DIAGNOSTIX TM 922 MERCURIAL BLOOD PRESSURE INSTRUMENT. Use, Care, & Maintenance

Similar documents
USE, CARE & MAINTENANCE DIAGNOSTIX TM 972 MERCURIAL BLOOD PRESSURE INSTRUMENT

Aneroid Sphygmomanometer. Use, Care, & Maintenance

General Warnings! A warning statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately could lead

Diagnostix TM Aneroids

ADC. Sphygmomanometer Accessories. Directions for Use, Care, & Maintenance AMERICAN DIAGNOSTIC CORPORATION

INSTRUCTION MANUAL MANUAL INFLATION BLOOD PRESSURE MONITOR

INSTRUCTIONS. Manual Home Blood Pressure Monitor MODEL Stethoscope Chestpiece. Cuff. Bulb

Wallmax Professional Aneroid Sphygmomanometer Model 222 User Manual

HOME BLOOD PRESSURE KIT INSTRUCTIONS TRILINGUAL INSTRUCTION GUIDE MODEL UA-100

Model ASSEMBLY and OPERATING INSTRUCTIONS

Semi-Automatic Blood Pressure Monitor with Memory

INSTRUCTION MANUAL MANUAL INFLATION BLOOD PRESSURE MONITOR

STANDARD OPERATING PROCEDURE ARTSANA MERCURY SPHYGMOMANOMETER

Model AUTOMATIC UPPER ARM Blood Pressure Monitor

RADAR. User Manual EASYCLIK MODEL Vascular. Compression. Device.

Introduction. Table of Contents. Automatic Wrist Blood Pressure Monitor With Voice-Guided Operation. Model No.: BP5K

5 series Blood Pressure Monitor

Noninvasive Blood Pressure Measurement. D. J. McMahon rev cewood

G7S Hand Pump Owner s Manual

J Air and Water Kit Instructions Part# 02584

Manual. FRESH AERO OneOP Model PC (Portable Compressors) Tire Inflator. Where Imagination Meets Innovation

Instruction Manual. Wrist Blood Pressure Monitor with Advanced Positioning Sensor (APS ) Model HEM-650 ESPAÑOL ENGLISH

Instruction Manual LIMITED 1 YEAR WARRANTY. Hydraulic Punch Driver Read this material before using this product.

FRESH AERO OneOP Model RC (Reservoir/tank) Tire Inflating System

TECHNICAL CHARACTERISTIC CT 071

Portable Compressed CO 2 Regulator

BLOOD PRESSURE SENSOR 0377i

INSTRUCTION MANUAL. OMRON IntelliSense Blood Pressure Monitor. ModelHEM-907

CONTENTS SPECIFICATIONS GENERAL INFORMATION RECOMMENDED USE OPERATING PRINCIPLE TIPS ON TAKING YOUR BLOOD PRESSURE 3-4 BATTERY INSTALLATION

3.1.3 Weight. Frequency. Weight is obtained at the baseline examination and annually. Equipment

Auto Arm Blood Pressure Monitor

4-in-1 Professional Inflator Kit

AIR INLINE METAL SHEAR

Operating Instruction Manual for Liquid Nitrogen Containers. CRY/CRYC Series. Hagavish st. Israel Tel: , Fax:

Blood Pressure Training System S415

BLOOD PRESSURE SENSOR BT17i USER S GUIDE

CSA Sample Draw Aspirator Adapter Operator s Manual

Please Do Not Return This Product To The Store!

3 Scotch-Weld Polyurethane Reactive Adhesive Applicator Replacement Part Installation Guide

Windy City II. User instruction manual WARNING: Fill pressure: 200 bar (approximately 2,900 psi)

D Series Air and Water Kit Part# 02550

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

TactAir PCP Fill Kit/Adapter. Operation Manual

Sports Wrist Digital Blood Pressure Monitor Item #

VER: A00. Model No:BA2010

Tactical Sniper air rifle & carbine

Oxygen Concentrator. Installation and Operation Manual. ClearWater Tech, LLC. Integrated Ozone Systems

Digital Blood Pressure Monitor. Model UA-705. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni 1WMPD B

