Agilent G Rebuilt Electron Capture Detector

Similar documents
Agilent Inlet EPC Leak Test

6890 GC Site Preparation

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

Best Practice for Identifying Leaks in GC and GC/MS Systems

3 Flow and Pressure Control

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Retention Time Locking: Concepts and Applications. Application

Mass Spec will not Autotune

Customer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist

7820A GC Site Preparation Checklist

Agilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist

Agilent G3180B Two-Way Splitter Kit

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Agilent Dimension Software for ELSD User Manual

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Instant Connect Helium Saver Injector Module

Impact of Air Leaks on the Productivity of GC and GC/MS Systems

Agilent 7890 Series Gas Chromatograph

Recent Advances to Ensure Simple, Leak Free GC Column Connections. Ken Lynam R&D/Application Chemist Agilent Consumables and Supplies Division

Super-Clean Gas Filters from Restek

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Tips and Tricks of Injector Maintenance. Abby Folk Agilent GC Technical Support August 2009

Rapid and Reliable Detection of Dissolved Gases in Water

6850 Series Control Module User Information

For additional information about our solutions, please visit our web site at US/Pages/HomePage.

6890 GC Not Ready Messages (A16010)

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

no peaks missing peaks too small peaks increasing retention decreasing retention

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

GCMSD-Headspace Analysis SOP

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

CP-3800 GC Operator s Manual

Please refer to the Hydrogen Safety Guide which is shipped with the Instrument. 5977B Series GC/MS Systems Site Preparation Checklist

Multiple Gas#5 GC configuration Jan 2016

Agilent 1290 Infinity Pump Head Maintenance

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

CP-3800 GC Getting Started Manual

Zebron Gas Management

5890II GC Standard Operating Procedure 9/2/2005

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

Agilent 7697A Headspace Sampler

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Agilent 7697A Headspace Sampler

Agilent 7697A Headspace Sampler

The Open-Split Capillary Interface. Part No

A New GC/MS System Designed for Helium Carrier Gas Conservation

Agilent G3185B QuickSwap Accessory

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

Agilent Flowmeter ADM1000

Agilent 7820A Gas Chromatograph

Installation, Operation, and Maintenance Manual

Capillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions

be present throughout these services; otherwise, they will miss important operational, maintenance and safety information.

Installation, Operation, and Maintenance Manual

Agilent 6890N, Option 652 Oral Fluids Analyzer

Products Designed to Solve Nagging GC Problems

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Carrier Gases in Capillary GC

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

Agilent Auxiliary Gas Module AGM 2. User s Guide

490 Micro GC Site Preparation Checklist

Agilent 7667A. Mini Thermal Desorber. Site Preparation

Experiment GC : Analysis of BTEX by GC-FID

User manual. TC-20 Sample conditioning and dry-purge rig. QUI-0003 Version 6.3 September

System Overview. TCD Detector temperature setpoint Regulator

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Automated Determination of Dissolved Gases in Water Anne Jurek

Agilent 490 Micro GC Natural Gas Analyzer

Advances in Capillary Column Backflushing

Anti-flood device Model B-1

M/s. Thermo fisher Scientific

THE BP-690 SERIES. Operating and Service Manual. Series includes all variants of BP-LF/MF-690/691

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

ACV-10 Automatic Control Valve

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

28 Site Preparation. Supplying valve actuator air

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector

Agilent 5975T LTM GC/MSD

Helium-Oxygen Blender

TECHNICAL DATA ANTI-FLOOD DEVICE MODEL B-1 1. DESCRIPTION

Beverage Systems. Component Inspection and Service TSB-Y Date: Subject: Products Affected: Description: July Inspection of CO 2

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Transcription:

Agilent G1533-90790 Rebuilt Electron Capture Detector User Information Agilent Technologies

Notices Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number G1533-90790 Edition Third edition, March 2002 Printed in USA Agilent Technologies, Inc. 1601 California Street Palo Alto, CA 94304 USA Safety Notices CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

