Recording established using the VO2 and RER graph template file (h19.gtl)

Size: px
Start display at page:

Download "Recording established using the VO2 and RER graph template file (h19.gtl)"

Transcription

1 Updated BSL PRO Lesson H19: VO 2 and RER Measurement Recording established using the VO2 and RER graph template file (h19.gtl) Overview Oxygen Consumption (VO 2 ) and Respiratory Exchange Ratio (RER) Real-time Oxygen Consumption (VO 2 ) and Respiratory Exchange Ratio (RER) measurements can be obtained using the MP3x System with the Gas-System2 module and some airflow accessories. When performing tests of this kind, many factors exist to confound the measurement. For example, it may be reasonable to think that the volume of expired air is always the same as inspired air, but this is only true when the volume of carbon dioxide expired is equal to the volume of oxygen consumed. The inspired and expired volumes are equal only when the RER equals one. RER is defined as VCO 2 produced divided by VO 2 (where V is volume). Accordingly, when performing these measurements, precise determination of inspired and expired volumes and accurate assessment of the gas concentration level are required. The Gas-System2 measures O 2 and CO 2 concentrations in vapor. When the subject inspires, air will be drawn into the Gas-System2 through the SS11LA airflow transducer. The SS11LA is placed on the inspiration side to eliminate any effects associated with Page 1 of 16

2 expired air humidity (see Hardware Setup for more detail). When the subject expires, air will be directed to the Gas-system2 module. The Gas-System2 is designed to work with saturated expired air. When using a mixing chamber to average O 2 and CO 2 concentrations over several breaths and measuring these changes for arbitrarily high breathing rates, there is no performance degradation. Accordingly, a mixing chamber (such as the Gas-System2 utilizes) is typically recommended for quick, accurate and easy metabolic analysis. IMPORTANT CONCEPT! The non-rebreathing T valve directs only expired air to the Gas-System2. Because only expired air is directed to the module, the system acts to average respiratory outflows. This averaging effect causes the CO 2 and O 2 concentrations to vary in accordance to the mean values resident in a few expired breaths. The size of the system s chamber determines the extent of the averaging effect. For example, if the subject s expired breath volume is 0.5 liters, the Gas-System2 s chamber will average about 10 expired breaths. Definitions Click here for definitions of abbreviations used in this lesson and the calculation channel Expression formulas used for VO 2 and RER. Objectives 1. To obtain absolute VO 2 and RER values for a subject at rest (optional segments include hyperventilation and recovery from exercise). 2. To demonstrate the relationship between VO 2 and RER. NOTE: The Gas-System2 module is intended for VO2 and RER studies ranging from resting to moderate exercise. For prolonged or vigorous exercise studies where CO2 levels may exceed 5%, BIOPAC recommends the AcqKnowledge Research system with O2100C and CO2100C gas sensing modules. Equipment BIOPAC Gas Analysis System (GASSYS2-EA) BIOPAC Pneumotach Airflow Transducer [High Flow] (SS11LA) BIOPAC Calibration Syringe (AFT26 or AFT6) BIOPAC 22 mm Non-Rebreathing T Valve (AFT22) BIOPAC Rigid Coupler (AFT11E) for T-valve to tubing BIOPAC Coupler (AFT11C) for T-valve to calibration syringe BIOPAC Smooth Bore Tubing-35 mm (AFT7) BIOPAC Disposable Nose Clip (AFT3) BIOPAC Disposable Mouthpiece (AFT2) BIOPAC Disposable Bacterial Filters (AFT1) requires 2 filters Computer running Windows XP or higher, or Mac OS X Biopac Student Lab PRO software BIOPAC Data Acquisition Unit (MP36/MP35/MP30) OPTIONAL: BIOPAC Calibration Gas Cylinder (16% O 2, 4% CO 2 ), with regulator and tubing The following measures are also required: Ambient temperature in C Ambient barometric pressure in mmhg Pressure of water vapor at ambient temperature = 22.4 mmhg Page 2 of 16

3 Setup Hardware The Gas-System2 sensors are factory calibrated prior to shipping. The O 2 sensor is a zirconia solid electrolyte with a % sensing range and an estimated 5-year lifetime. It runs hot, which helps to burn off humidity. The CO 2 sensor uses a humidity-repellant (hydroponic) membrane and has a sensing range of 0-5%. It uses non-dispersive infra-red diffusion with single-beam IR and a self-calibrating algorithm. It also runs hot, which helps to burn off humidity. 1. The Gas-System2 module is supplied with a 5 4 amp wall adapter. Plug the adapter into the main power supply and insert the adapter plug into the DC Input on front of the Gas-System2 module. 2. Flip up the power switch on the Gas-System2 to turn it "ON." o The unit should warm up for at least five minutes before calibration, so it is suggested that you turn it on now and then complete setup. 3. Plug the transducers into the MP3X as follows: Transducer MP3X CH # SS11LA CH 1 O2 line (from the Gas-System2) CH 2 CO2 line (from the Gas-System2) CH 3 Page 3 of 16

4 4. Turn on the MP unit (assuming the AC100A power adapter has already been connected). 5. Connect the airflow accessories as follows: Air Flow Accessory AFT1 filter SS11LA transducer AFT22 non-rebreathing T-valve AFT11E coupler AFT7 tubing AFT26 or AFT6 calibration syringe AFT11C coupler Connects to SS11LA transducer, on the "Inlet" side AFT22 non-rebreathing T-valve AFT11E coupler AFT7 tubing Gas-System2 inlet (back side) AFT11C coupler AFT22 non-rebreathing T-valve Note: The above setup uses the AFT26 or AFT6 calibration syringe. For recording, you must replace the calibration syringe with an AFT1 filter and an AFT2 disposable mouthpiece (connected to each other and to the AFT22 T-Valve). Page 4 of 16

