1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor)

Size: px
Start display at page:

Download "1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor)"

Transcription

1 1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor) 1. According to NFPA 99, SCFM is an acronym for. a) Surface Conditions at Fahrenheit Mercury b) Standard Conditions at Fahrenheit Mercury c) Standard Cubic Feet per Minute d) Surface Conditions For Meters 2. According to NFPA 99, a pressure not exceeding 200 psig is considered a pressure. a) High b) Partial c) Remote d) Working 3. When brazing joints in a piped vacuum system, must be blown through the lines as a method of purging. (blow down ) a) Oil-free dry nitrogen b) Liquid nitrogen c) Soapy water d) Nitrous oxide 4. After installing a level I piped vacuum system, but before the vacuum source is connected, the system must have a standing pressure test for 24 hours at a minimum test pressure of. a) 30 psig b) 45 psig c) 60 psig d) 300 mm (12 in.) gage HgV 5. With regard to pressure differentials, the alternate test pressure for Nitrous Oxide is what? a) 20 psi b) 30 psi c) 40 psi d) 50 psi 1

2 6. The intake to a medical air compressor which uses an outside source of air shall be located above the ground at least. a) 10 feet b) 15 feet c) 20 feet d) 25 feet 7. Piping shall be supported from the building structure in accordance with. a) Local Code b) Plans c) MSS SP-58 d) MSS SP Which of the following is not true about oxygen cylinders. a) Grease shall not come in contact with valves or fittings b) Valves should be left open on empty stored cylinders c) Cylinders shall not be stored near heat ducts d) Oxygen shall never be substituted for compressed air. 9. Each section of the piping in medical gas and vacuum systems shall be subject to an initial pressure test and examined for leakage. The test pressure for gases shall be 1.5 times the system working pressure but not less than a gage pressure of. a) 60 psi b) 100 psi c) 125 psi d) 150 psi 10. When cleaning the interior surfaces of level I distribution piping and controls on site, clean, hot wateralkaline solution, such as sodium carbonate or trisodium phosphate may be used at a rate of. a) 1 gallon to 3 gallon of potable water b) 1 gallon to 4 gallons of potable water c) 1 gallon to 5 gallons of potable water d) 1 pound to 3 gallons of potable water 11. Which of the following is not a level 1 pipe labeling requirement? a) At intervals of not more than 20 feet b) At least once in or above every room c) On both sides of walls or partitions penetrated by the piping d) At least once every other story height traversed by risers 2

3 12. Flexible connectors of other than all-metal construction used to connect outlets of pressure regulators to fixed piping shall not exceed in length, for a level 3 piping system. a) 24" b) 48" c) 60" d) 72" 13. 1" nominal copper tubing for level 3 shall be supported from the building structure a maximum of feet on center. a) 6' b) 7' c) 8' d) 9' 14. A level I pressure relief valve will be set at psig for a gas central supply system operating at 110 psig. a) 110 b) 165 c) 175 d) Which of the following is a correct statement about medical air compressor systems? a) Medical air receivers shall be provided with a 3-valve bypass. b) Dryer systems shall be at least duplexed and valved. c) In no case shall there be fewer than 2 compressors. d) All of the above are correct. 16. The NFPA 99 scope includes the installation of nonflammable medical gas and vacuum systems with operating pressures below psi. a) 150 b) 200 c) 250 d) A cylinder of non-flammable medical gas, in storage, shall be protected or located so that the temperature of the cylinder, in no case exceeds degrees C. a) 54 b) 72 c) 130 d) 212 3

4 18. The enclosure surrounding stored level 1 central supply system oxygen or nitrous oxide cylinders shall be constructed of building materials with a fire-resistive rating of at least minutes. a) 15 b) 30 c) 45 d) Electrical devices or switches, in accordance with NFPA 70, shall be installed in locations not less than inches above finished floor to avoid physical damage. Assume a Level l system. a) 36" b) 42" c) 60" d) 72" 20. A level I system cylinder not connected for use shall have its valve protection cap. a) Removed until connection with the piping b) Secured tightly in place c) Labeled with red letters, minimum 2" in height d) Hand tight for ready removal 21. The compressor air intake for level 1 shall be located outdoors above roof level, at a minimum distance of from any door, window, exhaust, other intake, or opening in the building and a minimum distance of above the ground. a) 10', 10 b) 25', 25' c) 25', 10' d) 10', 20' 22. When a level 1 storage location's capacity for non-flammable gases is less than 3,000 cubic foot (connected and in storage), its natural ventilation means shall provide a minimum of square inches of total free area. a) 36 b) 72 c) 144 d) A shall be installed between each cylinder lead and the manifold header to prevent the loss of gas from the cylinders in the event the pressure relief device fails on cylinder systems without reserve supply. a) Shutoff valve b) Cylinder valve c) Check valve d) Flex hose 4

5 24. Where provided, that portion of the source equipment that automatically supplies the systems in the event of failure of the primary and secondary operation supply is called what? a) primary supply b) secondary supply c) reserve supply d) backup supply 25. Pressure relief valves for a gas central supply system that are set at 50 percent above normal line pressure shall vented to the outside if the total capacity of the supply system is cubic feet or more of gas. a) 3,000 b) 2,000 c) 1,500 d) 1, The organization, office, or person responsible for approving equipment, installations, or procedures is by definition. a) The Technical Correlating Committee b) The Authority Having Jurisdiction c) The Fire Marshall's office d) The Area Administrator 27. Evacuation waste gas is referenced in the section called. a) Suction tubing b) Indirect waste systems c) Exhaust hoods d) Waste Anesthetic Gas Disposal (WAGD) 28. In a level I gas distribution system, the master alarm warning system shall consist of alarm panels in two locations. a) 2 b) 2 or more c) 3 d) 3 or more 29. Alarm areas for anesthetizing gas delivery shall have the sensors installed for each such area. a) On the source side of any of the individual room zone valve box assemblies b) On the control panel c) Prior to the main trunk line supplying d) Upstream from the junction box 5

