FUME CUPBOARD PERFORMANCE CHECK
|
|
- Kelley West
- 6 years ago
- Views:
Transcription
1 WS2 FUME CUPBOARD PERFORMANCE CHECK This exercise is adapted from Appendices E & F of Australian Standard Safety in laboratories: Fume cupboards. 1. PURPOSE This exercise determines how well a fume cupboard is meeting Australian Standard specifications. 2. SUMMARY The face velocity of air through the cupboard window is measured with a vane anemometer, while the presence of turbulence is checked using a Dragar smoke tube. The three operational requirements to be tested are: an average face velocity of m/s for the fully opened window, with a maximum variation of 20% from average a minimum turnover of 5 cupboard volumes per minute no turbulence causing air to escape into the laboratory 3. REAGENTS & EQUIPMENT 3.1 Tape measure 3.2 Dragar smoke tube 3.3 Vane anemometer 4. PROCEDURE 4A. Face velocity measurements 4.1 Turn the fume cupboard on and allow one minute to stabilise. 4.2 Lift the window to fully opened. 4.3 Measure the dimensions of the opening. 4.4 Measure the face velocity at each point shown on the Results Sheet. The anemometer will take around 15 seconds to read correctly, and will fluctuate, so take an average reading over 10 seconds. 4B. Volumetric capacity 4.5 Measure the internal dimensions of the fume cupboard. Assume it is a regular shape do not attempt to compensate for small gaps/ledges etc. 4.6 Lower the window to fully closed. 4.7 Measure the dimensions of the opening. 4.8 Measure the face velocity at each point shown on the Results Sheet. 4C. Smoke test Record all results as in the table below. Good Satisfactory Dangerous Smoke travels straight to exhaust without reverse flows or eddies Reverse flows or eddies may be present but not to the extent that smoke escapes from the fume cupboard Reverse flows or eddies can escape from the fume cupboard
2 4.9 Lift the window to fully opened Release smoke around the perimeter of the opening Release smoke along the floor of the fume cupboard, starting at the front Release smoke around the internal walls Release smoke near any items in the fume cupboard Lower the window to half-way Release smoke around the perimeter of the opening Lower the window to fully closed Release smoke around the perimeter of the opening. 4D. Location of fume cupboards 4.18 Check the siting of the fume cupboard, based on the requirements of AS Section 4.1, excerpted in Appendix REQUIREMENTS Complete the Results Sheet provided. 6. REPORT Calculations 4A. Face velocity measurements calculate the average face velocity for the seven measurements with window fully open calculate the maximum variation in the individual measurements from the average 4B. Volumetric capacity calculate the area of the opening with the window fully open calculate the area of the opening with the window closed to minimum calculate the average face velocity based on the three measurements made with the window closed to minimum calculate the volumetric capacity (in m 3 /min) with the window fully open and fully closed calculate the internal volume of the fume cupboard calculate the number of cupboard air volume changes per minute fully open and fully closed Question Are the operating requirements met? Tabulate the results, requirements and a PASS/FAIL for each. WS2.2
3 WS2. Fume Cupboard Performance Check RESULTS SHEET Location of fume cupboard Serial or identification number Date tested Items in fume cupboard when tested 4a. Face velocity Window fully open Height Width WS2.3
4 4B. Volumetric capacity Internal dimensions of fume cupboard Height (m) Width (m) Depth (m) Window fully closed Height 4C. Smoke test Fully open Floor Walls Around items Half open Closed WS2.4
5 4D. Location of fume cupboards AS Section 4.1 Clause Satisfactory (Y/N)? Height of working surface Traffic routes Opposing bench tops Opposing walls Other fume cupboards Architectural obstacles Comments WS2.5
Mott Manufacturing Limited
