MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 40 Ultra-Sensitive Kit

Similar documents
MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 42 Ultra-Sensitive Kit

MSD 96-Well MULTI-ARRAY CRP Assay

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

ab IgG1 Human ELISA Kit

RayBio Human vwf ELISA Kit

ab FGF basic (FGF2) Human ELISA Kit

RayBio Human TNF-alpha ELISA Kit

ab99968 Adiponectin Human ELISA Kit

ab Sonic Hedgehog Human ELISA Kit

biosensis Mouse CXCL10/IP-10 ELISA Kit Protocol

ab HB EGF Human ELISA Kit

HUMAN IL6 KITS PROTOCOL

ab IL-5 (Interleukin-5) Mouse ELISA Kit

RayBio Mouse bfgf ELISA Kit

ab VEGF Human ELISA Kit

biosensis Human IGF-II, Insulin-like growth factor II, Somatomedin-A ELISA Kit Protocol

biosensis Rat Fibronectin ELISA Kit Protocol

biosensis Human Soluble Tumor Necrosis Factor receptor type II (stnfrii) ELISA Kit Protocol

biosensis Human TNFα/Cachectin/TNFSF2 ELISA Kit Protocol

biosensis Mouse Vascular endothelial growth factor A/VEGF-A/VEGF-164/VEGF-1/VEGF- 120/VEGF-2 ELISA Kit Protocol

RayBio Human Adiponectin ELISA Kit

RayBio Human BDNF ELISA Kit

RayBio Human IFN alpha/beta R2 ELISA Kit

biosensis Mouse Brain-derived neurotrophic factor (BDNF) ELISA Kit Protocol

ab TCA-3 (CCL1) Mouse ELISA Kit

biosensis Rat Interleukin-1 beta, IL-1β ELISA Kit Protocol

biosensis Human Lipocalin-2/NGAL ELISA Kit Protocol

biosensis Mouse Interleukin-1 beta (IL-1β) ELISA Kit Protocol

biosensis Rat Glial cell line-derived neurotrophic factor/gdnf total /ATF ELISA Kit Protocol

Human ABCD1 ELISA KIT

Rat myeloperoxidase-antineutrophil cytoplasmic antibody(mpo-anca) ELISA Kit

CGMP KITS PROTOCOL ASSAY PRINCIPLE

ab IL-4 (Interleukin-4) Mouse ELISA Kit

biosensis Rat IGF-1/Somatomedin/Insulin-like growth factor ELISA Kit Protocol

Human Factor Xa Chromogenic Activity Kit

R-PLEX Antibody Sets Multiplex Assays

R-PLEX Antibody Sets Singleplex Assays

F7 (Human) Chromogenic Activity Assay Kit

CspB ELISA Kit DAS ELISA for the detection of CspB protein Catalog number: PSP 35500

KNG1 (Human) ELISA Kit

Bt-Cry2Ab and Bt-Cry3Bb1 Multi-trait ELISA Kit DAS ELISA for the detection of Bt-Cry2Ab and Bt-Cry3Bb1 proteins Catalog number: PSM 26800

Human Factor X Chromogenic Activity Kit

Bt-Cry1Ac & Bt-Cry2A Multi-trait ELISA Kit DAS ELISA for the detection of Bt-Cry1Ac & Bt-Cry2A proteins Catalog number: PSM 14900

Peninsula Laboratories International, Inc. 305 Old County Rd., San Carlos, CA 94070, USA Tel: (650) Fax: (650)

ab Haptoglobin Human ELISA Kit

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

Pipetting and Determining Protein Concentration

96-well Serum/Plasma Fatty Acid Kit Non-Esterified Fatty Acids Detection 100 point kit Cat# SFA-1

Factor X Human Chromogenic Activity Assay Kit

SAM510: SAM Methyltransferase Assay A Non Radioactive Colorimetric Continuous Enzyme Assay

96-well Serum/Plasma Fatty Acid Kit Non-Esterified Fatty Acids Detection 500 Point Kit Cat# SFA-5

TF (Human) Chromogenic Activity Assay Kit

Cat. No. S-1328 Hepcidin-25 (human)

NGAL Rapid ELISA Kit. NGAL Rapid ELISA Kit. For Research Use Only Not for use in diagnostic procedures KIT 037. Diagnostics

