The use of accelerometry to detect heel contact events for use as a sensor in FES assisted walking

Size: px
Start display at page:

Download "The use of accelerometry to detect heel contact events for use as a sensor in FES assisted walking"

Transcription

1 Medical Engineering & Physics 25 (2003) Technical note The use of accelerometry to detect heel contact events for use as a sensor in FES assisted walking Avril Mansfield, Gerard M. Lyons Biomedical Electronics Laboratory, Department of Electronic and Computer Engineering, University of Limerick, Limerick, Ireland Received 1 November 2002; received in revised form 31 March 2003; accepted 20 June 2003 Abstract Current sensors for the control of functional electrical stimulation (FES) assisted walking in hemiplegic individuals are not wholly satisfactory, as they are either not implantable or ineffectual in the detection of heel contact events. This study describes the use of an accelerometer placed on the trunk to detect heel contact events of both legs based on the examination of the anterior posterior horizontal acceleration signal. Four subjects wore an accelerometer over their lumbar spine. Footswitches placed on the sole of one foot recorded the heel contact and heel off times for that foot. The acceleration signal was reduced to a series of pulses by studying the negative positive changes in acceleration. It was found that there was approximately a 150 ms delay between heel contact and the negative positive change in acceleration. This delay was consistent across different walking speeds, but was different between subjects and when hemiplegic gait was simulated. Therefore, accelerometers placed on the trunk are valid sensors for the detection of heel contact events during FES assisted walking IPEM. Published by Elsevier Ltd. All rights reserved. Keywords: FES; Accelerometry; Hemiplegia; Gait events; Sensors 1. Introduction Since the 1960s, functional electrical stimulation (FES) has been used to correct foot drop in stroke patients [1]. It works by delivering an electrical stimulus to the tibialis anterior during the swing phase of gait to allow proper foot clearance and prevent the need for hip circumduction or other unstable gait compensations. While FES systems have proved successful in improving walking ability in stroke patients by improving walking speed, functional mobility, step length on the affected side, ankle flexion, muscle strength, motor coordination and balance [2 8], there are some fundamental problems associated with their use. Baker [9] has identified two such problems; the complexity of the gait cycle and the need for implantation of the system, and both problems put demands on the sensor and control system used. For any FES system to work in an effective manner, Corresponding author. Tel.: ; fax: address: gerard.lyons@ul.ie (G.M. Lyons). it requires a sensor to detect when the stimulus should be delivered. At first, the system can be initiated manually by an external switch. However, during the gait cycle itself, stimulation should be automatic. Therefore, a sensor is required to detect the phases of the gait cycle, so that the stimulation can be delivered during the swing phase of gait on the affected side. In order to do this effectively, the sensor must be able to detect heel off of the affected limb to trigger stimulation just prior to the swing phase and either heel contact or heel off of the contra-lateral leg to stop stimulation. Originally, footswitches, usually based on force sensitive resistors, were placed on the sole of the foot and used as sensors for FES [2]. This seems like the natural choice of sensor; heel loading and unloading can be detected with a footswitch. However, the footswitch as a sensor for FES assisted walking correction has many problems associated with it, most notably, reliability and the lack of implantation potential. The reliability of footswitches is often diminished due to their tendency to detect heel contact events during the swing phase as small forces are exerted on the heel during swing. Similarly, the reliability of the footswitch is reduced in /$ IPEM. Published by Elsevier Ltd. All rights reserved. doi: /s (03)

2 880 A. Mansfield, G.M. Lyons / Medical Engineering & Physics 25 (2003) patients who shuffle their feet when they walk [10,11]. The implantability of the stimulator is important as patients often reject systems that require a lot of donning time or are not aesthetic. This problem may be addressed through the use of implanted cuff electrodes placed on the sural nerve to use the patient s own sensory nerves to detect heel contact events [12]. However, such sensors are still ineffective in patients who shuffle their feet when they walk [10,11]. Hand switches have also been used to trigger stimulation, which may be preferable in the event of encountering an obstacle or uneven terrain [13]. However, hand switches increase the attentional demands of walking, which may reduce the patient s capacity to respond to perturbations to their balance [14], leading to an increased risk of a fall event. Accelerometers have the potential to be used as sensors for FES control systems. Accelerometers have been used previously to detect phases of the gait cycle [15], although these authors have placed the accelerometer on the leg, which requires more involved data processing. Implantation of the accelerometer could be carried out in the abdominal cavity of the patient, which would allow for implantation of both stimulator and sensor at the same site. Placement of the accelerometer on the trunk facilitates detection of phases of the gait cycle using changes in measured acceleration patterns and a trunk-mounted accelerometer has also been shown to measure instantaneous walking speed [16], which could provide another control input for stimulus intensity. From the placement of the accelerometer on the trunk, it may be possible to detect phases of the gait cycle by changes in acceleration patterns. It was the purpose of this study to determine the potential of accelerometry to detect gait events for their prospective use in a FES stimulator for the correction of drop foot in hemiplegic patients. 2. Theoretical background Changes in the kinematic parameters of gait are cyclical. That is, the shape of the acceleration curve recorded for one step is repeated for every step during continuous walking. Therefore, by studying the acceleration signal throughout the gait cycle, it is possible to determine gait events. This occurrence has already been exploited by Currie et al. [17] to detect gait cycle events by examining the acceleration time curve during walking. However, these authors used manual inspection to determine heel contact events rather than an automated computer algorithm. Fig. 1A shows a sample of an anterior posterior (AP) approximated centre of mass (CoM) acceleration signal to illustrate that this signal is cyclical during continuous walking. In order to mark the beginning and end of a step on this figure, a simultaneous footswitch output is shown. The footswitch signal measured heel contact events from one foot alone. According to this output, the heel is in contact with the ground, when it reads 1 and unloaded when it reads 0. It is evident from this graph that positive peaks in AP acceleration during walking occur just prior to unloading of the heel, whereas negative peaks in AP acceleration occur immediately following heel contact. By observing the positive/negative changes in AP acceleration throughout the gait cycle, one can predict heel contact and heel unloading times. The acceleration signal as presented in Fig. 1A is difficult to interpret without using complicated rules. Therefore, in order to simplify the algorithm, the acceleration signal can be reduced to a pulse-like signal similar to the footswitch signal. This is done by assigning a value 1 if the acceleration signal is increasing (a negative positive change) and a value 0 if the acceleration signal is decreasing (representing a positive-negative change). This results in the acceleration pulse presented in Fig. 1B. The data presented in this figure is the same as that presented in Fig. 1A, with the acceleration signal reduced to pulses. There are a few important points evident upon inspection of Fig. 1B. The first is that there are two acceleration pulses for every footswitch pulse. Therefore, every alternate acceleration pulse recorded coincides with the stance phase of alternate feet. Furthermore, it seems that analysis of the negative positive change in the acceleration signal is most relevant to the prediction of heel contact events, with each negative positive change occurring after heel contact of the alternating feet. We can therefore assume that each negative positive change occurs at a constant time after the heel contact as measured by the footswitch. Moreover, if this constant can be determined experimentally, it is possible to predict heel contact events from both feet from the accelerometer output. 3. Methods This study aimed to assess the potential use of accelerometry as a sensor for the control of FES assisted walking for the correction of drop foot in hemiplegic patients. Therefore, the ability of the accelerometer to detect heel contact events was compared to that of a footswitch. In order to test the ability of the accelerometer to detect heel contact events, experimental trials were performed on four subjects. Ethical approval for this study was first obtained from the University of Limerick Research and Ethics Committee. All subjects gave their informed consent to participation. The subjects were aged between 22 and 24 years and had no musculoskeletal impairment or injury that affected their balance or gait. Table 1 details the subject information.

