1 1234 Hach Hall PE 2400 II CHN/S Operating Instructions Revised 05/29/2015 S.V. Location: Contact: 1240 Hach Hall Steve Veysey, 1234 Hach Hall Safety All researchers working in 1240 Hach Hall must complete the EH&S course: Fire Safety and Extinguisher Training. Be aware that the CHN/S analyzer operates at very high temperatures. When weighing samples, very small excess quantities (typically microgram quantity spillage onto the weighing paper) may be disposed of in the waste container next to the instrument. It is your responsibility to remove the unweighed remainder of your samples from the lab. Samples left behind will be disposed. Personal protective equipment (safety glasses and gloves) is available in the lab, or in 1238A Hach Hall. NOTE: The guidelines in this document require that you understand the reason for each action. Quite often it will be necessary to deviate from the example protocol given below. Unless you understand the purpose of each action you will not be able to modify the actions intelligently.

2 The Measurement Principle The instrument may be operated in either CHN or CHNS mode. Small quantities of sample (typically about 2 milligrams) are accurately weighed into small tin capsules. At elevated temperatures, in the presence of excess oxygen, organic materials combust to form CO 2, H 2 O, various N x O y compounds, and SO 2 if sulfur is present. These reactions are facilitated by solid catalysts packed in the combustion tube; other combustion products like HCl are generally removed by silver gauze and other solid materials packed in the tube. The N x O y combustion products are reduced by fine copper in the reduction tube to N2. The excess oxygen used to support combustion is also scrubbed out in the copper reduction tube. The only gases collected for quantitative analysis are: CO 2 representative of carbon content, H 2 O representative of hydrogen content, N 2 representative of nitrogen content SO 2 representative of sulfur content (only with the CHNS configuration) The catalysts in the combustion tube have a long life; however they can be poisoned over time. The effective lifetime of a combustion tube is about 750 runs. The protective quartz liner used in the combustion tube to collect the tin slag from the sample capsules should be cleaned or replaced every 100 runs. Because the copper in the reduction tube is converted to copper oxide as N x O y and O 2 are reduced, the reduction tube lifetime is finite. For CHN operation, there are separate combustion and reduction tubes. The reduction tube lifetime is about 120 runs. For CHNS operation, combustion and reduction functions occur in one tube. The CHNS combined-tube lifetime is therefore limited by the amount of copper in the tube; the lifetime is typically about 80 runs. The combustion gases are collected automatically in a fixed-volume in the instrument, and then dosed over time onto a GC column designed specifically for the quantitative analysis of gases. The GC uses a thermal conductivity detector. Because the gases are dosed in over a period of time (as opposed to being instantly injected), discrete, well-separated peaks are not observed. Instead, plateaus are observed, and the steady-state values of each plateau are recorded by the instrument at pre-determined times. This is referred to as frontal chromatography. The plateau values for each component are obtained by subtracting the value of the previous plateau. In the example shown below: N R = = 1456 C R = = H R = = 4445

3 In order to convert the detector signals (baseline, nitrogen, carbon dioxide, water, sulfur dioxide) into %C, %H, %N, and %S, it is necessary to do the following. Each will be discussed in detail. 1. Establish steady state conditions in the instrument 2. Determine blank values 3. Run pure standards (known %C, %H, %N, %S) to establish K-factors 4. Run accurately weighed representative replicates of each sample. 1. Achieve steady-state conditions Adsorption and desorption of gases on all glass and metal surfaces occurs continuously in the entire flow path of the instrument. These gases come from the combustion of samples, but also from impurities in the oxygen support gas and the helium carrier gas, and from small leaks at seals and valves that can introduce air into the system. Clearly, if the instrument has been unused for a period of time, desorption will exceed adsorption. The result of running a sample when the instrument is in a desorbed state will be arbitrarily low readings, since the combustion gases formed will first need to fill adsorption sites in the instrument before reaching the chromatograph detector. Similarly, the result of running a high-organic content sample after running a series of low-organic content samples (or blanks) could be low values. This is especially evident with

4 hydrogen content as the water adsorption fluctuates. To minimize these effects, the following should be done: Always use high-purity oxygen and high purity helium gases. Since the inherent impurity content is lower, the fluctuation from steady-state conditions will be less when blanks or low-organic content samples are run. Always run the leak diagnostic (diagnostics-2-1-2) when the columns have been changed, or you suspect there may be a leak in the system. Always condition the system by running several unweighed capsules of acetanilide (CHN) or cysteine (CHNS) at the beginning of the run sequence. Adjust your samples sizes based upon the expected organic content. Weigh more if the organic content is low. This is just one of several factors to consider when deciding the target weight of your samples. 2. Determine blank values There will be a detector reading for zero, nitrogen, carbon, and hydrogen, even when an empty tin capsule is run. These are called the blanks, eg. Nbl, Cbl, Hbl. The instrument firmware keeps track of blank readings and uses these readings in a very specific manner, one that I do not totally agree with. When you run a blank, the immediate run values are given equal weight to the previous average blank values, and a new average blank value is calculated and used by the firmware calculations of %C, %H, %N, %S. To establish starting blank values independent of historical values, the firmware recognizes that when two blank runs occur in a row, the second one is to be considered correct and the values are not averaged with previous values; they are used as the new average blank value settings. Additional blanks need to be run throughout the run sequence so that trends in the blanks can be incorporated into the final recalculation of your results. 3. Establish K-factors The equations that convert detector readings into %C, %H, %N, %S require a conversion factor for each element. This is established by running accurately weighed standards; normalizing the detector reading by weight; then dividing the normalized reading by the theoretical weight-percent of the C, H, N, S in the standard. These are called K-factors, i.e. Kf C, Kf H, Kf N, Kf S. Several duplicates of the standard are run at the beginning of the run sequence, spaced by blanks. Then additional duplicates of the standard (always preceded by a blank) are run throughout the run sequence. How often the standards are run depends upon several considerations, but generally a duplicate of standards should be run after the first 5-10 sample runs, then duplicates should be run after ever sample runs. NOTE: Although the firmware allows you to run these standards as K- factors, we ALWAYS acquire them as sample runs. It is then easier to use the results in the recalculation trend analysis.

