RM-80 respiration monitor

Size: px
Start display at page:

Download "RM-80 respiration monitor"

Transcription

1 RM-80 respiration monitor User Manual September 18, M 950 North Hague Avenue Columbus, Ohio USA Sales: Service: Phone: (614) Fax: (614) Toll Free:

2

3 Table of Contents Chapter 1 Overview Overview Specification Inputs Front panel controls and indicators... 5 Chapter 2 Tubing connections Tubing connections... 6 I Cable connections Chapter 3 Software Setup Software installation Chapter 4 Operation Chamber setup Experiment setup Running Experiment

4 Chapter 1 Overview 1.1 Overview The Columbus Instruments RM-80 Respiration Frequency Monitor measures breaths per minute (BPM) in unrestrained mice and rats. Each subject is placed into a small chamber and the system measures changes in chamber pressure that result from the subject s respiration. The Respiration rate is then captured by the CI Multi-device software on a user based interval schedule The essentials of an RM-80 System consist of a pressure sensor, animal chamber, reference chamber, and a fresh air delivery system. The pressure sensor measures differential pressure between the animal chamber and the reference chamber with each breath. A microprocessor inside the RM-80 analyzes the resulting waveform of these small pressure changes and reports the frequency as BPM. The fresh air delivery system refreshes the air in both chambers for proper animal ventilation but does so through a special buffering chamber to greatly reduce air turbulence and background noise. The RM-80 has a buffered analog signal direct from the pressure sensor. In typical setups the BPM output is connected to a PC via USB. Software is included with the counter interface and BPM are recorded on an interval basis. Collecting data in this fashion allows the researcher to ignore the initial sniffing behavior and focus on resting BPM. The analog signal is provided for those adventurous enough to perform their own analysis on the pressure waveform (requires connection to a 3rd party A/D data acquisition system). 2

5 1.2 Specification RM-80 Basic Specifications Number of Channels: Range: Accuracy: Sensitivity: 1 to 4 depending upon Configuration BPM(Breaths Per Minute) 2% BPM Standard Rm-80 system will detect mouse Respiration in a 3 liter sealed cage volume Power Requirements Maximum of 50 watts ( VAC, HZ) Size and Weight Control unit size: Control unit weight: 14" x 11.5" x 4" (35cm x 29cm x 10cm) 6.2 pounds (2.8 kg) 3

6 1.3 Inputs The only inputs to the RM-80 are the Transducer Inputs and Reference connections. Each channel will have its own Transducer Input as well as its own Reference Connection. These connect to the animals chambers with ¼ tubing. Be very careful when connecting the hoses. The Transducers that these tubes connect to will be permanently damaged if they are pressurized to a pressure more than 0.2 PSIG. 1 Reference input This is where the tube from the reference chamber connects for each corresponding channel. 2 Cage input This is where the tube from the reference chamber connects for each corresponding channel. 3 Power input The supplied 12V dc wall adapter connects here to supply the RM- 80 power 4 CI bus This is where the standard RJ-45 network cable connects the RM- 80 to the Columbus Instruments hub box. The hub box is an interface to the computer. 4

7 1.4 Front panel controls and indicators A list of the front panel controls and table of their functions is shown below 1 Audio select switch This selects which channel s breath detection is output the the internal speaker. Each breath detected by RM-80 for the currently selected channel will be indicated by a beep from the speaker. This is useful for adjusting the breath signal sensitivity control for the currently selected channel 2 Breath signal sensitivity This control adjust the gain (sensitivity) of the pressure sensor amplifier for the corresponding channel. Adjustment of this control is described in the operation section. 3 Respiration indicator This led blinks with each detected breath 4 Ref overpressure indicator 5 Ref overpressure indicator This led lights if overpressure is indicated in the reference chamber. This is caused by the overflow tube or port being blocked. If this occurs the blockage must be resolved This led lights if overpressure is indicated in the reference chamber. This is caused by the overflow tube or port being blocked. If this occurs the blockage must be resolved 5

8 Chapter 2 Tubing connections 2.1 Tubing connections Quick-connect fittings are used to make an air-tight connection. They are commonly found near components of the system that may require frequent maintenance (such as filters) or around components which modify the operation of the system (such as flow restrictors). To make a connection, ensure there are no frayed edges on the end of the tube. If required, trim the end of the plastic tube with a sharp implement (tube cutter preferred) as square as possible. Fully insert the tube into the plastic fitting, through the resistance, until it stops (about 1.5cm for most fittings). The tube should not pull free from the fitting (the dark gray collar will travel and rotate with the tube). To remove the tube from the connector, press and hold the dark gray collar against the body of the fitting. Now pull the tube free from the fitting. 6

9 Connect the fresh air supply pump to the damping chamber as shown below. 7

10 I If the system is used for rats, connect 2 outputs from the pump together with a T fitting to supply the animal(s) with more air as shown below, 8

11 Connect the tube(s) from the damping chamber to the animal chamber and reference chamber as shown below. 9

12 10 Connect short (18 ) pieces to the overflow fitting on the animal chamber and reference chamber as shown below. These provide some restriction to the air coming out of the chamber to give better signal to the pressure sensor.

