Is deproteinization necessary in the determination of human plasmatic steroids by GC/IT-MS/MS analysis?

Similar documents
Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

WADA Technical Document TD2014EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

May 1, By John Heim,Doug Staples

Longterm Detection of Anabolic Steroid Metabolites in Urine

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

LCMSMS Solutions For Steroid Analysis

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

Now it is. easy to switch. CHS TM Steroid Profiling kit and software. Brochure not for distribution in the USA and Canada

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

WADA Technical Document TD2016EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites

Extended Application Note

National Institute for Public Health and the Environment Annual CRL workshop 22 October Update on natural Hormone studies

Complete. Male Hormones SAMPLE PATIENT. Patient: Order Number: Age: 47 Sex: M MRN:

Gas Chromatography-Mass Spectrometry Analysis Of Organics In Drinking Water: Concentrates And Advanced Waste Treatment Concentrates, Vol.

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

SIMPLE, EFFICIENT AND SIMULTANEOUS DETERMINATION OF ANABOLIC STEROIDS IN HUMAN URINE

Sample purification procedure for confirmation of 3 -hydroxy stanozolol by gas chromatography/high resolution mass spectrometry

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Methylprednisolone detection in urine following local and oral administrations

Physician Copy. Patient: SAMPLE PATIENT DOB: Sex: MRN: ACH24.1

WADA Technical Document TD2018EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

1. INTRODUCTION CH 3 CH3

EFFECTS OF CLEAROGEN ACNE LOTION ON TESTOSTERONE METABOLISM IN RECONSTRUCTED HUMAN EPIDERMIS

Analysis of Casein and Whey Protein in Whole, 2%, and Skim Milk by Capillary Gel Electrophoresis

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

3. Coupled Techniques in Thermal Analysis

SUPPORTING INFORMATION

Chemical Analysis for Methyltestosterone

WADA Technical Document TD2017NA

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

SPECIFIC DETECTION OF ANDROST-4-ENE-3,17-DIONE

EFFECTS OF REVIVOGEN SCALP THERAPY ON TESTOSTERONE METABOLISM IN RECONSTRUCTED HUMAN EPIDERMIS

Gas chromatographic and mass spectrometric analysis of conjugated steroids in urine

Androgen Measurements

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform

MV5: Automated concentration system

biosensis Human IGF-II, Insulin-like growth factor II, Somatomedin-A ELISA Kit Protocol

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

biosensis Human Lipocalin-2/NGAL ELISA Kit Protocol

WADA Technical Document TD2010EAAS. Endogenous Anabolic Androgenic Steroids Testing, Reporting and Interpretive Guidance

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

WADA Technical Document TD2018MRPL

Introduction. Saurabh Dubey 1, * and Alka Beotra 2

The instrument should have a mass range from 2 to 2000 m/z or better Interface

Validation of the MRM pesticides from the Working Document SANCO/12745/2013 of November 2017 rev. 9(1)

Analysis of prohormones in animal feed by GC-MS

FAST EVAPORATION AT NORMAL PRESSURE

Renzo BAGNATI Mario Negri Institute of Pharmacological Research- IRCCS Milano - Italy

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS

biosensis Mouse Interleukin-1 beta (IL-1β) ELISA Kit Protocol

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

biosensis Human Soluble Tumor Necrosis Factor receptor type II (stnfrii) ELISA Kit Protocol

CORESTA RECOMMENDED METHOD Nº 67

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

biosensis Rat Interleukin-1 beta, IL-1β ELISA Kit Protocol

biosensis Mouse Vascular endothelial growth factor A/VEGF-A/VEGF-164/VEGF-1/VEGF- 120/VEGF-2 ELISA Kit Protocol

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

Introduction. Experimental

Retention Time Locking: Concepts and Applications. Application

We will begin momentarily at 2pm ET. Slides available now! Recordings available as an exclusive ACS member benefit.

