1 LIQUID HELIUM SAFETY DOCUMENTS CV/APRILIO
2 Department of Physics and Astronomy University of Sheffield These notes and instructions form the generic Safe Working Practice for the use of liquid helium Liquid Helium User's Only experienced and properly instructed persons should handle liquid helium. Before using Liquid Helium a) Read the following guidelines b) Know and understand properties and hazards associated with it c) Establish plans to cover any emergency situations d) Understand your cryostat and correct operation If in doubt - ask your supervisor Liquid Helium - Special Precautions To know what precautions to take is important to recognise that at 4 kelvin all other gases solidify. Therefore, systems and helium dewars must prevent backflow of air, as this constitutes a major safety hazard. Small volumes of liquid evaporate into large volumes of gas and must be allowed to vent safely, therefore: 1) Always connect to helium recovery line and open appropriate vent valves 2) Never lea ve a dewar open to atmosphere Dewars open to atmosphere for prolong periods can cause "ice plugs" causing pressure build up which can lead to disastrous consequences. Liquid Helium Supplies Requirements for supplies of liquid helium should be arranged with a member of the liquefier staff. Removal to Laboratorv 1. Note how much liquid the dewar contains 2. Before removal: (i) Check bladder valve is open and in good condition (ii) Close both recovery line valves and dewar return valve. (iii) Proceed immediately to the laboratory and re-connect dewar to helium recovery line. (iv) Note gas meter reading and liquid helium in the dewar before commencing the experiment.
3 Liquid Helium - Safe Handling and Usage a) Minimum protection recommended is ( i) Cryogenic gloves (ii) Face visor or safety glasses b) Always In spect the Dewar High Pressure in the dewar wilj be indicated by: (i) An inflated rubber bladde r when the conne ction valve is opened or (ii) Pressure gauge on dewar (if fitted) Vent any gas slowly through the helium rec overy line if found in this condition. c) Never Drop Objects into the Liquid Beware of cold gas and rapid boiling of helium when lowering equipment at ambient temperature into the Dewar. THIS OPERATI ON MUST BE CARRIED OUT SLGWLY TO MINIMISE BOlL OFF AND POTENTIAL COLD VAPOUR BURNS! d) Never Accompany a Dewar in a Lift A sudden vapour release in a co nfined space could be fatal! e) Always Use Correct Syphon and Fittings A vacuum insulated syphon is the only method of transferring helium from a dewar. It consists of a vacuum shielded tube which dips below the liquid level in the de war. The dewar is then lightly pressurised which forces the liquid up and out thr ough the syphon tube. It is important that the correct order of assembly of a dewar head syp hon is foll owed to prevent leaks. Remove brass plu g and lock nut used to seal dewar whe n not in use. Insert syphon 'slowly ) and tighten brass flange with the lock nut to seal the ' 0 ' ring. ENSURE ALL PRESSURE HAS BEEN REMOVED 'BEFORE PERFORlVIING THIS OPERATION f) Always Transfer Liquid Slowly SEE FIG. 1 To prevent thermal shock, avoid high pressure build up (back pressure) and EFFICIENT USE OF THE LI QUID HELTIJM. g) Never Pressurise with Gas other than Heli um Pressuri sation with a bladder is suffic ient for most purp oses and this is ob tained by sgueezing the bladder to create a slight over-pressure. Do not use extern al regula ted supplies unless compe tent to do so!
4 HELIUM SYPHOt~ SEILS for COMMONLY USED TRANSFER LI ES Frer.I. TYPICAL CONTINUOU S FLOW SYPHON SEAL TYPICAL HELIUM TRAN SFER LINE SEAL ~ 'c51 STEM SEAL/ STEM S EAL~.. 5] 0"" RE OUCER/~ I R:==::::;---:".--'O' RIN GS HELIUM DEW AR HELIUM DE WAR
5 h) Pre-cooling Equipment Any liquid nitrogen used to pre-cool liquid helium space in cryostats must be fully removed prior to adding liquid helium. i) Purging Purging of syphons and cryogenic equipment for liquid helium service should only be done with dry helium gas j) Always Thaw Equipment with Hot Air This is by far the quickest and safest method. Liauid Helium - Transfer Efficiency a) Syphon and Cryostat Ensure good vacuum is maintained in both. b) Syphon and Cryostat cooldown Complete this operation slowly to prevent thermal sho ck or high back pressures. c) Boil off due to pressurisation gas (external supplies only ) Slowly apply the helium pressurisation gas as it is hot compared to the liquid d) Depressurisation Loss Do not pressurise the dewar more than is necessary to perform the transfer and always try to fiii in one go. Ice Plugs in Dewar Necks On rare occasions an ice plug may form in the neck of the dewar. This must be dealt with quickly since pressure build up is potentially dangerous. If you discover a dewar which has been left open to atmosphere for a period of time e.g. via syphon entry port, helium recovery valve or bladder pressurisation valve. Initially, probe inside dewar with helium level dipstick to establish it is clear and able to vent. Report occurrence to a member of the liquefier staff. If the Dewar is BlockedJPartialIv Blocked 1) Clear laboratory of all personnel 2) Report immediately to the liquefier staff and/or your supervisor
