AUTOMATIC GAS MANIFOLDS

Size: px
Start display at page:

Download "AUTOMATIC GAS MANIFOLDS"

Transcription

1 AUTOMATIC GAS MANIFOLDS Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) Bird Hall Lane, Fax No.: 44 (0) Stockport, U.K., SK3 0SB

2 Contents Page Section Page Product Description General 3 Automatic Control Panel 3 Manifold Headers 3 Operation General 3 Control Panel Layout 4 Pneumatic 5 Electrical 6 Pressure Settings 6 Safety General 7 Installation General 7 Mechanical 7 Installation Drawings 8 Electrical 8 Electrical Schematic 9 Monitoring & Status Indicator Panel 10 Testing General 10 Testing 10 Commissioning 11 Maintenance General 11 Preventative Maintenance 11 Spare Parts General Automatic Gas Manifolds Rev.3

3 1. Product Description 1.1 General Automatic Gas Manifolds comprise a control panel, manifold headers & racks, tailpipes, pigtails, an isolation valve, a non-return valve and a pressure relief valve. The manifold provides a continuous supply of gas from two banks of cylinders to the medical gas pipeline system by changing from the duty bank to the standby bank automatically when the duty bank has become depleted. Manifolds are available for oxygen, nitrous oxide, 50% oxygen/50% nitrous oxide mixture, medical air and surgical air. 1.2 Automatic Control Panel Automatic control panels comprise of regulators, changeover solenoid valves, pressure switches, gauges and an integral monitoring and status panel, all housed in a moulded enclosure with transparent cover. The monitoring & status panel indicates which bank of cylinders is running, which is empty and which is running low if the other bank is empty. The panel will also indicate if the distribution pipeline pressure has risen or fallen beyond acceptable limits. The display incorporates an alarm status indicator and includes volt free contacts for connection to the central alarm system and BMS if required. An isolation and pressure relief valve assembly is provided with copper stub pipes for on site copper to copper joints to the hospital distribution pipeline and exhaust pipeline. 1.3 Manifold Headers Manifold headers comprise of high-pressure headers with gas specific tailpipe connections, cylinder tailpipes, support racks and restraining chains for the cylinders. A non return valve is incorporated into each point where the tailpipe connects into the header to prevent a whole bank of cylinders becoming depleted in the event of a cylinder being disconnected or a tailpipe rupture. Tailpipes are available with pin-index, bull nose or CGA cylinder connections. 2. Operation 2.1 General All cylinders connected to the manifold headers should be open. Gas is from the left hand and right hand banks of cylinders to the automatic control panel. Under normal working conditions, gas is supplied to the hospital distribution pipeline from only one bank at a time. When the cylinder contents falls below the changeover pressure, the solenoid valve for the duty bank closes and the solenoid valve for the standby bank opens ensuring continuity of the gas supply. In the event of an electrical failure, both solenoid valves open and gas will be supplied from the duty and standby banks simultaneously. Upon restoration of the electrical supply, the control panel will supply gas from the bank of cylinders that was running immediately prior to the loss of power. Both solenoid valves will also open in the event of the distribution pipeline pressure falling to a dangerously low level. All regulators and pressure sensors within the control panel are preset and should require no further adjustment. The monitoring & status panel incorporates a duty bank push button to enable manual selection of the left or right hand cylinders as the duty bank. The manifold is fully automatic in operation and requires no manual intervention, other than the replace empty cylinders Automatic Gas Manifolds Rev.3

4 2.2 Control Panel Layout ID Description Qty ID Description Qty 1 Isolation Valve 1 12 Backplate 1 2 Pressure Relief Valve 1 13 Monitoring & Status Panel 1 3 Outlet Stub Pipe 1 14 Regulator 2 4 Solenoid Valve 2 15 Regulator Mounting Block 2 5 Pressure Switch, High 1 16 Header Inlet Connector 2 6 Pressure Switch, Low 1 17 Contents Pressure Switch 2 7 Override Pressure Switch 1 18 LH Regulator Outlet Pipe 1 8 Line Pressure Gauge 1 19 RH Regulator Outlet Pipe 1 9 Outlet Block 1 20 Contents Pressure Gauge 2 10 Earth Stud 1 21 Outlet Pressure Gauge 2 11 ST18 Connector (Power) 1 22 Enclosure Automatic Gas Manifolds Rev.3

5 2.3 Pneumatic Operation The gas from the left hand (LH) bank of cylinders is connected to the LH inlet of the control panel. This is connected to gauge G1, pressure switch P1 and passes through regulator R1, pressure gauge G3 to the solenoid valve S1. Similarly, the gas from the right hand (RH) bank of cylinders is connected to the RH inlet of the control panel. This is connected to gauge G2, pressure switch P2 and passes through regulator R2, pressure gauge G4 to the solenoid valve S2. Pressure gauges G1 and G2 indicate the cylinder contents pressure in the LH and RH banks respectively. Pressure gauges G3 and G4 indicate the outlet pressure from regulators R1 and R2 respectively. With the LH bank selected as the duty bank, the solenoid valve S1 is de-energised and opens allowing gas to pass to the distribution pipeline. Solenoid valve S2 is energised and closes to prevent the RH from supplying gas. When the pressure in the LH bank falls below the setting of pressure switch P1, the solenoid valve S1 is energised and closes, whilst the solenoid valve S2 is de-energised and opens thus effecting the changeover from the LH duty bank to the RH standby bank. The empty cylinders on the LH bank should now be replaced with full cylinders, ready to continue the supply of gas when the RH bank is depleted. When the pressure in the RH bank falls below the setting of pressure switch P2, the solenoid valve S2 is energised and closes and solenoid valve S1 is de-energised and opens thus allowing the LH bank (now full) to continue with the gas supply. The empty cylinders on the RH bank should now be replaced with full cylinders. This sequence will continue. Pressure gauge G5 indicates the distribution pipeline pressure, whilst pressure switches P3 and P4 monitor the distribution pipeline pressure for high or low pressure faults respectively. Pressure switch P5 will de-energise both solenoid valves S1 and S2 in the event of the distribution pipeline pressure falling to a dangerously low level allowing both banks to feed the system simultaneously Automatic Gas Manifolds Rev.3