5 4a. Run representative sample replicates sample composition Properly prepared solid samples must be very dry, and pulverized into fine particles. Failure to do so may affect measurement precision due to incomplete combustion (insufficient surface area), or inhomogeneity of the sample. Without establishing measurement precision (repeatability) there can be no confidence in measurement accuracy, even when the instrument is well calibrated. The precision specification of the instrument is +/-0.4 weight-% for each element when at least triplicates are run. If your samples are properly prepared (dry, homogeneous mixture of fine particles, reasonable sample weight for the expected organic content) you might very well achieve measurement precisions better than this. A recent set of four ACE standards showed a measurement precision (one standard deviation) of 0.06 %C, 0.05 %H, and 0.15 %N. 4b. Run representative sample replicates target sample weight When establishing the target sample size to weigh, several factors apply. As you consider these, you will recognize that one target weight might not be optimal for all the information you are trying to get from your sample. Understanding these factors will allow you to make the best choices depending upon which element is most important to you. Is your sample 100% organic, with %C not exceeding 80%? Then your target weight should be 2.0 mg, +/- 0.5 mg. Is your sample 100% organic, with %C exceeding 80%? Your target weight should be much less, perhaps 1.5 mg +/-0.3 mg. Is your sample between 10-50% organic content? Your target weight should be 2.5 mg +/-0.5 mg. Is your sample less than 10% organic content? Your target weight should be 3.5 mg +/- 0.5 mg. Is your sample known to be difficult to combust? Your target weight should be 1.0 mg +/- 0.2 mg. Consider performing a combustion optimization run sequence. For CHNS, is sulfur greater than 5%? The previous suggestions all apply. For CHNS, is sulfur less than 5%, but not as important as the CHN results? The previous suggestions apply, but also run several sulfur blanks. For CHNS, is sulfur less than 5% and precise %S is most important? Multiply the weights suggested above by a factor of 1.5; be sure to run sulfur blanks; consider running low-sulfur content standards similar in nature to your compounds. Does your sample contain silicon? Adjust the target weights downward; be ready to accept less precise measurements; consider performing a combustion optimization run sequence. Does your sample contain halogens? The CHN combustion tube has chemicals to scrub out the halogens; performance depends upon the compound. The CHNS tube does not have compounds to scrub out halogens. In either case, use lower weights; be prepared for less precise results. In several of the above situations, running additional replicates (as many as four or five) might be necessary to achieve reasonable confidence in the results.

6 Acquiring Data Instrument Preparation The system should be checked for leaks whenever a tube is replaced; diagnostic is run. The system should be purged (He = 300 seconds; O2 = 150 seconds) whenever a tube or gas cylinder is replaced, and just before a run sequence is initiated. CHN Operation The normal operating procedure includes running conditioners, blanks, calibrants, and standards throughout the run sequence. Acetanilide (ACE) is used for CHN studies. The default optimizecombustion settings for CHN measurements are 3,5,1,0. Check them before starting the run; often students change them to suit specific needs. CHNS Operation Cysteine (CYS) is used for CHNS studies. For CHNS measurement the optimize-combustion parameters are normally 1,1,1,0. A special sulfur conditioner (sulfonic acid - unweighed) is run prior to initiating the normal run sequence described above. You can run this as a Single Run, or as the first entry in the Autorun sequence. Cysteine (CYS) is used instead of acetanalide for all conditioning and K-factor adjustment runs. Operational Procedure Check gases He >200 psi in the tank; regulator set at 20 psi O2 >200 psi in the tank; regulator set at 18 psi Air >80 psi Verify Instrument Mode PARAMETER 6 CHN mode or CHNS mode. If wrong mode for the runs you are planning, tubes must be changed. Verify Run Capacity PARAMETER 4 Run counters. All three must have values in excess of your expected total number of runs, including conditioners, blanks, calibrants, and samples. Because the run capacity is just an estimate, factor in a margin of safety. If more aggressive optimizecombustion parameters are used, the run capacity of the reduction tube decreases. Check for Leaks (required after columns have been changed; otherwise optional) DIAGNOSTIC Resulting printout should pass. If failed, evaluate severity of failure. If pressure drops less than 20 mm Hg during the test, you may proceed if you wish as if the test had passed. Report the small leak to me.

7 Optimize Combustion PARAMETER 9 Optimize-combustion parameters. For CHN work, choose constant, unless you plan to burn some samples differently. Verify that the settings are 3,5,1,0 for CHN work. To use different values, choose select and set appropriate values in one of the four available programs. For CHNS, choose constant and verify that the values are 1,1,1,0. Purge Gases PURGE Purge helium for 300 seconds, and oxygen for 150 seconds. Autorun Sequence Move sample carousel clockwise to position #1. AUTORUN If sample counter is not #1, choose 4RP -> 1 Reset -> Reset All Y. This removes the existing data from the instrument memory. Be sure that the previous data has been fully processed before you do this. AUTORUN #1,2,3 Choose 3S for each of these to run the unweighed ACE conditioners (CHN mode), or CYS conditioners (CHN/S mode). AUTORUN #4,5 Choose 1B to run two blanks. AUTORUN #6,7 Choose 3S to run weighed ACE (or CYS) calibrants. AUTORUN #8 Choose 1B to run one blank AUTORUN #9,10 Choose 3S to run weighed ACE (or CYS) calibrants. AUTORUN #11 Choose 1B to run one blank AUTORUN #12,13 Choose 3S to run weighed ACE (or CYS) calibrants. AUTORUN #14->18 Choose 3S to run the first five samples AUTORUN #19,20,21 Choose 1B,3S,3S to run a blank and two weighed ACE (CYS) AUTORUN #22->31 Choose 3S to run the next ten samples AUTORUN #32,33,34 1B,3S,3S to run a blank and two weighed ACE (CYS) AUTORUN #35->44 Choose 3S to run the next ten samples. AUTORUN #xxx Finish by running a blank and two ACE (CYS) In summary, a typical CHN run sequence would be: Run # Compound Run as... Purpose 1-3 ace (unweighed) sample condition the system 4-5 empty tin capsule blank set initial blank values 6-7 ace ( mg) sample K-factor adjustment 8 empty tin capsule blank blank adjustment 9-10 ace ( mg) sample K-factor adjustment 11 empty tin capsule blank blank adjustment ace ( mg) sample K-factor adjustment sample runs 1-5 sample unknown samples 19 empty tin capsule blank blank adjustment

