Analysis of prohormones in animal feed by GC-MS

Size: px
Start display at page:

Download "Analysis of prohormones in animal feed by GC-MS"

Transcription

1 Analysis of prohormones in animal feed by GC-MS Document: SOP ARO/514 versie 1 Auteur: Klaas van Twillert Documentgebied: ARO Methods Autorisator: Leen van Ginkel (labhoofd) Status: Definitief Pagina 1 van 10

2 1 Introduction This SOP describes the method of analysis for prohormones, listed in table 1, in animal feed by GC- MS. The prohormones are extracted from the animal feed samples by liquid-liquid extraction with tertiair-butylmethylether (TBME). The TBME is evaporated and the product is redissolved in 40/60 v/v- % methanol/water. This solution is washed with heptane to remove fatty compounds. Subsequently the prohormones are extracted by using a SPE C18 column and a NH SPE column. After derivatisation with heptafluorobutyric anhydride the prohormones are analysed with GC-MS. The method is a validated as a quantitative screening method (table 1). TABLE 1 OVERVIEW OF PROHORMONES nr Name trivial name molecular formula 1 4-estren-3β,17β-diol 19-nor-4-androsten- 3β,17β-diol C18H8O 5(10)-estren-3β,17βdiol 19-nor-5-androsten- C18H8O 3β,17β-diol 3 4-estren-3,17-dione 19-nor-4-androsten- C18H4O 3,17-dione 4 4-estren-17β-ol-3-one (β) 19-nortestosterone C18H6O 5 4-androsten-17β-ol-3- (β) testosterone C19H8O one 6 4-androsten-3β, 17βdiol C19H30O 7 5-androsten-3β, 17βdiol C19H30O 8 4-androsten-3, 17- C19H6O dione 9 5β-androstan-17β-ol- 5βdihydrotestosterone C19H30O 3-one 10 5-androsten-3β-ol-17- DHEA C19H8O one 11 4-estren-17α-ol-3-one (α) 19-nortestosterone C18H6O 1 4-androsten-17α-ol-3- (α) testosterone C19H8O one 13 5α-androstan-3β-ol- epiandrosterone C19H30O 17-one 14 1, 4-androstadien-3. C19H4O 17-dione 15 5-androsten-3, 17- C19H6O dione 17 5α-androstan-17β-ol- C19H30O 3-one 18 4-estren-4-chloro- androstanolone (5αdihydrotestosterone) C19H30O 17β-ol, 3-one 19 5-androsten-3β-ol-7, 7-keto DHEA C19H6O 17-dione 3 0 1, (5α)-androsten- 1- C19H8O 17β-ol-3-one dehydroandrostanolon mol weight (g/mol) CCa (ng/g ) CCβ (ng/g ) U (%) Pagina van 10

3 1 1, (5α)-androsten-3, 17-dione e 1-androstenedione C19H6O Apparatus Reference to a company and/or product is for purpose of identification and information only..1 Bottles, 10 ml, brown glass, with screw cap.. SPE C18 column, 1 g, 6 ml, Alltech # SPE NH column, 1 g, 6 ml, Alltech # Aspec XL, an automatic sample preparation system, Gilson. The Aspec consists of a model 401 dilutor, a sample processor and racks to handle SPE columns..5 GC-MS equipment;.5.1 A Hewlett Packard 6890 GasChromatograph equipped with a 6890 injector and a Mass Selective Detector (MSD 5973)..5. Injector: injection vol. : μl. injection mode : pulsed splitless injector temp : 50 C pressure : 14.0 psi pulse time : 1.10 min purge flow : 40.0 ml/min. purge time : 1.00 min. total flow : 43.8 ml/min. gas saver : On saver flow : 0.0 ml/min. saver time : 6.00 min..5.3 Column: Factor Four capillary column, VF-1MS, 60 m x 0.5 mm i.d., DF 0.5 μm, Varian # CP Max. temp.: 345 C Mode: constant flow Initial flow: 0.7 ml/min. Temperature program: initial temp: 100 C, initial time min. Rate ( C/min.) Final temp. Final time (Off) Run time: 37.1 min. Pagina 3 van 10

4 .5.4 Detector: ions are recorded in SIM mode..6 Plastic centrifuge tubes, 50 ml. 3 Safety and environment The RIVM has defined rules to work safe. These rules can be found on the intranet. Waste is disposed according to SOP ARO/487. More information how to handle waste is found on the RIVM intranet. 4 Chemicals and reagents Reference to a company and/or product is for purpose of identification and information only. All reagents must be of analytical grade, unless otherwise specified. 4.1 Water; Milli-Q, Waters. 4. Tris(hydroxymethyl)aminomethane 4..1 Tris buffer, 0.1 molair, ph 9.5. Dissolve 1.1 g Tris(hydroxymethyl)aminomethane in 800 ml water. Adjust the ph to 9.5 ± 0.1 with hydrochloric acid and add water to a volume of 1000 ml. 4.3 Tertiair-ButylMethylEther (TBME). 4.4 Methanol Methanol/water, 40/60 v/v-%. Mix 400 ml methanol with 600 ml water. 4.5 Heptane. 4.6 Ethanol. 4.7 Acetone. 4.8 Heptafluorobutyric Anhydride (HFBA); Pierce # /75 v/v-% HFBA acetone mixture: mix 50 μl HFBA and 750 μl acetone. 4.9,, 4-Trimethylpentane (iso-octane) Stock solution: Weight 3 (=x) mg prohormone in a 10 ml bottle. Dissolve the prohormone in x ml ethanol. This is the 1 mg/ml stock solution and is stored in the dark at -180C. Working solution: Dilute the stock solution 1:10 with ethanol. This is the 100 μg/ml working solution and is stored at 40C Prohormone enrichment mixture. Bring 100 μl of all the prohormone working solutions in a 10 ml bottle and fill up to 10 ml with ethanol. This is the 1 μg/ml (1 ng/μl) enrichment mixture Internal standard mixture. A mixture of internal standards is used in this experiment: Prepare as described in.10 and.10.1 working solutions of Testosteron-d and Nortestosteron-d3. Pagina 4 van 10

