Accuracy and Stability of Sea-Bird s Argo CTD and Work Towards a Better CTD for Deep Argo. David Murphy Sea-Bird Electronics

Size: px
Start display at page:

Download "Accuracy and Stability of Sea-Bird s Argo CTD and Work Towards a Better CTD for Deep Argo. David Murphy Sea-Bird Electronics"

Transcription

1 Accuracy and Stability of Sea-Bird s Argo CTD and Work Towards a Better CTD for Deep Argo David Murphy Sea-Bird Electronics

2 The Sea-Bird Team This work represents efforts of many people over many years: Norge Larson, Kristi Anson, Dick Guenther, Rick Beed, Genevieve Howell, Jonathon Boyle, Carol Janzen, Mike Vorkapich, Joel Reiter, Ken Lawson

3 The Points of this Presentation To inform the Argo program of Sea-Bird s Argo CTD accuracy When it leaves the factory Lifetime performance of temperature and pressure To discuss conductivity drift in the context of climatological data corrections To inform the Argo program of Sea-Bird s expectations of performance of the Deep Argo CTD

4 Motivation Greg Johnson told us that 25% of his floats required salinity correction and many were drifting salty Dean Roemmich told us that the Deep Argo program depended on a CTD with better stability

5 The Data Sets Over 4 Float data sets downloaded from US GODAE site, /dac/aoml Float data was sorted by PI, 4 PIs are included Selected only floats with delayed mode processing for salinity, used profiles that had only 1 designations Ended up with 24 float data sets Effort required: Coding an ftp application Netcdf format read using intrinsic matlab functions About 3 man weeks to understand the data format and render the data into a state appropriate for this analysis Sea-Bird s calibration history for products that share sensor electronics and characteristics with the Argo CTD

6 Calibration Terminology: Slope (Span) and Offset Error Y = mx + b b is the offset, an error here is constant across the whole measurement range m is the slope or span Error is smaller at the low end of the range Error is larger at the high end of the range Error = Sensor - Truth Measurement Range Slope or Span Error Offset Error No Error

7 Physical Standards are the Foundation of Temperature Accuracy Standards grade platinum thermometer (SPRT) is calibrated against physical standards Each data point represents 1 2 days work and more than 1 measurements 2 µc jump in Ga melt is caused by NIST calibration of standard resistor Temperature (Deg.C.) Temperature (Deg.C.) SPRT Data at Triple Point of Water mdegree Full Scale SPRT Data at the Gallium Melt Point 1 mdegree Full Scale

8 Estimating Temperature Stability SBE 38 Proxy: Calibration history of 411 SBE 38s Average error over degree range Shares thermistor and circuit with Argo CTD Argo CTD: 96 calibrations of Argo CTD This one will never see the ocean Residual Error (C) Residual, (Degrees C) SBE 38 Drift (C) Years After Manufacture SBE41cp Temperature, Degrees C

9 What are We Doing For Deep Argo? (Temperature) Currently meeting requirement.3 Acquisition technique improves temperature noise performance Calibration apparatus Minor Improvement Temperature Noise (Crms) Single Measurement Multiple Measurement ~ 2 µc Improvement Measurement Time ms

10 Estimating Pressure Stability Past hurdles Electro-static discharge sensitivity Failed glass metal seals Plots below are for an initial 424 floats Most have very small offset after 72 profiles Mode at 4 decibars possibly induced by seal failure Pressure Offset at Profile 1 Pressure Offset at Profile Number of Floats Number of Floats Pressure Offset Decibars Pressure Offset Decibars

11 Pressure Drift Very difficult to document for Argo CTD No good comparable use case in other SBE products.5 SBE41cp SBE SBE41cp Residual, (% F S R) ~1 dbar Drift 5 years Residual, (% F S R) ~1 dbar Drift 4 years Residual, (% F S R) <1 dbar Drift 2 years Pressure (PSIA) Pressure (PSIA) Pressure (PSIA).5.25 SBE ~2 dbar Drift 5 years.5.25 SBE SBE Residual, (% F S R) Pressure (PSIA) Residual, (% F S R) ~5 dbar Drift 4 years Pressure (PSIA) Residual, (% F S R) <1dBar Drift 2 years Pressure (PSIA)

12 CTD Pressure Sensor Screening Process Starting in 29 Accelerated testing for seal failure method developed in collaboration with Druck 5 pressure cycles 3 psia 15 hours at 4 psia and 4 C Twice a week pressure vented Offset measured against barometric pressure Returned to 4 psia Hand selected after infant drift Drift must be within +/-.25 decibar +.25 dbar -.25 dbar HOURS - 15

13 Reducing Pressure Span Error Over Operating Temperature Range Dana Swift reports errors of as much as 3 dbars at 2 dbars Provoked investigation with CTD 2795 Currently sensor span temperature error is corrected with data at -1 and 35 degrees C with data from Druck Residual, (% F S R) dbar dbar SBE Pressure (PSIA) 3 psia 26-Sep Sep Sep Sep Sep-11.1 No Error at Room Temp ~1 dbar Error at 1 Degree

14 Pressure Sensor Span Sensitivity Not Try multi-temperature Point Pressure Span Correction Improves error for mid range temperatures Linear Residual, (% F S R) dbar SBE Pressure (PSIA) 26-Sep Sep Sep Sep Sep ~1 dbar Error at 1 Degree

15 Normalizing to Calibration Room Reform calibration equation to T Tr where Tr is room temperature at time of sensor calibration Temperature R e s id u a l, (% F S R ) dbar SBE Sep Sep Sep Sep Sep Pressure (PSIA)

16 What are We Doing for Deep Argo? (Pressure) Extend screening to Deep Argo pressure Custom 7 decibar sensor from Kistler Many currently in field Have received second generation for evaluation with improved internal construction Implement improved calibration of pressure sensor span for temperature sensitivity