13 Connect the output from the animal chamber and reference chamber to the RM-80 input as shown below. 11

14 2.2 Cable connections Connect the RJ-45 cable between the CI bus hub box and the RM-80 as shown below. Do not connect the USB cable between the CI bus hub box to the computer until after the software is installed 12

15 Chapter 3 Software Setup 3.1 Software installation The host computer is expected to be a IBM-PC compatible running a Windows operating system (98, ME, 2000, NT, XP, Vista, 7). It must have a minimum of: 128MB of memory, 1GB of free hard drive space, CD-ROM drive (for installation), and 1024 x 768 VGA color display Insert the supplied CD ROM into the CD ROM drive Navigate to the CD-ROM. Double click on setup.exe The following screens will appear. Close all other programs and Click OK to continue 13

16 It is strongly recommended to install the program in the default location. Click on the computer icon box to continue. Next the USB driver will be installed: Click the install button. This completes the software installation. The USB cable from the CI hub box to the computer can now be connected 14

17 Chapter 4 Operation 4.1 Chamber setup Set the control on top of the fresh air pump (if equipped) to it s maximum value (fully clockwise) Place the animal(s) into the RM-80 chambers and close the lid. Bedding can be placed in the chamber to help acclimate the animal. For each channel used make the breathing sensitivity adjustment as follows: Turn the Audio Select control switch to the chamber (cage) to be adjusted Turn the breathing sensitivity control for the channel being adjusted to it s lowest setting (fully counter clockwise) With the animal calm and not moving, observe it s respiration. Slowly increase the breathing sensitivity control until the beeping from the RM-80 corresponds to the breathing of the animal. Lock the control. Typically a value of 1 on the control dial will correspond to a rat and 2 for a mouse. Repeat for each channel used. 4.2 Experiment setup From the main menu choose Experiment Setup 15

18 The experiment setup screen will be displayed Click the box on the right side of the experiment filename box to open the save file dialog. Select a location where the experiment data is to be saved and file name then click open. Select Session Interval. This determines how frequently the respiration rate of each chamber is saved to the data file. If desired, Staggered Start can be enabled. The purpose of staggered start is to allow the start of each chamber independently. This is useful for example when dosing animals with a compound then placing them in a chamber. If Staggered Start is not enabled, all chambers will start collecting data immediately after starting the experiment. 16

19 Each chamber can be enabled or disabled by clicking in the enable column for the channel(s) that are to be disabled Double click in the Cage label column for each enabled cage to enter identification information about the respective subject 17

20 4.3 Running Experiment From the experiment setup screen click run on the bottom of the screen. If staggered start is enabled, the following screen will be displayed; otherwise the system will start collecting data at the specified interval. When it is desired to start a chamber, click in the box in the status column that corresponds to the channel to be started. It will be queued, then started. 18

21 The data format consists of the interval number (sample number) cage number, time and date that the reading was taken, and the respiration rate in breaths per minute. Note the data file format is CSV which stands for comma separated value. This is a standard format and will import directly to excel or other data analysis programs 19

22 20 The graphing tab is useful to look at the respiration rate over time to identify resting period which give the best respiration rate data.

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 QUICK START GUIDE P/N 6006154, REVISION C MAY 2013 Model 8120 Mask Integrity Tester is patented under U.S. Patent No. 8,312,761. Additional patents are pending.

More information

Oxygen Meter User Manual

Oxygen Meter User Manual Oxygen Meter User Manual Monday, July 23, 2007 1. Outline...2 2. Program...3 2.1. Environment for program execution...3 2.2. Installation...3 2.3. Un installation...3 2.4. USB driver installation...3 2.5.

More information

Technology. In the My Files [My Files] submenu you can store all the programs that you have made on the NXT or downloaded from your computer.

