Micro GC Fusion Pre-Installation Checlist

Size: px
Start display at page:

Download "Micro GC Fusion Pre-Installation Checlist"

Transcription

1 Site Preparation Proper site preparation prior to installing is very important. Contact INFICON ( with any questions regarding site preparation. Checkboxes ( ) identify hardware and actions that must be addressed when installing. List of Required Tools 1 1/8" wrench recommended, or adjustable large wrench (for attaching regulator to gas tank) Teflon tape for regulator connections The following tools are included in the optional Install Kit (PN G1). Open-ended wrench, 1/4 in. and 5/16 in. Open-ended wrench, 7/16 in. and 9/16 in. Open-ended wrench, 3/8 in. and 7/16 in. Hex driver, 2.5 mm ball-point long shank Tubing cutter Carrier Gas....Carrier gas (e.g., helium, argon) purity of %, or minimum % when the appropriate carrier gas trap is installed. Contact INFICON for carrier gas trap assistance. A pressurized [>16,547 kpa (>2,400 psi)] 50 liter tank of carrier gas will last approximately one year or more.....dual Stage Regulator for pressurized carrier gas tank with outlet pressure regulated to kpa (58-62 psi). A Swagelok 1/8 in. fitting is preferred on the regulator outlet to connect the carrier gas feed tube to. In North America, use the CGA 580 inlet connection (PN P1) for helium, nitrogen and argon carrier gas. Use the CGA 350 inlet connection (PN P1) for hydrogen carrier gas. For installations outside of North America, use the appropriate inlet connections as standard for the region. Exceeding 427 kpa (62 psi) carrier gas pressure may result in instrument damage. 1 of 5

2 Calibration Gas....Moisture/hydrocarbon trap (PN G2870A-01) required for the carrier gas when using Molsieve or Alumina columns. To protect and to enhance performance, carrier gas traps should be used to purify the carrier gas before introduction into. A hydrocarbon/moisture trap (PN G2870A-01) is usually sufficient.....copper or stainless steel tubing (pre-cleaned) 1/8 in. OD for carrier gas connections between the tank regulator output and carrier gas inlet. has up to two 1/8 in. Swagelok male carrier gas inlet connections. 3 m (10 ft.) of pre-cleaned GC grade copper tubing is included with. Additional copper tubing is available in the installation kit (PN G1), or can be ordered in 15 m (50 ft.) lengths (PN ). Do not clean copper or stainless steel tubing with solvent. Do not use Teflon or Tygon tubing for carrier gas.....assortment of male and female Swagelok fittings for regulator, tubing, etc., including nuts and front and back ferrules. Typical size is 1/8 in. but adapters for 1/4 in. to 1/8 in. may be needed if 1/4 in. tubing is used.....1/8 in. tee fitting and 1/8 in. ball valve, depending on connection requirements.....supply of calibration gas specific to the analysis from a commercial gas supplier. Calibration gas must be available at the time of installation. Calibration gas must closely resemble the composition of the actual process gas.....a regulator, if necessary, to reduce calibration gas pressure below 69 kpa (10 psi) at the inlets. The regulator must be clean, grease-free and non-venting. Checkout gas mixtures are only used for instrument installation and start up. Checkout gas mixtures supplied by INFICON are not calibration gases. 2 of 5

3 Sample Gas Pressure Range Sample gas pressure greater than 172 kpa (25 psi) will damage.....maximum pressure for direct sample injection into either the front or rear sample inlet is 172 kpa (25 psi). To prevent damage to, limit input sample pressure to 69 kpa (10 psi) or lower.....when the Integrated Sample Conditioner is installed, sample pressures up to 6,895 kpa (1,000 psi) can be introduced into either the front or rear sample inlet. Sample pressures higher than 6895 kpa (1,000 psi) must be reduced.....when a Genie Filter Assembly (PN G1) is installed, sample pressures up to 3447 kpa (500 psi) can be introduced into either the front or rear sample inlet. Sample pressures higher than 3447 kpa (500 psi) must be reduced. Installation Space and Venting Module instrument Dimension (L x W x H): 43.3 x 20 x 26.5 cm (17 x 7.9 x 10.4 in.) Weight: 6.8 kg (11 lb.)....install in a location with easy access to the instrument and associated computer.....for optimum performance and to dissipate heat, do not restrict the airflow around. WARNING Safely vent sample streams potentially toxic, noxious, or flammable gases outside the instrument and away from the operating area. Vent toxic gases to a fume hood, chemical trap, or reaction medium.....avoid venting into drafty areas, such as in front of a heating/cooling vent. 3 of 5

