Scanning Electron Microscope JEOL JSM F

Size: px
Start display at page:

Download "Scanning Electron Microscope JEOL JSM F"

Transcription

1 Scanning Electron Microscope JEOL JSM F

2 How to Use This Manual You will be promised to obtain successful results if you follow this step by step manual gently and carefully. In order to that, you must understand following precautions. Caution Measure height of the sample holder on your specimen correctly Locate the sample holder correctly into sample exchange chamber Be gently to the rod; do not push or pull it in a disorderly manner Understand difference between actual height (Z) and Working Distance(WD) Do not use your Power, Force and Strength, because this is the Fragile, Sensitive and Delicate instrument. Proceed every single step GENTLY and CAREFULLY Never set any magnetic specimens. Magnetic properties damage FE-SEM-J seriously. Use FE-SEM-F for magnetic specimens 1

3 Operational flow FE-SEM-J Observation Flow 2

4 3 FE-SEM-J Observation Flow Step By Step

5 26mm Offset 38mm 0 ~ 30mm 1. Confirm error messages appears on the screen Make sure there are no error message on the SEM user interface. 2. Mount the specimen on the sample holder The height of the specimen must not exceed the cylinder top surface!! *If your specimen protrudes above the cylinder s top surface (or the top face of another holder), you need to measure the offset between the specimen and holder top surfaces. In order to read the offset, use the scale (see Fig. 1). This offset value is necessary when loading your sample into the SEM chamber. Take a note of it. The protrude height of specimen must be less than 30mm from the Cylinder top. SEM Column Maximum height of specimen Height of sample holder 12.5mm Cylinder s top surface Fig.1. Scale for measuring the protrude height from the specimen holder. The protrude height of specimen must be less than 30 mm from the cylinder top surface. The distance between 4 cylinder s top surface and SEM column must be always 38mm, which is equal as the specimen exchange position.

6 5 FE-SEM-J Observation Flow Step By Step

7 Select an appropriate sample holder for your material. Mount material correctly on the specimen holder with carbon tape. For example, the specimen holder image below, specimen surface and cylinder s top surface must be equal; specimen surface never protrude the height of cylinder s top surface (Fig.2). In case when the specimen surface need protrude the cylinder s top surface, you must obey the height limitation of 30mm. Never mount specimen with height more than 30 mm from the cylinder top. Otherwise, you will damage the SEM column. Fig.2. Specimen positioning on 12.5 and 26 mm holders. (Diagram taken from JEOL s manual.) 6

8 7 FE-SEM-J Observation Flow Step By Step

9 3. Insert Specimen into SEM Chamber Insert Your Specimen into the Specimen Exchange Chamber (SEC) 1 Make sure the user interface is presented in the monitor (Fig.3.1) (*1). 2 The stage position must be at the Specimen Exchange Position of (X, Y, Z) = (0.0, 0.0, 38.0), R=0.0, T=0.0 (Fig.3.2.h). 3 Confirm the Specimen Exchange Chamber (SEC) must be locked (Fig.3.2.g) and the illumination of EVAC and EXCH POSN buttons below the SEC must be ON (Fig.3.2.i, j) (*2). Also, the airlock isolation valve between the SEM chamber and the SEC is an open state (Fig.3.3.e), and both areas are in the high vacuum condition. 4 Click VENT of the user interface (Fig.3.1.g-1) VENT button below the SEC starts blinking. At the same time, the airlock isolation valve between the SEM chamber and SEC is closed (Fig.3.3.k). The SEC will be in air condition after N 2 gas purge into that room; meanwhile, the SEM chamber keeps high vacuum condition (Fig.3.3.l). 5 When VENT button stop blinking, remove the clamp (Fig. 3.2.h), then, open the door of the SEC. (*1)When the user interface is not running, click SEM icon on the desktop. Then, activate SEM with guest account. Pass word is not necessary. (*2) When the lights are not illuminated, a. Click OFF button below the observation tab in the user interface (Fig.3.1.a). b. Confirm the SEM column gate is closed (Fig.3.1.d). * Please call our staff if the SEM Column Gate are not able to close. c. Click Observation button (Fig.3.1.a). d. Click Specimen Exchange button(fig.3.1.f) 8

10 9 FE-SEM-J Observation Flow Step By Step

11 a b c Airlock isolation valve Open d f e g-1 g-2 Fig.3.1. FE-SEM-J User interface h i Airlock isolation valve k l Close SEC SEM chamber Air condition High vacuum condition 10 Fig.3.2. Specimen Exchange chamber (SEC) and control panel of SEC j Fig.3.3. Vacuum condition of the SEM chamber and the SEC: (k) the airlock isolation valve, (l) the SEC is in air condition but the SEM chamber keeps high vacuum condition