6 Emergency Procedure for Liquid Helium Low Temperature Research Laboratories Forword Liquid helium is delivered in special dewars which unlike liquid nitrogen cannot be dispensed into a container via a liquid withdrawal valve.all the evaporation gas is returned via pipework connected to the recovery system and therefore should not poise an asphyxiation hazard. It has been calculated that under normal operation of a helium experiment and the rooms natural ventilation that it is not possible to lower the oxygen content below 18% even if the gas was not being returned to recovery. The exceptions are as follows: a) Catastrophic failure ofthe dewars vacuum insulation. b) Multiple dewars and experiments within a single laboratory c) Large volumes in bath cryostat i.e. D.S8 large magnet d) More than one type of cryogen It has been determined than c) and d) only applies to D.S8(or the large magnet should it ever be moved) Oxygen Deficiency alarm (if fitted) I.Never enter a laboratory if the Oxygen Deficiency alarm is sounding/flashing.report immediately to a member of the liquefier staff 2. If the alarm sounds whilst in the room. Immediately, tum off the flow pump, open windows and exit the room. Report immediately to a member of the liquefier staff. 3. Uncontrollable releases of gas i.e. catastrophic dewar failure. Exit immediately and sound the fire alarm Air/ice blockages in the dewar 1. If you are unable to remove a transfer tube from a dewar OR insert a transfer tube, clear the laboratory and adjacent laboratories of personnel. Report immediately to a member of the liquefier staff. 2. If the dewar top is missing or the dewar has been left with the return valve open to atmosphere. Initially, probe with a dip stick to confirm that it is not totally blocked. Connect to recovery and/or replace dewar plug. Report immediately to a member of the liquefier staff 3. On insertion or removal of transfer tubes you experience any restrictions or "Sticking" please report to the liquefier staff
7 Department of Physics and Astronomy Safe Working Procedure This form should be completed for each procedure that may be undertaken in the designated area, which has a hazard or potential hazard associated with it. Each procedure an undergraduate, postgraduate, research associate, staff member, technician or secretary is expected to perform, must be adequately explained, (with the aid ofthis form) at the time of their induction into the working area, by a suitably qualified person. Date: October 2005 Location; Experimental Physics & Astronomy Research laboratories Name of supervisor Details of the procedure: Use of Continuous Flow Helium Cryostats. 1. User 's must have appropriate PPE and experimental training prior to using these systems. 2. It is assumed that the sample is fixed inside the cryostat and evacuation of the system. 3. Connect the dewar to the helium recovery line. Open both the dewar and recovery line li4 swing valves. 4. Remove the dewar blanking plug and keep in a safe place 5. Open the transfer tube needle valve and slowly insert the syphon into the dewar. It is preferred that you get assistance whilst attempting this procedure.insert to a depth of approximately 30 ems and lightly screw the gland seal on the syphon inlet. Gently ease down the tube as it enters the liquid, taking care to minimise the frost on the flexible dewar return line 6. De-frost the syphon gland seal and check, ensuring a gas tight connection. 7. Connect to cryostat ensuring that the white PTFE seal is in position. 8. Connect the helium pump/gas controller to the syphon return connector. Open the valve in the laboratory for the helium pump return line. 9. Switch on the helium pump. 10. When the base temperatures have been reached adjust the flow via the syphon needle valve and gas control panel for efficient use of the helium. End of experiment If the experiment is to continue the next day, simply switch off the helium pump and close the helium pump recovery valve.for re-start open the helium pump return valve and then switch on the helium pump. Remove transfer tubes before the weekend or end of the experiment. Always replace the dewar plug and check both recovery valves are open to the dewar!!! Current safety precautions for this procedure 1.Guidance notes in the Laboratory Safety folder 2. Trained operators only 3. Use PPE 4. Know the Emergency Procedure
8 Recommended procedure for transportation of Cryogenic Lig uids using the Goods lift Liquid Nitrogen Open top dewars - typically the 4L Polystyrene containers should NOT be transported in the lift. In the event of spillage the liquid expansion volume would cause an asphyxiation environment. All such movements should be made using the staircase and no more than two at a time. Low Pressure sealed dewars (160L volume) - rarely, if ever should these vessels need to be transported in the lift. Ifrequired to do so, it is important that the gas over pressure is reduced to 50% ofthe reliefvalve pressure in a well ventilated area. This would be typically 10psi/ 0.7 barg. In the event of breakdown, use the emergency alarm for assistance, do not panic since the relief valve setting will take more than 12 hours to be reached. It is recommended that such a procedure is undertaken by a member ofthe liquefier staff. Liquid Helium lool sealed dewars - commonly transported between C&D floors. Only vessels that have been previously connected to the helium recovery system may be transported. Under these circumstances the dewar overpressure is ~ atmospheric. Upon removal, proceed immediately to your destination, reconnecting the dewar back to the recovery system. Visually check the pressure gauge prior to entering the lift or open bladder inflation valve (if fitted). Please ensure all valves remain closed during transportation and the dewar plug is fitted. In the event of breakdown, use the emergency alarm for assistance, do not panic since the relief valve setting will take more than 12 hours to be reached. Under no circumstances transport dewars that were not correctly connected to the recovery line or exhibit any frosting on the top flange. Seek advice from a member ofthe liquefier staff. General 1. Do not to use the passenger lifts unless the goods lift is out of order 2. Always travel alone 3. Try to avoid transportation during Out of Hours April 2008
Cryogenics is the science of ultra low temperatures. Low temperatures are achieved by the liquefaction of gases.
