Commissioning of Elekta 6MV FFF Versa HD and Pinnacle
|
|
- Florence Gray
- 5 years ago
- Views:
Transcription
1 Commissioning of Elekta 6MV FFF Versa HD and Pinnacle Poster No.: R-0044 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit L. Bendall, I. Patel, N. McGrath, C. Rowbottom; MANCHESTER/ UK Radiotherapy techniques, VMAT and Tomotherapy, Physics, Radiation therapy / Oncology, CT, Radiation physics, Oncology /ranzcr2014/R-0044 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply RANZCR/AIR/ACPSEM's endorsement, sponsorship or recommendation of the third party, information, product or service. RANZCR/AIR/ ACPSEM is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold RANZCR/AIR/ACPSEM harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies,.ppt slideshows,.doc documents and any other multimedia files are not available in the pdf version of presentations. Page 1 of 8
2 Aim To commission the first 6MV FFF beam within Pinnacle 9.6 at The Christie, Manchester, UK. Methods and materials Linac matched to manufacturer "golden dataset" A 3D water tank with dimensions 50cmx50cmx50cm was used with Farmer chambers with vented sensitive volumes of 0.125cm 3 (for fields #10x10cm 2 ) and 0.015cm 3 (for fields #10x10cm 2 ) Data was collected across range of field sizes 2x2-40x40cm 2, including Open field profiles (Figure 1) Percentage depth doses (Figure 2) Output factors Tissue Standard Ratios (Tissue Phantom Ratio normalised to standard set up; 10x10cm field, 10cm deep, 90cm FSD) cgy-mu factor For Pinnacle model, we started with a conventional flattened beam model previously produced at The Christie for the same linac and adjusted (in order): Arbitrary profile Energy spectrum (Figure 3) Small change to build up (electron contamination) Off-axis softening factor set to zero MLC and jaw transmission reduced FFF flag updated Maximum dose rate increased to 1500MU/min Model validation measurements Chamber measurements, EDR2 film and Delta 4 used for model validation All Pinnacle calculations made with Collapsed Cone Convolution algorithm, with a dose grid resolution of 1mm. Page 2 of 8
3 Tissue Standard Ratios; measured compared to calculated in Pinnacle. Test fields including those listed in Table 1. Point doses for test fields were measured in a water phantom and compared to Pinnacle calculated doses. Test fields were delivered to film and compared to Pinnacle fluences Clinical plan verification 4 field brick dose measurements, off-axis, various depths etc IMRT plans - fluence and dose point comparisons Images for this section: Fig. 1: 40x40cm^2 profiles at varying depths Page 3 of 8
4 Fig. 2: Comparison of measured Percentage Depth Doses for 6MV and 6MVFFF Page 4 of 8
5 Fig. 3: Comparison of 6MV and 6MVFFF modelled energy spectrum Fig. 4: Test fields used and the parameters they verify Page 5 of 8
6 Results A model was produced in Pinnacle for use with the 6MVFFF beam. Generally good agreement was seen between modelled and measured data, an example can be seen in Figure 5 and Table 2. The maximum difference between calculated (Pinnacle) and measured TSRs over a range of areas (16-900cm 2 ) and depths (1.6-20cm) was found to be <1%. All model validation measurements were within acceptable values. The department is currently working on implementing Lung SABR VMAT treatments with 6MVFFF. Images for this section: Fig. 5: Bar test results for 6MVFFF at 10cm deep Page 6 of 8
7 Fig. 6: Bar test comparison data measured vs. calculated in Pinnacle Page 7 of 8
8 Conclusion The 6MVFFF model was successfully produced and commissioned. The introduction of FFF treatment delivery into clinical use at The Christie will include a change in technique from conventional 11-field Lung SABR planning with 6MV to Lung SABR VMAT planning with 6MVFFF. The work required to introduce the new technique is currently in progress. Further work may involve the commissioning of paediatric treatment sites and 10MVFFF. Personal information L. Bendall worked at The Christie NHS Foundation Trust, Manchester, UK, where this work was undertaken, until the end of June She now works with the Institute of Medical Physics, University of Sydney, the Crown Princess Mary Cancer Care Centre, Westmead Hospital, and Nepean Cancer Centre, Nepean Hospital. References. Page 8 of 8
Quality control of breast ultrasound: How to use a breast ultrasound quality assurance phantom?