17 Estimating Salinity / Conductivity Stability Number of Floats Climatology measured salinity offsets from 24 floats deployed between Histograms for 18 floats at profile 72 of salinity error at next to bottom bin 87.5% of floats within +/-.2 PSU at profile Salinity Offset Float Profile Number of Floats Salinity Offset from Climatology Salinity Offset PSU Salinity Offset PSU

18 Salinity / Conductivity Performance in the Laboratory 97 calibrations Error less than +/-.5 PSU Residual, (S /m ) PSU. -.5 PSU -.5 SBE41cp 746 equivalent -.1 Conductivity Range 7 S/m Conductivity (Siemens/m)

19 Investigating Conductivity (Salinity) Stability Salinity performance measured against climatology PMEL reports 25% of their floats require salinity correction UW reports 5% require correction Salinity drifting fresh or negative typical of cell fouling Salinity drifting salty or positive not well understood Error at 2. degrees.323 PSU.263 S/m

20 Salinity Error Sources Error evaluated at potential temperature of 2. degrees, P and C are dependent variables Possible sources of salinity error Pressure error Temperature error Conductivity error Will examine each in turn

21 Pressure as Source of Salinity Error In our example: At 2. degrees and 2 dbars, climatologic salinity is PSU The example CTD has an error of.3 PSU If temperature is correct and conductivity is correct, Pressure would have to have an error of 6 decibars This error would be evident in surface pressure measurement

22 Temperature as Source of Salinity Error Same example: At 2. degrees and 2 dbars, climatologic salinity is PSU The example CTD has an error of.3 PSU If pressure and conductivity are correct, Temperature would have to be 3 milli-degrees off 3 mc drift in the SP6 thermistor is unlikely, SBE 38 has same thermistor and circuit topology Error Degrees C SBE 38 Drift (C) / Year Years of Service

23 Conductivity as Source of Salinity Error Error sources Conductivity Cell Fouling Negative Drift Reduces cell radius Electrode degradation Cell damage Can show positive drift Typically magnitude of error is high Conductivity Sensor Circuit Frequency of oscillation dependent on circuit components (resistors and capacitors) Indeterminate direction of drift

24 Plausible Mechanism Presume that fouling dominates negative drift and will not discuss further Identifying conductivity circuitry as a likely source for the positive salinity drift does not explain why some investigators are more impacted. Strategies: Investigate difference in climatology and CTD for clues for plausible mechanism for positive drift Use next to the bottom bin for each profile of each float If bin depth greater than 9 decibars Next to bottom bin assures flushed and equilibrated cell

25 Salinity Correction Criteria All 4 PIs are making corrections based on the same criteria and in the following order Salinity is corrected for pressure error based on the surface pressure of the preceding profile Cell thermal mass correction is applied to the entire profile.1 PSU magnitude is the correction threshold Once this is crossed the float salinity is corrected back to where the drift started Comments: Some subjectivity in correction criteria Quite a few cooks in the kitchen

26 ARGO Salinity Error Profile 1 Compared to Profile 72 Histograms are scaled to show distribution of corrections, number of floats corrected +/-.2 ranges to UW Profile 1 for 747 floats UW Profile 72 for 687 floats Scripps Profile 1 for 861 floats Scripps Profile 72 for 531 floats

27 ARGO Salinity Error Profile 1 Compared to Profile 72 Histograms are scaled to show distribution of corrections, number of floats corrected +/-.2 ranges to WHOI Profile 1 for 397 floats PMEL Profile 1 for 418 floats WHOI Profile 72 for 253 floats PMEL Profile 72 for 376 floats

28 Positive Conductivity Drift Correlates with Low Latitude Deployments More likely, correlation is with mixed layer temperature 3% 25% 2% 15% 1% 5% % Percent of Positive Drifters by Latitude Floats, 1 Positive Drifter Number of Floats Number of Floats All Floats by Latitude Latitude Positive Drifters by Latitude Latitude

29 Possible Mechanism More water vapor inside floats at warmer latitudes Components on CTD circuit board effected by higher humidity Desiccants hold less water at warmer temperatures Sud-Chemie, Inc 211

30 What are We Doing for Deep Argo? (Conductivity) Verify mechanism through experiments with desiccant Hermitically enclose conductivity circuit board Test in-situ calibration of circuit excluding cell

31 Work in Progress Work in progress Humidity experiments to verify mechanism of positive conductivity drift Acquisition of environmental chamber to calibrate pressure sensor span sensitivity to temperature Qualification of Kistler 7 dbar pressure sensors Work planned Redesign of calibration bath to improve thermal noise Comparison of SBE Microcat post deployment calibration to climatological correction

32 Fini

Sensor Stability. Summarized by Breck Owens David Murphy, Sea-Bird Electronics AST-14 Wellington March 18-21

Sensor Stability. Summarized by Breck Owens David Murphy, Sea-Bird Electronics AST-14 Wellington March 18-21 Sensor Stability Summarized by Breck Owens David Murphy, Sea-Bird Electronics AST-14 Wellington March 18-21 Physical Standards are the Foundation of Temperature Accuracy Standards grade platinum thermometer

More information

SBE61 CTD calibration: results from the June 2014 Tangaroa Voyage

SBE61 CTD calibration: results from the June 2014 Tangaroa Voyage SBE61 CTD calibration: results from the June 2014 Tangaroa Voyage Phil Sutton, Matt Walkington NIWA Dave Murphy, Sea Bird Electronics Dean Roemmich, Nathalie Zilberman Scripps Institution of Oceanography

More information

Stability of Temperature and Conductivity Sensors of Argo Profiling Floats

Stability of Temperature and Conductivity Sensors of Argo Profiling Floats Journal of Oceanography, Vol. 6, pp. 5 to 58, Short Contribution Stability of Temperature and Conductivity Sensors of Argo Profiling Floats EITAROU OKA * and KENTARO ANDO Frontier Observational Research