5 Bring 100 μl of these working solutions in a 10 ml bottle and fill up to 10 ml with ethanol. This is the 1 ng/μl internal standard mixture. 5 Procedure Sometimes the animal feed samples, suspensed in the Trisbuffer, form a gel when they are shaken with TBME. It concerns the samples with a white, milkpowder like appearance. These samples are handled in an alternative way. The lines hereunder are then replaced with lines Normal clean up 5.1 Samples are stored at -0 C until analysis. 5. Weight a gram portion of the sample in a 50 ml plastic tube. 5.3 Weight 6 portions of blank animal feed in plastic tubes to prepare the calibration line. 5.4 Spike these samples animal feed with a mixture of prohormones standards according to table. TABLE PREPARATION OF CALIBRATION CURVE Code of Standard Volume (μl) enrichment mixture (.10.) concentration in ng/g Add 40 μl internal standard mixture (.11) to all tubes. 5.6 Add 5 ml Trisbuffer to all tubes and shake the tubes. 5.7 Add 10 ml Tertiair-ButylMethylEther (TBME) to the tubes. 5.8 Extract the prohormones into the TBME-layer by rotating the tubes during 5 min. 5.9 Centrifuge the tubes for 10 min with 900 g Transfer the upper (organic) layer into a glass testtube and evaporate the TBME at 60 C under a gentle stream of nitrogen Bring another 5 ml of TBME in the plastic tubes and repeat the extraction procedure ( ). 5.1 Bring this TBME layer too in the glass testtube and evaporate Dissolve the dry residue in 5 ml 40/60 v/v-% methanol/water Add 1 ml heptane and shake vigorously on a Vortexmixer to remove fatty components Centrifuge the testtubes during 10 min with 900 g Discard the heptane layer Filter the remaining methanol water fraction via a 5 ml pipette tip filled with a small wad of cotton wool into another glass testtube (small, 75 x 1 mm) Clean this solution by use of SPE columns (C18, NH). The SPE clean up is performed by using the Aspec XL. The complete method is described in the ASPEC-method METHPROHORMONEN. See appendix 1. A short description of the method: a C18 SPE column is activated with methanol and flushed with water. The sample is passed through the column. The column is washed with water, followed by a 40/60 v/v-% methanol/water wash. The components are eluted from the C18 column into a testtube with acetone. The NH column is preconditioned with acetone. The sample is passed through this column and collected by air push in a testtube Evaporate the acetone at 60 C under a gentle stream of nitrogen. Pagina 5 van 10

6 5.0 Dissolve the residue in 1 ml ethanol and transfer this into a ml derivatisation vial. 5.1 Evaporate the ethanol under N flow at 60 C. 5. Derivatise the prohormones with 5 μl of 5/75 v/v-% HFBA acetone mixture during 1 hour at 60 C. 5.3 Remove the HFBA acetone mixture under N flow at 60 C. 5.4 Redissolve the product in 5 μl iso-octane and transfer it into an insert which is thereupon placed in this vial and mount a new screwcap. 5.5 Place the vials in the GC-MS autosampler and inject μl on the GC-MS. 5.6 Detect the components according to m/z values in table Add another 50 μl iso-octane in the insert and reinject the sample when the peaks have a bad shape (e.g. broader, tailing) compared to the standards. Alternative clean up 5.8 Add 10 ml Tertiair-ButylMethylEther (TBME) to the tubes. 5.9 Rotate these tubes during 90 min head over head Centrifuge the tubes during 10 min with 900 g Transfer the upper TBME layer through a paper Schleicher & Schüll filter (S&S whiteband, diam. 110 mm, ref.# ) into a 10 ml pipette tip equipped with a clad cotton wool in the tip. 5.3 Collect the TBME fraction in a glass test tube Evaporate the TBME fraction at 60 C under a gentle stream of nitrogen 6 Calculation Check The first step is a performance check of the GC-MS system after tuning. This is done by injecting a 10 ng/g standard (table ). The S/N ratio of the components is determined. This must be >100. Calculation Areas of the selected ion of the standard and of the internal standard are calculated. The ratio is the response variable. A calibration curve is constructed by linear curve fitting using least squares linear regression calculation. Unknown concentrations are calculated by interpolation. TABLE 3 DIAGNOSTIC IONS HFBA DERIVATIVES nr name m/z prohormone Other ions Internal standard M/z int. stand. 1 4-estren-3β,17β-diol nortestosteron(d3) (10)-estren-3β,17β-diol nortestosteron(d3) estren-3,17-dione nortestosteron(d3) estren-17β-ol-3-one nortestosteron(d3) androsten-17β-ol-3-one testosteron(d) androsten-3β, 17β-diol testosteron(d) androsten-3β, 17β-diol testosteron(d) androsten-3, 17-dione testosteron(d) β-androstan-17β-ol-3-one testosteron(d) androsten-3β-ol-17-one testosteron(d) estren-17α-ol-3-one nortestosteron(d3) androsten-17α-ol-3-one testosteron(d) α-androstan-3β-ol-17-one testosteron(d) 68. Pagina 6 van 10

