Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

Similar documents
CALIBRATION CERTIFICATE # 500

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

DeFelsko PosiTector CMM IS Probe

Certificate of Accreditation

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

DeFelsko PosiTector Dew Point Meter DPM & DPMD Probes

PPC4E Pressure Controller/Calibrator Superior rangeability and reliability

2014 Programme of Training Courses in Measurement & Calibration

UKAS calibration laboratory 8809 Temperature, pressure and electrical signals calibration guide

CTAG:b9927e36a4b3f34a9138ce0bb2426b22 Certificate of Calibration

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

PRESSURE TRANSDUCERS WITH UNIFIED (4-20) ma OR (10-50) ma OUTPUT SIGNAL

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

MET-101 Introduction to Metrology

CTAG:e43bda4f8e48714fbb aefb192 Certificate of Calibration

Pneumatic High-Speed Pressure Controller Model CPC3000

HUMOR 20 HUMOR 20. High-precision Humidity Calibrator. Water T, p1

Additel 761 Automated Pressure Calibrators

GMC-I Service GmbH Calibration Center

COSMO INSTRUMENTS CO., LTD.

Gas mixtures. Individual solutions specifically for your application

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)

AIR PRESSURE CONTROLLED MASS MEASUREMENT SYSTEM

7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications

High-speed pressure sensor Model CPT6140

Additel 761 Automated Pressure Calibrators Selection Guide

AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY

CORESTA RECOMMENDED METHOD N 6

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

Pressure Metrology in Mexico

THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST

Datasheet TM40. HUBER INSTRUMENTE AG, Grellingerstrasse 23, CH-4208 Nunningen. All rights reserved Paige 1 / 5 Datasheet_TM40_

Additel 761 Automated Pressure Calibrators Selection Guide

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex.

Additel 780 Series Pressure Controller

Additel 761 Automated Pressure Calibrators

CRITERIA FOR LABORATORY ACCREDITATION IN MASS METROLOGY

WBEA Standard Operating Procedure

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

UNDERSTANDING, ACCELERATED

Mass Flowmeters. Mass Flowmeters for Gases

Additel 780 Series Pressure Controller

Technical Procedure for General Laboratory Equipment

The Discussion of this exercise covers the following points:

Calibration Procedure

Series 7250 Ruska High-Speed Digital Pressure Controller. GE Sensing. Features

Unique mounting features for Thermal Mass Flow Meters. By Martin Raab in cooperation with CaTeC b.v.

Calibrating a Micropipette

NMISA New Gas Flow Calibration Facility

Additel 761 Automated Pressure Calibrators

Hart Scientific 5626/5628. Platinum Resistance Thermometer User s Guide. Rev. 6B0901

Calibration equipment

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

EURAMET P1061 COMPARISON OF AIR TEMPERATURE CALIBRATIONS

Pressure calibration. 20

Series Ruska Digital Pressure Indicator. GE Sensing & Inspection Technologies. Features

Pressure Calibration Solutions

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

Light Lab. Reliable Measurements through Regular Calibration. DAkkS

Advanced Test Equipment Rentals ATEC (2832)

5624 Platinum Resistance Thermometer User s Guide

TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION

Reference Material Institute for Clinical Chemistry Standards (ReCCS) JCCRM 621-4

PG9607 Piston Gauge. Automated primary pressure reference for absolute and gauge pressures to 500 kpa

Calibration equipment

THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS

LM LUMEN MAINTENANCE TEST REPORT

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF MASS METROLOGY

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

ali-q TM Gravimetric Verification Procedure

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

Performance Overview calibration Laboratory EP Instruments Messtechnik und Kalibrierung GmbH

Pressure Calibrators. Series PC..., PM...

Rotary Viscometers Rotary viscosimeters ST-2020

Guide No. 2/2009. Design, training and equipment Martin Firlus. Physikalisch Technische Bundesanstalt Braunschweig und Berlin

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

FPG8601 Force Balanced Piston Gauge

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR PRESSURE GAUGE. This publication replaces T.O. 33K dated 30 January 2012.

