1 A LAZAR LABORATORIES HANDBOOK DISSOLVED OXYGEN PRIMER Answers to your basic dissolved oxygen measurement questions and helpful hints on how to optimize measuring dissolved oxygen. Copyright Lazar Research Laboratories, Inc

2 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer 2 About Lazar Research Laboratories, Inc. Lazar Research Laboratories, Inc. have been coaching researchers on measuring dissolved oxygen for over 27 years. Our knowledge has helped researchers in all scientific arenas including biotechnology, biomedical research, microchemistry, chemical engineering, environmental science, geology and more. Our family of standard and micro dissolved oxygen electrodes are the premier product for measuring DO in various samples in all sizes down to one micro liter. Free Consultation Got any dissolved oxygen questions? Call us at ( international), fax us at , or us at service@lazarlab.com and we will put our 24 years of experience in the dissolved oxygen field at your disposal. There is no cost or obligation for this service. Contact Lazar Research Laboratories, Inc. service@lazarlab.com Website: Fax: Phone: (toll free in the U.S. only) Address: Lazar Research Laboratories, Inc. 731 North Labrea Avenue, Building 5 Los Angeles, California U.S.A.

3 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer 3 TABLE OF CONTENTS 1. Comparing the dissolved oxygen electrode to the Winkler colorimetric method 2. How an oxygen electrode works A. Clark polarographic oxygen electrode B. Galvanic oxygen electrode 3. Zeroing and calibrating an oxygen electrode 4. Effect of temperature, atmospheric pressure, salinity 5. Effect of stirring 6. Oxygen electrode maintenance 7. Oxygen electrode that can be used with ph/mv meter, chart recorder, and PC 8. Standard size oxygen electrode 9. Micro oxygen electrode 10. Needle oxygen electrode 11. Micro flow through oxygen electrode 12. Micro biological po2 monitoring system 13. Oxygen electrodes for fermentors and bioreactors 14. Ordering contact information

4 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer 4 1. COMPARING THE OXYGEN ELECTRODE TO THE WINKLER COLORIMETRIC METHOD The Winkler method, which involves measuring a color change in a sample when oxygen reacts with a special reagent, is inherently inaccurate for several reasons. The method involves adding a special reagent and waiting for the oxygen-reagent reaction to come to completion resulting in a yellowish tint. The amount of dissolved oxygen is related to the intensity of the yellow color. The color can be quantified by using a color comparison chart with the naked eye or by measuring the color with a colorimeter. The results can be affected by the subjective color comparison from one person to the next when using a color comparison chart, or by residual color or turbidity in the sample which affect the colorimeter results. The tedious setup procedure using the special reagent can also introduce errors due to user carelessness. Besides being inaccurate, the Winkler method is often very tedious especially when doing a large series of samples. Another factor which adds to the inaccuracy of this method is that the sample is first taken from the source (river, lake, etc.) and put into a sample container which is capped to prevent oxygen in the sample from escaping or oxygen from the atmosphere from entering the sample. The sample then has to be processed with the Winkler reagent prior to measuring the yellow tint. The removal of the water sample from the source prior to measurement can cause a rapid change in the oxygen level due to exposure to the atmosphere giving another cause for inaccuracy. With the electrode method, dissolved oxygen is measured directly at the source without having to take a sample and risking that the sample will either absorb or lose oxygen from the atmosphere. Since the oxygen electrode measurement does not depend on any change in color, this measurement is free of any interferences from other color bearing components in the water or from any effect of turbidity. Furthermore, the electrode method does not involve any introduction of reagents which are dependent on operator error and reagent aging. The readout from the oxygen meter is a quantitative digital concentration reading as opposed to a qualitative color comparison. The Lazar Research Standard Dissolved Oxygen Probe

5 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer 5 2. HOW AN OXYGEN ELECTRODE WORKS All oxygen electrodes are based on a simple electrochemical reaction that occurs in water when oxygen is present. The oxygen reacts with water at an electrode surface producing hydroxyl ion. A. THE CLARK POLAROGRAPHIC OXYGEN ELECTRODE Oxygen is reduced at a cathode by receiving electrons and is converted into hydroxyl ion. This reaction occurs at the cathode and is the basis of the Clark type oxygen electrode. In the Clark electrode the cathode is made out of platinum and the anode is made out of silver. The anode is necessary to complete the electrical circuit so that electrons are available at the cathode to reduce the oxygen. In order to prevent other electrochemical reactions from interfering with the oxygen reaction and giving a false signal at the cathode, the anode and cathode are usually inside an electrode body which is covered with a semipermeable membrane. The job of the membrane is to prevent most components present in water except for oxygen from entering the electrode interior. The membrane is what gives the oxygen electrode great selectivity for oxygen with hardly any interferences. The Clark electrode is based on a polarographic principle where the current running between the cathode and anode is generated by imposing a known voltage difference (potential) between these two electrodes. The consumption of electrons at the cathode during the oxygen reaction, results in a small current between the anode and the cathode. This current is directly related to the concentration of dissolved oxygen in the water sample being measured. The current which is in the low nanoampere range (1 billionth of an amp) is converted into a voltage output by a special current amplifier and is then read using a digital voltmeter or chart recorder. The voltage output can be calibrated to directly indicate the concentration of dissolved oxygen in the sample. B. THE GALVANIC OXYGEN ELECTRODE Another common type of oxygen electrode, called the galvanic oxygen electrode uses a similar principle where oxygen is reduced at a cathode, but instead of having an electronic circuit produce a voltage difference between the cathode and anode, the two electrodes have dissimilar electrode potentials which produces the voltage (potential) difference. This is similar to the operation of a standard battery. The cathode in this case is usually silver or an alloy of silver, and the anode is composed of lead. The current produced at the cathode is identical to the current produced at the Clark type electrode. The current is again amplified and converted to a voltage to be displayed on a digital voltmeter or chart recorder.

6 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer 6 3. ZEROING AND CALIBRATING AN OXYGEN ELECTRODE In an ideal electrode system for measurement of dissolved oxygen the residual current present at the cathode should be zero when the dissolved oxygen concentration is zero. For practical oxygen electrodes, either Clark polarographic type or galvanic there is always a small amount of residual current present even when the dissolved oxygen level is zero. In order to compensate for this residual current, there is an adjustment available on most oxygen amplifiers (which amplify the current signal and convert it to a voltage output) where the zero baseline can be adjusted so when the oxygen level is zero the output from the amplifier will also read zero. The zero adjustment is done by dipping the oxygen electrode into an environment which is totally devoid of oxygen and adjusting the amplifier output to zero. The calibration control on an oxygen meter (CAL or span) adjusts the amplifier output to correspond to the actual value of dissolved oxygen in a standard being measured. This calibration is accomplished by exposing the oxygen electrode to a known value of dissolved oxygen and then adjusting the CAL control so that oxygen value is read at the output of the amplifier or the digital voltmeter connected to the amplifier. A simple practical method to calibrate an oxygen electrode is to expose it to atmospheric oxygen or a solution of water in equilibrium with atmospheric oxygen. The calibration point is then set to correspond to the known saturation value of oxygen dissolved in water which is in equilibrium with air. There are tables available which give the saturation value in air for various temperatures and atmospheric pressures.