6 2.4 Electrical Operation The control panel requires a V, 50/60Hz, single-phase supply. The monitoring & status panel incorporates two sections, one labeled Manifold Indicator Unit and one labeled Alarm Signal Status Unit. The manifold indicator unit displays the status of the cylinder banks and the distribution pipeline pressure, whilst the alarm signal status unit displays the main central alarm system conditions. The Manifold Status Unit includes the following indicators. (a) For each bank, a green LED indicator (RUNNING) illuminates to display which bank is currently running. (b) For each bank, a yellow LED indicator (LOW PRESSURE) illuminates to display that the bank that is running has fallen to the low pressure setting and the standby bank is still empty. (c) For each bank, a yellow LED indicator (EMPTY) illuminates to display that a bank of cylinders is empty and changeover to the standby bank has occurred. (d) A red LED indicator (HIGH PRESSURE) illuminates to display when the distribution pipeline pressure has risen above the pressure sensor setting. (e) A red LED indicator (LOW PRESSURE) illuminates to display when the distribution pipeline pressure has fallen below the pressure sensor setting. The Alarm Signal Status Unit includes the following indicators. (a) A green LED indicator (POWER ON) illuminates to show that the power to the manifold is on. (b) A green LED indictor (NORMAL) that illuminates when the manifold is operating correctly and no faults exist. (c) A yellow LED indicator (CHANGE CYLINDERS) which illuminates when changeover from the duty to the standby bank has occurred. (d) A yellow LED indicator (CHANGE CYLINDERS IMMEDIATELY) which illuminates when the bank that is running has fallen below the pressure sensor setting and the standby bank is empty. (e) A red LED indicator (RESERVE LOW) which illuminates when the duty bank of the emergency reserve manifold (ESM) has fallen below the low pressure setting. (f) A red LED indicator (PRESSURE FAULT) which illuminates when the distribution pipeline pressure has risen or fallen beyond the respective pressure sensor settings. (g) A red LED indicator (SYSTEM FAULT) which illuminates in the event of a cabling fault. 2.5 Pressure Settings Component O2 N2O O2/N2O Medical Air (400 kpa) Surgical Air (700 kpa) Regulators (static) 4.4 bar 4.4 bar 4.4 bar 4.0 bar 7.5 bar Changeover Pressure 14.0 bar 14.0 bar 14.0 bar 14.0 bar 14.0 bar Pressure Relief Valve 5.5 bar 5.5 bar 5.5 bar 5.5 bar 9.0 bar Pipeline Pressure, High 5.0 bar 5.0 bar 5.0 bar 5.0 bar 8.0 bar Pipeline Pressure, Low 3.7 bar 3.7 bar 3.7 bar 3.7 bar 6.5 bar Override Switch 3.9 bar 3.9 bar 3.9 bar 3.9 bar 6.7 bar Automatic Gas Manifolds Rev.3

7 3. Safety 3.1 General This equipment should be installed, operated and maintained by personnel who are suitably trained, are fully conversant with HTM 2022 and are familiar with this product. This equipment should be kept clean and be free from oil and grease at all times. Oil and grease will ignite spontaneously in the presence of oxygen. If you suspect that any equipment is contaminated, do not use it. No attempt should be made to use or modify this equipment for use with a gas other than as identified. This equipment should not be operated at pressures exceeding those stated in HTM 2022 and this manual. Care should be taken when drawing conclusions about the contents of cylinders from their pressures. In the case of compressed gases, such as oxygen and medical air, the cylinder pressure is a good indication of the cylinder contents, for example, if the cylinder is half full, it will show half pressure. Liquefied gases, such as nitrous oxide, show full cylinder pressure as long as there is liquid in the cylinder. This may represent a small percentage of the full cylinder contents. It should not be assumed that a cylinder of nitrous oxide is full simply because the gauge shows full pressure. 4. Installation 4.1 General The automatic control panel is supplied with all internal electrical connections prewired. The only external wiring required for the installation is the connection of the mains power supply, the connection of the ESM pressure switch (if required) and the connection of the central alarm system and BMS wiring (if required). All regulators and pressure sensors are preset and no further adjustment of these settings should be required. 4.2 Mechanical (a) Remove the backplate assembly from the automatic control panel enclosure and store in a safe place. The enclosure will be easier to handle and fix. (b) Secure the automatic control panel enclosure to the wall using suitable fixings at the height indicated in figure 4.3. Ensure that there is sufficient space either side of the control panel to install the manifold headers. (c) Replace the backplate assembly and secure with the six retaining screws. (d) Secure the manifold headers at the height and position indicated in figure 4.3 using suitable fixings. Use the header connecting pipes to assist in the accurate alignment of the racks. (e) Install the outlet stub pipe, pressure relief valve and isolation valve. (f) Install the exhaust pipeline from the pressure relief valve to a safe external location. (g) Install all tailpipes onto the manifold headers. (h) Check and tighten all mechanical joints. (i) Connect cylinders, pressurise the manifold and check for leaks. Make sure that the isolation valve is closed if the connection to the distribution pipeline has been made Automatic Gas Manifolds Rev.3

8 4.3 Installation Drawings 4.4 Electrical (a) Connect the mains power supply to the male ST18 connector mounted on the backplate using the female ST18 connector provided. The cable should be connected as follows; Red or Brown wire to the LIVE (S+) terminal. Blue or Black wire to the NEUTRAL (N) terminal. Green/Yellow wire to the EARTH terminal. The mains power supply should be via an unswitched fused spur rated at 5 amps Automatic Gas Manifolds Rev.3

9 (b) Connect the emergency standby manifold (ESM) pressure switch into terminals 1 and 2 of CON1 on the monitoring and status panel PCB. Refer to figure 4.5. If no ESM is available, ensure that a 3K3Ω resistor (supplied) is fitted into these terminals to prevent a system fault from illuminating. (c) Connect the alarm conditions displayed on the alarm signal status unit to the central alarm system by connecting to CON1. Refer to figure 4.5 for details of each alarm conditions position in the 12-way connector. (d) Connect the alarm conditions to the building management system (BMS) by connecting to CON2. All alarm conditions have one common (COM) terminal, a normally closed (NC) terminal and a normally open (NO) terminal. Refer to figure 4.5 for details of each alarm conditions position in the 12-way connector. 4.5 Electrical Connections Diagram Automatic Gas Manifolds Rev.3

