The Preparation of Low Concentration Hydrogen Sulfide Standards Gulf Coast Conference September 12, 2002 Galveston, TX Paper 050,12:45 pm Moody F

Similar documents
Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

APPLICATION NOTE. GC Integrated Permeation Device

METHOD 3A - DETERMINATION OF OXYGEN AND CARBON DIOXIDE CONCENTRATIONS IN EMISSIONS FROM STATIONARY SOURCES (INSTRUMENTAL ANALYZER PROCEDURE)

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Multiple Gas#5 GC configuration Jan 2016

Gas Diluter Portable Gas Dilution System User s Manual

Generating Calibration Gas Standards

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Protocol Gas Verification For Compressed Gas Cylinders Containing Either SO 2, NO or CO. Quality Assurance Plan/Standard Operating Procedure

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin

Ozone. Sensoric O3 3E 1

CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO Walnut Lake Rd th Street Houston TX Garner, IA 50438

H 2. Sensor Series Take no risk! Install a Hydrogen Sensor in your Gaschromatograph

Automated Determination of Dissolved Gases in Water Anne Jurek

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

During today's webinar we're going to look at the following topics:

AND EQUIPMENT FOR ENVIRONMENTAL APPLICATIONS (800) B Polk Street Houston, TX

Experiment GC : Analysis of BTEX by GC-FID

System Overview. TCD Detector temperature setpoint Regulator

Zebron Gas Management

Carrier Gases in Capillary GC

Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

HiQ laboratory gas generators.

Svensk läkemedelsstandard

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

6200E Installation Requirements

Nonrefillable Cylinders

Application Note AN-107

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Retention Time Locking: Concepts and Applications. Application

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Title: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator

Carrier Gases in Capillary GC

LUMAZOTE T ECHNICAL BRO CHU RE

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

Precision Calibration Gas

Gas Mixture Two Components. Gas Mixtures Two Components

Thermo Scientific Model 146i Principle of Operation

Compatible with a Wide Range of Gases or Liquids for Laboratory or Plant Applications

METHOD 204F--VOLATILE ORGANIC COMPOUNDS CONTENT IN LIQUID INPUT STREAM (DISTILLATION APPROACH) 1.1 Applicability. This procedure is applicable for

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

DSC 204 HP Phoenix. High-Pressure Differential Scanning Calorimetry up to 15 MPa Method, Technique and Applications

ANALYSIS OF GASES PRODUCED BY THREE UNDERWATER CUTTING DEVICES

CHEM 355 EXPERIMENT 7. Viscosity of gases: Estimation of molecular diameter

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

REFINERY GAS ANALYSIS BY GAS CHROMATOGRAPHY

Specialty Gases & Equipment ISO 9001: 2008 CERTIFIED. A Division of Norco, Inc.

SOUTH COAST AIR QUALITY MANAGEMENT DISTRICT METHOD INSTRUMENTAL ANALYZER PROCEDURES FOR CONTINUOUS GASEOUS EMISSION SAMPLING

MMS Mercury Monitoring System for Natural Gas

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

3. Coupled Techniques in Thermal Analysis

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

Svensk läkemedelsstandard

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

Gravimetric preparation of NO 2 primary reference gas mixtures at the NMISA. Tshepiso Mphamo 03 September 2012

490 Micro GC Site Preparation Checklist

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Gas-Detection Instruments

Elimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU.

May 1, By John Heim,Doug Staples

Adsorption Unit Operating Procedures: CO2 Adsorption on to Activated Carbon

GAS DETECTION: CALIBRATION GAS

Phosgene Sensor. Sensoric COCl2 3E 1

Gases and welding distributors continue

The analysis of complex multicomponent

Determination of Zn using Atomic Absorption with Multiple Standard Additions

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Enbridge Pipelines (NE Texas) LP EXAMINER S REPORT AND RECOMMENDATION STATEMENT OF THE CASE

SRI Multi Gas Analyzer 2016


Basic Concepts of Chemistry Notes for Students [Chapter 10, page 1] D J Weinkauff - Nerinx Hall High School. Chapter 10 Gases

PERFORMANCE SPECIFICATION PROPELLANT, HYDROGEN

PRECISION GAS MIXTURES

Svensk läkemedelsstandard

Name Chemistry Pre-AP

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

Kinetic Molecular Theory

acrolein, acetaldehyde and acetone( cm -1 ); methanol (1306 cm -1 ); ethylene (949 cm -1 ); and isoprene (893 cm -1 ).

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

490 Micro GC Site Preparation Checklist

Agilent 490 Micro GC Natural Gas Analyzer

Vehicle and Engine Emissions Testing

Central scientific research automobile and automotive engine institute

STATE OF CALIFORNIA AIR RESOURCES BOARD AIR MONITORING QUALITY ASSURANCE VOLUME V AUDIT PROCEDURES MANUAL FOR AIR QUALITY MONITORING APPENDIX E

Determination of Zn using Atomic Absorption with Multiple Standard Additions

Calibration Gases. Intrial provides mix of special calibration gases for analyzers and detectors.

Calibration Gases. Intrial provides mix of special calibration gases for analyzers and detectors.

