BS EN :2003 Containment Factor deciphered

Similar documents
ASHRAE-110 Type-Test Report for 1.5m Wide Bench-Type fumetop1500 FH-150 Fume Hood of TopAir, USA Ref: ANSI/ASHRAE

Improving the Effectiveness of LEV A HSL perspective

PROPAGATION OF LONG-PERIOD WAVES INTO AN ESTUARY THROUGH A NARROW INLET

Mott Manufacturing Limited

Riverboat Simulator Activity

Measuring range Δp (span = 100%) Pa

CONSIDERATION OF DENSITY VARIATIONS IN THE DESIGN OF A VENTILATION SYSTEM FOR ROAD TUNNELS

Maximum precision for quantifiable success.

CORESTA RECOMMENDED METHOD N 6

Keeping It Simple. Gas Mixtures for Stack Emissions Monitoring.

ESTATES OFFICE OPERATIONS WORKS INSTRUCTIONS WI 6. Fume Cupboards, Ventilated Cabinets, Local Exhaust Ventilation

Comparative Tests on Differential Pressure Control Valves

CISAP-2 2 nd International Conference on Safety & Environment in Process Industry CISAP-2. May, 21-24, 2006, Naples, Italy

Riverboat Simulator Activity Sheet

Biography. Abstract. Keywords. Body box, garments, validation, calibration, clean garments, IEST-RP-CC003.3, garment. Introduction and Scope

TOXIC GAS DETECTORS IN THE WORKPLACE UK REGULATIONS AND STANDARDS

Standing Waves in a String

Energy capture performance

FUME CUPBOARD PERFORMANCE CHECK

Fumehood Performance Test Report

Statesville, NC. Results of AM ASHRAE 110 Laboratory Fume Hood Performance Evaluations Venturi Hood

Rotary air valves used for material feed and explosion protection are required to meet the criteria of NFPA 69 (2014)

c) Loss of product or replacement when units are being serviced All of these are factors that need to be considered by the user.

Teaching Notes. Contextualised task 35 The 100 Metre Race

Physics of Sports. Homewood Campus, JHU Hodson Hall 213 Muhammad Ali Yousuf Assignment 1. Part 1

AN INTRODUCTION TO COMPRESSED AIR QUALITY for BREATHABLE PURPOSES

Naval Postgraduate School, Operational Oceanography and Meteorology. Since inputs from UDAS are continuously used in projects at the Naval

Understanding the How, Why, and What of a Safety Integrity Level (SIL)

You should know how to find the gradient of a straight line from a diagram or graph. This next section is just for revision.

I.CHEM.E. SYMPOSIUM SERIES NO. 97 BUOYANCY-DRIVEN NATURAL VENTILATION OP ENCLOSED SPACES

EDEXCEL NATIONALS UNIT 6 MECHANICAL PRINCIPLES and APPLICATIONS. ASSIGNMENT No. 4

EWGAE 2010 Vienna, 8th to 10th September

DEVELOPMENT AND TESTING OF AIR EXCHANGE RATE METHODS TO IMPROVE COST EFFECTIVENESS WHILE MAINTAINING ACCURACY

NCERT solution Decimals-2

Implementing Provisions for Art. 411 of the ICR Ski Jumping

REAL LIFE GRAPHS M.K. HOME TUITION. Mathematics Revision Guides Level: GCSE Higher Tier

Advanced Hydraulics Prof. Dr. Suresh A. Kartha Department of Civil Engineering Indian Institute of Technology, Guwahati

Marking Guidelines 2009 examination June series. Physics Investigative Skills Assignment (ISA) Q. General Certificate of Education

Figure 2: Principle of GPVS and ILIDS.

Analysis of the Article Entitled: Improved Cube Handling in Races: Insights with Isight

Field verification of on-line HC dew point measurements

STUDY ON TSUNAMI PROPAGATION INTO RIVERS

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Vibration of floors and footfall analysis

Promoting Active Travel to School: Progress and Potential

1. (a) (i) 0.049, 0.39, 0.4, 0.409, B1 for correct order. (b) B1 for (a) B1 for (b) B1 for 0.08.

