More Speed and Resolution for Current Methods

Size: px
Start display at page:

Download "More Speed and Resolution for Current Methods"

Transcription

1 More Speed and Resolution for Current Methods A Simple How to Guide for Easier Method Transfer in the Current Lab Environment Edward Kim Application Engineer June 15, 2010 Page 1

2 Key Points for Successful Transfer 1. Set Your Goals--Speed, Resolution, Both? 2. Choosing The Column 3. Reducing The Column Length 4. Changing The Column i.d. 5. Adjusting The Injection Volume 6. Transferring to 1.8um Columns - Is and Grad 7. Increase Resolution 8. Speeding It Up Page 2

3 Rule of Thumb Maybe not always 100% Technically Correct but a very good easy alternative Page 3

4 Column Choice is Key to Overall Success Choose Bonded Phase as Close to Original Method as Possible Try Alternate Bonded Phases to Solve Peak Placement Problems Choose Column Chemistry for Optimum ph Stability and Column Life Page 4

5 Bonded Phase Structures Affect Selectivity and ph Ruggedness StableBond C18 Extend C18 C18 Si O C18 Si O Eclipse C18 StableBond CN Page 5

6 Start Method Development with RRHT Columns: Different ZORBAX RRHT C18 Bonded Phases for Max Selectivity 1 st choice Best Resolution Eclipse Plus C18 & Peak Shape 2 nd choice Good alternate selectivity due to non-endcapped StableBond SB-C18 Mobile phase: (69:31) ACN: water Flow 1.5 ml/min. Temp: 30 C Detector: Single Quad ESI positive mode scan Columns: RRHT 4.6 x 50 mm 1.8 um 3 rd choice Good efficiency & peak shape Resolution could be achieved 4 th choice Resolution not likely, Other choices better, for this separation ,3 4 4 Eclipse XDB-C min Extend-C18 Sample: 1. anandamide (AEA) 2. Palmitoylethanolamide (PEA) 3. 2-arachinoylglycerol (2-AG) 4. Oleoylethanolamide (OEA) Multiple bonded phases for most effective method development. Match to one you are currently using. Page 6

7 How To Match a Column to a ZORBAX RRHT Column General Phase Type Typical endcapped C18 or C8 bonded phases, newer columns Endcapped C18 or C8 columns, older generation Non-endcapped columns Older types of columns Aqueous type columns CN or Phenyl Starting ZORBAX Choice Eclipse Plus C18 or C8 Eclipse XDB-C18 or C8 StableBond C18 StableBond C18, C8 etc. SB-AQ SB-CN, SB-Phenyl Page 7

8 Choose Suitable Column Phase Acidic Stable Bond ph 1 6 SB can use increased temperature up to 100C Mobile Phase ph Intermediate Eclipse Plus /Eclipse XDB + ph 2 9 Eclipse can use increased temperature up to 60C Basic Extend ph Extend can use increased temperature up to 45C Page 8

9 The Complete List of RRHT 600 bar Columns Jan 1 Dimensions Eclipse Plus C18 Eclipse Plus C8 Eclipse XDB-C18 Eclipse XDB-C8 Extend-C x x x x x x x x x x x x x x x Dimensions SB-C18 SB-C8 SB-Phenyl SB-CN SB-AQ Rx-Sil 4.6 x x x x x x x x x x x x x x x (Main product page) (P/N List) 2 Recommended Starting Choices: They are Eclipse Plus C18 Columns. Page 9

10 First, Let s Work on Easy Speed Gains! Page 10

11 Page 11 IS O C R A T IC E L U T IO N

12 Simplified Method Transfer for Increased Isocratic Speed 1. Shorten the Column Length 2. Reduce Particle Size 3. Maintain Flow Rate 4. For More Speed. Wait, Hold the Line - Easy Gains First! Very Fast is Just a Few Slides Away! Page 12

13 Reducing Column Length and Particle Size Maintains Resolution While Reducing Assay Time Column Length (mm) Column Efficiency N(5 µm) Column Efficiency N(3.5 µm) Column Efficiency N(1.8 µm) ,500 21,000 35, ,500 14,000 23, ,500 17, ,200 7,000 12, N.A. 4,200 6, N.A. 2,100 2,500 Efficiency Analysis (N) Time Pressure Peak Volume Solvent Usage Analysis Time* - -33% -50% -67% -80% -90% Reduction in analysis time compared to 150 mm column; all columns 4.6-mm i.d. Shorter columns with small particles provide the efficiency of longer columns with larger particles Page 13

14 First, Reduce Column Length 5um 1.8um Reduce column length by factor of 3 Quite often original method will have more resolution than is actually needed and a reduction by 5 may be possible 3.5 um 1.8um Reduce column length by factor of 2 Quite often original method will have more resolution than is actually needed and a reduction by 3 is possible Page 14

15 Reduced Column Length and Particle Size Reduce Analysis Time Both Separations Performed on Same Agilent 1200 SL HPLC mau Original method ZORBAX LC column Extend-C x 150 mm, 5 μm min mau x faster Mobile phase: (70:30) MeOH: 50 mm pyrrolidine buffer Flow = 1.0 ml/min, Temp. : ambient ZORBAX RRHT column Extend C x 50 mm, 1.8 μm min Page 15

16 Reached Flow Rate Limit? Change Column I.D. Changing Column I.D. Will Allow Lower Flow Rate to Match Linear velocity Can improve speed potential lower initial flow rate used allows higher overheads within the flow rate range of the instrument Reduce Mobile Phase Use Can improve sensitivity for same mass loading, but must optimize the flow path volume to minimize dispersion. Page 16

17 How conversion works for flow Flow modification, for columns of different diameters Flow col. 1 Diam. Diam. column2 column1 2 Flow col. 2 i.e mm 1.0ml/min 0. 21ml/min 4.6mm Page 17