Transcription:

The Preparation of Low Concentration Hydrogen Sulfide Standards Gulf Coast Conference September 12, 2002 Galveston, TX Paper 050,12:45 pm Moody F Robert Benesch, Malik Haouchine, and Tracey Jacksier, Ph.D. Air Liquide, Chicago Research Center 5230 S. East Ave, Countryside, IL 60525 RECHERCHE ET DEVELOPPEMENT

Previous Focus Low level H 2 S gas mixtures (<10 ppm) can exhibit reactivity to materials Reactivity may be observed in flowing systems Long contact times Low flow rates in tubing Small contact amounts Injection of sample into carrier gas stream Non-flowing systems Long contact times between gas and material Storage ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /2

Introduction 6 feet ¼ inch tubing Linear velocity 2 cm/sec Teflon - Du Pont ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /3

Previous Focus Storage of 100 ppb H 2 S in 500 ml sample cylinders 30 day stability test Satisfactory - Sulfinert, Silcosteel - No experimentally noticeable loss in concentration Unsatisfactory -Aluminum, Stainless Steel, Carbon Steel Immediate loss of total concentration Restek Corp. ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /4

Present Focus Introducing Alphatech TM New patent pending, commercial cylinder treatment for specialty gas applications Provides Storage of low level sulfur compounds ( 100 ppb) in a commercial compressed gas cylinder Stability of H 2 S for at least 9 months ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /5

Experimental To determine the effectiveness of Alphatech TM Cylinders prepared, filled, and analyzed Cylinders filled by pressure from source cylinders 9.9 ppm H 2 S 99.999 % N 2 Analysis Utilized 2 different instruments (1) Varian 3800 GC (Initial testing instrument) Sievers 355 SCD (Sulfur Chemiluminescence) Restek 1/16 RT-Sulfur packed column ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /6

Experimental Analysis (Con t) (2) Agilent 6890 GC Antek 7090 SCD Restek SPB-1 capillary column Calibration of instruments Calibration curves generated within range of testing Dilution system used to generate four point calibration curve Concentration determination Averaging four points Experimental error (~10 %) ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /7

Untreated Cylinder Typical compressed gas cylinder Varied behavior New aluminum cylinder Vacuum baking Untreated Aluminum Cylinder 100 ppb H 2 S 150 Conc (ppb) 100 50 0 0 10 20 30 40 50 60 70 80 Time elapsed (days) ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /8

Alphatech Stability Treated with Alphatech TM process Three cylinders 2 filled simultaneously ALPHATECH TM Stability Test H 2 S in Nitrogen 150 Conc (ppb) 100 50 0 0 50 100 150 200 250 300 350 400 450 Time elapsed (days) ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /9

Alphatech TM Stability Typical Results Initial (calculated) Final Avg RSD Cylinder 1 79.4 76.4 73.1 5.8% Cylinder 2 104 111.8 96.6 5.7% Cylinder 3 104 99.6 99.7 6.4% All cylinders exhibit greater than 9 month stability ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /10

Adsorption Adsorption of H 2 S onto cylinder walls Is it occurring in Alphatech TM cylinders? Difference between fill concentration and first test concentration Difference observed is within experimental error Fill 106.2 106.2 106.2 106.2 Test 111.4 105.7 108.8 111.8 % Diff 4.90 0.47 2.45 5.27 Consistency of concentration at various pressures Increase in concentration at lower pressures if desorption is occurring ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /11

Consistency of Concentration vs. Pressure Cylinder depletion Cylinder filled 97 ppb H 2 S in N 2 at production facility using Alphatech TM preparation 3 % difference between 2150 and 63 psia Within experimental error H 2 S in Nitrogen 150 Conc (ppb) 100 50 0 0 500 1000 1500 2000 2500 Cylinder Pressure Depleted (psia) ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /12

Operation development Currently undergoing a production test with our specialty gas fill plant Alphatech TM cylinders prepared at specialty gas manufacturing facility Cylinders filled and analyzed at facility Different size cylinders (16 & 30) Different concentrations (nominal 50 and 100 ppb) ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /13

Other Testing Application of Alphatech TM to other analytes and balance gases 150 Alphatech TM Stability Test 100 ppb COS in Nitrogen Conc (ppb) 100 50 0 0 10 20 30 40 50 60 70 80 90 Elapsed Time (days) ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /14

Other Testing Alphatech TM Stability Test 50 ppb H 2 S in Gaseous Ethylene Conc (ppb) 80 60 40 20 0 0 50 100 150 Elapsed Time (days) ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /15

Conclusions Developed a cylinder treatment / process suitable for storage of low level H 2 S compounds Provide stability for at least 9 months Currently in operational development phase for production scale up Reliable - Easily repeatable Alphatech TM application to other sulfur analytes and balance gases underway ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /16

Acknowledgements Production & operations support: Theo Sadler Josh Mermelstein John Kuhn Ben Duong ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /17

Gulf Coast Conference Booth #130-132 Product Specialists Technical Experts ELECTRONICS Chicago Research Center / R. Benesch, M. Haouchine, T. Jacksier /Sept 02 /18