Technology. In the My Files [My Files] submenu you can store all the programs that you have made on the NXT or downloaded from your computer. NXT Main Menu My Files Files are automatically placed into the appropriate folders. When you download a program using a Sound file to the NXT, the program will be placed under Software files while the

More information

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300 Technical Note GA-300 Overview The GA-300 CO2 and O2 Gas Analyzer is easy to use, robust, and adaptable to human, animal, and plant applications. The GA-300 has two analog outputs to allow recording and

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Background There are two main sources of energy available for animal metabolism: carbohydrates (CHO) and fats. These molecules are broken

More information

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Background There are two main sources of energy available for animal metabolism: carbohydrates (CHO) and fats. These molecules are broken

More information

Air Ball! Evaluation copy

Air Ball! Evaluation copy Air Ball! Computer 24 Do you ever wonder how the National Basketball Association (NBA) decides how much air should be in the basketballs used during a game? The NBA measures the pressure inside the ball

More information

Race Screen: Figure 2: Race Screen. Figure 3: Race Screen with Top Bulb Lock

Race Screen: Figure 2: Race Screen. Figure 3: Race Screen with Top Bulb Lock Eliminator Competition Stand Alone Mode - Instruction Manual Main Menu: After startup, the Eliminator Competition will enter the Main Menu. Press the right/left arrow buttons to move through the menu.

More information

Oxymax for Windows. User Manual September 17, M

Oxymax for Windows. User Manual September 17, M Oxymax for Windows User Manual September 17, 2014 0233-122M 950 North Hague Avenue Columbus, Ohio 43204-2121 USA Sales: sales@colinst.com Service: service@colinst.com Phone: (614) 276-0861 Fax: (614) 276-0529

More information

Section 1: Overview... 6 Features... 6 Supported Software Part 1: Getting Started 7

Section 1: Overview... 6 Features... 6 Supported Software Part 1: Getting Started 7 Table of Contents About this Manual... 3 Intended Audience... 3 How this Manual is Organized... 4 Conventions Used in this Manual... 5 Related Information... 5 Where To Get Help... 5 Section 1: Overview...

More information

CONSUMER MODEL INSTALLATION GUIDE

CONSUMER MODEL INSTALLATION GUIDE CONSUMER MODEL INSTALLATION GUIDE System requirements Windows System Requirements To use your TOMI and its software, your system should have: A Microsoft Windows compatible PC with a Pentium IV processor

More information

Instruction Manual. BZ7002 Calibration Software BE

Instruction Manual. BZ7002 Calibration Software BE Instruction Manual BZ7002 Calibration Software BE6034-12 Index _ Index Index... 2 Chapter 1 BZ7002 Calibration Software... 4 1. Introduction... 5 Chapter 2 Installation of the BZ7002... 6 2. Installation

More information

MULTITUBE BUBBLE TUBE BY ROMPA (U)

MULTITUBE BUBBLE TUBE BY ROMPA (U) Page 1 of 16 MULTITUBE BUBBLE TUBE BY ROMPA 19781 (U) CONTENTS 1 x Base Unit (chassis) 1 x Base Cover (white plastic cover) 1 x Multitube Bubble Tube Column 1 x Bubble Tube Cap 1 x Power Supply (black

More information

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS WARNING THE WARRANTIES MADE BY MSA WITH RESPECT TO THE PRODUCT ARE VOIDED IF THE PRODUCT IS NOT USED AND MAINTAINED IN ACCORDANCE

More information

TR Electronic Pressure Regulator. User s Manual

TR Electronic Pressure Regulator. User s Manual TR Electronic Pressure Regulator Page 2 of 13 Table of Contents Warnings, Cautions & Notices... 3 Factory Default Setting... 4 Quick Start Procedure... 5 Configuration Tab... 8 Setup Tab... 9 Internal

More information

FOOD INTAKE MONITOR. Hardware and Software User s Manual DOC-047 Rev 3.0. Copyright 2009 All Rights Reserved

FOOD INTAKE MONITOR. Hardware and Software User s Manual DOC-047 Rev 3.0. Copyright 2009 All Rights Reserved Hardware and Software User s Manual DOC-047 Rev 3.0 Copyright 2009 All Rights Reserved MED Associates, Inc. P.O. Box 319 St. Albans, Vermont 05478 www.med-associates.com - ii - TABLE OF CONTENTS Chapter

More information

Diver-Office. Getting Started Guide. 2007, Schlumberger Water Services

Diver-Office. Getting Started Guide. 2007, Schlumberger Water Services Diver-Office Getting Started Guide 2007, Schlumberger Water Services Copyright Information 2007 Schlumberger Water Services. All rights reserved. No portion of the contents of this publication may be reproduced

More information

AX5000 Operational Manual

AX5000 Operational Manual MIYACHI AMERICA CORPORATION The World Leader in Hermetic Sealing Systems AX5000 Operational Manual 0 Document #107-00092-001 Dec, 2013 AX5000 Operational Manual Miyachi America Corporation 1820 S. Myrtle

More information

ELIMINATOR COMPETITION DRAG RACE Program Manual Firm Ver 4.11

ELIMINATOR COMPETITION DRAG RACE Program Manual Firm Ver 4.11 ELIMINATOR COMPETITION DRAG RACE Program Manual Firm Ver 4.11 The Portatree Eliminator Super 2000 Competition Track Timer can be used with an IBM Compatible Personal Computer connected through Com Port

