Englisch. Operating Manual. Dissolved Oxygen Measuring transducer OXY 3610 MP .3X.6C Keep for later use.

Size: px
Start display at page:

Download "Englisch. Operating Manual. Dissolved Oxygen Measuring transducer OXY 3610 MP .3X.6C Keep for later use."

Transcription

1 .3X.6C -07 Englisch Operating Manual Dissolved Oxygen Measuring transducer OXY 3610 MP Companies / Brands of the GHM Keep for later use.

2 T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 2 von 14 Contents 1 SAFETY GENERAL NOTE INTENDED USE QUALIFIED STAFF SAFETY SIGNS AND SYMBOLS REASONABLY FORESEEABLE MISUSE SUITABILITY During configuration During normal operation PRODUCT DESCRIPTION SCOPE OF SUPPLY TYPE PLATE INSTALLATION AND COMMISSIONING GENERAL INSTALLATION INSTRUCTIONS ASSIGNMENT OF ELBOW-TYPE PLUG DISPLAY FUNCTIONS CALIBRATION OF THE OXYGEN MEASURING TRANSDUCER CONFIGURATION OF THE DEVICE CONFIGURATION PROCEDURE OFFSET OFFS Average ambient absolute pressure PABS Display damping FiLT Save the last parameter values ERROR AND SYSTEM MESSAGES OFFSET SETTINGS FOR DIFFERENT APPLICATIONS AMBIENT PRESSURE AS A FUNCTION OF THE ALTITUDE ABOVE SEA LEVEL OXYGEN ELECTRODE CONSTRUCTION OF THE SENSOR GWO General Design WHEN IS ELECTRODE MAINTENANCE REQUIRED ELECTRODE MAINTENANCE HOW TO OPERATE DECOMMISSIONING, RESHIPMENT AND DISPOSAL DECOMMISSIONING RESHIPMENT AND DISPOSAL SPECIFICATION... 13

3 T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 3 von 14 1 Safety 1.1 General note Read this document carefully and get used to the operation of the device before you use it. Keep this document within easy reach near the device for consulting in case of doubt. Mounting, start-up, operating, maintenance and removing from operation must be done by qualified, specially trained staff that have carefully read and understood this manual before starting any work. The manufacturer will assume no liability or warranty in case of usage for other purpose than the intended one, ignoring this manual, operating by unqualified staff as well as unauthorized modifications to the device. The manufacturer is not liable for any costs or damages incurred at the user or third parties because of the usage or application of this device, in particular in case of improper use of the device, misuse or malfunction of the connection or of the device. Bestimmungsgemäße Verwendung 1.2 Intended use Dissolved Oxygen-Measuring transducer converts an air oxygen-value in a 4-20 ma signal. 1.3 Qualified staff For operating, running, maintaining and commissioning the device, the staffs needs to have an adequate level of knowledge of both, the measuring and the meaning of measured values. This manual can be a valuable contribution. All instructions have to be well-understood and complied with. To be sure that there is no risk arising due to misinterpretation of measured values, the operator must have further knowledge in case of doubt - the user is liable for any harm/damage resulting from misinterpretation due to insufficient knowledge. The manufacturer will assume no liability or warranty in case of usage for other purpose than the intended one, ignoring this manual, operating by unqualified staff as well as unauthorized modifications to the device. 1.4 Safety signs and symbols Warning notices are marked in this manual as shown below: This symbol warns of imminent danger, death, serious injuries and significant damage to property at non-observance. This symbol warns of danger to living tissue and many materials, that can be damaged or destroy by exposure to that chemical! Achtung! Symbol warnt vor möglichen Gefahren oder schädlichen Situationen, die bei Nichtbeachtung Schäden am Gerät bzw. an der Umwelt hervorrufen. This symbol points out processes which can indirectly influence operation or provoke unforeseen reactions at non-observance. This symbol refers of using a eye protection, which protects the eyes from strong light, UV radiation, laser, chemicals, dust, splitter or weather conditions. This symbol refers of using protective gloves, which protects for mechanical-, thermal-, chemical-, biological- or electrical hazards.

4 T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 4 von Reasonably foreseeable misuse This device must not be used at potentially explosive areas! Do not use these products as safety or emergency stop devices or in any other application where failure of the product could result in personal injury or material damage. Failure to comply with these instructions could result in death or serious injury and material damage. This device must not be used at a patient for diagnostic or other medical purpose 1.6 Suitability This device has been designed and tested in accordance with the safety regulations for electronic devices. However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using the device During configuration Caution, acid! The electrode contains KOH KOH can cause severe chemical burns! If there was contact: To skin: Flush contacted area with large amounts of water for several minutes. To clothing: remove contaminated clothing. To eyes: Flush with large amounts of water for at least 15 minutes, obtain medical treatment. After swallowing: Give large volumes of water. DO NOT induce vomiting! Obtain medical treatment. General instructions and safety regulations for electric, light and heavy current plants, including domestic safety regulations (e.g. VDE), have to be observed. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those stated under "Specification". If the device is transported from a cold to a warm environment condensation may cause in a failure of the function. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up. This device only serves as supervision by the monitoring of essential or other for the customer important systems. It must not be used instead of compulsory approval monitoring devices and it is not designed for that purpose. If this device is used for the monitoring of such systems on its own, the manufacturer will not assume liability for damages whatsoever.

5 T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 5 von During normal operation 2 Product description Do not use this product as safety- or emergency- stop devices or in any other application where failure of the product could result in personal injury or material damage. Failure to comply with these instructions could result in death or serious injury and material damage. Whenever there may be a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked accordingly to avoid re-starting. Operator safety may be a risk if: -there is visible damage to the device -the device is not working as specified -the device has been stored under unsuitable conditions for a longer time In case of doubt, please return device to manufacturer for repair or maintenance. With costs. The conceptual design, particularly when connecting with other devices has to be done thoroughly. Internal connections of other devices e.g. from ground with protective earth may lead to prohibited voltage levels that could disturb the function, damage or even destruct the device or any connected equipment. Never run the device with a damaged or faulty power supply. 2.1 Scope of supply 1 x Dissolved Oxygen-Measuring transducer 2 x GWO 3600-L04_MU 3 x KOH x Mounting and operating manual OXY 3610 MP Dissolved Oxygen-Measuring transducer GWO 3600-L04_MU 2.2 Fehler! Verweisquelle konnte nicht gefunden werden. Production date code Options Measuring range Output signal Power supply Power supply Article number CE-mark see below: Disposal (s. Chapter 12.2) Please refer to manual. Manufacturer address CE mark: With the CE-Sign declares the manufacturer, that the Product is conform with the prevailing requirements of EU.