Contents 1 About Your Refurbished Detector 2 Typical Operation of an ECD Over Long Periods Figure 1. Typical operation of an ECD over extended periods of time 8 3 Checkout Conditions for Common Configurations Table 1. Checkout conditions for 5890 and 6890 instruments with S/S Inlets and ECDs 9 6890 Noise Test 10 Table 2. 6890 parameters for detector noise test 10 Figure 2. ECD with capillary split/splitless inlet noise test 10 Figure 3. µecd with capillary split/splitless inlet noise test 11 Table 3. 6890 detector noise, wander and drift test limits 11 6890 Performance 12 Table 4. 6890 chemical test parameters 12 ECD 13 Table 5. ECD with capillary split/splitless inlet 13 Figure 4. 6890 ECD with capillary split/splitless inlet checkout chromatogram 14 Table 6. Typical expected checkout results 15 µecd 16 Table 7. µecd with capillary split/splitless inlet 16 Figure 5. µecd with capillary split/splitless inlet checkout chromatogram 17 Table 8. Typical expected checkout results 18 5890 Noise Test 19 Table 9. 5890 parameters for detector noise test 19 Table 10. ECD noise specifications 19 Figure 6. ECD noise test 20 5890 Performance 21 Table 11. ECD with split-only or split/splitless capillary inlet 21 Figure 7. ECD with capillary inlet checkout chromatogram 22 Table 12. Typical expected checkout results 22 Rebuilt Electron Capture Detector 3

4 Troubleshooting Contamination Problems on ECD Insure clean gas supplies 23 Isolate problem to carrier or makeup gas supplies 23 Figure 1. Correct installation of ECD makeup adapter 24 Evaluate the carrier side 25 4 Rebuilt Electron Capture Detector

Agilent G1533-90790 Rebuilt Electron Capture Detector User Information 1 About Your Refurbished Detector The enclosed Electron Capture Detector (ECD) was refurbished and checked to meet a set of specifications that will guarantee that the detector will perform as well as a brand new ECD on a new instrument. The enclosed chromatogram shows the typical response and noise expected from any of the refurbished ECDs that pass through our process and are run under the typical check-out conditions as specified in the Performance Verification Procedures for this model of new chromatographic instrument, or an instrument that is clean, leak free and supplied with clean gases. Since the condition of the instrument to be used to mount this refurbished detector is unknown, we cannot predict its performance when installed. Several instrument conditions may prevent this detector from performing to the desired specifications such as contamination entering the detector from any source, i.e., the column, the injection port, the support gases, etc. If the enclosed detector does not meet your requirements when installed, please use proper troubleshooting procedures to try to determine and isolate the source of the problem. It is likely that some of the other hardware that plays a vital role in the proper operation of the analytical system may be the source of contamination to the ECD. If the performance of this rebuilt ECD is questionable, please perform the performance verification (PV) according to the instructions for your instrument and compare that response to the expected response and noise levels for your configuration of the instrument. If the performance does not meet the guidelines given in the PV procedures, please refer to the enclosed troubleshooting guidelines or the troubleshooting guidelines in the service section of your instrument manual as an aid in your troubleshooting procedures. You may also wish to refer extended troubleshooting of a difficult problem to your Agilent service representative. If a detector is to be left unused for a period of time, purge it with a small flow of inlet or purge gas (helium or nitrogen, etc.). This will prevent other contaminants (such as laboratory air) from entering the detector and degrading the ultimate performance. Agilent Technologies 5