Utilization of Whey as One Dairy Industrial Waste in the Production of Alcohol

CHEM254 #4 The Synthesis of a Tertiary Alcohol Using a Pre-Made Grignard Reagent 1

DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178

Sensitive Analysis of Genotoxins by HPLC-ECD Using Boron-Doped Diamond Electrochemical Cell

Rapid and Reliable Detection of Dissolved Gases in Water

Steroids. Analytical Reference Standards /08

biosensis Rat Fibronectin ELISA Kit Protocol

Carrier Gases in Capillary GC

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Hair analyses. Hair analysis beyond workplace testing. Ia: Fatal steroid abuse. Identification of stanozolol in hair

biosensis Rat Glial cell line-derived neurotrophic factor/gdnf total /ATF ELISA Kit Protocol

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

biosensis Rat IGF-1/Somatomedin/Insulin-like growth factor ELISA Kit Protocol

N Korea Japan China Thailand Taiwan Philipines EP-5 EP km. Source: Google Maps and

Determination of Capsaicin Content and Pungency Level of Five Different Peppers Grown in Nigeria.

biosensis Mouse CXCL10/IP-10 ELISA Kit Protocol

CHAPTER 3 Development and Validation of a RP-HPLC method for Dutasteride and its impurities in bulk drug

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

Experiment GC : Analysis of BTEX by GC-FID

Setting a New Standard in Flash Chromatography Performance

Determination of capsaicinoid profile of some peppers sold in Nigerian markets

biosensis Human TNFα/Cachectin/TNFSF2 ELISA Kit Protocol

Pros and Cons of Hormone Testing in Different Body Fluids with Different Routes of Hormone Delivery

Studies on the determination of chlorotestosterone and its metabolites in bovine urine

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

SMART 4 Sports Panel. Drug Free Sport 1 7/13/07

More Speed and Resolution for Current Methods

IJPAR Vol.6 Issue 1 Jan - Mar Journal Home page:

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Monitoring Hormone Therapy Mark Newman, M.S. President of Precision Analytical Inc.

SCREENING OF TRENBOLONE-17β IN MILK SAMPLES AFTER APPLICATION OF TRENBOLONE ACETATE TO A CULL COW

TECHNICAL BULLETIN. TriControls. Control Materials for Liquid QC and Calibration Verification

Background Statement for SEMI Draft document 4657B NEW STANDARD: SPECIFICATIONS FOR TUNGSTEN HEXAFLUORIDE (WF 6 )

Transcription:

JOURNAL OF APPLIED BIOANALYSIS, Jan. 2015, p. 10-18. http://dx.doi.org/10.17145/jab.15.003. (ISSN 2405-710X) Vol. 1, No. 1 RESEARCH ARTICLE Is deproteinization necessary in the determination of human plasmatic steroids by GC/IT-MS/MS analysis? Antonio F. Toribio-Delgado 1, Maria C. Robles-Gil 2, Marcos Maynar-Mariño 2, Guillermo Olcina-Camacho 2, Juan.I Maynar-Mariño 1 1 Department of Analytical Chemistry. University of Extremadura, Badajoz, Spain. 2 Department of Physiology. University of Extremadura, Cáceres, Spain. (Received: 12 October 2014, Revised 8 November 2014, Accepted 13 November 2014). Human plasma is composed on average of 7% of proteins, where the main one is albumin, with concentrations range of 3.2-4.8 g/dl. Many studies consider plasmatic proteins to be a problem in relation to plasma analysis, so they must be eliminated. In this work, we have performed several deproteinization tests with acetonitrile, methanol, ethanol, HCl 0.5 M/methanol (1:1) (v/v), trichloroacetic acid (10%), sulfosalicylic acid (20%) and perchloric acid, proteases and activated carbon before analyzing twelfe plasmatic steroids hormones using gas chromatography directly coupled ion-trap mass spectrometry. Signal/noise ratios of tests were determine to assess whether a previous deproteinization treatment is necessary or not prior to their analysis. The use of chemical deproteinization methods, proteases and active carbon did not improve the signal/noise ratio. Deproteinization is not required for the analysis of steroid hormones in plasma using gas chromatography-directly coupled ion-trap mass spectrometry. Keywords: deproteinization, protease, activated carbon, human plasma, steroids. Introduction For many years concentrations of functional steroid hormones or their precursors have been determined mainly in plasma or serum samples of patients, occasionally in saliva or tissues, or as their metabolites, in urine or amniotic fluid [1]. Gas Chromatography/Mass Spectrometry (GC/MS) is well-suited for the identification of a large number of potential steroids and metabolites due to its high chromatographic resolution capacity and reproducible ionization efficiency [2]. Albumin is dominant in the binding, transport, and delivery of a range of endogenous ligands and pharmaceuticals because it accounts for approximately 60% of the total proteins in blood serum with a typical concentration of 3.4 5.0 g/dl [3]. Biblio graphy shows that protein precipitation for human plasma analysis performed by GC/MS can be *Correspondence: Department of Analytical Chemistry. Faculty of Sciences. University of Extremadura. Avd de Elvas s/n. CP 06071, Badajoz, Spain. Phone. +34 660 990 149, E-mail: afertd@gmail.com performed successfully with methanol by centrifugation [4, 5, 6] with acetonitrile [7, 8, 9] with 70% perchloric acid and n-hexane [10], with 5% sulfosalicylic [11], with ethanol [12], with acetone [13], and with trichloroacetic acid [14]. Other methods for plasma deproteinization focus on physical methods as described [15]. These methods include ultrafiltration and centrifugation with varied molecular weight cut-offs, or using other simple methods such as heating the plasma. However, it has not been shown that these deproteinization methods interfere with gaschromatography-ion trap-tandem mass spectrometry (GC/IT-MS/MS) analysis of plasmatic steroids. Hence, the aim of this work was to identify if it is necessary and useful to apply a deproteinization treatment to human plasma before its analysis in order to determine steroids hormones using GC/IT-MS/MS. Materials and methods Following human plasmatic steroids: testosterone, methyltestosterone, androsterone, 5β-androsterone, dihydrotestosterone (DHT), cortisone, cortisol, tetrahydro- 10