6 30. With regard to level I distribution, tubes shall be hard-drawn seamless copper ASTM B819 medical gas tube, Type L, except that where operating pressures are above a gauge pressure of 185 psi., Type K shall be used for sizes larger than O.D. a) 1/4" b) 3/8" c) 112" d) 3 1/8 in. 31. Turns, offsets, and other changes in direction in medical gas and vacuum piping shall be made with wrought copper capillary fittings or complying with ASME B a) Flared fittings b) Coupled unions c) Brazed fittings d) Compression type fittings 32. The minimum backfilled earth cover required above the top of buried piping located outside of buildings shall be, where physical damage to the piping is not likely to occur. a) 12" b) 18" c) 24" d) 36" 33. Medical gas riser piping shall be permitted in which location? a) Pipe shafts if protected from physical damage, heat, etc. b) Electrical service equipment of 660 volts c) Elevator shafts d) Kitchens 34. Flexible connectors, metallic or nonmetallic, shall have a minimum burst pressure of. a) 1,000 psia b) 1,000 psi c) 1,500 psia d) 1,500 psig 35. Level 1 tubes shall be hard-drawn seamless copper ASTM B 819 medical gas tube, Type L, except that where operation pressures are above a gauge pressure of 185 psi, shall be used for sizes larger than 3 1/8 in. O. D. a) Type M b) Type L c) Type K d) Soft 6

7 36. Mains, branches, and drops to individual service outlets in level 3 oxygen piping systems shall be not less than O. D. a) 1/8" b) 1/4" c) 3/8" d) 1/2" 37. A level 1 piped vacuum system source shutoff valve is placed. a) Immediately downstream of the receiver tank b) On the outlet side of the receiver tank, by the pump c) Between the zone valve and the patient outlet d) At the immediate connection of each source system to the distribution piping to permit the entire source, including all accessory devices. 38. During pressure testing of a level 1 vacuum system piping a brazed joint is found to be defective. The joint was repaired, testing resumed, and again the joint was found to be defective. Which of the following is acceptable, regarding the defective joint. a) Repair the joint again; it was probably due to flux residue b) Repair the joint again; it was probably due to excessive oxidation c) Repair the joint again, a crack was discovered in the tube d) Replace the joint, no joint shall reheated more than once before being replaced 39. In a level 1 piped gas system, the piping purge test shall be conducted by obtaining the appropriate adapter from the manufacturer and subjecting the pipe to a high purge rate of at least on each outlet. a) 30 gpm b) 8 SCFM c) 12 fpm d) 125 L / hour 40. Level 1 vacuum pumps and collection piping shall be capable of maintaining a vacuum of of mercury at the station inlet farthest away from the central vacuum source. a) 6 inches b) 12 inches c) 14.7 inches d) 6 hg 41. Generally, piping materials for field-installed positive pressure medical gas systems shall. a) Be seamless copper ASTM B 819 medical gas tube, Type L b) Be either aluminum or stainless steel c) May be made from cast iron d) Not be required to be brazed 7

8 42. Union joints may be made for medical gas and vacuum distribution pipeline systems when which of the following is accomplished? a) The unions are flared b) The unions are compression-type c) The unions are straight-threaded d) Unions are prohibited joints 43. Welded joints for medical gas and vacuum systems shall be permitted to be made using. a) Silver solder filler b) Copper-Phosphorus-Silver filler c) a Certified Plumbing Contractor d) Gas tungsten arc welding 44. Initial pressure tests for piping involving medical gases are done how? a) Using air or carbon dioxide at 1.5 times the system pressure not to be less than ISO psi b) Can be tested with 0 2 at 2 times the system pressure c) Using oil-free, dry nitrogen at 1.5 times the system pressure not to be less than ISO psi d) May be tested with CO at 2.0 times the system pressure not to be less than 250 psi 45. Which of the following is not correct with regard to standing pressure tests involving medical gases? a) The tests must maintain pressure for 24 hours b) The tests must be at least 20% above normal system pressure c) Test pressure must not be less than 300 mm (12 in.) gage HgV d) Test medium may be air or carbon dioxide 46. Where natural ventilation is permitted, it shall consist of 2 louvered openings, each having a minimum free area of with one located within 1 foot of the floor and one located within 1 foot of the ceiling. a) 50 square inches b) 72 square inches c) 100 square inches d) 144 square inches 47. Flexible connectors of other than all-metal construction used to connect outlets of pressure regulators to fixed piping shall not exceed in length and shall not penetrate walls, floors, ceilings, or partitions. a) 2 feet b) 4 feet c) 5 feet d) 10 feet 8

9 48. A piping verifier purity test's dew point shall not exceed 500 ppm at? a) 12 degrees C. b) 10 degrees F. c) 59 degrees F. d) 15 degrees C 49.Medical gas piping shall be permitted to be installed in the same service trench or tunnel with fuel gas lines, fuel oil lines, electrical lines, steam lines and similar utilities provided that the space is ventilated (naturally or mechanically) and the ambient temperature around the medical gas piping is limited to degrees F maximum. a) 54 b) 130 c) 250 d) Medical-surgical vacuum inlets shall draw without reducing the vacuum pressure below 12 inches gauge HgV at any adjacent station inlet a) 0.3 SCFM b) 3.0 SCFM c) 12 SCFM d) 30 SCFM 9

10 1 Exam Prep NFPA 99 Health Care Facilities Answers Answer: C Answer: D Answer: A , Answer: D Answer: C & Table & Table (B) 6 Answer: C (C) 7 Answer: D Answer C: Answer: B Answer: D Answer: D Answer: D (1)(2)(3)(4) (1)(2)(3)(4) 12 Answer: C Answer: C Table Table Answer: B (2) (2) 110 psig x 1.50 = 165 psig pressure relief valve is set 50% above the line pressure 15 Answer: D , , , (B), (D), , (E), (F) 16 Answer: D (1) (1) 17 Answer: A and and Answer: D (4) (4) 19 Answer: C (5) A (5) 20 Answer: B Answer: D (C) & (D) 22 Answer: C (E) 72 x2 = Answer: C Appendix A, Figure A (a) Figure A (a) Caption: "Gas-specific cylinder lead connector contains a check valve" 24 Answer: C Answer: A (4) (4) 10