7 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer 7 4. EFFECT OF TEMPERATURE, ATMOSPHERIC PRESSURE, AND SALINITY The effect of temperature is most important because the reaction which takes place at the cathode where oxygen reacts to produce a current is directly dependent on the temperature of the water which is being measured. Electronic amplifiers which convert the current signal to a voltage output can have an adjustment potentiometer which is set to correspond to the temperature of the water being measured (temperature needs to be measured separately). An alternate approach which is usually done in most modern oxygen meters is to have an automatic temperature adjustment where a thermister is embedded at the tip of the oxygen probe which changes resistance with temperature. The thermister is connected to the amplifier circuit and changes the output signal corresponding to the temperature of the sample. Atmospheric pressure is important, but to a lesser degree than temperature, because changes in atmospheric pressure will effect the amount of oxygen which is dissolved in the water. The effect of atmospheric pressure can be compensated for in the initial calibration of the oxygen electrode-amplifier where the saturation of oxygen is established based on well defined tables which give different saturation values as a function of atmospheric pressure. When measuring dissolved oxygen in sea water the salinity of the water also becomes a factor. Salinity level in water will affect the amount of oxygen which can dissolve in the water.

8 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer 8 5. EFFECT OF STIRRING Stirring can have an effect on the dissolved oxygen reading with a dissolved oxygen electrode. The reason for this is that a very small amount of oxygen is actually consumed by the electrode as it measures the concentration of dissolved oxygen. This oxygen consumption produces a depletion of oxygen at the membrane surface as compared to the oxygen level in the water sample. In other words, the oxygen concentration found at the membrane which is in direct contact with the oxygen electrode can be lower than the oxygen level in the sample itself. Stirring the sample reduces this effect whereby the oxygen level at the membrane now becomes equal to that of the sample. Stirring can be accomplished by using a stirrer or by measuring oxygen in water which is moving or flowing. Often the water sample being measured is not totally still but there is turbulence present. The turbulence can act as a stirring mechanism so that external stirring is not needed. Another instance of a natural stirring mechanism is measuring oxygen in a pipe where liquid is flowing by. In instances where stirring cannot be accomplished such as in very small samples, the effect of oxygen depletion can be compensated for in the calibration procedure.

9 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer 9 6. OXYGEN ELECTRODE MAINTENANCE In order to get the most accurate readings with an oxygen electrode certain maintenance procedures need to be carried out. The semipermeable membrane on the oxygen electrode is usually constructed of a fluoropolymer and is susceptible to tearing, pinholes, scratching etc. It is important to periodically check the membrane visually to make sure that the membrane is still in good condition. If any type of damage is seen, it is advisable to replace the membrane. A membrane which is leaking can produce very erratic oxygen readings. Another expendable item in oxygen electrodes is the internal filling solution. This solution is an elecrolyte which maintains solution conductivity between the cathode and anode. This solution should be replaced periodically in order to maintain the oxygen electrodes at optimal condition. Some manufacturers of oxygen electrodes also recommend a periodic polishing of the cathode using a very fine emery paper. The polishing removes any metal oxide films which can reduce the overall signal output of the electrode. When and if to polish is something to ask the manufacturer of the oxygen electrode. Recalibration of the oxygen electrode should also be done periodically to make sure that the electrode is still within calibration. Calibration intervals depend somewhat on the type of oxygen electrode and the application it is being used for. In most instances an air calibration can be done which does not require any special reagents or other items. Zeroing of the oxygen electrode is usually done at the factory for most oxygen electrodes and needs to be done by the user on an infrequent basis. Zeroing is usually done by adding a chemical such as sodium bisulfite to water. The sodium bisulfite reacts with oxygen depleting the water of any oxygen. Another convenient method for zeroing which can be done in a laboratory is to use a gas like nitrogen (N2) which is bubbled through a water sample removing the dissolved oxygen by gas displacement. Maintenance procedures for portable oxygen meters should include a check of the batteries. If the batteries read a low voltage they should either be replaced or recharged (for rechargeable batteries). Low battery voltage can give erroneous and erratic oxygen readings.

10 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer OXYGEN ELECTRODES WHICH CAN BE USED WITH A ph/mv METER OR CHART RECORDER Oxygen electrodes requires an amplifier circuit to read oxygen concentration and cannot be used directly with a regular ph/mv meter or chart recorder. Most manufacturers incorporate the oxygen electrode as part of an oxygen instrument which often costs many times the price of a ph/mv meter. Oxygen measurements are therefore normally restricted to users with a large enough budget to be able to afford a dedicated oxygen instrument which incorporates the oxygen electrode, amplifier circuit, and readout device. Lazar Research Laboratories, Inc. has developed a better solution to the high expense of dissolved oxygen meters. We have included the amplifier circuit directly with the dissolved oxygen probe, and made the output compatible with any standard ph/mv meter or chart recorder. Instead of using an expensive oxygen readout device, you can use the oxygen electrode with the most common instrument in any laboratory, namely a ph meter. Using the oxygen electrode now becomes no more complicated or expensive than using a standard ph electrode. Our family of oxygen electrodes can all be used with any laboratory or portable ph/mv meter, various chart recorders, and even directly with a PC having an A to D board. Given below is a description of the Lazar dissolved oxygen family which includes a standard size oxygen electrode for measuring water quality in lakes, rivers, the ocean etc, and also measuring oxygen in BOD bottles directly. Other oxygen electrodes includes a micro model for measuring dissolved oxygen in very small samples down to 20 microliters including 96 well plates, micro centrifuge tubes, petri dishes etc; a needle oxygen probe for measuring tissue samples; and a micro flow through oxygen electrode for measuring oxygen in micro flowing streams in HPLC, FIA, and organ perfusion. Lazar also sells a broad selection of dissolved oxygen electrodes for fermentation and bioreactors.

