OXYGEN TRANSFER CAPABILITIES CANADIANPOND.CA PRODUCTS LTD.

Size: px
Start display at page:

Download "OXYGEN TRANSFER CAPABILITIES CANADIANPOND.CA PRODUCTS LTD."

Transcription

1 EVALUATION OF THE OXYGEN TRANSFER CAPABILITIES OF CANADIANPOND.CA PRODUCTS LTD. ¾ BUBBLE TUBING AERATION SYSTEM ½ BUBBLE TUBING AERATION SYSTEM OctoAir-10 AERATION SYSTEM BY GSEE, INC. LA VERGNE, TN. SEPTEMBER, 2011

2 TABLE OF CONTENTS 1. INTRODUCTION DESCRIPTION OF THE AERATION TEST BASIN Horsepower TEST PROCEDURES DATA ANALYSIS METHODS Standard ASCE Data Analysis Method DISCUSSION OF RESULTS CONCLUSIONS CERTIFICATION APPENDIX Regression Analysis Printouts...24 TABLE OF FIGURES Figure 2-1 GSEE Test Facility... 3 Figure 2 - Bubble Tubing Arrangement... 6 Figure 5-1 SCFM/100 of tubing v. K La 20 - ½ Bubble Tubing Figure 5-2 SCFM/100 of tubing v. SOTR/100 of tubing - ½ Bubble Tubing Figure 5-3 SCFM/100 of tubing v. SOTE - ½ Bubble Tubing Figure 5-4 SCFM/100 of tubing v. SOTE/Ft. of Submergence - ½ Bubble Tubing Figure 5-5 SCFM/100 of tubing v. SOTE - ½ Bubble Tubing Figure 5-6 SCFM/100 of tubing v. SAE - ½ Bubble Tubing Figure 5-7 SCFM/100 of tubing v. K La 20 - ¾ Bubble Tubing Figure 5-8 SCFM/100 of tubing v. SOTR/100 of tubing - ¾ Bubble Tubing Figure 5-9 SCFM/100 of tubing v. SOTE - ¾ Bubble Tubing Figure 5-10 SCFM/100 of tubing v. SOTE/Ft. of Submergence - ¾ Bubble Tubing Figure 5-11 SCFM/100 of tubing v. SOTE - ¾ Bubble Tubing Figure 5-12 SCFM/100 of tubing v. SAE - ¾ Bubble Tubing Figure 5-13 SCFM/100 of tubing v. K La 20 - OctoAir Figure 5-14 SCFM/100 of tubing v. SOTR/100 of tubing - OctoAir Figure 5-15 SCFM/100 of tubing v. SOTE - OctoAir Figure 5-16 SCFM/100 of tubing v. SOTE/Ft. of Submergence - OctoAir Figure 5-17 SCFM/100 of tubing v. SOTE - OctoAir Figure 5-18 SCFM/100 of tubing v. SAE - OctoAir Figure 5-19 SCFM/100 of tubing v. SOTE/Ft. of Submergence All Tubing WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 1

3 LIST OF TABLES Table 5-1 Summary of ½ Bubble Tubing Test Results Table 5-2 Summary of ¾ Bubble Tubing Test Results Table 5-3 Summary of OctoAir-10 Test Results WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 2

4 1. INTRODUCTION During June-August, 2011, CanadianPond.ca Products Ltd. retained GSEE, Inc., to perform unsteady state clean water shop oxygen transfer tests on the CanadianPond.ca Products Ltd. ½ & ¾ Bubble Tubing and the OctoAir-10 aeration systems. Tests were performed on each system at airflow rates of 2.5, 5 and 10 SCFM at water depths of 5, 10 and 15 feet. Oxygen transfer is determined using the ASCE clean water non-steady state test procedures. The ASCE standard requires a regression analysis on the data from each sample location and then averages the obtained results. Test results are reported at standard conditions of 20 C liquid temperature, one (1) atmosphere barometric pressure, zero (0) dissolved oxygen, and alpha () and beta (ß) equal to 1.0 (clean tap water). All test results are calculated using the ASCE non-linear regression analysis method for the determination of the mass transfer coefficient K La T, the steady-state D.O. saturation value C *, and the D.O. concentration at time zero C 0. The final results are corrected to a standard TDS value of 1,000 mg/l. 2. DESCRIPTION OF THE AERATION TEST BASIN All testing occurred in the 21' diameter x 31' tall GSEE, Inc. test basin located in LaVergne, TN. as shown in Figure '. MOVEABLE. ELECTRIC HOIST. GUARD. 26' BRIDGE. SUPPORTS. 31'. LADDER. SUPPORTS FOR VARIABLE. HEIGHT BRIDGE STRUCTURE. PLATFORM. 6 Flange Penetration JIB CRANE. 599 Waldron Rd. LaVergne, TN Aeration Test Facility 21' Diameter Tank 30' Max Liquid Depth STEEL GRID FOR ANCHORING AERATION PIPING LADDER. GROUND LEVEL. 4'. 6 Flange Penetration 4" ANR 73 ANNUBAR. ORIFICE PLATE. TEMPERATURE & PRESSURE GAGES. Blower SCFM 40 HP. Blower SCFM 40 HP. Pulsation. Arrestor. Access. Hatch. 4" AIR HEADER. Window. Sulfite. Dissolution. Tank. NOTE: 4" Ø FLEXIBLE HOSE AVAILABLE TO CONNECT AIR HEADER TO AERATION GRID PIPING. SCALE - FEET. DRAIN. Figure 2-1 GSEE Test Facility The air source is a 1 HP blower. A bypass valve is used to obtain airflow rates from 0 to 10 SCFM. Exact measurement of the airflow was determined using a Dwyer 0-10 SCFM rotometer installed in the header pipe. A Dwyer digital manometer is used to monitor the flowing air line pressure. A temperature probe monitors line temperature during each test. A mercury barometer is used to monitor local atmospheric pressure. Airflow rate is defined as follows: SCFM = ELEVATION VIEW Standard Cubic Feet per Minute (14.7 PSIA, 68ºF, 36% RH and a density of Lb/Ft³) 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 3

5 The air flow rate observed using the Dwyer flowmeter must be corrected to standard conditions using the following equations 1 : To correct for temperature and pressure: ACFM 14.7 BP LP 460 LT 528 LT = Line Temperature, F LP = Line Pressure, PSIG BP = Site Barometric Pressure, PSIA Flowmeter Reading Eq. 2-1 The air flow rate in ACFM is then corrected for relative humidity using Eq. A-1 from the ASCE Standard: Q S Rh Pva BP LP1 BP 36.2 ACFM Eq LT where: Rh = Relative Humidity, % Pva = Vapor Pressure of Water at ambient temperature Q S = Humidity corrected air flow rate, SCFM SCFM = Standard Cubic Feet per Minute ( PSIA, 68ºF, 36% RH and a density of Lb/Ft³) The observed airflow rate may be converted to SI units as follows: Nm 3 SCFM hr Eq Nm³/h = Normal Cubic Meters per Hour (1013 hpa, 0ºC, 0% Rh) 2.1. Horsepower The blower HP motor can be determined as follows: HP motor GHP MHP Eq. 2-3 GHP = Gas Horsepower (From adiabatic compression, shown below) MHP = Mechanical Horsepower (Loss in gear reducers, etc.) The basic equation for determining the gas horsepower is: 1 Spink, L.K. "Principles and Practice of FLOW METER ENGINEERING" Ninth Edition, 1975 The FOXBORO COMPANY, Foxboro, Mass. 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 4