10 4.7 Monitoring & Status Indicator Panel Phoenix Pipeline Product Oxygen 5. Testing 5.1 General Prior to testing the manifold installation, please check the following. (a) All components have been installed are tightened. (b) The mains power supply has been installed and power is available. (c) The isolation valves on the automatic manifold and ESM are closed. (d) Full cylinders have been fitted to all tailpipes. 5.2 Testing (a) Slowly open one cylinder on each bank and allow the system to pressurise. (b) Check all tailpipes and header interconnecting pipes for leaks. (c) Slowly open one cylinder on the ESM (if available) and check for leaks. (d) Apply mains power to the panel. The LH bank RUNNING indicator will illuminate, the POWER ON indicator will illuminate and the NORMAL indicator will illuminate. (e) Close the cylinder on the LH bank. Create a leak between the control panel and the distribution pipeline and very slowly depressurise the system. (f) At changeover pressure (refer to the settings in table 2.5), the LH bank RUNNING indicator will extinguish and the EMPTY indicator will illuminate. The RH bank RUNNING indicator simultaneously illuminates as the control panel changes to run from the RH bank. The NORMAL indicator on the alarm signal status unit will extinguish and the CHANGE CYLINDERS will illuminate. (g) Open the cylinder again on the LH bank. The CHANGE CYLINDERS indicator will extinguish and the NORMAL indicator will illuminate. The LH bank EMPTY indicator will extinguish. (h) Close the cylinder on the RH bank. Create a leak again and slowly depressurise the system Automatic Gas Manifolds Rev.3

11 (i) At changeover pressure, the RH bank RUNNING indicator will extinguish and the EMPTY indicator will illuminate. The LH bank RUNNING indicator will illuminate. The NORMAL indicator will extinguish and the CHANGE CYLINDERS indicator will illuminate. (j) Close the cylinder on the LH bank. Create a leak and slowly depressurise the system. (k) At changeover pressure, the LH bank RUNNING indicator will extinguish and the LOW PRESSURE indicator will illuminate. Simultaneously, the CHANGE CYLINDERS IMMEDIATELY indicator will illuminate. (l) Close the cylinder on the ESM (if available). Open the isolation valve on the ESM. Continue to depressurise the system. At the setting of the ESM pressure switch, the RESERVE LOW indicator will illuminate. (m) Continue to depressurise the system. When the pressure falls to the setting of the low line pressure switch, the LOW PRESSURE indicator on the manifold indicator unit will illuminate and the PRESSURE FAULT indicator on the alarm signal status unit will illuminate. (n) Close all leaks and isolation valves and re-pressurise the system. All alarm signal indicators will extinguish and the NORMAL indicator will illuminate. The LH bank LOW PRESSURE indicator will extinguish and the RUNNING indicator will illuminate. The RH bank EMPTY indicator will extinguish. (o) Check all alarm system outputs using an ohmmeter. All outputs are normally closed and as each alarm condition occurs, the respective contact opens circuit. (p) If any pressure settings differ from those as stated in table 2.5, adjust as necessary. 5.3 Commissioning Demonstrate the manifold installation to the client by repeating the abovementioned test procedure. It is advisable to confirm that a central alarm system (if available) is operating correctly at the same time. The manifold must not be used until all testing & commissioning procedures for the pipeline system as detailed in HTM 2022 have been satisfactorily completed and accepted. 6. Maintenance 6.1 General Maintenance of the manifold should be restricted to periodic checking and adjustment and if necessary the replacement of faulty components. It should be noted that although many of the components appear to be standard items, many have been selected and treated to make them suitable for the gases carried and the pressures involved. 6.2 Preventative Maintenance Regular inspections and maintenance of the manifold will prolong it s life and reduce the possibility of sudden, inconvenient component failures. Automatic Manifolds should be subjected to regular inspection and testing as detailed below. Monthly; (a) Visually inspect the manifold for signs of damage. (b) Check all mechanical joints for leaks. (c) Check that changeover from the duty bank to the standby bank is operating correctly Automatic Gas Manifolds Rev.3

12 Annually; (a) Visually inspect the manifold system for signs of damage. (b) Check that one bank of cylinders on the ESM is open ready for use. (c) Check all mechanical joints for leaks. (d) Test the manifold system as detailed in section 5.2 to confirm correct operation. (e) Observe that all indicators are working correctly. 6.3 All maintenance should be carried out with the knowledge of the hospital engineer and in accordance with the Permit-to-Work system. 7. Spare Parts When ordering spare parts, please quote the batch number of the equipment and a description of the component required to ensure that you receive the component that you require! Item Description Part Number 1 Tailpipe, O2, Pin-index Tailpipe, O2, Bull-nose Tailpipe, O2, CGA Tailpipe, N2O, BS 341 No Tailpipe, N2O, CGA Tailpipe, 50% O2/50% N2O, Pin-index Tailpipe, Medical Air, Pin-index Tailpipe, Medical Air, Bull-nose Tailpipe, Medical Air, CGA Pressure Relief Valve, 5.5 Bar Pressure Relief Valve, 9.0 Bar Regulator Assembly, 4 Bar N2O Regulator Assembly, 4 Bar Regulator Assembly, 7 Bar Contents Pressure Switch, 14 bar Contents Pressure Gauge, Bar Contents Pressure Gauge, Bar Outlet Pressure Gauge, 0 6 Bar Outlet Pressure Gauge, 0 16 Bar Bodok Seal for pin-index tailpipes O-ring for bull-nose tailpipes Solenoid Valve Automatic Gas Manifolds Rev.3

13 Item Description Part Number 23 Pipeline Pressure Switch, High, 5.0 Bar Pipeline Pressure Switch, High, 8.0 Bar Pipeline Pressure Switch, Low, 3.7 Bar Pipeline Pressure Switch, Low, 6.5 Bar Override Switch, 3.9 Bar Override Switch, 6.7 Bar Line Pressure Gauge, 0-10 Bar Automatic Manifold PCB Manifold Header Link Pipe c/w o-rings Header O-ring Service Kit (per cylinder) Auto Control Panel Seal Service Kit Tailpipe, Medical Air, CGA Due to our policy of continuous product improvement, Phoenix Pipeline Products Ltd. reserve the right to alter dimensions and/or specification without notice although every effort will be made to advise of any such modifications. Certain aspects of P³ Ltd. designs are protected Automatic Gas Manifolds Rev.3

MANUAL GAS MANIFOLDS

MANUAL GAS MANIFOLDS MANUAL GAS MANIFOLDS Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com