More information

Development of the Deep Argo Program

Development of the Deep Argo Program Development of the Deep Argo Program Dean Roemmich, Nathalie Zilberman, Philip Sutton, Russ Davis, John Gilson, Kyle Grindley, Carol Janzen, Gregory Johnson, Taiyo Kobayashi, Nordeen Larson, Serge Le Reste,

More information

Argo Float Pressure Offset Adjustment Recommendations. 1: IORGC/JAMSTEC, 2-15, Natsushima-cho, Yokosuka, , Japan

Argo Float Pressure Offset Adjustment Recommendations. 1: IORGC/JAMSTEC, 2-15, Natsushima-cho, Yokosuka, , Japan Argo Float Pressure Offset Adjustment Recommendations Taiyo Kobayashi 1, 2 and Gregory C. Johnson 3 1: IORGC/JAMSTEC, 2-15, Natsushima-cho, Yokosuka, 237-0061, Japan taiyok@jamstec.go.jp 2: Now visiting

More information

Sea-Bird Scientific Carol Janzen, Ph.D. Oceanography David Murphy, M.S.E.E. Dan Quittman, M.S.E.E.

Sea-Bird Scientific Carol Janzen, Ph.D. Oceanography David Murphy, M.S.E.E. Dan Quittman, M.S.E.E. Sea-Bird Scientific Carol Janzen, Ph.D. Oceanography David Murphy, M.S.E.E. Dan Quittman, M.S.E.E. Sea-Bird Scientific Sensor Offering Fundamental feature of each kind of sensing technology Pumping sensors

More information

Procedures for correcting in situ CTD data and results obtained during the NEAR-GOOS Cross-Basin Climate Monitoring Section project

Procedures for correcting in situ CTD data and results obtained during the NEAR-GOOS Cross-Basin Climate Monitoring Section project Procedures for correcting in situ CTD data and results obtained during the NEAR-GOOS Cross-Basin Climate Monitoring Section project Dmitry Kaplunenko 1, Aleksander Lazaryuk 1, 2, Vyacheslav Lobanov 1,

More information

Deep SOLO. Nathalie Zilberman, Dean Roemmich, and SIO float lab. 1. Deep SOLO float characteristics. 2. Deep SOLO float Deployment

Deep SOLO. Nathalie Zilberman, Dean Roemmich, and SIO float lab. 1. Deep SOLO float characteristics. 2. Deep SOLO float Deployment Deep SOLO Nathalie Zilberman, Dean Roemmich, and SIO float lab 1. Deep SOLO float characteristics 2. Deep SOLO float Deployment 3. Deep SOLO temperature and salinity Deep SOLO Float deployment, R/V Tangaroa

More information

Module 9. Advanced Data Processing

Module 9. Advanced Data Processing Module 9 Advanced Data Processing 2 Module 9: Advanced Data processing Overview Advanced Data Processing or Why Doesn t My Data Look Like the Examples in Class? Sensor alignment, matching measurements

More information

QC for Hydrographic Data GEOSECS to GO-SHIP. Susan M. Becker Scripps Institution of Oceanography Oceanographic Data Facility

QC for Hydrographic Data GEOSECS to GO-SHIP. Susan M. Becker Scripps Institution of Oceanography Oceanographic Data Facility QC for Hydrographic Data GEOSECS to GO-SHIP Susan M. Becker Scripps Institution of Oceanography Oceanographic Data Facility Global Hydrographic Efforts Geo-chemical Ocean Sections (GEOSECS) World Ocean

More information

Oxygen measurements: sensors accuracy and scientific needs

Oxygen measurements: sensors accuracy and scientific needs Oxygen measurements: sensors accuracy and scientific needs L. Coppola (CNRS-INSU), E.Diamond (CNRS-INSU), L.Delauney (IFREMER), D.Hydes (NOC), M.Haller (HZG), J.Karstensen (IFM-GEOMAR), D.Lefevre (CNRS-INSU),

More information

Overview. 2 Module 13: Advanced Data Processing

Overview. 2 Module 13: Advanced Data Processing 2 Module 13: Advanced Data Processing Overview This section of the course covers advanced data processing when profiling. We will discuss the removal of the fairly gross effects of ship heave and talk

More information

Argo-O2 data Data Management Real-time and Delayed-mode QC Where we are/ Where we go

Argo-O2 data Data Management Real-time and Delayed-mode QC Where we are/ Where we go Argo-O2 data Data Management Real-time and Delayed-mode QC Where we are/ Where we go LPO: V. Thierry, T. Bouinot ALTRAN: J. P. Rannou GEOMAR: A. Kortzinger, H. Bittig MBARI: K. Johnson UW: S. Emerson,

More information

MIL-STD-883G METHOD

MIL-STD-883G METHOD STEADY-STATE LIFE 1. PURPOSE. The steady-state life test is performed for the purpose of demonstrating the quality or reliability of devices subjected to the specified conditions over an extended time

More information

Troubleshooting Basics: Profiling CTDs

Troubleshooting Basics: Profiling CTDs Technical Note Troubleshooting Basics: Profiling CTDs Methods for Diagnosing Erratic and Unreasonable Data Greg Ikeda, October 2018 Introduction Any CTD or sensor may eventually exhibit data spikes and

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

WBEA Standard Operating Procedure

WBEA Standard Operating Procedure Page 1 WBEA Standard Operating Procedure SOP Title Temperature and Relative Humidity Author Gary Cross Implementation date March 2, 2013 Revision History Revision # Date Description Author Page 2 Table

More information

Payload CTD. Carol D. Janzen 1 Research and Development 1 Sea-Bird Electronics, Inc. Elizabeth L. Creed 2 MaritimeSystems 2 irobot Corporation