11 Answer: B Answer: D , , see evacuation - waste gas, Answer: B Answer: A (2) (2) 30 Answer: D Answer: C Answer: B Answer: A , , Answer: B , Answer: C Answer: D (1) 37 Answer: D Answer: D Answer: B Answer: B Answer: A , , Answer: D Answer: D Answer: C , , Answer: D , , , , , , Answer: B , Answer: C Answer: B (5) Answer: B Answer: B

Safety Policy Manual Policy No. 133

Safety Policy Manual Policy No. 133 Policy: Installation, Testing, and Maintenance of Medical Gas Page 1 of 2 APPLICATION NYU Langone Medical Center (NYULMC) PURPOSE To provide guidelines for safe and proper installation, testing and maintenance

More information

MEDICAL EQUIPMENT CATALOG

MEDICAL EQUIPMENT CATALOG THE MOST TRUSTED MEDICAL EQUIPMENT IN NORTH AMERICA MEDICAL EQUIPMENT CATALOG Total Components Including Bulk Oxygen Supply, Back up and Delivery Equipment Designs for High Pressure, Liquid or Bulk Applications

More information

Section 1.3: PRESSURE VESSEL COMPONENTS AND SYSTEMS AND COMPRESSED GAS CYLINDERS

Section 1.3: PRESSURE VESSEL COMPONENTS AND SYSTEMS AND COMPRESSED GAS CYLINDERS Section 1.3: PRESSURE VESSEL COMPONENTS AND SYSTEMS AND COMPRESSED GAS CYLINDERS TABLE OF CONTENTS Page A. Regulations, Standards and References 1.3-2 B. Scope 1.3-2 C. Design Features for the Storage

More information

Installation Operating Instructions for Duplex Manual Manifolds PX-TMD Series

Installation Operating Instructions for Duplex Manual Manifolds PX-TMD Series Introduction Powerex manifolds are cleaned, tested and prepared for the indicated gas service and are built in accordance with the Compressed Gas Association guidelines. The manifold consists of a regulator

More information

Cal/OSHA T8 CCR 1536 Cal/OSHA T8 CCR 4799 Cal/OSHA T8 CCR 4845 Cal/OSHA T8 CCR 4848

Cal/OSHA T8 CCR 1536 Cal/OSHA T8 CCR 4799 Cal/OSHA T8 CCR 4845 Cal/OSHA T8 CCR 4848 Cal/OSHA Gas Systems for Welding GAS WELDING Cal/OSHA T8 CCR 1536 Cal/OSHA T8 CCR 4799 Cal/OSHA T8 CCR 4845 Cal/OSHA T8 CCR 4848 When performing gas welding, the following precautions, work procedures,

More information

CHAPTER 13 HEALTH CARE FACILITIES AND MEDICAL GAS AND VACUUM SYSTEMS

CHAPTER 13 HEALTH CARE FACILITIES AND MEDICAL GAS AND VACUUM SYSTEMS CHAPTER 13 HEALTH CARE FACILITIES AND MEDICAL GAS AND VACUUM SYSTEMS < Part I Special Requirements for 1301.0 Application. Health Care Facilities. 1301.1 Construction and equipment requirements shall be

More information

1 Exam Prep. Tabs and Highlights

1 Exam Prep. Tabs and Highlights 1 Exam Prep NFPA 14: Standard for the Installation of Standpipe and Hose Systems Tabs and s These 1 Exam Prep tabs are based on the NFPA 14: Standard for the Installation of Standpipe and Hose Systems,

More information

I. Subject: Compressed gas cylinders, regulators, and portable liquid systems.

I. Subject: Compressed gas cylinders, regulators, and portable liquid systems. I. Subject: Compressed gas cylinders, regulators, and portable liquid systems. II. Objective: To provide safe and effective use, storage, and transport of therapeutic gases. II. Procedure: A. Compressed

More information

Tri-Tech Medical Inc.

Tri-Tech Medical Inc. Introduction Tri-Tech Medical manifolds are cleaned, tested and prepared for the indicated gas service and are built in accordance with the Compressed Gas Association guidelines. The manifold consists

More information

SECTION DOMESTIC WATER PIPING

SECTION DOMESTIC WATER PIPING PART 1 - GENERAL 1.1 RELATED DOCUMENTS A. Drawings and general provisions of the Contract, including General and Supplementary Conditions Specification Sections, apply to this Section. B. Related Sections

More information

First Correlating Revision No. 2-NFPA [ Section No ] Submitter Information Verification. Committee Statement. 1 of 39 1/29/2016 8:53 AM

First Correlating Revision No. 2-NFPA [ Section No ] Submitter Information Verification. Committee Statement. 1 of 39 1/29/2016 8:53 AM 1 of 39 1/29/2016 8:53 AM First Correlating Revision No. 2-NFPA 99-2015 [ Section No. 1.3.4 ] 1.3.4 Patient Care Spaces. Global FR-113 1.3.4.1 The governing body health care facility s governing body of

More information

CHAPTER 13 HEALTH CARE FACILITIES AND MEDICAL GAS AND VACUUM SYSTEMS

CHAPTER 13 HEALTH CARE FACILITIES AND MEDICAL GAS AND VACUUM SYSTEMS CHAPTER 13 HEALTH CARE FACILITIES AND MEDICAL GAS AND VACUUM SYSTEMS Part I Special Requirements for Health Care Facilities. 1301.0 Application. 1301.1 Where Required. Construction and equipment requirements

More information

FLAMMABLE GASES AND FLAMMABLE CRYOGENIC FLUIDS

FLAMMABLE GASES AND FLAMMABLE CRYOGENIC FLUIDS CHAPTER 35 FLAMMABLE GASES AND FLAMMABLE CRYOGENIC FLUIDS SECTION 3501 GENERAL 3501.1 Scope. The storage and use of flammable gases shall be in accordance with this chapter. Compressed gases shall also

More information

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series Introduction Powerex manifolds are cleaned, tested and prepared for the indicated gas service and are built in accordance with the Compressed Gas Association guidelines. The manifold consists of a regulator

More information

Installation Instructions for NP100 Series Nitrogen Control Panels

Installation Instructions for NP100 Series Nitrogen Control Panels Installation Instructions for NP100 Series Nitrogen Control Panels Installation of the Nitrogen Control Panel involves installing the rough-in box and making the necessary piping connections in accordance

More information

CHAPTER 13. HEAlTH CARE FACiliTiEs And MEdiCAl GAs And VACuuM systems. Part i special Requirements for Health Care Facilities.

