MARK III Manometric Perfusion Pump Ve rtical Deck

Size: px
Start display at page:

Download "MARK III Manometric Perfusion Pump Ve rtical Deck"

Transcription

1 MARK III Manometric Perfusion Pump Ve rtical Deck Dentsleeve International Ltd. Manufactured by Mui Scientific 45 Traders Blvd. East, Unit Mississauga, Ontario, CANADA L4Z 3L3 Tel: (905) Fax: (905) Toll Free: (866) Website: Handbook September, 006

2 Index In Section In Section A A - A - A - 3 Section B Section C C - C - C - 3 C - 4 C - 5 C - 6 C - 7 C - 8 C - 9 C - 0 C - C - C - 3 Section D D - D - D - 3 Section E E - E - E - 3 E - 4 E - 5 E - 6 E - 7 Index Overview Schematic of pump Major pump components Air Water System Precautions & Warnings Normal Use Assembly & filling of perfusate reservoir Check, connect & set air supply Connection & flushing of water manifold Check flow values of each hydraulic resistor Turn on water perfusion to fill each transducer Transducer calibration (if required) Connect catheter to transducers Water injection procedure - each channel Check flow value for air perfusion manometry Observation & refilling of perfusate reservoir Maintain correct perfusate reservoir pressure Prevention, recognition and correction of hydraulic resistor blockage Compressor Steps on completion of measurements Perfusate reservoir Water manifold Air supply Set Up Procedures & Connection of Components Air supply Perfusate reservoir prior to first use Installation of pressure transducers Water perfusion manifold removal Standard hydraulic resistors Compact resistors Air perfusion circuit

3 In Index Section F F - F - F - 3 F - 4 F - 5 Section G G - G - G - 3 G - 4 Section H H - H - Section I I - I - I - 3 I - 4 Cleaning & Disinfection Cleaning of plastic parts Sterilization of perfusate reservoir Sterilization of manifolds Hydraulic resistors Cleaning of compressor Regular Maintenance Every 3 months (or as needed): perfusate water filter Every year (or as needed): gas filters Service of control box by Dentsleeve Compressor dessicant Problem Solving Diagnosis of abnormal air consumption Abnormal air consumption - perfusate reservoir leakage Specifications, Support, Spart Parts & Accessories Specifications Technical Support Spare parts Regulatory information Notes on the use of this manual References within the manual are shown in brackets eg (C 4.) = section C, part 4, instruction Part numbers (#) given in the text are unique for Dentsleeve The technical information and illustrations in this manual reflect specifications and operating procedures at the time of drafting. Some specifications and operating procedures differ from earlier manuals for Mark III perfusion pump models. Dentsleeve reserves the right to vary specifications and operating procedures as part of its continuous product improvement process. 3

4 Overview A Schematic of pump A 00 kpa or 5 psi Regulated air supply to perfusate reservoir Water perfusate Air supply Pump control box Manifold purge clamp Manifold side-arm pinch valves 3 Way stop cocks Pressure transducers Catheter connector tubes Catheter Note: Schematic diagram of water perfusion circuit. Only critical components are shown. 4

5 A Overview A Major pump components Reservoir pressurization line Perfusate reservoir Pump control box Reservoir outflow control device Pressure transducer rack Air inlet Air perfusion line Standard hydraulic resistors Manifold flush valve Calibration bowl Main perfusate manifold Pinch valves Note: Pump length, channel numbers and spacings vary according to individual specifications. Standard deck version shown. Transducer types vary: Abbott Transpac transducers shown. 5

6 Overview A A 3 MKIII Pump Air Water System #ZP-GAU/TR/ #ZPFIT/SW/SS-4CPA--3 #ZP-SEA/OR/5 Mini Female Quick Connect #ZPFIT/SW/SS-QM-B-00 6

7 B Precautions & Warnings! B!! Never drop. Do not use if visible damage. Only pressurise reservoir with this line (C 3.).! Water level never less than cm (C 0).! Air pressures never more than 300kpa / 45psi! Air perfusion channel. Only use for pharyngeal swallow monitoring (C 9). 7

8 Normal Use C Assembly & filling of perfusate reservoir C Note: See E for Dentsleeve part # s Fill with particle free, degassed, distilled H 0 Add water, then place perfusion float barrier on top Blow-off valve Perfusate outlet hole O ring Reservoir outflow control device! Always use diffusion float barrier! Always use water filter Part #: APFIL/WA/ 3 Tighten firmly to seal May 003 8

9 C Normal Use For set-up and first use see E to E 7 C Check, connect & set air supply Click! Air Air Set to 00 kpa / 5 psi To gas supply 9 May 003

10 Normal Use C Connection & flushing of water manifold C 3 Connect to water manifold Part # ZPFIT/SW/SS-QM-B-B-00 Air line from control box Filter 3 Open 4 Open, vent and close May 003 0

11 C Normal Use C 4 Check flow values of each hydraulic resistor Standard resistor shown : flow is colour coded Flow ml/min Tube Colour Screw Colour Part # Alternative: Compact hydraulic resistor 0.0 Black White R0HRE/ST/(3) 0.0 Black Black R0HRE/ST/(3) 0.04 Black Red R0HRE/ST/4(3) 0.08 Red White R0HRE/ST/8(3) 0.5 Red Black R0HRE/ST/5(3) 0.3 Green White R0HRE/ST/30(3) 0.45 Green Red R0HRE/ST/45(3) 0.6 Green Black R0HRE/ST/60(3) Water flow on case The flow value is... Correct, go to Changed, go to C 5 E 5 or 6 May 003

12 Normal Use C Turn on water perfusion to fill each transducer C 5 Open pinch valve or clamp Fill transducers Fill transducers Prefered position for 3 way stopcock - other than PVB DPT-600 Note: PVB DPT-600 Transducer shown Note: Abbott Transpac Transducer Shown

13 C Normal Use C 6 Transducer calibration (if required) Turn perfusion off - (C 5) 3 or use gravity calibration Set tranducer gain by applying standard external pressure Shows APCMA/TR/ 68cm H O, 50mm Hg Connect to transducers Pressurise with syringe Note: Shows Dentsleeve calibration manifold; made according to channel numbers and spacings Part # s APCMA/TR/to5 Note: Catheter must be attached to transducers (C 7) C 7 Connect catheter to transducers Select correct channel Twist & push on firmly Note: Catheter detail is for Dentsleeve product 3