Payload CTD. Carol D. Janzen 1 Research and Development 1 Sea-Bird Electronics, Inc. Elizabeth L. Creed 2 MaritimeSystems 2 irobot Corporation Physical Oceanographic Data from Seaglider Trials in Stratified Coastal Waters Using a New Pumped Payload CTD Carol D. Janzen 1 Research and Development 1 Sea-Bird Electronics, Inc. Bellevue, WA, USA cjanzen@seabird.com

More information

WHOTS Mooring Subsurface Instrumentation

WHOTS Mooring Subsurface Instrumentation UH Contributions to WHOTS-13 Cruise Report by Fernando Santiago-Mandujano, Daniel McCoy, Jefrey Snyder, R. Walter Deppe, Kellen Rosburg, Glenn Carter, Katrina Berry, and Roger Lukas WHOTS Mooring Subsurface

More information

HOBO MX Water Level Logger (MX2001-0x) Manual

HOBO MX Water Level Logger (MX2001-0x) Manual HOBO MX Water Level Logger (MX2001-0x) Manual HOBO MX Water Level Logger Models: MX2001-01 or MX2001-01-Ti (Titanium), 9-meter/30-foot depth) MX2001-02, 30-meter/100-foot depth MX2001-03, 76-meter/250-foot

More information

The 2008 North Atlantic Bloom Experiment. Calibration Report #3

The 2008 North Atlantic Bloom Experiment. Calibration Report #3 The 2008 North Atlantic Bloom Experiment Calibration Report #3 Calibration of the Dissolved Oxygen Sensors on Float 48 and on The Knorr CTD with Winkler bottle samples Eric D Asaro Applied Physics Laboratory,

More information

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson A Journal of Practical and Useful Vacuum Technology From By Phil Danielson Thermal Conductivity Gauges Thermal conductivity pressure gauges are extremely common in vacuum technology, but an understanding

More information

Argo National Data Management Report Italy (2016)

Argo National Data Management Report Italy (2016) Argo National Data Management Report Italy (2016) 1. Status Data acquired from floats: 335 floats were deployed in the Mediterranean and in Black Seas between 2001 and 2016 (the floats temporal distribution

More information

RSKtools for Matlab processing RBR data

RSKtools for Matlab processing RBR data Table of Contents Introduction... 1 RSKtools help... 1 Getting set up... 1 Remove atmospheric pressure from measured total pressure... 2 Correct for A2D zero-order hold... 2 Low-pass filtering... 3 Alignment

More information

Methods Used in Measuring Surface Seawater pco 2 Aboard RVIB Nathaniel B. Palmer

Methods Used in Measuring Surface Seawater pco 2 Aboard RVIB Nathaniel B. Palmer F:\NBPalmer\Palmer_methods.doc Page 1 09/16/02 11:55 AM 1. Introduction: Methods Used in Measuring Surface Seawater pco 2 Aboard RVIB Nathaniel B. Palmer Taro Takahashi, Colm Sweeney, Stewart C. Sutherland

More information

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated. Asset #: 123 Calibration - Certificate No: 2083.01 Customer: Condition: Action Taken: Sample Inc. 123 Sample Rd. Sample, MA 01234 The instrument was operational upon receipt. The instrument was adjusted

More information

BUYER S GUIDE AQUAlogger 520

BUYER S GUIDE AQUAlogger 520 OCEAN & ENVIRONMENTAL BUYER S GUIDE AQUAlogger 520 Mini Temperature and Pressure Logger AQUAlogger 520 Aquatec s primary temperature and depth model, the AQUAlogger 520, combines years of product development

More information

Best Practice for Calibrating LTH Conductivity Instruments

Best Practice for Calibrating LTH Conductivity Instruments Application Note Best Practice for Calibrating LTH Conductivity Instruments As accurate process measurement becomes an everyday requirement it is vital to be able to calibrate conductivity instruments

More information

Engineering Note. Algorithms. Overview. Detailed Algorithm Description. NeoFox Calibration and Measurement. Products Affected: NeoFox

Engineering Note. Algorithms. Overview. Detailed Algorithm Description. NeoFox Calibration and Measurement. Products Affected: NeoFox Engineering Note Topic: NeoFox Calibration and Measurement Products Affected: NeoFox Date Issued: 04/18/2011 Algorithms Overview NeoFox is a dynamic measurement system that has been designed to work with

More information

TR Test Report Complementary Testing RKN e1 Container (+5 C) Protection by Insulation Version: 01 Page: 2 (22)

TR Test Report Complementary Testing RKN e1 Container (+5 C) Protection by Insulation Version: 01 Page: 2 (22) RKN e1 Container (+5 C) Protection by Insulation Version: 01 Page: 2 (22) Document History Version Date of issue Revision(s) Issued by 01 16OCT2012 N/A (First version of document) David Graaf, Envirotainer

More information

PERSPECTIVES FOR ACCURACY AND QUALITY ASSURANCE OF CTD & BIOGEOCHEMICAL DATA STREAMS FROM AUTONOMOUS PLATFORMS

PERSPECTIVES FOR ACCURACY AND QUALITY ASSURANCE OF CTD & BIOGEOCHEMICAL DATA STREAMS FROM AUTONOMOUS PLATFORMS PERSPECTIVES FOR ACCURACY AND QUALITY ASSURANCE OF CTD & BIOGEOCHEMICAL DATA STREAMS FROM AUTONOMOUS PLATFORMS Ian D. Walsh (1), Jochen Klinke (2), Kim Martini (3), David Murphy (4) (1) Sea-Bird Scientific,

More information

2017 Float and Sensor Workshop. Conveners: Brian King, Steve Riser, Susan Wijffels Local Host: Steve Riser, U. Washington

2017 Float and Sensor Workshop. Conveners: Brian King, Steve Riser, Susan Wijffels Local Host: Steve Riser, U. Washington 2017 Float and Sensor Workshop Conveners: Brian King, Steve Riser, Susan Wijffels Local Host: Steve Riser, U. Washington Objectives The goals of the workshop were to Increase the overall efficiency of

