Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Size: px
Start display at page:

Download "Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator"

Transcription

1 Journal of BUON 8: 55-59, Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator G. Kemikler Istanbul University, Oncology Institute, Department of Medical Radiophysics, Istanbul, Turkey Summary Purpose: To investigate the field size and depth dependence of wedge factors for internal wedge in the isocentrical setup for 6 and 15 MV photon beams. Materials and methods: Measurements were performed at d max 5, 10, and 15 cm depths for fields 5 5 cm 2 to cm 2. The source-isocenter distance (SID was 100 cm. Relative wedge factors (s were calculated to describe the dependence of the wedge factor on the field size and depth. Results: The dependence of wedge factors on field size and depth were less than 2% at 6 MV and 1% at 15 MV X- ray beams for 15 0 and 30 0 wedges. When the field size increased from 5 5 to cm 2, the field size dependence increased by 2.7% and 5.2% for 45 0 and 60 0 wedge, respectively, at 6 MV. The dependence on the field size was less than 2.5% for 45 0 and 60 0 wedges at 15 MV. The depth dependence of 60 0 wedge was up to 6.5% for large field size at the depths at 6 MV. For 15 MV beam, s showed only a small change with increasing depth. Conclusion: The magnitude of error in using a single wedge factor was minimal for thin wedges. However, for 6 MV use of a single wedge factor produced errors up to 6% with 60 0 wedge filter. Therefore, in the presence of 45 0 and 60 0 wedges in the beam, a field size and depth dependence wedge factor or a separate output factor (OF curve for wedged beams should be used. Key words: depth dependence, field size dependence, internal wedge, wedge transmission factor, X-ray beam Introduction Received ; Accepted Author and address for correspondence: Gönül Kemikler, PhD Istanbul University Oncology Institute Department of Medical Radiophysics 34390, Topkapi Istanbul Turkey Tel: ext: Fax: gkemikler@ixir.com In external photon beam treatment, beam modifiers including wedges, blocks, or compensators are frequently used to shape the treatment fields and to modify dose distributions in photon fields. Wedge filters have been commonly used in radiation oncology to improve the dose uniformity in the target volume [1]. The wedge shaped isodoses have been achieved by physical wedges, universal motorized wedges, and dynamic wedges. Universal motorized wedge of 60 0 is used for a portion of the treatment to achieve the effective desired wedge angle [2]. When a wedge is present in a photon field, a wedge factor is used conventionally to describe a reduction of the radiation dose in a wedged field compared to the open field, due to the wedge attenuation. Wedge factors are usually measured at a fixed depth along the central axis in a reference field, such as cm 2 at 100 cm source-skin distance (SSD. These factors are used directly in dose calculations based on the calibrated dose in the reference open field. However, wedges not only attenuate but also harden the original photon beams. They also produce additional scattered radiation downstream. Therefore, dose in a wedged field is more complicated than a single wedge factor could describe. The reported wedge factor variations with field size are up to 10% [3-5]. Wedge factors are affected by many factors including photon beam energy, wedge material, field geometry and depth. Therefore, measurements of wedge factors for various fields and depths