7 14 1,4-androstadien dione testosteron(d) androsten-3, 17-dione testosteron(d) α-androstan-17β-ol-3-one testosteron(d) estren-4-chloro-17β-ol, 3-one testosteron(d) androsten-3β-ol-7, 17-dione testosteron(d) , (5α)-androsten-17β-ol-3-one testosteron(d) , (5α)-androsten-3, 17-dione testosteron(d) 68. Note: other ions can be used for confirmation purposes to fulfil the confirmation criteria as described in commission decision 00/ Validation and Measurement uncertainty The method described in this SOP was validated during routine control experiments. Samples were spiked at 0, 5, 10, 0, 5 and 50 μg kg-1. The additions of three separate experiments were combined and further evaluated by use of weighted regression and the model Y=aX+b. From these data is conform CD 00/657 the CCα and CCβ calculated according the following formulas. See table 1 for an overview of the CCa and CCß. The reproducibility is determined by combining the results from the three separate series and also combining the results of the different spiking levels (the data is normalized). In table 4 is the results for the determination of the reproducibility shown. TABLE 4 RESULT DETEMINATION REPRODUCIBILITY Concentration (ng/g) 4-estren-3b,17b-diol 5(10)-estren-3b,17b-diol 4-estren-3,17-dione * 4-estren-17b-ol-3-one 4-estren-17a-ol-3-one 4-androsten-17b-ol-3-one 4-androsten-3b,17b-diol 5-androsten-3b,17b-diol 4-androsten-3,17-dione/5- androsten-3,17-dione 5b-androstan-17b-ol-3-one 5 16,6 13,8 103,4 116,6 116, 18, 138,8 117,0 111,8 103, ,5 115,0 94,8 98,9 95,0 105,4 103,4 98,0 99,6 98,9 0 16,5 10,5 100,7 106,7 106,7 11,0 131,1 101,9 106,3 98,8 5 79,3 88,4 8,9 89, 88,5 9,0 77,9 86,1 84,9 83, ,8 105,8 91,1 101,9 104,6 101,3 103,5 99,0 90,0 104,0 Pagina 7 van 10

8 5 90,6 113, 96,4 10,8 10,8 114,8 91,0 100, 99,0 98, ,9 10,8 89,5 100, 104,6 104,9 104,1 103,4 9,3 98, ,8 97,3 87,7 97,4 10,6 98,5 114,5 9, 87,9 88,1 5 93,0 101,4 93,0 100,4 101,3 100,5 96,1 99,4 9, 91, , 97,4 78,6 88, 88, 101,6 97,4 111,8 90,0 95, 10 87,3 107,0 90,6 99,6 99,5 110,6 10,1 101,0 97,9 109,7 0 89,1 106,6 88,0 100, 100,9 104, 89,7 99,9 95,5 97, , 103,8 91,6 97,7 99, 100,9 95,8 103,1 95,1 109, ,1 101,8 90,8 98,8 99,1 98,7 9,9 100,9 9,1 104,1 average 11,1 104,8 91,4 99,9 100,7 105,3 10,7 101,0 95,3 98,8 CV (%) 3,4 8, 7,0 6,8 7,1 8,5 15,7 7,3 7,6 7,6 U 64,8 16,4 14,1 13,6 14,1 16,9 31,4 14,5 15, 15,1 Concentration (ng/g) 5-estrene-3a--ol-17-one 4-androsten-17a-ol-3-one 5a-androstan-3b-ol-17-one 1,4-androstadien-3,17-dione* 5a-androstan-17b-ol-3-one 4-estren-4-chloro-17b-ol,3-one 5-androsten-3b-ol-7,17-dione 1,(5a)-androsten-17b-ol-3-one 1,(5a)-androsten-3,17-dione 5 18,8 1,4 11,0 15,8 113,4 8,8 93,6 18, 11, ,1 91,4 107, 115,0 97,8 76,1 13,9 13,0 103, ,1 103,6 101,7 11,4 101,4 74,4 88,6 116,8 98,7 5 78,7 88, 84,4 90,5 83, 63,1 77,7 95,8 84, ,9 99,6 10,3 94,7 10,7 57,9 61,9 106,9 101, , 105, 107,6 11,0 86,0 60,8 118,0 14,4 78, ,9 109,3 9,1 99,8 90,3 54,9 78,6 14,7 80,5 0 9,7 10,5 86,0 94,5 87,7 54,0 61,3 1,8 78,7 5 90,0 101,3 93,4 94,6 91, 59,8 67,8 103,7 74, ,6 10,6 91,6 10, 98,6 65,4 11, 114, 97, ,5 114,7 96,4 111,0 108,1 67,5 90,1 118,0 99, ,4 108,9 10,6 114,1 103,8 65, 7,6 104,0 103,9 5 96, 108, 98,7 10,1 99,4 71,8 78, 116,9 100, ,0 103,6 97,8 98,0 100,5 67, 7,7 110,0 99,8 average 101,4 104,4 98,1 104,8 97,4 65,8 86, 115,6 93,7 CV (%) 1,9 8, 8,3 9,8 8,9 1,7 4,6 9,1 1,8 U 5,9 16,5 16,5 19,7 17,9 5,3 49, 18,1 5,5 * Compounds are not separated on the GC and share therefore the validation results 8 Quality control Quantification is only valid if: The maximum of the signal originating from the analyte has an S/N ratio >3. In the blank control samples all the internal standards are visible. In the spiked control samples all components are visible (S/N ratio >10 for internal standards, >6 for the non-deuterated compounds). 9 Responsibilities See ARO/MIS FUNC for the organization of responsibilities within ARO. 10 Relating documents Pagina 8 van 10