More information

Institute For Thermal Processing Specialists

Institute For Thermal Processing Specialists Institute For Thermal Processing Specialists TEMPERATURE DISTRIBUTION PROTOCOL FOR PROCESSING IN STILL, WATER IMMERSION RETORTS, INCLUDING AGITATING SYSTEMS OPERATED IN A STILL MODE Various methods and

More information

Argo Quality Control Manual For dissolved oxygen concentration

Argo Quality Control Manual For dissolved oxygen concentration Argo data management http://dx.doi.org/10.13155/46542 1 Argo Quality Control Manual For dissolved oxygen concentration Version 1.1 28 th September 2016 Argo data management quality control manual version

More information

NOTES ON CTD/O 2 DATA ACQUISITION AND PROCESSING USING SEA-BIRD HARDWARE AND SOFTWARE (AS AVAILABLE)

NOTES ON CTD/O 2 DATA ACQUISITION AND PROCESSING USING SEA-BIRD HARDWARE AND SOFTWARE (AS AVAILABLE) NOTES ON CTD/O 2 DATA ACQUISITION AND PROCESSING USING SEA-BIRD HARDWARE AND SOFTWARE (AS AVAILABLE) K.E. McTaggart 1, G.C. Johnson 1, M.C. Johnson 2, F.M. Delahoyde 2, and J.H. Swift 2 1 NOAA/Pacific

More information

Dissolved Oxygen Guide

Dissolved Oxygen Guide Educat i onser i es Di ssol vedoxygengui de Dissolved Oxygen Guide Introduction Dissolved oxygen probes provide a convenient approach to essentially direct measurement of molecular oxygen. The membrane

More information

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text. Hach LDO Sensor Instruction Sheet 00745589 Safety Precautions Please read this entire instruction sheet before operating this sensor. Pay particular attention to all danger and caution statements. Failure

More information

Quality control of data streams at ANFOG

Quality control of data streams at ANFOG Quality control of data streams at ANFOG Christine Hanson & Claire Gourcuff Australian National Facility for Ocean Gliders (ANFOG), School of Environmental Systems Engineering (SESE) and UWA Oceans Institute

More information

OPERATING AND MAINTENANCE MANUAL SBE 23 DISSOLVED OXYGEN SENSOR

OPERATING AND MAINTENANCE MANUAL SBE 23 DISSOLVED OXYGEN SENSOR OPERATING AND MAINTENANCE MANUAL SBE 23 DISSOLVED OXYGEN SENSOR Sea-Bird Electronics, Inc. 1808-136th Place NE Bellevue, Washington 98005 USA Phone: (206) 643 9866 Telex: 292915 SBEI vers. 000 L I M I

More information

USING THE CS400/CS405 (KPSI SERIES 169/173) SUBMERSIBLE PRESSURE TRANSDUCER WITH CAMPBELL SCIENTIFIC DATALOGGERS INSTRUCTION MANUAL

USING THE CS400/CS405 (KPSI SERIES 169/173) SUBMERSIBLE PRESSURE TRANSDUCER WITH CAMPBELL SCIENTIFIC DATALOGGERS INSTRUCTION MANUAL USING THE CS400/CS405 (KPSI SERIES 169/173) SUBMERSIBLE PRESSURE TRANSDUCER WITH CAMPBELL SCIENTIFIC DATALOGGERS INSTRUCTION MANUAL REVISION: 11/02 COPYRIGHT (c) 2000-2002 CAMPBELL SCIENTIFIC, INC. This

More information

CORESTA RECOMMENDED METHOD N 6

CORESTA RECOMMENDED METHOD N 6 CORESTA RECOMMENDED METHOD N 6 DETERMINATION OF VENTILATION DEFINITIONS AND MEASUREMENT PRINCIPLES (2015 Revision September 2016) 1. SCOPE This CORESTA Recommended Method specifies a method for the determination

More information

Systematic Errors in the Hydrographic Data and Their Effect on Global Heat Content Calculations

Systematic Errors in the Hydrographic Data and Their Effect on Global Heat Content Calculations Systematic Errors in the Hydrographic Data and Their Effect on Global Heat Content Calculations Titelmasterformat durch Klicken bearbeiten Viktor Gouretski KlimaCampus, University of Hamburg Frascati,

More information

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The

More information

Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004

Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004 Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004 INTRODUCTION Sontek/YSI has introduced new firmware and software for their RiverSurveyor product line. Firmware changes

More information

TRIAXYS Acoustic Doppler Current Profiler Comparison Study

TRIAXYS Acoustic Doppler Current Profiler Comparison Study TRIAXYS Acoustic Doppler Current Profiler Comparison Study By Randolph Kashino, Axys Technologies Inc. Tony Ethier, Axys Technologies Inc. Reo Phillips, Axys Technologies Inc. February 2 Figure 1. Nortek

More information

CORIOLIS, A FRENCH PROJECT FOR IN SITU OPERATIONAL OCEANOGRAPHY. S. Pouliquen, A. Billant, Y. Desaubies, G. Loaec, F. Gaillard, G.

CORIOLIS, A FRENCH PROJECT FOR IN SITU OPERATIONAL OCEANOGRAPHY. S. Pouliquen, A. Billant, Y. Desaubies, G. Loaec, F. Gaillard, G. CORIOLIS, A FRENCH PROJECT FOR IN SITU OPERATIONAL OCEANOGRAPHY S. Pouliquen, A. Billant, Y. Desaubies, G. Loaec, F. Gaillard, G. Maudire IFREMER, BP70, 29280 Plouzané, France Sylvie.Pouliquen@ifremer.fr

More information

GLIDERS Aanderaa oxygen optode (models 3830 & 3835)

GLIDERS Aanderaa oxygen optode (models 3830 & 3835) Kiel Oxygen Projects GLIDERS Aanderaa oxygen optode (models 3830 & 3835) FLOATS Webb Slocum Glider NEMO, Optimare, Germany MOORINGS Provcarbon, Martec, France Provor CTS3 DO, Martec, France APEX Float,

More information

UNIK 5600/5700 Marine Certified Pressure Sensing Platform

UNIK 5600/5700 Marine Certified Pressure Sensing Platform GE Measurement & Control UNIK 5600/5700 Marine Certified Pressure Sensing Platform The new UNIK 5600/5700 carries marine certification for most zones on-board ship, as well as Intrinsically Safe certifications.

