TRACE LEVEL DETERMINATION OF PER- AND POLYFLUOROALKYL SUBSTANCES (PFASs) IN WATER USING THE AGILENT 6460 LC/MS/MS

Similar documents
Occurrence of perfluoroalkyl compounds in surface waters from the North Pacific to the Arctic Ocean

Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083

TREATMENT MITIGATION STRATEGIES FOR POLY- AND PERFLUOROALKYL CHEMICALS. December 14, Objective

Perfluorinated compounds in marine surface waters: data from the Baltic Sea and methodological challenges for future studies

Supporting Information. (32 Pages) Bioconcentration of aqueous film-forming foam (AFFF) in juvenile rainbow. trout (Oncorhyncus mykiss)

The Analysis of Per (poly) Fluorinated Compounds Beyond UCMR3

Method Code: NH0278 Method Ref: SOP 454-PFAAS-ISOTOPE DILUTION Revision: Date:

Sampling & Analytical Considerations. Jim Occhialini Alpha Analytical

ASSESSMENT OF HEALTH RISKS AND HUMAN EXPOSURE ASSOCIATED WITH PERFLUORINATED COMPOUNDS (PFCs) IN TAP WATER FROM CHINA AND OTHER COUNTRIES

Benetton PFCs investigation report December 2015

Supplermentary Date to IEAM vol 7. no. 4, pp SUPPLEMENTAL DATA

City of Satellite Beach Per and Polyfluoroalkyl Substances (PFAS) Water Sampling Satellite Beach, Florida

Benetton Group srl PFCs Elimination Progress

Brick Utilities PFOA Source Trackdown Study

Introduction. Experimental

Automated Low Background Solid Phase Extraction of Perfluorinated Compounds in Water. Ruud Addink Fluid Management Systems Watertown MA

Perfluorerade organiska ämnen i serum från förstföderskor i Uppsala tidstrend

Case Study Phase out of Perfluorinated Compounds (PFCs) H&M Group April 2017

PFOS, PFOA and Other Fluorinated Compounds: Sampling and Analy<cal Challenges

Environmental Management of Firefighting Foam

Overview of Perfluorinated Chemicals (PFCs) and Related International Initiatives

PFCs/PFAS: Emerging Environmental Contaminants & NJDEP Update

Perfluorinated Compounds Treatment and Removal

Perfluorinated Compounds Prevalence and Assessment in Drinking Water

Perfluorinated Compounds (PFCs) Perfluorooctanoic Acid (PFOA) Perfluorooctane Sulfonate (PFOS) Focus On Change 2017

N Korea Japan China Thailand Taiwan Philipines EP-5 EP km. Source: Google Maps and

ANALYSIS REPORT. Report Date: April 27, :34. Project: Hoosick Falls WTP

Maine Department of Environmental Protection (MEDEP) Stone Farm, Arundel Sample Collection - Data Report Summary Report Date: June 15, 2017

Perfluorinated compounds in red-throated divers from the German Baltic Sea: new findings from their distribution in 10 different tissues

ANALYSIS REPORT. Report Date: February 16, :56. Project: SGPP - McCaffrey Street

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

DES Waste Management Division 29 Hazen Drive; PO Box 95 Concord, NH

Addendum 1 Lab Testing Services April 4, 2018

The ABCs of PFCs in Water Supplies

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis

PFC Analysis Kit for ACQUITY UPLC System Guide

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

Extended Application Note

Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform

LCMSMS Solutions For Steroid Analysis

May 1, By John Heim,Doug Staples

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

River Water Intake Collected: 20 June WW Loading Collected: 20 June 2018

The instrument should have a mass range from 2 to 2000 m/z or better Interface

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

Longterm Detection of Anabolic Steroid Metabolites in Urine

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

May 31, Mr. Thomas Touseau SAU 26, Merrimack School District Facilities Director Facilities Management 36 McElwain Street Merrimack, NH 03154

FINAL SITE INSPECTION REPORT PATRICK AIR FORCE BASE, FL

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Retention Time Locking: Concepts and Applications. Application

Renzo BAGNATI Mario Negri Institute of Pharmacological Research- IRCCS Milano - Italy

Rapid and Reliable Detection of Dissolved Gases in Water

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Study on the Generation of Perfluorooctane Sulfonate from the Aqueous Film-Forming Foam