6 T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 6 von 14 3 Installation and Commissioning 3.1 General installation instructions Loosen screws (A). Remove cover (B) Wall mounting screws (C) are not included. The mounting-hole-diameter is 4.2 mm. See pattern drilling above (50 x 70 mm). Release the screw (a). Unplug the connector. Release the connection sealing, cable fitting (b), shim (c) and sealing-ring (d). Pull out the connection socket (f). Thread cable through parts (b), (c), (d) and into connector-housing (e). To mount the connection cable 2-, 3-, or 4- wire depending on type of device see chapter 4 Assignment of elbow-type plug, the angle plug screw has to be loosened. Reattach part (f) in (e). Reattach parts (d) and (c) and tighten (b). Plug in the connector with connection sealing and tighten (a). Connect the ph electrode at the BNC socket (I). Connect the sensor at the 5 pole diode Pattern drilling for wall mounting 50 x 70 mm

7 T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 7 von 14 4 Assignment of elbow-type plug Description of the contacts. 1 1= supply voltage +Uv 2= GND / signal 2-wire connection (4-20mA) 2 Uv A 3- wire connection (voltage) 3 1 V Uv 1 = signal + 3 = supply voltage +Uv (4) = supply voltage -Uv signal - 4- wire connection (voltage) V Uv 1= signal + 2= signal - 3= supply voltage +Uv (4)= supply voltage -Uv 5 Display functions During normal operation the oxygen content is displayed in the unit [mg/l]. By pressing the Key 2 (down) the temperature at the electrode can be shown, pressing key 3 (up) shows the state of the electrode in [%]. The state is calculated and stored during the calibration (see below). The displays of the state and the temperature of the electrode are marked by small arrows at the upper end of the display. After 5 seconds the transducer display switches automatically back to the oxygen value. display dissolved oxygen display electrode temperature display state of electrode

8 T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 8 von 14 6 Calibration of the oxygen measuring transducer Expose electrode to the ambient air, wait until the temperature has adapted to the ambient conditions. Parameter button 1 < 5sec Description CAL After approx. 10 sec s the transducer is calibrated or a referring error message will be shown (see below): Display Meaning Possible fault causes Remedy CFE.1 CFE.3 Temperature out of allowed range Electrode error Wrong signal: to low Temperature has to be within 5 to 40 C Temperature sensing defective Membrane dried out/electrode used up Calibrate again at correct temperature Check cable & connection, replace electrode if necessary Electrode maintenance / replace CFE.4 Wrong signal: to high Calibration surrounding not valid Check Calibration surrounding CFE.6 Instable signal Calibration surrounding not valid Check Calibration surrounding Oxygen electrodes are sensitive components. Please read the referring manual prior to use. During calibration the rating of the electrode is evaluated: 100% means perfect state, 40 % means e.g. sensor element has a weak output signal and needs to be replaced soon (p.r.t. sensor manual). The electrode rating can be displayed with key 3 (down). 7 Configuration of the device For the correct function of the device in its application, it has to be configured to meaningful settings. For doing this the cover has to be removed. Then the jumper J1 right above of key 2 has to be set. Select the value with button 2 (down) and 3 (up). To save the value confirm with button 1 (set). Concurrent the new parameter value will be invoked. The device will start again after saving the last parameter value. 7.1 Configuration procedure OFFSET OFFS Parameter button 1 < 5sec Value button 2 / 3 Description OFFS The offset is entered in mg/l. It can be used to compensate different effects, e.g. special approaching flow conditions of the application. Please refer to Appendix A.

9 T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 9 von Average ambient absolute pressure PABS Parameter button 1 Value button 2 / 3 Description PABS Necessary for a correct evaluation of the state of the electrode. If the application works for example at an altitude of 350 m above sea level, 980mbar is a correct setting. Please refer to chapter 10 Ambient pressure as a function of the altitude above sea level Display damping FiLT Parameter button 1 Value button 2 / 3 Description FiLT By this factor the speed of the display can be adopted to the application. This is intended to reduce instabilities in the measured value as they appear in some specialised applications Save the last parameter values Enter the desired value by using keys 2 (down) and 3 (up). Input range: (0: fast display, 15: slow display. Setting ex works: 10. Parameter button 1 Value button 2 / 3 Description The values will be saved. The device will start again (The display is shown 8888). Bring back the jumper j1 to the parking position after configuration (one contact connected, the second contact on air)! Then the device can be calibrated by key 1 and the configuration data is protected. 8 Error and System Messages Display Description Possible fault cause Remedy FE 1 measuring range exceeded Calibration is wrong Recalibrate the transducer. FE 2 Measuring values below measuring range Wrong signal Check the connections, cable and electrode. FE 7 System fault Error in device Disconnect from supply and reconnect. If error remains: return to manufacturer FE 9 Input signal is not valid Sensor not connected or cable defective Temperature out of range Check electrode, cable and connections Check temperature Segment test The transducer performs a display test for 2 seconds after power up. After that it will change to the display of the measuring. Sensor not connected Check sensor, cable and connections Input signal is not valid Permissible input range is exceeded Check if not a wrong sensor is connected. Replace sensor.

10 Offset [mg/l] T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 10 von 14 9 Offset settings for different applications The Offset setting is supposed to correct measuring deviations caused by different flow speed (v) around the electrode: -1E-15 For activated sludge basins and at flow speeds > 1m/s an offset setting of ca 0.6 should be entered. Fü ein -0,2-0,4-0,6 0,3 0,5 0,7 0,9 1,1 1,3 1,5 1,7 1,9 v [m/s] In between 0.3 and 1m/s the offset can be read from the diagram. 0.3m/s correspond to the speed that is achieved by stirring by hand. Flow speeds below 0.3 m/s are not sufficient to get a correct and reproducible measuring. Zw ge die Str ge 10 Ambient pressure as a function of the altitude above sea level Altitude über NN Pabs [mbar) Altitude über NN Pabs [mbar) [m] [m] Values between are to be interpolated. 11 Oxygen electrode 11.1 Construction of the sensor GWO General The oxygen electrode is an active electrode consisting of a silver cathode and a lead anode using potassium hydroxide (KOH) as an electrolyte. In case of oxygen being present it will be reduced at the silver cathode and the electrode produces electric current. In case of no oxygen being present not current will be produced. Both the silver cathode and the lead anode will be used up during oxygen measuring. The electrode ages. We, therefore, recommend regular maintenance of the electrode at monthly intervals (see 11.3 Electrode maintenance). Attention when working with electrolyte! The electrolyte is caustic! (strong base, KOH) Avoid contact to skin, protect Your eyes. Protective equipment is necessary! Irritates the skin and the mucous membranes, caustic. When contact with the eyes immediately clean with water at least for 15 minutes. Consult a doctor! The membrane is delicate if damaged, caustic electrolyte gets lost and the sensor shows wrong signal. Make it a rule to always store the electrode in a humid environment. - in the storage flask filled with water - in another container filled with water If electrode has not been used for some time, clean membrane with soft cloth and remove deposits, if any (algae, bacteria etc.).