18 Reduce Column Diameter and Flow Rate by Same Factor 4.6 mm 3 mm Reduce flow rate by factor of 2.4 Reducing to 3 mm i.d. column allows use of higher linear flow rates as the 1200 SL pump will pump up to 5ml/min. Compared to 4.6 mm id column at 1 ml/min, you can operate a 3 mm id column at 0.42 ml/min and save solvent or reduce flow rate to MS detector, or you can keep the flow rate at 1 ml/min, and have 2.4-fold increase in linear velocity 4.6 mm 2.1 mm Reduce flow rate by factor of 4.8 Page 18

19 Quick Reference for Changing to Common Column Diameters Maintains Equivalent Linear Velocity for Different Column IDs: Column Type Column ID Flow Rate Analytical 4.6 mm 1.0 ml/min Solvent Saver 3.0 mm 0.42 ml/min Narrow Bore 2.1 mm 0.21 ml/min MicroBore 1.0 mm 47 ml/min Capillary 0.5 mm 12 ml/min Capillary 0.3 mm 4.2 ml/min Nano 0.1 mm 472 nl/min Nano mm 266 nl/min Flow rate column 2 = (diameter column 2) 2 /(diameter column 1) 2 x Flow rate column 1 Maintain equivalent mobile phase linear velocity when scaling down in column diameter. Page 19

20 Changing Both Column Length and Diameter Change from a 4.6 x 250 mm (5 um) to a 3.0 x 100 mm (3.5 um) Column. mau Mobile Phase: 25% methanol in 0.4% Formic Acid ZORBAX SB-C18, 4.6 x 250 mm, 5 mm, 1 ml/min Solvent Used: 34 ml mau min ZORBAX SB-C18, 3.0 x 100 mm, 3.5 mm, ml/min Solvent Used: 5.7 ml, decrease of 83% (decrease in analysis time of 57%) min Page 20

21 OK, Let s Run More Isocratic Speed Increase flow rate If flow rate limit of instrument* is reached, reduce column diameter and reduce flow rate to maintain linear velocity If pressure becomes problematic, increase mobile phase/column temperature Increase flow rate as necessary or desired to within 80-90% of flow or pressure limit * Flow Rate Limit is 5mL/min for Agilent Binary Pump, 10mL/min Agilent Quaternary Pump (P < 200 bar) Page 21

22 HETP (cm/plate) Speeding It Up-How Fast can We Go? Increase Linear Velocity (Flow Rate) mL/min 5.0 mm 3.5 mm 1.8 mm ml/min Interstitial linear velocity (u e - cm/sec) Step 1 (2x Original Flow) Increase the flow rate by 100% Reduce run time by 50% Reduce gradient time segments by 50% Example 1ml/min > 2ml/min 30 minutes > 15 minutes 35-65% over 30 mins > over 15 mins Page 22

23 HETP (cm/plate) Speeding It Up Faster! Increase Linear Velocity (Flow Rate) mL/min 5.0 mm 3.5 mm 1.8 mm ml/min Interstitial linear velocity (u e - cm/sec) Step 2 (3 x original flow) Increase the flow rate by another 100% Reduce original run time to 1/3 Reduce original Gradient time to 1/3 Example 1ml/min > 3ml/min 30 minutes > 10 minutes 35-65% over 30 mins > over 10 mins Is resolution OK? Pressure OK? Keep Going! Page 23

24 HETP (cm/plate) Speeding It Up-FASTER!! Increase Linear Velocity (Flow Rate) mm 3.5 mm 1.8 mm Step 3 (4 x original flow) Increase the flow rate by another 100% Reduce original run time to 1/4. Reduce original Gradient time to ¼ mL/min Example ml/min > 4ml/min 4.0 ml/min 30 minutes > 7.5 minutes Interstitial linear velocity (u e - cm/sec) 35-65% over 30 mins > over 7.5 mins Is resolution OK? Pressure OK? Keep Going! Page 24

25 Agilent 1200 Rapid Resolution System Speed Conventional LC UFLC RRLC F= 1.20ml/min T = 40 C Analysis Time = 11min Solvent Cons. = 13.2ml F = 4.80ml/min T = 40 C Analysis Time = 1.05min Solvent Cons. = 5.1ml R s = R s = 2.3 High Resolution: 4.6mm x 150mm 5.0µm min High Speed: 4.6mm x 50mm 5.0µm R s will be ~60% lower though min F= 1.00ml/min T = 40 C Analysis Time = 1.1min Solvent Cons. = 1.1ml High Speed & Resolution: 2.1mm x 50mm 1.8µm min F= 2.40ml/min T = 95 C Analysis Time: 0.4min Solvent Cons. = 1.0ml PW HH = 197msec > 20x faster! Max Speed at T = 95 o C 2.1mm x 50mm 1.8um min Page 25

26 8X Faster Analysis on RRHT Columns with Easy Method Transfer High Pressure not Required mau RRHT 4.6 x 50 mm 1.8 μm Flow Rate: 3 ml/min Pressure = 229 bar Column: ZORBAX Eclipse XDB-C18 Mobile Phase: 95% ACN: 5% Water Temp: Ambient Injection volume: 1 ul Conventional 4.6 x 150 mm 5 μm Flow Rate: 1 ml/min P = 37 bar min 13.5 min Sample: Vitamin E a, b, g-tocopherols in gel cap Eclipse XDB-C18 is a good first choice for many methods. min Page 26

27 Reduce Analysis Time by up to 95% using Rapid Resolution HT Columns R s (1,2) = N= x 250 mm, 5 mm N= R s (1,2) = x 100 mm, 3.5 mm N=11691 min N= R s (1,2) = 3.3 N= x 30 mm, 1.8 mm 1 ml/min min min N= N=6460 R s (1,2) = x 30 mm, 1.8 mm 2 ml/min min olumns: ZORBAX SB-C18 Mobile Phase: 50% 20 mm NaH 2 PO 4, ph 2.8: 50% ACN Flow Rate: 1 ml/min Temperature: RT Detection: UV 230 nm Sample: 1. Estradiol 2. Ethinyl estradiol 3. Dienestrol 4. Norethindrone Page 27

28 Viscosity Viscosity When to Stop When flow limit of pump is reached Consider using a smaller i.d. column When pressure is approaching maximum (e.g. 550 bar) Remember to allow for increase in pressure at viscosity maximum for mobile phase mixture When resolution is no longer satisfactory ACN/water 40C Methanol/water 40C %ACN %Methanol Page 28