9 10.1 SOP ARO/475, Method validation using ResVal, revision, 1 april 004, RIVM. 10. Multi residue analysis anabolic agents, SOP ARO/113, revision 4, 1 january 1997, RIVM Commission Decision 00/657/EC of 1 August 00 implementing Council Directive 96/3/EC concerning the performance of analytical methods and the interpretation of results Annex The cleanup method with the Aspec: Methprohormonen. Pagina 9 van 10

10 Documentbeheer Algemeen Invoerdatum: 0 februari 007 Wijzigingsdatum: 14 augustus 007 Controledatum: 31 december 01 Publicatiedatum: januari 008 Wijzigingen ten opzichte van vorige versie: Omzetting naar HTML vervangt SOP ARO/481 Beoordelaars Marco Blokland (Onderzoeker) Saskia Sterk (afdelingshoofd) Hyperlinks Pagina 10 van 10

1. INTRODUCTION CH 3 CH3

1. INTRODUCTION CH 3 CH3 1. ITRODUCTIO TITLE: Analysis of stanozolol and metabolites S.O.P. : ARO/446 in urine Page : 1 of 12 This method describes the analysis of stanozolol, 4ß-hydroxy-stanozolol and 16ß-hydroxy-stanozolol in

More information

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15 Chemicals needed: Precipitate solution, Methanol Derivatization Solution, Amplifex Keto Reagent Kit Mobile phases, H 2 0 with 0.1% Formic Acid LC/MS grade and Acetonitrile with 0.1% Formic Acid LC/MS grade

More information

CORESTA RECOMMENDED METHOD Nº 67

CORESTA RECOMMENDED METHOD Nº 67 CORESTA RECOMMENDED METHOD Nº 67 DETERMINATION OF WATER IN THE MAINSTREAM SMOKE OF CIGARS BY GAS CHROMATOGRAPHIC ANALYSIS (November 2005) 1. FIELD OF APPLICATION The method is applicable to the particulate

More information

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS UCT Part Numbers: CUQAX22Z Clean-Up C8+QAX, 2mg/1mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQID21-3UM

More information

GCMSD-Headspace Analysis SOP

GCMSD-Headspace Analysis SOP Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to

More information

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative

More information

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion: Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion

More information

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood UCT Part Numbers: ECQUUS1015CT- Enviro-Clean 15 ml centrifuge tube with 400 mg MgSO 4 and 100 mg NaCl CUMPS2CT - Enviro-Clean

More information

Rapid and Reliable Detection of Dissolved Gases in Water

Rapid and Reliable Detection of Dissolved Gases in Water Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,

More information

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Outline of Topics What is a ghost peak? How to identify ghost peaks and their source How to correct

More information

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National

More information

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS Abstract Highly sensitive analysis of metanephrines in plasma is critical in the diagnosis and treatment of pheochromocytoma and paraganglioma. Here, a HILIC LC-MS/MS method was developed using a Raptor

More information

May 1, By John Heim,Doug Staples

May 1, By John Heim,Doug Staples 1 of 7 7/11/2014 3:39 PM May 1, 2010 By John Heim,Doug Staples Anabolic steroid screening analysis in urine is complex and labor intensive requiring sensitive instrumentation and optimized chromatographic

More information

Chemical Analysis for Methyltestosterone

Chemical Analysis for Methyltestosterone APPLIED CONSUMER SERVICES, INC. Chemical Analysis for Methyltestosterone Report #: 27584 Report to: Mr. Mitch Ross 24 Dockside Lane # 131 Key Largo, Florida 33017 2016 1 1 8 9 0 N W 8 7 t h C o u r t,

More information

METHOD: 1024, Issue 2 EVALUATION: FULL Issue 1: 15 August 1987 Issue 2: 15 August 1994

METHOD: 1024, Issue 2 EVALUATION: FULL Issue 1: 15 August 1987 Issue 2: 15 August 1994 1,3-BUTADIENE 1024 H 2 C=CHHC=CH 2 MW: 54.09 CAS: 106-99-0 RTECS: EI9275000 METHOD: 1024, Issue 2 EVALUATION: FULL Issue 1: 15 August 1987 Issue 2: 15 August 1994 OSHA : NIOSH: ACGIH: 1000 ppm lowest feasible;

More information

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation Using Nitrogen instead of Helium as carrier gas, while maintaining exact the same separation efficiency, retention times, peak elution order, without changing the oven temperature conditions. Jaap de Zeeuw

More information

WADA Technical Document TD2014EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

WADA Technical Document TD2014EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting Endogenous Anabolic Androgenic Steroids Measurement and Reporting 1.0 Introduction The purpose of this Technical is to harmonize the approaches to the measurement and reporting of endogenous anabolic androgenic

More information

EVALUATION OF PROFICIENCY TESTING PROGRAM, SETTING PRIORITIES FOR THE FUTURE.

