Application Note 13: Electron Microscope Decontamination

Size: px
Start display at page:

Download "Application Note 13: Electron Microscope Decontamination"

Transcription

1 The Chlorine Dioxide People Providing you with gaseous chlorine dioxide solutions for your decontamination needs Application Note 13: Electron Microscope Decontamination M Uses: An electron microscope is sometimes used for studying dangerous biological organisms. Occasionally the organism can be sucked into the internals of the microscope making it hazardous to repair the microscope for concern of the organism causing health problems for the repair technician. To mitigate the concerns, decontaminating the inside components of the microscope can be accomplished using gaseous chlorine dioxide. Benefits: Quicker cycles with Chlorine Dioxide (CD) Gas than Vapor Phase Hydrogen Peroxide (VPHP) 1.5 to 3 hours depending on desired level of kill and sensitivity of components vs. 8 to 12 hours for VPHP. No cycle development required for CD gas CD: 1 mg/liter for 2 hours. VPHP: Cycle parameters must be developed for every specific application. If ambient temperatures change, the cycle parameters most likely need to be changed. Better distribution with a true gas like CD gas CD gas is a true gas which naturally fills the space it is contained within, no matter the shape or amount of items inside the space. VPHP is a liquid at room temperature and as such has limited natural diffusion. The small internal tubing diameters within the microscope are not large enough for vapors to flow and decontaminate the internals. Better & Consistent Kill with CD gas CD gas has better and more consistent kill. The standard error or variability for CD gas is much lower than all other methods of decontamination (see figure 1).

2 Insufficient Kill Not Tested Error Bars represent variation or amount of error from cycle to cycle (Repeatability of process) Dashed Line represents 4-log reduction Figure 1: Observed median Log reduction by fumigation system and location Ref. Beswick Alan J., et al, Comparison of Multiple Systems for Laboratory Whole Room Fumigation, Applied Biosafety Vol. 16, No. 3, CD gas is less harsh to electronics Chlorine dioxide gas has a lower oxidation potential than ozone, peracetic acid, bleach and hydrogen peroxide, making it scientifically less corrosive than those other decontaminating agents (see Table 1). Biocidal Agent Oxidation Potential (volts) Oxidation Capacity (electrons) Ozone e Peracetic acid e Hydrogen peroxide e Bleach e Chlorine dioxide e Table 1: Oxidation Potential of various decontamination agents Real-life testing performed by the US EPA confirms that CD gas is less corrosive on computers than hydrogen peroxide (see figure 2). Testing on electron microscopes found significant corrosion with VHP but only damage to a phosphorous screen with CD. CD Gas Features: Sterilization at ambient temperatures short cycle times precise concentration monitoring Uses a true gas excellent distribution into hard to reach areas simple to validate detailed cycle reporting no liquids in process does not require tight control of dew point quick aeration (can literally be minutes) non-carcinogenic non-flammable no measurable residuals does not condense out or breakdown during the process Figure 2: Computer Failures as reported by US EPA Snyder, Emily, Indoor and Outdoor Decontamination Presentation at EPA Region 9 / ORD Homeland Security Research Workshop, July 14, 2011 San Francisco, CA. Accessed from Accessed on

3 Equipment Required: The equipment required to decontaminate the inside of an electron microscope consists of: Minidox-M Portable CD Generator ClorDiSys Microscope Decontamination System Microscope Manufacturer Valve Sequencing System Electron Microscope Equipment Setup: The setup of the equipment is per the below diagram: The setup shown allows the gas generator to cycle through the sterilization steps, described under system operation. All components below the dotted line are supplied by ClorDiSys. The components above the dotted line is the microscope itself and also the microscope valve sequencing system which allow isolation of and sequencing specific column sections, etc. This has been done by the microscope manufacturer in the past, but can be done by ClorDiSys if desired.

4 Minidox-M Configuration: The setup of the Minidox-M CD Generator is as follows: Minidox-M Configurable settings FC-1 Pump This FC is connected to an AC/DC control box that turns on/off the pump that drives the loop In the configurable parameters this is turned on in every step of the cycle. FC-2 Signal to microscope This is a signal that is sent to the microscope that tells it when we are in exposure and aeration. The microscope will then open vaives to let the mixture into the system. In the configurable parameters, this is turned on during exposure and aeration. FC-3 Pre-Loop Valve This creates a loop that allows the mix box to circulate the humidity and gas mixture before it is sent into the microscope. In the configurable parameters this is turned on during pre-condition, condition, and charge. FC-4 Force Aeration Valve This valve is open during normal circulation but closes for pre-purge of nitrogen and aeration. (These two steps are the same vaive sequences, but one is before the cycle and one is at the end.) When the valve is closed it will force nitrogen into the system and expel everything out through the scrubber. (FC-5 must be turned on for this to work.) In the configurable parameters this is turned on during pre-condition, condition, charge, and exposure. FC-5 Aeration Valves Includes Nitrogen inlet valve and Scrubber Valve During the nitrogen pre-purge and aeration these valves are opened to allow nitrogen to enter the system and to aiiow all the gasses to exit through the scrubber. (FC-4 should be off to disrupt the closed loop system. With FC-4 off the pump will pull the nitrogen in and force everything out through the scrubber. In the configurable parameters this is turned on during leak test (we are using this step as the nitrogen pre-purge) and during aeration. Steam inject Turns on Humidifier Configurable parameter 3 (RHSP- X) is 15 Configurable parameter 4 (RH pause) is 120 seconds Configurable parameter 6 (RH burst) is 20 seconds Pressure Relief Relieves pressure Relief SP set to 600 (maximum) High pressure alarm set to 1500 CD Inject parameters Configurable parameter (CD pause) 120 sec Configurable parameter (Charge CD Burst) 0.5 seconds Configurable parameter (Exp CD Burst) 0.5 seconds