5 AFT7 tubing to Gas-System2 inlet 6. Pump the calibration syringe through enough cycles to fill the chamber with ambient air. o AFT26, pump 3-4 times. o AFT6, pump times. Software 1. Launch the BSL PRO software on the host computer. The program should create a new "Untitled1" window. 2. Open the VO 2 and RER Template by choosing File > Open > choose Files of type: Graph Template (*GTL) > File name: H19 V02 & RER.gtl The template will establish the required settings. This particular setup will provide a reading that, at any point in time, indicates the absolute volume of oxygen consumed, carbon dioxide produced, and the respective respiratory exchange ratio in the last 60 seconds. Units are liters because it is a volume measure, not a flow rate. 3. Save As the desired file name. Calibration The following calibration steps are outlined below: A. Calibrate flow transducer B. Calibrate O 2 channel (CH2) C. Calibrate CO 2 channel (CH3) D. Normalize moist gas to volume occupied by dry gas at 0 C, 760 mm. E. Optional: Calibrate for time interval (template is based on 60-second interval). Note: In the calculation channel formulas, the following assumed values were used for gas concentrations in normal atmosphere (chamber flooded with ambient air): 0 2 = 20.93% CO 2 = 0.04% N 2 = 79.03% The optional Stage 2 gas calibration is not required, but can be performed using the small line inlet on the back panel of the Gas-System2 module. If you introduce calibration gases into the chamber, adjust the calculation channel expressions for the real gas values. If you perform a gas calibration, use the following ranges: Flow (inspired): calibrate within the range of 0 to 13 liters/sec O 2 concentration: calibrate in range of 21% to 16% O 2 CO 2 concentration: calibrate in range of.04% to 4% CO 2 Page 5 of 16

6 Flow 1. Select MP3X > Setup Channels. 2. Click the wrench icon for CH Click the Scaling button. 4. Hold the Calibration Syringe such that the SS11LA hangs from it in an upright position. 5. Click on the Cal1 button. 6. Subtract 3,000 from the Cal1 input value and enter the result in Cal2 input value (Cal2 input value should be 3,000 less than Cal1 input value). 7. Confirm Cal1 scale value= 0 and Cal2 scale value= Click OK. 9. Set the equipment down. Page 6 of 16

7 O 2 Per Setup, the chamber should have been flooded with ambient air prior to O 2 calibration! 1. Click the wrench icon for CH Click the Scaling button. 3. Click the Cal2 button. 4. Enter into the Cal2 scale value. 5. Confirm both Cal1 input value= 0 and Cal1 scale value=0. 6. Click OK. CO 2 Per Setup, the chamber should have been flooded with ambient air prior to CO 2 calibration! 1. Click the wrench icon for CH Click the Scaling button. 3. Click the Cal1 button. 4. Enter 0.04 into the Cal1 scale value. 5. Add 10 to the Cal1 input value and enter the result as Cal2 input value. 6. Confirm Cal2 scale value= Click OK. Note: The sensor output is 10 mv per 1% C0 2. With proper calibration, if the sensor rails, it will read as 5% CO 2. Page 7 of 16

8 OPTIONAL Stage 2 Calibration (BIOPAC Calibration Gas Cylinder required) O 2 Per Setup, the chamber should NOW be flooded with a calibrated gas (16% O 2, 4% CO 2 )! 1. Click the wrench icon for CH Click the Scaling button. 3. Click the Cal1 button. 4. Enter 16 into the Cal1 scale value. 5. Click OK to dismiss scaling dialog. 6. Click OK to dismiss Input Channel Parameters dialog for CH 2. CO 2 Per Setup, the chamber should NOW be flooded with a calibrated gas (16% O 2, 4% CO 2 )! 1. Click the wrench icon for CH Click the Scaling button. 3. Click the Cal2 button. 4. Enter 4 into the Cal2 input value. 5. Click OK to dismiss the scaling dialog. 6. Click OK to dismiss Input Channel Parameters dialog for CH 3. Page 8 of 16

9 Gas Normalization Normalize moist gas to volume occupied by dry gas at 0 deg C, 760mm 1. Use this table to find the normalization factor. 2. Click the wrench icon for C3 Vis (STPD). 3. Enter the normalization factor in the formula: C2*(normalization factor) Be sure to enter the value as 0.x, with the leading decimal. 4. Click OK. Optional: Interval Only required if you change the time interval from the template, which is set for a 60-second interval. 1. Click the wrench icon for C2. Page 9 of 16

10 3. Calculate the Samples entry as follows: Integration Samples = Time Interval x Sample rate (Template setting: Integration Samples = 60 sec x 100 samples/sec = 6000 samples) 4. Enter the new Samples value. 5. Click OK. Recording This lesson records absolute volume of O 2 and CO 2 in a 60-second interval. Units are liters because this provides a volume measure, not a flow rate. If you change the time interval, you must adjust the Integrate formula. Hints for minimizing measurement error: 1. Ensure that the connections among the various pieces of equipment are fairly tight to prevent contamination via room air. 2. Try to start the recording immediately after a full exhale. This will prevent receiving results with an O 2 inspiration value less than that of the total expiration value (although the averaging will minimize the inaccuracies that this would cause). 3. It's very important that any extraneous volumes are minimized between the subject and the T-valve. Additional volumes affect the effective expired gas concentration levels. 4. The tubing must be connected to the SS11LA on the unlabeled side (that does NOT say "Inlet"). Preparation 1. Remove the calibration syringe from the transducer setup. 2. Establish the recording connections as shown below: Airflow Accessory Connects to AFT1 Filter AFT22 T-Valve (in the existing setup) AFT2 disposable mouthpiece AFT1 Filter Page 10 of 16