More information

SDM-2012 Docking Station Standalone Configuration Operator s Manual

SDM-2012 Docking Station Standalone Configuration Operator s Manual SDM-2012 Docking Station Standalone Configuration Operator s Manual Part Number: 71-0254RK Revision: P5 Released: 10/5/12 www.rkiinstruments.com Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

QRAE II Quick Reference Covers Diffusion & Pump Models with Firmware Version 3.12 and higher

QRAE II Quick Reference Covers Diffusion & Pump Models with Firmware Version 3.12 and higher QRAE II Quick Reference Covers Diffusion & Pump Models with Firmware Version 3.12 and higher WARNINGS Use only RAE Systems rechargeable battery pack part number 020-3402-000, or alkaline battery pack part

More information

Quadruple mass spectrometers (transducer type)

Quadruple mass spectrometers (transducer type) Quadruple mass spectrometers that meet the need for sensitive analysis and low outgassing which were specifically designed to be controlled by PCs Quadruple mass spectrometers (transducer type) M-070QA-TDF,

More information

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc. The HumiSys RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The new HumiSys with single or dual RH probes capabilities is

More information

Blackwave Dive Table Creator User Guide

Blackwave Dive Table Creator User Guide Blackwave Dive Table Creator User Guide Copyright 2002-2009 Blackwave. All rights reserved. These materials (including without limitation all articles, text, images, logos, compilation, and design) are

More information

KEM Scientific, Inc. Instruments for Science from Scientists

KEM Scientific, Inc. Instruments for Science from Scientists KEM Scientific, Inc. Instruments for Science from Scientists J-KEM Scientific, Inc. 6970 Olive Blvd. St. Louis, MO 63130 (314) 863-5536 Fax (314) 863-6070 E-Mail: jkem911@jkem.com Precision Vacuum Controller,

More information

Mapping a Magnetic Field. Evaluation copy. Figure 1: Detecting the magnetic field around a bar magnet

Mapping a Magnetic Field. Evaluation copy. Figure 1: Detecting the magnetic field around a bar magnet Mapping a Magnetic Field Experiment 16 The region around a magnet where magnetic forces can be detected is called a magnetic field. All magnets, no matter what their shape, have two poles labeled north

More information

Quick Start Guide. A. Hardware installation B. Software installation C. Start the software for the first time D. Do your first measurement

Quick Start Guide. A. Hardware installation B. Software installation C. Start the software for the first time D. Do your first measurement Quick Start Guide This Quick Start Guide describes the hardware and software installation process and the Measurement feature in a simple way. Please follow the sequence of the steps to avoid problems

More information

TOMI PRO MODEL PERSONAL MODEL

TOMI PRO MODEL PERSONAL MODEL TOMI PRO MODEL PERSONAL MODEL System requirements Windows System Requirements To use your TOMI and its software, your system should have: A Microsoft Windows compatible PC with a Pentium IV processor (1.5

More information

User Manual MFCS TM -EZ

User Manual MFCS TM -EZ User Manual MFCS TM -EZ Version 6A_December 2016 1. INTRODUCTION 5 2. GENERAL INFORMATION 6 2.1 General functioning 6 2.2 Different MFCS TM 6 3. PACKAGE CONTENT 8 4. MFCS TM -EZ DESCRIPTION 9 5. CONNECTIONS

More information

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual Air Bubbler Depth Gauge DG2200 Installation and Reference Manual Rev. 3.2.2 06/2011 Installation and Reference Manual Page 2 Contents Chapter 1: Installation and Overview... 3 1.1 DG2200 Location Diagram...

More information

SENSUS PRO MANAGER (for SENSUS or SENSUS PRO devices) User s Guide -- Windows. Version 2.0 Published October 17, ReefNet Inc.

SENSUS PRO MANAGER (for SENSUS or SENSUS PRO devices) User s Guide -- Windows. Version 2.0 Published October 17, ReefNet Inc. SENSUS PRO MANAGER (for SENSUS or SENSUS PRO devices) User s Guide -- Windows Version 2.0 Published October 17, 2002 2002 ReefNet Inc. 1.0 Introduction The SENSUS PRO data recorder captures the time, depth,

More information

AKTA pure 25 New Owner s Intro

AKTA pure 25 New Owner s Intro AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps

More information

HyperSecureLink V6.0x User Guide

HyperSecureLink V6.0x User Guide HyperSecureLink V6.0x User Guide Note: This software works with the LS-30 Version (06.0x or later) 1, Hardware Installation: 1-1, Connection Diagram for USB or RS-232 Computer Interface To LS-30 CM1 To