CRYOGENICS SAFETY Introduction Cryogenics is the science of ultra low temperatures. Low temperatures are achieved by the liquefaction of gases. The gases which are most widely used in industry and research
More informationSchool of Plant Biology - Liquid Nitrogen Induction
School of Plant Biology - Liquid Nitrogen Induction Liquid Nitrogen (LN 2 ) Usage Guidelines If you wish to use LN 2, there are a few rules that have been put in place to ensure your safety and to ensure
More informationEHS Laboratory Guidance: Cryogenic Material
EHS Laboratory Guidance Cryogenic Liquids 1 Introduction/Purpose Cryogenic liquid is defined by the Idaho Fire Code as having a boiling point of less than -90º C (-130º F) at 14.7 psia (1 bar). All cryogenic
More informationHazardous Substances Policy - Control Measures POLICY
Hazardous Substances Policy - Control Measures Aims of the policy Enhanced Good Chemical Practice for Work with Cryogenic Liquids INTRODUCTION Cryogenic liquids are liquefied gases which have very low
More informationMANAGEMENT OF LIQUID NITROGEN
MANAGEMENT OF LIQUID NITROGEN Version 01 / November 2014 Page 1 of 9 TSP/S007 This is a controlled document. It should not be altered in any way without the express permission of the author or their representative.
More informationJANIS OPERATING INSTRUCTIONS FOR SUPERCONDUCTING MAGNET CRYOSTATS
OPERATING INSTRUCTIONS FOR SUPERCONDUCTING MAGNET CRYOSTATS INTRODUCTION The Janis Research Company's Superconducting Magnet/Cryostat System is one of the most versatile tools available to the scientist
More informationCryogenic Materials Operating Procedure and Safety Requirements
VERSION 2.0 JANUARY 2, 2018 Cryogenic Materials Operating Procedure and Safety Requirements UNC CHARLOTTE 9201 UNIVERSITY CITY BLVD., CHARLOTTE, NC 28223 1 I. Overview of Liquid Nitrogen and other Cryogenic
More informationFV Flash Vessel Installation and Maintenance Instructions
4041050/5 IM-P404-10 EMM Issue 5 FV Flash Vessel Installation and Maintenance Instructions 1. Safety information 2. Specific product safety information 3. Product information 4. Installation 5. Commissioning
More informationEXPERIENCE WITH CRYOGENICS SAFETY, PROBLEMS AND SOLUTIONS
EXPERIENCE WITH CRYOGENICS SAFETY, PROBLEMS AND SOLUTIONS Rajiv Sharma and Cryogenic Division SST-1 IPR, Gandhinagar 33rd DAE Safety & Occupational Health Professionals Meet at IPR, Gandhinagar, 23-25
More informationProcedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat. Version 5.1
Project: Novel Electron Bubble Particle Detector Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat Version 5.1 (Using side filling line) Hand Processed Changes HPC No. Date
More informationCompressed Gas Properties. Introduction. Compressed Gas Properties. Definition of a Compressed Gas
1. Properties of gases 2. Cylinders 3. Storage of Gases 4. Regulators 5. Cryogenic Liquids Introduction All compressed gases by definition can be hazardous Know the properties of the gases you are working
More informationOperating Manual ALU 7 - ALU 10 - ALU 20 - ALU 26 ALU 35 - ALU 60 - ALU 100
Operating Manual ALU 7 - ALU 10 - ALU 20 - ALU 26 ALU 35 - ALU 60 - ALU 100 KGW - ISOTHERM Gablonzer Str. 6 76185 Karlsruhe Germany Tel: 0049 / 721 95897-0 Fax: 0049 / 721 95897-77 Internet: www.kgw-isotherm.com
More informationManual Actuated Boiler Blowdown Valves
Manual Actuated Boiler Blowdown Valves Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Operation 5. Maintenance 6. Spare parts p.1 1. Safety
More informationALU-CD 12, ALU-CD 25, ALU-CD 35, ALU-CD
Operating Manual ALU-CD 12, ALU-CD 25, ALU-CD 35, ALU-CD 50 und ALU-CD 60 KGW - ISOTHERM Gablonzer Straße 6 76185 Karlsruhe Germany Tel: 0049 / 721 95897-0 Fax: 0049 / 721 95897-77 Internet: www.kgw-isotherm.com
More informationTable of Contents Want full page pictures with most important parts labeled in the beginning.