More information

IRceL Operating Instructions

IRceL Operating Instructions IRceL Operating Instructions IRceL CO2 IRceL CH4 Carbon dioxide sensor Methane sensor Contents Introduction... 2 Evaluation... 2 Performance... 2 Instrument Development... 3 1. Lamp drive circuit... 3

More information

Equation 1: F spring = kx. Where F is the force of the spring, k is the spring constant and x is the displacement of the spring. Equation 2: F = mg

Equation 1: F spring = kx. Where F is the force of the spring, k is the spring constant and x is the displacement of the spring. Equation 2: F = mg 1 Introduction Relationship between Spring Constant and Length of Bungee Cord In this experiment, we aimed to model the behavior of the bungee cord that will be used in the Bungee Challenge. Specifically,

More information

Low Profile Load Cells

Low Profile Load Cells Low Profile Load Cells World Leading Performance With Improved Linearity, Hysteresis, Creep & Thermal Performance Introducing Interface s Enhanced Low Profile Load Cell Interface Inc. pioneered the first

More information

Applying Hooke s Law to Multiple Bungee Cords. Introduction

Applying Hooke s Law to Multiple Bungee Cords. Introduction Applying Hooke s Law to Multiple Bungee Cords Introduction Hooke s Law declares that the force exerted on a spring is proportional to the amount of stretch or compression on the spring, is always directed

More information

General Accreditation Guidance. User checks and maintenance of laboratory balances

General Accreditation Guidance. User checks and maintenance of laboratory balances General Accreditation Guidance User checks and maintenance of laboratory balances January 2018 Copyright National Association of Testing Authorities, Australia 2010 All intellectual property rights in

More information

SDX Submersible Depth Transmitter User Manual

SDX Submersible Depth Transmitter User Manual SDX Submersible Depth Transmitter User Manual July 2017 USER INFORMATION Stevens makes no warranty as to the information furnished in these instructions and the reader assumes all risk in the use thereof.

More information

Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017

Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017 PARTNERSHIP FOR THE DELAWARE ESTUARY Science Group Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017 Prepared By: Kurt Cheng Suggested Citation: Cheng,

More information

Preliminary Cruise Report

Preliminary Cruise Report R/V Melville 4 August 213-23 August 213 Seattle, Washington Seattle, Washington Chief Scientist: Dr. Dennis A. Hansell Rosenthal School of Marine & Atmospheric Science Preliminary Cruise Report Data processed

More information

Auto-Zero Calibration Technique for Pressure Sensors

Auto-Zero Calibration Technique for Pressure Sensors Auto-Zero Calibration Technique for Pressure Sensors A Technical Note 1.0 INTRODUCTION This technical note describes how to implement Auto-Zero, a calibration technique for pressure sensors based on sampling

More information

AMT-Ex Dewpoint Transmitter

AMT-Ex Dewpoint Transmitter AMT-Ex Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk

More information

Observational approaches to oxygen depletion

Observational approaches to oxygen depletion Observational approaches to oxygen depletion Denis Gilbert Institut Maurice-Lamontagne Pêches et Océans Canada Mont-Joli, Québec Workshop on Ocean Biology Observatories Venezia, Italia, September 17, 2009

More information

CHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller

CHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller CHEMICAL ENGINEERING LABORATORY CHEG 239W Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller Objective The experiment involves tuning a commercial process controller for temperature

More information

Argo quality control manual

Argo quality control manual Argo data management January 3 rd 2013 Argo quality control manual Version 2.8 1 Table of contents 1. INTRODUCTION 6 2. REAL-TIME QUALITY CONTROLS 7 2.1. ARGO REAL-TIME QUALITY CONTROL TEST PROCEDURES

More information

SDX Submersible Depth Transmitter User Manual

SDX Submersible Depth Transmitter User Manual SDX Submersible Depth Transmitter User Manual October 2007 USER INFORMATION Stevens makes no warranty as to the information furnished in these instructions and the reader assumes all risk in the use thereof.

More information

Water Samplers, Water Sampling with Internal Recording, Cabling, and Deployment

Water Samplers, Water Sampling with Internal Recording, Cabling, and Deployment Water Samplers, Water Sampling with Internal Recording, Cabling, and Deployment Water Samplers Care and feeding SeatermAF User interface for internally recording instruments firing water samplers with

More information

GEA FOR ADVANCED STRUCTURAL DYNAMIC ANALYSIS

GEA FOR ADVANCED STRUCTURAL DYNAMIC ANALYSIS SMART SOLUTIONS FOR VIBRATION MONITORING GEA FOR ADVANCED STRUCTURAL DYNAMIC ANALYSIS ANALYSIS OF CIVIL STRUCTURES - EXPO MERLATA PEDESTRIAN BRIDGE ABSTRACT Civil structures and in particular bridges and

More information

RDO Optical Dissolved Oxygen Sensor

RDO Optical Dissolved Oxygen Sensor Manual 51A-RDO Sensor April 2010 Model RDO Sensor RDO Optical Dissolved Oxygen Sensor RDO is a registered trademark of In-Situ Inc. All rights reserved. RDO PROBE SPECIFICATIONS Wetted Materials: Delrin

More information

Next Generation Quartz Pressure Gauges

Next Generation Quartz Pressure Gauges Next Generation Quartz Pressure Gauges Rick Puccio, Ph.D. Senior Scientist Quartzdyne Houston, TX March 2, 2016 Outline Quartz Pressure Gauge Introduction Gauge Size Reduction Long Term Pressure Drift