6 W a H GHP a 33,000 W a = Weight flow of gas, Lbs./Min H a = Adiabatic Head, Ft-Lb f/lb m o = Blower Overall Efficiency, % 33,000 = Units Conversion Factor, Ft-Lb f/min/hp The Weight Flow of gas (W a) can be determined as follows: o Eq. 2-4 W a ACFM Eq. 2-5 ACFM = Actual flow of gas at discharge conditions = Density of flowing gas at discharge conditions = MW m BP LP LT 460 MW m = Molecular Weight of Moist flowing gas = MW d G MW d = Molecular Weight of dry gas G = Specific Gravity of moist gas The Adiabatic Head (H a) is calculated as follows: H a = Rh Pv BP BP LP R Ta BP k 1 k R = Gas Constant, Ft-Lb f/lb m ºR = 1,545 MW m k = Ratio of specific heats, C p/c v a k 1 k 1 Eq. 2-6 Four (4) Yellow Springs Instruments (YSI) D.O. probes are installed in the test basin to monitor the dissolved oxygen concentrations during each test. A YSI Model 556 MPS multi parameter meter and probe are used to monitor water quality (Temp, TDS, D.O., and Barometric Pressure) during each test. The following sketch shows the arrangement used for the ½ & ¾ Bubble Tubing in the GSEE, Inc. test basin. For both sizes of Bubble Tubing 100-feet of tubing was installed in the GSEE test basin. The OctoAir-10 was installed centered in the GSEE test basin. The OctoAir-10 consists of 100-feet of ½ tubing installed in a compact spiral configuration on an easily retrievable stainless steel frame. 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 5

7 Blank Diffuser 100 of Tubing Installed Window 49.1 in Access Hatch Bridge Bar Spacing 1 starting 1 above floor 16.7 in in Figure 2 - Bubble Tubing Arrangement 12.8 in 3. TEST PROCEDURES Before testing, the aeration basin is cleaned and filled to the required test depth with potable water. Four (4) YSI dissolved oxygen meters and probes are placed in the test basin and later used to monitor the dissolved oxygen concentration during each test. The probes are located as follows: Overall test procedures include: Probe Probe Depth 1 2 Above Floor 2 2 Below Surface 3 Mid-Depth 4 Mid-Depth YSI MPS 556 Mid-Depth After installing the diffusers in the test basin, add water to a depth even with the top surface of the diffusers. Check diffusers to insure that the level of all is within ±¼". 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 6

8 After filling the test basin with tap water, add enough cobalt chloride catalyst to obtain a concentration of cobaltous ion less than 0.1 mg/l. Dissolve the catalyst into the basin contents by running the aeration system a minimum of thirty minutes before testing. Add enough ( % of stoichiometric) sodium sulfite to deoxygenate the tap water in the basin to start each test. Monitor the dissolved oxygen concentration as it depletes then starts to rise, using the in-situ dissolved oxygen probes. Record the water temperature using the YSI MPS 556 thermister. With the aeration system operating at the specified liquid depth and air flow rate, start monitoring as the oxygen concentration increases. Collect data to cover a range of dissolved oxygen concentrations from 1.0 mg/l to 98% of saturation, obtaining a minimum of 300 data points for each probe. I. The general test procedures are: II. 1. Thoroughly clean the aeration basin before testing and fill with tap water to the desired liquid depth. 2. Operate the aeration system in potable water at the test airflow rate and operating liquid depth for 30 minutes before testing to obtain temperature and mixing equilibrium. Record the liquid temperature a minimum of three times during each test run. Maintain the required airflow rate during testing by monitoring manometers connected across the airflow devices. Monitor operating air pressure and headloss via a mercury manometer. Measure the operating line temperature. 3. Install 4 dissolved oxygen probes with integral stirrers at locations in the test tank as required. 4. Use Cobalt Chloride (CoCl 2 * 6H 2O) as a catalyst at a concentration of 0.1 mg/l. 5. Use anhydrous sodium sulfite technical grade (Na 2SO 3) to deoxygenate the test liquid. Add sulfite solution before each test run to decrease the oxygen concentration to zero (0.50 mg/l or less D.O.) and maintained zero for 1 minute. 6. Use the azide modification of the Winkler method to calibrate the D.O. probes. Collect a minimum of three-hundred (300) D.O. observations for each D.O. probe between 10 and 98% of saturation. Detailed test procedures: A. Initial setup 1. Inspect aeration basin for adequate cleanliness, level of diffusers and correct water depth. 2. Check installation of airflow monitoring device. 3. Check D.O. probe thermisters for liquid temperature monitoring. 4. Prepare YSI (Yellow Springs Instruments) Dissolved Oxygen (D.O.) probes for installation. a) Replace electrolyte solution and membranes on each D.O. probe b) Connect probes to YSI D.O. meters c) Check each probe for functioning stirrer mechanism d) Connect all D.O. meters to computer for data logging 5. Check the placement of each D.O. probe in the test basin. 6. Start the blowers and begin aerating the test tank. 7. Check all air flow meters, gages, valves, and fittings on the air supply system for air leakage. 8. Collect at least two samples from the oxygen saturated aeration basin for analysis using the Winkler titration method to determine the D.O. concentration. 9. Calibrate all D.O. probes and meters to the saturation value determined by the Winkler method. 10. Check installation of the temperature gage in the aeration header piping system for the accurate determination of flowing air temperature. 11. Dissolve Cobalt Chloride into a container of water. 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 7

9 12. Pour Cobalt solution into the aeration basin. 13. Allow a minimum of thirty minutes mixing of the cobalt into the aeration basin before the start of testing. B. Procedure for clean water aeration testing. 1. Adjust the airflow rate to the test basin to the required test airflow. 2. Read and record the following data: a) Site barometric pressure (PSIA) b) Operating line pressure from the mercury manometer connected to the header system (PSIG) c) Operating line temperature ( F) d) h from the Annubar and Orifice, (Inches of H 2O) e) Liquid Temperature ( C) f) Aeration basin oxygen saturation value C so (mg/l) [Winkler Analysis] g) Ambient temperature h) Relative Humidity, % 3. Pump sodium sulfite slurry into the aeration test basin. 4. Begin observing D.O. meters. 5. Monitor D.O. on each of the YSI meters as it drops to 1.0 mg/l. 6. Continue recording D.O. values versus time for each of the D.O. probes, obtaining a minimum of 300 D.O. values for each probe. 7. Stop all recording of D.O. values when the aeration basin has reached 6/K La. 8. Collect a basin water sample and analyze for Total Dissolved Solids (TDS mg/l) 9. Perform linear regression analysis on the collected data. a) Determine K La 20 values for each probe b) Calculate SOTR and SOTE. 10. Repeat steps 1-9 for each test run. 4. DATA ANALYSIS METHODS 4.1. Standard ASCE Data Analysis Method The basic mass-transfer model used to determine oxygen transfer is as follows: dc dt K a C L * C Eq. 4-1 Which, upon integration, with initial condition C = C o at t =0, becomes: (logarithmic form) C ln C * * C K C o L a t Eq. 4-2 or (exponential form) * kla t C C e * C C Eq. 4-3 O C = D.O. Concentration, mg/l C * = Equilibrium D.O. concentration, the concentration obtained as time approaches infinity, mg/l 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 8

10 C 0 = D.O. concentration at time zero, mg/l K La = Apparent volumetric mass transfer coefficient, t -1 The overall mass transfer coefficient (K La T) is obtained experimentally by aerating deoxygenated water and observing the rate of change of dissolved oxygen (D.O.) concentration about time. A non-linear regression of D.O. about time is used to determine K La T, C *, and C 0. The logarithmic form of the mass transfer model can be rearranged to determine K La T using a logdeficit linear regression as follows: K L a T 60 t t 2 1 C ln C * * 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 9 C C 1 2 Eq. 4-4 K La T = Apparent volumetric mass transfer coefficient at test liquid temperature T, hr -1 C * = The observed saturation concentration of oxygen in the test basin at test temperature and barometric pressure at equilibrium, mg/l after an aeration period equal to 6/K La T C 1 and C 2= Dissolved oxygen concentration at time t 1 and t 2 respectively, mg/l For purposes of comparison, K La T must be corrected to standard temperature, 20 C. The appropriate correction has been found empirically to be: T = test liquid temperature ( C) = for all T 20T K La 20 K L a T Eq. 4-5 With the value of K La 20 known, it is possible to calculate the pounds of oxygen transferred to the test liquid at standard conditions of 20 C, maximum oxygen deficit (dissolved oxygen equal to zero), one atmosphere barometric pressure, and alpha and beta equal to 1.0 (clean tap water) for each sample point. SOTR i * K a C V Eq. 4-6 L 20i 20i SOTR i = pounds of oxygen transferred to the test liquid, lb. O 2 /hr., for Probe i V = Liquid volume of water in the test tank with aerators turned off C * 20i = * 1 C = Temperature correction factor, C * st/c * s20 C * s20 = mg/l, standard D.O. concentration at 20 C and one atmosphere C * st = oxygen saturation concentration from Standard Methods, mg/l, at test liquid temperature T = Pressure correction factor, P b/p s P b = Site barometric pressure, PSIA P s = Standard barometric pressure, PSIA The overall average value of SOTR is then calculated as the average of the individual SOTR i values determined for each sample point. Calculate the standard percent oxygen transfer (SOTE - %) once the oxygen transfer rate is known using the following equation:

11 SOTR 100 SOTE SCFM Eq. 4-7 The blower wire HP is determined using the adiabatic compression formula as described in Section 2 of this report. Calculate the standard aerator efficiency (SAE) using the following equation: SOTR SAE NO Eq. 4-8 HP No = aerator efficiency, lb. O 2/hr-Hp SOTR = Qo, standard oxygen transfer rate, lb. O 2/hr. HP wire = Blower HP determined by the adiabatic compression formula Finally, the reported values are corrected to a standard TDS concentration of 1,000 mg/l using the following: K L a T TDS Corrected K L a T observed e TDS TDS = Observed Total Dissolved Solids concentration for each run, mg/l All values are then recalculated based on the corrected K La T. Eq WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 10

12 5. DISCUSSION OF RESULTS The following table and graphs summarize the results obtained for the various CanadianPond.ca Products Ltd. tests. Individual computer printouts of the data analysis including time versus D.O. plots for each test run using the specified data analysis method are contained in Appendix. Table 5-1 Summary of ½ Bubble Tubing Test Results DATE: 20-Jun Jun Jun Jun Jun Jun Jun Jun Jun-11 RUN: Barometric Pres. (PSIA) Ambient Temperature ( F) Relative Humidity (%) Line Pressure (PSIG) Line Temperature ( F) Water Temp. ( C) Number Of Aeration Devices Air Release Depth (ft) Blower HPwire Average Air Flow (SCFM) Average Air Flow (SCFM/ft) C*20 Standard Conditions Tank Volume (Ft³) 5,195 5,195 5,195 3,464 3,464 3,464 1,905 1,905 1,905 TDS (mg/l) KLa20 (hr -1 ) TDS Corrected Results KLa20 (hr -1 ) SOTR (Lb O 2 /Hr) SOTE OBSERVED (%) 30.3% 33.5% 35.7% 27.1% 21.6% 24.6% 18.8% 14.0% 15.7% SOTE (%/Ft) 2.02% 2.24% 2.38% 2.71% 2.16% 2.46% 3.76% 2.79% 3.14% SAE (Lb O 2 /Hr/HP) Metric Results (TDS Corrected) Air Flow (Nm³/h) SOTR (g/hr) SOTE - %/m 6.63% 7.34% 7.80% 8.90% 7.08% 8.07% 12.32% 9.16% 10.31% 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 11

13 Figure 5-1 is a plot of air flow rate (SCFM/100 of tubing) versus the Mass Transfer Coefficient (K La 20 hr -1 ) K L a 20 - Hr Figure 5-1 SCFM/100 of tubing v. K La 20 - ½ Bubble Tubing Figure 5-2 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Oxygen Transfer Rate (SOTR Lb. O2/Hr./100 of tubing) SOTR - Lb O 2/Hr Figure 5-2 SCFM/100 of tubing v. SOTR/100 of tubing - ½ Bubble Tubing 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 12

14 SOTE - %/Ft SOTE - % Figure 5-3 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Oxygen Transfer Efficiency (SOTE %). 40% 35% 30% 25% 20% 15% 10% 1 5% 0% Figure 5-3 SCFM/100 of tubing v. SOTE - ½ Bubble Tubing Figure 5-4 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Oxygen Transfer Efficiency (SOTE %/Ft. of Submergence). 4.0% 3.5% 3.0% 2.5% 2.0% 1.5% 1.0% 0.5% 1 0.0% Figure 5-4 SCFM/100 of tubing v. SOTE/Ft. of Submergence - ½ Bubble Tubing 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 13

15 SAE - Lb O2/Hr/HPwire SOTE - %/m Figure 5-5 is a plot of air flow rate (Nm³/h/100 of tubing) versus Standard Oxygen Transfer Efficiency (SOTE %/m of Submergence). 14.0% 12.0% 10.0% 8.0% 6.0% 4.0% 1 2.0% 0.0% Total Airflow - Nm³/h Figure 5-5 SCFM/100 of tubing v. SOTE - ½ Bubble Tubing Figure 5-6 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Aerator Efficiency (SAE Lb. O 2/Hr./HP wire) Figure 5-6 SCFM/100 of tubing v. SAE - ½ Bubble Tubing 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 14

16 Table 5-2 Summary of ¾ Bubble Tubing Test Results DATE: 30-Jun Jun-11 1-Jul-11 5-Jul-11 5-Jul-11 6-Jul-11 7-Jul-11 8-Jul-11 8-Jul-11 RUN: Barometric Pres. (PSIA) Ambient Temperature ( F) Relative Humidity (%) Line Pressure (PSIG) Line Temperature ( F) Water Temp. ( C) Number Of Aeration Devices Air Release Depth (ft) Blower HPwire Average Air Flow (SCFM) Average Air Flow (SCFM/ft) C*20 Standard Conditions Tank Volume (Ft³) 5,310 5,310 5,310 3,464 3,464 3,464 1,963 1,963 1,963 TDS (mg/l) KLa20 (hr -1 ) TDS Corrected Results KLa20 (hr -1 ) SOTR (Lb O 2 /Hr) SOTE OBSERVED (%) 31.8% 37.3% 42.1% 22.7% 26.5% 34.6% 16.8% 14.4% 21.3% SOTE (%/Ft) 2.12% 2.49% 2.81% 2.27% 2.65% 3.46% 3.26% 2.78% 4.11% SAE (Lb O 2 /Hr/HP) Metric Results (TDS Corrected) Air Flow (Nm³/h) SOTR (g/hr) SOTE - %/m 6.95% 8.15% 9.21% 7.46% 8.69% 11.36% 10.69% 9.13% 13.50% Figure 5-7 is a plot of air flow rate (SCFM/100 of tubing) versus the Mass Transfer Coefficient (K La 20 hr -1 ) K L a 20 - Hr Figure 5-7 SCFM/100 of tubing v. K La 20 - ¾ Bubble Tubing 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 15

17 Figure 5-8 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Oxygen Transfer Rate (SOTR Lb. O2/Hr./100 of tubing) SOTR - Lb O 2/Hr SOTE - % Figure 5-8 SCFM/100 of tubing v. SOTR/100 of tubing - ¾ Bubble Tubing Figure 5-9 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Oxygen Transfer Efficiency (SOTE %). 50% 45% 40% 35% 30% 25% 20% 15% 10% 1 5% 0% Figure 5-9 SCFM/100 of tubing v. SOTE - ¾ Bubble Tubing 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 16

18 SOTE - %/m SOTE - %/Ft Figure 5-10 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Oxygen Transfer Efficiency (SOTE %/Ft. of Submergence). 4.5% 4.0% 3.5% 3.0% 2.5% 2.0% 1.5% 1.0% 1 0.5% 0.0% Figure 5-10 SCFM/100 of tubing v. SOTE/Ft. of Submergence - ¾ Bubble Tubing Figure 5-17 is a plot of air flow rate (Nm³/h/100 of tubing) versus Standard Oxygen Transfer Efficiency (SOTE %/m of Submergence). 16.0% 14.0% 12.0% 10.0% 8.0% 6.0% 4.0% 1 2.0% 0.0% Total Airflow - Nm³/h Figure 5-11 SCFM/100 of tubing v. SOTE - ¾ Bubble Tubing 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 17

