BekkTech s Procedures For Performing In Plane Membrane Conductivity Testing

Size: px
Start display at page:

Download "BekkTech s Procedures For Performing In Plane Membrane Conductivity Testing"

Transcription

1 BekkTech s Procedures For Performing In Plane Membrane Conductivity Testing Important Equipment and Software Considerations in Conductivity Testing Work Performed Under US DOE Contract # DE FC36 06GO16028 February 2008

2 Contents Reproducing BekkTech Conductivity Test Results... 3 Gas, Equipment & Software Requirements... 3 Gas Flow & System Pressure... 3 Equipment Requirements (BekkTech Recommendations)... 3 Software... 4 Preparing a Membrane Sample for Testing... 5 Equilibrate Sample to Room Conditions... 5 Cut Sample... 5 Verify Sample Width... 5 Sample Length... 5 Measure Sample Thickness... 6 Assemble Sample in the Conductivity Cell... 6 Insert Thermocouple... 7 Assemble Conductivity Cell with Cell Hardware... 7 Connect to Heated Gas Lines & Back Pressure Regulator... 8 Connect to Keithley SourceMeter... 9 Measuring Resistance... 9 Connect Cell Hardware to Heater Power and Heater Thermocouple Test Protocol for Samples Submitted to BekkTech Under the DOE Program Prepare Equipment & Software for 30C Test C Test Prepare Equipment & Software for 80C Test C Test C Test Data Analysis Calculating Conductivity Sample Graphs Generated by the Reporting Template Files Nafion 212 at 30C Nafion 212 at 80C Nafion 212 at 120C BekkTech Product Listing Contact Information: Page 2

3 Reproducing BekkTech Conductivity Test Results Because various types of equipment may be used to perform membrane conductivity testing, BekkTech does not offer a detailed procedure that is equipment specific, except for procedures specific to the equipment we sell. However, if you understand your equipment and your equipment can operate at the conditions we discuss, the following information includes all the parameters you will need to reproduce our tests. Please remember that the equipment used to perform conductivity testing may have a significant impact on the time it takes you to perform a test. It may also have a significant impact on your test results. We have tried to point out the areas where your choice of equipment will have the greatest impact on your test results. Of course, if you want to ensure that you can reproduce our test results, use BekkTech s BT 512 Membrane Conductivity Test System. If the option of purchasing this system is not available to you, use this information to set the parameters of your equipment to perform tests as similar to the following as possible. Gas, Equipment & Software Requirements BekkTech uses the BT 512 Membrane Conductivity Test System which includes all the capabilities, equipment, and software listed below. Gas Flow & System Pressure Gas Hydrogen 30 C Tests o Gas Flow Control Two Mass Flow Controllers are required for gas mixing to achieve dew points for 30 C testing SCCM flow controllers with accuracy of + / 1% of set point. o System Pressure 100 kpa absolute. This is approximate 15 kpa gauge (2 psig) in Colorado. 80 C Tests o Gas Flow Control One 1000 SCCM Mass Flow Controller with accuracy of + / 10% of set point. o System Pressure 100 kpa absolute. This is approximate 15 kpa gauge (2 psig) in Colorado. 120 C Tests o Gas Flow Control One 500 SCCM Mass Flow Controller with accuracy of + / 10% of set point. o System Pressure 230 kpa absolute. This is approximate 145 kpa gauge (21 psig) in Colorado. Equipment Requirements (BekkTech Recommendations) Fuel cell hardware capable of testing at 120C (Product #ACC 920 Fuel Cell Technologies 5cm 2 fuel cell test hardware) BekkTech Conductivity Cell for testing up to 120C (Product #BT 112 BekkTech Conductivity Cell) Saturator capable of controlling dew points to + / 1 degree absolute accuracy. Requires prompt reaction to temperature changes to reproduce relative humidity steps every 15 minutes. (Product #BT 104 BekkTech Saturator) Page 3

4 Potentiostat / Voltage Current Supply Must be capable of making a true 4 electrode measurement of a resistor in the 1 to 15 kohm range and allow at least 2 Volts between the current source and voltage sense leads. (Product #ACC 980 Keithley 2400 Sourcemeter with BekkTech Scanning DC Software) Temperature Controllers & Thermocouples Combined unit must have absolute accuracy of + / 1.5C with 0.2C reproducibility. (Product #BT 201 BekkTech Temperature Control System, includes Love Temperature Controllers and Type T Thermocouples) Heated Gas Inlet Line from the Saturator to the Cell Hardware which maintains the gas at a temperature above the dew point temperature. Care must be taken to avoid over heating the gas. (Product #BT 125 BekkTech Heated Gas Line) Water Management on the Gas Exit For accurate pressure control, the gas must be either heated to above the dew point through the back pressure regulator or water must be removed from the gas stream so that it does not interfere with the accurate control of back pressure. (Product #BT 130 BekkTech Heated Gas Exit Line and Back Pressure Regulator) Back Pressure Regulator sized appropriately for the gas and water flow associated with these tests. (Product #BT 130 BekkTech Heated Gas Exit Line and Back Pressure Regulator) Software To reproduce the graphs generated by the BekkTech procedures, you will need to know conductivity and relative humidity as a function of time. BekkTech s Conductivity Testing Software controls and records the following parameters, as a function of time: Gas Flow (includes wet & dry gas flow in gas mixing) Cell Temperature Saturator Temperature (Dew Point) Heated Gas Inlet Temperature Heated Gas Exit Temperature Voltage Current Pairs to Calculate Resistance Values BekkTech s Data Analysis Software takes the thousands of data points recorded in the Conductivity Testing Software, reduces the data, and generates the following graphs: Conductivity as a function of time Conductivity as a function of relative humidity Temperature of each component as a function of time Gas flow as a function of time Analysis of these graphs allows the researcher to determine the various factors affecting the conductivity of the sample under test. Important factors to consider if you are developing software for Conductivity Testing: 1) Cell temperature and dew point must be recorded as a function of time so that accurate relative humidity values can be calculated and trends can be observed. Page 4