29 Decreased Column Volume May Require Conversion for Injection Volume Keep Injection volume proportional to column volume Inj.Vol. col. 1 Volume Volume column2 column1 Inj.Vol. col. 2 Zorbax column volume = 3.14 x r 2 x L x 0.6 (r and L in cm) i.e. 20μl 0.4ml column2 col ml column1 μl col. 2 Page 29

30 Reduce injection volume Reduction to allow for diameter change 4.6 mm 3 mm = 0.4 x Original 4.6 mm 2.1 mm = 0.2 x Original x Reduction to allow for length change 150 mm 50 mm = 0.33 x Original 150 mm 100 mm = 0.67 x Original 100 mm 50 mm = 0.5 x Original e.g. Original 4.6mm x 150mm transferred to 2.1mm x 100mm = 0.2 x 0.67 = 0.13 x original injection volume Page 30

31 mau mau mau R = 16.5 N/m= Reduce Injection Volume as Column Volume is Reduced 4.6 x 150 mm, 5 um 5 ul inj. USP Requirements: L7 column, R s > 8 T f (5%) < 2.0 for each N=3000 or 20000/m R = 17.3 N/m=117,000 Rapid Resolution 4.6 x 100 mm, 3.5 umtailing factor for each of these six peaks is < ul inj R = 15.8 N/m= Rapid Resolution HT 4.6 x 50 mm, 1.8 um 1.7 ul inj. Mobile phase: ( 500:496:4) acetonitrile: water: H 3 PO 4 Flow = 2.0ml/min. isocratic Temp: ambient Detection: UV 272nm LC: Agilent 1100 Sample: resolution solution, fenoprofen (peak #1) with gemfibrozil prepared as described in USP High resolution and exceptional efficiency maintained for low cost updating to fast LC methods min min min Page 31

32 Page 32 GRADIENT ELUTION

33 Simplified Method Transfer for Increased Gradient Speed 1. Shorten the Column Length 2. Adjust Gradient Time by same Factor 3. Maintain Flow Rate 4. For More Speed Just Wait a Little Longer. Remember, Easy Gains First! Faster is Just a Few Slides Away! Page 33

34 What is Gradient Elution? Is It More Difficult to Increase Gradient Speed? Increasing the solvent strength = Increasing the % organic in the mobile phase Linear solvent strength gradient = % B per min is a constant %ACN 90 90% 70% 50% 30% min. } }} } t 1 = t 2 = t 3 = t 4 = 80% gradient change t G = 40 min. gradient time t G = 2%/min. gradient slope For every 20% change in ACN, t is 10 min. Page 34

35 0% B 0% B 100% B t g = 5 100% B t g = 10 Changing Gradient Time to Affect Retention (k*) and Resolution 100% B k* t g F S %B V m 1/k* = gradient steepness = b 0% B t g = % B F = change in volume fraction of B solvent S = constant F = flow rate (ml/min.) t g = gradient time (min.) V m = column void volume (ml) 0% B t g = Time (min) P1.PPT Group/Presentation Title Agilent Restricted Month ##, 200X

36 How conversion works for time Run Time or Gradient segment Time Adjustment* Time col. 1 Length Length column2 column1 Time col. 2 i.e. 150mm 25min. 250mm 15min. *assumes diameter is equal for columns 1 and 2 Page 36

37 Convert run time (and gradient) to shorter column with no flow increase Flow Rate Injection Volume Run time Gradient Times = As converted = As converted = Reduced by length change reduction factor = Reduced by length change reduction factor Reduction to allow for length change 150 mm 75 mm = 1/2 x Reduction e.g. 18mins 9mins 45-90% over 18mins 45-90% over 9mins Page 37

38 Short Columns Reduce Total Gradient Analysis Time A. 4.6 x 150 mm, 5 mm Eclipse XDB-C8 t G = 18 min 1 Gradient Separation of Cardiac Drugs B. 4.6 x 75 mm, 3.5 mm Eclipse XDB-C8 5 t G = 9 min 4.6 x 150 mm 4.6 x 75 mm Run Time 13 min 8 min Equilibration 15 min 7 min Time Total Analysis 28 min 15 min Time 3 48 samples/day 96 samples/day Time (min) Time (min) Page 38

39 Time to Have Fun More Gradient Speed Reduce Particle Size to Increase Efficiency and Peak Capacity Decrease Gradient Time and Increase Flow Rate by Same Factor If Flow Rate limit of instrument* is reached, reduce diameter of column and reduce flow rate to maintain linear velocity Repeat Flow Rate Increase to Flow or Pressure Limit If pressure becomes problematic, increase mobile phase/column temperature * Flow Rate Limit 5mL/min for Agilent Binary Pump, 10mL/min Agilent Quaternary Pump (P < 200 bar) Page 39

40 Faster Gradient Analysis of Cardiac Drugs What Changes Did We Make to Accomplish This? Optimized Column and Gradient 5 Run Time Equilibration 1 min Time Total Analysis 2.8 min Time 480 Samples/day min Column: Rapid Resolution Eclipse XDB- C8, 4.6 x 50 mm, 3.5 mm Mobile Phase: A: 25 mm Na 2 HPO 4, ph 3 B: MeOH Gradient: 42 90% B in 3 min Flow Rate: 3 ml/min Temperature:35 C Sample: Cardiac Drugs 1. Diltiazem 2. Dipyridamole 3. Nifedipine 4. Lidoflazine 5. Flunarizine Time (min) Page 40

41 Column Efficiency under Gradient Conditions Concept of Peak Capacity. Standard concept of column efficiency, N (plates), is only appropriate for isocratic conditions. A more useful concept for the case of gradient conditions is Peak Capacity - the number of peaks that can be separated (at a specified resolution) in a given amount of time. It is another measure of column efficiency. 4 P c = (1+ t G /w) for gradient conditions 5 mm 1.8 mm 5 2 peaks fit R s = +50% 3 peaks fit 50% more!! Page 41