2 56 should be done. However, wedge factors for several treatment conditions such as asymmetric fields or different SSDs for a SID technique also may be required. The purpose of this study was to investigate the field size and depth dependence of wedge factors for internal motorized wedge in the isocentrically setup for 6 and 15 MV photon energies. Materials and methods All measurements were performed with 6-15 MV X-ray beams (GE- Saturne 42, General Electric. The Saturne 42 accelerator uses an internal motorized wedge with 60 0 nominal wedge angle. This internal wedge ( made of tungsten can be positioned by remote control. The wedge angles ranged from 15 0 to 60 0 and were obtained by weighed superposition of the wedged and open beams [2]. Measurements were made in a RWA solid water phantom using a NE 2571, 1007 serial number, 0.6 cc ion chamber with a PTW Unidos electrometer. The reference point for the cylindrical chamber was at the chamber centre. Measurements were performed at d max 5, 10, and 15 cm depths in solid water phantom for field sizes (defined at SID ranging from 5 5 cm 2 to the maximum square field size available for a particular wedge. The maximum-wedged field size was cm. Sourcechamber distance (SCD or SID was 100 cm for all measurements. So, phantom SSDs were cm for 6 MV and cm for 15 MV photon beam when the depth of interest in the patient varied from d max to 15 cm for isocentrical treatment. The long axis of the chamber was always along the nonwedged plan. Measurements were made for two collimator orientations (180º apart and determined the average of the two measurements. The calculated wedge transmission factor (WTF was the ratio of the average ionization with the wedge in place to the average open field ionization: Dwdg ( d, fs WTF ( d, fs = (1 Dopen ( d, fs where D wdg and D open are the doses in a wedged field and in open field, respectively. d, and f s are depth and field size, respectively. In this study, s also were calculated to describe the dependence of the wedge factor on the field size and depth. f was defined as a ratio of the WTF at various depths to that under the reference conditions for field size dependence. For a specific wedge, f was calculated as: (, s f( d, fs = WTF d f WTF ( d, f (2 ref sref where reference depth and field size are d ref =5 cm for 6 MV, d ref =10 cm for 15 MV, and f ref =10 10 cm 2. These reference depths have been chosen because daily dose calibrations have been performed at these depths. Relative wedge factors ( d also were calculated to describe the dependence of the wedge factor on the depth. d was defined as: WTF ( d, fs d( d, fs = WTF ( d, f (3 where d is the ratio of the WTF at various depths to that at the depth of maximum dose (1.5 cm and 3 cm for 6 and 15 MV, respectively. The equation (2 can be rewritten as max Dwdg ( d, fs Dwdg ( dref, fsref f( d, fs = / D ( d, f D ( d, f open s open ref sref Dwdg ( d, fs Dopen( d, fs f( d, fs = / D ( d, f D ( d, f s wdg ref sref open ref sref OFwdg ( d, fs f( d, fs = OF ( d, f (4 open This equation (4 shows that f is a ratio of the OF of a wedged field to the relative OF of an open field. For OFs, doses for both open fields and wedged fields were normalized to the dose at the reference depth in the cm 2 field size. The dose at the depth of 5 cm and 10 cm in the cm 2 field was used as the reference dose at 6 and 15 MV, respectively. Thus, attenuation of the dose with depth is also included by the OF. Results The field size and depth dependence of the internal wedge transmission factors were determined for 6 and 15 MV beams for isocentrical setup. Wedge transmission measurements at 15 0 and 30 0 and relative wedge calculations indicated that the dependence of wedge factors on field size and depth were less than 2% at 6 MV and 1% at 15 MV X-ray beams. Therefore, only 45 0 and 60 0 wedge data are shown. Figures 1 and 2 present the f, which shows the dependence on the field size for 6 and 15 MV X-ray beams. The wedge factors were normalized to the wedge factors at the depth of 5 cm and 10 cm of the 10x10 cm 2 for 6 and 15 MV, respectively. Figure 3 presents a comparison of depth dependence of d for 45 0 and 60 0 for the 6 MV beam. s

3 57 Figure 1. A comparison of field size dependence of relative wedge factors (f for the 45 0 (A and 60 0 (B wedge for the 6 MV beam. Wedge factors are normalized to that at the depth of 5 cm in the cm 2 field. SID=100 cm. Figure 2. A comparison of field size dependence of relative wedge factors (f for the 45 0 (A and 60 0 (B wedge for the 15 MV beam. Wedge factors are normalized to that at the depth of 10 cm in the cm 2 field. SID=100 cm. Figure 3. The relative output factors as a function of field size and depth in open and wedged fields for the 45 0 (A and 60 0 (B wedge for the 6 MV beam. Output factors are normalized to that at the depth of 5 cm in the cm 2 field. SID=100 cm. Figure 4. The relative output factors as a function of field size and depth in open and wedged fields for the 45 0 (A and 60 0 (B wedge for the 15 MV beam. Output factors are normalized to that at the depth of 10 cm in the cm 2 field. SID=100 cm.