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor

Technical Procedure for General Laboratory Equipment

ab Androstenedione ELISA Kit

Biological Activity Assay for Gibco Recombinant Proteins Chemotaxis Assay

ab Testosterone ELISA Kit

QIAGEN Supplementary Protocol

Multiplate Platelet Function Analyzer: An Overview

Best Practices for Infinium Assays

Ocular Irritation: Dilution Method

Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts

MATERIAL SAFETY DATA SHEET. Known Hazardous Components CAS Number Percent. Stop Solution 99% Known Hazardous Components CAS Number Percent

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

PROTOCOLS. Coomassie Brilliant Blue G (Sigma 98%) ,4 mg. Ethanol 95% ml. Phosphoric acid 85%.(w/v) ml

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

Mixing Methacholine Dilutions for Bronchoprovocation Challenge Testing

Lab Equipment ANALYTICAL BALANCE

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs

Fungi-Plex Kit HOW TO USE IT F E C C. Wash Buffer [F] Standard Diluent Buffer [E] Assay Diluent Buffer [D] Mycotoxin Standard Mix

A Flexible, Multi-Purpose, Multi-Detector HTS Automated Platform

Degassing ERC KK Saitama, Japan

Analysis of Casein and Whey Protein in Whole, 2%, and Skim Milk by Capillary Gel Electrophoresis

Convergent Technologys product overview

Artel MVS as a Tool For Measuring Liquid Mixing Efficacy in Microplates

MVS Volume Verification Using Any Microtiter Plate or Small Volume Container, Such as a Tube or Vial

High-Throughput. Faster, Better 96 & 384-Well Pipetting Ultra Efficient and Extremely Versatile

Testosterone ELISA Kit

ANDROSTENEDIONE SALIVA ELISA Direct immunoenzymatic determination of Androstenedione in saliva

Intelligent Pipette Automation made Simple

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Microprojectile Bombardment Protocol

Name:. Correct Questions = Wrong Questions =.. Unattempt Questions = Marks =

VOLUMETRIC TECHNIQUES

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

Liquid Handling Systems

26-G Keewaydin Drive, Salem, NH P: (800) F: (603)

n Budget constraints n Limited benchspace n Interest in running magnetic n User-friendly data management n Benefits resulting from a

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

Performance Verification of BioTek s Precision 2000 using Artel s Multichannel Volume Verification (MVV ) System

CHM250 Calibration and Measurement Lab. Balance Calibration

CORESTA RECOMMENDED METHOD Nº 67

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

Experiment 1 Use of Micropipettes: measuring volume accurately

GCMSD-Headspace Analysis SOP

Triturus Analyzer Instructions for Use

TECHNICAL BULLETIN. TriControls. Control Materials for Liquid QC and Calibration Verification

Transcription:

MSD 96-Well MULTI-ARRAY Human (6E10) Abeta 40 Ultra-Sensitive Kit The first protocol has been optimized for quantifying Aβ 1-40 peptide in human cerebrospinal fluid (CSF). The second protocol has been optimized for quantifying Aβ 1-40 peptide in human plasma samples. Storage Read Buffer T (4X) RT Blocker A (dry powder) RT MULTI-SPOT 96-well, 4-Spot Human Abeta Peptide Plate(s) 2-8ºC Tris Wash Buffer (10X) 2-8ºC SULFO-TAG Anti-Aβ 40 Antibody (50X) 2-8ºC Dimethyl Sulfoxide (DMSO) -20ºC Aβ1-40 peptide (Lyophilized) -20ºC 6E10 The SECTOR Imager data file will identify spots according to their well location, not by the coated capture antibody name. FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC OR THERAPEUTIC PROCEDURES.

Notes: Other Materials & Equipment (not supplied) Deionized water for diluting Tris Wash Buffer and Read Buffer One 250 ml bottle Two 50 ml tubes Two 15 ml tubes Microcentrifuge tubes for making dilutions of peptide standards Automated plate washer, Multidrop, or other efficient multichannel pipetting equipment for washing 96-well plates Appropriate liquid handling equipment for desired throughput that must accurately dispense 25 µl and 150 µl into a 96-well micro plate Protocol at a Glance The CSF protocol takes approximately 3 hours to complete if each reagent is prepared during the preceding incubation. All reagents, with the exception of the diluted peptide, can be prepared ahead of time. This lengthens the overall time required but frees up time during the incubation steps. Read the entire detailed instructions before beginning work. For other sample types, the second protocol should be used as a starting point for optimization of the assay in the chosen matrix. 1. Add Blocker A solution, incubate 30 minutes. 2. Add peptide or samples, incubate 1 hour, wash. 3. Add SULFO-TAG Detection Antibody, incubate 1 hour, wash. 4. Add Read Buffer and analyze plate. Detailed instructions for use with Human Cerebrospinal Fluid (CSF) samples Prepare a stock of 1X Tris Wash Buffer: a) 1X Tris Wash Buffer is used throughout the assay to make other reagents as well as to wash plates. Approximately 250 ml per plate is required more if using an automatic plate washer to account for waste. b) In a 250 ml bottle combine: 25 ml 10X Tris Wash Buffer 225 ml deionized water Prepare 10% Blocker A Solution: a) In a 50 ml tube combine (per plate): 20 ml 1X Tris Wash Buffer 2 g Blocker A A larger amount of Tris Wash Buffer may be prepared at once and stored at room temperature for later use. Solutions containing Blocker A should be kept at 4ºC and discarded after 14 days.