42 Gradient Resolution and Shorter Columns - Faster Analyses for Complex Samples Higher Peak Capacity, P c = (1+ t G /w): 2.1x150mm, 5mm P/N min gradient 0.2 ml/min k* = 12.0 Detection: UV 214 nm Sample: HSA Tryptic Digest 120 peaks 60 min x50mm, 1.8mm P/N min gradient 0.5mL/min k* = 12.9 min 125 peaks 10 min.! x50mm, 1.8mm P/N min gradient 0.5mL/min k* = 38.6 min 156 peaks! 25 min Conditions: Mobile Phase A: Water w/ 0.1% TFA, B: ACN w/0.1% TFA, Gradient 2%B to 50%B, Temperature: 50C min Page 42

43 Peptide Map of Tryptic Digest of BSA run on Agilent RRHT Zorbax SB-C18, 2.1x50mm, 1.8µ Gradient time 10min Peak Capacity min 398 Longer Gradient Times Increase Peak Capacity 22.5 min 441 Page 43

44 Peak Capacity as a function of Gradient Time (t g ) and Column Length on 1.8 µm RRHT Columns Gradient Time 50 mm 100 mm 150 mm Tripling column length increases peak capacity by 55%. Increasing gradient time by 225% increase peak capacity by 25% Page 44

45 Flow Rate Limitations Due to Pressure? Higher Temperature Can Help! Page 45

46 Higher Temperature as an Aid to Faster Operation Higher Temperature: Temperature should always be considered as a parameter during speed optimization Provides more rapid mass transfer: Improves efficiency enhances resolution Decreases analysis time faster separations with no loss in resolution Decreases Mobile Phase Viscosity Lowers backpressure allows for higher flow rates, faster separations, greater efficiency and use of sub 2-micron columns Can change selectivity optimize resolution Be wary of on-column decomposition Faster flow rates shorten analyte residence time at elevated temperature and lead to less decomposition for labile compounds Page 46

47 Elevated Temperature Reduces Pressure Expands Column Choices Cardiac drugs Column: SB-C18, As described below Mobile Phase: A: 0.1% TFA, 5% MeCN, (v/v) B: 0.08% TFA, 95% MeCN (v/v) Sample: 0.1 mg/ml of cardiac drugs Temperature: 70 C Flow: 2 ml/min. gradient Detection: 230,16 nm R s = 4.3 Very high N column 4.6 x 250 mm, 3.5 µm P=221 bar R s = min R s = 3.6 R s = x 150 mm, 1.8 µm P=418 bar R s = R s = 3.4 % B mm mm mm x 50 mm, 1.8 µm P=164 bar min min Page 47

48 INCREASED RESOLUTION! When Do You Need It? How Do You Get It? Complex Samples Large No. of Peaks in Short Time Frame Closely Related Compounds Changes in Bonded Phase Have Not improved Resolution Changes in Mobile Phase Have Not Improved Resolution Temperature Has Not Helped Change Selectivity What Is Left That Will Improve Resolution? Page 48

49 mau Norm Reduce Particle Size and Maintain Column Length Increased N in Isocratic Separation DAD1 A, Sig=254,4 Ref =of f (051119A\SIG10003.D) DAD1 A, Sig=254,4 Ref=off (051004D\LC_J0002.D) 4.6 x 150, 5um 93 bar N = 7259 R s = 1.15 S/N = min min Column:150x4.6 mm 5µm Pressure: 93 bar N: 8213 Height: 1.25 mau S/N: 42.3 R s = 1.15 Height = 1.25 Noise = 24uAU t r = 14.9 min (1 st epimer) N ptp : mau 4.6 x 150, 3.5um 165 bar Column:150x4.6 mm 3.5µm N = Pressure: 165 bar N: R s = 1.37 S/N = 50 Height:1,34 mau S/N: 50.7 R s = 1.37 Height = 1.34 t r = 15.3 min (1 st epimer) Noise = 20uAU N ptp : mau N Norm N 1 d p DAD1 A, Sig=254,4 Ref=off (051007D\LC_X D) 10 1/dp 4.6 x 150, 1.8um 490 bar min Column:150 x 4.6 mm 1.8µm Pressure: 490 bar N: N = Height:1.78 R s = 1.80 (+57%) S/N: 43.6 S/N = 44 R s = 1.80 t r = 17.2 (1 st epimer) Height = 1.80 N ptp : Noise = 30uAU Page 49

50 1.8u Particles Reveal More Information and Improve Detection and Integration Customer Sample, Translation of Isocratic Impurity Methods, Zoom Critical Time Range (t = 7min) 4 Impurities 2 Not Baseline Separated! 7 Impurities 6 Not Baseline Separated! 7 Impurities All 7 Baseline Separated! 4.6 x 150, 5μm 93 bar N = 7259 R_S = 1.15 S/N = x 150, 3.5μm 165 bar N = R_S = 1.37 S/N = x 150, 1.8μm 490 bar N = R_S = 1.80 (+57%) S/N = 44 Up to 60% higher resolution without loss in sensitivity Page 50

51 u Particle Size Increases Gradient Resolution mau mau Agilent 1100 Agilent 1200 RRLC Increased Peak Capacity Zorbax 2.1x150mm SB C-18, 5µm min Zorbax 2.1x150mm SB C-18, 1.8µm Conditions for both experiments Pumps Solvent A: H 2 O + 0.1% TFA Solvent B: ACN + 0.1% TFA Gradient: 10% to 95% ACN in 40min, hold for 1min Flow Rate: 0.4ml/min Autosamplers Injection volume: 3µl Thermostatted Column Comp. Temperature: 50 C Detectors DAD 2µl cell and 20Hz, 220nm, Ref: Appl. Note by Edgar Naegele min Page 51

52 Particle Size - More Peak Capacity with 1.8 um RRHT Columns - Peptide Map of BSA mau Conditions: Columns: as listed, Mobile Phase: A:0.1% TFA in Water B:0.08% TFA in ACN Gradient: 5% B to 60%B in 25 min. Temperature: 80 C Sample: BSA tryptic digest Column used SB-C18, 2.1x150mm, 1.8µ Peak Capacity 673 Starting Pressure 380 bar 35% More peak capacity, more resolution mau SB-C18, 2.1x150mm, 3.5µ Peak Capacity 502 Starting Pressure 105 bar min Less peak capacity, less resolution Smaller particle size = sharper peaks = greater peak capacity min Page 52