EVALUATION OF PROFICIENCY TESTING PROGRAM, SETTING PRIORITIES FOR THE FUTURE. Click to edit Master title style EVALUATION OF PROFICIENCY TESTING PROGRAM, SETTING PRIORITIES FOR THE FUTURE. Saskia S. Sterk Henk A. Herbold, Marco H. Blokland October 2005. Introduction. One of the

More information

Figure 1: Plug in ion inlet of mass spec

Figure 1: Plug in ion inlet of mass spec Starting procedure: by Vicky - is the vacuum pump running (should run all time, otherwise it ll need 4-5 hours to have a adequate vacuum for the MS) otherwise something has gone wrong- F.T - Remove plug

More information

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent Gas Chromatograph interfaced to the Agilent

More information

Experiment GC : Analysis of BTEX by GC-FID

Experiment GC : Analysis of BTEX by GC-FID 46 Experiment GC : Analysis of BTEX by GC-FID Learning Goals: Familiarity with gas chromatography Gain experience in temperature programming and method development Correctly use an internal standard, and

More information

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis PO-CON695E High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis MSACL 6 EU Neil Loftus, Stephane Moreau, Mikaël Levi 3, Anja Grüning

More information

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS) Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent 6890 GC interfaced

More information

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS Application Note: Analysis of and Cyanuric Acid in Food Matrices by LC-MS/MS PeterVarelis, National Center for Food Safety and Technology, Illinois Institute of Technology. Jonathan Beck, Kefei Wang, and

More information

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries. Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s):

More information

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013 The SRI ASTM 3612/TOGA/ Dissolved Gas GC configuration ( TOGA GC ) permits measurement of gasses dissolved in water, oil or other liquids, and is based on the requirements of ASTM method 3612, option C,

More information

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking

More information

no peaks missing peaks too small peaks increasing retention decreasing retention

no peaks missing peaks too small peaks increasing retention decreasing retention GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering

More information

Methylprednisolone detection in urine following local and oral administrations

Methylprednisolone detection in urine following local and oral administrations Susana MS Simões, Marta Calçada, Luís Horta, Xavier de la Torre Methylprednisolone detection in urine following local and oral administrations Laboratório de Análises e Dopagem. Instituto do Desporto de

More information

Prep LC Training Manual S.V.

Prep LC Training Manual S.V. 1234 Hach Hall 515-294-5805 www.cif.iastate.edu Prep LC Training Manual 11-6-2013 S.V. Location: Contact: 1238 Hach Hall Kamel Harrata, 1236 Hach Hall and Steve Veysey, 1234 Hach Hall Safety: All researchers

More information

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Gas Chromatography Author Matthew S. Klee Agilent Technologies 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract

More information

National Institute for Public Health and the Environment Annual CRL workshop 22 October Update on natural Hormone studies

National Institute for Public Health and the Environment Annual CRL workshop 22 October Update on natural Hormone studies 2008 Annual CRL workshop 22 October 2008 Update on natural Hormone studies Natural hormone studies: update Possible approaches - C12/C13 ratio: can result in proof of abuse - Determination of intact esters

More information

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

Acclaim RSLC. (Rapid Separation Liquid Chromatography) Acclaim RSLC (Rapid Separation Liquid Chromatography) Product Manual for Acclaim RSLC Columns Page 1 of 9 Product Manual for Acclaim RSLC Columns (Rapid Separation Liquid Chromatography) Analytical Columns

More information

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek Abstract: Due to the expansion of natural gas drilling through horizontal fracturing, there has been

More information

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS Section A: Intro to the spectrophotometer A commonly used instrument in the analysis of cellular extracts is the Spectrophotometer. Today you

More information

Carrier Gases in Capillary GC

Carrier Gases in Capillary GC Carrier Gases in Capillary GC LC Columns and Consumables Mark Sinnott Application Engineer January 15, 2009 CARRIER GAS Mobile Phase Carries the solutes down the column Selection and velocity influences

More information

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind

More information

SOP: Buck Scientific BLC-20P HPLC Operation

SOP: Buck Scientific BLC-20P HPLC Operation Page 1 of 11 Approvals Preparer: John Buford Reviewer: Tim Kull Reviewer: Dr. Margaret Bryans Date: 16OCT13 Date: 30OCT13 Date: 31OCT13 1. Purpose 1.1. Basic operation of the Buck Scientific BLC-20P isocratic

More information

MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 42 Ultra-Sensitive Kit

MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 42 Ultra-Sensitive Kit MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 42 Ultra-Sensitive Kit The first protocol has been optimized for quantifying Aβ 1-42 peptide in human cerebrospinal fluid (CSF). An additional general protocol

More information

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay Summary This assay measures Human Granulocyte Colony Stimulating Factor (G-CSF) in a 96-well

More information

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation User s Booklet for the Wyatt minidawn Light Scattering Instrumentation The Wyatt minidawn Light Scattering instrument is able to measure the weight average molecular weight of a synthetic polymer or a

More information

MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 40 Ultra-Sensitive Kit

MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 40 Ultra-Sensitive Kit MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 40 Ultra-Sensitive Kit The first protocol has been optimized for quantifying Aβ 1-40 peptide in human cerebrospinal fluid (CSF). The second protocol has been

More information

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more! After years of development and testing, DPS Instruments is pleased to present the newest, most expandable and versatile Gas Chromatography Systems in history. The DPS 600 Series GC systems are the world

More information

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON' DISSOLUTION ASSAY Release & Stability Studies Ed. 04. General Index: 1. PRODUCT SPECIFICATIONS (USP MONOGRAPH) 2. IDENTIFICATION BY HPLC 3. ASSAY - HPLC SETUP 4. STANDARD PREPARATION 5. SYSTEM SUITABILITY