Mott Manufacturing Limited 452 Hardy Road, PO Box 1120, Brantford, Ontario, Canada, N3T 5T3. Telephone: 519-752-7825 Fax: 519-752-2895 Web Site: www.mott.ca Fume Hood Performance Test Test Date: January
More informationMott Manufacturing Limited
Mott Manufacturing Limited 452 Hardy Road, Brantford, Ontario, Canada, N3T 5L8. Telephone: 519-752-7825 Fax: 519-752-2895 Web Site: www.mott.ca Fume Hood Performance Test Test Date: 9-Sep-08 Model Number:
More informationStatesville, NC. Results of AM ASHRAE 110 Laboratory Fume Hood Performance Evaluations Venturi Hood
Statesville, NC Results of AM ASHRAE 11 Laboratory Fume Hood Performance Evaluations Venturi Hood January, 217 Prepared By: Exposure Control Technologies Inc. 231-C East Johnson Street Cary, NC 27513 Phone:
More informationLEV Inspection & Test Report
LEV Inspection & Test Report CLIENT: SITE: XXXXXXX LTD XXXXXXX Report Reference: Hinge Room-LEV System Identification: BeCu ESTA Survey Conducted By: Name: Position: Signed: XXXX Environmental Consultant
More informationFumehood Performance Test Report
Fumehood Performance Test Report 6' Protector ECHO IPA Modified NIH Test Prepared For Labconco Corporation Prepared By Report Date: 02-12-2015 Project No.: 142327 AccuTec Services, Inc*320 NW Capital Dr*Lee's
More information"LABORATORY SAFETY SERIES"
PRESENTER'S GUIDE "LABORATORY HOODS" Part of the "LABORATORY SAFETY SERIES" Quality Safety and Health Products, for Today...and Tomorrow OUTLINE OF MAJOR PROGRAM POINTS OUTLINE OF MAJOR PROGRAM POINTS
More informationASTM E1886 and ASTM E1996 TEST REPORT. Report No.: E Rendered to: 3M COMPANY St. Paul, Minnesota 55144
ASTM E1886 and ASTM E1996 TEST REPORT Report No.: E6587.01 201 44 Rendered to: 3M COMPANY St. Paul, Minnesota 55144 PRODUCT TYPE: Safety and Security Window Film SERIES/MODEL: 3M Safety and Security Film
More informationThorough Examination and Test of Local Exhaust Ventilation Systems to Regulation 9 of the COSHH Regulations 2002 (as amended)
Thorough Examination and Test of Local Exhaust Ventilation Systems to Regulation 9 of the COSHH Regulations 2002 (as amended) Test of LEV Systems 6 Monthly Test Report Ref: LEV16012014 Date: 16 th January
More informationLABORATORY SAFETY SERIES: Laboratory Hoods
ERI Safety Videos Videos for Safety Meetings 2956 LABORATORY SAFETY SERIES: Laboratory Hoods Leader s Guide Marcom Group Ltd. 1 INTRODUCTION TO THE PROGRAM Structure and Organization Information in this
More informationShuttlE. Schools Badminton
Schools Badminton LESSON PLANS Learn to Win Schools Badminton LESSON PLANS Learn to Win Badminton World Federation 2011. First published November 2011. Overview 1. Introduction Learn to Win Before starting
More informationCustomer Responsibilities
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationIH Report: SUFMO 08/09-CC21 1/27/2009 Stanford University Gas Cabinet Installation and Performance Criteria
Stanford University Gas Cabinet Installation and Performance Criteria I. Application Gas cabinets are ventilated enclosures that are used to isolate hazardous gas cylinders from the surrounding work place.
More informationUniversity of North Florida Performance Requirements For Laboratory Fume Hoods. As Manufactured Performance Tests. As Installed Performance Tests
University of North Florida Performance Requirements For Laboratory Fume Hoods As Manufactured Performance Tests As Installed Performance Tests As Used Performance Tests October 27, 2010 University of
More informationAPPENDIX A SAMPLE FHT SPECIFICATION (This recommended FHT specification is available from
SAMPLE FHT SPECIFICATION (This recommended FHT specification is available from www.nebb.org) NEBB recommends that these FHT Specifications be referenced as related documents in other appropriate sections
More informationvento specification sheet
type purpose application linear diffuser 18mm diffuser for air distribution for false ceiling the diffuser is used for the supply and exhaust of cooled or heated air in facilities such as offices, shopping
More informationJWS732, aerodynamics and propulsion laboratory users notes