1 About Your Refurbished Detector 6 Rebuilt Electron Capture Detector

Agilent G1533-90790 Rebuilt Electron Capture Detector User Information 2 Typical Operation of an ECD Over Long Periods Figure 1 illustrates how ECD performance can change with typical operation over extended periods of time. The Y axis represents the frequency or response of the detector and the X axis is a pseudo time axis. Please note that neither axis is meant to represent a linear scale but only the general concept of the response or time. Region A represents the operation of the detector when the instrument or detector is new and/or recently cleaned or refurbished. The peak would represent the chromatographic response to a standard sample under those conditions. The successive peaks across the time direction (in the B region) represent a typical response to the same standard sample at various times over the lifetime of the ECD. Note that the response tends to increase as the baseline increases with time. The baseline increase may result from a contamination build-up inside the detector over time with either sample components or other contamination fed to the detector via the support gases or column bleed. Note that the noise of the baseline increases over time as a direct result of the contamination build-up and baseline increase. However in some customers practice the increased noise may be unnoticeable because of the range and attenuation that the instrument is being operated at, as well as the fact that the noise increase is quite gradual and does not affect the quantitation from a calibrated detector response. Region C represents the approach to the end of the useful lifetime of the ECD due to the excessive response as well as the excessive noise of the baseline. In these cases the baseline noise begins to interfere with the peak identification (especially at low levels of detection for the desired components). The response from higher levels of sample may even saturate the frequency output of the detector resulting in flat-topped peaks. It is in the C region that the necessity of cleaning the detector becomes obvious and a user may remove the ECD from service and return it to the factory for refurbishment. When received, Agilent will clean the detector, replace any necessary components and test it for functionality to the same specifications as the original detector. We will return a detector to you in a similar condition as represented in the A region. Agilent Technologies 7

2 Typical Operation of an ECD Over Long Periods The relative difference between the peak height response of a detector operating in the C region and the A region can be as great as a factor of 8 10. The signal noise will also be increased by the same (or greater) amount, so the sensitivity (as defined by the signal to noise ratio) is not greater in the C region and the overall operation of the detector is improved with the cleaner detector. Saturation A B C Frequency (Scale not linear) Time (several months to several years) Figure 1 Typical operation of an ECD over extended periods of time 8 Rebuilt Electron Capture Detector

Agilent G1533-90790 Rebuilt Electron Capture Detector User Information 3 Checkout Conditions for Common Configurations This section lists the checkout conditions and typical expected performance for 5890 and 6890 GCs with split/splitless inlets. For other configurations please see the latest revision of the Standard Operating Procedures (SOP) documents or Performance Verification Procedures (PV) documents. See Table 1 for the appropriate document. Table 1 Checkout conditions for 5890 and 6890 instruments with S/S Inlets and ECDs SOP PV 5890 instruments 05890-90390 5965-6628 6890 instruments G1530-61210 5963-7184 Agilent Technologies 9

3 Checkout Conditions for Common Configurations 6890 Noise Test The tables below summarize the test conditions and typical expected results. For the detailed test procedure and more information, see the SOP or PV. Table 2 6890 parameters for detector noise test Detector Conditions ECD µecd 6890 Detector offset <100 * <400 6890 range 0 6 3396B/3396C/3397A ATTN^2 0 3 Zero 50 10 Chart speed 1 1 * 100 display units = 500 Hz at Ref. current of 1 na Figure 2 and Figure 3 show typical noise test results. Figure 2 ECD with capillary split/splitless inlet noise test 10 Rebuilt Electron Capture Detector

Checkout Conditions for Common Configurations 3 Figure 3 µecd with capillary split/splitless inlet noise test Table 3 6890 detector noise, wander and drift test limits Detector Range of Operation ECD µecd Eval time (min) 10 11 Noise (min) <9 <37 Wander (mm) <90 <90 Drift <90 <90 Rebuilt Electron Capture Detector 11