4 Power Requirement....Standard V (ac) electric outlet.....maximum power for 2-Module instrument: 260 W....Uninterruptible power supply with surge protection is recommended. Computer Hardware Requirement Minimum hardware configuration required to run software. Tablet....RAM Gb....Memory Gb....CPU GHz....Screen Size in.....resolution x 768 pixels....network Connectivity Wi-Fi wireless Computer....RAM Gb....Hard Drive Gb....CPU GHz....Screen Size in.....resolution x 768 pixels....network Connectivity Wi-Fi wireless or RJ-45 wired 4 of 5

5 Network Preparation....If will be connected to a local area network (LAN) using a static IP address, the hostname and static IP address must be assigned by the LAN administrator.....it is not necessary for a LAN administrator to set up the hostname and static IP address when Dynamic Host Configuration Protocol (DHCP) is used to assign the IP address. Web Browser Requirement.... supports the following tablet web browsers: Android 4+ stock browser Google Chrome or Firefox for Android Safari on ios 6 for ipad Symbian is not supported..... supports the following computer web browsers: Google Chrome for Mac OS and Windows XP or above Firefox for Mac OS and Windows XP or above Internet Explorer version 10 and above for Windows 7 or above 2014 INFICON reachus@inficon.com Due to our continuing program of product improvements, specifications are subject to change without notice. All trademarks are the property of their respective owners. 5 of 5

6890 GC Site Preparation

6890 GC Site Preparation 6890 GC Site Preparation (a16011) This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information

More information

490 Micro GC Site Preparation Checklist

490 Micro GC Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A) Purpose of Procedure Your site must meet this specification or set of requirements to assure a successful and timely installation of your 7820MSD Series Mass Selective Detector (MSD). This checklist is

More information

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A Purpose of Procedure Your site must meet this specification or set of requirements to assure a successful and timely installation of your 5975C Series Mass Selective Detector (MSD). This checklist is designed

More information

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,

More information

For additional information about our solutions, please visit our web site at US/Pages/HomePage.

For additional information about our solutions, please visit our web site at  US/Pages/HomePage. Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Agilent 7697A Headspace Sampler

Agilent 7697A Headspace Sampler Agilent 7697A Headspace Sampler Site Preparation Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Customer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist

Customer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies Agilent 7820A GC Site Preparation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2009 No part of this manual may be reproduced in any form or by any means (including electronic storage and

More information

Customer Responsibilities

Customer Responsibilities Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist Thank you for purchasing an Agilent system. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is

More information

7820A GC Site Preparation Checklist

7820A GC Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Agilent 7667A. Mini Thermal Desorber. Site Preparation

Agilent 7667A. Mini Thermal Desorber. Site Preparation Agilent 7667A Mini Thermal Desorber Site Preparation Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval

More information

490 Micro GC Site Preparation Checklist

490 Micro GC Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Manual Solvent Purification System. Innovative Technology, Inc.

Manual Solvent Purification System. Innovative Technology, Inc. Innovative Technology, Inc. 2 New Pasture Road, Newburyport, MA 01950-4054 USA Email: Info@gloveboxes.com TEL: +1 978 462 4415 FAX: +1 978 462 3338 Manual Solvent Purification System Copyright Innovative

More information

Agilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist

Agilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Installing a Probe. Typical Installation

Installing a Probe. Typical Installation for hazardous locations. Prior to any installation in a Classified Hazardous Location, verify installation methods by the Control Drawing referenced on the product s name tag 2. The Feed Through Assemblies

More information

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

Markes Thermal Desorption Systems Site Preparation Page 1 of 8 Thermal Desorption Systems Site Preparation Purpose of site preparation Your site must meet this specification to assure a successful and timely installation of your Agilent thermal desorption equipment.