Validation of the MRM pesticides from the Working Document SANCO/12745/2013 of November 2017 rev. 9(1)

Methylprednisolone detection in urine following local and oral administrations

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

Experiment GC : Analysis of BTEX by GC-FID

Video Notes. LC Troubleshooting Series Retention Time Shifts

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

Carrier Gases in Capillary GC

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

1. INTRODUCTION CH 3 CH3

GCMSD-Headspace Analysis SOP

Automated Determination of Dissolved Gases in Water Anne Jurek

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Agilent Dimension Software for ELSD User Manual

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

Gas Chromatography-Mass Spectrometry Analysis Of Organics In Drinking Water: Concentrates And Advanced Waste Treatment Concentrates, Vol.

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

PROJECT ID: AP0220 LABORATORY REPORT NUMBER: L PO NUMBER:

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Artificial sweeteners as tracers of septic wastewater in the environment. Collaborators

Fast method for Ginseng Analyses using Agilent Poroshell 120 Columns Scaled from a Traditional Method

ABSTRACT. Perfluorinated compounds (PFCs) are surfactants used in the manufacturing of numerous

CORESTA RECOMMENDED METHOD Nº 67

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

Setting a New Standard in Flash Chromatography Performance

Supporting information for J. Med. Chem., 1994, 37(5), , DOI: /jm00031a018

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

SUPPORTING INFORMATION

ABSTRACT. bonds are replaced with carbon-fluorine bonds. PFCs are used in the production of

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Transcription:

ENVIRONMENTAL MARKET TRACE LEVEL DETERMINATION OF PER- AND POLYFLUOROALKYL SUBSTANCES (PFASs) IN WATER USING THE AGILENT 6460 LC/MS/MS A method for the determination of per- and polyfluoroalkyl substances (PFASs) in various water types has been developed on the Agilent 6460 LC/MS/MS. Samples are prepared using solid phase extraction. Author Lutz Ahrens SLU, Swedish University of Agricultural Sciences, Uppsala, Sweden

ABSTRACT A method for the determination of per- and polyfluoroalkyl substances (PFASs) in various water types has been developed on the Agilent 6460 LC/MS/MS. Samples are prepared using solid phase extraction. The analytical suite contains 26 PFASs including 4 perfluoroalkane sulfonates (PFSAs) (PFBS, PFHxS, PFOS, PFDS), 13 perfluoroalkyl carboxylates (PFCAs) (PFBA, PFPeA, PFHxA, PFHpA, PFOA, PFNA, PFDA, PFUnDA, PFDoDA, PFTriDA, PFTeDA, PFHxDA, PFOcDA), 3 perfluorooctane sulfonamides (FOSAs) (FOSA, MeFOSA EtFOSA), 2 perfluorooctane sulfonamidoethanols (FOSEs) (MeFOSE, EtFOSE), 3 perfluorooctane sulfonamidoacetic acids (FOSAAs) (FOSAA, MeFOSAA, EtFOSAA) and fluorotelomer carboxylate (6:2 FTSA). The range of application for this method is 0.05 40 ng/ml of concentrated extract, which is equivalent to 0.05 40 ng/l of unextracted water. Method detection limits are less than 2 ng/l for all compounds. INTRODUCTION Per- and polyfluoroalkyl substances (PFASs), which are also commonly known as PFCs (per- and polyfluorinated chemicals), comprise a diverse group of man-made chemicals that are widely used as processing additives during fluoropolymer production and as surfactants in consumer applications, including surface coatings for textiles, furniture and paper products, as well as in some types of fire-fighting foam [1]. PFASs are known to be persistent, bioaccumulative and have possible adverse effects on humans and wildlife [2]. As a result, PFOS (perfluorooctane sulfonate) and its salts were added to the Stockholm Convention on Persistent Organic Pollutants (POPs) in May 2009 [3]. However, their homologues, neutral precursor compounds and new PFAS classes continue to be produced. The Swedish Drinking Water Guideline includes 11 PFASs. These are perfluorobutane sulfonate (PFBS), perfluorohexane sulfonate (PFHxS), PFOS, 6:2 fluorotelomer sulfonate (6:2 FTSA), perfluorobutanoate (PFBA), perfluoropentanoate (PFPeA), perfluorohexanoate (PFHxA), perfluoroheptanoate (PFHpA), perfluorooctanoate (PFOA), perfluorononanoate (PFNA) and perfluorodecanoate (PFDA). The action value for the sum of these 11 PFASs is 90 ng/l and the health risk limit is 900 ng/l (Swedish National Food Agency, 2015). The analytical technique of choice for the determination of PFASs in water is high performance liquid chromatography coupled with tandem mass spectrometry using the Agilent 6460 (LC/MS/MS). This solution note outlines a robust, sensitive method that has been developed for the routine determination of 26 PFASs for a variety of water types, including marine water, surface water from lakes and streams, groundwater, wastewater (influent and effluent), landfill leachate, surface run-off, precipitation, snow and ice. ANALYTICAL TECHNIQUE Standards Calibration standards prepared in methanol containing 26 target compounds and 16 mass-labelled internal standards Concentration of calibration standards 0.05, 0.25, 1, 4, 8 and 40 ng/ml Concentration of mass-labelled internal standards 2 ng/ml Mass-labelled injection standard 2 ng/ml Sample Preparation 0.5 L of water sample extracted using solid phase extraction (SPE) and eluent concentrated to 1 ml Instrumentation Agilent 1200 Series HPLC System Agilent 6460 Triple Quadrupole System 2