JWS732, aerodynamics and propulsion laboratory users notes These notes are intended to be used as basic guidelines for operating procedures and processes for the JWS732, aerodynamics and propulsion laboratory
More informationASHRAE-110 Type-Test Report for 1.5m Wide Bench-Type fumetop1500 FH-150 Fume Hood of TopAir, USA Ref: ANSI/ASHRAE
ASHRAE-110 Type-Test Report for 1.5m Wide Bench-Type FH-150 Fume Hood of TopAir, USA Ref: ANSI/ASHRAE 110-1995 by Dr A F Bicen Invent UK Ltd 85 Southdown Road Harpenden AL5 1PR Tel: +44 (0) 1582 461666
More informationAppendix FRANCE Lyon Croix-Rousse Tunnel
Appendix 2.9 - FRANCE Lyon Croix-Rousse Tunnel 1. SUMMARY A NEW MULTIMODAL TUNNEL The Croix-Rousse tunnel is located in Lyons (France). The City of Lyons has about 475.000 inhabitants with 1,31 million
More informationBS EN :2003 Containment Factor deciphered
BS EN 14175-3:2003 Containment Factor deciphered Waldner Knowledge Base January 2011 Please let us have your feedback on this publication: steffen.springer@waldner.co.uk +44 7795010353 Content 1. Definition
More informationCONSIDERATION OF DENSITY VARIATIONS IN THE DESIGN OF A VENTILATION SYSTEM FOR ROAD TUNNELS
- 56 - CONSIDERATION OF DENSITY VARIATIONS IN THE DESIGN OF A VENTILATION SYSTEM FOR ROAD TUNNELS Gloth O., Rudolf A. ILF Consulting Engineers Zürich, Switzerland ABSTRACT This article investigates the
More informationTExT Thorough Examination & Test of LEV Local Exhaust Ventilation Systems
ILEVE -Institute of LEV Engineers CPD & Information Day TExT Thorough Examination & Test of LEV Local Exhaust Ventilation Systems Jane Bastow CMIOSH Chartered Safety and Health Practitioner OSCHR Registered
More informationVENTILATION DURING TUNNEL CONSTRUCTION INDUSTRY CONSIDERATIONS
VENTILATION DURING TUNNEL CONSTRUCTION INDUSTRY CONSIDERATIONS Air Quality Working Group Information Package - Part 7 of 12 December 2018 Doc No. AQWG_5_0.08 Page 1 of 12 Ventilation during Tunnel Construction
More informationWorkshop 1: Bubbly Flow in a Rectangular Bubble Column. Multiphase Flow Modeling In ANSYS CFX Release ANSYS, Inc. WS1-1 Release 14.
Workshop 1: Bubbly Flow in a Rectangular Bubble Column 14. 5 Release Multiphase Flow Modeling In ANSYS CFX 2013 ANSYS, Inc. WS1-1 Release 14.5 Introduction This workshop models the dispersion of air bubbles
More informationATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN
ATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN (This page intentionally blank) ATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN TABLE OF CONTENTS O- Definitions... O- Objective... O- Design
More informationFAN ENGINEERING. Field Testing of Fans FE-900. Introduction. Testing Equipment. Testing Standards. Prior to Testing
FAN ENGINEERING Information and Recommendations for the Engineer Introduction There are many reasons why a fan test may be performed. AMCA publications cover three categories of tests. 1. General Fan System
More informationMethods The experiments were carried out in the wind tunnel in the Air physics laboratory at the Engineering Centre Bygholm as shown in figure 1.
1 Standardisation of test method for salt spreader: Air flow experiments Report 2: Visualization of airflow patterns by Jan S. Strøm, Consultant Aarhus University, Engineering Centre Bygholm, Test and
More informationModule 15 : Grit Chamber. Lecture 20 : Grit Chamber
1 P age Module 15 : Grit Chamber Lecture 20 : Grit Chamber 2 P age 15.6 Square Grit Chamber The horizontal flow rectangular grit chamber faces the problem of sedimentation of organic matter along with
More informationARCA Guidance Note. Guidance on Air Management in Asbestos Enclosures
Guidance on Air Management in Asbestos Enclosures 1 GUIDANCE ON AIR MANAGEMENT IN ASBESTOS ENCLOSURES 1. BACKGROUND 1.1 This guidance note is intended to clarify expectations with regard to air management,
More informationSeries 3000 Layout Guidelines
Series 3000 Layout Guidelines Open circuit cooling towers, closed circuit cooling towers, and evaporative condensers all depend upon an adequate supply of fresh, ambient air to provide design capacity.