1 Table of Contents Want full page pictures with most important parts labeled in the beginning. Precool Procedures (Perform before cooling down): 1. Check N2 and 4He gas tanks to be sure that there is
More informationRev. 1. Cryogenics Thermoflasks. 2122, 2123, 2124, 2129, and 2130
Operating Manual Application Catalog Number 056-720-00 Rev. 1 Cryogenics Thermoflasks 2122, 2123, 2124, 2129, and 2130 Important Read these operating instructions. Failure to read, understand and follow
More informationKGW-ISOTHERM. Information and Operating Instructions for Dewar Vessels Type 26 to 29 B or B-E
KGW-ISOTHERM Germany 76185 Karlsruhe Gablonzer Strasse 6 Tel. 0049 / 721 / 95897-0 Fax. 0049 / 721 / 95897-77 E-Mail: info@kgw-isotherm.de Internet: www.kgw-isotherm.de(com) Information and Operating Instructions
More informationINSTALLATION & MAINTENANCE INSTRUCTION
ARCHON Industries, Inc Liquid Level Gauges Models: BT-LLG ND-LLG INSTALLATION & MAINTENANCE INSTRUCTION Instruction No.: 1014.2 Revision Issued: 3/01/03 Approved: Engineering Manager Warning ONLY QUALIFIED
More informationCRYOGENIC LIQUID HANDLING STANDARD
CRYOGENIC LIQUID HANDLING STANDARD Last updated: January 22, 2018 TABLE OF CONTENTS 1.0 Purpose... 2 2.0 Scope... 2 3.0 Operational Requirements... 2 3.1 Training for users of cryogens... 2 3.2 Training:
More informationSA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Dial Adjustment)
3820050/4 IM-P382-01 CH Issue 4 SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Dial Adjustment) Installation and Maintenance Instructions 1. Safety information 2. Use 3.
More informationJohnston Invar Coupling Johnston Invar Coupling
Johnston Invar Coupling Rev.10 (2017.07) AvdS Page 1 of 8 Table of contents I. STRUCTURE OF THE MANUAL / CLARIFICATION... 3 II. SAFETY AND HEALTH CONCERNS... 4 1 GENERAL REMARK... 5 2 ASSEMBLY INSTRUCTIONS...
More informationPV4 and PV6 Piston Valves
1181250/1 IM-P118-05 ST Issue 1 PV4 and PV6 Piston Valves Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Commissioning 5. Operation 6.
More informationLiquid Helium Bath Cryostat
Liquid Helium Bath Cryostat Operating Manual Model TK1865M CONTENTS First A Word of Warning Lifting and Handling the Cryostat Using Cryogens Introduction The liquid helium cryostat. Type TK1865M Serial
More informationDUQUESNE UNIVERSITY. Liquid Nitrogen Safety
DUQUESNE UNIVERSITY Liquid Nitrogen Safety Prepared by: Environmental Health and Safety Department TABLE OF CONTENTS Page Introduction 1 Handling Liquid Nitrogen 1 First Aid Notice 5 Introduction The safe
More informationProduct Manual. Ultra-Helium Dewar Cryogenic Tanks 60, 100, 250 & 500 Liter Models. Designed and Built by: Chart Inc.
Product Manual Ultra-Helium Dewar Cryogenic Tanks 60, 100, 250 & 500 Liter Models Designed and Built by: Chart Inc. 1300 Airport Dr. Ball Ground, GA 30107 USA (800) 400-4683 Part Number 10576741 Rev. B
More informationSA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Knob Adjustment)
3810050/5 IM-P381-01 CH Issue 5 SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Knob Adjustment) Installation and Maintenance Instructions 1. Safety information 2. Use 3.
More informationMSC-P and MSC-N Manifolds for Steam Distribution and Condensate Collection
1170850/1 IM-P117-36 ST Issue 1 MSC-P and MSC-N Manifolds for Steam Distribution and Condensate Collection Installation and Maintenance Instructions 1. Safety information 2. General product information
More informationKBV21i and KBV40i Key Operated Boiler Blowdown Valves Installation and Maintenance Instructions
4059051/3 IM-P405-48 EMM Issue 3 KBV21i and KBV40i Key Operated Boiler Blowdown Valves Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4.