8 20-21 ace ( mg) sample K-factor adjustment sample runs 6-15 sample unknown samples continue samples in groups of 10 If this were a CHNS run sequence, #1 would be an unweighed sulfamic acid; runs 2,3 would be unweighed cysteine; cysteine would be used in all cases instead of acetanilide. REPROCESSING DATA The instrument prints out the results from each run as it is acquired. The data is also stored in the microprocessor memory, as long as it was acquired using the AUTORUN program. Data acquired as SINGLE RUN is printed, but not stored in memory. A typical printout of blank data will look like the results below. Note that since AVERAGE RESULTS are the same as the measured values, this must be the second of two consecutive blank runs. Recall that when blanks are run consecutively, the firmware assumes that you wish to assign the current blank measurement as the historical average. DATE TIME OPERATOR ID SV RUN 8 BLANK RUN SIGNALS AVERAGE RESULTS ZR CARBON -9 NR CARBON 9 HYDROGEN 153 CR HYDROGEN 153 NITROGEN 55 HR NITROGEN 55 A typical printout of an ACE standard run as a sample will look like: DATE TIME OPERATOR ID SV RUN 10 ID ACE WEIGHT SIGNALS ZR CARBON 71.46% NR HYDROGEN 6.49% CR NITROGEN 10.09% HR BLANKS KFACTORS OXFILL COMB BOOST1 BOOST FILL TIME 47 SECONDS Using nitrogen as an example, the equation used by the firmware to calculate the %N is: %N = ((NR-ZR-Nbl)*100) / (wt ugms*kf N ) =( )*100 / (1870*5.000) =10.09

9 Similarly the equation used for this same data to recalculate the Kf N, knowing that the theoretical weight-% for acetanilide is 10.36, would be: Kf N = ((NR-ZR-Nbl)*100) / (wt ugms*10.36) =( )*100 / (1870*10.36) =4.868 Clearly, as long as we have a printed copy of the original measurements for each run in the run sequence, we can use Excel to recalculate results from scratch. However, it is laborious to type in all of the original weights and measurements into Excel. We normally adopt a hybrid approach which requires some work in Excel, followed by some use of the instrument firmware recalculation capabilities. Hybrid recalculation approach The measured blank values are transferred to Excel. When trends are not discernible, blank corrections are established by taking the average of all readings obtained throughout the run, or by allowing the system-calculated running average values to remain. Either method will usually suffice for the C, N, and S readings. Hydrogen blank values normally require adjustment based upon a curve. For example: NOTE: The procedure to enter the reprocessing part of the firmware is a bit complicated, and is presented in the Appendix. The corrected blank values are then entered into the "reprocessing" function of the analyzer firmware, and a blank-corrected printout is obtained for all runs. From the blank-corrected printout, the measured values for all acetanilide runs (cysteine if it is a CHNS sequence) are transferred to Excel. The average %C, %H, %N, (and %S) values are obtained for the ACE (CYS) standards. Using simple ratio and proportion, corrected K-values are obtained.

10 newk C = oldk / Theory% X Average = ( / 71.09) X = newk H = oldk / Theory% X Average = ( / 6.71) X 6.32 = newk N = oldk / Theory% X Average = (5.050 / 10.36) X = Note that measurements not used in the averages should be crossed out. In this example, there were no obvious outliers to exclude from the %C or %N averages, but each of the first of the duplicate %H measurements was low. This is a result of water desorbing from the system during the immediately preceding blank run. While somewhat troublesome, this is normal. The effect can be minimized by using very pure helium carrier gas so that the change from adsorbed water steady-state conditions is small, but it is still necessary to correct based upon a curve. The standard deviations for the ACE measurements are reasonable; well below the instrument specification of +/- 0.4%. The new K value to be used in the final reprocessing of the data is calculated (using Kf C as an example) from: The corrected Kf values are then entered into the "reprocessing" function of the analyzer firmware, and a final printout using the best possible blank and K-factor values is obtained for all runs.

11 APPENDIX Precision and Accuracy The instrument specification is +/- 0.3% for each element as the standard deviation of four replicates. So, what does it mean if the precision of the measurements on your sample is not as good as the values for the standard? What does it mean if the accuracy of the measurements (based upon your expected formula) is not as good as the values for the standard? First of all, accuracy can never be better than precision. Precision should ideally be established with at least four replicates, but most students just do two replicates. If and when you notice that two measurements differ by more than 0.8% (the standard deviation would then be 0.6%), you should run a third replicate if sample is available. If the values for the third replicate group well with the higher valued of the first two runs, you can exclude the low outlier. There are several reasons why we can rationalize an outlier being low, but we cannot rationalize an outlier being high. Some of the factors you can control that affect precision are sample homogeneity and particle size. A factor you may not be able to control is the inherent combustibility of your sample. If that is an issue, a full combustion study should be done to optimize combustion parameters, and possibly combustion additives like vanadium pentoxide, or thermolite. Rule out homogeneity and particle size first as the cause of poor precision. If good measurement precision is achieved but the values are not as you expected, be aware that accuracy will depend upon factors such as the purity, dryness, and trapped solvent molecules in your sample. This last effect can be determined by running an NMR of your sample and quantifying the nature and percent of solvent in your sample. The measured %C, %H, %N values can then be corrected. Combustion and Reduction Tubes Tubes are packed in-house using reagents purchased from Perkin Elmer or from other thirdparty vendors. Depleted reduction tubes are regenerated using a special furnace we have constructed; a 6% mixture of hydrogen in argon gas (Praxair Hydrostar H-6) is used. Quartz tubes are reused whenever possible; new tubes are purchased from Perkin Elmer, from other third-party vendors, or manufactured by our glassblower. Ultra-clean tin capsules (6mm x 4mm) are used. UHP Plus grade oxygen is used to support combustion UHP Plus grade helium is used as the carrier gas All standards are from Perkin Elmer or from other third-party vendors.