WING PROFILE WITH VARIABLE FLAP

WAVE FORECASTING FOR OFFSHORE WIND FARMS

Combining Unlike Integers

Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

Aerodynamic Shape Design of the Bow Network Monitoring Equipment of High-speed Train

UNIT 7 PRACTICE PROBLEMS

Opening and closing force guidelines

By far the majority of New Zealand s fisheries are performing well

Universities of Leeds, Sheffield and York

Table of contents. I Drawings 33. A1 Design criteria 1 A1.1 Local thermal discomfort indices... 1

The role of diffuser arrangement and mixing for efficiency optimisation of fine bubble aeration systems

EXPERIMENTAL STUDY OF WIND PRESSURES ON IRREGULAR- PLAN SHAPE BUILDINGS

Name Date. 5. In each pair, which rational number is greater? Explain how you know.

Guidance on room integrity testing and its interpretation

The Daily Record. 42 The National Strategies Secondary Mathematics exemplification: Y7 NUMBER. Place value, ordering and rounding

Mott Manufacturing Limited

SHOT ON GOAL. Name: Football scoring a goal and trigonometry Ian Edwards Luther College Teachers Teaching with Technology

APPENDIX A SAMPLE FHT SPECIFICATION (This recommended FHT specification is available from

Measurement of Velocity Profiles in Production Wells Using Spinner Surveys and Rhodamine WT Fluorescent Tracer; Geothermal Field (California)

WANAKA SWIMMING FACILITIES STATEMENT OF PROPOSAL

2. Write a decimal that is equivalent. 4. Which of the following is NOT equal to ½? a. 0.5

Legendre et al Appendices and Supplements, p. 1

Homeostasis and Negative Feedback Concepts and Breathing Experiments 1

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

Traceable calibration of automatic weighing instruments in dynamic operation

Quantifying the impact of cover depth and panel width on longwall shield-strata interactions

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS General Certificate of Education Ordinary Level

ACCUMULATION OF HYDROGEN RELEASED INTO A VENTED ENCLOSURE - EXPERIMENTAL RESULTS

CFD SIMULATIONS OF GAS DISPERSION IN VENTILATED ROOMS

P2a Force, Acceleration and Graphs

Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA

Chapter 2 Displaying and Describing Categorical Data

Drilling Efficiency Utilizing Coriolis Flow Technology

Determination of the wind pressure distribution on the facade of the triangularly shaped high-rise building structure

Brooks Flow Controllers for Gas and Liquid Service Model FC 8744, Series FC 8800 and FC 8900

Flow in a shock tube

National Unit Specification: General Information

Your local gas generation partner. Precision series Modular gas generation solution for GC.

RESEARCH OF LEAK DETECTION SYSTEMS WITH ACCUMULATIVE MEMBRANES

STIFFNESS INVESTIGATION OF PNEUMATIC CYLINDERS. A. Czmerk, A. Bojtos ABSTRACT

High Speed Rail Tunnel Aerodynamics: Transient pressure and loadings on fixed tunnel equipment

REVIEW TEST Find the least common multiple (LCM) of the numbers 4, 18. A) 4 B) 2 C) 72 D) 1 E) 36

of 6. Module 5 Ratios, Rates, & Proportions Section 5.1: Ratios and Rates MAT001 MODULE 5 RATIOS, RATES, & PROPORTIONS.

MECHANICAL EQUIPMENTS: COMPRESSORS, PUMPS, SEALS, SPEED DRIVES, CONTROL VALVES & ACTUATORS & SAFETY RELIEF VALVES

2 Asymptotic speed deficit from boundary layer considerations

SPIRIT III Radiometer Saturation Effect

Instrument Performance Evaluation, comparing ISO and NORSOK I-106 Dr Paul Holland, EffecTech Group & Jon Carlsen, Norsk Analyse North Sea Flow

Technical Paper. Selection of Particulate Monitors. Author: William Averdieck, PCME Ltd

MICROPHONE WIND SPEED LIMITS DURING WIND FARM NOISE MEASUREMENTS

SEASONDE DETECTION OF TSUNAMI WAVES

Numerical Simulations of a Train of Air Bubbles Rising Through Stagnant Water

3. A shirt is priced at $32.00 now. If the shirt goes on sale for 30% off the current price, what will be the sale price of the shirt?