11 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer STANDARD SIZE DISSOLVED OXYGEN PROBES FEATURES USE WITH ANY ph METER. Probe can be plugged directly into any standard ph meter for dissolved oxygen measurements. USE WITH PC COMPUTER. Use with Lazar low cost datalogger for monitoring and recording with your PC. COVERS FULL DO RANGE. Electrode can measure dissolved oxygen from 0 to 20 ppm or 0 to 150% saturation. ELIMINATES WINKLER METHOD. Probe measures dissolved oxygen directly without involved wet chemistry or titrations needed with the Winkler method. BOD MEASUREMENTS. BOD probe fits directly into BOD bottle for 5 day or continuous BOD measurements. PORTABLE. On-the-spot field measurements of dissolved oxygen can be made without a trip back to the laboratory. RAPID RESPONSE. Typical response time at room temperature is 12 seconds. USE FOR BOTH BOD MEASUREMENTS IN THE LAB OR FOR FIELD USE. SPECIFICATIONS Oxygen Range: 0 to 20 ppm dissolved oxygen Accuracy: +/- 0.2 ppm at calibration temperature Temperature Range: 0 to 50 degrees C Response Time: Response time at 25 C is 12 seconds for 90% of reading. Connector: Available in either U.S. standard or BNC connector (please specify when ordering). Power: 9 volt disposable battery or 110/220 AC. Sample Size: Minimum sample size is about 2 ml.

12 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer 12 ORDERING INFORMATION DO-166 OXYGEN ELECTRODE including spare membranes and bottle of filling solution. Can be used for both BOD and field measurements. Cat. No. DO with 4 ft. cable Cat. No. DO with 15 ft. cable Cat. No. DO with 25 ft. cable Cat. No. SM-166 Spare DO Membranes- Package of 10 membranes Cat. No. SF-166 Filling Solution (2 ounce bottle of filling solution)

13 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer MICRO DISSOLVED OXYGEN ELECTRODE 96 well plates micro test tubes serum cups capillary tubes NMR tubes microcentrifuge tubes ADVANTAGES MEASURE ph IN 1/5 OF A DROP. This electrode enables you to measure samples as small as 10 microliters (1/5 of a drop). USE WITH ANY ph METER. The micro po2 electrode utilizes a self contained amplifier circuit which makes it useable with any standard ph/mv meter for dissolved oxygen measurements. The po2 electrode can be plugged directly into any ph meter and is available with all standard ph meter connectors including BNC, U.S. standard, Radiometer(R), and Metrohm (R). USE WITH CHART RECORDER. Output voltage of 0 to 2 volts DC allows connection to any standard chart recorder for continuous monitoring of po2 level. USE WITH PC COMPUTER. Can be used with any PC that has an A to D data card, or use the Lazar low cost A to D datalogger which plugs into the serial port of a PC. PORLAROGRAPHIC PRINCIPLE. The po2 electrode works on polarographic principle where oxygen is reduced at a tiny cathode sensor which is covered with a semipermeable membrane. EXTREMELY SMALL SENSOR SIZE. Tip diameter of the micro po2 electrode is only 2 mm. LOW IMMERSION DEPTH. Immersion depth for the micro po2 electrode is less than 1 mm. *CALIBRATE IN CONVENIENT UNITS. The micro po2 probe can be calibrated in mm Hg po2 partial pressure, % oxygen, or ppm dissolved oxygen. BROAD po2 CONCENTRATION RANGE. The micro po2 has a range from 0 to 100% oxygen or 0 to 760 mm Hg po2 partial pressure. LOW O2 DETECTION LIMIT. The micro po2 has a detection limit of 1 mm Hg po2 partial pressure. BROAD TEMPERATURE RANGE. Temperature range is 0 to 60 oc.

14 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer 14 APPLICATION AREAS Biotechnology Biological sciences Medicine Physiology Environmental studies Micro chemical analysis Forensic science Pharmaceuticals Radioactive samples NMR samples ORDERING INFORMATION Cat. No. DO-166MT-1 micro po2 probe. Comes complete with po2 electrode, internal amplifier circuit, connector cable to ph meter or PC, spare membranes, and spare filling solution. Specify type of connector needed for ph meter being used (BNC, US Standard, Radiometer(R), Metrohm(R), other) when ordering. Cat. No. SM-166 Spare DO Membranes - Package of 10 membranes. Cat. No. SF-166 Filling Solution - 2 ounce bottle of filling solution.

15 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer MICRO NEEDLE po2 PROBE YOU CAN NOW MEASURE po2 IN TISSUE SAMPLES The needle probe inserts directly into tissue samples for oxygen concentration measurements in various kinds of tissues including various plant and animal tissues and cell cultures. USE WITH ANY ph METER. The needle po2 electrode utilizes a self contained amplifier circuit which makes it useable with any standard ph/mv meter for dissolved oxygen measurements. The po2 needle probe can be plugged directly into any ph meter and is available with all standard ph meter connectors including BNC, U.S. standard, Radiometer(R), and Metrohm (R). USE WITH CHART RECORDER. Output voltage of 0 to 2 volts DC allows connection to any standard chart recorder for continuous monitoring of po2 level. USE WITH PC COMPUTER. Can be used with any PC that has an A to D data card, or use the Lazar low cost A to D datalogger which plugs into the serial port of a PC. PORLAROGRAPHIC PRINCIPLE. po2 needle probe works on polarographic principle where oxygen is reduced at a tiny cathode sensor housed inside a hypodermic needle. po2 DEPTH PROFILES. Determine po2 as a function of tissue depth. SMALL SENSOR SIZE. The po2 needle diameter is 16 gauge. LONG NEEDLE LENGTH. Needle length is 38 mm giving it ample size to insert into thick tissue samples. CALIBRATE IN CONVENIENT UNITS. The micro po2 probe can be calibrated in mm Hg po2 partial pressure, % oxygen, or ppm dissolved oxygen. BROAD po2 CONCENTRATION RANGE. The micro po2 has a range from 0 to 100% oxygen or 0 to 760 mm Hg po2 partial pressure. LOW O2 DETECTION LIMIT. The micro po2 has a detection limit of 1 mm Hg po2 partial pressure. BROAD TEMPERATURE RANGE. Temperature range is 0 to 50 oc. ORDERING INFORMATION Cat. No. DO-166NP po2 needle probe. Comes complete with needle po2 electrode, electronic amplifier circuit, and connector cable to fit ph meter. Specify type of connector needed for ph meter being used (BNC, US Standard, Radiometer(R), Metrohm(R), other) when ordering.