4 58 This Figure shows that the wedge factors increased with depth. For the 15 MV beam, the s showed only a small change with increasing depth (Figure 4. Figures 5 and 6 show the OF in the wedged and open fields at 6 MV and 15 MV beams, respectively. OF wdg increased more significantly with field size than OF open at all depths. Discussion Figure 1 implies that when the field size increases from 5 5 to cm 2, the f increases by 2.7% and 5.2% for 45 0 and 60 0 wedge at 6 MV, respectively. This is consistent with the literature data, showing an increase of wedge field with field size [4, 6-9]. Also, Figure 1B shows that the wedge factor increases with field size pre-eminently at the shorter SSD due to the increase of the wedge scatter. The increase of the wedgegenerated dose with SSD can be clinically important particularly for a SID treatment. In general, field size dependence of wedge factors is attributed to the introduction of scattered-photon fluence by the wedge [9]. The number of scattered photons increases with the irradiated wedge volume that increases with the field size. As shown in Figure 2 the dependence of wedge factor on field size is 1% and 2.5% for 45 0 and 60 0 wedge, respectively, at 15 MV for large field at the depth of 15 cm. Figure 3 presents a variation of depth dependence of d for 45 0 and 60 0 for the 6 MV, and shows that the wedge factors increase with depth. McCullough et al. found that these d related to the percent depth doses [3]. In this study, the SID setup was used for the measurements. Therefore, the d related to the tissue-maximum ratio (TMR. The d accounts for TMR differences between those for wedged and open fields. Figure 3 shows that the wedge factors have a linear dependence with depths for 45 0 and 60 0 wedges at 6 MV, respectively. The linear increase in wedge factors with depth arises from beam hardening [8]. The s at a depth greater than 10 cm differs by up to 3% from that measured at d max for 45 0 wedge. The variation for depths greater than 10 cm is up to 6.5% for large field size with 60 0 wedge. This would be an error more than 6% in dose calculation, if the single wedge factor measured at d max is used. These variations are related to TMRs which increase with increasing depth and wedge angle. An increase in beam penetration for wedge fields is expected in view of the beam hardening which results from the much more attenuation of low energy photon Figure 5. A comparison of depth size dependence of relative wedge factors for the 45 0 (A and 60 0 (B wedge for the 15 MV beam. Wedge factors are normalized to that at the depth of 1.5 cm (d max in the cm 2 field. Figure 6. A comparison of depth size dependence of relative wedge factors, d for the 45 0 (A and 60 0 (B wedge for the 15 MV beam. Wedge factors are normalized to that at the depth of 3 cm (d max in the cm 2 field.