cortisone (THE), nandrolone, dehydroepiandrosterone (DHEA), tetrahydrocortisol (THF), estriol, and epiandrosterone were analyzed using GC/IT-MS/MS under different conditions: receiving and not receiving a previous treatment with commonly used deproteinization methods. The signal/noise (S/N) ratios of the analytes studied were assessed to determine which are the best pre-treatments for the analysis of human plasmatic steroid profile. Chemicals and steroids Testosterone, methyltestosterone, androsterone, 5β-androsterone, dihydrotestosterone (DHT), cortisone, cortisol, tetrahydrocortisone (THE), nandrolone, dehydroepiandrosterone (DHEA), estriol, epiandrosterone, N-methyl N-trimethylsilyl-trifluoroacetamide (MSTFA), proteinase N from Bacillus subtilis and protease from saitoi, were purchased from Sigma-Aldrich (Barcelona, Spain). Tetrahydrocortisol (THF) was supplied by Steraloids. β-glucoronidase (type Escherichia coli K 12) was obtained from Boeheringer Mannheim (Mannheim, Germany). Dithioerythritol was obtained from Serva (Heidelberg, Germany). Ammonium iodide (NH 4 I) was supplied by Panreac. Human albumin at 20% was obtained from Grifols. Active carbon Norit A was obtained from Serva (Heidelberg, Germany). All other reagents and solvents were of analytical grade and were mainly supplied by Scharlau (Hamburg, Germany), JT Baker (Tarragona, Spain) and Panreac (Barcelona, Spain). Instrumentation The GC-IT-MS/MS analysis of the samples was conducted on a Varian 3800 gas chromatograph directly coupled to a MS/MS Saturn 2000 ion-trap mass spectrometer and equipped with an auto-sampler Varian 8200 with a capacity for 48 samples. Separation was performed using a HP-5MS (crosslinked 5% Ph-Me-silicone) capillary column having 30 m x 0.25 mm i.d.; film thickness, 0.25μm. Helium was used as the carrier gas (1mL/min) and the column oven temperature was programmed to maintain at 100 ºC for 1.5 min, rise at 25 ºC per minute until it reached 190 ºC, increase of 3 ºC per minute until 280 ºC, maintaining the temperature for 5 minutes, the temperature rose 25 ºC per minute until 325 ºC and maintained this temperature for 3 minutes. Analysis was performed in splitless mode. The injection volumen was 3µL. Sample preparation In this article, the schema of procedure was: A. For deproteinized plasma, the procedures were based in three steps: - Deproteinization of human plasma with different methods of deproteinization. - Steroids extraction with organic solvent mixture. - Derivatization of extracted steroids. B. For non-deproteinized plasma, the procedures were based in two steps [16]: - Steroids extraction with organic solvent mixture. - Derivatization of extracted steroids. A. Deproteinization of human plasma - Deproteinization by chemical deproteinizing agents. as internal standard were added. The deproteinization was carried out with 1 ml of the following reagents; acetonitrile, methanol, ethanol, HCl 0.5 M/methanol (1:1) (v/v ), trichloroacetic acid (10%), sulfosalicylic acid (20%) and perchloric acid after 30 minutes of shaking. - Deproteinization by heat as internal standard were added. The deproteinization was carried out by heat at 60ºC for 30 minutes. - Deproteinization by proteases Deproteinization was performed by using following proteases: saitoi and Bacillus subtilis. - Deproteinization by saitoi to 2.8 using HCl 0.1N. The deproteinization was carried out with 1 ml of the saitoi (1mg/mL) in a thermoblock for 24 hours at 36 ºC. - Deproteinization by Bacillus subtilis to 7 using acetic acid 1M (CH 3 COOH) and a phosphate buffer (K 3 /Na 2 H ) at ph 7. The deproteinization was carried out with 1 ml of the Bacillus subtilis (1mg/ ml) in a thermoblock for 24 hours at 36 ºC. Deproteinization by a mixture of saitoi and Bacillus subtilis without adjusting the ph as internal standard were added. The deproteinization was carried out with 1 ml of a mixture of saitoi (1mg/mL) and the Bacillus subtilis (1mg/mL) in a thermoblock for 24 hours at 36 ºC. 11