12 CHN/S Microprocessor The instrument is controlled from the soft keys on the front panel. These interface with the embedded microprocessor and memory. Each soft key should beep when pressed. Note that there are no alpha characters. Alpha characters are entered by pressing the DOT key followed by a two-digit numeric code associated with each letter. Generally, each program function is entered or exited by pressing the program key. In cases where the program has multiple levels, exiting may require pressing the function key multiple times to roll back through each program level. If you mistype something several levels deep in a program, first try the CE (clear entry) key; otherwise simply exit the program and start over. The two-line LCD display may show current state, selection choices, results, or keypad entries. The microprocessor programs include: Diagnostics: A suite of tests, including a leak test you must be familiar with. Monitor: Allows viewing of various transducers, including furnace temperature. Purge Gas: Both the Helium and Oxygen lines should be purged prior to a run sequence. Parameters: Allows viewing and selection of all instrument parameters. Autorun: Used to set up the run sequence and also reprocess data stored in memory Single Run: Not recommended; data is printed but not stored in memory Start: Used to start or continue execution of the Autorun sequence or Single Run. Standby: Not recommended; this key is more for emergency use. <- and ->: Move the LCD flashing cursor position left or right. Yes and No: Response to Y/N selection questions.

13 Alpha Chart Parameters Menu (1) Parameters Menu (2) Normal instrument state when no programs are running. 42 different parameter settings are used to define the instrument configuration and elements of the run sequence and data processing. Parameters you will likely review and/or set include: 4. Run Counters Set to 120 for new CHN reduction tube; set to 90 for CHNS Don t reset. Notify me when it drops below 100. Set to 120 for CHN; set to 90 for CHNS Used to set countdown warnings for the combustion/reduction tubes and vial receptacle. 6. Operating Mode. Used to select between CHN and CHN/S operation. We do not have the OXY option. 9. Optimize Combustion. 1.-> 2.-> Choosing constant prompts you to set static values for Oxyfill, Comb, OxyB1, and OxyB2 that will be used automatically in the Autorun program. Choosing select prompts you to establish four

14 different sets of values (Opt 1, 2, 3, 4) each with different Oxyfill, Comb, OxyB1, and OxyB2 settings; Autorun will then ask for a selection each time you add a run to the run sequence. 12. Furnace ON/OFF Use to start furnace cool-down so that a column can be replaced. 42. Blanks Selection Normal setting is OFF. ON results in Autorun presenting a choice. Normally Autorun establishes CHN/S blank values based upon running empty tins, but it is known empirically that the Sbl reading is slightly higher when a CHN standard, containing no sulfur, is run. For low-content sulfur samples, a more precise sulfur reading can be obtained by using a Sbl value derived from this special sulfur-blank method. Setting this parameter to ON results in Autorun presenting a selection choice when a blank run is requested. A range of instrument diagnostics can be selected from a multi-layer menu. Students should only run 2-1-2, the combustion leak program. 2.-> 1.-> 2.-> Runs comb. Tube leak test. Note that when the diagnostic starts, the tube should pressurize to about 760 torr within seconds, and then should stay relatively constant at that value, dropping no more than 5-10 torr over the duration of the test. The diagnostic will then report Pass or Fail. If the test fails catastrophically, you must consult with me and you should not begin your measurements. NOTE: The furnace must be completely warmed up and stable at the operating temperature before you begin the diagnostic. When cooling or heating the furnace, the temperature can be monitored. Choose NO to the Print List question, then select Monitor Code 1. NO -> Both the helium and the oxygen lines should be purged before a run sequence is started. Typically 300 seconds and 150 seconds are used. YES -> YES -> All runs should be programmed as part of an Autorun sequence. Of the three acquisition types, blanks, K-factor and sample runs, we only use

15 (1)Blanks, and (3) Samples. This has been discussed previously. The 4RP selection is used to reprocess data stored in memory after it has been acquired, and will be discussed in the next section. 1. Blanks Offers a choice only if Parameter 42 is set to ON. Offers a choice if Parameter 9 is set to SELECT. 3. Samples Use the numeric keypad to enter the coded alpha characters and/or the numeric characters of the sample ID. Pressing ENTER prompts you to enter the weight. Note that the weight can then be transferred directly from the balance by pressing ENTER, or it can be typed in, followed by pressing ENTER. If Parameter 9 is set to SELECT, you will be presented with the choice of optimize combustion programs 1-4. NOTE: Be aware that after the final ENTER is pressed, the entry is added to the queue and the system assumes there is now a sample (or blank) physically in the assigned position in the autosampler carousel. If an Autorun is in progress, this sample (or blank) will be run without further operator intervention. Reprocessing Data When you enter Autorun, selecting RP allows you to reset and begin a new run sequence, print sample information, print sample results, or recalculate the data.. (1) RESET -> is used only when you are setting up a new run sequence to begin at position/sample 1. This will erase all of the results from memory, so be sure that data has already been reprocessed. The system will ask if you wish to RESET ALL runs; select YES. (2) PRINT INF This is used to print a summary list of all the runs entered in the run sequence. You must provide this list to me at the end of your session so that proper billing can occur. (3) PR RESLTS This choice can be used to reprint the entire sample information, analysis conditions, and results for each entry in the run sequence. It will duplicate the report printed as each run finishes during the analysis. (4) RECALC To recalculate data, first enter the AUTORUN program. The display will show the next position in the queue available to be programmed: position 47. To access earlier data in the queue, use the left arrow key. In this example it is to move the flashing cursor so it is under the 4 in 47. Then press the CE (clear entry) key, and type in the first run number you wish to reprocess, followed by ENTER. In this example, it is run number 37.

