1 CLARIFICATIONS AND RESPONSES IN THE PRE BID MEETING HELD ON 19 TH AUGUST 2013 IN OUR COMMITTEE ROOM FOR THE PROCUREMENT OF DIOXIN ANALYSER (GC-MS/MS) File No : PUR/IMP/178/14 Description of Item : DIOXIN ANALYSER (GC-MS/MS) M/s. Thermo fisher Scientific SL. No NIIST Point No 1 Point No 8 2 Point No 13 3 Point No 23 NIIST Programmable Inlet upto 10 Ramps Autoinjector / Sampler to handle minimum 15 samples at one time. Mass filter: Heated maintenance free quadrapole rods & fast quadrapole scanning Amendment required by Thermo fisher scientific Programmable temperature vaporizer inlet up to 3 ramps Autoinjector / Sampler to handle minimum 8 samples at one time. Mass filter Unheated /Efficient maintenance free Quadrupole rods with pre filter and fast quadrapole scanning Amendments if any Accepted Accepted Adequate mechanism to avoid contamination of quadrapole and should be demonstrated during training to the personnel and before commissioning.
2 II. M/s. Agilent Technologies Sensitivity Please note the correct method to evaluate the sensitivity of a Mass spectrometer is by the detection limit & not by Signal-to-noise ratio. (Refer to attached technical note about the same). Due to this reason Agilent does not stipulate S/N ratio as a parameter for sensitivity estimation of mass spectrometers. S/N ratio can be easily manipulated. Also we request to change the detection limit to 1 fg or lesser for the reasons below: Dioxin and furans analysis requires an ultrasensitive equipment in order to quantify Pico- gram level analyte in complex matrix. a. EU keep reducing MRL for dioxin and furans, refer attached two recent regulations on food which require ultra low sensitivity of about 0.5 fg : Refer find attached slide and compare the MRL had reduced in 2006 and 2013 by EU In all food matrices MRL reduced by 3 times. For environment very soon the MRL is expected to be reduced substantially so you certainly need an ultra sensitive mass spectrometer. b. Refer our application note(attached) Where Feed with lesser MRL Comparison of HRMS versus GCMS/MS was done at European reference lab for Dioxin as a collaborative project by Agilent. The agreement between the results obtained on the GC-HRMS and 7000 GC- MS/MS for food and feed samples at levels between 0.5 pg TEQ/g were within the range of +/- 10% +/_ 20%. Only those animal feed stuff samples with results of pg TEQ/g ( below EU action level) of 0.5 pg TEQ/g gave results > 20% between HRMS and GCMSMS equipment. Even for feed analysis deviations lesser action level would be minimized by higher sensitive equipment like 7010 Agilent GC-MS/MS. To summarize, We need higher sensitive equipment in order to achieve detection limit of 0.5 fg and considering future reduced MRL by EU & other regulations.
3 Reply: Instrument detection limit is accepted as a better method to evaluate sensitivity than signal to noise ratio. 3fg or less of OFN derived at the 99% confidence level from area precision of eight sequential injections of 1 ul. For eight sequential injection of 1 ul of 3 fg OFN standard, the signal must be 10 times higher than the rms noise value. M/S. SHIMADZU SL. No NIIST Point No 1 Point No 8 2 Point No 13 3 Point No 17 4 Point No. 23 NIIST Programmable Inlet upto 10 Ramps & 450 degree celcius heating Autoinjector / Sampler to handle minimum 15 samples at one time. Mode of Operation: Electron ionization Mass filter: Heated maintenance free quadrapole rods & fast quadrapole scanning Amendment required by M/s. Shimadzu Programmable temperature vaporizer inlet up to 3 ramps or better & 450 degree celcius heating Autoinjector / Sampler to handle minimum 12 samples or more at one time. Mode of operation : EI (Electron Ionization) with CI (Chemical Ionization) with Ion Gauze for confirmation Mass Filter : Prefilter or Pre-rods maintenance free quadrupole rods Amendments if any Accepted Autoinjector / Sampler to handle minimum 8 samples at one time will remain unchanged. As per European Union regulations chemical ionization mode is not recommended. Adequate mechanism to avoid contamination of quadrapole and should be demonstrated during training to the personnel and before
4 5 Point 27 Collision cell : For the effective bombardment of precursor ions using quantified collision energies and effective MRM Collision cell : For the effective bombardment of precursor ions using Argon gas and effective MRM commissioning. will remain unchanged 6 Point 28 Software for data handling and reporting Sensitivity Software for data handling and reporting with smart MRM database to create optimal MRM methods of different matrices automatically 7 Pont 30 EI MRM OFN 100 fg EI MRM m/z OFN 100 fg S/N > m/z :1 272 or -222 S/N > higher. 3fg or less 8,000: OFN 1 or higher. derived at the IDL 99% 4fg or less OFN confidence level from derived area at the 99% precision of eight sequential confidence level from injections of 1 ul. area precision of eight (n = 8) sequential injections of 1 ul will remain unchanged - We require an appropriate software to provide us the required data specified in the tender. IDL of 3fg or less OFN derived at the 99% confidence level from area precision of eight sequential injections of 1 ul. For eight sequential injection of 1 ul of 3 fg OFN standard, the signal must be 10 times higher than the rms noise value. Dioxin analysis requires lower detection limits especially in the
