1 Elimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU. Control Analytics, Inc 6017 Enterprise Dr Export, PA (724) 387 CEMS (2367) sales@controlanalytics.com Control Analytics, Inc Page 1

2 Application Note Elimination of Percent Level H 2 S Calibration Gas standards from Flare Gas Monitoring Systems Measuring Total Reduced Sulfur. What is the problem? In November 2015 most refineries were required to be in compliance with the USEPA NSPS Subpart Ja rules which mandate the continuous measurement of H 2 S and Total Reduced Sulfur or Total Sulfur in Flare Gas samples (Pending compliance regulations, will add BTU to this measurement requirement). The regulation established an emission limit for H 2 S of 162 ppm (3 hour rolling average) and established the span for the measurement at ppm. The regulation also requires that a continuous TRS (or Total Sulfur) measurement be made during flaring events. Companies must establish the span of this Instrument based on "Maximum Anticipated Sulfur Concentration" in the flare gas. At many refineries, the flare gas, under certain flaring conditions, can contain sulfur in percent concentrations resulting in instrument span settings in percent concentrations, some at 0 100% by volume. The regulation also requires that the TRS/Total Sulfur Monitor follow USEPA Performance Specification 5 (TRS), which in turn references calibration drift requirements found in Performance Specification 2. PS 2 is familiar to most users as the daily calibration specification required by the NO X /O 2 priority pollutant CEMs operating in the plant. If the rule is followed as written, the TRS/TS instruments must be verified on daily span gas containing H 2 S concentrations between 50% 100% of the analyzer span. Companies employing these percent range analyzer spans are therefore required to operate H 2 S calibration gas cylinders, or liquid H 2 S standards, on a daily basis that are thousands or tens of thousands times more concentrated than the 100 ppm OSHA IDLH (Immediately Dangerous to Life and Health) limit for H 2 S. The use of these cylinders has created an extremely hazardous situation for engineers and technicians operating the CEMs, changing gas cylinders, as well as dangerous cylinder storage operations. What is the solution? The solution to the problem is to obtain from US EPA approval of an Alternative Monitoring Plan (AMP) that employs specific technology and methodology in such a manner that your system can be validated without the use of these dangerous high range H 2 S standards. The US EPA has approved such site specific AMPs which employ measurement techniques that are inherently linear and are accurate from ppm to percent levels. The AMP defines the verification methodology that provides validation of the analyzer ability to measure the components of interest in the flare from the low ppm levels to the high percent levels by using a combination of low levels of H 2 S in conjunction with surrogate gases for high range response. Control Analytics, Inc Page 2

3 In these applications, once approved, the use of the same daily gas for the ppm measurement can also be employed for daily verification of the TRS/Total Sulfur measurement. Generally, the highest hazardous daily calibration gas that will be needed is 275 ppm H 2 S, which is being used routinely on fuel gas H 2 S CEMs throughout the plant. Not all instruments have the ability to be used in this application. In fact, the successful instrument must have some unique capabilities if it is intended to be used to support this an AMP method. The instrument capabilities must include: A detection technique that is inherently linear based on the physics of the measurement. Ability to measure the surrogate gas as well as the TRS/H 2 S found in these applications using the one sample path, sampling system and detector. A detection technique capable of measuring low ppm to high percent levels without applying changing sample volume (multiple sample loops), and using a common detector for all analysis. Ability to measure all constituents in the sample, including the sulfur species as well as the surrogate gas, with the same linearity and accuracy throughout the entire measurement range. The only analyzer that meets these stringent requirements and currently demonstrate the capability operating in an EPA AMP is the Extrel MAX 300 Mass Spectrometer. The reasons for the use of this instrument in flare gas applications where clients are seeking EPA AMP approval are: The Extrel IMAX 300 can detect H 2 S as well as the surrogate gas (CO 2 ) with a dynamic measurement range from 1ppm to 100% by volume. The Extrel IMAX 300 can detect all flare components, including the CO 2 surrogate gas as well as all sulfur species and all hydrocarbon species, and produce a measurement for the H 2 S, 0 100% TRS and BTU, all with one instrument, on one sample using a single detector and sample system. The Extrel IMAX 300 is inherently linear in response across its entire dynamic range. How does the measurement work without high range gas validations? US EPA requires that all measurements for environmental compliance have a high degree of accuracy, defined by their performance specifications, and validated to insure correct readings to source gas concentrations. Techniques employed in the AMPs currently in use provide this assurance without validating the measurement on the high range H 2 S gas, and have obtained EPA AMP approval as such. The verification method is largely based upon the demonstrated linearity of the Mass Spec instrument methodology, and in particular the extensive H 2 S measurement data provided by Extrel on the method. This data includes multiple gas standards containing sulfur blends that simulate various flare gas conditions. The high degree of linearity demonstrated in the Extrel Mass Spec method ensures that a single point calibration will be accurate across the instrument's entire dynamic range, and it meets the requirement for an alternative validation method for the high H 2 S measurement using a low level H 2 S standard in place of dangerous, highpercent H 2 S blends. Due to the demonstrated linearity of the technique, an accurate validation at a low concentration validates the measurement at a high concentration as well. Control Analytics, Inc Page 3