11 T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 11 von Design Replenishing hole Electrode shaf Diaphragm Protective flask The electrode housing is made of ABS. With the exception of the electrode shaft all parts need to be maintained regularly and be replaced if necessary. o Protective flask: The protective flask is used to moisten the membrane. This prolongs service life of the electrode. The protective flask contains water. Attention! Use water only; never use potassium chloride (KCI); this is only required for storage of ph-electrode. o Diaphragm head: The membrane head is covered with a Teflon membrane. It will be filled with KOH electrolyte and screwed onto the electrode shaft (no air bubbles). Damages in the membrane, large air bubbles or air bubble rings in the membrane head will result in erroneous measurements. This may also be the reason for errors in the calibration. The membrane head is a spare part and can be ordered individually. o Filling hole: If the electrode is used at high temperatures or if it has been stored without its protective flask for a longer period of time, some electrolyte will be lost due to evaporation. Please refer to Refilling description below When is electrode maintenance required Attention when working with electrolyte! The electrolyte is caustic! (strong base, KOH) Avoid contact with skin and eyes. Protective equipment is necessary! Irritates skin and mucous membranes, caustic. In the case of contact with the eyes, immediately clean with clear water about 15 minutes Eye protective is necessary! Protective gloves are necesarry when filling the electrode! Do not touch the electrolyte with your bare hands. When contact with the skin immediately clean with water at least for 15 minutes.

12 T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 12 von 14 Unless it is used the electrode is to be stored in the protective flask filled with water or directly in a vessel filled with dechlorinated water. Residues of bacteria, fungi or algae must be removed prior to measuring using a soft paper towel. If a calibration can no longer be carried out or if the membrane is damaged the electrode needs maintenance. In the course of the time air bubbles may accumulated underneath the membrane. As long as they are small and do not affect the silver cathode the measurements will not be influenced. If, however, there is a large ring of air bubbles underneath the membrane covering the silver cathode, the electrode needs maintenance Electrode maintenance Prepare liquid absorbing paper towel. Remove membrane head and wipe off electrolyte solution with paper cloth. Do not touch electrolyte. If your skin has come into contact with electrolyte, rinse thoroughly with clear water. Clean silver cathode using sand paper (grain size 240). Do not shine silver cathode as its surface needs to be rough so that the electrolyte can be dispersed evenly. Remove any dust from grinding. If necessary, take out replenishing screw and top up with electrolyte till it is spilling out (e.g. using a disposable syringe). Put back replenishing screw. (Normally not necessary). Put new membrane head on the paper towel and fill up with electrolyte making sure there are no air bubbles. Put electrode into head vertically from above till the thread catches. Then take up head with paper towel and screw onto electrode from underneath. Electrolyte will be displaced from the membrane head, flowing over. Mop up excess electrolyte using paper towel. Turn electrode around and check for air bubbles. If there are no air bubbles or only tiny ones the maintenance has been completed. If there are large air bubbles the process has to be repeated. If O-ring has been damaged, it has to be replaced. Prior to its being calibrated again the electrode needs to be lying on a table for at least 3 hours. If electrode can no longer be calibrated although it has been properly maintained, the electrode needs to be returned to manufacturer for check-up and may have to be replaced How to Operate Store the oxygen electrode always in water, either in the protective flask or directly in a vessel filled with dechlorinated water. For storage in the protective flask first push the cap on the electrode second the o-ring and thereafter put the electrode in the flask and screw the cap down the flask. If the diaphragm has dried up the electrode must be soaked in water for approx. 2 h prior to measuring start-up. Then a calibration can be carried out without any problems. The electrode is only suitable for the devices OXY3610MP. Unsuitable devices may lead to the destruction of the measuring device and the oxygen electrode! For measuring remove the protective flask of the electrode. The electrode measurement is sensitive against shocks! By stirring of the electrode in the measured liquid be careful that the electrode does not hit the container. A vibration of the electrode has an effect to the measured value.

13 T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 13 von Decommissioning, Reshipment and disposal 12.1 Decommissioning Always disconnect the device before from its supply before decommission e.g. at fuse. Valid general safety requirements shall be observed. Please also make sure that connected Loads are disconnected also and are in a safe state Reshipment and disposal Alle Geräte, die an den Hersteller zurückgeliefert werden, müssen frei von Gefahrstoffen sein. Gefahrstoffe am Gehäuse können Personen oder die Umwelt gefährden. Verwenden Sie zur Rücksendung des Geräts, insbesondere wenn es sich um ein noch funktionierendes Gerät handelt, eine geeignete Transportverpackung. Legen Sie dem Gerät das ausgefüllte Rücksendeformular der GHM-Homepage unter bei. Das Gerät darf nicht über die Restmülltonne entsorgt werden. Senden sie das Gerät an uns zurück (ausreichend frankiert, siehe auch Hinweise oben). Wir entsorgen das Gerät sachgerecht und umweltschonend. 13 Specification Measuring range mg/l dissolved oxygen Output signal See 2.2 Fehler! Verweisquelle konnte nicht gefunden werden. Accuracy: (at nominal temperature) Approx. 2 years at nominal condition Display ±1.5% of measured value ±0.2 mg/l Output signal ±0.2 % FS Connection 4-20 ma (Zweileiter) Spannung (Dreileiter bzw. Vierleiter) Add. output signal ±0.2 % FS Auxiliary energy: (supply voltage) Uv = V DC (4-20mA) Uv = V DC (0-10V) or see 2.2 Fehler! Verweisquelle konnte nicht gefunden werden. Reverse voltage protection 50V permanent Perm. impedance (at 4-20mA) RA(Ohm) < (Uv - 12V) / 0.02A ) Example: for Uv = 18V RA < (18V - 12V) / 0.02A => RA < 300 Ohm Permissible load (at 0-...V) R L (Ohm) > 3000 Ohm Electrode Special oxygen electrode GWO3600MU (in scope of supply!) Electrode type active diaphragm type, with integrated NTC resistor Response time 95% in 10 sec., depending on temperature Operation life 3 years or more, depending on proper maintenance Operating pressure max. 3 bar Mounting diameter Ø 12,0 ±0,2 mm (also suitable for ½" glanding) Length approx. 220 mm (incl. anti-buckling glanding), Mounting length: approx. 110 mm Elektrodenanschluss 5 polige verschraubbare Diodenbuchse Temperature compensation automatic, C Electrode Special oxygen electrode GWO3600MU (in scope of supply!) Electrode type active diaphragm type, with integrated NTC resistor Response time 95% in 10 sec., depending on temperature Operation life 3 years or more, depending on proper maintenance Operating pressure max. 3 bar Mounting diameter Ø 12,0 ±0,2 mm (also suitable for ½" glanding) Length approx. 220 mm (incl. anti-buckling glanding), Mounting length: approx. 110 mm