12 11 FE-SEM-J Observation Flow Step By Step

13 6 Mount specimen holder on the base of holder in the SEC (Fig.3.4~3.6) (*3) (*3) a. Specimen with the protrude height less than 30mm from the Cylinder top is allowed to mount. b. Make sure flat side of the specimen holder set perpendicular to the insertion direction. Caution!! The specimen holder will drop off into SEM chamber and damaged SEM if it is not mounted correctly. Thus, Gently and carefully insert the specimen holder to the base. Do not make a gap between the specimen holder and the base. m n o Flat side Fig.3.4.Specimen holder :( m) top view (a diagram from JEOL user Manual), (n) top view of the actual specimen holder, (o) side view of specimen holder, indication of flat side Flat side Attached correctly Insert direction the base of holder in the SEC Gently insert the specimen holder to the base. Keep the flat side perpendicular to the insertion direction Flat side of the specimen holder must be attached to flat side of the base Fig.3.5. Correct way of locking specimen holder with the holder s flat side perpendicular to the insertion direction. 12

14 13 FE-SEM-J Observation Flow Step By Step

15 O ring 7 Before closing the door of the SEC, Confirm whether the O ring is set to the correct position without contamination (Fig.3.6).Do not touch O ring by your bear hands. 8 Gently close the door of the SEC and lock it with the cramp (Fig.3.2.h). Click EVAC button of the user interface (Fig.3.1.g-2). EVAC button below the SEC start blinking. Do not unlock the clamp during the blinking because the airlock is pumping to the high vacuum. 9 Confirm the lights of EVAC and EXCH POSN buttons have lit up (Fig.3.2.i,j). 10 Confirm the airlock isolation valve between the SEM chamber and the SEC is an open state (Fig.3.1.e), and both areas are in the high vacuum condition. Fig The specimen holder that is installed correctly to the base. p q Stopper 14 Fig.3.7. The overview of the SEC with the rod: (p) the rod at the neutral position, (q) the rod stopper. The rod stopper must be in the standing state when the rod is at the neutral position.

16 15 FE-SEM-J Observation Flow Step By Step

17 Transform Specimen Holder from the SEC to the SEM Chamber 11 Gently and carefully lower the rod horizontally without pulling along its axis from the neutral position (Fig.3.7.p). Keep the rod fully horizontal position(fig.3.8) 12 Gently and carefully insert the rod by pushing horizontally along the rod s axis until you feel it come to a firm stop (Fig.3.9.1,2). Then, apply little more force to lock the holder in the SEM stage. The rod is now completely contained in the SEM chamber. (Fig.3.9.3,4). 13 After confirming that the HDLR lights has lit up (Fig.3.8.s), fully retract the rod horizontally until the rod stopper snaps and comes up (Fig.3.9.6). 14 Lifting of the rod to the neutral position without using any force along the rod s axis (Fig.3.8.r). 15 Once the Specimen holder located into the SEM stage correctly, pop-up menu of the specimen holder selection appears on the user interface(fig.3.9.5). 16 Select the specimen holder currently in use (Fig.3.9.t). Enter offset value from 0mm to 30 mm in the specimen surface offset field (Fig. 3.9.u), which was measured with the scale (Fig.1). Once the rod reach to the SEM stage, it firmly stopped(fig.3.9.2). However, you must apply little more force to lock the holder in the SEM stage. When the specimen holder is not locked correctly to the stage, pop-up menu of the specimen holder selection is not appeared (Fig.3.9). When that pop-up menu is not appeared, Make sure whether you completely insert the rod to the SEM chamber Do not push the rod forcibly when it stops at the unexpected area. There might be problem. Make sure the lights of SEC are all illuminated except VENT after the specimen is able to locate to the SEM stage correctly. s r Keep flat Fig.3.8. (r) Fully horizontal position. The rod is held extended due to friction in sliding metal parts. The rod stopper (Fig.3.7.q) is pushed down, (s) HDLR light on after the specimen holder was is set up to the SEM stage correctly. 16 When the rod reach the fully horizontal position, you may feel that the rod will be slightly drawn into the SEM chamber; although, that slight movement will stop automatically by the friction.

18 17 FE-SEM-J Observation Flow Step By Step

19 The rod come to a firm stop here 4 The rod stopper snaps and come up 5 t u Fig.3.9. process to insert the rod into the SEM chamber: 1Fully horizontal position, 2the rod come to a firm stop around here, 3the rod is completely inside the SEM chamber, 4 top view of 3, 5pop-up menu of the specimen holder selection, (t) Selected specimen holder, (u) input the offset value you measured with the scale (Fig.1) 18