More informationStandard Operating Procedure (SOP)
Standard Operating Procedure (SOP) For using Cryogenics (Liquid Nitrogen) #1 General Process Description Liquid nitrogen is used for many different applications in the department of Materials Science and
More informationBCV31 DN40 - Blowdown Control Valve
4034850/5 IM-P403-71 AB Issue 5 BCV31 DN40 - Blowdown Control Valve Installation and Maintenance Instructions 1. Safety information 2. Application 3. Technical data 4. Operation 5. Installation 6. Flow
More informationTHE UNIVERISTY OF NEWCASTLE - SCHOOL of BIOMEDICAL SCIENCES
THE UNIVERISTY OF NEWCASTLE - SCHOOL of BIOMEDICAL SCIENCES STANDARD OPERATING PROCEDURE PROCEDURE NO: SBS-GDP 016 MOD: 2nd Issue Page: 1 of 10 Procedure Type: General Discipline Procedure 1. Purpose:
More informationDri-Line Mk2 Spirax-Monnier Compressed Air Drain Trap
0509950/2 IM-P050-21 CH Issue 2 Dri-Line Mk2 Spirax-Monnier Compressed Air Drain Trap Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation and
More informationRDK-408D2 Cold Head. Technical Manual. SHI-APD Cryogenics Inc Vultee Street Allentown, PA U.S.A. Revision A: September 2005
RDK-408D2 Cold Head Technical Manual SHI-APD Cryogenics Inc. 1833 Vultee Street Allentown, PA 18103-4783 U.S.A. Revision A: September 2005 (Reference SHI Manual: December 18, 2003 266404A CD32ZZ-160A)
More informationCRP INSTALLATION, OPERATING AND MAINTENANCE INFORMATION FOR INLINE SAMPLING VALVES
CRP INSTALLATION, OPERATING AND MAINTENANCE INFORMATION FOR INLINE SAMPLING VALVES Sampling Valves Installation Commissioning and Operating Instructions SD IL 300 & SD IL 400 Inline Sampling Valve This
More informationAssembly-, installation- and maintenance instruction manual for bladder accumulators IBV / EBV , top reparable
Page 1 von 8 Assembly-, installation- and maintenance instruction manual for bladder accumulators IBV / EBV 100-575, top reparable Content Seite 0 Legend 2 1 Overview 2 2 Accumulator assembly and installation
More informationStandard Operating Procedure Inert Vacuum Line. Ryan Mulvenna, May 2012
Standard Operating Procedure Inert Vacuum Line Ryan Mulvenna, May 2012 Description of Process The inert vacuum lines located in the fume hoods one (Conjugated Polymer Synthesis: CPS) and two (ATRP polymer
More informationDilution Refrigerator manual
Dilution Refrigerator manual Jaseung Ku November 24, 2011 Abstract This document provides a practical guide to dilution refrigerator operation and describes technical details about the hardware. It is
More informationType 131. Roller base. Information and Operating Instructions for Dewar Vessels Type 131 up to 135 boxed shaped KGW-ISOTHERM
KGW-ISOTHERM Germany 76185 Karlsruhe Gablonzer Strasse 6 Tel. 0049 / 721 / 95897-0 Fax. 0049 / 721 / 95897-77 E-Mail: info@kgw-isotherm.de Internet: www.kgw-isotherm.de(com) Information and Operating Instructions
More informationB. Working with liquid nitrogen local rules when using the X-ray facility, room 5.20, Flowers building
B. Working with liquid nitrogen local rules when using the X-ray facility, room 5.20, Flowers building Contents: Author: Liz Carpenter Date: 31 st March, 2003 1. Properties of liquid nitrogen 2. Hazards
More informationKBV21i and KBV40i Air Actuated Boiler Blowdown Valves
4059051/1 IM-P405-48 AB Issue 1 KBV21i and KBV40i Air Actuated Boiler Blowdown Valves Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Commissioning
More informationBCV31 DN20 - Blowdown Control Valve
4034650/8 IM-P403-37 AB Issue 8 BCV31 DN20 - Blowdown Control Valve Installation and Maintenance Instructions 1. Safety information 2. Application 3. Technical data 4. Operation 5. Installation 6. Flow
More informationSpiratec ST14, ST16 and ST17 Sensor Chambers and sensors
0862050/1 IM-P086-18 MI Issue 1 Spiratec ST14, ST16 and ST17 Sensor Chambers and sensors Installation and Maintenance Instructions 1. Safety Information 2. General product information 3. Installation 4.
More informationGlove Box Installation Manual
Glove Box Installation Manual 1998 by M. Braun Company File: GB-UNI-INS.DOC! Edition 08-00 by M. Boutin! Subject to be changed without notice Glovebox Installation Your Glove box has been fully assembled,
More informationEach cryogenic liquid has its own specific properties, but most cryogenic liquids can be placed into one of three groups:
1.0 Purpose and Applicability 1.1 The purpose of this guideline is to ensure the proper management of cryogenic liquids in order to minimize exposure, accidental release, damage to equipment and injury.