53 Tools Rapid Resolution Compendium CD-ROM (Please Ask For A Copy Pub No EN) Application Examples Technology Overview Rapid Resolution in The Scientific and Trade Press and Conference Posters System Configurator (guide to reduce dead volume and extra column effects) Method Translator Program Page 53

54 Agilent HPLC Method Translator A Simple to Use Tool to Move Methods to RRHT Mini-Demo Method Translator Page 54

55 Summary Shorter 1.8um columns allow faster analysis with same resolving power as longer 5um and 3.5 um columns Change column diameter if necessary to allow for lower solvent consumption and higher accessible linear velocity Adjust the injection volume to maintain the same mass loading on the smaller column Adjust run time and gradient to allow for the shorter column to match k* for longer column or to increase k* vs. longer column. Speed up the flow rate until max flow or max pressure or loss of resolution Page 55

56 Summary Shorter 1.8um columns allow faster analysis with same resolving power as longer 5um and 3.5 um columns Change column diameter if necessary to allow for lower solvent consumption and higher accessible linear velocity Adjust the injection volume to maintain the same mass loading on the smaller column Adjust run time and gradient to allow for the shorter column to match k* for longer column or to increase k* vs. longer column. Speed up the flow rate until max flow or max pressure or loss of resolution Page 56

57 Page 57 Questions?

Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think! There Are Many Reasons to Reduce Analysis Time

Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think! There Are Many Reasons to Reduce Analysis Time Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think! Agilent Technologies, Inc. Fall 8 Page There Are Many Reasons to Reduce Analysis Time Run More Samples/Day

More information

Agilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit

Agilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit Agilent 19 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit Technical Overview Author Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany mau 15 1 1 3 1 All components

More information

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC Evaluation of linear and step gradient performance and retention time precision of the Agilent 12 Series Rapid Resolution LC This document is believed to be accurate and up-to-date. However, Agilent Technologies,

More information

Video Notes. LC Troubleshooting Series Retention Time Shifts

Video Notes. LC Troubleshooting Series Retention Time Shifts Video Notes LC Troubleshooting Series Retention Time Shifts Introduction John Henderson is an Applications Scientist for Agilent Chemistries and Supplies Division. Maureen Joseph is the HPLC Columns Product

More information

Fast method for Ginseng Analyses using Agilent Poroshell 120 Columns Scaled from a Traditional Method

Fast method for Ginseng Analyses using Agilent Poroshell 120 Columns Scaled from a Traditional Method Fast method for Ginseng Analyses using Agilent Poroshell Columns Scaled from a Traditional Method Application Note Pharmaceutical Author Rongjie Fu Agilent Technologies (Shanghai) Co. Ltd. 4 Ying Lun Road

More information

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations GUIDE TO ULTRA-FAST HPLC Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations GUIDE TO ULTRA-FAST HPLC Quick Tips for Converting Conventional Reversed-Phase

More information

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION Standard OQ Test Suite This document describes the test program for qualifying LC analytical-scale Agilent systems; the following table lists

More information

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

Acclaim RSLC. (Rapid Separation Liquid Chromatography) Acclaim RSLC (Rapid Separation Liquid Chromatography) Product Manual for Acclaim RSLC Columns Page 1 of 9 Product Manual for Acclaim RSLC Columns (Rapid Separation Liquid Chromatography) Analytical Columns

More information

Carrier Gases in Capillary GC

Carrier Gases in Capillary GC Carrier Gases in Capillary GC LC Columns and Consumables Mark Sinnott Application Engineer January 15, 2009 CARRIER GAS Mobile Phase Carries the solutes down the column Selection and velocity influences

More information

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation Using Nitrogen instead of Helium as carrier gas, while maintaining exact the same separation efficiency, retention times, peak elution order, without changing the oven temperature conditions. Jaap de Zeeuw

More information

Carrier Gases in Capillary GC

Carrier Gases in Capillary GC Carrier Gases in Capillary GC GC Columns and Consumables Abby Folk Application Scientist January 15, 2008 CARRIER GAS Mobile Phase Carries the solutes down the column Selection and velocity influences

More information

Fused-Core TM Particle Columns provide UHPLC performance on any HPLC

Fused-Core TM Particle Columns provide UHPLC performance on any HPLC Fused-Core TM Particle Columns provide UPLC performance on any PLC Major Contributors and Co-Authors. Bell, W. Campbell, W. Way, K. Buckendahl Supelco a ivision of Sigma Aldrich Content Influence on Resolution

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310 www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople

More information

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS Abstract Highly sensitive analysis of metanephrines in plasma is critical in the diagnosis and treatment of pheochromocytoma and paraganglioma. Here, a HILIC LC-MS/MS method was developed using a Raptor

More information

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind

More information

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries. Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s):

More information

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Gas Chromatography Author Matthew S. Klee Agilent Technologies 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract

More information

Ascentis Express HPLC Columns with Fused-Core Technology

Ascentis Express HPLC Columns with Fused-Core Technology Ascentis Express HPLC Columns with Fused-Core Technology Extreme Performance on Any LC System Twice the Speed and Efficiency Extreme Performance on ANY LC System Half the Backpressure Super-Sensitivity

More information

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood UCT Part Numbers: ECQUUS1015CT- Enviro-Clean 15 ml centrifuge tube with 400 mg MgSO 4 and 100 mg NaCl CUMPS2CT - Enviro-Clean

More information

Prep LC Training Manual

Prep LC Training Manual Prep LC Training Manual (11-6-2013) Location: 1238 Hach Hall Contacts: kamel Harrata and Steve Veysey, 1234 Hach Hall Safety: Introduction: The Agilent PrepStar SD-1 preparative HPLC is a student operated

More information

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction Application Note Rapid performance verification of AZURA systems with refractive index detector Method Keywords ID HPLC Quality control, system verification, refractive index, AZURA Analytical HPLC Plus

More information

Setting a New Standard in Flash Chromatography Performance

Setting a New Standard in Flash Chromatography Performance Setting a New Standard in Flash Chromatography Performance What s New Faster Flow rates, up to 300 ml/min Higher pressures, up to 300 psi (20 bar) Bigger touchscreen, 12 or 15 inches Wider Dynamic Range

More information

Extended Application Note

Extended Application Note Extended Application Note Harmful Substances in Dietary Supplements HPLC Columns with APP A-337 www.mtc-usa.com 1-732-578-1777 INTRODUCTION A dietary supplement is a substance to be consumed for the purpose

More information

The instrument should have a mass range from 2 to 2000 m/z or better Interface

The instrument should have a mass range from 2 to 2000 m/z or better Interface Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for

More information

Prep LC Training Manual S.V.