20 19 FE-SEM-J Observation Flow Step By Step

21 4. Observe your specimen 1 Wait until the chamber vacuum is at 5x10-4 Pa or lower. Open the Gun Isolation Valve by clicking the ON button under Observation. 2 Make sure the ZFC is activated (Fig.3.10.v) (*4). If ZFC is not activated, click the ZFC icon. 3 Select High mag mode 4 Click WD(Working Distance) and select 8mm from the pull down menu (Fig.3.10.x-1) (*5) or Home Position (Fig.3.10.x-2)(*6) 5 Open the Gun Isolation Valve by clicking the ON button under Observation (Fig.3.10.w) (*4) When ZFC is activated, the stage height (Z) mechanically move to (8mm + Offset) position. As a result, image will be focused approximately WD=8mm area. (*5) Do not select WD at Low Mag mode. You have to proceed this at High Mag mod. (*6) When you click Home position, the SEM stage automatically moves to (x, y, z) = (0.0, 0.0, 8.0), R=0, T=0. Therefore, you have to make sure that you input the offset correctly; otherwise, you will seriously damage the SEM column. w z-3 z-2 z-1 v x-1 y x-2 Fig Main window of User interface: (v) ZFC mode, (w) electron beam on / off button, (x- 1) pull down menu of WD, (x-2) Home position button, (y) Specimen Exchange button, (z-1) camera button for capturing image,(z-2) Freeze button also for capturing image,(z-3) 20 Setting tab for integration set up

22 21 FE-SEM-J Observation Flow Step By Step

23 6 Adjust focus. The image will be focused at around WD = 8mm, if the correct offset number was set up. Stigma, contrast and brightness also need to adjust. 7 Search an area of interest. The Low mag mode is suitable to observe the over view of your specimen. 8 Increase magnification little by little, if required. 9 Save image. There are three process to save images a. Click Camera button (Fig.3.10.z-1). After that, Save as dialog box appears after the image once captured. b. Click Freeze button (Fig.3.10.z-2) once. The image is frozen after scanning currently ongoing finished. Select Save as from a pull down menu of File in the main menu. c. If you prefer to capture images with integration mode, you need to set up integration before clicking CF button. You can set up it with Setting (Fig.3.10.z- 3) basic setting view/ scan tab. You are allowed to create a holder within the SEM J Datastasion only. Pick up your data from the Datastation PC in the room 009 by your pen drive. After that, delete your data from the Datastation PC. Protect your data by yourself. Keep in mind that data within the Datastation will be completely deleted without notice. Basic Condition for Observation Emmision current / LC6 / 0.5~30 kv Acceleration Voltage SEM mode Minimum magnification 500x LM mode Minimum magnification 25x (depending on the spot size) Image shift Avilable over x Axisi alignemnt Wobbler starts when beam alignement is clicked. Minimize movement of the image using X and Y knob of Stigma. Capturing image Camear CF Charging effect free mode (reduce Dwell Time) Freeze Click once /capture a image with current setting scan speed // Click twice / immediately stop scanning. Integration Setting basic setting view/scan tab check use integration. Image is captured by clicking Freeze. 22

24 23 FE-SEM-J Observation Flow Step By Step

25 5. Exchange Specimen or Finishing Observation 1 Click the OFF button under Observation tab in the user interface (Fig.3.10.w) 2 Click the Exchange Position button (see Fig.3.10.y). SEM stage will be placed to the Specimen exchange position of (Z) = 38mm. 3 Gently insert the rod into the SEM chamber and transfer the specimen holder to the specimen exchange chamber (SEC) (Fig.3.9). 1. Gently and carefully lower the rod horizontally without pulling along its axis from the neutral position(fig.3.7.p) 2. Gently but completely insert the rod into the SEM chamber. 3. Gently and carefully retract the rod until the rod pass the stopper. 4. Lifting of the rod to the neutral position without using any force along the rod s axis (Fig.3.8.r). 4 Click VENT button on the user interface (Fig.3.1). 5 After VENT button below the SEC illuminated, open the door of SEC and remove the specimen holder For specimen exchange, set up a new specimen and repeat 3. Insert specimen into SEM chamber, 4.Observe your specimen 2. When you finish observation, close the door of the SEC and lock it with the cramp. Then, click EVAC button (Fig.3.1). 7 You are free to go after confirming following; No error message appeared on the user interface Illumination of EVAC and EXCH POSN below the SEC are now switch on Illumination of HLDR are off, which means there are no specimen holder on the SEM stage 24

TYPICAL OPERATION PROCEDURES

TYPICAL OPERATION PROCEDURES JSM-7500F TYPICAL OPERATION PROCEDURES Reminders! Remember to sign in on the sign up sheet. WEAR GLOVES: always wear gloves when handling your samples, the sample holders and the specimen exchange rod.

More information

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager. LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager. Please contact in case of emergency: SEM microscopist: Mr. Eric Miller, 7-0789

More information

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS Fabrication Laboratory Revision: A Rev Date: 09/29/03 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users

More information

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i LEO SEM SOP Page 1 of 9 LEO 440 SEM SOP Gun (Filament) Column Manual Valves Chamber Window Chamber Stage Movement Leica Leo Stereoscan 440i 1. Scope 1.1 This document provides the procedure for operating

More information

SEM LEO 1550 MANUAL. I. Introduction. Generality

SEM LEO 1550 MANUAL. I. Introduction. Generality SEM LEO 1550 MANUAL RESERVATION POLICY: 2 booking slots maximum per day and per person (ie. 1h). 6 booking slots maximum per week and per person (ie. 3h). Reservation names must correspond to the operators.