11 Segment 1 1. Have the subject put a nose clip on. 2. Click START in the PRO software. 3. Have the subject breathe normally for at least 2 minutes. o Subject must breathe long enough to ensure that the 5-liter mixing tank has been filled, thereby enabling it to average those breaths. o The subject should breathe into the disposable bacterial filter (AFT1) through a disposable mouthpiece attachment (AFT2). 4. Click STOP in the PRO software. Breathe through the mouthpiece Page 11 of 16

12 Optional Segments Recovery from Exercise 1. Have the subject perform several minutes of medium-to-strenuous exercise. o If desired, record the heart rate and correlate VO 2 to heart rate. 2. Click START in the PRO software. 3. Have the subject breathe normally for at least 2 minutes. 4. Click STOP in the PRO software. Hyperventilation 1. Click START in the PRO software. 2. Have the subject breathe normally for at least 1 minute. 3. Have the subject hyperventilate for at least 1 minute; enter a marker at start of hyperventilation (Esc on PC, F9 on Mac). 4. Have the subject breathe normally for at least 1 minute. 5. Click STOP in the PRO software. Notes To save recorded data, choose File menu > Save As > file type: BSL PRO files (*.ACQ) File name: (Enter Name) > Save button To erase all recorded data (make sure you have saved it first), and begin from Time 0, choose: MP menu > Setup Acquisition > Click Reset Page 12 of 16

13 Analysis The VO 2 and RER template will display channels as follows: Channel Displays Measure CH 1 Fi (inspired Air Flow) Value CH 2 O 2 E (expired O 2 ) Value CH 3 CO 2 E (expired CO 2 ) Value CH 40 (C1) N 2 E (expired N 2 ) Value CH 41 (C2) Vi (ATP) Value CH 42 (C3) VIS (STPD; inspired) Value CH 43 (C4) VES (STPD; expired) Value CH 44 (C5) VO 2 (STPD) Value CH 45 (C6) VCO 2 (STPD) Value CH 46 (C7) RER Value 1. Channels are not displayed because they are calculation channels required for conversions. You can show/hide channels by Ctrl-clicking the channel box. 2. Note that the VO 2 and RER measurements (bottom two channels) vary smoothly with time. The graphical and continuous nature of this recording and calculation method provides significant information regarding the changes of these variables over time. 3. To display measurements, click the measurement icon. 4. To display markers, click the flag icon. 5. To display grids, click the grid icon. Page 13 of 16

14 Appendix GRAPH TEMPLATE SETTINGS Click here to open a PDF of the graph template file settings. The BSL PRO Graph Template file automatically establishes the settings shown in the table. Definitions These abbreviations are used in the discussion and calculation channel Expression formulas for VO 2 and RER: ATP Concentration of gas at ambient temperature and pressure CO 2 Fi Carbon dioxide Est. concentration ambient environment: 0.04% CO 2 e = Carbon Dioxide fractional concentration in expired air Inspired air flow (ATP) N 2 Nitrogen N 2 e = Nitrogen fractional concentration in expired air N 2 e = 100 (CO 2 e + O 2 e) O 2 Pb PH 2 0 RER STPD Oxygen Est. concentration in ambient air: 20.93% O 2 e = Oxygen fractional concentration in expired air Est. concentration in expired breath: 16% oxygen Ambient barometric pressure (e.g. 745 mmhg) Ambient pressure of water vapor (e.g mmhg) Respiratory Exchange Ratio RER = VCO 2 / VO 2 Concentration of gas at standard temperature and pressure, dry Ta Ambient temperature (e.g. 24 deg. C) VCO 2 VO 2 Vi Volume of carbon dioxide produced (STPD) per 60-sec interval Real-time carbon dioxide production VCO 2 = (1/100)*[(Ves* CO 2 e) (Vis*.04)] Volume of oxygen consumed (STPD) per 60-sec interval Real-time oxygen consumption VO 2 = (1/100)*[(Vis*20.93) (Ves*O 2 e)] Inspired air volume (ATP) per 60-sec interval Vi = Integrate (Fi) over 60 seconds. The number of samples is equal to the time interval x sample rate (template is 60 sec. x 100 samples/sec = 6000 samples). Select "Average over samples." Vis Inspired air volume (STPD) per 60-sec interval Vis = C2 * Normalization Factor (see Table of Normalization Factors) If normalization factors are unavailable, you can use this formula: Vis = Vi*(273/(273+Ta))*((Pb-PH 2 0)/760) where Ta = ambient temperature Pb = barometric pressure PH 2 O = 22.4 mmhg (assumed value) Ves Expired air volume (STPD) per 60-sec interval Ves = (Vis*79.03) / N 2 e Note: If you change the time interval from 60 seconds, you must adjust the Vi formula. Page 14 of 16

15 Normalization Factors The following table lists the factors required to reduce volume of moist gas to volume occupied by dry gas at 0 C, 760 mm. OBR is an abbreviation for observed barometric reading, uncorrected for temperature. Fahrenheit values are roughly converted and rounded from Celsius values as T F =(9/5*T C )+32ºF. Important! All factors are 0.x, with x being the value from the table. Be sure to include the leading decimal when you complete your calculation. ºC ºF OBR Page 15 of 16

