PERFORMANCE OF NEEDLE VALVES

Size: px
Start display at page:

Download "PERFORMANCE OF NEEDLE VALVES"

Transcription

1 Br. J. Anaesth. (1986), 58, PERFORMANCE OF NEEDLE VALVES P. HUTTON AND R. W. BOADEN The use of needle valves to control theflowof gas in anaesthetic breathing systems is almost universal. We became interested in their detailed performance when assessing the suitability of various types offlowregulator for incorporation to a digital valve which was to be used to control the flow of anaesthetic gases. This is described in detail elsewhere (Boaden and Hutton, 1986). An extensive search of the literature and enquiries at several government laboratories produced no publication describing the behaviour of needle valves. In practice, their design appears to be largely empirical and devoid of any theoretical basis. Consequently, it was decided to investigate their properties and see whether, over a clinical range of flows, they behaved as simple flow resistances or whether they possessed nonlinearities and discontinuities in their performance curves which depended upon the upstream or downstream pressures. MATERIALS AND METHODS The basic apparatus required to investigate the effects of changes in upstream and downstream pressures on a needle valve is shown diagrammatically in figure 1. In the experiments described the high pressure gas supply was from Size J oxygen, medical gas cylinders fitted with an adjustable two-stage regulator. The upstream and downstream pipeline pressures were measured by Bourdon gauges which had themselves been calibrated against a Budenberg deadweight pressure gauge tester (Type 240). The downstream or "back" pressure on the needle valve under test could be increased by adjusting a variable resistance placed before theflowmeter (fig. 1). All P. HUTTON, B.SC., PH.D., M.B. CH.B., F.F-A.R.CS., Sir Humphry Davy Department of Anaesthesia, University of Bristol, Bristol Royal Infirmary, Bristol. R. W. BOADEN, B.SC., M.B., BJ., F.F.A.R.C.S., Bristol Royal Infirmary, Bristol. SUMMARY The pressure-flow characteristics of needle valves used on anaesthetic equipment were investigated. It was found that, under normal conditions of use, the valves behaved in a manner similar to convergent nozzles with sonic velocity at the throat. This means that, once the valve has been set, the mass flow rate of gas is effectively independent of the downstream changes in resistance and compliance encountered in normal anaesthetic practice. The error in a preset flow induced by the permissible variations in the upstream pipeline pressure was measured. the flows were measured on a, Platon Gapmeter Lab Kit. This is a set of metering tubes and floats which can be selected in combination to produce the optimum working range. Specially commissioned calibration charts for oxygen and nitrous oxide were supplied on request by the manufacturers. The accuracy quoted by the manufacturers was ± 1.25 % of full scale deflection. The anaesthetic needle valve, when mounted in its usual place below the rotameter block, cannot have its characteristics tested realistically because increases in downstream pressure cause the seals around the rotameters to leak and the glass to crack. Therefore, a standard, new BOC valve was fitted into a gas-tight brass block for our purposes (fig. 2). The outlet pressure of all pressure regulators depends upon cylinder pressure and flow (Hill, 1980). Consequently, when each set of readings was taken, the downstream variable resistance and the cylinder pressure regulator had to be adjusted together to achieve the required upstream and downstream pressures on the calibrated Bourdon gauges. Two sets of investigations were carried out.

2 920 BRITISH JOURNAL OF ANAESTHESIA Adjustable two-stage pressure regulator Upstream calibrated pressure gauge Downstream calibrated pressure gauge High accuracy (lowmeter 1 Needle valve under test Variable resistance FIG. 1. Diagrammatic representation of test apparatus. The effect of changes in downstream pressure on the flow through the needle valve over a range of clinical flow rates The needle valve (fig. 2) was fitted into the apparatus (fig. 1) and, with the valve discharging directly to atmosphere via the flow meter at the required nominal flow rate (zero value of variable resistance (fig. 1)), the upstream pressure was adjusted to 414 kpa gauge. The flow rate was then recorded accurately and measurements made of the discharging gas temperature and the atmospheric pressure. The variable resistance was then adjusted so that the back pressure on the needle valve increased and the differential pressure drivingflow through it decreased. If there was any change in the upstream pressure, this was readjusted to 414 kpa gauge. The flow was then allowed to stabilize and the upstream and downstream pressures, the discharging flow rate and the discharging gas temperature were recorded. The value of the variable resistance was then increased incrementally until the back pressure approached the driving pressure, at which point the flow was, by definition, zero. At each incremental step the upstream pressure was re-set to 414 kfa and the measurements described above recorded. Readings were taken with six different initial flow rates spanning the range 1-9 litre min" 1. The effect of changes in upstream pressure on the flow through the needle valve over a range of clinical settings With the needle valve in its test block (fig. 2) fitted into the apparatus (fig. 1) it was adjusted, with an upstream pressure of 414 kpa, to discharge directly to atmosphere via the flow meter at a nominal flow rate of 1 litre min" 1 (zero value of variable resistance (fig. 1)). The upstream pressure was then varied over the range kpa gauge and the resulting flow rates were measured. This was repeated for nominal initial flow rates of 3, 5 and 7 litre min" 1. RESULTS The results of these two experiments with the volume flow rates corrected to ATP (20 C, FIG. 2. Needle valve in brass block.

