1 OPERATING MANUAL GILMONT ACCUCAL FLOWMETERS 28W092 Commercial Ave. Barrington, IL U.S.A (847) (847) (Fax) A Edition 01
2 TEFLON- Reg TM E.I. DuPont de Nemours, Inc. Trademarks bearing the symbol in this publication are registered in the U.S. and in other countries.
3 DESCRIPTION Variable area Flowmeters (rotameters) are used to measure the flow rate of liquids and gases in a variety of laboratory and industrial applications. These meters consist of a spherical float moving vertically in a glass flowtube with a tapered inside diameter. As the flow through the flowtube increases, the float rises in the flowtube. A scale marked on the flowtube (along with calibration charts and correction factors for fluid density, temperature and pressure) is used to obtain accurate measurements of flow rate. These Flowmeters contain the following features: Correlated and direct reading all in one easy to read meter. Accuracy as good as ± 2% of reading. Also available with NISTtraceable calibration ±1% of reading accuracy at calibrated points. Rugged construction ideal for industrial and laboratory use. Stainless steel, PCTFE or PTFE and glass wetted parts. Interchangeable dual, direct reading scales. Available with or without 14 turn precision valve. Two scale cards are supplied for direct reading of air and water under standard operating conditions (70 F, 1 ATM) and a blank scale is supplied for use with the correlated scale of the meter. These scales are installed over the process connections at the back of the Flowmeter and held in place by the two palnuts (see Figure 2). An additional float made of stainless steel is supplied with each Flowmeter. This float can be used in place of the black glass float installed in the Flowmeter to extend the range of the Flowmeter by approximately 1.75 times for air and 2.12 times for water. UNPACKING Carefully unpack and check contents. If any components are missing or damaged, save the packing list and material and advise the carrier of the discrepancy. ACCESSORIES Flowmeter Stand Kit. Includes one base, one rod and two mounting clamps Flowmeter Base Rod, 18 in L Mounting Clamp, includes wood screws GF-4004 GF-4001 GF-4002 GF
4 INSTALLATION 1. Unpack the Flowmeter and retain all packing material until proper product operation has been verified. NOTE: If used for vacuum applications or when back pressure compensation is required (as in gas blending), the metering valve must be positioned at the outlet (top of the Flowmeter) by inverting the frame assembly in relation to the flowtube assembly. NOTE: Meters are supplied with the scale divisions on the flowtube set up for use with the printed tables. If direct reading scales are to be used, see the instructions at the back of the manual. 2. Before installing, open valve of the Flowmeter (if equipped) and check that the float moves freely in the flowtube. In some cases, floats may appear sluggish or do not move due to condensation resulting from temperature changes during shipping. It is advisable in such cases to blow a dry clean gas through the meter until the float is freed. 3. Leave the Flowmeter valve partly open during installation (it is shipped in the partly open position). 4. It is important that all lines to be connected to the Flowmeter are purged of any dust or other residual contamination prior to installing the meter. NOTE: In some applications a particulate filter should be installed at the inlet of the Flowmeter. Select a suitable filter for the particulate matter to be removed. 5. Isolation or shut-off valves should be installed in the system to permit removal of the meter for cleaning or servicing. 6. Mounting: a. The meter must be mounted in a vertical position. A deviation of 5 degrees or more from vertical will affect the accuracy of the meter. b. Connectors on the meter are supplied with wrench flats which must be held firmly when threading mating connectors, to prevent loosening of the connectors. c. Paste sealants are generally preferred. However, applications which are sensitive to contamination and require inert sealants, TEFLON PTFE tape is recommended. CAUTION: Care must be taken to avoid shredding of TEFLON PTFE tape which can foul meter operation. 2