More information

TA Instruments Installation Requirements for the Thermal Discovery Series

TA Instruments Installation Requirements for the Thermal Discovery Series Notice TA Instruments Installation Requirements for the Thermal Discovery Series Thank you for ordering a Discovery Series thermal analysis system from TA Instruments. To ensure that installation of your

More information

be present throughout these services; otherwise, they will miss important operational, maintenance and safety information.

be present throughout these services; otherwise, they will miss important operational, maintenance and safety information. Thank you for purchasing an Agilent Instrument. To get you started and to assure a successful and timely installation of your 9000 GC, please refer to this site prep checklist. Correct site preparation

More information

UltraLo 1800 Alpha Particle Counter

UltraLo 1800 Alpha Particle Counter XIA LLC UltraLo 1800 Alpha Particle Counter Site Requirements & Planning Version: 0.4 Friday, February 15, 2013 Table of Contents I. Site Preparation Tasks... 3 A. Instrument Overview... 3 B. Installation

More information

Customer Responsibilities. Important Customer Information. Agilent Infinity II Preparative HPLC System Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Agilent Infinity II Preparative HPLC System Site Preparation Checklist Agilent Site Preparation Infinity II Checklist Preparative HPLC System Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer

More information

28 Site Preparation. Supplying valve actuator air

28 Site Preparation. Supplying valve actuator air 28 Site Preparation Temperature and humidity ranges Ventilation requirements Venting oven exhaust Venting toxic or noxious gases Benchtop space requirements Electrical requirements Grounding Line voltage

More information

Please refer to the Hydrogen Safety Guide which is shipped with the Instrument. 5977B Series GC/MS Systems Site Preparation Checklist

Please refer to the Hydrogen Safety Guide which is shipped with the Instrument. 5977B Series GC/MS Systems Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Varian 240-MS GC/MS Ion Trap Mass Spectrometer Pre-installation Instructions

Varian 240-MS GC/MS Ion Trap Mass Spectrometer Pre-installation Instructions Varian 240-MS GC/MS Ion Trap Mass Spectrometer Pre-installation Instructions Checklist NOTE: Do not unpack the boxes. Place a check in the box after satisfying each requirement as described in the instructions.

More information

Discovery HP-TGA 75/750. Site Preparation Guide

Discovery HP-TGA 75/750. Site Preparation Guide Discovery HP-TGA 75/750 Site Preparation Guide Revision A Issued August 2018 Table of Contents Table of Contents... 2 Ideal Setup... 3 System Components... 4 Instrument Measurements... 5 Utility Requirements...

More information

Super-Clean Gas Filters from Restek

Super-Clean Gas Filters from Restek Super-Clean Gas Filters from Restek Chromatography Products www.restek.com 800-356-688 84-353-300 regulator instrument-grade tubing on/off valve GC carrier gas moisture trap hydrocarbon trap high-capacity

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere! The portable Gas Chromatograph is now more versatile than ever. DPS Instruments is pleased to present the newest Portable Companion 2 Gas Chromatograph with room for 2 Detectors. The Companion 2 GC was

More information

Customer Responsibilities

Customer Responsibilities Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Storage Systems and Accessories

Storage Systems and Accessories Storage Systems and Accessories High Pressure Air and Inert Gases 4500-6000 PSIG DOT and ASME Cylinders High Pressure Storage Systems If an application requires storage, either in the form of a single

More information

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS Using state-of-the-art measuring and microprocessor technologies, Advanced has assembled a series of systems which can measure mass

More information

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more! After years of development and testing, DPS Instruments is pleased to present the newest, most expandable and versatile Gas Chromatography Systems in history. The DPS 600 Series GC systems are the world

More information

Gas Management Supplies for GC Labs

Gas Management Supplies for GC Labs Restek GC Accessories Gas Management Supplies for GC Labs Gas Generators Pressure Regulators Tubing and Fittings Gas Purifiers www.restek.com Gas Management for Your Lab A clean gas stream is critical

