Porter Conscious Sedation Flowmeters and Bag Tee (Accessory)

Size: px
Start display at page:

Download "Porter Conscious Sedation Flowmeters and Bag Tee (Accessory)"

Transcription

1 Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA Office Fax Porter Conscious Sedation Flowmeters and Bag Tee (Accessory) Instructions and Safety Considerations MXR MODELS / REF: C3000 C3050 DTL-146W DTL-164W Bag Tee Accessory: P1407E This product complies with the Medical Device Directive (93 / 42 / EEC). A Declaration of Conformity in accordance with the directive has been made and is on file. European Communities should contact the Authorized Representative listed below regarding any Medical Device Directive (MDD) inquiries. Contact Name: Mailing Address: Phone: Fax: Parker Hannifin Ltd Instrumentation Products Division Riverside Road, Pottington Business Park Barnstaple, EX 31 1NP, England +44 (0) (0) The Quality System for Porter Instrument is certified to ISO The scope of our registration is: The design, manufacture, distribution and servicing of Nitrous Oxide Oxygen Sedation Flowmeters, Gas Scavenging Systems, Gas Distribution Systems and Office Communication Systems for use by a physician, dentist or licensed healthcare Professional. Check our website: for additional information. To register your product: choose Warranty tab. To download a User s Manual: choose Manuals tab. FM-825-CE-English Rev. L 4/2013

2 IMPORTANT: READ MANUAL COMPLETELY BE F O R E O P E R AT I N G T H I S D E V I C E Basic delivery technique is described. Also, this manual contains instructions on periodically required checks to be performed by the user. These checks are necessary to insure the proper performance of this device and its safety features. Retain this manual for future reference. W AR N I N G S AN D P R E C AU T I O N S These warnings and precautions are to help you to understand how to safely operate the MXR Flowmeter. A WARNING alerts you to a possible hazard to people. A CAUTION alerts you to the possibility of equipment damage. WARNING: Do not use this device for the administration of general anesthesia or as a part of, or in conjunction with, a general anesthesia administration system. WARNING: Workers are exposed to N 2 O during administration of N 2 O/O 2 conscious sedation analgesia. NIOSH has recommended that exposures should be minimized. Contact NIOSH ( NIOSH) to receive NIOSH Publications on Control of Nitrous Oxide in Dental Operatories. Exposure can be minimized by effective controls. National Institute for Occupational Safety and Health (NIOSH) publications state that controls, including System Maintenance, Ventilation and Work Practices can effectively reduce N 2 O concentrations in patient procedures. Your flowmeter accessory Porter scavenger system is an important part of the system of controls. WARNING: Porter Instrument equipment utilizes the cross+protection system. The flexible hose and connectors that connect to the flowmeter are diameter indexed; 3/8 O.D. for Nitrous Oxide and ½ for Oxygen. The cross+protection system is designed to prevent misconnection of Oxygen and Nitrous Oxide piping. DO NOT ATTEMPT TO CHANGE THE DIAMETERS OR CONNECTORS OF THE FLOWMETER! Tampering with the cross+protection system constitutes acceptance of liability by the installer. For your own protection, as well as that of the Doctor and the patients, use 3/8 O.D. tubing for all Nitrous Oxide lines and ½ O.D. tubing for all Oxygen lines. To assure safe operation and conformation to local fire codes, all Porter Instrument flowmeter systems are designed to be used with sedation delivery systems mounted inside walls and they meet or exceed the guidelines established by the National Fire Protection Association for Nonflammable Medical Gas Systems, NFPA 99. Copies of NFPA 99 or portions thereof may be obtained by writing to: National Fire Protection Association, Batterymarch Park, Quincy, MA ; or call WARNING: New or modified installations - properly connected gas pipelines are absolutely essential to patient safety. The authorized distributor or contractor should provide written documentation that all gas pipelines are connected properly and that the system has been pressure tested prior to use. While this is a good business practice, it is important that the user verify by their own test, independent of the authorized distributor or contractor, that all gas pipelines are connected correctly prior to using the system. The ultimate responsibility of assuring that lines are not crossed rests with the user. WARNING: During any power outage, remember to turn OFF the flowmeter and manually turn OFF the tank valves. With centralized, electrically powered gas systems, if gas was flowing when the power went out and the flowmeter is left ON, gas will be flowing when the power is restored. CAUTION: This device is for use by or on the order of a physician, dentist or licensed healthcare professional. CAUTION: Always use clean, dry medical grade gases. Introduction of moisture or other contaminants into this device may result in defective operation. CAUTION: Do not attempt to repair, alter or calibrate this device. Unauthorized repair, alteration or misuse of this device is likely to adversely affect the performance and will void the warranty. CAUTION: Never oil or grease any part of this system (minimize fire or explosion potential). 1

3 FIGURE 1 FLOWMETER FEATURES Oxygen Flowmeter Tube indicates the f low of O 2 in L/min ±5%. 2. Nitrous Oxide Flowmeter Tube indicates flow of N 2 O in L/min ±5%. 3. Nitrous Oxide Failsafe System. Duals e a l o x y g e n - p iloted v a lve s y s t e m t h a t a u t o m a t ically m a intains t h e % N 2 O c o n c e n t r a t ion s e t t ing with a n y c h a n g e in t h e O 2 f low o r p r e s s u r e. N 2 O f low is p r o p o r t ionately r e d u c e d if O 2 is s h u t o f f o r t h e p r e s s u r e is r e d u c e d ( c h e c k b e f o r e e a c h u s e ). 4. Flow Control Knob c o n t r o ls t h e c o m b ined f low o f O 2 a n d N 2 O ( o r O 2 f l o w o n ly when t h e Concentra t ion Control Knob is s e t t o 0 %). 5. Concentration Control Knob c o n t r o ls t h e % c o n c e n t r a t ion o f N 2 O b u t d o e s n o t c h a n g e t h e O 2 f low r a t e. ( 7 0 % M a x imum N 2 O d e livery.) ( Check Monthly) 6. Emergency Air Valve a u t o m a t ically p r o v ides t h e p a t ient with a m b ient a ir if g a s f low is interrupted. (Check Monthly) 7. Non-Rebreathing Valve g u a r d s a g a inst CO2 b u ild-up a n d r e b r e a t h ing o f u s e d g a s e s. ( Check Monthly) 8. Positive On - Off Switch Reduces t h e p o s s ibility o f a c c identally leaving t h e m a c h ine O N. 9. Power Flush p r o v ides e x t r a d e livery o f O 2 to the breathing bag. (Check Monthly) 10. O 2 Quick Connect ( left s ide o f m a c h ine), f a c ilitates c o n n e c t ion o f p o s itive p r e s s u r e / d e m a n d v a lve for emergency Oxygen. 11. Bag Tee Outlet c o n n e c t s t o f r e s h g a s t u b ing L and 2L Bag Connection 2