5 59 beam than the high energy one [8, 10, 11]. Depth dependence of wedge factors has been observed, particularly for the 6 MV photon beam with a large wedge angle and large field sizes. It is generally accepted that the depth dependence of wedge factors is due to beam hardening, especially for low energy megavolt photon beams such as 6 MV [8, 9]. For the 15 MV beam, s show only a small change with increasing depth (Figure 4. The variation with depth is within 1% for both 45 0 and 60 0 wedge. At a high photon energy the wedge scatter becomes more significant, thus the effects of the beam hardening are completely overcome by the wedge scatter. Thus, the variation of the wedge factor with depth at all field sizes is decreased [12]. OF wdg changes more significantly than OF open because s increase with depth. For 45 0 wedge, the OFs for open small field are either equal or larger than that of the wedged field at smaller depth. However, as the depth increases, the OF for wedged field comes bigger than that of the open field. This is especially more obvious for 60º wedged fields, where large variations of output ratios with field size are observed. Figure 5B shows that OF wdg increases more significantly with field size than OF open at all depths except for depths less than 5 cm. This increase is a result of wedge-generated dose [12]. On the other hand, the effect of beam hardening by the wedge can be seen in Figure 5. The beam output changes in the presence of wedge filter is due to changes in fluence of scattered photons reaching the isocenter [4]. At 15 MV the variation of the relative OFs is less than that of the 6 MV (Figure 6. Knöss et al. claimed that at this energy, decreasing beam hardening can be the eventual result after heavy filtration in wedge, and this was often the case for dual-energy machine equipped with one permanent install wedge (60º [11]. Conclusion In this study, field and depth dependence of motorized wedge was examined in 6 MV and 15 MV photon beam for the SID setup. For 6 MV, the dependence of 45 0 wedge on field size and depth was 3% at maximum field size and 15 cm depth, while the dependence of 60 0 wedge on field size and depth was about 5%. This variation of the wedge factor can result in large errors for many of the SID treatments, especially involving obese patients. For 15 MV, field size dependence was between 1% and 2% for 45 0 and 60 0 wedge filters, respectively. At 6 MV, it had the stronger depth dependence because of beam hardening with a large wedge angle and large field sizes. The depth dependence of the wedge factor was related to differences in percent depth dose (PDD or TMR between open and wedged fields in relation to SSD and SID setup. The results have shown that OFs increased more significantly with wedged field size than in the open fields. s have indicated that there are significant differences in PDD (or TMR and OFs between open and wedged fields, especially for 6 MV beam with large field sizes and large wedge. Thus, severe errors could be done in dose calculation if the depth dose data for open field are use for wedged field. It may not be valid to use a single wedge factor in combination with the open beam OF for dose calculation. The magnitude of error in using single wedge factor measured at reference depth for a 10x10 cm field is minimal for thin wedges. Nevertheless, for 6 MV X-rays the use of a single wedge factor introduces error of up to 6% with 60 0 wedge filter. Therefore, in the presence of a 45º and 60º wedge filters in the beam, a field size and depth dependence wedge factor or a separate OF curve for wedged beams should be used for treatment planning, especially at 6 MV photon beams. References 1. Bentel EG, Nelson CE, Noell KT (eds. Treatment planning and dose calculation in Radiation Oncology. Pergamon Press, Tatcher M. A method for varying effective angle of wedge filters. Radiology 1970; 97: McCullough EC, Gortney J, Blackwell CR. A depth dependence determination of the wedge transmission factor for 4-10 MV photon beams. Med Phys 1988; 15: Palta JR, Daftari I, Suntharalingam N. Field size dependence of wedge factors. Med Phys 1988; 15: Thomas SL. The variation of wedge factors with field size on a linear accelerator. Br J Radiol 1990; 63: Georg D, Garibaldi C, Dutreix A. Measurements of basic parameters in wedged high- energy photon beams using a miniphantom. Phys Med Biol 1997; 42: Niroomand-Rad A, Haleem M, Rodgers J, Obcemea C. Wedge factor dependence on depth and field size for various beam energies using symmetric and half-collimated asymmetric jaw settings. Med Phys 1992; 19: Popescu A, Lai K, Singer K, Phillips M. Wedge factor dependence with depth, field size, and nominal distance. A general computational rule. Med Phys 1999; 26: Tailor RC, Followill S, Hanson WF. A first order approximation of field- size and depth dependence on wedge transmission. Med Phys 1998; 25: Kalend AM, Wu A, Yoder V, Maitz A. Separation of dosegradient effect from beam-hardening effect on wedge factors in photon fields. Med Phys 1990; 1: Knöss T, Wittgren L. Which depth dose data should be used for dose planning when wedge filters are used to modify the photon beams? Phys Med Biol 1991; 36: Liu HH, Mackie TR, McCullough EC. Calculating dose and output factors for wedged photon radiotherapy fields using a convolution/superposition method. Med Phys 1997; 24:

Relative Dosimetry. Photons

Relative Dosimetry. Photons Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge

More information

Review of fundamental photon dosimetry quantities

Review of fundamental photon dosimetry quantities Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited

More information

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose

More information

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities

More information

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt

More information

CHAPTER-2. Wedge filters are the most important beam modifiers used in

CHAPTER-2. Wedge filters are the most important beam modifiers used in CHAPTER-2 2. WEDGE FLTERS 2.1. NTRODUCTON Wedge filters are the most important beam modifiers used in radiotherapy which are placed in the path of photon beams to modify their isodose diistributions. The

More information

Dosimetric Calculations. Lonny Trestrail

Dosimetric Calculations. Lonny Trestrail Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations

More information

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical

More information

1. Question Answer cgy / MU cgy / MU 2. Question Answer

1. Question Answer cgy / MU cgy / MU 2. Question Answer GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom

More information

Math Review. Overview

Math Review. Overview Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY

More information

High Speed Direct SAD Radiosurgery Beam Scanner

High Speed Direct SAD Radiosurgery Beam Scanner Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife

More information

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

The MC commissioning of CyberKnife with MLC (Tips and Tricks) The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed

More information

Introduction. Materials and Methods

Introduction. Materials and Methods STANDARD WEDGE AND TRAY TRANSMISSION VALUES FOR VARIAN, SEIMENS, ELEKTA/PHILIPS ACCELERATORS; A QUALITY ASSURANCE TOOL David S. Followill, Nadia Hernandez, and William F. Hanson Introduction The Radiological

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results

More information

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics. FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic

More information

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11 MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction

More information

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN 22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA

More information

Constancy checks of well-type ionization chambers with external-beam radiation units

Constancy checks of well-type ionization chambers with external-beam radiation units JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew

More information

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS

More information

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08. Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51

More information

A beam-matching concept for medical linear accelerators

A beam-matching concept for medical linear accelerators Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,

More information

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB

More information

Radiotherapy physics & Equipments

Radiotherapy physics & Equipments Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education

More information

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the

More information

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams Iranian Journal of Medical Physics Vol. 10, No. 3, Summer 2013, 147-155 Received: April 22, 2013; Accepted: July 21, 2013 Original Article Effective Point of Measurement in Cylindrical Ion Chamber for

More information

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the

More information

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS Acta Radiologica Oncology 23 (1984) Fusc. 6 FROM THE RADIATION THERAPY DEPARTMENT, ANTON1 VAN LEEUWENHOEK HOSPITAL, THE NETHERLANDS CANCER INSTITUTE, AMSTERDAM, THE PHILIPS MEDICAL SYSTEMS DIVISION, BEST,

More information

SnapShot IMRT with compensators and FFF beams

SnapShot IMRT with compensators and FFF beams SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy

More information

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Comparison of ionization chambers of various volumes for IMRT absolute dose verification Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical

More information

Monitor Unit Verification for Small Fields

Monitor Unit Verification for Small Fields Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the

More information

THE development of more advanced techniques in radiotherapy,

THE development of more advanced techniques in radiotherapy, 1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,

More information

TG-51: Experience from 150 institutions, common errors, and helpful hints

TG-51: Experience from 150 institutions, common errors, and helpful hints TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of

More information

Dynamic wedges dosimetry and quality control

Dynamic wedges dosimetry and quality control Original Paper Received: 2005.06.07 Accepted: 2006.02.28 Published: 2006.04.28 Authors Contribution: A Study Design B Data Collection C Statistical Analysis D Data Interpretation E Manuscript Preparation

More information

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,

More information

Zoubir Ouhib Lynn Cancer Institute

Zoubir Ouhib Lynn Cancer Institute Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy

More information

Abstract: Introduction:

Abstract: Introduction: Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston

More information

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE

More information

Clinical Implementation of Volumetric Modulated Arc Therapy

Clinical Implementation of Volumetric Modulated Arc Therapy Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy

More information

7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(-y,n)H reaction at

7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(-y,n)H reaction at Am 7Roentgenolia6:ia6oia65, 1976 CENTRAL AXIS DEPTH DOSE FOR A 2.5 MV VAN DE GRAAFF GENERATOR DAVID W. ANDERSON, DAVID E. RAESIDE, REBA I. ADAMS, AND MYRON R. GOEDE 2 ABSTRACT: Central axis percentage

More information

Scientific Journal of Nuclear Medicine & Radiation Therapy

Scientific Journal of Nuclear Medicine & Radiation Therapy Scientific Journal of Nuclear Medicine & Radiation Therapy Research Article Radiotherapy Dose Calculation Program, for Monitoring and Commissioning Treatment Planning Systems - Ioannis Vamvakas*, Stella

More information

Commissioning and periodic tests of the Esteya electronic brachytherapy system

Commissioning and periodic tests of the Esteya electronic brachytherapy system Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,

More information

Proton Therapy QA Tools

Proton Therapy QA Tools Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially

More information

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions Wendel Dean Renner a) Math Resolutions, LLC, 5975 Gales Lane, Columbia,

More information

Planning Assignment (3 field rectum)