16 ºF ºC From Peters and Van Slyke, Quantitative Clinical Chemistry, vol. 11. (Methods) Baltimore: Williams and Wilkins, 1932, reprinted Page 16 of 16

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805)

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805) APPLICATION NOTE 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 7.10.18 Application Note 183 VO 2 and RER Measurement For simplified VO 2 measurement

More information

Application Note 268 Metabolic Analysis Systems and Propane-based Verification

Application Note 268 Metabolic Analysis Systems and Propane-based Verification Application Note 268 Metabolic Analysis Systems and Propane-based Verification 8.10.11 This application note examines a conventional equipment setup for metabolic measurements. The setup is exposed to

More information

BSL PRO Lesson H19: V O 2, V CO 2, RER, O 2 deficit and EPOC Measurement. Objectives: Equipment:

BSL PRO Lesson H19: V O 2, V CO 2, RER, O 2 deficit and EPOC Measurement. Objectives: Equipment: HIGHER EDUCATION 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com Updated 05.30.18 : V O 2, V CO 2, RER, O 2 deficit and EPOC Measurement This BSL PRO

More information

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V O2100C Oxygen Measurement Module The O2100C module measures the partial pressure of O2 and thus the module output is proportional to the pressure in the sample cell. Gas sampled must be free of liquids

More information

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300 Technical Note GA-300 Overview The GA-300 CO2 and O2 Gas Analyzer is easy to use, robust, and adaptable to human, animal, and plant applications. The GA-300 has two analog outputs to allow recording and

More information

iworx Sample Lab Experiment HE-4: Respiratory Exchange Ratio (RER)

iworx Sample Lab Experiment HE-4: Respiratory Exchange Ratio (RER) Experiment HE-4: Respiratory Exchange Ratio (RER) Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to perform the experiment

More information

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Background There are two main sources of energy available for animal metabolism: carbohydrates (CHO) and fats. These molecules are broken

More information

Lesson 12 New Procedure PULMONARY FUNCTION I

Lesson 12 New Procedure PULMONARY FUNCTION I Physiology Lessons for use with the Biopac Student Lab Lesson 12 New Procedure PULMONARY FUNCTION I Volumes and Capacities For Windows 98SE, Me, 2000 Pro, XP or Mac OS X 10.3-10.4 Richard Pflanzer, Ph.D.

More information

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates Experiment HE-9: Resting, Active, and Exercising Metabolic Rates Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to

More information

O 6 = 6 CO H 2 O H 2 = 16 CO 2. consumed during the oxidation of fatty acids, a ratio of 0.7.

O 6 = 6 CO H 2 O H 2 = 16 CO 2. consumed during the oxidation of fatty acids, a ratio of 0.7. Experiment HE-4: Respiratory Exchange Ratio (RER) Background The two sources of energy available for human metabolism are carbohydrates (CHO) and fats. These molecules are broken down, or catabolized,

More information

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Background There are two main sources of energy available for animal metabolism: carbohydrates (CHO) and fats. These molecules are broken

More information

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR)

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR) Experiment HE-5: Resting Metabolic Rate (RMR) Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to perform the experiment

More information

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR)

iworx Sample Lab Experiment HE-5: Resting Metabolic Rate (RMR) Experiment HE-5: Resting Metabolic Rate (RMR) Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to perform the experiment

More information

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates Experiment HE-9: Resting, Active, and Exercising Metabolic Rates Before Starting 1. Read the procedures for the experiment completely before beginning the experiment. Have a good understanding of how to

More information

Pulmonary Function I (modified by C. S. Tritt, April 10, 2006) Volumes and Capacities

Pulmonary Function I (modified by C. S. Tritt, April 10, 2006) Volumes and Capacities I. Introduction Pulmonary Function I (modified by C. S. Tritt, April 10, 2006) Volumes and Capacities The volume of air a person inhales (inspires) and exhales (expires) can be measured with a spirometer

More information

Understanding the Issues Affecting the Accuracy of Measuring Oxygen Consumption Utilizing Metabolic Carts

Understanding the Issues Affecting the Accuracy of Measuring Oxygen Consumption Utilizing Metabolic Carts Understanding the Issues Affecting the Accuracy of Measuring Oxygen Consumption Utilizing Metabolic Carts Thank you for joining us. The webinar will begin shortly. Tweet #LifeScienceWebinar #ISCxAEITech

More information

SS32L Dissolved O 2 Probe

SS32L Dissolved O 2 Probe BIOPAC WWW.biopac.com Systems, Inc. Application Note PH-185 SS32L Dissolved O 2 Probe SS32L Dissolved O 2 Probe BIOPAC Software SS32L Specifications BSL PRO v. 3.6.6 BIOPAC Hardware SS32L Dissolved O 2

More information

Air Flow & Gas Analysis Accessories

Air Flow & Gas Analysis Accessories AFT Series Air Flow & Gas Analysis Accessories AFT1 AFT2 AFT3 AFT4 AFT6 AFT7 AFT8 AFT 9 AFT10 Disposable Bacterial Filter 22mm ID/OD; designed to remove airborne bacteria; for use between the TSD117 and

More information

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance Introduction Expired Gas Analysis or indirect calorimetry, can be used to measure ventilation and the fractions of oxygen and carbon dioxide in expired air. From these measurements, the body's oxygen consumption

More information

Resting, Active, and Exercising Metabolic Rates

Resting, Active, and Exercising Metabolic Rates iworx Physiology Lab Experiment Experiment HE-9 Resting, Active, and Exercising Metabolic Rates Background Set-up Lab Appendix Note: The lab presented here is intended for evaluation purposes only. iworx