More information

Improve Process Reliability

Improve Process Reliability Convenient Conductivity Calibration to Meet Pharmacopeia Regulations Pharmacopeias require that conductivity measurement systems be periodically calibrated to ensure their precision. The range of accessories

More information

Environmental Protection Division

Environmental Protection Division Operational Policy Manual Section Subsection Environmental Protection Division 6.0 6.01.02 Name of Policy: Replaces: Standard Auditing Procedure for Continuous Emission Monitors and Ambient Air Monitoring

More information

Chapter 7. SCIIB Pressure Sensor Performance Evaluations: Experiments, Results and Discussions

Chapter 7. SCIIB Pressure Sensor Performance Evaluations: Experiments, Results and Discussions : Experiments, Results and Discussions This chapter will summarize the experiments and results associated with the development of the single-mode fiber-based SCIIB pressure sensor system. We will start

More information

QUALIFICATION OF SUB-ATMOSPHERIC PRESSURE SENSORS FOR THE CRYOMAGNET BAYONET HEAT EXCHANGERS OF THE LARGE HADRON COLLIDER

QUALIFICATION OF SUB-ATMOSPHERIC PRESSURE SENSORS FOR THE CRYOMAGNET BAYONET HEAT EXCHANGERS OF THE LARGE HADRON COLLIDER EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH European Laboratory for Particle Physics Large Hadron Collider Project LHC Project Report 980 QUALIFICATION OF SUB-ATMOSPHERIC PRESSURE SENSORS FOR THE CRYOMAGNET

More information

Glider and shipboard observations for underwater optical detection and communications

Glider and shipboard observations for underwater optical detection and communications Glider and shipboard observations for underwater optical detection and communications By: LCDR Alec Cullen LCDR Brian Breshears LT Ross Hammerer Thesis Advisor: Peter C. Chu Co-Advisor: Tetyana Margolina

More information

7.4 Temperature, Salinity and Currents in Jamaica Bay

7.4 Temperature, Salinity and Currents in Jamaica Bay 7.4 Temperature, Salinity and Currents in Jamaica Bay Arnold Gordon, Bruce Huber and Robert Houghton 7.4.1 INTRODUCTION Jamaica Bay stratification is weakly indicative of a shallow, tidally active environment,

More information

Floats in Polar Oceans. Olaf Boebel and Eberhard Fahrbach, AWI-Bremerhaven, Germany

Floats in Polar Oceans. Olaf Boebel and Eberhard Fahrbach, AWI-Bremerhaven, Germany Floats in Polar Oceans Olaf Boebel and Eberhard Fahrbach, AWI-Bremerhaven, Germany Floats in Polar Oceans: Strategy Development of a system for bipolar use in the Antarctic because of easier conditions:

More information

Thermo Probe, Inc. TP-5 TP-7 TP-8. Instruction Manual English. Switch. Display. Carry Strap. Probe. Ground Clip. Switch. Display.

Thermo Probe, Inc. TP-5 TP-7 TP-8. Instruction Manual English. Switch. Display. Carry Strap. Probe. Ground Clip. Switch. Display. Display Thermo, Inc. Instruction Manual English TP-5 Display TP-7 Display TP-8 1-1 Thermo, Inc. Instructions for use- Models TP-5, TP-7, and TP-8 The manual describes basic function and use of Thermo instruments

More information

The Gas Laws: Boyle's Law and Charles Law

The Gas Laws: Boyle's Law and Charles Law Exercise 6 Page 1 Illinois Central College CHEMISTRY 130 Name The Gas Laws: Boyle's Law and Charles Law Objective The simple laws governing the properties of gases can be readily demonstrated experimentally.

More information

Long term stability tests of INO RPC prototypes

Long term stability tests of INO RPC prototypes Long term stability tests of INO RPC prototypes While the problem of sudden aging in the glass RPC prototypes is being investigated, a few RPCs of dimension 40 cm 30 cm were fabricated, using glass procured

More information

Drift Characteristics of Paroscientific pressure sensors

Drift Characteristics of Paroscientific pressure sensors Drift Characteristics of Paroscientific pressure sensors by Randolph Watts, Maureen Kennelly, Karen Tracey, and Kathleen Donohue (University of Rhode Island) PIES + current meter & CPIES arrays Paroscientific

More information

Implementation of Height Measurement System Based on Pressure Sensor BMP085

Implementation of Height Measurement System Based on Pressure Sensor BMP085 017 nd International Conference on Test, Measurement and Computational Method (TMCM 017) ISBN: 978-1-60595-465-3 Implementation of Height Measurement System Based on Pressure Sensor BMP085 Gao-ping LIU

More information

Applications of Collected Data from Argos Drifter, NOAA Satellite Tracked Buoy in the East Sea

Applications of Collected Data from Argos Drifter, NOAA Satellite Tracked Buoy in the East Sea Applications of Collected Data from Argos Drifter, NOAA Satellite Tracked Buoy in the East Sea Young-Sang Suh (yssuh@nfrdi.re.kr) Ocean Research Team, National Fisheries Research and Development Institute,

More information

Succeeding with Production Air Leak Testing Methods. Paul Chamberlain President, CEO

Succeeding with Production Air Leak Testing Methods. Paul Chamberlain President, CEO Succeeding with Production Air Leak Testing Methods Paul Chamberlain President, CEO 2 Overview Air Leak Testing Overview Pressure Decay Air Leak Testing Factors impacting leak rate measurement Test Data:

More information

Pressure Measurement. Introduction. Engr325 Instrumentation. Dr Curtis Nelson 3/12/18

Pressure Measurement. Introduction. Engr325 Instrumentation. Dr Curtis Nelson 3/12/18 3/12/18 Pressure Measurement Engr325 Instrumentation Dr Curtis Nelson Introduction A cluster of 72 helium-filled balloons over Temecula, California in April of 2003. The helium balloons displace approximately