5 7. Base Mounting: Secure the Flowmeter to the base by means of a 1/4 in-20 screw threaded in the hole in the bottom block. (See ACCES- SORIES for Flowmeter Base model number GF-4001.) Panel Mounting: Locate mounting holes per Figure 1. Secure the Flowmeter to the panel by means of nuts threaded over the inlet and outlet fittings. Hold the fittings across the flats with an open end wrench when removing the nuts so that the O-ring seal remains tight. 8. Be sure that the piping is adequately supported to prevent undue strain on the Flowmeter. WARNING: A leak test must be performed when hazardous fluids are involved. Perform the leak test using a non-hazardous fluid. Start with a low pressure and gradually increase. FLOWMETER FIGURE 1 MOUNTING DIMENSIONS DIMENSIONS-INCHES (mm) A B C D E F G H 65 mm scale 5.65 in 0.60 in 4.50 in 1/ in 1.00 in 1.75 in 1.25 in (All Flowtubes) (143.5 mm) (15.2 mm) (114.3 mm) NPT (19.8 mm) (25.4 mm) (44.5 mm) (31.8 mm) 150 mm scale in 0.60 in 8.81 in 1/ in 1.00 in 1.75 in 1.25 in ( Flowtubes)(254.5 mm)(15.2 mm) (223.8 mm) NPT (19.8 mm) (25.4 mm) (44.5 mm) (31.8 mm) 150 mm scale in 0.79 in in 1/ in 1.50 in 2.13 in 1.75 in ( Flowtubes)(321.1 mm)(20.1 mm) (279.4 mm) NPT (24.1 mm) (38.1 mm) (54.1 mm) (44.5 mm) 3
6 WARNING: Close inlet valve prior to start-up to avoid sudden pressure surges and/or shock. OPERATION 1. Gradually open the metering valve and adjust the flow to the desired rate using the valve of the Flowmeter. The impact of the float at the top of the flowtube can damage the meter if exposed directly to full line pressure. Avoid sudden pressure surges and shock by closing inlet valve prior to start-up. WARNING: Do not exceed the maximum operating pressure 200 PSIG (1379 kpag) or maximum temperature of 250 F (121 C). 2. Flow rate for air and water at standard operating conditions 1 ATM (98 kpag), 70 F (21.1 C) can be determined from reading the scale at the center of the float and appropriate direct reading scale or supplied flow tables. INSTALLATION INSTRUCTIONS FOR DIRECT READING SCALES Proceed as follows: 1. Select desired direct reading scale, i.e., desired range, float and media. 2. Position scale such that the desired scale faces the front of the Flowmeter. 3. Align zero mark on scale with zero mark on flowtube and tighten the palnuts to retain in position. See Figure Remove the four screws holding the Front Shield in place. 5. Remove the Front Shield. 6. Using a 3/16 in hex wrench, loosen the Cylinder Screw at the top of the Flowmeter, approximately 1 to 2 turns. 7. Rotate the flowtube to position the scale on the flowtube to the side. 8. Using a 3/16 in hex wrench, tighten the Cylinder Screw to secure the flowtube. 9. Reinstall the Front Shield and the four screws. 4
7 FIGURE 2 SPECIFICATIONS Accuracy: Greater of ±2% of reading or ±1 scale division using correlated flow tables. Greater of ±5% of reading or ±3 mm of scale using direct reading scale. Maximum input pressure: 200 PSIG (1379 kpag) Maximum fluid temperature: 250 F (121 C) Maximum external temperature: 200 F (93.3 C) Process Connections: 65 mm scale: All flowtubes 1/8-27 NPT (Female) 150 mm scale: flowtubes 1/8-27 NPT (Female) flowtubes 1/4-18 NPT (Female) Reproducibility: 5 Greater of ±1% of reading or ±1/2 scale division.
8 WARRANTY Gilmont Instruments warrants its product to be free of material or manufacturing defects for the period noted on the warranty card from the date received. If any such defect is noticed, please return the instrument directly to our factory for repair, in which case, it will be repaired at no charge. PRODUCT RETURN To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instructions before returning the product, either within or outside the warranty period. When returning the product, please state the reason for return. For your protection, pack the product carefully and insure against possible damage or loss. Any damages resulting from improper packaging are your responsibility. TECHNICAL ASSISTANCE If you have any questions about the use of this product, contact the Manufacturer or authorized seller. Printed in U.S.A
Laboratory Flowmeters
Laboratory Flowmeters Wide selection of flow ranges measure air from 0.02 ml/min to 675 LPM or water from 0.0002 ml/min to 20 LPM High-accuracy correlated flowmeters ±2% of reading! Excellent chemical
More informationInstallation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for the installation, operation, and
More informationOperation Manual Piston Sensed Gas Pressure Regulators
687 Technology Way Napa, CA 94558 Phone: (707) 259-0102 FAX: (707) 259-0117 www.aptech-online.com Operation Manual Piston Sensed Gas Pressure Regulators (Models KT9, KT10, Welded KT10, KT12) Table of Contents:
More informationTHE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)
THE BP-301 SERIES Operating and Service Manual Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv) Issue B October 2015 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation...