More information

WELCOME TO THE REVOLUTION

WELCOME TO THE REVOLUTION USER GUIDE WELCOME TO THE REVOLUTION THANK YOU FOR CHOOSING THE GCQUAD We listened to what you wanted - and created the most accurate, versatile and game-enhancing ball and club analysis solution available

More information

XC2 Client/Server Installation & Configuration

XC2 Client/Server Installation & Configuration XC2 Client/Server Installation & Configuration File downloads Server Installation Backup Configuration Services Client Installation Backup Recovery Troubleshooting Aug 12 2014 XC2 Software, LLC Page 1

More information

SPLIT HOPKINSON PRESSURE BAR OPERATIONS MANUAL 5/9/2014

SPLIT HOPKINSON PRESSURE BAR OPERATIONS MANUAL 5/9/2014 B Calibration of Split Hopkinson Pressure Bar Setup: Calibration is performed by comparing the wave speed of the initial pulse versus the reflected pulse and calculating the Energy Ratio. This ratio gives

More information

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02 ! Read this first Zetasizer nano series Self installation and Quick start guide I N S T R U M E N T S MRK825-02 Zetasizer Nano series Self installation and Quick start guide MAN0383 Issue 1.1 July 2007

More information

UNIVERSAL MASK TEST SYSTEM

UNIVERSAL MASK TEST SYSTEM f UNIVERSAL MASK TEST SYSTEM Quotation #110811 Automated Motion Inc. 225 NW Victoria Drive Lee s Summit, Missouri 64086-4709 www.automatedmotion.com SYSTEM OVERVIEW AMI s Universal Mask Test System is

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

Air Ball! LabQuest Vernier Gas Pressure Sensor Vernier Motion Detector basketball stopper with needle, stopper stem and tubing attached meter stick

Air Ball! LabQuest Vernier Gas Pressure Sensor Vernier Motion Detector basketball stopper with needle, stopper stem and tubing attached meter stick Air Ball! LabQuest 24 Do you ever wonder how the National Basketball Association (NBA) decides how much air should be in the basketballs used during a game? The NBA measures the pressure inside the ball

More information

Introduction WELCOME TO LEGO MINDSTORMS EDUCATION

Introduction WELCOME TO LEGO MINDSTORMS EDUCATION NXT User Guide Introduction WELCOME TO LEGO MINDSTORMS EDUCATION LEGO MINDSTORMS Education is the next generation in educational robotics, enabling students to discover Science, Technology, Engineering

More information

Operating Manual. SUPREMA Calibration. Software for Fire and Gas Warning Units. Order No.: /01. MSAsafety.com

Operating Manual. SUPREMA Calibration. Software for Fire and Gas Warning Units. Order No.: /01. MSAsafety.com Operating Manual Software for Fire and Gas Warning Units Order No.: 10154656/01 MSAsafety.com MSA Europe GmbH Schlüsselstrasse 12 8645 Rapperswil-Jona Switzerland info.ch@msasafety.com www.msasafety.com

More information

Boyle s Law: Pressure-Volume Relationship in Gases

Boyle s Law: Pressure-Volume Relationship in Gases Boyle s Law: Pressure-Volume Relationship in Gases The primary objective of this experiment is to determine the relationship between the pressure and volume of a confined gas. The gas we will use is air,

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information

HumiSys HF High Flow RH Generator

HumiSys HF High Flow RH Generator HumiSys HF High Flow RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The HumiSys HF is a high flow version of the previously

More information

SHIMADZU LC-10/20 PUMP

SHIMADZU LC-10/20 PUMP SHIMADZU LC-10/20 PUMP Clarity Control Module ENG Code/Rev.: M091/70C Date: 24.10.2017 Phone: +420 251 013 400 DataApex Ltd. Fax: +420 251 013 401 Petrzilkova 2583/13 clarity@dataapex.com 158 00 Prague

More information

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 The front door of the main body of the instrument may be open when you arrive. Take the sample chamber and slide it into position with the

More information

Overview. Front Panel: Keypad and Display

Overview. Front Panel: Keypad and Display Overview The GA-200B is an analyzer that integrates a gas sampling system with sensors to measure and display the concentrations of oxygen and carbon dioxide in a sample as the percentage of a gas in the

More information

MCQ Gas Blender 100 Series 3 Channels Gas Mixer

MCQ Gas Blender 100 Series 3 Channels Gas Mixer Channels Gas Mixer Revision 1.8 October 2016 High Performance Digital Gas flow Gas Dilutor Blender & Gas for Mixing flow Mixing Gases System and Calibration Routines Technical Data Sheet Lab in a box From

More information

Revision: 23 March 2008 (nk)

Revision: 23 March 2008 (nk) CCAP Air Monitoring Protocol #01 Secondhand Smoke Drift Between Apartment Units Version 1.0 California Clean Air Project March 2008 CCAP AIR MONITORING PROTOCOL #01 SECONDHAND SMOKE DRIFT BETWEEN APARTMENTS