ADT 761 Automated Pressure Calibrator

HumiSys HF High Flow RH Generator

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

Rigel 601 CHECKBOX. Instruction Manual. 348A551 Issue 2.0. April Seaward Electronic Ltd. Issue 2.0

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

KENYA ACCREDITATION SERVICE

CHAPTER 3 RESULTS AND DISCUSSION

Recommended Practices for Calibration Laboratories

Additel 780 Series Pressure Controller

Relative Humidity Calibration Kit

628 Differential Pressure Decay Leak Tester

TEXAS DEPARTMENT OF PUBLIC SAFETY Breath Alcohol Laboratory

Carbo Online CO 2 Analyzer for Beverages. ::: Great People Great Instruments

Automatic Isokinetic Sampler. ST5 Evo

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

Transferpette -8/-12

Transcription:

Asset #: 123 Calibration - Certificate No: 2083.01 Customer: Condition: Action Taken: Sample Inc. 123 Sample Rd. Sample, MA 01234 The instrument was operational upon receipt. The instrument was adjusted and calibrated. Date Received: November 29, 2016 Due Date: * November 30, 2017 RH Calibrated By: Approved By: Dan Soave Quality Manager T, P Calibrated By: Hun Khun Calibration Technician The measurement results on the certificate are traceable to the SI via NIST or another National Metrology Institute. The results of this calibration relate only to the items being calibrated. This certificate may not be reproduced, except in full, without the prior written approval of the issuing laboratory. The certificate and all measurements (unless otherwise specified) comply with the requirements of ISO/IEC 17025:2005. The calibration laboratory is controlled at 22 C ± 3 C and 40 %RH ± 20 %RH. Special Limitations: None. *Any due date given is based on a customer provided calibration interval. A number of factors may cause drift prior to the due date. Monitor all devices and calibrate when measurement error is suspected. Page 1 of 8

Accredited Relative Humidity Calibration Procedure #: PI213878 Rev. I Instrument Range: 0 to 100 %RH Lab Environment: Relative Humidity 36.6 %RH, Temperature 22.4 C As Found Data Out Of Tolerance As Received: NO Relative Humidity, %RH 11.50 11.53 0.03 1.00 0.42 30.14 29.99-0.15 1.00 0.60 50.15 49.94-0.21 1.00 0.77 80.11 79.42-0.69 1.00 0.79 Temperature, C 22.03 21.97-0.06 0.21 0.13 As Left Data Relative Humidity, %RH 11.50 11.65 0.15 1.00 0.42 30.10 30.17 0.07 1.00 0.60 50.10 50.35 0.25 1.00 0.77 80.10 80.25 0.15 1.00 0.79 Temperature, C 22.05 22.02-0.03 0.21 0.13 Page 2 of 8

Accredited Relative Humidity Calibration Error (%RH) 0.40 0.20 0.00-0.20-0.40-0.60-0.80 Relative Humidity Error As Found As Left 12 30 50 80 Test Point (%RH) Reference Standards Calibration Information Model Serial Number Asset Number Calibration Date Due Date Thunder Scientific 2500 1007799 5011-0089 Jun. 02, 2016 Jun. 02, 2017 Agilent 34970A MY44081004 3011-0377 Jun. 01, 2016 Jun. 01, 2017 Page 3 of 8

Accredited Temperature Calibration Procedure #: 11603135 Rev. A Instrument Range: -40 to 60 C Lab Environment: Relative Humidity 35.0 %RH, Temperature 22.0 C As Found & As Left Data Out Of Tolerance As Received: NO Temperature, C 70.00 70.01 0.01 0.33 0.05 0.02 Temperature Error As Found & As Left Error ( C) 0.01 0.00 70 Testpoint ( C) Reference Standards and Measurement Equipment Model Serial Number Asset Number Calibration Date Due Date Fluke 1560/2560/5615-12 A34186 / A34691 / 624556 1011-0270-1 Jun. 24, 2016 Dec. 24, 2016 Fluke 6331 A53056 6011-0001 N/A N/A Page 4 of 8

