Anesthesiologists work in remote locations both

Similar documents
Ventilators can be described as pneumatically or. Oxygen Consumption of a Pneumatically Controlled Ventilator in a Field Anesthesia Machine

Draeger Fabius GS Pre-Use Check

Narkomed MRI-2. A perfect fit for the MRI environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

Anes Machine Use and Pre-use Check - CA1 focus

Fabius Tiro TM. Designed for your environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

What is new in ventilation - New modes of ventilation and new Anesthesia Machine Technology

Flow meter. bellow vaporizer. APL valve. Scavenging system

Fabius Tiro TM. Designed for your environment. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

ROUTINE PREOXYGENATION

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution

New Anesthesia Delivery Systems What Do I Need to Know About My New Machine?

Indications for Mechanical Ventilation. Mechanical Ventilation. Indications for Mechanical Ventilation. Modes. Modes: Volume cycled

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

Automatic Transport Ventilator

Anesthesia Machine Intro

V8600 Ventilator. Integrated Invasive & Noninvasive Ventilation

Anesthesia Delivery Systems What Do I Need to Know About My Machine? A Workshop James H. Philip, MEE, MD, CCE

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

Mechanical Ventilation. Which of the following is true regarding ventilation? Basics of Ventilation

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

S/5 Aespire. Quick Reference Guide

Disclosures. The Pediatric Challenge. Topics for Discussion. Traditional Anesthesia Machine. Tidal Volume = mls/kg 2/13/14

Pressure Controlled Modes of Mechanical Ventilation

Mechanical Ventilation

Selecting the Ventilator and the Mode. Chapter 6

MACS Mask CPAP System

Operating Instructions for Microprocessor Controlled Ventilators


Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

COMPARISON OF PORTABLE EMERGENCY VENTILATORS USING A LUNG MODEL

AUTOVENT 4000 VENTILATOR

3100A Competency Exam

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge

Ventilating the Sick Lung Mike Dougherty RRT-NPS

Recommendations on Checking Anaesthesia Delivery Systems

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo

GE Healthcare. Aestiva/5. More than superior ventilation

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES

Initiation and Management of Airway Pressure Release Ventilation (APRV)

VIMA. (Volatile Induction and Maintenance Anesthesia) How and Why. James H. Philip M.E.(E), M.D.

Flight Medical presents the F60

Anaesthetic infrastructure and supplies

Selecting and Connecting Breathing Systems

Key words: intrahospital transport; manual ventilation; patient-triggered ventilation; respiratory failure

Panther 5 Acute Care Ventilator

ASV. Adaptive Support Ventilation

Your customized anesthesia workstation DRÄGER FABIUS GS PREMIUM. Dräger. Technology for Life

Mechanical Ventilation

ekom Compressor for Medical Air Sonata with Ekom

Operation and Maintenance of the EPV200 Portable Ventilator

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine

VENTILATION STRATEGIES FOR THE CRITICALLY UNWELL

Medical Instruments in the Developing World

Effects of Fresh Gas Flow, Respiratory Rate, and Mode of Ventilation on Recovery of Tidal Volume after Circuit Disconnection

Minimum size for maximum performance

Aespire 7100 Essential performance Compact design

Fabius plus. The plus that makes all the difference. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Hospital and Transport for Controlled Breathing

Automatic Transport Ventilators. ICU Quality Ventilation on the Street.