More information

JEDEC STANDARD. Temperature Cycling. JESD22-A104C (Revision of JESD22-A104-B) JEDEC SOLID STATE TECHNOLOGY ASSOCIATION MAY 2005

JEDEC STANDARD. Temperature Cycling. JESD22-A104C (Revision of JESD22-A104-B) JEDEC SOLID STATE TECHNOLOGY ASSOCIATION MAY 2005 JEDEC STANDARD Temperature Cycling JESD22-A104C (Revision of JESD22-A104-B) MAY 2005 JEDEC SOLID STATE TECHNOLOGY ASSOCIATION NOTICE JEDEC standards and publications contain material that has been prepared,

More information

FAQ to Optimare Precision Salinometer (OPS)

FAQ to Optimare Precision Salinometer (OPS) 1 FAQ to Optimare Precision Salinometer (OPS) collected and answered by Gereon Th. Budéus, AWI Dec 2015 What is the average sample time/throughput? Experience shows that a measurement of a sample takes

More information

WEBB RESEARCH CORPORATION

WEBB RESEARCH CORPORATION Use spacebar or arrow keys to advance slide show WEBB RESEARCH CORPORATION Falmouth, MA APEX Profiler SLOCUM Glider Rev: 06/22/08 WEBB RESEARCH CORPORATION Specialists in neutrally buoyant drifters and

More information

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Lab 1c Isentropic Blow-down Process and Discharge Coefficient 058:080 Experimental Engineering Lab 1c Isentropic Blow-down Process and Discharge Coefficient OBJECTIVES - To study the transient discharge of a rigid pressurized tank; To determine the discharge coefficients

More information

Argo Quality Control Manual For Biogeochemical Data

Argo Quality Control Manual For Biogeochemical Data Argo data management DOI: http://dx.doi.org/10.13155/40879 1 Argo Quality Control Manual For Biogeochemical Data Version 1.0 1 st March 2016 Argo Quality Control Manual For Biogeochemical Data Authors:

More information

Well Test Design. Dr. John P. Spivey Phoenix Reservoir Engineering. Copyright , Phoenix Reservoir Engineering. All rights reserved.

Well Test Design. Dr. John P. Spivey Phoenix Reservoir Engineering. Copyright , Phoenix Reservoir Engineering. All rights reserved. Well Test Design Dr. John P. Spivey Phoenix Reservoir Engineering Copyright 2006-2008, Phoenix Reservoir Engineering. All rights reserved. Objectives After completing this chapter, you will be able to:

More information

Geostrophic and Tidal Currents in the South China Sea, Area III: West Philippines

Geostrophic and Tidal Currents in the South China Sea, Area III: West Philippines Southeast Asian Fisheries Development Center Geostrophic and Tidal Currents in the South China Sea, Area III: West Philippines Anond Snidvongs Department od Marine Science, Chulalongkorn University, Bangkok

More information

Please note that there was an error in the initial proposal: samples should be nominally 1 inch in diameter (see below).

Please note that there was an error in the initial proposal: samples should be nominally 1 inch in diameter (see below). Test schedule for Inter-lab testing Perm/Strength/Vel... October 9, 2008 from David Lockner Dear colleagues, Here is the revised protocol for Inter-lab testing of Strength/Permeability/Wave

More information

Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip.

Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip. Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip. CEE 3310 - Summer 2012 SAFETY The major safety hazard in this laboratory is a shock hazard. Given that you will be working

More information

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide

More information

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER Features ±0.01% FS Measurement & Control Accuracy ±0.001% /ºC Thermal Stability Pressure Ranges from ±1 psid to 1200 psia Applications Barometric Measurement

More information

Kathleen Dohan. Wind-Driven Surface Currents. Earth and Space Research, Seattle, WA

Kathleen Dohan. Wind-Driven Surface Currents. Earth and Space Research, Seattle, WA Updates to OSCAR and challenges with capturing the wind-driven currents. Wind-Driven Surface Currents Kathleen Dohan Earth and Space Research, Seattle, WA ENSO OSCAR Surface currents from satellite fields

More information

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service? Pipette specifications according to ISO 8655 Chapter 5 5.1 The ISO 8655 standard gives the accuracy and precision limits as both absolute and relative values. Specifications will depend on the technique

More information

of Carbon Dioxide (CO 2

of Carbon Dioxide (CO 2 CHAPTER 10 Phase Changes of Carbon Dioxide (CO 2 ) Objectives This experiment is an introduction to phase changes of pure substances and an introduction to some simple microtechniques for doing experiments

More information

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality

More information

FPG8601 Force Balanced Piston Gauge

FPG8601 Force Balanced Piston Gauge FPG8601 Force Balanced Piston Gauge Reference Level Calibration System for very low pressure Pressure range: 0 to 15 kpa gauge, absolute and absolute differential Standard resolution: 0.010 Pa, high resolution

More information

BS12 Piezoresistive OEM Pressure Sensor

BS12 Piezoresistive OEM Pressure Sensor BS12 Piezoresistive OEM Pressure Sensor Feature Pressure range (-0.1~60)Mpa; Pressure reference: Gauge, absolute and sealed gauge; Constant current power supply; Isolated construction to measure various

More information

Kube Engineering Quinn E. Eddins, B.S.E.E Phone

Kube Engineering Quinn E. Eddins, B.S.E.E Phone Process Information: There are four ph assemblies used to monitor the ph value on a Sulfur Dioxide absorber. The chemical composition is approximately 20-30% Sodium Bisulfite, 4% Sodium Sulfite, and the

More information

Technical Information Conducal CLY421

Technical Information Conducal CLY421 TI00496C/07/EN/01.12 71142150 Products Solutions Services Technical Information Conducal CLY421 Conductivity calibration set for ultrapure water applications Factory calibrated measurement box Certified

More information