19 SAE - Lb O2/Hr/HPwire Figure 5-18 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Aerator Efficiency (SAE Lb. O 2/Hr./HP wire) Figure 5-12 SCFM/100 of tubing v. SAE - ¾ Bubble Tubing Table 5-3 Summary of OctoAir-10 Test Results DATE: 1-Aug-11 1-Aug-11 2-Aug-11 8/84/ Aug-11 3-Aug Jul Jul Jul-11 RUN: Barometric Pres. (PSIA) Ambient Temperature ( F) Relative Humidity (%) Line Pressure (PSIG) Line Temperature ( F) Water Temp. ( C) Number Of Aeration Devices Air Release Depth (ft) Blower HPwire Average Air Flow (SCFM) Average Air Flow (SCFM/ft) C*20 Standard Conditions Tank Volume (Ft³) 5,195 5,195 5,195 3,464 3,464 3,464 1,934 1,934 1,934 TDS (mg/l) KLa20 (hr -1 ) TDS Corrected Results KLa20 (hr -1 ) SOTR (Lb O 2 /Hr) SOTE OBSERVED (%) 18.6% 21.6% 32.7% 15.0% 19.6% 26.3% 14.9% 19.4% 11.1% SOTE (%/Ft) 1.24% 1.44% 2.18% 1.50% 1.96% 2.63% 2.97% 3.88% 2.23% SAE (Lb O 2 /Hr/HP) Metric Results (TDS Corrected) Air Flow (Nm³/h) SOTR (g/hr) SOTE - %/m 4.07% 4.72% 7.14% 4.91% 6.45% 8.62% 9.75% 12.72% 7.32% 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 18

20 Figure 5-13 is a plot of air flow rate (SCFM/100 of tubing) versus the Mass Transfer Coefficient (K La 20 hr -1 ) K L a 20 - Hr Figure 5-13 SCFM/100 of tubing v. K La 20 - OctoAir-10 Figure 5-14 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Oxygen Transfer Rate (SOTR Lb. O2/Hr./100 of tubing) SOTR - Lb O 2/Hr Figure 5-14 SCFM/100 of tubing v. SOTR/100 of tubing - OctoAir WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 19

21 SOTE - %/Ft SOTE - % Figure 5-15 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Oxygen Transfer Efficiency (SOTE %). 35% 30% 25% 20% 15% 10% 5% 1 0% Figure 5-15 SCFM/100 of tubing v. SOTE - OctoAir-10 Figure 5-16 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Oxygen Transfer Efficiency (SOTE %/Ft. of Submergence). 4.5% 4.0% 3.5% 3.0% 2.5% 2.0% 1.5% 1.0% 1 0.5% 0.0% Figure 5-16 SCFM/100 of tubing v. SOTE/Ft. of Submergence - OctoAir WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 20

22 SAE - Lb O2/Hr/HPwire SOTE - %/m Figure 5-17 is a plot of air flow rate (Nm³/h/100 of tubing) versus Standard Oxygen Transfer Efficiency (SOTE %/m of Submergence). 14.0% 12.0% 10.0% 8.0% 6.0% 4.0% 1 2.0% 0.0% Total Airflow - Nm³/h Figure 5-17 SCFM/100 of tubing v. SOTE - OctoAir-10 Figure 5-18 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Aerator Efficiency (SAE Lb. O 2/Hr./HP wire) Figure 5-18 SCFM/100 of tubing v. SAE - OctoAir WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 21

23 Figure 5-19 is a plot of air flow rate (SCFM/100 of tubing) versus Standard Oxygen Transfer Efficiency (SOTE %/Ft. of Submergence) for all tubing tested. 4.5% 4.0% 3.5% 3.0% SOTE - %/Ft 2.5% 2.0% 1.5% 1.0% 0.5% ½'' - ½'' - 1 ½'' - ¾'' - 1 ¾'' - ¾'' - Octo-10-1 Octo-10 - Octo % Figure 5-19 SCFM/100 of tubing v. SOTE/Ft. of Submergence All Tubing 6. CONCLUSIONS Based on the results obtained testing the CanadianPond.ca Products Ltd. aeration systems, the following conclusions are offered: 1. Over the airflow range tested, SOTR increases with increasing airflow rate and increasing air release depth. 2. SOTE and SAE decrease with increasing airflow rate. 3. SOTE increases with increasing air release depth. 4. SAE only marginally changes as the air release depth increases. 5. SOTE per Ft. of diffuser submergence decrease with increasing airflow rate over the range of airflows tested. 6. SOTE per Ft. of diffuser submergence decreases with increasing air release depth over the range of depths tested. 7. The ¾ Bubble Tubing results indicate a performance advantage compared to the other systems tested. 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 22

24 8. The OctoAir-10 produced lower results than the other systems tested. Note that this is expected due to the concentration of the aeration system in a small area of the tank as opposed to the Bubble Tubing which has more complete floor coverage of the basin. Overall, the results obtained for the CanadianPond.ca Products Ltd. aeration systems were uniformly excellent and produced some of the highest SOTE values GSEE, Inc. has observed. 7. CERTIFICATION GSEE, Inc., certifies that the results presented in this report are accurate and were obtained using the test procedures described above. Gerald L. Shell, PE 599 WALDRON RD., LAVERGNE, TN (615) FAX (615) PAGE 23

Fine Pore Aeration Systems Testing Michael K. Stenstrom Professor, Civil and Environmental Engineering Department

Fine Pore Aeration Systems Testing Michael K. Stenstrom Professor, Civil and Environmental Engineering Department Fine Pore Aeration Systems Testing Michael K. Stenstrom Professor, Civil and Environmental Engineering Department Fine Pore Diffusers Fine pore aeration systems are the most energy conserving alternative

More information

ACFM vs. SCFM vs. ICFM Series of Technical White Papers from Squire-Cogswell

ACFM vs. SCFM vs. ICFM Series of Technical White Papers from Squire-Cogswell ACFM vs. SCFM vs. ICFM Series of Technical White Papers from Squire-Cogswell Squire Cogswell / Aeros Instruments, Inc. 1111 Lakeside Drive Gurnee, IL 60031 Phone: (800) 448-0770 Fax: (847) 855-6304 info@squire-cogswell.com

More information

Measurements of k L a. Steady-State Mass Balance Method:

Measurements of k L a. Steady-State Mass Balance Method: Measurements of k a Steady-State Mass Balance Method: In theory, the K a in an apparatus that is operating continuously under steady-state conditions could be evaluated from the flow rates and the concentrations

More information

UTILIZATION OF LEAKY PIPES AS AERATION SYSTEM FOR AQUACULTURE

UTILIZATION OF LEAKY PIPES AS AERATION SYSTEM FOR AQUACULTURE UTILIZATION OF LEAKY PIPES AS AERATION SYSTEM FOR AQUACULTURE SAMIR AHMAD ALI ABSTRACT This paper aims to study the possibility of using the leaky pipes as a diffused-air aeration in the aquacultural systems.

More information

Quick Reference Technical Data

Quick Reference Technical Data Bulletin 127C 2 Quick Reference Technical Data For over 100 years, The Spencer Turbine Company has specialized in innovative solutions to air and gas handling problems. Spencer's product line includes

More information

Chapter 13 Gases, Vapors, Liquids, and Solids

Chapter 13 Gases, Vapors, Liquids, and Solids Chapter 13 Gases, Vapors, Liquids, and Solids Property is meaning any measurable characteristic of a substance, such as pressure, volume, or temperature, or a characteristic that can be calculated or deduced,

More information

The Use of a Process Simulator to Model Aeration Control Valve Position and System Pressure

The Use of a Process Simulator to Model Aeration Control Valve Position and System Pressure The Use of a Process Simulator to Model Aeration Control Valve Position and System Pressure Matthew Gray 1 * and Steve Kestel 1 1 BioChem Technology, King of Prussia, PA *Email: mgray@biochemtech.com ABSTRACT

More information

Cover Page for Lab Report Group Portion. Pump Performance

Cover Page for Lab Report Group Portion. Pump Performance Cover Page for Lab Report Group Portion Pump Performance Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 March 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:

More information

Device Description. Operating Information. CP Q (eq. 1) GT. Technical Bulletin TB-0607-CFP Hawkeye Industries Critical Flow Prover

Device Description. Operating Information. CP Q (eq. 1) GT. Technical Bulletin TB-0607-CFP Hawkeye Industries Critical Flow Prover A compressible fluid traveling at subsonic velocity through a duct of constant cross section will increase velocity when passing through a region of reduced cross-sectional area (in this case, an orifice)

More information

Technical Committee on LP-Gas at Utility Gas Plants

Technical Committee on LP-Gas at Utility Gas Plants Technical Committee on LP-Gas at Utility Gas Plants Addendum to the Agenda Sheraton Denver Downtown 1550 Court Place Denver, CO 80202 August 7-8, 2013 The following items relate to item 5.B of the Agenda:

More information

METHOD 2E - DETERMINATION OF LANDFILL GAS PRODUCTION FLOW RATE. NOTE: This method does not include all of the

METHOD 2E - DETERMINATION OF LANDFILL GAS PRODUCTION FLOW RATE. NOTE: This method does not include all of the 287 METHOD 2E - DETERMINATION OF LANDFILL GAS PRODUCTION FLOW RATE NOTE: This method does not include all of the specifications (e.g., equipment and supplies) and procedures (e.g., sampling and analytical)

More information

The Discussion of this exercise covers the following points: Pumps Basic operation of a liquid pump Types of liquid pumps The centrifugal pump.