4 INTENDED USE: Flowmeter for Analgesia Delivery System This device is intended for patient use by an attending physician, dentist or licensed professional properly trained in its use. Porter Instrument recommends the user be thoroughly familiar with the use of Nitrous Oxide - Oxygen Conscious Sedation for patient analgesia and be properly trained in its administration prior to using this product. For training requirements on the administration of Nitrous Oxide - Oxygen Conscious Sedation, contact the appropriate regulatory authority in your country, state, or province. Training is recommended to provide a practical, hands-on capability and an understanding of the behavioral aspects of Nitrous Oxide Sedation and will complement the safety features of this device. USE SCAVENGING Monitor for N 2 O in the patient treatment area to insure that controls are effective in achieving low levels of PPM (parts per million) exposure. Contact your Porter dealer for details on monitors and testing. GAS SUPPLY CONNECTION After installation of the flowmeter, connect the nitrous oxide and oxygen supply lines to the Diameter Indexed Safety System (DISS) fittings located on the back of the flowmeter unit. It is important that the regulators for both gases be set to give pressures in the range of 53 PSIG ±2 PSIG. Confirm the absence of leaks at pressure connections on the unit. Bubbles will appear at leaking locations when a soap / water solution is used. This procedure is recommended each time a cylinder is changed. MONTHLY LEAK CHECK (or if connections to flowmeter are disconnected and then reconnected, such as after a flowmeter service at the factory or if the DISS/DISS hose is replaced): Leak test the flowmeter system for working pressure leaks. After all hose connections are tightened, turn both control knobs to the off position and the on/off switch to the off position. Confirm that the DISS Shut-Off Valves are in the open position. Pressurize the sedation gas supply lines with 50 PSI. Observe any pressure decay after an overnight time period (5 PSI drop allowed). Monitor O 2 gas pressures at the beginning of each procedure to assure sufficient gas remains in the tank to complete the procedure. SAFETY FEATURE CHECKS IMPORTANT: These are safety features, which you should routinely check to assure proper function. If any of these safety features are not functioning properly, contact your authorized distributor or Porter Instrument and arrange for the necessary repairs. Porter Instrument recommends the repairs be made before reusing the device. NITROUS OXIDE FAILSAFE SYSTEM CHECK BEFORE EACH USE 1. Set the right-hand total Flow Control Knob to zero and set the left-hand Concentration Control Knob to zero. (See Figure 1-Items 4 and 5: Rotate control knobs to stop.) 2. Be sure O 2 and N 2 O are connected to your MXR and line pressure for both gases is 50 PSIG (which is standard). 3. Turn the Concentration Control Knob to 50%. There should be no flow of N 2 O. This is a check of the static position of the Nitrous Oxide Failsafe System valve. NOTE: A momentary low flow of N 2 O (about 1 L/min for about a second) may be seen if the Concentration Control Knob is turned to zero before turning the Flow Control Knob to zero. This is N 2 O gas trapped between the Nitrous Oxide Failsafe System valve and the Concentration Control Knob valve and is a normal occurrence. No other N 2 O flow should be observed. 4. With the Concentration Control Knob still set to 50%, turn the Flow Control Knob to achieve a flow of 3 to 4 L/min of O 2. You should observe an equal amount of N 2 O flowing by gradually turning the Flow Control Knob. (Refer to Monthly Calibration Check on Page 4, to read ball float on flowmeter tubes.) 5. Interrupt the flow of O 2. This will check the dynamic status of the Nitrous Oxide Failsafe System valve. This can be done by either disconnecting the oxygen hose from the wall or shutting off the oxygen at the tank. The Nitrous Oxide flow should drop as the Oxygen flow decreases, stopping completely before the Oxygen float drops to zero. WARNING If the Nitrous Oxide Failsafe System fails to perform as indicated, do not use this product prior to repair. Improper function of this safety feature may permit Nitrous Oxide to flow independently of the flow control knob, potentially allowing Nitrous Oxide to flow to the patient without Oxygen. 3