3 A. Mansfield, G.M. Lyons / Medical Engineering & Physics 25 (2003) Fig. 1. (A) (top) shows changes in AP horizontal acceleration during walking. The broken line shows the output from the footswitch (with 1 indicating on and 0 indicating off ) marking the beginning and end of each stride. The lower figure 9 (B) shows the reduction of the acceleration curve presented in (A) into a pulse-like signal. A value of 1 indicates a negative positive change (or increase) in the acceleration signal, while a value of 0 indicates a positive-negative change (or decrease) in the acceleration signal. The footswitch output is also shown for comparative purposes. Table 1 Information on subjects tested Subject Gender Age (years) Height (m) A Female B Female C Male D Female One accelerometer (ADXL202, Analog Devices Inc.) was attached to the lumbar region of the subjects. This accelerometer is a dual-axis accelerometer, but acceleration from only one axis was considered, that representing anterior posterior (AP) acceleration of the subject during walking. A 100 Hz anti-aliasing filter was applied to the accelerometer signal prior to sampling. The accelerometer was placed in a plastic box for stability, which was mounted on a plastic board and attached to some Velcro straps, allowing secure fastening to the subject with minimal movement. The accelerometer was placed in alignment with the lumbar spine, with the active axis perpendicular to the spine. The beginning of each stride was marked with a footswitch. This consisted of a force sensing resistor, configured in a voltage divider arrangement, placed on the heel of one foot; the affected side in simulated hemiplegic walking. The footswitch signal was processed using a threshold value set at 97.5%, so that all values, 97.5% or greater of the maximum value represented on, and all values less than 97.5% of the maximum represented off. It was found that this threshold level resulted in no false heel off detections (i.e. heel off being detected during stance) and fewer false heel contact detections (i.e. heel on being detected during swing). The heel contact time for each stride was then found by recording the time at which each on condition occurred. The accelerometer and footswitch were connected to a portable data logger (BM42, 32MB version, Biomedical Monitoring Ltd.), which allowed for the collection of data from walking trials. The three channels were sampled at 500 Hz for subject C, and 200 Hz for all other subjects. The data was then transferred via an external disk drive (Securmate SDDR-33, Sandisk) onto a CPU for post processing using MATLAB (version 4.2b, The Mathworks, Inc.). Gait trials were performed along a 40 m length of corridor. The surface of the corridor was covered in shallow carpeting. The subject was asked to walk in a straight line during the trials. Three different groups of trials were performed which, for the purposes of this analysis, were grouped as either steady, changing or hemiplegic walking. Steady trials involved the subject walking either as normal, or at a constant speed determined by a metronome. Cadences from 60 to 140 steps

4 882 A. Mansfield, G.M. Lyons / Medical Engineering & Physics 25 (2003) per minute were tested using the metronome. During changing trials, subjects were asked to alternate their walking speed between normal, faster than normal or slower than normal about half way down the corridor. Hemiplegic walking was simulated by halving the step length on the unaffected side. Subjects were also asked to walk slower than normal during these trials and as if they had a limp. Forty trials in all were performed with each subject; 24 steady trials, 8 changing trials and 8 hemiplegic trials. Post processing using MATLAB low-pass filtered the acceleration data at 2 Hz and compensated for the component of gravity measured during the trials by measuring the average angle of tilt of the accelerometer during quiet stance at the start of the trial. The timing of negative positive changes in acceleration values were calculated by reducing the acceleration signal to a series of pulses as described in the previous section. The times of heel contact as determined by the footswitch were also calculated. However, there were two negative positive acceleration changes for every stride, as each negative positive change occurs synchronously with foot contact of each foot. Therefore, the MATLAB routine deleted every second value from the negative positive timing vector produced. It did this by detecting the first acceleration pulse immediately following heel contact as detected by the footswitch. This was the first event of note from the accelerometer, and all acceleration pulses recorded previous to this were assumed to be due to postural sway or noise. The acceleration pulse immediately following this first event was deleted, and every second subsequent acceleration pulse was deleted to give unilateral information. Initially, the gait timing data was not synchronised exactly. In some instances, either the footswitch or the accelerometer falsely detected a heel contact event (i.e. detected more than one event per gait cycle). These values were removed manually using Miscrosoft Excel (Version 9.0, Microsoft Corporation). The percentage of incorrect foot event detections for both the footswitch and the accelerometer was recorded. Following this, the delay between foot contact timing as calculated by the footswitch and the negative positive acceleration changes was found for each step. A oneway ANOVA on SPSS (release , SPSS Inc.) was used to determine if these differences were the same for all subjects and for the three different types of trial (i.e. steady, changing and hemiplegic). Post-hoc analysis using Tukey s HSD determined which, if any, of the three groups of trials were different to each other. 4. Results A total of four subjects completed 40 gait trials each. Six trials were removed from analysis due to a fault with the recording device, leaving 154 trials for analysis. This resulted in a total of 2626 strides. Both the accelerometer and the footswitch were not wholly accurate in the prediction of heel contact events. In some instances, they falsely indicated a heel contact event. The error in detecting heel contact events was determined manually on Microsoft Excel, and the results of this determination are found in Table 2. From this table, it is evident that the accelerometer was between 98.2% and 99.8% reliable in the detection of heel contact events, whereas the footswitch is between 92.4% and 98.7% reliable when all the trials are considered together. Analysis of steady trials alone for all subjects pooled, revealed that there was on average a 147 ms (±91 ms standard deviation) delay in heel contact as measured from the footswitch and the negative positive change in AP acceleration. Therefore, there was a consistent and repeatable relationship between the output from the footswitches and the output from the accelerometer in terms of predicting heel contact events. Analysis of the mean difference in timing of rise in footswitch voltage and negative positive change in the accelerometer output between subjects revealed that there is large variation in this mean difference between subjects (p 0.001). This analysis included all trials. Table 3 shows the average difference for each subject. While the variation in the average delay for all subjects pooled was quite large (standard deviation of 91 ms), this variation greatly reduces when the results for each different subject are presented individually. It is possible that a large proportion of this variation is due to differing placement of the footswitch during the trials. While every effort was made to ensure that the footswitch was placed in approximately the same place for every subject, differences in foot and shoe size and the type of shoe worn between subjects may have meant that there were slight delays in the loading of the footswitch. However, it is also likely that the mean difference is due to differences in walking strategy between subjects. The three different trial conditions, steady, changing and hemiplegic, were then compared within subjects in terms of the delay in heel contact identification and negative positive acceleration change. Analysis revealed that these delays were statistically different across trial conditions (p 0.001). However, post-hoc using Tukey s HSD analysis revealed that the steady and changing conditions were similar to each other (p = 0.991). Fig. 2 illustrates this for all trials for all subjects. 5. Discussion Results from this study indicate that there was a negative positive change in acceleration approximately 150 ms after heel contact as identified by the foot switch. However, there was large variation observed in this figure, and while this variation was reduced when individ-