CHAPTER 13. HEAlTH CARE FACiliTiEs And MEdiCAl GAs And VACuuM systems. Part i special Requirements for Health Care Facilities. CHAPTER 13 HEAlTH CARE FACiliTiEs And MEdiCAl GAs And VACuuM systems Part i special Requirements for Health Care Facilities. 1301.0 Application. 1301.1 Where Required. Construction and equipment requirements

More information

WELDING AND OTHER HOT WORK

WELDING AND OTHER HOT WORK CHAPTER 26 WELDING AND OTHER HOT WORK SECTION 2601 GENERAL 2601.1 Scope. Welding, cutting, open torches and other hot work operations and equipment shall comply with this chapter. Exception: Hot work on

More information

READ ONLY. Adopting Agency BSC SFM. Adopt Entire Chapter Adopt Entire Chapter as amended (amended sections listed below) X X X X

READ ONLY. Adopting Agency BSC SFM. Adopt Entire Chapter Adopt Entire Chapter as amended (amended sections listed below) X X X X CALIFORNIA PLUMBING CODE MATRIX ADOPTION TABLE CHAPTER 13 - HEALTH CARE FACILITIES AND MEDICAL GAS AND VACUUM SYSTEMS (Matrix Adoption Tables are non-regulatory, intended only as an aid to the user. See

More information

University of Delaware

University of Delaware SECTION 22 60 00_AIR, GAS AND VACUUM PIPING SYSTEMS University Contact: Energy & Engineering Group (302) 831-1744 PART 1 GENERAL 1.1 SUMMARY A. This standard includes the requirements for compressed air,

More information

Storage Systems and Accessories

Storage Systems and Accessories Storage Systems and Accessories High Pressure Air and Inert Gases 4500-6000 PSIG DOT and ASME Cylinders High Pressure Storage Systems If an application requires storage, either in the form of a single

More information

Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions

Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions FM-171 Rev. B 8/00 Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions MODEL 6300-1 PORTER INSTRUMENT COMPANY, INC. 245 TOWNSHIP LINE RD. P.O. BOX 907 HATFIELD, PA 19440-0907

More information

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855

More information

Piped Air SCBA Refilling System Standard

Piped Air SCBA Refilling System Standard WALT WHITE Fire Chief 5770 Freeport Blvd., Suite 200 Sacramento, CA 95822-3516 Ph: (916) 808-1300 Fax: (916) 808-1629 www.sacfire.org Piped Air SCBA Refilling System Standard SCOPE: This specification

More information

Manifolds & pressure reducing sets

Manifolds & pressure reducing sets Manifolds & pressure reducing sets Automatic Changeover Manifold (Electronic) Providing continuous supply of medical gases Pneumatech Medical Gas Solutions (PMGS) PureGAS Manifolds provide a duty or standby

More information

UTMB RESPIRATORY CARE SERVICES GUIDELINES Respiratory Care Services Medical Gas Cylinder Storage, Use and Transport

UTMB RESPIRATORY CARE SERVICES GUIDELINES Respiratory Care Services Medical Gas Cylinder Storage, Use and Transport Page 1 of 4 I. Title Medical Gas. II. III. IV. Policy This policy applies to all departments and areas where medical gases are used, stored or transported for patient care activities. Respiratory Care

More information

INTERNATIONAL ASSOCIATION OF PLUMBING AND MECHANICAL OFFICIALS

INTERNATIONAL ASSOCIATION OF PLUMBING AND MECHANICAL OFFICIALS INTERNATIONAL ASSOCIATION OF PLUMBING AND MECHANICAL OFFICIALS IAPMO GUIDE CRITERIA FOR WATER HEATER SUPPLY SHUTOFF VALVE WITH A / THERMAL EXPANSION CONTROL VALVE IGC 128-2006a8 1. PURPOSE 1.1 The purpose

More information

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1 Dry 131h 1. DESCRIPTION The Viking Model G-1 Maintenance Air Compressor is an electric motor-driven, aircooled, single-stage, oil-less compressor. The unit is equipped with a check valve and provides a

More information

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS Annex 3, page 2 ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS The text of existing chapter 15 is replaced by the following: "1 Application This

More information

Types 749B and R130 Changeover Manifolds

Types 749B and R130 Changeover Manifolds Instruction Manual MCK-1179 Types 749B and R130 June 2012 Types 749B and R130 Changeover Manifolds TYPE HSRL-749B TYPE 64SR/122 TYPE R130/21 TYPE 749B/21 Figure 1. Changeover Manifolds and Regulator Assemblies

More information

Flexible Metal Hose Products

Flexible Metal Hose Products Flexible Metal Hose Products Specialty Flexible Hose Solutions of Stainless Steel The AEROCOM Advantage Aerocom offers engineered solutions that address specific flexible piping challenges such as vibration,

More information

Installation & Maintenance Manual. Economy Manifold

Installation & Maintenance Manual. Economy Manifold Installation & Maintenance Manual Economy Manifold Contents User Responsibility 3 Introduction 4 Features 4 Description of the Manifold 5 Shipment Details 5 Description of Parts 5-6 Pressure Regulators

More information

GUIDELINES FOR CONDUCTING HYDROGENATION REACTIONS

GUIDELINES FOR CONDUCTING HYDROGENATION REACTIONS Page 1 of 5 GUIDELINES FOR CONDUCTING 1. Definition of Hydrogenation and Hydrogenolysis Reactions Hydrogenation reactions are synthetic transformations involving hydrogen, metallic catalysts (such as nickel,

More information

CALIFORNIA PLUMBING CODE MATRIX ADOPTION TABLE CHAPTER 13 - HEALTH CARE FACILITIES AND MEDICAL GAS AND VACUUM SYSTEMS 1 2 1/AC AC SS SS/CC

