating, Installation, & Maintenance Manual Corrosion Resistant Fluid and Air Handling Systems. Dated 04-26-12 PRESSURE DROP SIMTECH strainers are engineered to offer the lowest
More informationSOTR Special Operations Tactical Respirator
SOTR Special Operations Tactical Respirator OPERATOR S MANUAL FOR INDIVIDUAL RESPIRATORY PROTECTION OPS-CORE 2018 OMM G055-1000 REV. B INTRODUCTION ABOUT YOUR SOTR The Ops-Core Special Operations Tactical
More informationHallowell EMC Vacuum / Scavenging Interface
Hallowell EMC Vacuum / Scavenging Interface For use in veterinary medical applications only. Operating Manual and Scavenging Primer with Receiving Information, Installation and Maintenance Instructions
More informationINSTRUCTION MANUAL. January 23, 2003, Revision 0
INSTRUCTION MANUAL Model 810A In-Vitro Test Apparatus for 310B Muscle Lever January 23, 2003, Revision 0 Copyright 2003 Aurora Scientific Inc. Aurora Scientific Inc. 360 Industrial Parkway S., Unit 4 Aurora,
More informationInterface Devices, Inc. Hydraulic Mini Mule
Interface Devices, Inc. Hydraulic Mini Mule INSTALLATION, OPERATION & MAINTENANCE MANUAL IMPORTANT! FILE THIS MANUAL IN A SAFE PLACE FOR FUTURE SERVICE & PARTS NEEDS ALWAYS REFERENCE THE SERIAL NUMBER
More informationModel No Product Name Handpiece Only Complete System
Ney QC 700 E 30,000 RPM IECE 30,000 R N HANDPIE IVEN HA R DRIV AIR D Owner & Operator's Manual Model No Product Name 7201326 Handpiece Only 7201325 Complete System Description Page Safety... 2 Setup...
More informationSPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1.
2 GALLON TWIN STACK AIR COMPRESSOR Model: 9526 DO NOT RETURN TO STORE. Please CALL 800-348-5004 for parts and service. CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or
More informationUser s Guide. Pressure Controller for liquid delivery systems
Flow Control User s Guide Pressure Controller for liquid delivery systems Precise Pressure Control throughout the experiment Flow control Compatible with any perfusion system Ideal for Small Volume Delivery
More informationABRASIVE REGULATOR II
ABRASIVE REGULATOR II 13096 Abrasive Regulator II TABLE OF CONTENTS Contact Information and Customer & Technical Service... 1 1 Introduction... 1 2 Abrasive Regulator... 1 Description:... 1 2.1 Regulator
More informationMEDICAL BREATHING CIRCUIT SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS
MEDICAL BREATHING CIRCUIT SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS Examination All components should be removed from their respective shipping containers and examined. Any omissions and/or damage
More informationSASK-A-POLE OWNERS AND USERS MANUAL
SASK-A-POLE OWNERS AND USERS MANUAL GENERAL INFORMATION The Saskatchewan Abilities Council s Sask-a-Pole accessibility and transfer aid is designed to help provide safe and easy access to chairs, beds,
More informationSPECIFICATIONS TABLE Maximum Air Pressure 115 PSI Air Tank Capacity 20 Gallons
SPECIFICATIONS TABLE Maximum Air Pressure 115 PSI Air Tank Capacity 20 Gallons Air Flow Capacity 6.2 CFM at 40 PSI 5.2 CFM at 90 PSI Motor 2.5 HP Working / 4 HP Peak 120 Volt / 60 Hz / 13 A / 1-Phase Required
More informationABRASIVE REGULATOR II
ABRASIVE REGULATOR II 13096 Abrasive Regulator II TABLE OF CONTENTS 1 Introduction... 1 2 Abrasive Regulator... 2 2.1 Regulator Installation... 2 Location... 2 Attaching the Regulator to the Machine...
More informationXpressFill XF4500 Counter Pressure Filler. Operating Instructions
XpressFill XF4500 Counter Pressure Filler Operating Instructions ii XPRESSFILL Systems LLC lling machine. ank you for choosing our handcrafted bottle filler as the technology to bottle your passion. We
More informationINSTRUCTIONS FOR CONVERTING CONDENSING GAS MOBILE HOME FURNACES MODEL SERIES: CMA3*, CMC1*, VMA3*, VMC1*
INSTRUCTIONS FOR CONVERTING CONDENSING GAS MOBILE HOME FURNACES MODEL SERIES: CMA3*, CMC1*, VMA3*, VMC1* THIS KIT CONTAINS: AOPS7741 (NATURAL TO PROPANE CONVERSION PARTS FOR MODEL SIZE, 50) AOPS7742(NATURAL
More informationSSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR
S S F U SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR MACHINERY DIVISION OWNER S MANUAL UNIT INSTRUCTIONS Please follow all SSFU Safety Instructions. Contact your Duro Dyne Tech Service if you have
More informationisup MANUAL THANK YOU FOR PURCHASING A KAHUNA isup, WE HOPE YOU HAVE MANY YEARS OF PADDLING ENJOYMENT.