Offset head dual chuck

Parts and Owners Manual

Digital Ear Thermometer

MEDICARE 60. Semi-automatic upper arm blood pressure monitor. Instruction manual

Research Grade Blood Pressure Transducer User's Manual

The RICH LIFE Project: Blood Pressure Training Certification Checklists

8675/8677/8678 OWNER S MANUAL

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR

Blood Pressure Monitor

Vevazz. Operation Manual. Vevazz, LLC Rancho California Road. Suite Temecula, CA Phone: (773) Fax: (888)

Contents. Safety Notice. Wrist Type

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

TOP VALVE. Pat. #5,857,486 & 5,944,050. High Temperature: max. 300 F (149 C) Back Pressure And Pressure Relief Valves. Instruction Manual

Accu-Tab PowerBase 3012AT by Axiall Corporation

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure?

MODEL 100 NITROGEN INFLATION CART

Digital Blood Pressure Monitor for the Upper Arm INSTRUCTION MANUAL

premium wrist blood pressure monitor

SPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1.

2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL

Vessel Installation and Operating Manual

450P- RV AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

Model UB-328. Wrist Digital Blood Pressure Monitor. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

TOP VALVE. Pat. #5,857,486 & 5,944,050. Mid-Range Pressure PSIG Back Pressure and Pressure Relief Valves. Instruction Manual

450P AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

MODEL NUMBER: P PORTABLE COMPRESSOR

Monster Windy City Leopard

Owner s Manual & Safety Instructions

Beeman Precision Airguns R-Series Air Rifle Owner s Manual

IZH 53M air pistol owner s manual

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

MODEL 1410 OWNER S MANUAL

Blood Pressure Monitor

42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only)

92831 TEL: (714) FAX:

Owner s Manual & Safety Instructions

MODEL NUMBER: P PORTABLE COMPRESSOR

RULES FOR SAFE FIRING:

92831 TEL: (714) FAX:

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

TL-12 FLUSH TANK. Manual Ref. # MN-??? Polyurethane Machinery Corp. REVISION 1.0

SPECIFICATIONS APCEPH1

User Instruction Manual

Circulation and Respiration: Vital Signs Student Version

Interface Devices, Inc. Hydraulic Mini Mule

MANUAL BE SERIES Test Benches

INSTRUCTION MANUAL. Automatic Blood Pressure Monitor with Fit Cuff. =Fit Cuff=!"#$% IA1B. Model

BLOOD PRESSURE SIMULATOR LF01095U INSTRUCTION MANUAL

6 digital caliper with case

Transcription:

DIAGNOSTIX TM 922 MERCURIAL BLOOD PRESSURE INSTRUMENT Use, Care, & Maintenance 1

A Special Thank You... Thank you for choosing an ADC blood pressure instrument. We re proud of the care and quality that goes into the manufacture of each and every sphygmomanometer that bears our name. Only the finest materials are used to assure you of a timeless instrument designed for optimum performance. You ll quickly appreciate the results for you now own the finest sphygmomanometer that money can buy. With proper care and maintenance your ADC blood pressure instrument is sure to provide you with many years of dependable service. Please read the following instructions and general information which will prove helpful in allowing you to enjoy your ADC product. Thank you for your patronage. It is indeed our pleasure to serve you. 2016 AMERICAN DIAGNOSTIC CORP. Sincerely, American Diagnostic Corp. Symbols Symbol Definition Important Warning/Caution Symbol Definition Manufacturer s Information Not made with natural rubber latex Circumference Size F Temperature Limit Humidity Limitation 2