Fraction of Inspired Oxygen Delivered by Elisée 350 Turbine Transport Ventilator With a Portable Oxygen Concentrator in an Austere Environment

HAMILTON-C3 HAMILTON-C3. The compact high-end ventilator

ONLINE DATA SUPPLEMENT. First 24 hours: All patients with ARDS criteria were ventilated during 24 hours with low V T (6-8 ml/kg

O-Two Self-Study Guide. e600 Transport Ventilator Ventilation Modes

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

PROBLEM SET 8. SOLUTIONS April 15, 2004

Technical Data and Specifications

What our equipment teaches us

Blease900 Ventilator PATIENT CENTERED VENTILATION

An intuitive and innovative answer to your anaesthesia challenges

Ventilators. Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course

Fabius plus. The plus that makes all the difference. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Working version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee

Virginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P

Quick Response Pneupac VR1 Emergency Ventilator AIRWAYMANAGEMENT

Update to RS-232 commands. Changing patient from NIV to INVASIVE Vent Type. SNDF command

Apollo Safety Features

SUPPLEMENTARY APPENDIX. Ary Serpa Neto MD MSc, Fabienne D Simonis MD, Carmen SV Barbas MD PhD, Michelle Biehl MD, Rogier M Determann MD PhD, Jonathan

NOTE: If not used, provider must document reason(s) for deferring mechanical ventilation in a patient with an advanced airway

Emergency Transport and Ventilation

Mechanical Ventilation

Anesthesia Workstation

Provide Comfort to your Patients is our commitment PRODUCT CATALOGUE HOSPTIAL SUPPLIES. Copyright ReadMed Company Ltd

Oxylog 3000 plus Emergency & Transport Ventilation

Materials and Methods

NEONATAL/INFANT VENTILATOR

Solutions. Digital pressure management solutions. Copyright 2013 GaleMed Corporation

IMPACT. Uni-Vent. Eagle. Model 754 Portable Ventilator

PICU Resident Self-Study Tutorial The Basic Physics of Oxygen Transport. I was told that there would be no math!

Comparison Of The Efficacy Of Two Low Fresh Gas Flow Techniques In Low Flow Anaesthesia

C O R P O R AT E B R O C H U R E

RESPIRATORY MUSCLES IN HEALTH AND EMPHYSEMA *

patient education program 4156 South 52 nd Street, Omaha, NE ChildrensOmaha.org/HomeHealthcare

The Internet Journal of Anesthesiology 2010 : Volume 22 Number 2

Supporting you in saving lives!


Transcription:

TECHNICAL COMMUNICATIONS Oxygen Consumption with Mechanical Ventilation in a Field Anesthesia Machine Dale F. Szpisjak, MD*, Charles L. Lamb, MD, and Kenneth D. Klions, MD *Department of Anesthesiology, Uniformed Services University of the Health Sciences, Bethesda, Maryland, and the Anesthesia Department, Fleet Hospital, Bremerton, Washington Field anesthesia machines (FAM) with gas-powered ventilators have been developed for remote locations that may not have a central supply of oxygen. These ventilators may rapidly deplete oxygen cylinders, especially in patients with decreased pulmonary compliance. Our goal in this study was to determine oxygen consumption rates with a contemporary FAM in models of high (HC) and low (LC) pulmonary compliance. Oxygen consumption rates were tested using D cylinders (initial pressure 1700 psig) and the Narkomed M FAM, which uses an air injector to decrease compressed gas consumption by entraining room air as part of the drive gas. Three different tidal volumes (Vt) were tested (500, 750, and 1000 ml) with HC and LC lung models, and the fresh gas flow rate was 1 L/min. Respiratory rate was constant at 10 breaths/min. Oxygen consumption varied directly with Vt and inversely with compliance, increasing from 4.8 0.07 L/min with the HC-500 ml Vt model to 6.2 0.05 L/min with the LC-1000 ml Vt model. D cylinder duration ranged from 56.8 0.4 to 73.6 1.0 minutes. Assuming oxygen fresh gas flow of 1 L/min, calculating tank duration with the fastest consumption rate underestimated the tank duration for more compliant and smaller Vt models but provided a greater margin of patient safety. (Anesth Analg 2005;100:1713 7) Anesthesiologists work in remote locations both inside and outside the fixed hospital setting, providing anesthetics for office-based, humanitarian, and military interventions. Space limitations often preclude the use of anesthesia systems present in the fixed hospital operating room (OR), especially when working in austere field conditions. These space limitations led to the development of alternate anesthesia machines. These so-called field anesthesia machines (FAM) have evolved significantly during the last decade (1). Contemporary FAMs are functionally equivalent to the anesthesia machines in the fixed hospital OR, which is a requirement of the ASA Guidelines for Nonoperating Room Anesthetizing Locations (2). This equivalence includes the ability to provide gas-powered mechanical ventilation. Gas-powered ventilators consume drive gas at a high rate (3). As adverse outcomes in office-based settings have been linked to a lack of supplemental The views expressed in this manuscript are those of the authors and do not reflect the official policy or position of the Department of the Navy, Department of Defense, or the United States Government. Accepted for publication October 21, 2004. Address correspondence and reprint requests to Dale F. Szpisjak, MD, Department of Anesthesiology, Uniformed Services University of the Health Sciences, 4301 Jones Bridge Road, Bethesda, MD 20814. Address e-mail to dszpisjak@usuhs.mil. DOI: 10.1213/01.ANE.0000149897.87025.A8 oxygen (4), the use of gas-powered ventilation in settings with limited gas supplies directly impacts on patient safety. Because the backup oxygen system should include the equivalent of at least a full E cylinder (2), knowing the drive gas consumption rate is essential to estimating cylinder duration. A previous study (3) investigating tank depletion rates with the Narkomed 2B (North American Draeger, Telford, PA) and the Ohmeda 7000 (Datex-Ohmeda, Helsinki, Finland) ventilators found a slower depletion rate with the former because its air injector decreases compressed gas consumption by entraining room air as part of the drive gas. However, that study only tested a high compliance (HC) lung model with a tidal volume (Vt) of 500 ml. The data generated in that previous study may not be directly applicable to patients requiring larger Vt or with decreased pulmonary compliance. Because cylinder duration affects patient safety, we measured drive gas consumption rates with a contemporary FAM in models of both HC and low pulmonary compliance (LC). Methods After obtaining institutional approval, the oxygen consumption rates were determined with the Narkomed M FAM (North American Draeger ) when mechanically 2005 by the International Anesthesia Research Society 0003-2999/05 Anesth Analg 2005;100:1713 7 1713