5 2) Conductivity measurements must be recorded as a function of time and plotted with the temperature data so that trends can be observed. Preparing a Membrane Sample for Testing Equilibrate Sample to Room Conditions Place sample on clean counter surface for 24 hours to allow equilibration to room temperature and room humidity. BekkTech Facility Room Conditions: Room temperature of approximately 21C and average room relative humidity of 15% 25%. Cut Sample Use the Sample Punch to cut a piece from the membrane for testing. Sample to be tested measures approximately 5mm x 25 mm. Figure 1: Sample Punch BekkTech Product #ACC 960 Alternative Cutting Method: Use an Exacto knife and a straight edge to cut the sample. The advantage of using the Sample Punch is more consistent sample sizes with less trauma to the entire sample. Verify Sample Width Use the Width Measurement Tool to verify the width of the sample you will be testing. The Width Measurement Tool has magnification of 11x and a reticule with 0.1mm gradients. Figure 2: Width Measurement Tool BekkTech Product #ACC 940 Alternative Width Measurement: Using the Width Measurement Tool or loupe will provide accurate measurements. Other loupes may provide accurate results. Rulers may also be used. But, are much less accurate. Sample Length Sample length is set by the BekkTech Conductivity Cell. We use a length of 4.2mm for all conductivity calculations. Page 5

6 Measure Sample Thickness The dimensional measurement with the greatest chance of error is sample thickness. The challenge of this measurement is in the accuracy of the gauge which is significantly affected by the skill of the operator. Operator error is dramatically reduced by using a gauge with a non rotating stem and consistently applied force. We use a Mitutoyo Gauge which applies a constant pressure of 3.5 Newtons to the sample using a non rotating stem. Rated accuracy is + / 3 microns. Even by the ratings of this high quality gauge, there is an error of + / 6% on a 50 micron sample. Careful training and use will allow users to achieve gauge accuracy. Figure 3: Mitutoyo Gauge BekkTech Product #ACC 950 Assemble Sample in the Conductivity Cell Remove the top clamp of the Conductivity Cell. Use tweezers to gently lift the platinum wires up from the bottom clamp. Insert the membrane sample under the platinum wires. Place the sample under the platinum wires of the Conductivity Cell. Replace the top clamp. Figure 5: Assemble Sample Under Pt Wires Figure 4: Place Sample in Conductivity Cell Conductivity Cell is BekkTech Product #BT 112 Figure 6: Replace Top Clamp Page 6

7 Insert Thermocouple The thermocouple (TC) should be placed so that the TC Junction (location where temperature is measured) is directly below the membrane. The TC Junction is generally 0.5 mm to 1.0 mm from the end of the TC. Figure 7: Internal Thermocouple Placement Equipment Recommendations: BekkTech uses Type T thermocouples. For accurate measurement, do not use a TC with a floating junction. Use a TC with a grounded junction. Assemble Conductivity Cell with Cell Hardware The Conductivity Cell shown is designed to be used with Fuel Cell Technologies 5cm 2 fuel cell test hardware. Place the Conductivity Cell in the Fuel Cell Technologies cell hardware. Re assemble the cell hardware. You will need longer bolts than the standard set that is included with the cell hardware. BekkTech provides the longer bolts with the Conductivity Cell. You will use only 4 bolts to secure the assembly. The Conductivity Cell makes a good seal with the 4 bolts. Figure 8: Assemble Conductivity Cell Into Cell Hardware Cell Hardware is BekkTech Product #ACC 920 Figure 9: Assembled Page 7

8 Connect to Heated Gas Lines & Back Pressure Regulator Connect the heated gas line from your Saturator to the gas inlet port on your cell hardware. You will use only one of the two gas inlet ports on your cell hardware. Plug the 2 nd gas inlet port with a ¼ stainless steel cap. Gas Inlet Port Plugged Gas Inlet Port Open Figure 10: Connect to Gas Inlet Line Heated Gas Inlet Lines: BekkTech Product #BT 125 Connect the gas line from the cell hardware to your Back Pressure Regulator. We heat the gas exit line as added precaution against condensation. Gas Exit Port Plugged Gas Exit Port Open Figure 11: Connect to Gas Exit Line Heated Gas Exit Line & Back Pressure Regulator: Product #BT 130 You must pressurize your system for conductivity tests at dew points above approximately 100 C to prevent the water in your Saturator from boiling. The exact boiling point will vary depending on the atmospheric pressure of your location. Page 8

9 Connect to Keithley SourceMeter Connect cable with 4 connections to the front of the Keithley Sourcemeter in the ports labeled 4 Wire Sense and Input/Output. The clips on the end of this cable will connect to the platinum wires in the Conductivity Cell, from left to right, as follows: Large Red Clip Current Small Red Clip Voltage Small Black Clip Voltage Large Black Clip Current Figure 12: Keithley SourceMeter Cables Connected to Conductivity Cell Platinum Leads Keithley SourceMeter with Scanning DC Software: BekkTech Product #ACC 980 Measuring Resistance BekkTech uses a Keithley SourceMeter to set a voltage and measure the resulting current for conductivity tests. We perform a Scanning DC Sweep from 0.1V to 0.1V. The linear Voltage Current data is fit for resistance. Alternative Equipment: A potentiostat may also be used to perform a linear sweep. The sweep must be linear even as current flow changes polarity in order for the data to be valid. Or, a potentiostat and frequency response analyzer (FRA) may be used to measure AC Impedance. Impedance should remain constant from the mhz to khz range. And, the phase angle should be very small. Page 9