Device Description and Intended Use ADC aneroid sphygmomanometers are used by professional healthcare providers and individuals trained in auscultatory blood pressure technique to determine systolic and diastolic blood pressure in humans and animals. A warning statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately could lead to patient injury, illness, or death. CAUTION: Federal law restricts this device to sale by or on the order of a physician or licensed healthcare practitioner. WARNING: If luer lock connectors are used in the construction of tubing, there is a possibility that they might be inadvertently connected to intra-vascular fluid systems, allowing air to be pumped into a blood vessel. Immediately consult a physician if this occurs. WARNING: Do not allow a blood pressure cuff to remain on patient for more than 10 minutes when inflated above 10 mmhg. This may cause patient distress, disturb blood circulation, and contribute to the injury of peripheral nerves. WARNING: Safety and effectiveness with neonate cuff sizes 1 through 5 is not established. WARNING: Do not twist connectors when joining the blood pressure cuff to tubing as this may cause the connectors to stick and make changing the blood pressure cuff difficult. WARNING: When using with an infant or child cuff, extra care must be taken to prevent overinflation. With smaller cuffs (infant or child) the cuff can inflate to over 300mmHg with just two full compressions of the bulb. To prevent discomfort or injury to the patient and damage to the instrument, bulb should only be partially squeezed, so that each stroke inflates the cuff in 40mmHg to 60mmHg increments until inflated to the desired level. WARNING: This product may contain a chemical known to the state of California to cause cancer, birth defects, or other reproductive harm. Parts and Assembly This booklet contains operating and maintenance information for the DIAGNOSTIX 922 series desktop mercurial blood pressure instrument. Please read and retain. Your DIAGNOSTIX 922 consists of a main unit with 4mm calibrated unbreakable plastic cartridge tube, inflation system (which includes the ADCUFF nylon cuff with Size Guide marking system, latex-free inflation bladder, bulb, and the ADFLOW valve), luer connector, and operating instructions. Cartridge Tube Main Unit Reservoir Cuff Bulb & Valve Locking Mechanism 3

Connecting the Inflation System to Manometer Remove the red cap which seals the reservoir containing mercury. Attach the free end of reservoir tube securely to air inlet. Save red safety cap for later use. When transporting mercury instrument we recommend replacing the red safety cap. Please note: It is normal for negligible amounts of residual mercury droplets to accumulate around the air inlet beneath the red safety cap during transportation. Warning: Before transporting this instrument, mercury must be stored in the tank and the reservoir lock MUST be switched to the off position to prevent mercury spills. Operating Instructions Your DIAGNOSTIX 922 Desktop Mercury Sphygmomanometer incorporates a special safety feature that locks the mercury within the reservoir during storage, transport, or maintenance. To Operate the 922: To release the mercury from the reservoir, move the locking lever to the left. Mercury will flow up into the plastic cartridge tube and rest at the 0 mark. If the mercury doesn t rise within the tube, tilt the entire unit 45 to the left. If mercury fails to rise, or doesn t reach the 0 level, have unit serviced. To Lock Mercury within the Reservoir: Tilt the entire unit back 45 towards the reservoir to permit mercury to flow out of cartridge tube and into reservoir. When cartridge tube is completely emptied of ALL mercury (and while it is still tilted 45 ), move locking lever to the right. Mercury should be locked within the reservoir during maintenance or transport. PLEASE NOTE: MERCURY SHOULD BE SECURED WITHIN THE RESERVOIR BEFORE CLOSING LID, DURING TRANSPORT, OR WHEN SERVICING THE CARTRIDGE TUBE, DI- APHRAGMS, OR FILTERS. NEVER DISASSEMBLE UNIT UNLESS MERCURY IS FIRST LOCKED WITHIN RESERVOIR. MEASUREMENT PROCEDURE 1. Patient Position The patient should sit or lie comfortably. The arm should be fully supported on a flat surface at heart level. (If the arm s position varies, or is not level with the heart, measurement values obtained will not be consistent with the patient s true blood pressure.) Observer should view manometer in a direct line and at eye level to avoid parallax error. 4