16 Press the Parameters key! You will be presented with the RP display. Select (4) RECALC. You will be presented with the recalculation choices dialog box. You can recalculate any single run based upon a different weight. This is helpful if you notice a transcription error in the weight recorded on your sample sheet (or in the logbook) and the weight you actually entered. You can also recalculate a series of runs based upon new blank values or new K-factor values. As discussed earlier, when reprocessing a series of data, we first correct the C, N and S blanks based upon the average of all of the blanks acquired during the run sequence. The H blank is taken from the plotted curve of Hbl vs run# and applied sequentially to smaller groups of runs. Select (1) BLANKS. You will be presented sequentially with the current C, H, N, and S blank values, and given the opportunity to enter new values for each one. In each case, the value is changed by moving the cursor that is flashing until it is under one of the digits, pressing CE, and typing in the new value. You will then be asked how many runs the new blank values should be applied to, beginning from (in this example) run 37. The recalculated results will print. After reprocessing the run sequence with more correct blank values, the data needs to be reprocessed again with more correct K-factors. The protocol for establishing new K-factors was discussed previously. After entering the calculation program at the first run to be reprocessed, select (2) KFACTORS. You will be presented sequentially with the current C, H, N, and S K-factors, and in each case given the opportunity to enter new values. The value is changed by pressing CE, and typing in a new value. You will then be asked how many runs the new K-factor values should be applied to, beginning from (in this example) run 37. The recalculated results will print.

Organic Elemental Analysis Series II Nitrogen Analyzer. proven performance. reliable results

Organic Elemental Analysis Series II Nitrogen Analyzer. proven performance. reliable results Organic Elemental Analysis 2410 Series II Nitrogen Analyzer proven performance reliable results PerkinElmer 2410 Series II Nitrogen Analyzer Advanced Combustion Method Q U I C K G L A N C E Advanced combustion

More information

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module TECHNICAL NOTE High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module TN42256 Dr. Liliana Krotz, Dr. Francesco Leone, Walter Galotta and Dr. Guido Giazzi

More information

Organic Elemental Analysis Series II CHNS/O Elemental Analyzer. proven performance. reliable results

Organic Elemental Analysis Series II CHNS/O Elemental Analyzer. proven performance. reliable results Organic Elemental Analysis 2400 Series II CHNS/O Elemental Analyzer proven performance reliable results 2400 Series II CHNS/O Elemental Analyzer raising the standard Q U I C K G L A N C E One analyzer

More information

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide

More information

Operation of the Perkin Elmer TGA-GC/MS

Operation of the Perkin Elmer TGA-GC/MS Operation of the Perkin Elmer TGA-GC/MS Summary of the TGA-GC/MS: The TGA-GC/MS allows a user to decompose a sample by heating, measure its loss of mass, and simultaneously analyze the chemical composition

More information

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE. Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV 16-1 INTRODUCTION 16-1.1 Purpose. The purpose of this chapter is to familiarize divers with the techniques used to mix divers breathing gas. 16-1.2 Scope. This chapter

More information

GCMSD-Headspace Analysis SOP

GCMSD-Headspace Analysis SOP Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to

More information

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD) Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet

More information

CORESTA RECOMMENDED METHOD Nº 67

CORESTA RECOMMENDED METHOD Nº 67 CORESTA RECOMMENDED METHOD Nº 67 DETERMINATION OF WATER IN THE MAINSTREAM SMOKE OF CIGARS BY GAS CHROMATOGRAPHIC ANALYSIS (November 2005) 1. FIELD OF APPLICATION The method is applicable to the particulate

More information

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER Michael T Tang, Ph.D. Grace Feng Greg Merideth Rui Huang Matheson Gas Applied Lab

More information

Gas Laws Chapter 14. Complete the following pressure conversion. Be sure to show how units cancel.

Gas Laws Chapter 14. Complete the following pressure conversion. Be sure to show how units cancel. Gas Laws Chapter 14 Complete the following pressure conversion. Be sure to show how units cancel. 1 atm = 760 mm Hg = 760 torr = 101.3 kpa = 14.7 psi = 1.013 bar 1. The air pressure for a certain tire

More information

no peaks missing peaks too small peaks increasing retention decreasing retention

no peaks missing peaks too small peaks increasing retention decreasing retention GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering

More information

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System) Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System) Revision date October 2014 2 Emergency Plan for Nano Furnace Standard Operating Procedures for Emergencies Contact information Person

More information

CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO Walnut Lake Rd th Street Houston TX Garner, IA 50438

CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO Walnut Lake Rd th Street Houston TX Garner, IA 50438 CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO. 10 Jerry Paul Smith Joel Clancy JPS Measurement Consultants, Inc Colorado Engineering Experiment Station, Inc (CEESI) 13002 Walnut Lake

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

Prep LC Training Manual S.V.

Prep LC Training Manual S.V. 1234 Hach Hall 515-294-5805 www.cif.iastate.edu Prep LC Training Manual 11-6-2013 S.V. Location: Contact: 1238 Hach Hall Kamel Harrata, 1236 Hach Hall and Steve Veysey, 1234 Hach Hall Safety: All researchers

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor Anwar Sutan, Metco Services Ltd. Charles Johnson, Metco Services Ltd. Jason Laidlaw,

More information

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Quantitative Analysis of Hydrocarbons by Gas Chromatography Quantitative Analysis of Hydrocarbons by Gas Chromatography Introduction Gas-liquid chromatography (GLC) accomplishes a separation by partitioning solutes between a mobile gas phase and a stationary liquid

More information

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cover Page for Lab Report Group Portion. Head Losses in Pipes Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section