More information

LCMSMS Solutions For Steroid Analysis

LCMSMS Solutions For Steroid Analysis LCMSMS Solutions For Steroid Analysis Dan Blake Manager, Clinical Applications For Research Use Only. Not for use in diagnostic procedures. RUO-MKT-18-1166-A Agenda Special considerations for steroid analysis

More information

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve USER'S GUIDE Nanoflow Probe 3-4mm Red Stripe ptfe Sleeve To Injector ZU1XC Fused Silica 20µm 90µm TTF115 ptfe Sleeve Fused Silica 25µm 375µm Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: +44

More information

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column Before you connect your trap and column You need to provide your own trap and column (Note our fittings accommodate trap and column made with fused silica 360 µm OD). Check our guide for packing traps

More information

PROTOCOLS. Coomassie Brilliant Blue G (Sigma 98%) ,4 mg. Ethanol 95% ml. Phosphoric acid 85%.(w/v) ml

PROTOCOLS. Coomassie Brilliant Blue G (Sigma 98%) ,4 mg. Ethanol 95% ml. Phosphoric acid 85%.(w/v) ml PROTOCOLS I. Bradford s method to quantify total protein concentration (Bradford, D, M. M. A rapid and sensitive method for the quantitaton of microgram quantities of protein utilizing the principle of

More information

RayBio Human vwf ELISA Kit

RayBio Human vwf ELISA Kit RayBio Human vwf ELISA Kit Catalog #: ELH-vWF User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross, GA

More information

Validation of the MRM pesticides from the Working Document SANCO/12745/2013 of November 2017 rev. 9(1)

Validation of the MRM pesticides from the Working Document SANCO/12745/2013 of November 2017 rev. 9(1) Validation of the MRM pesticides from the Working Document SANCO/12745/2013 of 21-22 November 2017 rev. 9(1) CONTENTS 1. Aim and scope... 2 2. Short description... 2 3. Apparatus and consumables... 2 4.

More information

The instrument will process the samples one by one till all the samples have been cleaned up.

The instrument will process the samples one by one till all the samples have been cleaned up. Table of Contents SPE-01 Cleanup Station... 1 LC-04SP Valve System... 5 LC-04SP Standalone Valve... 10 LC-05 Auto Injector... 12 LC-06 Dosing Pump... 16 LC-08 Column Selector... 17 SPE-01 Cleanup Station

More information

Prep LC Training Manual

Prep LC Training Manual Prep LC Training Manual (11-6-2013) Location: 1238 Hach Hall Contacts: kamel Harrata and Steve Veysey, 1234 Hach Hall Safety: Introduction: The Agilent PrepStar SD-1 preparative HPLC is a student operated

More information

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide

More information

Longterm Detection of Anabolic Steroid Metabolites in Urine

Longterm Detection of Anabolic Steroid Metabolites in Urine Longterm Detection of Anabolic Steroid Metabolites in Urine Application Note Forensics/Doping Control Author Stephan Baumann Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara, CA 95051 USA

More information

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS Analysis Conducted BY CC RESEARCH (P) Ltd., For Nanoceutical Solutions Inc. USA 2950 Thousand Oaks drive, Suite 3, San Antonio, Texas EXECUTIVE SUMMARY

More information

SPECIFIC DETECTION OF ANDROST-4-ENE-3,17-DIONE

SPECIFIC DETECTION OF ANDROST-4-ENE-3,17-DIONE 7 SPECIFIC DETECTION OF ANDROST-4-ENE-3,17-DIONE ADMINISTRATION BY PROFILING OXYGENATED METABOLITES Abstract For the selective detection of androst-4-ene-3,17-dione ( 4-AEDIONE) administration, the analysis

More information

COMMUNITY REFERENCE LABORATORIES IN THE FIELD OF VETERINARY PUBLIC HEALTH WITHIN THE EUROPEAN UNION

COMMUNITY REFERENCE LABORATORIES IN THE FIELD OF VETERINARY PUBLIC HEALTH WITHIN THE EUROPEAN UNION COMMUNITY REFERENCE LABORATORIES IN THE FIELD OF VETERINARY PUBLIC HEALTH WITHIN THE EUROPEAN UNION CRL for residues RIVM-ARO at Bilthoven, NL --------------------------------- Annual Report January 2003

More information

FAST EVAPORATION AT NORMAL PRESSURE

FAST EVAPORATION AT NORMAL PRESSURE FAST EVAPORATION AT NORMAL PRESSURE FAST EVAPORATION BY BLOWING-DOWN... The VAPORNADO Plus blow-down system is an innovative and efficient instrument for the evaporation of samples and fractions at normal

More information

Figure. Experimental set up of catalyst, reactants and product syringes.

Figure. Experimental set up of catalyst, reactants and product syringes. Construction and Use of the Gas Reaction Catalyst Tube** As shown in the figures, the experimental apparatus consists of a 60 ml syringe containing the gaseous reactants connected to a gas phase catalyst

More information

CHM250 Calibration and Measurement Lab. Balance Calibration

CHM250 Calibration and Measurement Lab. Balance Calibration CHM250 Calibration and Measurement Lab Green Profile Balance Calibration Introduction: Balances that are properly operated, calibrated and maintained are crucial for laboratory operations. The accuracy

More information

Analysis of Casein and Whey Protein in Whole, 2%, and Skim Milk by Capillary Gel Electrophoresis

Analysis of Casein and Whey Protein in Whole, 2%, and Skim Milk by Capillary Gel Electrophoresis Analysis of Casein and Whey Protein in Whole, 2%, and Skim Milk by Capillary Gel Electrophoresis Marcia Santos, Staff Applications Scientist, Beckman Coulter Life Sciences, Brea, CA USA Mark Lies, Marketing

More information

Gas chromatographic determination of 1,4-dioxane at the

Gas chromatographic determination of 1,4-dioxane at the J. Soc. Cosmet. Chem., 42, 97-104 (March/April 1991) Gas chromatographic determination of 1,4-dioxane at the parts-per-million level in consumer shampoo products MARK P. ITALIA and MATHEWS A. NUNES, Johnson

More information

This standard is due for Five Year Review. This process is required by the SEMI Regulations to ensure that the standard is still valid.