3 Checkout Conditions for Common Configurations 6890 Performance Table 4 summarizes the chemical test method parameters. Table 4 6890 chemical test parameters Temperatures, o C Flow/Pressure Rates Other Inlet Detector Detector ECD µecd Inlet Cap S/SL PP COC PTV Cap S/SL PP COC PTV Oven temp. 80 80 80 80 80 80 80 80 Initial temp. 80 80 80 80 80 80 80 80 Initial time, min 0 0 0 0 0 0 0 0 Rate, o C/min 15 15 15 15 15 15 15 15 Final temp. 180 180 180 180 180 180 180 180 Final time, min 10 10 10 10 10 10 10 10 Detect temp. 300 300 300 300 300 300 300 300 Inlet temp. 200 200 OT 80 * 200 200 OT 80 * Pressure (EPC) (psi) 25 25 25 25 25 25 25 25 Mode SL/CP CF CP SL/CP SL/CP CF CP SL/CP Purge time, min 0.75 N/A N/A 0.75 0.75 N/A N/A 0.75 Purge flow, ml/min 60 N/A N/A 60 60 N/A N/A 60 Ref gas N 2 N 2 N 2 N 2 N 2 N 2 N 2 N 2 Ref flow (ml/min) 6±1 6±1 6±1 6±1 Makeup gas N 2 N 2 N 2 N 2 N 2 N 2 N 2 N 2 C. mkup flow (ml/min) 60±3 60±3 60±3 60±3 30±3 30±3 30±3 30±3 Offset ** (display units) <100 <100 <100 <100 <400 <400 <400 <400 Range 0 0 0 0 6 6 6 6 Column 19091J-413 HP-5 30 m x 0.32 mm x 0.25 mm Sample ECD ECD ECD ECD ECD ECD ECD ECD * See the 6890 PV for PTV temperature program 7.5 to 7.9 ml/min (manual) at oven temperature of 40 C ECD anode purge flow ** ECD 0.2 x Hz ECD checkout sample - part no. 18713-60040 SL/CP - Splitless Constant/Pressure for EPC inlets CF - Constant Flow for EPC inlets OT - Oven Track 12 Rebuilt Electron Capture Detector

Checkout Conditions for Common Configurations 3 ECD Table 5 lists the checkout method and typical expected 6890 ECD performance. Table 5 ECD with capillary split/splitless inlet Oven Initial temp 80 C Initial time Rate 1 0 min 15 C/min Final temp 1 180 C Final time 1 Inlet 10 min Temperature 200 C Inlet pressure 25 psi (cons. press. for EPC inlets), 7.5 7.9 ml/min (manual) at an oven temp of 40 C Mode Purge time Purge flow Detector Splitless 0.75 min 60 ml/min Temperature 300 C Anode purge flow (N 2 ) Cons. mkup flow (N 2 ) Ref. current Offset Column and sample 6 ± 1 ml/min 60 ± 3 ml/min 1 na Should be < 100 display counts (< 500 Hz) Type HP-5 30 m x 0.32 mm x 0.25 mm -- part no. 19091J-413 Sample ECD checkout sample -- part no. 18713-60040 Composition Injection volume 1 µl 6890 signal parameters Range 0 33 pg/ml (0.33 ppm) of each lindane and aldrine in isooctane Rebuilt Electron Capture Detector 13

3 Checkout Conditions for Common Configurations Table 5 ECD with capillary split/splitless inlet (continued) 3396B or 3396C or 3397A integrator conditions with analog Set date and time ZERO 10 ATT 2^ 7 CHT SP 1 PK WD 0.04 THRSH 7 STOP TIME 11 Chromatogram specifications Sensitivity Lindane Area counts >757,000 Reproducibility RSD of Lindane retention times, for inj. 1 to 6, <1% * RSD of Lindane Area, for injections 1 to 6, <3% * MDL MDL (Lindane pg/s) = (3366 x Noise/(Area Lindane) with the area measured at range 2^) and the noise in mm at attn 2^0 (noise measure peak to peak) <=0.4 pg/s * 6890 Series Automatic Liquid Sampler only Figure 4 and Table 6 show typical 6890 ECD checkout chromatogram and results. 1.170 2.524 0.789 4.070 5.266 7.805 10.753 Figure 4 6890 ECD with capillary split/splitless inlet checkout chromatogram 14 Rebuilt Electron Capture Detector