More information

DUPLEX PRESSURE REDUCING SET

DUPLEX PRESSURE REDUCING SET DUPLEX PRESSURE REDUCING SET Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com

More information

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855

More information

AREA VALVE SERVICE UNITS

AREA VALVE SERVICE UNITS AREA VALVE SERVICE UNITS. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com U.K., SK3 0SB www.p3-phoenix.com

More information

Manifolds & pressure reducing sets

Manifolds & pressure reducing sets Manifolds & pressure reducing sets Automatic Changeover Manifold (Electronic) Providing continuous supply of medical gases Pneumatech Medical Gas Solutions (PMGS) PureGAS Manifolds provide a duty or standby

More information

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855

More information

Specialists in HTM Medical Gas Pipeline Equipment AVSU/Module Operating and Maintenance Instructions

Specialists in HTM Medical Gas Pipeline Equipment AVSU/Module Operating and Maintenance Instructions The Area Value Service Units and Module format (AVSU) provide a local gas isolation facility for use during normal installation and maintenance or in the event of an emergency. They are built in accordance

More information

PureGAS Manifold Control Panel - HTM02-01

PureGAS Manifold Control Panel - HTM02-01 0088 PureGAS Manifold Control Panel - HTM02-01 Manifold Systems Operation and Maintenance Manual Published by Pneumatech Medical Gas Solutions All possible care has been taken in the preparation of this

More information

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units Models: EPS Single Pump EPT Twin Pump EPS-HP EPT-HP Single Pump High Pressure Twin Pump High Pressure

More information

Budget Range Operators Handbook

Budget Range Operators Handbook Budget Range Operators Handbook BAMBI AIR COMPRESSORS LTD 152 Thimble Mill Lane Heartlands Birmingham B7 5HT United Kingdom Tel: 0121 322 2299 Fax: 0121 322 2297 Email: sales@bambi-air.co.uk www.bambi-air.co.uk

More information

Pressure Differential Automatic Manifold For LPG. PDH1000 Series. LPG Installation & Operations Instructions

Pressure Differential Automatic Manifold For LPG. PDH1000 Series. LPG Installation & Operations Instructions Pressure Differential Automatic Manifold For LPG PDH1000 Series LPG Installation & Operations Instructions Precise Equipment Company 3822 Market St., STE 101 Denton, Texas Phone : 800.795.8388 Precise

More information

Installation & Maintenance Manual. Economy Manifold

Installation & Maintenance Manual. Economy Manifold Installation & Maintenance Manual Economy Manifold Contents User Responsibility 3 Introduction 4 Features 4 Description of the Manifold 5 Shipment Details 5 Description of Parts 5-6 Pressure Regulators

More information

ECONORESS ELECTRONIC EPS & EPT - ENHANCED PRESSURISATION SET INSTALLATION OPERATION & MAINTENANCE DOCUMENTATION

ECONORESS ELECTRONIC EPS & EPT - ENHANCED PRESSURISATION SET INSTALLATION OPERATION & MAINTENANCE DOCUMENTATION ECONORESS ELECTRONIC EPS & EPT - ENHANCED PRESSURISATION SET INSTALLATION OPERATION & MAINTENANCE DOCUMENTATION OCT2010 STOKVIS ENERGY SYSTEMS 96R WALTON ROAD EAST MOLESEY SURREY KT8 0DL TEL: 020 87833050

More information

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series Introduction Powerex manifolds are cleaned, tested and prepared for the indicated gas service and are built in accordance with the Compressed Gas Association guidelines. The manifold consists of a regulator

More information

Installation Operating Instructions for Duplex Manual Manifolds PX-TMD Series

Installation Operating Instructions for Duplex Manual Manifolds PX-TMD Series Introduction Powerex manifolds are cleaned, tested and prepared for the indicated gas service and are built in accordance with the Compressed Gas Association guidelines. The manifold consists of a regulator

More information

System Pressure Manager Standard & System Pressure Manager Plus

System Pressure Manager Standard & System Pressure Manager Plus System Pressure Manager Standard & System Pressure Manager Plus Installation, Commissioning & Servicing Instructions Note: THESE INSTRUCTIONS MUST BE READ AND UNDERSTOOD BEFORE INSTALLING, COMMISSIONING,

More information

Installation & Operating Instructions for Genesys Series Manifolds

Installation & Operating Instructions for Genesys Series Manifolds Manufacturer of Medical Gas Pipeline Equipment Installation & Operating Instructions for Genesys Series Manifolds Model LLU Model CCU - 1 - Features & Benefits Features & Benefits Fully automatic changeover

More information

Installation & Operating Instructions for Genesys Analog Series Manifolds

Installation & Operating Instructions for Genesys Analog Series Manifolds Manufacturer of Medical Gas Piping Equipment Installation & Operating Instructions for Genesys Analog Series Manifolds Models NPCU & TMCU Model TMLU - 1 - Genesys Analog Manifold Systems Features & Benefits

More information

Tri-Tech Medical Inc.

Tri-Tech Medical Inc. Introduction Tri-Tech Medical manifolds are cleaned, tested and prepared for the indicated gas service and are built in accordance with the Compressed Gas Association guidelines. The manifold consists

More information

Economy Manifold Alert -2. Installation and Maintenance Manual V1.6

Economy Manifold Alert -2. Installation and Maintenance Manual V1.6 Economy Manifold Alert -2 Installation and Maintenance Manual V1.6 Corporation Table of Contents User Responsibility 4 Introduction 4 Features 4 Description of the Manifold 5 Shipment Details 5 The Collapsible

More information

SERIES-O REGULATOR. Instructions for use

SERIES-O REGULATOR. Instructions for use SERIES-O REGULATOR Instructions for use 522460-05-708 Contents Page 1. INTRODUCTION 2 2. INTENDED USES 3 3. GENERAL DESCRIPTION 4 4. PARTS IDENTIFICATION 5 5. SAFETY PRECAUTIONS 6 6. EQUIPMENT USE 8 7.