More information

Applications Note: Use of "pentane equivalent" calibration gas mixtures

Applications Note: Use of pentane equivalent calibration gas mixtures Introduction The gas that is used to verify accuracy is every bit as important as the detector itself when it comes to worker safety. Choosing (and using) the right mixture is critical to the success of

More information

UNIT 10 - GASES. Notes & Worksheets - Honors

UNIT 10 - GASES. Notes & Worksheets - Honors Ideal Gas Equation 1 WKSHT 1.) What is the pressure exerted by 2.0 moles of an ideal gas when it occupies a volume of 12.0 L at 373 K? 2.) A flashbulb of volume 2.6 cm 3 contains O 2 gas at a pressure

More information

Standard Operating Procedure

Standard Operating Procedure Page 1 of 10 Vigor Glove Box Standard Operating Procedure Lab: ESB 154 Department: Materials Science and Engineering PI: Paul V. Braun Written By: Subing (Alan) Qu Section 1: Overview Type of SOP: Process

More information

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12 LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/ This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the rates with which selected

More information

Experiment GC : Analysis of BTEX by GC-FID

Experiment GC : Analysis of BTEX by GC-FID 46 Experiment GC : Analysis of BTEX by GC-FID Learning Goals: Familiarity with gas chromatography Gain experience in temperature programming and method development Correctly use an internal standard, and

More information

Zebron Gas Management

Zebron Gas Management Zebron www.phenomenex.com/gasmanagement What are the key benefits of Zebron filters and traps? The high adsorption capacity of the Zebron Filters removes contaminants from mobile phase gases, reducing

More information

FLUORESCENCE DETERMINATION OF OXYGEN

FLUORESCENCE DETERMINATION OF OXYGEN FLUORESCENCE DETERMINATION OF OXYGEN Objectives This experiment will familiarize the students with the principles of fluorescence quenching as well as the Ocean Optics FOXY Fiber Optic Sensor System. Spectra

More information

Retention Time Locking: Concepts and Applications. Application

Retention Time Locking: Concepts and Applications. Application Retention Time Locking: Concepts and Applications Application Gas Chromatography December 1997 Authors Vince Giarrocco Bruce Quimby Matthew Klee Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

The Ideal Gas Constant

The Ideal Gas Constant Chem 2115 Experiment # 8 The Ideal Gas Constant OBJECTIVE: This experiment is designed to provide experience in gas handling methods and experimental insight into the relationships between pressure, volume,

More information

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information

HiQ laboratory gas generators.

HiQ laboratory gas generators. Product information HiQ laboratory gas generators. The convenient alternative for laboratory environments. 02 HiQ laboratory gas generators HiQ laboratory gas generators. Designed with safety and convenience

More information

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA BRIEFING 846 Specific Surface Area, USP 27 page 2385. The European Pharmacopoeia is the coordinating pharmacopeia for the international harmonization of the Specific Surface Area General Chapter, as part

More information

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

Equipment Operating Procedure Glove Box

Equipment Operating Procedure Glove Box Equipment Operating Procedure Glove Box Page 1 0.0 Changing the Compressed Gas Cylinder 1. Complete Compressed Gas Cylinder training from EHS website before manually exchanging gas cylinders. In order

More information

Diagnosing EVAP systems By: Bernie Thompson

Diagnosing EVAP systems By: Bernie Thompson Diagnosing EVAP systems By: Bernie Thompson Perhaps one of the toughest jobs in the automotive field is diagnosing small leaks, especially A/C and Fuel Vapor (EVAP) leaks. Finding these leaks can be time

More information

Kinetic-Molecular Theory

Kinetic-Molecular Theory GASES Chapter Eleven Kinetic-Molecular Theory! Recall that our only previous description of gases stated that gases completely fill and take the shape of their containers.! The Kinetic-Molecular Theory

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

4.) There are no forces of attraction or repulsion between gas particles. This means that

4.) There are no forces of attraction or repulsion between gas particles. This means that KINETIC MOLECULAR (K-M) THEORY OF MATTER NOTES - based on the idea that particles of matter are always in motion - assumptions of the K-M Theory 1.) Gases consist of large numbers of tiny particles that

More information

Quick Start Guide for SA-9600 Series Surface Area Analyzer

Quick Start Guide for SA-9600 Series Surface Area Analyzer Quick Start Guide for SA-9600 Series Surface Area Analyzer Multi-Point Measurement HORIBA Scientific www.horiba.com/us/particle labinfo@horiba.com 1-888-903-5001 SA-9603, Surface Area Analyzer, Multi-Station

More information

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Lab 1c Isentropic Blow-down Process and Discharge Coefficient 058:080 Experimental Engineering Lab 1c Isentropic Blow-down Process and Discharge Coefficient OBJECTIVES - To study the transient discharge of a rigid pressurized tank; To determine the discharge coefficients

More information

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the

More information

Experiment 18 Properties of Gases

Experiment 18 Properties of Gases Experiment 18 Properties of Gases E18-1 E18-2 The Task In this experiment you will investigate some of the properties of gases, i.e. how gases flow, their phase changes and chemical reactivity. Skills

More information

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

CHM Gas Chromatography: Our Instrument Charles Taylor (r10) 1/8 Introduction You should already be familiar with the basic concepts of gas chromatography. The purpose of this note pack is to familiarize you with the specific instrument that we will use in our labs.