5 CONCENTRATION CONTROL KNOB VALVE CHECK MONTHLY Set the Concentration Control Knob to 50%, and Flow Control Knob to 2 to 3 L/min. The ball indicators will be at about the same height (If not, refer to Monthly Calibration Check section for the procedure on checking the accuracy of the meter.). Turn the Concentration Control Knob to zero. The N 2 O flow should drop to zero. You essentially perform the check at the end of every procedure when you oxygenate the patient with 100% O 2. POWER OXYGEN FLUSH CHECK MONTHLY (O 2 SUPPLY PRESSURE AT 50 PSIG) Disconnect the corrugated rubber tubing from the bag tee outlet (Figure 1, Item 11). With both control knob valves OFF, depress the power (O 2 ) flush button while blocking the flow from the front of the bag tee. For proper operation, the gas reservoir bag should fill within about 5 seconds. Also test for bag / rubber goods leak following steps 1 through 6. Quick Test to Check Reservoir Bag / Rubber Goods for Leaks 1. With the flowmeter, bag tee and Porter rubber goods in place, remove the nosepiece and one of the two plastic connectors from the Porter rubber goods. 2. With the other plastic connector, join the two duplex hoses together making a closed system. 3. Taking care not to fill the bag too much (bag could burst), open the oxygen control valve until the reservoir bag starts to over-inflate or balloon, then close the valve. 4. Observe the reservoir bag for five minutes. 5. The bag should stay inflated. If so, the test has been successful and there are no excessive leaks. If the bag does not stay inflated, the reservoir bag or rubber goods have an excessive leak. Replace any parts that leak and retest until results are successful. 6. Disconnect one of the duplex hoses from the plastic connector and re-install the nosepiece. BAG TEE The Bag Tee assembly features a Non-rebreathing Valve and an Emergency Air Intake located on the Bag Tee (no carbon dioxide buildup). The emergency air intake has a valve that allows room air to be inhaled into the breathing circuit by the patient. A breathing bag is attached to the metal portion of the flowmeter. Bag Tee Installation to Flowmeter: Screw knurled seal down tight onto flowmeter making sure the rubber washer is inside the seal nut. When tight, the bag tee should not rotate. NON-REBREATHING VALVE CHECK MONTHLY With unit turned OFF, disconnect the corrugated rubber tubing from the Rubber Goods and breathe into the corrugated tubing connected to the bag tee. You should not be able to fill the bag with exhalation gas. If the bag fills, the system s Non-Rebreathing Valve is not functioning properly and should be replaced. EMERGENCY AIR VALVE CHECK MONTHLY With unit turned OFF, disconnect the corrugated rubber tubing from the Rubber Goods and draw air with your mouth through the corrugated tubing connected to the bag tee. You should be able to draw ambient air through the Emergency Air Valve (the gas bag may have to fully collapse first). Air going through the valve sounds different than normal gas flow. MONTHLY CALIBRATION CHECK The MXR Flowmeter is designed to maintain its accuracy and performance without routine user maintenance being required. The flowmeter tubes and ball floats are very resistant to accuracy changes over time such that the direct readings of the L/min on the scales maintain their accuracy. However, the user can check the relative accuracies of the % concentration and total flow valve system by performing a simple check. A calibration check of the % concentration can be done by setting the % concentration knob to 50% and the flow control knob to 3 to 4 L/min. Check to see if the tube readings are within 0.5 L/min of each other. Servicing is indicated if the readings are out of this tolerance. NOTE: Adequate and safe conscious sedation can be achieved even if the % concentration is outside of the listed tolerance, since tube scale accuracies are maintained. However, a change in the % concentration calibration is an indication of overall flowmeter condition. Porter servicing is available, including recalibration, pressure testing, internal component checking and replacement, and final factory testing by contacting your authorized distributor. It is advisable, on a two (2) year cycle, to have the MXR Flowmeter factory checked and serviced. 4

6 D I R E C T I O N S F O R U S E NOTE: These directions detail a basic delivery technique. However, this is not a comprehensive description and not a substitute for a training course that emphasizes a practical, hands-on approach together with instruction on safe administration techniques. 1. Maintain patient observation during procedure. 2. Turn ON unit by pushing in the ON / OFF switch. 3. Open N 2 O / O 2 tank valves. 4. Using Flow Control knob, set flow rate of O 2 to desired rate, keep bag about ¾ s full. Rotate flow control knob upwards (clockwise) to increase flow. (See Figure 1, Item 4) Flow Control Knob 7. When the procedure is nearing completion, amounts of N 2 O should be decreased. Terminate the flow of N 2 O and deliver 100% O 2 to begin a minimum postoxygenation period of 3 to 5 minutes. Assess the patient for appropriate recovery. Administer additional O 2 if necessary. Titration and post-procedure 100% O 2 will minimize nitrous exposure to the patient treatment area, potential patient side effects of lethargy, headache, or nausea, and any potential adverse effects of nitrous diffusion into air filled cavities. 8. When procedure is finally completed, turn off both control knob valves for gas shut off. 9. Place the ON / OFF switch (primary shut-off mechanism) in the OFF position. (Push from back of ON / OFF switch.) NOTE: If control valves are still open, gas flows should stop at this point. 10. Turn OFF the gas supply at the tank at the end of the day. 5. Set N 2 O concentration to desired level by rotating Concentration Control Knob (See Figure 1-Item 5) upwards (counterclockwise) to increase concentration, as read by percentages inscribed on the control knob. Rotate slowly until desired level is achieved. Practice titration* with 10% nitrous upward movements every 60 seconds until endpoint achieved. Patients may typically experience relief of anxiety, tingling in extremities, and euphoria. Patients typically require less than 50% nitrous. Concentration Control Knob Setting Shown at 60% 6. Flow Control knob may be re-adjusted to bring the total flow of gases back to desired level, when concentration is increased or decreased. Total flow is equal to the sum of right and left tube readings. Read center of ball float on flowmeter tube. BASIC DELIVERY TECHNIQUE: Practice titration. Titration is a method of administering a substance by adding definitive amounts of a drug until an endpoint is reached. For nitrous oxide / oxygen (N 2 O) / O 2 ) sedation, N 2 O is given in incremental doses until a patient has reached a comfortable relaxed state of sedation. The ability to titrate N 2 O is a significant advantage because it limits the amount of drug to that which is required by the patient. If titration is done properly, the patient does not receive any more of the drug than is necessary. The amount of N 2 O required by a patient on any given day or time varies. For information on titration, a most valuable resource for the practitioner is the Handbook of Nitrous Oxide and Oxygen Sedation, written by Clark and Brunswick and published by Mosby ( This text is a concise and contemporary guide for nitrous oxide / oxygen administration. IMPORTANT NOTE: When the Concentration Control knob is open, the Flow Control knob is closed, and there is no N 2 O flow indicated in the flow tube, the Nitrous Oxide Failsafe System will stop the flow of N 2 O. However, this safety feature should not be used as the primary shut off mechanism. The control knobs are for primary shut-off. 5