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

Angstrom Dielectric Sputterer Operation Manual

Angstrom Dielectric Sputterer Operation Manual Angstrom Dielectric Sputterer Operation Manual I. System overview The Angstrom Dielectric Sputterer (ADS) has a similar interface as the Angstrom metal sputterer. It has two screens, the process screen

More information

Login to ilab Kiosk. Revised 05/22/2018. Load your sample:

Login to ilab Kiosk. Revised 05/22/2018. Load your sample: Login to ilab Kiosk Load your sample: 1. Check: The analysis chamber pressure is

More information

[needs to be discussed with faculty if any new material to be done]

[needs to be discussed with faculty if any new material to be done] Tool Name: EVO 18 Version 1 Authorized Users: Gayatri, Vishnu kant Materials Used: Sample holder has facility to mount 9 stubs: Si, Glass, Materials allowed : [needs to be discussed with faculty if any

More information

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: MJB4 Mask Aligner Operating Procedure Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Karl Suss MJB4 Mask Aligner, Optical Lithography,

More information

Standard Operating Procedure. For. PVD E-Beam

Standard Operating Procedure. For. PVD E-Beam P a g e 1 Standard Operating Procedure For PVD E-Beam P a g e 2 Introduction The PVD Electron-Beam Evaporator (E-Beam) thin film deposition machine uses a magnetically guided and collimated stream of electrons

More information

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test. Angstrom E-Beam Instructions Tool Manager: Joe Palmer (jpalmer@princeton.edu) Office: 8-4706; Cell:609-751-1353 Backup: David Barth (dbarth@princeton.edu) Office: 8-4626; Cell: 610-405-8227 PROCESS CHECKS:

More information

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER CAUTION: THE CHAMBER PLATE GETS EXTREMELY HOT Start Up Procedure 1) Open bottle and regulator for Helium,

More information

Approved by BNC Managing Director Paul Lum Date: Approved by PI (s) /Super User (s): _Peter Hosemann/Andy Minor_Date:

Approved by BNC Managing Director Paul Lum Date: Approved by PI (s) /Super User (s): _Peter Hosemann/Andy Minor_Date: Approved by BNC Managing Director Paul Lum Date: Approved by PI (s) /Super User (s): _Peter Hosemann/Andy Minor_Date: Standard Operating Procedure BNC FEI Quanta 3D FEG/FIB ESEM Version 2011 Aug 30 I.

More information

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 The front door of the main body of the instrument may be open when you arrive. Take the sample chamber and slide it into position with the

More information

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100 PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100 June 2013 Interface Overview. The Oxford software is divided into 5 main screens. 1) Pump Control page 2) Recipe page 3)

More information

OPERATION OF THE DIMPLER

OPERATION OF THE DIMPLER OPERATION OF THE DIMPLER After thinning your sample to ~80 μm you can now do a dimpling process to thin the center up to 10 μm. When you walk in and use the DIMPLER it should already be calibrated and

More information

1)! DO NOT PROCEED BEYOND THIS MARK

1)! DO NOT PROCEED BEYOND THIS MARK Operating Instructions for X-ray Photoelectron Spectrometer: Physical Electronics Model 555 XPS/AES (John H. Thomas, III, Ph.D., Electron Spectroscopy) Sample Insertion: figure 1. Sample insertion rod

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Branson IPC 3000 O 2 Asher Copyright 2014 by Hong Kong University of Science & Technology. All rights reserved. Page 1 Contents 1. Picture and Location 2. Process Capabilities

More information

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves

More information

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure PPMS Service Note 1099-412 In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure Introduction: Loss of electricity to the PPMS EverCool II should not cause damage to

More information

Linkam Scientific Instruments USER GUIDE DSC600. Temperature Controlled Stage

Linkam Scientific Instruments USER GUIDE DSC600. Temperature Controlled Stage Linkam Scientific Instruments DSC600 Temperature Controlled Stage USER GUIDE Contents Before Setting Up Your Equipment... 3 Important Notice... 4 Warranty... 4 Technical Support... 4 Equipment Maintenance...

More information

David Cockayne Centre for Electron Microscopy.

David Cockayne Centre for Electron Microscopy. David Cockayne Centre for Electron Microscopy. Depart of Materials, University of Oxford JEOL JEM-2100 TEM USER GUIDE November 2014 JLH / NPY 1 Contents Control panels...4 Tracker Ball...7 Aperture controls....8

More information

Warnings: Notes: Revised: October 5, 2015

Warnings: Notes: Revised: October 5, 2015 Karl Suss MA6 Mask Aligner Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any campus

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Branson IPC 3000 O 2 Asher Page 1 of 14 Contents 1 Picture and Location 2 Process Capabilities 2.1 Cleanliness Standard 2.2 Recipes 2.3 Performance of Branson IPC 3000 O 2 Asher

More information

Karl Suss MA6 Mask Aligner SOP

Karl Suss MA6 Mask Aligner SOP Page 1 of 11 Karl Suss MA6 Mask Aligner SOP Safety UV Exposure: The high energy light produced by the high pressure Mercury Xenon lamp can cause eye damage and skin burns. Be sure that the light guards

More information

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work.....2

More information

Thermo K-Alpha XPS Standard Operating Procedure

Thermo K-Alpha XPS Standard Operating Procedure Thermo K-Alpha XPS Standard Operating Procedure Quick Guide Draft v.0.1 Procedure overview 1. Vent the loadlock 2. Secure your sample to the stage using clips, check the height of the final assembly. 3.