5 Equipment Operation: The operation of the equipment is as follows: The normal sterilization process is automated and consists of 5 steps: 1. Precondition: Raising of humidity to make spores susceptible to gas. This is achieved by using the RH probe in the loop to read humidity and then turning on the steam generator located in the mix box as needed to adjust the RH. 2. Condition: Holding of raised humidity level for spore softening. 3. Charge: Injection of gas into chamber This is achieved by injecting CD gas into the CD Gas Inject Tee until the photometer measures that the concentration is reached. 4. Exposure: Holding of gas concentration for the set amount of time. 5. Aeration: Expulsion of gas and humidity. Some microscope manufactures add a sixth step which is a pre-purge of the system with nitrogen. The sequencing of the valves are shown for the various steps of the cycle: If a Pre-Purge step is used the valves shown are opened and nitrogen is passed through the system. This should not be a necessary step. Note that the microscope is only open to the loop during the exposure and aeration steps. The microscope could be open during the Pre-Purge step if desired.

6 During the Pre-condition, Condition, and Charge steps the shown valves are opened. During these steps the microscope is not part of the loop. The reason for doing this is for some of the microscopes chambers there are small orifices that greatly restrict flow, so most choose to premix the humidity and gas to the desired amounts prior to sending the mixture into the microscope. When the volume of the selected microscope chamber is small relative to the volume of the mixbox and the rest of the loop, the adding of the microscope into the loop for exposure has little effect on the concentration levels of the mixture.

7 During the Exposure step the microscope s valves are opened to the loop. This loop is kept recirculating until the desired ppm hours are reached.

8 During the Aeration step the loop goes from circulating to a once through with air or nitrogen going into the system and the gas mixture leaving through a scrubber.

9 Microscope Valve Sequencing System Equipment Setup: The setup of the microscope valve sequencing system equipment is per the below diagram. This system is the interface between the CD generation system and the microscope. The valves within this system are sequenced to both get and remove the gas from the microscope.

Effects of Relative Humidity, Concentration, and Exposure Time on the Efficacy of Chlorine Dioxide Gas Decontamination

Effects of Relative Humidity, Concentration, and Exposure Time on the Efficacy of Chlorine Dioxide Gas Decontamination Revision Date: October 15, 2006 American Biological Safety Association 49th Annual Biological Safety Conference Effects of Relative Humidity, Concentration, and Exposure Time on the Efficacy of Chlorine

More information

H 2 O 2 Bio-Decontamination and Aseptic Transfer of Heat Sensitive Materials Sergio Mauri, Director, Global Marketing and Business Intelligence

H 2 O 2 Bio-Decontamination and Aseptic Transfer of Heat Sensitive Materials Sergio Mauri, Director, Global Marketing and Business Intelligence H 2 O 2 Bio-Decontamination Process and Aseptic Transfer of Heat Sensitive Materials Sergio Mauri, Director, Global Marketing and Business Intelligence Presentation Outline Surface bio-decontamination

More information

DESIGN AND INSTALLATION CONSIDERATIONS FOR TSI AEROTRAK VHP-RESISTANT REMOTE PARTICLE COUNTERS

DESIGN AND INSTALLATION CONSIDERATIONS FOR TSI AEROTRAK VHP-RESISTANT REMOTE PARTICLE COUNTERS DESIGN AND INSTALLATION CONSIDERATIONS FOR TSI AEROTRAK VHP-RESISTANT REMOTE PARTICLE COUNTERS APPLICATION NOTE CC-111 (A4) Introduction A widely used method to inactivate bio-contamination on surfaces

More information

Mold Remediation of a Research Facility in a Hospital

Mold Remediation of a Research Facility in a Hospital Mold Remediation of a Research Facility in a Hospital Kevin Lorcheim* and Paul Lorcheim ClorDiSys Solutions, Inc., Lebanon, New Jersey Abstract A hospital s life science research facility experienced flooding