More information

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

MATHESON s LaserPro Gases and Gas Delivery Equipment

MATHESON s LaserPro Gases and Gas Delivery Equipment MATHESON s LaserPro Gases and Gas Delivery Equipment MATHESON s LaserPro Support Gases and Services MATHESON delivers the LaserPro line of gases specifically designed to meet the stringent purity requirements

More information

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies Agilent 7820A GC Site Preparation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2010-2015 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Cryogenic Liquid Ultimate Liquid Cylinder Hose

Cryogenic Liquid Ultimate Liquid Cylinder Hose Cryogenic Liquid Ultimate Liquid Cylinder Hose Flexible Components brand liquid cylinder hose is considered the highest quality liquid cylinder hose on the market. It effectively transfers cryogenic liquids

More information

Features. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing)

Features. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing) 1 X-Flow Mass Flow Controller Flow Range from 0.8 ml/min to 20 l/min Target Markets Laboratory and Process Instrumentation Pharma / Bio-Pharma Equipment Air Quality Monitoring Systems Furnace and Coatings

More information

Pressure Dump Valve Service Kit for Series 3000 Units

Pressure Dump Valve Service Kit for Series 3000 Units Instruction Sheet Pressure Dump Valve Service Kit for Series 000 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00, 400, 500, and 700 applicator

More information

Preparing Your Laboratory for the Optima 7300 V and 8300 ICP-OES Spectrometers

Preparing Your Laboratory for the Optima 7300 V and 8300 ICP-OES Spectrometers LABORATORY PREPARATION Preparing Your Laboratory for the Optima 7300 V and 8300 ICP-OES Spectrometers TABLE OF CONTENTS Section Subject 1 Suitable Working Area 2 Exhaust Vent 3 Vent Positions 4 Handling

More information

TA Instruments Installation Requirements for Thermal Analysis Systems

TA Instruments Installation Requirements for Thermal Analysis Systems TA Instruments Installation Requirements for Thermal Analysis Systems Thank you for ordering a Thermal Analysis system from TA Instruments. To ensure that installation of that system goes as smoothly as

More information

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT APPLICATION NOTE Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT The INFICON 3000 Micro GC provides two fast and accurate solutions for the analysis of coal mine gas components.

More information

Discovery SDT 650. Site Preparation Guide. Revision G Issued December Page 1

Discovery SDT 650. Site Preparation Guide. Revision G Issued December Page 1 Discovery SDT 650 Site Preparation Guide Revision G Issued December 2018 Page 1 Table of Contents Table of Contents... 2 Ideal Setup... 3 System Components... 4 With Discovery Mass Spectrometer... 4 With

More information

How to order guide. Spectronus FTIR Trace Gas Analyser. Version: 1.0

How to order guide. Spectronus FTIR Trace Gas Analyser.   Version: 1.0 How to order guide Spectronus FTIR Trace Gas Analyser Version: 1.0 www.ecotech.com How to order guide Spectronus FTIR Tr ace Gas Analyser 1.0 This page is intentionally blank Page 2 Table of Contents Table

More information

WHAT ARE THE SPECIFICATIONS FOR THE ARGON GAS OPTION USED IN JOULE THOMSON REFRIGERATORS?

WHAT ARE THE SPECIFICATIONS FOR THE ARGON GAS OPTION USED IN JOULE THOMSON REFRIGERATORS? MMR Technologies, Inc. WHAT ARE THE SPECIFICATIONS FOR THE ARGON GAS OPTION USED IN JOULE THOMSON REFRIGERATORS? Joule Thomson refrigerators operate using high pressure gas to cool a sample stage with

More information

Thermo Scientific FOCUS GC. Gas Chromatograph Site Preparation and Installation Manual. PN , Revision May 2010

Thermo Scientific FOCUS GC. Gas Chromatograph Site Preparation and Installation Manual. PN , Revision May 2010 Thermo Scientific FOCUS GC Gas Chromatograph Site Preparation and Installation Manual PN 317 094 11, Revision May 2010 FOCUS GC Site Preparation and Installation Manual May 2010 Edition Part Number 317

More information

TM8 On-Line Gas Chromatography DGA Monitor Site Preparation Guide Model TMx Series