10 Connect Cell Hardware to Heater Power and Heater Thermocouple Plug the Internal Cell Thermocouple (TC) into the Thermocouple Outlet. Plug the power cable from the Cell Hardware Heaters into the Power Outlet on the test stand. CAUTION: We recommend that you always wait to connect the Cell Hardware Heaters to Power until after you have verified that the Cell Hardware Internal TC is properly placed in the Cell Hardware. If this TC is accidentally left out of the Cell Hardware or improperly installed, the TC will communicate an incorrect temperature to the Cell Hardware Temperature Controller. Irreparable damage to the Cell Hardware and Conductivity Cell may result. Test Protocol for Samples Submitted to BekkTech Under the DOE Program Each sample is tested for approximately 24 hours first at 30C, then 80C, then 120C. We use the BekkTech Conductivity Testing Software to automatically control each 8 hour test. Other methods of controlling the test and acquiring data are available. We will only be describing the procedures we use. A membrane sample is assembled into the Conductivity Cell/Cell Hardware fixture and tested at 30C, 80C, and 120C without removal from the Cell Hardware. Prepare Equipment & Software for 30C Test Ensure that the valve on the dry gas line which connects into the wet gas line is open. Open the BekkTech Conductivity Test Software. Click on the arrow in the upper left corner of the page. This will change the arrow from white to black. The black arrow indicates that the software is running. Ensure that the system is under operator control. The System Control button will read System Under Operator Control. If it does not, click on it to toggle to System Under Operator Control. Go to the Voltage Current tab. Click on Activate Keithley. Confirm that linear voltage current data is being collected. Go to the DataFile tab. Enter a new file name for the test to be performed. Click on Open Data File. Data is now being recorded. Go to the ScriptMode tab. Click on Upload Script File. From the drop down menu, select 30C Conductivity. The 30C script will upload. Review the script to ensure that you have uploaded the correct file. The script file includes gas flow rate, temperatures, dew points, and time for each step. Click on the System Control button so that it toggles to System Under Script Control. Set the back pressure of the system to 100kPa absolute. This is approximate 15 kpa gauge (2 psig) in Colorado. The system will now automatically run the 30C Conductivity Test. Page 10

11 30C Test The 30C script will control the system to the following parameters. Gas Flow 1000 SCCM Hydrogen Gas Mixing The dew points are set using a mixture of dry gas ( 50C) with gas saturated to a dew point of 45C. System Pressure 100 kpa absolute. This is approximate 15 kpa gauge (2 psig) in Colorado. Sample Conditioning Wet up at 30C, 70% RH for 2 hours. After 2 Hours of Conditioning Operating conditions are adjusted every 15 minutes. Note that 2 Mass Flow Controllers are required for this temperature. One flows gas through the Saturator (Wet Gas). The second mixes with the saturated gas after the Saturator (Dry Gas). Relative Humidity ( %RH) Wet Gas Flow Rate (45 C dew point) (SCCM) Dry Gas Flow Rate (-50 C dew point) (SCCM) 70% % % % % % % % % % % % % % % % % When the test has completed the script, take the system out of Script Mode. Then, go to the DataFile tab and click on Close Data File. Page 11

12 Prepare Equipment & Software for 80C Test Ensure that the valve on the dry gas line which connects into the wet gas line is closed. Gas mixing is not needed for controlling dew points at 80C. Go to the DataFile tab and enter a new data file name for the 80C test. Click on Open Data File. Go to the ScriptMode tab. Click on Upload Script. Select the 80C Conductivity Script. The 80C script will upload. Review the script to ensure that you have uploaded the correct file. Click on the System Control button so that it toggles to System Under Script Control. Set the back pressure of the system to 100kPa absolute. This is approximate 15 kpa gauge (2 psig) in Colorado. 80C Test The system will now run the 80C test at the following parameters: Gas Flow 1000 SCCM Hydrogen Gas Mixing None System Pressure 100 kpa absolute. This is approximate 15 kpa gauge (2 psig) in Colorado. Sample Conditioning Wet up at 80C, 70% RH for 2 hours. After 2 Hours of Conditioning Operating conditions are adjusted every 15 minutes, as follows: Dew Point of 67.9 C for 60% RH Dew Point of 63.8 C for 50% RH Dew Point of 58.9 C for 40% RH Dew Point of 52.9 C for 30% RH Dew Point of 49.2 C for 25% RH Dew Point of 44.8 C for 20% RH Dew Point of 49.2 C for 25% RH Dew Point of 52.9 C for 30% RH Dew Point of 58.9 C for 40% RH Dew Point of 63.8 C for 50% RH Dew Point of 67.9 C for 60% RH Dew Point of 71.4 C for 70% RH Dew Point of 74.6 C for 80% RH Dew Point of 77.4 C for 90% RH Dew Point of 78.7 C for 95% RH Dew Point of 80.0 C for 100% RH When the system has run through the end of this script, take the system out of script control. Then, go to the DataFile tab and click on Close Data File. Page 12

13 120C Test Go to the Data File field and enter a new data file name for the 120C test. Click on Open Data File. Go to the ScriptMode tab. Click on Upload Script. Select the 120C Conductivity Script. The 120C script will upload. Review the script to ensure that you have uploaded the correct file. Click on the System Control button so that it toggles to System Under Script Control. Set the back pressure to 230 kpa absolute. This is approximate 145 kpa gauge (21 psig) in Colorado. The system will now run the 120C test at the following parameters: Gas Flow 500 SCCM Hydrogen Gas Mixing None System Pressure 230 kpa absolute. This is approximate 145 kpa gauge (21 psig) in Colorado. Sample Conditioning Wet up at 80C, 70% RH for 2 hours. After 2 Hours of Conditioning Operating conditions are adjusted every 15 minutes, as follows: Dew Point of C for 60% RH Dew Point of 99.2 C for 50% RH Dew Point of 93.0 C for 40% RH Dew Point of 85.4 C for 30% RH Dew Point of 81.1 for 25% RH Dew Point of 75.2 C for 20% RH Dew Point of 81.1 for 25% RH Dew Point of 85.4 C for 30% RH Dew Point of 93.0 C for 40% RH Dew Point of 99.2 C for 50% RH Dew Point of C for 60 % RH Dew Point of C for 70% RH Dew Point of C for 80% RH Dew Point of C for 90 % RH Dew Point of C for 95% RH Dew Point of C for 100% RH When the test has completed running through the script, take the system out of script control. Go to the DataFile tab and close the data file. Purge the system with Nitrogen. Cool the system by reducing the temperatures on the temperature controllers. De activate the Keithley SourceMeter. End the Program. Close the Conductivity Testing Software. Turn the Temperature Control Box power to Off. Turn the Keithley Sourcemeter power to Off. Page 13