Planning Assignment (3 field rectum) Megan Sullivan Clinical Lab Rectum April 25, 2016 Planning Assignment (3 field rectum) Use a CT dataset of the pelvis. Create a CTV by contouring the rectum (start at the anus and stop at the turn where

More information

Surface buildup dose dependence on photon field delivery technique for IMRT

Surface buildup dose dependence on photon field delivery technique for IMRT Surface buildup dose dependence on photon field delivery technique for IMRT Shigeru Yokoyama, Peter L. Roberson, Dale W. Litzenberg, Jean M. Moran, and Benedick A. Fraass Department of Radiation Oncology,

More information

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli

More information

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF

More information

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER Proven Rad-Check technology specifically designed to provide you with the ultimate in versatility and cost-effective operation. Fast and easy

More information

Exposure System Selection

Exposure System Selection Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others

More information

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,

More information

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water M R McEwen, A J Williams and A R DuSautoy Centre for Ionising Radiation Metrology National Physical

More information

Kill-painting of hypoxic tumors in charged particle therapy

Kill-painting of hypoxic tumors in charged particle therapy Supplementary Information: Kill-painting of hypoxic tumors in charged particle therapy Walter Tinganelli 1,2,, Marco Durante 1,3 *, Ryoichi Hirayama 2, Michael Krämer 1, Andreas Maier 1, Wilma Kraft-Weyrather

More information

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy University of South Florida Scholar Commons Graduate Theses and Dissertations Graduate School 7-14-2010 Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

More information

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Radiotherapy and Oncology 71 (2004) 235 241 Technical Note A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Nadine Linthout*,

More information

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation This intercomparison will be carried out under the lead of KRISS with the cooperation of ARPANSA. This comparison has been approved

More information

Investigation of Suction Process of Scroll Compressors

Investigation of Suction Process of Scroll Compressors Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering 2006 Investigation of Suction Process of Scroll Compressors Michael M. Cui Trane Jack Sauls

More information

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization Basics of Proton Therapy Proton Treatment Planning and Beam Optimization SAM Educational Session, WE D BRB 2 Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center Today s objectives Review the

More information

1. BEAM MODIFIERS IN RADIOTHERAPY

1. BEAM MODIFIERS IN RADIOTHERAPY CHAPTER-1 1. BEAM MODIFIERS IN RADIOTHERAPY INTRODUCTION In radiotherapy, tumours under varying body contour are very often encountered. In such cases, due to the irregular body surface, the dose distribution

More information

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/

More information

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61: Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca

More information

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto

More information

Digiquartz Water-Balanced Pressure Sensors for AUV, ROV, and other Moving Underwater Applications

Digiquartz Water-Balanced Pressure Sensors for AUV, ROV, and other Moving Underwater Applications Digiquartz Water-Balanced Pressure Sensors for AUV, ROV, and other Moving Underwater Applications Dr. Theo Schaad Principal Scientist Paroscientific, Inc. 2002 Paroscientific, Inc. Page 1 of 6 Digiquartz

More information

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Received: 21 November 2016 Revised: 7 April 2017 Accepted: 5 June 2017 DOI: 10.1002/acm2.12160 RADIATION ONCOLOGY PHYSICS Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Jeffrey

More information

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2017, 6, 57-72 http://www.scirp.org/journal/ijmpcero ISSN Online: 2168-5444 ISSN Print: 2168-5436 Validation of Treatment

More information

User Manual. RW3 Slab Phantom T29672 and T

User Manual. RW3 Slab Phantom T29672 and T User Manual RW3 Slab Phantom Contents Operating Manual Technical Manual Service Manual D188.131.00/07 en 2015-07 Hn General Information General Information The product bears the CE-mark "CE" in accordance

More information

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and

More information

COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC

COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC by TAN ZI XIANG Thesis submitted in fulfillment of the requirements for the degree of Master of Science (Medical Physics) February 2015 ACKNOWLEDGEMENT

More information

3D Treatment Planning and verification with hand calculations

3D Treatment Planning and verification with hand calculations 3D Treatment Planning and verification with hand calculations Randy Holt, PhD, CEO Pacific Crest Medical Physics, Inc Chico, CA Dosimetric Calculations Calibration and Measurement Conditions Patient Conditions