CALIFORNIA PLUMBING CODE MATRIX ADOPTION TABLE CHAPTER 13 - HEALTH CARE FACILITIES AND MEDICAL GAS AND VACUUM SYSTEMS 1 2 1/AC AC SS SS/CC CALIFORNIA PLUMBING CODE MATRIX ADOPTION TABLE CHAPTER 13 - HEALTH CARE FACILITIES AND MEDICAL GAS AND VACUUM SYSTEMS Adopting Agency BSC SFM HCD DSA OSHPD CSA DPH AGR DWR CA 1 2 1/AC AC SS SS/CC 1 2 3

More information

IAPMO GUIDE CRITERIA FOR BALL VALVES IAPMO IGC PURPOSE

IAPMO GUIDE CRITERIA FOR BALL VALVES IAPMO IGC PURPOSE INTERNATIONAL ASSOCIATION OF PLUMBING AND MECHANICAL OFFICIALS IAPMO GUIDE CRITERIA FOR BALL VALVES IAPMO IGC 157-20067 1 PURPOSE 1.1 The purpose of this standard is to establish an acceptable standard

More information

CITY AND COUNTY OF DENVER CR&CF RISK UNIT Compressed Gas Safety Standard

CITY AND COUNTY OF DENVER CR&CF RISK UNIT Compressed Gas Safety Standard CITY AND COUNTY OF DENVER CR&CF RISK UNIT 65.5.11 Compressed Gas Safety Standard 1.0 Introduction 2.0 Scope This standard has been developed to protect all City and County of Denver employees and contractors

More information

TECHNICAL DATA. Q= Cv S

TECHNICAL DATA. Q= Cv S Page 1 of 13 1. DESCRIPTION The Viking 4 inch Model G-4000 Dry Valve Riser Assembly consists of a small profile, light weight, pilot operated valve that is used to separate the water supply from the dry

More information

Inert Air (N2) Systems Manual

Inert Air (N2) Systems Manual INSTRUCTION MANUAL Inert Air (N2) Systems Manual N2-MANUAL 2.10 READ AND UNDERSTAND THIS MANUAL PRIOR TO OPERATING OR SERVICING THIS PRODUCT. GENERAL INFORMATION Positive pressure nitrogen gas pressurizing

More information

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855

More information

Cash Valve TYPE KP PILOT OPERATED BACK PRESSURE VALVE. ISSUED - DECEMBER 2000 CAVMC-0518-US-0208 ISO 9001 Certified

Cash Valve TYPE KP PILOT OPERATED BACK PRESSURE VALVE. ISSUED - DECEMBER 2000 CAVMC-0518-US-0208 ISO 9001 Certified Cash Valve PILOT OPERATED BACK PRESSURE VALVE ISSUED - DECEMBER 2000 CAVMC-0518-US-0208 ISO 01 Certified KP HIGH CAPACITY PILOT OPERATED BACK PRESSURE VALVE DESCRIPTION The Cash Valve Type KP is a pilot

More information

PIP PNE00012 Piping Examination and Leak Test Guide

PIP PNE00012 Piping Examination and Leak Test Guide October 2017 Piping PIP PNE00012 Piping Examination and Leak Test Guide PURPOSE AND USE OF PROCESS INDUSTRY PRACTICES In an effort to minimize the cost of process industry facilities, this Practice has

More information

Flow meter. bellow vaporizer. APL valve. Scavenging system

Flow meter. bellow vaporizer. APL valve. Scavenging system Introductory Lecture Series: The Anesthesia Machine PORNSIRI WANNADILOK Objectives Anesthesia Machine Ventilators Scavenging Systems System Checkout 1 Flow meter ventilator bellow vaporizer Corrugated

More information

SECTION GAS SYSTEMS FOR LABORATORY AND HEALTHCARE FACILITIES

SECTION GAS SYSTEMS FOR LABORATORY AND HEALTHCARE FACILITIES SECTION 22 63 00 GAS SYSTEMS FOR LABORATORY AND HEALTHCARE FACILITIES PART 1 - GENERAL 1.1 DESCRIPTION SPEC WRITER NOTES: 1. Delete between // ---- // if not applicable to project. Also delete any other

More information

TECHNICAL DATA 3 MODEL G-3000 DRY VALVE RISER ASSEMBLY

TECHNICAL DATA 3 MODEL G-3000 DRY VALVE RISER ASSEMBLY Page 1 of 13 1. DESCRIPTION The Viking 3 Model G-3000 Dry Valve Riser Assembly is equipped with a small profile, light weight, pilot operated valve that is used to separate the water supply from the dry

More information

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS COMPACT TRIPLE WALL OUTLET STATION MODEL 6255-1 INSTALLATION AND OPERATING INSTRUCTIONS To assure safe operation and conformation to local fire codes, all Porter Outlet Stations are designed to be used

More information

WaterSource Heat Pump Hose Kit Performance Specification-Summary

WaterSource Heat Pump Hose Kit Performance Specification-Summary WaterSource Heat Pump Hose Kit Performance Specification-Summary AUTOMATIC & MANUAL BALANCING HOSE KITS Contractor shall provide and install Hays Hose Kits. Each kit, shall include a Hays 2500 Series Mesurflo

More information

COMPRESSED GASES CHAPTER 53

COMPRESSED GASES CHAPTER 53 CHAPTER 53 COMPRESSED GASES SECTION 5301 GENERAL 5301.1 Scope. Storage, use and handling of compressed gases in compressed gas containers, cylinders, tanks and systems shall comply with this chapter, including

More information

TECHNICAL DATA. than the water inlet pressure to the concentrate

TECHNICAL DATA. than the water inlet pressure to the concentrate Foam102a 1. DESCRIPTION The Viking Low Flow Foam/Water proportioning system, is a UL Listed and FM Approved system, for use with 3M foam concentrates. This system consists of a standard wet pipe sprinkler

More information

TECHNICAL DATA Q = C. v P S. 2 Model G-2000 Dry valve. Page 1 of 13

TECHNICAL DATA Q = C. v P S. 2 Model G-2000 Dry valve. Page 1 of 13 Page 1 of 13 1. Description The Viking 2 Model G-2000 Dry Valve Riser Assembly consists of a small profile, light weight, pilot operated valve that is used to separate the water supply from the dry sprinkler