More information

Ventam 85 Installation & Commissioning Instructions

Ventam 85 Installation & Commissioning Instructions Ventam Systems Ltd Unit D4 Seedbed Business Centre Vanguard Way Shoeburyness Essex SS3 9QY Phone 01702 382 307 Fax 01702 382 340 Ventam 85 Installation & Commissioning Instructions 1 General The Ventam

More information

Installation & Maintenance Manual. C2 & C3 Manifolds

Installation & Maintenance Manual. C2 & C3 Manifolds Installation & Maintenance Manual C2 & C3 Manifolds w w w. a m i c o. c o m Table of Contents User Responsibility 3 Introduction 4 Features 4 Description of the Manifold 5 Shipment Details 5 Description

More information

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units Digital Range

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units Digital Range Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units Digital Range Models: DS 126 Single Pump / Single System DT 126 Twin Pump / Single System DS 160

More information

Pressure Regulators. Instructions for Use

Pressure Regulators. Instructions for Use Pressure Regulators Instructions for Use 702-0032.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

Dialflow Regulator. Instructions for Use

Dialflow Regulator. Instructions for Use Dialflow Regulator Instructions for Use 702-0030.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

MEDICAL EQUIPMENT CATALOG

MEDICAL EQUIPMENT CATALOG THE MOST TRUSTED MEDICAL EQUIPMENT IN NORTH AMERICA MEDICAL EQUIPMENT CATALOG Total Components Including Bulk Oxygen Supply, Back up and Delivery Equipment Designs for High Pressure, Liquid or Bulk Applications

More information

MD R a n g e O p e r a t o r s H a n d b o o k

MD R a n g e O p e r a t o r s H a n d b o o k MD R a n g e O p e r a t o r s H a n d b o o k Covering Models:- MD35/20 MD75/80 MD75/0 MD75/0V MD75/250 MD75/250V MD0/500 MD225/1000 BAMBI AIR COMPRESSORS LTD 2 Thimble Mill Lane Heartlands Birmingham

More information

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves

More information

Installation & Operating Instructions for Genesys Analog Series Manifolds

Installation & Operating Instructions for Genesys Analog Series Manifolds Manufacturer of Medical Gas Piping Equipment Installation & Operating Instructions for Genesys Analog Series Manifolds Models NPC & TMC Model TML e6388brd 5/04/18 Tri-Tech, 35401 Avon Commerce Pkwy., Avon,

More information

571 Series Medical TM. IntelliSwitch Electronic Switchover

571 Series Medical TM. IntelliSwitch Electronic Switchover ADI 9505-H 571 Series Medical TM IntelliSwitch Electronic Switchover INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Carefully Read These Controls Corporation

More information

Technical Specification

Technical Specification Medical Gas Terminal Units (Gas Outlet Points) Technical Specification EN 9170-1 Terminal Unit Product Description CPX Terminal Units comprise a first fix, second fix and check valve. The first fix assembly

More information

Pressure Regulator. Instructions for Use

Pressure Regulator. Instructions for Use Pressure Regulator Instructions for Use 702-0083.9 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient, the

More information

Manual Actuated Boiler Blowdown Valves

Manual Actuated Boiler Blowdown Valves Manual Actuated Boiler Blowdown Valves Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Operation 5. Maintenance 6. Spare parts p.1 1. Safety

More information

Protocol Switchover Station

Protocol Switchover Station ADI 3215-B Certified ISO 9001 Protocol Switchover Station INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls Corporation of America

More information

Fully Automatic Gas Manifold. Patent US 7,013,906. Installation & Operation Manual

Fully Automatic Gas Manifold. Patent US 7,013,906. Installation & Operation Manual Fully Automatic Gas Manifold Patent US 7,013,906 Installation & Operation Manual P/N 255491 (Rev.1) 11/2009 FOR YOUR SAFETY: Installation and utilization of manifold and piping systems must conform to

More information

AutoChanger Installation & User Guide Issue 2

AutoChanger Installation & User Guide Issue 2 1 INDEX Page 1. Introduction... 3 2. AutoChanger Components Guide 4 3. Installation. 5-11 a) Installation Guidelines.. 5 b) Installation Retrofit... 6-11 c) Installation New... 11 4. AutoChanger User Guide

More information

AIR COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: CHAMP 3 PART NO: LS0115

AIR COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: CHAMP 3 PART NO: LS0115 AIR COMPRESSOR MODEL NO: CHAMP 3 PART NO: 2225222 OPERATION & MAINTENANCE INSTRUCTIONS LS0115 INTRODUCTION Thank you for purchasing this CLARKE Air Compressor. Please read this manual fully before use

More information

Operating Instructions

Operating Instructions Operating Instructions ProMinent CO2 Feeder CO2 - ProMinent P/N: 7746944 (Revision C), Date 09-14-12 Please completely read through these operating instructions first! Do not discard! The warranty shall

More information

CENTRAL GAS SUPPLY SYSTEMS : CYLINDER BATTERIES AND HIGH PRESSURE HOSES

CENTRAL GAS SUPPLY SYSTEMS : CYLINDER BATTERIES AND HIGH PRESSURE HOSES P0170E-1.DOCX/08.16/MG CENTRAL GAS SUPPLY SYSTEMS : CYLINDER BATTERIES AND HIGH PRESSURE HOSES With a Central Gas Supply System, individual work places or gas take-off stations are supplied with gas through

More information

Flamco Flamco-Fill STA Flamco-Fill STM

Flamco Flamco-Fill STA Flamco-Fill STM STA STM Typ STA 5936 Typ STM 5937 GB Installation and operating instructions 1999, 2 Contents GB UK Limited P.O. Box 9, Washway Lane, St. Helens, Merseyside WA10 6FE United Kingdom Telephone: 01744 744744

More information

ServicePlus S21R & S27R Series 2 Installation Instructions

ServicePlus S21R & S27R Series 2 Installation Instructions ServicePlus S21R & S27R Series 2 Installation Instructions Horstmann Controls ServicePlus two channel service interval programmers have been developed to facilitate compliance with the Gas Safety (Installation

More information

TABLE OF CONTENTS TECHNICAL DOCUMENTATION 03

TABLE OF CONTENTS TECHNICAL DOCUMENTATION 03 INSPIRED EF F ICIENCY PRESSURISATION UNITS MAXI P R ESSUR I SAT I O N U NI T S T E CH N I CAL D O CUMEN T A T I O N I SSUE 1 0 /1 7 R e v. 0 1 TABLE OF CONTENTS 1. 2. 3. 4. 5. 6. 7. 8. 9. Technical Data

More information

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units Micro-S / Micro-S Digital Wall Hung Unit Please fulfil all listed requirements prior to and during

More information

USER S MANUAL BASIC HORIZONTAL OIL-LUBRICATED MICROCOMPRESSOR Reference: RCMPBH

USER S MANUAL BASIC HORIZONTAL OIL-LUBRICATED MICROCOMPRESSOR Reference: RCMPBH USER S MANUAL BASIC HORIZONTAL OIL-LUBRICATED MICROCOMPRESSOR Reference: RCMPBH THIS MANUAL IS INTENDED FOR TECHNICAL STAFF IN CHARGE OF THE INSTALLATION, THE OPERATION AND THE MAINTENANCE OF THIS PRODUCT

