Sample Draw Oxygen Monitoring System (OMS) User Manual

Size: px
Start display at page:

Download "Sample Draw Oxygen Monitoring System (OMS) User Manual"

Transcription

1 Sample Draw Oxygen Monitoring System (OMS) User Manual

2 ETS-Lindgren Inc. reserves the right to make changes to any product described herein in order to improve function, design, or for any other reason. Nothing contained herein shall constitute ETS-Lindgren Inc. assuming any liability whatsoever arising out of the application or use of any product or circuit described herein. ETS-Lindgren Inc. does not convey any license under its patent rights or the rights of others. Copyright by ETS-Lindgren Inc. All Rights Reserved. No part of this document may be copied by any means without written permission from ETS-Lindgren Inc. Trademarks used in this document: The ETS-Lindgren logo and OMS are trademarks of ETS-Lindgren Inc.; NEMA is a registered trademark of NEMA; 3M and Scotch-Brite are trademarks of 3M. Revision Record MANUAL,OXYGEN DEFICIENCY MONITOR Part #399374, Rev. C Revision Description Date A Initial Release August, 2013 B Updated Introduction; added installation location information; added Sample Draw Sensor Tubing December, 2013 C Updated remote display alarm indicator information February, 2017 ii

3 Table of Contents Notes, Cautions, and Warnings... vii 1.0 Introduction... 9 Technology... 9 Long-Life Zirconium Oxide Sensor Cell... 9 Smart Electronics Standard Configuration Optional Remote Display Other Optional Items ETS-Lindgren Product Information Bulletin Maintenance Recommended Routine Maintenance Every 6-12 Months Annually Replacement and Optional Parts Service Procedures Specifications Electrical Specifications Physical Specifications Performance Specifications Gas Detection System Specifications Signal Outputs Default Factory Settings Component Views Front View Exterior Front View, Cover Removed Transmitter Interior Alarm Relay Board Before You Begin Installation Site Requirements Connect to 24 VDC Regulated Power Connect OMS and Sensor Before Powering iii

4 Do Not Exchange the Sensor Electronics Use Proper Calibration Steps Keep Away From Silicone Compounds Keep Away From High Air Flow Keep Away From a Water Stream Installation Mounting the OMS Transmitter and Sensor Enclosure Mounting Feet Dust Filter Sample Draw Sensor Tubing Using an Available Waveguide Using Kit to Install Waveguide Wiring Connecting a Remote Horn and Strobe to OMS Connecting a Remote Fan Contactor to OMS Initial Startup Operation Joystick Operation Main Operation Mode Internal Sample Flow Rate Signal Outputs Instrument Faults Loss of Power Indicator Alarm Reset Programming the OMS Program Flowchart Passwords Enter Password Change Password Enable/Disable Password Function Main Menus and Submenus Set 4 20 ma Loop Set Formats Set Alarm Threshold Parity iv

5 Set Latching Reset Latching Alarm Set Alarm Delay Set Zero Suppression Set Alarm Thresholds Set Alarm Hysteresis Set Sensor Adjust Sensor Verification Required Gas and Equipment Sensor Verification Procedure Sensor Verification to Nitrogen Sensor Verification to Known Concentration of Oxygen Remote Display Alarm Indicator How to Mount the Remote Display Alarm Indicator How to Wire the Remote Display Alarm Indicator Identification of Switches and Controls How to Set up and Operate the Remote Display How to Set the Zero and Full Scale Range How to Set the Internal Alarms on the Remote Display Appendix A: Warranty v

6 This page intentionally left blank. vi

7 Notes, Cautions, and Warnings Note: Denotes helpful information intended to provide tips for better use of the product. Caution: Denotes a hazard. Failure to follow instructions could result in minor personal injury and/or property damage. Included text gives proper procedures. Warning: Denotes a hazard. Failure to follow instructions could result in SEVERE personal injury and/or property damage. Included text gives proper procedures. See the ETS-Lindgren Product Information Bulletin for safety, regulatory, and other product marking information. vii

8 This page intentionally left blank. viii

9 1.0 Introduction The ETS-Lindgren Sample Draw Oxygen Monitoring System (OMS ) is a self-contained oxygen deficiency detection system suitable for remote sampling of confined spaces; this single-point monitor is designed for the continuous detection and measurement of ambient oxygen concentration levels. As a compact gas monitoring system, the OMS is ideal for the continuous monitoring of the air in MRI rooms, labs, freezers, confined spaces, and other shielded enclosures where inert gases may displace the oxygen and create a danger for patients and staff. The OMS is suitable for indoor and outdoor use, and is intended to be installed outside of the shielded enclosure. Each system consists of a long-life zirconium oxide sensor cell and 3-wire transmitter. Unlike electrochemical sensor cells, the zirconium oxide sensor cell provides stable oxygen readings even in areas where temperature and humidity levels are changing. The OMS may be used as a standalone gas detector or connected to your own centralized control and surveillance system. Technology LONG-LIFE ZIRCONIUM OXIDE SENSOR CELL The heart of the monitoring system is a zirconium sensor, which responds to low oxygen conditions within seconds and provides accurate measurements over a wide temperature and humidity range. The zirconium sensor will operate continuously for eight or more years, requiring minimum maintenance. There are no zero or span calibration pots to adjust, and when compared to disposable sensors, the long-life zirconium sensor can save hundreds of dollars in annual maintenance. Unlike concentration cells, the zirconium sensor does not need an oxygen reference gas for proper operation. The OMS can detect low oxygen levels in confined spaces and process tools without the need of a reference gas. Warranty 9

10 SMART ELECTRONICS The OMS incorporates a special electronic circuit that continuously monitors sensor operation. With the addition of the alarm relay option, any cell degradation or complete failure will be detected immediately. This smart circuitry alerts the user to sensor faults and other electrical problems that may interrupt surveillance through the standard ma signal output signal or through the optional fault relay option. Ideal for continuously monitoring oxygen levels in confined spaces or areas where inert gases are used, the OMS does not drift or lose sensitivity with weather or temperature changes. The electronics are housed in a NEMA Type 4X enclosure. Standard Configuration The OMS must be installed outside of the shielded enclosure. Oxygen Monitoring System Power supply Dust filter Optional Remote Display The remote display receives oxygen concentration information from the OMS. Up to two remote displays can be connected. See page 40 for installation information. 10 Warranty

11 Other Optional Items Remote Horn and Strobe Waveguide Installation Kit ETS-Lindgren Product Information Bulletin See the ETS-Lindgren Product Information Bulletin included with your shipment for the following: Warranty information Safety, regulatory, and other product marking information Steps to receive your shipment Steps to return a component for service ETS-Lindgren calibration service ETS-Lindgren contact information Warranty 11

12 This page intentionally left blank. 12 Warranty

13 2.0 Maintenance Calibrate or challenge the OMS ONLY outside the MRI suite. Before performing any maintenance, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment. Maintenance of the OMS is limited to the procedures described in this manual, and should be performed only by qualified personnel. WARRANTY Warranty may be void if non-authorized procedures are performed, or if performed by non-qualified personnel. If you have any questions concerning maintenance, contact ETS-Lindgren Customer Service. Warranty 13

14 Recommended Routine Maintenance The Sample Draw Oxygen Monitoring System (OMS ) is a continuous gas detection system that measures and detects hazardous gas leaks in the workplace, and therefore requires periodic maintenance to ensure proper operation. The frequency with which this routine maintenance is performed depends on your environment and company policies. Following are recommendations intended as a general guideline. EVERY 6-12 MONTHS Visual checks: Check for power and proper operation. The OMS should output a ma signal when the oxygen level is at 20.9%. Also, the display should indicate 20.9% oxygen when the oxygen is at ambient levels. Sensor verification with nitrogen: The ambient oxygen level is 20.9%; therefore, under ambient conditions verification of the OMS to 20.9% oxygen is constantly performed. The OMS requires periodic testing with nitrogen only to verify the cell response to low oxygen levels. See Sensor Verification on page 73 for detailed steps. ANNUALLY Depending on the environment, the filter should be replaced every 12 months; in dusty environments, the filter may need to be replaced more frequently. If the filter becomes completely blocked, the internal flow sensor will detect the loss of flow and activate the fault relay and LED. 14 Warranty

15 Replacement and Optional Parts ETS-Lindgren may substitute a similar part or new part number with the same functionality for another part/part number. Contact ETS-Lindgren for questions about part numbers and ordering parts. Following are the part numbers for ordering replacement or optional parts for the OMS. Part Description Part Number Optional Remote Display Filter Replacement Service Procedures For the steps to return a system or system component to ETS-Lindgren for service, see the Product Information Bulletin included with your shipment. Warranty 15

16 This page intentionally left blank. 16 Warranty

17 3.0 Specifications Electrical Specifications A regulated 24 VDC power supply is required. Power: Consumption: 24 VDC external power Approximately 700 ma Physical Specifications Height: Width: Depth: Weight: Enclosure Type: 7.0 in (178 mm) 5.0 in (127 mm) 5.0 in (127 mm) 4.0 lb (2 kg) NEMA Type 4X wall mount general purpose; not intended for explosive atmospheres or electrically-classified areas Warranty 17