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual ARC12M Sputter Copyright 11.2015 by Hong Kong University of Science & Technology. All rights reserved. Page 1 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness

More information

Glove Box Installation Manual

Glove Box Installation Manual Glove Box Installation Manual 1998 by M. Braun Company File: GB-UNI-INS.DOC! Edition 08-00 by M. Boutin! Subject to be changed without notice Glovebox Installation Your Glove box has been fully assembled,

More information

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA Page 1 of 9 SSI Solaris 150 RTA The Solaris 150 RTA is a rapid thermal annealing system capable of handling sample sizes up to 100mm (4 diameter) or smaller. The system can anneal in N 2 and Forming gas

More information

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel Date Created: May 3, 2004 Date Modified: June 1, 2005 MA6/BA6 Mask Aligner Equipment Standard Operating Procedure Greg Allion and Kimberly Appel 1. Purpose 1.1. Photolithography involves transferring a

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Allwin21 AW610 RTP Page 1 of 18 Contents 1 Picture and Location 2 Process Capabilities 2.1 Cleanliness Standard 2.2 Recipes 2.3 Performance of Allwin21 AW610 RTP 3 Contact List

More information

The SPI Sputter Coater Handbook

The SPI Sputter Coater Handbook The SPI Sputter Coater Handbook Coating of Specimens SPI-Module Sputter Coater with Etch Mode 1. Mount the specimens onto the SEM stub. Keep in mind that many adhesives have high vapor pressure solvents

More information

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering Edge Isolation Tool Standard Operating Procedure Version 1.1 Date: 20-9-2012 Prepared by, Sandeep S S Department of Electrical Engineering IIT Bombay This document is meant for internal circulation only

More information

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE Purpose of this Instrument: This instrument is for patterning photosensitive polymers with UV light. Location: White Hall 410 Cleanroom Primary

More information

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI NanoSight NS300 Operation instructions Injection/flushing brief overview: 1. Do not exceed flow of 1 ml per 20 seconds. 2. Inject two 1 ml syringes with nano-pure or DI water. 3. If the water does not

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Xactix XeF2 Etcher. All users are expected to have read and understood this document.

More information

Linkam Scientific Instruments USER GUIDE. LTS420 and LTS120. Temperature Controlled Stage

Linkam Scientific Instruments USER GUIDE. LTS420 and LTS120. Temperature Controlled Stage Linkam Scientific Instruments LTS40 and LTS10 Temperature Controlled Stage USER GUIDE Contents Before Setting Up your Equipment... 3 Important Notice... 4 Warranty... 4 Technical Support... 4 Equipment

More information

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage Linkam Scientific Instruments TST350 Temperature Controlled Tensile Stress Testing Stage USER GUIDE Contents Before Setting Up Your Equipment... 3 Important Notice... 4 Warranty... 4 Technical Support...

More information

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

Nanofabrication Facility: PECVD SOP Rev. 00, April 24 Author: Charlie Yao & Mario Beaudoin Email: charlieyao@gmail.com; Beaudoin@physics.ubc.ca Phone: 604-822-1853(MB). Purpose This document outlines the standard operation for the Trion Plasma Enhanced Chemical

More information

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300 Technical Note GA-300 Overview The GA-300 CO2 and O2 Gas Analyzer is easy to use, robust, and adaptable to human, animal, and plant applications. The GA-300 has two analog outputs to allow recording and

More information

UNDERGROUND SURVEY WITH MINEMODELLER

UNDERGROUND SURVEY WITH MINEMODELLER UNDERGROUND SURVEY WITH MINEMODELLER A Tutorial Derek C. Diamond derek@primethought.biz CONTENTS Underground Survey with MineModeller... 2 Introduction... 2 Requirements... 2 Getting Started with MineModeller...