- Deproteinization by a mixture of saitoi and Bacillus subtilis adjusting the ph to the optimal ph of saitoi to 2.8 using HCl 0.1N. The deproteinization was carried out with 1 ml of a mixture of saitoi (1mg/ ml) and the Bacillus subtilis (1mg/mL) in a thermoblock for 24 hours at 36 ºC. - Deproteinization by a mixture of saitoi and Bacillus subtilis adjusting the ph to the optimal ph of Bacillus subtilis to 7 using acetic acid 1M (CH 3 COOH) and a phosphate buffer (K 3 /Na 2 H ) ph 7. The deproteinization was carried out with 1 ml of a mixture of saitoi (1mg/mL) and the Bacillus subtilis (1mg/mL) in a thermoblock for 24 hours at 36 ºC. - Deproteinization by successive addition of saitoi and Bacillus subtilis to 2.8 using HCl 0.1N. It was added 1 ml of the saitoi (1mg/mL) in a thermoblock for 24 hours at 36 ºC. After the ph was adjusted to 7 using acetic acid 1M (CH 3 COOH) and phosphate buffer (K 3 /Na 2 H- )at ph 7 and 1 ml of the Bacillus subtilis (1mg/ ml) was added keeping samples in a thermoblock for 24 hours at 36 ºC. - Deproteinization by activated carbon as internal standard were added. After we added 1 ml of a mix of activated carbon (1%) in buffer (K 3 / Na 2 H ) at ph 7 samle was shaken during 20 minutes, centrifugated at 8000 rpm during 10 minutes, the supernatant was collected. B. Steroids extraction The ph was adjusted to 9.5 using NaOH 1M and buffer (HCO 3 1.34M /CO 3 2-0.2M) at ph 9.5. The extraction was carried out with 2 ml of n-hexane/ethyl acetate (70/30%, v/v). After 30 minutes of shaking, the mixture was centrifuged for 5 minutes and the organic phase was dried under a stream of nitrogen. - Derivatization procedure To the dried plasma extracts 50 µl of a mixture of MST- FA:NH 4 I:dithioerythritol (1000:2:4) (v/w/w) was added. The reaction mixture is heated in a thermoblock for 30 minutes at 60 ºC to perform the reaction of derivatization, and after, it was encapsulated and injected into the chromatograph. Measurement of proteins contents In order to verify the effectiveness of the different deproteinization treatments, total proteins and albumin were determined in the extraction solvent and in the plasma after its deproteinization. The analysis of proteins and albumin contents was as follows. 1 ml of plasma was added to three sample tubes: - Test tube number 1 One extraction was performed with 2 ml of extraction solvent based in our method (N-hexane/ethyl acetate 70:30% (v/v). Organic layer, middle layer and aqueous layer were separated. The contents of total proteins and albumin were measured. - Test tube number 2 Two extraction were performed with 2 ml of extraction solvent based in our method (N-hexane/ethyl acetate 70:30% (v/v). Organic layer, middle layer and aqueous layer were separated. The contents of total proteins and albumin were measured. -Test tube number 3 Three extraction were performed with 2 ml of extraction solvent based in our method (N-hexane/ethyl acetate 70:30% (v/v). Organic layer, middle layer and aqueous layer were separated. The contents of total proteins and albumin were measured. To identify the deproteinization capacity, the assessment of plasmatic total protein and albumin were performed once the proteases had acted. Results and discussion The characteristics of the analytes are reflected in Table 1 & Figure 1, in which we described retention time of analytes, retention time relative to internal standard (IS) methyltestosterone, qualifications ions, quantification ion of each studied compound and the limits of detection (LOD) of our method so for androsterone and 5β-androsterone 0.20 ng/ml, for DHEA 0.05 ng/ml, for DHT and cortisol 0.02 ng/ml, for epiandrosterone 0.36 ng/ml, for nandrolone 0.33 ng/ml, for testosterone 0,48 ng/ml, for estriol 0.06 ng/ml, for tetrahydrocortisone (THE) 0.18 ng/ml, for tetrahydrocortisol (THF) 0.04 ng/ml and for cortisone 0.29 ng/ml. The signal/noise values for GC/IT-MS/MS detection of plasmatic steroids after performing different chemical deproteinization treatments are shown in Table 2. In 12