4 Figure 1 In addition, Extrel has demonstrated the linearity and accuracy of a calibrated system is not component specific. This means that instrument accuracy can be validated at, or near, full range on the detector (100%) using any high concentration standard, regardless of chemical composition. For Subpart Ja, or any regulation requiring daily validations using dangerous high concentration sulfur standards, the mass spectrometer operation principle allows for an alternative approach using common, safe to handle gases like CO 2, nitrogen or argon to validate analysis at the high range, and has been approved by EPS as an alternative validation method. Control Analytics, Inc Page 4

5 Figure 2 Surrogate Gas Calibration: The use of the manufacturers verified and published linearity and response data provides the basis for low level verification of the entire measuring range of H 2 S using only low level H 2 S gas. The Surrogate Gas calibration provides a secondary validation of the instrument response and accuracy by verifying the response of the instrument system at both ppm and percent concentrations using a safe, surrogate gas such as CO 2. In this method the instrument is validated by a bottle of calibration gas of 270 ppm H 2 S and 270 PPM CO 2, and another validation using a bottle containing 20% by volume CO 2. This allows the validation of the H 2 S as well as a surrogate at ppm concentrations, and providing verification of the instrument response and linearity at high concentrations. The use of the inherent linearity of the Mass Spec, the data provided by the manufacturer, and the site validation of H 2 S response and system linearity and accuracy provides an alternative validation method that has been accepted by EPA. Control Analytics, Inc Page 5

6 Figure 3 Control Analytics, Inc Page 6

7 Figure 4 What is the response time of the Mass Spec compared to my existing systems? The Mass spec measurement is much faster than any comparative analyzer being employed on flare gas monitoring systems. The added value that can be realized in enhanced response time is dependent on the type of system you employ now for the H 2 S, TRS/TS measurements. The total analysis time of the Mass Spec is between 5 10 seconds. Typical GC H 2 S measurement cycle times are 5 6 minutes, while Oxidative Total Sulfur (Thermo Sola II type) measurement cycles are closer to 3 5 minutes. The response time of GC's and Oxidative Total Sulfur analyzers is longer when responding to a flaring event where sulfur concentrations are rising or falling quickly, as these instruments must switch injection sample loops to keep the measurement on scale, thereby requiring multiple injections before a stable reading is obtained. This delay while "chasing" rising or falling sulfur levels and can lead to improper assessment of flaring events and improper reporting. The Mass Spec can complete and output measurements on all components in 5 10 second intervals. Control Analytics, Inc Page 7

8 Can I measure BTU along with the H 2 S and TRS/TS measurements? Yes. The Mass Spectrometer can make a comprehensive hydrocarbon species and BTU measurement. In addition, the Mass Spectrometer cycle times are typically 100 times faster than the typical GC cycle time of 15 minutes. How difficult is it to modify my system? In most cases the modifications needed to convert a system which employs typical GC's and Thermo Sola type Total Sulfur instruments are straightforward and can be completed on site in a few days. The modifications vary depending upon the type of sampling system employed, the design of the calibration gas cabinet, and other details. Control Analytics can provide a detailed list of system specific modifications, as well as a budgetary proposal for the costs associated with purchasing the hardware and making the system modifications onsite. Alternatively, CAI can provide the engineering and material, and work with your site personnel to make the proper modifications to the system. CAI can provide startup and commissioning, as well as certification assistance for the system. What's my take away? There are a number of refineries in different EPA regions that have been EPA approved to operate their flare monitoring systems using an Alternative Monitoring Plan and this same alternative may be applicable to your plant and your flare monitoring application. AMP approval will allow you to operate your flare gas monitoring system using cal gas validation techniques that eliminate the need for dangerous, high concentration H 2 S calibration gases in your flare monitoring systems. This application of Mass Spectrometry to your flare application will also allow you to measure speciated hydrocarbons and BTU, and characterize your flare composition more accurately. Control Analytics has helped other clients engineer and modify their systems, and can support you should you choose implement this approach. Control Analytics, Inc Page 8

9 Control Analytics, Inc. is an analyzer engineering and system integrator of on line analyzers and systems for industrial use. Our product specialties include Continuous Emission Monitoring Systems (for EPA reporting), Flare Gas Monitoring Systems, on line process control monitoring, analyzer engineering, sample system design, and instrumentation enclosure and shelter integration. For more information on how we can assist in reducing H2S Calibration Gases at your facility, schedule a site walk down or any other process monitoring questions you may have, please contact: Control Analytics, Inc Enterprise Drive Export, PA Phone: (724) Fax: (724) Inside Sales Control Analytics, Inc Page 9

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1 Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

APPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6

APPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6 APPENDIX C 40 CFR PART 60 App. B SPEC. 1-4, 6 Part 60, Appendix B - Performance Specification 2 Specifications and Test Procedures for SO2 and NOx Continuous Emission Monitoring Systems in Stationary Sources

More information

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology: Title: Comparison of Chemical Sensor and NDIR Technologies TSN Number: 25 File:\\MII- SRV1\Metron\Bridge_Analyzers\Customer_Service_Documentation\White_Papers\25_G en EGA NDIR vs Chemical Sensor.docx Created

More information

METHOD 3A - DETERMINATION OF OXYGEN AND CARBON DIOXIDE CONCENTRATIONS IN EMISSIONS FROM STATIONARY SOURCES (INSTRUMENTAL ANALYZER PROCEDURE)

METHOD 3A - DETERMINATION OF OXYGEN AND CARBON DIOXIDE CONCENTRATIONS IN EMISSIONS FROM STATIONARY SOURCES (INSTRUMENTAL ANALYZER PROCEDURE) METHOD 3A - DETERMINATION OF OXYGEN AND CARBON DIOXIDE CONCENTRATIONS IN EMISSIONS FROM STATIONARY SOURCES (INSTRUMENTAL ANALYZER PROCEDURE) 1.0 Scope and Application What is Method 3A? Method 3A is a

More information

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the 1151 METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS 1.0 Scope and Application. 1.1 Analytes. Analyte Volatile Organic Compounds (VOC) CAS No. No CAS number assigned 1.2 Scope. This method

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD) Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet

More information

Protocol Gas Verification For Compressed Gas Cylinders Containing Either SO 2, NO or CO. Quality Assurance Plan/Standard Operating Procedure

Protocol Gas Verification For Compressed Gas Cylinders Containing Either SO 2, NO or CO. Quality Assurance Plan/Standard Operating Procedure Page 1 of 4 Protocol Gas Verification For Compressed Gas Cylinders Containing Either SO 2, NO or CO Quality Assurance Plan/Standard Operating Procedure Page 2 of 4 Table of Contents 2.3.6 Protocol Gas

More information

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1250 METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Total Organic Compounds N/A

More information

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle,

More information

APPLICATION NOTE. GC Integrated Permeation Device

APPLICATION NOTE. GC Integrated Permeation Device GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords:

More information

ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS

ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS XIX IMEKO World Congress Fundamental and Applied Metrology September 6, 9, Lisbon, Portugal ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS Ana Madeira, Florbela A. Dias and Eduarda Filipe 3 Instituto Português

More information

Verification Of Calibration for Direct-Reading Portable Gas Monitors

Verification Of Calibration for Direct-Reading Portable Gas Monitors U. S. Department of Labor Occupational Safety and Health Administration Directorate of Science, Technology and Medicine Office of Science and Technology Assessment Verification Of Calibration for Direct-Reading

More information

DEPARTMENT OF ENVIRONMENTAL PROTECTION Air Quality. 35 P.S. Sec (Air Pollution Control Act) and Act 95 (75 PA C.S.

DEPARTMENT OF ENVIRONMENTAL PROTECTION Air Quality. 35 P.S. Sec (Air Pollution Control Act) and Act 95 (75 PA C.S. DOCUMENT NUMBER: 274-0300-001 DEPARTMENT OF ENVIRONMENTAL PROTECTION Air Quality TITLE: AUTHORITY: POLICY: PURPOSE: Continuous Source Monitoring Manual 35 P.S. Sec. 4001-4005 (Air Pollution Control Act)

More information

Reduce Turnaround Duration by Eliminating H 2 S from Flare Gas Utilizing VaporLock Scrubber Technology

Reduce Turnaround Duration by Eliminating H 2 S from Flare Gas Utilizing VaporLock Scrubber Technology Reduce Turnaround Duration by Eliminating H 2 S from Flare Gas Utilizing VaporLock Scrubber Technology Jim Woodard Vice President of Sales Bryant Woods Lead Project Engineer 4/3/18 Page 1 Objectives Impact

More information

Standard Operating Procedure for H 2 S and SO 2 Converter

Standard Operating Procedure for H 2 S and SO 2 Converter Procedure No: SOP-001 Revision No: 1.0 January 24, 2011 Page No.: 1 of 6 1. INTRODUCTION AND SCOPE This procedure describes the operation and configuration of heated converters used in continuous ambient

More information

MMS Mercury Monitoring System for Natural Gas

MMS Mercury Monitoring System for Natural Gas MMS Mercury Monitoring System for Natural Gas Natural gas often contains mercury at concentrations that vary from below 1 to above 10000 µg/m 3. Mercury is both toxic and potentially damaging: it can corrode

More information

Gas Mixture Two Components. Gas Mixtures Two Components

Gas Mixture Two Components. Gas Mixtures Two Components Gas Mixture Two Components Two Components Contents Two-Component Mixtures Ammonia...M4 Argon...M5 Benzene...M6 n-butane...m7 Carbon Dioxide...M8 Carbon Monoxide...M10 Chlorine...M12 Ethane...M13 Ethylene