More informationMicroscopy Cryogenic WORKSTATION. Performance by Design
CRYO Microscopy Cryogenic WORKSTATION Optical cryostat for use in microscopy and spectroscopy Performance by Design RC102-CFM Microscopy Cryostat offers fast cooldown, high efficiency, excellent temperature
More informationPC3_ and PC4_ Pipeline Connectors
1283050/4 IM-P128-06 ST Issue 4 PC3_ and PC4_ Pipeline Connectors Installation and Maintenance Instructions 1. Safety information 2. Description PC30 shown 3. Installation 4. Welding of pipeline connector
More informationAPPS-160 Gen 2 User Manual
APPS-160 Gen 2 User Manual Applicable Products: PN 11914406 - APPS-160 10 to 50 PSIG operating pressure PN 11918618 - APPS-160 40 to 85 PSIG operating pressure Rev 9/16/2004 Chart Ind. www-chart-ind.com
More information9.0 Equipment Cleaning and Repair
9.0 Equipment Cleaning and Repair Introduction The following sections provide an overview of procedures and techniques used to prepare phosgene equipment for maintenance and return the equipment to service,
More informationSuperconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)
Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures
More informationMETHODS TO PREVENT THE PREMATURE ACTIVATION OF RELIEF DEVICES ON TRANSPORT TANKS
METHODS TO PREVENT THE PREMATURE ACTIVATION OF RELIEF DEVICES ON TRANSPORT TANKS IGC Doc 184/14/E EUROPEAN INDUSTRIAL GASES ASSOCIATION AISBL AVENUE DES ARTS 3-5 B 1210 BRUSSELS Tel: +32 2 217 70 98 Fax:
More informationMB-SP Series Solvent Purification System
Operation Manual MB-SP Series Solvent Purification System Table of contents 1. System overview 2 2. Column activation.. 5 3. System purging.. 6 4. Glove box integration.... 7 5. Solvent reservoir installation....
More informationPOP Safety Valve. POP Safety Valve INTRODUCTION DEFINITIONS
POP Safety Valve POP Safety Valve INTRODUCTION The effects of exceeding safe pressure levels in an unprotected pressure vessel or system, can have catastrophic effects on both plant and personnel. Safety
More informationSpirax Compact FREME Flash Recovery Energy Management Equipment
IM-UK-cFREME UK Issue 1 Spirax Compact FREME Flash Recovery Energy Management Equipment Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4.
More informationMF2M Monnier Miniature Compressed Air Filter
5040050 / 2 IM-P504-02 CH Issue 2 MF2M Monnier Miniature Compressed Air Filter Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation and commissioning
More informationFaculty/School: Pharmacy Initial Issue Date: Oct 2016
Use this form to assist you to complete risk assessments for hazardous activities and processes. Any serious or ongoing hazards should be reported via RiskWare to ensure that appropriate corrective actions
More informationS1, S2, S3, S5, S6, S7, S8, S12 and S13 Separators Installation and Maintenance Instructions
PREVIOUS REFERENCE NO. IMP02355 0231150/13 IMF0501ENISS2 CMGT S1, S2, S3, S5, S6, S7, S8, S12 and S13 Separators Installation and Maintenance Instructions 1. Safety information 2. General product information
More informationWorking with Cryogenic Liquids
Standard Operating Procedure (SOP) Working with Cryogenic Liquids BUILDING: PREPARED BY: ROOM: REVISION DATE: Experimental Process Brief Description of the Operation/Experiment: Specialized Training Instructions:
More informationThermocirc Installation and Maintenance Instructions
1570072/2 IM-P157-36 ST Issue 2 Thermocirc Installation and Maintenance Instructions 1 General safety information 2 General product information 3 Installation 4 Commissioning 5 Operation 6 Maintenance
More informationANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS
Annex 3, page 2 ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS The text of existing chapter 15 is replaced by the following: "1 Application This
More informationCP10 Sensor Installation and Maintenance Instructions
4030150/9 IM-P403-26 EMM Issue 9 CP10 Sensor Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Maintenance 5. Spare parts Copyright 2017
More informationMEDICAL EQUIPMENT CATALOG
THE MOST TRUSTED MEDICAL EQUIPMENT IN NORTH AMERICA MEDICAL EQUIPMENT CATALOG Total Components Including Bulk Oxygen Supply, Back up and Delivery Equipment Designs for High Pressure, Liquid or Bulk Applications
More informationMaterial Handling and Storage of compressed gases and Air equipment
Material Handling and Storage of compressed gases and Air equipment Compressed gases can be toxic, flammable, oxidizing, corrosive, inert, or some combination of these hazards. Compressed gas cylinders
More informationCRYOGENIC PRESSURE REGULATOR
Bestobell Valves is a world leader in the design and manufacture of cryogenic valves, recognised globally for product quality and innovation. Our valves are used with numerous cryogenic liquefied gases
More informationCVI Valve Line. Exceeds the industry s highest standards for reliability and performance