More information

Introduction to Ethylene Oxide (EO/EtO) Sterilization

Introduction to Ethylene Oxide (EO/EtO) Sterilization Introduction to Ethylene Oxide (EO/EtO) Sterilization About Ethylene-Oxide EO is a gas that is 1.5 times heavier than air and tends to settle along the floor. At concentrations below 500 ppm, EO is colorless

More information

Using Vaporized Hydrogen Peroxide

Using Vaporized Hydrogen Peroxide for Bio-Decontamination D-111046-2013 Using Vaporized Hydrogen Peroxide In 1994, Dräger introduced the first electrochemical hydrogen peroxide (H2O2) sensor for monitoring low concentrations of vaporized

More information

Walk in Modular Material Transfer Chamber Including Integration of a Bioquell Clarus L2 hydrogen peroxide vapour (HPV) generator

Walk in Modular Material Transfer Chamber Including Integration of a Bioquell Clarus L2 hydrogen peroxide vapour (HPV) generator Walk in Modular Material Transfer Chamber Including Integration of a Bioquell Clarus L2 hydrogen peroxide vapour (HPV) generator Page 1 of 20 Walk in chamber and HPV integrated system - Product overview.

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

3M Steri-Vac Sterilizer/Aerator GSX Series. For Life Science Applications. Product. Profile

3M Steri-Vac Sterilizer/Aerator GSX Series. For Life Science Applications. Product. Profile M Steri-Vac Sterilizer/Aerator GSX Series Product Profile Product Profile M Steri-Vac Sterilizer/Aerator GSX Series ethylene oxide (EO) sterilizer/aerators are designed to be used with M Steri-Gas EO Gas

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

Steritech: A World Leader in Sterilisation and Decontamination

Steritech: A World Leader in Sterilisation and Decontamination ain Sterilization Technologies Ethylene Oxide (55%) Surface sterilant Typically 100% Sterilize under vacuum Process variables include Temperature RH Time Pressure EO concentration Preconditioning / Aeration

More information

Gas Injection Systems

Gas Injection Systems NSF International Recreational Water Program Component Certification Specification CCS-14530 Gas Injection Systems The Public Health and Safety Company. TM NSF Recreational Water Program Component Certification

More information

HVAC Considerations for Gaseous Decontamination of Laboratory Spaces using Hydrogen Peroxide Vapor

HVAC Considerations for Gaseous Decontamination of Laboratory Spaces using Hydrogen Peroxide Vapor HVAC Considerations for Gaseous Decontamination of Laboratory Spaces using Hydrogen Peroxide Vapor Copyright 2006 Phoenix Controls Corporation The material in this paper is for information purposes only

More information

Generating Calibration Gas Standards

Generating Calibration Gas Standards Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and

More information

Transfer Procedures. By the end of this chapter, you will be able to: The Transfer Process. Chapter 12

Transfer Procedures. By the end of this chapter, you will be able to: The Transfer Process. Chapter 12 Chapter 12 Transfer Procedures By the end of this chapter, you will be able to: l Explain the term transfer process l Name two types of sanitisation methods often used l Describe the spray and wipe technique

More information

ADVANCED OXIDATION PROCESS

ADVANCED OXIDATION PROCESS ADVANCED OXIDATION PROCESS APG OZONE + UV-C OXIDATION DESTROYS 99.9% OF DANGEROUS CONTAMINANTS THE MOST POWERFUL ADVANCED SANITIZER c-m-p.com/del AOP: THE MOST POWERFUL ADVANCED SANITIZER THE ADVANTAGES

More information

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD) Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet

More information

Technology issues regarding Blends of Refrigerants Buffalo Research Laboratory

Technology issues regarding Blends of Refrigerants Buffalo Research Laboratory Technology issues regarding Blends of Refrigerants Buffalo Research Laboratory Agenda Basic Characteristics Background Degree of Superheat, degree of subcooling, average coil temperature Handling and setting

More information

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information

Mass Split of Two-Phase-Flow

Mass Split of Two-Phase-Flow Mass Split of Two-Phase-Flow The split of two-phase-flow at horizontal side-t-junctions in unbalanced pipe systems for clean extinguishing agents Speaker: Gudrun Fay Minimax 2010 2013 Content 1. Applications

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW AC 2010-2145: MEASUREMENT OF HYDROGEN IN HELIUM FLOW Randy Buchanan, University of Southern Mississippi Christopher Winstead, University of Southern Mississippi Anton Netchaev, University of Southern Mississippi

More information

A Discussion of Biological Safety Cabinet Decontamination Methods: Formaldehyde, Chlorine Dioxide, and Vapor Phase Hydrogen Peroxide

A Discussion of Biological Safety Cabinet Decontamination Methods: Formaldehyde, Chlorine Dioxide, and Vapor Phase Hydrogen Peroxide A Discussion of Biological Safety Cabinet Decontamination Methods: Formaldehyde, Chlorine Dioxide, and Vapor Phase Hydrogen Peroxide Mark A. Czarneski* and Kevin Lorcheim ClorDiSys Solutions, Inc., Lebanon,