14 Normal Use C Water injection procedure - each channel C 8 H O - no bubbles Open side-port 3 Inject 5ml of H O 4

15 C Normal Use C 8 Water injection procedure - each channel (continued) 4 Close side-port then remove syringe Note: Arrangement for PVB DPT-600 transducer shown 5

16 Normal Use C Check flow value for air perfusion manometry C 9 Identify air perfusion channel pinch valve! Only use air perfusion for detection of swallowing in pharynx Ensure resistor for air perfusion channel is correct! Air flow rate through hydraulic resistor is x00 water flow rate. Air flow ml/min Part # Black Black R0HRE/ST/(3) 4 Black Red R0HRE/ST/4(3) 8 Red White R0HRE/ST/8(3)! Must be installed between manifold and transducer to limit air flow into manometric catheter to less than 0ml/min May 003 Suitable compact resistors give airflow on case Air flow ml/min Water Flow Part # ml/min 0.0 R0HRE/CO/ R0HRE/CO/ R0HRE/CO/8 6

17 C Normal Use C 0 Observation & refilling of perfusate reservoir! If perfusate exhausted, large volumes of gas may be perfused down catheter Refill when at cm Release pressure Pull Back 3 Refill 4 Restart perfusion Particle free, degassed, distilled H 0 Tighten Repressurise 7

18 Normal Use C Maintain correct perfusate reservoir pressure C Monitor air pressure. Maintain at 00kPa/5psi kpa psi 00 DENTSLEEVE INT L Mfd. by Mui Scientific Check if air pressure drops 3 Check that air supply is adequate 00 At least 30 kpa/ 0psi kpa psi 400 DENTSLEEVE INT L Mfd. by Mui Scientific OR open At least 000 kpa/ 50psi 8

19 C Normal Use C Prevention, recognition and correction of hydraulic resistor blockage Prevention of blockages - Always use perfusate filter Minimise perfusate particle load Double-distilled or de-ionised water No particles in reservoir (E-) Perfusate filter Part # APFIL/WA/ 3 Replace water perfusate filter regularly - (G-) 4 Check perfusate flow rates - assemble ml syringe barrel on transducer ml syringe 5 Start perfusion - (C -) Female/Female adapter Part # ZPFIT/VP/FTLLC-9 Transducer on rack 6 Measure time to deliver water Note: for resistor flow rates < 0.5ml/min use ml syringe barrels 9

20 Normal Use C Prevention, recognition and correction of hydraulic resistor blockage C 7 Flushing of blocked resistors - remove resistor from circuit Transducer end 8 Backflush forcibly 9 Observe water flow 0 Re-install - (E 5) OR Remove resistor from circuit Insert flush support tool - (E 6) 3 Same steps as C.7 to.0 above 0

21 C Normal Use C 3 Compressor Always ground Turn on power switch. Power is on when switch is illuminated 3 Connect compressor air outlet to pump 4 Plug in compressor to pump control

22 Steps On Completion of Measurements D Perfusate reservoir D Depressurise 3 Remove lid Close and disconnect from manifold 4 Drain Reservoir 5 Remove float & air dry Lint & particle free surface! Do not transfer particles to internal surfaces of reservoir

23 D Steps On Completion of Measurements D Water manifold Close D 3 Air supply Turn off compressor OR Close cylinder valve 3

24 Set Up Procedures & Connection of Components E Air supply E! Installation only by an approved, qualified biomedical engineer Use air Note: Air chosen as:. Air dissolution in perfusate prevented by perfusate float barrier.. More suitable for gas perfusion manometry than N. 3. Usually more available and cheaper than N. 4. Available from wall supply and simple compressors. Set up for compressor, wall or air bottle supply Compressor Wall Air bottle E.3 E.4 E.6 or.9 4

25 E Set Up Procedures & Connection of Components E Air supply (continued) 3 Install next to pump deck or place on bottom shelf of trolley OR 4 Identify air line Black Part # ZPTUB/NS/ Part # ZPFIT/SW/SS Connects to pump Air inlet (C.) Part # ZPFIT/SW/B-QC4-B-400ZN 5 Ensure air source is suitable: kpa or 0-35 psi no oil or moisture filtered to 0.5 micron 6 Make appropriate connection to supply 5

26 Set Up Procedures & Connection of Components E Air supply (continued) E 7 Identify air line with Dentsleeve supplied regulator Factory set by Dentsleeve to deliver 300 kpa / 45 psi Part # ZPREG/HP/ Connect to pump air inlet (C.) Cylinder pressure gauge Medical international air pin index (Size C cylinders only) 8 Identify air line if no high pressure regulator supplied Black! 9 Connect to suitable locally sourced high pressure regulator with lockable outlet pressure control! 0 Adjust regulator to deliver outlet pressure of 300 kpa / 45 psi 6

27 E Set Up Procedures & Connection of Components E Perfusate reservoir prior to first use Check correct assembly of perfusate reservoir Blow-off valve Part # ZPFIT/SW/SS-4CPA-3 Closure Knob Part # FAS/PR/ O ring Part # ZPSEA/OR/ Reservoir outflow control device Gasket under lid Part # ZPSEA/GA/ Part # ZPFIT/SW/SS-QM-S-PM Always use water filter (C.) Part # APFIL/WA/!! Always use diffusion float barrier Part # FLO/PR/ Gasket in channel Part # ZPSEA/GA/ Outlet O ring Part # ZPSEA/OR/ Flush reservoir with water to remove any particles 7

28 Set Up Procedures & Connection of Components E Installation of pressure transducers E 3 PVB DPT-600 Abbott Transpac Mount bar removable 8

29 E Set Up Procedures & Connection of Components E 3 Installation of pressure transducers (continued) 3 4 9

30 Set Up Procedures & Connection of Components E Installation of pressure transducers (continued) E

31 E Set Up Procedures & Connection of Components E 4 Water perfusion manifold removal Water perfusion manifold! Air perfusion (C 9 and E 7) 3

32 Set Up Procedures & Connection of Components E Water perfusion manifold removal (continued) E 4 Pinch valve 3 3

33 E Set Up Procedures & Connection of Components E 5 Standard hydraulic resistors Identify standard resistor with correct flow value. (C 4) Prime resistor with water. Use ml syringe for ml/min resistors 0.5ml syringe for lower flow rates Particle-free water 3 Inject till water flows from other end 4 Fill resistor connector as syringe withdrawn No bubbles trapped in luer 33

34 Set Up Procedures & Connection of Components E Standard hydraulic resistors (continued) E 5 5 Connect resistor to manifold side arm luer Ensure resistor connector remains water filled Manifold clamp Part # ZPFAS/RI/ 6 Push and rotate resistor firmly onto manifold luer 7 Open 34