The Discussion of this exercise covers the following points: Pumps Basic operation of a liquid pump Types of liquid pumps The centrifugal pump. Exercise 2-3 Centrifugal Pumps EXERCISE OBJECTIVE In this exercise, you will become familiar with the operation of a centrifugal pump and read its performance chart. You will also observe the effect that

More information

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Lab 1c Isentropic Blow-down Process and Discharge Coefficient 058:080 Experimental Engineering Lab 1c Isentropic Blow-down Process and Discharge Coefficient OBJECTIVES - To study the transient discharge of a rigid pressurized tank; To determine the discharge coefficients

More information

Figure Vapor-liquid equilibrium for a binary mixture. The dashed lines show the equilibrium compositions.

Figure Vapor-liquid equilibrium for a binary mixture. The dashed lines show the equilibrium compositions. Another way to view this problem is to say that the final volume contains V m 3 of alcohol at 5.93 kpa and 20 C V m 3 of air at 94.07 kpa and 20 C V m 3 of air plus alcohol at 100 kpa and 20 C Thus, the

More information

METHOD 2D - MEASUREMENT OF GAS VOLUME FLOW RATES IN SMALL PIPES AND DUCTS. NOTE: This method does not include all of the

METHOD 2D - MEASUREMENT OF GAS VOLUME FLOW RATES IN SMALL PIPES AND DUCTS. NOTE: This method does not include all of the 279 METHOD 2D - MEASUREMENT OF GAS VOLUME FLOW RATES IN SMALL PIPES AND DUCTS NOTE: This method does not include all of the specifications (e.g., equipment and supplies) and procedures (e.g., sampling)

More information

Exercise 4-2. Centrifugal Pumps EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Pumps

Exercise 4-2. Centrifugal Pumps EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Pumps Exercise 4-2 Centrifugal Pumps EXERCISE OBJECTIVE Familiarize yourself with the basics of liquid pumps, specifically with the basics of centrifugal pumps. DISCUSSION OUTLINE The Discussion of this exercise

More information

The most common terms rating air flow capacity are ICFM, FAD, ANR, SCFM or nl/min

The most common terms rating air flow capacity are ICFM, FAD, ANR, SCFM or nl/min Rating of Air Compressors and Air Equipment The most common terms rating air flow capacity are ICFM, FAD, ANR, SCFM or nl/min There is no universal standard for rating air compressors, air equipment and

More information

CHAPTER 3 : AIR COMPRESSOR

CHAPTER 3 : AIR COMPRESSOR CHAPTER 3 : AIR COMPRESSOR Robotic & Automation Department FACULTY OF MANUFACTURING ENGINEERING, UTeM Learning Objectives Identify types of compressors available Calculate air capacity rating of compressor

More information

The Discussion of this exercise covers the following points:

The Discussion of this exercise covers the following points: Exercise 3-2 Orifice Plates EXERCISE OBJECTIVE In this exercise, you will study how differential pressure flowmeters operate. You will describe the relationship between the flow rate and the pressure drop

More information

The Effects of Stir Speed, Airflow Rate, Diffusers, and Volume on. Water Aeration

The Effects of Stir Speed, Airflow Rate, Diffusers, and Volume on. Water Aeration The Effects of Stir Speed, Airflow Rate, Diffusers, and Volume on Water Aeration Cycle 3 - Group B1 Group Leader: David Nguyen Katarina Van Wonterghem Joseph Jimenez Justin Chung CBEMS 140A University

More information

Advanced Management of Compressed Air Systems Pre-Workshop Assignment

Advanced Management of Compressed Air Systems Pre-Workshop Assignment Advanced Management of Compressed Air Systems Page 1 In order to ensure that the Compressed Air Challenge Level II Training is most useful to you, it will be important for you to bring information about

More information

Technical Bulletin 148. Comparison of Flat Panel to MiniPanel. by:

Technical Bulletin 148. Comparison of Flat Panel to MiniPanel. by: Technical Bulletin 148 Comparison of Flat Panel to MiniPanel by: Environmental Dynamics International Published: 6/2005 BULLETIN BRIEF Environmental Dynamics International offers a full spectrum of fine

More information

Natural Gas Gathering

Natural Gas Gathering Natural Gas Gathering Course No: R04-002 Credit: 4 PDH Jim Piter, P.E. Continuing Education and Development, Inc. 9 Greyridge Farm Court Stony Point, NY 10980 P: (877) 322-5800 F: (877) 322-4774 info@cedengineering.com

More information

Generating Calibration Gas Standards

Generating Calibration Gas Standards Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and

More information

Fundamentals of Compressed Air Systems. Pre-Workshop Assignment

Fundamentals of Compressed Air Systems. Pre-Workshop Assignment Page 1 In order to ensure that the Compressed Air Challenge Fundamentals of Compressed Air Systems Training is most useful to you, it will be important for you to bring information about your plant s compressed

More information

Pro-V Multivariable Flowmeter Model M22 In-line Vortex

Pro-V Multivariable Flowmeter Model M22 In-line Vortex Pro-V Multivariable Flowmeter Model M22 In-line Vortex Pro-V TM Advantage: Volumetric or mass flow monitoring of most liquids, gases, and steam Multivariable meter delivers mass flow, temperature, pressure,

More information

KNOWN: Mass, pressure, temperature, and specific volume of water vapor.

KNOWN: Mass, pressure, temperature, and specific volume of water vapor. .0 The specific volume of 5 kg of water vapor at.5 MPa, 440 o C is 0.60 m /kg. Determine (a) the volume, in m, occupied by the water vapor, (b) the amount of water vapor present, in gram moles, and (c)

More information

Integrating Aeration Blowers with Most-Open-Valve

Integrating Aeration Blowers with Most-Open-Valve Integrating Aeration Blowers with Most-Open-Valve Sponsored by Tom Jenkins P.E., JenTech Inc. Keynote Speaker For your free subscription, please visit http://www.blowervacuumbestpractices.com/magazine/subscription.

More information

F&J SPECIALTY PRODUCTS The nucleus of quality air monitoring programs

F&J SPECIALTY PRODUCTS The nucleus of quality air monitoring programs F&J SPECIALTY PRODUCTS The nucleus of quality air monitoring programs AIR FLOW CALIBRATORS FOR MOST AIR MONITORING PROGRAMS World Calibrator PC Version World Calibrator VFD Version Compact Digital Calibrator

More information

Pressure Measurement

Pressure Measurement Pressure Measurement Manometers Sensors, Transducers Ashish J. Modi Lecturer, Dept. of Mech.Engg., Shri S.V.M. inst. Of Technology, Bharuch Pressure Pressure is a force per unit area exerted by a fluid

More information

Comments on Homework. Class 4 - Pressure. Atmospheric Pressure. Gauge vs. Absolute Pressure. 2. Gauge vs. Absolute Pressure. 1.