14 Data Analysis The data you collected is saved in the files you created at the beginning of each test. Open the BekkTech Data Conversion file. Enter the file name of the 30C data file you created during the 30C test. This is the file name you entered on the DataFile tab. Open the Reporting Template file. Go to the Data Conversion file. Click on the Convert Data File button. The program will reduce the raw data based on the parameters set in the Data Conversion file and write the reduced data into the Reporting Template file. Review the resulting tables and graphs in the Reporting Template file for reasonableness. Save the Reporting Template file using an appropriate file name. You will perform the procedures listed above for the 80C and 120C data files also. The Reporting Template files you create will generate graphs such as the following. Calculating Conductivity The formula for calculating conductivity is included in the BekkTech Software. To manually calculate conductivity, use the following information. R is calculated using a Least Squares Fit of the voltage current data. I V ΔV R = ΔI W T σ = L = R* A L R* W * T L σms = 1cm Lmm* 10mm 1000mS * 1cm 1cm Rohms * Wmm * * Tmicrons * 1S 10mm 10,000microns Page 14

15 Sample Graphs Generated by the Reporting Template Files Nafion 212 at 30C Four Electrode Conductivity 30 C at 100 kpa Relative Humidity (%RH) 110% 100% 90% 80% 70% 60% 50% 40% 30% NRE212 at 30 C ( ) Relative Humidity Select Relative Humidities NRE212 at 30 C ( ) Conductivity Select 4 Electrode Conductivities Conductivity Calculated based on dry dimensions and no swelling Conductivity (ms / cm) 20% Time (Minutes) Figure 13: Nafion 212 at 30C Conductivity as a function of time Comparing Nafion Samples at 30 C 100 kpa 1000 NRE212 at 30 C ( ) N112 at 30 C ( ) Conductivity (ms/cm) Conductivity Calculated based on dry dimensions and no swelling 1 10% 20% 30% 40% 50% 60% 70% 80% 90% 100% 110% Relative Humidity (%RH) Figure 14: Nafion 212 at 30C Conductivity as a function of relative humidity Page 15

16 Nafion 212 at 80C 110% Four Electrode Conductivity 80 C at 100 kpa 180 Relative Humidity (%RH) 100% 90% 80% 70% 60% 50% NRE212 at 80 C ( ) Relative Humidity Select Relative Humidities NRE212 at 80 C ( ) Conductivity Select 4 Electrode Conductivities 40% 40 Conductivity Calculated 30% based on dry dimensions and no swelling 20 20% Time (Minutes) Conductivity (ms / cm) Figure 15: Nafion 212 at 80C Conductivity as a function of time Comparing Four Electrode Conductivity of N C 100 kpa 1000 NRE212 at 80 C ( ) N112 at 80 C ( ) Conductivity (ms/cm) Conductivity Calculated based on dry dimensions and no swelling 1 10% 20% 30% 40% 50% 60% 70% 80% 90% 100% 110% Relative Humidity (%RH) Figure 16: Nafion 212 at 80C Conductivity as a function of relative humidity Page 16

17 Nafion 212 at 120C Four Electrode Conductivity 120 C at 230 kpa Relative Humidity (%RH) 110% 100% 90% 80% 70% 60% 50% 40% NRE212 at 120 C ( ) Relative Humidity Select Relative Humidities NRE212 at 120 C ( ) Conductivity Select 4 Electrode Conductivities Conductivity Calculated based on dry dimensions and no swelling Conductivity (ms / cm) 30% 40 20% Time (Minutes) 0 Figure 17: Nafion 212 at 120C Conductivity as a function of time Comparing Four Electrode Conductivity of N C 230 kpa 1000 NRE212 at 120 C ( ) N112 at 120 C ( ) Conductivity (ms/cm) Conductivity Calculated based on dry dimensions and no swelling 1 10% 20% 30% 40% 50% 60% 70% 80% 90% 100% 110% Relative Humidity (%RH) Figure 18: Nafion 212 at 120C Conductivity as a function of relative humidity Page 17

18 BekkTech Products Referenced in this Document ACC 920 Fuel Cell Technologies 5cm 2 fuel cell hardware ACC 940 Width Measurement Tool ACC 950 Thickness Measurement Tool ACC 960 Sample Punch ACC 980 Keithley 2400 SourceMeter with BekkTech Scanning DC Software BT 104 BekkTech Saturator BT 112 BekkTech Conductivity Cell for Fuel Cell Technologies cell hardware BT 125 BekkTech Heated Gas Inlet Line (includes heating pad & Type T thermocouple) BT 130 BekkTech Heated Gas Exit Line & Back Pressure Regulator BT 201 BekkTech Temperature Control System (includes temperature controllers and thermocouples) BT 512 BekkTech Membrane Conductivity Test System BT 3000 BekkTech Data Analysis and Report Template Software Note: The BekkTech Conductivity Testing Software is available with the purchase of the BT 512 Membrane Conductivity Test System. Contact Information: BekkTech LLC tech@bekktech.com Phone: +1 (970) Fax: +1 (970) Page 18

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

Aerobic Respiration. Evaluation copy

Aerobic Respiration. Evaluation copy Aerobic Respiration Computer 17 Aerobic cellular respiration is the process of converting the chemical energy of organic molecules into a form immediately usable by organisms. Glucose may be oxidized completely

More information

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Universal pycnometer, no

More information

Mass Spec will not Autotune

Mass Spec will not Autotune Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems

More information

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cover Page for Lab Report Group Portion. Head Losses in Pipes Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section

More information

Gas Sampling Fitting with MK2.1 and L-Probe

Gas Sampling Fitting with MK2.1 and L-Probe measuring controlling automation measuring controlling automation measuring controlling automation Page i von 8 Operation Manual Gas Sampling Fitting with MK2.1 and L-Probe Page ii von 8 Manufacturer MESA

More information

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003 CEE 331 Lab 1 Page 1 of 9 SAFETY Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003 Laboratory exercise based on an exercise developed by Dr. Monroe Weber-Shirk The major safety hazard in this

More information

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1 HPICAL-15000 Operation & Data Logging Procedures Click spacebar to advance through slides 1 WARNING Always wear proper safety equipment when using high pressure equipment. Do not exceed 125 psi air pressure.