Checkout Conditions for Common Configurations 3 Table 6 Typical expected checkout results RT Area Type Width Area% 0.789 1843432 BV 0.132 27.22585 1.080 196465 VV 0.064 2.90161 1.170 903974 VP 0.256 13.35089 2.524 146218 BP 0.097 2.15951 4.070 86832 VB 0.055 1.28243 5.190 113038 PV 0.032 1.66947 5.266 106328 VV 0.032 1.57037 7.805 1361074 PB 0.034 20.10185 10.753 2013529 PB 0.064 29.73802 Rebuilt Electron Capture Detector 15

3 Checkout Conditions for Common Configurations µecd Table 7 lists the checkout method and typical expected 6890 µecd performance. Table 7 µecd with capillary split/splitless inlet Oven Initial temp 80 C Initial time Rate 1 0 min 15 C/min Final temp 1 180 C Final time 1 Inlet 10 min Temperature 200 C Inlet pressure 25 psi (cons. press. for EPC inlets), 7.5 7.9 ml/min (manual) at an oven temp of 40 C Mode Purge time Purge flow Detector Splitless 0.75 min 60 ml/min Temperature 300 C Anode purge flow (N 2 ) Cons. mkup flow (N 2 ) Ref. current Offset Column and sample 6 ± 1 ml/min 30 ± 3 ml/min 1 na (fixed) Should be < 400 display counts Type HP-5 30 m x 0.32 mm x 0.25 mm -- part no. 19091J-413 Sample ECD checkout sample -- part no. 18713-60040 Composition Injection volume 1 µl 6890 Signal parameters Range 6 33 pg/ml (0.33 ppm) of each lindane and aldrine in isooctane 16 Rebuilt Electron Capture Detector

Checkout Conditions for Common Configurations 3 Table 7 µecd with capillary split/splitless inlet (continued) 3396B or 3396C or 3397A integrator conditions with analog Set date and time ZERO 10 ATT 2^ 5 CHT SP 1 AR REJ 100000 PK WD 0.04 THRSH 5 STOP TIME 11 Chromatogram specifications Sensitivity Lindane Area Counts > 52,000 X noise Reproducibility RSD of Lindane Retention Times, for inj. 1 to 6 * < 1% * MDL * 6890 Series Automatic Liquid Sampler only RSD of Lindane Area, for injections 1 to 6 < 3% * MDL (Lindane fg/sec) = (47,950 x Noise/(Area Lindane) (with the area measured at range 2^6), the noise in mm at attn 2^3 (noise measure peak to peak), and Range^0 should be <37 mm (1 min). < 8 fg/s Figure 5 and Table 8 show typical 6890 µecd checkout chromatogram and results. 7.279 9.779 Figure 5 µecd with capillary split/splitless inlet checkout chromatogram Rebuilt Electron Capture Detector 17

3 Checkout Conditions for Common Configurations Table 8 Typical expected checkout results RT Area Type Width Area% 0.653 556300 PV 0.103 16.93781 7.279 1272275 PB 0.028 38.73728 9.779 1455794 VB 0.055 44.32493 18 Rebuilt Electron Capture Detector

Checkout Conditions for Common Configurations 3 5890 Noise Test The tables below summarize the test conditions and typical expected results. For the detailed test procedure and more information, see the SOP or PV. Table 9 5890 parameters for detector noise test Setting 5890 Range Sig 1 or Sig 2 0 3396 ATT 2 Zero 50 Chart speed 1-2 for 2 min 0 for 10 min Table 10 Noise Wander Drift ECD noise specifications For any 2-minute interval over the entire plot, the noise in mm should be 0.415 x the observed signal offset For any 10-minute interval over the entire plot, the wanders should remain within the confines of the page For any 10-minute interval over the entire plot, the drift should remain within the confines of the page Rebuilt Electron Capture Detector 19