More information

Biology 13A Lab #12: The Respiratory System

Biology 13A Lab #12: The Respiratory System Biology 13A Lab #12: The Respiratory System Lab #12 Table of Contents: Expected Learning Outcomes.... 94 Introduction...... 95 Activity 1: Structures of the Respiratory System 96 Activity 2: Measuring

More information

Content Display. - Introduction. Laboratory Unit : Lab 2 - Measurement of Oxygen Consumption. KINE xxxx Exercise Physiology

Content Display. - Introduction. Laboratory Unit : Lab 2 - Measurement of Oxygen Consumption. KINE xxxx Exercise Physiology Content Display Laboratory Unit : Lab 2 - Measurement of Oxygen KINE xxxx Exercise Physiology 6 Laboratory Unit 2 Lab 2 - Measurement of Oxygen 1 LAB2P1 - Introduction The goal of lab 2 is to facilitate

More information

PRODUCT SHEET. Order probe only as RXPROBE02

PRODUCT SHEET. Order probe only as RXPROBE02 SS69L DISSOLVED OXYGEN PROBE TRANSDUCER Order probe only as RXPROBE02 Order interface only as BSL-TCI16 SS69L Components The SS69L transducer measures dissolved oxygen. The SS69L includes a dissolved oxygen

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

PRODUCT SHEET. See also: Student Accessory Pack BSL-ACCPACK and BSLACCPACK-11B AFT7 AFT7L AFT12 AFT6A AFT27 AFT16

PRODUCT SHEET. See also: Student Accessory Pack BSL-ACCPACK and BSLACCPACK-11B AFT7 AFT7L AFT12 AFT6A AFT27 AFT16 See also: Student Accessory Pack BSL-ACCPACK and BSLACCPACK-11B AFT SERIES AIRFLOW & GAS ANALYSIS ACCESSORIES Includes the following airflow accessories: Bacterial Filters Mouthpieces Calibration Airflow

More information

Respiration. Exercise 1A: Breathing in Resting Volunteers Aim: To measure breathing parameters in a resting individual.

Respiration. Exercise 1A: Breathing in Resting Volunteers Aim: To measure breathing parameters in a resting individual. Respiration Background The amount of air that moves in or out of the lungs during any one breathing cycle is called the tidal volume. Above and beyond normal inspiration, it is possible to breathe in additional

More information

Experiment HS-2: Breathing and Gravity

Experiment HS-2: Breathing and Gravity Experiment HS-2: Breathing and Gravity Background Over a long time, the amount of oxygen (O 2 ) taken up and carbon dioxide (CO 2 ) given off at the tissues is matched with the amount of O 2 taken up and

More information

Respiratory Response to Physiologic Challenges. Evaluation copy

Respiratory Response to Physiologic Challenges. Evaluation copy Respiratory Response to Physiologic Challenges Computer 20 The respiratory cycle of inspiration and expiration is controlled by complex mechanisms involving neurons in the cerebral cortex, brain stem,

More information

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage Technical Note ISE-730 and DO2-100 Overview In 1954, Dr. Leland Clark invented the first membrane-covered electrode designed to measure the concentration of oxygen in blood, or solution. This electrode

More information

Overview. Front Panel: Keypad and Display

Overview. Front Panel: Keypad and Display Overview The GA-200B is an analyzer that integrates a gas sampling system with sensors to measure and display the concentrations of oxygen and carbon dioxide in a sample as the percentage of a gas in the

More information

Human Respiration Laboratory Experiment By

Human Respiration Laboratory Experiment By Human Respiration Laboratory Experiment By Alison L., Thurow, Brittany Baierlein, Rachel C. Holsinger and Robin L. Cooper Department of Biology, University of Kentucky, Lexington, KY 40506 0225, USA. Purpose:

More information

Vapor Pressure of Liquids

Vapor Pressure of Liquids Vapor Pressure of Liquids In this experiment, you will investigate the relationship between the vapor pressure of a liquid and its temperature. When a liquid is added to the Erlenmeyer flask shown in Figure

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

Oxygen Meter User Manual

Oxygen Meter User Manual Oxygen Meter User Manual Monday, July 23, 2007 1. Outline...2 2. Program...3 2.1. Environment for program execution...3 2.2. Installation...3 2.3. Un installation...3 2.4. USB driver installation...3 2.5.

More information

Figure 1. A schematic diagram of the human respiratory system.

Figure 1. A schematic diagram of the human respiratory system. Introduction to Respiration In this experiment, you will investigate various aspects of normal breathing, hyperventilation, rebreathing the effect of changing airway resistance and ways in which to measure

More information

iworx Metabolic Accessories

iworx Metabolic Accessories Technical Note Overview A selection of iworx metabolic accessories used in the recording of exercise physiology data is presented in this tech note. A-GAA-MC 5 L mixing chamber LabScribe is a trademark

More information

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Bante821 Portable Dissolved Oxygen Meter Instruction Manual Bante821 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante821 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante821 portable dissolved oxygen

More information

Biology 347 General Physiology Lab Basic Pulmonary Functions: Respirometry and Factors that Effect Respiration

Biology 347 General Physiology Lab Basic Pulmonary Functions: Respirometry and Factors that Effect Respiration Biology 347 General Physiology Lab Basic Pulmonary Functions: Respirometry and Factors that Effect Respiration Objectives Students will measure breathing parameters in a resting subject. Students will

More information

Review of gas frac9ons in atmospheric air 2/2/16. Metabolic Calculations of Indirect Calorimetry. Expired gas analysis.

