CALIBRATION CERTIFICATE # 500
|
|
- Stuart Franklin
- 6 years ago
- Views:
Transcription
1 CALIBRATION CERTIFICATE # 5 Calibration date : -- Certificate issued : --8 Company name Company address City, Province, Canada CLAS 9- Calibration of Mass flow meter Brooks SLA585S S/N : ABCD QUALITY PROGRAM CONFORMANCE All calibrations are performed in accordance with Polycontrols Laboratory Quality Assurance Manual and conform to ISO/IEC 75 5, ISO 9 8 and/or other quality requirements defined in customers purchase descriptions. TRACEABILITY The traceability for flow standard to the National Institute of Standards and Technology, NIST, is maintained by Fluke DH Instruments of Phoenix, Arizona and conform to ISO/IEC 75, ANSI/NCSL Z5--99, ISO-- and MIL-STD 566A. The Calibration Laboratory Assessment Service (CLAS) of the National Research Council of Canada (NRC) has assessed and certified specific calibration capabilities of this laboratory and traceability to the International System of Units (SI) or to standards acceptable to the CLAS program. This certificate of calibration is issued in accordance with the conditions of certification granted by CLAS and the conditions of accreditation granted by the Standards Council of Canada (SCC). Neither CLAS nor SCC guarantee the accuracy of individual calibrations by accredited laboratories. CALIBRATION AND MEASUREMENT CAPABILITY References used for flow calibration have an uncertainty of ±.% of reading for a flow range between 5 SCCM to, ±.% of reading for a flow range between to, ±.% of reading for a flow range between to, ±.% of reading for a flow range above to 6 and ±.5% of reading for a flow range under 5 SCCM down to SCCM. Test and reference uncertainties are given with a coverage factor of for a 95% level of confidence, air or nitrogen equivalent. The test uncertainty ratio (TUR) of this calibration is at least : unless otherwise stated. CONDITION SUMMARY OF THE DEVICE UNDER TEST Initial conditions Work done Results Remarks In good condition Cleaning and calibration Initial readings: refer to graphic Final readings in tolerance Laboratory Manager Polycontrols This calibration document shall not be reproduced except in full, without written approval of Polycontrols inc. 65, Matte blvd. (Unit A-), Brossard (Quebec), Canada, JY Z Tel: (5) -6 Fax: (5) Page of 5
2 Calibration Date: ABCD -- Standard equipment used for initial calibration Description Model Serial # Traceability Due date DHI molbloc ( slpm) E-VCR-V-Q DHI molbloc ( slpm) E-VCR-V-Q DHI molbloc ( slpm) E-VCR-V-Q DHI molbox Molbox Fluke Calibrator Initial specifications of the device under test Operation temperature C Reference temperature. C Reference pressure.5 mbar Range - Input/Output signal - ma Supply VDC Calibration conditions Gas N Gas N Accuracy ± %O.R. OR ±. %F.S. Ambient temperature C Ambient pressure.5 mbar Gas temperature C Correction factor Orientation Horizontal Seals Valve Initial readings Device under test Flow reference Calculated Error Acceptable Error TUR ma > > > Page of 5
3 Calibration Date: ABCD -- Initial results Error (%) Flow of the device under test () The measurement result is within the specification limit when the measurement uncertainty is taken into account The measurement result is outside the specification limit when the measurement uncertainty is taken into account It is not possible to state compliance using a 95 % coverage probability for the expanded uncertainty although the measurement result is below the limit (ILAC G8) Page of 5
4 Calibration date: ABCD -- Standard equipment used for final calibration Description Model Serial # Traceability Due date DHI molbloc ( slpm H) 5E-VCR-V-Q DHI molbloc ( slpm) E-VCR-V-Q DHI molbox Molbox Fluke Calibrator Final specifications of the device under test Operation temperature C Reference temperature. C Reference pressure.5 mbar Range - Input/Output signal - ma Supply VDC Calibration conditions Gas N Gas N Accuracy ± %O.R. OR ±. %F.S. Ambient temperature C Ambient pressure.5 mbar Gas temperature C Correction factor Orientation Horizontal Seals Valve Final readings Device under test Flow reference Calculated Error Acceptable Error TUR ma > > > Page of 5
5 Calibration date: ABCD -- Final results Error (%) Flow of the device under test () The measurement result is within the specification limit when the measurement uncertainty is taken into account The measurement result is outside the specification limit when the measurement uncertainty is taken into account It is not possible to state compliance using a 95 % coverage probability for the expanded uncertainty although the measurement result is below the limit (ILAC G8) Page 5 of 5
Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.