1714 TECHNICAL COMMUNICATIONS SZPISJAK ET AL. ANESTH ANALG OXYGEN CONSUMPTION IN A FIELD ANESTHESIA MACHINE 2005;100:1713 7 Table 1. High Compliance Lung Model Ventilator Settings Inspiratory flow Peak pressure Mean pressure Compliance (ml/cm H 2 O) 500 ML (b) 16 5 36 750 HL (c) 19 6 39 1000 LM (d) 26 8 42 Compliance calculated as tidal volume/(peak pressure positive end-expiratory pressure). Positive end-expiratory pressure 2cmH 2 O in all groups. ML mid-low; HL high-low; LM low-medium. Table 2. Low Compliance Lung Model Ventilator Settings Inspiratory flow Peak pressure Mean pressure Compliance (ml/cm H 2 O) 500 LM (b) 24 9 23 750 LM 31 9 26 1000 MM (c) 53 14 20 Compliance calculated as tidal volume/(peak pressure positive end-expiratory pressure). Positive end-expiratory pressure 2cmH 2 O in all groups. LM low-medium; MM mid-medium. ventilating and supplied by D cylinders. Detailed product information and photographs of the FAM model used in this study are available at the manufacturer s Web site (5). A certified technician performed machine preventive maintenance, and all tests were done with the same machine. The anesthesia machine s cylinder pressure gauge reading corresponded with the D cylinders pressure of 1700 psig. Three different Vt (500 ml, 750 ml, and 1000 ml) were tested with both LC and HC lung models, for six groups. Four tanks were tested in each group. A new disposable circle anesthesia circuit was used for each group (Sims Portex, Fort Myers, FL). The lung model for each HC test was a new 3-L breathing bag (Sims Portex ). The Siemens Test Lung 190 (Siemens AG, Munich, Germany) was used for the LC model. Circuit pressures and Vt were measured using the FAM s airway pressure transducer and spirometer, respectively. The dynamic compliance (C DYN )ofthe lung model for each group was calculated as follows: C DYN Vt/(peak airway pressure positive endexpiratory pressure [PEEP]) (Tables 1 and 2). PEEP was2cmh 2 O in all groups. The Food and Drug Administration recommended machine check was performed before the first test of the day, and an abbreviated check was performed between tests. An H cylinder provided the pipeline oxygen supply while the ventilator and machine settings were adjusted and verified. Respiratory rate was 10 breaths/min and the I:E ratio was 1:2. Vt was set by moving the ventilator bellows adjustment plate until the measured exhaled Vt equaled 500 ml, 750 ml, or 1000 ml. Fresh gas flow (FGF) was limited to 1 L/min. Because drive gas in the Narkomed M vents to the atmosphere when ventilating with an inspiratory pause (6), the inspiratory flow was limited to the setting that fully compressed the bellows without an inspiratory pause. After verification of the settings, the pipeline supply hose was disconnected from the machine. When the low oxygen supply pressure alarm sounded, the D cylinder was opened, and the time until the low oxygen supply pressure alarm next sounded was recorded. To perform calculations, a full D cylinder s pressure and volume were assumed to be 1900 psig and 400 L, respectively (7). The low oxygen supply pressure alarm threshold was assumed to be 35 psig. Because the D cylinders available for this study were not consistently filled to a pressure of 1900 psig, a pressure of 1700 psig was chosen. The initial volume was therefore calculated as 400 L (1700 psig/1900 psig) 358 L. The volume used from the cylinders was then calculated as 358 L (400 L 35 psig/1900 psig) or 351 L. Statistical calculations were done using Excel2000 (Microsoft Corporation, Redmond, WA) and Primer of Biostatistics (Version 4.02, McGraw-Hill, New York, NY). Drive gas consumption rates were compared to minute ventilation (V e) with linear regression analysis. Between-group comparisons with the same Vt were analyzed using independent sample Student s t-tests. Among-group comparisons with the same lung model were analyzed using repeated-measures analysis of variance with the Student-Newman-Keuls test for multiple comparisons. The data were reported as mean ( sd), and P 0.05 was accepted as significant. Results The time until the low oxygen supply pressure alarm sounded for the HC lung model was more than the LC lung model at each tested volume (P 0.001, Table 3). When D cylinders containing 358 L supplied oxygen for FGF and powered the ventilator, the time until the