Review of gas frac9ons in atmospheric air 2/2/16. Metabolic Calculations of Indirect Calorimetry. Expired gas analysis. Metabolic Calculations of Indirect Calorimetry Understanding atmospheric air and gas volumes Calcula9ng VO2 The Haldane Transforma9on Calcula9ng VCO2 Calcula9ng RER Calcula9ng caloric expenditure Expired

More information

Fish Respiration and Q 10

Fish Respiration and Q 10 BIOPAC Systems, Inc., 42 Aero Camino, Goleta, CA 93117 Ph: 805/685-0066 * Fax: 805/685-0067 * Web: http://www.biopac.com/ * E-mail: info@biopac.com BSL PRO Lesson #A07 2004 BIOPAC Systems, Inc. Updated

More information

Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams)

Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams) Name Lab Partners Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams) Part 1. Lung Volumes and Capacities Objectives 1. Obtain graphical representation of lung capacities

More information

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Universal pycnometer, no

More information

Gas exchange measurement in module M- COVX

Gas exchange measurement in module M- COVX Gas exchange measurement in module M- COVX This document explains how the gas exchange measurement in the gas module M-COVX works. 1. Basic equations To obtain the oxygen consumption VO2 of a patient,

More information

RM-80 respiration monitor

RM-80 respiration monitor RM-80 respiration monitor User Manual September 18, 2015 0025-003M 950 North Hague Avenue Columbus, Ohio 43204-2121 USA Sales: sales@colinst.com Service: service@colinst.com Phone: (614) 276-0861 Fax:

More information

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives exercise 7 Respiratory System Mechanics Objectives 1. To explain how the respiratory and circulatory systems work together to enable gas exchange among the lungs, blood, and body tissues 2. To define respiration,

More information

KEM Scientific, Inc. Instruments for Science from Scientists

KEM Scientific, Inc. Instruments for Science from Scientists KEM Scientific, Inc. Instruments for Science from Scientists J-KEM Scientific, Inc. 6970 Olive Blvd. St. Louis, MO 63130 (314) 863-5536 Fax (314) 863-6070 E-Mail: jkem911@jkem.com Precision Vacuum Controller,

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information

CARDIOVIT AT-104 ergospirometry

CARDIOVIT AT-104 ergospirometry E D E R I T Z M A N U F A C T U R E D I N A N D S W I G N S E D L A N D CARDIOVIT AT-104 ergospirometry Cardiopulmonary exercise tests (CPET) with touch screen by SCHILLER it s the experience that counts.

More information

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung Susan Blonshine RRT, RPFT, FAARC, AE-C Joint ATS/ERS Taskforce Recent literature reviewed Surveyed current technical

More information

Vapor Pressure of Liquids

Vapor Pressure of Liquids Experiment 10 In this experiment, you will investigate the relationship between the vapor pressure of a liquid and its temperature. When a liquid is added to the Erlenmeyer flask shown in Figure 1, it

More information

SDM-2012 Docking Station Standalone Configuration Operator s Manual

SDM-2012 Docking Station Standalone Configuration Operator s Manual SDM-2012 Docking Station Standalone Configuration Operator s Manual Part Number: 71-0254RK Revision: P5 Released: 10/5/12 www.rkiinstruments.com Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

Specifications...Easy CLIA Classification: Sample Type: Blood gas analyzer operation has never been simpler Sample Size: Measured Parameters PCO2

Specifications...Easy CLIA Classification: Sample Type: Blood gas analyzer operation has never been simpler Sample Size: Measured Parameters PCO2 medicacorp.com easy inside and out With EasyBloodGas, Medica has redefined blood gas analyzer design The sophistication of traditional blood gas analyzers has been packaged in a new compact format with

More information

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings SomnoSuite FAQ V1.3 January 2015 Setup 1. How do I connect the SomnoSuite to my oxygen source? 2. Is there a way to speed up the downward movement of the pusher block when setting the empty position? 3.

More information

VT650. Gas Flow Analyzer. Technical data. Key benefits and features:

VT650. Gas Flow Analyzer. Technical data. Key benefits and features: VT650 Gas Flow Analyzer Technical data Accurately test gas flow equipment including ventilators with the all-in-one, portable Fluke Biomedical VT650 Gas Flow Analyzer. Accurate The VT650 Gas Flow Analyzer

More information

Evaluation copy. Vapor Pressure of Liquids. computer OBJECTIVES MATERIALS

Evaluation copy. Vapor Pressure of Liquids. computer OBJECTIVES MATERIALS Vapor Pressure of Liquids Computer 10 In this experiment, you will investigate the relationship between the vapor pressure of a liquid and its temperature. When a liquid is added to the Erlenmeyer flask

More information

INSTRUCTOR RESOURCES

INSTRUCTOR RESOURCES Gases: Dalton s Law INSTRUCTOR RESOURCES By Dale A. Hammond, PhD LEARNING OBJECTIVES introduce the concept of ideal gases. experimentally determine the relationship between pressure and amount of gas,

More information

Vapor Pressure of Liquids

Vapor Pressure of Liquids Vapor Pressure of Liquids Experiment 10 In this experiment, you will investigate the relationship between the vapor pressure of a liquid and its temperature. When a liquid is added to the Erlenmeyer flask

More information

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Lab 1c Isentropic Blow-down Process and Discharge Coefficient 058:080 Experimental Engineering Lab 1c Isentropic Blow-down Process and Discharge Coefficient OBJECTIVES - To study the transient discharge of a rigid pressurized tank; To determine the discharge coefficients

More information

Cardiovascular and respiratory adjustments to exercise

Cardiovascular and respiratory adjustments to exercise Cardiovascular and respiratory adjustments to exercise Additional notes on breathing and use of respiratory belt and pulse transducers Notes on breathing The metabolic activities of tissues use up oxygen

More information

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante810 Benchtop Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante810 benchtop dissolved oxygen

More information

Experiment B-3 Respiration

Experiment B-3 Respiration 1 Experiment B-3 Respiration Objectives To study the diffusion process of oxygen and carbon dioxide between the alveoli and pulmonary capillaries. To determine the percentage of oxygen in exhaled air while

More information

Lung Volumes and Capacities

Lung Volumes and Capacities Lung Volumes and Capacities Normally the volume of air entering the lungs during a single inspiration is approximately equal to the volume leaving on the subsequent expiration and is called the tidal volume.