Comments on Homework. Class 4 - Pressure. Atmospheric Pressure. Gauge vs. Absolute Pressure. 2. Gauge vs. Absolute Pressure. 1. Class 4 - Pressure 1. Definitions 2. Gauge Pressure 3. Pressure and Height of Liquid Column (Head) 4. Pressure Measurement and Manometers Please don t forget the special problem for the next HW assignment

More information

Quiz #1 Thermodynamics Spring, 2018 Closed Book, Open Appendices, Closed Notes, CLOSED CALCULATORS

Quiz #1 Thermodynamics Spring, 2018 Closed Book, Open Appendices, Closed Notes, CLOSED CALCULATORS Quiz #1 Closed Book, Open Appendices, Closed Notes, CLOSED CALCULATORS An astronaut has a mass of 161 lbm on the surface of the earth. Calculate his weight (in lbf) on planet Rigel 4 where g = 20.0 ft/s

More information

Performance Characteristics of Airlift Pumps with Vortex Induced by Tangential Fluid Injection

Performance Characteristics of Airlift Pumps with Vortex Induced by Tangential Fluid Injection Bucknell University Bucknell Digital Commons Honors Theses Student Theses 2011 Performance Characteristics of Airlift Pumps with Vortex Induced by Tangential Fluid Injection Jacob Riglin Bucknell University

More information

12mm Dissolved Oxygen Sensor Care and Use Instructions

12mm Dissolved Oxygen Sensor Care and Use Instructions 12mm Dissolved Oxygen Sensor Care and Use Instructions How Dissolved Oxygen Sensors Work Your Dissolved Oxygen (DO) Probe is a galvanic electrochemistry device; i.e. it does not require power from your

More information

CHM Basics of Gases (r14) Charles Taylor 1/9

CHM Basics of Gases (r14) Charles Taylor 1/9 CHM 110 - Basics of Gases (r14)- 2014 Charles Taylor 1/9 Introduction The gas phase is noticeably different from the other two phases of matter. Here are some of the more obvious differences. Gases are

More information

Gas Turbine Performance Analysis

Gas Turbine Performance Analysis Gas Turbine Performance Analysis Gas turbines may seem too complicated or overwhelming at first glance, but for regular field monitoring on a relative basis, it is not all that difficult. A regular check

More information

PRODUCT SHEET. Order probe only as RXPROBE02

PRODUCT SHEET. Order probe only as RXPROBE02 SS69L DISSOLVED OXYGEN PROBE TRANSDUCER Order probe only as RXPROBE02 Order interface only as BSL-TCI16 SS69L Components The SS69L transducer measures dissolved oxygen. The SS69L includes a dissolved oxygen

More information

North American Orifice Metering System

North American Orifice Metering System Combustion North American Orifice Metering System 8697 and M8697 Metering Orifices accurately measure gas flows to industrial burners. They are compact, inexpensive, and can be installed in individual

More information

Comments on Homework. Quiz. Class 3 - Pressure. Atmospheric Pressure. 2. Gauge vs. Absolute Pressure. 1. Definitions. Temperature conversion

Comments on Homework. Quiz. Class 3 - Pressure. Atmospheric Pressure. 2. Gauge vs. Absolute Pressure. 1. Definitions. Temperature conversion Comments on Homework Quiz Temperature conversion T ( R) = T (K) 1.8 T ( C) = T(K) - 273.15 T ( F) = T( R) - 460 However, difference in temperature is: T ( C) = T (K) T ( F) = T ( R) T ( R) = 1.8 T ( C)

More information

Fluid Machinery Introduction to the laboratory measurements

Fluid Machinery Introduction to the laboratory measurements Fluid Machinery Introduction to the laboratory measurements Csaba H s (csaba.hos@hds.bme.hu) Ferenc Hegedus (hegedusf@hds.bme.hu) February 21, 2014 1 Requirements related to the measurement part of the

More information

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003 CEE 331 Lab 1 Page 1 of 9 SAFETY Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003 Laboratory exercise based on an exercise developed by Dr. Monroe Weber-Shirk The major safety hazard in this

More information

GLOSSARY OF TERMS. Adiabatic Compression Compression process when all heat of compression is retained in the gas being compressed.

GLOSSARY OF TERMS. Adiabatic Compression Compression process when all heat of compression is retained in the gas being compressed. GLOSSARY OF TERMS Absolute pressure Total pressure measured from absolute zero i.e. a perfect vacuum. As a practical matter, gauge pressure plus atmospheric pressure. Absolute temperature Temperature measured

More information

Noth American 7347 High Pressure Gas Regulators

Noth American 7347 High Pressure Gas Regulators Combustion Noth American 7347 High Pressure Gas s Bulletin 7347 7347 s reduce high gas supply pressures to practical use levels. Since capacities will vary with the pressure drop across the regulator (see

More information

S.A. Klein and G.F. Nellis Cambridge University Press, 2011

S.A. Klein and G.F. Nellis Cambridge University Press, 2011 16-1 A flow nozzle is to be used to determine the mass flow rate of air through a 1.5 inch internal diameter pipe. The air in the line upstream of the meters is at 70 F and 95 psig. The barometric pressure

More information

Pro-V Multivariable Flowmeter

Pro-V Multivariable Flowmeter Pro-V Multivariable Flowmeter Model M23 Insertion Meter M23-V flowmeter provides cost-effective volumetric flow monitoring solution for most liquids M23-VT incorporates temperature sensing to provide a

More information

AIR FLOW CALIBRATORS Single Sensor Designs

AIR FLOW CALIBRATORS Single Sensor Designs F&J SPECIALTY PRODUCTS, INC. The Nucleus of Quality Air Monitoring Programs AIR FLOW CALIBRATORS Single Sensor Designs 100 240 VAC World Calibrator PC Version World Calibrator VFD Version Compact Digital

More information

Runs Solution Temp. Pressure zero reset

Runs Solution Temp. Pressure zero reset IAEA-SM-367/8/02/P High Quality Tank Calibration Study JSGO H.Higuchi, S.Takeda NMCC--S. Uchikoshi, Y. Watanabe, K. Kaieda IAEA--D. Sellinschegg, R.Binner 1. INTRODUCTION: The Japanese Nuclear Material

More information

Applied Fluid Mechanics

Applied Fluid Mechanics Applied Fluid Mechanics 1. The Nature of Fluid and the Study of Fluid Mechanics 2. Viscosity of Fluid 3. Pressure Measurement 4. Forces Due to Static Fluid 5. Buoyancy and Stability 6. Flow of Fluid and

More information

Pump-Fan-Compressor Sizing

Pump-Fan-Compressor Sizing Pump-Fan-Compressor Sizing Introduction This program determines the fluid transportation characteristics of dominant pipeline process systems containing no loops. In addition, it determines the yearly

More information

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Universal pycnometer, no

More information

Unit A-2: List of Subjects

Unit A-2: List of Subjects ES312 Energy Transfer Fundamentals Unit A: Fundamental Concepts ROAD MAP... A-1: Introduction to Thermodynamics A-2: Engineering Properties Unit A-2: List of Subjects Basic Properties and Temperature Pressure

More information

MATH AND MAINTENANCE FOR PUMPS AND BLOWERS TRAINING SEMINAR

MATH AND MAINTENANCE FOR PUMPS AND BLOWERS TRAINING SEMINAR MATH AND MAINTENANCE FOR PUMPS AND BLOWERS TRAINING SEMINAR FORCE WEIGHT OF WATER MEASURED IN POUNDS PRESSURE = FORCE PER UNIT AREA WEIGHT PER UNIT AREA PRESSURE (PSI) = FORCE (LBS) / AREA (IN 2 ) P (PSI)

More information

Process Simulator Evaluates Blower and Valve Control Strategies for WWTP Aeration

Process Simulator Evaluates Blower and Valve Control Strategies for WWTP Aeration Process Simulator Evaluates Blower and Valve Control Strategies for WWTP Aeration inshare By Steve Kestel and Tilo Stahl (BioChem Technology) and Matthew Gray (Keystone Engineering Group) Introduction

More information

PRO-50 Instrument Supply Regulator

PRO-50 Instrument Supply Regulator Features CRN Approved The PRO-50 Regulator has been granted a Canadian Registration Number. Sour Service Capability Available in NACE configurations that comply with NACE MR0175/MR0103. Environmental limits

More information

Pressure Measurement

Pressure Measurement Pressure Measurement Absolute and Gage Pressure P abs = P gage + P atm where P abs = Absolute pressure P abs = Gage pressure P abs = atmospheric pressure A perfect vacuum is the lowest possible pressure.