More informationHigh Performance BSP Dial Up Pattern Pressure Reducing Valve
See our full range online at: www.vip-ltd.co.uk High Performance BSP Dial Up Pattern Pressure Reducing Valve VIP Product Code 31/0103 FUNCTION Pressure reducing valves are devices which, when installed
More informationKrantz Components. Displacement ventilation for indoor firing ranges. Air distribution systems
Displacement ventilation for indoor firing ranges Air distribution systems Low-turbulence displacement ventilation protects the breathing zone of the shooter 2 When shooting with firearms, pollutants in
More informationWind Flow Validation Summary
IBHS Research Center Validation of Wind Capabilities The Insurance Institute for Business & Home Safety (IBHS) Research Center full-scale test facility provides opportunities to simulate natural wind conditions
More informationResults and conclusions of a floating Lidar offshore test
Results and conclusions of a floating Lidar offshore test J. Gottschall, G. Wolken-Möhlmann, Th. Viergutz, B. Lange [Fraunhofer IWES Wind Lidar Buoy next to FINO1 met. mast] EERA DeepWind'2014 Conference,
More informationKansas State University Fume Hood Operation
FUME HOODS 2009, PAGE 1 FUME HOODS Kansas State University Fume Hood Operation One of the primary safety devices in a laboratory is a chemical fume hood. A well-designed hood, when properly installed and
More informationAirborne Activity Sampling and Monitoring at Hinkley Point A Site. David Williams, Hinkley Point A Site
Airborne Activity Sampling and Monitoring at Hinkley Point A Site David Williams, Hinkley Point A Site Presentation objectives Ventilation controls in use at Hinkley Point A Site The air sampling and monitoring
More informationContact person Date Reference Page Michael Lindblad F025843A 1 (1) Sustainable Built Environment
issued by an Accredited Testing Laboratory Contact person Michael Lindblad 2016-02-03 5F025843A 1 (1) Sustainable Built Environment +46 10 516 58 58 Michael.Lindblad@sp.se Accred. No. 1002 Testing ISO/IEC
More informationReport on Pre-improvement Air Leakage Testing, in compliance with ATTMA TSL1 (2010)
54 Melville Road Churchdown Gloucester GL3 2RG Mob: 07866 948200 Report on Pre-improvement Air Leakage Testing, in compliance with ATTMA TSL1 (2010) Result: Unsatisfactory Site address: 8 Benson Place,
More informationWIND LOADS / MOORING & FISH TAILING. Arjen Koop, Senior Project Manager Offshore Rogier Eggers, Project Manager Ships
WIND LOADS / MOORING & FISH TAILING Arjen Koop, Senior Project Manager Offshore Rogier Eggers, Project Manager Ships OVERVIEW Wind Loads Wind shielding Fish tailing? 2 WIND LOADS FOR OFFSHORE MARS TLP
More informationLaboratory Test Results Fire Safety Conveyor Belts
Laboratory Test Results Fire Safety Conveyor Belts ALEXANDRU MARIAN PULBERE* Inspectorate for Emergency Situations Tara Barsei Brasov county, 11 Mihai Viteazul Str., Brasov, Romania Testing of fire safety
More informationWIND DATA REPORT. Paxton, MA
WIND DATA REPORT Paxton, MA September 1 2003 November 30 2003 by James F. Manwell Anthony F. Ellis Kai Wu April 15, 2004 Renewable Energy Research Laboratory 160 Governors Drive, www.ceere.org/rerl (413)
More informationAIDGC PRESENTATION MARCH 2006 HAZARDOUS AREAS STANDARDS & CLASSIFICATION
AIDGC PRESENTATION MARCH 2006 HAZARDOUS AREAS STANDARDS & CLASSIFICATION Jeff Strath Principal Engineer Paterson Flood Engineers Pty Ltd www.pfe.com.au INTRODUCTION This presentation provides an overview
More informationCore Velocity 3.0. Different Ways To Connect & Setup The Core Velocity Belt The Safest and Most Effective Way Possible.