Prepare Antibody Dilution Buffer: a) Prepare 5 ml per plate. b) In a 15 ml tube combine: 500 µl 10% Blocker A Solution 4.5 ml 1X Tris Wash Buffer Notes: Solutions containing Antibody Dilution Buffer should be kept at 4 o C Prepare 1X Detection Antibody solution: a) Prepare 3 ml of Detection Antibody per plate: 60 µl 50X SULFO-TAG Anti-Aβ 40 Antibody 2940 µl Antibody Dilution Buffer Dilute Read Buffer: a) Prepare 2X Read Buffer T solution. b) In a 50 ml tube combine (per plate): 10 ml deionized water 10 ml 4X Read Buffer T Prepare Aβ40 peptide dilutions: a) Check the label on the peptide vial/tube for the correct volume of DMSO to dissolve the peptide vial in to yield a 0.1 mg/ml stock. b) Vortex peptides to ensure that it is entirely dissolved. c) Prepare the following dilution of the stock peptide: Prepare a 10 µg/ml stock by adding 10µL of the 0.1 mg/ml stocks in 90µL of 10% Blocker A Solution Make a 0.1µg/mL stock of each peptide by adding 5µL of the 10 µg/ml solution in 495µL of 10% Blocker A Solution d) Prepare the assay dilution series: Prepare the highest Calibrator by adding 40 µl of the 0.1µg/mL Aβ40 peptide solution and 360µL 10% Blocker A Solution Diluted Read Buffer may be kept in a tightly sealed container at room temperature for later use. The 0.1 mg/ml peptide stock can be aliquotted and stored at -20 o C for up to 2 months. MSD recommends making 20 aliquots of 25 µl/tube. Refreezing, storing more dilute peptides, or storing for longer times are not recommended. To avoid the possibility of aggregation of the standard peptide and/or sticking of the peptide to the dilution tubes, the dilutions should be prepared immediately before use. Prepare the next Calibrator by transferring 100 µl of the highest Calibrator to 300 µl 10% Blocker A Solution Repeat 4-fold serial dilutions 5 additional times to generate 7 Calibrators. It is recommended that both peptide standards and samples be assayed in duplicate.

Notes: This yields the following Calibrator concentrations: Calibrator (pg/ml) Aβ40 Cal 7 10,000 Cal 6 2,500 Cal 5 625 Cal 4 156 Cal 3 39 Cal 2 9.8 Cal 1 2.4 0 0 Use 10% Blocker A Solution for Cal 0. These Calibrators will be sufficient to run an 8-point calibration curve for multiple plates. Do not store diluted Calibrators. STEP 1 Block Plate: a) Add 25µL/well of 10% Blocker A Solution. b) Incubate at room temperature for 30 minutes with shaking. c) Do not wash plate before addition of standards or samples STEP 2 Sample Addition: a) Dispense 25 µl/well of diluted peptide standards or samples. c) Wash plate three times with Tris Wash Buffer. Abeta 40 peptide levels exhibit a broad dynamic range in CSF. Samples may need to be diluted for optimal peptide sensitivity. Incubation may also be done at 4 o C for 24 Hr with parallel assay performance STEP 3 Detection Antibody Addition: a) Add 25 µl/well of 1X SULFO-TAG Detection Antibody. c) Wash plates three times with Tris Wash Buffer. STEP 4 Read Plate: a) Add 150 µl/well of diluted 2X Read Buffer T, being careful to avoid bubbles in the wells. b) Analyze with SECTOR Imager. Read Buffer T contains surfactant, which will easily form bubbles. If bubbles are present in the wells when imaged, the results will be inaccurate.