35 E Set Up Procedures & Connection of Components E 5 Standard hydraulic resistors (continued) 8 Debubble downstream resistor luer Inject particle-free water into bottom of luer 9 Ensure no bubbles are trapped 0 Connect to pressure transducer Preferred position for 3 way stopcock Use special high pressure stopcock Push and rotate resistor firmly onto 3 way stopcock Abbott transpac shown Part # ZPSTO/HP/ Water filled Note PVB DPT-600 transducer has integral stop-cock at transducer outlet 3 Final setup: Follow instructions from (C 7.) 35

36 Set Up Procedures & Connection of Components E Compact resistors E 6 Check flow values for each hydraulic resistor Prime resistor with water - Use flush tool 3 Place flush tool on silicone rubber connector 4 Use ml syringe for ml/min resistors syringe for lower flow rates 5 Inject until water flows from other end 6 Fill resistor connector as syringe withdrawn 7 Remove flush tool 36

37 E Set Up Procedures & Connection of Components E 6 Compact resistors (continued) 8 Connect as shown DENTSLEEVE 0.60mL/min Water Only #HRE/CO/60 To water manifold Note: Connectors are self-debubbling. PVB DPT-600 transducer shown. 37

38 Set Up Procedures & Connection of Components E Compact resistors (continued) E 6 9 Connect resistor to manifold sidearm DENTSLEEVE 0.60mL/min Water Only #HRE/CO/60 Manifold clamp Part # ZPFAS/RI/ 0 Open Final set-up - follow from (C 5.). 38

39 E Set Up Procedures & Connection of Components E 7 Air perfusion circuit! N perfusion into pharynx may dilute inspired oxygen - use only air for perfusate reservoir pressurisation in small children, as this gas is also used for gas perfusion circuit Identify circuit Locate air manifold and check correctly connected 3 Resistor must always be installed To transducer 39

40 Set Up Procedures & Connection of Components E Air perfusion circuit (continued) E 7 4 Select resistor that controls airflow to less than 0ml/min! 5 Airflow is x00 waterflow Air flow Waterflow ml/min ml/min Part # 0.0 Black Black R0HRE/ST/(3) Black Red R0HRE/ST/4(3) R0HRE/ST/8(3) Red White Appropriate compact resistors show airflow on case DENTSLEEVE Air ml/min WATER 0.0mL/min HRE/CO/ DENTSLEEVE Air 4mL/min WATER 0.04mL/min HRE/CO/4 DENTSLEEVE Air 4mL/min WATER 0.04mL/min HRE/CO/4 Air = ml/min Air = 4ml/min Air = 8ml/min 6 Install air flow resistor between air manifold luer and transducer No 3 way stopcock 7 No water prime required! Air perfusion to be used only for pharyngeal manometry To manifold luer 40

41 Air 4mL/min WATER 0.04mL/min F Cleaning & Disinfection Cleaning of plastic parts F Clean with cloth and mild detergent Sterilization of perfusate reservoir F Gas sterilize only Sterilization of manifolds F 3 Water manifold Air manifold Autoclavable Hydraulic resistors F 4 DENTSLEEVE HRE/CO/4! Not autoclavable Autoclavable Cleaning of compressor F 5 Wipe with a dry cloth 4

42 Regular Maintenance G G Every 3 months (or as needed): perfusate water filter Replace Part # APFIL/WA/ 4

43 G Regular Maintenance Every year or as needed: gas filters G! To be done only by an approved, qualified biomedical engineer Disconnect gas supply then open control box (G 3.) Use correct filter Part # APFIL/GA/ Gas (Black) 3 Replace filter 4 Check for air-tightness Note: Gas filter life depends on quality of air always use medical grade 43

44 Regular Maintenance G G 3 Service of control box by Dentsleeve Undo screw that holds control box Send control box to Dentsleeve 44

45 G Regular Maintenance Compressor Dessicant G 4 Check dessicant is orange. When it becomes green, replace. Orange GREEN OR Spread granules evenly onto tray, one granule deep. Heat granules for approximately 5 hours. (or until it turns back to its original orange color) at 5 o C (50 o F) in a conventional oven. Cool dessicant before replacing back into canister. 45

46 Problem Solving H Diagnosis of abnormal air consumption H Check compressor connection For compressor malfunction, please contact Dentsleeve OR Air Check regulator/cylinder connection Not connected Tighten Pressurise air circuit - Turn on Air Adjust air pressure 4 Turn off Air kpa psi 00 DENTSLEEVE INT L Mfd. by Mui Scientific 46

47 H Problem Solving H Diagnosis of abnormal air consumption (continued) 5 Recognise leakage - Wait hour either or 6 Drop less than 33 kpa/5psi No significant leak 7 Drop more than 33 kpa/5psi Significant leak 8 Follow (H-.9) below 47

48 Problem Solving H Diagnosis of abnormal air consumption (continued) H 9 Check for leakage outside control box If there are bubbles, replace fitting - Part # ZPFIT/SW/SS-QM-B-00 kpa psi 00 DENTSLEEVE INT L Mfd. by Mui Scientific! Consult approved, qualified biomedical engineer to do check (H-. to H-.7) below Air Leakage? 3 Open pump control box. 4 Remove control box lid/back 48

49 H Problem Solving H H Diagnosis of abnormal air consumption (continued) 5 Filter leakage - tighten or replace 6 Check rest of air circuit 49

50 Problem Solving H Abnormal air consumption - perfusate reservoir leakage H First exclude air circuit leakage - (H ) Set pressure kpa psi 00 DENTSLEEVE INT L Mfd. by Mui Scientific Tighten 3 Set up reservoir. Remove blow-off valve cap 4 No water 5 Pressurise 6 Locate site of leakage - observe underwater Still pressurised 7 Locate site of bubbling 50

51 H Problem Solving H H Abnormal Air consumption - perfusate reservoir leakage (cont.) ZPFIT/SW/B-QM-B-00ZN 8 If leak elsewhere, dismantle reservoir 9 Identify cause of leakage at base 0 Gasket clean and intact? Identify cause of leakage at top Underside Flip Check O ring Part # ZPSEA/OR/ Gasket underneath clean and intact? Part # ZPSEA/GA/ 5