More information

Operating Procedures for Metal Evaporator I

Operating Procedures for Metal Evaporator I Operating Procedures for Metal Evaporator I Metal Evaporator I is intended as a tool and a training device. Understanding the operation of this equipment should give you a basic knowledge of vacuum and

More information

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the

More information

r β tr t t s t r s t s tr tt r t r t r r r tt r e s tt e tt r e s tt r e r r e r 2 t ö t t r s s t t r r r ts r rr t

r β tr t t s t r s t s tr tt r t r t r r r tt r e s tt e tt r e s tt r e r r e r 2 t ö t t r s s t t r r r ts r rr t tr µs 2 r β t r tr t s r t t rst 1 2 s t s tr tt r t r t r r r tt r e s tt e r 2 st e tt r e r2 e r s tt r e r r e r 2 t ö t t r s s t t r r r ts r rr t t s e s r r tt rs r LaB 6 ss s r t s r t3 t r t3

More information

CircuFlow Quick Setup Guide

CircuFlow Quick Setup Guide CircuFlow 5200 Quick Setup Guide LB04.0001 Rev B 20140314 TABLE OF CONTENTS Setup Therapy Setup Gradient Mode Pressure Mode Lock Device Unlock Device Treatment 03 04 05 08 11 12 13 02 SETUP 01 01 Open

More information

Standard Operational Procedure for Capillary Electrophoresis

Standard Operational Procedure for Capillary Electrophoresis Page 1 of 14 Standard Operational Procedure for Capillary Electrophoresis *Please ask an experienced operator for training before you use the CE if you can. A demo is most probably more efficient than

More information

SPLIT HOPKINSON PRESSURE BAR OPERATIONS MANUAL 5/9/2014

SPLIT HOPKINSON PRESSURE BAR OPERATIONS MANUAL 5/9/2014 B Calibration of Split Hopkinson Pressure Bar Setup: Calibration is performed by comparing the wave speed of the initial pulse versus the reflected pulse and calculating the Energy Ratio. This ratio gives

More information

Quintel Mask Aligner

Quintel Mask Aligner Quintel Mask Aligner (quintel) 1.0 1.0 Title Quintel Q4000 Mask Aligner 2.0 2.0 Purpose The Quintel Q4000 MA (Mask Aligner) is a top and bottom side contact lithography printer with the video-view split

More information

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents Title: CHA E-Beam Evaporator Page 1 of 7 Table of Contents Purpose/Scope... 2 2.0 Reference Documents... 2 3.0 Equipment/Supplies/Material... 2 4.0 Safety... 2 5.0 Set Up Procedures... 2 5.1 PC Logon and

More information

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300 Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300 The Thermo Scientific icap 6300 Inductively Coupled Plasma Optical Emission

More information

Mass Spec will not Autotune

Mass Spec will not Autotune Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems

More information

Quick Start Guide. A. Hardware installation B. Software installation C. Start the software for the first time D. Do your first measurement

Quick Start Guide. A. Hardware installation B. Software installation C. Start the software for the first time D. Do your first measurement Quick Start Guide This Quick Start Guide describes the hardware and software installation process and the Measurement feature in a simple way. Please follow the sequence of the steps to avoid problems

More information

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the STS ASE. All users are expected to have read and understood this document. It is

More information

Napa Technology Trouble Shooting. For Premier & Premier PLUS Models

Napa Technology Trouble Shooting. For Premier & Premier PLUS Models Napa Technology Trouble Shooting For Premier & Premier PLUS Models Before contacting Napa Technology for support, please check if the problem and solution are found below: Machine Is Off & All LCD s are

More information

Using the Akta Prime plus October 22, 2012

Using the Akta Prime plus October 22, 2012 Some starting precautions: 1. Vacuum filter all buffers. Removes any large particles/debris that may clog your column De-gases the buffers 2. Clarify lysates first by centrifugation and then filtration

More information

OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference.

OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference. English Lubrication Free Air Turbine Handpiece with Water Spray OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference. OM-T0286E 001 Thank you for purchasing

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC OAI 200 Lithographic Mask Aligner (Aligner 3) Version 2011 June 2 I. Purpose This Standard Operating

More information

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cover Page for Lab Report Group Portion. Head Losses in Pipes Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section

More information

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 QUICK START GUIDE P/N 6006154, REVISION C MAY 2013 Model 8120 Mask Integrity Tester is patented under U.S. Patent No. 8,312,761. Additional patents are pending.

More information

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity Rev. B 2017,Janyary,20 Next > Preconditions 7900, 7800, 7700, or 8800 ICP-MS is used. Sample introduction type is PeriPump. If you are

More information

Updated: Variable Temperature

Updated: Variable Temperature Updated: 2016-11-02 Variable Temperature Sample Preparation Make sure the probe is rated for the ultimate/final temperature Fully degas the sample, particularly if going to low temperature Be aware of

More information

[USING THE VITROBOT April 9, Using the Vitrobot

[USING THE VITROBOT April 9, Using the Vitrobot Using the Vitrobot The Vitrobot Mk 3 (aka Mark III, Mark 3, etc.) can control both the local temperature and the local humidity during the process of plunge-freezing a grid for cryoem. It does require

More information

Cryo-Evaporator Operation

Cryo-Evaporator Operation Cryo-Evaporator Operation Cara Ricci November 25, 2003 Thin Film Deposition THE UNIVERSITY OF TEXAS AT DALLAS ERIK JOHNSON SCHOOL OF ENGINEERING DOCUMENT NUMBER: FA2003-TF-008 EDITION: 1.1 PAGE: 1 of 28