Bulletin Category: GAS Bulletin: G-23 Page: 1 of 7 The copy of this document located on Measurement Canada s website is considered to be the controlled copy. Recognition of data from specialty gas suppliers
More informationVaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.
Asset #: 123 Calibration - Certificate No: 2083.01 Customer: Condition: Action Taken: Sample Inc. 123 Sample Rd. Sample, MA 01234 The instrument was operational upon receipt. The instrument was adjusted
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 FLUKE CALIBRATION, PHOENIX PRIMARY PRESSURE AND FLOW LABORATORY 4765 East Beautiful Lane Phoenix, AZ 85044 Mike DiGioia Phone: 602 773
More informationCERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Antibus Scales & Systems, Inc. 1919 Research
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 FIRSTENERGY BETA LABORATORY 6670 Beta Drive Mayfield Village, OH 44143 William Coll Phone: 440 604 9889 CALIBRATION Valid To: November
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 THE L.S. STARRETT COMPANY 121 Crescent Street Athol, MA 01331 Harold J. Bacon Phone: 978 249 3551 CALIBRATION Valid To: October 31,
More informationEN ISO/IEC Technical Requirements
EN ISO/IEC 17025 - Technical Requirements Wolfgang Garche Environmental Protection Agency Department Air Quality Monitoring, Information, Assessment 1 What means accreditation in accordance to EN ISO/IEC
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ACME SCALE COMPANY 1801 Adams Ave San Leandro, CA 94577 Barbara Byrd Phone: 510 638 5040 CALIBRATION Valid To: August 31, 2019 Certificate
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 APPLIED TEST SYSTEMS 154 East Brook Lane Butler, PA 16002 Rob Carroll Phone: 724 283 1212 CALIBRATION Valid To: March 31, 2018 Certificate
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 CAPSTONE CALIBRATION 7016 Baker Blvd Richland Hills, TX 76118 Jon T. Cappa Phone: 817 284 2707 CALIBRATION Valid To: January 31, 2020
More informationCTAG:e43bda4f8e48714fbb aefb192 Certificate of Calibration
CTAG:ebdaf8e87fbb5996aefb9 Certificate of Calibration Data Type: FOUND-LEFT Certificate No: 6858 Instrument Description Manufacturer Model Serial Number Inventory Number CABLE ANALYZER FLUKE NETWORKS DTX-8
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 ANSI/NCSL Z540-1-1994 ISO-CAL NORTH AMERICA 36610 N 20 th Street Phoenix, AZ 85086 Erik Naranen Phone: 480 336 8133 Fax: 480 336 8737 enaranen@isocalnorthamerica.com
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 FLUKE CALIBRATION, PHOENIX PRIMARY PRESSURE AND FLOW LABORATORY 4765 East Beautiful Lane Phoenix, AZ 85044 Mike DiGioia Phone: 602 773
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 APPLIED TEST SYSTEMS 154 East Brook Lane Butler, PA 16002 Rob Carroll Phone: 724 283 1212 CALIBRATION Valid To: February 29, 2020 Certificate
More informationCTAG:b9927e36a4b3f34a9138ce0bb2426b22 Certificate of Calibration
CTAG:b997e6abfa98cebb6b Certificate of Calibration Data Type: FOUND-LEFT Certificate No: 79 Instrument Description Manufacturer Model Serial Number Inventory Number CABLE ANALYZER FLUKE NETWORKS DTX-8
More informationFlow Meters and Flow Controllers
Flow Meters and Flow Controllers HFM-205 Flow Meter HFC-207 Flow Controller HIGH CAPACITY FLOW METERS AND FLOW CONTROLLERS FEATURES ± 1% full scale accuracy 1 Input Power: +/- 15 VDC or +24 VDC (specify
More informationGas mixtures. Individual solutions specifically for your application
Gas mixtures Individual solutions specifically for your application For many applications besides pure gases also defined gas mixtures are required. For many routine applications, from operating gases
More informationAGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY
AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY Agilent Vacuum Products: Our Committment to Quality Agilent s commitment to quality and Measurement s philosophy Reference Standards (PNEUROP, ISO)
More informationMass Flowmeters. Mass Flowmeters for Gases
Mass Flowmeters Mass Flowmeters for Gases Fast, Accurate, Low Pressure Drop for critical measurement applications! Flow (Std l/min) Response to a step change in flowrate 300 200 100 4 Milliseconds 63%
More informationBrooks TMF-Zone 1 Models 5816-Ex/5864-Ex
Data Sheet DS-TMF-586Ex-MFC-eng TMF-Zone Brooks TMF-Zone Models 586-Ex/5864-Ex Customer Benefits Assured process accuracy and repeatability Long-term reliability Reduced maintenance CE certified Improved
More informationDate of Shipment: Xxxxx 00, 20xx 3402c National Institute of Advanced Industrial Science and Technology National Metrology Institute of
Date of Shipment: Xxxxx 00, 20xx 3402c01151112-170831 National Metrology Institute of Japan Reference Material Certificate NMIJ CRM 3402-c01 Sulfur Dioxide This certified reference material (CRM) was produced
More informationNMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013
NMISA NEW GAS FLOW CALIBRATION FACILITY Deona Jonker 7 October 2013 Introduction Flow the total volume of a fluid that flows past a fixed point in a given time Flow measurements used in industry as a means
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PRECISION MEASURING CORP Groesbeck Hwy. Fraser, MI Robert Meyer Phone:
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 PRECISION MEASURING CORP. 33250 Groesbeck Hwy. Fraser, MI 48026 Robert Meyer Phone: 586 293 4420 CALIBRATION Valid To: September 30, 2019 Certificate Number:
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z CALIBRATION
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006 ROBB PRECISION TOOL SERVICES, INC. 2100 196th St SW, Suite 144 Lynnwood, WA 98036 Robert Robb Phone: 425 672
More informationmolbox1+ Flow Terminal ± % of reading lowest uncertainty for gas flow calibration
molbox1+ Flow Terminal ± 0.125 % of reading lowest uncertainty for gas flow calibration Technical Data The new molbox1+ flow terminal from Fluke s DH Instruments Division represents a significant update
More informationMass Flow Controllers
Mass Flow Meters & Controllers HFM-D-3A & 305A Mass Flow Meters HFC-D-3A & 307A Mass Flow Controllers FEATURES Range 0-25 slm to 0-2500 slm (N 2 Equivalent) Excellent Accuracy ±(0.5% of Reading + 0.2%
More informationLight Lab. Reliable Measurements through Regular Calibration. DAkkS
Light Lab Reliable Measurements through Regular Calibration QM DAkkS ARS 85 EYREIENCEG QM EXP Benefits of Calibration PERFECTION THROUGH CALIBRATION RIN EASU HT M OLOGY IG L IN ECHN T LEGAL CERTAINTY Avoidance
More informationCALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)
CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE (+34) 968 162 005 www.sensovant.com comercial@sensovant.com www.epluse.com CALIBRATION AT THE HIGHEST LEVEL The
More informationCertified Reference Materials FOR REFRACTOMETERS AND POLARIMETERS
Certified Reference Materials FOR REFRACTOMETERS AND POLARIMETERS Regular verification of laboratory instrumentation is of primary importance in a modern manufacturing facility, not only for reasons of
More informationTechnical Data Sheet High Performance Digital Gas Mass Flow Meters and Controllers Features Description
Smart-Trak Model 100 High Performance Digital Gas Mass Flow Meters and Controllers Features Measure & control gas mass flow rates up to 1000 slpm True digital performance provides high accuracy and great
More informationNMISA New Gas Flow Calibration Facility
NMISA New Gas Flow Calibration Facility Speaker / Author: D. Jonker Co-author: E.P. Tarnow National Metrology Institute of South Africa (NMISA) Private Bag X34, Lynnwood Ridge, Pretoria, 0040, South Africa
More informationKeeping It Simple. Gas Mixtures for Stack Emissions Monitoring.