7 NOTE: If your unit does not operate as described in Steps 1 through 10 under Directions for Use, please contact your authorized distributor Porter Instrument. MAINTENANCE AND SERVICE It is advisable, on a two (2) year cycle, to have the MXR Flowmeter factory checked and serviced. Inspect and maintain the analgesia delivery system to prevent N 2 O leaks in all hoses, connections and fittings. Repair all leaks immediately. CLEANING METHODS We recommend the use of an approved disinfectant for the healthcare environment for cleaning the outside of the flowmeter. Do not spray disinfectant directly onto meter. Spray disinfectant into disposable towel and wipe unit thoroughly removing excess disinfectant to eliminate buildup. Control knobs can be autoclaved. Remove knobs from flowmeter. Clean using an appropriate disinfectant. Rinse with water then autoclave. Do not exceed 275 F for 15 minutes maximum. TROUBLESHOOTING CHART FOR MXR FLOWMETERS SYMPTOM POSSIBLE CAUSE REMEDY No f low o f O 2 o r N 2 O when O N / O F F s witch is O N a n d left k n o b is s e t a t a c o n c e n t r a t ion o f N 2 O o r t h e r ight k n o b is r o t a t e d t o g ive O 2 o r m ixed g a s f low. Can g e t O 2 f low b u t c a n n o t g e t N 2 O f low. With N 2 O c o n c e n t r a t ion s e t, b o t h f lows v a r y p r o p o r t ionally with no change in flow setting. M e t e r will f low N 2 O without a n y O 2 f low in the O 2 tube. Cannot g e t 9 ½ L / m in O 2 f low with c o n c e n t r a t ion c o n t r o l k n o b O F F a n d f low c o n t r o l k n o b f u ll O N. Ballooning of the gas bag. G a s is leaking f r o m t h e O N/OFF switch. Patient not feeling effects of g a s e s. Bag is going flat during p r o c e d u r e. 1. O2 s u p p ly not turned ON. 2. M a c h ine n o t c o n n e c t e d t o p ipeline system. 3. Empty O 2 c y linders T u r n O 2 r e g u lator of Cylinder Valve O N. 2. Connect to wall outlet. 3. Replace with full cylinder. 1. N2 O s u p p ly not turned ON. 2. N2 O c y linder empty. 1. T u r n O N N 2 O c y linder. 2. Replace with full cylinder. O 2 r e g u l a t o r is v a r y ing p ipeline Check O 2 r e g u lator. Be s u r e p r e s s u r e. O 2 m a n ifold p r e s s u r e is 5 0 PSIG ± 2 PSIG. I f n o t, c a ll Dealer for service. Failsafe failure. T a k e o u t o f s e r v ice a n d r e t u r n t o Porter. L o w O2 pressure setting. Check O 2 r e g u lator. Be s u r e O 2 m a n ifold p r e s s u r e is 5 0 PSIG ± 2 PSIG. I f n o t, c a ll Dealer for service. 9 0 Elbow c o n n e c t e d o n t o b a g Remove r u b b e r g o o d s a n d 9 0 t e e is p u s h e d o n t o o f a r, e lbow f r o m b a g t e e. Reconnect blocking movement of the non - e lbow and rubber goods. r e b r e a t h ing valve. Nick o r c u t in o rings o n t h e O N/OFF switch. 1. O u t e r m a s k is n o t f it p r o p e r ly to patient s face. 2. G a s f lows d o n o t m e e t p a t ient s requirement. 3. I n n e r m a s k s e x h a lation v a lve is missing. 4. I n n e r m a s k is m issing ( m u s t h a v e inner a n d o u t e r m a s k together). 1. G a s f lows d o n o t m e e t p a t ient s requirement. 2. O u t e r m a s k is n o t f it p r o p e r ly to patient s face. Call Dealer for service. 1. F it s o inner m a s k is s e c u r e t o f a c e b u t o u t e r m a s k is just o f f f a c e. 2. Re-adjust g a s f lows t o o b t a in acceptable bag action. 3. Replace valve. 4. Replace inner mask. 1. Re-adjust g a s f lows t o o b t a in acceptable bag action. 2. F it s o inner m a s k is s e c u r e t o f a c e b u t o u t e r m a s k is just o f f f a c e.

8 REF C3000 C3050 DTL-146W DTL-164W P1407E VARIANTS / COMPARED TO THE BASIC MODEL Basic Model MXR: 1-10 L/min O 2, Green Bands on flow tube, green buttons, 1-7 L/min N 2 O, Blue Bands on flow tube. 70% Max Concentration N 2 O Similar to basic Model MXR design, except, different color-coding for Oxygen white bands on flow tube and white buttons. Flowmeter body is painted white and has a failsafe block cover that is also painted white. The overall length of meter body and tube assemblies is reduced in length by approximately % Max Concentration N 2 O Similar to basic Model MXR design, except, different color-coding for Oxygen white bands on flow tube and white buttons. Flowmeter body is painted white and has a failsafe block cover that is also painted white. The overall length of meter body and tube assemblies is reduced in length by approximately % Max Concentration N 2 O Similar to basic Model MXR design, except, different color-coding for Oxygen white bands on flow tube and white buttons. Flowmeter body is painted white and has a failsafe block cover that is also painted white. The overall length of meter body and tube assemblies is reduced in length by approximately Also has special bag tee adapter. Note: The bag tee and breathing circuit accessories are supplied by the user. Similar to basic Model MXR design, except, different color-coding for Oxygen white bands on flow tube and white buttons. Flowmeter body is painted white and has a failsafe block cover that is also painted white. The overall length of meter body and tube assemblies is reduced in length by approximately Also has special Swedish connectors. 60% MXR Basic Model Bag Tee which includes a 22mm connector, serial number and warranty card with device. Check out our website for the following topics. Flowmeter Mounting Options Flowmeter Accessories Product W arranty and Registration Downloading User Manuals 7

PORTER INSTRUMENT COMPANY, INC.

PORTER INSTRUMENT COMPANY, INC. PORTER INSTRUMENT COMPANY, INC. MXR FLOWMETER Instructions and Safety Considerations MXR Models of Porter Conscious Sedation Flowmeters and Bag Tee (Accessory) Wall Mount Models: 2000 and 2000-OS, 3000

More information

STANDARD FLOWMETER USER S MANUAL

STANDARD FLOWMETER USER S MANUAL STANDARD FLOWMETER USER S MANUAL Instructions and Safety Considerations MXR Models of Porter Conscious Sedation Flowmeters and Bag Tee (Accessory) Flowmeter Models: 1000 1100 Bag Tee Model: P1407A Porter

More information

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions PORTER Parker Hannifin Corporation Porter Instrument Division 245 Township Line Rd. P.O. Box 907 Hatfield, PA 19440-0907 USA (215) 723-4000 / fax (215) 723-5106 Compact Triple Cabinet Outlet Station Model

More information

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions Compact Triple Cabinet Outlet Station Model 6258-1 Installation and Operating Instructions The Porter Compact Triple Outlet Station (6258-1) provides a quick, safe, and reliable method of connection to