18 Performance Specifications Sensor Type: Response Time: Repeatability: Fault Indicators: Long-life zirconium oxide sensor cell, 0% 25% Within 2 seconds of any change in oxygen ± 2% of reading Loss of VDC power (analog signal drops to 0 ma). Sensor cell failure: fault relay activated Operating Temperature: Humidity: -40 C to + 40 C (40 F to 104 F) Contact ETS-Lindgren for lower or higher operating temperatures 0% to 95% relative humidity (RH) Contact ETS-Lindgren for sensors which can operate in 100% condensing RH environments Gas Detection System Specifications Type: Sensor Life: Transmitter: Long-life zirconium oxide sensor cell, 0% 25% 8 to10 years, under normal conditions Microprocessor electronics with built-in 3-digit backlit LCD display Joystick-operated menus 18 Warranty

19 Signal Outputs Local Display: Digital display calibrated for oxygen. The range is stated on the serial number and can be accessed via the joystick on the front panel. In the measurement mode push the joystick down to scroll the gas and range on the display. Push the joystick down again to stop the scrolling and re-display the gas. Standard: Optional: Analog Output: DC 4 20 ma Relay Output: Dual level user-selectable alarm relays and one fault relay Rated: 2 30 VDC; VAC Default Factory Settings The built-in relay settings are user-changeable; see Main Menus and Submenus on page 58 for more information. Menu Function / Description Set 4 20 ma loop To set the OMS 4 ma (zero) and 20 ma (span) to your PLC or distributive control system. Set Formats; LED and Alarm Relay State To set the relays to energize (normal) or de-energize (fail safe) when the alarm activates. Factory Default The ma output is set at the factory using a calibrated Fluke meter. Alarm 1 = Normal Alarm 2 = Normal Fault = Normal Note: The LED indicators on the front panel are connected directly to the alarm relays. Warranty 19

20 Menu Function / Description Set Alarm Threshold Polarity To set to alarm at a level higher (normal) or lower (inverted) than the alarm threshold. Note: The Audio Alarm feature is optional. Set Latching To set the alarm to automatically reset (non-latching) or to manually reset (latching). Alarm Delay Factory Default Alarm 1 = Inverted Alarm 2 = Inverted Audio = Inverted Alarm 1 = Non-latching Alarm 2 = Non-latching Audio = Non-latching Alarm = 5 seconds To set the time to wait until the relay alarms activate. Zero Suppression 000 = 0.00% Note: This function is not enabled on the OMS. Set Alarm Thresholds To set the level to alarm. Note: The Audio Alarm feature is optional. Set Alarm Hysteresis Use when utilizing the OMS for control or valves and process; see page 69 for more information. Sensor Adjustment Alarm 1 = 19.5 % Alarm 2 = 18.0 % Audio = 19.5% Alarm 1 = 0.0 % Alarm 2 = 0.0 % Audio = 0.0 % No factory default For use when dynamically gas calibrating the OMS to a known span gas; see page 71 for more information. Manage Passwords Factory default is 557 To change the password from the factory default to a new password. 20 Warranty

21 4.0 Component Views Front View Exterior 1. Front cover Removable, waterproof cover that protects the interior of the transmitter; fastened by four captive screws. 2. Joystick Used to select and adjust the built-in menus. The OMS features dual level user-selectable alarms. 3. Digital display Displays the oxygen levels in percentage; the normal oxygen level on Earth is 20.9%. Warranty 21

22 4. Alarm relay LEDs Three multi-colored LED indicators, from left to right: Alarm 2 Red Alarm 1 Orange Fault Yellow 5. Mounting feet (4) 6. Sample inlet Permits the flow of oxygen to enter the sensor. 7. Cable strain relief The sealed opening in the transmitter housing for connecting the input power, 4 20 ma output, and relay wiring. 8. Sample exhaust Permits the flow of oxygen to exit the enclosure. 9. Front cover fasteners (4) Four captive screws attach the front window to the base unit. 10. Front panel fastening screws (2) 22 Warranty

23 Front View, Cover Removed 1. Sample pump Brings in a sample to the sensor. Flow rate is preset at the factory, and is continuously protected with a built-in flow sensor. For more information, see Instrument Faults on page Sensor assembly A zirconium oxide sensor cell that detects and measures the level of oxygen. When exposed to oxygen, the sensor outputs an electrical signal proportional to the actual concentration of oxygen. 3. Retaining screws for electronics panel (2) These mount the display front panel and electronics to the enclosure. When unfastening the front panel electronics, apply upwards pressure on the back of the front panel to release the captive screw. Warranty 23

24 Front view with electronics panel rotated 24 Warranty

25 Transmitter Interior The fuse is field-replaceable. The power analog output terminal block is a 5-pin terminal block where the 24 VDC power and 4 20 ma analog output connections are made. Warranty 25

26 Alarm Relay Board For the relays, from left to right: NC C NO 26 Warranty

27 5.0 Before You Begin Installation Before installing or connecting any components, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment. Failure to follow the instructions in this chapter may damage the OMS and/or the sensor. Site Requirements The OMS is not designed for installation in a hazardous area. Contact ETS-Lindgren for information on special enclosures for use in a hazardous environment. The OMS must be installed outside of the MRI room; otherwise, the OMS may interfere with the normal operation of the MRI system. The OMS must be installed outside of the shielded enclosure. The Sample Draw Oxygen Monitoring System (OMS ) enclosure should be mounted in an area free of vibration and electrical noise or interference. If possible, avoid areas with high temperatures or condensing humidity. Warranty 27

28 Connect to 24 VDC Regulated Power The OMS requires 24 VDC regulated power. WARRANTY Do not connect the OMS to any voltage that exceeds 24 Volts DC. Do not connect the OMS to any AC voltage. Connect OMS and Sensor Before Powering To avoid damage to the sensor: Do not power the OMS with the sensor unplugged from the main PC board. Do not connect the sensor to the PC board while the OMS is powered. Do Not Exchange the Sensor Electronics The sensor is matched to the electronics. Never exchange the electronics with a sensor from a different OMS. Use Proper Calibration Steps Calibrate or challenge the OMS ONLY outside the MRI suite. 28 Warranty

29 When calibrating or challenging the OMS: Do not expose the OMS to flow rates that exceed a half-liter per minute (500 cc per minute). When testing sample draw monitors use an on-demand regulator to expose the span gas to the OMS. Expose the OMS to span gas blends that only consist of oxygen and nitrogen. Do not expose the OMS to any combustible gas (methane, hydrogen, etc.). Exposure to combustible span gases can damage the sensor. Keep Away From Silicone Compounds Do not expose the OMS to silicone compounds. Exposure to silicone compounds can cause a loss of sensitivity. Keep Away From High Air Flow Do not expose the OMS to high air flow or install it directly in front of fans. High air flow can cool the sensor and cause inaccurate readings. Warranty 29

30 Keep Away From a Water Stream Do not expose the sensor directly to a water stream. Cover and protect the OMS and power supply if located in an area that requires wash downs. Contact ETS-Lindgren for information on purchasing a waterproof enclosure. 30 Warranty

31 6.0 Installation Before connecting any components, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment. Prior to performing any of the steps in this chapter, see Before You Begin Installation on page 27. The OMS must be installed outside of the MRI room; otherwise, the OMS may interfere with the normal operation of the MRI system. The OMS must be installed outside of the shielded enclosure. Mounting the OMS The Sample Draw Oxygen Monitoring System (OMS ) is designed primarily for wall mounting and should be installed at a height convenient for operation, maintenance, and viewing of the instrument display. The transmitter and sensor should be installed in a location where gas leaks are likely to occur or where released gases may accumulate. Warranty 31

32 TRANSMITTER AND SENSOR Leave clearance room on the bottom of the OMS for the filter, which adds an additional 3.75 inches. The OMS should be mounted no closer than 12 inches above floor level. Airflow within the monitored area, the characteristics of the gas (lighter or heavier than air), and the position of workstations and personnel should all be considered in determining the most suitable installation location. 32 Warranty

33 ENCLOSURE MOUNTING FEET The four mounting feet can be oriented in any direction. The feet can also be removed for mounting the OMS flush with a wall or other surface. Warranty 33

34 DUST FILTER To protect the pump from dust, a particulate filter is provided. 1. Attach the filter to the sample inlet by inserting the filter into the 1/4-in tube compression fitting. 2. Verify the correct flow by aligning the arrow towards the OMS. 34 Warranty

35 Sample Draw Sensor Tubing A penetration panel area contains active electrical equipment and connections; contact with this equipment can be life threatening. Installation must be performed by a qualified individual. Run transparent polyurethane 1/4-in (7-mm) tubing (not included) a maximum of 100 feet (30 meters) from the display unit, through a waveguide, and into the interior of the MRI room or shielded enclosure. The OMS flow pump continuously samples air drawn from end points up to a maximum of 100 feet (30 meters) from the unit. For optimal sensing, position the end point of the tube near the ceiling of the monitored space. A tuned pipe waveguide allows the tubing to enter the shield but attenuates electromagnetic interference (EMI) from entering. Warranty 35