14 T02.0.3X.6C-07 Mounting and operating manual OXY 3610 MP Page 14 von 14 Storage temperature Housing Dimensions Mounting Mounting distance Electric connection Directives / standards C (electrode C) ABS (IP65 - with the exception of electrode and temperature probe connection sockets) 82 x 80 x 55 mm (without elbow-type plug and sensor sockets) With fixing holes for wall mounting (in housing - accessible after cover has been removed). 50 x 70mm, max. shaft diameter of mounting screws is 4 mm Elbow-type plug conforming to DIN (IP65), Max. wired cross-section 1.5 mm², cable diameter 4.5 to 7 mm. The device confirm to following European directives: 2014/30/EU EMC directive 2011/65/EU RoHS Applied harmonized standards EN : 2013 Emission level: Class B EMI immunity according to table 1 Additional error: <1% GHM GROUP Greisinger GHM Messtechnik GmbH Hans-Sachs-Str Regenstauf GERMANY Phone Fax info@greisinger.de

T02.0.3X.6C-06 Page 1 of 6. Operating Manual for Dissolved Oxygen-Measuring transducer OXY 3610 MP

T02.0.3X.6C-06 Page 1 of 6. Operating Manual for Dissolved Oxygen-Measuring transducer OXY 3610 MP T02.0.3X.6C-06 Page 1 of 6 Specification: Measuring range: Operating Manual for Dissolved Oxygen-Measuring transducer OXY 3610 MP 0.00.. 25.00 mg/l dissolved oxygen Output signal: refer to type plate Accuracy:

More information

OXY 3690 MP. Measuring transducer for air oxygen

OXY 3690 MP. Measuring transducer for air oxygen T02.0.4X.6C-11 Measuring transducer for air oxygen OXY 3690 MP as of version V1.3 operating manual OXY 3690 MP - LO WEEE-Reg.-Nr. DE 93889386 GHM Messtechnik GmbH Standort Greisinger Hans-Sachs-Str. 26

More information

Please read these instructions carefully before use! Please consider the safety instructions! Please keep for future reference!

Please read these instructions carefully before use! Please consider the safety instructions! Please keep for future reference! Z11.0.1X.6C-18 Mounting and Operating Manual Atmospheric Oxygen Sensor Element GOEL 370, GOEL 381 Please read these instructions carefully before use! Please consider the safety instructions! Please keep

More information

SensoDirect Oxi200 Dissolved Oxygen Meter Operating instructions

SensoDirect Oxi200 Dissolved Oxygen Meter Operating instructions SensoDirect Oxi200 Dissolved Oxygen Meter Operating instructions mmhg hpa mg/l %O 2 % C C MEAS HOLD F SAL OK Cal hpa Meas Cal Hold Rcl Store On Off Menu GB Declaration of CE-Conformity The manufacturer:

More information

Instruction Manual Contact Pressure Vacuum Gauge

Instruction Manual Contact Pressure Vacuum Gauge MS10 Instruction Manual Contact Pressure Vacuum Gauge Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance

More information

Operating Instructions Temperature Measuring System MF420-5T-100

Operating Instructions Temperature Measuring System MF420-5T-100 Operating Instructions Temperature Measuring System Dnhqdhnjah MF420-5T-100 Dmsklakla Read before use! Observe all safety instructions! Keep for future reference! Fon +49/7221/64103 Fax +49/7221/17103

More information

Mounting instructions. Strain transducer SLB-700A. B 26.SLB700A.10 e

Mounting instructions. Strain transducer SLB-700A. B 26.SLB700A.10 e Mounting instructions Strain transducer SLB-700A B 26.SLB700A.10 e SLB700A 3 Contents Page Safety instructions.............................................. 4 1 Applications..................................................

More information

Instruction Manual Differential Pressure Transmitter

Instruction Manual Differential Pressure Transmitter DE61 Instruction Manual Differential Pressure Transmitter Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance

More information

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument OxyScan Graphic Operating Instructions UMS Micro-oxygen sensor 501 Microprocessor instrument Introduction Thank you for choosing the UMS Micro Oxygen Sensor 501 - a highly advanced product! Please read

More information

Instruction Manual Differential Pressure Transmitter

Instruction Manual Differential Pressure Transmitter DE13 Instruction Manual Differential Pressure Transmitter Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance

More information

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante810 Benchtop Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante810 benchtop dissolved oxygen

More information

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature Operating manual Dissolved Oxygen Depth Armature ba25302e05 04/2009 Copyright Weilheim 2009, WTW GmbH Reprinting - even as excerpts - is only allowed with the explicit written authorization of WTW GmbH,

More information

PRAHER PLC-MP AQUASTAR MANAUL 2009

PRAHER PLC-MP AQUASTAR MANAUL 2009 PRAHER PLC-MP AQUASTAR MANAUL 2009 1. Copyrights This Operating Manual contains copyright-protected information. All rights reserved to Praher Kunststofftechnik GmbH. This Operating Manual is designed

More information

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

Operating Instructions Melt Pressure Sensor

Operating Instructions Melt Pressure Sensor Operating Instructions Melt Pressure Sensor DA Certified to ISO 9001:2008 Please read this instruction manual carefully before installing the transducer 1 Contents: 1. Introduction 2. Operating range and

More information

QPL... Pressure Switches. Siemens Building Technologies HVAC Products

QPL... Pressure Switches. Siemens Building Technologies HVAC Products 7 221 Pressure Switches QPL... The pressure switches are used for monitoring gas or air pressures. When the pressure falls below the switching point, the associated electrical circuit will open or switch

More information

STERILIZABLE DISSOLVED OXYGEN SENSORS

STERILIZABLE DISSOLVED OXYGEN SENSORS STERILIZABLE DISSOLVED OXYGEN SENSORS USER MANUAL Table of Contents 1. Introduction 2. Operation 3. Calibration 4. Electrode Verification 5. Mounting 6. Maintenance 7. Storage 8. Instructions for Exchange

More information

Dextens Thermo Conductivity Sensors. Type 52101, 54101, 55101

Dextens Thermo Conductivity Sensors. Type 52101, 54101, 55101 Dextens Thermo Conductivity Sensors Type 52101, 54101, 55101 Operating Manual May 2012 Table of Content 1 GENERAL INFORMATION...4 1.1 DISCLAIMER...4 1.2 SYMBOLS AND CONVENTIONS...4 1.3 SERVICES AND REPAIRS...4

More information

* * Data Sheet and Operating Manual. Digital manometer ME01 ## # 87 # HL S#### Dust explosion protection Zone 22.