52 Specifications, Support, Spare Parts & Accessories I Specifications I Air High pressure regulators Recommended first stage regulators (*if supplied) Medical grade only CONCOA Medical Air (Air pin index) Inlet pressure Pump inlet pressure (from gas supply first stage regulator) Air kPa (5-45 psi) Compressor 30-40kpa (0-35 psi) Supply lines (from gas bottles) Air (black) line Air Male Inlet Air female connector to perfusate reservoir Air supply on pump Driving air supply (pump) Air flow rate ± 0% Adjustable 0-50 kpa (0-36psi) Flow restricted to 30ml/min at 00kPa (5psi) Pressure relief valves Perfusate reservoir relief valve Control box inlet overpressure relief valve Preset to 00 kpa (9psi) Preset to 300 kpa (45psi) Filtration Air 0.5 micron male/female luer connection disposable disc filter Contact Dentsleeve for advice See contact details on front cover Technical Support I 5

53 I Specifications, Support, Spare Parts & Accessories H Spare Parts I 3 Spare Parts Kit x 5x x APFIL/GA/ Gas filters, for installation within pump control box. APFIL/WA/ Water filters, for filtration of water perfusate ZP-OCD/PR/ Reservoir outflow control device, for perfusate outflow. Regulatory Information I 4 EC REP Authorised European Representative SOLAL Rue du Travail Strasbourg FRANCE Tel: Fax: FDA Approved CE Marked Health Canada Licensed 53

54

Portable Pressurized Infusion Pump. and. Air Compressor. Operating and Service Manual

Portable Pressurized Infusion Pump. and. Air Compressor. Operating and Service Manual Portable Pressurized Infusion Pump and Air Compressor Operating and Service Manual Model # s: PIP-6-4SS PIP-6-6SS PIP-6-8SS PIP-6-10SS PIP-6-12SS PIP-6-14SS PIP-6-16SS PIP-6-18SS PIP-6-20SS PIP-6-22SS

More information

Manual Solvent Purification System. Innovative Technology, Inc.

Manual Solvent Purification System. Innovative Technology, Inc. Innovative Technology, Inc. 2 New Pasture Road, Newburyport, MA 01950-4054 USA Email: Info@gloveboxes.com TEL: +1 978 462 4415 FAX: +1 978 462 3338 Manual Solvent Purification System Copyright Innovative

More information

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings SomnoSuite FAQ V1.3 January 2015 Setup 1. How do I connect the SomnoSuite to my oxygen source? 2. Is there a way to speed up the downward movement of the pusher block when setting the empty position? 3.

More information

HANDBOOK. Squeeze Off Unit SOU 400. Please refer any queries to:

HANDBOOK. Squeeze Off Unit SOU 400. Please refer any queries to: HANDBOOK Squeeze Off Unit SOU 400 Please refer any queries to: Hy Ram Engineering Co Ltd Pelham Street Mansfield Nottinghamshire NG18 2EY Telephone No: (01623) 422982 Please note all queries should state

More information

InFlow Modular perfusion bioreactor

InFlow Modular perfusion bioreactor InFlow is an incubator compatible, modular bioreactor for the controlled, bidirectional and interstitial perfusion of up to 9 cell seeded scaffolds with culture medium. InFlow is ideal for general purpose

More information

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM IMI CORNELIUS INC One Cornelius Place Anoka, MN 55303-623 Telephone (800) 238-3600 Facsimile (612) 22-326 INSTALLATION INSTRUCTIONS FOR INSTALLING CO 2 BLENDER KIT (P/N 111612000 IN BEER SYSTEM SECONDARY

More information

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual BioAerosol Nebulizing Generator Operation and Maintenance User Manual INTRODUCTION The BANG or BioAerosol Nebulizing Generator is a unique nebulizer for the generation of aqueous aerosols at a low air

More information

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50 Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y OFFICE: (507) 843-4700 P.O. BOX 7079 Toll

More information

MF2M Monnier Miniature Compressed Air Filter

MF2M Monnier Miniature Compressed Air Filter 5040050 / 2 IM-P504-02 CH Issue 2 MF2M Monnier Miniature Compressed Air Filter Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation and commissioning

More information

Code AWC20HP Air Compressor

Code AWC20HP Air Compressor Code 951816 AWC20HP Air Compressor Index of Contents Index of Contents 02 Declaration of Conformity 02 What s Included 03 Safety Precautions 03 Specifications (AWC20HP Air Compressor) 04 Assembly Instructions

More information

Helium-Oxygen Blender

Helium-Oxygen Blender Helium-Oxygen Blender Service Manual Model No. PM5400 Series PM5500 Series (shown) SAVE THESE INSTRUCTIONS 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 www.precisionmedical.com

More information

Design of an Automated Perfusion Pump James J Higham

Design of an Automated Perfusion Pump James J Higham Design of an Automated Perfusion Pump James J Higham Abstract Department of Gastroenterology Repatriation General Hospital The design and construction of an automated perfusion pump commenced as part of

More information

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

MEDICAL REGULATOR RANGE (MEDIREG RANGE) MEDICAL REGULATOR RANGE (MEDIREG RANGE) Medireg Dialreg Flowreg INSPECTION Remove the Medireg from the packaging and inspect for damage. If there is any damage, DO NOT USE, and contact Therapy Equipment

More information

MANUAL GAS MANIFOLDS

MANUAL GAS MANIFOLDS MANUAL GAS MANIFOLDS Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com

More information

AUTOMATIC GAS MANIFOLDS

AUTOMATIC GAS MANIFOLDS AUTOMATIC GAS MANIFOLDS Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com

More information

User s Guide. Vacuum Controller for liquid delivery systems

User s Guide. Vacuum Controller for liquid delivery systems Flow Control User s Guide Vacuum Controller for liquid delivery systems Precise Vacuum Control throughout the experiment Flow control Compatible with any perfusion system Ideal for Small Volume Delivery

More information

SERIES-O REGULATOR. Instructions for use

SERIES-O REGULATOR. Instructions for use SERIES-O REGULATOR Instructions for use 522460-05-708 Contents Page 1. INTRODUCTION 2 2. INTENDED USES 3 3. GENERAL DESCRIPTION 4 4. PARTS IDENTIFICATION 5 5. SAFETY PRECAUTIONS 6 6. EQUIPMENT USE 8 7.