36 USING AN AVAILABLE WAVEGUIDE (Magnet penetration panel area shown) 36 Warranty

37 1. In the equipment room locate the penetration panel and determine if there is an available waveguide port. Any unused pipe waveguide with a 1/2-in diameter or greater can be used. If there is no waveguide available, go to page 37 to install Waveguide Kit If one of the following types of waveguides is available, then route the sensor tubing through and into the MRI room or other shielded enclosure: If a threaded pipe waveguide is available: Install standard threaded plastic bushings to ensure that the sensor tubing is not abraded over time. If a medical gas port that consists of a copper tube is available: Use a type of grommet to protect each end of the waveguide tube. USING KIT TO INSTALL WAVEGUIDE The Waveguide Kit includes the following: 1/2-in threaded brass pipe waveguide (1 each) 1/2-in brass bee nuts (2 each) Warranty 37

38 To prevent linen dust or other airborne particulate matter from clogging the sensor tubing, use a small filtered terminus that can be serviced easily. 1. Locate the area in the penetration panel to place the 1/2-in pipe waveguide. 2. Using a hole saw with the same inner diameter as the 1/2-in pipe waveguide outer diameter, gently cut through the copper shielding. This can be done by hand-rotating the hole saw in reverse, or by marking the copper and using a box cutter to remove the copper. 3. Cut through the plywood with the hole saw. 4. Use a 3M Scotch-Brite pad or fine steel wool to clean the copper surface where the waveguide will pass through the shield; do not use sandpaper. 38 Warranty

39 5. Run one of the brass bee nuts down to the approximate half-point of the waveguide. Slide the 12-oz copper washer onto the waveguide so that it will rest between the copper shield face and the bee nut. Insert the waveguide. 6. On the inside of the shield install the second bee nut and hand tighten. On the outside of the shield grip the waveguide and tighten the bee to a snug torque. 7. Route the 1/4-in sensor tubing through the waveguide. 8. For an MRI room: Inside the MRI room, route the sensor tubing along the inside of the finished ceiling to a point above the magnet, or to a point between the top of the magnet and a return air grill. Warranty 39

40 Wiring A 3-wire shielded cable: 3-conductor, 18 AWG stranded General Cable E2203S , or equivalent is recommended. The maximum permissible cable length is 0.62 miles (1 km). The OMS requires a single, 3-wire shielded cable for analog output and 24 VDC power input. The analog out and VDC power in connections are made on the terminal block inside the transmitter housing. Pin# Connection Description Common (Signal Ground) 0V Common (Signal Ground) 0V MA Signal Out DC 4 20 ma Output + Power DC + 24 V Input + Power DC + 24 V Input Additional contacts for +24 VDC Power and Common have been added to accommodate additional wiring for remote horns and strobes. 40 Warranty

41 Connecting a Remote Horn and Strobe to OMS The following illustrates how to connect a remote horn and strobe (federal signal horn/strobe, or equivalent) to the OMS with Alarm Relay #1 set to Normally Open. Warranty 41

42 Connecting a Remote Fan Contactor to OMS Use 24 VAC voltage inside the OMS; do not use 120 VAC or higher voltage. 42 Warranty

43 Initial Startup Once installation of the OMS is complete, it is ready for startup. Follow these steps prior to putting the OMS into operation: 1. Check the integrity of all wiring. 2. Apply 24 VDC power to the power supply PCB board. The sample pump will activate and the digital display second line will quickly display the digital transmitter module (DTM), serial number, and software version. To display the DTM numbers: Push the joystick to the left. To display the gas and range: Push the joystick down. The gas and range will continuously scroll. To stop the scrolling: Push the joystick down. The OMS should now be powered up. The display will show a 4-minute, (240-second) countdown as the current to the sensor stabilizes. The monitor will output a 4 ma signal during the entire warm-up period. Do not make any adjustments to the reading until after the OMS has been powered for at least one hour. After the countdown, the sensor will continue to reach its operating temperature for approximately 30 minutes and the reading that is displayed will slowly increase to ambient. Immediately after the countdown, the OMS will display FLOW FAILURE: OUT OF RANGE on the top line and disappear. This is a fault history message only and is generated at startup to verify the integrity of the system. The message will automatically clear and Oxygen will display on the top line and 20.9% will display on the bottom line. If the fault is still present, the actual fault code will display on the bottom line. Warranty 43

44 The OMS should now be operating properly and ready for operation. For information on adjusting the OMS to the ambient oxygen level, see Set Sensor Adjust on page 71. Pump Adjustment: The OMS has a built-in digitally controlled flow controlled sample pump. The pump flow is set at the factory and should not require adjustment. It has a range to accommodate tubing lengths of three feet to over 100 feet. If flow adjustment is necessary, contact ETS-Lindgren. Normal flow rate is between 0.10 and 0.25 liters per minute. 44 Warranty

45 7.0 Operation Before placing the OMS into operation, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment. Joystick Operation The joystick has a built-in delay to prevent accidental tampering with the menus; deliberate entries are required. The Sample Draw Oxygen Monitoring System (OMS ) uses an 8-position joystick with a center pushbutton for selecting menus and changing values. The joystick is programmed to standard protocol as follows: + Plus Push the joystick in this direction to increase the value. Minus Push the joystick in this direction to decrease the value. Next Previous Enter Push the joystick in this direction to moves to the next level of the menu hierarchy. Push the joystick in this direction to go back to the last level of menu hierarchy. Push the joystick directly in the center to enter the information. Warranty 45

46 Main Operation Mode Step To select the main menu from any submenu, push the joystick to the left until the main menu displays. Display Oxygen 20.9% Internal Sample Flow Rate The OMS has an internal sample pump flow rate that is programmed at the factory and cannot be changed in the field. A flow sensor on the relay board continually monitors flow rate and when a loss of flow is detected, a signal is sent to the fault relay and the front mounted LED will activate. Sample flow to the monitor is continuously monitored and controlled by the flow control microprocessor. If the sample line is blocked, the fault indicator will illuminate and the pump will accelerate and attempt to re-establish the proper flow rate. If the line is cleared, the pump will speed and slow down and the fault light will turn off when the flow rate is back to factory setting. For the recommended filter maintenance, see page 14. Signal Outputs The OMS outputs a continuous 4 20 ma analog signal proportional to the measured concentration of oxygen; 4 ma represents 0% oxygen, and 20 ma represents 25% oxygen, which is the full range. In the event of a system fault, a specific factory-defined code will be displayed on the display, indicating the exact nature of the system fault. 46 Warranty

47 Instrument Faults When using your own power supply, make sure that the voltage is regulated to 24 VDC ± 0.5 Volts. A voltage that is too low or too high will activate a Supply Voltage Out of Range fault and disable the OMS. The OMS incorporates a number of self-checking features to ensure reliable operation. If a fault condition is detected, the analog output signal is altered as follows: All system faults are displayed on the front panel. Each fault has a unique code to identify the specific problem. Contact ETS-Lindgren if a fault is displayed. Condition Supply Voltage Out of Range Analog Signal Analog output drops to 2 ma Fault Code 16 Transmitter cable cut O2 Cell Current Fault Fault Code 128 O2 System Warm Up Analog output drops to 0 ma Analog output drops to 2 ma (0 ma on request) Fault Relay activates (available with Relay Option only) Analog output drops to 2 ma Fault Relay activates and turns off when system is in Oxygen Operation Mode O2 Cell Voltage Fault Fault Code 64 No Flow to Oxygen Sensor Fault Code 32 Analog output drops to 2 ma Fault Relay activates (available with Relay Option only) Analog output drops to 2 ma Fault Relay activates Warranty 47

48 EEPROM Fault 08 Analog output drops to 2 A (0 ma on request) Loss of Power Indicator If the OMS loses 24 VDC power, the 4 20 ma analog output signal drops to 0 ma. The display will show a blank screen. Alarm Reset If the OMS is supplied with the optional alarm relays, when the monitors alarms are activated, the built-in alarm relays, panel mounted LEDs, and optional audio horn will also activate. When the relay settings are non-latching, the alarm relays, LEDs, and horn will automatically reset. If the relay settings are latching, then a manual reset of the alarms is required. Resetting the alarms can be performed by using the joystick or using the remote reset function. Joystick To reset the alarms, enter the password, and then push the joystick directly in the center to enter the information. Remote Reset The alarm relay board has a 2-pin connector for wiring to a remote switch. When connected to a switch, this remote reset will bypass the joystick and a password will not be needed to reset the alarms. The oxygen levels must recover above the alarm thresholds before the horn can be reset from the remote reset switch or joystick. 48 Warranty

49 8.0 Programming the OMS Before placing the OMS into operation, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment. Only qualified personnel should program the OMS. The Sample Draw Oxygen Monitoring System (OMS ) is supplied with user-selectable settings to adjust the alarm settings, 4 and 20 ma output, and minor sensor adjustments. The settings are arranged in menus that are accessed by moving the joystick. Use the password set at the factory to access the menus. The OMS will continuously monitor oxygen while accessing the menus. The alarm, fault relays, and ma output are all active and on line while making changes to the menus. Warranty 49

50 Program Flowchart All numerical values shown in the flowcharts on the following pages are examples; the values and are not factory defaults. The Zero Suppression function is not available for oxygen. 50 Warranty