More information

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1. Nanomaster NSC-3000 DC Magnetron Sputter Tool Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly

More information

SOLUBILITY OF A SOLID IN WATER

SOLUBILITY OF A SOLID IN WATER 1516L Experiment 1 SOLUBILITY OF A SOLID IN WATER Objectives In this experiment you will determine the solubility of potassium nitrate (KNO 3 ) in water at various temperatures. You will prepare a plot

More information

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1 HPICAL-15000 Operation & Data Logging Procedures Click spacebar to advance through slides 1 WARNING Always wear proper safety equipment when using high pressure equipment. Do not exceed 125 psi air pressure.

More information

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS WARNING THE WARRANTIES MADE BY MSA WITH RESPECT TO THE PRODUCT ARE VOIDED IF THE PRODUCT IS NOT USED AND MAINTAINED IN ACCORDANCE

More information

COBILT CA-800 Mask Aligner Equipment Operation

COBILT CA-800 Mask Aligner Equipment Operation COBILT CA-800 Mask Aligner Equipment Operation For the Micro-Electronics Laboratory At University of Notre Dame Department of Electrical Engineering This user manual is not be removed from room 247A. This

More information

University of Minnesota, MN Nano Center Standard Operating Procedure

University of Minnesota, MN Nano Center Standard Operating Procedure Equipment Name: HDPCVD Revision Number: 2 Badger Name: HDPCVD Revisionist: L. von Dissen Model: Advanced Vacuum Date: 10/25/2016 Apex SLR ICP Location: PAN, Bay 3 1 Description The Apex SLR ICP is a high

More information

FOURIER TRANSFORM INFRARED SPECTROSCOPY

FOURIER TRANSFORM INFRARED SPECTROSCOPY Purpose and Scope: This document describes the procedures and policies for using the MSE Bruker FTIR Spectrometer. The scope of this document is to establish user procedures. Instrument maintenance and

More information

General remarks regarding ALS robot system:

General remarks regarding ALS robot system: General remarks regarding ALS robot system: Preparing pucks at home. a) Preparing pins: - for ALS pucks use ONLY the special Hampton pins or spine pins - break off the 2 bottom sections of the standard

More information

University of Minnesota Nano Fabrication Center

University of Minnesota Nano Fabrication Center Equipment name: Karl Suss Mask Aligners Badger name: MA6 & Maba6 Revision number: 4 Models: Revisionist: Paul Kimani Location: Bay 2 Date: April 25, 2013 1. Introduction The Karl Suss MA6 and MABA6 are

More information

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating Plasma-Therm PECVD A PECVD (plasma enhanced chemical vapor deposition) reacts gases in a RF (radio frequency) induced plasma to deposit materials such as silicon dioxide and silicon nitride. This PECVD

More information

Industrial Pneumatics

Industrial Pneumatics Industrial Pneumatics Industrial Training Manual 1 Level 1 Version 3.0 April 2015 This manual was developed for use with the following products: FluidSIM Pneumatics 5.0 English DEPCO Pneumatics Training

More information

Using the Smart IRT diamond ATR.

Using the Smart IRT diamond ATR. 1. Installing the Smart ITR ATR Using the Smart IRT diamond ATR. 1.1. Remove any accessory in the Nicolet 6700. 1.2. Remove the cover of the FTIR s sample compartment using the thumbscrew at the center

More information

Pressure Dump Valve Service Kit for Series 3000 Units

Pressure Dump Valve Service Kit for Series 3000 Units Instruction Sheet Pressure Dump Valve Service Kit for Series 000 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00, 400, 500, and 700 applicator

More information

Oerlikon Sputtering Evaporator SOP

Oerlikon Sputtering Evaporator SOP Oerlikon Sputtering Evaporator SOP Short UNT Cleanroom 1. Taking out sample holder from Transport Chamber : Log in FOM to access the software Go to the software and log in with user1 and password user1

More information

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP Page 1 of 5 NRF Unaxis ICP/RIE Etch SOP Unaxis Shuttlelock Reactive Ion Etcher with Inductively Coupled Plasma Module. Etch Capabilities: SiO2, Si3N4, Al, dielectrics and other commonly used materials.

More information

Telescoping Attic Stair Operating Instructions

Telescoping Attic Stair Operating Instructions Telescoping Attic Stair Operating Instructions 1. Opening the stairs: a. The provided pull cord and pull rod will be used to open and close the attic stair. If your stair has a pull cord, pull on the cord

More information

Burner Management System DEMO Operating instructions

Burner Management System DEMO Operating instructions Burner Management System DEMO Operating instructions Burner Management System DEMO Operating Instructions Startup Summary - Normal startup is accomplished in four basic steps: 1. Leak Test a. Safety Valve

More information

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7 Title: Heat Pulse 610 Operating Procedure Page 1 of 7 Table of Contents 1.0 Purpose/Scope... 2 2.0 Reference Documents... 2 2.1 Heat Pulse Operations Manual... 2 2.2 Solaris software Operations Manual...