5 A. Mansfield, G.M. Lyons / Medical Engineering & Physics 25 (2003) Table 2 Error in heel contact event prediction from the footswitch and the accelerometer for all trials (steady, changing and hemiplegic), normal walking alone and simulated hemiplegic walking. The values displayed reflect the percentage of times either the footswitch or the accelerometer correctly detected a heel contact event Subject Correct footswitch detection (%) Correct accelerometer detection (%) All Normal Hemiplegic All Normal Hemiplegic A B C D Table 3 Average delay in heel contact as determined by the footswitch and the negative positive change in AP horizontal acceleration Subject Delay (ms) Standard deviation (ms) A B C D Fig. 2. The average delay in heel contact and negative positive acceleration change for all four subjects for each type of trial. Significant differences in these delays were found between subjects (p 0.001). Significant differences were also found between the simulated hemiplegic trials and the steady/changing trials (p 0.001). ual subjects data was considered, the standard deviations observed were still quite high (see Table 3). This variation may be largely due to the limitations of the use of thresholding for determination of heel contact events from the footswitch. That is, the use of an absolute voltage, above which is considered on and below which is considered off. The value used in this study (97.5% of maximum voltage) yielded fewer false gait event detections, although there were still more detected than with the accelerometer (from Table 2). More sophisticated algorithms may be used for processing the footswitch data and subsequently determining heel contact events, although the methods used here reflect those typically used in FES applications [5]. Furthermore, different timing in negative positive changes in acceleration was found when hemiplegic walking was simulated. These different timing patterns observed indicate that when gait is altered significantly (e.g. made asymmetrical), the timing of trunk acceleration/deceleration in relation to heel contact changes. This same difference is not observed with simply a change in walking speed. This occurrence has a significant effect on the clinical use of the accelerometer in an FES system. That is, in order to determine the temporal relationship between trunk acceleration changes and FES stimulation, pathological subjects will have to be used. However, further research is required to determine the exact nature of this relationship. The concept of the precise detection of heel contact events is somewhat misleading. The transitions between the phases of the gait cycle are gradual and very often two different and equally experienced investigators may disagree on the exact moment of, for example, heel contact based on video recordings. This proves problematic in the automatic detection of gait events using artificial sensors. While footswitches can provide a simple on/off output for that part of the foot they occupy, the relatively slow loading and unloading of that part of the foot can distort the results obtained. Therefore, no sensor can be deemed wholly accurate in the precise detection of gait events. The inadequacy of footswitches in detecting foot contact events is evident in this paper. Footswitches, on average, detected heel contact falsely % of the time (see Table 2). It is thought that pressure from the sole of the shoe on the footswitch caused some on footswitch signals being output during the swing phase. Furthermore, relatively slow unloading of the heel may be responsible for false heel contact detection immediately following heel off. Indeed, these false heel contact detections are evident in Figs. 1 and 2 immediately following the first and second stride. Shuffling of the feet, leading to an unclear heel contact, can also cause these false heel contact detections. By contrast, the accelerometer was

6 884 A. Mansfield, G.M. Lyons / Medical Engineering & Physics 25 (2003) less likely to output a negative positive acceleration change at an inappropriate time in the gait cycle. Noise in the acceleration signal can obviously cause such contamination of the signal, but filtering at 2 Hz removed most of this noise and resulted in % false negative positive changes detected for all trials. Furthermore, positive/negative changes recorded during quiet stance due to postural sway could result in the false detection of foot contact events. Therefore, some rule should be implemented in the sensor, stating that walking speed should be above a predetermined amount before initiation of the stimulus. Despite the shortcomings in footswitches for the detection of heel contact events, it is generally accepted in the research literature that footswitches are the gold standard in heel contact timing detection, particularly for use as FES sensors [2]. However, their use is limited in stroke patients and other such individuals whose gait is such that they shuffle their feet when they walk [10,11]. Indeed, this study supports this assertion, with improved gait event detection of the accelerometer in two subjects when compared with the footswitch during simulated hemilpegic walking (subjects C and D, Table 2). The improved performance was more prevalent in subject D, who from observation during the trials, seemed to have a less defined heel contact than the other subjects. Therefore, the accelerometer seems a better choice of sensor for individuals with a poorly defined heel contact, such as those with drop foot. Even during the normal walking in our non-pathological subjects, the accelerometer did not detect any false gait events, but the footswitch did, again possibly due to undefined heel contact or insole pressure during swing. Decreased performance of the accelerometer compared with the footswitch was observed for subjects A and B during simulated hemiplegic walking. While every effort was made to ensure that the subjects were familiar with the simulated hemiplegic walking patterns, this could be a result of unnecessary accelerations and decelerations of the trunk caused by the atypical walking patterns required. This is not the first study to use accelerometry to detect heel contact events during walking. As previously mentioned, Currie et al. [17] used manual inspection of acceleration patterns observed during walking to record the timing of heel contact events. Other studies [15,18,19] have used accelerometers placed on the leg to detect heel contact events. Indeed, Williamson and Andrews [15] have stated that a sensor which is capable of detecting heel contact events and measuring joint angle information may be necessary and the accelerometer is capable of meeting both of these criteria. However, this is the first study to report the automated determination of heel contact timing events based on the analysis of acceleration curves of an approximation of the centre of mass. By placing the accelerometer on the trunk, the detection algorithm is simplified; the acceleration signals presented in this study, in Figs. 1 and 2, are much easier to interpret than those reported by Williamson and Andrews [18]. Additionally, the accuracy of heel contact detection found in this study is comparable to accuracies obtained previously, when more involved rules for heel contact detection were used [15]. Furthermore, the accelerometer may be implanted in the abdomen to facilitate a fully implantable FES system. One of the most important issues when dealing with accelerometers is the need to correct for the gravitational component of measured acceleration. While this was corrected for in this study, compensation for the tilt of the accelerometer may not be necessary when the accelerometer is used solely for the detection of heel contact events as it is the shape of the acceleration curve, rather than the values obtained that are studied. 6. Conclusions The results of this study indicate the possibility of a lightweight, implantable sensor for the detection of gait events. An accelerometer may be placed on the trunk of an individual and a simple yet effective algorithm can be used to automatically detect heel contact of both feet. Acknowledgements The authors wish to acknowledge the assistance of Ms Karen Culhane and Ms Elizabeth Egan in the completion of this project and the preparation of this manuscript. References [1] Liberson WT, Holmquest HJ, Scott D, Dow M. Functional electrotherapy: stimulation of the peroneal nerve synchronized with the swing phase of the gait of hemiplegic patients. Arch Phys Med Rehabil 1961;42: [2] Brandell BR. Development of a universal control unit for functional electrical stimulation (FES). Am J Phys Med 1982;61(6): [3] Cozean CD, Pease WS, Hubbell SL. Biofeedback and functional electrical stimulation in stroke rehabilitation. Arch Phys Med Rehabil 1988;69: [4] Burridge J, Taylor P, Hagan S, Swain I. Experience of clinical use of the Odstock dropped foot stimulator. Artif Organs 1997;21(3): [5] Burridge JH, Taylor PN, Hagan SA, Wood DE, Swain ID. The effects of common peroneal stimulation on the effort and speed of walking. A randomised controlled trial with chronic hemiplegic patients. Clin Rehabil 1997;11(3): [6] Stefanovska A, Vodovnik L, Gros N, Rebersek S, Acimovic- Janezic R. FES and spasticity. IEEE Trans Biomed Eng 1989;36(7): [7] Glanz M, Klawansky S, Stason W, Berkey C, Chalmers TC. Functional electrostimulation in poststroke rehabilitation: a metaanalysis of the randomised controlled trials. Arch Phys Med Rehabil 1996;77(6):

7 A. Mansfield, G.M. Lyons / Medical Engineering & Physics 25 (2003) [8] Van Griethuysen CM, Paul JP, Andrews BJ, Nicol AC. Biomechanics of functional electrical stimulation. Prosthet Orthotics Int 1982;6(3): [9] Baker LL. Electrical stimulation to increase functional activity. In: Nelson RM, Hayes KW, Currier DP, editors. Clinical electrotherapy. 3rd ed. Stamford: Appleton and Lange; p [10] Aminian K, Rezakhanlou K, De Andres E, Fritsch C, Leyvraz PF, Robert P. Temporal feature estimation during walking using miniature accelerometers: an analysis of gait improvement after hip arthroplasty. Med Biol Eng Comput 1999;37(6): [11] Von Schroeder HP, Coutts RD, Lyden PD, Billings E, Nickel VL. Gait parameters following stroke: a practical assessment. J Rehabil Res Dev 1995;32(1): [12] Haugland MK, Sinkjaer T. Cutaneous whole nerve recording used for correction of footdrop in hemiplegic man IEEE. Trans Rehabil Eng 1995;3(4): [13] Bajd T, Kralj A, Turk R, Benko H, Šega J. The use of a fourchannel electrical stimulator as an ambulatory aid for paraplegic patients. Phys Ther 1983;63(7): [14] Lundin-Olsson L, Nyberg L, Gustafson Y. Stops walking when talking as a predictor of falls in elderly people. Lancet 1997;349:617. [15] Williamson R, Andrews BJ. Sensor systems for lower limb functional electrical stimulation (FES) control. Med Eng Phys 2000;22: [16] Lyons GM, Wilcox DJ, Grace PA. Real-time measurement of walking speed using an integrated accelerometer. In: Proceedings of the 6th Annual Conference of the International Functional Electrical Stimulation Society, Cleveland, USA p [17] Currie G, Rafferty D, Duncan G, Bell F, Evans AL. Measurement of gait by accelerometer and walkway: a comparison study. Med Biol Eng Comp 1992;30(6): [18] Williamson R, Andrews BJ. Gait event detection for FES using accelerometers and supervised machine learning. IEEE Trans Rehabil Eng 2000;8(3): [19] Willemsen AThM, Bloemhof F, Boom HBK. Automatic stance swing phase detection from accelerometer data for peroneal nerve stimulation. IEEE Trans Biomed Eng 1990;37(12):