51 Warranty 51

52 52 Warranty

53 Warranty 53

54 54 Warranty

55 Passwords If you forget your password, contact ETS-Lindgren for instructions on password recovery. You will need to provide the digital transmitter module (DTM) for the OMS, which is displayed by moving the joystick to the left. ENTER PASSWORD The OMS has a factory-set password to prevent unauthorized access to the menus; the password is 557. Following are the steps to enter the password: Step 2. Push the joystick once to the right. 3. Push the joystick again once more to the right to enter the input screen. Display Enter Password The letter A will appear and flash. The display has characters A Z and 0 9. Pushing the joystick up or down will scroll through the alphanumeric characters. 4. Push the joystick up or down to enter the first digit. 5. Push the joystick to the right to select the second entry, and then push the joystick up or down to select the second digit. The character to be entered will flash. The character being entered will flash and the first character entered will remain illuminated. Warranty 55

56 Step 6. Push the joystick to the right to select the third entry, and then push the joystick up or down to select the third and final digit. 7. Push the joystick in the center to enter the password. Display The character being entered will flash, and the first and second characters entered will remain illuminated. Password OK. If an incorrect password was entered, the display will show Password Failed. Push the joystick to the left to access the monitoring mode. From this mode you can re-enter the password. CHANGE PASSWORD The OMS has a factory-set password to prevent unauthorized access to the menus; the password is 557. Following are the steps to change the password: Step 1. Push the joystick down to access the Manage Passwords menu. 2. Push the joystick to the right to enter the input screen. 3. Push the joystick to the right to enter the input screen. Display Manage Passwords Enter New User Password The letter A will appear and flash. The display has characters A Z and 0 9. Pushing the joystick up or down will scroll through the alphanumeric characters. 4. Push the joystick up or down to enter the first digit. The character to be entered will flash. 56 Warranty

57 Step 5. Push the joystick to the right to select the second entry, and then push the joystick up or down to select the second digit. 6. Push the joystick to the right to select the third entry, and then push the joystick up or down to select the third and final digit. 7. Push the joystick in the center to enter the new password. 8. Repeat steps 4 7 to re-enter the new password. Display The character being entered will flash and the first character entered will remain illuminated. The character being entered will flash, and the first and second characters entered will remain illuminated. Re-Enter New Password New Password Entry OK If the re-entered password does not match the initial entry, Re-Enter Password will display, and you will need to repeat the entire process starting with step 2 on page Warranty 57

58 ENABLE/DISABLE PASSWORD FUNCTION This function allows you to activate or deactivate the password function on the OMS. Step Display 1. Push the joystick down. Enable User Password 2. Push the joystick to the right to display the status. 3. Push the joystick up or down to change the status to Enabled or Disabled, and then push the joystick in the center to enter the new status. If activated, Enabled will display. Enable User Password Main Menus and Submenus The OMS provides menus for adjusting ma outputs, alarm relay settings, sensor adjustments, and zero suppression for toxic and corrosive gas sensor cells. SET 4 20 MA LOOP This set of menus is used to adjust the 4 ma and 20 ma output from the OMS. It also provides a function to send an actual output between 4 ma and 20 ma to test any remote control and alarm system attached to the OMS. To read the ma output, the OMS must be connected to a remote PLC controller or to a SCADA system. You can also connect the OMS to a volt meter to read the ma output. Please contact ETS-Lindgren for more information. 58 Warranty

59 Step 1. From this main menu, push the joystick to the right to select the submenu used to adjust the 4 ma output. 2. To change the value, push the joystick to the right to display the 4 ma setting. 3. Push the joystick up to increase the value and push it down to decrease the value; Enter to accept the value. 4. Push the joystick to the left to go back to the previous main menu. 5. Push the joystick down to access the next submenu, which is used to adjust the 20 ma output. 6. To change the value, push the joystick to the right to display the 20 ma span setting. 7. Push the joystick up to increase the value and push it down to decrease the value; Enter to accept the value. 8. Push the joystick to the left to go back to the previous main menu. Display Set 4mA Zero The display will indicate a value between 0 and 255 counts. The 4 ma output sent from the OMS will change as the number on the display changes. Set 4mA Zero Set 20mA Span The display will indicate a value between 0 and 255 counts. The 20 ma output sent from the OMS will change as the number on the display changes. Set 20mA Span You will be able to access the Force Loop submenu, but because the Force Loop function was designed for toxic and corrosive gases, it is not available for use on the OMS. Warranty 59

60 SET FORMATS This set of menus is used to adjust the relay states for the two gas alarm relays and the individual instrument fault relay. To access this menu, the OMS must be installed with the relay module. If not installed, the display will indicate N/A (not available). Step 1. Push the joystick down to access the next main menu, Set Formats. This is used to set the two alarm relays and the fault relay settings from a normally de-energized state (Normal) to a normally energized state (Inverted). 2. Push the joystick to the right to select the submenu used to adjust the first level alarm relay state. 3. Push the joystick to the right to display the relay state. 4. Push the joystick down to change the relay state to NORMAL, and then Enter to accept the value. Display Set Formats Format Relay 1 INVERT NORMAL will display, and then will default to the Set Formats menu. 5. Push the joystick to the right. Format Relay 1 6. Push the joystick down to access the next menu, used to adjust the second level alarm relay state. Format Relay 2 60 Warranty

61 Step 7. Push the joystick to the right to display the relay state. 8. Push the joystick down to change the relay state to NORMAL, and then Enter to accept the value. Display INVERT NORMAL will display, and then will default to the Set Formats menu. 9. Push the joystick to the right. Format Relay Push the joystick twice to select the fault relay to be adjusted. 11. Push the joystick to the right to display the relay state. 12. Push the joystick down to change the relay state to NORMAL, and then Enter to accept the value. Format Fault Relay INVERT NORMAL will display, and then will default to the Set Formats menu. Warranty 61

62 SET ALARM THRESHOLD PARITY Alarm Threshold Polarity determines if an alarm concentration is set above or below a threshold value. For example, if an alarm of 19.0% for oxygen is selected, the Alarm Threshold Polarity must be set to INVERT for the alarm to activate when the reading goes below 19.0%. For toxic and corrosive gases, a NORMAL setting for the Alarm Threshold Polarity causes the system to alarm when the gas concentration exceeds/goes above an alarm set point. Step 1. Access this menu from the Set Formats menu by pushing the joystick down. Display Set Alarm Threshold Polarity 2. Push the joystick to the right. Set Alarm 1 Polarity 3. Push the joystick to the right to display the relay state. 4. Push the joystick down to change the relay state to NORMAL, and then Enter to accept the value. 5. Push the joystick down to access the next submenu. 6. Push the joystick to the right to display the relay state. 7. Push the joystick down to change the relay state to NORMAL, and then Enter to accept the value. INVERT NORMAL will display, and then will default to the Set Alarm 1 Polarity menu. Set Alarm 2 Polarity INVERT NORMAL will display, and then will default to the Set Alarm 2 Polarity menu. 62 Warranty

63 Step 8. Push the joystick down to access the next submenu. Display Set Audio Alarm Polarity To access this menu, the OMS must be installed with the audio option module. If not installed, the display will indicate N/A (not available). The optional built-in horn is designed to operate in only one alarm mode. It will activate in either a decreasing alarm or an increasing alarm mode. 9. Push the joystick to the right to display the relay state. 10. Push the joystick down to change the relay state to NORMAL, and then Enter to accept the value. INVERT NORMAL will display, and then will default to the Set Audio Alarm Polarity menu. Warranty 63

64 SET LATCHING This sets the relay alarm state for the two gas alarm relays and the individual instrument fault relay to a latching or non-latching state. In a latching state, the relay will remain activated until the user manually selects the Enter key. In a non-latching state, the alarm relay will automatically reset once the gas concentration has returned to 20.9% for oxygen. To access this menu, the OMS must be installed with the relay module. If not installed, the display will indicate N/A (not available). Step 1. Push the joystick to the right to select the submenu to adjust the first level alarm relay state. 2. Push the joystick to the right to display the relay state. 3. Push the joystick down to change the relay state to NON-LATCHING, and then Enter to accept the value. 4. Push the joystick to the right to select the submenu. 5. Push the joystick down to adjust the second level alarm relay state. 6. Push the joystick to the right to display the relay state. Display Set Latching Relay 1 LATCHING NONLATCH will display, and then will default to the Set Latching menu. Set Latching Relay 1 Set Latching Relay 2 LATCHING 64 Warranty

65 Step 7. Push the joystick down to change the relay state to NON-LATCHING, and then Enter to accept the value. 8. Push the joystick to the right to select the submenu. 9. Push the joystick twice to adjust the audio alarm relay state. Display NONLATCH will display, and then will default to the Set Latching menu. Set Latching Relay 1 Set Latching Audio Alarm To access this menu, the OMS must be installed with the audio alarm option module. If not installed, the display will indicate N/A (not available). 10. Push the joystick to the right to display the relay state. 11. Push the joystick down to change the relay state to NON-LATCHING, and then Enter to accept the value. LATCHING NONLATCH will display, and then will default to the Set Latching menu. RESET LATCHING ALARM To reset a latching alarm relay, you must enter the password correctly and then push the joystick down to send the reset command. The OMS also has an internal 2-pin terminal block for connecting a remote reset switch. Warranty 65