3 PERFORMANCE OF NEEDLE VALVES 921 g Q 0. -= 4 e o g o Upstream pressure (kpa gauge) Downstream pressure (kpa gauge) 400 FIG. 3. The effect of changes in downstream pressure on the flow through the needle valve over a range of clinical settings. Driving pressure = 414 kpa gauge. Atmospheric pressure = kpa kpa) are shown in figures 3 and 4. Correcting all the oxygen flow rates (V) to the same conditions of temperature and pressure implies that all the points plotted will be proportional to the massflowrate (M) of gas. This becomes relevant when the needle valve is compared with a frictionless nozzle. DISCUSSION Figure 3 indicates that, as the downstream pressure was increased from atmospheric, there was initially a plateau phase on the flow-pressure curve before there was any change in the flow through the valve. The extent of the plateau for eachflowis shown approximately by line A (which was drawn in by eye). It can be seen that, at 9 litre min" 1, the downstream pressure had to increase to 80 kpa gauge before flow decreased, and that the plateau increased to considerably more than 1 atm at the four lowest flow rates. The presence of this plateau implies that, under certain conditions, the mass flow of a given gas through a needle valve may be unaltered, despite changes in the pressure drop across it. This contrasts with flow in pipes where, as soon as the pressure gradient changes, the flow (whether laminar or turbulent) follows part passu (although not FIG. 4. The effect of changes in upstream pressure on the flow through the needle valve over a range of clinical settings. The vertical dotted line shows the normal supply pressure of 414 kpa gauge. Atmospheric pressure = 102 kpa. necessarily linearly). To explain this, perhaps unexpected, characteristic of needle valves it is useful to consider the properties of a frictionless convergent nozzle. A convergent nozzle is shown schematically in figure 5A and its behaviour under variations of back pressure is described in many standard textbooks (Shapiro, 1953). In summary, the nozzle (fig. 5A) is considered to be supplied with gas at an absolute static upstream pressure of Pu and the absolute downstream or back pressure on the nozzle Pb is controlled by a valve. The actual absolute pressure at the exit of the nozzle throat ispe. Figures 5B and c show, respectively, the variations in pressure down the nozzle and mass flow through it as Pb is varied. Initially, when Pb = Pu there is no flow (point a) then when Pb is reduced, theflowbegins (point b). With further reductions in Pb this trend is continued until point c is reached. After this, further decreases in Pb to attain points d and e produce no subsequent increase in mass flow through the nozzle. The nozzle in this condition is said to be " choked". In addition, Pe (the true throat pressure) never decreases to less than that at condition c, the further reduction in pressure to Pb occurring after the gas has left the nozzle. The condition of maximum or choked flow occurs when the gas has achieved the local speed of sound (sonic velocity)

4 922 BRITISH JOURNAL OF ANAESTHESIA Maximum e d c Mass flow rate of gas Distance down nozzle Pb/Pu FIG. 5. Schematic diagram of nozzle characteristics. in the narrowest part of the nozzle. This is often termed the "critical flow condition". The minimum pressure drop across the nozzle necessary to produce chokedflowoccurs at a given Pb: Pu ratio (in absolute pressures) which depends on the gas in use. It is denned (Shapiro, 1953) as jtor choking I_Y+ 1 ' where y = the ratio of the specific heats for the gas in question. Critical flow devices have been used previously in the Pneumotron ventilator (Mushin et al., 1980), in gas analysers (Hill, 1980), in scavenging systems (Vaughan et al., 1977) and to meter accurately the flow of natural gas and the output from compressors (Ower and Pankhurst, 1966). We do not, however, know of any previous publications which have identified needle valves as possessing these properties. Comparing figure 3 withfigure 5c, the similarity in performance between a needle valve and a convergent nozzle is obvious. The critical pressure ratio for oxygen (y = 1.4) calculated from equation (1) is 0.53 and is shown plotted as line B on figure 3. Above 2.5 litre min" 1 the plateau region of the valve characteristic decreases gradually from 171 kpa gauge (where Pb/Pu = 0.53) to approximately 80 kpa gauge at 9 litre min" 1 (where Pb/Pu = 0.35). This decrease in the value of Pb/Pu from the ideal of 0.53 is almost certainly the result of the increased frictional losses in the input and output sections of the needle valve. These increase dramatically with increases in gas velocity. When the pressure decrease caused by these becomes of the same order of magnitude as the nozzle loss, then the actual input pressure to the nozzle is not Pu, but Pu minus the frictional loss of pressure head up to the entrance of the nozzle. Consequently, the downstream pressure for choking flow based on Pb/Pu = 0.53 is reduced. Despite these frictional losses, even at the high flow rates, the nozzle flow is invariant with increases of downstream pressure up to 80 kpa gauge. Since the least compliant lungs one would ever ventilate in practice would almost never need an inflation pressure greater than 10 kpa gauge, any anaesthetic ventilating system fed by a needle valve from normal pipe line or cylinder pressures effectively becomes a constant flow generator. In addition, if the pressure in the patient breathing system or ventilator supply line increases, any change in the rotameter readings does not imply a change in the massflowrate of gas through the needle valves. The increase in downstream pressure does of course reduce the volume flow rate. In addition, the rotameters (which are very sensitive to downstream pressure) are calibrated at a specific downstream gas pressure of one standard atmosphere. When the downstream pressure increases, they are effectively being forced onto a different calibration curve. The effect of changes in upstream pressure on the mass flow rate of gas through the nozzle when the downstream pressure is constant at atmospheric are shown in figure 4. It can be seen that, for any given nominal setting, increasing the upstream

5 PERFORMANCE OF NEEDLE VALVES 923 pressure increases the flow of gas through the needle valve and decreasing it does the reverse. This is in agreement with the theory of critical flow devices (see Appendix) which predicts that, under given initial conditions, the mass flow rate of gas per unit cross sectional area of nozzle is proportional to the absolute upstream pressure (Shapiro, 1953). For there to be perfect agreement with the theory, all the lines on figure 4 should, when extrapolated backwards, intersect the abscissa at an absolute upstream pressure of zero and they do not do so. This discrepancy is the result of pressure losses in the piping between the pressure gauges and needle valve (fig. 1), viscosity effects in real fluids which are omitted from the theoretical solution, and the changes in the geometry of the orifice of the needle valve as it is adjusted to give the initial set flow. The data of figure 4 can be replotted with the gradient dp^/dpu against flow (P^) to produce an empirical expression to predict the deviation of any given initial flow with variations in upstream pressure. Using a least squares best fit, this results (for our experimental model and needle valve) in the equation: Change in flow = change in pressure (kpa) (litre min- 1 ) x ( (initial flow rate (litre min- 1 )) lo" 3 (2) Statistical analysis shows that r = and the 95% confidence limits for the coefficients are ((1.8±0.9) x 10" 3 litre min- 1 kpa" 1 ) and ((2.2±0.2)xl0" 3 kpa- 1. The variations in gas flow rate as a result of permitted variations in gas supply pressure (393-^434 kpa gauge) (D.H.S.S., 1977) can be calculated easily. For example, a flow rate of 7 litre min" 1 (ATP) set at a pipeline pressure of 414 kpa gauge would increase to 7.34 litre min" 1 at a pipeline pressure of 434 kpa gauge and decrease to 6.64 litre min" 1 at a pipeline pressure of 393 kpa gauge. Several different types of needle valve were subjected to the type of bench testing described above. They did not all approximate to a perfect nozzle to the same degree. A general conclusion was that the physically bigger the valve, the more closely it conformed to theory at higherflow rates. This was thought to be mainly a result of the reduction in frictional losses in the entrance piping of the larger valves. The whole series of tests was repeated with nitrous oxide (y = 1.3) and the results were essentially the same as those described above for oxygen. They are not shown since they would merely repeat the data in figures 3 and 4. APPENDIX The theory of critical flow devices (Shapiro, 1953) predicts that, under conditions of choking flow, the mass flow rate per unit area of nozzle (M) is given as: where Pu = absolute upstream pressure of static gas, y = ratio of specific heats, R = universal gas constant and T = absolute temperature of upstream gas. For a given gas with a given T, the quantity under the square root sign is constant (k). Therefore: M ^M and, since M is proportional to V (see Results above), Pu~ dpu (3) Using the data of figure 4, the absolute upstream pressure at the initial flow rate is = 516 kpa. Re-arranging equation (3) results in: change in flow _ change in pressure (kpa) (litre min" 1 ) 516 kpa x initial flow rate (litre min" 1 ) = change in pressure (kpa) x initial flow rate (litre min" 1 ) x 1.94 x 10"* Comparing this expression with our empirical results (see above, equation (2)) the experimental coefficient of (1.8 x 10"') litre min" 1 kpa" 1 differs significantly from zero at the P < 0.05 level, but there is no significant difference between the experimental coefficient of (2.2 x 10"*) kpa" 1 and the theoretical value of (1.94 x 10"') kpa" 1. It can be seen (equation (2)) that the major deviation from theoretical behaviour occurs at low flow rates where the error in the predicted change in flow approaches 50%. To achieve a low flow through the needle valve requires the needle to be advanced a relatively long distance into its sleeve. This simultaneously reduces the cross sectional area available to the flow and increases the length of the throat section. Both these factors increase viscous losses and encourage deviation from theoretical predictions based on a perfect gas. ACKNOWLEDGEMENTS The authors acknowledge the loan of a Platon Gapmeter Lab Kit from Mr A. Burleton and a Budenberg deadweight pressure gauge tester from the Department of Aeronautical Engineering, University of Bristol. REFERENCES Boaden, R. W., and Hutton, P. (1986). The digital control of anaesthetic gas flow. Anaesthesia, 41, 413. Department of Health and Social Security (1977). Piped medical gases, medical compressed air and medical vacuum installations: and supplement. Permit to work lystem. Hospital Technical Memorandum No. 22.