Clinical Application of Acceleration Sensor to Detect the Swing Phase of Stroke Gait in Functional Electrical Stimulation

Clinical Application of Acceleration Sensor to Detect the Swing Phase of Stroke Gait in Functional Electrical Stimulation Tohoku J. Exp. Med., 2005, 207, 197-202 Acceleration Sensor in FES 197 Clinical Application of Acceleration Sensor to Detect the Swing Phase of Stroke Gait in Functional Electrical Stimulation YOICHI SHIMADA,

More information

An investigation of kinematic and kinetic variables for the description of prosthetic gait using the ENOCH system

An investigation of kinematic and kinetic variables for the description of prosthetic gait using the ENOCH system An investigation of kinematic and kinetic variables for the description of prosthetic gait using the ENOCH system K. OBERG and H. LANSHAMMAR* Amputee Training and Research Unit, University Hospital, Fack,

More information

APPLICATION OF THREE DIMENSIONAL ACCELEROMETRY TO HUMAN MOTION ANALYSIS

APPLICATION OF THREE DIMENSIONAL ACCELEROMETRY TO HUMAN MOTION ANALYSIS APPLICATION OF THREE DIMENSIONAL ACCELEROMETRY TO HUMAN MOTION ANALYSIS INTRODUCTION Ken'ichi Egawa, T. Tsuboi, T. Satoh, and M. Miyazaki Graduate School of Human Sciences, Waseda University Three dimensional

More information

Normal and Abnormal Gait

Normal and Abnormal Gait Normal and Abnormal Gait Adrielle Fry, MD EvergreenHealth, Division of Sport and Spine University of Washington Board Review Course March 6, 2017 What are we going to cover? Definitions and key concepts

More information

(received 13 December 2011, accepted 28 December 2011)

(received 13 December 2011, accepted 28 December 2011) Akita J Med 38 : 105-110, 2011 (15) Takenori Tomite 1), Yoichi Shimada 1), Toshiki Matsunaga 2), Kana Sasaki 1), Takayuki Yoshikawa 1) and Takehiro Iwami 3) (received 13 December 2011, accepted 28 December

More information

Mobility Lab provides sensitive, valid and reliable outcome measures.

Mobility Lab provides sensitive, valid and reliable outcome measures. Mobility Lab provides sensitive, valid and reliable outcome measures. ith hundreds of universities and hospitals using this system worldwide, Mobility Lab is the most trusted wearable gait and balance

More information

NIH Public Access Author Manuscript Gait Posture. Author manuscript; available in PMC 2009 May 1.

NIH Public Access Author Manuscript Gait Posture. Author manuscript; available in PMC 2009 May 1. NIH Public Access Author Manuscript Published in final edited form as: Gait Posture. 2008 May ; 27(4): 710 714. Two simple methods for determining gait events during treadmill and overground walking using

More information

Gait Analyser. Description of Walking Performance

Gait Analyser. Description of Walking Performance Gait Analyser Description of Walking Performance This brochure will help you to understand clearly the parameters described in the report of the Gait Analyser, provide you with tips to implement the walking

More information

C-Brace Reimbursement Guide

C-Brace Reimbursement Guide Reimbursement Guide Information for practitioners and payers Product Information Effective September 24, 2018 The The is the first microprocessor stance and swing phase controlled orthosis (SSCO ). This

More information

Analysis of stroke technique using acceleration sensor IC in freestyle swimming

Analysis of stroke technique using acceleration sensor IC in freestyle swimming Analysis of stroke technique using acceleration sensor IC in freestyle swimming Y. Ohgi, M. Yasumura Faculty of Environmental Information, Keio Univ., Japan H. Ichikawa Doctoral Prog. of Health and Sport

More information

Towards determining absolute velocity of freestyle swimming using 3-axis accelerometers

Towards determining absolute velocity of freestyle swimming using 3-axis accelerometers Towards determining absolute velocity of freestyle swimming using 3-axis accelerometers Author Stamm, Andy, Thiel, David, Burkett, Brendan, James, Daniel Published 2011 Journal Title Procedia Engineering

More information

As a physiotherapist I see many runners in my practice,

As a physiotherapist I see many runners in my practice, When rubber meets road Mark Richardson reveals the story that our running shoes can tell us, and how it can help you avoid running injury at a glance This article: Shows you how to analyse the sole of

More information

PURPOSE. METHODS Design

PURPOSE. METHODS Design 7 Murrary, M.P.; Sepic, S.B.; Gardner, G.M.; and Mollinger, L.A., "Gait patterns of above-knee amputees using constant-friction knee components," Bull Prosthet Res, 17(2):35-45, 1980. 8 Godfrey, C.M.;

More information

INTERACTION OF STEP LENGTH AND STEP RATE DURING SPRINT RUNNING

INTERACTION OF STEP LENGTH AND STEP RATE DURING SPRINT RUNNING INTERACTION OF STEP LENGTH AND STEP RATE DURING SPRINT RUNNING Joseph P. Hunter 1, Robert N. Marshall 1,, and Peter J. McNair 3 1 Department of Sport and Exercise Science, The University of Auckland, Auckland,

More information

C-Brace Reimbursement Guide

C-Brace Reimbursement Guide C-Brace Reimbursement Guide Information for practitioners and payers Product Information The C-Brace The C-Brace is the first microprocessor stance and swing phase controlled orthosis (SSCO). This highly

More information

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications Application Description AG/266PILD-EN Rev. C 2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic Pressure Measurement Engineered solutions for all applications Increase plant productivity

More information

Analysis of Foot Pressure Variation with Change in Stride Length

Analysis of Foot Pressure Variation with Change in Stride Length IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-853, p-issn: 2279-861.Volume 13, Issue 1 Ver. IV (Oct. 214), PP 46-51 Dr. Charudatta V. Shinde, M.S. MCh ( Orthopaedics ), Dr. Weijie

More information

Posture influences ground reaction force: implications for crouch gait

Posture influences ground reaction force: implications for crouch gait University of Tennessee, Knoxville From the SelectedWorks of Jeffrey A. Reinbolt July 14, 2010 Posture influences ground reaction force: implications for crouch gait H. X. Hoang Jeffrey A. Reinbolt, University

More information

Ranger Walking Initiation Stephanie Schneider 5/15/2012 Final Report for Cornell Ranger Research

Ranger Walking Initiation Stephanie Schneider 5/15/2012 Final Report for Cornell Ranger Research 1 Ranger Walking Initiation Stephanie Schneider sns74@cornell.edu 5/15/2012 Final Report for Cornell Ranger Research Abstract I joined the Biorobotics Lab this semester to gain experience with an application

More information

Fall Prevention Midterm Report. Akram Alsamarae Lindsay Petku 03/09/2014 Dr. Mansoor Nasir

Fall Prevention Midterm Report. Akram Alsamarae Lindsay Petku 03/09/2014 Dr. Mansoor Nasir Fall Prevention Midterm Report Akram Alsamarae Lindsay Petku 03/09/2014 Dr. Mansoor Nasir Updates to timeline Last semester we created a timeline that included important milestones. We have narrowed down

More information

Walking speemtmmkubjects and amputees: aspects of validity of gait analysis

Walking speemtmmkubjects and amputees: aspects of validity of gait analysis Prostheticsand Orthoti~Inte~national, 1993, 17, 78-82 Walking speemtmmkubjects and : aspects of validity of gait analysis A. M. BOONSTRA*, V. FIDLER** and W. H. EISMA* *Department of Rehabilitation Medicine,

More information

The use of electrical stimulation for correction of dropped foot in subjects with upper motor neurone lesions.