16 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer MICRO FLOW THROUGH po2 AND pco2 ELECTRODES FEATURES USE IN BIOREACTOR, CELL CULTURE, AND ORGAN PERFUSION. The po2 micro flow through system (180 microliter internal volume) includes a special geometry dissolved oxygen electrode housed within a micro flow cell and is ideally suited for monitoring micro flowing streams in biotechnology, organ perfusion, cell culture, and HPLC. The po2 probe includes an internal preamplifier making it useable with any ph/mv meter. A micro stir bar is also included in the flow cell. The pco2 micro flow through system has a flow cell volume of only 50 microliters. ATTACHES DIRECTLY TO SILASTIC TUBING. End fittings on po2 & pco2 probes attach directly to standard micro silastic tubing. USE WITH ANY ph METER. The flow through po2 and pco2 electrodes plug directly into any standard ph/mv meter. Electrodes are available with all standard ph meter connectors including BNC, U.S. standard, Radiometer(R), and Metrohm(R). USE WITH CHART RECORDER. Output voltage from 0 to 2 volts allows connection of both po2 and pco2 flow through electrodes to any standard chart recorder for continuous monitoring of po2 and pco2 levels. USE WITH PC COMPUTER. Can be used with any PC that has an A to D data card, or use the Lazar low cost A to D datalogger which plugs into the serial port of a PC. ORDERING INFORMATION Model DO-166FT micro flow through po2 electrode. Comes complete with micro flow through po2 probe, micro flow through cell with input and output ports, silastic tubing for port connections, electronic amplifier circuit, and connector cable to fit ph meter. Specify type of connector needed for ph meter being used (BNC, US Standard, Radiometer(R), Metrohm(R), other) when ordering. Model GS-136CO-1 micro flow through pco2 electrode. Comes complete with micro flow through pco2 probe, micro flow through cell with input and output ports, silastic tubing for port connections, and connector cable to fit ph meter. Specify type of connector needed for ph meter being used (BNC, US Standard, Radiometer(R), Metrohm(R), other) when ordering. Model PP-166 peristaltic pump. A variable speed peristaltic pump with flow rates from 0 to 25 ml/min.

17 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer COMPLETE MICRO BIO po2 MONITORING SYSTEM You can now continuously monitor oxygen uptake or oxygen evolution of cells or tissue samples. Output from oxygen probe can be fed directly to chart recorder for plotting oxygen uptake or evolution curves. SYSTEM INCLUDES THE FOLLOWING: 1. Specially designed micro po2 electrode with built in amplifier. 2. micro chamber with variable volume from 200 to 1000 microliters 3. micro stir bar for stirring sample (when needed) 4. Miniature magnetic stirrer 5. Sealed septum sample injection system using syringe 6. Digital ph/mv meter for monitoring po2 and pco2 level 7. Optional pco2 electrode from monitoring pco2 level ORDERING INFORMATION Model RMS-166 bio po2 monitoring system including micro po2 electrode, micro variable volume sample chamber, septum sample injection system, digital ph/mv meter, micro magnetic stir bar, and miniature magnetic stirrer. Model GS-136-CO pco2 electrode designed to be used with sample chamber above.

18 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer OXYGEN ELECTRODES FOR FERMENTATION AND BIOREACTORS Please refer to the following web page for our extensive line of stainless steel and glass dissolved oxygen electrodes for fermentation and bioreactors.

19 Lazar Research Laboratories, Inc. Dissolved Oxygen Measurement Primer HOW TO PLACE AN ORDER OR REQUEST PRICING INFORMATION by phone: in the U.S International by by fax: by mail: Lazar Research Laboratories, Inc. 731 North LaBrea Avenue, Suite 5 Los Angeles, CA U.S.A. Note: Specifications and prices subject to change without notice To get a brochure on the Lazar micro ion electrodes contact us via at service@lazarlab.com giving us your fax number and request brochure FIS. METHODS OF PAYMENT net 30 to companies with recognized credit rating Visa or Mastercharge COD, check, money order Note: Specifications and prices subject to change without notice To get a full brochure on the micro ph electrode including pictures contact Lazar Research Labs, Inc. at service@lazarlab.com giving your fax number and asking for brochure FPHR.

STERILIZABLE DISSOLVED OXYGEN SENSORS

STERILIZABLE DISSOLVED OXYGEN SENSORS STERILIZABLE DISSOLVED OXYGEN SENSORS USER MANUAL Table of Contents 1. Introduction 2. Operation 3. Calibration 4. Electrode Verification 5. Mounting 6. Maintenance 7. Storage 8. Instructions for Exchange

More information

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE GENERAL INFORMATION This family of dissolved oxygen electrodes is of the steam sterilizable polarographic type, designed to be interchangeable with NBS

More information

12mm Dissolved Oxygen Sensor Care and Use Instructions

12mm Dissolved Oxygen Sensor Care and Use Instructions 12mm Dissolved Oxygen Sensor Care and Use Instructions How Dissolved Oxygen Sensors Work Your Dissolved Oxygen (DO) Probe is a galvanic electrochemistry device; i.e. it does not require power from your

More information

Dissolved Oxygen Guide

Dissolved Oxygen Guide Educat i onser i es Di ssol vedoxygengui de Dissolved Oxygen Guide Introduction Dissolved oxygen probes provide a convenient approach to essentially direct measurement of molecular oxygen. The membrane

More information

PRODUCT SHEET. Order probe only as RXPROBE02

PRODUCT SHEET. Order probe only as RXPROBE02 SS69L DISSOLVED OXYGEN PROBE TRANSDUCER Order probe only as RXPROBE02 Order interface only as BSL-TCI16 SS69L Components The SS69L transducer measures dissolved oxygen. The SS69L includes a dissolved oxygen

More information

LogIT Dissolved Oxygen Probe

LogIT Dissolved Oxygen Probe LogIT Dissolved Oxygen Probe INTRODUCTION The LogIT Dissolved Oxygen Probe (DO2) has been designed especially for LogIT and is easy to set up and use. It features simple single control calibration and

More information

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage Technical Note ISE-730 and DO2-100 Overview In 1954, Dr. Leland Clark invented the first membrane-covered electrode designed to measure the concentration of oxygen in blood, or solution. This electrode

More information

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +. The Oxygen sensor is a maintenance- free galvanic oxygen electrode capable of measuring Oxygen in solutions. With built-in temperature compensation, this sensor provides an easy way for students to receive

More information

SS32L Dissolved O 2 Probe

SS32L Dissolved O 2 Probe BIOPAC WWW.biopac.com Systems, Inc. Application Note PH-185 SS32L Dissolved O 2 Probe SS32L Dissolved O 2 Probe BIOPAC Software SS32L Specifications BSL PRO v. 3.6.6 BIOPAC Hardware SS32L Dissolved O 2