5 8 Point fg on-column GC injection of OFN should give IDL of less than 3.0 fg 100 fg on-column GC injection of OFN should give IDL of less than 4.0 fg case of vegetable oils and fats. IDL should be 3fg or less for such samples. S/N ratio is removed as IDL is more practical. Rejected (See reply to Point 30) 9 Point 32 For ten sequential injections For ten sequential %RSD should be <5% injections of OFN standards %RSD should be <5% 10 Point 37 Detection limit of 3 fg or less as specified in No. 30 is mandatory Detection limit of 4 fg or less as specified in No. 30 is mandatory For ten sequential injections of dioxin standards %RSD should be <5% Rejected (See reply to point 30) 1. The Title of the indent should be modified as Triple Quadrupole based Gas Chromatograph- Mass Spectrophotometer for Screening & confirmation of Dioxins with EI in environmental matrices & with CI for additional capability of handling Food, Feed, Herbs & Spices as per EPA & EU. Since as discussed, no Global reputed Vendors make any dedicated Dioxin Analyzer based on TQ-GCMS as such nor we do manufacture any dedicated DIOXIN ANALYZER in Shimadzu facility in Japan. Reply: The equipment is dedicated for dioxin analysis and hence the end user has every right to name it as Dioxin Analyzer. In addition
6 it is clearly mentioned in the detailed specificiations that Dioxin analyzer should be Triple Quadrupole based Gas Chromatograph- Mass Spectrophotometer. As per European Union regulations electron ionization (EI) method is only being recommended. Hence CI mode is not required. 2. All of our Pre-configured TQ-GC-MS system comes (as per the purchase order) after thorough, stringent Quality Checking without fail before dispatched to any countries along with certifications with standards like OFN (EI) & NCI & BZN for PCI. For additional testing at factory for Dioxin analysis report of the same model can be submitted from our Q.C. Lab or would be done and possible at your installation site only. Reply: End user has every right to scrutinize the efficiency of the equipment before shipment. 3. The system set up and validation would be done and possible of our offered model at installation site in presence of the end user at an additional charges apart from the cost of the instrument. To do it at other site or any deviation to do it would unnecessarily attract additional charges at consignee s account. Reply: Instrument training and compliance justification at vendor s facility is one of the requirements in the tender. The end user should be ready to bear the charges for 10 working days training for the personnel at factory site, charges if any for the training should be quoted separately. Additional Amendments/ Additions by the indentor Point No 5: Amendment: Necessary columns for dioxin analysis has to be provided Three columns for dioxin analysis should be quoted. Point No 11: Split/ splitless (Capillary) injector. Split ratio upto 7500:1 to
7 avoid column overload Amendment: Split/ splitless (Capillary) injector. Split ratio upto 7500:1 or better to avoid column overload. Point number 20: Ion source temperature: 150 to 350 Degrees Amendment: Ion source temperature: 150 to 350 Degrees Celsius Point number 43: 3 year comprehensive warranty for the complete system and free maintenance after warranty period for three years. Amendment: 3 year comprehensive warranty for the complete system and Comprehensive maintenance contract (CMC) for next three years including consumables except gases during warranty period as well in CMC period. Please quote for the additional warranty asked for. Quote separately for CMC for next 3 years. Additionally please quote for Annual Maintenance contract (Labour Only) for 3 years after the 3 years of comprehensive warranty. Additions Addition 1: Addition 2 : Addition 3: Our personnel should be provided minimum of 10 working days training for instrument handling and analysis at the factory laboratory facility. Charges if any should be quoted separately. Additional consumables to be supplied a. syringes 5 Nos b. Liner Split/ splitless, PTV liner 20 Nos each c. O- rings 20 Nos d. Septa 20 Nos e. Screw capped vials 20 Nos f. Gold Seal 10 Nos g. Nuts and ferrules suitable for columns h. vacuum pump oil 3 litres Operation qualification/ Performance Verification Services (OQPV) should be done every year during warranty period and CMC.