The use of electrical stimulation for correction of dropped foot in subjects with upper motor neurone lesions. The use of electrical stimulation for correction of dropped foot in subjects with upper motor neurone lesions. Paul Taylor Department of Medical Physics and Biomedical Engineering, Salisbury District Hospital,

More information

Ambulatory monitoring of gait quality with wearable inertial sensors

Ambulatory monitoring of gait quality with wearable inertial sensors Ambulatory monitoring of gait quality with wearable inertial sensors Dr. Philippe Terrier, PhD June 2016 Summary 1. Why? Reasons for measuring gait in real life conditions 2. What? Real-life assessment

More information

Figure 1 betois (bending torsion insole system) system with five measuring points and A/D- converter.

Figure 1 betois (bending torsion insole system) system with five measuring points and A/D- converter. Bending and Torsional Moments - A new measuring system for gait analysis Nora Dawin & Annette Kerkhoff & Klaus Peikenkamp, University of Applied Sciences Münster, Germany Abstract The Laboratory of Biomechanics

More information

CHAPTER IV FINITE ELEMENT ANALYSIS OF THE KNEE JOINT WITHOUT A MEDICAL IMPLANT

CHAPTER IV FINITE ELEMENT ANALYSIS OF THE KNEE JOINT WITHOUT A MEDICAL IMPLANT 39 CHAPTER IV FINITE ELEMENT ANALYSIS OF THE KNEE JOINT WITHOUT A MEDICAL IMPLANT 4.1 Modeling in Biomechanics The human body, apart of all its other functions is a mechanical mechanism and a structure,

More information

Assessments SIMPLY GAIT. Posture and Gait. Observing Posture and Gait. Postural Assessment. Postural Assessment 6/28/2016

Assessments SIMPLY GAIT. Posture and Gait. Observing Posture and Gait. Postural Assessment. Postural Assessment 6/28/2016 Assessments 2 SIMPLY GAIT Understanding movement Evaluations of factors that help therapist form professional judgments Include health, palpatory, range of motion, postural, and gait assessments Assessments

More information

THE INFLUENCE OF SLOW RECOVERY INSOLE ON PLANTAR PRESSURE AND CONTACT AREA DURING WALKING

THE INFLUENCE OF SLOW RECOVERY INSOLE ON PLANTAR PRESSURE AND CONTACT AREA DURING WALKING March 12, 2015 5:39:44pm WSPC/170-JMMB 1540005 ISSN: 0219-51942nd Reading Journal of Mechanics in Medicine and Biology Vol. 15, No. 2 (2015) 1540005 (6 pages) c World Scientific Publishing Company DOI:

More information

Analysis of Gait Characteristics Changes in Normal Walking and Fast Walking Of the Elderly People

Analysis of Gait Characteristics Changes in Normal Walking and Fast Walking Of the Elderly People IOSR Journal of Engineering (IOSRJEN) ISSN (e): 2250-3021, ISSN (p): 2278-8719 Vol. 08, Issue 7 (July. 2018), V (V) 34-41 www.iosrjen.org Analysis of Gait Characteristics Changes in and Of the Elderly

More information

Development of Wearable Sensor Combinations for Human Lower Extremity Motion Analysis

Development of Wearable Sensor Combinations for Human Lower Extremity Motion Analysis Proceedings of the 2006 IEEE International Conference on Robotics and Automation Orlando, Florida - May 2006 Development of Wearable Sensor Combinations for Human Lower Extremity Motion Analysis Tao Liu,

More information

Clinical view on ambulation in patients with Spinal Cord Injury

Clinical view on ambulation in patients with Spinal Cord Injury Clinical view on ambulation in patients with Spinal Cord Injury Sasa Moslavac Spinal Unit, Special Medical Rehabilitation Hospital, Varazdinske Toplice,, Croatia 1 Spinal Cord Injury (SCI) to walk again

More information

SIMULTANEOUS RECORDINGS OF VELOCITY AND VIDEO DURING SWIMMING

SIMULTANEOUS RECORDINGS OF VELOCITY AND VIDEO DURING SWIMMING Portuguese Journal of Sport Sciences. 6:supl. 2, 32-35, 2006 SIMULTANEOUS RECORDINGS OF VELOCITY AND VIDEO DURING SWIMMING Albert B. Craig 1, Budd Termin2, and David R. Pendergast 2 1University of Rochester,

More information

empower Reclaim your power. Information for technicians empower Ottobock 1

empower Reclaim your power. Information for technicians empower Ottobock 1 empower Reclaim your power. Information for technicians empower Ottobock 1 empower Powered propulsion for more freedom in life The empower Ankle is an innovation in the field of prosthetic feet. It is

More information

Neurorehabil Neural Repair Oct 23. [Epub ahead of print]

Neurorehabil Neural Repair Oct 23. [Epub ahead of print] APPENDICE Neurorehabil Neural Repair. 2009 Oct 23. [Epub ahead of print] Segmental Muscle Vibration Improves Walking in Chronic Stroke Patients With Foot Drop: A Randomized Controlled Trial. Paoloni M,

More information

Basic walker-assisted gait characteristics derived from forces and moments exerted on the walker s handles: Results on normal subjects

Basic walker-assisted gait characteristics derived from forces and moments exerted on the walker s handles: Results on normal subjects Medical Engineering & Physics 29 (2007) 380 389 Basic walker-assisted gait characteristics derived from forces and moments exerted on the walker s handles: Results on normal subjects Majd Alwan a,, Alexandre

More information

Control Strategies for operation of pitch regulated turbines above cut-out wind speeds

Control Strategies for operation of pitch regulated turbines above cut-out wind speeds Control Strategies for operation of pitch regulated turbines above cut-out wind speeds Helen Markou 1 Denmark and Torben J. Larsen, Risø-DTU, P.O.box 49, DK-4000 Roskilde, Abstract The importance of continuing

More information

The Effect of Military Load Carriage on Ground Reaction Forces. Technical Note. Stewart A Birrell 1 Robin H Hooper 1 Roger A Haslam 1

The Effect of Military Load Carriage on Ground Reaction Forces. Technical Note. Stewart A Birrell 1 Robin H Hooper 1 Roger A Haslam 1 The Effect of Military Load Carriage on Ground Reaction Forces Technical Note Stewart A Birrell 1 Robin H Hooper 1 Roger A Haslam 1 1 Department of Human Sciences, Loughborough University, Leicestershire,

More information

C-Brace Orthotronic Mobility System

C-Brace Orthotronic Mobility System C-Brace Orthotronic Mobility System You ll always remember your first step Information for practitioners C-Brace Orthotics reinvented Until now, you and your patients with conditions like incomplete spinal

More information

SPRINTING CHARACTERISTICS OF WOMEN S 100 METER FINALS AT THE IAAF WORLD CHAMPIONSHOPS DAEGU 2011

SPRINTING CHARACTERISTICS OF WOMEN S 100 METER FINALS AT THE IAAF WORLD CHAMPIONSHOPS DAEGU 2011 9:30-9:45 am Sang-Kyoon Park. Sprinting characteristics of women's 100 meter finals at the IAAF world championships Daegu 2011. (228) SPRINTING CHARACTERISTICS OF WOMEN S 100 METER FINALS AT THE IAAF WORLD

More information

Motion Control of a Bipedal Walking Robot

Motion Control of a Bipedal Walking Robot Motion Control of a Bipedal Walking Robot Lai Wei Ying, Tang Howe Hing, Mohamed bin Hussein Faculty of Mechanical Engineering Universiti Teknologi Malaysia, 81310 UTM Skudai, Johor, Malaysia. Wylai2@live.my

More information

Inertial compensation for belt acceleration in an instrumented treadmill

Inertial compensation for belt acceleration in an instrumented treadmill Inertial compensation for belt acceleration in an instrumented treadmill Sandra K. Hnat, Antonie J. van den Bogert Department of Mechanical Engineering, Cleveland State University Cleveland, OH 44115,