Core Velocity 3.0 Different Ways To Connect & Setup The Core Velocity Belt The Safest and Most Effective Way Possible. Criss Cross Ideally the Criss Cross setup is designed for stationary exercises inside
More informationAppendix ITALY Valsassina Tunnel near Lecco (Lombardy)
Appendix 2.12 - ITALY Valsassina Tunnel near Lecco (Lombardy) 1. SUMMARY The Valsassina tunnel is located along the National Road 36 (SS36), from Milan towards Valsassina in the town of Lecco with about
More informationField Measurement for Aerodynamic Mitigation of Wind Pressure on Gable-roofed Low-rise Building
The Eighth Asia-Pacific Conference on Wind Engineering, December 10 14, 2013, Chennai, India Field Measurement for Aerodynamic Mitigation of Wind Pressure on Gable-roofed Low-rise Building Peng Huang 1,
More informationInstruction Manual. Pipe Friction Training Panel
Instruction Manual HL 102 Pipe Friction Training Panel 100 90 80 70 60 50 40 30 20 10 HL 102 Instruction Manual This manual must be kept by the unit. Before operating the unit: - Read this manual. - All
More informationCORESTA RECOMMENDED METHOD N 6
CORESTA RECOMMENDED METHOD N 6 DETERMINATION OF VENTILATION DEFINITIONS AND MEASUREMENT PRINCIPLES (2015 Revision September 2016) 1. SCOPE This CORESTA Recommended Method specifies a method for the determination
More informationSchool of Engineering & Technology University of Hertfordshire Hatfield, AL10 9AB. Wind Tunnel Testing of Crossgrip TPO Roof Walkway Matting
School of Engineering & Technology University of Hertfordshire Hatfield, AL10 9AB Wind Tunnel ing of Crossgrip TPO Roof Walkway Matting Consultant: Dr. Ken Hart (k.j.hart@herts.ac.uk) Consultancy Report
More informationCustomer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationOIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply
OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply 4.1.1 General The primary function of a system for handling fuel oil is to transfer oil from the storage tank to the oil burner at specified conditions
More informationAlberta Police Recruit Selection Standards
Alberta Police Recruit Selection Standards SUPPLEMENTAL VISION REPORT FORMS For Police Constable Applicants Alberta Solicitor General and Public Security Police Recruit Selection Standards Corneal Refractive
More informationLaboratory Self Inspection Checklist LAB01
Laboratory Self Inspection Checklist LAB01 General Chemical/Biological Laboratories Location Assessor Date This core checklist is provided by the Health and Safety Department as a template, which provides
More informationBUCKET WHEEL RECLAIMERS. AUTHOR: MR. W KNAPPE MAN TAKRAF Fördertechnik GmbH
BUCKET WHEEL RECLAIMERS AUTHOR: MR. W KNAPPE MAN TAKRAF Fördertechnik GmbH Performance of Bucket Wheel Reclaimers (Method of Calculation in Principle) A: Introduction The objective of the following is
More informationVertex VP Operation and Maintenance
AIRBENCH LTD Vertex VP Operation and Maintenance This Edition: 2 October 2018 AirBench Ltd, 14 Grange Farm Road, Colchester CO2 8JW www.airbench.com +44 (0) 1206 791191 Copyright AirBench Ltd 2016 Page
More informationEFFECTS OF SIDEWALL OPENINGS ON THE WIND LOADS ON PIPE-FRAMED GREENHOUSES
The Seventh Asia-Pacific Conference on Wind Engineering, November 8-12, 29, Taipei, Taiwan EFFECTS OF SIDEWALL OPENINGS ON THE WIND LOADS ON PIPE-FRAMED GREENHOUSES Yasushi Uematsu 1, Koichi Nakahara 2,
More informationAdvanced Test Equipment Rentals ATEC (2832)
Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) DOCUMENT NO: STP 2203 VERSION: 2008-07-01 Page 1 of 6 1.0 PURPOSE This test procedure is used by the Electrical Safety
More informationESTATES OFFICE OPERATIONS WORKS INSTRUCTIONS WI 6. Fume Cupboards, Ventilated Cabinets, Local Exhaust Ventilation
ESTATES OFFICE OPERATIONS WORKS INSTRUCTIONS WI 6 ISSUE RECORD Revision Description Updated Date Compiled by Review date 01 Modified Mar 2010 J Cummins Mar 2011 02 Reviewed Aug 2011 J Cummins Aug 2012
More informationExperiment Instructions. Circulating Pumps Training Panel
Experiment Instructions Circulating Pumps Training Panel Experiment Instructions This manual must be kept by the unit. Before operating the unit: - Read this manual. - All participants must be instructed
More informationCHAPTER 1 STANDARD PRACTICES
CHAPTER 1 STANDARD PRACTICES OBJECTIVES 1) Functions and Limitations 2) Standardization of Application 3) Materials 4) Colors 5) Widths and Patterns of Longitudinal Pavement Marking Lines 6) General Principles
More informationRiver Bridge - Test Shaft-1 Clarke County, MS, 4/2/2012
Page 1 of 9 The enclosed report contains the data and analysis summary for the SoniCaliper shaft caliper, performed at River Bridge (Test Shaft-1), Clarke County, MS on Monday, April 02, 2012 by Dave Jakstis.
