Air Removal from the Sterilization Chamber

Similar documents
What is the D-ifference in D-value?

Discussion and guidance on the definition and qualification of porous loads

Shukla Medical. TITLE: Manual Surgical Orthopedic Instruments Recommendations for Care, Cleaning, Maintenance and Sterilization

Can Temperature And Pressure Datalogging Replace The Bowie and Dick Test? Dr. Brian Kirk

DEFENSE LOGISTICS AGENCY LAND AND MARITIME P.O. BOX 3990 COLUMBUS, OHIO

Application Note 13: Electron Microscope Decontamination

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

User Guide. Pall Laboratory Manifold. For laboratory use. Not for use in a manner other than indicated. Introduction. Regulatory References

Leak Checking Large Vacuum Chambers

Risk Analysis and Hazard Reduction Report V1.0

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Steam penetration in steam sterilization processes

Related Products: Auto Drain Valve. Model/Specifications. Model

H 2 O 2 Bio-Decontamination and Aseptic Transfer of Heat Sensitive Materials Sergio Mauri, Director, Global Marketing and Business Intelligence

Aseptic Vial and Syringe Filling Narlin B. Beaty, Ph.D. Adelaide Nye

Fig 12, Fig 14HP, Fig 16, Fig 16HP and Fig 16L Strainers

BT6HC Hygienic Sanitary Balanced Pressure Steam Trap for High Capacity and CIP/SIP Applications

Prof AH Basson, Pr Eng March Departement Meganiese en Megatroniese Ingenieurswese. Department of Mechanical and Mechatronic Engineering

Assembly and Installation Procedures

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

Fig 1, Fig 12, Fig 13, Fig 14HP, Fig 16, Fig 16HP and Fig 16L Strainers

EtO Micro-Dose Sterilization

EVALUATING KEGGING OPERATIONS: SEEING INTO THE BLACK BOX. Eric Johann Samp, Ph.D. Chris Nimptsch

MOLEBIO LAB #1: Microquantity Measurement

CONVECTION SECTION FAILURE ANALYSIS AND FITNESS-FOR-SERVICE ASSESSMENT

IC67 - Pre-Instructional Survey

COMMON PRECAUTIONS Data Sheet - Be Sure To Read Before Handling Page 1 of 6

Sterilization Liquid Loads Description. Daniel Clifford, Product Manager Life Science Australia and New Zealand

8.1 Calibration of gas sensors

MIL-STD-883G METHOD

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

Repair Alert Never use compressed air to dismantle a hydraulic cylinder

Steritech: A World Leader in Sterilisation and Decontamination

Natural Gas Conditioning Skids & Equipments

Oxygen Contamination

longitudinal wave Figure 1 amplitude transverse wave Figure 2

Suction Regulators. Recommendations for Cleaning, Disinfection and Reprocessing: Boehringer 3800 &3900 Series Suction Regulators

Installation and Operation 370ESP (Electric Single-Point Purger) & 370ESPR (Retrofit Purger)

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

Bulletin TCR-104 & 109 Filling and adding to the Glycol pressure system

Operating Guidelines for PermSelect Modules Liquid Contacting

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

3M Steri-Vac Sterilizer/Aerator GSX Series. For Life Science Applications. Product. Profile

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

ACT Prep - Experiments Round 1!

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

The latest automotive systems require innovative leak test methods and fixturing.

Fig 1, Fig 12, Fig 13, Fig 14, Fig 16 and Fig 16L Strainers Installation and Maintenance Instructions

ASAHI Neurovascular Guide Wire

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

LABORATORY FUME HOOD SPECIFICATION AND MAINTENANCE PROGRAM

Operating Manual VSA100A. tina29e1 ( )

ASPHALT PLANT LEVEL 1

Biological Safety Cabinet Guidelines

Kenyana Wesley, Luke Grater, Katie Dionne

Hygea Ultrasonic Validation Kit

Both physical and chemical effects come into play either separately or in combination.

Equipment and Accessories. ROMIL LiQuiSure. Designed for High Purity Containment & Dispensing PURE CHEMISTRY

Corning HYPERStack Cell Culture Vessel Closed System

STANDARD OPERATING PROCEDURES

DESIGN AND INSTALLATION CONSIDERATIONS FOR TSI AEROTRAK VHP-RESISTANT REMOTE PARTICLE COUNTERS

9.0 Equipment Cleaning and Repair

Work Method Statement. Prepared for: The Bolton NHS Foundation Trust Prepared by: Mark Fentiman Date:

Effects of Relative Humidity, Concentration, and Exposure Time on the Efficacy of Chlorine Dioxide Gas Decontamination

Rubber Dock Hose care, Use & Maintenance

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

Institute For Thermal Processing Specialists

Measuring Mass and Volume

HOW TO MANAGE VAPORIZATION IN AN ANALYTICAL SYSTEM By Dean Slejko and Tony Waters

RESOLUTION MEPC.209(63) Adopted on 2 March GUIDELINES ON DESIGN AND CONSTRUCTION TO FACILITATE SEDIMENT CONTROL ON SHIPS (G12)