More information

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,

More information

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi

More information

Chapter 3 EXPERIMENTAL: EQUIPMENT, SET-UP, AND PROCEDURE

Chapter 3 EXPERIMENTAL: EQUIPMENT, SET-UP, AND PROCEDURE Chapter 3 EXPERIMENTAL: EQUIPMENT, SET-UP, AND PROCEDURE 72 3.0 Introduction The experimental work was carried out in three stages. First stage: a laboratory scale single nozzle horizontal jet ejector

More information

FCTechnik. flox[control] Gas Control Units for Aluminium Production. for aluminium production flox[control] flox[on] FC Technik 1

FCTechnik. flox[control] Gas Control Units for Aluminium Production. for aluminium production flox[control] flox[on] FC Technik 1 FCTechnik flox[control] Gas Control Units for Aluminium Production for aluminium production flox[control] flox[on] FC Technik 1 THE company flox[on] series FC Technik is a company with headquarters in

More information

Phosgene Sensor. Sensoric COCl2 3E 1

Phosgene Sensor. Sensoric COCl2 3E 1 Phosgene Sensor 1 FEATURES Amperometric 3 electrode sensor cell Fixed organic electrolyte High reliability TYPICAL APPLICATIONS Ambient monitoring of TLV levels Chemical Industry, Homeland Security PART

More information

Protective Atmospheres, Measurement Technologies and Troubleshooting Tools

Protective Atmospheres, Measurement Technologies and Troubleshooting Tools Protective Atmospheres, Measurement Technologies and Troubleshooting Tools Furnace atmospheres are critical to meet metallurgical specifications defined by control processes. The makeup of a furnace s

More information

Pressure Automated Calibration Equipment

Pressure Automated Calibration Equipment GE Measurement & control Pressure Automated Calibration Equipment Safety Instructions and User Guide - K0447 PACE5000 PACE6000 K0447 Issue No. 9 1 10 1 PACE5000 1 2 3 4 5 PACE6000 2 6 7 8 3 4 5 6 7 8 9

More information

What is the D-ifference in D-value?

What is the D-ifference in D-value? Volume 3, Number 3 September 2006 Garrett Krushefski Scientific & Technical Services Manager What is the D-ifference in D-value? Occasionally our Customer Service Department receives a request for biological

More information

International 17/10/14 WORKING PAPER. Rio de NITRIC OXIDE. SUMMARY The DGP. is requested. Oxide Quantities. (5 pages)

International 17/10/14 WORKING PAPER. Rio de NITRIC OXIDE. SUMMARY The DGP. is requested. Oxide Quantities. (5 pages) International Civil Aviation Organization DGP-WG/14-IP/4 17/10/14 WORKING PAPER DANGEROUS GOODS PANEL (DGP) MEETING OF THE WORKING GROUP OF THE WHOLE Rio de Janeiro, Brazil, 20 to 244 October 2014 Agenda

More information

Fundamentals of NGL Sampling Systems. By Dominic Giametta & Jim Klentzman

Fundamentals of NGL Sampling Systems. By Dominic Giametta & Jim Klentzman Fundamentals of NGL Sampling Systems By Dominic Giametta & Jim Klentzman The purpose of this paper is to discuss in depth the systems we use as a standard to sample natural gas liquids, or NGL s. Before

More information

ProSys Laboratory Analytical Isolators

ProSys Laboratory Analytical Isolators Custom and standard isolators for analytical and QA laboratories Contained handling of samples for dispensing and analysis Integration of a wide range of different weigh scale systems and other instrumentation

More information

Air Removal from the Sterilization Chamber

Air Removal from the Sterilization Chamber BIOLOGICAL INDICATOR SPECIALISTS TM Volume 5, Number 2 March 2008 Garrett Krushefski Scientific & Technical Services Manager Air Removal from the Sterilization Chamber I recently received a question from

More information

SAMPLING SYSTEMS FOR LIQUIDS, GASES AND OTHERS APPLICATIONS.

SAMPLING SYSTEMS FOR LIQUIDS, GASES AND OTHERS APPLICATIONS. SAMPLING SYSTEMS FOR LIQUIDS, GASES AND OTHERS APPLICATIONS. SAMPLING SYSTEMS CONTENTS 1. SAMPLING SYSTEMS OVERVIEW...Pag. 3-4 2. LIQUID SAMPLING SYSTEM ESV SERIES ESV STANDARD CONFIGURATION...Pag. 5-6

More information

What I have learned about SF 6 gas testing.a Practical explanation

What I have learned about SF 6 gas testing.a Practical explanation What I have learned about SF 6 gas testing.a Practical explanation Over the past few years, testing of SF6 gas quality has become increasingly important and with the EPA in USA, setting guidelines for

More information

Japan s RIKEN BSI: Whole Facility Chlorine Dioxide Gas Decontamination Approach for a Barrier Facility A Case Study