Quality control of breast ultrasound: How to use a breast ultrasound quality assurance phantom? Poster No.: C-2282 Congress: ECR 2013 Type: Educational Exhibit Authors: S. J. Oh, B. K. Seo, S. E. Song,
More informationPitfalls and artifacts in FDG oncologic PET/CT: A pictorial essay to the thoracic radiologist
Pitfalls and artifacts in FDG oncologic PET/CT: A pictorial essay to the thoracic radiologist Poster No.: P-0120 Congress: ESTI 2015 Type: Scientific Poster Authors: G. Meirelles, J. Capobianco, M. Oliveira;
More informationSpectrum of imaging findings of local adverse effects of anabolic steroid use
Spectrum of imaging findings of local adverse effects of anabolic steroid use Poster No.: C-2350 Congress: ECR 2010 Type: Educational Exhibit Topic: Musculoskeletal Authors: H. Gupta, V. Bhat, S. Edward,
More informationRevalidating the importance of anatomical side markers and annotations on radiographic images
Revalidating the importance of anatomical side markers and annotations on radiographic images Poster No.: C-2464 Congress: ECR 2015 Type: Authors: Keywords: DOI: Educational Exhibit H. Khosa, C. Thomas,
More informationShock Wave Lithotripsy (ESWL) results aren't improving. What can Radiographers do to improve outcomes with better kidney stone fragmentation
Shock Wave Lithotripsy (ESWL) results aren't improving. What can Radiographers do to improve outcomes with better kidney stone fragmentation Poster No.: C-1094 Congress: ECR 2015 Type: Scientific Exhibit
More informationRevaluation of sonographic criteria of acute appendicitis in children
Revaluation of sonographic criteria of acute appendicitis in children Poster No.: C-1615 Congress: ECR 2011 Type: Scientific Exhibit Authors: A. Alvarez Vazquez, N. Fernandez Garcia, G. Anes, C. J. 1 2
More informationMonitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription
Monitor Unit Calculations Part 1 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Return to our first patient 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt
More informationCommissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN
22 Turkish Journal of Cancer Volume 37, No.1, 2007 Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN SATISH PELAGADE 1, KALPANA
More informationCommissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale
Commissioning an IMRT System for MLC Delivery Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale Taking the broad view of commissioning Commissioning elements Validating the dosimetry system Commissioning the
More informationSnapShot IMRT with compensators and FFF beams
SnapShot IMRT with compensators and FFF beams Vladimir Feygelman, PhD (1) Moffitt Cancer Center, Tampa, FL, USA Disclosure VF has a sponsored research agreement with.decimal. A short history of radiotherapy
More informationCHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS
47 CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS 4.1 INTRODUCTION Advanced treatment techniques use optimized radiation beam intensities to conform dose distribution to the
More informationIMRT in clinical practice at the UMC-Utrecht. Utrecht
3 IMRT in clinical practice at the UMC-Utrecht Utrecht Clinical use of IMRT Improvement of local control without increased incidence of normal tissue complications + Dose escalation in the prostate Step-and
More informationTG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009
TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009 DISCLAIMER: This publication and associated spreadsheets and digital files are based on sources and
More informationReview of fundamental photon dosimetry quantities
Review of fundamental photon dosimetry quantities Narayan Sahoo Main sources of the materials included in this lecture notes are: (1) Radiation Oncology Physics: A Handbook for Teachers and Students Edited
More informationImplementation of respiratory-gated VMAT on a Versa HD linear accelerator
Received: 21 November 2016 Revised: 7 April 2017 Accepted: 5 June 2017 DOI: 10.1002/acm2.12160 RADIATION ONCOLOGY PHYSICS Implementation of respiratory-gated VMAT on a Versa HD linear accelerator Jeffrey
More informationMonitor Unit Verification for Small Fields
Monitor Unit Verification for Small Fields Patrick Higgins, Ph.D University of Minnesota Department of Radiation Oncology October 10, 2013 The Issues: How do we verify the monitor units calculated by the
More informationConstancy checks of well-type ionization chambers with external-beam radiation units
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Constancy checks of well-type ionization chambers with external-beam radiation units Sara L. Hackett, 1a Benjamin Davis, 2 Andrew
More informationThe Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.