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No.. Innovative Instruments, Inc. http://www.2in.com Dissolved Oxygen Meter ino2 Instruction Manual Serial No.. Innovative Instruments, Inc. 8533 Queen Brooks Ct. Tampa, FL 33637. USA Phone: 813-727-0676. Fax:

More information

COM280. Technical Information. Oxygen measurement Battery-driven pocket meter for dissolved oxygen measurement

COM280. Technical Information. Oxygen measurement Battery-driven pocket meter for dissolved oxygen measurement Technical Information COM280 Oxygen measurement Battery-driven pocket meter for dissolved oxygen measurement Application COM280 is a pocket meter for measuring dissolved oxygen. It is suitable for the

More information

NEULOG OXYGEN LOGGER SENSOR GUIDE

NEULOG OXYGEN LOGGER SENSOR GUIDE NeuLog oxygen logger sensor NUL-205 The oxygen sensor can be used for any science experiment or activity where oxygen levels, dissolved or gaseous, are required. Some fields of study include Chemistry,

More information

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante810 Benchtop Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante810 benchtop dissolved oxygen

More information

Introduction The Smart Q Oxygen Adaptor enables an Oxygen electrode to be connected to an EasySense unit.

Introduction The Smart Q Oxygen Adaptor enables an Oxygen electrode to be connected to an EasySense unit. The Oxygen Adaptor & Electrode Arbor Scientific Oxygen Adapter (Product No PC-3130) Comes with Oxygen Electrode (PC-3131) 3 membrane modules 50 ml electrolyte Polishing paper for cathode Oxygen Electrode

More information

Oxygen Sensor with Electrode

Oxygen Sensor with Electrode imagine explore learn Oxygen Sensor with Electrode Product Number: ENOXY-A222 Overview The Oxygen sensor is a maintenance-free galvanic oxygen electrode, capable of measuring Oxygen both in the air and

More information

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Bante820 Portable Dissolved Oxygen Meter Instruction Manual Bante820 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante820 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante820 portable dissolved oxygen

More information

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter. Instruction Manual HI 9142 Portable Waterproof Dissolved Oxygen Meter www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully

More information

DISSOLVED OXYGEN SENSOR BT34i

DISSOLVED OXYGEN SENSOR BT34i DISSOLVED OXYGEN SENSOR BT34i USER S GUIDE CENTRE FOR MICROCOMPUTER APPLICATIONS http://www.cma-science.nl Short description The CMA Dissolved Oxygen (DO) sensor BT34i measures the concentration of dissolved

More information

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Bante821 Portable Dissolved Oxygen Meter Instruction Manual Bante821 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante821 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante821 portable dissolved oxygen

More information

Instruction Manual HI Portable Dissolved Oxygen Meter. DO meter HI 8043 STB C OFF O 2 ZERO SLOPE C. ins rum n s

Instruction Manual HI Portable Dissolved Oxygen Meter. DO meter HI 8043 STB C OFF O 2 ZERO SLOPE C. ins rum n s Instruction Manual HI 8043 Portable Dissolved Oxygen Meter DO meter HI 8043 ZERO SLOPE C ins rum n s This Instrument is in Compliance with the CE Directives Dear Customer, Thank you for choosing a Hanna

More information

Aqua-DP. Aqua-DP Waterproof Dissolved Oxygen - ph -mv Temperature Meter. TPS Australia - Quality hand-made instruments since 1968.

Aqua-DP. Aqua-DP Waterproof Dissolved Oxygen - ph -mv Temperature Meter. TPS Australia - Quality hand-made instruments since 1968. TPS Australia - Quality hand-made instruments since 1968. Waterproof Dissolved Oxygen - ph -mv Temperature Meter Dual readout of Dissolved Oxygen and ph/mv Waterproof to IP65 Easy to use One-touch calibration

More information

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual BANTE INSTRUMENTS CO., LTD DOscan10 Pocket Dissolved Oxygen Tester 1 Thank you for selecting the DOscan10 pocket dissolved oxygen tester. This

More information

Optical Dissolved Oxygen Meter

Optical Dissolved Oxygen Meter Optical Dissolved Oxygen Meter HI764113 Rugged Optical Dissolved Oxygen Probe for Fresh and Saltwater Applications Digital, weighted probe No membranes No electrolytes No oxygen consumption No flow dependence

More information

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter INSTRUCTION MANUAL Loop Powered Dissolved Oxygen Transmitter Contents Contents...2 Introduction...3 Range and Display Settings...4 Installation...5 Electrical Installation...6 Calibration...7 EasyCal...8

More information

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111 INTRODUCTION Water vapor in natural gas has more than a substantial effect on the quality of the gas stream. Without quality measurement of water vapor the gas is basically not saleable. Contracts are

More information

QUICK START GUIDE TO DRINKING WATER MONITORING

QUICK START GUIDE TO DRINKING WATER MONITORING APPENDIX F QUICK START GUIDE TO DRINKING WATER MONITORING Introduction: This Quick Start guide is designed to allow users who are already familiar with the basic use of YSI 6-series sondes for surface

More information

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter Overview C % ppm 4~20mA Analog Output ppb Power Supply 100~240 VAC mg/l Dimensions 96 x 96 x 132mm RS-485 Digital Output (for DC-5310-RS only)

More information

MODEL GT820 OXYGEN SENSOR

MODEL GT820 OXYGEN SENSOR INSTRUCTION MANUAL MODEL GT820 OXYGEN SENSOR 70046 The information and technical data disclosed by this document may be used and disseminated only for the purposes and to the extent specifically authorized

More information

Improve Process Reliability

Improve Process Reliability Convenient Conductivity Calibration to Meet Pharmacopeia Regulations Pharmacopeias require that conductivity measurement systems be periodically calibrated to ensure their precision. The range of accessories

More information

D-OptoLogger Dissolved Oxygen Logger Operation Manual

D-OptoLogger Dissolved Oxygen Logger Operation Manual D-OptoLogger Dissolved Oxygen Logger Operation Manual Software version 3.4 Contents 1. Introduction..... 1 Optical Sensor Technology.....1 Features...... 1 2. D-OptoLogger Specifications.2 Specifications...