More information

ATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN

ATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN ATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN (This page intentionally blank) ATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN TABLE OF CONTENTS O- Definitions... O- Objective... O- Design

More information

Dean Pump Self-Priming Chemical Process Pumps

Dean Pump Self-Priming Chemical Process Pumps Bulletin C 1.2.34.7 Dean Pump Self-Priming Chemical Process Pumps php Series HEAD CAPACITY RANGE CHARTS php Self Primer - 2 Pole 3500 RPM 500 CAPACITY M 3 /HR 2900 RPM 50 HERTZ 25 50 75 125 150 400 TOTAL

More information

SPH 4C Unit 4 Hydraulics and Pneumatic Systems

SPH 4C Unit 4 Hydraulics and Pneumatic Systems SPH 4C Unit 4 Hydraulics and Pneumatic Systems Properties of Fluids and Pressure Learning Goal: I can explain the properties of fluids and identify associated units. Definitions: Fluid: A substance that

More information

MEASUREMENT BEST PRACTICES FORIMPROVEDREFINERY SAFETY, AVAILABILITY & EFFICIENCY

MEASUREMENT BEST PRACTICES FORIMPROVEDREFINERY SAFETY, AVAILABILITY & EFFICIENCY MEASUREMENT BEST PRACTICES FORIMPROVEDREFINERY SAFETY, AVAILABILITY & EFFICIENCY Mark Menezes, PlantWeb Marketing Manager Rosemount Inc. Eric Wickberg, Pressure Applications Marketing Manager Rosemount

More information

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow Georgia Institute of Technology School of Earth and Atmospheric Sciences EAS 4641, Spring 2007 Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow Purpose of Lab 3: 1) To gain a

More information

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cover Page for Lab Report Group Portion. Head Losses in Pipes Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section

More information

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text. Hach LDO Sensor Instruction Sheet 00745589 Safety Precautions Please read this entire instruction sheet before operating this sensor. Pay particular attention to all danger and caution statements. Failure

More information

Cover Page for Lab Report Group Portion. Lift on a Wing

Cover Page for Lab Report Group Portion. Lift on a Wing Cover Page for Lab Report Group Portion Lift on a Wing Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 17 January 2017 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:

More information

pumping gases JET PUMP TECHNICAL DATA

pumping gases JET PUMP TECHNICAL DATA Section 1000 Bulletin 100 Issued 9/87 Replaces 7/84 JET PUMP TECHNICAL DATA pumping gases This technical bulletin includes general information about Penberthy Jet Pumps plus specific details for selecting

More information

Chapter 13 Fluids. Copyright 2009 Pearson Education, Inc.

Chapter 13 Fluids. Copyright 2009 Pearson Education, Inc. Chapter 13 Fluids Phases of Matter Density and Specific Gravity Pressure in Fluids Atmospheric Pressure and Gauge Pressure Pascal s Principle Units of Chapter 13 Measurement of Pressure; Gauges and the

More information

Irrigation &Hydraulics Department lb / ft to kg/lit.

Irrigation &Hydraulics Department lb / ft to kg/lit. CAIRO UNIVERSITY FLUID MECHANICS Faculty of Engineering nd Year CIVIL ENG. Irrigation &Hydraulics Department 010-011 1. FLUID PROPERTIES 1. Identify the dimensions and units for the following engineering

More information

Crab and Lobster Live Holding System Part II (Water Pumps, Air Pumps and Pipework)

Crab and Lobster Live Holding System Part II (Water Pumps, Air Pumps and Pipework) Crab and Lobster Live Holding System Part II (Water Pumps, Air Pumps and Pipework) 1990/2/SF/AM The increased demand for crabs and lobsters, and the requirement to export live shellfish has led to the

More information

The Application of Temperature and/or Pressure Correction Factors in Gas Measurement

The Application of Temperature and/or Pressure Correction Factors in Gas Measurement The Application of Temperature and/or Pressure Correction Factors in Gas Measurement COMBINED BOYLE S CHARLES GAS LAWS To convert measured volume at metered pressure and temperature to selling volume at

More information

AERATION & MIXING SYSTEMS

AERATION & MIXING SYSTEMS SWIM and Horizon 2020 Support Mechanism Working for a Sustainable Mediterranean, Caring for our Future SWIM-H2020 SM Regional Activities 14 Presented by: MOHAMMD SUTARI, MEHSIP RESIDENT EXPERT-JORDAN SWIM

More information

Cover Page for Lab Report Group Portion. Drag on Spheres

Cover Page for Lab Report Group Portion. Drag on Spheres Cover Page for Lab Report Group Portion Drag on Spheres Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 29 September 2017 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:

More information

A TYPICAL AERATION UPGRADE CONSIDERATIONS AND OUTCOMES. Michael Young. Michael Young, Director, Faraga Pty Ltd and Aer-Force Pty Ltd

A TYPICAL AERATION UPGRADE CONSIDERATIONS AND OUTCOMES. Michael Young. Michael Young, Director, Faraga Pty Ltd and Aer-Force Pty Ltd A TYPICAL AERATION UPGRADE CONSIDERATIONS AND OUTCOMES Paper Presented by: Michael Young Author: Michael Young, Director, Faraga Pty Ltd and Aer-Force Pty Ltd 6 th Annual WIOA NSW Water Industry Engineers

More information

Aerobic Respiration. Evaluation copy

Aerobic Respiration. Evaluation copy Aerobic Respiration Computer 17 Aerobic cellular respiration is the process of converting the chemical energy of organic molecules into a form immediately usable by organisms. Glucose may be oxidized completely

More information

COMPAFLOW. Compressed Air. Volumetric flow. Gas. Mass flow. Steam. Net volumetric flow. Liquid

COMPAFLOW. Compressed Air. Volumetric flow. Gas. Mass flow. Steam. Net volumetric flow. Liquid Volumetric flow Compressed Air Mass flow Gas Net volumetric flow Steam Liquid Universal compact orifice flow meter combines a compact orifice flow sensor and sotiphicated converter One-piece flow sensor,

More information

BI-680 Online Dissolved Oxygen Controller Instruction Manual

BI-680 Online Dissolved Oxygen Controller Instruction Manual BI-680 Online Dissolved Oxygen Controller Instruction Manual BANTE INSTRUMENTS CO., LTD BI-680 Online Dissolved Oxygen Controller 1 Introduction Thank you for selecting the BI-680 online dissolved oxygen

More information

Case 12 Multistage Centrifugal Refrigeration System Halocarbon Refrigerant

Case 12 Multistage Centrifugal Refrigeration System Halocarbon Refrigerant Case 12 Multistage Centrifugal Refrigeration System Halocarbon Refrigerant Copy Right By: Thomas T.S. Wan 温 ) April 15, 2011 All Rights Reserved Case Background: This case is to show how to achieve the

More information

Hydrostatic pressure Consider a tank of fluid which contains a very thin plate of (neutrally buoyant) material with area A. This situation is shown in Figure below. If the plate is in equilibrium (it does

More information

CTB3365x Introduction to Water Treatment

CTB3365x Introduction to Water Treatment CTB3365x Introduction to Water Treatment W4e Aeration Merle de Kreuk There is one last thing of the biological part that needs some design. As you know, nitrifyers need oxygen, and also the heterotrophs,

More information

RCD - Ultra-Rugged Differential Pressure Flowmeter

RCD - Ultra-Rugged Differential Pressure Flowmeter 0.15-1 to 85-600 GPM Liquids 2.5-25 to 300-1700 SCFM Gases 1/2 to 3 Line Sizes Bronze or 316-Ti Stainless Steel Mechanical Indicator, Digital Indicator, Transmitters and Switches Available Custom Calibrations

More information

8. Now plot on the following grid the values of T (K) and V from the table above, and connect the points.

8. Now plot on the following grid the values of T (K) and V from the table above, and connect the points. Charles s Law According to Charles s law, the volume of a fixed mass of gas varies directly with its Kelvin temperature if its pressure is constant. The following table contains Celsius temperature and

More information

Phys101 Lectures Fluids I. Key points: Pressure and Pascal s Principle Buoyancy and Archimedes Principle. Ref: 10-1,2,3,4,5,6,7.