More information

How to Design Medical Accelerator Systems for Reliability: IBA PT System

How to Design Medical Accelerator Systems for Reliability: IBA PT System How to Design Medical Accelerator Systems for Reliability: IBA PT System Yves Jongen Founder & Chief Research Officer Ion Beam Applications s.a. Belgium 1 Outline A short introduction of a proton therapy

More information

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens The characteristic ionizing radiation response is defined by the relationship between the stimulus

More information

IPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford

IPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 48 (2003) 2929 2970 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(03)65222-X The IPEM code of practice for electron dosimetry for radiotherapy beams of

More information

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media N Tyagi 1, B H Curran, P L Roberson, J M Moran, E Acosta and B A Fraass Department

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Verification of Monitor Unit Calculations for Eclipse Treatment Planning System by in-house Developed Spreadsheet Athiyaman Mayivaganan 1*, Hemalatha

More information

Evaluation of six TPS algorithms in computing entrance and exit doses

Evaluation of six TPS algorithms in computing entrance and exit doses JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Evaluation of six TPS algorithms in computing entrance and exit doses Yun I. Tan, 1,2a Mohamed Metwaly, 3 Martin Glegg, 3 Shaun P.

More information

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry Bryan R. Muir a) and Malcolm R. McEwen Measurement Science and Standards, National Research Council of

More information

Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation

Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation By Ruth Ann M. Good, CMD. R.T. (R)(T) Virginia Commonwealth University, Richmond,

More information

TEST REPORT BEAMSCAN WITH HALCYON

TEST REPORT BEAMSCAN WITH HALCYON TEST REPORT BEAMSCAN WITH HALCYON UZ Leuven, November 20 th, 2017 + = Ideal BEAMSCAN tank dimensions to fit perfectly into the Halcyon bore Fast tank positioning using the Halcyon lasers, no electronics

More information

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning

More information

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI Electronic Oncology Systems the Management and the Safety Measures Hansen Chen, Director, TDSI Radiation Oncology Department People Place Equipments Procedures Patients Hansen Chen 2 Radiation Oncology

More information

IMRT in clinical practice at the UMC-Utrecht. Utrecht

IMRT in clinical practice at the UMC-Utrecht. Utrecht 3 IMRT in clinical practice at the UMC-Utrecht Utrecht Clinical use of IMRT Improvement of local control without increased incidence of normal tissue complications + Dose escalation in the prostate Step-and

More information

Breast Treatment System REFERENCE MANUAL

Breast Treatment System REFERENCE MANUAL Breast Treatment System REFERENCE MANUAL This manual may not be reproduced in whole or in part, by mimeograph or any other means, without the written permission of Diacor, Inc. The MammoRx Breast Treatment

More information

Perturbation Correction of a Cylindrical Thimble-type Chamber. in a Graphite Phantom for 60Co Gamma Rays*

Perturbation Correction of a Cylindrical Thimble-type Chamber. in a Graphite Phantom for 60Co Gamma Rays* Rapport BIPM-86/l4 Perturbation Correction of a Cylindrical Thimble-type Chamber in a Graphite Phantom for 60Co Gamma Rays* by C.E. de Almeida**, A.-M. Perroche-Roux and M. Boutillon Bureau International

More information

Optimization in high dose rate vaginal cylinder for vaginal cuff irradiation

Optimization in high dose rate vaginal cylinder for vaginal cuff irradiation Rep Pract Oncol Radiother, 2008; 13(1): 35-48 Original Paper Received: 2007.07.09 Accepted: 2007.12.13 Published: 2008.02.29 Authors Contribution: A Study Design B Data Collection C Statistical Analysis

More information

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS A9744593 INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSLS Ladislav Czap, Georg Matscheko and Pedro Andreo osimetry and Medical Radiation Physics Section, ivision

More information

Calculation of Trail Usage from Counter Data

Calculation of Trail Usage from Counter Data 1. Introduction 1 Calculation of Trail Usage from Counter Data 1/17/17 Stephen Martin, Ph.D. Automatic counters are used on trails to measure how many people are using the trail. A fundamental question