Agilent 1200/6460 LC/MS/MS Operating Conditions LC Conditions Analytical Column Betasil C18 50 x 2.1 mm, 5 µm Guard Column Hypersil Gold 10 x 2.1 mm, 5 µm Column Temperature 25 C Injection Volume 10 µl Mobile Phase Flow Rate Run Time A = MilliQ + 10 mm NH 4 OH B = Methanol + 10 mm NH 4 OH 0.35 0.40 ml/min 20 min MS Conditions Acquisition Mode MRM, ESI Negative Gas Temperature 300 C Gas Flow 7 L/min Nebuliser Pressure 30 psi Sheath Gas Temperature 250 C Sheath Gas Flow 11 L/min Capillary Voltage 3500 V Nozzle Voltage 0 V RESULTS AND DISCUSSIONS An Agilent 6460 LC/MS/MS system was used to determine a group of 26 PFASs, consisting of 4 perfluoroalkane sulfonates (PFSAs) (PFBS, PFHxS, PFOS, PFDS), 13 perfluoroalkyl carboxylates (PFCAs) (PFBA, PFPeA, PFHxA, PFHpA, PFOA, PFNA, PFDA, PFUnDA, PFDoDA, PFTriDA, PFTeDA, PFHxDA, PFOcDA), 3 perfluorooctane sulfonamides (FOSAs) (FOSA, MeFOSA EtFOSA), 2 perfluorooctane sulfonamidoethanols (FOSEs) (MeFOSE, EtFOSE), 3 perfluorooctane sulfonamidoacetic acids (FOSAAs) (FOSAA, MeFOSAA, EtFOSAA) and fluorotelomer carboxylate (6:2 FTSA) (Tables 1 and 2). Class Name PFSAs perfluoroalkane sulfonates PFCAs perfluoroalkyl carboxylates FOSAs perfluorooctane sulphonamides FOSEs perfluorooctane sulfonamidoethanols FOSAAs perfluorooctane sulfonamidoacetic acids X:2 FTSAs X:2 fluorotelomer carboxylates Table 1. Classes of Compounds within the PFAS Analytical Suite. 3