8 The tender stands amended to the extent as given above. The due date for the tender is extended up to 5 th June HRS (IST) and tenders would be opened at HRS(IST) on the same day. Bidders who are interested to participate in the tender opening can be present by HRS (IST) on the same day. Stores & Purchase Officer
Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople
More informationThe instrument should have a mass range from 2 to 2000 m/z or better Interface
Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for
More informationUser guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection
User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind
More informationComparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS
TECHNICAL NOTE 1598 Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS Authors Amit Gujar, 1 Tim Anderson, 1 Daniela
More informationExtend the time window for the detection of low level anabolic androgenic steroids and their metabolites
Application Note Doping Control Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites Steroid profiling, screening, and confirmatory analysis using the
More informationAgilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS
Agilent 7800 ICP-MS Specifications and Typical Performance Fast track metals analysis with Solution-Ready ICP-MS The Solution-Ready Agilent 7800 Quadrupole ICP-MS combines proven, robust hardware, auto-optimization
More information6890 GC Site Preparation
6890 GC Site Preparation (a16011) This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions
More informationRapid and Reliable Detection of Dissolved Gases in Water
Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,
More informationDETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY
DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY Nengbing Xu, Hongmei Ying and Libo Zhu Ningbo Environmental Monitoring
More informationQuick Reference Guide. OPTIC 3 - PTV on Shimadzu GC-2010 Gas Chromatograph
Quick Reference Guide OPTIC 3 - PTV on Shimadzu GC-2010 Gas Chromatograph ATAS GL International B.V. De Run 4441 5503 LS Veldhoven The Netherlands Tel: +31 40 254 9531 Fax: +31 40 254 9779 e-mail : info@atasgl.com
More informationSite Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)
Purpose of Procedure Your site must meet this specification or set of requirements to assure a successful and timely installation of your 7820MSD Series Mass Selective Detector (MSD). This checklist is
More informationRetention Time Locking: Concepts and Applications. Application
Retention Time Locking: Concepts and Applications Application Gas Chromatography December 1997 Authors Vince Giarrocco Bruce Quimby Matthew Klee Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More informationCustomer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationSimplified Backflush Using Agilent 6890 GC Post Run Command Application Note
Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Gas Chromatography Author Matthew S. Klee Agilent Technologies 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.
Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography
More informationAutomated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:
Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion
More informationTechnical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)
Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent 6890 GC interfaced
More informationJaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation
Using Nitrogen instead of Helium as carrier gas, while maintaining exact the same separation efficiency, retention times, peak elution order, without changing the oven temperature conditions. Jaap de Zeeuw
More informationextraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.
Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s):
More informationApplications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples
Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Outline of Presentation Introduction Process Gas Analysis Magnetic Sector Mass Spectrometer Standard
More informationFINAL REVISED TECHNICAL SPECIFICATION OF INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER (ICP-MS) WILL BE AS FOLLOWS:- GENERAL SPECIFICATIONS
Ref No. INST/12(121)/2018-Pur Date: 31/12/2018 CORRIGENDUM Reference to NIT no. INST/12(121)/2018-Pur published in national newspapers on 07/09/2018 for purchase of equipment: ICP-MS WITH ACCESSORIES.