More information

Walking Simulator Mechanism

Walking Simulator Mechanism The Downtown Review Volume 2 Issue 2 Article 4 2015 Walking Simulator Mechanism Titus Lungu Cleveland State University Igor Tachynskyy Cleveland State University Omri Tayyara Cleveland State University

More information

INCLINOMETER DEVICE FOR SHIP STABILITY EVALUATION

INCLINOMETER DEVICE FOR SHIP STABILITY EVALUATION Proceedings of COBEM 2009 Copyright 2009 by ABCM 20th International Congress of Mechanical Engineering November 15-20, 2009, Gramado, RS, Brazil INCLINOMETER DEVICE FOR SHIP STABILITY EVALUATION Helena

More information

Empower. Reclaim your power. Information for technicians. Empower Ottobock 1

Empower. Reclaim your power. Information for technicians. Empower Ottobock 1 Empower Reclaim your power. Information for technicians Empower Ottobock 1 Empower Powered propulsion for more freedom in life The Empower is an innovation in the field of prosthetic feet. It is equipped

More information

Analysis of Pressure Rise During Internal Arc Faults in Switchgear

Analysis of Pressure Rise During Internal Arc Faults in Switchgear Analysis of Pressure Rise During Internal Arc Faults in Switchgear ASANUMA, Gaku ONCHI, Toshiyuki TOYAMA, Kentaro ABSTRACT Switchgear include devices that play an important role in operations such as electric

More information

The Effect of a Seven Week Exercise Program on Golf Swing Performance and Musculoskeletal Screening Scores

The Effect of a Seven Week Exercise Program on Golf Swing Performance and Musculoskeletal Screening Scores The Effect of a Seven Week Exercise Program on Golf Swing Performance and Musculoskeletal Screening Scores 2017 Mico Hannes Olivier Bachelor of Sport Science Faculty of Health Sciences and Medicine Bond

More information

Stride Time Calculation from EMG and Foot Switch Data and Finding Corelation between Them for Prosthetic Control

Stride Time Calculation from EMG and Foot Switch Data and Finding Corelation between Them for Prosthetic Control 2012 International Conference on Environment Science and Engieering IPCBEE vol.3 2(2012) (2012)IACSIT Press, Singapoore Stride Time Calculation from EMG and Foot Switch Data and Finding Corelation between

More information

Author s Name Name of the Paper Session. Positioning Committee. Marine Technology Society. DYNAMIC POSITIONING CONFERENCE September 18-19, 2001

Author s Name Name of the Paper Session. Positioning Committee. Marine Technology Society. DYNAMIC POSITIONING CONFERENCE September 18-19, 2001 Author s Name Name of the Paper Session PDynamic Positioning Committee Marine Technology Society DYNAMIC POSITIONING CONFERENCE September 18-19, 2001 POWER PLANT SESSION A New Concept for Fuel Tight DP

More information

GROUND REACTION FORCE DOMINANT VERSUS NON-DOMINANT SINGLE LEG STEP OFF

GROUND REACTION FORCE DOMINANT VERSUS NON-DOMINANT SINGLE LEG STEP OFF GROUND REACTION FORCE DOMINANT VERSUS NON-DOMINANT SINGLE LEG STEP OFF Sara Gharabaghli, Rebecca Krogstad, Sara Lynch, Sofia Saavedra, and Tamara Wright California State University, San Marcos, San Marcos,

More information

A4 Operation Manual. Fig.1-1 Controller Socket Diagram

A4 Operation Manual. Fig.1-1 Controller Socket Diagram A4 Operation Manual Safety Instruction Please read this manual carefully, also with related manual for the machinery before use the controller. For installing and operating the controller properly and

More information

1. Outline of the newly developed control technologies

1. Outline of the newly developed control technologies This paper describes a vertical lifting control and level luffing control design for newly developed, fully hydraulicdriven floating cranes. Unlike lattice boom crawler cranes for land use, the floating

More information

THE EFFECT OF BINDING POSITION ON KINETIC VARIABLES IN ALPINE SKIING

THE EFFECT OF BINDING POSITION ON KINETIC VARIABLES IN ALPINE SKIING THE EFFECT OF BINDING POSITION ON KINETIC VARIABLES IN ALPINE SKIING H. SCHWAMEDER 1, B. M. NIGG 2, V. v. TSCHARNER 2, D. STEFANYSHYN 2 1 Institute of Sports Sciences, University of Salzburg 2 Human Performance

More information

Analysis of Skip Motion as a Recovery Strategy after an Induced Trip

Analysis of Skip Motion as a Recovery Strategy after an Induced Trip 2015 IEEE International Conference on Systems, Man, and Cybernetics Analysis of Skip Motion as a Recovery Strategy after an Induced Trip Kento Mitsuoka, Yasuhiro Akiyama, Yoji Yamada, and Shogo Okamoto

More information

A4s Operation Manual

A4s Operation Manual A4s Operation Manual Safety Instruction Please read this manual carefully, also with related manual for the machinery before use the controller. For installing and operating the controller properly and

More information

Foot side detection from lower lumbar spine acceleration

Foot side detection from lower lumbar spine acceleration This is an Author's Original Manuscript of an article published by Elsevier in Gait and Posture Published online 6 June 2015 and available at: doi:10.1016/j.gaitpost.2015.05.021 Foot side detection from

More information

Technical Note. Determining the surface tension of liquids by measurements on pendant drops

Technical Note. Determining the surface tension of liquids by measurements on pendant drops Technical Note Pendant Drop Measurements Technical note: TN316e Industry section: all Author: FT, TW Date: 12/2010 Method: Drop Shape Analyzer DSA100 Keywords: Methods, surface tension, interfacial tension,

More information

Knee Kinematic Improvement after Total Knee Replacement Using a Simplified Quantitative Gait Analysis Method

Knee Kinematic Improvement after Total Knee Replacement Using a Simplified Quantitative Gait Analysis Method Iranian Rehabilitation Journal, Vol. 11, No. 18, October 2013 Original Article Knee Kinematic Improvement after Total Knee Replacement Using a Simplified Quantitative Gait Analysis Method Hassan Sarailoo

More information

3 people 3 unique lifestyles 3 advanced foot solutions

3 people 3 unique lifestyles 3 advanced foot solutions 3 people 3 unique lifestyles 3 advanced foot solutions Reclaim your life Information for practitioners Shelby Hans Intelligent feet are made for more than just walking Today s advanced microprocessor controlled

More information

Spasticity in gait. Wessex ACPIN Spasticity Presentation Alison Clarke

Spasticity in gait. Wessex ACPIN Spasticity Presentation Alison Clarke Spasticity in gait Clinicians recognise spasticity but the elements of spasticity contributing to gait patterns are often difficult to identify: Variability of muscle tone Observation/recording General

More information

PREDICTING THE ABILITY OF SURVIVAL AFTER DAMAGE IN TANKERS. José Poblet Martínez - SENER, (Spain) José Javier Díaz Yraola - SENER, (Spain)

PREDICTING THE ABILITY OF SURVIVAL AFTER DAMAGE IN TANKERS. José Poblet Martínez - SENER, (Spain) José Javier Díaz Yraola - SENER, (Spain) 8 th International Conference on 477 PREDICTING THE ABILITY OF SURVIVAL AFTER DAMAGE IN TANKERS Abstract José Poblet Martínez - SENER, (Spain) José Javier Díaz Yraola - SENER, (Spain) To meet the demand

More information

Available online at ScienceDirect. Procedia Engineering 112 (2015 )

Available online at  ScienceDirect. Procedia Engineering 112 (2015 ) Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 112 (2015 ) 540 545 7th Asia-Pacific Congress on Sports Technology, APCST 2015 Movement variability of professional pool billiards