66 SET ALARM DELAY The Alarm Delay is the amount of time an alarm level concentration of oxygen must be present before the gas concentration alarm(s) will be activated. Use this menu to set a user-selected time delay for activating alarm relays 1 and 2, from 0 seconds to 255 seconds. When an alarm level has been exceeded, the alarm relays will activate after this period of time. Step 1. Push the joystick down to access the Alarm Delay main menu. 2. Push the joystick to the right to display the time setting function. Push the joystick up to increase the value or down to decrease the value, and then Enter to accept the value. Display Alarm Delay The display will indicate a value between 0 and 255 seconds, and then will default to the Alarm Delay menu. The alarm delay is only available for alarm 1 and alarm 2; there is no delay for the fault relay. Any system fault will immediately activate the fault relay. SET ZERO SUPPRESSION Although these settings can be changed, this function is disabled on the OMS; it is only used to decrease the sensitivity of selected gas sensors. The factory default is set at Warranty

67 SET ALARM THRESHOLDS To activate the audio alarm, the OMS must be installed with the audio alarm option module. This is used to adjust the oxygen concentration percentage that will activate alarm levels 1 and 2. If the audio alarm output module is installed, it will also be used to set the level at which the audio alarm will activate. Step 1. Push the joystick to the right to select the first submenu. 2. This is the gas concentration at which the first level alarm will be activated. To display the setting, push the joystick to the right. Push the joystick up to increase the value or down to decrease the value, and then Enter to accept the value. 3. Push the joystick to the right to select the next submenu. 4. This is the gas concentration at which the second level alarm will be activated. To display the setting, push the joystick to the right. Push the joystick up to increase the value or down to decrease the value, and then Enter to accept the value. 5. Push the joystick down to select the next submenu. Display Set Relay 1 Alarm Threshold The display will indicate a value between 00.0% and 25.5%, and then will default to the Set Relay 1 Alarm Threshold menu. Set Relay 2 Alarm Threshold The display will indicate a value between 00.0% and 25.5%, and then will default to the Set Relay 2 Alarm Threshold menu. Set Audio Alarm Threshold Warranty 67

68 Step 6. This is the gas concentration at which the audio alarm will be activated. To change the displayed value, push the joystick to the right to display the second level alarm setting. Push the joystick up to increase the value or down to decrease the value, and then Enter to accept the value. Display The display will indicate a value between 00.0% and 25.5%, and then will default to the Set Audio Alarm Threshold menu. To access this menu, the OMS must be installed with the audio alarm option module. If not installed, the display will indicate N/A (not available). The audio can be set to only one alarm level. Choose between alarm level 1 and alarm level 2, or select a different setting. 68 Warranty

69 SET ALARM HYSTERESIS The OMS may be used as a control system. When used to regulate oxygen levels, a dead band, hysteresis may be required. This menu sets the alarm hysteresis to a desired concentration of oxygen. When using hysteresis, the alarm set point becomes an average alarm setting for an action to occur. When adding the hysteresis value to the alarm set point, this then defines the alarm and dead band for an action to occur. For example, to require a valve to close at 14.9% oxygen level and to reopen at 15.5% oxygen level, set the alarm threshold at 15.2% and the hysteresis value at 0.3%. Valve off: Average alarm set point = 15.2% - Hysteresis 0.3% = 14.9% Valve on: Average alarm set point = 15.2% + Hysteresis 0.3% = 15.5% Warranty 69

70 Step 1. Push the joystick down to access the menu. Display Set Alarm Hysteresis 2. Push the joystick to the right. Set Alarm 1 Hysteresis 3. Push the joystick up to increase the percentage; the maximum value is 2.5%. Push the joystick down to decrease the percentage. Enter to accept the value. 4. Push the joystick down to access the next submenu. 5. Push the joystick to the right to display a value 0.0%. Push the joystick up to increase the percentage; the maximum value is 2.5%. Push the joystick down to decrease the percentage. Enter to accept the value. 6. Push the joystick down to access the next submenu. 7. Push the joystick to the right to display a value 0.0%. Push the joystick up to increase the percentage; the maximum value is 2.5%. Push the joystick down to decrease the percentage. Enter to accept the value. The display will indicate the selected value, and then will default to the Set Alarm 1 Hysteresis menu. Set Alarm 2 Hysteresis The display will indicate the selected value, and then will default to the Set Alarm 2 Hysteresis menu. Set Alarm Audio Hysteresis The display will indicate the selected value. 70 Warranty

71 SET SENSOR ADJUST Warm up the OMS for two hours prior to making any adjustments to the sensor. Sensor Adjust sets the oxygen readout to a known concentration. It is recommended to adjust the oxygen display to ambient oxygen levels of 20.9%. Step 1. Push the joystick down to access the menu. 2. Push the joystick to the right to access the submenu. 3. Push the joystick up to increase the counts and decrease the percent oxygen value displayed. Push the joystick down to decrease the counts and increase the oxygen value displayed. Adjust to a value of 20.9% ± 0.1%, and then Enter to accept the value. Display Sensor Adjustment Set Sensor Span The display will indicate a value between 0 and 255 counts, and then will display the selected value of 20.9% ± 0.1%. Then the display will default to Set Sensor Span. The actual oxygen reading will fluctuate from 20.8% to 21.0% 4. Push the joystick to the left to go back to the main menu. Sensor Adjustment Set Module Zero is not available for the OMS; it was designed for toxic and corrosive gas monitors. Push the joystick to the left to exit this menu. Warranty 71

72 This page intentionally left blank. 72 Warranty

73 9.0 Sensor Verification Calibrate or challenge the OMS ONLY outside the MRI suite. Before performing any maintenance, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment. Observe all safety guidelines when generating and using nitrogen. Sensor verification should be performed only by qualified personnel. Warranty may be void if performed by non-qualified personnel. WARRANTY If you have any questions concerning maintenance, contact ETS-Lindgren Customer Service. Earth is a source of calibrated oxygen at 20.9%, and therefore under ambient conditions. Verification of the Sample Draw Oxygen Monitoring System (OMS ) to 20.9% oxygen is constantly being performed. As the sensor ages over time, it may require a slight adjustment to 20.9%. The OMS also requires periodic testing with nitrogen to verify the cells response to 0% oxygen. Warranty 73

74 Required Gas and Equipment The required gas and equipment are available from your gas supplier or from Air Liquide at Nitrogen and an on-demand regulator are recommended to expose the OMS to a known span gas. You may purchase both items from your gas supplier or from Air Liquide. Air Liquide Part Number Description CZF6D Nitrogen 105 liter cylinder, 99.99% Regulator, On-Demand Sensor Verification Procedure Prior to performing the procedure, place the OMS in a clean, non-oxygen deficient environment. Warm up the OMS for two hours prior to making any adjustments to the sensor. If the OMS is connected to a controller, put the controller in standby mode to avoid accidental alarms. If required by your safety protocol, you may subject the OMS to different concentrations of oxygen span gas. 74 Warranty

75 Under ambient non-oxygen deficient environments, the OMS will display 20.9%. As the sensor ages, the reading may decrease in value. To adjust the reading to 20.9%: Step 1. Enter the password, and then push the joystick down to select the Sensor Adjustment menu. 2. Push the joystick to the right to access the submenu. 3. Push the joystick up to increase the counts and decrease the percent oxygen value displayed. Push the joystick down to decrease the counts and increase the oxygen value displayed. Adjust to a value of 20.9% ± 0.2%, and then Enter to accept the value. 4. Push the joystick to the left to return to the main menu. Display Sensor Adjustment Set Sensor Span The display will indicate a value between 0 and 255 counts, and then will display the selected value of 20.9% ± 0.2%. Then the display will default to Set Sensor Span. Sensor Adjustment Sensor Verification to Nitrogen ETS-Lindgren recommends challenging the OMS with nitrogen every 6 to 12 months. The filter on the OMS has a 1/4-in male tube fitting designed to connect sample tubing from a nitrogen cylinder. Expose the OMS to nitrogen using the on-demand regulator. The reading will drop off to 1% or below in less than one minute when the sensor is exposed to pure nitrogen. The system will recover to 20.9% when the nitrogen is removed. Warranty 75

76 Sensor Verification to Known Concentration of Oxygen Protect the OMS from wind and high airflow when gas calibrating with test gas. Test the OMS in an upright position to allow the span gas to fully saturate the sensor. To see the exact span gas concentration the OMS must be completely immersed into a span gas environment. To see a true zero, the OMS must be completely immersed into a zero oxygen environment. The sensor inlet on the OMS has a 1/4-in compression tube fitting designed to connect to the filter. Connect 1/4-in diameter sample tubing from a nitrogen cylinder directly to the filter. Expose the OMS directly from the nitrogen cylinder and the reading will drop off to the span gas concentration in less than one minute. The final reading should be within ±0.3% of the span gas concentration. 76 Warranty

77 10.0 Remote Display Alarm Indicator The Remote Display Alarm Indicator is designed to display remote oxygen concentration information from PureAire oxygen monitors. All PureAire O2 monitors have a built in 4-20mA output. The remote display alarm easily connects to the monitor s input power and ma output connection. An 18 AWG, three conductor, shielded control and instrumentation cable, Belden 8770 or equivalent, is recommended for the connection. The maximum permissible cable length is 250 feet. You can simultaneously connect one remote display to one Air Check Oxygen monitor and one programmable logic controller, PLC. Warranty 77