Table 1. Quantification- and qualification ions of different steroids Compound Retention time (min.) Rel. Retention time (min.)* Parent Ion (m/z) Qualifier Ions (m/z) Quantifier Ion (m/z) LOD (ng/ml) Androsterone 21.942 0.776 420 420-329-239 329 0.20 5β-androsterone (etiocholanolone) 22.207 0.785 420 420-329-239 329 0.20 Dehydroepiandrosterone (DHEA) 23.530 0.832 418 418-327-237 327 0.05 Epiandrosterone 23.786 0.841 420 420-329-239 329 0.36 Nandrolone 24.491 0.866 418 418-403 403 0.33 Dihydrotestosterone (DHT) 24.491 0.866 434 434-419-405-377 419 0.02 Testosterone 25.311 0.895 432 432-417-327-301 417 0.48 Estriol 29.752 1.052 504 504-311-295-281-269 311 0.06 Tetrahyidrocortisone (THE) 33.035 1.168 635 635-619-530 619 0.18 Tetrahydrocortisol (THF) 34.232 1.210 636 636-531-430 531 0.04 Cortisone 36.009 1.273 617 617-525-435 525 0.29 Cortisol 38.155 1.349 632 632-543-437 543 0.02 * Relative retention time is refered to internal standard (IS) methyltestosterone; LOD: limit of detection. 13