ANESTH ANALG TECHNICAL COMMUNICATIONS SZPISJAK ET AL. 1715 2005;100:1713 7 OXYGEN CONSUMPTION IN A FIELD ANESTHESIA MACHINE Table 3. Narkomed M Ventilator D cylinder Duration High compliance Low compliance P value 500 73.6 1.0* 64.8 1.2* 0.001 750 68.0 0.9* 61.5 1.2* 0.001 1000 63.4 0.6* 56.8 0.4* 0.001 Values are mean sd (min). * P 0.05 compared with other same compliance groups. low oxygen supply pressure alarm sounded ranged from 56.8 0.4 to 73.6 1.0 min (Table 3). For each Vt tested, the cylinder gas consumption rate with the HC lung model was less than the cylinder gas consumption rate with LC lung model (P 0.001) (Fig. 1) and increased from 4.8 0.07 L/min with the HC-500 ml Vt model to 6.2 0.05 L/min with the LC-1000 ml Vt model. The cylinder gas consumption rate correlated highly with V e, with r 2 values of 0.980 and 0.993 for the HC and LC lung models, respectively (Fig. 1). Discussion This study demonstrated that the oxygen consumption rates in a contemporary FAM using mechanical ventilation increased with both increasing V e and decreasing pulmonary compliance. Using the fastest consumption rate, obtained in the LC-1000 ml Vt model, a full D cylinder will supply oxygen for approximately 1 hour when mechanically ventilating with a FGF of 1 L/minute. These results compare favorably to data reported elsewhere (3,6,8). Cicman et al. (6) reported the ventilator s oxygen consumption rate with a medium inspiratory flow control setting as 12.5 L/min but did not describe a lung model or other ventilator settings. Taenzer et al. (3), using a HC-500 ml Vt lung model, demonstrated that with a FGF of 1 L/minute and a respiratory rate of 10 breaths/minute, the E cylinder duration was 99 minutes. Assuming a 650-L oxygen supply with the E cylinder, the total oxygen consumption (from both FGF and the Narkomed 2B ventilator) was 6.6 L/min. This cylinder depletion rate was approximately 38% more than that obtained with the Narkomed M FAM using the same lung model. Klemenzson and Perouansky (8) used a different HC lung model (ventilator tester, BIO-TEK, Winooski, VT) with the Narkomed AV-2 ventilator and demonstrated a total oxygen consumption (FGF 1 L/minute, Vt 700 ml, V e 7 L/minute) of 7.0 L/minute. This cylinder depletion rate was approximately 35% more than that obtained with the Narkomed M FAM with a larger Vt of 750 ml. The reason for these differences may be the lower inspiratory flow control setting with the FAM (midlow versus low-medium), which was selected to eliminate the inspiratory pause. When the Narkomed Figure 1. Cylinder gas consumption versus minute ventilation with fresh gas flow of 1 L/min. Results are means sd P 0.001 for all same minute ventilation group comparisons, and P 0.05 for all same lung model group comparisons. LC low compliance, HC high compliance. ventilator delivers an inspiratory pause, drive gas continues to flow to the valve case assembly en route to the bellows canister. When the bellows canister has reached its volume limit, any additional drive gas is vented to the atmosphere via the valve case assembly (6). Therefore, eliminating the inspiratory pause should conserve drive gas. Strategies for minimizing the use of compressed oxygen by the ventilator include maintaining spontaneous ventilation or using manual ventilation, although these may not be practical in an austere environment depending on the surgical procedure or necessity for the clinician to multitask. Because drive gas consumption is proportional to the V e, setting a smaller V e can reduce it, but this may not be practical if hyperventilation is indicated. Minimizing FGF via closed-circuit anesthesia will also conserve cylinder oxygen for use as drive gas. Another strategy is to use a different gas to power the ventilator. The Narkomed M FAM can use air, instead of oxygen, as the drive gas. Selecting air as drive gas requires simply changing the drive gas selection toggle switch from oxygen to air, but implementation of this strategy depends on the availability of a compressed air supply. Finally, although not an option with the FAM used in this study, an electrically driven ventilator would maximize medical gas availability for the patient (9,10). Although the LC lung model was not as sophisticated as other commercially available models, the C DYN values obtained with it were comparable to those reported for patients with adult respiratory distress syndrome (ARDS) (11) and status asthmaticus (12) (25 and 20 ml/cm H 2 O, respectively). The C DYN of the LC lung model was also comparable to that reported in a group of 118 intensive care unit patients with respiratory failure (13). The etiologies of respiratory failure in 65% of those patients were pneumonia, chronic obstructive pulmonary disease,

