Discovery TGA 55, TGA 550, TGA Site Preparation Guide

Similar documents
Discovery SDT 650. Site Preparation Guide. Revision G Issued December Page 1

Discovery HP-TGA 75/750. Site Preparation Guide

TA Instruments Installation Requirements for the Thermal Discovery Series

TA Instruments Installation Requirements for Q Series Thermal Analysis Systems

TA Instruments Installation Requirements for Thermal Analysis Systems

TA Instruments Installation Requirements for Thermal Analysis Systems

Customer Responsibilities

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist

Customer Responsibilities

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist

UltraLo 1800 Alpha Particle Counter

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

DD2 Site Preparation Checklist Hardware Site Preparation Specification

490 Micro GC Site Preparation Checklist

Customer Responsibilities

Customer Responsibilities. Important Customer Information. Agilent Infinity II Preparative HPLC System Site Preparation Checklist

Customer Responsibilities

Customer Responsibilities

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

Agilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist

TA Instruments Installation Requirements for Ther

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

490 Micro GC Site Preparation Checklist

HumiSys HF High Flow RH Generator

Features. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing)

For additional information about our solutions, please visit our web site at US/Pages/HomePage.

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual

Agilent 5100 and 5110 ICP-OES. Site Preparation Guide

Recirculating Coolers

Preinstallation Checklist and Instructions for the TriStar These Preinstallation Checklist and Instructions were reviewed and approved by:

Benefits. Rev.III_032018_Oxybaby 6.0_engl subject to technical alterations

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

6890 GC Site Preparation

Vacuum. Dry Turbo Pumping Stations. Roughing Pumps for 1 K Helium Pots and Lambda Refrigerators. Hermetically Sealed Pumps for He3 Recirculation

BGA244 Binary Gas Analyzer

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

Flow and Energy Measurement F-5400 THERMAL MASS FLOW METER

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

analytical bulk flow distance FLOW level pressure temperature industrial communication

GAS ANALYSER OXYBABY 6.0 for O 2. or O 2 /CO 2. premium version

Varian 240-MS GC/MS Ion Trap Mass Spectrometer Pre-installation Instructions

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Procedures for operation of the TA Instruments DSC

STATISTICS AT IIHF EVENTS Hardware / Infrastructure / Staff Requirements 2017 Gemini Light Configuration

DSC Differential Scanning Calorimeter

Gas Switching Accessory

A Pulmonary Function Product of Rare Distinction

MEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #:

Operation of the Perkin Elmer TGA-GC/MS

Series Environmental Chambers

LUMAZOTE T ECHNICAL BRO CHU RE

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

OMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.

Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline

Agilent Auxiliary Gas Module AGM 2. User s Guide

Oxford Instruments Plasma Technology

Best Practice Guide, Servomex 2700

In-Water Mass Spectrometry for Characterization of Light Hydrocarbon Seeps and Leaks

Varian 325-MS LC/MS Quadrupole Mass Spectrometer Pre-installation Instructions

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

GAS ANALYSER OXYBABY M+ for O 2. or O 2 /CO 2. standard version

General Specifications

Pre-installation Checklist

FOURIER TRANSFORM INFRARED SPECTROSCOPY

Bubble Point / Pressure Decay Test / Gas Transport Rates through Membranes

GAS ANALYSER OXYBABY 6.0 for O 2. or O 2 /CO 2

The University of Hong Kong Department of Physics Experimental Physics Laboratory

Information Sheet. About this information sheet. Floor and wall loadings. Attachment of process module to floor. PlasmaPro Estrelas100

Compatible with a Wide Range of Gases or Liquids for Laboratory or Plant Applications

RM-80 respiration monitor

OXYBABY - MOBILE HAND HELD GAS ANALYSER

THE PMI AUTOMATED GAS/HELIUM PYCNOMETER PYC-G100A

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

SUPERIOR RESULTS WITH UNIQUE FEATURES

GFM. Typical Aluminum GFM Mass Flow Meter NIST MASS FLOW METERS. Principles of Operation. Design Features

Table of Content IMPORTANT NOTE: Before using this guide, please make sure you have already set up your settings in

Micro GC Fusion Pre-Installation Checlist

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Portable Gas Analyzers

Oxidation Stability of Gasoline and Aviation Fuels

Purpose and Scope: Responsibilities: Warnings and precautions: Background Beers law