* * Data Sheet and Operating Manual. Digital manometer ME01 ## # 87 # HL S#### Dust explosion protection Zone 22. *09005087* DB_BA_EN_ME01_S Rev.A 12/16 *09005087* d e v e l o p i n g s o l u t i o n s Data Sheet and Operating Manual ME01 Digital manometer Table of Contents ME01 ## # 87 # HL S#### Dust explosion protection

More information

Technical Data Sheet MF010-O-LC

Technical Data Sheet MF010-O-LC Technical Data Sheet MF010-O-LC - 1 - 1. Properties The oxygen measuring system MF010-O-LC determines the oxygen content in gas mixtures up to a temperature of 250 C. It is particularly suitable for the

More information

BI-680 Online Dissolved Oxygen Controller Instruction Manual

BI-680 Online Dissolved Oxygen Controller Instruction Manual BI-680 Online Dissolved Oxygen Controller Instruction Manual BANTE INSTRUMENTS CO., LTD BI-680 Online Dissolved Oxygen Controller 1 Introduction Thank you for selecting the BI-680 online dissolved oxygen

More information

OXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES

OXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES PURAPOOL OXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES 1. PARTS IDENTIFICATION LIST 1. 5. 2. 4. 3. 6. 6. Figure 1.1 1. OXYGEN Minerale Power pack with timer option (x1) 2.

More information

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073 Operating instructions Safety Rope Emergency Stop Switches UK ZB0052 / ZB0053 ZB0072 / ZB0073 7390878 / 02 03 / 2011 Contents 1 Safety instructions...3 2 Installation / set-up...4 2.1 Applications...4

More information

DTG - LCD Digital Temperature Gauge USER MANUAL

DTG - LCD Digital Temperature Gauge USER MANUAL DTG - LCD USER MANUAL Page 1 of 7 USERS GUIDE page 1. Description 3 2. Function 3 3. Safety Instruction 3 3.1. Safety Conventions 3 3.2. Proper Use 3 4. Installation 4 4.1. Unpacking 4 4.2. Storage 4 4.3.

More information

Instruction Manual Contact Pressure Gauge

Instruction Manual Contact Pressure Gauge MS11 Instruction Manual Contact Pressure Gauge Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance 7.

More information

3G Operating instructions. Digital Purge Gas Valve 3G Type: Document no.: D0001 Version: 01. July 2011/Rev.

3G Operating instructions. Digital Purge Gas Valve 3G Type: Document no.: D0001 Version: 01. July 2011/Rev. 3G Operating instructions Type: 03-5110-00.. Document no.: 03-5110-7D0001 Version: 01. July 2011/Rev. 0 Operating Instructions Type: 03-5110-00.. Document no.: 03-5110-7D0001 Version: 1 July 2011 / Rev.

More information

HBLT-A1 LIQUID LEVEL SENSOR

HBLT-A1 LIQUID LEVEL SENSOR Instruction manual HBLT-A1 LIQUID LEVEL SENSOR for measuring liquid level in refrigerant vessels Instruction manual HBLT - A1 - Liquid level Transmitter (HBLT-A1-008-UK) 1 / 16 Contents Safety instructions...

More information

JUMO dtrans p30 Pressure Transmitter

JUMO dtrans p30 Pressure Transmitter 609 Fulda, Germany Postal address: 605 Fulda, Germany Phone: +49 661 600-0 Fax: +49 661 600-607 Phone: +44 179 6 55 Data Sheet 40466 Page 1/6 JUMO dtrans p0 Pressure Transmitter General application Pressure

More information

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator Self-operated Pressure Regulators Type 2357-31 Pressure Build-up Regulator with safety function and integrated excess pressure valve Type 2357-31 with non-return unit at port C Ports A and B with soldering

More information

Operating Manual VSA100A. tina29e1 ( )

Operating Manual VSA100A. tina29e1 ( ) Operating Manual Incl. EU Declaration of Conformity Vacuum Switch VSA100A tina29e1 (2017-05) 1 Product Identification In all communications with INFICON, please specify the information on the product nameplate.

More information

P499 Heavy Duty Pressure Transducer

P499 Heavy Duty Pressure Transducer P499 Heavy Duty Pressure Transducer Product Bulletin Code No. LIT-??? E Issued 11 2006 The P499 Series is a new global Pressure Transducer with an excellent price performance ratio. The P499 exceeds the

More information

Digital Melting Point Apparatus

Digital Melting Point Apparatus Digital Melting Point Apparatus Heating Plateau Ramping Start/Stop Plateau set Ramp stop Hold User Guide Version 1.1 Heating Viewing tube Sample Chamber IEC power inlet socket Power on/off Temperature

More information

Safety. Operating instructions Solenoid valve for gas VG 6 VG 15/10 DANGER. Contents WARNING CAUTION. Changes to edition 09.14

Safety. Operating instructions Solenoid valve for gas VG 6 VG 15/10 DANGER. Contents WARNING CAUTION. Changes to edition 09.14 15 Elster GmbH Edition 7.15 Translation from the German 519 D F NL I E DK S N P GR TR CZ PL RUS H www.docuthek.com Operating instructions Solenoid valve for gas VG VG 15/1 Contents Solenoid valve for gas

More information

OWNER S MANUAL NMB SERIES

OWNER S MANUAL NMB SERIES OWNER S MANUAL NMB SERIES NMB Manual V2.0.doc 1. PARTS IDENTIFICATION LIST 1. 5. 2. 4. 3. 6(a). 6(b). Figure 1.1 1. Powerpack with timer option (x1) 2. Cell (x1) 3. Cell Housing (x1) 4. O-ring (x1) 5.