2. Apply the cuff Select Larger Cuff CORRECT ADCUFF nylon cuffs, with proprietary Size Select Smaller Cuff Guide marking system, are specially designed to promote the precise, accurate determination of blood pressure. Index and range markings ensure use of the correct cuff size. The artery mark indicates proper cuff positioning. Place the cuff over the bare upper arm with artery mark positioned directly over the brachial artery. The bottom edge of the cuff should be positioned approximately one inch above the antecubital fold. Wrap the end of the cuff, not containing the bladder, around arm snugly and smoothly and engage adhesive strips. To verify a correct fit, check that the INDEX line falls between the two RANGE lines. 3. Inflate the cuff Close the ADFLOW valve by turning thumbscrew clockwise. Palpate the radial artery while inflating the cuff. Be sure to inflate cuff quickly by squeezing bulb rapidly. Inflate cuff 20-30mmHg above the point at which the radial pulse disappears. 4. Position the Stethoscope Position the chestpiece in the antecubital space below the cuff, distal to the brachium. Do not place chestpiece underneath the cuff, as this impedes accurate measurement. Use the bell side of a combination stethoscope for clearest detection of the low pitched Korotkoff (pulse) sounds. 5. Deflate the cuff Open the valve to deflate the cuff gradually at the AHA recommended rate of 2-3mmHg per second. 6. Measurement Record the onset of Korotkoff sounds (Phase 1) as the systolic pressure, and the disappearance of these sounds (Phase 5) as the diastolic pressure. (Some healthcare professionals prefer recording diastolic 1 and diastolic 2. Diastolic one occurs at phase 4). After measurement is completed, open valve fully to release any remaining air in the cuff. Remove cuff. s s s Care and Maintenance Warning: Before transporting this instrument, mercury must be stored in the tank and the reservoir lock MUST be switched to the off position to prevent mercury spills. Storage After measurement, wrap cuff around bulb and store in base of main unit. IT IS NOT NEC- ESSARY TO CLOSE UNIT AFTER EACH USE - HOWEVER, ALWAYS REMEMBER TO LOCK MERCURY WITHIN THE RESERVOIR BEFORE CLOSING LID. 5

Manometer Periodically replace the filter plug and diaphragm, and clean the inside of the plastic cartridge tube using alcohol and a lint free pipe brush. Units should be cleaned and serviced periodically depending on frequency of use, or whenever inside of cartridge tube appears dirty. Proper servicing of your mercury unit according to a strict maintenance schedule will assure maximum measurement accuracy and extend the life of your instrument. For directions on disassembling the unit refer to the enclosed schematic diagram and accompanying instructions. PLEASE NOTE: MERCURY IS TOXIC. SECURE MERCURY WITHIN THE RESERVOIR BEFORE SERVICING UNIT. ALWAYS WEAR PROTECTIVE CLOTHING (GLOVES AND GOGGLES) WHEN CLEANING OR SERVICING THE UNIT. CLEAN ANY SPILLED MERCURY PROMPTLY USING AP- PROVED MERCURY CONTAINMENT KITS. ONLY PROPERLY TRAINED SERVICE TECHNICIANS SHOULD ATTEMPT TO SERVICE THE UNIT. CONSULT THE ENCLOSED MANUAL ENTITLED JUST IN CASE FOR PROCEDURES ON DEALING WITH A MERCURY SPILL. Limited Warranty American Diagnostic Corporation (ADC) warrants its products against defects in materials and workmanship under normal use and service as follows: 1. Warranty service extends to the original retail purchaser only and commences with the date of delivery. 2. The entire sphygmomanometer is warranted for one year. 3. The inflation system is warranted for three years. 4. The calibrated plastic cartridge tube is warranted for life and will be replaced FREE OF CHARGE when returned freight prepaid with the complete unit. What Is Covered: Calibration, repair or replacement of parts, and labor. What is Not Covered: Transportation charges to and from ADC. Damages caused by abuse, misuse, accident, or negligence. Incidental, special, or consequential damages. Some states do not allow the exclusion or limitation of incidental, special, or consequential damages, so this limitation may not apply to you. To Obtain Warranty Service: Send item(s) postage paid to ADC, Attn: Repair Dept., 55 Commerce Drive, Hauppauge, NY 11788. Please include your name and address, phone no., proof of purchase, with a brief note explaining the problem, and $2.00 to cover the cost of return shipping and handling. Implied Warranty: Any implied warranty shall be limited in duration to the terms of this warranty and in no case beyond the original selling price (except where prohibited by law). This warranty gives you specific legal rights and you may have other rights which vary from state to state. 6

To register your product visit us at www.adctoday.com/register FOR QUESTIONS, COMMENTS, OR SUGGESTIONS CALL TOLL FREE: 1-800-ADC-2670 OR VISIT www.adctoday.com/feedback This manual is available online in a variety of languages, follow the links for language options. www.adctoday.com/care 55 Commerce Drive Hauppauge, N.Y. 11788, U.S.A. www.adctoday.com IB p/n 9353-00 - rev 9 Printed in the U.S.A 7