78 How to Mount the Remote Display Alarm Indicator 1. Open the two side doors to expose the front panel screws. 2. Loosen the 4 screws to separate the front panel from the case. 78 Warranty

79 Mount the case to a wall or other flat surface. There are 4ea. through holes for fastening the case. The case is designed to be flush mounted. The front cover can be removed from the case by gently pulling on the plastic hinges. (They will stay connected to the front cover). To connect the front cover, just push the plastic hinge into the case. Warranty 79

80 How to Wire the Remote Display Alarm Indicator Ground and shield are tied together. ma When connecting to PLC, remove jumper and connect MA to PLC+ and SHIELD to PLC-. Do not remove jumper unless connecting to PLC. 18 AWG, three conductor, shielded control and instrumentation cable, Belden 8770 or equivalent Max distance 250 feet Make all wire terminations with the power OFF. Failure to do so can damage the oxygen monitor and remote display and void your warranty. 80 Warranty

81 Identification of Switches and Controls 1. Alarm Adjusting Switches These switches are used select the alarm level to be adjusted. They are factory set at 19.5% for Alarm 1 and 18% for Alarm 2. (Normally set at the factory) 2. Alarm Polarity Switches These switches are used to select when the alarms will activate. For oxygen the alarm switches are set for Below. If the readings go below ambient, 20.9% then the internal horn and LED s will activate. When used with our toxic gas monitors, the polarity is set to Above. (Normally set at the factory) 3. Range Auto Adjustment Switches These switches are used to adjust the zero and full scale reading on the Remote Display Alarm Indicator. When sending a 4mA and 20mA signal from your oxygen monitor, pushing these switches will automatically adjust the zero and full scale readings on the Remote Digital Display. 4. Alarm Adjusting Buttons These buttons are used to increase, (UP) or decrease (DOWN) the alarm settings. 5. Horn On/Off Switch This switch is used to turn the internal audio horn on and off. In the On position, any alarm will activate the audible horn. To disable the internal horn, turn the switch to Off. 1. Alarm Adjusting Switches 2. Alarm Polarity Switches 3. Range auto adjustment switches Do not remove jumper unless connecting to PLC. 4. Alarm Adjusting Buttons 5. Horn On Off Switch Warranty 81

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information

Datasheet: K-30 ASCII Sensor

Datasheet: K-30 ASCII Sensor Datasheet: K-30 ASCII Sensor The K30 ASCII sensor is a low cost, infrared and maintenance free transmitter module intended to be built into different host devices that require CO2 monitoring data. The

More information

MODEL GT820 OXYGEN SENSOR

MODEL GT820 OXYGEN SENSOR INSTRUCTION MANUAL MODEL GT820 OXYGEN SENSOR 70046 The information and technical data disclosed by this document may be used and disseminated only for the purposes and to the extent specifically authorized

More information

Installation and Operation Manual

Installation and Operation Manual Manual Static pressure transducer with controller Differential static pressure transducer with analog output and optional PI control mode Large diaphragm element with differential transformer Transducer

More information

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

Portable Gas Monitor GX User Maintenance Manual (H4-0050) H4E-0050 Portable Gas Monitor GX-8000 User Maintenance Manual (H4-0050) Need of Maintenance and Servicing This gas monitor must be maintained in a normal state at all times to prevent accidents due to

More information

Electro-Pneumatic Converter YT-940 SERIES

Electro-Pneumatic Converter YT-940 SERIES Electro-Pneumatic Converter YT-940 SERIES PRODUCT MANUAL VERSION 1.00 Contents 1. Introduction 3 1.1 General information for the users. 3 1.2 Manufacturer Warranty 3 1.3 Explosion Proof Warning. 4 2. Product

More information

Stand-Alone Bubble Detection System

Stand-Alone Bubble Detection System Instruction Sheet P/N Stand-Alone Bubble Detection System 1. Introduction The Bubble Detection system is designed to detect air-bubble induced gaps in a bead of material as it is being dispensed. When

More information

210 Series Transmitter with External Electrochemical Sensor

210 Series Transmitter with External Electrochemical Sensor 210 Series Transmitter with External Electrochemical Sensor INSTRUCTIONS Installation and Operation of the AMC-210 Series Transmitter with External Electrochemical Sensor IMPORTANT: Please read these installation

More information

Ultima. X Series Gas Monitor

Ultima. X Series Gas Monitor Ultima X Series Gas Monitor Safety Manual SIL 2 Certified " The Ultima X Series Gas Monitor is qualified as an SIL 2 device under IEC 61508 and must be installed, used, and maintained in accordance with

More information

Met One E-BAM Particulate Monitor

Met One E-BAM Particulate Monitor STANDARD OPERATING PROCEDURES Met One E-BAM Particulate Monitor AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION PROJECT LOWER MANHATTAN DEVELOPMENT CORPORATION 1 Liberty Plaza

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

Air Check Lite Toxic Gas Monitor

Air Check Lite Toxic Gas Monitor Air Check Lite Toxic Gas Monitor Instruction Manual PureAire Monitoring Systems, Inc. 557 Capital Drive Lake Zurich, Illinois 60047 Phone: 847-726-6000 Fax: 847-726-6051 Toll-Free: 888-788-8050 www.pureairemonitoring.com

More information

RPS900W Redundant Power Supply. Installation Guide.

RPS900W Redundant Power Supply. Installation Guide. RPS900W Redundant Power Supply Installation Guide www.edge-core.com Installation Guide RPS900W Redundant Power Supply Single DC Output Port with Dual Output Voltages RPS900W E10013-CS-R01 1500000081A

More information

TR Electronic Pressure Regulator. User s Manual

TR Electronic Pressure Regulator. User s Manual TR Electronic Pressure Regulator Page 2 of 13 Table of Contents Warnings, Cautions & Notices... 3 Factory Default Setting... 4 Quick Start Procedure... 5 Configuration Tab... 8 Setup Tab... 9 Internal

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

BUBBLER CONTROL SYSTEM

BUBBLER CONTROL SYSTEM BUBBLER CONTROL SYSTEM Description: The HDBCS is a fully automatic bubbler system, which does liquid level measurements in water and wastewater applications. It is a dual air compressor system with, air

More information

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE 1 These instructions generally describe the installation, operation, and maintenance of subject equipment. The manufacturer reserves the right

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 36-751 XenoGard 133 Xe Monitor Operators Manual March 2005 Manual No. 128011 Rev. 3 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

Best Practice Guide, Servomex 2700

Best Practice Guide, Servomex 2700 For full installations details refer to the. Best Practice Guide, Servomex 2700 Mounting: General Guidelines: Servomex 2700 Control Units and air supplies (utilities units) should, ideally, be mounted

More information

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Installation, operating and maintenance Instructions for Seemag bypass level indicator Issue: S Date: 05-09-14 Type G35 General information The Seetru bypass magnetic level indicator, abbreviate SEEMAG, serves to show the filling level of fluids in tanks, basins, tubes etc. The Seemag operates

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel TANK MANAGER FOR TWO TANKS OPERATING MANUAL 10/31/11 C-More T6C L color touch panel 1 TABLE OF CONTENTS GENERAL...3 INSTALLATION...4 STONE TEST PROCEDURE...7 OPERATIONAL SUMMARY...7 AUTO CARBONATION...10

More information

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin Electronic Control Manual 216 Sensors and Stats Section S Technical Bulletin CDS-2000 Issue Date 0393 CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin Introduction 3 Pre-Verification

More information

Model TX-FM/FN Sensor Head

Model TX-FM/FN Sensor Head Model TX-FM/FN Sensor Head Instruction Manual PureAire Monitoring Systems, Inc. 557 Capital Drive Lake Zurich, Illinois 60047 Phone: 847-726-6000 Fax: 847-726-6051 Toll-Free: 888-788-8050 Pureairemonitoring.com

More information

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021-PR Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-PR For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4

More information

Aquavar SOLO 2 Frequently Asked Questions

Aquavar SOLO 2 Frequently Asked Questions Aquavar SOLO 2 Frequently Asked Questions How do I size the Aquavar SOLO 2 for the appropriate pump/motor combination? Can I use a 208 Volt motor? Can I run the Aquavar SOLO 2 up to 80HZ? What are the

More information

OC Panel High Limit Aquastat Kit, Manual Reset p/n

OC Panel High Limit Aquastat Kit, Manual Reset p/n OC Panel High Limit Aquastat Kit, Manual Reset p/n 233202 Instruction Sheet APPLICATION The OC (Option Control) Panel High Limit Aquastat Kit provides electronic temperature sensing in a UL limit-rated

More information

RAM Operation Manual

RAM Operation Manual RAM 4021-1 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-1 For Your Safety... 2 Description... 2 Setup Mode... 3 Lights/Alarms... 3 Operation... 4 Calibration...