More information

Sensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual

Sensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual Sensor for Air Bubble Detection at Liquid Filled Tubes SONOCHECK Type ABD06.xx Operating Manual Manufacturer: Model: Type: SONOTEC Ultraschallsensorik Halle GmbH Air Bubble Detector ABD06.xx SONOTEC Ultraschallsensorik

More information

OxiQuant S. Operating Instructions

OxiQuant S. Operating Instructions ENVITEC -WISMAR GMBH Environmental & Medical Sensor Technology OxiQuant S Operating Instructions P/N: 46-00-0021 EnviteC-Wismar GmbH Alter Holzhafen 18, D-23966 Wismar, Germany Tel.: +49-(0)-3841-360-1

More information

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual BANTE INSTRUMENTS CO., LTD DOscan10 Pocket Dissolved Oxygen Tester 1 Thank you for selecting the DOscan10 pocket dissolved oxygen tester. This

More information

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Bante820 Portable Dissolved Oxygen Meter Instruction Manual Bante820 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante820 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante820 portable dissolved oxygen

More information

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter INSTRUCTION MANUAL Loop Powered Dissolved Oxygen Transmitter Contents Contents...2 Introduction...3 Range and Display Settings...4 Installation...5 Electrical Installation...6 Calibration...7 EasyCal...8

More information

PLAQUE MFM18240 UNDERWATER LIGHT. Operating Instructions

PLAQUE MFM18240 UNDERWATER LIGHT. Operating Instructions PLAQUE MFM18240 UNDERWATER LIGHT Operating Instructions CONTENTS Preface... 3 Features... 3 Precautions... 3 Controls and Connectors... 4 Underwater Light PLAQUE MFM18240... 4 Power Supply Unit MPS061000

More information

Oxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications

Oxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications Technical Information Oxymax W COS41 Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications Application The continuous measurement of the dissolved

More information

PLAQUE MFM0680 UNDERWATER LIGHT. Operating Instructions

PLAQUE MFM0680 UNDERWATER LIGHT. Operating Instructions PLAQUE MFM0680 UNDERWATER LIGHT Operating Instructions CONTENTS Preface... 3 Features... 3 Precautions... 3 Controls and Connectors... 4 Underwater Light PLAQUE MFM0680... 4 Power Supply Unit MPS021000

More information

User Manual PTW-DensiX. D / Di/Zi

User Manual PTW-DensiX. D / Di/Zi User Manual PTW-DensiX D148.131.0/3 2005-05 Di/Zi General Information General Information The product bears the CE-mark "CE-0124" in accordance with the Council Directive 93/42/EEC about Medical Devices

More information

Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar

Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar EN - english Instructions for installation and operation Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar Dear customer, Thank you for deciding in favour of the METPOINT

More information

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17 Operation Manual O2 Quickstick Oxygen Analyzer 08.17 If you have any questions on this equipment please contact Technical Support at: Nuvair 1600 Beacon Place Oxnard, CA 93033 Phone: 805-815-4044 FAX:

More information

Air Sensor. FCS Ex. Manual. AQ Elteknik AB

Air Sensor. FCS Ex. Manual. AQ Elteknik AB Air Sensor FCS Ex Manual AQ Elteknik AB Air Sensor Model FCS Ex Manual version 2.0 Oct 2010 AQ Elteknik AB AQ Elteknik AB Air Sensor Manual 2 Table of contents 1. Manufacturer information... 4 Manufacture

More information

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING Please Note: Ranges above 500mbar are designed and manufactured in accordance with sound engineering

More information

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual OxyCheq Expedition-X Oxygen Analyzer Operator s Manual OxyCheq 3528 Russell Road Marianna, Florida 32446 USA Tel: 850.482.0385 Email: oxy@oxycheq.com Web: http://oxycheq.com CONTENTS 1.0 Introduction..

More information

Precision pressure sensor Basic version Model CPT6020

Precision pressure sensor Basic version Model CPT6020 Calibration Precision pressure sensor Basic version Model CPT6020 WIKA data sheet CT 25.13 Applications Calibration technology High-accuracy pressure monitoring Pressure sensing in critical applications

More information

Operating Instructions. Ball valve fitting according to ZB For pressure transmitter VEGABAR 82. Document ID: 50027

Operating Instructions. Ball valve fitting according to ZB For pressure transmitter VEGABAR 82. Document ID: 50027 Operating Instructions Ball valve fitting according to ZB 2553 For pressure transmitter VEGABAR 82 Document ID: 50027 Contents Contents 1 About this document 1.1 Function... 3 1.2 Target group... 3 1.3

More information

Instruction Manual BA90 Precision Barometer

Instruction Manual BA90 Precision Barometer Instruction Manual BA90 Precision Barometer halstrup-walcher GmbH Stegener Straße 10 D-79199 Kirchzarten Germany Phone: +49 (0) 76 61/39 63 0 Fax: +49 (0) 76 61/39 63 99 E-Mail: info@halstrup-walcher.com

More information

Transmitter CS 21 Operation Manual

Transmitter CS 21 Operation Manual Transmitter CS 21 Operation Manual 1194 Oak Valley Drive, Suite 20, Ann Arbor, MI 48108 800-959-0573 734-769-1888 Content Page For your Safety 3 General Description 3 Detection Principle 4 Operation 4

More information

E8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm)

E8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm) Pressure Sensor (Stainless Steel Diaphragm) CSM DS_E_3_1 Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications Incorporates double diaphragms consisting of SUS316L stainless

More information

OSAT Series. Intrinsically Safe Infrared Temperature Sensor Operator's Guide

OSAT Series. Intrinsically Safe Infrared Temperature Sensor Operator's Guide OSAT Series Intrinsically Safe Infrared Temperature Sensor Operator's Guide Issue L Jan 2017 Introduction OSAT Series intrinsically safe non-contact infrared temperature sensors measure the temperature

More information

Pressure maintenance valve SDV-P / SDV-P-E

Pressure maintenance valve SDV-P / SDV-P-E Translation of the original manual Pressure maintenance valve SDV-P / SDV-P-E Assembly and Operating Manual Superior Clamping and Gripping Imprint Imprint Copyright: This manual remains the copyrighted

More information

FPD20 Series. Gear wheel flow sensor

FPD20 Series. Gear wheel flow sensor . FPD20 Series Gear wheel flow sensor Series FPD20-2 - M-5455/0315 Series FPD20 Table of contents page 0 About this operating manual... 4 1 Device description... 5 1.1 Intended use... 5 1.2 Exclusion of

More information

Operation Manual. O2 Quickstick. Oxygen Analyzer

Operation Manual. O2 Quickstick. Oxygen Analyzer Operation Manual O2 Quickstick Oxygen Analyzer 01.11 www.nuvair.com If you have any questions on this equipment please contact Technical Support at: Nuvair 2949 West 5 th St. Oxnard, CA 93030 Phone: 805-815-4044

More information

Installation and Maintenance Instruction Manual

Installation and Maintenance Instruction Manual Installation and Maintenance Instruction Manual Bi-metal Thermometer Model A and E in a configuration - pursuant to ASME B40.200: ##=E#=### or ##=ERT#=### - pursuant to EN 13190: ###=A#=### or ##=ART#=###

More information

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05 IDL0 Battery Powered Precision Digital Gauge for Leak Testing Stainless Steel Sensor class 0.05 Nominal pressure from 0 00 mbar up to 0... 00 bar Special characteristics modular sensor concept data logger