More information

MXR FLOWMETER USER S MANUAL

MXR FLOWMETER USER S MANUAL MXR FLOWMETER USER S MANUAL Instructions and Safety Considerations MXR Models of Porter Conscious Sedation Flowmeters And Bag Tee (Accessory) Wall Mount Models: 2000 2000-OS 3000 3000-OS Bag Tee Models:

More information

DIGITAL MXR-D FLOWMETER USER S MANUAL

DIGITAL MXR-D FLOWMETER USER S MANUAL DIGITAL MXR-D FLOWMETER USER S MANUAL Instructions and Safety Considerations Cabinet Mount MXR-D Porter Conscious Sedation Flowmeter and Bag Tee (Accessory) FM-1061 Rev. B 4/11 IMPORTANT: READ MANUAL COMPLETELY

More information

Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions

Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions FM-171 Rev. B 8/00 Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions MODEL 6300-1 PORTER INSTRUMENT COMPANY, INC. 245 TOWNSHIP LINE RD. P.O. BOX 907 HATFIELD, PA 19440-0907

More information

2 CYLINDER MOBILE CART USER S MANUAL / INSTRUCTIONS

2 CYLINDER MOBILE CART USER S MANUAL / INSTRUCTIONS Parker Hannifin Corporation Precision Fluidics Division Porter Instrument 245 Township Line Road Hatfield, PA 9440 Office 25-723-4000/Fax 25-723-506 www.porterinstrument.com 2 CYLINDER MOBILE CART USER

More information

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS COMPACT TRIPLE WALL OUTLET STATION MODEL 6255-1 INSTALLATION AND OPERATING INSTRUCTIONS To assure safe operation and conformation to local fire codes, all Porter Outlet Stations are designed to be used

More information

MOBILE E TANK STAND INSTALLATION AND INSTRUCTIONS

MOBILE E TANK STAND INSTALLATION AND INSTRUCTIONS Parker Hannifin Corporation Precision Fluidics Division Porter Instrument 245 Township Line Road Hatfield, PA 19440 Office 215-723-4000/Fax 215-723-5106 MOBILE E TANK STAND INSTALLATION AND INSTRUCTIONS

More information

PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual

PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual ! WARNING IMPORTANT: READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE This manual contains instructions on periodically required checks to be

More information

Yoke Block Instruction Manual

Yoke Block Instruction Manual Yoke Block Instruction Manual ! WARNING IMPORTANT: READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE This manual contains instructions on periodically required checks to be performed by the user. These

More information

NITRONOX SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS

NITRONOX SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS NITRONOX SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS Examination All components should be removed from their respective shipping containers and examined. Any omissions and/or damage should be reported

More information

Yoke Block Instruction Manual

Yoke Block Instruction Manual Yoke Block Instruction Manual ! WARNING IMPORTANT: READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE This manual contains instructions on periodically required checks to be performed by the user. These

More information

MEDICAL BREATHING CIRCUIT SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS

MEDICAL BREATHING CIRCUIT SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS MEDICAL BREATHING CIRCUIT SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS Examination All components should be removed from their respective shipping containers and examined. Any omissions and/or damage

More information

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref Rev.

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref Rev. Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA 19440 USA Tel: 215-723-4000 Ref. 10049700 Rev. L Table of Contents Warnings...2 Examination...2 Description and

More information

Digital MDM Operations Manual

Digital MDM Operations Manual Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA 19440 Office 215 723 4000 Fax 215 723 5106 Digital MDM Operations Manual AA173200i This product complies with

More information

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified U S E R M A N U A L AIR FLOWMETER MODELS: 1MFA2001 (shown) 1MFA9001 ISO 13485 Certified Authorized Your EU local Representative: distributor: EMERGO EUROPE, INC. Molenstraat 15 2513866 BH -624 The -3952

More information

USER INSTRUCTIONS. S I L - C O N N - K I T S m a l l e r h o s e t o F l o w m e t e r L a r g e r h o s e t o V a c u u m C o n t r o l D e v i c e

USER INSTRUCTIONS. S I L - C O N N - K I T S m a l l e r h o s e t o F l o w m e t e r L a r g e r h o s e t o V a c u u m C o n t r o l D e v i c e USER INSTRUCTIONS FOR PORTER Silhouette D isposable Breath ing Circui t Silhouette -PD, -S M, - M D, - LG D ispos abl e Section R eusable Secti on S I L - C O N N - K I T S m a l l e r h o s e t o F l

More information

NITRONOX USER S MANUAL

NITRONOX USER S MANUAL NITRONOX USER S MANUAL Nitronox Field Unit Analgesia System www.porterinstrument.com/nitronox REF 10223000 Rev. J Safety Summary s and Precautions : The Nitronox Inhalation Analgesia apparatus is a specialized

More information

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator Operating Instructions Table of Contents Resource v.2 Conserving Regulator Safety Information Device Precautions Introduction Product Features Product Specifications Feature Illustrations Set Up Usage

More information

Rev. B. Operating Instructions

Rev. B. Operating Instructions 780745-00 Rev. B Operating Instructions 780745-00 Rev. B TABLE OF CONTENTS PAGE Warnings, Cautions, and Notes..............................4-7 Definition of Symbols......................................

More information

Flow meter. bellow vaporizer. APL valve. Scavenging system

Flow meter. bellow vaporizer. APL valve. Scavenging system Introductory Lecture Series: The Anesthesia Machine PORNSIRI WANNADILOK Objectives Anesthesia Machine Ventilators Scavenging Systems System Checkout 1 Flow meter ventilator bellow vaporizer Corrugated

More information

FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series SAVE THESE INSTRUCTIONS

FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series SAVE THESE INSTRUCTIONS USER MANUAL FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series 1MFA3001 (Shown) 8MFA1001 (Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

More information

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. U S E R M A N U A L 1600 SERIES OXYGEN REGULATOR 168715G (Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090

More information

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS TM controlled ventilation Equinox II NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E USER TABLE OF CONTENTS CHAPTER TITLE PAGE 1 INTRODUCTION 1.1 Equinox II Administration

More information

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series Introduction Powerex manifolds are cleaned, tested and prepared for the indicated gas service and are built in accordance with the Compressed Gas Association guidelines. The manifold consists of a regulator