More information

FIG: 27.1 Tool String

FIG: 27.1 Tool String Bring up Radioactive Tracer service. Click Acquisition Box - Edit - Tool String Edit the tool string as necessary to reflect the tool string being run. This is important to insure proper offsets, filters,

More information

ETT Olfactometer: Getting started guide 2013a. ETT Olfactometer quick start

ETT Olfactometer: Getting started guide 2013a. ETT Olfactometer quick start ETT Olfactometer quick start Table of Contents Safety Information... 3 Contact to ETT... 3 Un-boxing / Installation... 3 Resources overview... 7 Working with the ETT Olfactometer... 8 Creating the first

More information

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere! The portable Gas Chromatograph is now more versatile than ever. DPS Instruments is pleased to present the newest Portable Companion 2 Gas Chromatograph with room for 2 Detectors. The Companion 2 GC was

More information

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI NanoSight NS300 Operation instructions Injection/flushing brief overview: 1. Do not exceed flow of 1 ml per 20 seconds. 2. Inject two 1 ml syringes with nano-pure or DI water. 3. If the water does not

More information

Microsoft Windows Software Manual for FITstep Stream Version 4

Microsoft Windows Software Manual for FITstep Stream Version 4 Thank you for purchasing this product from Gopher. If you are not satisfied with any Gopher purchase for any reason at any time, contact us and we will replace the product, credit your account, or refund

More information

AUTOMATIC HOSE TEST UNIT, TYPE SPU

AUTOMATIC HOSE TEST UNIT, TYPE SPU VALVES AND FITTINGS UP TO 14,000 BAR TEST AND CONTROL EQUIPMENT H IGH PRESSURE TECHNOLOGY AUTOMATIC HOSE TEST UNIT, TYPE SPU Pressure range from 1 up to 10,000 bar User-friendly touch panel operation HIGH-PRESSURE

More information

BLOOD PRESSURE SENSOR 0377i

BLOOD PRESSURE SENSOR 0377i BLOOD PRESSURE SENSOR 0377i USER S GUIDE CENTRE FOR MICROCOMPUTER APPLICATIONS http://www.cma-science.nl Short description The Blood Pressure sensor 0377i is used to measure arterial blood pressure in

More information

Hallowell EMC Anesthesia WorkStation Operating Manual

Hallowell EMC Anesthesia WorkStation Operating Manual Hallowell EMC Anesthesia WorkStation Operating Manual PN 000a2770 Anesthesia WorkStation A compact (9" W x 9" D x 15" H), convenient system for use on animals under 7 kg (15 lb.) Hallowell Engineering

More information

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1 HPICAL-15000 Operation & Data Logging Procedures Click spacebar to advance through slides 1 WARNING Always wear proper safety equipment when using high pressure equipment. Do not exceed 125 psi air pressure.

More information

LEGO Engineering Conferences ROBOLAB and MINDSTORMS Education Version 4.5 March 2008

LEGO Engineering Conferences ROBOLAB and MINDSTORMS Education Version 4.5 March 2008 LEGO Engineering Conferences ROBOLAB and MINDSTORMS Education Version 4.5 March 2008 NXT-G Program Book II: Intermediate Robotics Activities for use with the NXT 2008 Tufts Center for Engineering Education

More information

This addendum describes the back pressure control implementation if FuelCell Version 4.1a and later.

This addendum describes the back pressure control implementation if FuelCell Version 4.1a and later. FuelCell Addendum - Back Pressure Controls D. Johnson, Scribner Associates, Inc. 7/23/2015, Ver. 6 Introduction This addendum describes the back pressure control implementation if FuelCell Version 4.1a

More information

Users Guide for the CableSoft Software Package

Users Guide for the CableSoft Software Package Issue 1.4: January, 2007 REF: CABLESOFT SOFTWARE MANUAL.DOC Users Guide for the CableSoft Software Package For Use with the Modified IEC 60332-3 Chamber for Measuring Heat Release Rate and Smoke Production

More information

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA Software version 2.6 Table of contents 1 Introduction...3 1.1 Optical Sensor Technology...3 1.2 Installation...4 1.3 Operation...5

More information

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003 CEE 331 Lab 1 Page 1 of 9 SAFETY Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003 Laboratory exercise based on an exercise developed by Dr. Monroe Weber-Shirk The major safety hazard in this

More information

User s Guide. Vacuum Controller for liquid delivery systems

User s Guide. Vacuum Controller for liquid delivery systems Flow Control User s Guide Vacuum Controller for liquid delivery systems Precise Vacuum Control throughout the experiment Flow control Compatible with any perfusion system Ideal for Small Volume Delivery