Prep LC Training Manual S.V. 1234 Hach Hall 515-294-5805 www.cif.iastate.edu Prep LC Training Manual 11-6-2013 S.V. Location: Contact: 1238 Hach Hall Kamel Harrata, 1236 Hach Hall and Steve Veysey, 1234 Hach Hall Safety: All researchers

More information

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic. 458 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com LC TROUBLESHOOTING Dwell Volume Revisited Gradient transfer does not have to be problematic. John W. Dolan LC Troubleshooting

More information

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON' DISSOLUTION ASSAY Release & Stability Studies Ed. 04. General Index: 1. PRODUCT SPECIFICATIONS (USP MONOGRAPH) 2. IDENTIFICATION BY HPLC 3. ASSAY - HPLC SETUP 4. STANDARD PREPARATION 5. SYSTEM SUITABILITY

More information

Agilent Dimension Software for ELSD User Manual

Agilent Dimension Software for ELSD User Manual Agilent Dimension Software for ELSD User Manual Agilent Dimension Software for ELSD User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in

More information

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative

More information

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS UCT Part Numbers: CUQAX22Z Clean-Up C8+QAX, 2mg/1mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQID21-3UM

More information

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column Before you connect your trap and column You need to provide your own trap and column (Note our fittings accommodate trap and column made with fused silica 360 µm OD). Check our guide for packing traps

More information

Ensuring Best Possible UHPLC Performance Through an Innovative Capillary and Fitting Design

Ensuring Best Possible UHPLC Performance Through an Innovative Capillary and Fitting Design Ensuring Best Possible UHPLC Performance Through an Innovative Capillary and Fitting Design Frank Steiner, Thomas Piecha, Tobias Fehrenbach, and Fraser McLeod, Dionex Corporation, Germering, Germany Abstract

More information

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction A Column-Flow Independent Configuration for QuickSwap Application Authors Matthew S. Klee and Bruce Quimby Agilent Technologies, Inc. 80 Centerville Road Wilmington, DE 08 USA Abstract As can be seen from

More information

SOP: Buck Scientific BLC-20P HPLC Operation

SOP: Buck Scientific BLC-20P HPLC Operation Page 1 of 11 Approvals Preparer: John Buford Reviewer: Tim Kull Reviewer: Dr. Margaret Bryans Date: 16OCT13 Date: 30OCT13 Date: 31OCT13 1. Purpose 1.1. Basic operation of the Buck Scientific BLC-20P isocratic

More information

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler Optimizing Vial Pressurization Parameters for the Analysis of Residual Solvents Using the 7697A Headspace Sampler Application Note Pharmaceuticals Author Roger L Firor Agilent Technologies, Inc.

More information

GCMSD-Headspace Analysis SOP

GCMSD-Headspace Analysis SOP Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to

More information

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Supporting information for J. Med. Chem., 1994, 37(5), , DOI: /jm00031a018

Supporting information for J. Med. Chem., 1994, 37(5), , DOI: /jm00031a018 Supporting information for J. Med. Chem., 1994, 37(5), 674 688, DOI: 10.1021/jm00031a018 Terms & Conditions Electronic Supporting Information files are available without a subscription to ACS Web Editions.

More information

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI Low-Pressure Retention Locking with the 789A GC Application HPI Authors Courtney Milner and Russell Kinghorn BST International 41 Greenaway Street Bulleen, VIC 315 Australia Matthew S. Klee Agilent Technologies,

More information

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS Analysis Conducted BY CC RESEARCH (P) Ltd., For Nanoceutical Solutions Inc. USA 2950 Thousand Oaks drive, Suite 3, San Antonio, Texas EXECUTIVE SUMMARY

More information

Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform

Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform POCON3E Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LCMS/MS platform ASMS 2 ThP2 Tetsuo Tanigawa, Toshikazu Minohata Shimadzu Corporation, Kyoto, Japan analysis in single

More information

BETTER SAMPLE CARE. Spark Holland introduces. SPH1299 UHPLC Gradient Pump

BETTER SAMPLE CARE. Spark Holland introduces. SPH1299 UHPLC Gradient Pump BETTER SAMPLE CARE Spark Holland introduces SPH1299 UHPLC Gradient Pump SPH1299 Spark Holland introduces SPH1299 UHPLC Gradient Pump Robust, Precise and Easy to use Innovative pumping technology created

More information

LC, RSLC, Bio RS and Dual Gradient Systems

LC, RSLC, Bio RS and Dual Gradient Systems Thermo Scientific Dionex UltiMate 3000 LC, RSLC, Bio RS and Dual Gradient Systems P/N 4823.0205 Revision B December 2014 Automated Method Scouting Considerations All information in this manual is subject

More information

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Your local gas generation partner. Precision series Modular gas generation solution for GC. Your local gas generation partner Precision series Modular gas generation solution for GC www.peakscientific.com Perform with Precision Specifically designed and engineered for GC laboratory applications,

More information

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS B.Lakshmi,Kallam Haranadhareddy Institute of Technology, Guntur Prof T.V.Reddy,Prof Mallareddy College of Engineering, Secunderabad

More information

no peaks missing peaks too small peaks increasing retention decreasing retention

no peaks missing peaks too small peaks increasing retention decreasing retention GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering

More information

During today's webinar we're going to look at the following topics:

During today's webinar we're going to look at the following topics: Alternative Carrier Gas for GC and GC-MS Presentation Transcript Hello and welcome to the Peak scientific webinar which today will concern alternative carrier gas for GC and GC-MS. My name is Ed Connor,

More information

Running HPLC Methods on an ACQUITY UPLC System

Running HPLC Methods on an ACQUITY UPLC System Running HPLC Methods on an ACQUITY UPLC System Note: This document is an addendum to the ACQUITY UPLC System Operator s Guide. Contents Topic Page System considerations 1-1 Choosing fittings 1-3 System

More information

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist Thank you for purchasing an Agilent system. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is

More information

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS Peter Alden and Michael Jones Waters Corporation, Milford, MA, U.S. APPLICATION BENEFITS UPLC combines

More information

UltiMate 3000 Viper Capillary Kits for Intelligent LC (LCi) System Solutions

UltiMate 3000 Viper Capillary Kits for Intelligent LC (LCi) System Solutions UltiMate 3000 Viper Capillary Kits for Intelligent LC (LCi) System Solutions January 2010 Rev. 02 Table of Contents 1. Introduction... 1 1.1 How to Use this Manual... 1 1.2 Definition of LCi... 2 1.3

More information

AKTA pure 25 New Owner s Intro

AKTA pure 25 New Owner s Intro AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps

More information

Gradients in ion chromatography. Optimal separation with shorter analysis time

Gradients in ion chromatography. Optimal separation with shorter analysis time Gradients in ion chromatography Optimal separation with shorter analysis time Gradients yield precise results faster 02 The goal of ion chromatography in a nutshell reads: optimum separation within an

More information

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation User s Booklet for the Wyatt minidawn Light Scattering Instrumentation The Wyatt minidawn Light Scattering instrument is able to measure the weight average molecular weight of a synthetic polymer or a

More information

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve USER'S GUIDE Nanoflow Probe 3-4mm Red Stripe ptfe Sleeve To Injector ZU1XC Fused Silica 20µm 90µm TTF115 ptfe Sleeve Fused Silica 25µm 375µm Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: +44

More information

Dionex IonPac AS19-4µm

Dionex IonPac AS19-4µm Total Thermo Scientific Dionex IonPac AS19-4µm Column Product Manual P/N: March 2015 Part of Thermo Fisher Scientific Product Manual for Dionex IonPac AS19-4µm Column Dionex IonPac AG19-4µm Capillary Guard

More information

Sensitive Analysis of Genotoxins by HPLC-ECD Using Boron-Doped Diamond Electrochemical Cell

Sensitive Analysis of Genotoxins by HPLC-ECD Using Boron-Doped Diamond Electrochemical Cell Sensitive Analysis of Genotoxins by HPLC-ECD Using Boron-Doped Diamond Electrochemical Cell Bruce Bailey, Marc Plante, Christopher Crafts, Paul Gamache, John Waraska, and Ian Acworth ESA A Dionex Company,

More information

HPLC systems. Overview

HPLC systems. Overview HPLC systems Overview An hplc system is just a way to get a sample onto the end of a column, to pump solvents through the column until the analytes drop out the end, and to detect them as they appear.

More information

Customer Responsibilities

Customer Responsibilities Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Retention Time Locking: Concepts and Applications. Application

Retention Time Locking: Concepts and Applications. Application Retention Time Locking: Concepts and Applications Application Gas Chromatography December 1997 Authors Vince Giarrocco Bruce Quimby Matthew Klee Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Your local gas generation partner. Precision series Modular gas generation solution for GC. Your local gas generation partner Precision series Modular gas generation solution for GC www.peakscientific.com Perform with Precision Specifically designed and engineered for GC laboratory applications,

More information

DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178

DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178 174 DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS PAGE 175 APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 178 SEE STARRED PRODUCTS FOR TOP SELLERS! Series DEBUBBLERS & DEGASSERS

More information

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage: Research Article CODEN: IJRPJK ISSN: 2319 9563 International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage: www.ijrpns.com SIMULTANEOUS ESTIMATION AND VALIDATION OF TAMSULOSIN

More information

Rapid and Reliable Detection of Dissolved Gases in Water

Rapid and Reliable Detection of Dissolved Gases in Water Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,

More information

Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin

Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin HPLC Hints, Tips, Guidelines and Important Considerations Presenters: Wilhad Reuter / Tom Breslin Important HPLC Solvent Characteristics: ALL injected compounds/diluents must be soluble High purity and

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen Jim McCurry Agilent Technologies Wilmington, DE 18951 USA 1 Market Situation The world of He supply is not reliable, prices are increasing

More information

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15 Chemicals needed: Precipitate solution, Methanol Derivatization Solution, Amplifex Keto Reagent Kit Mobile phases, H 2 0 with 0.1% Formic Acid LC/MS grade and Acetonitrile with 0.1% Formic Acid LC/MS grade

More information

Next Generation Quartz Pressure Gauges

Next Generation Quartz Pressure Gauges Next Generation Quartz Pressure Gauges Rick Puccio, Ph.D. Senior Scientist Quartzdyne Houston, TX March 2, 2016 Outline Quartz Pressure Gauge Introduction Gauge Size Reduction Long Term Pressure Drift

More information

HPLC troubleshooting guide

HPLC troubleshooting guide CASE STUDY 70303 HPLC troubleshooting guide A free app that will save time and frustration Introduction During the development of a new HPLC method, there were unexpected results all peaks, except for

More information

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems Inorganic Analysis Chromatography Molecular Spectroscopy Thermal/Elemental Analysis Informatics Gas Chromatography MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass

More information

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Your local gas generation partner. Precision series Modular gas generation solution for GC. Your local gas generation partner Precision series Modular gas generation solution for GC www.peakscientific.com Perform with Precision Specifically designed and engineered for GC laboratory applications,

More information

Agilent G Rebuilt Electron Capture Detector

Agilent G Rebuilt Electron Capture Detector Agilent G1533-90790 Rebuilt Electron Capture Detector User Information Agilent Technologies Notices Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any form or by any means