More informationFLOW MEASUREMENT (SEMI-SUBMERGIBLE OBJECT METHOD)
Page 83 of 232 OKLAHOMA CONSERVATION COMMISSION WATER QUALITY DIVISION STANDARD OPERATING PROCEDURE FLOW MEASUREMENT (SEMI-SUBMERGIBLE OBJECT METHOD) Page 84 of 232 1.0 PROCEDURAL SECTION 1.1 Scope and
More informationScience in Sport. 106 How fast is the wind speed? Read. Count/Tachometer Any EASYSENSE. Sensors: Loggers: Logging time: EasyLog
Sensors: Loggers: Count/Tachometer Any EASYSENSE Science in Sport Logging time: EasyLog 106 How fast is the wind speed? Read When an Olympic or world record is set, the rules for it to stand are quite
More informationBritish Showjumping Course Designers Education System COMPETENCIES REQUIRED
British Showjumping Course Designers Education System COMPETENCIES REQUIRED Competencies for all British Showjumping Course Designers Course Designers Course Designers should be able to demonstrate a good
More informationPurge Star & Ring Purge Systems
March 2013 Operator s Manual Purge Star & Ring Purge Systems Additional instruction for Purge Star Single Hose, Purge Star Double (bypass) Hose and Ring Purge with bypass hose and gas dump valve feature.
More informationPROCESS CONTROL HANDBOOK Z-6
Reversal Bath This step prepares the film for the color-developer step. In the reversal bath, a chemical reversal agent is absorbed into the emulsion. No chemical reaction takes place until the film enters
More informationTiling And Filling The Pool Student Materials Page 1 of 12
The Pool Part A At last, your parents have finally agreed that it might be possible to put a small pool in the back yard. The only catch is, your parents are very busy and don t have time to spend figuring
More informationWind tunnel tests of a non-typical stadium roof
Wind tunnel tests of a non-typical stadium roof G. Bosak 1, A. Flaga 1, R. Kłaput 1 and Ł. Flaga 1 1 Wind Engineering Laboratory, Cracow University of Technology, 31-864 Cracow, Poland. liwpk@windlab.pl
More informationThe principal mechanism for taking medium air velocity measurement is the Anemometer.
1 Introduction Face Ventilation Stoppings Check Curtains Air Measurements Principals of Airflow Respirable Dust Control Use of Scrubbers Actions for Excessive Methane Rock Dusting Permissibility 2 The
More informationLEV GUIDANCE FOR SAFE USAGE, TESTING AND MAINTENANCE OF LOCAL EXHAUST VENTILATION SYSTEMS WITHIN ANGLIA RUSKIN UNIVERSITY RISK MANAGEMENT
2016 LEV GUIDANCE FOR SAFE USAGE, TESTING AND MAINTENANCE OF LOCAL EXHAUST VENTILATION SYSTEMS WITHIN ANGLIA RUSKIN UNIVERSITY RISK MANAGEMENT This guidance is in addition to AR-RMD-HSMS38 ANGLIA RUSKIN
More informationDrying ovens FD series Drying ovens with forced convection Leistungsmerkmale/Ausstattung:
Drying ovens www.matronic.se FD series Drying ovens with forced convection FD series units are primarily used in applications needing rapid drying and sterilization. Totally homogenous temperature distribution,
More informationEFFECTS OF CHEMICAL ADDITIVES ON THE PRESSURE DROP IN THE PIPES
International Journal of Bio-Technology andresearch (IJBTR) ISSN(P): 2249-6858; ISSN(E): 2249-796X Vol. 4, Issue 1, Feb 2014, 1-6 TJPRC Pvt. Ltd. EFFECTS OF CHEMICAL ADDITIVES ON THE PRESSURE DROP IN THE
More informationSUMMARY PROBLEMS CAUSED BY BACKFLOW IN PIPE SYSTEMS.
Page 1 of 11 SUMMARY There are many problems derived from a reverse flow in a piping system. A solution presented in this paper is the WaStop inline check valve. The paper aims to explain some of the important
More informationProtec P3000 (XL) Helium Sniffer Leak Detector
Protec P3000 (XL) Helium Sniffer Leak Detector AC C U R AT E, R E L I A B L E L E A K DETECTION AT A DISTANCE Sniffer leak detection you can rely on INFICON Protec P3000 and Protec P3000XL Helium Sniffer
More informationGlobal EHS Powered Industrial Truck and Pedestrian Safety
Global EHS Powered Industrial Truck and Pedestrian Safety Control No: Policy Name: Powered Industrial Truck and Pedestrian Safety Policy Responsible Corporate Function/Business Segment: Global Environmental
More informationAIC RUGBY BY LAWS 2018
AIC RUGBY BY LAWS 2018 Clause 1: GRADES AND AIC AGGREGATE* a) Grades catered for shall be: 1 st XV 2 nd XV 3 rd XV 4 th XV 10A 10B Yr 9A Yr 9B Yr 8A Yr 8B Yr 7A Yr 7B Yr 6A Yr 6B Yr 5A Yr 5B *PLEASE NOTE
More informationHOW WISE TECHNOLOGY WORKS INFICON Protec P3000 sensor powered by Wise Technology, consists of an evacuated gas cell covered by a quartz membrane.