More information

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

ACRYLIC FLOW METERS 0-30, 0-70 & LPM ACRYLIC FLOW METERS 0-30, 0-70 & 0-120 LPM ASSEMBLY INSTRUCTIONS & INSTRUCTIONS FOR USE R219P86 (0-120LPM) R219P87 (0-70LPM) R219P88 (0-30 LPM) R138P11 Rev. D TABLE OF CONTENTS: 1.0 PRODUCT OVERVIEW...1

More information

Dialflow Regulator. Instructions for Use

Dialflow Regulator. Instructions for Use Dialflow Regulator Instructions for Use 702-0030.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

AND OPERATION INSTRUCTIONS

AND OPERATION INSTRUCTIONS ADI 2111-D Certified ISO 9001 MEDICAL FLOWMETER (Back Pressure Compensated Thorpe Tube) INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls

More information

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200 Operators Manual Supera Floor Stand Anesthesia Machine Models: M1000 Basic Non-rebreathing M1200 Basic M1000 M1200 (All shown with optional equipment) Copyright Supera, LLC 2017 Examination and Preparation

More information

Digital MDM Mixer Operations Manual

Digital MDM Mixer Operations Manual Parker Hannifin Corporation Precision Fluidics Division Porter Instrument 245 Township Line Road Hatfield, PA 19440 Office 215 723 4000 Fax 215 723 5106 Digital MDM Mixer Operations Manual This product

More information

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER)

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER) DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER) 0-3, 0-15, 0-30 & 0-70 LPM ASSEMBLY INSTRUCTIONS & INSTRUCTIONS FOR USE R219P87-400 (0-70 LPM) R219P88-400 (0-30 LPM) R219P79-400 (0-15 LPM) R219P99-400

More information

M5400 Operators Manual

M5400 Operators Manual M5400 Operators Manual For Veterinary Use Only Supera, LLC Copyright 2012 Examination and Preparation for Use Thank you for ordering an Supera Anesthesia Innovations anesthesia machine! We are delighted

More information

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION DUAL SUPPLY REGULATOR ESCAPE CYLINDER VORTEX MAIN INLET PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION WARNING THIS MANUAL MUST BE READ CAREFULLY BY ALL PERSONS WHO HAVE OR WILL HAVE THE RESPONSIBILITY

More information

O-Two Self-Study Guide. Equinox Relieve

O-Two Self-Study Guide. Equinox Relieve O-Two Self-Study Guide Equinox Relieve PRODUCT DESCRIPTION/PRINCIPLES OF OPERATION The Equinox Relieve provides spontaneously breathing patients with a preset mixture of oxygen and nitrous oxide for analgesia

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.12 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient,

More information

AUTOVENT 4000 VENTILATOR

AUTOVENT 4000 VENTILATOR OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on

More information

Peninsula Dental Social Enterprise (PDSE)

Peninsula Dental Social Enterprise (PDSE) Peninsula Dental Social Enterprise (PDSE) Medical Gas Policy Version 3.0 Date approved: August 2018 Approved by: The Board Review due: August 2021 Policy will be updated as required in response to a change

More information

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021-PR Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-PR For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4

More information

RAM 4021 Operation Manual

RAM 4021 Operation Manual RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6

More information

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC. MANUFACTURERS AND DISTRIBUTORS OF: O-Two eseries Electronic Transport Ventilators O-Two Single Use Open CPAP Systems O-Two CAREvent Range of Automatic Transport Ventilators O-Two CAREvent Range of Automatic

More information

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician. U S E R M A N U A L Model: 19MFA1001 Series SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090 Northampton,

More information

SERVICE MANUAL MODEL BA600AMST-S1 BREATHING AIR PANEL

SERVICE MANUAL MODEL BA600AMST-S1 BREATHING AIR PANEL SERVICE MANUAL MODEL BA600AMST-S1 BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device. 2/05 GENERAL

More information

Pressure Regulator. Instructions for Use

Pressure Regulator. Instructions for Use Pressure Regulator Instructions for Use 702-0083.9 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient, the

More information

Budget Range Operators Handbook

Budget Range Operators Handbook Budget Range Operators Handbook BAMBI AIR COMPRESSORS LTD 152 Thimble Mill Lane Heartlands Birmingham B7 5HT United Kingdom Tel: 0121 322 2299 Fax: 0121 322 2297 Email: sales@bambi-air.co.uk www.bambi-air.co.uk

More information

RAM Operation Manual

RAM Operation Manual RAM 4021-1 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-1 For Your Safety... 2 Description... 2 Setup Mode... 3 Lights/Alarms... 3 Operation... 4 Calibration...

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

Pressure Regulators. Operating Instructions. Instrumentation

Pressure Regulators. Operating Instructions. Instrumentation Pressure Regulators Operating Instructions FAILURE OR IMPROPER SELECTION OR IMPROPER USE OF THIS PRODUCT CAN CAUSE DEATH, PERSONAL INJURY AND PROPERTY DAMAGE. This document and other information from the

More information

45A, Armthorpe Road, Brampton, Ontario, Canada, L6T 5M4. Made in Canada by O-Two Medical Technologies Inc.

45A, Armthorpe Road, Brampton, Ontario, Canada, L6T 5M4. Made in Canada by O-Two Medical Technologies Inc. O-Two eseries Electronic Transport Ventilators O-Two Single Use Open CPAP Systems O-Two CAREvent Range of Automatic Transport Ventilators O-Two CAREvent Range of Automatic Resuscitators O-Two SMART BAG

More information

453 Series Steam Heated Vaporizing Regulator

453 Series Steam Heated Vaporizing Regulator ADI 0453A Certified ISO 9001:2000 453 Series Steam Heated Vaporizing Regulator INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls Corporation

More information

AMP Oil Free Manual AMP 50-8-TC AMP 50-6-D AMP General User and Maintenance Instructions

AMP Oil Free Manual AMP 50-8-TC AMP 50-6-D AMP General User and Maintenance Instructions AMP Oil Free Manual AMP 50-8-TC AMP 50-6-D AMP 50-24 General User and Maintenance Instructions Silentaire Technology 8614 Veterans Memorial Dr. Houston, TX 77088 800-972-7668 Fax 832-327-0669 www.silentaire.com

More information

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

AIR/OXYGEN BLENDER INSTRUCTION MANUAL BIO-MED DEVICES AIR/OXYGEN BLENDER INSTRUCTION MANUAL CATALOG #2120 REV 042203 BIO-MED DEVICES INC. 1445 BOSTON POST ROAD, GUILFORD, CT 06437 (203) 458-0202 FAX (203) 458-0440 www.biomeddevices.com TABLE

More information

AIR/OVER HYDRAULIC JACK 20 TON

AIR/OVER HYDRAULIC JACK 20 TON AIR/OVER HYDRAULIC JACK 0 TON 4487 ASSEMBLY AND OPERATING INSTRUCTIONS 349 Mission Oaks Blvd., Camarillo, CA 930 Visit our Web site at http://www.harborfreight.com Copyright 999 by Harbor Freight Tools.