More information

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide

More information

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR) Procedure No: SOP-025 Revision No: 1.0 (January 21, 2011) Page No.: 1 of 8 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to

More information

Permitted MSS Emissions Tracking, Recordkeeping, and Reporting. Presented by ACES April 24, 2008

Permitted MSS Emissions Tracking, Recordkeeping, and Reporting. Presented by ACES April 24, 2008 Permitted MSS Emissions Tracking, Recordkeeping, and Reporting Presented by ACES April 24, 2008 1 Agenda Introductions Overview of Current MSS Permitting Program Status Discussion of Model MSS Permit Requirements

More information

Sr. No. General points CPCB Comments 1. Print option, acknowledgement of submission for CPCB OCEMS Protocol

Sr. No. General points CPCB Comments 1. Print option, acknowledgement of submission for CPCB OCEMS Protocol Dated: - 18-06-2018 General Points Discussed with CPCB on OCEMS Protocol Sr. No. General points CPCB Comments 1. Print option, acknowledgement of submission for CPCB OCEMS Protocol Agreed. After final

More information

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER Michael T Tang, Ph.D. Grace Feng Greg Merideth Rui Huang Matheson Gas Applied Lab

More information

Proposed Uniform Standards for Equipment Leaks

Proposed Uniform Standards for Equipment Leaks 12 th LDAR Symposium May 15-16, 2012 New Orleans, LA Proposed Uniform Standards for Equipment Leaks Standards Certification Education & Training Publishing Conferences & Exhibits 1 Proposed LDAR Uniform

More information

HiQ Specialty Gases and Specialty Equipment. Precision matters. In everything we do.

HiQ Specialty Gases and Specialty Equipment. Precision matters. In everything we do. HiQ Specialty Gases and Specialty Equipment Precision matters. In everything we do. 02 Introduction Petrochemical refinery complex Our industry is highly complex. But our vision is simple. We provide precisely

More information

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring.

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring. Product Data Sheet CODEL Continuous Emission Monitoring GCEM4100 Extractive Gas Analyser CO, CO2, NO, NO2, SO2, HCL, CH4 & H2O Hot extractive multi-channel gas analyser for continuous process and emissions

More information

Trace Environmental Systems

Trace Environmental Systems Trace Environmental Systems Providers of Continuous Stack Monitoring Systems www.traceenv.com Trace Overview Jim Toolen Sales Manager/Co-Owner 201-670-7077 jtoolen@traceenv.com Trace Overview Kevin Ramazan

More information

PIQCS HACCP Minimum Certification Standards

PIQCS HACCP Minimum Certification Standards PIQCS HACCP Minimum Certification Standards In the EU, requirements for the hygiene of food is laid down in Regulation (EC) 852/2004. This regulation establishes general hygiene procedures for food at

More information

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring.

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring. Product Data Sheet CODEL Continuous Emission Monitoring GCEM4100 Extractive Gas Analyser CO, CO2, NO, NO2, SO2, HCL, CH4 & H2O Hot extractive multi-channel gas analyser for continuous process and emissions

More information

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality

More information

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015 page 1 of 6 TNI Standard EL-V1M4 Sections 1.7.1 and 1.7.2 September 2015 Description This TNI Standard has been taken through all of the voting stages and has received consensus approval by the TNI membership.

More information

The Differences Between Part 60, and 75 Regulations

The Differences Between Part 60, and 75 Regulations ESC SPRING USERS GROUP The Differences Between Part 60, and 75 Regulations Taking energy to heart. Applicability 2 Part 60 regulations mostly apply to Limit monitoring Subparts D, Da, and Db Often required

More information

HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION

HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION SCENTROID is the global leader in sensory technology providing intelligent environmental solutions across a wide range of industries

More information

Thermo Scientific Model 146i Principle of Operation

Thermo Scientific Model 146i Principle of Operation Thermo Scientific Model 146i Principle of Operation The Model 146i Multi-gas Calibrator dilutes calibration gases to precise concentrations. The diluted gases are used to perform zero, precision and Level

More information

The Gold Standard in Calibration Mixtures

The Gold Standard in Calibration Mixtures The Gold Standard in Calibration Mixtures Calibration standards are used in a wide variety of industries. The need for accurate and precise standards has become increasingly important because of the consequences

More information

LFE OEM TCD - Thermal Conductivity Detector

LFE OEM TCD - Thermal Conductivity Detector LFE OEM TCD - Thermal Conductivity Detector Key features Quick response - T90 3 sec Extremely suppressed ranges High corrosion resistance High temperature capability up to 180 C Infallible containment

More information

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Outline of Presentation Introduction Process Gas Analysis Magnetic Sector Mass Spectrometer Standard

More information

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi

More information

Vehicle and Engine Emissions Testing

Vehicle and Engine Emissions Testing Vehicle and Engine Emissions Testing Calibration gases and equipment for compliance Benefits and Features Guaranteed analytical accuracy facilitates precise instrument calibration to ensure Part 1065,

More information

Gases and welding distributors continue

Gases and welding distributors continue Specialty Gas Instrumentation By Ray Borzio, Mike Lee, and Scott Varian Gases and welding distributors continue to have a significant future in the production, delivery, and related services for specialty

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

The revised XML schemas have been posted on the ECMPS Support Website.