More information

OXY-VIVA 3 RESUSCITATOR OPERATING INSTRUCTIONS

OXY-VIVA 3 RESUSCITATOR OPERATING INSTRUCTIONS OXY-VIVA 3 RESUSCITATOR OPERATING INSTRUCTIONS COMWELD MEDICAL TABLE OF CONTENTS 1.0 INTRODUCTION 3 2.0 GENERAL DESCRIPTION 3 3.0 SAFETY PRECAUTIONS 3 4.0 PARTS IDENTIFICATION 4 5.0 OPERATIONAL CHECK-OUT

More information

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200 Operators Manual Supera Floor Stand Anesthesia Machine Models: M1000 Basic Non-rebreathing M1200 Basic M1000 M1200 (All shown with optional equipment) Copyright Supera, LLC 2017 Examination and Preparation

More information

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , & REDQ Regulators Model 411B Barstock Design Powreactor Dome Regulator OPERATION AND MAINTENANCE Contents Scope..............................1 Installation..........................1 General Description....................1

More information

Budget Range Operators Handbook

Budget Range Operators Handbook Budget Range Operators Handbook BAMBI AIR COMPRESSORS LTD 152 Thimble Mill Lane Heartlands Birmingham B7 5HT United Kingdom Tel: 0121 322 2299 Fax: 0121 322 2297 Email: sales@bambi-air.co.uk www.bambi-air.co.uk

More information

Product Sheet Spraymist Unit

Product Sheet Spraymist Unit Product Sheet Spraymist Unit DESCRIPTION The Spraymist unit is perfectly adapted for use with differents liquids for applications as metalcutting, drilling, grinding, cooling or wetting processes and chain

More information

Instruction Manual Updated 7/26/2011 Ver. 2.2

Instruction Manual Updated 7/26/2011 Ver. 2.2 4-Unit Model MB HTHP Filter Press #171-50-4: 115-Volt #171-51-4: 230-Volt Instruction Manual Updated 7/26/2011 Ver. 2.2 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. Houston, Texas 77065 U.S.A. Tele:

More information

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator Operating Instructions Table of Contents Resource v.2 Conserving Regulator Safety Information Device Precautions Introduction Product Features Product Specifications Feature Illustrations Set Up Usage

More information

Dialflow Regulator. Instructions for Use

Dialflow Regulator. Instructions for Use Dialflow Regulator Instructions for Use 702-0030.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP LIMITED WARRANTY Getz Equipment Innovators warrants its products, and component parts of any product manufactured by

More information

P5523. Users Manual. Liquid to Gas Separator. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

P5523. Users Manual. Liquid to Gas Separator. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot. Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com P5523 Liquid to Gas Separator Users Manual PN 3952355 November 2010 2010 Fluke Corporation. All rights

More information

Beamex PG PRESSURE GENERATORS. Beamex PGM PGV PGC PGHH PGPH PGL

Beamex PG PRESSURE GENERATORS. Beamex PGM PGV PGC PGHH PGPH PGL Beamex PG PRESSURE GENERATORS 97 Beamex PGM PGV PGC PGHH PGPH PGL Beamex PG pressure generators are portable hand-operated pressure generators and ultimate field calibration pumps 98 Calibration pumps

More information

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels 12.09 product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels Introduction: This manual covers the installation and operation of Humboldt FlexPanels for Triaxial and Permeability

More information

Instructions for Use

Instructions for Use Select-380 T-Auto T-Auto Contents Page Instructions for Use Ref: 3.0 IFU 380 T-Auto Mar 18 2 Schematic layout of the doser 3 Quick-fit instructions 4 Description/Installation/Operation Pump tubes & Water

More information

Pressure Automated Calibration Equipment

Pressure Automated Calibration Equipment GE Measurement & control Pressure Automated Calibration Equipment Safety Instructions and User Guide - K0447 PACE5000 PACE6000 K0447 Issue No. 9 1 10 1 PACE5000 1 2 3 4 5 PACE6000 2 6 7 8 3 4 5 6 7 8 9

More information

HANDBOOK. Squeeze Off Unit SOU 250. Please refer any queries to:

HANDBOOK. Squeeze Off Unit SOU 250. Please refer any queries to: HANDBOOK Squeeze Off Unit SOU 250 Please refer any queries to: Hy Ram Engineering Co Ltd Pelham Street Mansfield Nottinghamshire NG18 2EY Telephone No: (01623) 422982 Please note all queries should state

More information

Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators

Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators Genstar Technologies Co., Inc. 4525 Edison Avenue Chino, CA 91710 USA TEL 909-0-272 FAX 909-0-485

More information

WELLS JOHNSON HIGH VOLUME CANISTERS

WELLS JOHNSON HIGH VOLUME CANISTERS WELLS JOHNSON HIGH VOLUME CANISTERS USER MANUAL DISASSEMBLY OF HARVESTING CANISTERS Please follow the instructions provided to disassemble 5L, 3L, 2L, and 1L, 250mL and 500mL harvesting canisters. 1. Unscrew

More information

Instructions for use Compressed air injector with water trap

Instructions for use Compressed air injector with water trap Instructions for use Compressed air injector with water trap Table of Contents Introduction 1. Description of Functions 2. Putting into Operation 3. Instructions for Cleaning and Care 3.1 Disassembling

More information

Pressure Dump Valve Service Kit for Series 3000 Units

Pressure Dump Valve Service Kit for Series 3000 Units Instruction Sheet Pressure Dump Valve Service Kit for Series 000 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00, 400, 500, and 700 applicator

More information

1020 Industrial Drive, Orlinda, TN fax

1020 Industrial Drive, Orlinda, TN fax Operation Manual Ultrafiltration for High Purity Distribution K-A-HPTUF Series 615-654-4441 sales@specialtyh2o.com 615-654-4449 fax TABLE OF CONTENTS Section 1 GENERAL 1.1 Warnings and Cautions... 1 1.2

More information

Pressure Dump Valve Service Kit for Series 2300 Units

Pressure Dump Valve Service Kit for Series 2300 Units Instruction Sheet Pressure Dump Valve Service Kit for Series 00 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00 applicator tanks when the unit

More information

RD(H)20/25 Pressure-Reducing Regulator User Manual

RD(H)20/25 Pressure-Reducing Regulator User Manual RD(H)20/25 Pressure-Reducing Regulator User Manual Read the complete manual before installing and using the regulator. 2 Safe Product Selection When selecting a product, the total system design must be

More information

Assembly-, installation- and maintenance instruction manual for bladder accumulators IBV / EBV , top reparable

Assembly-, installation- and maintenance instruction manual for bladder accumulators IBV / EBV , top reparable Page 1 von 8 Assembly-, installation- and maintenance instruction manual for bladder accumulators IBV / EBV 100-575, top reparable Content Seite 0 Legend 2 1 Overview 2 2 Accumulator assembly and installation