Japan s RIKEN BSI: Whole Facility Chlorine Dioxide Gas Decontamination Approach for a Barrier Facility A Case Study Japan s RIKEN BSI: Whole Facility Chlorine Dioxide Gas Decontamination Approach for a Barrier Facility A Case Study Eiki Takahashi 1, Mark A. Czarneski 2, and Akihoki Sugiura 3 1RIKEN Brain Science Institute,

More information

Ozone. Sensoric O3 3E 1

Ozone. Sensoric O3 3E 1 Ozone Sensoric O3 3E 1 1 FEATURES Amperometric 3 electrode sensor cell Long life time High reliability High resolution Fast response time Fixed organic gel electrolyte TYPICAL APPLICATIONS Environmental

More information

Simplicity in VRU by using a Beam Gas Compressor

Simplicity in VRU by using a Beam Gas Compressor Simplicity in VRU by using a Beam Gas Compressor By Charlie D. McCoy and Mark Lancaster Abstract: Vapor Recovery Units are often expensive, complicated to operate and unable to deal with High H2S and liquids.

More information

Low Coking Temperatures & Recent Site Experience

Low Coking Temperatures & Recent Site Experience Delayed Coker Low Coking Temperatures & Recent Site Experience Presented by Mitch Moloney of ExxonMobil mitchell.j.moloney@exxonmobil.com @ Coking.com April-2010 MJ a Moloney - ExxonMobil April-2010 coking.com

More information

Best Practices for Installation of Gaseous Sample Pumps

Best Practices for Installation of Gaseous Sample Pumps Best Practices for Installation of Gaseous Sample Pumps Goal of the Gas Sample Pump Transport a clean, reliable, sample of the process gas to the analyzer, in the safest and easiest way possible, to ensure

More information

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR) Procedure No: SOP-025 Revision No: 1.0 (January 21, 2011) Page No.: 1 of 8 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to

More information

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING Blair Chalpin Charles A. Matthews Mechanical Design Engineer Product Support Manager Scanivalve Corp Scanivalve Corp Liberty Lake, WA

More information

Mitigation Starts With Design! Continues with Safe Operation, Mechanical Integrity and Training!

Mitigation Starts With Design! Continues with Safe Operation, Mechanical Integrity and Training! P.S.I.G. Mitigation Starts With Design! Continues with Safe Operation, Mechanical Integrity and Training! Last Unfortunately is When It Goes Wrong! The Failure! The Release! Loss of Product or Production

More information

Active Control of Vapor Pressurization (VaPak) Systems

Active Control of Vapor Pressurization (VaPak) Systems Active Control of Vapor Pressurization (VaPak) Systems Ralph Ewig, PhD Holder Consulting Group, Renton, WA 98059, USA Vapor pressurized (VaPak) propellant feed systems hold great promise for the development

More information

FAST EVAPORATION AT NORMAL PRESSURE

FAST EVAPORATION AT NORMAL PRESSURE FAST EVAPORATION AT NORMAL PRESSURE FAST EVAPORATION BY BLOWING-DOWN... The VAPORNADO Plus blow-down system is an innovative and efficient instrument for the evaporation of samples and fractions at normal

More information

O3 3E 1 F Gas Sensor Module

O3 3E 1 F Gas Sensor Module Product Information Pack O3 3E 1 F Gas Sensor Module (Ozone) CONTENTS Product Data Sheet Product Specification 2 Poisoning and Cross Sensitivities 3 Operating Instructions Introduction 4 Electrostatic

More information

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere! The portable Gas Chromatograph is now more versatile than ever. DPS Instruments is pleased to present the newest Portable Companion 2 Gas Chromatograph with room for 2 Detectors. The Companion 2 GC was

More information

ISOARC project. Samoylov water vapour isotopes analyser raw data

ISOARC project. Samoylov water vapour isotopes analyser raw data ISOARC project Samoylov water vapour isotopes analyser raw data Within the ISOARC project, the water vapour analyser on Samoylov is dedicated to the continuous measurements in ambient air of humidity level

More information

University of Minnesota Nano Center Standard Operating Procedure

University of Minnesota Nano Center Standard Operating Procedure University of Minnesota Nano Center Standard Operating Procedure Equipment Name: Controlled Atmosphere Glove Box Model: Labconco Protector Location: PAN 185 Badger Name: Not on Badger Revision Number:

More information

Process Control Loops

Process Control Loops In this section, you will learn about how control components and control algorithms are integrated to create a process control system. Because in some processes many variables must be controlled, and each

More information

Multiple Gas#5 GC configuration Jan 2016

Multiple Gas#5 GC configuration Jan 2016 History: Unfortunately there is no single column that can separate: Hydrogen Oxygen Nitrogen Methane CO CO2 Ethane Water Propane Butane Pentane Over the years SRI Instruments has devised several solutions