The revised XML schemas have been posted on the ECMPS Support Website. I. Introduction The purpose of this document is to provide guidance for complying with the Emission Collection and Monitoring Plan System (ECMPS) reporting requirements related to the amendments to the

More information

HiQ laboratory gas generators.

HiQ laboratory gas generators. Product information HiQ laboratory gas generators. The convenient alternative for laboratory environments. 02 HiQ laboratory gas generators HiQ laboratory gas generators. Designed with safety and convenience

More information

IDEM Indiana Department of Environmental Management Office of Land Quality P.O. Box 6015 Indianapolis, IN OLQ PH: (317)

IDEM Indiana Department of Environmental Management Office of Land Quality P.O. Box 6015 Indianapolis, IN OLQ PH: (317) IDEM Indiana Department of Environmental Management Office of Land Quality P.O. Box 6015 Indianapolis, IN 46206-6015 OLQ PH: (317) 232-8941 Information SUBPART CC AIR EMISSION STANDARDS (40 CFR 264.1080

More information

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy. Bulletin Category: GAS Bulletin: G-23 Page: 1 of 7 The copy of this document located on Measurement Canada s website is considered to be the controlled copy. Recognition of data from specialty gas suppliers

More information

CO, NO, NO 2, NO x, SO 2, CH 4, HCl, CO 2 & H ² O. Extractive low cost, low maintenance devices for continuous process and emissions monitoring.

CO, NO, NO 2, NO x, SO 2, CH 4, HCl, CO 2 & H ² O. Extractive low cost, low maintenance devices for continuous process and emissions monitoring. Product Data Sheet CODEL Continuous Emission Monitoring GCEM 40 Series Extractive Gas Analyser CO, NO, NO 2, NO x, SO 2, CH 4, HCl, CO 2 & H ² O Extractive low cost, low maintenance devices for continuous

More information

HCL MONITORING. ICAC MARAMA WORKSHOP May 18-19th, Presented by: Ty Smith. CEMTEK

HCL MONITORING. ICAC MARAMA WORKSHOP May 18-19th, Presented by: Ty Smith. CEMTEK HCL MONITORING ICAC MARAMA WORKSHOP May 18-19th, 2011 Presented by: Ty Smith. CEMTEK 1 HCL MONITORING Applications Existing Measurement Technologies New Technologies Calibration and issues 2 APPLICATION

More information

Protocol Gas Verification Program (PGVP)

Protocol Gas Verification Program (PGVP) Protocol Gas Verification Program (PGVP) Michael Hayes ICAC/ MARAMA Workshop Advances in Control Technologies July 9-10, 2008 Exton, PA PGVP Protocol Gases from an historic perspective Historical Audit

More information

Nitrogen subtraction on reported CO 2 emission using ultrasonic flare gas meter

Nitrogen subtraction on reported CO 2 emission using ultrasonic flare gas meter Nitrogen subtraction on reported CO 2 emission using ultrasonic flare gas meter Kjell-Eivind Frøysa, Christian Michelsen Research AS, Bergen, Norway Henning Ekerhovd, StatoilHydro ASA, Kollsnes, Norway

More information

GCMSD-Headspace Analysis SOP

GCMSD-Headspace Analysis SOP Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to

More information

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT APPLICATION NOTE Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT The INFICON 3000 Micro GC provides two fast and accurate solutions for the analysis of coal mine gas components.

More information

VIC offers a variety of Calibrated Gas Leaks

VIC offers a variety of Calibrated Gas Leaks CALIBRATED GAS LEAKS VIC offers a high quality line of calibrated gas leaks traceable to National Institute of Standards Technology (NIST). Calibrated gas leaks are used to provide a controlled delivery

More information

TEMPLE UNIVERSITY ENVIRONMENTAL HEALTH AND RADIATION SAFETY

TEMPLE UNIVERSITY ENVIRONMENTAL HEALTH AND RADIATION SAFETY Page 1 of 9 ISSUED: 5/00 REVISED: 08/06 Introduction Purpose: In accordance with applicable regulations and Temple University, this policy was developed to minimize exposure to Ethylene Oxide. Applicability

More information

Multiple Gas#5 GC configuration Jan 2016

Multiple Gas#5 GC configuration Jan 2016 History: Unfortunately there is no single column that can separate: Hydrogen Oxygen Nitrogen Methane CO CO2 Ethane Water Propane Butane Pentane Over the years SRI Instruments has devised several solutions

More information

The Feasibility of Flare Gas Recovery for Ethylene Plants Dr. John Joyce, PhD, PE

The Feasibility of Flare Gas Recovery for Ethylene Plants Dr. John Joyce, PhD, PE The Feasibility of Flare Gas Recovery for Ethylene Plants Dr. John Joyce, PhD, PE 3/7/16 Page 1 Flare Gas Recovery Concept and technology for flare gas recovery (FGR) has bloomed (especially compressor

More information

It s Just Calibration How Hard Can It Be?