This standard is due for Five Year Review. This process is required by the SEMI Regulations to ensure that the standard is still valid. Background Statement for SEMI Draft Document 5211 REVISION OF SEMI C3.58-0303 SPECIFICATION FOR OCTAFLUOROCYCLOBUTANE, C 4 F 8, ELECTRONIC GRADE IN CYLINDERS, 99.999% QUALITY Note: This background statement

More information

NMR Service. How to Prepare Samples for NMR

NMR Service. How to Prepare Samples for NMR NMR Service How to Prepare Samples for NMR In NMR, unlike other types of spectroscopy, the quality of the sample has a profound effect on the quality of the resulting spectrum. If you follow a few simple

More information

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler Optimizing Vial Pressurization Parameters for the Analysis of Residual Solvents Using the 7697A Headspace Sampler Application Note Pharmaceuticals Author Roger L Firor Agilent Technologies, Inc.

More information

Extended Application Note

Extended Application Note Extended Application Note Harmful Substances in Dietary Supplements HPLC Columns with APP A-337 www.mtc-usa.com 1-732-578-1777 INTRODUCTION A dietary supplement is a substance to be consumed for the purpose

More information

Pipetting and Determining Protein Concentration

Pipetting and Determining Protein Concentration Pipetting and Determining Protein Concentration Background Information: When performing experiments in Cell Biology, it is often necessary to use very small volumes of reagents sometimes because the reagents

More information

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE. Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to

More information

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer application Note Gas Chromatography/ Mass Spectrometry Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer Overview A mass spectrometer

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 5.1 Subatmospheric

More information

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin Development of DMS and Acetonitrile, and Formaldehyde gas standards Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin The amount of chemical substances in cylinder container Purity uncertainty Gravimetry

More information

RayBio Human TNF-alpha ELISA Kit

RayBio Human TNF-alpha ELISA Kit RayBio Human TNF-alpha ELISA Kit Catalog #: ELH-TNFa User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross,

More information

Automated Determination of Dissolved Gases in Water Anne Jurek

Automated Determination of Dissolved Gases in Water Anne Jurek Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: Tapping the natural gas reservoirs throughout the United States has long been a viable solution for energy independence; however

More information

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone Development of a Clinical Research Method for the Measurement of Testosterone and Dihydrotestosterone Martijn van Faassen, Maarten van den Noort and Ido Kima University Medical Center, Groningen, Netherlands

More information

AKTA pure 25 New Owner s Intro

AKTA pure 25 New Owner s Intro AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps

More information

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by: 1 AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by: Materials/Reagents/Equipment Vendor 1. 0.2 um filtered Water + 0.05% sodium azide 2. 0.2 um filtered

More information

Impact of Air Leaks on the Productivity of GC and GC/MS Systems

Impact of Air Leaks on the Productivity of GC and GC/MS Systems Impact of Air Leaks on the Productivity of GC and GC/MS Systems Application Note Environmental Author Ken Lynam Agilent Technologies, Inc. Abstract Agilent 7890A GC dual-channel FID, Agilent 7890B GC,

More information

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service? Pipette specifications according to ISO 8655 Chapter 5 5.1 The ISO 8655 standard gives the accuracy and precision limits as both absolute and relative values. Specifications will depend on the technique

More information

Introduction. Experimental

Introduction. Experimental PO-CON162E Development of Automated Screening and Quantitation Approach on Novel On-Line SFE-SFC-MS/MS Platform (I) For 23 Restricted Perflurocompounds in Textiles ASMS 2016 MP 283 Jie Xing, Jun Xiang

More information

CHEMTRACE Fremont Blvd. Fremont, CA 94538, Tel: (510) Fax: (510)

CHEMTRACE Fremont Blvd. Fremont, CA 94538, Tel: (510) Fax: (510) Report Date: 2/12/04 Page 1 of 5 Analysis Requested: Organic Outgassing Analysis by GC-MS at 100 C for 30 minutes. Sample Information: Sample ID: Wafer Carrier Material Samples were received wrapped in

More information

Liquid Handling Systems

Liquid Handling Systems Liquid Handling Systems 1.0 Purpose - This procedure specifies the required elements for the maintenance and use of the Hamilton- Microlab 1000 Plus and the Microlab 500 Liquid Handling Systems. 2.0 Scope

More information

Determination of Oxygenates in Gasoline by Gas Chromatography and Oxygen Selective Flame Ionization Detector Method

Determination of Oxygenates in Gasoline by Gas Chromatography and Oxygen Selective Flame Ionization Detector Method United States Environmental Protection Agency Office of Transportation and Air Quality National Vehicle and Fuel Emissions Laboratory 2565 Plymouth Road Ann Arbor, MI 48105 Determination of Oxygenates

More information

Agilent G Rebuilt Electron Capture Detector

Agilent G Rebuilt Electron Capture Detector Agilent G1533-90790 Rebuilt Electron Capture Detector User Information Agilent Technologies Notices Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any form or by any means