3 Checkout Conditions for Common Configurations Figure 6 shows a typical noise test result. Figure 6 ECD noise test 20 Rebuilt Electron Capture Detector

Checkout Conditions for Common Configurations 3 5890 Performance Table 11 ECD with split-only or split/splitless capillary inlet Temperatures Oven 170 C Oven equilib. time 3 min Initial temp 170 C Initial Time 8 min Rate 0 Final temp 0 Final time 0 Detector 300 C Inlet 200 C 5890 Signal Parameters Range 0 Flow Rates Carrier (Helium) Makeup (N 2 ) Split vent Septum purge 15 ± 1 ml/min 60 ± 2 ml/min 60 ± 2 ml/min 5 ± ml/min Sample Type ECD Sample Part number 18713-60040 Composition 33 pg/µl (0.33) ppm) of each of Lindane and Aldrine in Isooctane Injection volume 1 µl 3396 Integrator Conditions ZERO 10 Attn 5 Chart speed 1 Peak width 0.04 Threshold 5 Area reject 0 Rebuilt Electron Capture Detector 21

3 Checkout Conditions for Common Configurations Table 11 ECD with split-only or split/splitless capillary inlet (continued) Stop time 8 Chromatogram Specifications Sensitivity Lindane Area Counts >11,400 x Noise (mm) Reproducibility RSD of Aldrin Retention Times, for inj. 2 to 6 < 1% (7673 only) RSD of Aldrin Area, for injections 2 to 6 < 3% (7673 only) Figure 7 and Table 12 show typical 5890 ECD checkout chromatogram and results. Figure 7 ECD with capillary inlet checkout chromatogram Table 12 Typical expected checkout results RT Area Type Width Area% 0.167 2195754 SBB 0.033 84.70592 0.300 32279 TBB 0.027 1.24523 2.980 172491 BB 0.116 6.65422 6.933 191684 BB 0.261 7.39463 22 Rebuilt Electron Capture Detector

Agilent G1533-90790 Rebuilt Electron Capture Detector User Information 4 Troubleshooting Contamination Problems on ECD Persistent problems with high ECD background noise or Ghost Peaks during temperature program are almost always due to contamination introduced to the GC system from outside the GC, either from dirty samples, consumables, or from the carrier/make-up gas supplies. Insure clean gas supplies Before beginning the troubleshooting process, make sure the GC system is installed with supply gases of sufficient purity. The carrier and makeup gas purity should be >99.999%. See the site prep section in the 6890 user documentation. After you are sure of the purity of the carrier gas, insure that the tank regulators have STAINLESS STEEL diaphragms, equivalent to Agilent p/n 8507-0407 (CGA 580 in USA). Install new 1/8-inch copper supply tubing Agilent p/n 5180-4196. (Many times clean tubing from other vendors has caused high ECD background). At the same time install new traps in both the carrier and makeup supplies moisture trap p/n 5060-9084 closest to the tank, and indicating O 2 trap p/n 3150-0528 closest to the GC. Be very careful to leak test the entire plumbing manifold. Isolate problem to carrier or makeup gas supplies Determine what components of the apparent ECD contamination are from the carrier vs. makeup systems in the GC. Typically, sharp, well-resolved peaks that elute during a temperature program with no injection are from the carrier/inlet system. Broader humps in the baseline are usually from the makeup system. Overall high background can be contaminated gas from either the carrier or makeup gas supply or a contaminated ECD cell. To determine the source of contamination, perform the following procedure: 1 Remove the column from the ECD, and inspect the installation of the makeup gas adapter in the ECD cell. It is very common to install the adapter too low in the detector. To determine proper placement of the adapter measure from the bottom of the 1/4-inch swagelok nut to the bottom of the hex of the makeup gas adapter. The measurement should be 19-20 mm. If it is >22 mm, then the adapter was installed incorrectly. A ridge that is machined into the entrance of the ECD cell prevents the Agilent Technologies 23