It s Just Calibration How Hard Can It Be? It s Just Calibration How Hard Can it Be? Deever Bradley, P.E. Environmental Resources Management Bart Leininger, P.E. Principal Ashworth Leininger Group Camarillo, CA 4/3/18 Page 1 Objectives Proper instrument

More information

Thermo Probe, Inc. TP-5 TP-7 TP-8. Instruction Manual English. Switch. Display. Carry Strap. Probe. Ground Clip. Switch. Display.

Thermo Probe, Inc. TP-5 TP-7 TP-8. Instruction Manual English. Switch. Display. Carry Strap. Probe. Ground Clip. Switch. Display. Display Thermo, Inc. Instruction Manual English TP-5 Display TP-7 Display TP-8 1-1 Thermo, Inc. Instructions for use- Models TP-5, TP-7, and TP-8 The manual describes basic function and use of Thermo instruments

More information

Sensepoint XCD. Calibration Guide

Sensepoint XCD. Calibration Guide Sensepoint XCD Calibration Guide Sensepoint XCD Calibration Calibration Handbook It is recommended to periodically carry out a gas response check on the Sensepoint XCD to ensure correct operation. This

More information

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The

More information

Generating Calibration Gas Standards

Generating Calibration Gas Standards Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and

More information

Analyzer Specifications Technical Bulletin

Analyzer Specifications Technical Bulletin 1. Introduction At Tiger Optics, we provide comprehensive specifications to give a detailed understanding of our instruments performance. Each analyzer goes through a rigorous qualification procedure to

More information

Every things under control High-Integrity Pressure Protection System (HIPPS)

Every things under control High-Integrity Pressure Protection System (HIPPS) Every things under control www.adico.co info@adico.co Table Of Contents 1. Introduction... 2 2. Standards... 3 3. HIPPS vs Emergency Shut Down... 4 4. Safety Requirement Specification... 4 5. Device Integrity

More information

3 HiQ Specialty Gases and Specialty Equipment. Precision matters in everything we do

3 HiQ Specialty Gases and Specialty Equipment. Precision matters in everything we do 3 HiQ Specialty Gases and Specialty Equipment Precision matters in everything we do 02 Introduction Introduction 03 Contents. 4-5 What we can do for you 6-7 Specialty gases products 8-9 How we measure

More information

Calibration Requirements for Direct Reading Confined Space Gas Detectors

Calibration Requirements for Direct Reading Confined Space Gas Detectors : Calibration Requirements for Direct Reading Confined Space Gas Detectors However, the definition of bump test has always been a little slippery. Some manufacturers differentiate between a bump test that

More information

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module TECHNICAL NOTE High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module TN42256 Dr. Liliana Krotz, Dr. Francesco Leone, Walter Galotta and Dr. Guido Giazzi

More information

Improve Process Reliability

Improve Process Reliability Convenient Conductivity Calibration to Meet Pharmacopeia Regulations Pharmacopeias require that conductivity measurement systems be periodically calibrated to ensure their precision. The range of accessories

More information

Finding leaks in Main Cryogenic Heat Exchangers without shutdown

Finding leaks in Main Cryogenic Heat Exchangers without shutdown Finding leaks in Main Cryogenic Heat Exchangers without shutdown Liquefaction is the most fundamental step in the LNG production process. To achieve this there are several process solutions available,

More information

Applications Note: Use of "pentane equivalent" calibration gas mixtures

Applications Note: Use of pentane equivalent calibration gas mixtures Introduction The gas that is used to verify accuracy is every bit as important as the detector itself when it comes to worker safety. Choosing (and using) the right mixture is critical to the success of

More information

Specialty Gases + World-Class Performance

Specialty Gases + World-Class Performance Specialty Gases + World-Class Performance A breakthrough in specialty gases. PurityPlus specialty gases deliver the purity you need plus the availability and service you demand. Purity. PurityPlus Specialty

More information

AUSTRALIA ARGENTINA CANADA EGYPT NORTH SEA U.S. CENTRAL U.S. GULF. SEMS HAZARD ANALYSIS TRAINING September 29, 2011

AUSTRALIA ARGENTINA CANADA EGYPT NORTH SEA U.S. CENTRAL U.S. GULF. SEMS HAZARD ANALYSIS TRAINING September 29, 2011 AUSTRALIA ARGENTINA CANADA EGYPT NORTH SEA U.S. CENTRAL U.S. GULF SEMS HAZARD ANALYSIS TRAINING September 29, 2011 Purpose The purpose of this meeting is to provide guidelines for determination of hazard