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC OAI Lithographic Mask Aligner (Aligner 2) Version 2008 October 31 I. Purpose This Standard Operating

More information

A Development of Mechanical Arm for Weight Scale Calibration

A Development of Mechanical Arm for Weight Scale Calibration A Development of Mechanical Arm for Weight Scale Calibration Pattaraweerin Woraratsoontorn Department of Industrial Physics and Medical Instrumentation, Faculty of Applied Science, King Mongkut s University

More information

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT ATTENTION SYRINGE REPLACEMENT Please read through the document completely before starting any repairs. Refer to the proper section in the service manual for complete removal and replacement procedures.

More information

CONSUMER MODEL INSTALLATION GUIDE

CONSUMER MODEL INSTALLATION GUIDE CONSUMER MODEL INSTALLATION GUIDE System requirements Windows System Requirements To use your TOMI and its software, your system should have: A Microsoft Windows compatible PC with a Pentium IV processor

More information

Pump out insert (condenser and still lines) overnight with turbo pump

Pump out insert (condenser and still lines) overnight with turbo pump Running the Dilution Fridge Before starting a run: Pump out insert (condenser and still lines) overnight with turbo pump Make sure LN and LHe traps are clean pump out with He4 rotary pump Start He3 rotary

More information

SOP for Karl Suss MJB3 #1 Mask Aligner

SOP for Karl Suss MJB3 #1 Mask Aligner SOP for Karl Suss MJB3 #1 Mask Aligner Rev. 5 (30/11/2016) Safety UV Exposure: The high-energy light produced by the high-pressure Mercury Xenon lamp can cause eye damage and skin burns. Be sure that the

More information

Standard Operating Manual

Standard Operating Manual NANOELECTRONICS FABRICATION FACILITY (NFF), HKUST Standard Operating Manual STS ICP-DRIE Copyright 2014 by Hong Kong University of Science & Technology. All rights reserved. Page 1 NANOELECTRONICS FABRICATION

More information

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner Revision date September 2014 2 Emergency Plan for Karl Suss MA6 Aligner Standard Operating Procedures for Emergencies Contact information

More information

SOLUBILITY OF A SOLID IN WATER

SOLUBILITY OF A SOLID IN WATER 1516L Experiment 2 SOLUBILITY OF A SOLID IN WATER Objectives In this experiment you will determine the solubility of potassium nitrate (KNO 3 ) in water at various temperatures. You will prepare a plot

More information

Linkam Scientific Instruments USER GUIDE DSC600. Temperature Controlled Stage

Linkam Scientific Instruments USER GUIDE DSC600. Temperature Controlled Stage Linkam Scientific Instruments DSC600 Temperature Controlled Stage USER GUIDE Contents Before Setting Up Your Equipment...3 Important Notice...4 Warranty...4 Technical Support...4 Equipment Maintenance...4

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC Commonwealth Dual Ion Beam Deposition System (CDIBS) Version 2010 February 14 I. Purpose This Standard

More information

HAI UPB Wall Switch and Auxiliary Switch Installation and Operating Instructions

HAI UPB Wall Switch and Auxiliary Switch Installation and Operating Instructions HAI UPB Wall Switch and Auxiliary Switch Installation and Operating Instructions For the following Models: 35A00-1 HAI 600W Dimmer Switch, 35A00-3 HAI 600W Non-Dimming Switch (collectively referred to

More information

Karl Suss MJB4 Mask Aligner

Karl Suss MJB4 Mask Aligner Karl Suss MJB4 Mask Aligner Tool Manager: Yong Sun ( yongs@princeton.edu; Office 8-8234; Cell 609-917-5076 ) Backup: George Watson ( gwatson@princeton.edu; Office 8-4626; Cell 732-996-2713 ) ******************************************************************************

More information

3. Align ratchet drive with female insert on assembly, and raise ratchet handle until the ratchet drive is firmly seated into the assembly.

3. Align ratchet drive with female insert on assembly, and raise ratchet handle until the ratchet drive is firmly seated into the assembly. Remote Bung Cap Removal Tool Assembly Instructions 1. Pull collar down on ratchet to fully extend. 2. Insert ratchet head and neck under spring. 3. Align ratchet drive with female insert on assembly, and

More information

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02 ! Read this first Zetasizer nano series Self installation and Quick start guide I N S T R U M E N T S MRK825-02 Zetasizer Nano series Self installation and Quick start guide MAN0383 Issue 1.1 July 2007

More information

Equipment Operating Procedure Glove Box

Equipment Operating Procedure Glove Box Equipment Operating Procedure Glove Box Page 1 0.0 Changing the Compressed Gas Cylinder 1. Complete Compressed Gas Cylinder training from EHS website before manually exchanging gas cylinders. In order

More information