FRANZCR Examination Part I Radiation Oncology Radiotherapeutic Physics Candidate No.: The Royal Australian and New Zealand College of Radiologists FRANZCR Examination Part I Radiation Oncology Radiotherapeutic
More informationRelative Dosimetry. Photons
Relative Dosimetry Photons What you need to measure! Required Data (Photon) Central Axis Percent Depth Dose Tissue Maximum Ratio Scatter Maximum Ratio Output Factors S c & S cp! S p Beam profiles Wedge
More informationValidation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2017, 6, 57-72 http://www.scirp.org/journal/ijmpcero ISSN Online: 2168-5444 ISSN Print: 2168-5436 Validation of Treatment
More informationThe MC commissioning of CyberKnife with MLC (Tips and Tricks)
The MC commissioning of CyberKnife with MLC (Tips and Tricks) Alain Guemnie Tafo, PhD UPMC Pinnacle, Harrisburg, PA Institute Gustave Roussy, Villejuif, France Disclaimer & Disclosure The views expressed
More informationTHE development of more advanced techniques in radiotherapy,
1 A Geant4-based simulation of an accelerator s head used for Intensity Modulated Radiation Therapy F. Foppiano, B. Mascialino, M.G. Pia, M. Piergentili. Abstract We present a Geant4-based simulation,
More informationTG219: IT'S USE, STRENGTHS AND WEAKNESSES
1 CONFLICT OF INTEREST TG219: IT'S USE, STRENGTHS AND WEAKNESSES SOTIRI STATHAKIS, PHD, DABR None ROSTER STATEMENT OF THE PROBLEM Timothy C. Zhu, University of Pennsylvania, Philadelphia, PA (Co- Chair)
More information13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.
13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS. 13.1.1 TRAINING GOAL 13.1.1.1 Among the responsabilities of a medical
More informationANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL
ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL FOR THE DEGREE MASTER OF ARTS BY CHARLES TRAVIS WEBB
More informationCLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions
CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions Rebecca M. Howell, Ph.D., D.A.B.R, The University of Texas at M.D. Anderson Cancer Center Overview 2 Commissioning
More informationAbsorption measurements for a carbon fiber couch top and its modelling in a treatment planning system
Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system G. Kunz, F. Hasenbalg, P. Pemler 1 1 Klinik für Radio-Onkologie und Nuklearmedizin, Stadtspital Triemli
More informationClinical Implementation of Volumetric Modulated Arc Therapy
Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter Definition 2 Intensity Modulated Arc therapy
More informationCOMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy
COMPASS 3D IMRT Verification in Patient Anatomy AK IMRT Bamberg April 2010 Dr. Lutz Müller Patient-specific Verification? CT Treatment unit Gantry =0 Treatmt unit Patient couch X-ray beam Patient couch
More informationBeam modeling and VMAT performance with the Agility 160-leaf multileaf collimator
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator James L. Bedford, a Michael D. R. Thomas and Gregory
More informationZoubir Ouhib Lynn Cancer Institute
Zoubir Ouhib Lynn Cancer Institute April 2015 Speaker for ELEKTA Brachytherapy using miniature X-ray sources Energy
More informationOutline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface
Chapter 11 reatment Planning Single Beams Radiation Dosimetry I Outline Basic terminology Curved contour surface correction (bolus, compensators, wedges) Oblique beam incidence Correction for tissue inhomogeneities
More informationMonitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting
Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose
More informationGEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY
The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF
More informationA dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions
A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions Wendel Dean Renner a) Math Resolutions, LLC, 5975 Gales Lane, Columbia,
More informationTreatment plan complexity metrics for predicting IMRT pretreatment. assurance results. Scott Crowe, QUT
Treatment plan complexity metrics for predicting IMRT pretreatment quality assurance results Scott Crowe, QUT Pre-treatment QA Verification of fluence delivery to ensure that dose can be delivered accurately
More informationClinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry
Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Tom Jordan Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002 Comparison of monitor unit calculations performed with a 3D computerized planning system and independent hand calculations: Results
More informationDosimetric Calculations. Lonny Trestrail
Dosimetric Calculations Lonny Trestrail 20 October 2008 Objectives Dose Distribution Measurements PDD, OCR TAR, SAR, TPR, TMR, SPR, SMR Arc or Rotational Therapy Isodose Curves Point Dose Calculations
More informationPractical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11
MV Photon Dosimetry in the Clinic This document discusses both the scientific methods and the system of work required to calibrate a field instrument expanding upon the practical details of both.. Introduction