Keeping It Simple Gas Mixtures for Stack Emissions Monitoring www.uk.airliquide.com Stack Emissions Monitoring The Stack Emissions Monitoring market has undergone substantial legislative and quality changes
More informationGFM. Typical Aluminum GFM Mass Flow Meter NIST MASS FLOW METERS. Principles of Operation. Design Features
Design Features Rigid metallic construction. Maximum pressure of 1000 psig (70 bars). Leak integrity 1 x 10-7 of helium. NIST traceable certification. Built-in tiltable LCD readout. 0-5 Vdc and 4-20 ma
More informationUNDERSTANDING, ACCELERATED
UNDERSTANDING, ACCELERATED MEASURE FLOW, PRESSURE, AND TEMPERATURE ALL IN ONE INSTRUMENT! Designed for Performance TSI thermal mass flowmeters incorporate a proprietary platinum film sensor design for
More informationRecommended Practices for Calibration Laboratories
Recommended Practices for Calibration Laboratories May 1999 General This document recommends environmental practices and documentary and operational protocols for calibration laboratories that provide
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z TROXLER ELECTRONICS LABORATORIES, INC. James Pratt Phone:
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 TROXLER ELECTRONICS LABORATORIES, INC. James Pratt Phone: 919 314 2707 CALIBRATION Valid To: See Footnote 5 Certificate Number: 3260.01
More informationENGINEERING TECHNICAL REPORT
SSP DUOLOK FITTINGS ON 1.000 x 0.139 WALL 316SS TUBING ENGINEERING TECHNICAL REPORT SSP 8250 Boyle Parkway Twinsburg, Ohio 44087 U.S.A. mysspusa.com 1.330.425.4250 1 TABLE OF CONTENTS 1.0 Introduction
More informationUKAS calibration laboratory 8809 Temperature, pressure and electrical signals calibration guide
ABB MEASUREMENT & ANALYTICS PRODUCT GUIDE PG/SMD140 EN REV. B UKAS calibration laboratory 8809 Temperature, pressure and electrical signals calibration guide Measurement made easy UKAS calibration laboratory
More informationService ISO17025 explained
ABB MEASUREMENT & ANALYTICS ARTICLE Service ISO17025 explained The ISO17025 standard covers any organization offering testing and calibration services Measurement made easy Accredited laboratory for testing
More informationVT900A + VAPOR. Technical data. One Solution. Auto-detection to Ensure Patient Safety. Key benefits and features
Technical data VT900A + VAPOR The Fluke Biomedical VT900A + VAPOR is designed to efficiently and reliably perform a full anesthesia machine PM, from ventilators to vaporizers. One Solution VT900A + VAPOR
More informationGMC-I Service GmbH Calibration Center
2016 GMC-I Service GmbH Calibration Center DAkkS, ISO & Factorycalibration www.gmci-service.com From spare parts, reparation, rental service, maintenance of industrial health and safety standards, calibration
More informationTechnical Procedure for General Laboratory Equipment
Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all
More informationTitle: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator
Procedure No: SOP-012 Revision No: 1.2, January 27, 2011 Page No.: 1 of 7 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Sabio model 2010 and 4010 dilution calibrators.