More information

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA Software version 2.6 Table of contents 1 Introduction...3 1.1 Optical Sensor Technology...3 1.2 Installation...4 1.3 Operation...5

More information

Optical Dissolved Oxygen Meter

Optical Dissolved Oxygen Meter Optical Dissolved Oxygen Meter HI764113 Rugged Optical Dissolved Oxygen Probe for Fresh and Saltwater Applications Digital probe Fast and stable readings No membranes Not affected by sunlight No electrolytes

More information

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving Technical Information Oxymax H COS21 Sensor with long-term stability for frequent sterilization and autoclaving Application Process control in enzyme production Control of culture growth Biotechnological

More information

Dissolved Oxygen and measurement possibilities. Berno Lüpkes, 15 th March 2017

Dissolved Oxygen and measurement possibilities. Berno Lüpkes, 15 th March 2017 Dissolved Oxygen and measurement possibilities Berno Lüpkes, 15 th March 2017 Content 1. Introduction to Dissolved Oxygen 2. Amperometric measurement principle 3. Optical measurement principle 4. Optical

More information

D-Opto Dissolved Oxygen Sensor Operation Manual

D-Opto Dissolved Oxygen Sensor Operation Manual D-Opto Dissolved Oxygen Sensor Operation Manual Software version 2.6 Contents 1. Introduction..... 1 Optical Sensor Technology.... 1 Features...... 1 2. D-Opto Specifications. 2 Specifications... 2 3.

More information

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax DH-40i, Rev. 04.03.02 Micro-Incubation Platform For RC-40 Chambers Model DH-40i 1125 Dixwell Avenue, Hamden, CT 06514 (800) 599-4203 / (203) 776-0664 (203) 776-1278 - fax DH-40i, Rev. 04.03.02 Table of

More information

BI-680 Online Dissolved Oxygen Controller Instruction Manual

BI-680 Online Dissolved Oxygen Controller Instruction Manual BI-680 Online Dissolved Oxygen Controller Instruction Manual BANTE INSTRUMENTS CO., LTD BI-680 Online Dissolved Oxygen Controller 1 Introduction Thank you for selecting the BI-680 online dissolved oxygen

More information

PRECISION DISSOLVED OXYGEN MEASUREMENT SYSTEMS

PRECISION DISSOLVED OXYGEN MEASUREMENT SYSTEMS PRECISION DISSOLVED OXYGEN MEASUREMENT SYSTEMS Strathkelvin Instruments precision dissolved oxygen measurement systems and accessories are successfully operating in teaching and research laboratories worldwide.

More information

Pool and Spa. Water Testing

Pool and Spa. Water Testing Pool and Spa Water Testing Pocket Digital Testers for Pools HI-701 Checker: Free Chlorine Hand-held Colorimeter The HI-701 is a simple, accurate and cost effective way to measure free chlorine. Free Chlorine:

More information

Technical Data Sheet MF010-O-LC

Technical Data Sheet MF010-O-LC Technical Data Sheet MF010-O-LC - 1 - 1. Properties The oxygen measuring system MF010-O-LC determines the oxygen content in gas mixtures up to a temperature of 250 C. It is particularly suitable for the

More information

Blood Parameter Monitoring System 550

Blood Parameter Monitoring System 550 Technical Compendium CDI Blood Parameter Monitoring System 550 An overview of the CDI System 550 and its industry leading technology. CDI System 550 Measures or Calculates 12 Critical Blood Parameters

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT

OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT BACKGROUND READING Animal Physiology by Hill, Wyse & Anderson, 2004: pp. 130 139 & 198 201. Living material Freshwater fish, probably goldfish.

More information

D-Opto. Dissolved Oxygen Sensor. Operation Manual (Software Version 1.0)

D-Opto. Dissolved Oxygen Sensor. Operation Manual (Software Version 1.0) D-Opto Dissolved Oxygen Sensor Operation Manual (Software Version 1.0) Table of contents 1 Introduction...3 1.1 Optical Sensor Technology...3 2 Software...4 2.1 Installation...4 2.2 Operation...4 3 Operation...9

More information

Dissolved Oxygen Probe (Order Code DO-BTA)

Dissolved Oxygen Probe (Order Code DO-BTA) Dissolved Oxygen Probe (Order Code DO-BTA) The Dissolved Oxygen Probe can be used to measure the concentration of dissolved oxygen in water samples tested in the field or in the laboratory. You can use

More information

TSE Blood Pressure Monitor Invasive series

TSE Blood Pressure Monitor Invasive series TSE Blood Pressure Monitor Invasive 209100series For Invasive Measurements in Laboratory Animals 2 Specifications subject to change without notice USA / Canada / Mexico Phone 1-989-698-3067 Fax 1-989-698-3068

More information

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors Abstract Boiler water in the low parts per billion dissolved oxygen concentration range can be significantly contaminated by oxygen

More information

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text. Hach LDO Sensor Instruction Sheet 00745589 Safety Precautions Please read this entire instruction sheet before operating this sensor. Pay particular attention to all danger and caution statements. Failure

More information

Dissolved Oxygen Measurement. Dissolved Oxygen

Dissolved Oxygen Measurement. Dissolved Oxygen Dissolved Oxygen Measurement Dissolved Oxygen Some oxygen is dissolved in practically every liquid. For example, at a temperature of 68 F (20 C) and an atmospheric pressure of 1013 mbar saturated water

More information

D i s s o l v e d O x y g e n M e a s u r e m e n t

D i s s o l v e d O x y g e n M e a s u r e m e n t D i s s o l v e d O x y g e n M e a s u r e m e n t Dissolved Oxygen Dissolved oxygen is present in virtually every liquid. For example, at a temperature of 20 C (68 F) and an atmospheric pressure of 1013

More information

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc. Advanced Calibration Designs, Inc. Calibration Gas Instrument INSTRUCTION MANUAL Release I www.goacd.com Instruction Manual Gas Generator Release I TABLE OF CONTENTS I. General Description Page 2 II. Start-Up

More information

Bioreactors Prof G. K. Suraishkumar Department of Biotechnology Indian Institute of Technology, Madras. Lecture - 15 Bioreactor env.par.

Bioreactors Prof G. K. Suraishkumar Department of Biotechnology Indian Institute of Technology, Madras. Lecture - 15 Bioreactor env.par. Bioreactors Prof G. K. Suraishkumar Department of Biotechnology Indian Institute of Technology, Madras Lecture - 15 Bioreactor env.par. (DO) Welcome to lecture 15 NPTEL online certification course on Bioreactors.