1716 TECHNICAL COMMUNICATIONS SZPISJAK ET AL. ANESTH ANALG OXYGEN CONSUMPTION IN A FIELD ANESTHESIA MACHINE 2005;100:1713 7 Figure 2. Estimated D cylinder duration in a high compliance (HC) 750 ml tidal volume with fresh gas flow of 1 to 5 L/min. Estimates extrapolated from data generated with the Narkomedâ M field anesthesia machine with a HC 750 ml tidal volume model and fresh gas flow of 1 L/min. Figure 3. Estimated D cylinder duration in a low compliance (LC) 750 ml tidal volume with fresh gas flow of 1 to 5 L/min. Estimates extrapolated from data generated with the Narkomedâ M field anesthesia machine with a LC 750 ml tidal volume model and fresh gas flow of 1 L/min. and ARDS, whereas the C DYN was in the 22 to 24 ml/cm H 2 O range. A limitation of this study was that it did not model clinical events (e.g., induction and emergence) or disease states requiring a higher FGF. In addition to the increased oxygen needed to power a ventilator, a patient with poorly compliant lungs may have injury or illness that will increase metabolic oxygen requirements. To compensate for these limitations, graphs of D cylinder duration with various cylinder pressures and FGF were extrapolated from the HC and LC 750 ml Vt models and included as a guide for clinicians using this FAM (Figs. 2 and 3). In conclusion, oxygen consumption rates in a contemporary FAM with gas-powered mechanical ventilation were directly proportional to Vt and inversely proportional to pulmonary compliance. With a FGF of 1 L/minute, calculating tank duration with the fastest consumption rate (6.2 L/minute, obtained from the LC-1000 ml Vt model) underestimates the tank duration for more compliant and smaller Vt models, while providing a greater margin of patient safety. References 1. Petty WC. Military Anesthesia Machines. In: Zajtchuk R, Grande CM, eds. Anesthesia and perioperative care of the combat casualty. Washington, DC: Office of the Surgeon General, 1995:141 74. 2. American Society of Anesthesiologists. Guidelines for nonoperating room anesthetizing locations. Available at: http://www. asahq.org/publicationsandservices/ standards/14.pdf. Accessed April 30, 2004. 3. Taenzer AH, Kovatsis PG, Raessler KL. E cylinder-powered mechanical ventilation may adversely impact anesthetic management and efficiency. Anesth Analg 2002;95:148 50. 4. Morrell RC. OBA questions, problems just now recognized, being defined. APSF Newsl 2000;15:1 3. 5. Draeger Medical. Narkomed M product brochure. Available at: http://www.draeger.com /MT/ internet/pdf/careareas/ ORAnesthesia/or_narm_br_us_PL00049.pdf. Accessed July 20, 2004. 6. Cicman JH, Skibo VF, Yoder JM. Anesthesia systems. Part II: operating principles of fundamental components. J Clin Monit 1993;9:104 11. 7. Dorsch JA, Dorsch SE. Medical gas cylinders and containers. In: Understanding anesthesia equipment: Construction, care and complications. 3rd ed. Baltimore: Lippincott, Williams and Wilkins, 1999:10. 8. Klemenzson GK, Perouansky M. Contemporary anesthesia ventilators incur a significant oxygen cost. Can J Anaesth 2004; 51:616 20.

ANESTH ANALG TECHNICAL COMMUNICATIONS SZPISJAK ET AL. 1717 2005;100:1713 7 OXYGEN CONSUMPTION IN A FIELD ANESTHESIA MACHINE 9. Draeger Medical. Fabius GS Product Information. Available at: http://www.draeger.com/mt/internet/pdf/careareas/ ORAnesthesia /or_fabiusgs_br_us_9049128.pdf. Accessed April 25, 2004. 10. Draeger Medical. Virtual Fabius GS. Available at: http:// www.simanest.org/vfgs3.html. Accessed July 16, 2004. 11. Shimada Y, Yoshiya I, Tanaka K, et al. Evaluation of the progress and prognosis of adult respiratory distress syndrome. Chest 1979;76:180 6. 12. Afessa B, Morales I, Cury J. Clinical course and outcome of patients admitted to an ICU for status asthmaticus. Chest 2001; 120:1616 21. 13. Afessa B. Predicting 3-day and 7-day outcomes of weaning from mechanical ventilation. Chest 1999;116:456 61.