More information

SPECIFICATIONS TABLE Maximum Air Pressure 115 PSI Air Tank Capacity 20 Gallons

SPECIFICATIONS TABLE Maximum Air Pressure 115 PSI Air Tank Capacity 20 Gallons SPECIFICATIONS TABLE Maximum Air Pressure 115 PSI Air Tank Capacity 20 Gallons Air Flow Capacity 6.2 CFM at 40 PSI 5.2 CFM at 90 PSI Motor 2.5 HP Working / 4 HP Peak 120 Volt / 60 Hz / 13 A / 1-Phase Required

More information

Hi-Force Limited Prospect Way Daventry Northants NN11 8PL United Kingdom Tel: +44(0) : Fax: +44(0) : Website:

Hi-Force Limited Prospect Way Daventry Northants NN11 8PL United Kingdom Tel: +44(0) : Fax: +44(0) : Website: 1.0 Inspection of the product upon receipt: On receipt of the product, visually inspect the item for any evidence of shipping damage. Please note shipping damage is not covered by warranty. If shipping

More information

LV LOADER DIRECTIONAL CONTROL VALVES INSTALLATION & USER GUIDE

LV LOADER DIRECTIONAL CONTROL VALVES INSTALLATION & USER GUIDE LV LOADER DIRECTIONAL CONTROL VALVES INSTALLATION & USER GUIDE SPECIFICATIONS: 10 gpm (38 lpm) Nominal Capacity. Rated up to 4000 psi (275 bar). Port Sizes-Inlet/Outlet #8SAE (3/4-16). Work Ports #8SAE

More information

System Pressure Manager Standard & System Pressure Manager Plus

System Pressure Manager Standard & System Pressure Manager Plus System Pressure Manager Standard & System Pressure Manager Plus Installation, Commissioning & Servicing Instructions Note: THESE INSTRUCTIONS MUST BE READ AND UNDERSTOOD BEFORE INSTALLING, COMMISSIONING,

More information

INSTRUCTION MANUAL & PARTS

INSTRUCTION MANUAL & PARTS AIR LINE PURIFIERS FOR BREATHING EQUIPMENT AP-C RADEX INSTRUCTION MANUAL & PARTS 01-0-11 introduction CANABLAST BREATHING EQUIPMENT PURIFIER CAN BE USED IN CONBINATION WITH WITH OTHER SAFEGUARDS TO SUPPLY

More information

Tridak Model 1050 User Guide

Tridak Model 1050 User Guide Tridak Model 1050 User Guide Syringe Filling System Instructions for Safe Use Setup and Operation Maintenance Ordering Spare Parts and Accessories 2 Tridak Model 1050 Syringe Filling System User Guide

More information

TJF-Q180V Cleaning and Disinfection Checklist

TJF-Q180V Cleaning and Disinfection Checklist TJF-Q180V Cleaning and Disinfection Checklist TJF-Q180V Cleaning and Disinfection Checklist This checklist is used to evaluate and confirm if cleaning and disinfection of the TJF-Q180V has been performed

More information

SEMEM3-73 Maintaining medical gas pipeline systems and equipment

SEMEM3-73 Maintaining medical gas pipeline systems and equipment Maintaining medical gas pipeline systems and equipment Overview This unit identifies the competences you need to carry out maintenance activities on medical gas pipeline systems and equipment, in accordance

More information

Multi-Probe Sampling System User Guide & Manual

Multi-Probe Sampling System User Guide & Manual Multi-Probe Sampling System User Guide & Manual Rev. 06/05/12 Part #22200012 Table of Contents CHAPTER 1: SYSTEM DESCRIPTION... 2 FUNCTION AND THEORY... 3 SYSTEM COMPONENTS... 4 CHAPTER 2: SYSTEM INSTALLATION...

More information

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC. 4075 East Market Street York, PA 17402 Manual P/N 035-80589-02 Table of Contents Components... 2 Test... 4 Recovery/Recycling... 5 Evacuation... 7 Charging...

More information

NGP-250/500 Nitrogen Generator Quick Start Guide

NGP-250/500 Nitrogen Generator Quick Start Guide NGP-250/500 Nitrogen Generator Quick Start Guide Version: A July 2013 Potter Electric Signal Company, LLC 5757 Phantom Dr., Suite 125 P. O. Box 42037 Hazelwood, MO 63042 Phone: (314) 595-6900 Document

More information

Installation Operation Maintenance. Bermad Level Control Valve with Modulating Horizontal Float Pilot valve One Way Flow IOM.

Installation Operation Maintenance. Bermad Level Control Valve with Modulating Horizontal Float Pilot valve One Way Flow IOM. Bermad Level Control Valve with Modulating Horizontal Float Pilot valve One Way Flow Model: FP 450-80 Installation Operation Maintenance PAGE 1 OF 5 1. Safety First BERMAD believes that the safety of personnel

More information

User's Manual. MixRite TF 10. Edition 05.08

User's Manual. MixRite TF 10. Edition 05.08 User's Manual MixRite TF 10 Edition 05.08 1 Tefen MixRite TF 10 fertilizer and chemicals Injector Congratulations on your purchase of one of Tefen s high quality products. To get the best results from

More information

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y P.O. Box 7079 OFFICE: (507)

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS Tempe Pressure Cell June 1995 Fig. 1a 1400 Tempe Pressure Cells with 3cm cylinders and 6cm cylinders) mounted on the Tempe Cell Stand Fig. 1b Disassembled 1400 Tempe Pressure Cell

More information

Alfred ECMO Circuit Priming and Storage Guide

Alfred ECMO Circuit Priming and Storage Guide Alfred ECMO Circuit Priming and Storage Guide 1 March 2012 Check List Page 2. Assembly, Priming and Storage Page 3. Prepare ECMO trolley and Circuit immediately before use Page 6. Connection to Patient

More information

Installation of Your SprayMaster System

Installation of Your SprayMaster System Installation of Your SprayMaster System 1. At the installation site, remove all equipment from the corrugated box and the polyethylene drum and replace the drum lid. Check the picture to identify each

More information

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1 ANA 74300 Oxygen Regulator with Yoke Connector User Manual Manual No. ANA 74300 001 Issue 1 Safety Thank you for purchasing this Clements Oxygen Regulator. For your safety it is imperative that this unit

More information

DIAGNOSTIC CARDIOLOGY

DIAGNOSTIC CARDIOLOGY DIAGNOSTIC CARDIOLOGY For Your Procedural Needs Manufactured to the Highest Standards, Customised to Your Specific Needs MEDLINE PROCEDURE PACKS The items contained within this brochure are perfect for