More information

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone Development of a Clinical Research Method for the Measurement of Testosterone and Dihydrotestosterone Martijn van Faassen, Maarten van den Noort and Ido Kima University Medical Center, Groningen, Netherlands

More information

Using the Akta Prime plus October 22, 2012

Using the Akta Prime plus October 22, 2012 Some starting precautions: 1. Vacuum filter all buffers. Removes any large particles/debris that may clog your column De-gases the buffers 2. Clarify lysates first by centrifugation and then filtration

More information

May 1, By John Heim,Doug Staples

May 1, By John Heim,Doug Staples 1 of 7 7/11/2014 3:39 PM May 1, 2010 By John Heim,Doug Staples Anabolic steroid screening analysis in urine is complex and labor intensive requiring sensitive instrumentation and optimized chromatographic

More information

Introduction. Experimental

Introduction. Experimental PO-CON162E Development of Automated Screening and Quantitation Approach on Novel On-Line SFE-SFC-MS/MS Platform (I) For 23 Restricted Perflurocompounds in Textiles ASMS 2016 MP 283 Jie Xing, Jun Xiang

More information

DOS in Food Science & Nutrition, Manasagangotri, Mysore

DOS in Food Science & Nutrition, Manasagangotri, Mysore Sl No UNIVERSITY OF MYSORE DOS in Food Manasagangotri, Mysore 570 006. TENDERS FOR THE SUPPLY OF EQUIPMENTS, NETWORKING & COMPUTATIONAL FACILITIES FOR UNIVERSITY OF MYSORE (UNDER UGC XI SCHEME) INVITATION

More information

Agilent 1100 Series Analytical and Preparative Scale Fraction Collectors. User s Guide

Agilent 1100 Series Analytical and Preparative Scale Fraction Collectors. User s Guide Agilent 1100 Series Analytical and Preparative Scale Fraction Collectors User s Guide s1 Notices Agilent Technologies, Inc. 2001, 2003 No part of this manual may be reproduced in any form or by any means

More information

INSTRUMENT INSTRUMENT YL INSTRUMENT

INSTRUMENT INSTRUMENT YL INSTRUMENT INSTRUMENT INSTRUMENT INSTRUMENT Nitrogen Generator Whisper-N2 Series Features Membrane Technology Compressed air is forced through a hollow fiber membrane, with selective permeation of the different components

More information

Rev. CO Packed Columns for Normal Phase Chromatography. TSKgel NH m INSTRUCTION MANUAL

Rev. CO Packed Columns for Normal Phase Chromatography. TSKgel NH m INSTRUCTION MANUAL Rev. CO067000 Packed Columns for Normal Phase Chromatography TSKgel NH2-100 3 m INSTRUCTION MANUAL Safety Precautions To help protect you and/or your property from potential damage, please read this manual

More information

Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083

Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083 Liquid Chromatography Mass Spectrometry No. LCMS-083 Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS-8060 Summary ASTM methods D7968-14 and D7979-15 are used to determine

More information

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion: Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion

More information

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Quantitative Analysis of Hydrocarbons by Gas Chromatography Quantitative Analysis of Hydrocarbons by Gas Chromatography Introduction Gas-liquid chromatography (GLC) accomplishes a separation by partitioning solutes between a mobile gas phase and a stationary liquid

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

A VALIDATED STABILITY-INDICATING RP-HPLC ASSAY METHOD FOR BOLDENONE UNDECYLENATE AND ITS RELATED SUBSTANCES

A VALIDATED STABILITY-INDICATING RP-HPLC ASSAY METHOD FOR BOLDENONE UNDECYLENATE AND ITS RELATED SUBSTANCES Academic Sciences International Journal of Pharmacy and Pharmaceutical Sciences ISSN- 0975-1491 Vol 6 suppl 2, 2014 Research Article A VALIDATED STABILITY-INDICATING RP-PLC ASSAY METD FR BLDENNE UNDECYLENATE

More information

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer application Note Gas Chromatography/ Mass Spectrometry Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer Overview A mass spectrometer

More information

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide

More information

Agilent 1260 Infinity Purification Solution

Agilent 1260 Infinity Purification Solution Agilent 1260 Infinity Purification Solution System User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2014-2015, 2016 No part of this manual may be reproduced in any form or by any means

More information

Experiment GC : Analysis of BTEX by GC-FID

Experiment GC : Analysis of BTEX by GC-FID 46 Experiment GC : Analysis of BTEX by GC-FID Learning Goals: Familiarity with gas chromatography Gain experience in temperature programming and method development Correctly use an internal standard, and

More information

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port So, Why Do I Have To Do Maintenance? 1. Things wear out septa, syringes, nuts, ferrules, o-rings, etc.

More information

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE. Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to

More information

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS Application Note: Analysis of and Cyanuric Acid in Food Matrices by LC-MS/MS PeterVarelis, National Center for Food Safety and Technology, Illinois Institute of Technology. Jonathan Beck, Kefei Wang, and

More information

Customer Responsibilities

Customer Responsibilities Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites Application Note Doping Control Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites Steroid profiling, screening, and confirmatory analysis using the

More information

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis PO-CON695E High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis MSACL 6 EU Neil Loftus, Stephane Moreau, Mikaël Levi 3, Anja Grüning

More information

IJPAR Vol.6 Issue 1 Jan - Mar Journal Home page:

IJPAR Vol.6 Issue 1 Jan - Mar Journal Home page: IJPAR Vol.6 Issue 1 Jan - Mar -2017 Journal Home page: ISSN:2320-2831 Research article Open Access Optimization of stability indicating RP-HPLC method for the estimation of an anti-cancer drug Sorafenib

More information

Peak Biotech - Company & Product Presentation

Peak Biotech - Company & Product Presentation Peak Biotech - Company & Product Presentation July 2005 Peak Biotech Company & Product Presentation Peak Biotech Company Profile Location Jacob Nøhr Peak Biotech A/S was founded Kvistgaard, Denmark QA

More information

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD) Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet

More information

Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results

Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results Nitrogen Generator Whisper-N2 Series Features Membrane Technology Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results

More information