TM8 On-Line Gas Chromatography DGA Monitor Site Preparation Guide Model TMx Series TMx Site Preparation Guide 810-1647-03 Rev B TM8 On-Line Gas Chromatography DGA Monitor Site Preparation Guide Model TMx Series June, 2017 Document ID: 810-1647-03 Rev B www.serveron.com Legal Information

More information

Avio 500 ICP-OES. ICP-Optical Emission Spectroscopy PREPARING YOUR LAB. Preparation Checklist. Suitable Working Area

Avio 500 ICP-OES. ICP-Optical Emission Spectroscopy PREPARING YOUR LAB. Preparation Checklist. Suitable Working Area PREPARING YOUR LAB Avio 500 ICP-OES ICP-Optical Emission Spectroscopy Preparation Checklist Suitable Working Area Exhaust Vent Vent Positions Handling of Gas Cylinders and Other Suggested Safety Practices

More information

Thermo Scientific. TRACE GC Ultra. Gas Chromatograph Site Preparation and Installation Manual. PN , Revision May 2010

Thermo Scientific. TRACE GC Ultra. Gas Chromatograph Site Preparation and Installation Manual. PN , Revision May 2010 Thermo Scientific TRACE GC Ultra Gas Chromatograph Site Preparation and Installation Manual PN 31709190, Revision May 2010 TRACE GC Ultra Site Preparation and Installation Manual May 2010 Edition Part

More information

Pre-installation Checklist

Pre-installation Checklist Pre-installation Checklist Varian 1200/1200L Mass Spectrometer Use this checklist as a guide while preparing your site for the installation of the Varian 1200 or 1200L MS system. Make sure that your site

More information

Agilent 5975T LTM GC/MSD

Agilent 5975T LTM GC/MSD Agilent 5975T LTM GC/MSD Site Preparation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Customer Responsibilities

Customer Responsibilities Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Varian 325-MS LC/MS Quadrupole Mass Spectrometer Pre-installation Instructions

Varian 325-MS LC/MS Quadrupole Mass Spectrometer Pre-installation Instructions Varian 325-MS LC/MS Quadrupole Mass Spectrometer Pre-installation Instructions Checklist NOTE: Do not unpack the shipping cartons. Check off each checklist box after satisfying each requirement as described

More information

Review of the Hall B Gas System Hardware. George Jacobs

Review of the Hall B Gas System Hardware. George Jacobs of the Hardware George Jacobs DSG Staff 2 Hall B Gas Utilities for detectors Drift Chamber (DC) Low Threshold Cherenkov Counter (LTCC) Micromegas Vertex Tracker (MVT) Forward Tagger (FT) Ring Imaging Cherenkov

More information

Agilent Intuvo 9000 Gas Chromatograph

Agilent Intuvo 9000 Gas Chromatograph Agilent Intuvo 9000 Gas Chromatograph Installation and First Startup Agilent Technologies Notices Agilent Technologies, Inc. 2016 No part of this manual may be reproduced in any form or by any means (including

More information

Gas delivery systems. How to ensure uninterrupted high quality gas delivery from the source to the apparatus? Dr. Gastons Vereskuns Trakai,Lithuania

Gas delivery systems. How to ensure uninterrupted high quality gas delivery from the source to the apparatus? Dr. Gastons Vereskuns Trakai,Lithuania Gas delivery systems. How to ensure uninterrupted high quality gas delivery from the source to the apparatus? Dr. Gastons Vereskuns Trakai,Lithuania May 7, 2014 What it is gas delivery system? It is a

More information

21700 Nordhoff St. Chatsworth, CA Phone

21700 Nordhoff St. Chatsworth, CA Phone 21700 Nordhoff St. Chatsworth, CA. 91311 Phone 818.882.2331 DILUTION SYSTEM DILUTION SYSTEM ~ 1 ~ 09 21 07 TABLE OF CONTENTS DEFINITIONS.. 3 INTRODUCTION.. 4 INSTALLATION & SETUP 5 CONVERSION TO PRESSURE

More information

Air Operated Hydraulic Pumping Systems to 50,000 psi

Air Operated Hydraulic Pumping Systems to 50,000 psi High Pressure Equipment Air Operated Hydraulic Pumping Systems to 50,000 psi PS-10: 10,000 psi PS-20: 20,000 psi PS-30: 30,000 psi PS-40: 40,000 psi PS-50: 50,000 psi PS-90: 90,000 psi High Pressure air