More information

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc. The HumiSys RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The new HumiSys with single or dual RH probes capabilities is

More information

Transpiration. DataQuest OBJECTIVES MATERIALS

Transpiration. DataQuest OBJECTIVES MATERIALS Transpiration DataQuest 13 Water is transported in plants, from the roots to the leaves, following a decreasing water potential gradient. Transpiration, or loss of water from the leaves, helps to create

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Temperature controlled, multi-technique volumetric

More information

HOBO U20 Water Level Logger (U x and U x-Ti) Manual

HOBO U20 Water Level Logger (U x and U x-Ti) Manual HOBO U20 Water Level Logger (U20-001-0x and U20-001-0x-Ti) Manual The HOBO U20 Water Level Logger is used for monitoring changing water levels in a wide range of applications including streams, lakes,

More information

Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip.

Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip. Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip. CEE 3310 - Summer 2012 SAFETY The major safety hazard in this laboratory is a shock hazard. Given that you will be working

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

Met One E-BAM Particulate Monitor

Met One E-BAM Particulate Monitor STANDARD OPERATING PROCEDURES Met One E-BAM Particulate Monitor AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION PROJECT LOWER MANHATTAN DEVELOPMENT CORPORATION 1 Liberty Plaza

More information

HumiSys HF High Flow RH Generator

HumiSys HF High Flow RH Generator HumiSys HF High Flow RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The HumiSys HF is a high flow version of the previously

More information

Exercise 3. Power Versus Wind Speed EXERCISE OBJECTIVE DISCUSSION OUTLINE. Air density DISCUSSION

Exercise 3. Power Versus Wind Speed EXERCISE OBJECTIVE DISCUSSION OUTLINE. Air density DISCUSSION Exercise 3 Power Versus Wind Speed EXERCISE OBJECTIVE When you have completed this exercise, you will know how to calculate the power contained in the wind, and how wind power varies with wind speed. You

More information

CCT-7320/ROC-2313 Reverse Osmosis Controller

CCT-7320/ROC-2313 Reverse Osmosis Controller CCT-7320/ROC-2313 Reverse Osmosis Controller 1 General The instrument is a combined control instrument of a reverse osmosis controller and an on-line conductivity instrument. It can perform the operation

More information

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW AC 2010-2145: MEASUREMENT OF HYDROGEN IN HELIUM FLOW Randy Buchanan, University of Southern Mississippi Christopher Winstead, University of Southern Mississippi Anton Netchaev, University of Southern Mississippi

More information

AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP

AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP USER MANUEL AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP TIRE EQUIPMENT MANUFACTURER 1.866.409.RACK WWW.MARTINSINDUSTRIES.COM info@martinsindustries.com PARTS Verify that the following components

More information

GCMSD-Headspace Analysis SOP

GCMSD-Headspace Analysis SOP Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to

More information

Onset Computer Corporation Inside this package: HOBO U20 Water Level Logger Calibration Certificate

Onset Computer Corporation Inside this package: HOBO U20 Water Level Logger Calibration Certificate Onset Computer Corporation Inside this package: HOBO U20 Water Level Logger Calibration Certificate Models The Water Level logger is available in the following models: U20-001-01 (30 foot depth) and U20-001-01-Ti

More information

OC Panel High Limit Aquastat Kit, Manual Reset p/n

OC Panel High Limit Aquastat Kit, Manual Reset p/n OC Panel High Limit Aquastat Kit, Manual Reset p/n 233202 Instruction Sheet APPLICATION The OC (Option Control) Panel High Limit Aquastat Kit provides electronic temperature sensing in a UL limit-rated

More information

A4s Operation Manual

A4s Operation Manual A4s Operation Manual Safety Instruction Please read this manual carefully, also with related manual for the machinery before use the controller. For installing and operating the controller properly and

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

Glove Box Installation Manual

Glove Box Installation Manual Glove Box Installation Manual 1998 by M. Braun Company File: GB-UNI-INS.DOC! Edition 08-00 by M. Boutin! Subject to be changed without notice Glovebox Installation Your Glove box has been fully assembled,

More information

HOBO U20L Water Level Logger (U20L-0x) Manual

HOBO U20L Water Level Logger (U20L-0x) Manual Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com HOBO U20L Water Level Logger (U20L-0x) Manual The HOBO U20L Water Level Logger is used for monitoring

More information

Operating Instructions for Cobra and Lynx Duct Leakage Testers

Operating Instructions for Cobra and Lynx Duct Leakage Testers Operating Instructions for Cobra and Lynx Duct Leakage Testers Table of Contents Introduction Introduction... 2 Safety Precautions Safety Precautions.. 2 Prior to Testing System Preparation.. 4 Power Requirements...

More information

FPG8601 Force Balanced Piston Gauge

FPG8601 Force Balanced Piston Gauge FPG8601 Force Balanced Piston Gauge Reference Level Calibration System for very low pressure Pressure range: 0 to 15 kpa gauge, absolute and absolute differential Standard resolution: 0.010 Pa, high resolution

More information

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Background There are two main sources of energy available for animal metabolism: carbohydrates (CHO) and fats. These molecules are broken

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work.....2

More information

Operator Quick Guide ORBISPHERE 3654

Operator Quick Guide ORBISPHERE 3654 Operator Quick Guide ORBISPHERE 3654 Revision H - 14/03/2008 Operating Information About this Guide The information in this guide has been carefully checked and is believed to be accurate. However, Hach

More information

A4 Operation Manual. Fig.1-1 Controller Socket Diagram

A4 Operation Manual. Fig.1-1 Controller Socket Diagram A4 Operation Manual Safety Instruction Please read this manual carefully, also with related manual for the machinery before use the controller. For installing and operating the controller properly and