general, the signal/noise ratio decreased preventing most of the analytes from beying analyzed qualitatively. When using acetonitrile as deproteinization method, androstenedione and progesterone were detected, whereas the other steroid hormones failed to improve their signal/noise ratio. The use of ethanol for deproteinization allowed detection of progesterone, whereas for the other steroid hormones, the signal/noise ratio decreased. The heat treatment allowed deproteinized progesterone to be detected, but the signal/noise ratio for the rest of detecting steroid hormones impaired. The use of 0.5 M HCl/ MeOH (1:1) (v/v) improved the detection of progesterone but the signal/noise ratio for other steroid hormones got worse. The detection got MeOH progesterone, but does not improve the signal/noise ratio of the other steroid hormones. The TCA (10%) and sulfosalicylic acid (20%) impaired the signal/noise ratio of all steroid hormones. The use of perchloric acid caused all analytes under study to remain undetectable. In Table 3 are shown the results of using proteases as a deproteinization method. The use of a mixture of and Bacillus without ph adjustment improved the signal/noise ratio of epiandrosterone. Digestion with for 24 hours followed by another digestion with Bacillus during another 24 hours and the use of, allowed detection of progesterone. The signal/ noise ratios of the other compounds under study did not improve. The use of activated carbon as deproteinization method (Table 4) decreased all the signal/noise ratios of the compounds analyzed in this study. Most of them could hardly be detected. In relation to the analysis of the extraction solvent, neither proteins nor albumin appeared there, however in the middle layer, some proteins appeared as aqueous layer fractions (Table 5). The results presented in Table 6 show how total protein values and albumin values in human plasma decreased after a treatment with proteases. According to the results of this study, using any standard deproteinization method to remove the protein content from human plasma did not improve the signal/noise ratio for the detection of a range of steroids by GC/IT-MS/MS. Furthermore, the investigated deproteinization methods caused a loss of detection in many of the tested compounds, maybe due to coagulation-flocculation produced by the addition of these solvents. In relation to the use of proteases, proteins were degraded into their corresponding amino acids, but it is not effective to liberate the steroids under study. This might be due to the residual amino acid digestion products attaching to the studied compound, preventing a free hydroxyl group for subsequent derivatization reaction, and thus preventing analysis by GC/IT-MS/MS. The use of activated carbon for deproteinization of plasma emerged as an idea of their use in radioim munoassay; however the results have not been good enough for the detection of human plasmatic steroid hormones of steroids by GC/IT-MS/MS. Its use decreased the signal/noise ratio for all compounds under study, maybe because although the carbon absorbs the steroids, our solvent extractor was not effective enough to remove them. When studied compounds were extracted, all proteins remained in the aqueous phase, so the protein doesn t pass to organic phase.this Figure 1. Steroids in human plasma. (1) Androsterone; (2) Etiocolanolone; (3) DHEA; (4) Epiandrosterone; (5) Nandrolone; (6) DHT; (7) Testosterone; (8) Methyltestosterone (IS); (9) Estriol; (10) THE; (11) THF; (12) cortisone; (13) Cortisol. 14

Table 2. Signal/noise values for GC-IT-MS/MS detection of plasmatic steroids performing chemical deproteinization treatments Compound ACN EtOH Heat HCl/MeOH MeOH TCA (10%) SS (20%) HClO 4 No deproteinization Androsterone 5 20 7 7 26 ND ND ND 1184 5β-androsterone 9 14 ND 12 18 ND 13 ND 523 Dehydroepiandrosterone 1 5 5 2 1 ND 5 ND 315 Epiandrosterone 2 4 ND ND ND ND ND ND 4 Nandrolone ND ND ND ND ND ND ND ND 56 Dihydrotestosterone ND 7 ND 17 10 3 ND ND 24 Testosterone 6 6 ND ND ND ND ND ND 74 Methyltestosterone (IS) 34 593 3 7 14 2 8 ND 195 Estriol ND ND ND ND ND ND ND ND 69 Tetrahydrocortisone 3 1 1 3 1 ND 7 ND 105 Tetrahydrocortisol ND ND ND ND ND ND ND ND 47 Cortisone 4 46 4 17 ND ND ND ND 137 Cortisol 20 36 41 10 4 1 ND ND 484 ND=signal/noise< 3; ACN:acetonitrile; EtOH:ethanol; MeOH: methanol; TCA: trichloroacetic acid; SS: sulfosalicylic acid; HClO 4 : perchloric acid 15

Table 3. Signal/noise values for GC/IT-MS/MS detection of plasmatic steroids after using proteases. Compound & Bacillus a Protease(s) from Bacillus & Bacillus b & Bacillus c & Bacillus d Healthy subject Androsterone 12 10 12 10 2 13 1184 5β-androsterone 15 44 17 9 3 79 523 Dehydroepiandrosterone ND 1 1 3 ND 4 315 Epiandrosterone ND ND 2 3 3 13 4 Nandrolone ND ND ND ND ND ND 56 Dihydrotestosterone 12 40 66 ND ND 12 24 Testosterone ND ND ND ND 22 ND 74 Methyltestosterone (IS) 3 29 15 3 2 8 195 Estriol ND 10 ND ND ND ND 69 Tetrahydrocortisone 2 ND 3 ND ND ND 105 Tetrahydrocortisol ND ND ND ND 2 ND 47 Cortisone 16 ND 5 17 ND ND 137 Cortisol 11 24 18 37 22 22 484 a 24h; b at ph ; c at ph Bacillus; d without ph adjustment; ND: Signal/Noise <3 Table 4. Signal/noise values for GC/IT-MS/MS detection of plasmatic steroids after using activate carbon. Compound Activated carbon Healthy subject Androsterone 13 1184 5β-androsterone ND 523 Dehydroepiandrosterone 4 315 Epiandrosterone ND 4 Nandrolone ND 56 Dihydrotestosterone ND 24 Testosterone ND 74 Methyltestosterone (IS) 1 195 Estriol ND 69 Tetrahydrocortisone ND 105 Tetrahydrocortisol ND 47 Cortisone ND 137 Cortisol ND 484 might be due to a greater polarity of the aqueous compared with the organic phase, causing the proteins, which have a polar character, to remain in the aqueous phase. Conclusions Even though the use of a mixture of proteases ( saitoni & Bacillus subtilis 1 mg/ml) reached a 100% elimination of plasma albumin and 95% of plasma total protein, the signal/noise ratio for GC/ IT-MS/MS detection of plasmatic steroids did not improve, or was impaired. This was true for the performance of the other deproteinization meth- 16