More information

Workhorse-15 Oxygen Generator

Workhorse-15 Oxygen Generator 11436 Sorrento Valley Road, San Diego CA 92121 USA 858-558-0202 Fax 858-558-1915 Workhorse-15 Oxygen Generator Model No. 5719, 5727, 5735 for Domestic and International Units Part number 1432 Revision

More information

Alphasense 4-20 ma transmitters offer convenience and easy maintenance for toxic sensors:

Alphasense 4-20 ma transmitters offer convenience and easy maintenance for toxic sensors: ALPHASENSE USER MANUAL Page 1 4-20mA Transmitter for Toxic Sensors UMTOX-1 Issue 4 1 INTRODUCTION The Transmitter PCB includes circuitry for a three electrode toxic sensor to convert the µa output signal

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

OEM Manual. MODEL ½ Digit DRUM SCALE

OEM Manual. MODEL ½ Digit DRUM SCALE OEM Manual MODEL 4020-3 ½ Digit DRUM SCALE Scaletron Industries, Ltd. Bedminster Industrial Park 53 Apple Tree Lane P.O. Box 365 Plumsteadville, PA 18949 USA Toll Free: 1-800-257-5911 (USA & Canada) Phone:

More information

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL TABLE OF CONTENTS PRODUCT OVERVIEW 2 I - USE AND CARE: 3 II - INSTALLATION: 5 III - GENERAL MAINTENANCE TIPS: 5 IV - APPENDIX: A-1 2-WIRE CURRENT LOOP

More information

H-DIL Hydrogen/Oxygen Sample-Draw Detector Operator s Manual

H-DIL Hydrogen/Oxygen Sample-Draw Detector Operator s Manual 35-3001-06H-DIL Hydrogen/Oxygen Sample-Draw Detector Operator s Manual Part Number: 71-0346 Revision: A Released: 8/7/17 www.rkiinstruments.com WARNING Read and understand this instruction manual before

More information

Pressure Automated Calibration Equipment

Pressure Automated Calibration Equipment GE Measurement & control Pressure Automated Calibration Equipment Safety Instructions and User Guide - K0447 PACE5000 PACE6000 K0447 Issue No. 9 1 10 1 PACE5000 1 2 3 4 5 PACE6000 2 6 7 8 3 4 5 6 7 8 9

More information

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272) Manual Addendum Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272) P/N M07273 DATE 06/11/13 TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments 16830 Chestnut Street City of Industry,

More information

CSA Sample Draw Aspirator Adapter Operator s Manual

CSA Sample Draw Aspirator Adapter Operator s Manual 30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

Operation & Maintenance Manual GDS-68XP

Operation & Maintenance Manual GDS-68XP Operation & Maintenance Manual GDS-68XP Process Gas Monitor for Low Oxygen Applications GDS Corporation 2513 Hwy 646 Santa Fe, Texas 77510 (409) 927-2980 (409) 927-4180 (Fax) www.gdscorp.com CAUTION: FOR

More information

LVU2800 Series. Ultrasonic Level Transmitter

LVU2800 Series. Ultrasonic Level Transmitter LVU2800 Series Ultrasonic Level Transmitter 2 of 23 INTRODUCTION / TABLE OF CONTENTS Step One The LVU2800 Series is a general purpose ultrasonic level transmitter that provides a loop powered 4 20 ma output.

More information

GAS FUEL VALVE FORM AGV5 OM 8-03

GAS FUEL VALVE FORM AGV5 OM 8-03 ALTRONIC AGV5 OPERATING MANUAL GAS FUEL VALVE FORM AGV5 OM 8-03 WARNING: DEVIATION FROM THESE INSTALLATION INSTRUCTIONS MAY LEAD TO IMPROPER ENGINE OPERATION WHICH COULD CAUSE PERSONAL INJURY TO OPERATORS

More information

HLM-2000-TX Gas Transmitter User Manual

HLM-2000-TX Gas Transmitter User Manual HLM-2000-TX Gas Transmitter User Manual Technical Support Continental North America Toll Free 1-(800) 387-9487 Ph: +1 (905) 829-2418 Fx: +1 (905) 829-4701 A Product of Arjay Engineering Ltd. Oakville,

More information

GG-VL2-NH3 AMMONIA VENT LINE SENSOR. Installation and Operation Manual

GG-VL2-NH3 AMMONIA VENT LINE SENSOR. Installation and Operation Manual GG-VL2-NH3 AMMONIA VENT LINE SENSOR Installation and Operation Manual 2 GG-VL2-NH3 Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified

More information

UBEC 1AT. AUTO TANK Fill System Installation, Operation, & Setup Instructions

UBEC 1AT. AUTO TANK Fill System Installation, Operation, & Setup Instructions Document Number: XE-ATA5PM-R1A UBEC 1AT AUTO TANK Fill System 08899155 Installation, Operation, & Setup Instructions Rev170906-EB-FRC PHYSICAL: 1302 WEST BEARDSLEY AVE ELKHART, IN 46514 WWW.ELKHARTBRASS.COM

More information

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL Operating and Installation Manual Warning Use this product only in the manner described in this manual. If the equipment is used in a

More information

RAM 4021-DPX Operation Manual

RAM 4021-DPX Operation Manual RAM 4021-DPX Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS ABL 4021-DPX / RAM 4021-DPX For Your Safety... 3 Description... 3 Setup Mode... 4 Lights/Alarms... 4 Operation...

More information

User Manual. asense VAV. CO 2 / temperature sensor with built-in general purpose controller

User Manual. asense VAV. CO 2 / temperature sensor with built-in general purpose controller Gas and Air Sensors User Manual asense VAV CO / temperature sensor with built-in general purpose controller General The IAQ-sensor product asense VAV is used to measure indoor air carbon dioxide concentration

More information

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2. FANs 216, 1628.3 Technical Bulletin CDS4000 Issue Date 0797 CDS4000 CO 2 Sensor Calibration and Troubleshooting Table of Contents Introduction 1 Handling Information 1 Calibration 2 Preparation 2 Cautions

More information

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values Operations Manual Please read, understand, and follow these instructions before operating this equipment.

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0578 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

780S Ultra High Purity Series

780S Ultra High Purity Series Smart Ultra High Purity Thermal Gas Mass Flow Meter Features Measures mass flow directly, no seperate temperature or pressure inputs required Field adjustment of critical flow meter settings via on-board

More information

Overview. Front Panel: Keypad and Display

Overview. Front Panel: Keypad and Display Overview The GA-200B is an analyzer that integrates a gas sampling system with sensors to measure and display the concentrations of oxygen and carbon dioxide in a sample as the percentage of a gas in the

More information

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1)

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1) ENGLISH ENGLISH ENGLISH ENGLISH Installation user s manual Warning! electrical scheme modified JANUARY 2005 version 3.3 JOLLY2 DATA TO BE FILLED OUT BY THE INSTALLER (Page 1) 6 different operating modes

More information

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES PROCESS ANALYSERS DF-310E The DF-310E is a Coulometric sensor based oxygen analyzer designed to measure trace and percent level oxygen in pure and multi-gas backgrounds for process and quality control

More information

700 SERIES BMS II BUBBLER MONITORING SYSTEM

700 SERIES BMS II BUBBLER MONITORING SYSTEM 1 700 SERIES BMS II BUBBLER MONITORING SYSTEM VISIT OUR WEBSITE AT SIGMACONTROLS.COM 700 SERIES BMS II MANUAL 062114 2 TABLE OF CONTENTS INTRODUCTION 3 Physical Specifications WIRING 5 PROGRAMMING & SETUP..

More information

571 Series Medical TM. IntelliSwitch Electronic Switchover

571 Series Medical TM. IntelliSwitch Electronic Switchover ADI 9505-H 571 Series Medical TM IntelliSwitch Electronic Switchover INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Carefully Read These Controls Corporation

More information

ECS Protector Nitrogen Generator PGEN-30 (PGEN-30E)

ECS Protector Nitrogen Generator PGEN-30 (PGEN-30E) ECS Protector PGEN-30 (PGEN-30E) Specifications For use under U.S. Patents 8,720,591, 9,144,700 and 9,186,533 Dimensions (cabinet): 24.5 (W) x 52.5 (H) x 8.5 (D) (622mm(W) x 1,334mm(H) x 216mm(D)) Dimensions

More information

DPC-30 DPC-100. Reference Manual

DPC-30 DPC-100. Reference Manual DPC-30 DPC-100 Reference Manual 1. Introduction 1.1 Description The Martel DPC Digital Pneumatic Calibrator improves upon traditional dial gauge pneumatic calibrators. The Martel DPC improves accuracy,

More information

BUBBLER CONTROL SYSTEM

BUBBLER CONTROL SYSTEM BUBBLER CONTROL SYSTEM Description: The LDBCS is a fully automatic bubbler system, which does liquid level measurements in water and wastewater applications. It is a dual air compressor system with, air

More information

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual Air Bubbler Depth Gauge DG2200 Installation and Reference Manual Rev. 3.2.2 06/2011 Installation and Reference Manual Page 2 Contents Chapter 1: Installation and Overview... 3 1.1 DG2200 Location Diagram...