More informationGreen 0-3% Acceptable Orange 3.1-5% Alert Red 5.1 and up Fail. Calibrated By:
Miller Thermal 3620 Certificate of Calibration Cert. No. : 2009-031 Page: 1 of 10 Customer: ABC Console Miller 3620 Hopper 1262 Power Supply PS100 Address: 1365 Newton S/N: KB26001B S/N: KC560100 S/N:
More informationAFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features
ANALOG MASS FLOW CONTROLLERS Model AF mass fl ow controllers are designed to indicate fl ow rates and control set fl ow rates of gases. Each of these units incorporates an advanced straight tube sensor
More informationGas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)
Int. J. Metrol. Qual. Eng. 5, 303 (2014) c EDP Sciences 2014 DOI: 10.1051/ijmqe/2014012 Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA) D. Jonker and E.P. Tarnow
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: 12100 West 6 th Avenue, Lakewood, CO 80225 (Hereinafter
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 LEDFORD GAGE LAB, INC. 227 Industrial Dr. Mulvane, KS 67110 Blaine Johnson Phone: 316 777 9300 CALIBRATION Valid To: December 31, 2019
More informationUnique mounting features for Thermal Mass Flow Meters. By Martin Raab in cooperation with CaTeC b.v.
Unique mounting features for Thermal Mass Flow Meters By Martin Raab in cooperation with CaTeC b.v. Development and production of sensors and transmitters for: humidity dew point CO2 O2 air velocity mass
More informationVT900. Gas Flow Analyzer. Technical data. Key benefits and features:
VT900 Gas Flow Analyzer Technical data The Fluke Biomedical VT900 is designed to accurately and reliably test all types of medical gas flow equipment ventilators, insufflators, oxygen meters especially
More informationNIST Traceability Document
Page 1 of 9 Gas Flowmeter Calibration Using the Restek Designed Calibration System July 10, 2010 Document # 02-16-11 Procedure# 02-16-06 1.0 General Description The flow meter is a volumetric flow meter.
More informationSideTrak 830/840. Process Gas Mass Flow Meters and Controllers. FEATURES DESCRIPTION
Process Gas Mass Flow Meters and Controllers FETURES n Direct monitoring of mass flow eliminates need for ancillary pressure and temperature sensing n Patented straight sensor tube with access ports permits
More informationGUIDANCE FOR THE APPLICATION OF EQUIPMENT CALIBRATION AND TRACEABILITY OF MEASUREMENTS ON INSPECTION BODIES
GUIDANCE FOR THE APPLICATION OF EQUIPMENT CALIBRATION AND TRACEABILITY OF MEASUREMENTS ON INSPECTION BODIES ESYD G-DIAKRIB/01/01/26-03-2015 1/6 ESYD G-DIAKRIB Issue: 01 Revision: 01 Issue Date: 28-11-2007
More informationDesign Features. General Description
Model thermal Mass Flow Controllers are designed to indicate and control set flow rates of gases. The combines the characteristics and accuracy of conventional mass flow devices into a unique compact design
More informationHUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0
High Accuracy Humidity Calibrator The role of humidity calibrations that are accurate, reproducible, and documentable is becoming more and more important. ISO quality guidelines and regulations according
More informationSmartTrak. High Performance Digital Gas Mass Flow Meters & Controllers. DESCRIPTION
High Performance Digital Gas Mass Flow Meters & Controllers n Measure and control gas mass flow rates up to 1000 slpm n Pressure up to 5000 psig (333 barg) nideal for OEM, industry or research applications
More informationHigh Performance Digital Gas Mass Flow Meters & Controllers
High Performance Digital Gas Mass Flow Meters & Controllers FEATURES n Measure and control gas mass flow rates up to 1000 slpm n Pressure up to 500 psig (34.5 barg). For pressures greater than 500 psig
More informationSpeciality Gases Practice 4. Our quality, your safety. Accredited Test and Calibration Laboratory Speciality Gases Centre in Hörstel.