More informationINSTRUCTIONS FOR FM4460N SERIES HIGH PRESSURE, METAL TUBE FLOWMETERS
INSTRUCTIONS FOR FM4460N SERIES HIGH PRESSURE, METAL TUBE FLOWMETERS THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY FOR
More informationInstallation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker Probe Instrument Regulator Model The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for
More informationBP48-1 AND SSP48-1 IN-GROUND POST INSTALLATION INSTRUCTIONS
BP48-1 AND SSP48-1 IN-GROUND POST INSTALLATION INSTRUCTIONS WARNING: THIS IN-GROUND POST IS NOT DESIGNED FOR USE WITH AN LP GAS CYLINDER. WARNING: SEE YOUR GRILL OWNER S MANUAL FOR PROPER LOCATION, MINIMUM
More informationTHE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301
THE HF-300 SERIES Operating and Service Manual Series includes all variants of HF-300/301 Issue A July 2015 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation... 4 3.1. Spring Loaded...
More informationPTFE-SINGLE GLASS FLOW METERS
PTFE-SINGLE GLASS FLOW METERS T GENERAL DESCRIPTION Model T fl ow meters incorporate the principles of traditional variable area fl ow technology. PTFE 150mm with MVT TM Valve These rugged PTFE-Glass flow
More informationCompatible with a Wide Range of Gases or Liquids for Laboratory or Plant Applications
Riteflow Flowmeters Compatible with a Wide Range of Gases or Liquids for Laboratory or Plant Applications Used for flow measurements in the lab, process flows in pilot plants, metering carrier gases in
More informationOPERATING MANUAL EASY-LOAD PUMP HEADS
OPERATING MANUAL EASY-LOAD PUMP HEADS Cole Parmer Cole-Parmer Instrument Co. 1-800-MASTERFLEX (627-8373) (U.S. and Canada only) 11 (847) 549-7600 (outside U.S.) (847) 549-7600 (Local) www.masterflex.com
More informationINSTRUCTIONS FOR SERIES 50 & 50K FLOWMETERS
INSTRUCTIONS FOR SERIES 50 & 50K FLOWMETERS THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY FOR USE IN CONJUNCTION WITH SERIES
More informationKOBOLD KSM FLOWMETER. User Instructions
KOBOLD KSM FLOWMETER User Instructions USA 1801 Parkway View Drive Pittsburgh, PA 15205 PH 412-788-2830 Canada 9A Aviation Point Claire, QC H9R 4Z2 PH 514-428-8090 www.koboldusa.com KSM_manual_08-10 Table
More informationshutoff valves 900 Series psig Manual Shutoff Valves
0 6000 psig Manual Shutoff Valves Features Zero leakage Protected o-ring Full flow passages Freedom from wire drawing Panel mount standard Hand wheel or toggle handle How it Works Closed O-ring seal on
More informationPressure Dump Valve Service Kit for Series 2300 Units
Instruction Sheet Pressure Dump Valve Service Kit for Series 00 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00 applicator tanks when the unit
More informationCSA Sample Draw Aspirator Adapter Operator s Manual
30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating
More informationPP-1, PP-1W, PP-2, & PP-3
Welker Pitot Probe Model PP-1, PP-1W, PP-2, & PP-3 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for the installation, operation, and maintenance
More informationLRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.
LRS(H)4 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to the
More informationINSTALLATION AND OPERATING INSTRUCTIONS
INSTALLATION AND OPERATING INSTRUCTIONS For 3D Instruments Direct Drive Pressure Gauges 3D Instruments Inc., 15542 Chemical Lane, Huntington Beach, CA, 92649, U.S.A. Tel. (714) 894-5351, Fax. (714) 895-4309
More informationTHE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401
THE MF-400 SERIES Operating and Service Manual Series includes all variants of MF-400/401 Issue A October 2013 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation... 4 4. Special
More informationINSTRUCTIONS FOR SERIES 150 TWO TUBE GAS BLENDER
INSTRUCTIONS FOR SERIES 150 TWO TUBE GAS BLENDER THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY FOR USE IN CONJUNCTION WITH
More information2 Sentry MVS Installation, Operation & Maintenance
Gas Liquid & Slurry Solid & Powder Steam & Water Installation, Operation & Maintenance Manual Original Instructions Liquid Sampling Manual Low-Emission Samplers S-LS-IOM-0026-4 -7 Sentry MVS 966 Blue Ribbon
More informationTESCOM 50-4X Series Safety, Installation & Start-Up Procedures
Operations & Service Manual TESCOM 50-4X Series Safety, Installation & Start-Up Procedures Do not attempt to select, install, use or maintain this product until you have read and fully understood this
More informationInstallation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model CP-30SI