1) The loop is almost completely in the field. 2) The speed of the loop is tripled. 3) The field direction is reversed.

Transcription:

BS EN 14175-3:2003 Containment Factor deciphered Waldner Knowledge Base January 2011 Please let us have your feedback on this publication: steffen.springer@waldner.co.uk +44 7795010353

Content 1. Definition of the Containment Factor C F1 according to BS EN14175 2. Interpretation of the Results 3. Discussion: Is the Containment Factor objective and conclusive when comparing fume cupboard performance (or: Is C F1 suitable as a specification performance / selection criteria)? 4. Summary and Conclusion 5. Sample Calculations Page 2

Definition of Containment Factor According to BS EN 14175-3:2003, the dimensionless containment factor C F1 can be optionally (the code does not demand the use of this factor as part of the test certification) calculated as: q... Flow rate of tracer gas (SF 6 note that the tracer gas flow rate is only 10% of the test gas flow rate because a mixture of 90% N 2 and 10% SF 6 is used) Q... Fume cupboard extract flow rate Φ 1... Mean tracer gas concentration as volume fraction rounded to the second decimal place Page 3

Interpretation of the Results Structure of the equation is very simple Theoretically, the result means: the higher the factor the higher the containment of the fume cupboard. Magnitude of containment factor is dependent in three values: flow rate of tracer gas, extract flow rate of fume cupboard and mean tracer gas concentration (the latter is the only value in this equation that relates to the ability of fume cupboard to retain). According to EN, C F1 should be rounded to the nearest integer and it must be indicated if the result is limited by the detection limit of the instrument Why? Just read on! Page 4

The Big Question Does the containment factor deliver an objective, conclusive and independent result to judge and compare the true containment of fume cupboards (specification selection criteria)? Page 5

Discussion The influence of the extract flow rate Q is in the denominator of the equation Hence, the larger Q the smaller C F1 will be Everything else being equal, this indicates a flaw. Surely, the containment of a fume cupboard will increase with increasing extract flow rates (volumes or face velocity). With q, the tracer gas flow rate, being always constant (stipulated in EN) and an assumed detection limit of 0.01ppm of the measuring equipment, the only variable remaining is the extract volume of the fume cupboard. Is this result (C F1 calculated as described before) suitable for an objective comparison of fume cupboard containment? Page 6

Discussion The influence of the detection limit The detection limit of the equipment is a crucial part of the EN testing regime as the market demands threshold values in the parts per million or even parts per billion range BS EN 14175 clearly stipulates that the mean tracer gas concentration as a volume fraction of the test period must be calculated and rounded to the second decimal place. Hence, if BS EN 14175 is applied, any mean tracer gas concentration value below 0.01 must be rounded to 0.01 and certified accordingly. In terms of the containment factor this means that certified results are dependent on the detection limit of the instrument used. Does this allow an objective (performance based) comparison / selection? Page 7

Summary and Conclusion The containment factor according to BS EN 14175 depends on three variables, whereby only one (mean tracer gas concentration) is related to the specific ability of the fume cupboard to contain. The actual magnitude of the containment factor also fluctuates with the extract flow rate used and the detection limit of the instrument applied. As the illustration on the next page shows, the containment factor is not suitable as an objective indication for fume cupboard containment unless all models in question are tested under the same conditions, with the same extract volumes and with same test equipment. Only one of these variables not being the same, will lead to different containment factors. In practical terms, the containment factor does not enable the user to select the best cupboard based on containment performance. Page 8

Sample Calculation STANDARD F/C (UK) SECUFLOW (UK) SECUFLOW (Europe 270m 3 /h/m width) 500 v =0.50m/s Q=1440m 3 /h v =0.30m/s Q=864m 3 /h v =0.15m/s Q=432m 3 /h 100 100 100 1800 1800 1800 C F1 C F1 C F1 q = 0.20 l/min 1) φ = 0.01ppm 2) 833 1,389 2,778 q = 0.20 l/min 1) φ = 0.10ppm 2) 83 139 278 1) Tracer gas flow rate to BS EN 14175 for inner grid measurement, 2.0 l/min test gas = 0.2 l/min tracer gas (90% N 2 +10% SF 6 ) 2) assumed detection limit Page 9