More information

December 16, I. Introduction

December 16, I. Introduction ECMPS Reporting Requirements Related to the Amendments to the Protocol Gas Verification Program and Minimum Competency Requirements for Air Emissions Testing I. Introduction The purpose of this document

More information

Field verification of on-line HC dew point measurements

Field verification of on-line HC dew point measurements Field verification of on-line HC dew point measurements Andy Benton, Technical Consultant, Michell Instruments Limited, Ely, UK andy.benton@michell.com www.michell.com Introduction: The Michell Condumax

More information

Rapid and Reliable Detection of Dissolved Gases in Water

Rapid and Reliable Detection of Dissolved Gases in Water Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,

More information

800.Keen.Gas

800.Keen.Gas C 2 H 2 ACETYLENE Acetylene is a colorless and tasteless gas with a garlic-like odor. It is flammable and can be an asphyxiant. In order to prevent explosive decomposition, acetylene is supplied in cylinders

More information

Ron Gibson, Senior Engineer Gary McCargar, Senior Engineer ONEOK Partners

Ron Gibson, Senior Engineer Gary McCargar, Senior Engineer ONEOK Partners New Developments to Improve Natural Gas Custody Transfer Applications with Coriolis Meters Including Application of Multi-Point Piecewise Linear Interpolation (PWL) Marc Buttler, Midstream O&G Marketing

More information

10 G Street, N.E. Suite 700 Washington, D.C

10 G Street, N.E. Suite 700 Washington, D.C 10 G Street, N.E. Suite 700 Washington, D.C. 20002 Environmental Protection Agency EPA Docket Center, Mailcode: 6102T 1200 Pennsylvania Avenue, NW Washington, DC 20460 Attention: Docket ID No. EPA-HQ-OAR-2006-0640

More information

Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program

Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program 1 Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program Jim Meacham, CTG Energetics Agenda 2 In Situ Testing Results Reduced Ventilation after

More information

ON736 Series: Oxygen/Nitrogen by Inert Gas Fusion

ON736 Series: Oxygen/Nitrogen by Inert Gas Fusion ON736 Series: Oxygen/Nitrogen by Inert Gas Fusion Delivering the Right Results 736 Series: Oxygen/Nitrogen by Inert Gas Fusion The 736 family of Elemental Analyzers is designed for routine measurement

More information

Equipment Accuracy / Selection

Equipment Accuracy / Selection Revision 5.0: Included the PPP program for clarification. Revision 6.0: Update references for CTL PDSH 251B to CTL DSH 251E" Added leakage current tolerances for values above 5KHz For Client Labs Purpose

More information

PERFORMANCE SPECIFICATION PROPELLANT, HYDROGEN

PERFORMANCE SPECIFICATION PROPELLANT, HYDROGEN METRIC 17 May 2013 SUPERSEDING MIL-PRF-27201D 07 February 2007 PERFORMANCE SPECIFICATION PROPELLANT, HYDROGEN This specification is approved for use by all Departments and Agencies of the Department of

More information

Spot the Hazards: Review Confined Space Hazards Missed by Most Safety Professionals, and the PPE Solutions that Meet OSHA Requirements

Spot the Hazards: Review Confined Space Hazards Missed by Most Safety Professionals, and the PPE Solutions that Meet OSHA Requirements Spot the Hazards: Review Confined Space Hazards Missed by Most Safety Professionals, and the PPE Solutions that Meet OSHA Requirements Why is Gas Detection Important? Spot the hazard Composition of Fresh

More information

(NIST Special Publication )

(NIST Special Publication ) (NIST Special Publication 260-126) THE NIST TRACEABLE REFERENCE MATERIAL PROGRAM FOR GAS STANDARDS by Franklin R. Guenther, William D. Dorko, Walter R. Miller and George C. Rhoderick Analytical Chemistry

More information

Subpart L NOTES: NOTES:

Subpart L NOTES: NOTES: Odorization of Gas Subpart L Wayne St. Germain Pipeline Safety Specialist Phone (405) 954 7219 Live Person (405) 954 8575 Direct Line Fax (405) 954 0206 Email wayne.stgermain@dot.gov Odorization of Gas

More information

The FLUX-METER: implementation of a portable integrated instrumentation for the measurement of CO 2 and CH 4 diffuse flux from landfill soil cover.