More information

NGP-250/500 Nitrogen Generator Quick Start Guide

NGP-250/500 Nitrogen Generator Quick Start Guide NGP-250/500 Nitrogen Generator Quick Start Guide Version: A July 2013 Potter Electric Signal Company, LLC 5757 Phantom Dr., Suite 125 P. O. Box 42037 Hazelwood, MO 63042 Phone: (314) 595-6900 Document

More information

A Perspective on Filter Cartridge Issues

A Perspective on Filter Cartridge Issues A Perspective on Filter Cartridge Issues Michigan Grape and Wine Conference February 24 th 26th HOW LONG WILL MY FILTERS LAST? UNCERTAIN Prefiltration Life Factors General rule of thumb: Prefiltration

More information

Reclaim Basic Set Up

Reclaim Basic Set Up This purpose of the document is to simplify the set up and understand the Gas Services reclaim system functions. The Gas Services Reclaim Manual is to be used for reference, maintenance, and servicing.

More information

SAMSON SOLUTIONS Energy and Power (Critical Applications)

SAMSON SOLUTIONS Energy and Power (Critical Applications) SAMSON SOLUTIONS Energy and Power (Critical Applications) Speakers: Ing. Gianluigi Rossi Ing. Stefano Salvadori Managing Director - SAMSON Italy Head of Sales and Project Department - SAMSON Italy MARKET

More information

EMISSION CHAMBERS. Bulk and Surface Emission Detection for the RAD7 User Manual

EMISSION CHAMBERS. Bulk and Surface Emission Detection for the RAD7 User Manual EMISSION CHAMBERS Bulk and Surface Emission Detection for the RAD7 User Manual TABLE OF CONTENTS INTRODUCTION 3 1 SYSTEM CONFIGURATION 4 1.1 Closed Loop Configuration 4 Fig. 1 Emission Chamber Closed Loop

More information

CLEANING AND SANITIZING

CLEANING AND SANITIZING CLEANING AND SANITIZING NATIONAL ICE CREAM MIX ASSOCIATION MEETING 1/17/2017 David Blomquist DFB Consulting LLC 651-324-6158 THE BIG THREE OF CLEANING Soil Cleaner Equipment THE BIG THREE OF CLEANING Soil

More information

Pool and Spa. Water Testing

Pool and Spa. Water Testing Pool and Spa Water Testing Pocket Digital Testers for Pools HI-701 Checker: Free Chlorine Hand-held Colorimeter The HI-701 is a simple, accurate and cost effective way to measure free chlorine. Free Chlorine:

More information

CLOSED LOOP SOLVENT EMISSIONS CONTROL. David C.H. Grant

CLOSED LOOP SOLVENT EMISSIONS CONTROL. David C.H. Grant CLOSED LOOP SOLVENT EMISSIONS CONTROL David C.H. Grant Abstract: An approach to closed loop solvent emissions control systems, useful for all conventional solvent processes and solvents is presented. The

More information

Model Description

Model Description LEAKHUNTER Plus LEAKHUNTER Plus Model 8066 Description The MATHESON LEAKHUNTER Plus is a truly universal leak detector. Extensive R&D has produced a multi-functional leak detector engineered to perform

More information

WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, PHONE #: (970) FAX #: (970) TO-Clean Manual

WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, PHONE #: (970) FAX #: (970) TO-Clean Manual WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, 80524 PHONE #: (970) 221-9179 FAX #: (970) 221-9364 TO-Clean Manual Copyright 2008 by Wasson ECE Instrumentation, Inc. (SJC 02/05/2008) All rights reserved

More information

Inert Air (N2) Systems Manual

Inert Air (N2) Systems Manual INSTRUCTION MANUAL Inert Air (N2) Systems Manual N2-MANUAL 2.10 READ AND UNDERSTAND THIS MANUAL PRIOR TO OPERATING OR SERVICING THIS PRODUCT. GENERAL INFORMATION Positive pressure nitrogen gas pressurizing

More information

Ozone PLUS Germicidal UV Advanced Oxidation Process (AOP) A Synergistic Sanitation Approach for Residential Pools

Ozone PLUS Germicidal UV Advanced Oxidation Process (AOP) A Synergistic Sanitation Approach for Residential Pools Ozone PLUS Germicidal UV Advanced Oxidation Process (AOP) A Synergistic Sanitation Approach for Residential Pools Types of Ozone & UV Systems Ozone gas can be produced by passing air (or oxygen) through

More information

Information Sheet. About this information sheet. Floor and wall loadings. Attachment of process module to floor. PlasmaPro Estrelas100

Information Sheet. About this information sheet. Floor and wall loadings. Attachment of process module to floor. PlasmaPro Estrelas100 Information Sheet PlasmaPro Estrelas100 About this information sheet This data sheet provides basic information about the installation of a PlasmaPro Estrelas100 tool. Some of the information in this sheet

More information

Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch

Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch Revision date September 2014 2 Emergency Plan for XeF 2 Si Etcher Standard Operating Procedures for Emergencies Contact information