EASY AND COMFORTABLE TO USE FOR FOOLPROOF OPERATION After the initial setup, there is no need to access the base unit, which frees the operator to concentrate on the sniffing process. All relevant messages
More informationNitrous Oxide as a Tracer Gas in the ASHRAE Standard
Nitrous Oxide as a Tracer Gas in the ASHRAE 110-1995 Standard Martin Burke 1, Larry Wong 2, Ben A. Gonzales 1, and Gerhard Knutson 3 1 TSS Inc., Berkeley, CA 2 University of California Office of the President
More informationOPTIMISATION OF VENTILATION IN THE CASE OF A FIRE IN ROAD TUNNELS
- 51 - OPTIMISATION OF VENTILATION IN THE CASE OF A FIRE IN ROAD TUNNELS ABSTRACT Stosch R. 1, Wehner M. 1, Brandt R. 2 HBI HAERTER AG, Tunnel Ventilation Heidenheim 1, D and Zurich 2, CH The control of
More informationESB TONAL NOISE ASSESSMENT. 30 May West Offaly Power Station. Report Author: Stephen Kearney
Allegro Acoustics Limited, Unit 2A Riverside, Tallaght Business Park, Tallaght, Dublin 24 Tel/Fax: +353 (0) 1 4140485 ESB West Offaly Power Station TONAL NOISE ASSESSMENT 30 May 2016 Report Author: Stephen
More informationDomestic ventilation systems
A BSRIA Guide www.bsria.co.uk A guide to measuring airflow rates EXPOSURE DRAFT In Association with ED46/2013 This publication has been printed on Nine Lives Silk recycled paper. The guidance given in
More informationLINE FOLLOWING ENHANCED RULES. Allar Aasjõe
LINE FOLLOWING ENHANCED RULES Allar Aasjõe +372 5346 7363 allar.aasjoe@robotex.ee Content 1 Introduction... 3 2 The field... 3 3 The Robot... 3 4 The Competition... 4 5 Obstacles... 4 5.1 Line break...
More informationDesign a grit chamber for population with water consumption of 135 LPCD.
1 Example:1 Design a grit chamber for population 50000 with water consumption of 135 LPCD. Solution Average quantity of sewage, considering sewage generation 80% of water supply, is = 135 x 50000 x 0.8
More informationSHORT- AND LONG- TERM MODELLING IN SUPPORT OF SEA LICE BATH TREATMENTS AZAMETHIPHOS, CYPERMETHRIN AND DELTAMETHRIN
ANDERSON MARINE SURVEYS LTD. Croabh, Tighnabruaich, Argyll. PA21 2DY Tel/Fax: 01700 811677 Email: amsl@sol. co.uk Report To: The Scottish Salmon Company Ltd Report No: 14/12.2 Issued By: Checked By: Status:
More informationMaintenance, examination and testing of local. exhaust ventilation
Maintenance, examination and testing of local exhaust ventilation Maintenance, examination and testing of local exhaust ventilation Crown copyright 1998 First published 1990 Second edition 1998 Reprinted
More informationFigure 1. Outer dimensions of the model trucks.
1 Standardisation of test method for salt spreader: Air flow experiments Report 3: Simulations of airflow patterns by Jan S. Strøm, Consultant Aarhus University, Engineering Centre Bygholm, Test and Development
More informationAIC RUGBY BY LAWS 2015
AIC RUGBY BY LAWS 2015 Clause 1: GRADES AND AIC AGGREGATE a) Grades catered for shall be: 1 st XV 2 nd XV 3 rd XV 16A 16B 15A 15B 14A 14B Yr 8A Yr 8B Yr 7A Yr 7B Yr 6A Yr 6B Yr 5A Yr 5B b) Colleges with
More informationMP 3.4 SWIMMING POOL BARRIERS. Table of Contents
Table of Contents Commencement... 2 Application... 2 Referral Agency... 2 Compliance with the QDC... 2 Relationship between the Standard and this part... 2 Associated Requirements... 2 Referenced Documents...