The FLUX-METER: implementation of a portable integrated instrumentation for the measurement of CO 2 and CH 4 diffuse flux from landfill soil cover. : implementation of a portable integrated instrumentation for the measurement of CO 2 and CH 4 diffuse flux from landfill soil cover. I. Minardi*, E. Giovenali*, L. Coppo*, G. Virgili*, D. Continanza*,

More information

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson A Journal of Practical and Useful Vacuum Technology From By Phil Danielson Thermal Conductivity Gauges Thermal conductivity pressure gauges are extremely common in vacuum technology, but an understanding

More information

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES PROCESS ANALYSERS DF-310E The DF-310E is a Coulometric sensor based oxygen analyzer designed to measure trace and percent level oxygen in pure and multi-gas backgrounds for process and quality control

More information

Instruction and Maintenance Manual

Instruction and Maintenance Manual Instruction and Maintenance Manual GRYF OXY ZM 02/100/2 E Contact GRYF HB, spol. s r.o. Cechova 314 Havlickuv Brod 580 01 tel.: +420 569 426 627 fax: +420 569 426 627 www.gryf.eu ver.: 11.2.2015 OEM module

More information

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience - Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience - Dr. Wolfgang Jockel, Martin Schneider, TÜV Rheinland Group, D-51105 Cologne / Germany 1. Introduction A new basis for the certification

More information

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV 16-1 INTRODUCTION 16-1.1 Purpose. The purpose of this chapter is to familiarize divers with the techniques used to mix divers breathing gas. 16-1.2 Scope. This chapter

More information

M-06 Nitrogen Generator (Nitrogen Making Machine)

M-06 Nitrogen Generator (Nitrogen Making Machine) Guideline No.M-06 (201510) M-06 Nitrogen Generator (Nitrogen Making Machine) Issued date: 20 th October, 2015 China Classification Society Foreword This Guideline is a part of CCS Rules, which contains

More information

MIL-STD-883G METHOD

MIL-STD-883G METHOD STEADY-STATE LIFE 1. PURPOSE. The steady-state life test is performed for the purpose of demonstrating the quality or reliability of devices subjected to the specified conditions over an extended time

More information

Electrical Solutions Corporation 2368 Eastman Ave, Suite 13, Ventura, CA Phone (805) ~ Fax (805)

Electrical Solutions Corporation 2368 Eastman Ave, Suite 13, Ventura, CA Phone (805) ~ Fax (805) Gas Chromatograph Operation Gas Company GC Initial setup - 6 Feb 2007 The Gas Company Gas Chromatograph (GC) data now come into Wonderware and can be accessed from the Gas Chrom button on the Wonderware

More information

Autocalibration Systems For In Situ Oxygen Analyzers

Autocalibration Systems For In Situ Oxygen Analyzers Product Data Sheet PDS 106-340AC SPS 4000, IMPS 4000, and MPS 3000 January, 1999 Autocalibration Systems For In Situ Oxygen Analyzers Cost-effective autocalibration systems for installations ranging from

More information

Instrument Performance Evaluation, comparing ISO and NORSOK I-106 Dr Paul Holland, EffecTech Group & Jon Carlsen, Norsk Analyse North Sea Flow

Instrument Performance Evaluation, comparing ISO and NORSOK I-106 Dr Paul Holland, EffecTech Group & Jon Carlsen, Norsk Analyse North Sea Flow Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech Group & Jon Carlsen, Norsk Analyse North Sea Flow Measurement Workshop 2017 Overview of presentation Importance

More information

Continuous Gas Analysis In situ laser gas analyzers TÜV and MCERTS add-on for LDS 6 operating instructions Compact Operating Instructions

Continuous Gas Analysis In situ laser gas analyzers TÜV and MCERTS add-on for LDS 6 operating instructions Compact Operating Instructions Continuous Gas Analysis In situ laser gas analyzers Compact Operating Instructions 1 LDS 6 TÜV/MCERTS certified analyzers for NH3/H2O, NH3 and H2O 1.1 Operation requirements Please observe the following:

More information

PENNSYLVANIA DEPARTMENT OF ENVIRONMENTAL PROTECTION

PENNSYLVANIA DEPARTMENT OF ENVIRONMENTAL PROTECTION Page 1 of 9 PENNSYLVANIA DEPARTMENT OF ENVIRONMENTAL PROTECTION COMPLIANCE DEMONSTRATION INSTRUCTIONS FOR THE CATEGORY NO. 38 AIR QUALITY PERMIT EXEMPTION CRITERIA The Category No. 38 exemption criteria

More information

PERFORMANCE SPECIFICATION PROPELLANT PRESSURIZING AGENT, ARGON

PERFORMANCE SPECIFICATION PROPELLANT PRESSURIZING AGENT, ARGON METRIC 07 August 2013 SUPERSEDING MIL-PRF-27415B 08 February 2007 PERFORMANCE SPECIFICATION PROPELLANT PRESSURIZING AGENT, ARGON Comments, suggestions, or questions on this document should be addressed

More information

Plasma. Argon Production Helium Liquefaction Truck Loading Pure Gas Bottling

Plasma. Argon Production Helium Liquefaction Truck Loading Pure Gas Bottling SERVOPRO Plasma The SERVOPRO Plasma is a Plasma Emission Detector based analyser designed to continuously measure Nitrogen in pure Argon for process and quality control in Argon production within Cryogenic

More information