More information

(AS AT 31 st MARCH, 2002)

(AS AT 31 st MARCH, 2002) ACACA PROTOCOL 2000 (AS AT 31 st MARCH, 2002) ACACA PROTOCOL 2000 INCLUDES (A) CODE OF PRACTICE FOR MANUFACTURERS AND/OR SUPPLIERS OF COMMERCIAL AIR COMPRESSORS AND METHOD FOR DETERMINING (B) RECIPROCATING

More information

Dissolved Oxygen Sensor

Dissolved Oxygen Sensor Instruction Manual 499ADO and 499ADO-VP LIQ_MAN_ABR_499ADO May 2014 Dissolved Oxygen Sensor CAUTION For additional information, please visit our website at www.rosemountanalytical.com SENSOR/PROCESS APPLICATION

More information

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii Operator s Manual Table of Contents Operator Safety... 1 Introduction... 2 Unpacking and Setup... 3 Unpacking... 3 Setup... 4 FLYER Overview... 5 FLYER Diagram... 5 Capabilities... 5 Control Panel... 6

More information

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary ADH 1/7/014 LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the

More information

MPM280 Piezoresistive OEM Pressure Sensor

MPM280 Piezoresistive OEM Pressure Sensor MPM280 Piezoresistive OEM Pressure Sensor MPM280 Pressure Sensor (V1.2) Features Pressure range 0kPa~20kPa 70MPa; Gauge, absolute and sealed gauge; Constant current / Constant Voltages power supply; Isolated

More information

SRI Model H2-40 Hydrogen Generator Operation and Maintenance Nov 2006

SRI Model H2-40 Hydrogen Generator Operation and Maintenance Nov 2006 Nov 2006 Place the H2-40 on a benchtop surface which will not be damaged by water in the event of a leak. Fill a bottle with clean water. Tapwater is OK in most large cities, but bottled drinking water

More information

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the

More information

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300 Technical Note GA-300 Overview The GA-300 CO2 and O2 Gas Analyzer is easy to use, robust, and adaptable to human, animal, and plant applications. The GA-300 has two analog outputs to allow recording and

More information

WP2 Fire test for toxicity of fire effluents

WP2 Fire test for toxicity of fire effluents Pagina 3 di 89 TRA SFEU VTT 22.6.2009 v.2 WP2 Fire test for toxicity of fire effluents Task 2.1.2 Development of small-scale test method for fire effluents Step 1: Use of modeling Plans according to DoW:

More information

E8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm)

E8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm) Pressure Sensor (Stainless Steel Diaphragm) CSM DS_E_3_1 Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications Incorporates double diaphragms consisting of SUS316L stainless

More information

TECHNICAL BULLETIN CRYOGENIC APPLICATION SERVICES. website:

TECHNICAL BULLETIN CRYOGENIC APPLICATION SERVICES.  website: TECHNICAL BULLETIN SERVICES website: email: www.pipingtech.com info@pipingtech.com copyright 2009 Piping Technology & Products, Inc. SERVICES Objective TECHNICAL BULLETIN To determine the amount of ice

More information

Roller AC Servo System

Roller AC Servo System Safely Instruction Roller AC Servo System HMI-15 User Manual Please read this manual carefully, also with related manual for the machinery before use the controller. For installing and operating the controller

More information

Pressure Independent Control Series

Pressure Independent Control Series Document No. 155-522 Pressure Independent Control Series Two-Way Cast Iron Flanged Bodies, ANSI 125 and 250 Description Siemens Pressure Independent Control Valves integrate three functions into a single

More information

LABORATORY INVESTIGATION

LABORATORY INVESTIGATION LABORATORY INVESTIGATION MEASURING THE RATE OF PHOTOSYNTHESIS Light and Photosynthesis About 2.5-3 billion years ago a new chemical process, photosynthesis, was evolved by a unicellular life form. This

More information

TR Electronic Pressure Regulator. User s Manual

TR Electronic Pressure Regulator. User s Manual TR Electronic Pressure Regulator Page 2 of 13 Table of Contents Warnings, Cautions & Notices... 3 Factory Default Setting... 4 Quick Start Procedure... 5 Configuration Tab... 8 Setup Tab... 9 Internal

More information

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI NanoSight NS300 Operation instructions Injection/flushing brief overview: 1. Do not exceed flow of 1 ml per 20 seconds. 2. Inject two 1 ml syringes with nano-pure or DI water. 3. If the water does not

More information

MPM280 Piezoresistive OEM Pressure Sensor

MPM280 Piezoresistive OEM Pressure Sensor MPM280 Piezoresistive OEM Pressure Sensor MPM280 Pressure Sensor (V2.0) Features Pressure range 0kPa~20kPa 100MPa; Gauge, absolute and sealed gauge; Constant current / Constant Voltages power supply; Isolated

More information

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12 LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/ This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the rates with which selected

More information

FTC130 Transmitter. Operating Manual

FTC130 Transmitter. Operating Manual Seite 1 von 11 FTC130 Transmitter Fast Thermal Conductivity Analyzer Operating Manual 1.09KD180724CWI1V05 Messkonzept GmbH Seite 2 von 11 1. Intended Use... 3 2. Description... 4 3. Measuring gases and

More information

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes Instruction Bulletin Addendum IB PB1000 Part no. Serial no. Order no. Oxymitter 4000 Applicability: Used with pressure-balancing

More information

Instructions for SMV 3000 Multivariable Configuration (MC) Data Sheets

Instructions for SMV 3000 Multivariable Configuration (MC) Data Sheets Instructions for SMV 3000 Multivariable Configuration (MC) Data Sheets Similar to the TC option for ST 3000 transmitters, the MC option for the SMV 3000 provides a service to our customers which results

More information

Oxygen Meter User Manual

Oxygen Meter User Manual Oxygen Meter User Manual Monday, July 23, 2007 1. Outline...2 2. Program...3 2.1. Environment for program execution...3 2.2. Installation...3 2.3. Un installation...3 2.4. USB driver installation...3 2.5.