Section 1: Overview... 6 Features... 6 Supported Software Part 1: Getting Started 7

Schedule of Requirements THERMODYNAMICS LABORATORY- CHEMICAL ENGINEERING DEPARTMENT

FPG8601 Force Balanced Piston Gauge

PHOENIX L300i Best-in-class Helium Leak Detectors

Plasma. Argon Production Helium Liquefaction Truck Loading Pure Gas Bottling

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Design Features. General Description

PURCHASE AND DELIVERY OF A HELIUM CLOSED-CYCLE OPTICAL CRYOSTAT. The present tender is only an unofficial translation provided as a courtesy.

Mass Flowmeters. Mass Flowmeters for Gases

Complete Range of Force Tensiometers

Customer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist

2 Switching off ECRIS 3. 4 Acquire Beam Spectrum 4. 6 Vent Instrument Chamber 5. 7 Pump down Chamber 5. 8 Baking out Chamber 5.

User and Installation Manual

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A

Certificate of Compliance

Transcription:

Discovery TGA 55, TGA 550, TGA 5500 Site Preparation Guide Revision E Issued March 2018

Table of Contents Table of Contents... 2 Ideal Setup... 3 System Components... 4 With Discovery Mass Spectrometer... 4 With Blending Gas Delivery Module... 5 Instrument Measurements... 6 Utility Requirements... 7 Computer Requirements... 8 Hardware Requirements... 8 Software Requirements... 9 Accessories (Discovery Mass Spectrometer)... 10 Accessories (Blending GDM)... 11 Site Preparation Checklist... 12 TA Instrument Offices... 13 Revision E Issued March 2018 Page 2

Ideal Setup IDEAL PLACEMENT AND BENCH MEASUREMENTS Select a location with adequate floor space and a rigid laboratory bench that is level and is in a vibration-free environment. For optimal performance, it is recommended that the instrument be placed by itself on a separate marble table. Bench length: 61 183 cm (24 72 in) Bench depth: 76 cm (30 in) Table length: 60 cm (24 in) Table depth: 76 cm (30 in) Distance from the wall: 30.5 cm (12 in) min. Revision E Issued March 2018 Page 3

System Components MAIN SYSTEM COMPONENTS WITH MASS SPECTROMETER B C A D A. Computer (Controller) B. Discovery Mass Spectrometer (Optional) C. Instrument D. Heat Exchanger Revision E Issued March 2018 Page 4

System Components MAIN SYSTEM COMPONENTS WITH BLENDING GDM A C B D A. Computer (Controller) B. Blending Gas Delivery Module (Optional) C. Instrument D. Heat Exchanger Revision E Issued March 2018 Page 5

Instrument Measurements MAIN INSTRUMENT Height: 61 cm (24 in) Width: 56 cm (22 in) Depth: 56 cm (22 in) Weight: 34 kg (76 lbs) Revision E Issued March 2018 Page 6

Utility Requirements POWER 100 240 VAC, 47 63 Hz, 1200 W Safety ground per local regulation GAS Purge Gases Acceptable purge gases: air, nitrogen, oxygen, argon, and helium Source gas pressure is a maximum 20 psig for all inlets at the back of the instrument Pressure regulator required Conditions: Must be dry Must be free from oil, dirt, and water Use dry nitrogen as the base purge gas when using a cooler Wire-wound furnace Flow Rates Purge gas flow rate: up to 500 ml/min. Recommended flow rates: Furnace Type Flow Rate for Sample Flow Rate for Balance Wire-wound furnace 60 ml/min 40 ml/min EGA furnace 90 ml/min 10 ml/min IR furnace 25 ml/min 10 ml/min EGA furnace IR furnace Revision E Issued March 2018 Page 7

Computer Requirements HARDWARE REQUIREMENTS Description Processor Memory Hard drive DVD Screen resolution Graphic memory Screen (LCD) size Requirement Intel Core 2 Duo or better 2.93 GHz with 3 MB L2 cache Minimum: 4 GB RAM (if the computer is only acting as the controller) Recommended: 8 GB RAM or greater (for best analysis performance) 80 GB free space 700 MB required for Full version of TRIOS 400 MB required for Lite version of TRIOS (without Online help) 48x CD-ROM or DVD Required: 1280 x 1024 with 24-bit colors Recommended: 1920 x 1080 with 24-bit colors 128 MB Required: 19 or greater Recommended: 24 wide screen Revision E Issued March 2018 Page 8