More information

Unisense In Situ UniAmp Connector System

Unisense In Situ UniAmp Connector System Unisense In Situ UniAmp Connector System Overview The Unisense In Situ UniAmp Connector system is designed for easy replacement of sensors. Unisense In Situ UniAmp Connector system is a dedicated system

More information

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL - Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL - Table of Contents 1. Introduction 3 2. Instrument Setup 4 2.1 Instrument Checklist 4 2.2 Optional Accessories 4 2.3 Gas Concentration Ranges Available

More information

Dissolved Oxygen/Temp Waterproof Tester

Dissolved Oxygen/Temp Waterproof Tester User's Guide AM-DO-01 Battery Compartment Cap Main Display Dissolved Oxygen/Temp Waterproof Tester Lanyard Connection old/max/min Button Function Mode Button Power / Calibration Button Probe Coller Probe

More information

Aerobic Respiration. Evaluation copy

Aerobic Respiration. Evaluation copy Aerobic Respiration Computer 17 Aerobic cellular respiration is the process of converting the chemical energy of organic molecules into a form immediately usable by organisms. Glucose may be oxidized completely

More information

InFlow Modular perfusion bioreactor

InFlow Modular perfusion bioreactor InFlow is an incubator compatible, modular bioreactor for the controlled, bidirectional and interstitial perfusion of up to 9 cell seeded scaffolds with culture medium. InFlow is ideal for general purpose

More information

Chemistry - Scale-up F.32. Lara TM - Controlled - Lab Reactor. The Concept

Chemistry - Scale-up F.32. Lara TM - Controlled - Lab Reactor. The Concept The Concept The concept behind Lara was to develop a flexible, affordable, easy-to-use CLR system. Lara incorporates many unique features that simplify the user s interaction saving time and money and

More information

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument OxyScan Graphic Operating Instructions UMS Micro-oxygen sensor 501 Microprocessor instrument Introduction Thank you for choosing the UMS Micro Oxygen Sensor 501 - a highly advanced product! Please read

More information

OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT

OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT BACKGROUND READING Animal Physiology by Hill, Wyse & Anderson, 2004: pp. 130 139 & 198 201. PRE-LAB (Due at the start of the lab) ** In your

More information

HI 9143, HI 9143M HI 9145

HI 9143, HI 9143M HI 9145 SAMPLE Instruction Manual HI 9143, HI 9143M HI 9145 Portable Waterproof Microprocessor Dissolved Oxygen Meters ppm C HI 9143 Microprocessor Auto Cal Dissolved Oxygen Meter ON/OFF CAL RANGE HOLD FACTOR

More information

Benefits. Rev.III_032018_Oxybaby 6.0_engl subject to technical alterations

Benefits. Rev.III_032018_Oxybaby 6.0_engl subject to technical alterations OXYBABY 6.0 for O 2 or O 2 /CO 2 Cordless hand held oxygen or combined oxygen and carbon dioxide analyser for checking modified atmospheres in food packs. The ideal instrument for portable, faster and

More information

Generating Calibration Gas Standards

Generating Calibration Gas Standards Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and

More information

PRACTICAL TIPS FOR TRACE-LEVEL OXYGEN MONITORING WITH ELECTRO-CHEMICAL SENSORS

PRACTICAL TIPS FOR TRACE-LEVEL OXYGEN MONITORING WITH ELECTRO-CHEMICAL SENSORS APPLICATION NOTE PRACTICAL TIPS FOR TRACE-LEVEL OXYGEN MONITORING WITH ELECTRO-CHEMICAL SENSORS Improved sensor performance yields longer intervals between service, consistent on-line monitoring, and increased

More information

Heavy Duty Dissolved Oxygen Meter

Heavy Duty Dissolved Oxygen Meter User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 FAX 781.665.0780 - TestEquipmentDepot.com Introduction Congratulations

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

Oxymax W COS51D. Technical Information. Digital sensor for measuring dissolved oxygen Membrane-covered amperometric sensor

Oxymax W COS51D. Technical Information. Digital sensor for measuring dissolved oxygen Membrane-covered amperometric sensor Technical Information Oxymax W COS51D Digital sensor for measuring dissolved oxygen Membrane-covered amperometric sensor Application The continuous measurement of the concentration of dissolved oxygen

More information

GAS ANALYSER OXYBABY M+ for O 2. or O 2 /CO 2. standard version

GAS ANALYSER OXYBABY M+ for O 2. or O 2 /CO 2. standard version GAS ANALYSER OXYBABY M+ for O 2 or O 2 standard version Cordless hand held oxygen or combined oxygen and carbon dioxide analyser for checking modified atmospheres in food packs. The ideal instrument for

More information

Model Description

Model Description LEAKHUNTER Plus LEAKHUNTER Plus Model 8066 Description The MATHESON LEAKHUNTER Plus is a truly universal leak detector. Extensive R&D has produced a multi-functional leak detector engineered to perform

More information

Instruction Manual DO 6. Economy Hand-held Dissolved Oxygen Meter. 68X Rev 2 Apr 08. Technology Made Easy... Part of Thermo Fisher Scientific

Instruction Manual DO 6. Economy Hand-held Dissolved Oxygen Meter. 68X Rev 2 Apr 08. Technology Made Easy... Part of Thermo Fisher Scientific Instruction Manual DO 6 Economy Hand-held Dissolved Oxygen Meter Technology Made Easy... 68X243622 Rev 2 Apr 08 Part of Thermo Fisher Scientific Preface This manual serves to explain the use of the DO

More information

Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017

Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017 PARTNERSHIP FOR THE DELAWARE ESTUARY Science Group Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017 Prepared By: Kurt Cheng Suggested Citation: Cheng,

More information

Dissolved Oxygen Probe (Order Code DO-BTA)

Dissolved Oxygen Probe (Order Code DO-BTA) Dissolved Oxygen Probe (Order Code DO-BTA) The Dissolved Oxygen Probe can be used to measure the concentration of dissolved oxygen in water samples tested in the field or in the laboratory. You can use

More information

A180 DO/BOD/OUR/SOUR Meter Instruction Manual

A180 DO/BOD/OUR/SOUR Meter Instruction Manual A180 DO/BOD/OUR/SOUR Meter Instruction Manual BANTE INSTRUMENTS CO., LTD A180 DO/BOD/OUR/SOUR Meter 1 Introduction Thank you for selecting the A180 multiparameter dissolved oxygen meter. This manual provides

More information

AUTOMATIC HOSE TEST UNIT, TYPE SPU

AUTOMATIC HOSE TEST UNIT, TYPE SPU VALVES AND FITTINGS UP TO 14,000 BAR TEST AND CONTROL EQUIPMENT H IGH PRESSURE TECHNOLOGY AUTOMATIC HOSE TEST UNIT, TYPE SPU Pressure range from 1 up to 10,000 bar User-friendly touch panel operation HIGH-PRESSURE