More information

SEMEM3-73 Maintaining medical gas pipeline systems and equipment

SEMEM3-73 Maintaining medical gas pipeline systems and equipment Maintaining medical gas pipeline systems and equipment Overview This unit identifies the competences you need to carry out maintenance activities on medical gas pipeline systems and equipment, in accordance

More information

24L OIL FREE AIR COMPRESSOR MODEL NO: TIGER 7/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS10/13

24L OIL FREE AIR COMPRESSOR MODEL NO: TIGER 7/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS10/13 24L OIL FREE AIR COMPRESSOR MODEL NO: TIGER 7/250 PART NO: 2244030 OPERATION & MAINTENANCE INSTRUCTIONS LS10/13 INTRODUCTION Thank you for purchasing this product. Before attempting to use this product,

More information

Technical Datasheet. Rigid & Retractable Pendants. Rigid Pendant

Technical Datasheet. Rigid & Retractable Pendants. Rigid Pendant Rigid & Retractable Pendants Rigid Pendant Description Rigid pendants are intended for use operating theatres, anaesthetic rooms and similar medical locations. They are an ergonomic solution for the provision

More information

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

MEDICAL REGULATOR RANGE (MEDIREG RANGE) MEDICAL REGULATOR RANGE (MEDIREG RANGE) Medireg Dialreg Flowreg INSPECTION Remove the Medireg from the packaging and inspect for damage. If there is any damage, DO NOT USE, and contact Therapy Equipment

More information

Electronic Single and Twin Pump System Pressure Managers

Electronic Single and Twin Pump System Pressure Managers Electronic Single and Twin Pump System Pressure Managers Installation, Commissioning & Servicing Instructions Note: THESE INSTRUCTIONS MUST BE READ AND UNDERSTOOD BEFORE INSTALLING, COMMISSIONING, OPERATING

More information

Spirax Compact FREME Flash Recovery Energy Management Equipment

Spirax Compact FREME Flash Recovery Energy Management Equipment IM-UK-cFREME UK Issue 1 Spirax Compact FREME Flash Recovery Energy Management Equipment Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4.

More information

EDUCATION DEPARTMENT ACCU-TEST

EDUCATION DEPARTMENT ACCU-TEST ABN 17 100 208 964 EDUCATION DEPARTMENT ACCU-TEST GAS SAFETY SHUT OFF SYSTEM Operating & Installation Manual Head Office (Melbourne) Sydney Brisbane (Distributor) 1/5 Samantha Court 2/14 Welder Road 17

More information

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Installation, operating and maintenance Instructions for Seemag bypass level indicator Issue: S Date: 05-09-14 Type G35 General information The Seetru bypass magnetic level indicator, abbreviate SEEMAG, serves to show the filling level of fluids in tanks, basins, tubes etc. The Seemag operates

More information

NGP-250/500 Nitrogen Generator Quick Start Guide

NGP-250/500 Nitrogen Generator Quick Start Guide NGP-250/500 Nitrogen Generator Quick Start Guide Version: A July 2013 Potter Electric Signal Company, LLC 5757 Phantom Dr., Suite 125 P. O. Box 42037 Hazelwood, MO 63042 Phone: (314) 595-6900 Document

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE15C150, VE18C150, VE25C150 PART NO: 2226010, 2226020, 2226025 OPERATION & MAINTENANCE INSTRUCTIONS LS0715 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

User manual. ZX trolley.

User manual. ZX trolley. User manual. ZX trolley. Date July 2011 02 User manual. ZX trolley. ZX trolley. Safe transportation for medical gas cylinders. Important Please read this manual carefully before attempting to use the BOC

More information

CS150 CAP STAPLER OWNER S MANUAL

CS150 CAP STAPLER OWNER S MANUAL Operation Revised 6/2013 www.stingerworld.com CS150 CAP STAPLER OWNER S MANUAL! Maintenance Safety Warranty PLEASE READ! This manual contains important information about product safety. WELCOME TO STINGER

More information

R E D I C O N T R O L S

R E D I C O N T R O L S R E D I C O N T R O L S Operation & Maintenance Manual Portable Service Purger for Low Pressure Chillers Model: PSP-LP-1B For Refrigerants R-11, R-113, R-114 & R-123 & Other Similar Refrigerants File Literature

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

24L AIR COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: RANGER 7/240 PART NO: LS0913

24L AIR COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: RANGER 7/240 PART NO: LS0913 24L AIR COMPRESSOR MODEL NO: RANGER 7/240 PART NO: 2242000 OPERATION & MAINTENANCE INSTRUCTIONS LS0913 INTRODUCTION Thank you for purchasing this CLARKE 24L Air Compressor. Please read this manual fully

More information

TECHNICAL MANUAL HEAVY DUTY HORIZONTAL MICRO-COMPRESSOR Reference: RCMPH

TECHNICAL MANUAL HEAVY DUTY HORIZONTAL MICRO-COMPRESSOR Reference: RCMPH TECHNICAL MANUAL HEAVY DUTY HORIZONTAL MICRO-COMPRESSOR Reference: RCMPH THIS MANUAL IS INTENDED FOR TECHNICAL STAFF IN CHARGE OF THE INSTALLATION, THE OPERATION AND THE MAINTENANCE OF THIS PRODUCT CAUTION!