More information

LABORATORY INVESTIGATION

LABORATORY INVESTIGATION LABORATORY INVESTIGATION MEASURING THE RATE OF PHOTOSYNTHESIS Light and Photosynthesis About 2.5-3 billion years ago a new chemical process, photosynthesis, was evolved by a unicellular life form. This

More information

Product Overview. Product Description CHAPTER

Product Overview. Product Description CHAPTER CHAPTER 1 This chapter provides a functional overview of the Cisco Redundant Power System 2300 and covers these topics: Product Description, page 1-1 Features, page 1-3 Supported Devices, page 1-4 Deployment

More information

How to set up and use DeWiggler Analyst

How to set up and use DeWiggler Analyst How to set up and use DeWiggler Analyst The most important mission of an instrument system is correctly reporting wind direction (see http://www.ockam.com/functrue.html). DeWiggler Analyst is an application

More information

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more! After years of development and testing, DPS Instruments is pleased to present the newest, most expandable and versatile Gas Chromatography Systems in history. The DPS 600 Series GC systems are the world

More information

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the

More information

UsER manual for Watersens ph -REDOX

UsER manual for Watersens ph -REDOX UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5

More information

Window and Door Measurement System

Window and Door Measurement System Inhaltsverzeichnis Window and Door Measurement System Air Permeability Test BlowerDoor GmbH Version: TECLOG2 for Windows January 2011 BlowerDoor GmbH März 2009 1 Instructions Window and Door Measurement

More information

FinePointe WBP. D a t a S c i e n c e s I n t e r n a t i o n a l TH STREET NW SUITE 100 ST. PAUL, MN USA (651) (800)

FinePointe WBP. D a t a S c i e n c e s I n t e r n a t i o n a l TH STREET NW SUITE 100 ST. PAUL, MN USA (651) (800) FinePointe WBP D a t a S c i e n c e s I n t e r n a t i o n a l 119 14TH STREET NW SUITE 100 ST. PAUL, MN 55112 USA (651)481-7400 1(800)262-9687 www.datasci.com PN: 007898-001 REV04 OVERVIEW This manual

More information

Wire Buddy Manual IOpath Software

Wire Buddy Manual IOpath Software Wire Buddy Manual IOpath Software Wire Buddy is a calculation reference tool for Electricians and Maintenance personnel. Wire Buddy was created for the Maintenance Electrician. There are many occasions

More information

Tru Flight TRUFLIGHT INSTALLATION GUIDE TRUGOLF.COM

Tru Flight TRUFLIGHT INSTALLATION GUIDE TRUGOLF.COM Tru Flight T R A C K I N G S Y S T E M TRUFLIGHT INSTALLATION GUIDE TRUGOLF.COM TruFlight Camera Tracking System Setup & Installation TruFlight : How It Works... 1 TruFlight Components... 1 TruFlight Installation...

More information

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering Edge Isolation Tool Standard Operating Procedure Version 1.1 Date: 20-9-2012 Prepared by, Sandeep S S Department of Electrical Engineering IIT Bombay This document is meant for internal circulation only

More information

Laboratory Mortar Mixer (Testing)

Laboratory Mortar Mixer (Testing) TomTom-Tools GmbH Zelgli 20 8905 Arni info@tomtom-tools.com Switzerland www.tomtom-tools.com User Manual Version February 22, 2015 Laboratory Mortar Mixer (Testing) 1 Introduction The Laboratory Mortar

More information

Technology. Using Bluetooth

Technology. Using Bluetooth Bluetooth is a communication technology that makes it possible to send and receive data without using wires. Using the Bluetooth features, you can set up a wireless connection between your NXT and other

More information

BUBBLER CONTROL SYSTEM

BUBBLER CONTROL SYSTEM BUBBLER CONTROL SYSTEM Description: The HDBCS is a fully automatic bubbler system, which does liquid level measurements in water and wastewater applications. It is a dual air compressor system with, air

More information

Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached

Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached Note: You must read the Primer manual Sections 1-6 before using the IRGA. The manual is located in the wall cabinet

More information

Strathkelvin Channel Oxygen System version 2.2 SAFETY AND TECHNICAL INFORMATION...1 THE 928 SYSTEM...2 GETTING STARTED...7

Strathkelvin Channel Oxygen System version 2.2 SAFETY AND TECHNICAL INFORMATION...1 THE 928 SYSTEM...2 GETTING STARTED...7 Strathkelvin 928 6-Channel Oxygen System version 2.2 CONTENTS SAFETY AND TECHNICAL INFORMATION...1 THE 928 SYSTEM...2 1.1 The 928 System...2 1.2 Minimum System Requirements...2 1.3 The 928 Interface...3

More information

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Universal pycnometer, no

More information

Using the Lego NXT with Labview.