Accredited Pressure Calibration Procedure #: 11603172 Rev. A Instrument Range: 500 to 1100 hpa Lab Environment: Relative Humidity 37.0 %RH, Temperature 21.0 C As Found Data Out Of Tolerance As Received: NO Pressure, hpa Reference Unit Under Test Error ± Tolerance ± Uncertainty 610.00 610.30 0.30 0.32 0.066 700.00 700.30 0.30 0.32 0.066 810.00 810.20 0.20 0.32 0.066 850.00 850.20 0.20 0.32 0.066 910.00 910.20 0.20 0.32 0.066 950.00 950.20 0.20 0.32 0.066 1000.00 1000.20 0.20 0.32 0.066 1060.00 1060.20 0.20 0.32 0.066 As Left Data Pressure, hpa Reference Unit Under Test Error ± Tolerance ± Uncertainty 510.00 510.00 0.00 0.10 0.066 610.00 610.00 0.00 0.10 0.066 700.00 700.00 0.00 0.10 0.066 810.00 810.00 0.00 0.10 0.066 900.00 900.00 0.00 0.10 0.066 1000.00 1000.00 0.00 0.10 0.066 1060.00 1060.00 0.00 0.10 0.066 1100.00 1100.00 0.00 0.10 0.066 Page 5 of 8

Accredited Pressure Calibration Error (hpa) 0.35 0.30 0.25 0.20 0.15 0.10 0.05 0.00 Pressure Error As Found As Left 610 700 810 850 910 950 1000 1060 Testpoint (hpa) Reference Standards and Measurement Equipment Model Serial Number Asset Number Calibration Date Due Date Fluke PPC4 A100Kp 439 PA-13451 May. 26, 2016 Feb. 26, 2017 Vaisala Shunt Resistor N/A ES-11112 May. 25, 2016 May. 25, 2017 Agilent 34970A MY44081004 3011-0377 Jun. 01, 2016 Jun. 01, 2017 Page 6 of 8

Description The calibration was performed in the Calibration Standards Laboratory of Vaisala, Inc. The instrument was first allowed to equilibrate to the laboratory environmental conditions for a period of at least 8 hours. Relative Humidity Calibration: The sensor of the instrument was placed in the chamber of a Thunder Scientific 2500. The instrument was allowed to stabilize at each testpoint. Temperature Calibration: The sensor of the instrument was put in a protective tube and immersed in a Fluke liquid bath. The reference probe was placed in the bath next to the unit under test. The instrument was allowed to stabilize for at least 30 minutes at each testpoint. Pressure Calibration: The instrument was allowed to warm up for at least 2 hours before the calibration. The instrument's input port was connected to the output of a Fluke PPC4 Pressure Controller/Calibrator and the connection was tested for leaks. The testpoints are measured from high to low then again from low to high. The instruments were allowed to stabilize for at least 2 minutes after each testpoint was reached. The reported readings are the average of the readings from the high to low cycle and the readings from the low to high cycle. References The Thunder Scientific 1200/2500 Two-Pressure Humidity Generator saturates a continuous stream of air with water vapor at a controlled pressure and temperature. The saturated high-pressure air then passes through an expansion valve to generate a specific humidity at the chamber pressure and temperature. Fluke liquid baths are stable and uniform systems that use digital control to maintain setpoint conditions. The reference probe is a Fluke platinum resistance thermometer. The Fluke Black Stack is a digital, high resolution, graphical indicator. The Fluke PPC4 Pressure Controller/Calibrator digitally controls the pneumatic pressure output using solenoid valves and differential pressure regulators. It measures the pressure with a quartz reference pressure transducter (Q-RPT). Page 7 of 8

Measurement results At least ten consecutive pairs of reference and unit under test measurements were recorded at each testpoint. Each measurement result on the certificate is the average of this set of readings. In or Out of Tolerance Decision Rule Out of tolerance conditions are determined by the product specification only. The calibration uncertainty is not tied in with the instrument's accuracy. Uncertainty The reported expanded uncertainty of the measurement is stated as the standard uncertainty of the measurement multiplied by the coverage factor of k=2, which corresponds to a coverage probability of approximately 95%. The standard uncertainty of the measurement has been determined in accordance with the ISO Guide to the Expression of Uncertainty in Measurement. DOC228428 Rev. D Page 8 of 8