Speciality Gases Practice 4 Our quality, your safety. Accredited Test and Calibration Laboratory Speciality Gases Centre in Hörstel. Accredited Westfalen quality. Speciality Gases Centre in Hörstel. They
More informationReference Material Institute for Clinical Chemistry Standards (ReCCS) JCCRM 621-4
Certified Reference Material for Blood Gases JCCRM 621-4 Handling Instructions Properties and Intended Use This Certified Reference Material (CRM) for Blood Gases is intended for use in verifying the accuracy
More informationIt s not like other pressure gauges
It s not like other pressure gauges Stronger. Tougher. Better. Dropped, Submerged, Over-Pressurized and it still reads accurately! Submersible: Sealed against dust and water. IP67 certified for 30 minutes
More informationSmartTrak 100HP. Ultra-High Pressure Digital Gas Mass Flow Meters & Controllers. FEATURES DESCRIPTION
Ultra-High Pressure Digital Gas Mass Flow Meters & Controllers FEATURES n Ideal for pilot plants, hydrogenation reactors, and autoclave processes n Measure and control gas mass flow rates over an inlet
More informationTELEDYNE HASTINGS INSTRUMENTS PRELIMINARY BULLETIN. Models HFM-D-301 & 305, HFC-D-303 & 307
TELEDYNE HASTINGS DIGITAL 300 MEDIUM/HIGH CAPACITY FLOWMETERS AND CONTROLLERS INSTRUMENTS Models HFM-D-301 & 305, HFC-D-303 & 307 FEATURES Flows to 2500 slm (N 2 Equivalent) Accuracy in Nitrogen ±(0.2%
More informationMax-Trak Model 180. Technical Data Sheet. NEMA 6 / IP67 Compliant Industrial Mass Flow Meters and Controllers For Gases
NEMA 6 / IP67 Compliant Industrial Mass Flow Meters and Controllers For Gases Features Industry s only NEMA 6/IP67 industrial MFC affords wash down protection eliminating the need for secondary enclosures
More informationMicro Motion 7812 Fiscal Gas Density Meter
Product Data Sheet PS-001162, Rev. B April 2009 Micro Motion 7812 Fiscal Gas Density Meter Micro Motion density and concentration meters are built to tackle the most demanding process and fiscal applications.
More informationCALIBRATION SYSTEM REQUIREMENTS. ESCC Basic Specification No
Page 1 of 8 CALIBRATION SYSTEM REQUIREMENTS ESCC Basic Specification Issue 4 February 2014 Document Custodian: European Space Agency see https://escies.org PAGE 2 LEGAL DISCLAIMER AND COPYRIGHT European
More informationTechnical Data for Alicat MS Series Mass Flow Meters
Technical Data for Alicat MS Series Mass Flow Meters Alicat MS instruments are built for use with aggressive For the most part, these instruments maintain the specifications of equivalently ranged M Series
More informationStability and traceability of calibration gases for EN 14181
Stability and traceability of calibration gases for EN 14181 Stephen Mellor Especialista de Produto Product Specialist Analytical, Air Products Plc, United Kingdom DOSSIER EMISSÕES E MONITORIZAÇÃO ESTABILIDADE
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 CORPORATE CONSULTING, SERVICE, & INSTRUMENTS, INC. University Park, 221 Beaver Street Akron, OH 44304 Michael Kent Warner Phone: 330 376 3600 CALIBRATION Valid
More informationmolbloc/molbox Low Mass Flow Calibration System AFMETCAL Configurations Installation and Operation Manual Supplement
molbloc/molbox Low Mass Flow Calibration System AFMETCAL Configurations Installation and Operation Manual Supplement High pressure liquids and gases are potentially hazardous. Energy stored in these liquids
More informationSpecific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities
Specific Accreditation Criteria Calibration ISO IEC 17025 Annex Mass and related quantities January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected
More informationHow to evaluate pressure measurement specifications
How to evaluate pressure measurement specifications Application Note There are thousands of different pressure measuring devices out there. How do you decide which instruments to invest in to calibrate
More informationPressure is our Business
Pressure is our Business Count on gauges and calibrators that perform Wherever you work Reduce Measurement Error Protect against diminished accuracy from environmental conditions, accidental damage, and
More informationTraceable calibration of automatic weighing instruments in dynamic operation
Traceable calibration of automatic weighing instruments in dynamic operation Matej Grum* Metrology Institute of the Republic of Slovenia, Grudnovo nabrežje 17, 1000 Ljubljana, Slovenia Abstract. The article
More informationBeamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments
Beamex Calibration White Paper info@beamex.com Weighing scale calibration - How to calibrate weighing instruments Weighing scale calibration - How to calibrate weighing instruments Weighing scales, weighing
More informationEngineering Report Crash Safety (Impulse) Shock Test. Crystal Group, Inc. for. Prepared by. Phillip M. Toftely, Test Engineer.