Table 5. Total protein and albumin values in different layers Layer Total protein (g/dl) Reference values (g/dl) Albumin (g/dl) Reference values (g/dl) Plasma 6.9 6.4 8.2 4.0 3.7-5.2 Plasma after 1 extraction 6.6 6.4 8.2 3.8 3.7-5.2 Plasma after 2 extractions 7.1 6.4 8.2 4.0 3.7-5.2 Plasma after 3 extractions 7.1 6.4 8.2 4.1 3.7-5.2 Middle layer after 1 extraction 6.4 6.4 8.2 3.4 3.7-5.2 Middle layer after 2 extractions 6.9 6.4 8.2 3.4 3.7-5.2 Middle layer after 3 extractions 1.1 6.4 8.2 ----- 3.7-5.2 Extract after 1 extraction 0.0 6.4 8.2 ----- 3.7-5.2 Extract after 2 extractions 0.0 6.4 8.2 ----- 3.7-5.2 Extract after 3 extractions 0.0 6.4 8.2 ----- 3.7-5.2 Table 6. Total protein and albumin values treated with different proteases. Sample Total proteins Total protein elimination Reference values Albumin Albumin elimination Reference values (g/dl) (%) (g/dl) (g/dl) (%) (g/dl) Plasma 8.0-6.4-8.2 5.2-3.2-4.8 Plasma + 10mg/10mL Protease Mixture of Proteases (10mg/10mL) Plasma + (/ Bacillus 24 h) 2.1 73.8 6.4-8.2 < 0.1 100 3.2-4.8 0.4 95.0 6.4-8.2 < 0.1 100 3.2-4.8 0.6 92.5 6.4-8.2 < 0.1 100 3.2-4.8 ods, i.e. temperature treatments or activated carbon. The analysis of plasmatic steroid hormones in this study do not carry any problem because of there are not any proteins in the extraction solvent. None of the deproteinization methods tested in this paper effectively improve the signal/noise ratio of steroid hormones in plasma detected by GC/IT-MS/MS, and furthermore, these methods increase the pre-treatment time of the samples. Abbreviations GC/IT-MS/MS: Gas chromatograph directly coupled ion-trap mass spectrometer. GC-MS: Gas Chromatography-Mass Spectrometry. LOD: Limit of Detection. ND: Not Detected. S/N: Signal-to-noise ratio. IS: Internal Standard. THE : Tetrahydrocortisone. DHT: Dihydrotestosterone DHEA: Dehydroepiandrosterone. THF: Tetrahydrocortisol. MSTFA: N-methyl N-trimethylsilyl-trifluoroacetamide. References 1. Wolthers BG, Kraan GP: Clinical applications of gas chromatography and gas chromatography-mass spectrometry of steroids. J Chromatogr A 843(1-2), 247-274 (1999). 2. Bowden JA, Colosi DM, Mora-Montero DC, Garrett TJ, Yost RA: Enhancement of chemical derivatization of steroids by gas chromatography/mass spectrometry (GC/MS). J Chromatogr B Analyt Technol Biomed Life Sci 877(27), 3237-3242 (2009). 3. Daykin CA, Foxall PJ, Connor SC, Lindon JC, Nicholson JK: The comparison of plasma deproteinization methods for the detection of low-molecular-weight metabolites by (1)H nuclear magnetic resonance spectroscopy. Anal Biochem 304(2), 220-230 (2002). 4. Diallo S, Lecanu L, Greeson J, Papadopoulos V: A capillary gas chromatography/mass spectrometric method for the quantification of hydroxysteroids 17