More information

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,

More information

3 Flow and Pressure Control

3 Flow and Pressure Control 3 Hydrogen shutdown Column shutdown Turning gas flows on and off EPC-controlled streams NonEPC-controlled streams Electronic Pneumatic Control (EPC) Interpreting flow and pressure readings Configuration

More information

Model Description

Model Description LEAKHUNTER Plus LEAKHUNTER Plus Model 8066 Description The MATHESON LEAKHUNTER Plus is a truly universal leak detector. Extensive R&D has produced a multi-functional leak detector engineered to perform

More information

Multiple Gas#5 GC configuration Jan 2016

Multiple Gas#5 GC configuration Jan 2016 History: Unfortunately there is no single column that can separate: Hydrogen Oxygen Nitrogen Methane CO CO2 Ethane Water Propane Butane Pentane Over the years SRI Instruments has devised several solutions

More information

Mass Spec will not Autotune

Mass Spec will not Autotune Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems

More information

VIC offers a variety of Calibrated Gas Leaks

VIC offers a variety of Calibrated Gas Leaks CALIBRATED GAS LEAKS VIC offers a high quality line of calibrated gas leaks traceable to National Institute of Standards Technology (NIST). Calibrated gas leaks are used to provide a controlled delivery

More information

CHEM 3351 Physical Chemistry I, Fall 2017

CHEM 3351 Physical Chemistry I, Fall 2017 CHEM 3351 Physical Chemistry I, Fall 2017 Problem set 1 Due 9/15/2017 (Friday) 1. An automobile tire was inflated to a pressure of 24 lb in -2 (1.00 atm = 14.7 lb in -2 ) on a winter s day when the temperature

More information

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative

More information

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary ADH 1/7/014 LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the

More information

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies IMPROVED CALIBRATION AND OPERATION OF A STABLE-GAS QUANTIFICATION MANIFOLD AS PART OF A RADIOXENON COLLECTION SYSTEM ABSTRACT Richard M. Williams, James C. Hayes, and Randy R. Kirkham Pacific Northwest

More information

Protective Atmospheres, Measurement Technologies and Troubleshooting Tools

Protective Atmospheres, Measurement Technologies and Troubleshooting Tools Protective Atmospheres, Measurement Technologies and Troubleshooting Tools Furnace atmospheres are critical to meet metallurgical specifications defined by control processes. The makeup of a furnace s

More information

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS Using state-of-the-art measuring and microprocessor technologies, Advanced has assembled a series of systems which can measure mass

More information

Paper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration

Paper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration Paper 2.2 Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration Mr William Freund, Daniel Measurement and Control Mr Klaus Zanker, Daniel Measurement and Control Mr Dale Goodson,

More information

Cover Page for Lab Report Group Portion. Pump Performance

Cover Page for Lab Report Group Portion. Pump Performance Cover Page for Lab Report Group Portion Pump Performance Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 March 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:

More information

Name Chemistry Pre-AP

Name Chemistry Pre-AP Name Chemistry Pre-AP Notes: Gas Laws and Gas Stoichiometry Period Part 1: The Nature of Gases and The Gas Laws I. Nature of Gases A. Kinetic-Molecular Theory The - theory was developed to account for

More information

Sampling Equipment Catalog. MicroPurge Low-Flow Pump Control

Sampling Equipment Catalog. MicroPurge Low-Flow Pump Control MicroPurge Low-Flow Pump Control Sampling Equipment Catalog 10 MicroPurge Controls The MicroPurge Controller (U.S. Patent Number 6,508,310) revolutionizes low-flow sampling with advanced logic control

More information

Technical Support Note

Technical Support Note Title: Air Dilution Correction using O2 and CO2 TSN Number: 17 File:S:\Bridge_Analyzers\Customer_Service_Documentation\Technical_Support_Notes\ 17 Air Dilution correction using O2 and CO2.docx Created

More information

You should be able to: Describe Equipment Barometer Manometer. 5.1 Pressure Read and outline 5.1 Define Barometer

You should be able to: Describe Equipment Barometer Manometer. 5.1 Pressure Read and outline 5.1 Define Barometer A P CHEMISTRY - Unit 5: Gases Unit 5: Gases Gases are distinguished from other forms of matter, not only by their power of indefinite expansion so as to fill any vessel, however large, and by the great

More information

School of Chemistry SOP For Operation Of Glove Boxes

School of Chemistry SOP For Operation Of Glove Boxes School SOP for Operation of Glove Boxes: The following SOP provides guidelines on how to adequately and safely operating a standard laboratory glove box and its associated devices. General Information:

More information

End of Chapter Exercises

End of Chapter Exercises End of Chapter Exercises Exercises 1 12 are conceptual questions that are designed to see if you have understood the main concepts of the chapter. 1. While on an airplane, you take a drink from your water

More information

Chapter 13 Gases, Vapors, Liquids, and Solids

Chapter 13 Gases, Vapors, Liquids, and Solids Chapter 13 Gases, Vapors, Liquids, and Solids Property is meaning any measurable characteristic of a substance, such as pressure, volume, or temperature, or a characteristic that can be calculated or deduced,

More information

Operating the DataMaster cdm VERSION 1.2

Operating the DataMaster cdm VERSION 1.2 Operating the DataMaster cdm VERSION 1.2 The DataMaster cdm is a scientific instrument designed to analyze a sample of a person s breath and determine the Breath Alcohol Concentration (BrAC) in that sample.

More information

Calcimeter Instruction Manual

Calcimeter Instruction Manual Hohner (UK - Canada - Texas) Calcimeter Instruction Manual The Hohner Calcimeter is based on industry standard versions, and is used to measure the calcium carbonate and magnesium carbonate in samples.

More information

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work.....2

More information

SOP: Buck Scientific BLC-20P HPLC Operation

SOP: Buck Scientific BLC-20P HPLC Operation Page 1 of 11 Approvals Preparer: John Buford Reviewer: Tim Kull Reviewer: Dr. Margaret Bryans Date: 16OCT13 Date: 30OCT13 Date: 31OCT13 1. Purpose 1.1. Basic operation of the Buck Scientific BLC-20P isocratic

More information

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions Varian 3300 Gas Chromatograph Abbreviated Operating Instructions 1. Introduction and Startup 2 2. Basic Configurations for the GC 3 3. Sample Preparation 3 4. Building a Method 3 5. Split/Splitless Injection

More information

5890II GC Standard Operating Procedure 9/2/2005

5890II GC Standard Operating Procedure 9/2/2005 5890II GC Standard Operating Procedure 9/2/2005 This procedure is for the analysis of natural gas using the HP 5890II GC with the Restek ShinCarbon column. Detector Type: TCD Column Type: Restek Micropacked,

More information

Manual for continuous distillation

Manual for continuous distillation Manual for continuous distillation 1. Week 1: Objectives: Run the column at total reflux. When steady state is reached, take the sample from the top and bottom of the column in order to determine the overall

More information

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1 Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC

More information

Gases and welding distributors continue

Gases and welding distributors continue Specialty Gas Instrumentation By Ray Borzio, Mike Lee, and Scott Varian Gases and welding distributors continue to have a significant future in the production, delivery, and related services for specialty

More information

This test shall be carried out on all vehicles equipped with open type traction batteries.