More information

video Purpose Pathological Gait Objectives: Primary, Secondary and Compensatory Gait Deviations in CP AACPDM IC #3 1

video Purpose Pathological Gait Objectives: Primary, Secondary and Compensatory Gait Deviations in CP AACPDM IC #3 1 s in CP Disclosure Information AACPDM 71st Annual Meeting September 13-16, 2017 Speaker Names: Sylvia Ounpuu, MSc and Kristan Pierz, MD Differentiating Between, Secondary and Compensatory Mechanisms in

More information

REPORT. A comparative study of the mechanical and biomechanical behaviour of natural turf and hybrid turf for the practise of sports

REPORT. A comparative study of the mechanical and biomechanical behaviour of natural turf and hybrid turf for the practise of sports REPORT A comparative study of the mechanical and biomechanical behaviour of natural turf and hybrid turf for the practise of sports Addressed to: PSF - PALAU TURF Date: May 2015 Table of Contents SHEET

More information

Artifacts Due to Filtering Mismatch in Drop Landing Moment Data

Artifacts Due to Filtering Mismatch in Drop Landing Moment Data Camenga et al. UW-L Journal of Undergraduate Research XVI (213) Artifacts Due to Filtering Mismatch in Drop Landing Moment Data Elizabeth T. Camenga, Casey J. Rutten, Brendan D. Gould, Jillian T. Asmus,

More information

Dynamic analysis and motion measurement of ski turns using inertial and force sensors

Dynamic analysis and motion measurement of ski turns using inertial and force sensors Available online at www.sciencedirect.com Procedia Engineering 6 ( 213 ) 355 36 6th Asia-Pacific Conference on Sports Technology Dynamic analysis and motion measurement of ski turns using inertial and

More information

Gait pattern and spinal movement in walking - A therapeutic approach in juvenile scoliosis

Gait pattern and spinal movement in walking - A therapeutic approach in juvenile scoliosis Gait pattern and spinal movement in walking - A therapeutic approach in juvenile scoliosis H. Fleiss**, O. Fleiss**, P. Holzer**, G. Ritter***,V. P. Stokes* University of Graz**, Medical University of

More information

AN EXPERIMENTAL INVESTIGATION ON GOLF SHOE DESIGN USING FOOT- PRESSURE DISTRIBUTION DURING THE GOLF SWING

AN EXPERIMENTAL INVESTIGATION ON GOLF SHOE DESIGN USING FOOT- PRESSURE DISTRIBUTION DURING THE GOLF SWING Proceedings of the 4 th BSME-ASME International Conference on Thermal Engineering 27-29 December, 2008, Dhaka, Bangladesh AN EXPERIMENTAL INVESTIGATION ON GOLF SHOE DESIGN USING FOOT- PRESSURE DISTRIBUTION

More information

Normative data of postural sway by using sway meter among young healthy adults

Normative data of postural sway by using sway meter among young healthy adults Normative data of postural sway by using sway meter Original Research Article ISSN: 2394-0026 (P) Normative data of postural sway by using sway meter among young healthy adults Tejal C Nalawade 1*, Shyam

More information

Simulation-based design to reduce metabolic cost

Simulation-based design to reduce metabolic cost Simulation-based design to reduce metabolic cost Overview: Lecture + Hands On Exercise 1. Generating and evaluating a muscledriven simulation of walking 2. Metabolics 101 3. Designing and evaluating devices

More information

Available online at ScienceDirect. Procedia Manufacturing 2 (2015 )

Available online at   ScienceDirect. Procedia Manufacturing 2 (2015 ) Available online at www.sciencedirect.com ScienceDirect Procedia Manufacturing 2 (2015 ) 490 494 2nd International Materials, Industrial, and Manufacturing Engineering Conference, MIMEC2015, 4-6 February

More information

What is the optimal design of a rocker shoe

What is the optimal design of a rocker shoe What is the optimal design of a rocker shoe University of Salford School of Health, Sport and Rehabilitation Sciences Deutsche Sporthochschule Köln J Chapman, S Preece, C Nester, B Braunstein, P Bruggerman

More information

Equine Cannon Angle System

Equine Cannon Angle System Equine Cannon System How to interpret the results December 2010 Page 1 of 14 Table of Contents Introduction... 3 The Sagittal Plane... 4 The Coronal Plane... 5 Results Format... 6 How to Interpret the

More information

The Mechanics of Modern BREASTSTROKE Swimming Dr Ralph Richards

The Mechanics of Modern BREASTSTROKE Swimming Dr Ralph Richards The Mechanics of Modern BREASTSTROKE Swimming Dr Ralph Richards Breaststroke is the least efficient of the four competition strokes because a large amount of water resistance is created due to body position

More information

Megan E. Krause, BSBSE, Young Hui Chang, Ph.D. Comparative Neuromechanics Laboratory. Georgia Institute of Technology

Megan E. Krause, BSBSE, Young Hui Chang, Ph.D. Comparative Neuromechanics Laboratory. Georgia Institute of Technology Megan E. Krause, BSBSE, Young Hui Chang, Ph.D. Comparative Neuromechanics Laboratory Sh School of Applied Physiology Georgia Institute of Technology 1 Variety of ankle constraints during locomotion: Fashion

More information

Gait. Kinesiology RHS 341 Lecture 12 Dr. Einas Al-Eisa

Gait. Kinesiology RHS 341 Lecture 12 Dr. Einas Al-Eisa Gait Kinesiology RHS 341 Lecture 12 Dr. Einas Al-Eisa Definitions Locomotion = the act of moving from one place to the other Gait = the manner of walking Definitions Walking = a smooth, highly coordinated,

More information

Ultrasonic motion analysis system - measurement of temporal and spatial gait parameters Huitema, RB; Hof, At L.; Postema, K

Ultrasonic motion analysis system - measurement of temporal and spatial gait parameters Huitema, RB; Hof, At L.; Postema, K University of Groningen Ultrasonic motion analysis system - measurement of temporal and spatial gait parameters Huitema, RB; Hof, At L.; Postema, K Published in: Journal of biomechanics IMPORTANT NOTE:

More information

Dynamic Positioning Control Augmentation for Jack-up Vessels

Dynamic Positioning Control Augmentation for Jack-up Vessels DYNAMIC POSITIONING CONFERENCE October 9-10, 2012 Design and Control Session Dynamic Positioning Control Augmentation for Jack-up Vessels By Bradley Deghuee L-3 Communications 1 Introduction Specialized

More information

Sensitivity of toe clearance to leg joint angles during extensive practice of obstacle crossing: Effects of vision and task goal

Sensitivity of toe clearance to leg joint angles during extensive practice of obstacle crossing: Effects of vision and task goal Sensitivity of toe clearance to leg joint angles during extensive practice of obstacle crossing: Effects of vision and task goal Sérgio Tosi Rodrigues 1, Valéria Duarte Garcia 1,2, Arturo Forner- Cordero

More information

Shock absorbing material on the shoes of long leg braces for paraplegic walking

Shock absorbing material on the shoes of long leg braces for paraplegic walking Prosthetics and Orthotics International, 1990, 14, 27-32 Shock absorbing material on the shoes of long leg braces for paraplegic walking F. BIERLING-SØRENSEN, H. RYDE, *F. BOJSEN-MØLLER AND **E. LYQUIST

More information

REPLACING REDUNDANT STABILOMETRY PARAMETERS WITH RATIO AND MAXIMUM DEVIATION PARAMETERS

REPLACING REDUNDANT STABILOMETRY PARAMETERS WITH RATIO AND MAXIMUM DEVIATION PARAMETERS Proceedings of the 12th IASTED International Conference Biomedical Engineering (BioMed 2016) February 15-16, 2016 Innsbruck, Austria REPLACING REDUNDANT STABILOMETRY PARAMETERS WITH RATIO AND MAXIMUM DEVIATION

More information

The Incremental Evolution of Gaits for Hexapod Robots

The Incremental Evolution of Gaits for Hexapod Robots The Incremental Evolution of Gaits for Hexapod Robots Abstract Gait control programs for hexapod robots are learned by incremental evolution. The first increment is used to learn the activations required

More information

WalkOn product range. Dynamic Ankle-Foot Orthoses. Information for specialist dealers