More information

POWER FLUSH KIT INSTRUCTIONS

POWER FLUSH KIT INSTRUCTIONS 1 GENERAL INFORMATION POWER FLUSH KIT INSTRUCTIONS ACE Power Flush is an ozone-safe solvent engineered specifically for flushing refrigeration and air conditioning systems. Used exactly like the old R-11

More information

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N)

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N) 1 APPLICABILITY Exhibit 4 Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N) Definitions common to all certification and test procedures

More information

Work Method Statement. Prepared for: The Bolton NHS Foundation Trust Prepared by: Mark Fentiman Date:

Work Method Statement. Prepared for: The Bolton NHS Foundation Trust Prepared by: Mark Fentiman Date: Work Method Statement Prepared for: The Bolton NHS Foundation Trust Prepared by: Mark Fentiman Date: 06.03.2014 Detailed Work Plan The Bolton NHS Foundation Trust Date: 06.03.2014 1 Contents: 1. The Deprox

More information

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS Precise and Stable SF 6 Gas Analyzer Simultaneous measurement of humidity, SF 6 purity and SO 2 concentration Integrated gas recovery system with automatic pump back Fully automated SF 6 gas testing Fundamental

More information

Experimental Verification of Integrated Pressure Suppression Systems in Fusion Reactors at In-Vessel Loss-of -Coolant Events

Experimental Verification of Integrated Pressure Suppression Systems in Fusion Reactors at In-Vessel Loss-of -Coolant Events Experimental Verification of Integrated Pressure Suppression Systems in Fusion Reactors at In-Vessel Loss-of -Coolant Events K. Takase 1), H. Akimoto 1) 1) Japan Atomic Energy Research Institute (JAERI),

More information

1 5 2 Definition of Isolator Technology Isolator Types and Applications 3 Critical Design Considerations 4 Biodecontamination Conclusion and Summary 2

1 5 2 Definition of Isolator Technology Isolator Types and Applications 3 Critical Design Considerations 4 Biodecontamination Conclusion and Summary 2 Isolator Technology for Aseptic Manufacturing May 29, 2015 1 1 5 2 Definition of Isolator Technology Isolator Types and Applications 3 Critical Design Considerations 4 Biodecontamination Conclusion and

More information

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805)

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805) APPLICATION NOTE 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 7.10.18 Application Note 183 VO 2 and RER Measurement For simplified VO 2 measurement

More information

Safe Handling. of Chlor Alkali Products

Safe Handling. of Chlor Alkali Products Safe Handling of Chlor Alkali Products Safe Handling of Chlorine Table of Contents 1. Regulatory Information 2. Production of Chlorine 3. Product Information 4. Fire Hazards 5. Health Concerns 6. Personal

More information

Overview. Front Panel: Keypad and Display

Overview. Front Panel: Keypad and Display Overview The GA-200B is an analyzer that integrates a gas sampling system with sensors to measure and display the concentrations of oxygen and carbon dioxide in a sample as the percentage of a gas in the

More information

SERIES 10. Pilot Operated Tank Blanketing Valve, 2" PROTECTOSEAL.

SERIES 10. Pilot Operated Tank Blanketing Valve, 2 PROTECTOSEAL. SERIES 10 PROTECTOSEAL 2" NPT inlet and outlet standard Pilot operated design provides tight operating band Optional inlet and outlet connections available Inlet gas pressures from 40 PSIG to 130 PSIG

More information

Using the UltraRAE. Firmware 2.35

Using the UltraRAE. Firmware 2.35 Using the UltraRAE Firmware 2.35 Training Agenda UltraRAE features Setting up the UltraRAE Turning on the UltraRAE Idle Operation RAE-Sep Tubes Prepping for a measurement Taking a measurement Alarm modes

More information

Gas Cabinets and Valve Manifold Boxes: Applying CGA G13- Rev 2015 to Gas Delivery Equipment

Gas Cabinets and Valve Manifold Boxes: Applying CGA G13- Rev 2015 to Gas Delivery Equipment Gas Cabinets and Valve Manifold Boxes: Applying CGA G13- Rev 2015 to Gas Delivery Equipment The Standard Governing the Storage & Use of Silane: Compressed Gas Association s Publication CGA G-13 2015: Storage

More information

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Temperature controlled, multi-technique volumetric

More information

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the 1151 METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS 1.0 Scope and Application. 1.1 Analytes. Analyte Volatile Organic Compounds (VOC) CAS No. No CAS number assigned 1.2 Scope. This method

More information

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc. The HumiSys RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The new HumiSys with single or dual RH probes capabilities is

More information

SAFETY POLICY AND PROCEDURE MANUAL QUESTIONNAIRE

SAFETY POLICY AND PROCEDURE MANUAL QUESTIONNAIRE The following questionnaire is designed to aid us in the creation of your policy and procedure manual. Please place a checkmark in the boxes that correspond to the sections that you would like to be included