6 924 BRITISH JOURNAL OF ANAESTHESIA Hill.D. W. (1980). Physics Applied to Anaesthesia, 4thEdn, pp. Shapiro, A. H. (1953). The Dynamics and Thermodynamics of 71, 189. London: Butterworths. Compressible Fluid Flow, Vol. 1, pp. 91, 84, 85. New York: Mushin, W. W., Rendell-Baker, L., Thompson, P. W., and The Ronald Press Co. Mapleson, W. W. (1980) Automatic Ventilation of the Lungs, Vaughan, R. S., Willis, B. A., Mapleson, W. W., and Vickers, 3rd Edn, p Oxford: Blackwell Scientific Publications. M. D. (1977). The Cardiff Aldavac anaesthetic scavenging Ower, E., and Pankhurst, R. C. (1966). The Measurement of system. Anaesthesia, 32, 339. Airflow, pp. 196, 250. New York: Pergamon Press.

MEMORANDUM. Investigation of Variability of Bourdon Gauge Sets in the Chemical Engineering Transport Laboratory

MEMORANDUM. Investigation of Variability of Bourdon Gauge Sets in the Chemical Engineering Transport Laboratory 1 MEMORANDUM TO: FROM: Prof. Davis Hubbard Prof. Faith A. Morrison DATE: 22 April 2014 RE: Investigation of Variability of Bourdon Gauge Sets in the Chemical Engineering Transport Laboratory Introduction

More information

RESISTANCE TO AIRFLOW IN ANAESTHETIC BREATHING SYSTEMS

RESISTANCE TO AIRFLOW IN ANAESTHETIC BREATHING SYSTEMS Br. J. Anaesth. (1989), 62, 456-461 RESISTANCE TO AIRFLOW IN ANAESTHETIC BREATHING SYSTEMS D. G. MARTIN, K. L. KONG AND G. T. R. LEWIS The resistance to airflow of the components of breathing systems have

More information

Applied Fluid Mechanics

Applied Fluid Mechanics Applied Fluid Mechanics 1. The Nature of Fluid and the Study of Fluid Mechanics 2. Viscosity of Fluid 3. Pressure Measurement 4. Forces Due to Static Fluid 5. Buoyancy and Stability 6. Flow of Fluid and

More information

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow Georgia Institute of Technology School of Earth and Atmospheric Sciences EAS 4641, Spring 2008 Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow Purpose of Lab 3: 1) To gain a

More information

S.A. Klein and G.F. Nellis Cambridge University Press, 2011

S.A. Klein and G.F. Nellis Cambridge University Press, 2011 16-1 A flow nozzle is to be used to determine the mass flow rate of air through a 1.5 inch internal diameter pipe. The air in the line upstream of the meters is at 70 F and 95 psig. The barometric pressure

More information

The Discussion of this exercise covers the following points:

The Discussion of this exercise covers the following points: Exercise 3-2 Orifice Plates EXERCISE OBJECTIVE In this exercise, you will study how differential pressure flowmeters operate. You will describe the relationship between the flow rate and the pressure drop

More information

TWO PHASE FLOW METER UTILIZING A SLOTTED PLATE. Acadiana Flow Measurement Society

TWO PHASE FLOW METER UTILIZING A SLOTTED PLATE. Acadiana Flow Measurement Society TWO PHASE FLOW METER UTILIZING A SLOTTED PLATE Acadiana Flow Measurement Society Gerald L. Morrison Presented by: Mechanical Engineering Department Daniel J. Rudroff 323 Texas A&M University Flowline Meters

More information

BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY

BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY Dr.R.Selvakumar.M.D.D.A.DNB Professor of Anaesthesiology, K.A.P.Viswanatham Govt medical college, Trichy. The current practice of Anaesthesiology demands knowledge

More information

Experimental Analysis on Vortex Tube Refrigerator Using Different Conical Valve Angles

Experimental Analysis on Vortex Tube Refrigerator Using Different Conical Valve Angles International Journal of Engineering Research and Development e-issn: 7-067X, p-issn: 7-00X, www.ijerd.com Volume 3, Issue 4 (August ), PP. 33-39 Experimental Analysis on Vortex Tube Refrigerator Using

More information

Characterizers for control loops

Characterizers for control loops Characterizers for control loops By: F. G. Shinskey (May 1999) Introduction Commercial controllers such as the PID series (proportional, integral, derivative, and their combinations) are linear devices