More information

TOTAL CAPABILITIES IN THE PIPELINE INDUSTRY. UTILITY TECHNOLOGIES INTERNATIONAL CORPORATION Cincinnati Columbus West Jefferson

TOTAL CAPABILITIES IN THE PIPELINE INDUSTRY. UTILITY TECHNOLOGIES INTERNATIONAL CORPORATION Cincinnati Columbus West Jefferson TOTAL CAPABILITIES IN THE PIPELINE INDUSTRY UTILITY TECHNOLOGIES INTERNATIONAL CORPORATION Cincinnati Columbus West Jefferson OGA SEMINAR DECEMBER 3, 2012 OGA Seminar 2012 2 Regulators Meters Valves Flanges

More information

Compressed Gases Storage and Handling Safety Policy. Individual Unit Function: Safety Procedure No.: SOP-0103 Page: 1 of 6

Compressed Gases Storage and Handling Safety Policy. Individual Unit Function: Safety Procedure No.: SOP-0103 Page: 1 of 6 Procedure No.: SOP-0103 Page: 1 of 6 MIDGA Reviewed: 07/09/18 Effective: 1/14/98 Supersedes: Original Preparer: Owner: Approver: Safety Safety Safety 1. PURPOSE The purpose of this procedure is to establish

More information

AIR DRIVEN. Gas Boosters & SYSTEMS. Accepting VISA, MasterCard and American Express

AIR DRIVEN. Gas Boosters & SYSTEMS. Accepting VISA, MasterCard and American Express AIR DRIVEN Gas Boosters & SYSTEMS Accepting VISA, MasterCard and American Express Gas Boosters AIR DRIVEN FROM 30 PSI TO 21,750 PSI MAXIMATOR Maximator gas boosters are an excellent alternative to high

More information

TECHNICAL DATA. Q = C v P S

TECHNICAL DATA. Q = C v P S Page 1 of 13 1. DESCRIPTION The Viking 6 Model G-6000 Dry Valve Riser Assembly consists of a small profile, light weight, pilot operated valve that is used to separate the water supply from the dry sprinkler

More information

Economy Manifold Alert -2. Installation and Maintenance Manual V1.6

Economy Manifold Alert -2. Installation and Maintenance Manual V1.6 Economy Manifold Alert -2 Installation and Maintenance Manual V1.6 Corporation Table of Contents User Responsibility 4 Introduction 4 Features 4 Description of the Manifold 5 Shipment Details 5 The Collapsible

More information

TYPICAL TOXIC GAS SYSTEM INSPECTION VIOLATIONS

TYPICAL TOXIC GAS SYSTEM INSPECTION VIOLATIONS TYPICAL TOXIC GAS SYSTEM INSPECTION VIOLATIONS The following is a list of typical violations often found by inspectors and a generic solution. You can use this list to improve the safety of your facility,

More information

29 CFR Ch. XVII ( Edition)

29 CFR Ch. XVII ( Edition) 1910.253 (ii) Field shop operations. Where field shop operations are involved for fabrication of fittings, river crossings, road crossings, and pumping and compressor stations the requirements of paragraphs

More information

SECTION PRESSURE TESTING OF PIPING

SECTION PRESSURE TESTING OF PIPING SECTION 15044 PRESSURE TESTING OF PIPING PART 1 - GENERAL 1.01 DESCRIPTION A. Scope of Work: This section specifies the leakage and pressure testing of piping. B. Testing Records: PART 2 - PRODUCTS 2.01

More information

Installation & Maintenance Manual. C2 & C3 Manifolds

Installation & Maintenance Manual. C2 & C3 Manifolds Installation & Maintenance Manual C2 & C3 Manifolds w w w. a m i c o. c o m Table of Contents User Responsibility 3 Introduction 4 Features 4 Description of the Manifold 5 Shipment Details 5 Description

More information

INTERLOCKED METAL HOSE

INTERLOCKED METAL HOSE INTERLOCKED METAL HOSE HOSE FUNDAMENTALS Penflex interlocked metal hoses are made for a variety of specialized industrial applications and are available in a wide range of metals, styles and sizes. Interlocked

More information

SPECIFICATIONS. Approximate Weight: 10 oz. Surface Finish: Ra micro inch or less MATERIALS OF CONSTRUCTION

SPECIFICATIONS. Approximate Weight: 10 oz. Surface Finish: Ra micro inch or less MATERIALS OF CONSTRUCTION DIFFUSION-RESISTANT, DIAPHRAGM SEAL, EXCESS FLOW SHUT-OFF VALVES (FS SERIES) FS Series Excess Flow Shut-Off Valves are designed to automatically shut-off the delivery of gas in a line if the flow exceeds

More information

Wet pipe low flow foam/water system

Wet pipe low flow foam/water system December 6, 2010 Foam 14a 1. The Viking Low Flow Foam/Water proportioning system, is a UL Listed and FM Approved system, for use with Viking supplied foam concentrates. This system consists of a standard

More information

SERIES 30. Spring Operated Tank Blanketing Valve PROTECTOSEAL.

SERIES 30. Spring Operated Tank Blanketing Valve PROTECTOSEAL. SERIES 30 PROTECTOSEAL 1 2" NPT inlet and outlet standard Direct acting valve mechanism Optional flanged or threaded inlet and outlet connections available Inlet gas pressures from 10 PSIG to 200 PSIG

More information

Process gas requirements (Core, VWI, and OptiMix gas connect consoles)

Process gas requirements (Core, VWI, and OptiMix gas connect consoles) Process gas requirements (Core, VWI, and OptiMix gas connect consoles) You must supply the process gases and supply gas plumbing for your cutting system. See Table 7 on page 46 for supply gas quality,

More information

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM ASSE STANDARD #1048 - REVISED: 2009 Double Check Detector Fire Protection Backflow Prevention Assemblies Separate, complete laboratory

More information

VACUUM TESTING PRECAST CONCRETE MANHOLES

VACUUM TESTING PRECAST CONCRETE MANHOLES 1 OF 5 testing is a quick, safe and practical way to validate manhole system integrity. Manhole sections can be tested at the precast concrete plant prior to delivery or on site prior to backfilling. Here