More information

VAPOMATIC ANESTHETIC VAPORIZER

VAPOMATIC ANESTHETIC VAPORIZER VAPOMATIC ANESTHETIC VAPORIZER INSTRUCTION MANUAL SOLD FOR VETERINARY USE ONLY BY: A.M. BICKFORD, INC. 12318 BIG TREE ROAD WALES CENTER, NY 14169 1-800-795-3062 Fax: 716-652-2046 www.ambickford.com sales@ambickford.com

More information

Operators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500

Operators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500 Operators Manual Supera Wall Mount Anesthesia Machines Models: M2300 M2500 (All shown with optional equipment) M2300 M2500 Examination and Preparation for Use Thank you for ordering an Supera Anesthesia

More information

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A BlenderBuddy 2 Operating Manual & Instructions For Use ENGLISH R233M01 REV. A Maxtec 2305 South 1070 West Salt Lake City, Utah 84119 USA TEL (800) 748.5355 FAX (801) 973.6090 www.maxtec.com E-mail: sales@maxtec.com

More information

Micro Dial-Flowmeter. Instructions for Use

Micro Dial-Flowmeter. Instructions for Use Micro Dial-Flowmeter Instructions for Use 702-0082.8 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician. USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. RECEIVING / INSPECTION Remove the Precision Medical,

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

SILENTAIRE TECHNOLOGY

SILENTAIRE TECHNOLOGY SILENTAIRE TECHNOLOGY General User and Maintenance Instructions Thank you and congratulations on the purchase of your PANTHER, the leader in the industry of portable air compressors. The PANTHER is built

More information

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1 ANA 74300 Oxygen Regulator with Yoke Connector User Manual Manual No. ANA 74300 001 Issue 1 Safety Thank you for purchasing this Clements Oxygen Regulator. For your safety it is imperative that this unit

More information

1020 Industrial Drive, Orlinda, TN fax

1020 Industrial Drive, Orlinda, TN fax Operation Manual Ultrafiltration for High Purity Distribution K-A-HPTUF Series 615-654-4441 sales@specialtyh2o.com 615-654-4449 fax TABLE OF CONTENTS Section 1 GENERAL 1.1 Warnings and Cautions... 1 1.2

More information

SILENTAIRE TECHNOLOGY

SILENTAIRE TECHNOLOGY SILENTAIRE TECHNOLOGY 8614 Veterans Memorial. Houston, Texas 77088 832/327-7452 800/972-7668 Fax: 832/327-0668 E-mail: silentaire@silentaire.com Thank you and congratulations on the purchase of your OILLESS

More information

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289 3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS Item #31289 The EASTWOOD 3 GALLON, OILLESS PANCAKE COMPRESSOR, with an Integral Air Regulator, efficiently supplies all compressed air requirements for

More information

RAM 4021-DPX Operation Manual

RAM 4021-DPX Operation Manual RAM 4021-DPX Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS ABL 4021-DPX / RAM 4021-DPX For Your Safety... 3 Description... 3 Setup Mode... 4 Lights/Alarms... 4 Operation...

More information

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL www.modsafe.com SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device.

More information

SPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1.

SPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1. 2 GALLON TWIN STACK AIR COMPRESSOR Model: 9526 DO NOT RETURN TO STORE. Please CALL 800-348-5004 for parts and service. CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or

More information

Equinox II NITROUS OXIDE / OXYGEN ADMINISTRATION EQUIPMENT 01EQ7000E

Equinox II NITROUS OXIDE / OXYGEN ADMINISTRATION EQUIPMENT 01EQ7000E MANUFACTURERS AND DISTRIBUTORS OF: CPR Bio-Barrier Face Shields Universal Pocket Ventilators Rescue Breather CPR Protective Barriers Demand Valve Resuscitators CAREvent Range of Automatic Transport Ventilators

More information

Medical Gas Control Panel

Medical Gas Control Panel Medical Gas Control Panel INSTALLATION & MAINTENANCE MANUAL 255105 (Rev. 6) 2018-04 2 255105 (Rev. 6) 2018-04 1.0 User Responsibilities The information contained in this installation and operation maintenance

More information

Code AWC20HP Air Compressor

Code AWC20HP Air Compressor Code 951816 AWC20HP Air Compressor Index of Contents Index of Contents 02 Declaration of Conformity 02 What s Included 03 Safety Precautions 03 Specifications (AWC20HP Air Compressor) 04 Assembly Instructions

More information

CPR MEDICAL DEVICES INC. OPERATING MANUAL

CPR MEDICAL DEVICES INC. OPERATING MANUAL CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR OXYLATOR EMX RESUSCITATION AND INHALATION SYSTEM Symbols used in the Operating Manual for the OXYLATOR EMX and/or on the OXYLATOR EMX unit: WARNING! Do not

More information

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight Technical Specifications Operating Instructions Maintenance Information Troubleshooting Guide Parts Diagrams AC1810 / AC1810-A THE EVOLUTION OF PERFECTION CAUTION: Before attempting to use or service this

More information

Pressure Regulators. Instructions for Use

Pressure Regulators. Instructions for Use Pressure Regulators Instructions for Use 702-0032.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR OXYLATOR FR-300 RESUSCITATOR Symbols used in the Operating Manual for the Oxylator FR-300 and/or on the Oxylator FR-300 unit: WARNING! Do not use grease or

More information

2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL

2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL 2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL Model: 52024 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement,