More information

Technical Data Sheet MF010-O-LC

Technical Data Sheet MF010-O-LC Technical Data Sheet MF010-O-LC - 1 - 1. Properties The oxygen measuring system MF010-O-LC determines the oxygen content in gas mixtures up to a temperature of 250 C. It is particularly suitable for the

More information

SPECIFICATIONS ATTENTION

SPECIFICATIONS ATTENTION VPS 504 S06 Installation Manual - P/N 80122 - Ed. 01/09 VPS 504 S06 and S05 Valve Proving System Installation Instructions VPS 1 6 Gases Natural gas, air and other inert gases. NOT suitable for butane

More information

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties Quick Installation Guide Doc nr.: 9.17.052H Date: 10-12-2015 SCOPE OF THIS GUIDE mini CORI-FLOW instruments are highly accurate instruments for measuring and controlling the mass flow rate of liquids and/or

More information

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel: PERMA PURE INSTRUCTION MANUAL FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model 3300 Version 4.06 Perma Pure LLC Tel: 732-244-0010 P.O. Box 2105, 8 Executive Drive Tel: 800-337-3762 (toll free US)

More information

RAM 4021 Operation Manual

RAM 4021 Operation Manual RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6

More information

HLM-2000-TX Gas Transmitter User Manual

HLM-2000-TX Gas Transmitter User Manual HLM-2000-TX Gas Transmitter User Manual Technical Support Continental North America Toll Free 1-(800) 387-9487 Ph: +1 (905) 829-2418 Fx: +1 (905) 829-4701 A Product of Arjay Engineering Ltd. Oakville,

More information

OXYGEN MONITORING SYSTEM OMS

OXYGEN MONITORING SYSTEM OMS User Manual OXYGEN MONITORING SYSTEM OMS 420 Read and observe before commissioning! Ex Situ oxygen analyzer based on ZrO 2 sensor ENGLISH Caution! Please inspect in the presence of the transporting agent

More information

Manual Weighingblock VB2 series and Uniscale

Manual Weighingblock VB2 series and Uniscale Manual Weighingblock VB2 series and Uniscale Note: At page 8 in this manual you will find a short form instruction. Normally the only instruction shipped together with the Scale. Overview different ranges.

More information

Installation and Operation Manual

Installation and Operation Manual Installation and Operation Manual WIKA FLR-SBDF / BLR-SBDF Magnetic Level Transmitter (Please retain for future usage) Contact: Gayesco-WIKA USA, L.P. 229 Beltway Green Boulevard Pasadena, TX 77503 www.wika.com

More information

Transmitter CS 21 Operation Manual

Transmitter CS 21 Operation Manual Transmitter CS 21 Operation Manual Content Page For your Safety 3 General Description 3 Detection Principle 4 Operational Notes 4 Design 4 Mounting Position of CS21 5 Mounting 6 Installation of Electrical

More information

Features. Description

Features. Description Features Fast response flow meter ideal for inert gas and liquid mass flow measurement applications Smart electronics permit field adjustment of critical flow meter settings Field validation of flow meter

More information

Sensoric 4-20 ma Transmitter Board Operation Manual

Sensoric 4-20 ma Transmitter Board Operation Manual Sensoric 4-20 ma Transmitter Board Operation Manual 1 Content Features Operation Manual Technical Data Mechanical Dimensions Remarks & Contact information 2 Features Soldered sensor cell (non replaceable)

More information

UsER manual for Watersens ph -REDOX

UsER manual for Watersens ph -REDOX UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5

More information

Calibration Requirements for Direct Reading Confined Space Gas Detectors

Calibration Requirements for Direct Reading Confined Space Gas Detectors : Calibration Requirements for Direct Reading Confined Space Gas Detectors However, the definition of bump test has always been a little slippery. Some manufacturers differentiate between a bump test that

More information

GasSense NDIR User Manual

GasSense NDIR User Manual INDEX INDEX... 1 1. OVERVIEW... 2 2. TECHNICAL DATA... 3 3. SPECIFICATIONS... 4 4. PRODUCT DESCRIPTION... 5 4.1 Mechanical details... 5 4.2 Piping... 7 4.3 Connections... 7 5. INSTALLATION... 10 6. CALIBRATION

More information

AMT-Ex Dewpoint Transmitter

AMT-Ex Dewpoint Transmitter AMT-Ex Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk

More information

with O 2 Controller Instruction Manual

with O 2 Controller Instruction Manual 14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Hypoxic Cabinets (In-Vivo / In-Vitro) with O 2 Controller Instruction Manual Index Page Warranty 2 Warnings

More information

RK-IR Sample Draw Aspirator Adapter Operator s Manual

RK-IR Sample Draw Aspirator Adapter Operator s Manual 30-0951RK-IR Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0018RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr A DIVISION OF THE FREDERICKS COMPANY 2400 PHILMONT AVE. HUNTINGDONVALLEY, PA 19006 PARTS LIST 1 3 4 2 # QTY ITEM DESCRIPTION

More information

CF8-W-Disp-CO. User manual. CO 2 / CO sensor with built-in general purpose controller

CF8-W-Disp-CO. User manual. CO 2 / CO sensor with built-in general purpose controller User manual CF8-W-Disp-CO CO 2 / CO sensor with built-in general purpose controller General The IAQ-sensor product CF8-W-Disp-CO is used to measure indoor air carbon dioxide and carbon monoxide concentrations.

More information

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES PROCESS ANALYSERS DF-550E The DF-550E Nano Trace oxygen analyser is a Coulometric sensor based analyser designed to measure oxygen as a contaminant at ultra trace levels in ultra high purity electronic

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

EASIDEW TRANSMITTER with Current Source Output

EASIDEW TRANSMITTER with Current Source Output EASIDEW TRANSMITTER with Current Source Output INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue March 2002 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element

More information

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1 Dry 131h 1. DESCRIPTION The Viking Model G-1 Maintenance Air Compressor is an electric motor-driven, aircooled, single-stage, oil-less compressor. The unit is equipped with a check valve and provides a

More information

Ventam 85 Installation & Commissioning Instructions

Ventam 85 Installation & Commissioning Instructions Ventam Systems Ltd Unit D4 Seedbed Business Centre Vanguard Way Shoeburyness Essex SS3 9QY Phone 01702 382 307 Fax 01702 382 340 Ventam 85 Installation & Commissioning Instructions 1 General The Ventam

More information

Model Description

Model Description LEAKHUNTER Plus LEAKHUNTER Plus Model 8066 Description The MATHESON LEAKHUNTER Plus is a truly universal leak detector. Extensive R&D has produced a multi-functional leak detector engineered to perform

More information

MODEL 400 CLD INSTRUCTION MANUAL

MODEL 400 CLD INSTRUCTION MANUAL MODEL 400 CLD INSTRUCTION MANUAL This manual describes installation, calibration and operation of oxides of nitrogen gas analyzer. To assure correct operation and accurate results, it is recommended that

More information

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS Using state-of-the-art measuring and microprocessor technologies, Advanced has assembled a series of systems which can measure mass

More information

Digital Vacuum Regulator

Digital Vacuum Regulator Temperature Control for Research and Industry Digital Vacuum Regulator User s Manual Model 300 INDEX SECTION PAGE 1. QUICK OPERATING INSTRUCTIONS........................... 3 Safety Notices.................................................

More information

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc. Advanced Calibration Designs, Inc. Calibration Gas Instrument INSTRUCTION MANUAL Release I www.goacd.com Instruction Manual Gas Generator Release I TABLE OF CONTENTS I. General Description Page 2 II. Start-Up

More information

MP15 Jockey Pump Controller

MP15 Jockey Pump Controller Setup and Operating Instructions MP15 Jockey Pump Controller This manual provides general information, installation, operation, maintenance, and system setup information for Metron Model MP15 Jockey Pump

More information

Detcon MicroSafe Model DM-534C Oxygen Deficiency Sensor (0-25% O2)

Detcon MicroSafe Model DM-534C Oxygen Deficiency Sensor (0-25% O2) Detcon MicroSafe Model DM-534C Oxygen Deficiency Sensor (0-25% O2) Operator s Installation and Instruction Manual DETCON, Inc. 4055 Technology Forest Blvd, Suite 100, The Woodlands, Texas 77381 Ph.281.367.4100

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0545 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER Features ±0.01% FS Measurement & Control Accuracy ±0.001% /ºC Thermal Stability Pressure Ranges from ±1 psid to 1200 psia Applications Barometric Measurement

More information

Quick Installation Guide Doc nr.: C Date:

Quick Installation Guide Doc nr.: C Date: Quick Installation Guide Doc nr.: 9.17.093C Date: 13-05-2016 SCOPE OF THIS GUIDE mini CORI-FLOW instruments are highly accurate instruments for measuring and controlling the mass flow rate of liquids and/or

More information

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V O2100C Oxygen Measurement Module The O2100C module measures the partial pressure of O2 and thus the module output is proportional to the pressure in the sample cell. Gas sampled must be free of liquids

More information

Operation Manual. Rev. C LPT Low Power Transmitter (Electrochemical Sensor)

Operation Manual. Rev. C LPT Low Power Transmitter (Electrochemical Sensor) Operation Manual Rev. C 2014.04 LPT Low Power Transmitter (Electrochemical Sensor) www.critical-environment.com LPT - Operation Manual Rev. C 2014.04 TABLE OF CONTENTS 1 POLICIES...4 1.1 Important Note...4

More information

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves

More information

Simple Set. Pressure Independent Control Valves 2 Way 1/2-2

Simple Set. Pressure Independent Control Valves 2 Way 1/2-2 The Bray Simple Set is a threaded pressure independent control (PIC) valve designed for a wide variety of hot water and chilled water control applications. The SS Series combines high rangeability control

More information