More information

Device Description. Operating Information. CP Q (eq. 1) GT. Technical Bulletin TB-0607-CFP Hawkeye Industries Critical Flow Prover

Device Description. Operating Information. CP Q (eq. 1) GT. Technical Bulletin TB-0607-CFP Hawkeye Industries Critical Flow Prover A compressible fluid traveling at subsonic velocity through a duct of constant cross section will increase velocity when passing through a region of reduced cross-sectional area (in this case, an orifice)

More information

Third measurement MEASUREMENT OF PRESSURE

Third measurement MEASUREMENT OF PRESSURE 1. Pressure gauges using liquids Third measurement MEASUREMENT OF PRESSURE U tube manometers are the simplest instruments to measure pressure with. In Fig.22 there can be seen three kinds of U tube manometers

More information

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Lab 1c Isentropic Blow-down Process and Discharge Coefficient 058:080 Experimental Engineering Lab 1c Isentropic Blow-down Process and Discharge Coefficient OBJECTIVES - To study the transient discharge of a rigid pressurized tank; To determine the discharge coefficients

More information

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS Brit. J. Anaesth. (1956), 28, 196 CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS BY RUSSELL M. DAVIES, I. R. VERNER Queen Victoria Hospital, East Grinstead AND A. BRACKEN Research and Development Centre,

More information

Figure 1 Schematic of opposing air bearing concept

Figure 1 Schematic of opposing air bearing concept Theoretical Analysis of Opposing Air Bearing Concept This concept utilizes air bearings to constrain five degrees of freedom of the optic as shown in the figure below. Three pairs of inherently compensated

More information

Flow in a shock tube

Flow in a shock tube Flow in a shock tube April 30, 05 Summary In the lab the shock Mach number as well as the Mach number downstream the moving shock are determined for different pressure ratios between the high and low pressure

More information

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12 LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/ This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the rates with which selected

More information

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features ANALOG MASS FLOW CONTROLLERS Model AF mass fl ow controllers are designed to indicate fl ow rates and control set fl ow rates of gases. Each of these units incorporates an advanced straight tube sensor

More information

Schedule of Requirements THERMODYNAMICS LABORATORY- CHEMICAL ENGINEERING DEPARTMENT

Schedule of Requirements THERMODYNAMICS LABORATORY- CHEMICAL ENGINEERING DEPARTMENT S. No 1 Description Calorimeter The Unit should be designed for the accurate determination of the calorific value of liquid and solid hydrocarbons and other fuels. Specifications: A temperature-controlled

More information

Impact of imperfect sealing on the flow measurement of natural gas by orifice plates

Impact of imperfect sealing on the flow measurement of natural gas by orifice plates Impact of imperfect sealing on the flow measurement of natural gas by orifice plates Rubens Silva Telles 1, Kazuto Kawakita 2 1 IPT Instituto de Pesquisas Tecnológicas, São Paulo, Brazil, rtelles@ipt.br

More information

Micro Motion Pressure Drop Testing

Micro Motion Pressure Drop Testing 12/2018 Micro Motion Pressure Drop Testing www.emerson.com/micromotion Introduction Micro Motion has traditionally taken a very conservative approach to pressure drop, with single pressure measurements

More information

MET 335W Fluid Mechanics Laboratory. Lab 1: Bourdon Tube Calibration. Nick Peak

MET 335W Fluid Mechanics Laboratory. Lab 1: Bourdon Tube Calibration. Nick Peak MET 335W Fluid Mechanics Laboratory Lab 1: Bourdon Tube Calibration Nick Peak January 15, 2018 Purpose The purpose of this experiment is to test the accuracy of a bourdon pressure gauge. This is accomplished

More information

OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply

OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply 4.1.1 General The primary function of a system for handling fuel oil is to transfer oil from the storage tank to the oil burner at specified conditions

More information

Lab # 03: Visualization of Shock Waves by using Schlieren Technique

Lab # 03: Visualization of Shock Waves by using Schlieren Technique AerE545 Lab # 03: Visualization of Shock Waves by using Schlieren Technique Objectives: 1. To get hands-on experiences about Schlieren technique for flow visualization. 2. To learn how to do the optics

More information

3 1 PRESSURE. This is illustrated in Fig. 3 3.

3 1 PRESSURE. This is illustrated in Fig. 3 3. P = 3 psi 66 FLUID MECHANICS 150 pounds A feet = 50 in P = 6 psi P = s W 150 lbf n = = 50 in = 3 psi A feet FIGURE 3 1 The normal stress (or pressure ) on the feet of a chubby person is much greater than

More information

Process Dynamics, Operations, and Control Lecture Notes - 20

Process Dynamics, Operations, and Control Lecture Notes - 20 Lesson 0. Control valves 0.0 Context Controller output is a signal that varies between 0 and 100%. Putting this signal to use requires a final control element, a device that responds to the controller

More information

THE WAY THE VENTURI AND ORIFICES WORK

THE WAY THE VENTURI AND ORIFICES WORK Manual M000 rev0 03/00 THE WAY THE VENTURI AND ORIFICES WORK CHAPTER All industrial combustion systems are made up of 3 main parts: ) The mixer which mixes fuel gas with combustion air in the correct ratio

More information

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary ADH 1/7/014 LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the

More information

AIR EJECTOR WITH A DIFFUSER THAT INCLUDES BOUNDARY LAYER SUCTION

AIR EJECTOR WITH A DIFFUSER THAT INCLUDES BOUNDARY LAYER SUCTION Engineering MECHANICS, Vol. 20, 2013, No. 3/4, p. 213 220 213 AIR EJECTOR WITH A DIFFUSER THAT INCLUDES BOUNDARY LAYER SUCTION Václav Dvořák* The article deals with axial-symmetric subsonic air-to-air

More information

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE TELEDYNE HASTINGS TECHNICAL PAPERS INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE Proceedings of FEDSM 98 1998 ASME Fluids Engineering

More information

Experiment 8: Minor Losses

Experiment 8: Minor Losses Experiment 8: Minor Losses Purpose: To determine the loss factors for flow through a range of pipe fittings including bends, a contraction, an enlargement and a gate-valve. Introduction: Energy losses

More information

CHARACTERISTICS OF LEAK DETECTION BASED ON DIFERENTIAL PRESSURE MEASUREMENT

CHARACTERISTICS OF LEAK DETECTION BASED ON DIFERENTIAL PRESSURE MEASUREMENT CHARACTERISTICS OF LEAK DETECTION BASED ON DIFERENTIAL PRESSURE MEASUREMENT Harus L.G..*, Maolin CAI**, Kenji KAWASHIMA** and Toshiharu KAGAWA** * Graduate student of Mechano-Micro Engineering Dept., Tokyo