More informationMath Review. Overview
Drawn by Steve Yan, CMD Math Review Shirley Ann Pinegar-Johnston MS RT(R)(T)CMD Overview Drawn by Steve Yan, CMD Divergence & Intensity DIVERGENCE X-rays travel in Straight but Divergent lines INTENSITY
More informationCommissioning of a new total body irradiation protocol
University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2012 Commissioning of a new total body irradiation protocol Zoe Baldwin University
More informationA beam-matching concept for medical linear accelerators
Acta Oncologica, 2009; 48: 192200 ORIGINAL ARTICLE A beam-matching concept for medical linear accelerators DAVID SJÖSTRÖM, ULF BJELKENGREN, WIVIANN OTTOSSON & CLAUS F. BEHRENS Copenhagen University Hospital,
More informationHigh Speed Direct SAD Radiosurgery Beam Scanner
Open Access Original Article DOI: 10.7759/cureus.20 High Speed Direct SAD Radiosurgery Beam Scanner Walter Nikesch 1, James M. Hevezi 2, Irene Monterroso 3, Daniel Navarro 4, James G. Schwade 5 1. CyberKnife
More informationThe Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007
The Effect of Immobilisation Devices on Radiotherapy Dose Distributions A thesis submitted in fulfilment of the requirements for the Degree of Master of Applied Science (Medical and Health Physics) By
More informationField size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
Journal of BUON 8: 55-59, 2003 2003 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator
More informationVMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 3, 2016 VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests Michael P. Barnes, 1,2a Pejman Rowshanfarzad,
More informationRESPIRATORY MOTION IN RADIATION THERAPY
RESPIRATORY MOTION IN RADIATION THERAPY Significant effect on targets in the chest and upper abdominal cavities Can cause difficulty in reproducing radiation planning protocols in radiation treatment sessions
More informationScientific Journal of Nuclear Medicine & Radiation Therapy
Scientific Journal of Nuclear Medicine & Radiation Therapy Research Article Radiotherapy Dose Calculation Program, for Monitoring and Commissioning Treatment Planning Systems - Ioannis Vamvakas*, Stella
More informationComparison of ionization chambers of various volumes for IMRT absolute dose verification
Comparison of ionization chambers of various volumes for IMRT absolute dose verification Leonid B. Leybovich, a) Anil Sethi, and Nesrin Dogan Department of Radiation Oncology, Loyola University Medical
More informationElectronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI
Electronic Oncology Systems the Management and the Safety Measures Hansen Chen, Director, TDSI Radiation Oncology Department People Place Equipments Procedures Patients Hansen Chen 2 Radiation Oncology
More informationInvestigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy
University of South Florida Scholar Commons Graduate Theses and Dissertations Graduate School 7-14-2010 Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy
More informationTG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:
Practical Implementation of TG-61: II. Guidelines for clinical implementation of TG-61 J.P. Seuntjens Medical Physics Unit McGill University, Montreal General Hospital Montréal, Canada jseuntjens@medphys.mcgill.ca
More informationExperimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media
Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media N Tyagi 1, B H Curran, P L Roberson, J M Moran, E Acosta and B A Fraass Department
More informationA simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation
Radiotherapy and Oncology 71 (2004) 235 241 Technical Note A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation Nadine Linthout*,
More informationA clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment techniques
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2009 A clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment
More informationDynamic wedges dosimetry and quality control
Original Paper Received: 2005.06.07 Accepted: 2006.02.28 Published: 2006.04.28 Authors Contribution: A Study Design B Data Collection C Statistical Analysis D Data Interpretation E Manuscript Preparation
More information1. Question Answer cgy / MU cgy / MU 2. Question Answer
GS020113: Introduction to Medical Physics III: Therapy s to home work problem set assigned on 3/22/11 1. Question A patient is set up at 100 cm SSD on a 6 MVX machine. The dose rate at 10 cm in phantom
More informationIntroduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components
Introduction The CyberKnife SRS/SBRT System James M. Hevezi,, Ph.D., FACR CyberKnife Unit San Antonio, Texas CyberKnife Components Treatment Delivery (G3 Platform San Antonio) Robot (modified German auto
More informationClinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas Current Implementation Status Current Implementation Status 1600 1400 TOTAL 1494 of 1623 ACTIVE INSTITUTIONS
More informationOnly 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.