This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following

More information

CHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller

CHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller CHEMICAL ENGINEERING LABORATORY CHEG 239W Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller Objective The experiment involves tuning a commercial process controller for temperature

More information

The Experts in Vacuum Solutions

The Experts in Vacuum Solutions By Woodrow Farrow - Reprinted with permission from Specialty Gas Report Vacuum: a space that is relatively empty of matter, especially when that void has been created through artificial means. The earth

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

Lab 1. Adiabatic and reversible compression of a gas

Lab 1. Adiabatic and reversible compression of a gas Lab 1. Adiabatic and reversible compression of a gas Introduction The initial and final states of an adiabatic and reversible volume change of an ideal gas can be determined by the First Law of Thermodynamics

More information

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking

More information

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY Nengbing Xu, Hongmei Ying and Libo Zhu Ningbo Environmental Monitoring

More information

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion: Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion

More information

Birck Nanotechnology Center

Birck Nanotechnology Center EFFECTIVE DATE: October 20, 2016 PAGE 1 of 8 This instruction covers the set-up and use of the bench-top for annealing and oxidizing samples within the cleanroom. 1. SAFETY REQUIREMENTS 1.1 Safety glasses

More information

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Temperature controlled, multi-technique volumetric

More information

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology: Title: Comparison of Chemical Sensor and NDIR Technologies TSN Number: 25 File:\\MII- SRV1\Metron\Bridge_Analyzers\Customer_Service_Documentation\White_Papers\25_G en EGA NDIR vs Chemical Sensor.docx Created

More information

Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA

Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA Dear Roger, Your Ref: Our Ref: Direct Dial: 020 7901 7468 Email:

More information

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013 The SRI ASTM 3612/TOGA/ Dissolved Gas GC configuration ( TOGA GC ) permits measurement of gasses dissolved in water, oil or other liquids, and is based on the requirements of ASTM method 3612, option C,

More information

UNIVERSITY OF WATERLOO

UNIVERSITY OF WATERLOO UNIVERSITY OF WATERLOO Department of Chemical Engineering ChE 524 Process Control Laboratory Instruction Manual January, 2001 Revised: May, 2009 1 Experiment # 2 - Double Pipe Heat Exchanger Experimental

More information

University of MN, Minnesota Nano Center Standard Operating Procedure

University of MN, Minnesota Nano Center Standard Operating Procedure Equipment name: STS Etcher Badger name: STS Revision number: 3 Model: 320 Revisionist: Paul Kimani Location: Bay 3 Date: 1 October 2013 A. Description The 320 is a manually loaded batch plasma etching

More information

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel TANK MANAGER FOR TWO TANKS OPERATING MANUAL 10/31/11 C-More T6C L color touch panel 1 TABLE OF CONTENTS GENERAL...3 INSTALLATION...4 STONE TEST PROCEDURE...7 OPERATIONAL SUMMARY...7 AUTO CARBONATION...10

More information

TABLE OF CONTENTS INTRODUCTION 3 SAFETY PRECAUTIONS 3 PACKAGE CONTENTS 4 DEVICE OVERVIEW 4 BUTTON OPERATION SUMMARY 5 BASIC OPERATION 6

TABLE OF CONTENTS INTRODUCTION 3 SAFETY PRECAUTIONS 3 PACKAGE CONTENTS 4 DEVICE OVERVIEW 4 BUTTON OPERATION SUMMARY 5 BASIC OPERATION 6 TABLE OF CONTENTS INTRODUCTION 3 SAFETY PRECAUTIONS 3 PACKAGE CONTENTS 4 DEVICE OVERVIEW 4 BUTTON OPERATION SUMMARY 5 BASIC OPERATION 6 CURRENT TIME MODE 7 FUNCTIONAL DISPLAY 7 WEATHER FORECAST FEATURE

More information

DRAFT. Operating Procedures for the NPDGamma Liquid Hydrogen Target in TA-53, Building MPF-35

DRAFT. Operating Procedures for the NPDGamma Liquid Hydrogen Target in TA-53, Building MPF-35 Operating Procedures for the NPDGamma Liquid Hydrogen Target V0.03 11/26/05 1 DRAFT Operating Procedures for the NPDGamma Liquid Hydrogen Target in TA-53, Building MPF-35 Version 0.03 November 26, 2005

More information

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries. Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s):

More information

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler. Exercise 5-2 Bubblers EXERCISE OBJECTIVE Learn to measure the level in a vessel using a bubbler. DISCUSSION OUTLINE The Discussion of this exercise covers the following points: Bubblers How to measure

More information

APPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6

APPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6 APPENDIX C 40 CFR PART 60 App. B SPEC. 1-4, 6 Part 60, Appendix B - Performance Specification 2 Specifications and Test Procedures for SO2 and NOx Continuous Emission Monitoring Systems in Stationary Sources

More information

FOURIER TRANSFORM INFRARED SPECTROSCOPY

FOURIER TRANSFORM INFRARED SPECTROSCOPY Purpose and Scope: This document describes the procedures and policies for using the MSE Bruker FTIR Spectrometer. The scope of this document is to establish user procedures. Instrument maintenance and

More information

Chem 110 General Principles of Chemistry

Chem 110 General Principles of Chemistry CHEM110 Worksheet - Gases Chem 110 General Principles of Chemistry Chapter 9 Gases (pages 337-373) In this chapter we - first contrast gases with liquids and solids and then discuss gas pressure. - review

More information