More information

Flow meter. bellow vaporizer. APL valve. Scavenging system

Flow meter. bellow vaporizer. APL valve. Scavenging system Introductory Lecture Series: The Anesthesia Machine PORNSIRI WANNADILOK Objectives Anesthesia Machine Ventilators Scavenging Systems System Checkout 1 Flow meter ventilator bellow vaporizer Corrugated

More information

A. M. Dalavi, Mahesh Jadhav, Yasin Shaikh, Avinash Patil (Department of Mechanical Engineering, Symbiosis Institute of Technology, India)

A. M. Dalavi, Mahesh Jadhav, Yasin Shaikh, Avinash Patil (Department of Mechanical Engineering, Symbiosis Institute of Technology, India) IOSR Journal of Mechanical and Civil Engineering (IOSR-JMCE) ISSN(e) : 2278-1684, ISSN(p) : 2320 334X, PP : 45-49 www.iosrjournals.org Modeling, Optimization & Manufacturing of Vortex Tube and Application

More information

Micro Channel Recuperator for a Reverse Brayton Cycle Cryocooler

Micro Channel Recuperator for a Reverse Brayton Cycle Cryocooler Micro Channel Recuperator for a Reverse Brayton Cycle Cryocooler C. Becnel, J. Lagrone, and K. Kelly Mezzo Technologies Baton Rouge, LA USA 70806 ABSTRACT The Missile Defense Agency has supported a research

More information

CORESTA RECOMMENDED METHOD N 6

CORESTA RECOMMENDED METHOD N 6 CORESTA RECOMMENDED METHOD N 6 DETERMINATION OF VENTILATION DEFINITIONS AND MEASUREMENT PRINCIPLES (2015 Revision September 2016) 1. SCOPE This CORESTA Recommended Method specifies a method for the determination

More information

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS Frank M. Gavila F&J SPECIALTY PRODUCTS, INC. 404 Cypress Road Ocala, Florida, USA 3447 fandj@fjspecialty.com INTRODUCTION The calibration

More information

Sign up to receive ATOTW weekly -

Sign up to receive ATOTW weekly - THE PHYSICS OF FLOW ANAESTHESIA TUTORIAL OF THE WEEK 84 9TH APRIL 2008 Paul Clements, SpR in Anaesthetics, Hope Hospital, Salford, UK. Carl Gwinnutt, Consultant Anaesthetist, Hope Hospital, Salford, UK.

More information

Operational Behaviour of Safety Valves with Constant Superimposed Backpressure

Operational Behaviour of Safety Valves with Constant Superimposed Backpressure Operational Behaviour of Safety Valves with Constant Superimposed Backpressure Arne Gastberg Albert Richter ARI-Armaturen GmbH & Co. KG When selecting spring loaded safety valves, an accurate knowledge

More information

Lesson 6: Flow Control Valves

Lesson 6: Flow Control Valves : Flow Control Valves Basic Hydraulic Systems Hydraulic Fluids Hydraulic Tank Hydraulic Pumps and Motors Pressure Control Valves Directional Control Valves Flow Control Valves Cylinders : Flow Control

More information

NUMERICAL INVESTIGATION OF THE FLOW BEHAVIOUR IN A MODERN TRAFFIC TUNNEL IN CASE OF FIRE INCIDENT

NUMERICAL INVESTIGATION OF THE FLOW BEHAVIOUR IN A MODERN TRAFFIC TUNNEL IN CASE OF FIRE INCIDENT - 277 - NUMERICAL INVESTIGATION OF THE FLOW BEHAVIOUR IN A MODERN TRAFFIC TUNNEL IN CASE OF FIRE INCIDENT Iseler J., Heiser W. EAS GmbH, Karlsruhe, Germany ABSTRACT A numerical study of the flow behaviour

More information

Smart Water Application Technologies (SWAT)

Smart Water Application Technologies (SWAT) Smart Water Application Technologies (SWAT) Turf and Landscape Irrigation Equipment PRESSURE REGULATING SPRAY HEAD SPRINKLERS Equipment Functionality Test Testing Protocol Version 3.0 (May 2012) Developed

More information

DPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE

DPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE DPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE INSTRUCTIONS FOR PORTABLE ANAESTHETIC MACHINE DPA 02 TM The Diamedica Portable Anaesthetic machine DPA 2 TM has three principal

More information

I. Subject: Compressed gas cylinders, regulators, and portable liquid systems.

I. Subject: Compressed gas cylinders, regulators, and portable liquid systems. I. Subject: Compressed gas cylinders, regulators, and portable liquid systems. II. Objective: To provide safe and effective use, storage, and transport of therapeutic gases. II. Procedure: A. Compressed

More information

2. Determine how the mass transfer rate is affected by gas flow rate and liquid flow rate.

2. Determine how the mass transfer rate is affected by gas flow rate and liquid flow rate. Goals for Gas Absorption Experiment: 1. Evaluate the performance of packed gas-liquid absorption tower. 2. Determine how the mass transfer rate is affected by gas flow rate and liquid flow rate. 3. Consider

More information

Gas Physics Pressure and Flow Topics Covered:

Gas Physics Pressure and Flow Topics Covered: Gas Physics Pressure and Flow Topics Covered: Molecular Theory of Gases Definition of Pressure The Gas Laws Definition of Flow Definition of Pressure Drop Gas Physics Pressure and Flow Topics Covered:

More information

Universities of Leeds, Sheffield and York

Universities of Leeds, Sheffield and York promoting access to White Rose research papers Universities of Leeds, Sheffield and York http://eprints.whiterose.ac.uk/ White Rose Research Online URL for this paper: http://eprints.whiterose.ac.uk/7819/

More information

Influence of rounding corners on unsteady flow and heat transfer around a square cylinder

Influence of rounding corners on unsteady flow and heat transfer around a square cylinder Influence of rounding corners on unsteady flow and heat transfer around a square cylinder S. K. Singh Deptt. of Mech. Engg., M. B. M. Engg. College / J. N. V. University, Jodhpur, Rajasthan, India Abstract

More information

The University of Hong Kong Department of Physics Experimental Physics Laboratory

The University of Hong Kong Department of Physics Experimental Physics Laboratory The University of Hong Kong Department of Physics Experimental Physics Laboratory PHYS2260 Heat and Waves 2260-1 LABORATORY MANUAL Experiment 1: Adiabatic Gas Law Part A. Ideal Gas Law Equipment Required:

More information

PIG MOTION AND DYNAMICS IN COMPLEX GAS NETWORKS. Dr Aidan O Donoghue, Pipeline Research Limited, Glasgow

PIG MOTION AND DYNAMICS IN COMPLEX GAS NETWORKS. Dr Aidan O Donoghue, Pipeline Research Limited, Glasgow PIG MOTION AND DYNAMICS IN COMPLEX GAS NETWORKS Dr Aidan O Donoghue, Pipeline Research Limited, Glasgow A model to examine pigging and inspection of gas networks with multiple pipelines, connections and

More information

Study on the Influencing Factors of Gas Mixing Length in Nitrogen Displacement of Gas Pipeline Kun Huang 1,a Yan Xian 2,b Kunrong Shen 3,c

Study on the Influencing Factors of Gas Mixing Length in Nitrogen Displacement of Gas Pipeline Kun Huang 1,a Yan Xian 2,b Kunrong Shen 3,c Applied Mechanics and Materials Online: 2013-06-13 ISSN: 1662-7482, Vols. 321-324, pp 299-304 doi:10.4028/www.scientific.net/amm.321-324.299 2013 Trans Tech Publications, Switzerland Study on the Influencing

More information

THEORETICAL EVALUATION OF FLOW THROUGH CENTRIFUGAL COMPRESSOR STAGE

THEORETICAL EVALUATION OF FLOW THROUGH CENTRIFUGAL COMPRESSOR STAGE THEORETICAL EVALUATION OF FLOW THROUGH CENTRIFUGAL COMPRESSOR STAGE S.Ramamurthy 1, R.Rajendran 1, R. S. Dileep Kumar 2 1 Scientist, Propulsion Division, National Aerospace Laboratories, Bangalore-560017,ramamurthy_srm@yahoo.com

More information

INTERACTION BETWEEN WIND-DRIVEN AND BUOYANCY-DRIVEN NATURAL VENTILATION Bo Wang, Foster and Partners, London, UK

INTERACTION BETWEEN WIND-DRIVEN AND BUOYANCY-DRIVEN NATURAL VENTILATION Bo Wang, Foster and Partners, London, UK INTERACTION BETWEEN WIND-DRIVEN AND BUOYANCY-DRIVEN NATURAL VENTILATION Bo Wang, Foster and Partners, London, UK ABSTRACT Ventilation stacks are becoming increasingly common in the design of naturally

More information

Pressure Control. where: p is the pressure F is the normal component of the force A is the area

Pressure Control. where: p is the pressure F is the normal component of the force A is the area Pressure Control First of all, what is pressure, the property we want to control? From Wikipedia, the free encyclopedia. Pressure is the application of force to a surface, and the concentration of that

More information

Methods of Rating Flow

Methods of Rating Flow Methods of Rating Flow What engineers need to know when using CV as a measurement of flow in pneumatic applications In the pneumatic industry, CV is one standard for expressing the flow capacity of devices

More information

Design Features. General Description

Design Features. General Description Model thermal Mass Flow Controllers are designed to indicate and control set flow rates of gases. The combines the characteristics and accuracy of conventional mass flow devices into a unique compact design

More information

OIL AND GAS INDUSTRY

OIL AND GAS INDUSTRY This case study discusses the sizing of a coalescer filter and demonstrates its fouling life cycle analysis using a Flownex model which implements two new pressure loss components: - A rated pressure loss

More information

Appendix A Frequently Asked Questions

Appendix A Frequently Asked Questions Appendix A Frequently Asked Questions This set of Frequently Asked Questions has been supplied courtesy of Mitech Spirex Sarco Ltd, South Africa, with acknowledgements and thanks to Michael Sessions. Questions

More information

Fluid Flow. Link. Flow» P 1 P 2 Figure 1. Flow Model

Fluid Flow. Link. Flow» P 1 P 2 Figure 1. Flow Model Fluid Flow Equipment: Water reservoir, output tubes of various dimensions (length, diameter), beaker, electronic scale for each table. Computer and Logger Pro software. Lots of ice.temperature probe on

More information

Research and optimization of intake restrictor for Formula SAE car engine

Research and optimization of intake restrictor for Formula SAE car engine International Journal of Scientific and Research Publications, Volume 4, Issue 4, April 2014 1 Research and optimization of intake restrictor for Formula SAE car engine Pranav Anil Shinde Mechanical Engineering,

More information

****** * EX * ****** DWN W.W.POWELL CALCULATION OF FLOW LOSSES IN INLET CHK D.PAPA AND DISCHARGE HEADERS ASSOCIATED WITH

****** * EX * ****** DWN W.W.POWELL CALCULATION OF FLOW LOSSES IN INLET CHK D.PAPA AND DISCHARGE HEADERS ASSOCIATED WITH BC ****** * EX * ****** DWN W.W.POWELL 4-17-75 CALCULATION OF FLOW LOSSES IN INLET CHK D.PAPA 4-17-75 AND DISCHARGE HEADERS ASSOCIATED WITH APPR APPR SAFETY RELIEF VALVES SIZE REV APPR A 02.0175.128 L

More information

Cooling Characteristics of GM-type Pulse Tube Refrigerator with Neon as Working Gas

Cooling Characteristics of GM-type Pulse Tube Refrigerator with Neon as Working Gas Cooling Characteristics of GM-type Pulse Tube Refrigerator with Neon as Working Gas J. Ko, Y.J. Hong, H. Kim, H. Yeom, S.J. Park, D.Y. Koh Korea Institute of Machinery & Materials Daejeon, Korea (S), 305-343

More information

Investigation on Divergent Exit Curvature Effect on Nozzle Pressure Ratio of Supersonic Convergent Divergent Nozzle

Investigation on Divergent Exit Curvature Effect on Nozzle Pressure Ratio of Supersonic Convergent Divergent Nozzle RESEARCH ARTICLE OPEN ACCESS Investigation on Divergent Exit Curvature Effect on Nozzle Pressure Ratio of Supersonic Convergent Divergent Nozzle Shyamshankar.M.B*, Sankar.V** *(Department of Aeronautical

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.12 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient,

More information

The Estimation Of Compressor Performance Using A Theoretical Analysis Of The Gas Flow Through the Muffler Combined With Valve Motion

The Estimation Of Compressor Performance Using A Theoretical Analysis Of The Gas Flow Through the Muffler Combined With Valve Motion Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering The Estimation Of Compressor Performance Using A Theoretical Analysis Of The Gas Flow Through