The mass-spectrometer was operated in MRM negative ionisation mode, and all compounds were determined in a single analytical run. Table 2 lists the compounds in the analytical suite, along with their retention times and method detection limits (MDLs) based on solid phase extraction (SPE) of 0.5 L of water and extract concentration to 1 ml. Peak RT (min) Compound Abbreviation Class MDLs (ng/l) 1 3.9 perfluorobutanoate PFBA PFCAs 0.25 2 5.3 perfluoropentanoate PFPeA PFCAs 0.03 3 5.6 perfluorobutane sulfonate PFBS PFSAs 0.22 4 6.5 perfluorohexanoate PFHxA PFCAs 0.09 5 7.6 perfluoroheptanoate PFHpA PFCAs 0.05 6 7.7 perfluorohexane sulfonate PFHxS PFSAs 0.15 7 8.6 6:2 fluorotelomer sulfonate 6:2 FTSA X:2 FTSAs 1.79 8 8.6 perfluorooctanoate PFOA PFCAs 0.40 9 9.6 perfluorononanoate PFNA PFCAs 0.08 10 9.6 perfluorooctane sulfonate PFOS PFSAs 0.21 11 10.3 perfluorooctane sulfonamidoacetic acid FOSAA FOSAAs 0.25 12 10.4 perfluorodecanoate PFDA PFCAs 0.19 13 10.7 N-methylperfluorooctane sulfonamidoacetic acid N-MeFOSAA FOSAAs 0.25 14 10.8 perfluorooctane sulfonamide FOSA FOSAs 0.11 15 11.1 perfluoroundecanoate PFUnDA PFCAs 0.16 16 11.1 N-ethylperfluorooctane sulfonamidoacetic acid N-EtFOSAA FOSAAs 0.25 17 11.1 perfluorodecane sulfonate PFDS PFSAs 0.25 18 11.7 perfluorododecanoate PFDoDA PFCAs 0.19 19 12.0 N-methylperfluorooctane sulfonamide N-MeFOSA FOSAs 0.05 20 12.0 N-methylperfluorooctane sulfonamido-ethanol N-MeFOSE FOSEs 1.00 21 12.3 perfluorotridecanoate PFTriDA PFCAs 0.05 22 12.4 N-ethylperfluorooctane sulfonamide N-EtFOSA FOSAs 0.05 23 12.4 N-ethylperfluorooctane sulfonamido-ethanol N-EtFOSE FOSEs 0.25 24 12.8 perfluorotetradecanoate PFTeDA PFCAs 0.05 25 13.5 perfluorohexadecanoate PFHxDA PFCAs 0.05 26 14.1 perfluorooctadecanoate PFOcDA PFCAs 0.25 Table 2. Retention Times for individual PFASs, Abbreviation, Compound Class and MDLs (see Figure 1 for the Chromatogram) The calibration range for each compound was 0.05 40 ng/ml, with standards at 0.05, 0.25, 1, 4, 8 and 40 ng/ml. The internal standards were spiked into the standards and samples at a concentration of 2 ng/ml. Figure 1 shows a typical chromatogram for a 16 ng/ml standard, and Figure 2 shows the compound information for PFOA (using 13 C 4 PFOA as the internal standard) alongside a typical PFOA calibration curve. 4

Figure 1. 16 ng/ml Standard Chromatogram (see Table 2 for the PFAS of the corresponding peak). Figure 2. PFOA Compound Information and Calibration Curve (0.05 40 ng/ml). A surface water sample chromatogram is shown in Figure 3. This sample contains a total of 10 different PFASs out of the 26 target compounds in this suite. Figure 3. Surface Water Sample Chromatogram (see Table 2 for the PFAS of the corresponding peak). 5

CONCLUSIONS A robust, sensitive method has been developed for the determination of 26 PFASs in water samples using SPE and analysis on the Agilent 6460 LC/MS/MS. This method demonstrates excellent linearity and sensitivity, with a calibration range of 0.05 40 ng/ml of concentrated extract, which is equivalent to 0.05 40 ng/l of unextracted water. The method achieves MDLs of less than 2 ng/l for all compounds. REFERENCES 1. L. Ahrens, Polyfluoroalkyl compounds in the aquatic environment: a review of their occurrence and fate, J. Environ. Monit., 2011, 13, 20 31. 2. L. Ahrens, M. Bundschuh, Fate and effects of poly- and perfluoroalkyl substances in the aquatic environment: A review, Environ. Toxicol. Chem., 2014, 33, 1921 1929. 3. Stockholm Convention on Persistent Organic Pollutants, Report of the Conference of the Parties of the Stockholm Convention on Persistent Organic Pollutants on the work of its fourth meeting, UNEP/POPS/COP.4/38, 8 May 2009. Solutions for your analytical business Markets & Applications Programs www.solutions-to-win.com This information is subject to change without notice. Agilent Technologies, Inc. 2016 Printed in EU, 2016-10-13 5991-7532EN