CVI Valve Line Exceeds the industry s highest standards for reliability and performance Available in the most standard models with the shortest lead times in the industry Exceptional quality every CVI
More informationDRAFT. Operating Procedures for the NPDGamma Liquid Hydrogen Target in TA-53, Building MPF-35
Operating Procedures for the NPDGamma Liquid Hydrogen Target V0.03 11/26/05 1 DRAFT Operating Procedures for the NPDGamma Liquid Hydrogen Target in TA-53, Building MPF-35 Version 0.03 November 26, 2005
More informationSCA Series Inverted Bucket Steam Traps
0770050/5 IM-P077-06 ST Issue 5 SCA Series Inverted Bucket Steam Traps Installation and Maintenance Instructions 1. General safety information 2. General product information 3. Installation 4. Commissioning
More informationTHE UNIVERSITY OF NEWCASTLE- DISCIPLINE OF MEDICAL BIOCHEMISTRY. STANDARD OPERATING PROCEDURE PROCEDURE NO: GLP 022 MOD: 1st Issue Page: 1 of 8
THE UNIVERSITY OF NEWCASTLE- DISCIPLINE OF MEDICAL BIOCHEMISTRY STANDARD OPERATING PROCEDURE PROCEDURE NO: GLP 022 MOD: 1st Issue Page: 1 of 8 Procedure Type: General Laboratory Procedure 1. Risk Assessment:
More informationTransferring and Storage
Organometallics Transferring and Storage A dip tube is available for Container 2 which can be screwed into the neck to permit safe transfer of product. Small quantities of metal alkyl can be withdrawn
More informationBT6HC Hygienic Sanitary Balanced Pressure Steam Trap for High Capacity and CIP/SIP Applications
1800350/6 IM-P180-12 ST Issue 6 BT6HC Hygienic Sanitary Balanced Pressure Steam Trap for High Capacity and CIP/SIP Applications Installation and Maintenance Instructions 1. Safety information 2. General
More informationGas Lines. Technical Manual. Sumitomo (SHI) Cryogenics of America, Inc Vultee Street Allentown, PA U.S.A.
Gas Lines Technical Manual Sumitomo (SHI) Cryogenics of America, Inc. 1833 Vultee Street Allentown, PA 18103-4783 U.S.A. Revision F: April 2008 261320A TABLE OF CONTENTS Page DESCRIPTION...1 SPECIFICATIONS...2
More informationStandard Operating Procedure
Standard Operating Procedure Use of liquid nitrogen filling station Date: 02-12-13 SOP Title: Principal Investigator: Room and Building: Lab Phone Number: Section 1 Process Use of liquid nitrogen filling
More informationTHE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY
THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY Page: 1 of 9 1. Risk Assessment: This Risk Assessment is to be used as a general guide and as such, cannot accommodate all the varying factors
More informationLiquid Gas Cylinder (LGC)
Liquid Gas Cylinder (LGC) www.soxal.com SOXAL Liquid Gas Cylinder (LGC) A Liquid Gas Cylinder (LGC) is a portable super-insulated vacuum vessel holding liquid nitrogen, oxygen, argon and carbon dioxide.
More informationRC102 & RC110 Sample in Vacuum Continuous Flow Cryogenic Workstation Cryostats
RC102 & RC110 Sample in Vacuum Continuous Flow Cryogenic Workstation Cryostats www.cryoindustries.com INDUSTRIES RC102 & RC110 Sample in Vacuum Continuous Flow Cryogenic Workstation Cryostats Designed
More informationSafe Handling & Use of Liquid Nitrogen and Other Cryogenic Gases
Safe Handling & Use of Liquid Nitrogen and Other Cryogenic Gases Laboratory Safety Training Office of Engineering Safety Texas Engineering Experiment Station (TEES) & The Dwight Look College of Engineering
More information2.0 Dates and Times Tests to be carried out starting Monday 28 Nov for completion Friday 2 Dec.
MS No: :. A 1.0 Scope of Work Preparation notes: delete column before issuing. 1.1 General Description of Works, Location & Boundaries Strength and tightness testing of Glycol System between plant room
More informationM33F ISO, M33S ISO and M33V ISO Ball Valves Installation and Maintenance Instructions
BAC13365 IM-P133-65 CMGT Issue 2 M33F ISO, M33S ISO and M33V ISO Ball Valves Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Commissioning
More informationTA10A and TA10P Steam Tracing Temperature Control Valves Installation and Maintenance Instructions
3500032/2 IM-P350-02 CH Issue 2 TA10A and TA10P Steam Tracing Temperature Control Valves Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4.
More informationD R A F T. Operating Procedures for the NPDGamma Liquid Hydrogen Target at the BL 13. Version 1.00
D R A F T Operating Procedures for the NPDGamma Liquid Hydrogen Target at the BL 13 Version 1.00 October 06, 2010 Operating Procedures for the NPDGamma Liquid Hydrogen Target V1.00 6/10/10 2 Table of Content
More informationSafe handling of cryogenic liquids
Safetygram 16 Safe handling of cryogenic liquids A cryogenic liquid is defined as a liquid with a normal boiling point below 130 F ( 90 C). The most commonly used industrial gases that are transported,
More informationDri-Line Mk3 Monnier Compressed Air Drain Trap
5044050/2 IM-P504-24 CH Issue 2 Dri-Line Mk3 Monnier Compressed Air Drain Trap Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation and Operation
More informationDeveloping Safety Guidelines for Cryogenic Liquids.