More information

Note: You will need a full tank of CO2 for connection and leak testing! Kit Number

Note: You will need a full tank of CO2 for connection and leak testing! Kit Number Owners Manual Design Engineering Inc. CryO2 Cryogenic Tank Installation Kit To be used with CryO2 components only! Persons experienced in the installation and proper operation of Performance systems like

More information

OPERATING INSTRUCTIONS AND PARTS LIST AUTOMATIC FILLING MACHINE DE-401-GB-AUTO

OPERATING INSTRUCTIONS AND PARTS LIST AUTOMATIC FILLING MACHINE DE-401-GB-AUTO OPERATING INSTRUCTIONS AND PARTS LIST AUTOMATIC FILLING MACHINE DE-401-GB-AUTO www.adceng.co.za P O BOX 5413, JOHANNESBURG 2000 Phone: 011 493-8209 Fax: 011 493-1459 Email: adceng@mweb.co.za P O BOX 83,

More information

100L AIR COMPRESSOR MODEL NO: TIGER 16/1010 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13

100L AIR COMPRESSOR MODEL NO: TIGER 16/1010 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13 100L AIR COMPRESSOR MODEL NO: TIGER 16/1010 PART NO: 2244025 OPERATION & MAINTENANCE INSTRUCTIONS LS01/13 INTRODUCTION Thank you for purchasing this product. Before attempting to use this product, please

More information

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM IMI CORNELIUS INC One Cornelius Place Anoka, MN 55303-623 Telephone (800) 238-3600 Facsimile (612) 22-326 INSTALLATION INSTRUCTIONS FOR INSTALLING CO 2 BLENDER KIT (P/N 111612000 IN BEER SYSTEM SECONDARY

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE11C150, VE15C150, VE18C150 PART NO: 2226005, 2226000, 2226015 OPERATION & MAINTENANCE INSTRUCTIONS LS0615 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

Oxygen Demand Valve. Instructions for Use

Oxygen Demand Valve. Instructions for Use Oxygen Demand Valve Instructions for Use 702-0059.11 May 2017 Contents 1. Symbols... 2 2. Warnings and Cautions... 2 2.1. Warnings!... 2 2.2. Cautions!... 3 2.3. Notes... 4 3. Functional Description...

More information

AUTOMATIC GAS MANIFOLD

AUTOMATIC GAS MANIFOLD Corporation AUTOMATIC GAS MANIFOLD 200-SERIES LED Andersen Medical Gas 12 Place Lafitte Madisonville, LA 70447 http://www.themedicalgas.com 1-866-288-3783 Operation Maintenance and Installation Manual

More information

24L AIR COMPRESSOR MODEL NO: TIGER 8/260 PART NO:

24L AIR COMPRESSOR MODEL NO: TIGER 8/260 PART NO: 24L AIR COMPRESSOR MODEL NO: TIGER 8/260 PART NO: 1499490 OPERATION & MAINTENANCE INSTRUCTIONS ORIGINAL INSTRUCTIONS LS0918 - ISS 2 INTRODUCTION Before attempting to use this product, please read this

More information

Air Mortar User Manual

Air Mortar User Manual Page 1/5 1 Introduction The Air Mortar is a fast-acting pneumatic system, designed to provide a rapid discharge of inert gas when triggered by 12VDC input signal. The standard system vessel has a volume

More information

Inert Air (N2) Systems Manual

Inert Air (N2) Systems Manual INSTRUCTION MANUAL Inert Air (N2) Systems Manual N2-MANUAL 2.10 READ AND UNDERSTAND THIS MANUAL PRIOR TO OPERATING OR SERVICING THIS PRODUCT. GENERAL INFORMATION Positive pressure nitrogen gas pressurizing

More information

FOGEX HIGH-PRESSURE WATER MIST CYLINDER SYSTEMS

FOGEX HIGH-PRESSURE WATER MIST CYLINDER SYSTEMS Manual FOGEX HIGH-PRESSURE WATER MIST CYLINDER SYSTEMS User Guide Issue 1 October 2004 Assessed to ISO 9001: 2000 Certificate number 018 Chapter 2 Safety and Environmental Considerations Safety is of paramount

More information

Installation and Maintenance Manual. Microprocessor Based Digital Manifold NFPA

Installation and Maintenance Manual. Microprocessor Based Digital Manifold NFPA Installation and Maintenance Manual Microprocessor Based Digital Manifold NFPA w w w. a m i c o. c o m Table of Contents User Responsibility 4 Introduction 5 Features 5 Description of the Manifold 6 Shipment

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0578 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

24L AIR COMPRESSOR MODEL NO: TIGER 11/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13

24L AIR COMPRESSOR MODEL NO: TIGER 11/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13 24L AIR COMPRESSOR MODEL NO: TIGER 11/250 PART NO: 2244010 OPERATION & MAINTENANCE INSTRUCTIONS LS01/13 INTRODUCTION Thank you for purchasing this product. Before attempting to use this product, please

More information

Installation Instructions / Operating & Maintenance Manual Midi-Fill Digital Pressurisation Unit Models MFD15 & MFD22

Installation Instructions / Operating & Maintenance Manual Midi-Fill Digital Pressurisation Unit Models MFD15 & MFD22 Arrow Valves Ltd Installation Instructions / Operating & Maintenance Manual Midi-Fill Digital Pressurisation Unit Models MFD15 & MFD22 Tel 01442 823 123 Fax 01442 823 234 www.arrowvalves.co.uk Model shown

More information

TOOL BOX 033- HOSEREEL SERVICE / FAULT FINDING

TOOL BOX 033- HOSEREEL SERVICE / FAULT FINDING In SANS 10105 you will find the Minimum water pressure for both the hydrant and the hose reel. Which is 300KPA Manufacturing Standard SANS 543 1 Front & Back plates 2 Curved Plate 3 Waterway assembly 4

More information

KMW3000C. User Manual. Water Chiller for Xcalibur and Gemini systems. Version 1.6, September 2010

KMW3000C. User Manual. Water Chiller for Xcalibur and Gemini systems. Version 1.6, September 2010 KMW3000C Water Chiller for Xcalibur and Gemini systems User Manual Version 1.6, September 2010 Agilent Technologies XRD Products 10 Mead Road, Yarnton, Oxfordshire. OX5 1QU, UK Tel: +44 (0)1865 291600

More information

Suitable for anyone who is required to maintain industrial pneumatic systems. No prior knowledge of pneumatic or electrical principles is necessary.