More information

Lab. Manual. Fluid Mechanics. The Department of Civil and Architectural Engineering

Lab. Manual. Fluid Mechanics. The Department of Civil and Architectural Engineering Lab. Manual of Fluid Mechanics The Department of Civil and Architectural Engineering General Safety rules to be followed in Fluid Mechanics Lab: 1. Always wear shoes before entering lab. 2. Do not touch

More information

PERFORMANCE OF A FLAPPED DUCT EXHAUSTING INTO A COMPRESSIBLE EXTERNAL FLOW

PERFORMANCE OF A FLAPPED DUCT EXHAUSTING INTO A COMPRESSIBLE EXTERNAL FLOW 24 TH INTERNATIONAL CONGRESS OF THE AERONAUTICAL SCIENCES PERFORMANCE OF A FLAPPED DUCT EXHAUSTING INTO A COMPRESSIBLE EXTERNAL FLOW P. R. Pratt, J. K. Watterson, E. Benard, S. Hall School of Aeronautical

More information

FUNDAMENTALS OF PRESSURE REGULATORS ROBERT BENNETT MANAGER OF TRAINING ELSTER AMERICAN METER

FUNDAMENTALS OF PRESSURE REGULATORS ROBERT BENNETT MANAGER OF TRAINING ELSTER AMERICAN METER FUNDAMENTALS OF PRESSURE REGULATORS ROBERT BENNETT MANAGER OF TRAINING ELSTER AMERICAN METER SUPPLY = DEMAND FUNCTION OF A REGULATOR A regulator may be defined as a "mechanism for controlling or governing

More information

An Experimental Performance Study of Vortex Tube Refrigeration System

An Experimental Performance Study of Vortex Tube Refrigeration System An Experimental Performance Study of Vortex Tube Refrigeration System Sankar Ram T. Department Of Industrial Refrigeration and Cryogenics T. K. M. College of Engineering Karicode, Kollam, Kerala sankarram9@gmail.com

More information

THE IMPLEMENTATION OF TOROIDAL THROAT VENTURI NOZZLES TO MAXIMIZE PRECISION IN GAS FLOW TRANSFER STANDARD APPLICATIONS

THE IMPLEMENTATION OF TOROIDAL THROAT VENTURI NOZZLES TO MAXIMIZE PRECISION IN GAS FLOW TRANSFER STANDARD APPLICATIONS Presented 2005 JUNE FLOMEKO Peebles, SCOTLAND THE IMPLEMENTATION OF TOROIDAL THROAT VENTURI NOZZLES TO MAXIMIZE PRECISION IN GAS FLOW TRANSFER STANDARD APPLICATIONS Pierre Delajoud, Martin Girard, Michael

More information

MEASUREMENT OF GAS DISCHARGE COEFFICIENT

MEASUREMENT OF GAS DISCHARGE COEFFICIENT MEASUREMENT OF GAS DISCHARGE COEFFICIENT Ilona Pasková, Ondřej Novák, Václav Koza RWE GasNet, ilona.paskova@rwe.cz ICT Prague, Department of gas, coke and air protection, ondrej.novak@vscht.cz, vaclav.koza@vscht.cz

More information

Cover Page for Lab Report Group Portion. Compressible Flow in a Converging-Diverging Nozzle

Cover Page for Lab Report Group Portion. Compressible Flow in a Converging-Diverging Nozzle Cover Page for Lab Report Group Portion Compressible Flow in a Converging-Diverging Nozzle Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 13 January 2012 Name 1: Name 2: Name

More information

Tutorial. BOSfluids. Relief valve

Tutorial. BOSfluids. Relief valve Tutorial Relief valve The Relief valve tutorial describes the theory and modeling process of a pressure relief valve or safety valve. It covers the algorithm BOSfluids uses to model the valve and a worked

More information

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855

More information

29 Pressure, Temperature relationship of a gas

29 Pressure, Temperature relationship of a gas Chemistry Sensors: Loggers: Gas Pressure, Temperature Any EASYSENSE Logging time: EasyLog Teacher s notes 29 Pressure, Temperature relationship of a gas Read The ideal gas laws tell us that if we keep

More information

Fluid-Structure Interaction Analysis of a Flow Control Device

Fluid-Structure Interaction Analysis of a Flow Control Device Abaqus Technology Brief Fluid-Structure Interaction Analysis of a Control Device TB-06-FSI-2 Revised: April 2007. Summary The Vernay VernaFlo flow controls are custom-designed fluid flow management devices

More information

Gerald D. Anderson. Education Technical Specialist

Gerald D. Anderson. Education Technical Specialist Gerald D. Anderson Education Technical Specialist The factors which influence selection of equipment for a liquid level control loop interact significantly. Analyses of these factors and their interactions

More information

TUTORIAL. NPSHA for those who hate that stuffy word. by Jacques Chaurette p. eng. copyright 2006

TUTORIAL. NPSHA for those who hate that stuffy word. by Jacques Chaurette p. eng.  copyright 2006 TUTORIAL NPSHA for those who hate that stuffy word by Jacques Chaurette p. eng. www.lightmypump.com copyright 2006 page.2 NPSHA for those who hate that stuffy word This article follows the same approach

More information

485 Annubar Primary Flow Element Installation Effects

485 Annubar Primary Flow Element Installation Effects ROSEMOUNT 485 ANNUBAR 485 Annubar Primary Flow Element Installation Effects CONTENTS Mounting hole diameter Alignment error Piping Geometry Induced Flow Disturbances Pipe reducers and expansions Control

More information

PROCESS MONITORING AND CONTROL

PROCESS MONITORING AND CONTROL Chemical Engineering Process Design PROCESS MONITORING AND CONTROL Keith Marchildon David Mody Monitoring means Measurement 1. There is no Control without Monitoring 2. Monitoring should not be done without

More information

Instruction Manual. Pipe Friction Training Panel

Instruction Manual. Pipe Friction Training Panel Instruction Manual HL 102 Pipe Friction Training Panel 100 90 80 70 60 50 40 30 20 10 HL 102 Instruction Manual This manual must be kept by the unit. Before operating the unit: - Read this manual. - All

More information

STANDARD FOR CONTROL VALVE SEAT LEAKAGE

STANDARD FOR CONTROL VALVE SEAT LEAKAGE 11-3 STANDARD FOR CONTROL VALVE SEAT LEAKAGE 1. PURPOSE 1.1 This standard establishes a series of seat leakage classes for control valves and defines the test procedures. 2. SCOPE & LIMITATIONS 2.1 Selection