More informationno peaks missing peaks too small peaks increasing retention decreasing retention
GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering
More informationAPPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT
APPLICATION NOTE Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT The INFICON 3000 Micro GC provides two fast and accurate solutions for the analysis of coal mine gas components.
More informationSite Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A
Purpose of Procedure Your site must meet this specification or set of requirements to assure a successful and timely installation of your 5975C Series Mass Selective Detector (MSD). This checklist is designed
More informationVarian 240-MS GC/MS Ion Trap Mass Spectrometer Pre-installation Instructions
Varian 240-MS GC/MS Ion Trap Mass Spectrometer Pre-installation Instructions Checklist NOTE: Do not unpack the boxes. Place a check in the box after satisfying each requirement as described in the instructions.
More informationGeneral Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!
The portable Gas Chromatograph is now more versatile than ever. DPS Instruments is pleased to present the newest Portable Companion 2 Gas Chromatograph with room for 2 Detectors. The Companion 2 GC was
More informationMajor Design Features
Major Design Features 3Flex Features Design Benefits Discussion 316 SS fittings and pneumatically actuated, hard-seal valves Provides virtually leak-free gas management with the lowest outgassing rate
More informationASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist
ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking
More informationOMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative
More informationInstant Connect Helium Saver Injector Module
Thermo Scientific Instant Connect Helium Saver Injector Module for TRACE 1300/1310 Gas Chromatographs User Guide PN 31709737 Revision B January 2016 2016 Thermo Fisher Scientific Inc. All rights reserved.
More informationPre-installation Checklist
Pre-installation Checklist Varian 1200/1200L Mass Spectrometer Use this checklist as a guide while preparing your site for the installation of the Varian 1200 or 1200L MS system. Make sure that your site
More informationEnvironmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!
After years of development and testing, DPS Instruments is pleased to present the newest, most expandable and versatile Gas Chromatography Systems in history. The DPS 600 Series GC systems are the world
More informationGas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems
Inorganic Analysis Chromatography Molecular Spectroscopy Thermal/Elemental Analysis Informatics Gas Chromatography MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass
More informationReal-time PCB Gas Monitoring System (MOHMS-21GP)
Real-time PCB Gas Monitoring System (MOHMS-21GP) SHINSAKU DOBASHI* 1 YOSHIHIRO DEGUCHI* 1 MASAHARU KIRA* 2 RYUICHIRO TANAKA* 3 YOSHINORI IZAWA* 3 This report describes the technical characteristics and
More informationNIT LIMITED NO. NC-185-SCS-P-NOCB ARGON & NITROGEN CYLINDER WITH REGULATOR
NIT LIMITED NO. NC-185-SCS-P-NOCB-15-16 ARGON & NITROGEN CYLINDER WITH REGULATOR NATIONAL INSTITUTE OF SCIENCE EDUCATION & RESEARCH JATNI CAMPUS, P.O.: BHIMPUR-PADANPUR, VIA:-JATNI Khurdha-752050 Tender
More informationGhost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager
Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Outline of Topics What is a ghost peak? How to identify ghost peaks and their source How to correct
More informationThermo Scientific TRACE 1300 Series Gas Chromatograph. Evolve into GC 2.0. powerful breakthroughs
Thermo Scientific TRACE 1300 Series Gas Chromatograph Evolve into GC 2.0 powerful breakthroughs Transcend the traditional and step into modern gas chromatography Unique and innovative breakthroughs featured
More informationUsing the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer
application Note Gas Chromatography/ Mass Spectrometry Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer Overview A mass spectrometer
More informationHiden QIC Series Gas Analysers Real-time gas analysis systems
Hiden QIC Series Gas Analysers Real-time gas analysis systems vacuum analysis surface science plasma diagnostics gas analysis Application Focussed Options Multistream valves 20, 40 and 80 way versions
More informationLCMSMS Solutions For Steroid Analysis
LCMSMS Solutions For Steroid Analysis Dan Blake Manager, Clinical Applications For Research Use Only. Not for use in diagnostic procedures. RUO-MKT-18-1166-A Agenda Special considerations for steroid analysis
More informationImprove Process Reliability
Convenient Conductivity Calibration to Meet Pharmacopeia Regulations Pharmacopeias require that conductivity measurement systems be periodically calibrated to ensure their precision. The range of accessories
More informationCommercial Practice Test Method Internal Vapor Analysis of Hermetic Devices
Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The
More informationANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION
ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION Standard OQ Test Suite This document describes the test program for qualifying LC analytical-scale Agilent systems; the following table lists
More informationCustomer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationSupporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial
Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle,
More informationOperation of the Perkin Elmer TGA-GC/MS
Operation of the Perkin Elmer TGA-GC/MS Summary of the TGA-GC/MS: The TGA-GC/MS allows a user to decompose a sample by heating, measure its loss of mass, and simultaneously analyze the chemical composition
More informationFor additional information about our solutions, please visit our web site at US/Pages/HomePage.