More information

The University of Hong Kong Department of Physics Experimental Physics Laboratory

The University of Hong Kong Department of Physics Experimental Physics Laboratory The University of Hong Kong Department of Physics Experimental Physics Laboratory PHYS2260 Heat and Waves 2260-1 LABORATORY MANUAL Experiment 1: Adiabatic Gas Law Part A. Ideal Gas Law Equipment Required:

More information

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc. The HumiSys RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The new HumiSys with single or dual RH probes capabilities is

More information

Exercise 3. Power Versus Wind Speed EXERCISE OBJECTIVE DISCUSSION OUTLINE. Air density DISCUSSION

Exercise 3. Power Versus Wind Speed EXERCISE OBJECTIVE DISCUSSION OUTLINE. Air density DISCUSSION Exercise 3 Power Versus Wind Speed EXERCISE OBJECTIVE When you have completed this exercise, you will know how to calculate the power contained in the wind, and how wind power varies with wind speed. You

More information

A180 DO/BOD/OUR/SOUR Meter Instruction Manual

A180 DO/BOD/OUR/SOUR Meter Instruction Manual A180 DO/BOD/OUR/SOUR Meter Instruction Manual BANTE INSTRUMENTS CO., LTD A180 DO/BOD/OUR/SOUR Meter 1 Introduction Thank you for selecting the A180 multiparameter dissolved oxygen meter. This manual provides

More information

Operation Manual. Pro 4 Warn Analyzer TM Rev 08.17

Operation Manual. Pro 4 Warn Analyzer TM Rev 08.17 Operation Manual Pro 4 Warn Analyzer TM Rev 08.17 If you have any questions on this equipment please contact Technical Support at: Nuvair 1600 Beacon Place Oxnard, CA 93033 Phone: 805-815-4044 FAX: 805-486-0900

More information

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003 CEE 331 Lab 1 Page 1 of 9 SAFETY Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003 Laboratory exercise based on an exercise developed by Dr. Monroe Weber-Shirk The major safety hazard in this

More information

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument OxyScan Graphic Operating Instructions UMS Micro-oxygen sensor 501 Microprocessor instrument Introduction Thank you for choosing the UMS Micro Oxygen Sensor 501 - a highly advanced product! Please read

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Bante820 Portable Dissolved Oxygen Meter Instruction Manual Bante820 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante820 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante820 portable dissolved oxygen

More information

Instruction and Maintenance Manual

Instruction and Maintenance Manual Instruction and Maintenance Manual GRYF OXY ZM 02/100/2 E Contact GRYF HB, spol. s r.o. Cechova 314 Havlickuv Brod 580 01 tel.: +420 569 426 627 fax: +420 569 426 627 www.gryf.eu ver.: 11.2.2015 OEM module

More information

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA Software version 2.6 Table of contents 1 Introduction...3 1.1 Optical Sensor Technology...3 1.2 Installation...4 1.3 Operation...5

More information

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual BANTE INSTRUMENTS CO., LTD DOscan10 Pocket Dissolved Oxygen Tester 1 Thank you for selecting the DOscan10 pocket dissolved oxygen tester. This

More information

HESTA Issues affecting the accuracy of Metabolic Carts. Danny Rutar

HESTA Issues affecting the accuracy of Metabolic Carts. Danny Rutar HESTA 2017 Issues affecting the accuracy of Metabolic Carts Danny Rutar Danny Rutar Managing Director, Redback Biotek Qualifications Biophysics / Instrumentation Consultant Sport Technologist Athletics

More information

Respiration Lab Instructions

Respiration Lab Instructions Respiration Lab Instructions This laboratory investigation can be performed in any order. Be sure to read all instructions for each section before performing the experiment. PART 1 STUDENT WET SPIROMETER

More information

Vapor Pressure of Liquids

Vapor Pressure of Liquids Vapor Pressure of Liquids In this experiment, you will investigate the relationship between the vapor pressure of a liquid and its temperature. When a liquid is added to the Erlenmeyer flask shown in Figure

More information

PRESSURE-TEMPERATURE RELATIONSHIP IN GASES

PRESSURE-TEMPERATURE RELATIONSHIP IN GASES PRESSURE-TEMPERATURE RELATIONSHIP IN GASES LAB PS2.PALM INTRODUCTION Gases are made up of molecules that are in constant motion and exert pressure when they collide with the walls of their container. The

More information

BREATH-BY-BREATH METHOD

BREATH-BY-BREATH METHOD BREATH-BY-BREATH METHOD COR-MAN-0000-005-IN / EN Issue A, Rev. 2 2013-07 INNOISION ApS Skovvænge DK-5620 Glamsbjerg Denmark Tel.: +45 65 95 91 00 Fax: +45 65 95 78 00 info@innovision.dk www.innovision.dk

More information

VT900. Gas Flow Analyzer. Technical data. Key benefits and features:

VT900. Gas Flow Analyzer. Technical data. Key benefits and features: VT900 Gas Flow Analyzer Technical data The Fluke Biomedical VT900 is designed to accurately and reliably test all types of medical gas flow equipment ventilators, insufflators, oxygen meters especially

More information

Experiment B-43 Respiration Rates

Experiment B-43 Respiration Rates 1 Experiment B-43 Rates Objectives To learn about the lungs and respiration s. To measure respiration s of the students. To compare female and male respiration s. To compare respiration s before and after

More information

Quality assurance of Modular Metabolic Systems. Danny Rutar, Redback Biotek, Ireland/Australia Jeffrey Swen(Intern), University of Ulster, Belfast.