Installation, Operation, and Maintenance Manual Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model The information in this manual has been carefully checked for accuracy and is intended
More informationLRS(H)4 Pressure-Reducing Regulator User Manual
LRS(H)4 Pressure-Reducing Regulator User Manual Read the complete manual before installing and using the regulator. 2 Safe Product Selection When selecting a product, the total system design must be considered
More informationTank Blanketing Pressure Regulators RHPS Series
www.swagelok.com Tank Blanketing Pressure Regulators RHPS Series Types: pressure reducing and vapor recovery 16L stainless steel construction 1/2, 1, and 2 in. end connections Working pressures up to 22
More informationRK-IR Sample Draw Aspirator Adapter Operator s Manual
30-0951RK-IR Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0018RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment
More informationINSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER CONSTANT PRESSURE CYLINDER
INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER CONSTANT PRESSURE CYLINDER MODEL CP-11 DRAWING NUMBERS AD157CO AD157CP.CE MANUAL NUMBER IOM-184 REVISION Rev. B, 6/7/2018 TABLE OF CONTENTS SAFETY
More informationInstallation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker Instrument Supply Pressure System Model WIC The information in this manual has been carefully checked for accuracy and is intended to be used as a
More informationPURA. Pure Gas Dewpoint Transmitter. Users Guide
PURA Pure Gas Dewpoint Transmitter Users Guide Issue January 2003 Page 2 Table of Contents SECTION PAGE 1 Product Overview 3 2 Preparation 3 3 Installation 3 3.1 PURA Premium & OEM Dewpoint Transmitter
More informationMultiple Tube - Teflon PTFE-Glass Flowmeters
Multiple Tube - Teflon PTFE-Glass Flowmeters xt Model xt Multiple Tube Teflon PTFE-Glass Flowmeters combine the convenience of multiple tube meters with the unique design features of single tube Teflon
More informationBULLETIN FOR ACCURATE INDICATION AND CONTROL OF GAS AND LIQUID FLOW IN LOW RANGES.
BULLETIN 20-7000 ROTAMETERS PURGE FOR ACCURATE INDICATION AND CONTROL OF GAS AND LIQUID FLOW IN LOW RANGES. INTRODUCTION McCrometer purge rotameters are precision-made instruments designed to accurately
More informationPRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.
PRS(TC)4,8 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to
More information6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS
6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS USER MANUAL -1- SI-BARNET DEADWEIGHT TESTER 6000 SERIES Release: S31-0250 Revision: A Date: 05/08/03 MANUFACTURED BY: GE RUSKA RUSKA INSTRUMENT CORPORATION 10311
More informationBPB26-1 AND SSPB26-1 PATIO BASE INSTALLATION INSTRUCTIONS
BPB26-1 AND SSPB26-1 PATIO BASE INSTALLATION INSTRUCTIONS WARNING: THIS PATIO BASE IS NOT DESIGNED FOR USE WITH AN LP GAS CYLINDER. WARNING: SEE YOUR GRILL OWNER'S MANUAL FOR PROPER LOCATION, MINIMUM CLEARANCES,
More informationModel VR6 System. Installation, Operation & Maintenance
Model VR6 System Installation, Operation & Maintenance General: All Archer Instruments chlorination systems are carefully designed and tested for years of safe, accurate field service. All Archer Instruments
More informationInstructions Manual. Variable area flowmeter 60 M-1. Series. instrumentation for fluids. Technical Data. C-MI-60M1 Rev.: 0 English Version
instrumentation for fluids Variable area flowmeter Series 60 M-1 Instructions Manual Technical Data Accuracy: ± 2,5% of full scale value. Based on VDE/VDI 3513 Scales: In l/h, l/min, m³/h, kg/h, %, mm.,
More informationMUELLER. A Wall Type. Indicator Post. Reliable Connections. General Information 2. Technical Data/ Dimensions 3. Installation 4-5.
Installation Instructions manual MUELLER table of contents PAGE A-20814 Wall Type General Information 2 Technical Data/ Dimensions Installation 4-5 Maintenance 6 Parts 7 Indicator Post! WARNING: 1. Read
More informationRS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.
RS(H)10,15 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to
More informationINSTRUCTIONS FOR MODEL SSC LOW DELIVERY PRESSURE GAS REGULATORS
INSTRUCTIONS FOR MODEL SSC LOW DELIVERY PRESSURE GAS REGULATORS THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY FOR USE IN
More informationPrecision and Instrument Regulators 1/4" port size
Section 8 Precision and Instrument Regulators /" port size Contents -8 Precision Air Pressure Regulator /" ports............ale-8-2 R38 Instrument Regulator Aluminum Model /" ports.........ale-8- -8 R38
More informationSPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1.