More information

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cover Page for Lab Report Group Portion. Head Losses in Pipes Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section

More information

H16 Losses in Piping Systems

H16 Losses in Piping Systems H16 Losses in Piping Systems The equipment described in this manual is manufactured and distributed by TECQUIPMENT LIMITED Suppliers of technological laboratory equipment designed for teaching. BONSALL

More information

MEMORIAL UNIVERSITY OF NEWFOUNDLAND Faculty of Engineering and Applied Science FLUID MECHANICS LABORATORY PIPE FRICTION

MEMORIAL UNIVERSITY OF NEWFOUNDLAND Faculty of Engineering and Applied Science FLUID MECHANICS LABORATORY PIPE FRICTION MEMORIAL UNIVERSITY OF NEWFOUNDLAND Faculty of Engineering and Applied Science FLUID MECHANICS LABORATORY PIPE FRICTION Objective To estimate the fluid pressure drops and roughness specifications for copper

More information

CFD Simulation and Experimental Validation of a Diaphragm Pressure Wave Generator

CFD Simulation and Experimental Validation of a Diaphragm Pressure Wave Generator CFD Simulation and Experimental Validation of a Diaphragm Pressure Wave Generator T. Huang 1, A. Caughley 2, R. Young 2 and V. Chamritski 1 1 HTS-110 Ltd Lower Hutt, New Zealand 2 Industrial Research Ltd

More information

Experiment Instructions. Circulating Pumps Training Panel

Experiment Instructions. Circulating Pumps Training Panel Experiment Instructions Circulating Pumps Training Panel Experiment Instructions This manual must be kept by the unit. Before operating the unit: - Read this manual. - All participants must be instructed

More information

By Syed Ahmed Amin Shah 4 th semester Class No 8 Submitted To Engr. Saeed Ahmed

By Syed Ahmed Amin Shah 4 th semester Class No 8 Submitted To Engr. Saeed Ahmed EXPERIMENT # 01 DEMONSTRATION OF VARIOUS PARTS OF HYDRAULIC BENCH. HYDRAULIC BENCH Hydraulic bench is a very useful apparatus in hydraulics and fluid mechanics it is involved in majority of experiments

More information

CHAPTER 3 RESULTS AND DISCUSSION

CHAPTER 3 RESULTS AND DISCUSSION CHAPTER 3 RESULTS AND DISCUSSION The research included the development and verification of the earlier apparatus, and the collection and validation of moisture diffusion data under both isothermal and

More information

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products Saturn Evo SATURN EVO COLOR A N A E S T H E S I A S Y S T E M i n t e l l i g e n t s i m p l i c i t y Manufacturer of high quality anaesthesia products SATURN EVO COLOR PATIENT SAFETY AND USER PATIENT

More information

AC : A LABORATORY EXERCISE TO DEMONSTRATE HOW TO EXPERIMENTALLY DETERMINE THE OPERATING POINT FOR A FAN

AC : A LABORATORY EXERCISE TO DEMONSTRATE HOW TO EXPERIMENTALLY DETERMINE THE OPERATING POINT FOR A FAN AC 2007-206: A LABORATORY EXERCISE TO DEMONSTRATE HOW TO EXPERIMENTALLY DETERMINE THE OPERATING POINT FOR A FAN Robert Edwards, Pennsylvania State University-Erie Robert Edwards is currently a Lecturer

More information

Gas Laws. Introduction

Gas Laws. Introduction Gas Laws Introduction In 1662 Robert Boyle found that, at constant temperature, the pressure of a gas and its volume are inversely proportional such that P x V = constant. This relationship is known as

More information

Seminar 65 Compression Challenges for Low GWP Refrigerants

Seminar 65 Compression Challenges for Low GWP Refrigerants Margaret Mathison, Ph.D. mm1@iastate.edu Seminar 65 Compression Challenges for Low GWP Refrigerants Design Improvements of the Spool Compressor for Various Working Fluids using Comprehensive Modeling Techniques

More information

CHEM 355 EXPERIMENT 7. Viscosity of gases: Estimation of molecular diameter

CHEM 355 EXPERIMENT 7. Viscosity of gases: Estimation of molecular diameter CHEM 355 EXPERIMENT 7 Viscosity of gases: Estimation of molecular diameter Expressed most simply, the viscosity of a fluid (liquid or gas) relates to its resistance to flow. The viscosity of a gas is determined

More information

Investigation of Suction Process of Scroll Compressors

Investigation of Suction Process of Scroll Compressors Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering 2006 Investigation of Suction Process of Scroll Compressors Michael M. Cui Trane Jack Sauls

More information

MODELLING OF FUME EXTRACTORS C. R.

MODELLING OF FUME EXTRACTORS C. R. LD8 19th International Symposium of Ballistics, 7 11 May 21, Interlaken, Switzerland MODELLING OF FUME EXTRACTORS C. R. Woodley WS4 Guns and Warheads Department, Defence Evaluation and Research Agency,

More information

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV 16-1 INTRODUCTION 16-1.1 Purpose. The purpose of this chapter is to familiarize divers with the techniques used to mix divers breathing gas. 16-1.2 Scope. This chapter

More information

I.CHEM.E. SYMPOSIUM SERIES NO. 97 BUOYANCY-DRIVEN NATURAL VENTILATION OP ENCLOSED SPACES

I.CHEM.E. SYMPOSIUM SERIES NO. 97 BUOYANCY-DRIVEN NATURAL VENTILATION OP ENCLOSED SPACES BUOYANCY-DRIVEN NATURAL VENTILATION OP ENCLOSED SPACES M. R. Marshall* and P. L. Stewart-Darling* A simple mathematical model for the buoyancy driven ventilation of an enclosed space, using a two-pipe

More information

Lab Problems. Lab Problems for Chapter Fluid Characterization by Use of a Stormer Viscometer L-1

Lab Problems. Lab Problems for Chapter Fluid Characterization by Use of a Stormer Viscometer L-1 Lab Problems This section contains end-of-the-chapter problems that involve data obtained from various simple laboratory experiments. These lab problems for any chapter can be obtained by clicking on the

More information

Theoretical Modelling of Rotary Valve Air Leakage for Pneumatic Conveying Systems

Theoretical Modelling of Rotary Valve Air Leakage for Pneumatic Conveying Systems University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 2002 Theoretical Modelling of Rotary Valve Air Leakage for Pneumatic Conveying

More information