More information

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION Standard OQ Test Suite This document describes the test program for qualifying LC analytical-scale Agilent systems; the following table lists

More information

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction A Column-Flow Independent Configuration for QuickSwap Application Authors Matthew S. Klee and Bruce Quimby Agilent Technologies, Inc. 80 Centerville Road Wilmington, DE 08 USA Abstract As can be seen from

More information

Pipette apparatus. Meet the difference. Manual. T E I

Pipette apparatus. Meet the difference. Manual. T E I Pipette apparatus Manual Meet the difference Eijkelkamp Soil & Water Nijverheidsstraat 30, 6987 EM Giesbeek, the Netherlands T +31 313 880 200 E info@eijkelkamp.com I www.eijkelkamp.com 2018-07 M-0816E

More information

Renzo BAGNATI Mario Negri Institute of Pharmacological Research- IRCCS Milano - Italy

Renzo BAGNATI Mario Negri Institute of Pharmacological Research- IRCCS Milano - Italy STEROIDS IN WASTE WATERS AS MARKERS OF ABUSE Renzo BAGNATI Mario Negri Institute of Pharmacological Research- IRCCS Milano - Italy A first attempttoapplythe sewage epidemiologyapproachin doping Compare

More information

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions Varian 3300 Gas Chromatograph Abbreviated Operating Instructions 1. Introduction and Startup 2 2. Basic Configurations for the GC 3 3. Sample Preparation 3 4. Building a Method 3 5. Split/Splitless Injection

More information

HUMAN IL6 KITS PROTOCOL

HUMAN IL6 KITS PROTOCOL HUMAN IL6 KITS PROTOCOL Part # 62HIL06PEG & 62HIL06PEH Test size: 500 tests (62HIL06PEG), 10,000 tests (62HIL06PEH) - assay volume: 20 µl Revision: 04 (Jan. 2018) Store at: -60 C or below This product

More information

CHEM254 #4 The Synthesis of a Tertiary Alcohol Using a Pre-Made Grignard Reagent 1

CHEM254 #4 The Synthesis of a Tertiary Alcohol Using a Pre-Made Grignard Reagent 1 CHEM254 #4 The Synthesis of a Tertiary Alcohol Using a Pre-Made Grignard Reagent 1 Background In this project, we will perform a Grignard reaction using a pre-made Grignard reagent. Grignard reagents can

More information

Retention Time Locking: Concepts and Applications. Application

Retention Time Locking: Concepts and Applications. Application Retention Time Locking: Concepts and Applications Application Gas Chromatography December 1997 Authors Vince Giarrocco Bruce Quimby Matthew Klee Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor Anwar Sutan, Metco Services Ltd. Charles Johnson, Metco Services Ltd. Jason Laidlaw,

More information

Now it is. easy to switch. CHS TM Steroid Profiling kit and software. Brochure not for distribution in the USA and Canada

Now it is. easy to switch. CHS TM Steroid Profiling kit and software. Brochure not for distribution in the USA and Canada Now it is easy to switch to Clinical MSMS CHS TM Steroid Profiling kit and software Brochure not for distribution in the USA and Canada 2 Meeting your needs for steroid assays PerkinElmer s new CHS MSMS

More information

Svensk läkemedelsstandard

Svensk läkemedelsstandard Svensk läkemedelsstandard 2018.2 Nitrogen Postadress/Postal address: P.O. Box 26, SE-751 03 Uppsala, SWEDEN Besöksadress/Visiting address: Dag Hammarskjölds väg 42, Uppsala Telefon/Phone: +46 (0)18 17

More information

Zebron Gas Management

Zebron Gas Management Zebron www.phenomenex.com/gasmanagement What are the key benefits of Zebron filters and traps? The high adsorption capacity of the Zebron Filters removes contaminants from mobile phase gases, reducing

More information

LEVANT WORMSEED FOR HOMOEOPATHIC PREPARATIONS CINA FOR HOMOEOPATHIC PREPARATIONS

LEVANT WORMSEED FOR HOMOEOPATHIC PREPARATIONS CINA FOR HOMOEOPATHIC PREPARATIONS LEVANT WORMSEED FOR HOMOEOPATHIC PREPARATIONS CINA FOR HOMOEOPATHIC PREPARATIONS Artemisia cina ad praeparationes homoeopathicas DEFINITION Dried, non-blooming capitulum of Artemisia cina Berg. Content

More information

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation CHE 4115 Chemical Processes Laboratory 2 Experiment 1 Batch Distillation BACKGROUND Distillation is one of the most commonly used unit operations in chemical engineering. In general, a distillation operation

More information

Using the Akta Prime plus October 22, 2012

Using the Akta Prime plus October 22, 2012 Some starting precautions: 1. Vacuum filter all buffers. Removes any large particles/debris that may clog your column De-gases the buffers 2. Clarify lysates first by centrifugation and then filtration

More information

Composition of Exhaled Breath

Composition of Exhaled Breath Composition of Exhaled Breath Exhaled Breath (~37 C) is mainly composed by gaseous substances: Water Vapor, CO 2, O 2, N 2, NO, CO, etc. Several volatile, low-volatile and non-volatile compounds of endogenous

More information

VOLUMETRIC TECHNIQUES

VOLUMETRIC TECHNIQUES REVISED 10/14 CHEMISTRY 1101L VOLUMETRIC TECHNIQUES Volume measurements are important in many experimental procedures. Sometimes volume measurements must be exact; other times they can be approximate.

More information