Computer Requirements SOFTWARE REQUIREMENTS Item Operating System Browser Other Other Network Conflicts Requirement Windows 7, 8, 10 Ultimate, Enterprise & Professional Home version not supported Minimum: 32-bit version (if the computer is only acting as the controller) Recommended: 64-bit version (for best analysis performance) Internet Explorer Microsoft Operating System Service Pack Must be up-to-date TA Instruments is not responsible for resolving issues associated with connections to your corporate network. TA Instruments is not responsible for resolving hardware/software conflicts created by the addition of third party hardware or software to the computer. Revision E Issued March 2018 Page 9

Accessories DISCOVERY MASS SPECTROMETER MEASUREMENTS Height: 41 cm (16 in) Width: 26 cm (10 in) Depth: 61.5cm (24 in) Weight: 28 kg (62 lbs) DISCOVERY MASS SPECTROMETER REQUIREMENTS Requirements Power adapter: Pump inlet: Backing pump: Turbo pump: Heaters: Mains inlet: All fuses: High Pressure Mass Spectrometer only: 240 VAC/2 A; 100 VAC/5 A; 50/60 Hz 24 VDC, 2.5 A 24 VDC, 3.15 AT 5.0 AT 8.0 AT 6.3 AT 250 VAC 5x20 mm Ceramic Fast Acting if applicable Acceptable purge gases: nitrogen or argon Must be dry and free of oil, dirt, and water Purge gas pressure: 7 14 kpa gauge (1 2 psig) Circulator Power Cooling Gas LN 2 Fluid Light Hardware Software Temp Lab Customer Revision E Issued March 2018 Page 10

Accessories DISCOVERY BLENDING GDM MEASUREMENTS Height: 13 cm (5 in) Width: 15 cm (6 in) Depth: 15 cm (6 in) Weight: 0.5 kg (1 lb) DISCOVERY BLENDING GDM REQUIREMENTS Requirements Power adapter: 100 240 VAC, 0.5A, 50/60 Hz Approved for operation on a 20 A branch circuit with protective conductor (ground). DC requirements: 24V, 0.8A. Use only TA Instruments-provided power adaptor. Must be dry and free of oil, dirt, and water Maximum inlet pressure: 140 kpa gauge (20 psig) Acceptable gases: Nitrogen, argon, helium, air, oxygen, carbon monoxide, carbon dioxide, forming gas. Circulator Power Cooling Gas LN 2 Fluid Light Hardware Software Temp Lab Customer Revision E Issued March 2018 Page 11

Site Preparation Checklist Discovery TGA 55, TGA 550, TGA 5500 Sufficient bench space for instrument, computer, and Mass Spectrometer (if needed) Length: 61 183 cm (24 72 in) Depth: 76 cm (30 in) Instrument power is 100 240 VAC, 47 63 Hz, 1200 W Purge gas: Is one of the following: air, nitrogen, oxygen, argon, or helium Is dry and free of oil, dirt, and water Pressure regulator is present Maximum 20 psig inlet pressure Gas source regulated pressure: up to 500 ml/min High Pressure Mass Spectrometer Purge Gas (if applicable): Is nitrogen or argon Pressure is 7 14 kpa gauge (1 2 psig) Is dry and free of oil, dirt, and water I hereby acknowledge that all utility requirements have been met per the checklist above and that they will be ready at the agreed time of installation. If all utility requirements are not met at the agreed time of installation, additional charges may be incurred for a return Service trip. Customer DD MM YYYY Company City State Country Please send a signed copy of the completed checklist to your local Service representative. Revision E Issued March 2018 Page 12

TA Instruments Offices For information on our latest products, contact information, and more, see our website at: http://www.tainstruments.com. For TGA training videos, visit http://www.tainstruments.com/videos/ To find your local TA Instruments office and contact information, visit http://www.tainstruments.com/contact/ta-directory/ TA Instruments Waters LLC Corporate Headquarters 159 Lukens Drive New Castle, DE 19720 USA Telephone: 302-427-4000 Fax: 302-427-4001 Email: info@tainstruments.com Revision E Issued March 2018 Page 13