Developing Safety Guidelines for Cryogenic Liquids. Purpose: Develop a reference document that can be used by a Principle Investigator or a designee to assist in the development of Safety Operating Procedures
More informationTHE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)
THE BP-301 SERIES Operating and Service Manual Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv) Issue B October 2015 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation...
More informationTSS21 Sealed Thermostatic Steam Tracer Trap
1255050/4 IM-P125-10 ST Issue 4 TSS21 Sealed Thermostatic Steam Tracer Trap Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Commissioning
More informationSEries 29 Hydrant valves installation, operation & maintenance manual
Instruction for use Thank you for selecting an AVK product. With correct use, it will give long and reliable service. This manual has been prepared to assist you install, operate and maintain the valve
More informationLIQUID OXYGEN. General. Health Effects. Safety Considerations
LIQUID OXYGEN General Oxygen is the second largest component of the atmosphere, comprising 20.8% by volume. Liquid oxygen is pale blue and extremely cold. Although nonflammable, oxygen is a strong oxidizer.
More informationUIB30 and UIB30H Sealed Inverted Bucket Steam Traps for use with Pipeline Connectors
1130050/4 IM-P113-02 ST Issue 4 UIB30 and UIB30H Sealed Inverted Bucket Steam Traps for use with Pipeline Connectors Installation and Maintenance Instructions 1. Safety information 2. General product information
More informationTHE HF-210 SERIES. Operating and Service Manual. Series includes all variants of HF-210/211. Issue A October 2014 *Section 6.
THE HF-210 SERIES Operating and Service Manual Series includes all variants of HF-210/211 Issue A October 2014 *Section 6.1 in development 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3.
More information776 Cryogenic Safety Valve
776 Cryogenic Safety Valve INTRODUCTION 776 Cryogenic Safety Valve The effects of exceeding safe pressure levels in an unprotected pressure vessel or system, can have catastrophic effects on both plant
More informationVACUUM & AIRLESS. Select from Kimble vacuum traps, manifolds, Schlenk tubes, adapters, and bubblers for your vacuum and airless glassware needs.
VACUUM & AIRLESS Select from Kimble vacuum traps, manifolds, Schlenk tubes, adapters, and bubblers for your vacuum and airless glassware needs. tions precision glassware solutions precision glassware solutions
More informationPressure Dump Valve Service Kit for Series 3000 Units
Instruction Sheet Pressure Dump Valve Service Kit for Series 000 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00, 400, 500, and 700 applicator
More informationOperating Procedures for GripTight 15.5 SDR & IPS Test Plugs
EST Group DC2518 08/01 REV 3 12/12 Page 1 of 6 Operating Procedures for GripTight SDR & IPS Test Plugs WARNING For proper operation, GripTight plugs must be assembled as shown in Figure 1. Pressure testing
More informationCompressed Gases and Cryogens
Compressed Gases and Cryogens University of Tennessee Safety Guide HM-011 Document Contact: EHS Date effective: January 1, 2009 Revision Date: October 1, 2015 Purpose This guideline adheres to the Occupational
More informationPump out insert (condenser and still lines) overnight with turbo pump
Running the Dilution Fridge Before starting a run: Pump out insert (condenser and still lines) overnight with turbo pump Make sure LN and LHe traps are clean pump out with He4 rotary pump Start He3 rotary
More informationAPT14, APT14HC and APT14SHC Spares 2
6120251/7 IM-P612-06 ST Issue 7 APT14, APT14HC and APT14SHC Spares 2 Installation and Maintenance Instructions 1. Safety information 2. Replacement of trap and outlet check valve mechanism 3. Replacement
More informationOxygen Dialflow Meter. Instructions for Use
Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user
More informationInstallation and commissioning instructions 255 series and 256 series
Installation and commissioning instructions 255 series and 256 series Table of contents 1 General information... 3 1.1 About these instructions... 3 1.2 About this product... 3 1.3 Appropriate usage...
More informationOxygen Contamination
MPMS Application Note 1014-210 Oxygen Contamination This application note describes potential sources for oxygen contamination in the sample chamber and discusses its possible effects. Molecular oxygen,
More informationGAS PROPERTIES ALWAYS REFER TO SAFETY DATA SHEETS
GAS PROPERTIES ALWAYS REFER TO SAFETY DATA SHEETS FLAMMABLE ACETYLENE C 2 H 2 PROPANE C 3 H 8 BUTANE C 4 H 10 HYDROGEN OXIDISING OXYGEN O 2 NITROUS OXIDE FLUORINE CHLORINE INERT NITROGEN N 2 ARGON Ar CARBON
More informationInert Air (N2) Systems Manual
INSTRUCTION MANUAL Inert Air (N2) Systems Manual N2-MANUAL 2.10 READ AND UNDERSTAND THIS MANUAL PRIOR TO OPERATING OR SERVICING THIS PRODUCT. GENERAL INFORMATION Positive pressure nitrogen gas pressurizing
More information