Phys101 Lectures Fluids I. Key points: Pressure and Pascal s Principle Buoyancy and Archimedes Principle. Ref: 10-1,2,3,4,5,6,7. Phys101 Lectures 21-22 Fluids I Key points: Pressure and Pascal s Principle Buoyancy and Archimedes Principle Ref: 10-1,2,3,4,5,6,7. Page 1 10-1 Phases of Matter The three common phases of matter are solid,

More information

SIZING THE EXTROL DIAPHRAGM-TYPE HYDRO-PNEUMATIC TANK

SIZING THE EXTROL DIAPHRAGM-TYPE HYDRO-PNEUMATIC TANK SIZING THE EXTROL DIAPHRAGM-TYPE HYDRO-PNEUMATIC TANK 1400 Division Road, West Warwick, RI 02893 T: 401.884.6300 F: 401.885.2567 www.amtrol.com Sizing the ExTrol Diaphragm-Type Hydro-Pneumatic Tank For

More information

AIAA Brush Seal Performance Evaluation. P. F. Crudgington Cross Manufacturing Co. Ltd. Devizes, ENGLAND

AIAA Brush Seal Performance Evaluation. P. F. Crudgington Cross Manufacturing Co. Ltd. Devizes, ENGLAND AIAA 98-3172 Brush Seal Performance Evaluation P. F. Crudgington Cross Manufacturing Co. Ltd. Devizes, ENGLAND BRUSH SEAL PERFORMANCE EVALUATION AIAA-98-3172 P. F. Crudgington Cross Manufacturing Co. Ltd

More information

Expansion Tank Design (Chilled Water)

Expansion Tank Design (Chilled Water) Expansion Tank Design (Chilled Water) Table of Contents 1.0 Introduction... 2 1.1 Units... 2 2.0 Disclaimer... 2 3.0 Expansion Tank Types... 2 3.1 Open Tank... 2 3.2 Closed Tank with No Bladder... 3 3.3

More information

DISSOLVED OXYGEN SENSOR BT34i

DISSOLVED OXYGEN SENSOR BT34i DISSOLVED OXYGEN SENSOR BT34i USER S GUIDE CENTRE FOR MICROCOMPUTER APPLICATIONS http://www.cma-science.nl Short description The CMA Dissolved Oxygen (DO) sensor BT34i measures the concentration of dissolved

More information

Chapter 5: Gases 5.1 Pressure Why study gases? An understanding of real world phenomena. An understanding of how science works.

Chapter 5: Gases 5.1 Pressure Why study gases? An understanding of real world phenomena. An understanding of how science works. Chapter 5: Gases 5.1 Pressure Why study gases? An understanding of real world phenomena. An understanding of how science works. A Gas Uniformly fills any container. Easily compressed. Mixes completely

More information

1 PIPESYS Application

1 PIPESYS Application PIPESYS Application 1-1 1 PIPESYS Application 1.1 Gas Condensate Gathering System In this PIPESYS Application, the performance of a small gascondensate gathering system is modelled. Figure 1.1 shows the

More information

Test Report # Rev 0. Adiabatic Compression With Constant Bleed Valve

Test Report # Rev 0. Adiabatic Compression With Constant Bleed Valve Test Report # 30016558 Rev 0 Adiabatic Compression With Constant Bleed Valve Scope: This test report contains test data on the effects of adiabatic compression in nitrogen service on AP3600 diaphragm valves,

More information

Gas Physics Pressure and Flow Topics Covered:

Gas Physics Pressure and Flow Topics Covered: Gas Physics Pressure and Flow Topics Covered: Molecular Theory of Gases Definition of Pressure The Gas Laws Definition of Flow Definition of Pressure Drop Gas Physics Pressure and Flow Topics Covered:

More information

Integral type Differential pressure flowmeter VNT Series

Integral type Differential pressure flowmeter VNT Series Integral type Differential pressure flowmeter VNT Series OUTLINE VH series Wafer-Cone differential pressure flowmeter and high precision differential pressure transmitter are integrated into one flowmeter.

More information

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No.. Innovative Instruments, Inc. http://www.2in.com Dissolved Oxygen Meter ino2 Instruction Manual Serial No.. Innovative Instruments, Inc. 8533 Queen Brooks Ct. Tampa, FL 33637. USA Phone: 813-727-0676. Fax:

More information

Exercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION

Exercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION Exercise 2-2 Second-Order Interacting Processes EXERCISE OBJECTIVE Familiarize yourself with second-order interacting processes and experiment with the finer points of controller tuning to gain a deeper

More information

Chapter 15 Fluid. Density

Chapter 15 Fluid. Density Density Chapter 15 Fluid Pressure Static Equilibrium in Fluids: Pressure and Depth Archimedes Principle and Buoyancy Applications of Archimedes Principle By Dr. Weining man 1 Units of Chapter 15 Fluid

More information

Pressure of the atmosphere varies with elevation and weather conditions. Barometer- device used to measure atmospheric pressure.

Pressure of the atmosphere varies with elevation and weather conditions. Barometer- device used to measure atmospheric pressure. Chapter 12 Section 1 Pressure A gas exerts pressure on its surroundings. Blow up a balloon. The gas we are most familiar with is the atmosphere, a mixture of mostly elemental nitrogen and oxygen. Pressure

More information

Unit 2 Kinetic Theory, Heat, and Thermodynamics: 2.A.1 Problems Temperature and Heat Sections of your book.

Unit 2 Kinetic Theory, Heat, and Thermodynamics: 2.A.1 Problems Temperature and Heat Sections of your book. Unit 2 Kinetic Theory, Heat, and Thermodynamics: 2.A.1 Problems Temperature and Heat Sections 10.1 10.2 of your book. Convert the following to Celsius and Kelvin temperatures: 1. 80.0 o F Early E. C.:

More information

Instructions for SMV 3000 Multivariable Configuration (MC) Data Sheets

Instructions for SMV 3000 Multivariable Configuration (MC) Data Sheets Instructions for SMV 3000 Multivariable Configuration (MC) Data Sheets Similar to the TC option for ST 3000 transmitters, the MC option for the SMV 3000 provides a service to our customers which results

More information

Lecture Outline Chapter 15. Physics, 4 th Edition James S. Walker. Copyright 2010 Pearson Education, Inc.

Lecture Outline Chapter 15. Physics, 4 th Edition James S. Walker. Copyright 2010 Pearson Education, Inc. Lecture Outline Chapter 15 Physics, 4 th Edition James S. Walker Chapter 15 Fluids Density Units of Chapter 15 Pressure Static Equilibrium in Fluids: Pressure and Depth Archimedes Principle and Buoyancy

More information

DS/PI-14 DIAPHRAGM SEAL TEMPERATURE ERROR

DS/PI-14 DIAPHRAGM SEAL TEMPERATURE ERROR Diaphragm Seals DS/PI-14 DIAPHRAGM SEAL TEMPERATURE ERROR Diaphragm seals can be used to protect pressure instruments from high temperature process media, cold ambient environments and corrosive process

More information

Turbine Flowmeters for Gas Applications

Turbine Flowmeters for Gas Applications Turbine Flowmeters for Gas Applications Description Flow Technology s FT Series turbine flowmeters utilize a proven flow measurement technology to provide exceptionally reliable digital outputs. Because

More information

Design of a Combined Propeller and Venturi Tube System for Aquaculture Ponds

Design of a Combined Propeller and Venturi Tube System for Aquaculture Ponds Design of a Combined Propeller and Venturi Tube System for Aquaculture Ponds Santi Laksitanonta 1 and Gajendra Singh 2 ABSTRACT The main objective of the study was to design and develop an aerator (a combined

More information

Experiment. THE RELATIONSHIP BETWEEN VOLUME AND TEMPERATURE, i.e.,charles Law. By Dale A. Hammond, PhD, Brigham Young University Hawaii

Experiment. THE RELATIONSHIP BETWEEN VOLUME AND TEMPERATURE, i.e.,charles Law. By Dale A. Hammond, PhD, Brigham Young University Hawaii Experiment THE RELATIONSHIP BETWEEN VOLUME AND TEMPERATURE, i.e.,charles Law By Dale A. Hammond, PhD, Brigham Young University Hawaii The objectives of this experiment are to... LEARNING OBJECTIVES introduce

More information