PRE LABORATORY ASSIGNMENT: Lab Section Score: /10 READ THE LAB TEXT BEFORE ATTEMPTING THESE PROBLEMS!

introduce Grade 10 Earth and Dynamics heat transfers 2). Materials: 3. A hot plate. Procedure: heat it to a boil. 3. Remove flask or table) and

Steam sterilization Hollow Load Process

Aseptic Techniques. Techniques for Sterile Compounding. Pharmacy Technician Training Systems Passassured, LLC

Florida Method of Test for MEASUREMENT OF WATER PERMEABILITY OF COMPACTED ASPHALT PAVING MIXTURES

Proton Monitor Thermal Vacuum Test Procedure GP-B P0635

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

1.0 PURPOSE The purpose of this procedure is to describe the use of basic aseptic technique.

6900 Maintenance Instruction System Flush

Lab Session #4 AN Physical Properties

Gleim ATP FAA Knowledge Test 2011 Edition, 1st Printing Updates December 6, 2010

Gases and Pressure. Main Ideas

Delayed Coker Safe Sampling Tar or Spalled Coke

HVAC Considerations for Gaseous Decontamination of Laboratory Spaces using Hydrogen Peroxide Vapor

DEALING WITH HAZARDOUS SPILLS

TA10A and TA10P Steam Tracing Temperature Control Valves Installation and Maintenance Instructions

1. If grams of the vapor of a volatile liquid can fill a 498 ml flask at o C and 775 mm Hg, what is the molecular mass of the gas?

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

White Paper. Overview of MAP Headspace Gas Measurement: Sample Volume Considerations:

Use and Care Instructions. Bakewell BubbleLevel

CH2250: Techniques in Laboratory Chemistry. Outline Measuring Mass Measuring Volume Significant figures. Mass Measurement

Bubble Chamber Detector System

SEMATECH Provisional Test Method for Pressure Cycle Testing Filter Cartridges Used in UPW Distribution Systems

Installation Operation Maintenance

HM and HM34 Inverted Bucket Steam Traps Installation and Maintenance Instructions

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

INSTruCTIONS FOr SOTa WaTEr O ZONaTOr model WOZ5 CONTENTS INTrOduCTION COmplEmENTary units

Acoustic Pulse Reflectometry Brings an End to Tube Inspection Sampling By Dr. Noam Amir, Chief Technology Officer, AcousticEye

Our technical information is based on Tip: calls attention to special information.

ASME and PED Requirements for Drum Level Instrumentation. Presented by: Amir Hedayatnia Clark-Reliance Corp.

Transcription:

BIOLOGICAL INDICATOR SPECIALISTS TM Volume 5, Number 2 March 2008 Garrett Krushefski Scientific & Technical Services Manager Air Removal from the Sterilization Chamber I recently received a question from a customer asking why we state in the EZTest instructions for use to Place the EZTest biological indicators in a horizontal position When sterilizing by steam, it is imperative to first remove ambient air not only from the chamber but also from the devices being sterilized. Failure to do so will impede steam contact with all surfaces which can result in non-sterile items being used from a seemingly acceptable sterilization cycle. There are two general means for removing ambient air from the sterilization chamber: dynamic air removal or gravity-displacement. Dynamic-air-removal sterilization cycles can be either a prevacuum cycle, in which case air is actively removed by a series of pressure and vacuum phases prior to admission of steam, or by steam-flush-pressure pulsing (SFPP) in which case a series of steam flushes and pulses of pressure are employed. Gravity-displacement is a passive means of air removal in which incoming steam displaces or pushes residual air out through the drain of the chamber. ANSI/AAMI ST79:2006 notes, The dynamic-air-removal cycle is generally preferred to a gravity-displacement cycle because of more efficient air removal, a shorter exposure time at higher temperatures, and a vacuum drying phase, resulting in an overall reduction in cycle time. While the above ANSI/AAMI note is accurate, do not think that you must start upgrading and replacing your gravity-displacement sterilizers with new machinery. Gravity displacement is still an effective means of sterilization as long as one employs a few precautions. Such precautions are not necessary in a dynamic-air-removal process because the air is actively sucked-out of the chamber and load. Without the aid of active air removal, one needs to facilitate removal of air from the load when using gravity-displacement systems. This is achieved by positioning items in such a manner that does not create any potential air traps. To illustrate this point, start 1

with a simple example such as a standard laboratory beaker. Of the three orientations shown in Figure 1, which are appropriate for a gravity-displacement steam sterilizer? Figure 1 A C B NOTE: In all of the photos non-packaged items are shown for visual clarity. In actual practice these items would be in sterilization pouches, allowing for post-exposure protection from contamination. The answer is (B & C). The ambient air that is contained within the beaker is heavier (i.e. more dense) than the steam that will be introduced into the chamber. In the upright orientation (A), the steam will not be able to displace or push the ambient air out of the beaker because the air has no means of escape (i.e. the beaker does not have a drain ). Note that the orientation shown for (B) does allow the air to drain from the beaker. The next example is an Erlenmeyer flask which is a bit trickier. Review the four orientations shown in Figure 2 and decide which are the most appropriate for a gravitydisplacement cycle. 2