4 Troubleshooting Contamination Problems on ECD adapter from easily seating all the way. Wiggling the adapter while installing will allow it to go all the way in. If the adapter is not all the way in, it can leak during a temperature program and can adversely affect linearity for pesticide methods. 1 mm 19 ± 1 mm Capillary column nut Capillary column Figure 1 Correct installation of ECD makeup adapter 2 Remove the makeup adapter from the ECD and then unscrew the tip and take out the Gigabore liner. Inspect the makeup adapter body for graphite deposits. Remove all graphite deposits and clean it thoroughly with methanol. Soak the removable tip in methanol. 3 Reassemble the adapter WITHOUT the Gigabore liner. Cap off the adapter using a 5-inch piece of narrow bore column, capped at one end with a new septum. The other end of the column should extend one mm past the tip of the adapter. When troubleshooting contamination problems, it is wise to use column ferrules of Vespel instead of graphite. Tighten the tip just past finger tight with a clean pair of pliers and clean the whole assembly with methanol before installing. 4 Reinstall the capped off makeup adapter, making sure it is fully inserted into the ECD cell. Use a new ¼ Vespel ferrule to allow for easy installation. Retighten after heating the ECD. 5 Reset the makeup flow to the original setpoint and bakeout the detector at 350 degrees for l hour. During this detector bakeout, put the inlet into split mode with 200-300 ml/min of split vent flow (gas saver off) and bakeout the injection port at 275 degrees and the column at its appropriate bakeout temperature. 24 Rebuilt Electron Capture Detector

Troubleshooting Contamination Problems on ECD 4 6 After the bakeout, run a series of blank runs using the customer method. If the baseline is acceptable free of humps or peaks, and stays under 1000 hz during the temperature program, then the detector and makeup side of the system is clean. 7 Any unacceptable baseline problems could indicate contaminated makeup gas, EPC module, makeup adapter or ECD cell. These must be addressed before proceeding. If the 6890 was manufactured prior to 6/97, then the EPC modules are suspect to having O-ring contamination. See service note G1530-14 and follow the procedure outlined the EPC module should also be replaced. The O-ring contamination problem was resolved after 7/97. Evaluate the carrier side Once the ECD and the makeup system are determined to be clean, then the carrier, inlet and column should be evaluated. 1 Reinstall the column in the makeup adapter by the same procedure used to install the piece of column used to cap the detector: a Remove the makeup adapter from the ECD and remove the short piece of column b Install new ferrule on the column and trim the column end c Install the column in the makeup adapter such that <1 mm of column extends past the tip of the adapter. It is best not to use the Gigabore liner. A better liner solution is the new mixing liner (p/n G2397-20540) from the µecd. It is backwards compatible to 5890 SER II and 6890 ECDs. Column installation is per 6890 µecd directions. d Wipe-off the entire makeup adapter with methanol and install the adapter fully inserted into the ECD. 2 Bakeout the entire system for another hour with the ECD at 350 degrees and the inlet in split mode at 275 degrees. The column should be baked at the manufacturer s recommended bakeout temperature. 3 After bakeout, reload the customer s method and make a series of blank runs (no injection) to determine if the contamination problem persists. Again a single, well resolved peak could be due to EPC module O-ring contamination mentioned previously. Address per service note. 4 If the contamination persists then perform a complete injection port maintenance, including a thorough cleaning of the shell weldment, replacing gold seal and liner. Also install a known good 30 meter/320 µm HP-5 checkout column to rule out column contamination. 5 Typically these peaks from the inlet side are due to contaminated carrier gas supply, EPC module, insert weldment, injection port/liner or column. 6 On the 6890 it is best to turn Auto-Prep Run to off so that the inlet is being purged when the GC is not in use. Rebuilt Electron Capture Detector 25

Troubleshooting Contamination Problems on ECD 4 Agilent Technologies, Inc. Printed in USA, March 2002 G1533-90790