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationAGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY
AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY Agilent Vacuum Products: Our Committment to Quality Agilent s commitment to quality and Measurement s philosophy Reference Standards (PNEUROP, ISO)
More informationNIT LIMITED NO. NC-105-P-SCS-NOCB CYLINDERS & REGULATORS
NIT LIMITED NO. CYLINDERS & REGULATORS NATIONAL INSTITUTE OF SCIENCE EDUCATION & RESEARCH JATNI CAMPUS, P.O.: JATNI, Khurdha-752050 Tender Notice No. NIT Limited NO. Sealed Tenders are invited on behalf
More informationToxicology Gas Chromatography-Mass Spectrometry (GC-MS)
Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent Gas Chromatograph interfaced to the Agilent
More informationHigh Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module
TECHNICAL NOTE High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module TN42256 Dr. Liliana Krotz, Dr. Francesco Leone, Walter Galotta and Dr. Guido Giazzi
More informationMass Spec will not Autotune
Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems
More information20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as
Instruments and Equipment Introduction 5.5. Authorization to use technical equipment 5.5.3 Use of instruments and equipment Instruction and maintenance manuals 5.4. 5.5.3 Use of nonlaboratory equipment
More informationAgilent G3185B QuickSwap Accessory
Agilent G3185B QuickSwap Accessory Reference Manual Agilent Technologies Notices Agilent Technologies, Inc. 2005 No part of this manual may be reproduced in any form or by any means (including electronic
More informationIndian Institute of Technology Kanpur Samtel Centre for Display Technologies
Fax: + 91-0512-2596620 Phones: + 91-0512-2596622,6088 Indian Institute of Technology Kanpur Samtel Centre for Display Technologies Enquiry number: SCDT/FlexE/2016-17/02 Date:05/05/2016 Sealed Quotations
More informationUSER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve
USER'S GUIDE Nanoflow Probe 3-4mm Red Stripe ptfe Sleeve To Injector ZU1XC Fused Silica 20µm 90µm TTF115 ptfe Sleeve Fused Silica 25µm 375µm Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: +44
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 9 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 3.1 Canister Receipt 3.2 Canister Storage 4.0 INTERFERENCE AND POTENTIAL PROBLEMS
More informationAgilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationNow it is. easy to switch. CHS TM Steroid Profiling kit and software. Brochure not for distribution in the USA and Canada
Now it is easy to switch to Clinical MSMS CHS TM Steroid Profiling kit and software Brochure not for distribution in the USA and Canada 2 Meeting your needs for steroid assays PerkinElmer s new CHS MSMS
More informationThermo Scientific Chromatography Well Plates
Thermo Scientific Chromatography Well Plates High-throughput sample handling solutions Well Plates for Chromatography A full selection of high quality sample well plates for your UHPLC, LC and GC applications.
More informationThermo Scientific Chromatography Well Plates
Thermo Scientific Chromatography Well Plates High-throughput sample handling solutions Well Plates for Chromatography A full selection of high quality sample well plates for your UHPLC, LC and GC applications.