More information

NTF-230 OPERATION MANUAL

NTF-230 OPERATION MANUAL NTF-230 OPERATION MANUAL MAHLE Aftermarket Inc., Service Solutions 10 Innovation Drive York, Pennsylvania 17402 USA Phone: 717-840-0678 Toll Free: 800-468-2321 Web-site: www.servicesolutions.mahle.com

More information

VACUUM REGULATORS CONTENTS

VACUUM REGULATORS CONTENTS CAD drawing data catalog is available. ACCESSORIES GENERAL CATALOG AIR TREATMENT, AUXILIARY, VACUUM, AND FLUORORESIN PRODUCTS CONTENTS Small Regulators Features 759 Specifications, Order Codes, Flow Rate

More information

TJF-Q180V Cleaning and Disinfection Checklist

TJF-Q180V Cleaning and Disinfection Checklist TJF-Q180V Cleaning and Disinfection Checklist TJF-Q180V Cleaning and Disinfection Checklist This checklist is used to evaluate and confirm if cleaning and disinfection of the TJF-Q180V has been performed

More information

The NAMIC Dream Kit. Single Operator Control. Reduce Contrast Waste. Less Force to Inject. Compensator * Manifold. How the Compensator Manifold Works:

The NAMIC Dream Kit. Single Operator Control. Reduce Contrast Waste. Less Force to Inject. Compensator * Manifold. How the Compensator Manifold Works: Dream Kit The NAMIC Dream Kit Single Operator Control. Reduce Contrast Waste. Less Force to Inject. For more than 40 years, NAMIC * Fluid Management products have been providing cardiologists, technicians

More information

Inert Air (N2) Systems Manual

Inert Air (N2) Systems Manual INSTRUCTION MANUAL Inert Air (N2) Systems Manual N2-MANUAL 2.10 READ AND UNDERSTAND THIS MANUAL PRIOR TO OPERATING OR SERVICING THIS PRODUCT. GENERAL INFORMATION Positive pressure nitrogen gas pressurizing

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

PRESSURE TEST PUMP OPERATING & MAINTENANCE INSTRUCTIONS. Model No: PTP100. Part No: GC01/09

PRESSURE TEST PUMP OPERATING & MAINTENANCE INSTRUCTIONS. Model No: PTP100. Part No: GC01/09 PRESSURE TEST PUMP Model No: PTP00 Part No: 06020 OPERATING & MAINTENANCE INSTRUCTIONS GC0/09 INTRODUCTION Thank you for purchasing this CLARKE Pressure Testing Pump. Before attempting to use the product,

More information

Installation and operating manual. Pneumatic control station LK product no: PCS 1-10

Installation and operating manual. Pneumatic control station LK product no: PCS 1-10 LK product no: PCS 1-10 Article no: 74503 Revision:8 Article no: 74503 Revision: 8 2 (23) Contents 1. General information... 5 2. Safety precautions... 5 2.1 Significance of symbols... 5 2.2 Explanatory

More information

INSTRUCTION MANUAL. January 23, 2003, Revision 0

INSTRUCTION MANUAL. January 23, 2003, Revision 0 INSTRUCTION MANUAL Model 810A In-Vitro Test Apparatus for 310B Muscle Lever January 23, 2003, Revision 0 Copyright 2003 Aurora Scientific Inc. Aurora Scientific Inc. 360 Industrial Parkway S., Unit 4 Aurora,

More information

PERFORM Operating Document

PERFORM Operating Document PERFORM Operating Document Use and Maintenance of CO 2 Incubator PC-POD-CA-007-v03 Revision History Version Reason for Revision Date 01 New POD 30-Sep-13 02 Minor revisions for section 2.3, 3.1, 4.3. 14-April-16

More information

Binks SG-2 Plus TM Rotary 2 QT. PRESSURE CUP

Binks SG-2 Plus TM Rotary 2 QT. PRESSURE CUP Binks SG-2 Plus TM Rotary 2 QT. PRESSURE CUP Model No. 80-651 Rotary Agitator INTRODUCTION Binks SG-2 TM Plus Rotary 2 Qt. Pressure Cup is ideal for component spraying and industrial applications where

More information

ATD LB PRESSURE BLASTER INSTRUCTION MANUAL

ATD LB PRESSURE BLASTER INSTRUCTION MANUAL ATD-8402 90LB PRESSURE BLASTER INSTRUCTION MANUAL SAVE THESE INSTRUCTIONS SAFETY INSTRUCTIONS FOR SANDBLASTER 1. Before opening the tank release the air pressure on the sand tank. To do this, turn off

More information

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax OPERATION MANUAL Zutron Medical, LLC 17501 W 98 th St #40-27 Lenexa, KS 66219 Phone 877-343-5873 Fax 913-967-5944 ZUT-Lab-004-30004 REV. 03312017 Table of Contents 2 Introduction 1. Intended Use 2. Labels,

More information

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot. Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com P5513 Pneumatic Comparison Test Pump Users Manual PN 3963372 November 2010 2010 Fluke Corporation. All

More information

3:1 High Ratio Oil Pump W. Extn. Kit

3:1 High Ratio Oil Pump W. Extn. Kit 3:1 High Ratio Oil Pump W. Extn. Kit OWNER S MANUAL WARNING: Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions may result

More information

1.060 Set-Up and Priming of the Bypass Circuit. Perfusion Technology Department. Perfusionists

1.060 Set-Up and Priming of the Bypass Circuit. Perfusion Technology Department. Perfusionists TITLE/DESCRIPTION: DEPARTMENT: PERSONNEL: 1.060 Set-Up and Priming of the Bypass Circuit Perfusion Technology Department Perfusionists EFFECTIVE DATE: 8/97 REVISED: 09/07, 04/09, 9/17 PURPOSE: The function

More information

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL www.modsafe.com SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device.