Figure 2 D G E F It is easy to see that (F) or (G) is the best option and (D) must be avoided. But what about (E)? This orientation is also problematic because there is a zone within the Erlenmeyer flask where ambient air will become trapped and steam contact to the surfaces in that zone will be impeded (see outline of dashed red zone in Figure 3). A very effective way to demonstrate the effects of trapped air is to run a standard sterilization cycle with three flasks oriented as shown in D, E and F (see Figure 3 below). Place an EZTest biological indicator at each location marked in the photo and incubate the units after exposure. The EZTest from locations 1 and 2 will show surviving spores while the EZTest from location 3 will have no growth. This is a perfect example to show that the parametric data for the cycle (time, temperature, and pressure readings) will indicate acceptable cycle performance but the biological indicator indicates non-sterility. 3

Figure 3 D E F 3 1 2 When conducting a validation several years ago, I performed the above recommended test in a 65-minute exposure. The penetration thermocouples (with attached EZTest BIs) recorded accumulated Fo values in excess of 80 minutes even from locations 1 and 2. The physical data indicated a lethal delivery that was well in excess of what was needed to kill the spore challenge, yet the EZTest from locations 1 and 2 were positive while location 3 was negative. How could the spores survive 80 Fo? The reason is the prevailing condition at locations 1 and 2 was NOT saturated steam; rather, it was 121-123 C dry heat. Without an escape route, the less-dense steam molecules could not displace the heavier ambient air and air pockets (dashed red zones in Figure 3) become trapped in the flasks. The penetration thermocouples recorded temperatures of 121 to 123 C for most of the 65-minute dwell time, but unlike the spores in the biological indicator, the thermocouples can not differentiate between dry heat conditions and saturated steam. When reviewing the cycle data, time, temperature and pressure were all within acceptable levels, thus indicating acceptable cycle performance, yet the spores survived. The biological results indicated non-sterility and did not agree with the physical data which indicated sterilization was achieved. We refer to this condition as a catastrophic sterilization failure. SPORES DON T LIE! Air pockets within the materials being sterilized can not be detected by any instrumentation available today. The only system available to detect these pockets is the biological indicator. Air pockets occur in the worst case to sterilize locations in the 4

product or load. All standards direct the user to challenge the worst case locations with biological indicators. As a final example, consider the lengths of tubing shown in Figure 4. When oriented as shown in (G) multiple air pockets will form at various low spots along the length of the tube. One must carefully coil the tube in a helix shape (H) to facilitate air removal and steam penetration. Figure 4 G H Using colored water, one can illustrate the multiple low spots that exist in (G) where air pockets will form. For Figure 5, both tubes were filled with purple water and placed on the perforated stainless tray. All of the colored water drained from (H). Compare that to the amount and location of water that was unable to escape the (G) configuration. The colored water allows one to visually discern where the trapped air will form pockets and thus, the areas in the tube where steam will not penetrate. 5

Figure 5 G H The water visualization method is a valuable tool that can be used when training personnel to use steam sterilizers. One must position items in the chamber such that if they were filled with water, the water would completely drain from the item. If any water were retained, the positioning is not suitable (as shown for the Erlenmeyer flask in Figure 6). 6

Figure 6 The colored water is unable to drain from the flask and shows where trapped air will collect and impede steam contact. Now back to the original question, Why is it necessary to place the EZTest biological indicators in a horizontal position? The EZTest, by virtue of its design is an excellent detector of trapped air. The EZTest itself is not unlike the Erlenmeyer flask in that it too can harbor ambient air which would impede steam contact with the spores on the inoculated paper strip. Thus, just like the beaker shown in (B), this is the correct orientation for the EZTest BI. Is it necessary to use the horizontal orientation in a dynamic-air-removal sterilization cycle? In dynamic-air-removal cycles, one is relying on the prevacuum or SFPP to remove air from the chamber, the load and the EZTest (which may or may not be in a test pack). Regardless, it is always a good idea to use the horizontal orientation for your EZTest as it does help avoid inadvertent trapping of air, even in dynamic-air-removal systems. 7

Finally, consider Figure 7; can you see the flaw in (I) and why (J) is the preferred set-up? I J Figure 7 Recall the earlier statement, the EZTest itself is not unlike the Erlenmeyer flask in that it too can harbor ambient air which would impede steam contact with the spores on the inoculated paper strip. In both (I & J) above, the Erlenmeyer flask is inverted to the extent that if filled with water, all of the water would drain from the flask. HOWEVER, in (I) the EZTest is in a slight cap-up orientation and if the EZTest biological indicator were filled with water, it would be trapped inside the EZTest. In (J) a slight cap-down orientation prevails (by virtue of the inverted positioning of the Erlenmeyer flask) and thus we have a drain which allows the ambient air from inside the EZTest to escape the unit. SPORES DON T LIE Please email us with topics you would like to see addressed in Spore News. 8