More information

OPERATION MANUAL NTF-15

OPERATION MANUAL NTF-15 OPERATION MANUAL NTF-15 Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81235-00 (Rev B) TABLE

More information

ACCU-PULSE Installation and Operation Instructions

ACCU-PULSE Installation and Operation Instructions ACCU-PULSE Installation and Operation Instructions Pump Discharge Installation: Chargeable Models Step 1: Mounting Position Mount ACCU-PULSE as close to the pump discharge as possible to absorb the pulse

More information

Installation and Maintenance Manual. Compact Medical Gas Outlets

Installation and Maintenance Manual. Compact Medical Gas Outlets Installation and Maintenance Manual Compact Medical Gas Outlets Contents Product Description 3 Cleaning and Lubricating 4 Inspection and Testing 4 Installation and Dimensions 5-6 Compact Outlet Indexing

More information

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST OIL-LESS TYPE IMPORTANT: PLEASE READ CAREFULLY BEFORE STARTING OPERATIONS. THE CONTENTS ARE FOR GENERAL INFORMATION OF ALL THE SIMILAR MODELS. Record

More information

User Operation Manual and Safety Instructions

User Operation Manual and Safety Instructions PCP Airgun Charging Station 74 cu.ft. / 98 cu. ft. Carbon Fiber Tank & Accessories User Operation Manual and Safety Instructions BRAND: Air Venturi DESCRIPTION: PCP Airgun Charging System FILL PRESSURE:

More information

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL Product Manual MST Portable Air Filtration System SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the

More information

Operating Manual. Model 25200B/25202B Refrigerant Recovery Unit. (Not for use in automotive R-134a applications)

Operating Manual. Model 25200B/25202B Refrigerant Recovery Unit. (Not for use in automotive R-134a applications) Operating Manual Model 25200B/25202B Refrigerant Recovery Unit (Not for use in automotive R-134a applications) ROBINAIR Refrigerant Recovery Unit Model: 25200B /25202B Series Type of Equipment: Recovery

More information

SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR

SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR S S F U SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR MACHINERY DIVISION OWNER S MANUAL UNIT INSTRUCTIONS Please follow all SSFU Safety Instructions. Contact your Duro Dyne Tech Service if you have

More information

accidents which arise due to non-observance of these instructions and the safety information herein.

accidents which arise due to non-observance of these instructions and the safety information herein. 3 GALLON PANCAKE COMPRESSOR Model: 50959 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement, or other masonry.

More information

CSA Sample Draw Aspirator Adapter Operator s Manual

CSA Sample Draw Aspirator Adapter Operator s Manual 30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

Welcome to the Specialized Medical Services Respiratory training webinar series!

Welcome to the Specialized Medical Services Respiratory training webinar series! Welcome to the Specialized Medical Services Respiratory training webinar series! SMS is your Long Term Care (LTC) facility single source for oxygen, medical equipment, respiratory care services and supplies

More information

Certified ISO 9000 INSTALLATION AND OPERATION INSTRUCTIONS. Before Installing or Operating, Read and Comply with These Instructions

Certified ISO 9000 INSTALLATION AND OPERATION INSTRUCTIONS. Before Installing or Operating, Read and Comply with These Instructions ADI 1219-M Certified ISO 9000 1600 Series Dual-Range Flowmeters INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls Corporation of America

More information

Product Information News

Product Information News Product Information News MMR SECOND STAGE REGULATOR VALVE CORE MSA has changed its recommended rebuilding of the MMR Second Stage Regulator s bypass sleeve assembly during the annual inspection. Based

More information

G92 Series BASOTROL Automatic Pilot Gas Valve

G92 Series BASOTROL Automatic Pilot Gas Valve Installation Instructions Issue Date September 17, 2008 G92 Series BASOTROL Automatic Pilot Gas Valve Installation IMPORTANT: Only qualified personnel should install or service BASO Gas Products. These

More information

Blendox AII5000F Inhalation Antalgic System. User s Manual :SP-Y Series No. Version :A1. Date of Review:

Blendox AII5000F Inhalation Antalgic System. User s Manual :SP-Y Series No. Version :A1. Date of Review: AII5000F Inhalation Antalgic System User s Manual Series No. :SP-Y53564 Version :A1 Date of Review:2012-04-29 To Users: Welcome to Blendox AII5000F Inhalation Antalgic System (hereafter referred to as

More information

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting This is a quick reference for setting up, using, and troubleshooting the Hallowell EMC 2000 and 2002 Small Animal Ventilator.

More information

- GAS REGULATOR - COMPACT STYLE

- GAS REGULATOR - COMPACT STYLE INSTRUCTIONS FOR USE - GAS REGULATOR - COMPACT STYLE Preset at 50 or 60 psi Rx ONLY GR - XXX - [(TT) (UU) - (XX)(Z)] GR XXX (TT) (UU) (XX) (Z) Gas Connection Outlet Flowrate Addition Outlet Additional

More information

80 Litre Suction Oil Drainer

80 Litre Suction Oil Drainer 80 Litre Suction Oil Drainer Please dispose of packaging for the product in a responsible manner. It is suitable for recycling. Help to protect the environment, take the packaging to the local amenity

More information

High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters

High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters ADI 3210-A ADI 3210-A Certified ISO 9001 Certified ISO 9001 High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters Warning: An appropriately sized pressure relief device downstream of the regulator

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE15C150, VE18C150, VE25C150 PART NO: 2226010, 2226020, 2226025 OPERATION & MAINTENANCE INSTRUCTIONS LS0715 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

OPERATION MANUAL NTF-60 Plus

OPERATION MANUAL NTF-60 Plus OPERATION MANUAL NTF-60 Plus Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81264-00 (Rev A)

More information

Protocol Switchover Station

Protocol Switchover Station ADI 3215-B Certified ISO 9001 Protocol Switchover Station INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls Corporation of America

More information

stormrider Instructions for Use Version 08/2017

stormrider Instructions for Use Version 08/2017 stormrider EN Instructions for Use Version 08/2017 INTRODUCTION DESCRIPTION SAFETY INSTRUCTIONS TECHNICAL DATA HANDLING & OPERATION CARE & MAINTENANCE TROUBLESHOOTING DESCRIPTION Front Sight Barrel Rear

More information