2 GALLON TWIN STACK AIR COMPRESSOR Model: 9526 DO NOT RETURN TO STORE. Please CALL 800-348-5004 for parts and service. CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or
More informationBioAerosol Nebulizing Generator. Operation and Maintenance User Manual
BioAerosol Nebulizing Generator Operation and Maintenance User Manual INTRODUCTION The BANG or BioAerosol Nebulizing Generator is a unique nebulizer for the generation of aqueous aerosols at a low air
More informationDiscontinued. Powers Controls. Technical Instructions Document No P25 RV Rev. 1, May, RV 201 Pressure Reducing Valves.
Powers Controls RV 201 Pressure Reducing Valves Description Features Product Numbers Dual Pressure PRV Technical Instructions Document No. 155-049P25 RV 201-1 Single Pressure PRV The RV 201 Pressure Reducing
More informationMultichannel Pump Head pages 1-8. Tête de pompe multicanal pages F1-F8. Mehrkanal- Pumpenkopf Seiten D1-D8
OPERATING MANUAL: Multichannel Pump Head pages 1-8 NOTICE D UTILISATION: Tête de pompe multicanal pages F1-F8 BEDIENUNGSANLEITUNG: Model Nos. Nº de modèle Modellnummern Modelo No. Modello No. 07534-04
More informationInstruction Manual - Diaframless TM Ejector Chlorine, Sulfur Dioxide and Ammonia
Instruction Manual - Diaframless TM Ejector Chlorine, Sulfur Dioxide and Ammonia - 1-122.6010.6 These instructions describe the installation, operation and maintenance of the subject equipment. Failure
More informationAnti-flood device Model B-1
December 4, 2009 Dry Systems 123a 1. DESCRIPTION The Anti-flood Device is required when Viking accelerators are installed on dry systems according to Viking Model E-1 Accelerator Trim Charts. In the SET
More informationINSTRUCTIONS FOR MODELS UPH & UPHS HIGH FLOW, HIGH PURITY GAS REGULATORS
INSTRUCTIONS FOR MODELS UPH & UPHS HIGH FLOW, HIGH PURITY GAS REGULATORS THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY
More informationWelker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker Sampler Model GSS-1 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide to operations. Correct
More informationSingle Tube Flowmeters
Single Tube Flowmeters P Interchangeable Designed for low flowrates, the model P flowmeter is a precision instrument embodying the inherent simplicity, versatility and economy of the classical variable
More informationYour safety and the safety of others are very important.
NATURAL GAS TO PROPANE CONVERSION KIT ALPKT57- INSTALLATION INSTRUCTIONS PROPANE CONVERSION KIT SAFETY... INSTALLATION REQUIREMENTS... Tools and Parts... LP Gas Requirements... Table of Contents INSTALLATION
More informationOperation Manual. Diaphragm Sensed Gas Pressure Regulators
687 Technology Way Napa, CA 94558 Phone: (707) 259-0102 FAX: (707) 259-0117 www.aptech-online.com Diaphragm Sensed Gas (AP/AZ/AK Models: 20, 100, 500, 1000, 1000T 10PA, 1100, 1200, 12PA, 1300, 1400T, 14PA,
More informationBinks SG-2 Plus TM Rotary 2 QT. PRESSURE CUP
Binks SG-2 Plus TM Rotary 2 QT. PRESSURE CUP Model No. 80-651 Rotary Agitator INTRODUCTION Binks SG-2 TM Plus Rotary 2 Qt. Pressure Cup is ideal for component spraying and industrial applications where
More informationEASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:
2 Table Of Contents 1. Introduction 3 2. About this Manual 3 3. Contacting YZ Systems 3 4. Vessel Components 4 5. Specifications 5 6. Application 6 7. Theory of Operation 7 8. DuraSite Installation & Use
More informationConstant Pressure Crude Oil Container Model CPCCP
Installation, Operations, and Maintenance Manual Constant Pressure Crude Oil Container Model CPCCP The information in this manual has been carefully checked for accuracy and is intended to be used as a
More informationRelated Products: Auto Drain Valve. Model/Specifications. Model
Related Products: Auto Drain Valve /600 Drainage is automatically discharged in a reliable manner, without requiring human operators. Highly resistant to dust and corrosion, operates reliably, and a bowl
More informationBinks SG-2 Plus TM 2 QT. PRESSURE CUPS Model No (Standard) Model No (with OscillatingAgitator)
INTRODUCTION Binks SG-2 Plus TM Pressure Cup is ideal for component spraying and industrial applications where small batch production spraying is required. The 2 qt. capacity is sufficient to complete
More informationTECHNICAL DATA ANTI-FLOOD DEVICE MODEL B-1 1. DESCRIPTION
Page 1 of 6 1. DESCRIPTION The Model B-1 Anti-flood Device is required when Viking accelerators are installed on dry systems according to Viking Model E-1 Accelerator Trim Charts. In the SET condition,
More informationGas Proportioners - Back Pressure Compensated
Gas Proportioners - Back Pressure Compensated G To blend two or three gases in homogeneous infinitely variable concentrations, directly at the end use point, this instrument is unsurpassed in convenience
More informationInstallation, Operation & Maintenance Manual
Original Instructions Installation, Operation & Maintenance Manual Sentry VREL Control Valve Pressure Conditioning S-SW-IOM-00277-15 1-17 Table of Contents Safety Information... 3 General Safety Precautions...