More information

GasTrak Pro TM Delivery/Control Systems

GasTrak Pro TM Delivery/Control Systems GasTrak Pro TM Delivery/Control Systems Regulator, Purifier and Outlet Sections Introduction GasTrak Pro TM Systems are a unique approach to controlling and delivering high-purity gases distributed in

More information

Pressure Dump Valve Service Kit for Series 2300 Units

Pressure Dump Valve Service Kit for Series 2300 Units Instruction Sheet Pressure Dump Valve Service Kit for Series 00 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00 applicator tanks when the unit

More information

Nonrefillable Cylinders

Nonrefillable Cylinders Nonrefillable Cylinders A DIVISION OF NORCO, INC. Norlab, the Specialty Gas Division of Norco, Inc., has been producing and distributing calibration mixtures and pure specialty gases in refillable and

More information

780S Ultra High Purity Series

780S Ultra High Purity Series Smart Ultra High Purity Thermal Gas Mass Flow Meter Features Measures mass flow directly, no seperate temperature or pressure inputs required Field adjustment of critical flow meter settings via on-board

More information

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW AC 2010-2145: MEASUREMENT OF HYDROGEN IN HELIUM FLOW Randy Buchanan, University of Southern Mississippi Christopher Winstead, University of Southern Mississippi Anton Netchaev, University of Southern Mississippi

More information

TITAN Oxygen Regulators

TITAN Oxygen Regulators EMS OXYGEN REGULATORS TITAN Oxygen Regulators EMS TITAN SERIES OXYGEN EMS AND FIRE SERVICE APPLICATIONS CLICK-STYLE REGULATOR Whether its EMS, ambulance, fire rescue or tactical medical use, Titan Oxygen

More information

NexION 1000/2000 ICP-MS

NexION 1000/2000 ICP-MS PREPARING YOUR LAB NexION 1000/2000 ICP-MS ICP - Mass Spectrometry Preparation Checklist Environmental conditions Electrical requirements Space requirements Exhaust ventilation Coolant requirements Argon

More information

Preinstallation Checklist and Instructions for the TriStar These Preinstallation Checklist and Instructions were reviewed and approved by:

Preinstallation Checklist and Instructions for the TriStar These Preinstallation Checklist and Instructions were reviewed and approved by: Title: Preinstallation Checklist and Instructions for TriStar 3000 Date: 08/04 300/42870/01 Revision ( A ) MICROMERITICS Preinstallation Checklist and Instructions for the TriStar 3000 These Preinstallation

More information

ATIS OPERATION MANUAL

ATIS OPERATION MANUAL ATIS OPERATION MANUAL Adsorbent Tube Injector System T702019A 1 Introduction The Adsorbent Tube Injector System (ATIS) is a sample preparation device that provides a quick and efficient way to transfer

More information

6200E Installation Requirements

6200E Installation Requirements Requirements When purchasing a standalone 6200E unit with the calibrator, the analyzer will be supplied as three 19 rack mount units. Analyzer Converter Calibrator Included with the unit are two lengths

More information

MB-SP Series Solvent Purification System

MB-SP Series Solvent Purification System Operation Manual MB-SP Series Solvent Purification System Table of contents 1. System overview 2 2. Column activation.. 5 3. System purging.. 6 4. Glove box integration.... 7 5. Solvent reservoir installation....

More information

Installation, Operation, and Maintenance Manual

Installation, Operation, and Maintenance Manual Installation, Operation, and Maintenance Manual Welker Probe Instrument Regulator Model The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for

More information

LUMAZOTE T ECHNICAL BRO CHU RE

LUMAZOTE T ECHNICAL BRO CHU RE by Chemiluminescence Standards AST D4629 / ASTM D6069 / ASTM D 5176 / ASTM D 5762 / NF EN 12260 / NF M 07-058 Licence TOTAL LUMAZOTE T ECHNICAL BRO CHU RE ERALY & Associés Phone: +33 (0) 1 77 04 80 97