More informationGAS FILLING MACHINES PROCESS MONITORING PRESSURE BALANCE
GAS FILLING MACHINES PROCESS MONITORING PRESSURE BALANCE www.helantec.de Gas filling - visibly better! 2 Managing partners Jochen Weiß and Gregor Witrofsky Advantage for you Worldwide, HELANTEC is the
More informationADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM. OPENING DATE & TIME: January 3, 2018 at 3:00PM
Procurement Services ADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM ITB NUMBER: 1708MC OPENING DATE & TIME: January 3, 2018 at 3:00PM ITB TITLE: X-Ray Photoelectron Spectroscopy System ADDENDUM
More informationSimultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS
Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS UCT Part Numbers: CUQAX22Z Clean-Up C8+QAX, 2mg/1mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQID21-3UM
More informationBest Practice for Identifying Leaks in GC and GC/MS Systems
Best Practice for Identifying Leaks in GC and GC/MS Systems Technical Overview Maintaining a leak-free GC or GC/MS system is critical for obtaining optimal system performance with reliable, reproducible,
More informationProtocol for the verification tests on ethylene oxide treatment technology
Pilot project for the environmental technology verification In the field of ethylene oxide treatment technology Protocol for the verification tests on ethylene oxide treatment technology Sept. 11, 2003
More informationZebron Gas Management
Zebron www.phenomenex.com/gasmanagement What are the key benefits of Zebron filters and traps? The high adsorption capacity of the Zebron Filters removes contaminants from mobile phase gases, reducing
More informationNIT LIMITED NO. RD-NC-40-SPS-PH
NIT LIMITED NO. RD-NC-40-SPS-PH1302-14-15 ARGON GAS & CO2 with CYLINDER NATIONAL INSTITUTE OF SCIENCE EDUCATION & RESEARCH INSTITUTE OF PHYSICS CAMPUS, P.O.: SAINIK SCHOOL, BHUBANESWAR-751 005 Tender Notice
More informationOMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.
OMNISTAR / THERMOSTAR The efficient solution for gas analysis. Intelligent software. Wide range of applications. OMNISTAR /THERMOSTAR The efficient solution for gas analysis. Intelligent software. Wide
More informationThe Helium Leak Detector
The Helium Leak Detector Helium Leak Detector Main Components The main components of a helium leak detector are: 1. The analyzed, which enables to separate the tracer gas from other gases inside leak detector.
More informationExperiment GC : Analysis of BTEX by GC-FID
46 Experiment GC : Analysis of BTEX by GC-FID Learning Goals: Familiarity with gas chromatography Gain experience in temperature programming and method development Correctly use an internal standard, and
More informationAgilent G Rebuilt Electron Capture Detector
Agilent G1533-90790 Rebuilt Electron Capture Detector User Information Agilent Technologies Notices Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any form or by any means
More informationLow-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI
Low-Pressure Retention Locking with the 789A GC Application HPI Authors Courtney Milner and Russell Kinghorn BST International 41 Greenaway Street Bulleen, VIC 315 Australia Matthew S. Klee Agilent Technologies,
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCE AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More informationTitle: INSTRUMENT AND EQUIPMENT DOCUMENTATION AND RECORDS
Page 1 of 7 Sections Included in this Document and Document History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References/(Deleted date) 6. Procedure 7. Definitions 8. Records 9. Supporting
More informationAWS C2.21M/C2.21:2003 An American National Standard. Specification for Thermal Spray Equipment Acceptance Inspection
AWS C2.21M/C2.21:2003 An American National Standard Specification for Thermal Spray Equipment Acceptance Inspection Key Words Thermal spray equipment, flame arc, plasma, HVOF, acceptance inspection AWS
More informationDate of Shipment: Xxxxx 00, 20xx 3402c National Institute of Advanced Industrial Science and Technology National Metrology Institute of
Date of Shipment: Xxxxx 00, 20xx 3402c01151112-170831 National Metrology Institute of Japan Reference Material Certificate NMIJ CRM 3402-c01 Sulfur Dioxide This certified reference material (CRM) was produced
More informationGo/no-go test for the acceptance of components for unbaked vacuum sectors
Go/no-go test for the acceptance of components for unbaked vacuum sectors Paolo Chiggiato (Thanks to Sophie Meunier for the preparation of the presentation) Introduction CERN policy: All vacuum components
More informationThe Tekran Series 2600-IVS Difference!