More informationSPECIFICATIONS. Approximate Weight: 10 oz. Surface Finish: Ra micro inch or less MATERIALS OF CONSTRUCTION
DIFFUSION-RESISTANT, DIAPHRAGM SEAL, EXCESS FLOW SHUT-OFF VALVES (FS SERIES) FS Series Excess Flow Shut-Off Valves are designed to automatically shut-off the delivery of gas in a line if the flow exceeds
More informationMUELLER A A Non-Adjustable. Vertical Indicator Posts. Reliable Connections. General Information 2. Technical Data 3.
Installation Instructions manual MUELLER table of contents PAGE A-20808 General Information 2 Technical Data 3 Dimensions 4 A-20809 Non-Adjustable Installation 5-6 Parts 7 Maintenance 8 Vertical Indicator
More informationINSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER FILTER DRYER
INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER FILTER DRYER MODEL F-4 DRAWING NUMBERS AD042C[ ] AD054C[ ] MANUAL NUMBER IOM-046 REVISION Rev. F, 4/23/2018 TABLE OF CONTENTS SAFETY 3 1. PRODUCT
More informationDPC-30 DPC-100. Reference Manual
DPC-30 DPC-100 Reference Manual 1. Introduction 1.1 Description The Martel DPC Digital Pneumatic Calibrator improves upon traditional dial gauge pneumatic calibrators. The Martel DPC improves accuracy,
More informationVariable Area Flow Meters
Variable Area Flow Meters Over sixty years of variable area meter expertise Brooks started designing and manufacturing variable area (VA) meters, also referred to as rotameters, in 1946. Now, with more
More informationInstallation/Care/Use Manual
Installation/Care/Use Manual EMASM Surface Mount Bottle Filling Station IMPORTANT THIS IS AN INDOOR APPLICATION ONLY! ALL SERVICE TO BE PERFORMED BY AN AUTHORIZED SERVICE PERSONNEL. TOOLS/ITEMS REQUIRED
More informationRD(H)20/25 Pressure-Reducing Regulator User Manual
RD(H)20/25 Pressure-Reducing Regulator User Manual Read the complete manual before installing and using the regulator. 2 Safe Product Selection When selecting a product, the total system design must be
More information1200B2 Series Service Regulators. Instruction Manual
00B Series Service Regulators Instruction Manual 00B Series Service Regulators 0 Elster American Meter 00B Series Service Regulators General Information The 00B Series Service Regulators are available
More informationFOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM
IMI CORNELIUS INC One Cornelius Place Anoka, MN 55303-623 Telephone (800) 238-3600 Facsimile (612) 22-326 INSTALLATION INSTRUCTIONS FOR INSTALLING CO 2 BLENDER KIT (P/N 111612000 IN BEER SYSTEM SECONDARY
More informationSection 3- Repair and Calibration Procedures
Section 3- Repair and Calibration Procedures 1. Refrigerant Removal Procedures 2. Scale/PC Board Recalibration Procedure 3. PC Board Configuration Procedure of AR400 4. Scale Tare Procedure 5. Gauge Calibration
More informationINSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER RELIEF VALVE
INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER RELIEF VALVE MODELS RV-1 RV-2 RV-2CP RV-3 DRAWING NUMBERS AD017A[ ] AD018A[ ] AD020A[ ] AD282BO MANUAL NUMBER IOM-033 REVISION Rev. E, 3/28/2016 TABLE
More informationBrooks Flow Controllers for Gas and Liquid Service Model FC 8744, Series FC 8800 and FC 8900