Engineering Report 47706-2 Crash Safety (Impulse) Shock Test for Crystal Group, Inc. Prepared by Phillip M. Toftely, Test Engineer Approved by David M. Gillen, Vice President This document shall not be
More informationHow we calibrate the Wave Height and Period Measurements from the Gravitational Acceleration Wave Buoys in RMIC/AP
How we calibrate the Wave Height and Period Measurements from the Gravitational Acceleration Wave Buoys in RMIC/AP YU Jianqing RMIC for the Asia-Pacific Region National Center of Ocean Standards and Metrology,
More informationmolbox1 molbloc Terminal (Ver and Higher) Operation and Maintenance Manual
molbox1 molbloc Terminal (Ver. 5.20 and Higher) Operation and Maintenance Manual High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases can be released unexpectedly
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: UNEX/HYTORC, 333 Route 17 North, Mahwah, NJ 07430
More informationali-q TM Gravimetric Verification Procedure
ali-q TM Gravimetric Verification Procedure Introduction This document describes the recommended procedure for verifying performance of the ali-q aliquoting pipet controller using the industry standard-
More informationMass Flow Meter (MFM) for Gases
873 Mass Flow Meter (MFM) for Gases Direct flow measurement by MEMS- Technology for nominal flow rates from 1 ml N /min to 8 l N /min (N 2 ) High accuracy Short response time Compact design and digital
More informationProtocol Gas Verification For Compressed Gas Cylinders Containing Either SO 2, NO or CO. Quality Assurance Plan/Standard Operating Procedure
Page 1 of 4 Protocol Gas Verification For Compressed Gas Cylinders Containing Either SO 2, NO or CO Quality Assurance Plan/Standard Operating Procedure Page 2 of 4 Table of Contents 2.3.6 Protocol Gas
More informationGuide No. 2/2009. Design, training and equipment Martin Firlus. Physikalisch Technische Bundesanstalt Braunschweig und Berlin
Guide No. 2/2009 THE ROAD TO a mass laboratory Design, training and equipment Martin Firlus Physikalisch Technische Bundesanstalt Braunschweig und Berlin Imprint Published by: Responsible: Text: Layout:
More informationTEXAS DEPARTMENT OF PUBLIC SAFETY Breath Alcohol Laboratory
TEXAS DEPARTMENT OF PUBLIC SAFETY Breath Alcohol Laboratory 5805 N LAMAR BLVD BOX 4087 AUSTIN, TEXAS 78773 512/424-5200 STEVEN C. McCRAW DIRECTOR DAVID G. BAKER CHERYL MacBRIDE DEPUTY DIRECTORS January
More informationMET-101 Introduction to Metrology
Pasig city Tel: 654-13-42 MET-101 Introduction to Metrology This 2-day course offers a basic introduction into Metrology, meant for people who are working in calibration and testing labs, maintenance services
More informationPrecision pressure sensor Basic version Model CPT6020
Calibration Precision pressure sensor Basic version Model CPT6020 WIKA data sheet CT 25.13 Applications Calibration technology High-accuracy pressure monitoring Pressure sensing in critical applications
More informationTEK-THERMAL 1700B. Thermal Mass Flowmeter. FLOW. Technology Solutions
Technology Solutions TEK-THERMAL 1700B Thermal Mass Flowmeter FLOW www.tek-trol.com Flow Level Temperature Pressure Valves Analyzers Accessories TekValSys Introduction Tek-Thermal 1700B Thermal Mass Flowmeters,
More informationAdditel 761 Automated Pressure Calibrators
Automated Calibrators UPDATED Fully automated pressure calibrator with built-in pressure generator / controller to as high as 600 psi (40 bar) or as low as 0.01 Pa (0.00004 inh 2 O) accuracy Dual pressure
More informationJOS 15 JOS 25 JOS 40. Manual and semiautomatic test benches from DN 25 to DN 40. Characteristics. Basic characteristics. Test bench using balances
TEST BENCHES JOS 15 JOS 25 JOS 40 Manual and semiautomatic test benches from DN 25 to DN 40 02 Characteristics Compact test benches - all in one room Designation - for residential and apartment meters
More informationCertification of AMS acc. EN 15267, Part 3 - Overview and First Experience -
Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience - Dr. Wolfgang Jockel, Martin Schneider, TÜV Rheinland Group, D-51105 Cologne / Germany 1. Introduction A new basis for the certification
More informationMACH ONE MASS FLOW CONTROLLER. MACH ONE SERIES flow control. MASS FLOW CONTROLLERS at the speed of sound.