More information

Kube Engineering Quinn E. Eddins, B.S.E.E Phone

Kube Engineering Quinn E. Eddins, B.S.E.E Phone Process Information: There are four ph assemblies used to monitor the ph value on a Sulfur Dioxide absorber. The chemical composition is approximately 20-30% Sodium Bisulfite, 4% Sodium Sulfite, and the

More information

Oxygen Measurement Systems High Reliability and Wide Application Coverage

Oxygen Measurement Systems High Reliability and Wide Application Coverage Introduction High Process Efficiency Under Any Condition Oxygen Measurement Systems High Reliability and Wide Application Coverage METTLER TOLEDO provides sensors to measure dissolved oxygen (DO) in demanding

More information

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES PROCESS ANALYSERS DF-550E The DF-550E Nano Trace oxygen analyser is a Coulometric sensor based analyser designed to measure oxygen as a contaminant at ultra trace levels in ultra high purity electronic

More information

PH01 Perfusion Cannula Manual

PH01 Perfusion Cannula Manual PH01 Perfusion Cannula Manual Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted without the express written permission of Multi

More information

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL TABLE OF CONTENTS PRODUCT OVERVIEW 2 I - USE AND CARE: 3 II - INSTALLATION: 5 III - GENERAL MAINTENANCE TIPS: 5 IV - APPENDIX: A-1 2-WIRE CURRENT LOOP

More information

Best Practice for Calibrating LTH Conductivity Instruments

Best Practice for Calibrating LTH Conductivity Instruments Application Note Best Practice for Calibrating LTH Conductivity Instruments As accurate process measurement becomes an everyday requirement it is vital to be able to calibrate conductivity instruments

More information

The New PHD 4 Portable Helium Detector

The New PHD 4 Portable Helium Detector The New PHD 4 Portable Helium Detector SPECIAL FEATURES: High Sensitivity to Helium Easy to Use Truly Portable Versatile Dependable The system is designed to allow easy replacement of sampling line components

More information

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars. Simply a question of better measurement SCHMIDT Flow Sensor SS 20.261 The cost-effective alternative in pressurised systems up to 10 bars. Compressed air technology Industrial processes A cost analysis

More information

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS Interdependence of Plants and Animals Computer 14 Plants and animals share many of the same chemicals throughout their lives. In most ecosystems, O 2, CO 2, water, food and nutrients are exchanged between

More information

Oxygen Meter User Manual

Oxygen Meter User Manual Oxygen Meter User Manual Monday, July 23, 2007 1. Outline...2 2. Program...3 2.1. Environment for program execution...3 2.2. Installation...3 2.3. Un installation...3 2.4. USB driver installation...3 2.5.

More information

Cover Page for Lab Report Group Portion. Pump Performance

Cover Page for Lab Report Group Portion. Pump Performance Cover Page for Lab Report Group Portion Pump Performance Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 March 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:

More information

Mass Flowmeters. Mass Flowmeters for Gases

Mass Flowmeters. Mass Flowmeters for Gases Mass Flowmeters Mass Flowmeters for Gases Fast, Accurate, Low Pressure Drop for critical measurement applications! Flow (Std l/min) Response to a step change in flowrate 300 200 100 4 Milliseconds 63%

More information

with O 2 Controller Instruction Manual

with O 2 Controller Instruction Manual 14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Hypoxic Cabinets (In-Vivo / In-Vitro) with O 2 Controller Instruction Manual Index Page Warranty 2 Warnings

More information

Radnoti Liver Perfusion System

Radnoti Liver Perfusion System Radnoti Liver Perfusion System 130003 Offset Offset Gain Gain Radnoti 2006 Description Qty Part # A Base only, for 4-bar stand 1 159950-B4 B Stabilizer Bar only, for 4-Bar stand 2 159950-C4 C Rod 24 Long

More information

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions L 100 Bubble-Tube Level System Installation, Operation and Maintenance Instructions Figure 1 Contents Section Description Page 1.0 Introduction 2 2.0 Specifications 3 3.0 Installation 3 4.0 Warranty 6

More information

PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa)

PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa) Instruction Sheet for the PASCO Model CI-6532A PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa) 012-06859B 10/98 $1.00 polyurethane tubing syringe cable with DIN s to computer interface quick release s (4) pressure

More information

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cover Page for Lab Report Group Portion. Head Losses in Pipes Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS Tempe Pressure Cell June 1995 Fig. 1a 1400 Tempe Pressure Cells with 3cm cylinders and 6cm cylinders) mounted on the Tempe Cell Stand Fig. 1b Disassembled 1400 Tempe Pressure Cell

More information

Simple, Robust, Reliable Measures ph Doses either acid (lower) or alkali (raise) Displays ph, Total Dose Count (DCT) and Dose Count per hour (DC/h)

Simple, Robust, Reliable Measures ph Doses either acid (lower) or alkali (raise) Displays ph, Total Dose Count (DCT) and Dose Count per hour (DC/h) Simple, Robust, Reliable Measures ph Doses either acid (lower) or alkali (raise) Displays ph, Total Dose Count (DCT) and Dose Count per hour (DC/h) Integrated peristaltic dosing pump When using automatic

More information

Dextens Thermo Conductivity Sensors. Type 52101, 54101, 55101

Dextens Thermo Conductivity Sensors. Type 52101, 54101, 55101 Dextens Thermo Conductivity Sensors Type 52101, 54101, 55101 Operating Manual May 2012 Table of Content 1 GENERAL INFORMATION...4 1.1 DISCLAIMER...4 1.2 SYMBOLS AND CONVENTIONS...4 1.3 SERVICES AND REPAIRS...4

More information

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature Operating manual Dissolved Oxygen Depth Armature ba25302e05 04/2009 Copyright Weilheim 2009, WTW GmbH Reprinting - even as excerpts - is only allowed with the explicit written authorization of WTW GmbH,

More information

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin Electronic Control Manual 216 Sensors and Stats Section S Technical Bulletin CDS-2000 Issue Date 0393 CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin Introduction 3 Pre-Verification

More information

OxyProbe II Instruction Manual

OxyProbe II Instruction Manual OxyProbe II Instruction Manual D500 T82 Pg13.5 threaded nut O-Ring Sensor Shaft D600 ESSENTIAL INSTRUCTIONS: READ THIS PAGE BEFORE PROCEEDING! This product has been designed, manufactured, and tested to

More information

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

Instrumentation & Data Acquisition Systems

Instrumentation & Data Acquisition Systems Instrumentation & Data Acquisition Systems Section 3 -Level Robert W. Harrison, PE Bob@TheHarrisonHouse.com Made in USA 1 Level Section Question Which level measuring technology is the best solution when

More information