More information

MSC-P and MSC-N Manifolds for Steam Distribution and Condensate Collection

MSC-P and MSC-N Manifolds for Steam Distribution and Condensate Collection 1170850/1 IM-P117-36 ST Issue 1 MSC-P and MSC-N Manifolds for Steam Distribution and Condensate Collection Installation and Maintenance Instructions 1. Safety information 2. General product information

More information

SILENTAIRE TECHNOLOGY

SILENTAIRE TECHNOLOGY SILENTAIRE TECHNOLOGY 8614 Veterans Memorial. Houston, Texas 77088 832/327-7452 800/972-7668 Fax: 832/327-0668 E-mail: silentaire@silentaire.com Thank you and congratulations on the purchase of your OILLESS

More information

instruction manual highpressure handpump manual

instruction manual highpressure handpump manual instruction manual highpressure handpump manual Introduction The FX pump is designed, manufactured and tested by FX Airguns AB in Sweden. There are two patented systems. The first of them is the 3 stage

More information

RIX COMPRESSORS. MICROBOOST Owner s Manual CONFIGURATION A MB-115 MB-230. November 21, 2002 APPLICABILITY [ A]

RIX COMPRESSORS. MICROBOOST Owner s Manual CONFIGURATION A MB-115 MB-230. November 21, 2002 APPLICABILITY [ A] RIX COMPRESSORS MICROBOOST Owner s Manual CONFIGURATION A MB-115 MB-230 November 21, 2002 APPLICABILITY [ A] INTRODUCTION Safety This electromechanical equipment is designed to produce high pressure gas.

More information

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref Rev.

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref Rev. Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA 19440 USA Tel: 215-723-4000 Ref. 10049700 Rev. L Table of Contents Warnings...2 Examination...2 Description and

More information

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the

More information

GREIG FILTERS, INC. OPERATION MANUAL

GREIG FILTERS, INC. OPERATION MANUAL GREIG FILTERS, INC. OPERATION MANUAL MODEL DCFH 3P 15/3 100 S4 DOE/222 DUAL CARTRIDGE FILTER HOUSING TABLE OF CONTENTS I. INTRODUCTION II. CARTRIDGE FILTER HOUSING UNIT A. OPERATION B. INSTALLATION C.

More information

AMS SUPER SILENT COMPRESSOR USER & MAINTENANCE GUIDE

AMS SUPER SILENT COMPRESSOR USER & MAINTENANCE GUIDE 625 Joyce Kilmer Avenue New Brunswick, New Jersey 08901 AMS SUPER SILENT COMPRESSOR USER & MAINTENANCE GUIDE Congratulations! You have purchased the finest specialty air compressor product in the world!

More information

User s Guide. CO2 & O2 Controller for miniature incubators

User s Guide. CO2 & O2 Controller for miniature incubators CO2 O2 Control User s Guide CO2 & O2 Controller for miniature incubators Precise CO2 & O2 Control throughout the experiment Media ph control Compatible with any perfusion system Miniature incubators for

More information

AMP Oil Free Manual AMP 50-8-TC AMP 50-6-D AMP General User and Maintenance Instructions

AMP Oil Free Manual AMP 50-8-TC AMP 50-6-D AMP General User and Maintenance Instructions AMP Oil Free Manual AMP 50-8-TC AMP 50-6-D AMP 50-24 General User and Maintenance Instructions Silentaire Technology 8614 Veterans Memorial Dr. Houston, TX 77088 800-972-7668 Fax 832-327-0669 www.silentaire.com

More information

Operation & Service Manual

Operation & Service Manual Operation & Service Manual Model: 20-4505-7000 Four Bottle Oxygen Cart with Booster CAUTION! It is mandatory that this instruction manual be read and understood by all persons operating this high pressure

More information

Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown)

Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown) Service Manual Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 www.precisionmedical.com

More information

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator. LRS(H)4 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to the

More information

Installation and commissioning instructions 255 series and 256 series

Installation and commissioning instructions 255 series and 256 series Installation and commissioning instructions 255 series and 256 series Table of contents 1 General information... 3 1.1 About these instructions... 3 1.2 About this product... 3 1.3 Appropriate usage...

More information

MoveRoll Conveyor Operating and Maintenance Manual

MoveRoll Conveyor Operating and Maintenance Manual MoveRoll Conveyor Operating and Maintenance Manual 1. Read this first! This manual contains information for protection of personnel in the roll handling area from possible injury and/or equipment damage.

More information

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010 UNITY 2 Uni-DAAMS Manual Version 1.3 (Changes to Section 1.1) March 2010 1. Introduction...2 1.1. Installing the DAAMS tube ovens...2 1.2. Swapping between tube types...5 1.3. Changing tube oven o-rings...6

More information

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage Linkam Scientific Instruments TST350 Temperature Controlled Tensile Stress Testing Stage USER GUIDE Contents Before Setting Up Your Equipment... 3 Important Notice... 4 Warranty... 4 Technical Support...

More information

easycomposites SETUP & DAILY USE INSTRUCTIONS NEVER SWITCH OFF YOUR PUMP WITHOUT FIRST CLOSING THE VALVE TO THE CHAMBER

easycomposites SETUP & DAILY USE INSTRUCTIONS NEVER SWITCH OFF YOUR PUMP WITHOUT FIRST CLOSING THE VALVE TO THE CHAMBER TM easycomposites share the knowledge SETUP & DAILY USE INSTRUCTIONS NEVER SWITCH OFF YOUR PUMP WITHOUT FIRST CLOSING THE VALVE TO THE CHAMBER 26L Vacuum Degassing Chamber 26L Starter Vacuum Degassing

More information

HPB25 Hydraulic Paving Breaker

HPB25 Hydraulic Paving Breaker INSTRUCTION MANUAL HPB25 Hydraulic Paving Breaker Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. Register this product at www.greenlee.com

More information

Training. Testor Training Manual

Training. Testor Training Manual Training Testor Training Manual Index Section 1 Introduction and Safety Warnings Section 2 Test Procedures Section 3 Test Hoses Section 4 Fault Location 1:1 1.1 Introduction The Dräger Testor test equipment

More information

RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL

RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL DESCRIPTION 1. No mar 2. No mar tip 3. Housing grip 4. Trigger 5. Air inlet 6. Air inlet plug 7. Plastic board Important! It is essential that you read the

More information

Lubricated TC Models

Lubricated TC Models Lubricated TC Models P 15 TC P 30 TC P 50 TC+AL General User and Maintenance Instructions Werther International, Inc 8614 Veterans Memorial Houston, TX 77088 832-327-9378 Fax 832-327-0669 Thank you and

More information

Operation and Maintenance of the EPV200 Portable Ventilator

Operation and Maintenance of the EPV200 Portable Ventilator Operation and Maintenance of the EPV200 Portable Ventilator 1 Applications of the EPV200 The EPV200 Portable Ventilator is a gas powered electronically controlled mechanical ventilator, designed to provide

More information

UltRo Dual Flow Reverse Osmosis Water System

UltRo Dual Flow Reverse Osmosis Water System UltRo Dual Flow Reverse Osmosis Water System Congratulations on this great investment to your health. **IMPORTANT NOTE BEFORE YOU BEGIN** We recommend you call your local friendly plumber to ensure proper

More information