More informationGEOL 106: Earthquake Country Activity 08: Discharge Measurements. Name: Date:
Name: Date: Discharge Calculation Discharge = Velocity X Area or Q = VA Calculate the Discharge for the measurements shown at the right. Q = Discharge Measurement Do not collect water samples in the area
More informationSEQUENTIAL SHOWER VALVE INSTRUCTION MANUAL PLEASE LEAVE THIS MANUAL WITH THE END USER
SEQUENTIAL SHOWER VALVE INSTRUCTION MANUAL PLEASE LEAVE THIS MANUAL WITH THE END USER CONTENTS 1. INTRODUCTION & SAFETY 1 2. DIMENSIONS 2 3. TECHNICAL DATA 3 4. OPERATION 4 5. COMPONENTS 5 6. SITE INSTALLATION
More informationTransferring and Storage
Organometallics Transferring and Storage A dip tube is available for Container 2 which can be screwed into the neck to permit safe transfer of product. Small quantities of metal alkyl can be withdrawn
More informationChapter 4: Implementing Effective Exposure Controls
Chapter 4: Implementing Effective Exposure Controls Implementing effective controls to protect workers from health hazards requires attention and dedication. For example, equipment needs to be purchased,
More informationOutside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program
1 Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program Jim Meacham, CTG Energetics Agenda 2 In Situ Testing Results Reduced Ventilation after
More informationTechniques to achieve wind comfort & wind loads on buildings and appurtenances
Vipac Engineers & Scientists Techniques to achieve wind comfort & wind loads on buildings and appurtenances including shades, verandahs, hoardings and walls. Mr Ian Jones, Managing Director Dr Seifu Bekele,
More informationThe use of the analytical balance, and the buret.
1211L Experiment 1. Density 2015 by H. Patterson Instructor Notes: Students make measurements individually then share data to make the graph. There are four volumetric measurements to be studied; 3.00
More informationUltraLo 1800 Alpha Particle Counter
XIA LLC UltraLo 1800 Alpha Particle Counter Site Requirements & Planning Version: 0.4 Friday, February 15, 2013 Table of Contents I. Site Preparation Tasks... 3 A. Instrument Overview... 3 B. Installation
More informationA. M. Dalavi, Mahesh Jadhav, Yasin Shaikh, Avinash Patil (Department of Mechanical Engineering, Symbiosis Institute of Technology, India)
IOSR Journal of Mechanical and Civil Engineering (IOSR-JMCE) ISSN(e) : 2278-1684, ISSN(p) : 2320 334X, PP : 45-49 www.iosrjournals.org Modeling, Optimization & Manufacturing of Vortex Tube and Application
More informationTHE WAY THE VENTURI AND ORIFICES WORK
Manual M000 rev0 03/00 THE WAY THE VENTURI AND ORIFICES WORK CHAPTER All industrial combustion systems are made up of 3 main parts: ) The mixer which mixes fuel gas with combustion air in the correct ratio
More informationFINAL REPORT. Wind Assessment for: NEW OFFICE BUILDING AT ESSENDON FIELDS Essendon, Victoria, Australia
FINAL REPORT Wind Assessment for: NEW OFFICE BUILDING AT ESSENDON FIELDS Essendon, Victoria, Australia Prepared for: Essendon Fields Pty Ltd Essendon Fields House Level 2, 7 English Street Essendon Fields
More informationPACIFIC SHORING, LLC ALUMINUM SHORING PRODUCTS PSH HEAVY DUTY HYDRAULIC WALER RAIL SYSTEM. TABULATED DATA Effective December 10, 2013
1 HEAVY DUTY HYDRAULIC WALER RAIL SYSTEM Effective December 10, 2013 PSH 265 Roberts Avenue Santa Rosa, Ca. 95407 (707) 575-9014 Construction Engineering Resource 1837 Wright Street Santa Rosa, Ca. 95404
More informationPACIFIC SHORING, LLC ALUMINUM SHORING PRODUCTS HEAVY DUTY HYDRAULIC WALER RAIL SYSTEM. TABULATED DATA Effective August 25, 2015
1 HEAVY DUTY HYDRAULIC WALER RAIL SYSTEM Effective August 25, 2015 PSH 265 Roberts Avenue Santa Rosa, Ca. 95407 (707) 575-9014 Construction Engineering Resource 1837 Wright Street Santa Rosa, Ca. 95404
More informationUNDERSTANDING, ACCELERATED
UNDERSTANDING, ACCELERATED MEASURE FLOW, PRESSURE, AND TEMPERATURE ALL IN ONE INSTRUMENT! Designed for Performance TSI thermal mass flowmeters incorporate a proprietary platinum film sensor design for
More information