More information

General Accreditation Guidance. User checks and maintenance of laboratory balances

General Accreditation Guidance. User checks and maintenance of laboratory balances General Accreditation Guidance User checks and maintenance of laboratory balances January 2018 Copyright National Association of Testing Authorities, Australia 2010 All intellectual property rights in

More information

Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator

Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator James L. Bedford, a Michael D. R. Thomas and Gregory

More information

Commissioning of a new total body irradiation protocol

Commissioning of a new total body irradiation protocol University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2012 Commissioning of a new total body irradiation protocol Zoe Baldwin University

More information

Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units

Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units INTRODUCTION Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units RADIATION SAFETY GUIDE IMAGING AND MICROSCOPY FACILITY University of California, Merced Training Document

More information

Underwater measurement of photosynthetically active radiation

Underwater measurement of photosynthetically active radiation Underwater measurement of photosynthetically active radiation Mark Blonquist, Damon Nitzel, and Bruce Bugbee Apogee Instruments, Inc., Logan, Utah Introduction Quantum sensors (photosynthetically active

More information

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007 The Effect of Immobilisation Devices on Radiotherapy Dose Distributions A thesis submitted in fulfilment of the requirements for the Degree of Master of Applied Science (Medical and Health Physics) By

More information

Pressure Sensitive Paint (PSP) / Temperature Sensitive Paint (TSP) Part 1

Pressure Sensitive Paint (PSP) / Temperature Sensitive Paint (TSP) Part 1 AerE 545 class notes #12 Pressure Sensitive Paint (PSP) / Temperature Sensitive Paint (TSP) Part 1 Hui Hu Department of Aerospace Engineering, Iowa State University Ames, Iowa 50011, U.S.A Introduction

More information

The dose distribution of medium energy electron boosts to the laryngectomy stoma

The dose distribution of medium energy electron boosts to the laryngectomy stoma JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 1, WINTER 2001 The dose distribution of medium energy electron boosts to the laryngectomy stoma Ellen D. Yorke* Memorial Sloan-Kettering Cancer

More information

OUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD

OUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD RAPHEX 2001 T1. All of the following are true regarding Percentage Depth Dose (PDD) except: A. Increases with increasing energy. B. Depends on field size. C. Is the dose at depth expressed as a % of the

More information

WIND-INDUCED LOADS OVER DOUBLE CANTILEVER BRIDGES UNDER CONSTRUCTION

WIND-INDUCED LOADS OVER DOUBLE CANTILEVER BRIDGES UNDER CONSTRUCTION WIND-INDUCED LOADS OVER DOUBLE CANTILEVER BRIDGES UNDER CONSTRUCTION S. Pindado, J. Meseguer, J. M. Perales, A. Sanz-Andres and A. Martinez Key words: Wind loads, bridge construction, yawing moment. Abstract.

More information

MIT Analytical X-Ray Safety Policies and Procedures

MIT Analytical X-Ray Safety Policies and Procedures 5th Edition 2014 MIT Analytical X-Ray Safety Issued by: MIT Radiation Protection Committee Contents 1.0 Program Overview...1 2.0 Purpose...1 3.0 Scope and Application...2 4.0 Radiation Protection Program

More information

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Commissioning and quality assurance for the treatment delivery components of the AccuBoost system Ileana Iftimia, 1,2a Mike Talmadge,

More information

The Gas Attenuator of FLASH

The Gas Attenuator of FLASH INTRODUCTION The Gas Attenuator of FLASH K. Tiedtke, N. von Bargen, M. Hesse, U. Jastrow, U. Hahn The experimental hall of the FLASH user facility is located approximately 30m behind the last dipole magnet

More information

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy COMPASS 3D IMRT Verification in Patient Anatomy AK IMRT Bamberg April 2010 Dr. Lutz Müller Patient-specific Verification? CT Treatment unit Gantry =0 Treatmt unit Patient couch X-ray beam Patient couch

More information

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS Frank M. Gavila F&J SPECIALTY PRODUCTS, INC. 404 Cypress Road Ocala, Florida, USA 3447 fandj@fjspecialty.com INTRODUCTION The calibration

More information