More information

Transmitter CS 21 Operation Manual

Transmitter CS 21 Operation Manual Transmitter CS 21 Operation Manual Content Page For your Safety 3 General Description 3 Detection Principle 4 Operational Notes 4 Design 4 Mounting Position of CS21 5 Mounting 6 Installation of Electrical

More information

Pressure Transmitter P41

Pressure Transmitter P41 SENSORS FOR AND BIOPHARMA. Product Information P41 Pressure Transmitter P41 Application/Specified usage Pressure measurement in pipes and tanks For process temperatures up to 250 C (428 F) with optional

More information

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE GENERAL INFORMATION This family of dissolved oxygen electrodes is of the steam sterilizable polarographic type, designed to be interchangeable with NBS

More information

PRODUCT SHEET. Order probe only as RXPROBE02

PRODUCT SHEET. Order probe only as RXPROBE02 SS69L DISSOLVED OXYGEN PROBE TRANSDUCER Order probe only as RXPROBE02 Order interface only as BSL-TCI16 SS69L Components The SS69L transducer measures dissolved oxygen. The SS69L includes a dissolved oxygen

More information

VS18/26 with PROFINET and EtherNet/IP Interface. ATEX Installation Instructions

VS18/26 with PROFINET and EtherNet/IP Interface. ATEX Installation Instructions VS18/26 with PROFINET and EtherNet/IP Interface ATEX Installation Instructions INDEX 1. INTENDED USAGE 3 2. OPERATING MANUAL ATEX 4 2.1 General conditions 4 2.2 Installation 5 2.3 Operating 5 2.4 Failures

More information

Datasheet: K-30 ASCII Sensor

Datasheet: K-30 ASCII Sensor Datasheet: K-30 ASCII Sensor The K30 ASCII sensor is a low cost, infrared and maintenance free transmitter module intended to be built into different host devices that require CO2 monitoring data. The

More information

ANALOX O2 PORTABLE OPERATORS MANUAL

ANALOX O2 PORTABLE OPERATORS MANUAL ANALOX O2 PORTABLE OPERATORS MANUAL Analox Sensor Technology Ltd 15 Ellerbeck Court Stokesley Business Park Stokesley N. Yorks TS9 5JY Tel: +44 (0)1642 711400 Fax: +44 (0)1642 713900 Email: info@analox.net

More information

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents AQUARIUS Operating Instructions for the Type 70 Electrolytic Gas Soldering / Welding Unit Table of Contents 1. Important General Remarks 2 2. Safety Instructions 2 3. Description of the Soldering Unit

More information

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05 Battery Powered Stainless Steel Sensor class 0.05 Nominal pressure from 0 00 mbar up to 0... 00 bar Special characteristics modular sensor concept data logger graphic display stainless steel housing Ø

More information

EC214 EC215 - EC215R Bench Conductivity Meters

EC214 EC215 - EC215R Bench Conductivity Meters Instruction Manual EC214 EC215 - EC215R Bench Conductivity Meters http://www.hannainst.com These Instruments are in Compliance with the CE Directives Dear Customer, Thank you for choosing a Hanna Instruments

More information

MASS-STREAM TM Instruction Manual. D-6200 Analog Mass Flow Meters / Controllers

MASS-STREAM TM Instruction Manual. D-6200 Analog Mass Flow Meters / Controllers MASS-STREAM TM Instruction Manual D-6200 Analog Mass Flow Meters / Controllers Doc. No.: 9.17.112A Date: 17-03-2017 ATTENTION Before installing and operating the instrument it is strongly recommended that

More information

DK48 - DK800. Quick Start. Variable area flowmeter

DK48 - DK800. Quick Start. Variable area flowmeter Quick Start Variable area flowmeter General safety notes You can find the newest and/or additional information in the handbook, the datasheet, special manuals, certificates and the download center at www.krohne.com.

More information

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Differential Pressure Regulators (opening) Type Type 42-25

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Differential Pressure Regulators (opening) Type Type 42-25 Self-operated Pressure Regulators Differential Pressure Regulators (opening) Type 42-20 Type 42-25 Type 42-20 Differential Pressure Regulator Type 42-25 Differential Pressure Regulator Mounting and Operating

More information

Type Operating Instructions. 2/2-way solenoid valve 2/2-Wege Magnetventil Electrovanne 2/2 voies.

Type Operating Instructions. 2/2-way solenoid valve 2/2-Wege Magnetventil Electrovanne 2/2 voies. 2/2-way solenoid valve 2/2-Wege Magnetventil Electrovanne 2/2 voies We reserve the right to make technical changes without notice. Technische Änderungen vorbehalten. Sous réserve de modifications techniques.

More information

Proportional pressure regulators VPPE

Proportional pressure regulators VPPE Proportional pressure regulators VPPE Proportional pressure regulators VPPE Product range overview Function Version Pneumatic connection Proportional pressure regulator Nominal size for pressurisation/

More information

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50 Rev.02 (2014.06) OL Page 1 of 8 Table of contents I. STRUCTURE OF THE MANUAL / CLARIFICATION... 3 II. SAFETY AND HEALTH CONCERNS... 4 1 INTRODUCTION... 5 1.1 Use

More information

OPTIBAR P 1010 C Technical Datasheet

OPTIBAR P 1010 C Technical Datasheet Technical Datasheet Pressure transmitter with recessed diaphragm for general applications Wide variety thanks to modular design Measuring ranges up to 250 bar / 3750 psi High overload and temperature stability

More information

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

DM01. Please visit our website:  Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05 DM0 Battery Powered Stainless Steel Sensor class 0.05 Nominal pressure from 0 00 mbar up to 0... 00 bar Special characteristics modular sensor concept data logger graphic display stainless steel housing

More information

x act i Precision Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor accuracy according to IEC 60770: 0.

x act i Precision Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor accuracy according to IEC 60770: 0. Precision Pressure Transmitter for Food ndustry, Pharmacy and Biotechnology Stainless Steel Sensor accuracy according to EC 60770: 0. % FSO Nominal pressure from 0... 400 mbar up to 0... 40 bar Output

More information

PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100

PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100 MPS-1100 User Manual Mega Pool Saver Ltd PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100 For further up to date instructions on how to install Mega Pool Saver MPS 1100, please

More information

Hurlcon KX Siesta Above Ground Salt Water Chlorinator by AstralPool Australia

Hurlcon KX Siesta Above Ground Salt Water Chlorinator by AstralPool Australia Hurlcon KX Siesta Above Ground Salt Water Chlorinator by AstralPool Australia OPERATING INSTRUCTIONS AstralPool Australia Pty. Limited. A.B.N. 97 007 284 504 Melbourne: Ph: (03) 8796 8600 Fax: (03) 8796