Suitable for anyone who is required to maintain industrial pneumatic systems. No prior knowledge of pneumatic or electrical principles is necessary. PNEUMATICS COURSE 660: 4 DAYS: Max 8 Candidates This course provides maintenance personnel and production operators etc with the skills and knowledge necessary to carry out maintenance tasks on pneumatic

More information

Chesil - Electronic Pressurisation Unit

Chesil - Electronic Pressurisation Unit Chesil - Electronic Pressurisation Unit Equipment For Commercial/Light Industrial Central Heating Installations. Installation, Commissioning and Servicing Instructions IMPORTANT NOTE THESE INSTRUCTIONS

More information

Installation & Operating Instructions for Genesys Analog Series Manifolds

Installation & Operating Instructions for Genesys Analog Series Manifolds Manufacturer of Medical Gas Piping Equipment Installation & Operating Instructions for Genesys Analog Series Manifolds Model TMCU Model TMLU Tri-Tech, 35401 Avon Commerce Pkwy., Avon, Ohio 44011-1 - Genesys

More information

2 CYLINDER MOBILE CART USER S MANUAL / INSTRUCTIONS

2 CYLINDER MOBILE CART USER S MANUAL / INSTRUCTIONS Parker Hannifin Corporation Precision Fluidics Division Porter Instrument 245 Township Line Road Hatfield, PA 9440 Office 25-723-4000/Fax 25-723-506 www.porterinstrument.com 2 CYLINDER MOBILE CART USER

More information

Peninsula Dental Social Enterprise (PDSE)

Peninsula Dental Social Enterprise (PDSE) Peninsula Dental Social Enterprise (PDSE) Medical Gas Policy Version 3.0 Date approved: August 2018 Approved by: The Board Review due: August 2021 Policy will be updated as required in response to a change

More information

Installation & Operating Instructions for Genesys Series Manifolds

Installation & Operating Instructions for Genesys Series Manifolds Manufacturer of Medical Gas Pipeline Equipment Installation & Operating Instructions for Genesys Series Manifolds Model LLU Model CCU - 1 - Features & Benefits Features & Benefits Fully automatic changeover

More information

PRODUCT OPERATING MANUAL

PRODUCT OPERATING MANUAL PRODUCT OPERATING MANUAL PANBLAST TM VISIFLO RESPIRATOR AIRLINE FILTERS Manual Number: ZVP PC 0084 01 Pressure Equipment Directive 97/23/EC SECTION 1. GENERAL INFORMATION 2. INTRODUCTION 3. PREPARATION

More information

INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May

INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May 2015 98-0520 S/N 2001 N. Indianwood Ave. Tulsa, Oklahoma 74012 U.S.A. TEL: (918) 250-7200 FAX: (918) 459-0165 E-mail: chandler.sales@ametek.com

More information

Propane Conversion Kit Instructions

Propane Conversion Kit Instructions 604 8 0/ US/CA For heating engineers Propane Conversion Kit Instructions Logano G4 X gas-fired boiler This conversion kit and the accompanying instructions are for conversion of G4 X gas-fired boilers

More information

200 PSI COMPRESSORS - MODEL NUMBERS

200 PSI COMPRESSORS - MODEL NUMBERS 200 PSI COMPRESSORS - MODEL NUMBERS 380C AIR COMPRESSOR KIT PART NO. 38033 480C AIR COMPRESSOR KIT PART NO. 48043 380C 480C IMPORTANT: It is essential that you and any other operator of this product read

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0545 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

100C Air Compressor Kit

100C Air Compressor Kit 10010 100C Air Compressor (standard mounting bracket, CE Spec) 10014 100C Air Compressor (no leader hose or check valve, CE Spec) 10016 100C Air Compressor (with Omega Bracket, CE Spec) IMPORTANT: It is

More information

Training. Testor Training Manual

Training. Testor Training Manual Training Testor Training Manual Index Section 1 Introduction and Safety Warnings Section 2 Test Procedures Section 3 Test Hoses Section 4 Fault Location 1:1 1.1 Introduction The Dräger Testor test equipment

More information

CS150 CAP STAPLER OWNER S MANUAL Operation Safety Maintenance Warranty! www.stingerworld.com PLEASE READ! This manual contains important information about product safety. REV 06/15 WELCOME TO STINGER Congratula*ons

More information

COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: WARRIOR 55 PART NO: (110V) , (230V) , LS0512

COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: WARRIOR 55 PART NO: (110V) , (230V) , LS0512 COMPRESSOR MODEL NO: WARRIOR 55 PART NO: (110V) 2323010, (230V) 2322020, OPERATION & MAINTENANCE INSTRUCTIONS LS0512 INTRODUCTION Thank you for purchasing this CLARKE Compressor. Before attempting to use

More information

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1 ANA 74300 Oxygen Regulator with Yoke Connector User Manual Manual No. ANA 74300 001 Issue 1 Safety Thank you for purchasing this Clements Oxygen Regulator. For your safety it is imperative that this unit

More information

MEDICAL GAS CYLINDERS SELECTION AND MAINTENANCE OF SEALS USED ON HIGH PRESSURE CYLINDERS

MEDICAL GAS CYLINDERS SELECTION AND MAINTENANCE OF SEALS USED ON HIGH PRESSURE CYLINDERS GUIDANCE NOTE 26 MEDICAL GAS CYLINDERS SELECTION AND MAINTENANCE OF SEALS USED ON HIGH PRESSURE CYLINDERS 2013 GUIDANCE NOTE 26 MEDICAL GAS CYLINDERS SELECTION AND MAINTENANCE OF SEALS USED ON HIGH PRESSURE

More information

IntelliSwitch II Electronic Switchover

IntelliSwitch II Electronic Switchover ADI 9515-I IntelliSwitch II Electronic Switchover TM INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Carefully Read These Controls Corporation of America 1501

More information

APPS-160 Gen 2 User Manual

APPS-160 Gen 2 User Manual APPS-160 Gen 2 User Manual Applicable Products: PN 11914406 - APPS-160 10 to 50 PSIG operating pressure PN 11918618 - APPS-160 40 to 85 PSIG operating pressure Rev 9/16/2004 Chart Ind. www-chart-ind.com

More information

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO 97C COMPRESSOR KIT 12V PART NO. 00097 97C COMPRESSOR KIT 24V PART NO. 02497 98C COMPRESSOR KIT PART NO. 00098 97C 98C IMPORTANT: It is essential that you and any other operator of this product read and

More information

Flow meter. bellow vaporizer. APL valve. Scavenging system

Flow meter. bellow vaporizer. APL valve. Scavenging system Introductory Lecture Series: The Anesthesia Machine PORNSIRI WANNADILOK Objectives Anesthesia Machine Ventilators Scavenging Systems System Checkout 1 Flow meter ventilator bellow vaporizer Corrugated

More information

HANDBOOK. Squeeze Off Unit SOU 400. Please refer any queries to:

HANDBOOK. Squeeze Off Unit SOU 400. Please refer any queries to: HANDBOOK Squeeze Off Unit SOU 400 Please refer any queries to: Hy Ram Engineering Co Ltd Pelham Street Mansfield Nottinghamshire NG18 2EY Telephone No: (01623) 422982 Please note all queries should state

More information