Data Sheet Flow Controllers for Gas and Liquid Service Brooks Flow Controllers for Gas and Liquid Service Model FC 8744, Series FC 8800 and FC 8900 Features and Benefits Flow controllers for high pressure
More informationINSTRUCTIONS FOR SINGLE STAGE LECTURE BOTTLE REGULATORS
INSTRUCTIONS FOR SINGLE STAGE LECTURE BOTTLE REGULATORS THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY FOR USE IN CONJUNCTION
More informationTECHNICAL DATA ACCELERATOR MODEL E-1
Page 1 of 10 1. PRODUCT DESCRIPTION The Viking Model E-1 Accelerator is a quick-opening device. When installed with the required external Anti-flood Device, the assembly is designed to increase the operating
More informationINSTRUCTIONS FOR MODELS APC & APG HIGH PURITY, CORROSIVE GAS REGULATORS
INSTRUCTIONS FOR MODELS APC & APG HIGH PURITY, CORROSIVE GAS REGULATORS THIS BOOKLET CONTAINS PROPRIETARY INFORMATION OF ADVANCED SPECIALTY GAS EQUIPMENT CORP. AND IS PROVIDED TO THE PURCHASER SOLELY FOR
More informationTECHNICAL DATA CAUTION
Page 1 of 6 1. DESCRIPTION The Viking Model D-2 Accelerator is a quick-opening device, with an integral anti-flood assembly, used to increase the operating speed of a differential type dry pipe valve.
More informationP5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.
Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com P5513 Pneumatic Comparison Test Pump Users Manual PN 3963372 November 2010 2010 Fluke Corporation. All
More informationMultiple Tube Flowmeters
Multiple Tube Flowmeters xp The model xp multiple tube flowmeter line offers, the convenience and simplicity of 2, 3, 4, 5 and 6 tube meters, retaining most of the unique design features associated with
More informationInstallation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series
Introduction Powerex manifolds are cleaned, tested and prepared for the indicated gas service and are built in accordance with the Compressed Gas Association guidelines. The manifold consists of a regulator
More informationTECHNICAL BULLETIN. Sigma-Aldrich Pure-Pac II System. Technical Bulletin AL-236
Sigma-Aldrich Pure-Pac II System Technical Bulletin AL-236 TECHNICAL BULLETIN Product The Sigma-Aldrich Pure-Pac II packaging/dispensing system provides a convenient method for storing and dispensing laboratory
More informationHP-104 & HP-106 Tapping Machine OPERATIONS MANUAL
30230 Los Alamos Rd. Murrieta, CA 92563 800-279-5659 FAX 951-926-2334 TappingMachines.com HP-104 & HP-106 Tapping Machine OPERATIONS MANUAL & OPERATING INSTRUCTIONS WARNING READ: Before using this product,
More informationPressure Dump Valve Service Kit for Series 3000 Units
Instruction Sheet Pressure Dump Valve Service Kit for Series 000 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00, 400, 500, and 700 applicator
More informationVarea-Meter Flowmeters. For Metering and Controlling Flows of Process Fluids
Varea-Meter Flowmeters For Metering and Controlling Flows of Process Fluids Product Descriptions Our broad line of rotametertype flowmeters provides reliable readout, transmission, and control of liquids
More informationDTG - LCD Digital Temperature Gauge USER MANUAL
DTG - LCD USER MANUAL Page 1 of 7 USERS GUIDE page 1. Description 3 2. Function 3 3. Safety Instruction 3 3.1. Safety Conventions 3 3.2. Proper Use 3 4. Installation 4 4.1. Unpacking 4 4.2. Storage 4 4.3.