More information

SOP: Buck Scientific BLC-20P HPLC Operation

SOP: Buck Scientific BLC-20P HPLC Operation Page 1 of 11 Approvals Preparer: John Buford Reviewer: Tim Kull Reviewer: Dr. Margaret Bryans Date: 16OCT13 Date: 30OCT13 Date: 31OCT13 1. Purpose 1.1. Basic operation of the Buck Scientific BLC-20P isocratic

More information

DN 50 REGULATOR SUTON 5000A1 REGULATOR SUTON 5000D1

DN 50 REGULATOR SUTON 5000A1 REGULATOR SUTON 5000D1 www.apq.com.es EU Product / Spain DN 50 REGULATOR SUTON 5000A1 REGULATOR SUTON 5000D1 Cod. RISUT5000A1 Cod. RISUT5000D1 DESCRIPTION The Suton 50001 Regulator is designed for use in distribution networks

More information

Lab 4: Transpiration

Lab 4: Transpiration Lab 4: Transpiration Water is transported in plants, from the roots to the leaves, following a decreasing water potential gradient. Transpiration, or loss of water from the leaves, helps to create a lower

More information

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS Precise and Stable SF 6 Gas Analyzer Simultaneous measurement of humidity, SF 6 purity and SO 2 concentration Integrated gas recovery system with automatic pump back Fully automated SF 6 gas testing Fundamental

More information

Constant Pressure Inlet (CCN) Operator Manual

Constant Pressure Inlet (CCN) Operator Manual Constant Pressure Inlet (CCN) Operator Manual DOC-0125 Revision J 2545 Central Avenue Boulder, CO 80301-5727 USA C O P Y R I G H T 2 0 1 1 D R O P L E T M E A S U R E M E N T T E C H N O L O G I E S, I

More information

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Bante821 Portable Dissolved Oxygen Meter Instruction Manual Bante821 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante821 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante821 portable dissolved oxygen

More information

#LZ400 LEAKALYZER. Water Loss Sensor INSTRUCTION MANUAL Country Dr. #190 St. Paul, MN

#LZ400 LEAKALYZER. Water Loss Sensor INSTRUCTION MANUAL Country Dr. #190 St. Paul, MN #LZ400 LEAKALYZER Water Loss Sensor INSTRUCTION MANUAL 2885 Country Dr. #190 St. Paul, MN 55117 800-348-1316 www.leaktools.com Your Partner in Swimming Pool Water Conservation Product Purpose: 2 The Leakalyzer

More information

Mass Flow Controller (MFC) for Gases

Mass Flow Controller (MFC) for Gases Mass Flow Controller (MFC) for Gases Type 8713 can be combined with... Direct flow measurement by MEMS- Technology for nominal flow rates from 1 ml N /min to 8 l N /min (N 2 ) High accuracy and repeatability

More information

TH-1800-T TRX-700-CNG Converter

TH-1800-T TRX-700-CNG Converter TH-1800-T TRX-700-CNG Converter low detector The CNG fuel gas flowmeter is so accurate as to meet a demand of measuring the CNG directly by making the most use of our well-established mini-thermal flowmeter

More information

Improve Process Reliability

Improve Process Reliability Convenient Conductivity Calibration to Meet Pharmacopeia Regulations Pharmacopeias require that conductivity measurement systems be periodically calibrated to ensure their precision. The range of accessories

More information

Vapor Pressure of Liquids

Vapor Pressure of Liquids Experiment 10 In this experiment, you will investigate the relationship between the vapor pressure of a liquid and its temperature. When a liquid is added to the Erlenmeyer flask shown in Figure 1, it

More information

Lab 1. Adiabatic and reversible compression of a gas

Lab 1. Adiabatic and reversible compression of a gas Lab 1. Adiabatic and reversible compression of a gas Introduction The initial and final states of an adiabatic and reversible volume change of an ideal gas can be determined by the First Law of Thermodynamics

More information

Equipment Accuracy / Selection

Equipment Accuracy / Selection Revision 5.0: Included the PPP program for clarification. Revision 6.0: Update references for CTL PDSH 251B to CTL DSH 251E" Added leakage current tolerances for values above 5KHz For Client Labs Purpose

More information

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the

More information

Updated: Variable Temperature

Updated: Variable Temperature Updated: 2016-11-02 Variable Temperature Sample Preparation Make sure the probe is rated for the ultimate/final temperature Fully degas the sample, particularly if going to low temperature Be aware of

More information

OFI MODEL 20 CONSTANT SPEED BLENDER INSTRUCTIONAL MANUAL

OFI MODEL 20 CONSTANT SPEED BLENDER INSTRUCTIONAL MANUAL OFI MODEL 20 CONSTANT SPEED BLENDER INSTRUCTIONAL MANUAL OPERATING INSTRUCTIONS OFI MODEL 20 CONSTANT SPEED BLENDER GENERAL OFI's Model 20 Constant Speed Blender was designed to prepare well cements for

More information

Operations Manual. Internal automatic Pressure control for orbital tube welding. By Exel Orbital Systems, Inc. Page

Operations Manual. Internal automatic Pressure control for orbital tube welding. By Exel Orbital Systems, Inc. Page Operations Manual Internal automatic Pressure control for orbital tube welding. By Exel Orbital Systems, Inc. Page SG1 Overview and Dimensions 5.25 / 133 mm PS Power Supply Connection 4.06 / 103 mm Top

More information

T208VR Series Tank Blanketing Vacuum Regulator

T208VR Series Tank Blanketing Vacuum Regulator February 2014 T208VR Series Tank Blanketing Vacuum Regulator Figure 1. Typical T208VR Series Vacuum Regulator Introduction The T208VR Series direct-operated vacuum regulators are used where a decrease

More information

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V O2100C Oxygen Measurement Module The O2100C module measures the partial pressure of O2 and thus the module output is proportional to the pressure in the sample cell. Gas sampled must be free of liquids

More information

Free Chlorine Sensor For additional information, please visit our website at

Free Chlorine Sensor For additional information, please visit our website at Instruction Manual 499ACL01 LIQ_MAN_ABR_499ACL01 April 2014 Free Chlorine Sensor For additional information, please visit our website at www.rosemountanalytical.com. CAUTION Sensor/Process Application