i isup MANUAL THANK YOU FOR PURCHASING A KAHUNA isup, WE HOPE YOU HAVE MANY YEARS OF PADDLING ENJOYMENT. THIS MANUAL IS A GENERAL GUIDELINE FOR SET-UP, CARE AND PROPER USE. PLEASE REFER TO OUR WEBSITE
More informationHallowell EMC Anesthesia WorkStation Operating Manual
Hallowell EMC Anesthesia WorkStation Operating Manual PN 000a2770 Anesthesia WorkStation A compact (9" W x 9" D x 15" H), convenient system for use on animals under 7 kg (15 lb.) Hallowell Engineering
More informationRG3100 and RG3100Ice Regulator System
RG3100 and RG3100Ice Regulator System User Guide www.diverite.com Date of purchase: www.diverite.com RG1208-5 & RG1208-5Ice www.diverite.com First Stage Regulator Product Description The RG1208-5 and RG1208-5Ice
More informationMEDICAL EQUIPMENT CATALOG
THE MOST TRUSTED MEDICAL EQUIPMENT IN NORTH AMERICA MEDICAL EQUIPMENT CATALOG Total Components Including Bulk Oxygen Supply, Back up and Delivery Equipment Designs for High Pressure, Liquid or Bulk Applications
More informationUser Manual MFCS TM -EZ
User Manual MFCS TM -EZ Version 6A_December 2016 1. INTRODUCTION 5 2. GENERAL INFORMATION 6 2.1 General functioning 6 2.2 Different MFCS TM 6 3. PACKAGE CONTENT 8 4. MFCS TM -EZ DESCRIPTION 9 5. CONNECTIONS
More informationModel ASSEMBLY and OPERATING INSTRUCTIONS
QUICK CHANGE AIR BRUSH KIT Model 93506 ASSEMBLY and OPERATING INSTRUCTIONS Due to continuing improvements, actual product may differ slightly from the product described herein. 3491 Mission Oaks Blvd.,
More informationGETZ EQUIPMENT INNOVATORS PART No. 4G59678 CARBON DIOXIDE MINI PUMP FOR EXTINGUISHER SERVICE (Revised )
GETZ EQUIPMENT INNOVATORS PART No. 4G59678 CARBON DIOXIDE MINI PUMP FOR EXTINGUISHER SERVICE (Revised 11-24-10) 2320 Lakecrest Drive, Pekin IL 61554 PH. (888) 747-4389 Fax (309) 495-0625 Website: www.getzequipment.com
More informationPRESSURE TEST PUMP OPERATING & MAINTENANCE INSTRUCTIONS. Model No: PTP100. Part No: GC01/09
PRESSURE TEST PUMP Model No: PTP00 Part No: 06020 OPERATING & MAINTENANCE INSTRUCTIONS GC0/09 INTRODUCTION Thank you for purchasing this CLARKE Pressure Testing Pump. Before attempting to use the product,
More informationVAC-U DRIVE Swinging float arm for Acrobat like agility 36 Fin disk with DEEP VEE Scoop
ACROBAT Vac-U-Drive Automatic Pool Cleaner VAC-U DRIVE Swinging float arm for Acrobat like agility 36 Fin disk with DEEP VEE Scoop Proven technology:diaphragm driven strength & quietness Unique hose design
More informationLUBRICATOR ASSEMBLY AND OPERATING INSTRUCTIONS
AIR FILTER, REGULATOR AND LUBRICATOR 4035 ASSEMBLY AND OPERATING INSTRUCTIONS 349 Mission Oaks Blvd., Camarillo, CA 930 Visit our Web site at http://www.harborfreight.com Copyright 004 by Harbor Freight
More informationSERVICE MANUAL MODEL BA020AMST BREATHING AIR PANEL
SERVICE MANUAL MODEL BA020AMST BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device. 5/08 2 GENERAL
More informationTL-12 FLUSH TANK. Manual Ref. # MN-??? Polyurethane Machinery Corp. REVISION 1.0
TL-12 FLUSH TANK Manual Ref. # MN-??? REVISION 1.0 Polyurethane Machinery Corp. Corporate: 1 Komo Dr, Lakewood, NJ 08701 Manufacturing: 2 Komo Dr, Lakewood, NJ 08701 Phone: 732-415-4400 Fax: 732-364-4025
More informationUNIVERSAL MASK TEST SYSTEM
f UNIVERSAL MASK TEST SYSTEM Quotation #110811 Automated Motion Inc. 225 NW Victoria Drive Lee s Summit, Missouri 64086-4709 www.automatedmotion.com SYSTEM OVERVIEW AMI s Universal Mask Test System is
More information