Clinical Implementation of the TG-51 Protocol David Followill Radiological Physics Center Houston Texas 2009 AAPM Summer School Educational Objectives Improve your understanding of how to implement TG-51
More informationProton Therapy QA Tools
Proton Therapy QA Tools A complete range of tools for absolute dosimetry, machine QA and patient plan verification Introduction During the past years more and more heavy particle therapy centers, especially
More informationFailure Mode and Effect Analysis (FMEA) for a DMLC Tracking System
Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System Amit Sawant 1, Sonja Dieterich 1, Michelle Svatos 2 and Paul Keall 1 1 Stanford University, Stanford, CA 2 Varian Medical Systems, Palo
More informationUltrasound Dose Calculations
Ultrasound Dose Calculations One of the advantages of ultrasound therapy remains the reasonably broad range of trials from which effective treatment doses can be established. In principle, there is no
More informationWEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS
Acta Radiologica Oncology 23 (1984) Fusc. 6 FROM THE RADIATION THERAPY DEPARTMENT, ANTON1 VAN LEEUWENHOEK HOSPITAL, THE NETHERLANDS CANCER INSTITUTE, AMSTERDAM, THE PHILIPS MEDICAL SYSTEMS DIVISION, BEST,
More informationSurface buildup dose dependence on photon field delivery technique for IMRT
Surface buildup dose dependence on photon field delivery technique for IMRT Shigeru Yokoyama, Peter L. Roberson, Dale W. Litzenberg, Jean M. Moran, and Benedick A. Fraass Department of Radiation Oncology,
More informationManchester City Stars Football Club Club Constitution
Manchester City Stars Football Club Club Constitution SECTION 1. CLUB DETAILS 1. The club name will be Manchester City Stars Football Club, hereafter referred to as the club for the remainder of this document.
More informationCommissioning and periodic tests of the Esteya electronic brachytherapy system
Review paper Educational Activity Commissioning and periodic tests of the Esteya electronic brachytherapy system Cristian Candela-Juan, PhD 1, Yury Niatsetski, MSc 2, Zoubir Ouhib, PhD 3, Facundo Ballester,
More informationEvaluation of six TPS algorithms in computing entrance and exit doses
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Evaluation of six TPS algorithms in computing entrance and exit doses Yun I. Tan, 1,2a Mohamed Metwaly, 3 Martin Glegg, 3 Shaun P.