More information

HYPOXIC AIR FIRE PREVENTION SYSTEM TECHNICAL SPECIFICATION

HYPOXIC AIR FIRE PREVENTION SYSTEM TECHNICAL SPECIFICATION HYPOXIC AIR FIRE PREVENTION SYSTEM TECHNICAL SPECIFICATION Introduction: Hypoxic Air Fire Prevention System is a revolutionary fire prevention technology. It has the unique ability to create a breathable

More information

21700 Nordhoff St. Chatsworth, CA Phone

21700 Nordhoff St. Chatsworth, CA Phone 21700 Nordhoff St. Chatsworth, CA. 91311 Phone 818.882.2331 DILUTION SYSTEM DILUTION SYSTEM ~ 1 ~ 09 21 07 TABLE OF CONTENTS DEFINITIONS.. 3 INTRODUCTION.. 4 INSTALLATION & SETUP 5 CONVERSION TO PRESSURE

More information

Optimizing Gas Supply for Industrial Lasers

Optimizing Gas Supply for Industrial Lasers Optimizing Gas Supply for Industrial Lasers Laser cutting of metals and other materials has grown rapidly due to developments in laser power, advancements in CNC automation, and decreasing costs. The industrial

More information

CO 2 and O 2 scrubbers

CO 2 and O 2 scrubbers CO 2 and O 2 scrubbers The CO2 scrubber Fruit breathes : it uses oxygen (O2) and produces carbon dioxide (CO2). If the concentration of carbon dioxide (carbonic acid) gets too high, the quality of the

More information

IMA SpA. Advancements in Isolation Technology for Aseptic Filling. Marco Preus, July 2016

IMA SpA. Advancements in Isolation Technology for Aseptic Filling. Marco Preus, July 2016 IMA SpA Advancements in Isolation Technology for Aseptic Filling Marco Preus, 11-12 July 2016 Agenda Isolation technology advantages Isolation technology cost saving Air flow circuits Case study: vials

More information

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions. Oxford Instruments Plasma Technology www.oxford-instruments.com About this data sheet This data sheet provides basic information about the installation of a tool. Floor and wall loadings Table 1 lists

More information

Usage Policies Notebook for Trion RIE / ICP Dry Etch

Usage Policies Notebook for Trion RIE / ICP Dry Etch Usage Policies Notebook for Trion RIE / ICP Dry Etch Revision date September 2014 2 Emergency Plan for Trion RIE/ICP Dry Etch Standard Operating Procedures for Emergencies Contact information Person Lab

More information

6200 Automatic Isoperibol Calorimeter Calorimeter with 1108 Oxygen Bomb, Bucket and 1757 Printer

6200 Automatic Isoperibol Calorimeter Calorimeter with 1108 Oxygen Bomb, Bucket and 1757 Printer 6200 Automatic Isoperibol Calorimeter 6200 Calorimeter with 1108 Oxygen Bomb, Bucket and 1757 Printer Model 6200 is a microprocessor controlled, isoperibol oxygen bomb calorimeter which is widely used

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types

Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types 19 th World Conference on Non-Destructive Testing 2016 Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types Klaus HERRMANN 1, Daniel WETZIG

More information

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax OPERATION MANUAL Zutron Medical, LLC 17501 W 98 th St #40-27 Lenexa, KS 66219 Phone 877-343-5873 Fax 913-967-5944 ZUT-Lab-004-30004 REV. 03312017 Table of Contents 2 Introduction 1. Intended Use 2. Labels,

More information

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes Instruction Bulletin Addendum IB PB1000 Part no. Serial no. Order no. Oxymitter 4000 Applicability: Used with pressure-balancing

More information

Focus on VOC Emissions Reduction Using an Oxygen Based Inerting Control System For Inert Gas Blanketing of Chemical Process Vessels

Focus on VOC Emissions Reduction Using an Oxygen Based Inerting Control System For Inert Gas Blanketing of Chemical Process Vessels Inerting Control Systems by NTRON PROCESS ANALYZER DIVISION of NEUTRONICS INC. EXTON. PA. 19341 RJN Rev. A June 2001 Focus on VOC Emissions Reduction Using an Oxygen Based Inerting Control System For Inert

More information

Ballast Water Management Technologies A Brief Review

Ballast Water Management Technologies A Brief Review Ballast Water Management Technologies A Brief Review The following information is a brief overview of some of the potential ship board ballast water treatment methods which have been brought to the attention

More information

Reduce Turnaround Duration by Eliminating H 2 S from Flare Gas Utilizing VaporLock Scrubber Technology

Reduce Turnaround Duration by Eliminating H 2 S from Flare Gas Utilizing VaporLock Scrubber Technology Reduce Turnaround Duration by Eliminating H 2 S from Flare Gas Utilizing VaporLock Scrubber Technology Jim Woodard Vice President of Sales Bryant Woods Lead Project Engineer 4/3/18 Page 1 Objectives Impact

More information