Using the Lego NXT with Labview. Using the Lego NXT with Labview http://www.legoengineering.com/component/content/article/105 The Lego NXT 32-bit ARM microcontroller - an Atmel AT91SAM7S256. Flash memory/file system (256 kb), RAM (64

More information

WF STEUERUNGSTECHNIK GMBH. INFORMATION and TECNICAL DESCRIPTION

WF STEUERUNGSTECHNIK GMBH. INFORMATION and TECNICAL DESCRIPTION 1 INFORMATION and TECNICAL DESCRIPTION 14 2 AIRLEADER Professional Compressor Management + Compressed Air Visualisation + Alarm Service Management + AIRLEADER - has been the effective answer for compressed

More information

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Bante821 Portable Dissolved Oxygen Meter Instruction Manual Bante821 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante821 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante821 portable dissolved oxygen

More information

[MYLAPS INTEGRATION]

[MYLAPS INTEGRATION] 2018 The Race Director MyLaps Integration Manual [MYLAPS INTEGRATION] This document explains how to manage the results data between your MyLaps readers and Race Director using manual file transfers. Contents

More information

Technology. Using Bluetooth

Technology. Using Bluetooth Bluetooth is a communication technology that makes it possible to send and receive data without using wires. Using the Bluetooth features, you can set up a wireless connection between your NXT and other

More information

IBU3 Manual Addendum 1

IBU3 Manual Addendum 1 This addendum covers the follow items. Connection of airsoft micro switch Connection of Taigen airsoft recoil units Selecting airsoft mode Settings for clutch style gearboxes Motor enable selection Confirmation

More information

Cover Page for Lab Report Group Portion. Flow Visualization in a Water Channel

Cover Page for Lab Report Group Portion. Flow Visualization in a Water Channel Cover Page for Lab Report Group Portion Flow Visualization in a Water Channel Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 08 September 2017 Name 1: Name 2: Name 3: [Name

More information

In addition to reading this assignment, also read Appendices A and B.

In addition to reading this assignment, also read Appendices A and B. 1 Kinematics I Introduction In addition to reading this assignment, also read Appendices A and B. We will be using a motion detector to track the positions of objects with time in several lab exercises

More information

Heating of Land and Water

Heating of Land and Water Heating of Land and Water Experiment 8 How fast land and water heat and cool influences our weather. In Part I of this experiment, you will heat sand and water. You will use two Temperature Probes to see

More information

TESLAGON. ShotHelper Manual. How to install and use the Program. Version /30/2014

TESLAGON. ShotHelper Manual. How to install and use the Program. Version /30/2014 TESLAGON ShotHelper Manual How to install and use the Program 11/30/2014 Version 1.11 Table of Contents Introduction... 3 Installation Process... 3 ShotHelper Quick Setup... 4 The Main Window... 6 The

More information

Oxidation Stability of Gasoline and Aviation Fuels

Oxidation Stability of Gasoline and Aviation Fuels Oxidation Stability of Gasoline (Induction Period Method) Oxidation Stability of Aviation Fuels (Potential Residue Method) Test Method Provides an indication of the tendency of gasoline and aviation fuels

More information

DST Host User Manual

DST Host User Manual For DST Host version 7.0 onwards, published October 2017 Cefas Technology Limited CONTENTS About this Manual... 2 Conventions used in this Manual... 2 Getting Started... 3 Installing the Host Software...

More information

STM-3. Surface Station. Technical Reference Manual. October Desert Star Systems Imjin Road. Marina, CA (831)

STM-3. Surface Station. Technical Reference Manual. October Desert Star Systems Imjin Road. Marina, CA (831) STM-3 Surface Station Technical Reference Manual October 2007 Desert Star Systems 3261 Imjin Road Marina, CA 93933 (831) 384-8000 (831) 384-8062 FAX http://www.desertstar.com Copyright 2007, Desert Star

More information

EXPERIMENT 12 GAS LAWS ( BOYLE S AND GAY-LUSSAC S LAW)

EXPERIMENT 12 GAS LAWS ( BOYLE S AND GAY-LUSSAC S LAW) EXPERIMENT 12 GAS LAWS ( BOYLE S AND GAY-LUSSAC S LAW) INTRODUCTION: In order to specify fully the condition of a gas it is necessary to know its pressure, volume, and temperature. This quantities are

More information

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work.....2

More information

Figure SM1: Front panel of the multipatcher software graphic user interface (GUI) at the beginning of multipatcher operation.

Figure SM1: Front panel of the multipatcher software graphic user interface (GUI) at the beginning of multipatcher operation. APPENDIX 2. Multipatcher Software Setup and Operation. The multipatcher program is organized into four panels. There are controls that allow the user to specify various parameters into the system. The

More information