Detailed instructions for use with Human Plasma Samples Prepare a stock of 1X Tris Wash Buffer: a) 1X Tris Wash Buffer is used throughout the assay to make other reagents as well as to wash plates. Approximately 250 ml per plate is required more if using an automatic plate washer to account for waste. b) In a 250 ml bottle combine: 25 ml 10X Tris Wash Buffer 225 ml deionized water Prepare 3% Blocker A Solution: a) In a 50 ml tube combine (per plate): 20 ml 1X Tris Wash Buffer 600 mg Blocker A (30 mg/ml or 3%) Notes: A larger amount of Tris Wash Buffer may be prepared at once and stored at room temperature for later use. Solutions containing Blocker A should be kept at 4ºC and discarded after 14 days. Prepare Antibody Dilution Buffer: a) Prepare 10.5 ml per plate. b) In a 15 ml tube combine: 3.5 ml 3% Blocker A Solution 7 ml 1X Tris Wash Buffer Prepare 1X SULFO-TAG Detection Antibody: b) Prepare 3 ml of Detection Antibody per plate: 60 µl 50X SULFO-TAG Anti Aβ 40 Antibody 2940 µl Antibody Dilution Buffer Dilute Read Buffer: c) Prepare 1X Read Buffer T solution. d) In a 50 ml tube combine (per plate): 5 ml 4X Read Buffer T 15 ml deionized water

STEP 1 Prepare Aβ peptide dilutions: a) Check the label on the peptide vial/tube for the correct volume of DMSO to dissolve the peptide vial in to yield a 0.1 mg/ml stock. b) Vortex peptides to ensure that it is entirely dissolved. c) Prepare a dilution series of the peptide. Prepare the following dilutions in the appropriate diluent: 10 µg/ml: 10µL of the 0.1 mg/ml stock in 90 µl of diluent 0.1µg/mL: 5µL of the 10 µg/ml solution in 495µL of diluent 10,000 pg/ml: 40 µl of the 0.1µg/mL solution plus 360 µl diluent 3,160 pg/ml: 100 µl of the 10,000 pg/ml solution plus 216 µl diluent 1,000 pg/ml: 100 µl of the 3,160 pg/ml solution plus 216 µl diluent 316 pg/ml: 100 µl of the 1,000 pg/ml solution plus 216 µl diluent 100 pg/ml: 100 µl of the 316 pg/ml solution plus 216 µl diluent 32 pg/ml: 100 µl of the 100 pg/ml solution plus 216 µl diluent 10 pg/ml: 100 µl of the 32 pg/ml solution plus 216 µl diluent 3.2 pg/ml: 100 µl of the 10 pg/ml solution plus 216 µl diluent 1 pg/ml: 100 µl of the 3.2 pg/ml solution plus 216 µl diluent 0.32 pg/ml: 100 µl of the 1 pg/ml solution plus 216 µl diluent 0.1 pg/ml: 100 µl of the 0.32 pg/ml solution plus 216 µl diluent 0 pg/ml: diluent alone Block Plate: a) Add 150µL/well of 3% Blocker A Solution. b) Incubate at room temperature for 1 hour with shaking. c) Wash plate three times with Tris Wash Buffer. Notes: The 0.1 mg/ml peptide stock can be aliquotted and stored at -20 o C for up to 2 months. MSD recommends making 20 aliquots of 25 µl/tube. Refreezing, storing more dilute peptides, or storing for longer times are not recommended. This dilution scheme gives a halflog titration, in which data points will be evenly spaced when plotted on a log scale. Peptides should be diluted in 10% MSD Blocker A in 1X Tris Wash Buffer. If the calibration curve is to be used to quantify peptide in human CSF, see Detailed Instructions for use with human Cerebrospinal Fluid (CSF) samples. For other complex matrices (culture supernatant, serum, etc.) a different diluent may be desired. For best results, the peptide should be prepared in a diluent that mimics the sample matrix as closely as possible. STEP 2 STEP 3 STEP 4 Sample Addition: a) Dispense 25 µl/well of diluted peptide standards or K 3 EDTA plasma samples. c) Wash plate three times with Tris Wash Buffer. Detection Antibody Addition: a) Add 25 µl/well of 1X SULFO-TAG Detection Antibody. c) Wash plates three times with Tris Wash Buffer. Read Plate: a) Add 150 µl/well of diluted 1X Read Buffer T, being careful to avoid bubbles in the wells. b) Analyze with SECTOR Imager. Read Buffer T contains surfactant, which will easily form bubbles. If bubbles are present in the wells when imaged, the results will be inaccurate.