Quality assurance of Modular Metabolic Systems. Danny Rutar, Redback Biotek, Ireland/Australia Jeffrey Swen(Intern), University of Ulster, Belfast. Quality assurance of Modular Metabolic Systems Danny Rutar, Redback Biotek, Ireland/Australia Jeffrey Swen(Intern), University of Ulster, Belfast. Background (1987) Danny Rutar: (technologist, scientist)

More information

New product release. Universal Rebreather Monitor (URBM) Three cell / Independent backup

New product release. Universal Rebreather Monitor (URBM) Three cell / Independent backup New product release Closed Circuit Research is pleased to announce the launch of our range of Universal Rebreather Monitors Universal Rebreather Monitor (URBM) Three cell / Independent backup Key features

More information

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Temperature controlled, multi-technique volumetric

More information

Lung Volumes and Capacities

Lung Volumes and Capacities BIOL242 Lung Volumes and Capacities Measurement of lung volumes provides a tool for understanding normal function of the lungs as well as disease states. The breathing cycle is initiated by expansion of

More information

You Are Really Full of Hot Air!

You Are Really Full of Hot Air! You Are Really Full of Hot Air! Student Information Page 5B Activity Introduction: You re just full of hot air! How many times has someone said this to you when they didn t quite believe what you were

More information

Flow Vision I MX Gas Blending Station

Flow Vision I MX Gas Blending Station Flow Vision I MX Gas Blending Station Alicat Scientific, Inc. 7641 N Business Park Drive Tucson, Arizona 85743 USA alicat.com 1 Notice: Alicat Scientific, Inc. reserves the right to make any changes and

More information

Biology Unit 2, Structure of Life, Lab Activity 2-3

Biology Unit 2, Structure of Life, Lab Activity 2-3 Biology Unit 2, Structure of Life, Lab Activity 2-3 Cellular respiration is the release of energy from organic compounds by metabolic chemical oxidation in the mitochondria within each cell. Cellular respiration

More information

VT900A + VAPOR. Technical data. One Solution. Auto-detection to Ensure Patient Safety. Key benefits and features

VT900A + VAPOR. Technical data. One Solution. Auto-detection to Ensure Patient Safety. Key benefits and features Technical data VT900A + VAPOR The Fluke Biomedical VT900A + VAPOR is designed to efficiently and reliably perform a full anesthesia machine PM, from ventilators to vaporizers. One Solution VT900A + VAPOR

More information

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation High-precision Humidity Calibrator The role of humidity calibrations that are accurate, reproducible, and documentable is becoming more and more important. ISO quality guidelines and regulations according

More information

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 QUICK START GUIDE P/N 6006154, REVISION C MAY 2013 Model 8120 Mask Integrity Tester is patented under U.S. Patent No. 8,312,761. Additional patents are pending.

More information

AKTA pure 25 New Owner s Intro

AKTA pure 25 New Owner s Intro AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

LABORATORY INVESTIGATION

LABORATORY INVESTIGATION LABORATORY INVESTIGATION MEASURING THE RATE OF PHOTOSYNTHESIS Light and Photosynthesis About 2.5-3 billion years ago a new chemical process, photosynthesis, was evolved by a unicellular life form. This

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

Section 1: Overview... 6 Features... 6 Supported Software Part 1: Getting Started 7

Section 1: Overview... 6 Features... 6 Supported Software Part 1: Getting Started 7 Table of Contents About this Manual... 3 Intended Audience... 3 How this Manual is Organized... 4 Conventions Used in this Manual... 5 Related Information... 5 Where To Get Help... 5 Section 1: Overview...

More information

Aerobic Respiration. Evaluation copy

Aerobic Respiration. Evaluation copy Aerobic Respiration Computer 17 Aerobic cellular respiration is the process of converting the chemical energy of organic molecules into a form immediately usable by organisms. Glucose may be oxidized completely

More information

Exploring the Properties of Gases. Evaluation copy. 10 cm in diameter and 25 cm high)

Exploring the Properties of Gases. Evaluation copy. 10 cm in diameter and 25 cm high) Exploring the Properties of Gases Computer 30 The purpose of this investigation is to conduct a series of experiments, each of which illustrates a different gas law. You will be given a list of equipment

More information

Exploring the Properties of Gases

Exploring the Properties of Gases Computer 30 The purpose of this investigation is to conduct a series of experiments, each of which illustrates a different gas law. You will be given a list of equipment and materials and some general

More information

Breeze Suite Quick Start Operator Manual. An Introduction to Pulmonary Function Testing

Breeze Suite Quick Start Operator Manual. An Introduction to Pulmonary Function Testing Breeze Suite Quick Start Operator Manual An Introduction to Pulmonary Function Testing The MedGraphics Breeze Suite Quick Start Operator Manual is Part No. 142143-001 Rev A. Caution: Federal (USA) law

More information