More information2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications
Application Description AG/266PILD-EN Rev. C 2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic Pressure Measurement Engineered solutions for all applications Increase plant productivity
More informationPositionMaster EDP300 Extended Diagnostics. Compact, well-proven, and flexible
Change from one to two columns Technical Information TI/EDP300_ED-EN Rev. A PositionMaster EDP300 Extended Diagnostics Compact, well-proven, and flexible The benefits of the extended diagnostics feature
More informationTG-51: Experience from 150 institutions, common errors, and helpful hints
TG-51: Experience from 150 institutions, common errors, and helpful hints Ramesh C. Tailor, Ph. D., William F. Hanson, Ph.D., and Geoffrey S. Ibbott, Ph.D. Department of Radiation Physics, University of
More informationReviewing Hyperbaric Oxygen Services: Consultation Guide
Information Reader Box (IRB) to be inserted on inside front cover for documents of 6 Reviewing Hyperbaric Oxygen Services: Consultation Guide NHS England INFORMATION READER BOX Directorate Medical Operations
More informationPicture This! 2016 Calendar Photo Contest
Picture This! 2016 Calendar Photo Contest For years we have utilized photos of our real guests (not models) in our marketing materials. So many guests have gotten a kick out of seeing themselves, or loved
More informationDisclosures. What Is So Hard? End of ITV: Gating is the Best ITV Killer 8/3/2016. Daniel A. Low, Ph.D. UCLA. Varian Grant Siemens Grant Accuray Grant
Real Timefullness 8/3/2016 End of ITV: Gating is the Best ITV Killer Daniel A. Low, Ph.D. UCLA Varian Grant Siemens Grant Accuray Grant Disclosures What Is So Hard? Light Ultrasound MRI X Ray Planar X
More informationRadiotherapy physics & Equipments
Radiotherapy physics & Equipments RAD 481 Lecture s Title: Beam Modification devices in Radiotherapy Dr. Mohammed Emam Vision :IMC aspires to be a leader in applied medical sciences, health care education
More informationPlanning Assignment (3 field rectum)
Megan Sullivan Clinical Lab Rectum April 25, 2016 Planning Assignment (3 field rectum) Use a CT dataset of the pelvis. Create a CTV by contouring the rectum (start at the anus and stop at the turn where
More informationProcessing SPECT VQ Studies
Processing SPECT VQ Studies Joseph O'Brien Clinical Scientist joseph.o'brien@nhs.net City Hospital, Birmingham How do we get from this to this? Perf Vent Topics QC Checks Processing Display Raw Data Verification
More informationIranian Journal of Medical Physics
Iranian Journal of Medical Physics ijmp.mums.ac.ir Verification of Monitor Unit Calculations for Eclipse Treatment Planning System by in-house Developed Spreadsheet Athiyaman Mayivaganan 1*, Hemalatha
More informationTEST REPORT BEAMSCAN WITH HALCYON
TEST REPORT BEAMSCAN WITH HALCYON UZ Leuven, November 20 th, 2017 + = Ideal BEAMSCAN tank dimensions to fit perfectly into the Halcyon bore Fast tank positioning using the Halcyon lasers, no electronics
More informationAalborg Universitet. Published in: Proceedings of Offshore Wind 2007 Conference & Exhibition. Publication date: 2007
Aalborg Universitet Design Loads on Platforms on Offshore wind Turbine Foundations with Respect to Vertical Wave Run-up Damsgaard, Mathilde L.; Gravesen, Helge; Andersen, Thomas Lykke Published in: Proceedings
More informationHost League. Website Complete Address INTERLEAGUE BOUTING AGREEMENT
Host League Website Complete Address INTERLEAGUE BOUTING AGREEMENT This contract serves as an agreement ( Agreement ) between League A, more commonly known as Abbreviation or nickname ( Hosting League
More informationdatasheet M25S 24VDC motor Features & benefits
datasheet M25S 24VDC motor Features & benefits Slow Start, Slow Stop Starting from any position, the M25S starts very slow for a better quality feel of movement. The end position is smoothly approached
More informationExradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers
EXRADIN Ion Chambers Exradin Ion Chambers Exradin (EXacting RADiation INstrumentation) Ion Chambers have been built for over 33 years, are recognized by top research institutes and standards laboratories,
More informationPHOTOGRAPHY SUBMISSION GUIDELINES FOR YEAR 2014 CALENDARS. August 3, 2012