More information

SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER

SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER SUPPLIMENT TO eom IMPORTANT READ THIS MANUAL BEFORE PRODUCT INSTALLATION, OPERATION, INSPECTION & MAINTENANCE Tougher and more rigid guidelines are being established

More information

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY FOR USE

More information

Optimizing Gas Supply for Industrial Lasers

Optimizing Gas Supply for Industrial Lasers Optimizing Gas Supply for Industrial Lasers Laser cutting of metals and other materials has grown rapidly due to developments in laser power, advancements in CNC automation, and decreasing costs. The industrial

More information

SINGLE STAGE ULTRA-HIGH PRESSURE REDUCING REGULATOR INSTALLATION AND OPERATION INSTRUCTIONS. Certified ISO 9001

SINGLE STAGE ULTRA-HIGH PRESSURE REDUCING REGULATOR INSTALLATION AND OPERATION INSTRUCTIONS. Certified ISO 9001 SINGLE STAGE ULTRA-HIGH PRESSURE REDUCING REGULATOR INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Warning: An appropriately sized pressure

More information

Installation and Operating Instructions For MANUAL MANIFOLDS SD and SDHP SERIES

Installation and Operating Instructions For MANUAL MANIFOLDS SD and SDHP SERIES Installation and Operating Instructions For MANUAL MANIFOLDS SD and SDHP SERIES INTRODUCTION Western manifold systems are cleaned, tested and prepared for the indicated gas service and are built in accordance

More information

Laboratory Gases CEU 234. Continuing Education from the American Society of Plumbing Engineers. April ASPE.ORG/ReadLearnEarn

Laboratory Gases CEU 234. Continuing Education from the American Society of Plumbing Engineers. April ASPE.ORG/ReadLearnEarn Laboratory Gases Continuing Education from the American Society of Plumbing Engineers April 2016 ASPE.ORG/ReadLearnEarn CEU 234 READ, LEARN, EARN Note: In determining your answers to the CE questions,

More information

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,

More information

1800CPB2 Service Regulators Maximum Inlet Pressure 125 PSIG

1800CPB2 Service Regulators Maximum Inlet Pressure 125 PSIG A C E D SB-8520.2 1800CPB2 Service Regulators Maximum Inlet 125 PSIG AMC Quality System QMI is Accredited by: C R E D I T ISO 9001 Certified Certificate #006697 R E ANSI RAB G I S T R A R Dutch Council

More information

AREA VALVE SERVICE UNITS

AREA VALVE SERVICE UNITS AREA VALVE SERVICE UNITS. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com U.K., SK3 0SB www.p3-phoenix.com

More information

Installation & Operating Instructions for Genesys Analog Series Manifolds

Installation & Operating Instructions for Genesys Analog Series Manifolds Manufacturer of Medical Gas Piping Equipment Installation & Operating Instructions for Genesys Analog Series Manifolds Models NPCU & TMCU Model TMLU - 1 - Genesys Analog Manifold Systems Features & Benefits

More information

Protocol Switchover Station

Protocol Switchover Station ADI 3215-B Certified ISO 9001 Protocol Switchover Station INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls Corporation of America

More information

632 Series Gas Distribution System

632 Series Gas Distribution System ADI3223-B Certified ISO 9001:2000 632 Series Gas Distribution System INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls Corporation

More information

DSEAR- Storage, Use and handling of gas cylinders guidance.

DSEAR- Storage, Use and handling of gas cylinders guidance. OHSS: Storage, Use and handling of gas cylinders guidance 202 DSEAR- Storage, Use and handling of gas cylinders guidance. Contents Introduction... 2 General recommendations for storage, use and handling....

More information

v. Size shall be specified on drawings.

v. Size shall be specified on drawings. SPECIFICATION FOR VACUUM INSULATED PIPING Part 1 General 1. Submittals a. After award of contract and before executing any manufacturing, shop drawings and specifications shall be submitted to the customer

More information

Eliminating Medical Gas Shutdowns for 10 years

Eliminating Medical Gas Shutdowns for 10 years Eliminating Medical Gas Shutdowns for 10 years Smart Tap is a tool used by hospitals. It has been used OVER 5000 TIMES to accomplish: Relocation of Bulk Gas Systems Relocation of zone valves Tie-ins of

More information

Gas Cabinets and Valve Manifold Boxes: Applying CGA G13- Rev 2015 to Gas Delivery Equipment

Gas Cabinets and Valve Manifold Boxes: Applying CGA G13- Rev 2015 to Gas Delivery Equipment Gas Cabinets and Valve Manifold Boxes: Applying CGA G13- Rev 2015 to Gas Delivery Equipment The Standard Governing the Storage & Use of Silane: Compressed Gas Association s Publication CGA G-13 2015: Storage

More information

SUBSTATION MAINTENANCE ELECTRICAL OPERATING PROCEDURE

SUBSTATION MAINTENANCE ELECTRICAL OPERATING PROCEDURE Page 1 of 6 A. PURPOSE AND SCOPE This procedure establishes safe working practices for handling sulfur hexafluoride (SF 6 ) gas and release minimization practices. This document also specifies methods

More information

ASHE Monograph. Medical Gas Cylinder and Bulk Tank Storage

ASHE Monograph. Medical Gas Cylinder and Bulk Tank Storage ASHE Monograph Medical Gas Cylinder and Bulk Tank Storage ASHE Monograph Medical Gas Cylinder and Bulk Tank Storage Susan B. McLaughlin, MBA, FASHE, CHFM, CHSP David A. Dagenais, CHSP, CHFM, SASHE JUNE

More information

ASSE International Product (Seal) Listing Program

ASSE International Product (Seal) Listing Program ASSE International Product (Seal) Listing Program ASSE 1070-2015 / ASME A112.1070-2015 / CSA B125.70-15 Performance Requirements for Water Temperature Limiting Devices Manufacturer: Contact Person: E-mail:

More information

NAVSEA STANDARD ITEM. 2.2 T9074-AS-GIB-010/271, Requirements for Nondestructive Testing Methods