Tel:+1-416-449-3084 Fax: +1-416-449-9298 The Tekran Series 2600-IVS Difference! Rev 021916 Method 1631 (Dual Gold Trap CVAFS) Method 245.7 (Direct CVAFS Analysis) US EPA IO-5 (Ambient Air) ASTM D-6350
More informationConsiderations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port
Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port So, Why Do I Have To Do Maintenance? 1. Things wear out septa, syringes, nuts, ferrules, o-rings, etc.
More informationBEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION
BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the
More informationDuring today's webinar we're going to look at the following topics:
Alternative Carrier Gas for GC and GC-MS Presentation Transcript Hello and welcome to the Peak scientific webinar which today will concern alternative carrier gas for GC and GC-MS. My name is Ed Connor,
More informationGases and welding distributors continue
Specialty Gas Instrumentation By Ray Borzio, Mike Lee, and Scott Varian Gases and welding distributors continue to have a significant future in the production, delivery, and related services for specialty
More informationSuperior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results
Nitrogen Generator Whisper-N2 Series Features Membrane Technology Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results
More informationAir Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions
Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions Introduction: There is a bit of explanation of the health effects of benzene in the Introduction section of the first part
More informationA Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction
A Column-Flow Independent Configuration for QuickSwap Application Authors Matthew S. Klee and Bruce Quimby Agilent Technologies, Inc. 80 Centerville Road Wilmington, DE 08 USA Abstract As can be seen from
More informationIntegration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform
POCON3E Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LCMS/MS platform ASMS 2 ThP2 Tetsuo Tanigawa, Toshikazu Minohata Shimadzu Corporation, Kyoto, Japan analysis in single
More informationAddressing the World Shortage of Helium
Addressing the World Shortage of Helium Introducing the Programmable Helium Conservation Module GPD Solutions June 2013 Market Situation Unreliable supply of helium worldwide and increasing prices have
More informationA New GC/MS System Designed for Helium Carrier Gas Conservation
A New GC/MS System Designed for Helium Carrier Gas Conservation Approaches for the Smarter Use of Helium Dale R. Walker GCMS Applications Specialist August, 7, 2013 1 Helium Facts How it's produced Where
More informationElimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU.
Elimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU. Control Analytics, Inc 6017 Enterprise Dr Export, PA 15632 (724) 387 CEMS (2367)
More information1)! DO NOT PROCEED BEYOND THIS MARK
Operating Instructions for X-ray Photoelectron Spectrometer: Physical Electronics Model 555 XPS/AES (John H. Thomas, III, Ph.D., Electron Spectroscopy) Sample Insertion: figure 1. Sample insertion rod
More informationFlare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1
Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC
More informationms NoVent II Instruction Booklet
ms NoVent II Instruction Booklet ms NoVent II The ms NoVent II interface is designed to allow column changing or column maintenance to be carried out without the need to shut down the mass spectrometer.
More information3 Flow and Pressure Control
3 Hydrogen shutdown Column shutdown Turning gas flows on and off EPC-controlled streams NonEPC-controlled streams Electronic Pneumatic Control (EPC) Interpreting flow and pressure readings Configuration
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 5.1 Subatmospheric
More informationResidual Gas Analysis Systems for Industry
HIDEN RC SYSTEMS QUADRUPOLE MASS SPECTROMETERS FOR RGA, GAS ANALYSIS AND PROCESS MONITORING The HAL RC systems are designed for RGA, gas analysis and process monitoring applications including leak detection,
More informationDOS in Food Science & Nutrition, Manasagangotri, Mysore
Sl No UNIVERSITY OF MYSORE DOS in Food Manasagangotri, Mysore 570 006. TENDERS FOR THE SUPPLY OF EQUIPMENTS, NETWORKING & COMPUTATIONAL FACILITIES FOR UNIVERSITY OF MYSORE (UNDER UGC XI SCHEME) INVITATION
More informationUNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008
UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method
More informationTrace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS
Abstract Highly sensitive analysis of metanephrines in plasma is critical in the diagnosis and treatment of pheochromocytoma and paraganglioma. Here, a HILIC LC-MS/MS method was developed using a Raptor
More information