More information! WARNING. Model PFC-1-G (direct acting) PFC-1-GR (reverse acting) Modulating Pneumatic Liquid Level Controls INSTRUCTION MANUAL MM-110B
INSTRUCTION MANUAL MM-110B Model PFC-1-G (direct acting) PFC-1-GR (reverse acting) Modulating Pneumatic Liquid Level Controls APPLICATIONS: Use with other pneumatic devices, for liquid level sensing in
More informationAPCO ARV CLEAN WATER AIR RELEASE VALVES. Model 50A
APCO ARV CLEAN WATER AIR RELEASE VALVES Model 50A Instruction D12013 February 2017 Instructions These instructions provide installation, operation and maintenance information for APCO ARV Clean Water Air
More informationInstallation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Welker Magnetic Indicator (With Gravity Mixer) Model CP2GM-HP
Installation, Operation, and Maintenance Manual Welker Constant Pressure Cylinder With Welker Magnetic Indicator (With Gravity Mixer) Model The information in this manual has been carefully checked for
More informationATD LB PRESSURE BLASTER INSTRUCTION MANUAL
ATD-8402 90LB PRESSURE BLASTER INSTRUCTION MANUAL SAVE THESE INSTRUCTIONS SAFETY INSTRUCTIONS FOR SANDBLASTER 1. Before opening the tank release the air pressure on the sand tank. To do this, turn off
More informationINSTALLATION / OPERATION / MAINTENANCE INSTRUCTION
ARCHON Industries, Inc. Armored Gauge Valves Model series 700 & 900 INSTALLATION / OPERATION / MAINTENANCE INSTRUCTION Instruction # 1011 Rev. B Issued: July 22, 2003 Approved: K. Mayer, Engineering Manager
More informationAir Operated Hydraulic Pumping Systems to 50,000 psi
High Pressure Equipment Air Operated Hydraulic Pumping Systems to 50,000 psi PS-10: 10,000 psi PS-20: 20,000 psi PS-30: 30,000 psi PS-40: 40,000 psi PS-50: 50,000 psi PS-90: 90,000 psi High Pressure air
More information!!!! SERVICE MANUAL PRESSURE POT 2 GALLON. Service Manual: LT Washington St 931 Progress Ave., #7
EXEL North America, Inc. EXEL Industrial Canada, Inc. 1310 Washington St 931 Progress Ave., #7 West Chicago, IL 60185 Scarborough ONT, M1G 3V5 Ph : (800) 573 5554 Ph : (800) 450 0655 Fx : (800) 664 1511
More informationModel PSI Compressor with 3-Gallon Air Tank 12VDC
Model 6350 150 PSI Compressor with 3-Gallon Air Tank 12VDC IMPORTANT: It is essential that you and any other operator of this product read and understandd the contents of this manual before installing
More informationModel Series 62 Constant Differential Relay
Siemens Industry, Inc. INSTALLATION AND SERVICE INSTRUCTION INTRODUCTION Model Series 62 Constant Differential Relay Rev 11 March 2011 Supersedes Rev 10 The Constant Differential Relay maintains a constant
More informationGETZ EQUIPMENT INNOVATORS RECOVERY SYSTEM CLEAN AGENT FE-36 / HALOTRON 1 PART NO. 4G59751 OPERATIONS MANUAL
GETZ EQUIPMENT INNOVATORS CLEAN AGENT FE-36 / HALOTRON 1 PART NO. 4G59751 OPERATIONS MANUAL GETZ EQUIPMENT INNOVATORS PEKIN, ILLINOIS, U.S.A. PHONE: (888) 747-4389 FAX: (309) 495-0625 Limited Warranty
More informationAIR/OVER HYDRAULIC JACK 20 TON
AIR/OVER HYDRAULIC JACK 0 TON 4487 ASSEMBLY AND OPERATING INSTRUCTIONS 349 Mission Oaks Blvd., Camarillo, CA 930 Visit our Web site at http://www.harborfreight.com Copyright 999 by Harbor Freight Tools.
More informationChoosing the right miniature instrument and control valve is a snap
Choosing the right miniature instrument and control valve is a snap INVALCO Micro Valve The Micro Valve is a unique miniature valve offering versatility and rugged construction. Since its introduction
More informationINSTALLATION & MAINTENANCE INSTRUCTIONS
DESCRIPTION / IDENTIFICATION The QBX series valve uses Proportion-Air closed loop technology for Pressure control. It gives an output pressure proportional to an electrical command signal input. The QB1X
More informationMUELLER. A A Adjustable. Vertical Indicator Posts. Reliable Connections. General Information 2. Technical Data 3.
Installation Instructions manual MUELLER table of contents PAGE A-20806 A-20807 Adjustable General Information 2 Technical Data 3 Dimensions 4 Installation 5-6 Parts 7 Maintenance 8 Vertical Indicator
More information1800C and 1800C-HC Series Service Regulators
1800C and 1800C-HC Series Service Regulators Installation Instructions www.elster-americanmeter.com General Information: The 1800C and 1800C-HC Regulators are available as Full Capacity Internal Relief
More informationINSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May
INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May 2015 98-0520 S/N 2001 N. Indianwood Ave. Tulsa, Oklahoma 74012 U.S.A. TEL: (918) 250-7200 FAX: (918) 459-0165 E-mail: chandler.sales@ametek.com
More information