More information

Mass Flowmeters. Mass Flowmeters for Gases

Mass Flowmeters. Mass Flowmeters for Gases Mass Flowmeters Mass Flowmeters for Gases Fast, Accurate, Low Pressure Drop for critical measurement applications! Flow (Std l/min) Response to a step change in flowrate 300 200 100 4 Milliseconds 63%

More information

Discovery TGA 55, TGA 550, TGA Site Preparation Guide

Discovery TGA 55, TGA 550, TGA Site Preparation Guide Discovery TGA 55, TGA 550, TGA 5500 Site Preparation Guide Revision E Issued March 2018 Table of Contents Table of Contents... 2 Ideal Setup... 3 System Components... 4 With Discovery Mass Spectrometer...

More information

Specialty Gases & Equipment ISO 9001: 2008 CERTIFIED. A Division of Norco, Inc.

Specialty Gases & Equipment ISO 9001: 2008 CERTIFIED. A Division of Norco, Inc. Specialty Gases & Equipment ISO 9001: 2008 CERTIFIED A Division of Norco, Inc. Located in Boise Idaho, Norlab has been producing and distributing calibration gas mixtures and specialty gases in refillable

More information

Process Micro-chromatograph PGC 490 ATEX explosion-proof rapid gas analyzer General user manual

Process Micro-chromatograph PGC 490 ATEX explosion-proof rapid gas analyzer General user manual Process Micro-chromatograph PGC 490 ATEX explosion-proof rapid gas analyzer General user manual p. 1 / 26 Table of contents 1. INTRODUCTION 4 2. SAFETY INSTRUCTIONS 4 2.1 Important general information

More information

TA Instruments Installation Requirements for Thermal Analysis Systems

TA Instruments Installation Requirements for Thermal Analysis Systems TA Instruments Installation Requirements for Thermal Analysis Systems Notice The material contained in this manual, and in the online help for the software used to support TA Instruments products, is believed

More information

VOLUMATIC WATER REGULATOR

VOLUMATIC WATER REGULATOR VOLUMATIC WATER REGULATOR For Cage Systems - PDS and Manual Adjustment Part Number 00-, 00-, 00- Introduction The Chore-Time VOLUMATIC Water Regulator is designed to regulate the water pressure for the

More information

Generating Calibration Gas Standards

Generating Calibration Gas Standards Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and

More information

Baffles and Traps Cryotrap

Baffles and Traps Cryotrap Baffles and Traps 362-6 Cryotrap Varian Low-Profile Water-Cooled Baffles combine 100% optical density with high conductance and unusually low overall height. They are especially useful in applications

More information

GAS DETECTION: CALIBRATION GAS

GAS DETECTION: CALIBRATION GAS ENVIRONMENTAL SURVEY/PURGING MONITORS GAS DETECTION: CALIBRATION GAS Gas Mixtures Non-Reactive Single & Two-Gas Mixtures 98 Reactive Single & Two-Gas Mixtures 98 Non-Reactive Multi Mixtures 99 Reactive

More information

Energy Management solutions for compressed air and industrial gases. vpinstruments.com

Energy Management solutions for compressed air and industrial gases. vpinstruments.com Energy Management solutions for compressed air and industrial gases vpinstruments.com 6 Introduction VPINSTRUMENTS Measure, Discover and Save! Compressed air is a very expensive energy source. In fact,

More information

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the

More information

Ammonia Treatment Instructions Page 1 of 5

Ammonia Treatment Instructions Page 1 of 5 Page 1 of 5 Original version by Douglas Call Updated January 21, 2009 by David Jones Department: Chemical Engineering Location: Fenske 102 (access thru room 103) Contact: BP Matthew Holbrook 161 Fenske

More information

Agilent 490 Micro GC Natural Gas Analyzer

Agilent 490 Micro GC Natural Gas Analyzer Agilent 490 Micro GC Natural Gas Analyzer User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic

More information

NGN Series nitrogen generator

NGN Series nitrogen generator NGN Series 8-64 nitrogen generator Applications how it works RK series 5-40 F Blanketing of Chemicals and Pharmaceuticals Gas Assisted Injection Moulding (GAIM) Laser Cutting Food Heat Treament of Ferrous