WalkOn product range. Dynamic Ankle-Foot Orthoses. Information for specialist dealers WalkOn product range Dynamic Ankle-Foot Orthoses Information for specialist dealers WalkOn Flex WalkOn WalkOn Trimable WalkOn Reaction WalkOn Reaction plus One range Many different applications The WalkOn

More information

Line Following with RobotC Page 1

Line Following with RobotC Page 1 Line Following with RobotC Page 1 Line Following with By Michael David Lawton Introduction Line following is perhaps the best way available to VEX Robotics teams to quickly and reliably get to a certain

More information

SCHEINWORKS Measuring and Analysis Systems by

SCHEINWORKS Measuring and Analysis Systems by Pressure Measurement Systems for standing and walking analysis Germany since 1879 Pressure Measurement Systems for standing and walking analysis Documentation of Gait image Stance Symmetry of all parameters

More information

ABSTRACT AUTHOR. Kinematic Analysis of the Women's 400m Hurdles. by Kenny Guex. he women's 400m hurdles is a relatively

ABSTRACT AUTHOR. Kinematic Analysis of the Women's 400m Hurdles. by Kenny Guex. he women's 400m hurdles is a relatively Study Kinematic Analysis of the Women's 400m Hurdles by IAAF 27:1/2; 41-51, 2012 by Kenny Guex ABSTRACT The women's 400m hurdles is a relatively new discipline and a complex event that cannot be approached

More information

BODY FORM INFLUENCES ON THE DRAG EXPERIENCED BY JUNIOR SWIMMERS. Australia, Perth, Australia

BODY FORM INFLUENCES ON THE DRAG EXPERIENCED BY JUNIOR SWIMMERS. Australia, Perth, Australia 1 BODY FORM INFLUENCES ON THE DRAG EXPERIENCED BY JUNIOR SWIMMERS Andrew Lyttle 1, Nat Benjanuvatra 2, Brian A Blanksby 2, Bruce C Elliott 2 1 Western Australian Institute of Sport, Perth, Australia 2

More information

siot-shoe: A Smart IoT-shoe for Gait Assistance (Miami University)

siot-shoe: A Smart IoT-shoe for Gait Assistance (Miami University) siot-shoe: A Smart IoT-shoe for Gait Assistance (Miami University) Abstract Mark Sullivan, Casey Knox, Juan Ding Gait analysis through the Internet of Things (IoT) is able to provide an overall assessment

More information

Ball impact dynamics of knuckling shot in soccer

Ball impact dynamics of knuckling shot in soccer Available online at www.sciencedirect.com Procedia Engineering 34 (2012 ) 200 205 9 th Conference of the International Sports Engineering Association (ISEA) Ball impact dynamics of knuckling shot in soccer

More information

INTERACTION BETWEEN WIND-DRIVEN AND BUOYANCY-DRIVEN NATURAL VENTILATION Bo Wang, Foster and Partners, London, UK

INTERACTION BETWEEN WIND-DRIVEN AND BUOYANCY-DRIVEN NATURAL VENTILATION Bo Wang, Foster and Partners, London, UK INTERACTION BETWEEN WIND-DRIVEN AND BUOYANCY-DRIVEN NATURAL VENTILATION Bo Wang, Foster and Partners, London, UK ABSTRACT Ventilation stacks are becoming increasingly common in the design of naturally

More information

Activity Overview. Footprints In The Sand Inquiry MO-BILITY. Activity 2E. Activity Objectives: Activity Description: Activity Background: LESSON 2

Activity Overview. Footprints In The Sand Inquiry MO-BILITY. Activity 2E. Activity Objectives: Activity Description: Activity Background: LESSON 2 Footprints In The Sand Inquiry Activity 2E Activity Objectives: Working in collaborative groups, students will be able to: assume assigned roles to complete a given task examine the various parts of the

More information

SHUFFLE TURN OF HUMANOID ROBOT SIMULATION BASED ON EMG MEASUREMENT

SHUFFLE TURN OF HUMANOID ROBOT SIMULATION BASED ON EMG MEASUREMENT SHUFFLE TURN OF HUMANOID ROBOT SIMULATION BASED ON EMG MEASUREMENT MASANAO KOEDA, TAKAYUKI SERIZAWA, AND YUTA MATSUI Osaka Electro-Communication University, Faculty of Information Science and Arts, Department

More information

The technique of reciprocal walking using the hip guidance orthosis (hgo) with crutches

The technique of reciprocal walking using the hip guidance orthosis (hgo) with crutches The technique of reciprocal walking using the hip guidance orthosis (hgo) with crutches P. B. BUTLER, R. E. MAJOR and J. H. PATRICK Orthotic Research and Locomotor Assessment Unit, The Robert Jones and

More information

Design of a Microcontroller-Based Pitch Angle Controller for a Wind Powered Generator

Design of a Microcontroller-Based Pitch Angle Controller for a Wind Powered Generator Journal of Engineering and Science Research 1 (2): 133-138, e-issn RMP Publications, DOI: Design of a Microcontroller-Based Pitch Angle Controller for a Wind Powered Generator Glenn V. Magwili, Michael

More information

Dynamically stepping over large obstacle utilizing PSO optimization in the B4LC system

Dynamically stepping over large obstacle utilizing PSO optimization in the B4LC system 1 Dynamically stepping over large obstacle utilizing PSO optimization in the B4LC system QI LIU, JIE ZHAO, KARSTEN BERNS Robotics Research Lab, University of Kaiserslautern, Kaiserslautern, 67655, Germany

More information

EXPERIMENTAL STUDY OF EXOSKELETON FOR ANKLE AND KNEE JOINT

EXPERIMENTAL STUDY OF EXOSKELETON FOR ANKLE AND KNEE JOINT EXPERIMENTAL STUDY OF EXOSKELETON FOR ANKLE AND KNEE JOINT PROJECT REFERENCE NO. : 37S0925 COLLEGE : NEW HORIZON COLLEGE OF ENGINEERING, BANGALORE BRANCH : MECHANICAL ENGINEERING GUIDES : DR GANESHA PRASAD

More information

HRC adjustable pneumatic swing-phase control knee

HRC adjustable pneumatic swing-phase control knee HRC adjustable pneumatic swing-phase control knee S. NAKAMURA and S. SAWAMURA Hyogo Rehabilitation Centre, Kobe, Japan Abstract Since 1972 the Hyogo Rehabilitation Centre has been developing a variable-resistance-type

More information

Body Stabilization of PDW toward Humanoid Walking

Body Stabilization of PDW toward Humanoid Walking Body Stabilization of PDW toward Humanoid Walking Masaki Haruna, Masaki Ogino, Koh Hosoda, Minoru Asada Dept. of Adaptive Machine Systems, Osaka University, Suita, Osaka, 565-0871, Japan ABSTRACT Passive

More information

Fail Operational Controls for an Independent Metering Valve

Fail Operational Controls for an Independent Metering Valve Group 14 - System Intergration and Safety Paper 14-3 465 Fail Operational Controls for an Independent Metering Valve Michael Rannow Eaton Corporation, 7945 Wallace Rd., Eden Prairie, MN, 55347, email:

More information

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW AC 2010-2145: MEASUREMENT OF HYDROGEN IN HELIUM FLOW Randy Buchanan, University of Southern Mississippi Christopher Winstead, University of Southern Mississippi Anton Netchaev, University of Southern Mississippi

More information

The effects of a suspended-load backpack on gait

The effects of a suspended-load backpack on gait Industrial and Manufacturing Systems Engineering Publications Industrial and Manufacturing Systems Engineering 2009 The effects of a suspended-load backpack on gait Xu Xu North Carolina State University

More information

Characteristics of ball impact on curve shot in soccer

Characteristics of ball impact on curve shot in soccer Available online at www.sciencedirect.com Procedia Engineering 60 (2013 ) 249 254 6 th Asia-Pacific Congress on Sports Technology (APCST) Characteristics of ball impact on curve shot in soccer Sungchan

More information