T H E C A L E N D A R C O M P A N Y PHOTOGRAPHY SUBMISSION GUIDELINES FOR YEAR 2014 CALENDARS August 3, 2012 BrownTrout Publishers, Inc. ( BrownTrout ) is pleased to present the following Photography Submission
More informationSFHGEN77 - SQA Code HD48 04 Perform first line calibration on clinical equipment to ensure it is fit for use
Perform first line calibration on clinical equipment to ensure it is fit for Overview This standard relates to first line calibration of specific equipment prior to clinical. Individuals will be able to
More informationAbstract: Introduction:
Are photon and electron beam calibrations more consistent with TG-51 than with TG-21? Ramesh C. Tailor, William F. Hanson, Nathan Wells, and Geoffrey S. Ibbott. U.T. MD Anderson Cancer Center, Houston
More informationREGULATIONS FOR THE BLACK & WHITE, COLOUR AND NATURE FIAP BIENNIALS
FIAP Biennial Service DOC 025 / 2018 E REGULATIONS FOR THE BLACK & WHITE, COLOUR AND NATURE FIAP BIENNIALS 1. ORGANIZATION International competitions called FIAP BIENNIALS are organized amongst all the
More informationAAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,
More informationMethods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning
Washington University School of Medicine Digital Commons@Becker Open Access Publications 2016 Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning
More informationBREEDING CONTRACT FROZEN SEMEN
Foothills Farm of Sedro-Woolley LLC Breeding Manager/Stallion Agent Kari Klingenberg 360-202-9008 BREEDING CONTRACT FROZEN SEMEN This contract dated this day of, 20 is between herein referred to as Mare
More informationFor Sale by Bid 2002 Chev Astro Van
For Sale by Bid 2002 Chev Astro Van FOR SALE BY BID Bid# 007-18 One (1) 2002 Chev Astro Van Engine: 4.3 V6 Automatic Transmission 244,118 Kms 1 set of (4) winter tires 95% 1 set of (4) summer tires 90%
More informationPerformance of SSTH-70 after Delivery and Future of SSTH Masahiro Itabashi, Ryoji Michida Ishikawajima-Harima Heavy Industries Co.,Ltd.
Performance of SSTH-70 after Delivery and Future of SSTH Masahiro Itabashi, Ryoji Michida Ishikawajima-Harima Heavy Industries Co.,Ltd. ABSTRUCT The SSTH-70 Ocean Arrow designed under the concept of the
More informationThe dose distribution of medium energy electron boosts to the laryngectomy stoma
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 1, WINTER 2001 The dose distribution of medium energy electron boosts to the laryngectomy stoma Ellen D. Yorke* Memorial Sloan-Kettering Cancer
More informationkvue IGRT Couch top One couch top, multiple treatment solutions For Varian, Elekta, Siemens, Proton and more The Best of Both Worlds
TM kvue IGRT Couch top One couch top, multiple treatment solutions For Varian, Elekta, Siemens, Proton and more kvue The Best of Both Worlds kvue IGRT COUCH TOP One couch top, multiple treatment solutions
More informationHow to Design Medical Accelerator Systems for Reliability: IBA PT System
How to Design Medical Accelerator Systems for Reliability: IBA PT System Yves Jongen Founder & Chief Research Officer Ion Beam Applications s.a. Belgium 1 Outline A short introduction of a proton therapy
More informationUSA Water Ski Event Sanction Agreement
USA Water Ski Event Sanction Agreement This is a contract (the Agreement ) between USA Water Ski ( USA-WS: ), a New Jersey not-for-profit corporation, the event promoter(s) or director(s) and all other
More informationCommissioning and quality assurance for the treatment delivery components of the AccuBoost system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Commissioning and quality assurance for the treatment delivery components of the AccuBoost system Ileana Iftimia, 1,2a Mike Talmadge,
More informationHenry K. Van Offelen a, Charles C. Krueger a & Carl L. Schofield a a Department of Natural Resources, College of Agriculture and
This article was downloaded by: [Michigan State University] On: 09 March 2015, At: 12:59 Publisher: Taylor & Francis Informa Ltd Registered in England and Wales Registered Number: 1072954 Registered office:
More informationSontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004
Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004 INTRODUCTION Sontek/YSI has introduced new firmware and software for their RiverSurveyor product line. Firmware changes
More information