NAVSEA STANDARD ITEM. 2.2 T9074-AS-GIB-010/271, Requirements for Nondestructive Testing Methods NAVSEA STANDARD ITEM ITEM NO: 009-53 DATE: 29 JUL 2011 CATEGORY: II 1. SCOPE: 1.1 Title: Bolted Bonnet Steam Valve; repair (shop) 2. REFERENCES: 2.1 S9086-CJ-STM-010/CH-075, Fasteners 2.2 T9074-AS-GIB-010/271,

More information

South-Tek Systems - Nitrogen Generation Corrosion Inhibiting System. Designed for: Dry or Preaction Fire Protection Systems (FPS)

South-Tek Systems - Nitrogen Generation Corrosion Inhibiting System. Designed for: Dry or Preaction Fire Protection Systems (FPS) South-Tek Systems - Nitrogen Generation Corrosion Inhibiting System Designed for: Dry or Preaction Fire Protection Systems (FPS) 1.0 Description of Work 1.1 The Fire Sprinkler Contractor shall provide

More information

September 2018 Pressure Test

September 2018 Pressure Test September 2018 Pressure Test The requirements for a pressure test are found in NFPA 54, Chapter 8, Sections 8.1-8.1.53 of the 2012 edition. A pressure test shall not be confused with a leak check. The

More information

Pressure Regulators. Operating Instructions. Instrumentation

Pressure Regulators. Operating Instructions. Instrumentation Pressure Regulators Operating Instructions FAILURE OR IMPROPER SELECTION OR IMPROPER USE OF THIS PRODUCT CAN CAUSE DEATH, PERSONAL INJURY AND PROPERTY DAMAGE. This document and other information from the

More information

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL www.modsafe.com SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device.

More information

623 Series Dome Load Regulators

623 Series Dome Load Regulators ADI 3202C Certified ISO 9001 623 Series Dome Load Regulators INSTALLATION AND OPERATION INSTRUCTIONS This Equipment is intended to be installed and operated by trained personnel. Before Installing or Operating,

More information

Basic & Controlled Atmosphere Glove Boxes

Basic & Controlled Atmosphere Glove Boxes I N T R O D U C T I O N Precise Basic and Controlled Atmosphere Glove Boxes provide a leak-tight environment for work with contamination-sensitive materials. Precise Basic Glove Boxes are simple, economical

More information

Oxygen and other Compressed Medical Gas Rules and Regulations Complying with the Law. Regulatory agencies:

Oxygen and other Compressed Medical Gas Rules and Regulations Complying with the Law. Regulatory agencies: Oxygen and other Compressed Medical Gas Rules and Regulations Complying with the Law BACKGROUND Cylinders containing gases (i.e. oxygen, nitrogen, nitrous oxide) are under high pressure and can be a combustion,

More information

MSC Guidelines for Refrigeration Machinery

MSC Guidelines for Refrigeration Machinery Procedure Number E1-32 Revision Date: 07/27/10 S. J. Kelly, CDR, Chief of Engineering Division References: a. 46 CFR Part 56.50-105 Low Temperature Piping b. 46 CFR Part 58.20 Refrigeration Machinery c.

More information

SEMASPEC Test Method for Determination of Cycle Life of Automatic Valves for Gas Distribution System Components

SEMASPEC Test Method for Determination of Cycle Life of Automatic Valves for Gas Distribution System Components SEMASPEC Test Method for Determination of Cycle Life of Automatic Valves for Gas Distribution System Components Technology Transfer 90120395B-STD and the logo are registered service marks of, Inc. 1996,

More information

AUTOMATIC GAS MANIFOLD

AUTOMATIC GAS MANIFOLD Corporation AUTOMATIC GAS MANIFOLD 200-SERIES LED Andersen Medical Gas 12 Place Lafitte Madisonville, LA 70447 http://www.themedicalgas.com 1-866-288-3783 Operation Maintenance and Installation Manual

More information

CASH VALVE TYPE KP BACK PRESSURE VALVES

CASH VALVE TYPE KP BACK PRESSURE VALVES A high capacity pilot operated back pressure valve that offers accurate control and dependable protection against overpressure conditions FEATURES Automatically maintains maximum pressure in a vessel or

More information

THE IMPACT ON ENERGY CONSUMPTION CAUSED BY PRESSURE DROP IN A COMPRESSED AIR SYSTEM

THE IMPACT ON ENERGY CONSUMPTION CAUSED BY PRESSURE DROP IN A COMPRESSED AIR SYSTEM THE IMPACT ON ENERGY CONSUMPTION CAUSED BY PRESSURE DROP IN A COMPRESSED AIR SYSTEM What is pressure drop in a compressed air utility? Pressure drop is a term used to characterize the reduction in air

More information

GAS DELIVERY PRESSURE AND MEASUREMENT

GAS DELIVERY PRESSURE AND MEASUREMENT GAS DELIVERY PRESSURE AND MEASUREMENT I. PURPOSE: To establish a uniform basis on which to provide gas measurement and pressure regulation facilities on customers premises. II. DELIVERY PRESSURES AND PROVISIONS:

More information

ECS Protector Nitrogen Inerting Vent (PAV-WN)

ECS Protector Nitrogen Inerting Vent (PAV-WN) ECS Protector (PAV-WN) U.S. Pat. No. 8,636,023, 9,526,933 and 9,717,935 B2 Specifications Stock Number: Service Pressure: Regulator Setting: System Connection: Temperature Range: Dimensions: Weight: Clear

More information

Installation & Operating Instructions for Genesys Analog Series Manifolds

Installation & Operating Instructions for Genesys Analog Series Manifolds Manufacturer of Medical Gas Piping Equipment Installation & Operating Instructions for Genesys Analog Series Manifolds Models NPC & TMC Model TML e6388brd 5/04/18 Tri-Tech, 35401 Avon Commerce Pkwy., Avon,

More information

NGP-250/500 Nitrogen Generator Quick Start Guide

NGP-250/500 Nitrogen Generator Quick Start Guide NGP-250/500 Nitrogen Generator Quick Start Guide Version: A July 2013 Potter Electric Signal Company, LLC 5757 Phantom Dr., Suite 125 P. O. Box 42037 Hazelwood, MO 63042 Phone: (314) 595-6900 Document

More information