MACH ONE MASS FLOW CONTROLLER MACH ONE SERIES flow control MASS FLOW CONTROLLERS at the speed of sound. MACH ONE SERIES MASS FLOW CONTROLLER FLOW CONTROL AT THE SPEED OF SOUND The Mach One revolutionary
More informationBattery Powered Thermal Mass Flow Meters for Gases
red-y compact series product information attery Powered hermal Mass Flow Meters for Gases icenta Controls Ltd, Unit 3 he Woodford Centre, Lysander Way, Old arum Park, alisbury, P4 6U, U Digital advantage:
More informationRecalibration of gas meters
12 November 2015 Recalibration of gas meters Experiences of GTS The value of Recalibration Robert Kruithof First European Recalibration Workshop NMi Euroloop, Rotterdam Robert Kruithof Agenda Introduction
More informationAccredited Laboratory
Accredited Laboratory A2LA has accredited CAL-RITE CORPORATION Naperville, IL for technical competence in the field of Calibration This laboratory is accredited in accordance with the recognized International
More informationDust & Water Test Report SMA SMA-J-C-XX-ST-SB1/SMA-P-C-XX-ST-CA1
Project Number: Design Verification Test Report Requested by: Tori Meek Date: 11/3/2010 Product Rev: 0 Part #:SMA-J-C-XX-ST-SB1/SMA-P-C-XX-ST-CA1 Lot #: na Tech: Gary Lomax Eng: Eric Mings Qty to test:
More informationBalancing and Settlement Code. Code of Practice Four
Balancing and Settlement Code Code of Practice Four CODE OF PRACTICE FOR THE CALIBRATION, TESTING AND COMMISSIONING REQUIREMENTS OF METERING EQUIPMENT FOR SETTLEMENT PURPOSES Issue 6 Version 11.0 Date:
More informationPressure Metrology in Mexico
Pressure Metrology in Mexico Jorge C. TORRES G., Benjamin SORIANO C. and Victor ARANDA C. Centro Nacional de Metrologia-CENAM km 4,5 Carretera a Los Cues, El Marques, Queretaro, MEXICO e-mail: jorge.torres@cenam.mx
More informationAccreditation for In-Service Inspection of Transportable Pressure Receptacles (TPRs)
RG 3 Edition 4 December 2018 Accreditation for In-Service Inspection of Transportable Pressure Receptacles (TPRs) RG 3 / Edition: 4 Page 1 of 8 Contents 1. Introduction 2 2. Scope - Inspection Services
More informationCALIBRATION AND TRACEABILITY: ITS IMPORTANCE IN THE GAS INDUSTRY
ICAST 2000 edited by Wan Ramli Wan Daud et. al 1079 Faculty of Engineering, Universiti Kebangsaan Malaysia CALIBRATION AND TRACEABILITY: ITS IMPORTANCE IN THE GAS INDUSTRY Rahmat Mohsin, Zulkifli Abdul
More informationInterlaboratory comparison in the pressure range from 0 to 2 MPa for accredited calibration laboratories
Int. J. Metrol. Qual. Eng. 6, 307 (2015) c EDP Sciences 2015 DOI: 10.1051/ijmqe/2015021 Interlaboratory comparison in the pressure range from 0 to 2 MPa for accredited calibration laboratories Lovorka
More information7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications
GE Sensing & Inspection Technologies 7250 Sys Ruska Multi-Range Pressure Calibration System Features Fully integrated multi-range pressure test and calibration systems Select either an eight range or the
More information