More information

Agilent GC, GC/MS, and ALS

Agilent GC, GC/MS, and ALS Agilent GC, GC/MS, and ALS Site Preparation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2017 No part of this manual may be reproduced in any form or by any means (including electronic

More information

BGA244 Binary Gas Analyzer

BGA244 Binary Gas Analyzer Quick Start Guide Revision 1.0 Certification Warranty Service certifies that this product met its published specification at the time of shipment. This product is warranted against defects in materials

More information

Gas Absorption Draft Standard Operating Procedure

Gas Absorption Draft Standard Operating Procedure Gas Absorption Draft Standard Operating Procedure R.C. 9-14 Scope Page 1 Equipment Overview Page 1 Safety Page 2 Operating Procedure Page 4 Equipment Description Page 16 Gas Absorption Experiment Draft

More information

IRwin Methane Leak Detector. Innovation makes the difference

IRwin Methane Leak Detector. Innovation makes the difference IRwin Methane Leak Detector Innovation makes the difference Innovative methane detector system for easy survey of gas pipelines IRwin is a portable methane leak detector for fast and easy survey of gas

More information

Technical Information Conducal CLY421

Technical Information Conducal CLY421 TI00496C/07/EN/01.12 71142150 Products Solutions Services Technical Information Conducal CLY421 Conductivity calibration set for ultrapure water applications Factory calibrated measurement box Certified

More information

Agilent G3180B Two-Way Splitter Kit

Agilent G3180B Two-Way Splitter Kit Agilent G3180B Two-Way Splitter Kit With Makeup Gas Installation and Operation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or

More information

Instruction Manual For Model Equilibrium Flash Separator P/N: D01 Revision E February 2016

Instruction Manual For Model Equilibrium Flash Separator P/N: D01 Revision E February 2016 Instruction Manual For Model 2353-805 Equilibrium Flash Separator P/N: 2353-1D01 Revision E February 2016 S/N: 2001 N. Indianwood Ave. Broken Arrow, Oklahoma 74012 U.S.A. Telephone: 918-250-7200 Fax: 918-459-0165

More information

SPECIFICATIONS. Approximate Weight: 10 oz. Surface Finish: Ra micro inch or less MATERIALS OF CONSTRUCTION

SPECIFICATIONS. Approximate Weight: 10 oz. Surface Finish: Ra micro inch or less MATERIALS OF CONSTRUCTION DIFFUSION-RESISTANT, DIAPHRAGM SEAL, EXCESS FLOW SHUT-OFF VALVES (FS SERIES) FS Series Excess Flow Shut-Off Valves are designed to automatically shut-off the delivery of gas in a line if the flow exceeds

More information

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1 Dry 131h 1. DESCRIPTION The Viking Model G-1 Maintenance Air Compressor is an electric motor-driven, aircooled, single-stage, oil-less compressor. The unit is equipped with a check valve and provides a

More information

Running HPLC Methods on an ACQUITY UPLC System

Running HPLC Methods on an ACQUITY UPLC System Running HPLC Methods on an ACQUITY UPLC System Note: This document is an addendum to the ACQUITY UPLC System Operator s Guide. Contents Topic Page System considerations 1-1 Choosing fittings 1-3 System

More information

MEDICAL EQUIPMENT CATALOG

MEDICAL EQUIPMENT CATALOG THE MOST TRUSTED MEDICAL EQUIPMENT IN NORTH AMERICA MEDICAL EQUIPMENT CATALOG Total Components Including Bulk Oxygen Supply, Back up and Delivery Equipment Designs for High Pressure, Liquid or Bulk Applications

More information

SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER

SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER SUPPLIMENT TO eom IMPORTANT READ THIS MANUAL BEFORE PRODUCT INSTALLATION, OPERATION, INSPECTION & MAINTENANCE Tougher and more rigid guidelines are being established

More information

NHR Powder Feed Hopper

NHR Powder Feed Hopper Instruction Sheet P/N 0804A0 NHR-8-50 Powder Feed Hopper Description This instruction sheet provides specifications, installation and operation instructions, and parts lists for the NHR-8-50 powder feed

More information