More information

Type BBS-03, BBS-05, BBS-06, BBS-25

Type BBS-03, BBS-05, BBS-06, BBS-25 Type BBS-03, BBS-05, BBS-06, BBS-25 Sterile connection elements Sterile Verbindungselemente Raccords union stériles Operating Instructions Bedienungsanleitung Manuel d utilisation 1. THE OPERATING INSTRUCTIONS

More information

Instruction Manual AD 1000 / BA 1000 Absolute Pressure Sensor

Instruction Manual AD 1000 / BA 1000 Absolute Pressure Sensor Instruction Manual AD 1000 / BA 1000 Absolute Pressure Sensor halstrup-walcher GmbH Stegener Straße 10 D-79199 Kirchzarten Phone: +49 (0) 76 61/39 63 0 Fax: +49 (0) 76 61/39 63 99 E-Mail: info@halstrup-walcher.com

More information

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter. Instruction Manual HI 9142 Portable Waterproof Dissolved Oxygen Meter www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully

More information

Operation Manual ARTICEL SunSun CUF SunSun CUF SunSun CUF-6000

Operation Manual ARTICEL SunSun CUF SunSun CUF SunSun CUF-6000 Operation Manual ARTICEL 50945 SunSun CUF-2500 50946 SunSun CUF-5000 50947 SunSun CUF-6000 Read and follow the operating instructions and safety information before using for the first time. Technical changes

More information

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. Type: ETd 04. class 0.05

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor.  Type: ETd 04. class 0.05 Battery Powered Stainless Steel Sensor class 0.05 Nominal pressure from 0 100 mbar up to 0... 400 bar Special characteristics modular sensor concept data logger graphic display stainless steel housing

More information

Transferpette -8/-12

Transferpette -8/-12 Transferpette -8/-12 Testing Instructions (SOP) May 2009 1. Introduction The standard DIN EN ISO 8655 describes both the design and the testing of the piston operated pipette Transferpette 8/ 12. The following

More information

Pressure Measurement Single-range transmitters for general applications

Pressure Measurement Single-range transmitters for general applications Siemens A 207 Overview Application The SITRANS P Compact pressure transmitter is designed for the special requirements of the food, pharmaceutical and biotechnology industries. The use of high-grade materials

More information

testo Leakage detector for refrigerants Instruction manual

testo Leakage detector for refrigerants Instruction manual testo 316-3 Leakage detector for refrigerants Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Safety and the environment... 4 2.1. About this document... 4 2.2. Ensure safety... 4 2.3. Protecting

More information

Instruction Manual EMA 84 Digital Pressure Gauge

Instruction Manual EMA 84 Digital Pressure Gauge Instruction Manual EMA 84 Digital Pressure Gauge halstrup-walcher GmbH Stegener Straße 10 D-79199 Kirchzarten Phone: +49 (0) 76 61/39 63 0 Fax: +49 (0) 76 61/39 63 99 E-Mail: info@halstrup-walcher.com

More information

LTB013EN. Operating Instruction Trimod Besta Level Switch types XA 5, XB 5 for use in potentially explosive atmospheres acc.

LTB013EN. Operating Instruction Trimod Besta Level Switch types XA 5, XB 5 for use in potentially explosive atmospheres acc. LTB013EN Operating Instruction Trimod Besta Level Switch types XA 5, XB 5 for use in potentially explosive atmospheres acc. to IECEx scheme Subject to technical modification Bachofen AG Ackerstrasse 42

More information

KERN EG/EW Version /02

KERN EG/EW Version /02 E KERN EG/EW Version 1.5 07/02 Operating Instructions Electronic Precision Balances Contents 1 TECHNICAL DATA... 20 2 UNPACKING AND STANDARD ACCESSORIES... 23 3 SETTING UP THE BALANCE... 23 4 EXTERNAL

More information

RAM 4021 Operation Manual

RAM 4021 Operation Manual RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6

More information

L3 Pressure and Level transmitter

L3 Pressure and Level transmitter SENSORS FOR AND BIOPHARMA. Product Information L3 L3 Pressure and Level transmitter Range of applications Hydrostatic Level measurement in dynamic temperature environments Pressure measurement in pipes

More information

Type 3709 Pneumatic Lock-up Valve. Translation of original instructions. Mounting and Operating Instructions EB 8391 EN

Type 3709 Pneumatic Lock-up Valve. Translation of original instructions. Mounting and Operating Instructions EB 8391 EN Type 3709 Pneumatic Lock-up Valve Translation of original instructions Mounting and Operating Instructions EB 8391 EN Edition July 2017 Note on these mounting and operating instructions These mounting

More information

Operating Manual BCP-O2. BCP X-large Aluminum IP65. BCP regular with PA 6 housing. Rev

Operating Manual BCP-O2. BCP X-large Aluminum IP65. BCP regular with PA 6 housing. Rev Operating Manual BCP-O2 BCP X-large Aluminum IP65 BCP regular with PA 6 housing Rev.140310001 Contents 1 ABOUT THIS DOCUMENT... 2 1.1 Function... 2 1.2 Target group... 2 1.3 Symbols used... 2 2 FOR YOUR

More information

Contents. Operating instructions. Pressure switch, heavy-duty version Model PSM-520

Contents. Operating instructions. Pressure switch, heavy-duty version Model PSM-520 Operating instructions Contents EN Pressure switch, heavy-duty version Model PSM-520 1. General information 2. Design and function 3. Safety 4. Transport, packaging and storage 5. Commissioning, operation

More information

CONUS MSR0680 UNDERWATER LIGHT. Registered Community Design. Operating Instructions

CONUS MSR0680 UNDERWATER LIGHT. Registered Community Design. Operating Instructions CONUS MSR0680 UNDERWATER LIGHT Registered Community Design Operating Instructions CONTENTS Preface... 3 Features... 3 Precautions... 3 Controls and Connectors... 4 Underwater Light CONUS MSR0680... 4 Power

More information

Solenoid gas valve. The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners

Solenoid gas valve. The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners 7 634 VGS1... with AGA67 connector VGS2... with AGA67 connector Solenoid gas valve VGS The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners The VGS and

More information

Air Sensor. SAC Ex. Manual ATEX. AQ Elteknik AB

Air Sensor. SAC Ex. Manual ATEX. AQ Elteknik AB Air Sensor SAC Ex Manual AQ Elteknik AB ATEX Air Sensor SAC Ex ATEX Certified model Manual version 2.1 April 2014 AQ Elteknik AB 2 Table of contents 1. Manufacturer information... 4 Manufacture Declaration

More information

x act i Precision- Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor

x act i Precision- Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor Precision- Pressure Transmitter for Food ndustry, Pharmacy and Biotechnology Stainless Steel Sensor accuracy according to EC 60770: 0. % FSO Nominal pressure from 0... 400 mbar up to 0... 40 bar Output

More information