More information

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante810 Benchtop Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante810 benchtop dissolved oxygen

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0545 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE TELEDYNE HASTINGS TECHNICAL PAPERS INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE Proceedings of FEDSM 98 1998 ASME Fluids Engineering

More information

PRESSURE-TEMPERATURE RELATIONSHIP IN GASES

PRESSURE-TEMPERATURE RELATIONSHIP IN GASES PRESSURE-TEMPERATURE RELATIONSHIP IN GASES LAB PS2.PALM INTRODUCTION Gases are made up of molecules that are in constant motion and exert pressure when they collide with the walls of their container. The

More information

AIR FLOW ANEMOMETER INSTRUCTION MANUAL

AIR FLOW ANEMOMETER INSTRUCTION MANUAL AIR FLOW ANEMOMETER INSTRUCTION MANUAL Thank you for purchasing our company Air Flow Anemometer. This manual provides relative information on how to use the Air Anemometer and warning in operation Please

More information

PRODUCT SHEET. Order probe only as RXPROBE02

PRODUCT SHEET. Order probe only as RXPROBE02 SS69L DISSOLVED OXYGEN PROBE TRANSDUCER Order probe only as RXPROBE02 Order interface only as BSL-TCI16 SS69L Components The SS69L transducer measures dissolved oxygen. The SS69L includes a dissolved oxygen

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

NGP-250/500 Nitrogen Generator Quick Start Guide

NGP-250/500 Nitrogen Generator Quick Start Guide NGP-250/500 Nitrogen Generator Quick Start Guide Version: A July 2013 Potter Electric Signal Company, LLC 5757 Phantom Dr., Suite 125 P. O. Box 42037 Hazelwood, MO 63042 Phone: (314) 595-6900 Document

More information

Cover Page for Lab Report Group Portion. Pump Performance

Cover Page for Lab Report Group Portion. Pump Performance Cover Page for Lab Report Group Portion Pump Performance Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 March 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:

More information

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars. Simply a question of better measurement SCHMIDT Flow Sensor SS 20.261 The cost-effective alternative in pressurised systems up to 10 bars. Compressed air technology Industrial processes A cost analysis

More information

Tex-616-J, Construction Fabrics

Tex-616-J, Construction Fabrics Contents: Section 1 Overview... Section s...3 Section 3 Archived Versions...10 Texas Department of Transportation 1 06/00 1/07 Section 1 Overview Effective dates: June 000 December 007. Section 1 Overview

More information

Exercise 4-2. Centrifugal Pumps EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Pumps

Exercise 4-2. Centrifugal Pumps EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Pumps Exercise 4-2 Centrifugal Pumps EXERCISE OBJECTIVE Familiarize yourself with the basics of liquid pumps, specifically with the basics of centrifugal pumps. DISCUSSION OUTLINE The Discussion of this exercise

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual ARC12M Sputter Copyright 11.2015 by Hong Kong University of Science & Technology. All rights reserved. Page 1 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness

More information

Gas Conversion Kit Pilot Ignition Heaters

Gas Conversion Kit Pilot Ignition Heaters INSTALLATION INSTRUCTIONS Gas Conversion Kit Pilot Ignition Heaters View these instructions online at www.lbwhite.com Kit Contents: DESCRIPTION QTY. Instructions 1 Burner Orifi ce 1 Pilot Orifi ce 1 Variable

More information

DTG - LCD Digital Temperature Gauge USER MANUAL

DTG - LCD Digital Temperature Gauge USER MANUAL DTG - LCD USER MANUAL Page 1 of 7 USERS GUIDE page 1. Description 3 2. Function 3 3. Safety Instruction 3 3.1. Safety Conventions 3 3.2. Proper Use 3 4. Installation 4 4.1. Unpacking 4 4.2. Storage 4 4.3.

More information

Installation, Operation, & Maintenance Manual. MODEL: Watlow Controller P53 Direct, P53 Remote

Installation, Operation, & Maintenance Manual. MODEL: Watlow Controller P53 Direct, P53 Remote Installation, Operation, & Maintenance Manual Mustang Sample Conditioning System MODEL: Watlow Controller P53 Direct, P53 Remote Conforms to UL STDS 61010-1 & 1203 Certified to CSA STD C22.2 Nos. 61010-1

More information

Oxidation Stability of Gasoline and Aviation Fuels

Oxidation Stability of Gasoline and Aviation Fuels Oxidation Stability of Gasoline (Induction Period Method) Oxidation Stability of Aviation Fuels (Potential Residue Method) Test Method Provides an indication of the tendency of gasoline and aviation fuels

More information

Mass Flow Meter (MFM) for Gases

Mass Flow Meter (MFM) for Gases 873 Mass Flow Meter (MFM) for Gases Direct flow measurement by MEMS- Technology for nominal flow rates from 1 ml N /min to 8 l N /min (N 2 ) High accuracy Short response time Compact design and digital

More information

STS PECVD Instructions

STS PECVD Instructions STS PECVD Instructions I. Introduction A PECVD (Plasma Enhanced Chemical Vapor Deposition) reacts gases in a RF- (Radio Frequency) - induced plasma to deposit materials such as SiO 2 and Si X N Y. This

More information

AMT-Ex Dewpoint Transmitter

AMT-Ex Dewpoint Transmitter AMT-Ex Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk

More information

Experiment 11: The Ideal Gas Law

Experiment 11: The Ideal Gas Law Experiment 11: The Ideal Gas Law The behavior of an ideal gas is described by its equation of state, PV = nrt. You will look at two special cases of this. Part 1: Determination of Absolute Zero. You will

More information

Unaxis PECVD. SiH4 (5% in He)

Unaxis PECVD. SiH4 (5% in He) Unaxis PECVD Table of Contents: I: Introduction II: Machine Specifications III: System Components IV: Deposited Materials and Precursor Gases V: Operating Instructions VI: Creating a Recipe VII: Troubleshooting

More information