in human plasma. Anal Biochem 324(1), 123-130 (2004). 5. Zarzycki PK, Kulhanek KM, Smith R, Clifton VL: Determination of steroids in human plasma using temperature-dependent inclusion chromatography for metabolomic investigations. J Chromatogr A 1104(1-2), 203-208 (2006). 6. Cecchi F, Ruggiero M, Cappelletti R, Lanini F, Vannucchi S: Improved method for analysis of glycosaminoglycans in glycosaminoglycan/protein mixtures: application in Cohn-Oncley fractions of human plasma. Clin Chim Acta 376(1-2), 142-149 (2007). 7. Nikolaou PD, Papoutsis, Ii, Atta-Politou J et al.: Validated method for the simultaneous determination of methadone and its main metabolites (EDDP and EMDP) in plasma of umbilical cord blood by gas chromatography-mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci 867(2), 219-225 (2008). 8. Anderson D, Kerr DJ, Blesing C, Seymour LW: Simultaneous gas chromatographic-mass spectrophotometric determination of alpha-fluoro-beta-alanine and 5-fluorouracil in plasma. J Chromatogr B Biomed Sci Appl 688(1), 87-93 (1997). 9. Wang L, Morris ME: Liquid chromatography-tandem mass spectroscopy assay for quercetin and conjugated quercetin metabolites in human plasma and urine. J Chromatogr B Analyt Technol Biomed Life Sci 821(2), 194-201 (2005). 10. Staerk U, Kulpmann WR: High-temperature solid-phase microextraction procedure for the detection of drugs by gas chromatography-mass spectrometry. J Chromatogr B Biomed Sci Appl 745(2), 399-411 (2000). 11. Xie S, Suckow RF, Mason BJ, Allen D, Cooper TB: Rapid and sensitive determination of nalmefene in human plasma by gas chromatography-mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci 773(2), 143-149 (2002). 12. Kountouri AM, Mylona A, Kaliora AC, Andrikopoulos NK: Bioavailability of the phenolic compounds of the fruits (drupes) of Olea europaea (olives): impact on plasma antioxidant status in humans. Phytomedicine 14(10), 659-667 (2007). 13. Wang S, Chen Y, Gao Z, Xiong M, Zhong Z, Ye L: Gas chromatographic-mass spectrometric analysis of d-limonene in human plasma. J Pharm Biomed Anal 44(5), 1095-1099 (2007). 14. Martinez-Ortega MV, Garcia-Parrilla MC, Troncoso AM: Comparison of different sample preparation treatments for the analysis of wine phenolic compounds in human plasma by reversed phase high-performance liquid chromatography. Anal Chim Acta 205, 49-55 (2004). 15. Ralston PB, Strein TG: A Study of Deproteinization Methods for Subsequent Serum Analysis with Capillary Electrophoresis. Microchem J 55(2), 270-283 (1997). 16. Toribio-Delgado AF, Maynar-Mariño M, Caballero-Loscos MJ, Robles-Gil MC, Olcina-Camacho G, Maynar-Mariño JI: Qualification and Quantification of Seventeen Natural Steroids in Plasma by GC Q- MS and GC-IT MS/MS. J Chromatogr Sci 50(4) 349-357 (2012). Citation: Toribio-Delgado AF, Robles-Gil MC, Maynar-Mariño M, Olcina-Camacho G, Maynar-Mariño JI. Is deproteinization necessary in the determination of human plasmatic steroids by GC/IT-MS/MS? J Appl Bioanal 1(1), 10-18 (2015). Open Access and Copyright: 2015 Toribio-Delgado AF et al. This article is an open access article distributed under the terms of the Creative Commons Attribution License (CC-BY) which permits any use, distribution, and reproduction in any medium, provided the original author(s) and source are credited. Funding/Manuscript writing assistance: The authors have no financial support or funding to report and also declares that no writing assistance was utilized in the production of this article. Competing interest: The authors have declared that no competing interest exist. 18